Technical Note: Model-based magnification/minification correction of patient size surrogates extracted from CT localizers

Size: px
Start display at page:

Download "Technical Note: Model-based magnification/minification correction of patient size surrogates extracted from CT localizers"

Transcription

1 Technical Note: Model-based magnification/minification correction of patient size surrogates extracted from CT localizers Christiane Sarah Burton Department of Radiology, University of Wisconsin-Madison, 1111 Highland Avenue, Madison WI 53705, USA Annie Malkus Department of Medical Physics, University of Wisconsin-Madison, 1111 Highland Avenue, Rm 1005, Madison WI 53705, USA Frank Ranallo Departments of Medical Physics and Radiology, University of Wisconsin-Madison, 1111 Highland Avenue, Madison WI 53705, USA Timothy P. Szczykutowicz a) Departments of Radiology, Medical Physics, and Biomedical Engineering, University of Wisconsin-Madison, 1111 Highland Avenue, Madison WI 53705, USA (Received January 018; revised August 018; accepted for publication 11 September 018; published 0 November 018) Purpose: Patient size-specific dose estimate (SSDE) calculations require knowledge of a patient s size. Errors in patient size propagate through SSDE calculations. AAPM Reports 04 and 0 recommend that a magnification correction be applied to patient size surrogates extracted from CT localizer radiographs. This technical note presents a novel approach for such a magnification correction. Methods: In our model-based magnification correction, we assume that the patient s cross sections are elliptical with minor and major axes defined using the anterior posterior (AP) and lateral (LAT) patient dimensions. We parameterize the problem by modeling a line emanating from the source, grazing the patient (i.e., the ellipse), and then terminating onto the detector plane. We model tangent lines on each side of the ellipse on both the LAT and AP CT localizer radiographs. We also account for vertical mispositioning with table offset. We compared our correction model to the actual AP and LAT dimensions to the vendor-supplied CT localizer images that only received a geometric magnification correction, and to other methods described in the literature. We compare our model to the others using direct size to size comparisons as well as SSDE conversion factor. Results: Our model-based method provides consistent accurate results (less than 1.8 error for absolute size and 1. error for SSDE for all measurement conditions) for all positions and patient sizes. Existing literature-based methods had maximum errors for absolute size and SSDE of 7.5 and 5., and for the vendor, they were 30.9 and 17.0, respectively. Conclusion: We presented a new model-based geometric size correction method that outperforms a simple geometric correction as well as other methods presented in the literature. By modeling the patient cross section and beam geometry using information all derived from the DICOM header and CT localizer views, we demonstrated SSDE correction factor improvements from 17.0 (vendor correction) to 1. (model base). These changes correspond directly into changes in SSDE itself and also represent clinically realistic patient sizes and mispositioning amounts. 018 American Association of Physicists in Medicine [ Key words: CT Localizer radiograph, SSDE, Geometric Magnification, CT Dose 1. INTRODUCTION The size-specific dose estimate (SSDE) uses a normalized dose coefficient to scale the CTDIvol and this scale depends on the patient s body size. Therefore, having an accurate estimation of SSDE depends on having and accurate estimate of the patient size. The dose received by the patient depends on the scanner output and on the patient s size. Turner et al. 11,1 demonstrated that by using CTDIvol as a normalization factor, organ dose estimates can be obtained for a specific patient size, 11 and the relationship between normalized dose and patient size was consistent across scanner models. 1 Hence, obtaining accurate information about patient size is crucial for estimating patient dose in CT. To determine the size-specific dose estimate (SSDE), the American Association of Physicists in Medicine (AAPM) Task Group (TG) Report 04 states that patient size be measured based on their dimensions including anterior posterior pffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi (AP), lateral (LAT), and effective diameter ð LAT APÞ. Estimating patient surrogates from axial CT images can only be performed after the CT scan is finished and tomographic images are reconstructed. However, knowledge of patient size may be desirable in order to calculate SSDE prior to tomographic scanning or in cases when only CT localizer radiograph 165 Med. Phys. 46 (1), January /019/46(1)/165/8 018 American Association of Physicists in Medicine 165

2 166 Burton et al.: CT localizer size correction 166 images and dose information are available such as with the American College of Radiology (ACR) dose index registry. AAPM TG Report 0 6 recommends that to accurately estimate AP and LAT from CT localizers (might also be referred to as Scanned Projection Radiograph, SPR) 10,13,14 a magnification/minification correction should be applied. To our knowledge, all known magnification correction methods proposed including the magnification correction believed to be used by all vendors are illustrated in Fig. 1. The vendor correction is a simple geometric magnification correction which assumes that the patients widest width is located at isocenter and that they have a thickness of zero extent. With these assumptions, the vendor correction is simply a similar triangles based on SID over SOD correction. Where SOD stands for source to object distance which is the difference from the source to the center of the idealized ellipse. The SID represents the source to image/detector distance. Therefore, the vendor applies the same magnification/ minification scaling factor to all images regardless of their actual position within the gantry. Methods by Christensen, 4 Li et al., 5,10 and Raupach et al. 9 (also implemented by McMillan et al. 7 ) aim to modify the vendor correction scheme by estimating the actual patient position within the gantry in order to obtain a true SID over SOD correction. This is commonly accomplished by assuming that the patient s center is offset from isocenter along the y-direction (same orientation as AP, shown in Fig. 1) by an amount determined using the table height and patient thickness. The patient thickness term allows the distance between the center of the patient and the table top to be taken into account. While the methods essentially use the same equation, they differ in their processing methods needed to obtain the table height and projected patient thickness (AP dimension). For example, Li et al. 5 use an orthogonal CT localizer radiograph to estimate the table height and correct for patient miscentering relative to the isocenter. Raupach et al. 7,9 estimate the lateral dimension taking into account the DICOM table height field to modify their geometric magnification as Li et al. did. The commonality of these methods is that they do not model the actual intersection of the x-ray beam with the patient (as shown in Fig. 1). These methods assumption of the x-ray intersecting the patient at their widest extent, solely defined using the knowledge of table height and patient thickness, is incorrect. This paper will investigate the effect of assuming a patient to be a flat line, and the x-ray beam first intersecting the patient at their widest point to provide a new model-based magnification/minification correction scheme as shown in Fig. 1. y x Christensen 1990 Raupach et al. 007 Li et al. 01 Vendor Model-based x 1, y 1 Isocenter Phantom Grazing occurs at this y value x, y FIG. 1. Illustration of the difference in lateral projections between the modelbase, Christensen , Raupach et al. 9, and Li et al. 5, and the vendor s geometric magnification correction approaches. In this illustration, the patient s major cross section defined by LAT is not located at isocenter. The table height, h, is defined here as the difference between the table top and isocenter. The vendor assumes that the patient s lateral dimension is located at isocenter. Previously reported geometric correction methods allow the maximum patient lateral dimension to be at a table position other than isocenter, but do not model the nonzero patient thickness (here defined by AP). Our model-based approach models the patient as an ellipse (or circle if AP = LAT) with a line representing x rays emanating from the source striking the first point of contact on the patient (grazing) and striking the detector which in the CT localizer image would be the edges of LAT M. [Color figure can be viewed at wileyonlinelibrary.com]. MATERIALS AND METHODS In our model-based magnification correction, the patient is assumed to have an elliptical cross section. We assume that there exists an x-ray line path that emanates from the source which touches the ellipse at only one point on each side of the patient and intersects the detector plane. We call this point of contact between the x-ray beam and the patient as grazing as shown in Fig. 1. Table I describes our method for obtaining the true lateral and anterior posterior patient TABLE I. Previously reported magnification correction methods and our model-based approach. In the publication by Li et al., the addition of a square root was an error in the original paper and that has been corrected here. The model-based method lists the general solution to the intersection between a line characterized by the focal spot and edge of the patient s extent on a localizer view and the patient parameterized by their offset from isocenter and extent in the vertical and horizontal directions. See the Section 6 for more details on the model-based method. Method Inputs Equations Geometric SID/ SOD Christensen/ Raupach/ Limethods Model-based (present study) SOD, SID h, SOD, AP M h, SOD, AP M ; LAT M LAT ¼ LAT M SOD SID LAT ¼ LAT SOD ðh 1 APMÞ M SOD System of equations for the intersection of a line (i.e., x-ray projection grazing the patient as shown in Fig. 1) and an ellipse (i.e., the patient parametrized by an offset from isocenter calculated using the table height and the AP thickness with major and minor dimensions given by the LAT and AP dimensions)

