Performance Evaluation of the Philips Gemini PET/CT System

Size: px
Start display at page:

Download "Performance Evaluation of the Philips Gemini PET/CT System"

Transcription

1 Performance Evaluation of the Philips Gemini PET/CT System Rebecca Gregory, Mike Partridge, Maggie A. Flower Joint Department of Physics, Institute of Cancer Research, Royal Marsden HS Foundation Trust, Downs Road, Sutton, Surrey, SM2 5PT, UK Abstract--Methods to test the practical performance of the Philips Gemini PET/CT system during clinical imaging are described and results presented. Measurements were performed using the EEC Emission phantom, alongside phantoms provided by Philips. The test methods used were based on EMA (U2 200) and those suggested by a task group of the Commission of European Communities (EEC). The clinical protocols in use were based on the single-pass emission-transmission whole body scanning protocols provided by Philips. Images are reconstructed using a 3D row action maximum likelihood algorithm (RAMLA). Attenuation correction was applied using both CT data (CTAC) and Cs-37 transmission scans (CsAC) to allow a comparison of the two methods. Images can be acquired with the PET and CT gantries adjacent (closed) or separated by 60 cm (open); geometric calibrations show that gantry alignment is most consistent in the closed position. Results show the non-uniformity of the images varied across the slices, for the non-attenuation corrected (OAC) images, but that uniformity was improved and more consistent in the both the CTAC and CsAC images. The reconstructed scatter fractions show that the RSF was better for OAC (0.5) than for CsAC (0.8) and CTAC (0.22), the sinogram derived scatter fraction was At the centre the transverse resolution was 5.3 mm in air (5.4 mm in water) and the axial resolution was 5.8 mm (6.5 mm). earer the FOV periphery the transverse radial resolution was 5.3 mm (5.3 mm), the transverse tangential resolution was 5.3 mm (5.5 mm) and the axial resolution was 7.5 mm (7.4 mm). The recovery coefficients show that partial volume only affects objects smaller than 22 mm. T I. ITRODUCTIO HE Philips Gemini PET/CT system consists of an Allegro 3D PET scanner in line with an Mx8000 EXP Dual-Slice CT Imaging System so that the patient can be imaged in both scanners on the same couch. The Gemini PET/CT system, at the Royal Marsden Hospital, has been in routine use for imaging cancer since in February A. Routine Acquisition Protocols The Allegro has no septa and operates exclusively in 3D mode. The Allegro scanner design has already been detailed in other publications []. Manuscript submitted October 5, A low dose CT scan (20kV, 80 mas/slice) is acquired during the patient study. The PET/CT scan allows for better localisation over single modality PET. The CT scan can be used for attenuation correction (CTAC). A Cs-37 source transmission scan is also acquired interleaved with the emission scan and stored using 4D rebinning (emission /transmission scan). This transmission scan is used for attenuation correction (Cs AC). B. Routine Reconstruction Protocols Currently the routine clinical images are reconstructed three times; without attenuation correction (AC), with CT AC and with Cs AC. All images are normalised to correct for the variation in detector efficiencies and distortion. All counts in each frame are decay corrected. The AC images use background subtraction which, from the level of background outside the body, estimates the background throughout the field of view and subtracts it, this background includes scatter and randoms that weren t removed using the energy window and delayed coincidence window respectively. To apply Cs AC the Cs-37 source transmission images are reconstructed using ordered subsets maximum estimated likelihood (OSEM). These images are corrected for emission contamination from the patient, scatter and then segmented. The segmented attenuation map is then forward projected into a sinogram of attenuation factors, which are then applied to the emission sinogram. CT AC is applied in the same way as in Cs AC, but the low dose CT image is used in place of the Cs-37 source transmission image and converted to PET attenuation coefficients. Because the CT image has better statistics, than the transmission image, it does not need to be segmented. All images are reconstructed routinely using 3D row-action maximum likelihood algorithm (RAMLA), a fast iterative algorithm [2], with a relaxation parameter of 0.006, for 2 iterations and a blob radius of 2.5. The image size is always 44 44, for a FOV of 576 mm, 4 mm thick slices. II. THE PERFORMACE TESTS Results of EMA tests, provided by the manufacturers, give an indication of the ideal performance of the system but not the performance in conditions for use during in vivo studies. The /04/$20.00 (C) 2004 IEEE

