Data-driven optimal binning for respiratory motion management in PET
|
|
- Steven Baldwin
- 5 years ago
- Views:
Transcription
1 Data-driven optimal binning for respiratory motion management in PET Adam L. Kesner a) Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, USA Joseph G. Meier Department of Imaging Physics, University of Texas MD Anderson Cancer Center, Houston, TX, USA Darrell D. Burckhardt Siemens Medical Solutions USA, Inc., Hoffman Estates, IL, USA Jazmin Schwartz Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, USA David A. Lynch Department of Radiology, National Jewish Health, Denver, CO, USA (Received 17 May 217; revised 24 October 217; accepted for publication 24 October 217; published 2 November 217) Purpose: Respiratory gating has been used in PET imaging to reduce the amount of image blurring caused by patient motion. Optimal binning is an approach for using the motion-characterized data by binning it into a single, easy to understand/use, optimal bin. To date, optimal binning protocols have utilized externally driven motion characterization strategies that have been tuned with populationderived assumptions and parameters. In this work, we are proposing a new strategy with which to characterize motion directly from a patient s gated scan, and use that signal to create a patient/instance-specific optimal bin image. Methods: Two hundred and nineteen phase-gated FDG PET scans, acquired using data-driven gating as described previously, were used as the input for this study. For each scan, a phase-amplitude motion characterization was generated and normalized using principle component analysis. A patient-specific optimal bin window was derived using this characterization, via methods that mirror traditional optimal window binning strategies. The resulting optimal bin images were validated by correlating quantitative and qualitative measurements in the population of PET scans. Results: In 53% (n = 115) of the image population, the optimal bin was determined to include 1% of the image statistics. In the remaining images, the optimal binning windows averaged 6% of the statistics and ranged between 2% and 9%. Tuning the algorithm, through a single acceptance window parameter, allowed for adjustments of the algorithm s performance in the population toward conservation of motion or reduced noise enabling users to incorporate their definition of optimal. In the population of images that were deemed appropriate for segregation, average lesion SUV max were 7.9, 8.5, and 9. for nongated images, optimal bin, and gated images, respectively. The Pearson correlation of FWHM measurements between optimal bin images and gated images were better than with nongated images,.89 and.85, respectively. Generally, optimal bin images had better resolution than the nongated images and better noise characteristics than the gated images. Discussion: We extended the concept of optimal binning to a data-driven form, updating a traditionally one-size-fits-all approach to a conformal one that supports adaptive imaging. This automated strategy was implemented easily within a large population and encapsulated motion information in an easy to use 3D image. Its simplicity and practicality may make this, or similar approaches ideal for use in clinical settings. 217 American Association of Physicists in Medicine [ Key words: adaptive imaging, automatic binning, data driven, gated PET, optimal binning, personalized imaging 1. INTRODUCTION Positron Emission Tomography (PET) is an established imaging technology that has useful application across diverse areas of medicine. A limitation of PET is that it suffers from degradation caused by the patient respiratory motion, which in turn diminishes its diagnostic potential. Gating can be used to minimize the effects of motion, but at the expense of obtaining count-limited images. Developing robust methods for overcoming this issue, in a clinically efficacious manner, is fundamentally important to the goal of implementing respiratory motion correction of PET images in a clinical setting. 1 3 Our challenge has been well articulated in the title of a recent publication Respiratory Motion Handling Is Mandatory to Accomplish the High-Resolution PET Destiny. 1 While we have seen more than a decade of research and commercial solutions introduced into the field, respiratory motion solutions have not been adopted on a large scale. 4 We speculate that the 277 Med. Phys. 45 (1), January /218/45(1)/277/1 217 American Association of Physicists in Medicine 277
2 278 Kesner et al.: Data driven optimal binning 278 additional efforts associated with acquiring gated data, along with the uncertainties associated with using gated images, present the major hurdles preventing its acceptance and application. In this context, there remains room for the development of alternative strategies to generate motion-corrected images. 5 Data-driven strategies are enabled by modern computing technologies and offer new prospects for implementing practical motion correction in PET. 6 Data-driven gating has been presented as a solution to help with the practical acquisition of gated data. Data-driven gating can be used to acquire a motion characterization of a patient that is similar to one collected from hardware device, but is achieved in an operator independent, fully automated manner, and requires no changes to clinical protocols. 7 Research is progressing in this field 8,9 and, since data-driven motion characterization requires nothing more than automated background processing, it is reasonable to envision a scenario in which motion characterization becomes ubiquitously available in clinical PET in the near future. In order to make such a situation reality, finding methods for using this information practically in clinical environments may be more relevant in coming years than it has been in the past, and is therefore gaining urgency. Many methods have been proposed for sorting motion-characterized PET data. 1,11 Motion correction in PET typically involves segregating raw data into bins, for each of which an image is reconstructed separately. The suitability of the sorting strategy is fundamentally important because it dictates the quality of the final images along with the risks of degrading the images diagnostic potential. Amplitude- or phase-based gating strategies are the most popular methods for dealing with motion-characterized PET data. These methods segregate data based on its relation to the amplitude or phase of the respiratory cycle. While the simplicity of these strategies is attractive, they are not ideal for use in the diagnostic clinic. First, data may be sorted into too many or too few bins. Patients have different tracer distributions and move to different extents, and the ideal number of gates is not fixed but rather unique to a patient/tracer/acquisition. In fact, a large portion of patients do not benefit at all from the gating effort 12,13 and their optimal image is nongated. Once 4D images are generated, physicians are asked to read motion-corrected image sets that are more numerous than the traditional nongated form, all of which have greater noise characteristics and may or may not be useful to a patient s diagnosis. Thus, physicians are weary of bringing motion correction technology, which comes with more work and more risk, into the clinic. An alternative strategy which has been presented for utilizing 4D data is optimal binning Optimal binning is a strategy aimed at using a subset of the originally acquired data to create a single, optimally binned, reduced motion image. Optimal binning has some practical advantages that make it appealing for clinical consideration and development: (1) Optimal binning is easy to understand: utilization of a subset of data (2) Optimal binning is easy to process: reconstruction of a subset of data (3) Optimal binning is easy to use clinically: output is a standard 3D image and works with any 3D image reading software (4) Optimal binning is practical to implement: processing is fully automated. (5) Optimal binning is easy to integrate in clinical procedures: presents an operator independent standardizable approach to handling motion The concept of optimal binning is that data are sorted such that the resultant image embodies a theoretical optimization between the resolution improvement and image statistics. Optimal binning has traditionally been implemented using all the data that are near the end expiration of the respiratory cycle, the quiescent period, as defined using an external hardware motion characterization device 17 such as a pressure belt or camera. The size of the optimal bin has in the past been determined from a population of respiratory traces; values of 35% 16 and 4% 15 of the total counts have been used. A problem with traditional optimal binning strategies is that they are based on theoretical optimization and not directly connected to specific images or acquisitions. Also they assume that the motion characterization is accurately associated with scan data. In this work, we introduce a new approach of applying data-driven principles to carry out an optimal binning strategy. This modification of an old strategy can potentially have significant implications because the optimization can now conform to the instance of data. The assumption that an external signal and internal motion correlate is no longer requisite. Also, by introducing an automated data-driven sorting algorithm, and combining it with data-driven gating, an entire datadriven motion correction workflow can be implemented and is exemplified in this work. The ease of a 1% fully automated software motion correction workflow option, as we present, could likely be integrated into clinics with ease. 18 In this work, we present data-driven optimal binning concepts, incorporated them into a data-driven motion correction workflow, and validated them in a population of scans to establish the proof of principle for automated adaptive motion correction in PET imaging. 