DEVELOPMENT AND CHARACTERISATION OF A HEAD CALIBRATION PHANTOM FOR IN VIVO MEASUREMENTS OF ACTINIDES

Size: px
Start display at page:

Download "DEVELOPMENT AND CHARACTERISATION OF A HEAD CALIBRATION PHANTOM FOR IN VIVO MEASUREMENTS OF ACTINIDES"

Transcription

1 DEVELOPMENT AND CHARACTERISATION OF A HEAD CALIBRATION PHANTOM FOR IN VIVO MEASUREMENTS OF ACTINIDES 1 G. Gualdrini, 1 P. Battisti, 2 R. Biagini, 2 P. De Felice, 2 A. Fazio, 3 P. Ferrari 1; Ente Nazionale per le Nuove Tecnologie, l Energia e l Ambiente - Istituto per la Radioprotezione v. Dei Colli 16 I Bologna (Italy) 2; Ente Nazionale per le Nuove Tecnologie, l Energia e l Ambiente - Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti v. Anguillarese, 301, I S. M. Galeria-Roma (Italy) 3; Degree in Physics, Physics Department-University of Bologna Viale C. Berti Pichat 6/2 I Bologna (Italy) Abstract: The investigation of actinides internal contamination in human body makes use of a variety of techniques. In large scale screening procedures the in vivo measurement of bone 241 Am burden via the determination of the nuclide activity in the skull is often used. To this purpose adequate calibration procedures and standard phantoms are needed. The present paper summarises the studies and technical procedures followed for the development of a calibration phantom based on a commercial Alderson angiographic head in which a set of Am point sources were imbedded. A theoretical study was first carried out, at the ENEA Institute for Radiation Protection, using the MCNP4-B Monte Carlo code to determine the point source distribution that closer approximate a homogeneous bone contamination. The numerical models were also used to evaluate the resulting degree of approximation. The point sources were prepared at the ENEA National Metrology Institute for ionising radiation quantities and are traceable to the Italian national standard of radionuclide activity. The sources were prepared quantitatively dispensing a liquid solution onto a plastic film. The activity of each source was checked by gamma-ray spectrometry and the reproducibility of the activity values was determined. Each source was then placed in the optimum position, given by the Monte Carlo modelling, by a precision mechanical device. The phantom was finally used to calibrate a counting system operating at the ENEA Institute for Radiation Protection.

2 The paper reports the main theoretical and experimental aspects of this work also discussing the results of the first calibrations. 1-INTRODUCTION The evaluation of 241 Am bone burden through in vivo measurement of the nuclide activity in the skull is one of the most used techniques in large scale screening of actinides previous internal contamination [1]. Taking into account that, for radiation protection purposes, very high sensitivity, accuracy and precision are requested in this kind of measurements, the major problem is often due to the difficulty to develop suitable standards for the calibration of the detection system. The present paper summarises the studies and the technical procedure followed for the development and characterisation of a head calibration phantom [2] for in vivo measurement of 241 Am based on a commercial one, normally employed for angiographic analysis calibration. The main purpose was to develop a phantom in which a limited number of point sources should well approximate the homogeneous distribution typical of an Americium internal contamination [3]. 2- PREPARATION OF THE HEAD PHANTOM A selection of commercial head phantoms was made, in order to find the one with the best anthropomorphic characteristics. The Alderson phantom for angiographic analysis was chosen (figure 1). Its constituents (as kindly supplied by the manufacturer [4] ) are: Alderson TM cortical bone [ρ= 1.88 (g/cm 3 ); 60 kev µ= (cm 1 )] Alderson TM trabecular bone [ρ= 1.17 (g/cm 3 ); 60 kev µ= (cm 1 )] Alderson TM soft tissue [ρ= 1.09 (g/cm 3 ); 60 kev µ= (cm 1 )]

3 Fig. 1: Lateral view of the Alderson TM phantom The following requirements were taken into account in the preparation and characterisation of the radioactive sources: small volume (quasi point-) of radioactive material, homogeneous activity values, precision source positioning inside the head phantom. The point source geometry was chosen as it minimises the source volume and source selfabsorption. The point sources were prepared depositing 5 µl of a liquid 241 Am standard solution (0.5M HCl, containing, at 10 March 1999, kbq/g of 241 Am, with a combined standard uncertainty of 1.5%) onto 5 mm diameter and 1 mm thick polystyrene disks and evaporating the solution under infrared lamp. The standard solution was calibrated by the ENEA well-type ionisation chamber, and was therefore traceable to the Italian national standard of radionuclide activity. The diameter of the dried radioactive deposit was about 1.5 mm. The active disks were then covered by similar inactive disks with a double adhesive paper in between the two disks. A first set of standard point sources were prepared by quantitative deposition of the radioactive solution by a polyethylene picnometer. This method allows an accurate determination (combined standard uncertainty of the mass measurements of 0.05%) of the deposited solution mass. By this method a number of 35 standard sources were prepared with average deposited solution mass of 5.4 mg and experimental standard deviation of 26%. The source activities ranged from 94.4 to Bq. This was not considered satisfactory with respect to the reproducibility needs. Each source was also counted on a gamma-ray spectrometry system, based on a p-type HPGe coaxial detector with

4 15% relative efficiency, on top of the detector window, for a counting time of 1000 s. The corresponding full-energy peak efficiency values showed an experimental standard deviation of 2.7%. A comparison of this value with the net peak area standard uncertainty based on counting statistics (2.2%), limited other important sources of variability. A second set of sources was then prepared. In order to obtain more reproducible source activity values, a Pipetman P20 Gilson pipette was used instead of the plastic picnometer. A constant volume of 5 µl was set and polypropylene tips were replaced after each single deposition as droplets frequently remained on the end of the tip. A number of 19 sources were prepared by this second procedure and gamma counted, as described before, in order to get the activity on the base of the average efficiency value previously determined. An average activity of Bq with experimental standard deviation of 2.9% was obtained. This is 4.3% lower than the activity value (128.4 Bq) calculated on the base of the activity concentration of the starting solution and its delivered volume. This discrepancy was only in part explained with the accuracy of the pipette (2% declared by the constructor). As a conclusion of this analysis it was decided to group the two sets of point sources and to rely the source characterisation on the spectrometric measurements. The frequency distribution of the measured activity values for the 54 (35+19) sources is reported in figure 2. Number of sources Range of selected sources (b) Activity (Bq) Figure 2: Frequency distribution of the activity values of the 54 standard point sources. The activity values are those obtained from gamma-ray spectrometry measurements.