3 167 Burton et al.: CT localizer size correction 167 thicknesses. The full mathematical description is given in Section 6. Variables AP M and LAT M (M meaning measured from the CT localizer with the vendor s SOD SID correction applied) are measured from the CT localizer radiograph using digital calipers (ImageJ, National Institutes of Health, Bethesda, MD, USA), and AP and LAT are the true dimensions. In Section 6, we also describe a source of error inherent to measuring flat objects in a flat plane using a curved detector. We experimentally validated our model-based method by imaging three phantoms with known dimensions at various table heights. The three phantoms included: (a) a 30 mm CTDI phantom, (b) a 15 mm cylindrical water phantom, and (c) a 3: ratio elliptical phantom with three differently sized regions: large 350 and 33 mm, medium 300 and 00 mm, and small 50 and mm, respectively, for the major and minor phantom diameters. 8 In our experimental validation, we kept all phantoms centered in the lateral direction while varying the table height in the y-direction. The images in this study were acquired using a 64 slice MDCT scanner (Discovery HD 750, GE Healthcare, Chicago, IL, USA). There are vendors where TCM is directly affected by patient centering, and others that are not. The accurate size calculation of the patient from the CT localizer may be more important for SSDE calculations than TCM control for some CT vendors. 8 From the CT localizer radiographs, we used digital calipers to measure the AP and LAT dimensions of the phantoms from AP and LAT. We repeated AP M and LAT M measurements five times and took the average of the AP M and LAT M dimensions measured for each phantom. As phantom edges were well-defined, there was not a significant variation in width measurements; therefore, we did not include error bars in this study. Measurement maximum deviations between the five measurements of AP M and LAT M were submillimeter for all phantom sizes and positioning. We compare our methods to other magnification corrections summarized in Table I for both absolute sizes and also for SSDE conversion factors. The latter allows for the dosimetric impact of the different geometric correction methods to be compared. Absolute size differences were quantified by calculating the percent difference in size between the geometric correction algorithm and the known ideal phantom size. SSDE was calculated for all methods using the equations provided in AAPM TG report 04 [Eq. (A1), parameters a p= ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi , b = ] for effective diameter ( LAT AP) to conversion factor. 3. RESULTS Figure shows LAT and AP estimation using magnification correction methods for 30 mm CTDI and 15 mm water cylinder phantoms. The dashed lines represent the actual AP and LAT dimensions. Figure 3 shows how close each magnification correction is to the actual measured diameter for elliptical phantoms. The dashed lines represent the actual AP and LAT dimensions. Figures and 3 demonstrate how the experimental points are closest to the dashed lines (actual thickness) for the modelbase approach, particularly for larger phantoms relative to the other methods. Table II summarizes the worst and best conditions for each phantom and correction method shown in Figs. and 3. Figure 4 shows the calculation pffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi of correction factors using the effective diameter ð LAT APÞ from the magnification correction approaches. Table III summarizes the worst and best conditions for SSDE determination for each phantom and correction method shown in Fig DISCUSSION We demonstrate that our magnification correction method gives a more accurate estimate of the phantom size for both the lateral (LAT) and anterior (a) 400 Lateral dimension [mm] mm 15 mm 150 (b) Anterior-posterior dimension [mm] mm 15 mm 150 FIG.. Magnification correction for 30 and 15 mm cylindrical phantom dimensions estimated from CT localizer images. Measurements are shown from a variety of table heights using the University of Wisconsin (UW) model-based, the geometric, and the methods. (a) and (b) show the lateral anterior posterior dimensions, respectively. The dashed lines across the graphs show the ideal 30 and 15 mm dimensions. [Color figure can be viewed at wileyonlinelibrary.com]