2. MATERIALS AND METHODS Our methods for creating a data-driven optimally binned representation of a 4D data set consist of three main steps. A flowchart of the data-driven optimal binning workflow is shown in Fig A. Data driven optimal binning workflow Step 1 Generating a data-driven phase amplitude relationship characterization from a 4D phase-gated data set. A 4D phase-gated data set can be created from any PET acquisition that has (prospective or retrospectively collected) motion characterization associated with it by sorting image data
3 279 Kesner et al.: Data driven optimal binning 279 FIG. 1. Flow chart of data-driven optimal binning workflow. [Color figure can be viewed at wileyonlinelibrary.com] evenly across phases of the cycle. 19 The set of phase-gated images can then be analyzed to quantify motion content measures within each phase image to generate what we call a phase amplitude transfer function (PATF). This function serves to define an amplitude of motion measure for each phase of a patient s respiratory cycle. There are theoretically several ways to do this, including using center of mass measures or edge detection algorithms. In our work, we used principle component analysis. 2 Specifically, a phase amplitude relationship was described for each scan by plotting the first principle component (PC) weight array against phase. The first PC should adequately provide this characterization because it represents the greatest coherent signal variation in the field of view over the gated image cycle a measure which conveniently correlates with respiratory motion if it exists in the image. In this way, an amplitude of motion characterization can be derived from a phase-gated data set and used to define a patient-specific PATF. Example PATFs are shown in Fig. 2. Our strategy of associating phase and amplitude of motion in this step is in fact inspired by similar work; 21,22 the advancement of our work, however, is that we used an entirely internal, data-driven method for capturing the amplitude of motion signal from phase-gated data. The algorithm is compartmentalized such that the source of the motion characterization is immaterial. Step 2 Determining a scan-specific optimal binning window from phase amplitude motion characterization and using it to assign sorting of data. The PATF curve generated in step 1 is useful for describing patient motion over the cycle. Its units represent a measure of the total coherent changes in signal across a collection of image voxels over phase gates. We know in theory that we want to more aggressively subjugate data of patients who exhibit significant amounts of motion in their scans, while at the other end of the spectrum, we want to retain count statistics in patient scans that have no useful motion information in them. All images processed will be assigned some PATF by mathematical definition of principle component analysis, and we need to generate an understanding of its seemingly arbitrary magnitude to understand its significance. In order to gain perspective on the amplitude of the first PC curve, we compare it to the third PC curve, which is derived from the data in the same manner as the first PC. We do this because we assume that the first (and possibly second) PC components describe any true cycle respiratory motion. It is assumed that the third PC represents the amplitude of PC fluctuations that result from nonrespiratory motion, for example, noise, and we can use it to normalize the first PC curve. The magnitude of third PC fluctuations is used to create an amplitude-based acceptance window. Specifically we used the third PC curve to calibrate the magnitude of nonrespiratory fluctuations in our first PC signal, with the reasoning that this is the size of the window for random fluctuations and values within this window cannot confidently be labeled as dissimilar. More specifically, we defined our amplitude acceptance window as 3 standard deviations (SDs) of the third PC curve. This value was chosen using the assumption that the third PC represents Poisson noise and first PC curve variation changes outside 3 SDs are likely caused by nonnoise phenomena (e.g., motion). This single variable (width of acceptance window) can be adjusted in future works depending on the user s desire to be more or less conservative with respect to resolution and noise, and is considered further in the Results and Discussion. Once the amplitude acceptance window is defined, the PATF (first PC) is analyzed, and the portion of the curve that contains the most data points within the window s amplitude is chosen to define the optimal bin phase inclusion set. This is achieved by brute-force processing all permutations of bins are calculated and the window with maximum number of phases included in it is used. Examples of third PC-derived acceptance windows and phase inclusion selection are shown in Figs. 2(c) and 2(d). Step 3 Combining sorted data to obtain optimal image. A data-driven optimal bin image can be generated by combining data from accepted phases in a single, optimal bin. In this work, sorting was achieved by adding accepted phase image data together. 2.B. Clinical image analysis A population of 219 FDG PET scans were processed and reviewed. The images were acquired in a previous study 8 from a population of 112 individual patients undergoing single or dual time point clinically indicated PET/CT for radiologic evidence of pulmonary nodules measuring 6 2 mm, for whom PET/CT is clinically indicated. All scanning was performed on a BioGraph TruePoint PET/CT scanner (Siemens, Knoxville, TN, USA) having a single bed field of view of 22 cm. Injected radioactivity ranged between 5 and 73 MBq. Patients underwent standard whole-body PET/CT
4 28 Kesner et al.: Data driven optimal binning 28 Rela ve amplitude (normalized rel PCA weight) (a) Example PATF from image with useful mo on informa on Gate (phase gates) (b) Rela ve amplitude (normalized rel PCA weight) Example PATF from image with no useful mo on informa on Gate (phase gates) (c) Rela ve amplitude (normalized rel PCA weight) Example PATF from image with useful mo on informa on Gate (phase gates) (d) Rela ve amplitude (normalized rel PCA weight) Example PATF from image with no useful mo on informa on Gate (phase gates) Phase-mo on amplitude rela onship (1st PC) Non first principle component fluctua ons phase-mo on amplitude rela onship (3nd PC) Phased included in op mal window Acceptance window (= ± 3 * standard devia ons of non first principle component fluctua ons) FIG. 2. Illustration of phase amplitude transfer functions in two example patients. (a) A curve representing a patient with a significant amount of detectable motion. (b) A curve representing a patient with a nonsignificant amount of detectable motion. (c and d) Illustrations showing how the third principle component of a signal is used to define an amplitude acceptance window. The window, overlaid on the 1st principle component in the figure, is used to determine the phases that are accepted in the optimal bin, in a patient with a significant amount of detectable motion (c); and in a patient with a nonsignificant amount of detectable motion (d). Y axis shown scaled between and 1 for display. [Color figure can be viewed at wileyonlinelibrary.com] examinations, followed by a hardware-gated PET examination of the chest. Acquisition duration was 6 min, and acquisitions occurred 6 and/or 12 min after injection. Approximately 2% of the study sample had noticeable FDG uptake in lung lesions. Full accounting of acquisition criteria and scan settings can be found in the original reference. 8 4D respiratorygated PET data sets were generated using data-driven gating and phase-based binning (1 bins per cycle) 8 and used for primary analysis. Additional image sets generated using hardware gating system, and random gating triggers were also acquired and assessed for comparison. Hardware-driven respiratory gating was implemented using the Anzai pressure belt (Anzai Medical, Tokyo, Japan). All scans, including gated scans, were reconstructed using Siemens reconstruction software, with attenuation-weighted, ordered subsets expectation maximization reconstruction, two iterations, six subsets, 6-mm isotropic Gaussian post reconstruction filter, and delayed randoms correction. For each gated scan, a datadriven phase amplitude motion characterization was generated in the form of a PATF using methods described in this paper. To improve processing, the principle component analysis was processed using a truncated field of view five axial slices at each edge of the scan were cut to remove the low sensitivity/higher noise region. Three images were generated for each patient for the purpose of comparison: (a) nongated, (b) optimal bin, and (c) phase gated. Nongated images were defined as the sum of the gated images so as to reduce the variables specific to reconstruction. For quantitative analysis, we looked at noise and uptake measurements in the different images. Specifically, noise