2 methods described within this paper aim to test the practical performance of the system during routine clinical imaging using locally available phantoms, and the routine clinical protocols. The test methods used are based on EMA U2 [3] and those suggested by a task group of the Commission of European Communities [4]. These tests have been performed using the EEC emission phantom. This phantom was designed to reflect the human shape, consisting of a whole-body phantom and a head-sized phantom, with various inserts that can be used to simulate small areas of activity uptake. The tests are designed to be reproducible, so that they can be repeated and initial results used as a base line for future performance checks. Uniformity and gantry alignment tests were performed using proprietary phantoms supplied by Philips. A. PET and CT Alignment Six 370 MBq a-22 point sources were carefully positioned using a jig provided by Philips. CT and PET images of the jig were acquired for the gantry in both the closed and open positions. The data were reconstructed, without AC and then analysed using software supplied by Philips, which measures the static offsets and rotation between the CT and PET image volumes. These offsets are stored, in the file header of subsequently acquired scans, and automatically applied to achieve image registration. Repeatability of these measurements was tested to see whether the changes may be due to the accuracy of the software or to the movement of the gantry between the open and closed positions. The alignment test was repeated 3 times with the gantry in the closed position. Then the test was repeated in the closed position and open positions, but moving the gantry between measurements. B. on-uniformity The uniformity phantom, supplied by Philips was filled with approximately 88 MBq of F-8. The phantom was then suspended from the end of the couch and then centred in the FOV. A low-dose CT scan was acquired followed by emission/ transmission scans. The images were reconstructed three times with each form of AC. on-uniformity is defined as the maximum relative deviation of counts from the mean. Positive and negative deviations were calculated, as + 00(max mu ) / mu and () + 00( mu min) / mu, where max is the maximum, min is the minimum and mu the mean of the counts, in a grid of small 2 mm square regions of interest (ROIs), defined within a circular useful region cm within the phantom, on each slice, as described in the EEC protocol [4]. C. Scatter Fractions The scatter source for the EEC phantom consists of a line source that can be positioned at 3 radial positions in the water filled cylindrical (head) phantom, at A=0, B=45 mm and C=90 mm. This scatter source was filled with approximately 4 MBq of Ga-68. At this activity the dead-time losses and the random coincidence rate were both below 5%. The head phantom was placed on the couch and centred in the field of view. Images were acquired with the scatter source at each of the radial positions. At each position a CT scan was acquired for attenuation correction (AC). Then emission/transmission scans were acquired for each source position. The PET protocol was modified to collect 0,000,000 counts per couch position. ) Reconstructed scatter fraction The images were reconstructed using the routine reconstruction algorithm, without AC, with CT AC and with Cs AC. The reconstructed scatter fraction was analysed using the EEC methodology [4]. In brief, the full width of half maximum (FWHM) of the source was determined by taking a profile across the image of the source. 3 ROIs were drawn on each slice of the image. ROI had a diameter of 4FWHM, and was centred on the image of the scatter source. The scatter within ROI was estimated from a concentric circular ROI (ROI 2 ), with diameter 4 FWHW plus 2 pixel widths. An ROI with the diameter of the head phantom was drawn centred on the source at the central position (ROI 3 ). These ROIs are shown in the diagram, Fig.. The reconstructed scatter fraction, RSFi, the scattered counts S i divided total counts, scattered plus true, T i, was calculated for each slice i and source position. RSFi = Si /( Si + Ti ) = ( C3 ( C A )) / C (2) 3 where C 3 is the counts in ROI 3, there for the total scattered plus true counts. C is the number of counts in ROI. A is the estimate of the scattered counts in ROI, calculated as, A ( ) ( /( ) = C2 C 2 ) (3) where C 2 is the counts in ROI 2, and 2 is the number of pixels in ROI 2 and is the number of pixels in ROI. These scatter fractions describe the scatter behaviour at each radial position of the source. ROI Source ROI 3 ROI 2 Fig.. A schematic diagram showing the position of the regions of interest for the source positioned at the centre of the cylinder. To give the average scatter fraction for the whole of the uniform cylinder, RSF cyl, the individual RSF for each source position are weighted and summed. The weighting corresponds to the fraction of the area of the cross-section that they present /04/$20.00 (C) 2004 IEEE