4 168 Burton et al.: CT localizer size correction 168 (a) 500 Lateral dimension [mm] mm 300 mm 50 mm 00 (b) Anterior-posterior dimension [mm] mm 00 mm mm FIG. 3. Magnification correction for three different regions of the elliptical phantom dimensions. Measurements are shown from a variety of table heights using the University of Wisconsin (UW) model-based, the geometric, and the methods. (a) and (b) show the lateral anterior posterior dimensions, respectively. The three different lateral (LAT) and anterior posterior (AP) regions of the elliptical phantom are as follows for the major to minor axis: large mm, medium mm, and small mm. The dashed lines across the graph show the actual size of these dimensions for both lateral and anterior posterior measurements. [Color figure can be viewed at wileyonlinelibrary.com] TABLE II. Summary of best and worst results for absolute size determination from the various correction schemes for each phantom. The subscripts max and min on LAT and AP indicate that the measurement is the worst or best, respectively. Offset of difference between center of patient and isocenter refers to the table heights corresponding to the best and worst results, respectively. LAT max LAT min AP max AP min Length Offset max Length Offset min Length h max Length h min Vendor (30 mm) vendor (15 mm) Vendor (large) Vendor (medium) Vendor (small) Christensen/Raupach/ Li (30 mm) Christensen/Raupach/ Li (15 mm) Christensen/Raupach/ Li (large) Christensen/Raupach/ Li (medium) Christensen/Raupach/ Li (small) Model-based (30 mm) Model-based (15 mm) Model-based (large) < Model-based (medium) < Model-based (small) posterior (AP) dimensions shown in Figs. and 3 relative to the other methods. Our model, as discussed in the Section 6, assumes that there is no vertical mispositioning present in the lateral CT localizer radiograph. This explains why our model shows a slight increase in error as the mispositioning increases for Figs. (b) and 3(b). This also explains the slight downward slope in our corrected LAT measurements as seen in Figs. (a) and 3(a). Our data show that the UW modelbased method provides consistent accurate results, less than 1.8 error for absolute size for all measurement conditions relative to 30.9 and 7.5 for the vendor and Christensen/ Raupach/Li approaches. Using the

5 169 Burton et al.: CT localizer size correction 169 (a). (b). Correction factor mm 30 mm Correction factor small medium large 0.8 table difference from isocenter [mm] 1 FIG. 4. SSDE correction factor estimated from CT localizer images using the University of Wisconsin (UW) model-based, the geometric, and the magnification corrections. Correction factors as shown as a function of table height for (a) the cylindrical phantoms and (b) the three different sized sections of the elliptical phantom. Dashed lines across the each graph show the ideal SSDE correction factor derived using the known phantom effective diameter. [Color figure can be viewed at wileyonlinelibrary.com] approach for the largest elliptical phantom, the SSDE correction factor differed by approximately 5.0 and for the vendor s correction the maximum table height gave a difference of Our model-based method only differed by 1. for the largest elliptical section for SSDE correction factor. While determining D W was not done in this technical note, our method can be applied to CT localizer-based methods for calculating D W. Li et al. 5 demonstrate how the effective diameter (D E ) can be related to the waterequivalent diameter (D W ) by measuring the mean pixel value of the phantom area with the product of the width and normalizing it to the attenuation coefficient of water. Anam et al. 1 have recently reported a method for calculating D W which involves determining a linear relationship between pixel value and water thickness and then solving for water-equivalent area (A W ) by integrating over the x- axis of the CT localizer. Both Li et al. and Aman et al. s methods rely on accurate geometric size information to TABLE III. Summary of best and worst results for SSDE CF, where the subscript CF stands for conversion factor, for the various correction schemes for each phantom. The subscripts max and min on LAT and AP indicate that the measurement is the worst or best, respectively. Offset of difference between center of patient and isocenter refer to the table heights corresponding to the best and worst results, respectively. SSDE max SSDE min SSDE CF Offset max SSDE CF Offset min Vendor (30 mm) Vendor (15 mm) Vendor (large) < Vendor (medium) Vendor (small) ( mm) ( mm) (large) (medium) (small) Model-based (30 mm) Model-based (15 mm) Model-based (large) Model-based (medium) Model-based (small)

6 170 Burton et al.: CT localizer size correction 170 obtain D W and should therefore benefit from a geometric correction such as that described in this note. A source of error in using this method for future patient studies would come from not accounting for miscentering in the lateral (i.e., x-direction). This is due to not being able to estimate the patient s miscentering along the lateral dimension using information from the DICOM header. Similar to the postprocessing method by Li et al., one could calculate largest center of mass along each column in the image and subtract that location from the center of the image to estimate patient offset. Not accounting for this miscentering in the magnification/minification correction would result in either an under or overdose to the patient for calculating the effective diameter or for TCM response. We should note that our methods may not be needed for correcting geometric magnification on AEC algorithms by all vendors. Some vendors, as shown by Merzan et al. 8, are not effected by patient mispositioning. Albeit a SSDE derived using a CT localizer radiograph from such vendors may still experience error from geometric magnification. 5. CONCLUSIONS We have shown that our model-based magnification correction method is more accurate than the current vendor geometric SID over SOD based correction algorithms. We have also compared our method to all previously known reported methods in the literature and demonstrated our method to be superior. Our method requires no actual tomographic CT images, it only uses CT localizer radiograph and DICOM header information, making it amenable to large dose registries. APPENDIX (A1) The model-based approach models the patient as an ellipse and takes into account when the x-ray beam first grazes the anatomy (Fig. 1). In this section, we show derivation of two common points on the ellipse and a line using a system of two equations. The patient is modeled as an ellipse which is parameterized as ðx bþ ðy kþ 1 ¼ þ (A1) LAT AP where variables AP and LAT represent the anterior posterior and lateral dimensions of the patient or phantom, respectively, and variables k and b represent the table offset from isocenter in the anterior posterior (i.e., y-direction or vertical) and lateral (i.e., x-direction or horizontal) direction, respectively. In this note, we assume that the patient is centered along the isocenter in the x-direction; therefore, b = 0. The distance between the source and isocenter is referred to as the sourceto-object distance (SOD). We solve for AP and LAT using a system of two equations. We refer to these equations as the anterior posterior and lateral equation, named because they are set up using the geometry of the anterior posterior and lateral CT localizer radiographs, respectively. The two equations are set up by modeling the geometry of a ray emanating from the source, grazing ( grazing is defined in Fig. 1) the patient at a single point, and then progressing to strike the detector. In such a geometry, a grazing ray can be parameterized using the source as the origin and defining the slope of the line as the ratio of half the projected patient width to SOD. The projected patient thickness is given by either LAT M or AP M depending on if the CT localizer radiograph is a AP/PA or lateral view, respectively. For the anterior posterior equation, the line grazing the side of the patient ellipse model can be represented as y = mx + c where c = 0 as the source location is taken at the origin and m ¼ SOD. For the lateral LATM equation, m ¼ SOD. APM We assume that the effect of geometric magnification on the size of the lateral CT localizer radiograph is negligible and therefore set k to SOD for the lateral equation. The error introduced by not modeling the vertical offset present in the lateral equation for the 30 mm phantom used in this study was quite small. For table offsets of 65.0, 3.5, 0.0, 31.5, and 63.5 mm from isocenter the measured size of the phantom using only the vendor geometric calibration was 33, 335, 335, 334, and 33 mm. In other words, while our model assumed a constant AP M, in reality, it changed by a maximum of 3 mm for the 30 mm phantom over the table heights we analyzed. For larger patients, this error will increase slightly; however, in such cases, the patient will quickly become truncated in the CT localizer radiograph view. Therefore, we feel that simplifying the mathematics by not modeling the vertical mispositioning in setting up the lateral equation is justified. For the anterior posterior equation, k is given by SOD þ h AP which is not an assumption, but the actual off set. The solution between a line and an ellipse has a known closed-form solution for intersection points x 1; and y 1;. The points x 1; and y 1; are intersection points in the x- and y- planes, and equations for these points are provided here. The solution is given by pffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi x 1; ¼ ba ml LA L m þ A d k þ dk L m þ A (A) and pffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi y 1; ¼ A d kl m LAm L m þ A d k þ dk L m þ A (A3) where e = c k, d = c + mb, L ¼ LAT AP, and A ¼. In our geometry, the y-value of the intersections is equal making solution for the y-coordinates simpler by forcing the term to zero. Exploiting this fact for both the anterior posterior and lateral equations provides two equations with only two unknowns. Therefore, our system of equations is constructed by using the anterior posterior and lateral equation geometry to set the inside of the square root term to zero. Equations (A) and (A3) are not reduced so that the entire equation may be used in future studies to account for patient offset in the x-