5 281 Kesner et al.: Data driven optimal binning 281 was studied by quantifying the %SD in liver volumes of interest (VOIs), defined with a 71 cc spherical volume of interest (VOI) and placed on all images with obvious liver presentation in the field of view. Uptake measurements were taken by defining 8% maximum threshold VOIs on the subset of images that had obvious lesion uptake. Fullwidth half-maximum (FWHM) measurements were acquired from lesion superior inferior profiles taken at the VOI s center of mass, and reported for cases that had high enough signal to background ratios to make the measurement appropriate. For qualitative analysis, we compared observed motion with expected distribution of statistics in optimally binned images. In our previous study, 8 a qualitative assessment of motion was graded for each scan using three independent blinded readers. Two nuclear medicine physicians and one medical physicist, all with 1+ years experience and ABR certification, reviewed looping animations of phase-gated images, presented in random order on the screen. The animations consisted of coronal slice images with slice locations that had been preselected per scan for their proximity to either lung lesions or heart regions when there were no lesions present, determined through in-house 4D image navigation software, in an effort to present the slice that would most likely be affected by respiratory motion. The readers were asked to record whether or not (1 or ) they observed detectable motion in the looping image set. This qualitative assessment was performed independently of the present work, approximately 2 yr prior and with no knowledge it would be used in this study. 8 In this work, we look at the distribution of the number of gates included in the optimal bin across the population and compared them with the independent reader analysis. 3. RESULTS Two hundred and nineteen gated PET scans were processed and each binned into a single, optimal, 3D image. Given input sets of phase-gated images, processing only required approximately 5 s on a standard desktop PC. Given the default parameters: data-driven gating and an acceptance window of three standard deviations, in 53% (n = 115) of the scans, the optimal bin was determined to include 1% of the image statistics, that is, the patients did not noticeably benefit from the gating effort and the optimal bin was the nongated embodiment. For the remaining images, the optimal statistics binning windows contained an average of 6% of the statistics found in the nongated image and ranged between 2% and 9%. A histogram of the population data is shown in Fig. 3(a). Also shown in this figure are histogram statistics for alternative acceptance window settings (1 SD, 2 SD, 3 SD, and 4 SD) [Fig. 3(b)] and other source gating methods (hardware gating and random gating) [Fig. 3(c)]. The random gating method was included to show how the algorithm handles images that we know contain no useful motion information (motion should not be captured randomly). 98% of these scans were determined optimal with 1% of the statistics. To quantitatively asses the method, we looked at the effects of optimal binning on noise in the liver and on lesion quantification. In total, 12 images had well-defined liver in the FOV that would fit the 71 cc Spherical VOI. Since our method adapts to patient data and segregates data differently for each patient, it is not possible to characterize the noise implications of the algorithm in one single value. We found that noise, defined as SD/mean (%SD), correlated with statistics. When more statistics were included in an optimal bin, relative noise levels were lower. It is likely that noise is simply a function of statistics, which are clearly defined in our algorithm as a percentage of total. To test this hypothesis, we plotted theoretical Poisson noise defined as (sum SUV in VOI/sum SUV in image 9 # net trues)^ (1/2) vs. measured %SD (Fig. 4). The figure shows that counts (/theoretical noise) may be a predictor of measured noise, although the relationship is not purely Poisson as demonstrated in the nonunity slope. This divergence from theory is likely due to the fact that we are using iterative reconstruction with is known to be nonlinear. 23 Lesion VOIs were placed in 44 patient images that had definable lesions as described in methods. Generally speaking, the population average of the mean VOI measurement was highest in the gated images, when averaged over gates. However, we may expect this by definition as the gated images had the highest noise and smallest threshold VOI volumes. A summary of population lesion, VOI volume, and FWHM measurements is shown in Table I. To qualitatively asses the performance of the algorithm, we asked three observers to review animated phase-gated data (a central coronal slice looping through gates) throughout the population of scans and in random order. 89% of images assessed as not displaying motion, through agreement with three blind readers, were presented as optimal using 1% statistics qualitatively validating the tendency of the algorithm to leave scans without noticeable motion information in their nongated embodiment. Of the scans that were observed to have motion by all three readers, 78% of them were segregated into less than 1 bins, again illustrating the tendency of the algorithm to appropriately segregate scans with useful information. Visual inspection of the optimally binned images consistently showed better resolution when compared to the nongated images, and better noise characteristics than the original phase-gated images (which by definition were noisier because they contain less data). Example patient images are shown in Fig. 5 and online animations 1 and 2. An example lesion profile is shown in Fig DISCUSSION Optimal binning is a strategy for implementing motion correction in the clinic that is simple to understand and straightforward to use. In the work presented here, we extended the concept of optimal binning to a data-driven
6 282 Kesner et al.: Data driven optimal binning 282 (a) Number of images (b) Number of images (c) Number of images Histogram of op mal bin sta s cs in popula on Software gating, Acc window = 3 SD 1% 2% 3% 4% 5% 6% 7% 8% 9% 1% Por on of total sta s cs in op mal bin Histogram of op mal bin sta s cs in popula on Software gating, Acc window = 1 SD Software gating, Acc window = 3 SD Software gating, Acc window = 2 SD Software gating, Acc window = 4 SD 1% 2% 3% 4% 5% 6% 7% 8% 9% 1% Por on of total sta s cs in op mal bin Histogram of op mal bin sta s cs in popula on So ware ga ng, Acc window = 3 SD Random ga ng, Acc window = 3 SD Hardware ga ng, Acc window = 3 SD 1% 2% 3% 4% 5% 6% 7% 8% 9% 1% Por on of total sta s cs in op mal bin FIG. 3. Histogram of optimal bin statistics showing the number of images in the population that retain specified statistics in their optimal embodiment, as determined by our algorithm. (a) Histogram created using default parameters. (b) Comparison of algorithm behavior when using different acceptance windows. (c) Histogram showing behavior of algorithm when phase-gated input data come from different sources. [Color figure can be viewed at wileyonlinelibrary.com] form, in a way that supports adaptive optimization of image data. Our approach can be tuned and employed in an automated and standardized fashion. We have demonstrated (a) motion amplitude information can be derived from phasegated images, (b) optimal binning strategies can use this characterization to adapt data segregation to the instance of data in a way that correlates with independent, blinded reader analysis, (c) data-driven motion correction workflows, that span acquisition to reconstruction, can be applied in large populations with ease, and (d) adaptive motion correction algorithms are relevant because patients/scans contain varying capacities to support motion correction. In our quantitative analysis, the noise measurements exhibited behavior that we might expect. The greater the counts in our images, the lower the noise levels measured in uniform regions, as seen in Fig. 4. The lesion analysis was somewhat less conclusive. When averaged across the population, the SUV mean, SUV max, and lesion FWHM measurement in the optimal bin image were between the values for the gated and nongated embodiments, as could be expected. On closer analysis, specific changes varied among cases. In some instances, SUVs increased between gated and nongated image sets, while others decreased. This varying trend demonstrates the limitations of lesion analysis in differing noise environments. In our previous work, we have shown that the mere process of gating subjugating data into different bins can increase or decrease SUV measurements when studied in simulations. 