5 The 24 sources needed for the phantom were then selected among all the sources available in the frequency peak between 115 and 135 Bq (figure 2) with an average activity of Bq and experimental standard deviation of 2.7%. Of the 24 sources selected, 6 were taken from the set obtained by the picnometer method and the remaining from that obtained by the pipette method. The calibration certificate of the point sources reported the activity value of each source, as obtained by the spectrometric measurement, with its associated uncertainty. The main uncertainty components were (relative standard uncertainty): 0.5% (dead time), 2.2% (counting statistics), 1.5% ( counting efficiency), giving a combined standard uncertainty of 2.7%. The head phantom was mounted on a spherical co-ordinates rotating base that was positioned onto the working plane of a precision milling machine by which a number of 24 cylindrical holes were made on the head phantom. The head position and the hole depth were chosen to reach the bottom of each hole at the co-ordinates given by the theoretical modelling (described later). As far as possible the drilling direction was set orthogonal to the local head surface. Each source was then mounted on a plastic piston that was inserted into the corresponding hole whose protruding part was finally cut. 3- CHARACTERISATION OF THE HEAD PHANTOM The main purpose of the work was to develop a head calibration phantom in which a heterogeneous distribution of 24 nearly identical point sources should approximate the homogeneous distribution that Americium really has in the bone. The two main steps of the procedure were Monte Carlo modelling of the plastic head and optimisation of source placing. 3-1 Monte Carlo modelling of the plastic head The MCNP (version 4B) [5] Monte Carlo code was used. This is a powerful 3-dimensional multipurpose code developed at the Los Alamos National Laboratory. It can solve transport problems of

6 neutrons, photons and electrons in a fully coupled way (i.e. it is possible to follow the secondary particles created during interactions). A powerful geometric package, based on boolean algebra operators, allows to define 3-D complex geometries using the first, second, fourth (torii) order surfaces that MCNP handles. Effective variance reduction techniques and a detailed statistical package make MCNP a reliable radiation transport code widely renowned for its many applications. The Monte Carlo model of the experimental phantom was the most critical point of the procedure. It implied the knowledge of the bone mass distribution throughout the skull. The problem was in fact to determine 24 nearly equal volume skull zones where to place the sources that had to be representative of the radioactive content of the stated bone volume. A first implemented mathematical model, based on experimental measurements of bone thicknesses and weights of a didactic skull phantom, put in evidence that a more rigorous technique was needed to model the Alderson TM phantom. Using a Computed Tomography (Ge Hi Speed Adv RP) a number of 248, 1 mm thick, phantom axial slices were obtained. Only a subset of them were used for the production of the 3-dimensional MCNP model of the skull, increasing the number of the slices, mathematically treated, when quick geometry variations, along the vertical axis, needed more details. Each slice of the subset was studied in order to determine which mathematical equations, among those handled by MCNP, could better fit the CT contours. An example of a CT slices together with its MCNP representation is shown in figure 3. The final geometry is made of 54 layers, 277 MCNP cells and 1929 surfaces. In figure 4 a SABRINA [6] 3-dimensional plot of the complete skull is shown. The capability of the MCNP code to evaluate spatial cell volumes was employed to determine the 241 Am sources positions. In fact each of the 24 sources should have been placed in the centroid of 24 equal volume sub-regions to obtain a nearly homogeneous distribution of the radioactivity.

7 Fig.3: A CT scan of head calibration phantom (left), at cheek height, and its MCNP representation 3-2 Optimisation of source positioning The positions chosen in the distribution of point sources strongly influence the counting system efficiency. As a matter of fact the closer to the head surface is a detector, the easier the condition of a homogeneous distribution can fail. For each measurement position a correction factor between practical heterogeneous and theoretical homogeneous configuration was defined through the Monte Carlo modelling of the experiment.

8 Fig.4: SABRINA plot of the MCNP Alderson TM phantom with soft tissue transparent MCNP allowed to obtain an azimuthal-polar mapping of the photon fluence within the detector ROI (Region Of Interest) in the closeness of the upper head surface. This fluence mapping was analysed at various distances from the phantom and allowed to: Optimise the two Ge detector positions for the calibration; Determine the behaviour of the correction factor for the fluence as a function of azimuthal-polar angle and of the distance from the head surface. The procedure adopted can be briefly summarised as follows: A quarter of a hollow sphere was modelled in the top front area of the head (figure 5); It is subdivided in 15 azimuthal and polar angles; Each of so defined directions represents a possible scoring area for the detector. This allowed subdividing the response (photon fluence contributing to the total absorption peak) in 13x6+1 circular areas of the same size of the detector window; The same method was applied both for homogeneous source distribution and for the 24 sources distribution; The results obtained were graphically compared (see figure 6) in which the 6 curves (from left to right) represent the increasing polar angle and the 13 points on each curve represent the ratio

9 between the two fluences (homogeneous / 24 points) calculated for that particular azimuthal angle. The study was repeated for different distances from the head surface. Fig.5 Sketch of the response spherical segment on the top of the head 1,35 1,3 1,25 1,2 1,15 1,1 1,05 1 0,95 0,9 0,85 0,8 0,75 ratio hom / 24-points about 1cm from head surface ratio hom / 24-points about 3cm from head surface Detectors polar-azimuthal distribution (detectors labels) Fig.6: Behaviour of the fluence ratios in the 79 detection positions.