7 171 Burton et al.: CT localizer size correction 171 Arc of Detectors SID d C' B C Isocenter SOD θ Focal Spot SOD/SID correction for that point at which B and C touch, the measured length of B will simply be the length of C. If h is the half angle of the Arc C. Then, C ¼ðSODÞh (A4) B ¼ðSODÞ tan h; (A5) where h ¼ arctan B ðsodþ; (A6) FIG. 5. Illustration of the geometric image distortion inherent to projecting a straight object onto a curved detector surface. The Arc C is centered on the focal spot and B is a straight line tangent to it at the iso-ray. Assume B is a straight metric ruler passing through the center of the scanned field of view (SFOV) and centered at the iso-ray. However B, as shown, does not necessarily pass through the isocenter. The SID/SOD correction applied to B to correct for its magnification at the detector is actually only correct at its center. The measured length of the cm increments on B will progressively decrease at positions farther away from its center. This is due to two effects; (1) the parts of B away from its center are closer to the detectors and their geometric magnification is less; () these parts of B are no longer parallel to the detector and are appear foreshortened to the detector. [Color figure can be viewed at wileyonlinelibrary.com] direction. Once the term was set to zero, we inserted Equations (A) and (A3) into a Matlabâ function (vpasolver, The Mathworks, Natwick, MA, USA) which provided us with a numerical approximation to the solution of the system of equations. In this system, AP and LAT are unknown, h is determined from the DICOM header (tag ID:0018,1130), SOD may be determined from the DICOM header (tag ID: 0018,940) or may be found in the vendor s technical manual if not provided in the DICOM header, and LAT M =AP M are measured from CT localizer radiograph images. In this note, we used digital calipers to manually determine LAT M =AP M ; however, in practice, they can be determined using automated methods. 3 In practice, the patient model-based magnification approach is useful for cases that require two CT localizers such as abdomen and chest cases for both adult and pediatric patients. However, since this approach requires two CT localizers, it might not be useful in determining size surrogates for extremity and head scans where only one CT localizer is more commonly acquired. In addition to modeling the grazing effect, there is an image distortion that should be taken into account for most scanners. It derives from the attempted mapping of a straight line through the scan field of view to the detectors which form an arc of a circle. Distances measured in the CT localizer radiograph are relative to this arc of detectors but with a minification correction factor of SOD/SID. This correction is accurate for distances along the Arc C. However, as shown in Fig. 5, the distances measured along B are progressively minified for distances measured farther from the iso-ray (the ray from the focal spot passing through the isocenter). At the isoray, measurements along B and C are equal. Using a single thus, C ¼ ðsodþarctanbðsodþ; (A7) gives the measured length of the ruler, which will always be less than its actual length, B. To somewhat counteract this overall minification of the measured length of the ruler, the length given above can be multiplied by a factor greater than one (f), or equivalently the SOD/SID correction can be increased (SOD is increased to SOD + d). In this case, the central part of the ruler is magnified while the outside remains minified, relative to the true distance. You can see this effect in increasing the SOD in the Arc C in Fig. 5. In this case, we can modify the measured length of B to be B : B 0 B ¼ ðfþðsodþarctan (A8) ðsodþ or B 0 B ¼ ðsod þ dþ arctan (A9) ðsodþ For the scanner used in this paper: f = or d = 1.3 cm. With these values B = B when B = 15 cm. In other words, for a ruler with a length of 30 cm, its measured length in the CT localizer will be 30 cm, if the center of the ruler is at isocenter. In the case of our system, for actual lengths ranging from 0 to 36 cm, the error in the measured lengths in the CT localizer will be less than mm. Thus, in this paper, no correction needed to be applied. If these corrections needed to be applied, one would simply adjust the measured distances from the localizer radiographs (LAT M and AP M ) using Eqs. (8) and (9). a) Author to whom correspondence should be addressed. Electronic mail: tszczykutowicz@uwhealth.org. REFERENCES 1. Anam C, Fujibuchi T, Toyoda T, et al. A simple method for calibrating pixel values of the CT localizer radiograph for calculating water-equivalent diameter and size-specific dose estimate. Radiat Prot Dosimetry. 017;179: Boone JMSK, Strauss K, Cody D. Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations: report of AAPM task group 04. American Association of Physicists in Medicine website; Christianson O, Li X, Frush D, Samei E. Automated size-specific CT dose monitoring program: assessing variability in ct dose. Med Phys. 01;39:

8 17 Burton et al.: CT localizer size correction Curry TS, Dowdey JE, Murry RC. Christensen's Physics of Diagnostic Radiology, Philadelphia, PA: Lippincott Williams & Wilkins; Li B, Behrman RH, Norbash AM. Comparison of topogram-based body size indices for CT dose consideration and scan protocol optimization. Med Phys. 01;39: McCollough CH, Donovan MB, Bostani M, et al. Use of water equivalent diameter for calculating patient size and size-specific dose estimates (ssde) in CT. Technical report, College Park, MD: American Association of Physicists in Medicine; McMillan K, Bostani M, Cagnon CH, et al. Estimating patient dose from CT exams that use automatic exposure control: development and validation of methods to accurately estimate tube current values. Med Phys. 017;44: Merzan D, Nowik P, Poludniowski G, Bujila R. Evaluating the impact of scan settings on automatic tube current modulation in CT using a novel phantom. BJR. 016;90: Raupach R, Schaller A, S uss C, Wolf H. Computerized method and device for calculating an orthogonal x-ray attenuation of a subject using a measured, x-ray attenuation reference. US Patent 7,89, Toth T, Ge Z, Daly M. The influence of patient centering on CT dose and image noise. Med Phys. 007;34: Turner AC, Zankl M, DeMarco JJ, et al. The feasibility of a scannerindependent technique to estimate organ dose from MDCT scans: using CTDIvol to account for differences between scanners. Med Phys. 010;37: Turner AC, Zhang D, Khatonabadi M, et al. The feasibility of patient size-corrected, scanner-independent organ dose estimates for abdominal CT exams. Med Phys. 011;38: Wang J, Christner JA, Duan X, Leng S, Yu L, McCollough C. Attenuation-based estimation of patient size for the purpose of size specific dose estimation in CT. Part ii. Implementation on abdomen and thorax phantoms using cross sectional CT images and scanned projection radiograph images. Med Phys. 01;39: Wang J, Duan X, Christner JA, Leng S, Yu L, McCollough C. Attenuation-based determination of patient size for the purpose of size specific dose estimation in CT: Part ii. Implementation on abdomen and thorax phantoms using cross sectional CT images and scanned projection tadiograph images. Med Phys. 01;39:

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

Automated Calculation of Water-equivalent Diameter (D W. ) Based on AAPM Task Group 220

Automated Calculation of Water-equivalent Diameter (D W. ) Based on AAPM Task Group 220 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 Automated Calculation of Water-equivalent Diameter (D W ) Based on AAPM Task Group 220 Choirul Anam, 1,2a Freddy Haryanto, 2 Rena

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

S. Guru Prasad, Ph.D., DABR

S. Guru Prasad, Ph.D., DABR PURPOSE S. Guru Prasad, Ph.D., DABR Director of Medical Physics IAEA Consultant NorthShore University Health System and University of Chicago, Pritzker School of Medicine Current TPS utilize more information

More information

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist CT Protocol Review: Practical Tips for the Imaging Physicist Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center August 8, 2013 AAPM Annual Meeting Goals Understand purpose and importance

More information

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE / Radiographic Tomosynthesis: Image Quality and Artifacts Reduction Baojun Li, Ph.D Department of Radiology Boston University Medical Center 2012 AAPM Annual Meeting Background Linear Trajectory Tomosynthesis

More information

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization

More information

Computed Tomography Imaging: CT Protocol Management. Caveat 8/3/2017

Computed Tomography Imaging: CT Protocol Management. Caveat 8/3/2017 Computed Tomography Imaging: CT Protocol Management Mark P. Supanich, Ph.D., DABR AAPM Annual Meeting 3 rd August, 2017 Slides at: goo.gl/k8n8jf Rush is an academic health system comprising Rush University

More information

Dose-volume product (DVP) as descriptor for estimating total energy imparted to patient undergoing CT examination

Dose-volume product (DVP) as descriptor for estimating total energy imparted to patient undergoing CT examination DIAGNOSTIC AND INTERVENTIONAL IMAGING Journal of Medical Physics and Biophysics, Vol. 3, No. 1, December 2016 Dose-volume product (DVP) as descriptor for estimating total energy imparted to patient undergoing

More information

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography

More information

Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset

Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset Andreas Maier 1, Patrick Kugler 2, Günter Lauritsch 2, Joachim Hornegger 1 1 Pattern Recognition Lab and SAOT Erlangen,

More information

Dose-Volume Product (DVP) as descriptor for estimating total energy imparted to patient undergoing CT examination

Dose-Volume Product (DVP) as descriptor for estimating total energy imparted to patient undergoing CT examination DIAGNOSTIC AND INTERVENTIONAL IMAGING Journal of Medical Physics and Biophysics, Vol. 2, No. 1, September 2016 Dose-Volume Product (DVP) as descriptor for estimating total energy imparted to patient undergoing

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT.

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT. Progress report: Development of assessment and predictive metrics for quantitative imaging in chest CT Subaward No: HHSN6801000050C (4a) PI: Ehsan Samei Reporting Period: month 1-18 Deliverables: 1. Deployment

More information

TESTING OF THE CIRCLE AND LINE ALGORITHM IN THE SETTING OF MICRO-CT

TESTING OF THE CIRCLE AND LINE ALGORITHM IN THE SETTING OF MICRO-CT SCA2016-080 1/7 TESTING OF THE CIRCLE AND LINE ALGORITHM IN THE SETTING OF MICRO-CT Alexander Katsevich 1, 2 and Michael Frenkel 1 1 itomography Corp., 2 University of Central Florida (UCF) This paper

More information

Implementation and evaluation of a protocol management system for automated review of CT protocols

Implementation and evaluation of a protocol management system for automated review of CT protocols JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Implementation and evaluation of a protocol management system for automated review of CT protocols Joshua Grimes, Shuai Leng, a Yi

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Effects of the difference in tube voltage of the CT scanner on. dose calculation Effects of the difference in tube voltage of the CT scanner on dose calculation Dong Joo Rhee, Sung-woo Kim, Dong Hyeok Jeong Medical and Radiological Physics Laboratory, Dongnam Institute of Radiological

More information

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),

More information

Non-Stationary CT Image Noise Spectrum Analysis

Non-Stationary CT Image Noise Spectrum Analysis Non-Stationary CT Image Noise Spectrum Analysis Michael Balda, Björn J. Heismann,, Joachim Hornegger Pattern Recognition Lab, Friedrich-Alexander-Universität Erlangen Siemens Healthcare, Erlangen michael.balda@informatik.uni-erlangen.de

More information

AAPM Standard of Practice: CT Protocol Review Physicist

AAPM Standard of Practice: CT Protocol Review Physicist AAPM Standard of Practice: CT Protocol Review Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center September 11, 2014 2014 Texas Radiation Regulatory Conference Goals Understand purpose

More information

UNCOMPROMISING QUALITY

UNCOMPROMISING QUALITY ION CHAMBERS UNCOMPROMISING QUALITY Designed with over 30 years of scientific integrity for a broad range of dosimetry measurements in diverse radiation beams Farmer-type Chambers For absolute dosimetry

More information

CT vs. VolumeScope: image quality and dose comparison

CT vs. VolumeScope: image quality and dose comparison CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,

More information

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

Improvement of Efficiency and Flexibility in Multi-slice Helical CT J. Shanghai Jiaotong Univ. (Sci.), 2008, 13(4): 408 412 DOI: 10.1007/s12204-008-0408-x Improvement of Efficiency and Flexibility in Multi-slice Helical CT SUN Wen-wu 1 ( ), CHEN Si-ping 2 ( ), ZHUANG Tian-ge