24 In our further work, using this same study population, we found that SUV mean and SUV max measurements will increase in the population 13% 8 and 4% 8, respectively, from the segregation of data and with no incorporation of motion information. 8 Patients have different uptake patterns, have different extents of breathing motion, have varying counts in their images, and different centers use different scanners reconstruction protocols. The specific number of gates used for sorting data also affects measurement outcomes. Gating will affect lesion SUV measurement in different patients differently and therefore should be used cautiously when assessing the success of a method. Our qualitative analysis was performed by comparing the amount of data in optimal bins in the population and with blinded reader assessment of the images and their motion content. The fact that we did not find absolute agreement between the readers and the algorithm likely illustrates this assessment had limitations. The blind review was limited by the assessment of only a single slice rather than the whole volume, and the optimal binning algorithm was calibrated at a specific sensitivity which we expect would not necessarily match that of the reviewers. Despite these limitations, the correlations we found between observed motion in the images and the data segregation produced by the algorithm across the population indicate that the algorithm is largely performing as predicted and far from random or incorrect. Further qualitative improvements are demonstrated in the images provided. In Fig. 5(a), we can see that the lesion is better defined (smaller) and greater small structure detail in the optimal bin image than the nongated image. At the same time, the noise in the liver is favorable in the optimal bin image as compared to the gated image. More examples of these improvements are demonstrated in online Figs. 1 and 2. Within the data-driven optimal binning algorithm, there is a question which arises as to how best define the acceptance
7 283 Kesner et al.: Data driven optimal binning 283 3% Measured vs theore cal noise in liver Gated Nongated Op mal Bin Linear fit (Gated) Linear fit (Nongated) Linear fit (Op man Bin) 25% Measured % SD 2% 15% y =.363x +.12 R² = % 5% y =.1688x +.49 R² =.91 y =.841x R² =.462 % % 5% 1% 15% 2% 25% 3% Theore cal % SD (Poisson noise) FIG. 4. Measured vs. theoretical Poisson noise in the liver for different reconstructions. Linear best fit lines, with equations and R 2 values, shown on chart. [Color figure can be viewed at wileyonlinelibrary.com] TABLE I. Summary of lesion measurements, averaged across population. N Nongated Optimally binned Average gated value Best gated value Worst gated value SUV mean SUV max Volume (cc) (8% max thresh) FWHM a FWHM Pearson correlation between.85 optimally binned and nongated FWHM Pearson correlation between optimally.89 binned and gated a Average FWHM derived from scans that had appropriate profiles, scans not amiable to FWHM characterization were not included. window for separating noise and motion [visualized by the gray stripe in Figs. 2(c) and 2(d)]. We used a window of 3 standard deviations of the assumed noise magnitude for much of our analysis. However, we also tested other window sizes ranging from 1 4 SDs, as shown in Fig. 3b, and found that the window size affected the results as theory predicts. It is likely that this parameter should be calibrated for the desired sensitivity. If limiting noise in output images outweighs concerns of recovering resolution, a wider window could be used so to include more signal, but less motion information. The opposite is true as well; that is, a narrower window could be used to ensure that more detailed motion information is preserved, but at the cost of introducing a greater amount of noise in the optimal bin data. It is important to clarify that when working with optimal binning, or any 4D signal utilization strategy, there is no absolute true or correct way to segregate data. The goal of optimal binning is to find an optimal trade-off, and its definition will likely depend on the desired use of the data. What we know for sure, however, is what we want from the conceptual extremes. Data sets that do not benefit from data segregation should be left in their nongated embodiment, and those with useful information in them should incorporate segregation. Our application of adaptive optimal binning was successful in that it spanned this spectrum in a conformal manner, and in a way that correlated with reader observation. Optimal binning is a well-established concept. To our knowledge, all optimal binning strategies published to date use a fixed width window for segregating data, applied uniformly across a population. Modifying the method to be data driven, as we did, has several benefits: it is automated, straightforward to use and, perhaps most importantly, it conforms to the capacity of data in each image instance, making maximal use of the information available for the given patient s PET scan. The process results in an amplitude-based optimally binned 3D image. The method is unique in that it is scan-specific and can be used to optimize any instance of a gated image, irrespective of how it was created (gating method, reconstruction, etc.). Our strategy for optimal binning is based on a data-driven approach. It can be compartmentalized and implemented with standard hardware gating systems, as we did for the secondary measurement as shown in Fig. 3(c), or it can be integrated with data-driven gating to enable an entire data-driven motion correction workflow, 6 as demonstrated in this work. Full data-driven workflows are a new concept in the field that
8 284 Kesner et al.: Data driven optimal binning 284 (a) Phase gated, bin 1 stats: 1% Op mal bin stats: 5% Nongated stats: 1% (b) Phase gated, bin 1 stats: 1% Op mal bin stats: 1% Nongated stats: 1% FIG. 5. Example coronal phase-gated, optimal bin, and nongated image reconstructions for a patient who had a significant amount of detectable motion (a), and a patient who did not have a significant amount of detectable motion (b). The stats statistic shows the fraction of nongated data used in the reconstruction. The figure illustrates how the optimal bin image can conform to the capacity of the data. The images in this figure correspond to the patient data used in Fig. 2. [Color figure can be viewed at wileyonlinelibrary.com] Voxel ac vity (SUV) Example Lesion Profile Rela ve loca on (mm) single gate (FWHM = 7.9 mm) op mal bin (FWHM = 8.2 mm) average bin (FWHM = 1. mm) FIG. 6. Example lesion profile [Coronal images shown in Fig. 4(a)]. [Color figure can be viewed at wileyonlinelibrary.com] may have significant practical benefits because they can be introduced to clinical environments with ease. Data-driven gating allows us to characterize motion and create phasegated images from PET scans with no extra effort and no changes to clinical protocols, and in a standardized way. 8 Data-driven 4D signal utilization, the topic of this paper, allows us to create patient-specific solutions that conform to the information contained in the patient scans. The two methods go together elegantly: count limited data will not have useful respiratory signal and not benefit from gating, and our algorithm will detect this and revert the image back to its nongated embodiment, whereas traditional hardware-based/ population-calibrated methods would deliver a less useful image. This may be a key feature of our algorithm. If we want to address the problem of motion in diagnostic PET, we need solutions that do not degrade the subpopulation of images that do not benefit from the gating process. A limitation of this work may be in the assumption that we can characterize patient motion and the PATF using the first PC weight array and relative noise using the 3rd. Theory tells us that cyclic motion, and noise, should be represented in the measured first PC signal, when the data represent a single cycle scan (e.g., a respiratory-gated PET scan). The third PC likely contains little motion information and is thus a similar measure of noise. These are first-order assumptions and are variables that can be studied, and possibly improved upon in future work. Integrating motion correction in PET imaging can help with diagnosis, 25 localization, 26 and quantification, 27 and aid with radiotherapy planning. 