10 Some conclusions can be drawn from these Monte Carlo calculations: The agreement between the two models (fully homogeneous and 24 sources heterogeneous) is bounded within maximum ± 20% The agreement is within ± 10% for higher polar angles. The agreement is improved, as expected, at higher distances from the skull surface 4-CALIBRATION MEASUREMENTS The measurement tests on the head phantom were carried out using the Whole Body Counter (WBC) operating at the ENEA Institute for Radiation Protection, in the Casaccia research centre. A detection system for low energy photons, equipped with 2 Canberra LEGe detectors (each one with active area of 38 cm 2 ), was used. The shieldeding is made (from external to internal) of 50 cm concrete, 20 cm iron, 3mm lead, 2 mm copper and 1 mm cadmium. This system is frequently used for in vivo monitoring of actinides. The detector position, with respect to the head phantom, was chosen as a compromise between the need of minimising the photon fluence variation due to the heterogeneous source distribution and that of reaching a Minimum Detectable Activity (MDA) [7] of the order of 3-4 Bq in the skull, in a reasonable measuring time (20-30 minutes). Head surface to detector distance (cm) MCNP (homogeneous source) Cps/Bq MCNP (24 points heterogeneous source) Cps/Bq WBC Cps/Bq ~ 1 (6.70± 0.01)10-3 (7.54± 0.01) 10-3 (8.7± 0.1) 10-3 ~ 10 (2.10± 0.01) 10-3 (2.21± 0.01) 10-3 (2.48± 0.03) 10-3 ~ 15 (1.37± 0.01) 10-3 (1.40± 0.01) 10-3 (1.41± 0.02) 10-3 ~ 20 (0.953± 0.003) 10-3 (0.969± 0.003) 10-3 (0.95± 0.01) 10-3

11 Head surface to detector distance (cm) MCNP homogeneous WBC measur. ratio MCNP heterogeneous WBC measur. ratio ~ ~ ~ ~ MCNP Homogeneous-heterogeneous correction factor Table1 and Table 2: MCNP and experimental results for some detector configurations Analysing the results achieved through the fluence mapping, it was decided to work in the neighbourhood of 60 and 75 polar angles. Three different evaluations of full energy peak efficiency for the 59.5 kev 241 Am photon emission, were performed at different distances of the detector from the phantom surface by: Monte Carlo calculation simulating the phantom experimental conditions; Monte Carlo calculation simulating the real man contamination; experimental WBC measurements on the phantom. A pulse height estimator was used in MCNP to evaluate the pulse number distribution as a function of the photon energy. The particles were scored only within the detector ROI, killing the photons below its lower limit, to reduce the computing time per tracked particle. The results are reported in tables 1 and 2. As expected the agreement among efficiency values obtained by the two Monte Carlo simulations (24 points-source, and homogeneous source) and the WBC measurements, progressively improves with the increase of the measurement distance, reaching a nearly perfect agreement at cm. Taking into account that the required values of MDA (3-4 Bq of 241 Am in the skull) have been reached only with a 30 minutes measurement and with a distance of 1 cm between detector and

12 head surface, the analysis of the results suggests the adoption of the following calibration procedure: 1- Perform the calibration measurements at 15 cm from the head surface of the phantom, in order to guarantee the accuracy, reducing the non-uniformity effects of 24 points distribution of sources. 2- Perform the in vivo measurement at 1 cm from the skull surface in order to reduce the counting time increasing detection efficiency. 3- Apply the proper correction coefficient CF for the calibration factor given by: CF = CF D x CF H where: CF d = correction factor for different head to detector distance, for the homogeneous source, obtained from two MCNP calculation at measurement (1 cm) and at calibration (15 cm) distance, for the selected detection position. CF H = correction factor for source heterogeneity, calculated comparing MCNP results, at calibration distance of 15 cm, for the homogeneous and heterogeneous source distribution, for the same detection position. In the adopted measurement configuration (15 cm distance), CF d = 4.89 ± 0.1 and CF H = 0.99 ±0.02 (Monte Carlo two standard errors) and therefore CF = 4.8 ± 0.5 (95% confidence interval including both the Monte Carlo precision and the MCNP complex head model accuracy that could be quantified within ± 10%). CONCLUSIONS The procedure here presented was used to develop a calibration phantom in which a heterogeneous distribution of 241 Am sources well approximates a homogeneous distribution of the nuclide in the bone. The Monte Carlo code MCNP was used to quantify the level of approximation introduced and to evaluate correction factors for detection position and source heterogeneity.

13 The comparison with the experimental measurements confirmed the reliability of the procedure, that could be adopted as a standard for the design of head calibration phantoms. Possible refinements of the methodology could be based on a larger number of point sources to be embedded in the head plastic matrix and a Monte Carlo model of the head based on a VOXEL phantom directly obtained from the CT scan Hounsfield raw data. Acknowledgement: the authors are grateful to Dr.L.Pierotti (Health Physics Service, S.Orsola Hospital, Bologna) and Dr. G.Turci (Radiology Service Canini, S.Orsola Hospital, Bologna) for the CT scan analyses. REFERENCES [1] Current status of the whole body counting as means to detect and quantify previous exposure to radioactive materials Health Ph. Vol 60. Supp. 1, pp 7-42, [2] Phantoms and Computational Models in Therapy, Diagnosis and Protection, ICRU report 48, [3] The Metabolism of Plutonium and Related Elements, Annals of ICRP publication 48, Vol. 16 o. 2/3 Pergamon Press [4] Alderson Private Communication. [5] J. F. Bresmeister (editor) MCNP A general Monte Carlo N-Particle Transport Code Version 4B. LA Los Alamos National Laboratory (March 1997). [6] Van Riper K. Sabrina User s Guide. White Rock Science, [7] Currie L.A. Limits for qualitative detection and quantitative determination: application to radiochemistry Analytical Chemistry, Vol.40, N.3, March 1998.

Improved Detector Response Characterization Method in ISOCS and LabSOCS

Improved Detector Response Characterization Method in ISOCS and LabSOCS P Improved Detector Response Characterization Method in ISOCS and LabSOCS *1 1 1 1 1 R. VenkataramanP P, F. BronsonP P, V. AtrashkevichP P, M. FieldP P, and B.M. YoungP P. 1 PCanberra Industries, 800 Research

More information

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry Sasha A. Philips, Frazier Bronson, Ram Venkataraman, Brian M. Young Abstract--Activity measurements require knowledge of the

More information

Monte Carlo Simulations for In Vivo Internal Dosimetry (including phantom development)

Monte Carlo Simulations for In Vivo Internal Dosimetry (including phantom development) Monte Carlo Simulations for In Vivo Internal Dosimetry (including phantom development) Gianfranco Gualdrini ENEA - Istituto di Radioprotezione Bologna - Italy IRPA11 Madrid - Spain 23-28 May 2004 Quantities

More information

Michael Speiser, Ph.D.