3 S Ai + 8 S Bi SC i = (4) RSFcyl C3Ai + 8 C3Bi C3C i where S Ai, S Bi, S Ci are the scattered counts in each slice for the source at positions A, B and C respectively, these are summed over slice. C 3Ai, C 3Bi and C 3Ci are the counts in ROI 3 in each slice for the source at positions A, B and C respectively. 2) Sinogram derived Scatter fraction For comparison, the scatter fraction was also calculated from the interpolated sinogram, for the line source at position B, which was rebinned into 2 mm slices. This is based on the test described in [3]. For each projection angle, the peak was shifted to align with the central pixel. Then for each slice, the counts within ±20 mm of this peak were summed over all the projections. For each peak profile, the counts below a line joining the counts at the edge of the 40 mm strip (C L and C R ) are considered to be scattered and the counts in the peak above to be unscattered, see Fig. 2. Therefore the scatter fraction for each slice i, SF i, is given by, 0.5p( CR + CL ) + C0 SFi = (5) CT where C T is the total counts under the peak, p is the number of pixels between the edges of the strip and C 0 the number of counts outside the strip. The system scatter fraction is weighted average of the slice scatter fractions. C L Unscattered counts Scattered counts C R -20mm 0mm +20mm Fig. 2. Illustration of scattered and un-scattered counts. D. Transaxial and Axial Resolution Two line-sources, consisting of glass tubes filled with F-8 were placed in the head phantom. One was aligned along the central axis and the other parallel and 9 cm away. The central axis of the head phantom was aligned with the z-axis and rotated so that images could be acquired with the second line source displaced along the x-axis and then displaced along the y-axis. Then the central axis of the head phantom was positioned along the x-axis and the 2 line-sources; one positioned at the centre and the other 4.5 cm away in the z- direction, so that the axial resolution could be measured. These measurements were then repeated for the head phantom both empty and filled with water. PET images were acquired over a small FOV of 256 mm and the reconstruction protocol was amended for a matrix size of to give a pixel size of mm (less than /3 full width half maximum (FWHM)). The resolution was calculated as the average tangential and axial FWHM s of a profile taken through and perpendicular to the images of the line sources, following the U 2 guidelines [3]. E. Recovery Coefficients EEC spheres, placed in the water-filled head phantom, were used to measure the recovery coefficients. All the spheres were filled with Ga-68, from the same stock vial, with an activity concentration of 0.57 MBq/ml. Another head phantom was filled with Ga-68 to give an activity concentration of MBq/ml. A CT scan of the head phantom with the spheres was acquired. To ensure that the spheres are centred on a reconstructed image slice, three emission and Cs-source transmission scans were acquired of the spheres, at axial intervals of 0.5 times the reconstructed slice width. A CT scan was then acquired of the head phantom, without the spheres, followed by emission and Cs-source transmission scans. The scans were reconstructed, using the routinely used clinical reconstruction protocol, with CT AC and Cs AC. The images of the spheres were viewed to find the axial position where the smallest sphere was brightest; the scan at this position was then used for the following analysis. Recovery coefficients were measured for both types of AC. Small regions of interest (ROIs), roughly the same diameter as the resolution of the system (2 mm diameter) were placed on the image of each of the spheres. And a large ROI (50 mm diameter) was placed on the image of the head phantom, without the spheres, on a slice at the same position as the spheres. The recovery coefficient, R Si, for each sphere was calculated using, R Si = ( CSi As ) ( CH AH ), (6) where C Si and C H are the count rate per second per pixel, in sphere i and the head phantom respectively. A S and A H are the activity concentrations in the spheres and the head phantom respectively, decay corrected for the time of each scan. III. RESULTS A. PET and CT Alignment Initial results, showing the variation in static offsets with time for both the open and closed gantry positions, are shown in Fig. 3. The results of the reproducibility tests are shown in table I. TABLE I RESULTS OF THE ALIGMET REPEATABILITY MEASUREMETS /04/$20.00 (C) 2004 IEEE

4 The coefficient of variation of the non-uniformity of the ROIs, defined on each slice are also plotted to demonstrate how the uniformity varies within, the useful region on each of the slices. Due to set up errors the last 2 cm of the CT attenuation corrected images are missing. C. Scatter Fractions ) Reconstructed scatter fraction The variation in reconstructed scatter fraction across the slices is shown in Fig. 5, for the on AC image. There is an apparent increase in the scatter fraction toward the 90 mm slice position, this is due to the presence of the Cs-source (which is stored at that end of the scanner), and is not scatter. In order to avoid this area the volume scatter fractions were calculated using the central values in the flatter region of the graph (from 65 to +65 mm). The volume reconstructed scatter fractions were 0.5, 0.8 and 0.22 for the non-ac, Cs AC and CT AC images respectively. Figure 3. Graphs to show the variation in static offsets since the PET/CT scanner went into clinical use. Graph a shows the values for the gantry in the closed position, b) shows the values for the gantry in the open position and c) the rotation about the z-axis for both the closed and open gantry position (rotation about the x and y axis is not determined) B. on-uniformity Results of the non-uniformity measurements are shown in Fig. 4. These are displayed as the maximum positive and negative deviations in each of the slices, to give an indication of how the maximum non-uniformity changes across the slices. Fig. 5. Reconstructed scatter fractions versus slice for each source position for the non-ac image. Fig. 4. Results of the non-uniformity test, on the top row the range of nonuniformity values measured in 45 slices, and on the bottom row the coefficient of variation of non-uniformity values for each slice, a) for non-ac, b) for CT AC and c) for Cs AC /04/$20.00 (C) 2004 IEEE