More information

An approximate cone beam reconstruction algorithm for gantry-tilted CT

An approximate cone beam reconstruction algorithm for gantry-tilted CT An approximate cone beam reconstruction algorithm for gantry-tilted CT Ming Yan a, Cishen Zhang ab, Hongzhu Liang a a School of Electrical & Electronic Engineering, Nanyang Technological University, Singapore;

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

Computational Algorithm for Estimating the Gradient Isocenter of an MRI Scanner

Computational Algorithm for Estimating the Gradient Isocenter of an MRI Scanner Computational Algorithm for Estimating the Gradient Isocenter of an MRI Scanner R Cai 1, KE Schubert 1, R Schulte 2 ritchiecai@r2labsorg, keith@r2labsorg, rschulte@dominionllumcedu 1 School of Computer

More information

AIDR 3D Iterative Reconstruction:

AIDR 3D Iterative Reconstruction: Iterative Reconstruction: Integrated, Automated and Adaptive Dose Reduction Erin Angel, PhD Manager, Clinical Sciences, CT Canon Medical Systems USA Iterative Reconstruction 1 Since the introduction of

More information

Estimating 3D Respiratory Motion from Orbiting Views

Estimating 3D Respiratory Motion from Orbiting Views Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng, Jeffrey A. Fessler, James M. Balter The University of Michigan Oct. 2005 Funding provided by NIH Grant P01 CA59827 Motivation Free-breathing

More information

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

C a t p h a n / T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 Internationally recognized for measuring the maximum obtainable performance of axial, spiral and multi-slice

More information

Micro-CT Methodology Hasan Alsaid, PhD

Micro-CT Methodology Hasan Alsaid, PhD Micro-CT Methodology Hasan Alsaid, PhD Preclinical & Translational Imaging LAS, PTS, GlaxoSmithKline 20 April 2015 Provide basic understanding of technical aspects of the micro-ct Statement: All procedures

More information

An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography

An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography 4th International Symposium on NDT in Aerospace 2012 - Tu.3.A.3 An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography Jonathan HESS *, Patrick KUEHNLEIN *, Steven OECKL

More information

Enhancement Image Quality of CT Using Single Slice Spiral Technique

Enhancement Image Quality of CT Using Single Slice Spiral Technique Enhancement Image Quality of CT Using Single Slice Spiral Technique Doaa. N. Al Sheack 1 and Dr.Mohammed H. Ali Al Hayani 2 1 2 Electronic and Communications Engineering Department College of Engineering,

More information

Performance Evaluation of CT Systems

Performance Evaluation of CT Systems 8/2/206 Performance Evaluation of CT Systems AAPM Task Group 233 Foundations, Methods, and Prospects Ehsan Samei Duke University Medical Center Disclosures Research grant: NIH R0 EB00838 Research grant:

More information

THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!

THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER! THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER! Confidence in complex treatments Modern radiation therapy uses complex plans with techniques such as IMRT, VMAT and Tomotherapy.

More information

LIGHT: Two-slit Interference

LIGHT: Two-slit Interference LIGHT: Two-slit Interference Objective: To study interference of light waves and verify the wave nature of light. Apparatus: Two red lasers (wavelength, λ = 633 nm); two orange lasers (λ = 612 nm); two

More information

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic

More information

Measurement of Skin Dose

Measurement of Skin Dose Measurement of Skin Dose Sources of Uncertainty Kenneth A. Fetterly, Ph.D. William Pavlicek, Ph.D. Dan Bednarek, PhD 2014 AAPM Annual Meeting, Austin Texas 2013 MFMER slide-1 Purpose 1. Present a framework

More information

An Investigation of a Model of Percentage Depth Dose for Irregularly Shaped Fields

An Investigation of a Model of Percentage Depth Dose for Irregularly Shaped Fields Int. J. Cancer (Radiat. Oncol. Invest): 96, 140 145 (2001) 2001 Wiley-Liss, Inc. Publication of the International Union Against Cancer An Investigation of a Model of Percentage Depth Dose for Irregularly

More information

8/2/2016. Acknowledgement. Common Clinical Questions. Presumption Images are Good Enough to accurately answer clinical questions

8/2/2016. Acknowledgement. Common Clinical Questions. Presumption Images are Good Enough to accurately answer clinical questions Image Quality Assessment using Model Observers: Clinical Implementation and Practical Considerations Shuai Leng, PhD Associate Professor, Department of Radiology Mayo Clinic, Rochester MN Acknowledgement

More information

Self-Calibration of Cone-Beam CT Using 3D-2D Image Registration

Self-Calibration of Cone-Beam CT Using 3D-2D Image Registration Self-Calibration of Cone-Beam CT Using 3D-2D Image Registration Sarah Ouadah 1 J. Webster Stayman 1 Grace Jianan Gang 1 Ali Uneri 2 Tina Ehtiati 3 Jeffrey H. Siewerdsen 1,2 1. Department of Biomedical

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

An approach for measuring the spatial orientations of a computed-tomography simulation system

An approach for measuring the spatial orientations of a computed-tomography simulation system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 2, 2014 An approach for measuring the spatial orientations of a computed-tomography simulation system Meng Chia Wu, a Ramani Ramaseshan Department

More information

AP Physics: Curved Mirrors and Lenses

AP Physics: Curved Mirrors and Lenses The Ray Model of Light Light often travels in straight lines. We represent light using rays, which are straight lines emanating from an object. This is an idealization, but is very useful for geometric

More information

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing ImPACT Information Leaflet No. 1: CT Scanner Acceptance Testing Version 1.02, 18/05/01 CONTENTS: 1. SCOPE OF LEAFLET 2. GENERAL PRINCIPLES OF ACCEPTANCE AND COMMISSIONING 2.1 PHANTOMS 2.2 EXPOSURE AND

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare

More information

USING cone-beam geometry with pinhole collimation,

USING cone-beam geometry with pinhole collimation, IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE 2009 687 A Backprojection-Based Parameter Estimation Technique for Skew-Slit Collimation Jacob A. Piatt, Student Member, IEEE, and Gengsheng L.

More information

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT CLINICAL COMMISSIONING AND ACCEPTANCE TESTING OF A 3D SURFACE MATCHING SYSTEM Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago Learning Objective Describe acceptance

More information

Quality control phantoms and protocol for a tomography system

Quality control phantoms and protocol for a tomography system Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing

More information

Moore Catholic High School Math Department

Moore Catholic High School Math Department Moore Catholic High School Math Department Geometry Vocabulary The following is a list of terms and properties which are necessary for success in a Geometry class. You will be tested on these terms during

More information

NIH Public Access Author Manuscript Med Phys. Author manuscript; available in PMC 2009 March 13.