Michael Speiser, Ph.D. IMPROVED CT-BASED VOXEL PHANTOM GENERATION FOR MCNP MONTE CARLO Michael Speiser, Ph.D. Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX September 1 st, 2012 CMPWG Workshop Medical

More information

Application of MCNP Code in Shielding Design for Radioactive Sources

Application of MCNP Code in Shielding Design for Radioactive Sources Application of MCNP Code in Shielding Design for Radioactive Sources Ibrahim A. Alrammah Abstract This paper presents three tasks: Task 1 explores: the detected number of as a function of polythene moderator

More information

MODELLING RADIATION TRANSPORT IN VOXEL GEOMETRIES WITH THE MONTE CARLO CODE MCNPX

MODELLING RADIATION TRANSPORT IN VOXEL GEOMETRIES WITH THE MONTE CARLO CODE MCNPX MODELLING RADIATION TRANSPORT IN VOXEL GEOMETRIES WITH THE MONTE CARLO CODE MCNPX G. Gualdrini 1, C. Daffara 2, K.W. Burn 3, M. Pierantoni 4, P. Ferrari 1 1 ENEA - ION-IRP, V. dei Colli 16, 40136 Bologna

More information

IMPLEMENTATION OF SALIVARY GLANDS IN THE BODYBUILDER ANTHROPOMORPHIC PHANTOMS

IMPLEMENTATION OF SALIVARY GLANDS IN THE BODYBUILDER ANTHROPOMORPHIC PHANTOMS Computational Medical Physics Working Group Workshop II, Sep 30 Oct 3, 2007 University of Florida (UF), Gainesville, Florida USA on CD-ROM, American Nuclear Society, LaGrange Park, IL (2007) IMPLEMENTATION

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!

More information

UNCOMPROMISING QUALITY

UNCOMPROMISING QUALITY ION CHAMBERS UNCOMPROMISING QUALITY Designed with over 30 years of scientific integrity for a broad range of dosimetry measurements in diverse radiation beams Farmer-type Chambers For absolute dosimetry

More information

NKS GammaUser

NKS GammaUser Uncertainties Marie-Christine Lépy Laboratoire National Henri Becquerel CEA Saclay, F-91191 Gif-sur-Yvette cedex, France E-mail: marie-christine.lepy@cea.fr NKS GammaUser 2014 Helsinki 6-8 October 2014

More information

Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations

Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations A. Al-Basheer, G. Sjoden, M. Ghita Computational Medical Physics Team Nuclear & Radiological Engineering University

More information

CT vs. VolumeScope: image quality and dose comparison

CT vs. VolumeScope: image quality and dose comparison CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,

More information

Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms.

Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms. Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms. D.Leone, A.Häußler Intitute for Nuclear Waste Disposal, Karlsruhe Institute for Technology,

More information

An Investigation into the Limitations of Tomographic Scanning for Quantitative Nondestructive Gamma-ray Measurements of High Density Radioactive Waste

An Investigation into the Limitations of Tomographic Scanning for Quantitative Nondestructive Gamma-ray Measurements of High Density Radioactive Waste An Investigation into the Limitations of Tomographic Scanning for Quantitative Nondestructive Gamma-ray Measurements of High Density Radioactive Waste S.C. Kane 1, S. Croft 1, P. McClay 1, W.F. Mueller

More information

Deliverable D10.2. WP10 JRA04 INDESYS Innovative solutions for nuclear physics detectors

Deliverable D10.2. WP10 JRA04 INDESYS Innovative solutions for nuclear physics detectors MS116 Characterization of light production, propagation and collection for both organic and inorganic scintillators D10.2 R&D on new and existing scintillation materials: Report on the light production,

More information

Voxel phantoms and Monte Carlo methods applied to internal and external dose calculations.

Voxel phantoms and Monte Carlo methods applied to internal and external dose calculations. Voxel phantoms and Monte Carlo methods applied to internal and external dose calculations. J.G.Hunt, E.G. Cavalcanti, D.S. dos Santos e A.M.G. Azeredo. Instituto de Radioproteção e Dosimetria IRD/CNEN

More information

Total Measurement Uncertainty Estimation For Tomographic Gamma Scanner R.Venkataraman 1, S.Croft 1, M.Villani 1, R.D. McElroy 1, and R.J.

Total Measurement Uncertainty Estimation For Tomographic Gamma Scanner R.Venkataraman 1, S.Croft 1, M.Villani 1, R.D. McElroy 1, and R.J. Total Measurement Uncertainty Estimation For Tomographic Gamma Scanner R.Venkataraman 1, S.Croft 1, M.Villani 1, R.D. McElroy 1, and R.J. Estep 2 1 Canberra Industries, Inc., 800 Research Parkway, Meriden,

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

Evaluation of RayXpert for shielding design of medical facilities

Evaluation of RayXpert for shielding design of medical facilities Evaluation of Raypert for shielding design of medical facilities Sylvie Derreumaux 1,*, Sophie Vecchiola 1, Thomas Geoffray 2, and Cécile Etard 1 1 Institut for radiation protection and nuclear safety,

More information

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information

arxiv: v2 [cond-mat.mtrl-sci] 5 Jan 2010

arxiv: v2 [cond-mat.mtrl-sci] 5 Jan 2010 Gamma scattering scanning of concrete block for detection of voids. Shivaramu 1, Arijit Bose 2 and M. Margret 1 1 Radiological Safety Division, Safety Group, IGCAR, Kalpakaam - 63 12 (India) 2 Chennai

More information

Investigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography

Investigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography Investigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography Natalia Grozmani 1, Andrea Buratti 1, Robert H. Schmitt 1 More info about this article:

More information

Modeling the ORTEC EX-100 Detector using MCNP

Modeling the ORTEC EX-100 Detector using MCNP Modeling the ORTEC EX-100 Detector using MCNP MCNP is a general-purpose Monte Carlo radiation transport code for modeling the interaction of radiation with materials based on composition and density. MCNP

More information

Limitations in the PHOTON Monte Carlo gamma transport code

Limitations in the PHOTON Monte Carlo gamma transport code Nuclear Instruments and Methods in Physics Research A 480 (2002) 729 733 Limitations in the PHOTON Monte Carlo gamma transport code I. Orion a, L. Wielopolski b, * a St. Luke s/roosevelt Hospital, Columbia

More information

ANALYSIS OF CT AND PET/SPECT IMAGES FOR DOSIMETRY CALCULATION

ANALYSIS OF CT AND PET/SPECT IMAGES FOR DOSIMETRY CALCULATION 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 ANALYSIS OF

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

Improvement in the calibration techniques of approved dosimetry services in Spanish nuclear power plants of whole body counters with NaI detectors.