5 2) Sinogram derived Scatter fraction The volume scatter fraction, calculated in the area unaffected by the transmission source was 0.36, the variation in scatter fraction across the slices is shown in Fig. 6. Fig.6 A graph of scatter fractions (calculated from the sinogram) vs. slice position. D. Transaxial and Axial Resolution The results of the resolution measurements are shown in table II. TABLE. U-2 RESOLUTIO E. Recovery Coefficients The recovery coefficients were plotted against the sphere diameters, these are shown in Fig. 7. As expected these values tend towards.0, a double exponential, tending to was fitted to the data (Trend line). Recovery Coefficient y = ( 5.5e 'CS AC' 'CT AC' Trend line Sphere Diameter (mm) Fig. 7. Recovery coefficients as a function of sphere diameter. IV. DISCUSSIO )( 0.9e 0.2x 0.7 x A major advantage of a PET/CT system is the simplification of image registration. Alignment of the images assumes that the position of the PET system relative to the CT system is known. Initial checks show that the PET and CT scanner alignment needs to be calibrated regularly, but over time the variation in alignment has settled, as the PET/CT system and building have settled. The reproducibility measurements show ) that the variation in offsets is not due to uncertainty in the measurements, and that moving the gantry does not affect the alignment of the inherently registered images. on-uniformity measurements tested the PET scanner's response to a homogeneous activity distribution. The EEC protocol [4] states that non-uniformity measurements should be made on the EEC head phantom. However, Philips provide a cylinder to be used for monthly uniformity checks, so this cylindrical phantom was used for consistency. These results show that, there is more non-uniformity towards the end slices for the non-ac images, than the AC images. The degree of non-uniformity is comparable to the GE Advance nonuniformity [5]. Scattering of photons, within the phantom and off the gantry components, causes coincident events to be incorrectly localised. The reconstructed scatter fractions show that the attenuation corrected images are more sensitive to scatter effects than images without attenuation correction. The CT AC images are effected more that the Cs AC images. The intrinsic scatter fraction derived from the sinogram is very different, but this is due to the different analysis technique. The spatial resolution measured in both air and water are very similar, the scatter of radiation in the water has caused very slightly larger FWHMs in some cases, but the difference is very small. As expected the transverse resolution is better than the axial resolution. Over the region measured, there was no apparent degradation in spatial resolution for sources away from the central axis, due to the energy absorbed in the scintillator being spread over a greater area, and therefore over more detector elements. The results of the recovery coefficient measurements quantify the apparent decrease in tracer concentration in an ROI, when the dimensions of the object in any direction are less than 3 FWHM. As the average FWHM is about 5.5 mm in water, it may be expected that the apparent tracer concentration will not recover until the object is over 6.5 mm in all dimensions, therefore full recovery would only occur after the 3rd data points, this is observed in the graph of recovery coefficients. The recovery coefficients are fairly similar for both forms of attenuation correction. V. REFERECES [] S. Surti and J.S. Karp, Imaging characteristics of a 3-Dimensional GSO whole-body PET camera J ucl Med., vol. 45, no. 6, pp , Jan [2] J. Browne, A row-action alternative to the EM algorithm for maximum likelihoods in emission tomography IEEE Trans. Med. Imag., vol. 5, no. 5, pp , Oct [3] EMA U 2-200, Performance measurements of positron emission tomographs, EMA standards publication 200. [4] R. Guzzardi, C.R. Bellina, B. Knoop, K. Jordan, H. Ostertag, H.W. Reist et al., Methodologies for perfomance evaluation of positron emission tomographs J. ucl. Biol. Med. Vol. 35, pp. 4-57, July 99. [5] T.K. Lewellen, S.G. Kohlmyer, R.S. Miyaoka, M.S. Kaplan, C.W. Staerns, S.F. Schubert, Investigation of the performance of the General Electric ADVACE positron emission eomograph in 3D mode. IEEE Trans. ucl. Sci, Vol. 43, o. 4, pp , 996, Aug /04/$20.00 (C) 2004 IEEE

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation

More information

Introduction to Positron Emission Tomography

Introduction to Positron Emission Tomography Planar and SPECT Cameras Summary Introduction to Positron Emission Tomography, Ph.D. Nuclear Medicine Basic Science Lectures srbowen@uw.edu System components: Collimator Detector Electronics Collimator

More information

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging

Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 3, JUNE 2003 405 Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging Edward J. Soares, Kevin W. Germino,

More information

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Review of PET Physics Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Chart of Nuclides Z (protons) N (number of neutrons) Nuclear Data Evaluation Lab.

More information

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC What is needed? Why? How often? Who says? QA and QC in Nuclear Medicine QA - collective term for all the efforts made to produce

More information

Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide

Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide 2005 IEEE Nuclear Science Symposium Conference Record M04-8 Characterization of a Time-of-Flight PET Scanner based on Lanthanum Bromide J. S. Karp, Senior Member, IEEE, A. Kuhn, Member, IEEE, A. E. Perkins,

More information

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Workshop on Quantitative SPECT and PET Brain Studies 14-16 January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Físico João Alfredo Borges, Me. Corrections in SPECT and PET SPECT and

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial 8//0 AAPM0 Scientific Symposium: Emerging and New Generation PET: Instrumentation, Technology, Characteristics and Clinical Practice Aug Wednesday 0:4am :pm Solid State Digital Photon Counting PET/CT Instrumentation

More information

Fits you like no other

Fits you like no other Fits you like no other BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase in room

More information

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET

Corso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 5 SPECT, PET Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes

More information

Iterative SPECT reconstruction with 3D detector response

Iterative SPECT reconstruction with 3D detector response Iterative SPECT reconstruction with 3D detector response Jeffrey A. Fessler and Anastasia Yendiki COMMUNICATIONS & SIGNAL PROCESSING LABORATORY Department of Electrical Engineering and Computer Science