NIH Public Access Author Manuscript Med Phys. Author manuscript; available in PMC 2009 March 13. NIH Public Access Author Manuscript Published in final edited form as: Med Phys. 2008 February ; 35(2): 660 663. Prior image constrained compressed sensing (PICCS): A method to accurately reconstruct dynamic

More information

Proton dose calculation algorithms and configuration data

Proton dose calculation algorithms and configuration data Proton dose calculation algorithms and configuration data Barbara Schaffner PTCOG 46 Educational workshop in Wanjie, 20. May 2007 VARIAN Medical Systems Agenda Broad beam algorithms Concept of pencil beam

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60601-2-44 2001 AMENDMENT 1 2002-09 Amendment 1 Medical electrical equipment Part 2-44: Particular requirements for the safety of X-ray equipment for computed tomography Amendement

More information

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings)

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings) Marquette University e-publications@marquette Biomedical Engineering Faculty Research and Publications Engineering, College of 1-1-1999 An Iterative Approach to the Beam Hardening Correction in Cone Beam

More information

Preparing for Part 2 of the ABR Diagnostic Physics Exam

Preparing for Part 2 of the ABR Diagnostic Physics Exam Preparing for Part 2 of the ABR Diagnostic Physics Exam Joseph Zambelli, Ph.D. Spectrum Health, Grand Rapids, MI Outline Exam Background Exam Preparation Suggested Resources Location, Time, and Format

More information

Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms.

Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms. Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms. D.Leone, A.Häußler Intitute for Nuclear Waste Disposal, Karlsruhe Institute for Technology,

More information

Light and the Properties of Reflection & Refraction

Light and the Properties of Reflection & Refraction Light and the Properties of Reflection & Refraction OBJECTIVE To study the imaging properties of a plane mirror. To prove the law of reflection from the previous imaging study. To study the refraction

More information

PHYS 219 General Physics: Electricity, Light and Modern Physics

PHYS 219 General Physics: Electricity, Light and Modern Physics PHYS 219 General Physics: Electricity, Light and Modern Physics Exam 2 is scheduled on Tuesday, March 26 @ 8 10 PM In Physics 114 It will cover four Chapters 21, 22, 23, and 24. Start reviewing lecture

More information

The effect of automatic exposure control with inappropriate scout image on radiation dose: a chest phantom experiment with three different CT machines

The effect of automatic exposure control with inappropriate scout image on radiation dose: a chest phantom experiment with three different CT machines The effect of automatic exposure control with inappropriate scout image on radiation dose: a chest phantom experiment with three different CT machines Poster No.: C-0891 Congress: ECR 2015 Type: Authors:

More information

Phys102 Lecture 21/22 Light: Reflection and Refraction

Phys102 Lecture 21/22 Light: Reflection and Refraction Phys102 Lecture 21/22 Light: Reflection and Refraction Key Points The Ray Model of Light Reflection and Mirrors Refraction, Snell s Law Total internal Reflection References 23-1,2,3,4,5,6. The Ray Model

More information

Correlation between Model and Human Observer Performance on a Lesion Shape Discrimination Task in CT

Correlation between Model and Human Observer Performance on a Lesion Shape Discrimination Task in CT Correlation between Model and Human Observer Performance on a Lesion Shape Discrimination Task in CT Yi Zhang, Shuai Leng, Lifeng Yu and Cynthia McCollough Department of Radiology Mayo Clinic, Rochester

More information

Advanced Image Reconstruction Methods for Photoacoustic Tomography

Advanced Image Reconstruction Methods for Photoacoustic Tomography Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic

More information

Limits of Dose Reduction in CT: How low is too low? Acknowledgement. Disclosures 8/2/2012

Limits of Dose Reduction in CT: How low is too low? Acknowledgement. Disclosures 8/2/2012 8/2/22 Limits of Dose Reduction in CT: How low is too low? Ehsan Samei Duke University Acknowledgement Disclosures Research grant: NIH Research grant: DHS Research grant: RSNA, QIBA Research grant: General

More information

Chapter 3 Image Registration. Chapter 3 Image Registration

Chapter 3 Image Registration. Chapter 3 Image Registration Chapter 3 Image Registration Distributed Algorithms for Introduction (1) Definition: Image Registration Input: 2 images of the same scene but taken from different perspectives Goal: Identify transformation

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc.

Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc. Unique Features of the GE Senoclaire Tomosynthesis System Tyler Fisher, M.S., DABR Therapy Physics, Inc. Conflict of Interest Disclosure I have no conflicts to disclose. Learning Objectives Overview of

More information

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory Introduction In this lab exercise, you will investigate the linearity of the DeskCAT scanner by making measurements

More information

THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS

THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS A. Campos a*, A. Zorrilla a, F. Lasagni a a Materials & Processes dept., Center for Advanced Aerospace Technologies

More information

Essential Physics I. Lecture 13:

Essential Physics I. Lecture 13: Essential Physics I E I Lecture 13: 11-07-16 Reminders No lecture: Monday 18th July (holiday) Essay due: Monday 25th July, 4:30 pm 2 weeks!! Exam: Monday 1st August, 4:30 pm Announcements 250 word essay

More information

Scaling Calibration in the ATRACT Algorithm

Scaling Calibration in the ATRACT Algorithm Scaling Calibration in the ATRACT Algorithm Yan Xia 1, Andreas Maier 1, Frank Dennerlein 2, Hannes G. Hofmann 1, Joachim Hornegger 1,3 1 Pattern Recognition Lab (LME), Friedrich-Alexander-University Erlangen-Nuremberg,

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

RADIOLOGY AND DIAGNOSTIC IMAGING

RADIOLOGY AND DIAGNOSTIC IMAGING Day 2 part 2 RADIOLOGY AND DIAGNOSTIC IMAGING Dr hab. Zbigniew Serafin, MD, PhD serafin@cm.umk.pl 2 3 4 5 CT technique CT technique 6 CT system Kanal K: RSNA/AAPM web module: CT Systems & CT Image Quality

More information

Fujifilm DR Solution. FDR AcSelerate. The new pinnacle in diagnostic imaging from Fujifilm ISS. CsI. Dynamic Visualization. Technology.