Improvement in the calibration techniques of approved dosimetry services in Spanish nuclear power plants of whole body counters with NaI detectors. Improvement in the calibration techniques of approved dosimetry services in Spanish nuclear power plants of whole body counters with NaI detectors. E.Sollet (*), A.Felipe(**), JM.Perez (**), S. De Maria

More information

Shielding factors for traditional safety glasses

Shielding factors for traditional safety glasses Shielding factors for traditional safety glasses Malcolm McEwen, Hong Shen and Ernesto Mainegra-Hing Ionizing Radiation Standards, National Research Council Canada Alan DuSautoy, Radiation and Health Sciences

More information

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation

More information

ELECTRON DOSE KERNELS TO ACCOUNT FOR SECONDARY PARTICLE TRANSPORT IN DETERMINISTIC SIMULATIONS

ELECTRON DOSE KERNELS TO ACCOUNT FOR SECONDARY PARTICLE TRANSPORT IN DETERMINISTIC SIMULATIONS Computational Medical Physics Working Group Workshop II, Sep 30 Oct 3, 2007 University of Florida (UF), Gainesville, Florida USA on CD-ROM, American Nuclear Society, LaGrange Park, IL (2007) ELECTRON DOSE

More information

NUMERICAL METHOD TO ESTIMATE TOLERANCES COMBINED EFFECTS ON A MECHANICAL SYSTEM

NUMERICAL METHOD TO ESTIMATE TOLERANCES COMBINED EFFECTS ON A MECHANICAL SYSTEM INTERNATIONAL DESIGN CONFERENCE - DESIGN 2004 Dubrovnik, May 18-21, 2004. NUMERICAL METHOD TO ESTIMATE TOLERANCES COMBINED EFFECTS ON A MECHANICAL SYSTEM D. Cambiaghi, A. Magalinia and D. Vetturi Keywords:

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory Introduction In this lab exercise, you will investigate the linearity of the DeskCAT scanner by making measurements

More information

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling nd International Symposium on NDT in Aerospace - We..A. Comparison of Probing Error in Dimensional Measurement by Means of D Computed Tomography with Circular and Helical Sampling Jochen HILLER, Stefan

More information

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti Monte Carlo methods in proton beam radiation therapy Harald Paganetti Introduction: Proton Physics Electromagnetic energy loss of protons Distal distribution Dose [%] 120 100 80 60 40 p e p Ionization

More information

SHIELDING DEPTH DETERMINATION OF COBALT PHOTON SHOWER THROUGH LEAD, ALUMINUM AND AIR USING MONTE CARLO SIMULATION

SHIELDING DEPTH DETERMINATION OF COBALT PHOTON SHOWER THROUGH LEAD, ALUMINUM AND AIR USING MONTE CARLO SIMULATION Research Article SHIELDING DEPTH DETERMINATION OF COBALT PHOTON SHOWER THROUGH LEAD, ALUMINUM AND AIR USING MONTE CARLO SIMULATION 1 Ngadda, Y. H., 2 Ewa, I. O. B. and 3 Chagok, N. M. D. 1 Physics Department,

More information

Monte Carlo Method for Medical & Health Physics

Monte Carlo Method for Medical & Health Physics Med Phys 774 Monte Carlo Method for Medical & Health Physics Chapter 5. MCNP Monte Carlo Code 1 MCNP stands for: A general-purpose M C N-P code Particles that can be transported??? See the references:

More information

Automated ADVANTG Variance Reduction in a Proton Driven System. Kenneth A. Van Riper1 and Robert L. Metzger2

Automated ADVANTG Variance Reduction in a Proton Driven System. Kenneth A. Van Riper1 and Robert L. Metzger2 Automated ADVANTG Variance Reduction in a Proton Driven System Kenneth A. Van Riper1 and Robert L. Metzger2 1 White Rock Science, P. O. Box 4729, White Rock, NM 87547, kvr@rt66.com Radiation Safety Engineering,

More information

Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.

Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran. DXRaySMCS First User Friendly Interface Developed for Prediction of Diagnostic Radiology X-Ray Spectra Produced by Monte Carlo (MCNP-4C) Simulation in Iran M.T. Bahreyni Toosi a*, H. Moradi b, H. Zare

More information

Code characteristics

Code characteristics The PENELOPE Computer code M.J. Anagnostakis Nuclear Engineering Department National Technical University of Athens The PENELOPE code system PENetration and Energy LOss of Positrons and Electrons in matter

More information

Construction of Voxel-type Phantom Based on Computed Tomographic Data of RANDO Phantom for the Monte Carlo Simulations

Construction of Voxel-type Phantom Based on Computed Tomographic Data of RANDO Phantom for the Monte Carlo Simulations Construction of Voxel-type Phantom Based on Computed Tomographic Data of RANDO Phantom for the Monte Carlo Simulations K. Minami 1, K. Ejiri 1, M. Shimo 1, M. Kato, Y. Takeuchi, K. Yonemochi, H. Toyama

More information

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters Monte Carlo Simulation for Neptun 1 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters M.T. Bahreyni Toossi a, M. Momen Nezhad b, S.M. Hashemi a a Medical Physics Research Center,

More information

Frequency-based method to optimize the number of projections for industrial computed tomography

Frequency-based method to optimize the number of projections for industrial computed tomography More Info at Open Access Database www.ndt.net/?id=18739 Frequency-based method to optimize the number of projections for industrial computed tomography Andrea Buratti 1, Soufian Ben Achour 1, Christopher

More information

ICRP Symposium on Radiological Protection Dosimetry

ICRP Symposium on Radiological Protection Dosimetry ICRP Symposium on Radiological Protection Dosimetry Tokyo February 18 th, 2016 M. Zankl a, J. Becker a, K.F. Eckerman b, W.E. Bolch c, R. Behrens d a Helmholtz Zentrum München, Neuherberg, Germany b Oak