More information

Fits you like no other

Fits you like no other Fits you like no other Philips BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase

More information

Introduction to Emission Tomography

Introduction to Emission Tomography Introduction to Emission Tomography Gamma Camera Planar Imaging Robert Miyaoka, PhD University of Washington Department of Radiology rmiyaoka@u.washington.edu Gamma Camera: - collimator - detector (crystal

More information

Performance Evaluation of radionuclide imaging systems

Performance Evaluation of radionuclide imaging systems Performance Evaluation of radionuclide imaging systems Nicolas A. Karakatsanis STIR Users meeting IEEE Nuclear Science Symposium and Medical Imaging Conference 2009 Orlando, FL, USA Geant4 Application

More information

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO Abstract C.S. Levin, Y-C Tai, E.J. Hoffman, M. Dahlbom, T.H. Farquhar UCLA School of Medicine Division

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 61675-2 First edition 1998-01 Radionuclide imaging devices Characteristics and test conditions Part 2: Single photon emission computed tomographs Dispositifs d imagerie par radionucléides

More information

Modeling and Incorporation of System Response Functions in 3D Whole Body PET

Modeling and Incorporation of System Response Functions in 3D Whole Body PET Modeling and Incorporation of System Response Functions in 3D Whole Body PET Adam M. Alessio, Member IEEE, Paul E. Kinahan, Senior Member IEEE, and Thomas K. Lewellen, Senior Member IEEE University of

More information

664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005

664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005 664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005 Attenuation Correction for the NIH ATLAS Small Animal PET Scanner Rutao Yao, Member, IEEE, Jürgen Seidel, Jeih-San Liow, Member, IEEE,

More information

Improvement of contrast using reconstruction of 3D Image by PET /CT combination system

Improvement of contrast using reconstruction of 3D Image by PET /CT combination system Available online at www.pelagiaresearchlibrary.com Advances in Applied Science Research, 2013, 4(1):285-290 ISSN: 0976-8610 CODEN (USA): AASRFC Improvement of contrast using reconstruction of 3D Image

More information

In the short span since the introduction of the first

In the short span since the introduction of the first Performance Characteristics of a Newly Developed PET/CT Scanner Using NEMA Standards in 2D and 3D Modes Osama Mawlawi, PhD 1 ; Donald A. Podoloff, MD 2 ; Steve Kohlmyer, MS 3 ; John J. Williams 3 ; Charles

More information

Simultaneous measurement of noise and spatial resolution in PET phantom images

Simultaneous measurement of noise and spatial resolution in PET phantom images IOP PUBLISHING Phys. Med. Biol. 55 (2010) 1069 1081 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/55/4/011 Simultaneous measurement of noise and spatial resolution in PET phantom images Martin

More information

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography

More information

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives 1. Be familiar with recommendations

More information

James R Halama, PhD Loyola University Medical Center

James R Halama, PhD Loyola University Medical Center James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives Be familiar with the tests recommended

More information

Evaluation of AutoQA Lite TM Image Quality Measurement Software

Evaluation of AutoQA Lite TM Image Quality Measurement Software Evaluation of AutoQA Lite TM Image Quality Measurement Software Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4 2XU Phone: 0131 537 1161

More information

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation 26 IEEE Nuclear Science Symposium Conference Record M6-349 Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation Habib Zaidi, Senior Member, IEEE and Mohammad

More information

S. Guru Prasad, Ph.D., DABR

S. Guru Prasad, Ph.D., DABR PURPOSE S. Guru Prasad, Ph.D., DABR Director of Medical Physics IAEA Consultant NorthShore University Health System and University of Chicago, Pritzker School of Medicine Current TPS utilize more information

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

Quality control phantoms and protocol for a tomography system

Quality control phantoms and protocol for a tomography system Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

ISO ISO ISO OHSAS ISO

ISO ISO ISO OHSAS ISO ISO 9001 ISO 13485 ISO 14001 OHSAS 18001 ISO 27001 Pro-NM Performance 08-101 - standard version 08-103 - version with the PET Lid Phantom for NM and PET systems performance evaluation (collimator, artifacts,

More information

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

C a t p h a n / T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 Internationally recognized for measuring the maximum obtainable performance of axial, spiral and multi-slice

More information

Comparison of 3D PET data bootstrap resampling methods for numerical observer studies

Comparison of 3D PET data bootstrap resampling methods for numerical observer studies 2005 IEEE Nuclear Science Symposium Conference Record M07-218 Comparison of 3D PET data bootstrap resampling methods for numerical observer studies C. Lartizien, I. Buvat Abstract Bootstrap methods have

More information

A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques

A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques IMAGING A Comparison of the Uniformity Requirements for SPECT Image Reconstruction Using FBP and OSEM Techniques Lai K. Leong, Randall L. Kruger, and Michael K. O Connor Section of Nuclear Medicine, Department