Fujifilm DR Solution. FDR AcSelerate. The new pinnacle in diagnostic imaging from Fujifilm ISS. CsI. Dynamic Visualization. Technology. Fujifilm DR Solution FDR AcSelerate The new pinnacle in diagnostic imaging from Fujifilm CsI Scintillator ISS Technology Dynamic Visualization Welcome to the X-ray room of the future! A streamlined solution

More information

International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN:

International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: COMPARISON OF RADIATION DOSE AND IMAGE QUALITY BETWEEN THE SINGLE DETECTOR CT AND MULTI DETECTOR CT ON HEAD EXAMINATION: STUDY PHANTOM Samburi Master of Physics, University of Diponegoro, Indonesia Agency

More information

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti Monte Carlo methods in proton beam radiation therapy Harald Paganetti Introduction: Proton Physics Electromagnetic energy loss of protons Distal distribution Dose [%] 120 100 80 60 40 p e p Ionization

More information

MOC Update for Medical Physicists ABR MOC Medical Physics 3/14/2017 Disclosures NOTE Maintenance of Certification Why do we have MOC?

MOC Update for Medical Physicists ABR MOC Medical Physics 3/14/2017 Disclosures NOTE Maintenance of Certification Why do we have MOC? Disclosures MOC Update for Medical Physicists AAPM Spring Clinical Meeting I am a part time paid employee of the American Board of Radiology G. Donald Frey Associate Executive Director American Board of

More information

Clinical implementation of photon beam flatness measurements to verify beam quality

Clinical implementation of photon beam flatness measurements to verify beam quality JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Clinical implementation of photon beam flatness measurements to verify beam quality Simon Goodall, a Nicholas Harding, Jake Simpson,

More information

Photon beam dose distributions in 2D

Photon beam dose distributions in 2D Photon beam dose distributions in 2D Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2014 Acknowledgments! Narayan Sahoo PhD! Richard G Lane (Late) PhD 1 Overview! Evaluation

More information

IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques

IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques Report of the Coordinated Research Project (CRP) on Development of Procedures

More information

Noise power spectrum and modulation transfer function analysis of breast tomosynthesis imaging

Noise power spectrum and modulation transfer function analysis of breast tomosynthesis imaging Noise power spectrum and modulation transfer function analysis of breast tomosynthesis imaging Weihua Zhou a, Linlin Cong b, Xin Qian c, Yueh Z. Lee d, Jianping Lu c,e, Otto Zhou c,e, *Ying Chen a,b a

More information

Basic Radiation Oncology Physics

Basic Radiation Oncology Physics Basic Radiation Oncology Physics T. Ganesh, Ph.D., DABR Chief Medical Physicist Fortis Memorial Research Institute Gurgaon Acknowledgment: I gratefully acknowledge the IAEA resources of teaching slides

More information

Thank-You Members of TG147 TG 147: QA for nonradiographic

Thank-You Members of TG147 TG 147: QA for nonradiographic Thank-You Members of TG147 TG 147: QA for nonradiographic localization and positioning systems Twyla Willoughby, M.S. Medical Physicist Clinical AAPM Meeting March 2013 Department of Radiation Oncology

More information

Determination of rotations in three dimensions using two-dimensional portal image registration

Determination of rotations in three dimensions using two-dimensional portal image registration Determination of rotations in three dimensions using two-dimensional portal image registration Anthony E. Lujan, a) James M. Balter, and Randall K. Ten Haken Department of Nuclear Engineering and Radiological

More information

Adaptive algebraic reconstruction technique

Adaptive algebraic reconstruction technique Adaptive algebraic reconstruction technique Wenkai Lua) Department of Automation, Key State Lab of Intelligent Technology and System, Tsinghua University, Beijing 10084, People s Republic of China Fang-Fang

More information

Optimisation of Toshiba Aquilion ONE Volume Imaging

Optimisation of Toshiba Aquilion ONE Volume Imaging Optimisation of Toshiba Aquilion ONE Volume Imaging Jane Edwards, RPRSG Royal Free London NHS Foundation Trust Dr Mufudzi Maviki, Plymouth Hospitals NHS Trust Background In 2011/12 Radiology at RFH was

More information

A simple method to test geometrical reliability of digital reconstructed radiograph (DRR)

A simple method to test geometrical reliability of digital reconstructed radiograph (DRR) JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 1, WINTER 2010 A simple method to test geometrical reliability of digital reconstructed radiograph (DRR) Stefania Pallotta, a Marta Bucciolini

More information

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Stefan C. Saur 1, Hatem Alkadhi 2, Luca Regazzoni 1, Simon Eugster 1, Gábor Székely 1, Philippe Cattin 1,3 1 Computer Vision

More information

Lorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc.

Lorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc. Lorad FFDM QC Procedures for Medical Physicist Tao Wu, Ph.D. Hologic, Inc. Lorad Selenia FFDM Tests Following Lorad QC Manual Collimation Assessment Artifact Evaluation System Resolution Phantom Image

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice

Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice Guido Hilgers, a Tonnis Nuver, André Minken Department

More information

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Wenjuan Sun and Stephen Brown 10 th May 2016 1 Possible factors influencing XCT measurements Components Influencing variables Possible

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Time: 3 hour Total Marks: 90

Time: 3 hour Total Marks: 90 Time: 3 hour Total Marks: 90 General Instructions: 1. All questions are compulsory. 2. The question paper consists of 34 questions divided into four sections A, B, C, and D. 3. Section A contains of 8

More information

A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY

A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY BACKSCATTER DATA Young S. Ham and C.F. Poranski Materials Chemistry Branch Naval Research Laboratory Washington DC 20375 E.C. Greenawald Geo-Centers,

More information

C quantities. The usual calibration method consists of a direct measurement of. Calibration of Storage Tanks B. SHMUTTER U. ETROG

C quantities. The usual calibration method consists of a direct measurement of. Calibration of Storage Tanks B. SHMUTTER U. ETROG B. SHMUTTER U. ETROG Geodetic Research Station Technion, Haifa, Israel Calibration of Storage Tanks Adequate accuracy is obtained in the application of terrestrial photographs. (Abstract on next page)

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

1. Learn to incorporate QA for surface imaging

1. Learn to incorporate QA for surface imaging Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago ***No disclosures*** 1. Learn to incorporate QA for surface imaging into current QA procedures for IGRT. 2. Understand

More information

Nicholas J. Giordano. Chapter 24. Geometrical Optics. Marilyn Akins, PhD Broome Community College

Nicholas J. Giordano.   Chapter 24. Geometrical Optics. Marilyn Akins, PhD Broome Community College Nicholas J. Giordano www.cengage.com/physics/giordano Chapter 24 Geometrical Optics Marilyn Akins, PhD Broome Community College Optics The study of light is called optics Some highlights in the history

More information