28 The scope of this work has been to establish proof of principle for a new strategy of motion correction that addresses practical motion correction. Ideally, fully automated motion correction may in effect extend the dimensional imaging capacity of PET to inherently include respiratory motion. Specifically, it could support protocols where the complimentary motion-corrected images are generated alongside standard PET images, for all acquisitions with no changes to procedures, using methods similar to those presented here. Alternatively, data-driven motion correction workflows can be integrated with extended acquisitions to get higher count/higher quality motioncorrected images. The problem of motion is a well-articulated problem in PET imaging. In our work here, we developed an algorithm that addresses the drawbacks of traditional one-size-fits-all optimal binning solutions as well as the need for clinically
9 285 Kesner et al.: Data driven optimal binning 285 practical protocols. It is reasonable to expect that it can be improved upon with further methodological developments. A possible opportunity for improvement may be to incorporate the data segregation with reconstruction. In this work, the reconstructed phased-gated images that fell within the reference window were averaged together. Future work can also assess the benefit of resorting data prior to reconstruction, which should benefit the final images because iterative reconstruction, used in PET, is non-linear. Further improvement may be found by forcing the optimal bin to align with the corresponding attenuation map, to reduce the occurrence of artifacts stemming from attenuation mismatch. Ultimately, the goal is to use these methods to improve clinical use of PET. For that to happen, clinical utility of these methods must be assessed. 5. CONCLUSION We have presented a proof-of-principle study showing that it is possible to derive an amplitude characterization of motion from a phase-gated data set using data-driven strategies. We have also shown that we can use that characterization to generate adaptive optimal bin images. When applied to a population of FDG PET images, we found large variations in optimal bin characteristics of different patients, indicating a need and an appropriateness of developing personalized imaging approaches. Future work will include assessing these strategies in clinical operations, quantifying their benefit, and extending these methods to other modalities (SPECT, CT, MR) and other types of motion including cardiac motion. ACKNOWLEDGMENTS We would like to acknowledge the support of Siemens (Hoffman Estates, IL, USA), who provided grant support for study recruitment and image analysis. We would also like to thank specifically Ramya Rajaram and James Hamill of Siemens for their help in the acquisition/processing of study data. We would also like to thank Dr. Phillip Koo and Dr. Jennifer Kwak for their aid in the analysis as blinded readers. CONFLICTS OF INTEREST The authors have no relevant conflicts of interest to disclose. a) Author to whom correspondence should be addressed. Electronic mail: kesnera@mskcc.org; Telephone: REFERENCES 1. Daou D. Respiratory motion handling is mandatory to accomplish the high-resolution PET destiny. Eur J Nucl Med Mol Imaging. 28;35: Chi A, Nguyen NP. 4D PET/CT as a strategy to reduce respiratory motion artifacts in FDG-PET/CT. Front Oncol. 214;4: Lamare F, Ledesma Carbayo MJ, Cresson T, et al. List-mode-based reconstruction for respiratory motion correction in PET using non-rigid body transformations. Phys Med Biol. 27;52: Weber WA, Gatsonis CA, Mozley PD, et al. Repeatability of 18F-FDG PET/CT in advanced non-small cell lung cancer: prospective assessment in two multicenter trials. J Nucl Med. 215;56: Rahmim A, Rousset O, Zaidi H. Strategies for motion tracking and correction in PET. PET Clin. 27;2: Kesner A, Schleyer P, Buther F, Walter M, Schafers K, Koo P. On transcending the impasse of respiratory motion correction applications in routine clinical imaging - a consideration of a fully automated data driven motion control framework. EJNMMI Phys. 214;1:8. 7. Kesner AL, Kuntner C. A new fast and fully automated software based algorithm for extracting respiratory signal from raw PET data and its comparison to other methods. Med Phys. 21;37: Kesner AL, Chung JH, Lind KE, et al. Validation of software gating: a practical technology for respiratory motion correction in PET. Radiology. 216;281: Buther F, Vehren T, Schafers KP, Schafers M. Impact of data-driven respiratory gating in clinical PET. Radiology. 216;281: Dawood M, B uther F, Lang N, Schober O, Sch afers KP. Respiratory gating in positron emission tomography: a quantitative comparison of different gating schemes. Med Phys. 27;34: Pepin A, Daouk J, Bailly P, Hapdey S, Meyer ME. Management of respiratory motion in PET/computed tomography: the state of the art. Nucl Med Commun. 214;35: Guerra L, De Ponti E, Elisei F, et al. Respiratory gated PET/CT in a European multicentre retrospective study: added diagnostic value in detection and characterization of lung lesions. Eur J Nucl Med Mol Imaging. 212;39: Liu C, Pierce LA 2nd, Alessio AM, Kinahan PE. The impact of respiratory motion on tumor quantification and delineation in static PET/CT imaging. Phys Med Biol. 29;54: Van Der Gucht A, Serrano B, Hugonnet F, Paulmier B, Garnier N, Faraggi M. Impact of a new respiratory amplitude-based gating technique in evaluation of upper abdominal PET lesions. Eur J Radiol. 214;83: Liu C, Alessio A, Pierce L, et al. Quiescent period respiratory gating for PET/CT. Med Phys. 21;37: van Elmpt W, Hamill J, Jones J, De Ruysscher D, Lambin P, Ollers M. Optimal gating compared to 3D and 4D PET reconstruction for characterization of lung tumours. Eur J Nucl Med Mol Imaging. 211;38: Daouk J, Fin L, Bailly P, Meyer ME. Improved attenuation correction via appropriate selection of respiratory-correlated PET data. Comput Methods Programs Biomed. 28;92: Kesner AL. The relevance of data driven motion correction in diagnostic PET. Eur J Nucl Med Mol Imaging. 217;44: Nehmeh SA, Erdi YE, Ling CC, et al. Effect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancer. Med Phys. 22;29: Smith LI. A tutorial on Principal Components Analysis; pdf Accessed 4/2/ Liu C, Alessio AM, Kinahan PE. Respiratory motion correction for quantitative PET/CT using all detected events with internal-external motion correlation. Med Phys. 211;38: Silin R, Xiao J, Chung C, et al. Data-driven event-by-event respiratory motion correction using TOF PET list-mode centroid of distribution. Phys Med Biol. 217;62: Schmidtlein CR, Beattie BJ, Bailey DL, et al. Using an external gating signal to estimate noise in PET with an emphasis on tracer avid tumors. Phys Med Biol. 21;55: Kesner AL. A Simulation of Noise Effects on SUV Measurements in Gated/Ungated Data - A Story of Apples and Oranges European Association on Nuclear Medicine Annual Conference; 21; Vienna, Austria 25. van der Vos CS, Koopman D, Rijnsdorp S, et al. Quantification, improvement, and harmonization of small lesion detection with state-ofthe-art PET. Eur J Nucl Med Mol Imaging. 217;44: Polycarpou I, Tsoumpas C, King AP, Marsden PK. Impact of respiratory motion correction and spatial resolution on lesion detection in PET: a
10 286 Kesner et al.: Data driven optimal binning 286 simulation study based on real MR dynamic data. Phys Med Biol. 214;59: Lupi A, Zaroccolo M, Salgarello M, Malfatti V, Zanco P. The effect of 18F-FDG-PET/CT respiratory gating on detected metabolic activity in lung lesions. Ann Nucl Med. 29;23: Bettinardi V, Picchio M, Di Muzio N, Gianolli L, Gilardi MC, Messa C. Detection and compensation of organ/lesion motion using 4D-PET/CT respiratory gated acquisition techniques. Radiother Oncol. 21;96: SUPPORTING INFORMATION Additional Supporting Information may be found online in the supporting information tab for this article. Fig. S1. Sample coronal slices from population of patient PET scans showing nongated, optimally binned, and gated (1 gates) images side by side. Gates are looped in animation so that motion can be observed on the gated images. Fig. S2. Sample coronal slices, zoomed in over lesion, from population of patient PET scans showing nongated, optimally binned, and gated (1 gates) images side by side. Gates are looped in animation so that motion can be observed on the gated images.
Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies
g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,
More informationWhite Paper. EQ PET: Achieving NEMAreferenced. Technologies. Matthew Kelly, PhD, Siemens Healthcare
White Paper EQ PET: Achieving NEMAreferenced SUV Across Technologies Matthew Kelly, PhD, Siemens Healthcare Table of Contents Introduction 1 Case Study 1 Cross-Scanner Response Assessment 2 Clinical Example
More informationUvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication
UvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication Citation for published version (APA): Kruis, M. F. (2014). Motion compensation for 4D PET/CT General
More informationImprovement of contrast using reconstruction of 3D Image by PET /CT combination system
Available online at www.pelagiaresearchlibrary.com Advances in Applied Science Research, 2013, 4(1):285-290 ISSN: 0976-8610 CODEN (USA): AASRFC Improvement of contrast using reconstruction of 3D Image
More informationQIBA PET Amyloid BC March 11, Agenda
QIBA PET Amyloid BC March 11, 2016 - Agenda 1. QIBA Round 6 Funding a. Deadlines b. What projects can be funded, what cannot c. Discussion of projects Mechanical phantom and DRO Paul & John? Any Profile
More informationImproving Positron Emission Tomography Imaging with Machine Learning David Fan-Chung Hsu CS 229 Fall
Improving Positron Emission Tomography Imaging with Machine Learning David Fan-Chung Hsu (fcdh@stanford.edu), CS 229 Fall 2014-15 1. Introduction and Motivation High- resolution Positron Emission Tomography
More information3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH
3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk
More informationGating, enhanced gating, and beyond: information utilization strategies for motion management, applied to preclinical PET
Kesner et al. EJNMMI Research 2013, 3:29 ORIGINAL RESEARCH Open Access Gating, enhanced gating, and beyond: information utilization strategies for motion management, applied to preclinical PET Adam Leon
More informationClassification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging
1 CS 9 Final Project Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging Feiyu Chen Department of Electrical Engineering ABSTRACT Subject motion is a significant
More informationQuantitative imaging for clinical dosimetry
Quantitative imaging for clinical dosimetry Irène Buvat Laboratoire d Imagerie Fonctionnelle U678 INSERM - UPMC CHU Pitié-Salpêtrière, Paris buvat@imed.jussieu.fr http://www.guillemet.org/irene Methodology
More informationWorkshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET
Workshop on Quantitative SPECT and PET Brain Studies 14-16 January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Físico João Alfredo Borges, Me. Corrections in SPECT and PET SPECT and
More informationRespiratory Motion Estimation using a 3D Diaphragm Model
Respiratory Motion Estimation using a 3D Diaphragm Model Marco Bögel 1,2, Christian Riess 1,2, Andreas Maier 1, Joachim Hornegger 1, Rebecca Fahrig 2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg 2
More informationUnconstrained image reconstruction with resolution modelling does not have a unique solution
Nuyts EJNMMI Physics 2014, 1:98 COMMENTARY OpenAccess Unconstrained image reconstruction with resolution modelling does not have a unique solution Johan Nuyts Correspondence: johan.nuyts@uz.kuleuven.be
More informationPMOD Features dedicated to Oncology Research
While brain research using dynamic data has always been a main target of PMOD s developments, many scientists working with static oncology data have also found ways to leverage PMOD s unique functionality.
More informationRADIOMICS: potential role in the clinics and challenges
27 giugno 2018 Dipartimento di Fisica Università degli Studi di Milano RADIOMICS: potential role in the clinics and challenges Dr. Francesca Botta Medical Physicist Istituto Europeo di Oncologia (Milano)
More informationRespiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data
Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data Christopher M Rank 1, Thorsten Heußer 1, Andreas Wetscherek 1, and Marc Kachelrieß 1 1 German Cancer
More informationAutomatic Lesion Detection for Measuring Response using Dynamic FDG-PET
Automatic Lesion Detection for Measuring Response using Dynamic FDG-PET Xiujuan Zheng a,b, Guangjian Tian a, Shaoli Song b, Gang Huang b, David Dagan Feng a,c a Department of Electronic and Information
More informationCT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION
CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology
More information8/3/2017. Contour Assessment for Quality Assurance and Data Mining. Objective. Outline. Tom Purdie, PhD, MCCPM
Contour Assessment for Quality Assurance and Data Mining Tom Purdie, PhD, MCCPM Objective Understand the state-of-the-art in contour assessment for quality assurance including data mining-based techniques
More informationIf it matters to you, it matters to us
If it matters to you, it matters to us Philips clinical innovations in nuclear medicine Innovation with insight We understand that clinical innovations are only as valuable as the day-to-day difference
More informationDetermination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 3, JUNE 2003 405 Determination of Three-Dimensional Voxel Sensitivity for Two- and Three-Headed Coincidence Imaging Edward J. Soares, Kevin W. Germino,
More information664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005
664 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005 Attenuation Correction for the NIH ATLAS Small Animal PET Scanner Rutao Yao, Member, IEEE, Jürgen Seidel, Jeih-San Liow, Member, IEEE,
More informationPET/CT multimodality imaging for radiotherapy planning in lung cancer The medical physicist point of view Isabelle Gardin Rouen CHB and Quant.I.
PET/CT multimodality imaging for radiotherapy planning in lung cancer The medical physicist point of view Isabelle Gardin Rouen CHB and Quant.I.F (EA4108 FR CNRS 3638) Outline Specific acquisition conditions
More informationBy choosing to view this document, you agree to all provisions of the copyright laws protecting it.
Copyright 2009 IEEE. Reprinted from 31 st Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2009. EMBC 2009. Sept. 2009. This material is posted here with permission
More informationPhilips SPECT/CT Systems
Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET
More informationNIH Public Access Author Manuscript J Nucl Med. Author manuscript; available in PMC 2010 February 9.