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER

COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER COUNT RATE AND SPATIAL RESOLUTION PERFORMANCE OF A 3-DIMENSIONAL DEDICATED POSITRON EMISSION TOMOGRAPHY (PET) SCANNER By RAMI RIMON ABU-AITA A THESIS PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY

More information

486 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 3, JUNE 2007

486 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 3, JUNE 2007 486 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 3, JUNE 2007 Count-Rate Dependent Component-Based Normalization for the HRRT Mario Rodriguez, Jeih-San Liow, Member, IEEE, Shanthalaxmi Thada, Merence

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

DUE to beam polychromacity in CT and the energy dependence

DUE to beam polychromacity in CT and the energy dependence 1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact

More information

Continuation Format Page

Continuation Format Page C.1 PET with submillimeter spatial resolution Figure 2 shows two views of the high resolution PET experimental setup used to acquire preliminary data [92]. The mechanics of the proposed system are similar

More information

2005 IEEE Nuclear Science Symposium Conference Record M Influence of Depth of Interaction on Spatial Resolution and Image Quality for the HRRT

2005 IEEE Nuclear Science Symposium Conference Record M Influence of Depth of Interaction on Spatial Resolution and Image Quality for the HRRT 2005 IEEE Nuclear Science Symposium Conference Record M03-271 Influence of Depth of Interaction on Spatial Resolution and Image Quality for the HRRT Stéphan Blinder, Marie-Laure Camborde, Ken R. Buckley,

More information

Investigation of Motion Induced Errors in Scatter Correction for the HRRT Brain Scanner

Investigation of Motion Induced Errors in Scatter Correction for the HRRT Brain Scanner Investigation of Motion Induced Errors in Scatter Correction for the HRRT Brain Scanner Jose M Anton-Rodriguez 1, Merence Sibomana 2, Matthew D. Walker 1,4, Marc C. Huisman 3, Julian C. Matthews 1, Maria

More information

NIH Public Access Author Manuscript J Nucl Med. Author manuscript; available in PMC 2010 February 9.

NIH Public Access Author Manuscript J Nucl Med. Author manuscript; available in PMC 2010 February 9. NIH Public Access Author Manuscript Published in final edited form as: J Nucl Med. 2010 February ; 51(2): 237. doi:10.2967/jnumed.109.068098. An Assessment of the Impact of Incorporating Time-of-Flight

More information

Time-of-Flight Technology

Time-of-Flight Technology Medical Review Time-of-Flight Technology Bing Bai, PhD Clinical Sciences Manager, PET/CT Canon Medical Systems INTRODUCTION Improving the care for every patient while providing a high standard care to

More information

QIBA PET Amyloid BC March 11, Agenda

QIBA PET Amyloid BC March 11, Agenda QIBA PET Amyloid BC March 11, 2016 - Agenda 1. QIBA Round 6 Funding a. Deadlines b. What projects can be funded, what cannot c. Discussion of projects Mechanical phantom and DRO Paul & John? Any Profile

More information

Quantitative capabilities of four state-of-the-art SPECT-CT cameras

Quantitative capabilities of four state-of-the-art SPECT-CT cameras Seret et al. EJNMMI Research 2012, 2:45 ORIGINAL RESEARCH Open Access Quantitative capabilities of four state-of-the-art SPECT-CT cameras Alain Seret 1,2*, Daniel Nguyen 1 and Claire Bernard 3 Abstract

More information

Medical Imaging BMEN Spring 2016

Medical Imaging BMEN Spring 2016 Name Medical Imaging BMEN 420-501 Spring 2016 Homework #4 and Nuclear Medicine Notes All questions are from the introductory Powerpoint (based on Chapter 7) and text Medical Imaging Signals and Systems,

More information

PURE. ViSION Edition PET/CT. Patient Comfort Put First.

PURE. ViSION Edition PET/CT. Patient Comfort Put First. PURE ViSION Edition PET/CT Patient Comfort Put First. 2 System features that put patient comfort and safety first. Oncology patients deserve the highest levels of safety and comfort during scans. Our Celesteion

More information

The Design and Implementation of COSEM, an Iterative Algorithm for Fully 3-D Listmode Data

The Design and Implementation of COSEM, an Iterative Algorithm for Fully 3-D Listmode Data IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 20, NO. 7, JULY 2001 633 The Design and Implementation of COSEM, an Iterative Algorithm for Fully 3-D Listmode Data Ron Levkovitz, Dmitry Falikman*, Michael Zibulevsky,

More information

PERFORMANCE MEASUREMENTS OF POSITRON EMISSION TOMOGRAPHS

PERFORMANCE MEASUREMENTS OF POSITRON EMISSION TOMOGRAPHS NEMA NU 2 PERFORMANCE MEASUREMENTS OF POSITRON EMISSION TOMOGRAPHS NEMA Standards Publication NU 2-2001 Performance Measurements of Positron Emission Tomographs Published by National Electrical Manufacturers

More information

3-D PET Scatter Correction

3-D PET Scatter Correction Investigation of Accelerated Monte Carlo Techniques for PET Simulation and 3-D PET Scatter Correction C.H. Holdsworth, Student Member, IEEE, C.S. Levin", Member, IEEE, T.H. Farquhar, Student Member, IEEE,

More information

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

USING cone-beam geometry with pinhole collimation,

USING cone-beam geometry with pinhole collimation, IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE 2009 687 A Backprojection-Based Parameter Estimation Technique for Skew-Slit Collimation Jacob A. Piatt, Student Member, IEEE, and Gengsheng L.