More information

Dosimetry Simulations with the UF-B Series Phantoms using the PENTRAN-MP Code System

Dosimetry Simulations with the UF-B Series Phantoms using the PENTRAN-MP Code System Dosimetry Simulations with the UF-B Series Phantoms using the PENTRAN-MP Code System A. Al-Basheer, M. Ghita, G. Sjoden, W. Bolch, C. Lee, and the ALRADS Group Computational Medical Physics Team Nuclear

More information

Applying Hounsfield unit density calibration in SkyScan CT-analyser

Applying Hounsfield unit density calibration in SkyScan CT-analyser 1 Bruker-microCT Method note Applying Hounsfield unit density calibration in SkyScan CT-analyser Hounsfield units (HU) are a standard unit of x-ray CT density, in which air and water are ascribed values

More information

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Tecniques of imaging in Overview Introduction Tecniques of imaging

More information

Attenuation Coefficients for Layered Ceiling and Floor Shields in PET/CT Clinics

Attenuation Coefficients for Layered Ceiling and Floor Shields in PET/CT Clinics Attenuation Coefficients for Layered Ceiling and Floor Shields in PET/CT Clinics Robert L. Metzger and Kenneth A. Van Riper Radiation Safety Engineering, Inc 3245 North Washington Street, Chandler, AZ

More information

PROVIDE A UNIFORM DOSE IN THE SMALL ANIMAL.

PROVIDE A UNIFORM DOSE IN THE SMALL ANIMAL. Considerations in the Use of the RS 2000 X ray Irradiator for Biological Research (Primarily Small Animal, tissue, and cells) and the fallacy of the High KV spectrum. The performance goal for a small animal

More information

Outline. Monte Carlo Radiation Transport Modeling Overview (MCNP5/6) Monte Carlo technique: Example. Monte Carlo technique: Introduction

Outline. Monte Carlo Radiation Transport Modeling Overview (MCNP5/6) Monte Carlo technique: Example. Monte Carlo technique: Introduction Monte Carlo Radiation Transport Modeling Overview () Lecture 7 Special Topics: Device Modeling Outline Principles of Monte Carlo modeling Radiation transport modeling with Utilizing Visual Editor (VisEd)

More information

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology Diagnostic imaging techniques Krasznai Zoltán University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology 1. Computer tomography (CT) 2. Gamma camera 3. Single Photon

More information

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY David K. Lehmann 1, Kathleen Brockdorf 1 and Dirk Neuber 2 1 phoenix x-ray Systems + Services Inc. St. Petersburg, FL, USA 2 phoenix x-ray Systems + Services

More information

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

C a t p h a n / T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 Internationally recognized for measuring the maximum obtainable performance of axial, spiral and multi-slice

More information

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Benedikt Lorch 1, Martin Berger 1,2, Joachim Hornegger 1,2, Andreas Maier 1,2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg

More information

Particle track plotting in Visual MCNP6 Randy Schwarz 1,*

Particle track plotting in Visual MCNP6 Randy Schwarz 1,* Particle track plotting in Visual MCNP6 Randy Schwarz 1,* 1 Visual Editor Consultants, PO Box 1308, Richland, WA 99352, USA Abstract. A visual interface for MCNP6 has been created to allow the plotting

More information

Computer aided error analysis for a threedimensional precision surface mapping system

Computer aided error analysis for a threedimensional precision surface mapping system Computer aided error analysis for a threedimensional precision surface mapping system M. Hill, J.W. McBride, D. Zhang & J. Loh Mechanical Engineering, U~riversity of Southampton, UK Abstract This paper

More information

TEP Hounsfield units. Related topics Attenuation coefficient, Hounsfield units

TEP Hounsfield units. Related topics Attenuation coefficient, Hounsfield units Hounsfield units TEP Related topics Attenuation coefficient, Hounsfield units Principle Depending on the type of CT scanner and the settings, the result of a CT scan of the same material can be different

More information

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 3, JUNE 2003 405 Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging Edward J. Soares, Kevin W. Germino,

More information

Proton dose calculation algorithms and configuration data

Proton dose calculation algorithms and configuration data Proton dose calculation algorithms and configuration data Barbara Schaffner PTCOG 46 Educational workshop in Wanjie, 20. May 2007 VARIAN Medical Systems Agenda Broad beam algorithms Concept of pencil beam

More information

SESAME: A NEW TOOL FOR THE NUMERICAL DOSIMETRIC RECONSTRUCTION OF A RADIOLOGICAL ACCIDENT CAUSED BY AN

SESAME: A NEW TOOL FOR THE NUMERICAL DOSIMETRIC RECONSTRUCTION OF A RADIOLOGICAL ACCIDENT CAUSED BY AN SESAME: A NEW TOOL FOR THE NUMERICAL DOSIMETRIC RECONSTRUCTION OF A RADIOLOGICAL ACCIDENT CAUSED BY AN EXTERNAL SOURCE. APPLICATION TO THE ACCIDENT IN CHILE IN DECEMBER 2005. C. Huet, A. Lemosquet, I.

More information

RADIOMICS: potential role in the clinics and challenges

RADIOMICS: potential role in the clinics and challenges 27 giugno 2018 Dipartimento di Fisica Università degli Studi di Milano RADIOMICS: potential role in the clinics and challenges Dr. Francesca Botta Medical Physicist Istituto Europeo di Oncologia (Milano)

More information

Development of a Radiation Shielding Monte Carlo Code: RShieldMC

Development of a Radiation Shielding Monte Carlo Code: RShieldMC Development of a Radiation Shielding Monte Carlo Code: RShieldMC Shenshen GAO 1,2, Zhen WU 1,3, Xin WANG 1,2, Rui QIU 1,2, Chunyan LI 1,3, Wei LU 1,2, Junli LI 1,2*, 1.Department of Physics Engineering,

More information

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct J. DHAENE* 1, E. PAUWELS 1, T. DE SCHRYVER 1, A. DE MUYNCK 1, M. DIERICK 1, L. VAN HOOREBEKE 1 1 UGCT Dept. Physics

More information

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Wenjuan Sun and Stephen Brown 10 th May 2016 1 Possible factors influencing XCT measurements Components Influencing variables Possible

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Monte Carlo Methods 1 Monte Carlo! Most accurate at predicting dose distributions! Based on

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

The Monte Carlo simulation of a Package formed by the combination of three scintillators: Brillance380, Brillance350, and Prelude420.