NIH Public Access Author Manuscript Published in final edited form as: J Nucl Med. 2010 February ; 51(2): 237. doi:10.2967/jnumed.109.068098. An Assessment of the Impact of Incorporating Time-of-Flight
More informationModeling and Incorporation of System Response Functions in 3D Whole Body PET
Modeling and Incorporation of System Response Functions in 3D Whole Body PET Adam M. Alessio, Member IEEE, Paul E. Kinahan, Senior Member IEEE, and Thomas K. Lewellen, Senior Member IEEE University of
More informationValidation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition
Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.
More informationBrilliance CT Big Bore.
1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.
More informationMonte-Carlo-Based Scatter Correction for Quantitative SPECT Reconstruction
Monte-Carlo-Based Scatter Correction for Quantitative SPECT Reconstruction Realization and Evaluation Rolf Bippus 1, Andreas Goedicke 1, Henrik Botterweck 2 1 Philips Research Laboratories, Aachen 2 Fachhochschule
More informationUvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication
UvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication Citation for published version (APA): Kruis, M. F. (2014). Motion compensation for 4D PET/CT General
More informationSlide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.
Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing
More informationMartin A. Lodge PET Center, Johns Hopkins Hospital, Nelson B1125, 1800 Orleans Street, Baltimore, MD 21287, USA
Impact of PET/CT system, reconstruction protocol, data analysis method, and repositioning on PET/CT precision: An experimental evaluation using an oncology and brain phantom Syahir Mansor* Department of
More informationEstimating 3D Respiratory Motion from Orbiting Views
Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng, Jeffrey A. Fessler, James M. Balter The University of Michigan Oct. 2005 Funding provided by NIH Grant P01 CA59827 Motivation Free-breathing
More informationPURE. ViSION Edition PET/CT. Patient Comfort Put First.
PURE ViSION Edition PET/CT Patient Comfort Put First. 2 System features that put patient comfort and safety first. Oncology patients deserve the highest levels of safety and comfort during scans. Our Celesteion
More informationA Generation Methodology for Numerical Phantoms with Statistically Relevant Variability of Geometric and Physical Properties
A Generation Methodology for Numerical Phantoms with Statistically Relevant Variability of Geometric and Physical Properties Steven Dolly 1, Eric Ehler 1, Yang Lou 2, Mark Anastasio 2, Hua Li 2 (1) University
More informationCS 229 Final Project Report Learning to Decode Cognitive States of Rat using Functional Magnetic Resonance Imaging Time Series
CS 229 Final Project Report Learning to Decode Cognitive States of Rat using Functional Magnetic Resonance Imaging Time Series Jingyuan Chen //Department of Electrical Engineering, cjy2010@stanford.edu//
More informationREMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO
REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO Abstract C.S. Levin, Y-C Tai, E.J. Hoffman, M. Dahlbom, T.H. Farquhar UCLA School of Medicine Division
More informationImplementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system
Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation
More informationQ.Clear. Steve Ross, Ph.D.
Steve Ross, Ph.D. Accurate quantitation (SUV - Standardized Uptake Value) is becoming more important as clinicians seek to utilize PET imaging for more than just diagnosing and staging disease, but also
More informationMotion artifacts occurring at the lung/diaphragm interface using 4D CT attenuation correction of 4D PET scans
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 4, fall 2011 Motion artifacts occurring at the lung/diaphragm interface using 4D CT attenuation correction of 4D PET scans Joseph H. Killoran,
More informationDigital Volume Correlation for Materials Characterization
19 th World Conference on Non-Destructive Testing 2016 Digital Volume Correlation for Materials Characterization Enrico QUINTANA, Phillip REU, Edward JIMENEZ, Kyle THOMPSON, Sharlotte KRAMER Sandia National
More informationReview of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA
Review of PET Physics Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Chart of Nuclides Z (protons) N (number of neutrons) Nuclear Data Evaluation Lab.
More information8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit
Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant
More informationThe Near Future in Cardiac CT Image Reconstruction
SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de
More informationSPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.
SPECT QA and QC Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC What is needed? Why? How often? Who says? QA and QC in Nuclear Medicine QA - collective term for all the efforts made to produce
More informationAdvanced Targeting Using Image Deformation. Justin Keister, MS DABR Aurora Health Care Kenosha, WI
Advanced Targeting Using Image Deformation Justin Keister, MS DABR Aurora Health Care Kenosha, WI History of Targeting The advance of IMRT and CT simulation has changed how targets are identified in radiation
More informationMagnetic Resonance Elastography (MRE) of Liver Disease
Magnetic Resonance Elastography (MRE) of Liver Disease Authored by: Jennifer Dolan Fox, PhD VirtualScopics Inc. jennifer_fox@virtualscopics.com 1-585-249-6231 1. Overview of MRE Imaging MRE is a magnetic
More informationA fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study
A fluence convolution method to account for respiratory motion in three-dimensional dose calculations of the liver: A Monte Carlo study Indrin J. Chetty, a) Mihaela Rosu, Neelam Tyagi, Lon H. Marsh, Daniel
More informationPublic Comment Form for QIBA Documents
Public Comment Form for QIBA Documents Notes: 1. identify the commenter to facilitate clarification of the issue and/or communication of the resolution.. : ow. Typo or other minor correction that an editor
More informationHelical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice Guido Hilgers, a Tonnis Nuver, André Minken Department
More informationMotion artifact detection in four-dimensional computed tomography images
Motion artifact detection in four-dimensional computed tomography images G Bouilhol 1,, M Ayadi, R Pinho, S Rit 1, and D Sarrut 1, 1 University of Lyon, CREATIS; CNRS UMR 5; Inserm U144; INSA-Lyon; University
More informationPerformance Evaluation of radionuclide imaging systems
Performance Evaluation of radionuclide imaging systems Nicolas A. Karakatsanis STIR Users meeting IEEE Nuclear Science Symposium and Medical Imaging Conference 2009 Orlando, FL, USA Geant4 Application
More informationFinite Element Simulation of Moving Targets in Radio Therapy
Finite Element Simulation of Moving Targets in Radio Therapy Pan Li, Gregor Remmert, Jürgen Biederer, Rolf Bendl Medical Physics, German Cancer Research Center, 69120 Heidelberg Email: pan.li@dkfz.de Abstract.