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!

More information

Non-Stationary CT Image Noise Spectrum Analysis

Non-Stationary CT Image Noise Spectrum Analysis Non-Stationary CT Image Noise Spectrum Analysis Michael Balda, Björn J. Heismann,, Joachim Hornegger Pattern Recognition Lab, Friedrich-Alexander-Universität Erlangen Siemens Healthcare, Erlangen michael.balda@informatik.uni-erlangen.de

More information

Machine and Physics Data Guide

Machine and Physics Data Guide WWW..COM Machine and Physics Data Guide STANDARD IMAGING, INC. 3120 Deming Way Middleton, WI 53562-1461 May / 2008 2008 Standard Imaging, Inc. TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 www.standardimaging.com

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

PET Quantification using STIR

PET Quantification using STIR PET Quantification using STIR STIR User s Meeting Charalampos Tsoumpas, PhD King s College London Hammersmith Imanet 1 PET Quantification Image elements should correspond to concentration of the injected

More information

VALIDATION OF GATE MONTE CARLO SIMULATIONS FOR INDIUM 111 IMAGING

VALIDATION OF GATE MONTE CARLO SIMULATIONS FOR INDIUM 111 IMAGING VALIDATION OF GATE MONTE CARLO SIMULATIONS FOR INDIUM IMAGING Karine Assié, Isabelle Gardin, Pierre Véra, and Irène Buvat Abstract-- Accurate quantification of indium ( In) Single Photon Emission Computed

More information

Fast Timing and TOF in PET Medical Imaging

Fast Timing and TOF in PET Medical Imaging Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part

More information

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Marc Kachelrieß, a Katia Sourbelle, and Willi A. Kalender Institute of Medical Physics, University of

More information

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Joonki Noh, Jeffrey A. Fessler EECS Department, The University of Michigan Paul E. Kinahan Radiology Department,

More information

STIR User s Meeting. Reconstruction of PET data acquired with the BrainPET using STIR. 31. October 2013 Liliana Caldeira

STIR User s Meeting. Reconstruction of PET data acquired with the BrainPET using STIR. 31. October 2013 Liliana Caldeira STIR User s Meeting Reconstruction of PET data acquired with the BrainPET using STIR 31. October 2013 Liliana Caldeira Outline 3T MR-BrainPET System Reading BrainPET data into STIR FBP reconstruction OSEM/OSMAPOSL

More information

Addition of Noise by Scatter Correction Methods in PVI

Addition of Noise by Scatter Correction Methods in PVI IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 41, NO. 4, AUGUST 1994 1551 Addition of Noise by Scatter Correction Methods in PVI.J. Scott Barneyt, Ronald Harropl and M. Stella Atkinsi t Div. of Nuclear Medicine,

More information

GPU implementation for rapid iterative image reconstruction algorithm

GPU implementation for rapid iterative image reconstruction algorithm GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial

More information

Effect of normalization method on image uniformity and binding potential estimates on micropet

Effect of normalization method on image uniformity and binding potential estimates on micropet Effect of normalization method on image uniformity and binding potential estimates on micropet Marie-Laure Camborde, Arman Rhamim, Danny F. Newport, Stefan Siegel, Ken R. Buckley, Eric Vandervoort, Thomas

More information

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing ImPACT Information Leaflet No. 1: CT Scanner Acceptance Testing Version 1.02, 18/05/01 CONTENTS: 1. SCOPE OF LEAFLET 2. GENERAL PRINCIPLES OF ACCEPTANCE AND COMMISSIONING 2.1 PHANTOMS 2.2 EXPOSURE AND

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60601-2-44 2001 AMENDMENT 1 2002-09 Amendment 1 Medical electrical equipment Part 2-44: Particular requirements for the safety of X-ray equipment for computed tomography Amendement

More information

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry Sasha A. Philips, Frazier Bronson, Ram Venkataraman, Brian M. Young Abstract--Activity measurements require knowledge of the

More information

Optimisation of Toshiba Aquilion ONE Volume Imaging

Optimisation of Toshiba Aquilion ONE Volume Imaging Optimisation of Toshiba Aquilion ONE Volume Imaging Jane Edwards, RPRSG Royal Free London NHS Foundation Trust Dr Mufudzi Maviki, Plymouth Hospitals NHS Trust Background In 2011/12 Radiology at RFH was

More information

A study on image quality provided by a kilovoltage cone-beam computed tomography

A study on image quality provided by a kilovoltage cone-beam computed tomography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 1, 2013 A study on image quality provided by a kilovoltage cone-beam computed tomography Julia Garayoa a and Pablo Castro Servicio de Radiofísica,

More information

Brilliance CT Big Bore.