The Monte Carlo simulation of a Package formed by the combination of three scintillators: Brillance380, Brillance350, and Prelude420. EURONS I3 506065 JRA9 RHIB Report made during stay IEM-CSIC Madrid december 2006 MINISTERIO DE ASUNTOS EXTERIORES Y DE COOPERACIÓN AECI VICESECRETARÍA GENERAL The Monte Carlo simulation of a Package formed

More information

COMPARISON OF DOSE CALCULATION ALGORITHMS FOR LEKSELL GAMMA KNIFE PERFEXION USING MONTE CARLO VOXEL PHANTOMS

COMPARISON OF DOSE CALCULATION ALGORITHMS FOR LEKSELL GAMMA KNIFE PERFEXION USING MONTE CARLO VOXEL PHANTOMS COMPARISON OF DOSE CALCULATION ALGORITHMS FOR LEKSELL GAMMA KNIFE PERFEXION USING MONTE CARLO VOXEL PHANTOMS Jan Pipek 1, Josef Novotný Jr. 1,2,3, Josef Novotný 1, Petra Kozubíková 1 1 Faculty of Nuclear

More information

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT More Info at Open Access Database www.ndt.net/?id=18754 Novel evaluation method of low contrast resolution performance of dimensional X-ray CT Makoto Abe 1, Hiroyuki Fujimoto 1, Osamu Sato 1, Kazuya Matsuzaki

More information

Simulation of Radiographic Testing for POD Assessment

Simulation of Radiographic Testing for POD Assessment 4th European-American Workshop on Reliability of NDE - Th.5.B.2 Simulation of Radiographic Testing for POD Assessment Gerd-Rüdiger JAENISCH, Carsten BELLON, Uwe EWERT, BAM, Berlin, Germany Abstract. Efficient

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

Novel techniques for the detection and localization of attenuated gamma-ray sources

Novel techniques for the detection and localization of attenuated gamma-ray sources University of New Mexico UNM Digital Repository Nuclear Engineering ETDs Engineering ETDs 9-3-2010 Novel techniques for the detection and localization of attenuated gamma-ray sources Tyler Alecksen Follow

More information

Scatter Correction Methods in Dimensional CT

Scatter Correction Methods in Dimensional CT Scatter Correction Methods in Dimensional CT Matthias Baer 1,2, Michael Hammer 3, Michael Knaup 1, Ingomar Schmidt 3, Ralf Christoph 3, Marc Kachelrieß 2 1 Institute of Medical Physics, Friedrich-Alexander-University

More information

CIVA Computed Tomography Modeling

CIVA Computed Tomography Modeling CIVA Computed Tomography Modeling R. FERNANDEZ, EXTENDE, France S. LEGOUPIL, M. COSTIN, D. TISSEUR, A. LEVEQUE, CEA-LIST, France page 1 Summary Context From CIVA RT to CIVA CT Reconstruction Methods Applications

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Advanced Image Reconstruction Methods for Photoacoustic Tomography

Advanced Image Reconstruction Methods for Photoacoustic Tomography Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic

More information

Continuation Format Page

Continuation Format Page C.1 PET with submillimeter spatial resolution Figure 2 shows two views of the high resolution PET experimental setup used to acquire preliminary data [92]. The mechanics of the proposed system are similar

More information

OPTIMISATION OF PIN FIN HEAT SINK USING TAGUCHI METHOD

OPTIMISATION OF PIN FIN HEAT SINK USING TAGUCHI METHOD CHAPTER - 5 OPTIMISATION OF PIN FIN HEAT SINK USING TAGUCHI METHOD The ever-increasing demand to lower the production costs due to increased competition has prompted engineers to look for rigorous methods

More information

MCNP Variance Reduction technique application for the Development of the Citrusdal Irradiation Facility

MCNP Variance Reduction technique application for the Development of the Citrusdal Irradiation Facility IYNC 2008 Interlaken, Switzerland, 20-26 September 2008 Paper No. 376 MCNP Variance Reduction technique application for the Development of the Citrusdal Irradiation Facility R Makgae Pebble Bed Modular

More information

I. INTRODUCTION. Figure 1. Radiation room model at Dongnai General Hospital

I. INTRODUCTION. Figure 1. Radiation room model at Dongnai General Hospital International Journal of Computational Engineering Research Vol, 04 Issue, 4 Simulation of Photon and Electron dose distributions 5 code for the treatment area using the linear electron accelerator (LINAC)

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

Suitability Study of MCNP Monte Carlo Program for Use in Medical Physics

Suitability Study of MCNP Monte Carlo Program for Use in Medical Physics Nuclear Energy in Central Europe '98 Terme Catez, September 7 to 10, 1998 SI0100092 Suitability Study of MCNP Monte Carlo Program for Use in Medical Physics R. Jeraj Reactor Physics Division, Jozef Stefan

More information

Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette

Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette Pipettes allow transfer of accurately known volume of solution from one container to another. Volumetric or transfer pipettes deliver a

More information

Quality control facilities for large optical reflectors at ENEA-Casaccia for physics application

Quality control facilities for large optical reflectors at ENEA-Casaccia for physics application Quality control facilities for large optical reflectors at ENEA-Casaccia for physics application S. Baccaro a,c, C. Bosio c, A. Maccari b, M. Montecchi a,c a FIS-ION, ENEA-Casaccia, Via Anguillarese 301,

More information

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

Artifact Mitigation in High Energy CT via Monte Carlo Simulation PIERS ONLINE, VOL. 7, NO. 8, 11 791 Artifact Mitigation in High Energy CT via Monte Carlo Simulation Xuemin Jin and Robert Y. Levine Spectral Sciences, Inc., USA Abstract The high energy (< 15 MeV) incident