Brilliance CT Big Bore. 1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam

Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam Home Search Collections Journals About Contact us My IOPscience Iterative and analytical reconstruction algorithms for varying-focal-length cone-beam projections This content has been downloaded from IOPscience.

More information

Characterizing the parallax error in multi-pinhole micro-spect reconstruction

Characterizing the parallax error in multi-pinhole micro-spect reconstruction Characterizing the parallax error in multi-pinhole micro-spect reconstruction B. Vandeghinste, Student Member, IEEE, J. De Beenhouwer, R. Van Holen, Member, IEEE, S. Vandenberghe, and S. Staelens Abstract

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

A publicly accessible Monte Carlo database for validation purposes in emission tomography

A publicly accessible Monte Carlo database for validation purposes in emission tomography Short communication A publicly accessible Monte Carlo database for validation purposes in emission tomography I. Castiglioni 1, I. Buvat 2, G. Rizzo 1, M. C. Gilardi 1, J. Feuardent 2, F. Fazio 1 1 IBFM-CNR,

More information

2007 IEEE Nuclear Science Symposium Conference Record M26-296

2007 IEEE Nuclear Science Symposium Conference Record M26-296 2007 IEEE Nuclear cience ymposium Conference Record M26-296 A Global and a egmented Plane catter Calibration: Improving the Quantitative Accuracy of Frames with High Random Fraction and/or Low Number of

More information

Symbia E and S System Specifications Answers for life.

Symbia E and S System Specifications Answers for life. www.siemens.com/mi Symbia E and S System Specifications Answers for life. Symbia E: Work with confidence. What does it mean to work with confidence? It is clarity not only image clarity but the clarity

More information

A new PET prototype for proton therapy: comparison of data and Monte Carlo simulations

A new PET prototype for proton therapy: comparison of data and Monte Carlo simulations A new PET prototype for proton therapy: comparison of data and Monte Carlo simulations V. Rosso, G.Battistoni, N. Belcari, N. Camarlinghi, A. Ferrari, S. Ferretti, A. Kraan, A. Mairani, N. Marino, J. E.

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT CLINICAL COMMISSIONING AND ACCEPTANCE TESTING OF A 3D SURFACE MATCHING SYSTEM Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago Learning Objective Describe acceptance

More information

Simulations in emission tomography using GATE

Simulations in emission tomography using GATE Simulations in emission tomography using GATE Irène Buvat buvat@imed.jussieu.fr Laboratory of Functional Imaging, U678 INSERM, Paris, France Outline Emission tomography and need for simulations GATE short

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory Introduction In this lab exercise, you will investigate the linearity of the DeskCAT scanner by making measurements

More information

The great interest shown toward PET instrumentation is

The great interest shown toward PET instrumentation is CONTINUING EDUCATION PET Instrumentation and Reconstruction Algorithms in Whole-Body Applications* Gabriele Tarantola, BE; Felicia Zito, MSc; and Paolo Gerundini, MD Department of Nuclear Medicine, Ospedale

More information

Constructing System Matrices for SPECT Simulations and Reconstructions

Constructing System Matrices for SPECT Simulations and Reconstructions Constructing System Matrices for SPECT Simulations and Reconstructions Nirantha Balagopal April 28th, 2017 M.S. Report The University of Arizona College of Optical Sciences 1 Acknowledgement I would like

More information

UvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication

UvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication UvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication Citation for published version (APA): Kruis, M. F. (2014). Motion compensation for 4D PET/CT General

More information

Qalitätssicherung an Multileafkollimatoren. Dr. Lutz Müller Würzburg jan 2004

Qalitätssicherung an Multileafkollimatoren. Dr. Lutz Müller Würzburg jan 2004 Qalitätssicherung an Multileafkollimatoren Dr. Lutz Müller Würzburg jan 2004 IMRT Verification - present Target Volume Constraints Inverse Planning Algorithm Fluence Map Leaf Sequencer Leaf & Gantry sequence

More information

Advanced Image Reconstruction Methods for Photoacoustic Tomography

Advanced Image Reconstruction Methods for Photoacoustic Tomography Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

Positron. MillenniumVG. Emission Tomography Imaging with the. GE Medical Systems

Positron. MillenniumVG. Emission Tomography Imaging with the. GE Medical Systems Positron Emission Tomography Imaging with the MillenniumVG GE Medical Systems Table of Contents Introduction... 3 PET Imaging With Gamma Cameras PET Overview... 4 Coincidence Detection on Gamma Cameras...

More information

in PET Medical Imaging

in PET Medical Imaging Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part

More information

RADIOMICS: potential role in the clinics and challenges

RADIOMICS: potential role in the clinics and challenges 27 giugno 2018 Dipartimento di Fisica Università degli Studi di Milano RADIOMICS: potential role in the clinics and challenges Dr. Francesca Botta Medical Physicist Istituto Europeo di Oncologia (Milano)

More information