More information

COMPUTATIONAL FLUID DYNAMICS ANALYSIS OF ORIFICE PLATE METERING SITUATIONS UNDER ABNORMAL CONFIGURATIONS

COMPUTATIONAL FLUID DYNAMICS ANALYSIS OF ORIFICE PLATE METERING SITUATIONS UNDER ABNORMAL CONFIGURATIONS COMPUTATIONAL FLUID DYNAMICS ANALYSIS OF ORIFICE PLATE METERING SITUATIONS UNDER ABNORMAL CONFIGURATIONS Dr W. Malalasekera Version 3.0 August 2013 1 COMPUTATIONAL FLUID DYNAMICS ANALYSIS OF ORIFICE PLATE

More information

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed

More information

Radiological Characterization and Decommissioning of Research and Power Reactors 15602

Radiological Characterization and Decommissioning of Research and Power Reactors 15602 Radiological Characterization and Decommissioning of Research and Power Reactors 15602 INTRODUCTION Faezeh Abbasi *, Bruno Thomauske *, Rahim Nabbi * RWTH University Aachen The production of the detailed

More information

A CT-based Monte Carlo Dose Calculations for Proton Therapy Using a New Interface Program

A CT-based Monte Carlo Dose Calculations for Proton Therapy Using a New Interface Program World Academy of Science, Engineering and Technology 53 29 A CT-based Monte Carlo Dose Calculations for Proton Therapy Using a New Interface Program A. Esmaili Torshabi, A. Terakawa, K. Ishii, H. Yamazaki,

More information

CALCULATION OF THE ACTIVITY INVENTORY FOR THE TRIGA REACTOR AT THE MEDICAL UNIVERSITY OF HANNOVER (MHH) IN PREPARATION FOR DISMANTLING THE FACILITY

CALCULATION OF THE ACTIVITY INVENTORY FOR THE TRIGA REACTOR AT THE MEDICAL UNIVERSITY OF HANNOVER (MHH) IN PREPARATION FOR DISMANTLING THE FACILITY CALCULATION OF THE ACTIVITY INVENTORY FOR THE TRIGA REACTOR AT THE MEDICAL UNIVERSITY OF HANNOVER (MHH) IN PREPARATION FOR DISMANTLING THE FACILITY Gabriele Hampel, Friedemann Scheller, Medical University

More information

CALIBRATION OF DIFFRACTOMETERS: A TEST METHOD TO MONITOR THE PERFORMANCE OF INSTRUMENTS. G. Berti, U. Bartoli, M. D Acunto and F.

CALIBRATION OF DIFFRACTOMETERS: A TEST METHOD TO MONITOR THE PERFORMANCE OF INSTRUMENTS. G. Berti, U. Bartoli, M. D Acunto and F. Materials Science Forum Online: 004-01-15 ISSN: 166-975, Vols. 443-444, pp 7-30 doi:10.408/www.scientific.net/msf.443-444.7 004 Trans Tech Publications, Switzerland CALIBRATION OF DIFFRACTOMETERS: A TEST

More information

Mesh Human Phantoms with MCNP

Mesh Human Phantoms with MCNP LAUR-12-01659 Mesh Human Phantoms with MCNP Casey Anderson (casey_a@lanl.gov) Karen Kelley, Tim Goorley Los Alamos National Laboratory U N C L A S S I F I E D Slide 1 Summary Monte Carlo for Radiation

More information

Muon imaging for innovative tomography of large volume and heterogeneous cemented waste packages

Muon imaging for innovative tomography of large volume and heterogeneous cemented waste packages Muon imaging for innovative tomography of large volume and heterogeneous cemented waste packages This project has received funding from the Euratom research and training programme 2014-2018 under grant

More information

NON-COLLIMATED SCATTERED RADIATION TOMOGRAPHY

NON-COLLIMATED SCATTERED RADIATION TOMOGRAPHY NON-COLLIMATED SCATTERED RADIATION TOMOGRAPHY Gorshkov V.A., Space Research Institute, Moscow, Russia Yumashev V.M., State corporation "Rosatom", Centre "Atom-innovation", Moscow, Russia Kirilenko K.V.,

More information

TWENTY YEARS OF VISUAL MONTE CARLO

TWENTY YEARS OF VISUAL MONTE CARLO IX Latin American IRPA Regional Congress on Radiation Protection and Safety - IRPA 2013 Rio de Janeiro, RJ, Brazil, April 15-19, 2013 SOCIEDADE BRASILEIRA DE PROTEÇÃO RADIOLÓGICA - SBPR TWENTY YEARS OF

More information

SIMULATION OF ARTIFICIAL SYSTEMS BEHAVIOR IN PARAMETRIC EIGHT-DIMENSIONAL SPACE

SIMULATION OF ARTIFICIAL SYSTEMS BEHAVIOR IN PARAMETRIC EIGHT-DIMENSIONAL SPACE 78 Proceedings of the 4 th International Conference on Informatics and Information Technology SIMULATION OF ARTIFICIAL SYSTEMS BEHAVIOR IN PARAMETRIC EIGHT-DIMENSIONAL SPACE D. Ulbikiene, J. Ulbikas, K.

More information

Chapter 13 RADIATION HEAT TRANSFER

Chapter 13 RADIATION HEAT TRANSFER Heat and Mass Transfer: Fundamentals & Applications Fourth Edition in SI Units Yunus A. Cengel, Afshin J. Ghajar McGraw-Hill, 2011 Chapter 13 RADIATION HEAT TRANSFER PM Dr Mazlan Abdul Wahid Universiti

More information

Loma Linda University Medical Center Dept. of Radiation Medicine

Loma Linda University Medical Center Dept. of Radiation Medicine Loma Linda University Medical Center Dept. of Radiation Medicine and Northern Illinois University Dept. of Physics and Dept. of Computer Science Presented by George Coutrakon, PhD NIU Physics Dept. Collaborators

More information

Non-Stationary CT Image Noise Spectrum Analysis

Non-Stationary CT Image Noise Spectrum Analysis Non-Stationary CT Image Noise Spectrum Analysis Michael Balda, Björn J. Heismann,, Joachim Hornegger Pattern Recognition Lab, Friedrich-Alexander-Universität Erlangen Siemens Healthcare, Erlangen michael.balda@informatik.uni-erlangen.de

More information

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube T. Donath* a, F. Pfeiffer a,b, O. Bunk a, W. Groot a, M. Bednarzik a, C. Grünzweig a, E. Hempel c, S. Popescu c, M. Hoheisel

More information