More information7/13/2015 EVALUATION OF NONLINEAR RECONSTRUCTION METHODS. Outline. This is a decades-old challenge
EVALUATION OF NONLINEAR RECONSTRUCTION METHODS Kyle J. Myers, Ph.D. Director, Division of Imaging, Diagnostics, and Software Reliability Office of Science and Engineering Laboratories, CDRH, FDA 2 Outline
More informationIntroduction to Positron Emission Tomography
Planar and SPECT Cameras Summary Introduction to Positron Emission Tomography, Ph.D. Nuclear Medicine Basic Science Lectures srbowen@uw.edu System components: Collimator Detector Electronics Collimator
More informationImage Registration. Prof. Dr. Lucas Ferrari de Oliveira UFPR Informatics Department
Image Registration Prof. Dr. Lucas Ferrari de Oliveira UFPR Informatics Department Introduction Visualize objects inside the human body Advances in CS methods to diagnosis, treatment planning and medical
More informationPET Quantification using STIR
PET Quantification using STIR STIR User s Meeting Charalampos Tsoumpas, PhD King s College London Hammersmith Imanet 1 PET Quantification Image elements should correspond to concentration of the injected
More informationAutomated segmentation methods for liver analysis in oncology applications
University of Szeged Department of Image Processing and Computer Graphics Automated segmentation methods for liver analysis in oncology applications Ph. D. Thesis László Ruskó Thesis Advisor Dr. Antal
More informationCLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0
Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed
More informationFour-dimensional (4D) Motion Detection to Correct Respiratory Effects in Treatment Response Assessment Using Molecular Imaging Biomarkers
Four-dimensional (4D) Motion Detection to Correct Respiratory Effects in Treatment Response Assessment Using Molecular Imaging Biomarkers Walter Curran, Emory University Ian Crocker, Emory University Eduard
More information3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction
3-D Monte Carlo-based Scatter Compensation in Quantitative I-131 SPECT Reconstruction Yuni K. Dewaraja, Member, IEEE, Michael Ljungberg, Member, IEEE, and Jeffrey A. Fessler, Member, IEEE Abstract- we
More informationSimultaneous measurement of noise and spatial resolution in PET phantom images
IOP PUBLISHING Phys. Med. Biol. 55 (2010) 1069 1081 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/55/4/011 Simultaneous measurement of noise and spatial resolution in PET phantom images Martin
More informationBasic fmri Design and Analysis. Preprocessing
Basic fmri Design and Analysis Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial filtering
More informationImage-based Monte Carlo calculations for dosimetry
Image-based Monte Carlo calculations for dosimetry Irène Buvat Imagerie et Modélisation en Neurobiologie et Cancérologie UMR 8165 CNRS Universités Paris 7 et Paris 11 Orsay, France buvat@imnc.in2p3.fr
More informationarxiv: v1 [cs.cv] 6 Jun 2017
Volume Calculation of CT lung Lesions based on Halton Low-discrepancy Sequences Liansheng Wang a, Shusheng Li a, and Shuo Li b a Department of Computer Science, Xiamen University, Xiamen, China b Dept.
More informationFactors influencing image quality and quantification: optimization of PET parameters
Factors influencing image quality and quantification: optimization of PET parameters Ronald Boellaard Department of Radiology & Nuclear Medicine, VU University Medical Center, Amsterdam Department of Nuclear
More informationRigid-body transformation of list-mode projection data for respiratory motion correction in cardiac PET
INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 50 (2005) 3313 3322 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/50/14/008 Rigid-body transformation of list-mode projection data for respiratory
More informationBME I5000: Biomedical Imaging
BME I5000: Biomedical Imaging Lecture 1 Introduction Lucas C. Parra, parra@ccny.cuny.edu 1 Content Topics: Physics of medial imaging modalities (blue) Digital Image Processing (black) Schedule: 1. Introduction,
More informationESTIMATING 3D DEFORMABLE MOTION FROM A SERIES OF FAST 2D MRI IMAGES. Jason M. Brown. Chapel Hill 2016
ESTIMATING 3D DEFORMABLE MOTION FROM A SERIES OF FAST 2D MRI IMAGES Jason M. Brown A thesis submitted to the faculty of the University of North Carolina at Chapel Hill in partial fulfillment of the requirements
More informationComparison of 3D PET data bootstrap resampling methods for numerical observer studies
2005 IEEE Nuclear Science Symposium Conference Record M07-218 Comparison of 3D PET data bootstrap resampling methods for numerical observer studies C. Lartizien, I. Buvat Abstract Bootstrap methods have
More informationFMRI Pre-Processing and Model- Based Statistics
FMRI Pre-Processing and Model- Based Statistics Brief intro to FMRI experiments and analysis FMRI pre-stats image processing Simple Single-Subject Statistics Multi-Level FMRI Analysis Advanced FMRI Analysis
More informationDepth-Layer-Based Patient Motion Compensation for the Overlay of 3D Volumes onto X-Ray Sequences
Depth-Layer-Based Patient Motion Compensation for the Overlay of 3D Volumes onto X-Ray Sequences Jian Wang 1,2, Anja Borsdorf 2, Joachim Hornegger 1,3 1 Pattern Recognition Lab, Friedrich-Alexander-Universität
More informationA rib-specific multimodal registration algorithm for fused unfolded rib visualization using PET/CT
A rib-specific multimodal registration algorithm for fused unfolded rib visualization using PET/CT Jens N. Kaftan a, Marcin Kopaczka a, Andreas Wimmer b, Günther Platsch a, and Jérôme Declerck a a Siemens
More informationANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma
ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma ORAL DEFENSE 8 th of September 2017 Charlotte MARTIN Supervisor: Pr. MP REVEL M2 Bio Medical
More informationSupplementary Figure 1. Decoding results broken down for different ROIs
Supplementary Figure 1 Decoding results broken down for different ROIs Decoding results for areas V1, V2, V3, and V1 V3 combined. (a) Decoded and presented orientations are strongly correlated in areas
More informationAn automated method for comparing motion artifacts in cine four-dimensional computed tomography images
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 An automated method for comparing motion artifacts in cine four-dimensional computed tomography images Guoqiang Cui, 1 Brian Jew,
More informationTomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram
TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization
More informationJoint CI-JAI advanced accelerator lecture series Imaging and detectors for medical physics Lecture 1: Medical imaging
Joint CI-JAI advanced accelerator lecture series Imaging and detectors for medical physics Lecture 1: Medical imaging Dr Barbara Camanzi barbara.camanzi@stfc.ac.uk Course layout Day AM 09.30 11.00 PM 15.30
More informationUNIVERSITY OF SOUTHAMPTON
UNIVERSITY OF SOUTHAMPTON PHYS2007W1 SEMESTER 2 EXAMINATION 2014-2015 MEDICAL PHYSICS Duration: 120 MINS (2 hours) This paper contains 10 questions. Answer all questions in Section A and only two questions
More informationPrinciples of PET Imaging. Positron Emission Tomography (PET) Fundamental Principles WHAT IS PET?
Positron Emission Tomography (PET) Fundamental Principles Osama Mawlawi Ph.D Department of Imaging Physics MD Anderson Cancer Center Houston TX. WHAT IS PET? Functional imaging modality as compared to
More informationRobust Clustering of PET data
Robust Clustering of PET data Prasanna K Velamuru Dr. Hongbin Guo (Postdoctoral Researcher), Dr. Rosemary Renaut (Director, Computational Biosciences Program,) Dr. Kewei Chen (Banner Health PET Center,
More information2007 IEEE Nuclear Science Symposium Conference Record M26-296
2007 IEEE Nuclear cience ymposium Conference Record M26-296 A Global and a egmented Plane catter Calibration: Improving the Quantitative Accuracy of Frames with High Random Fraction and/or Low Number of
More informationOptimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago
Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic
More informationEffect of normalization method on image uniformity and binding potential estimates on micropet
Effect of normalization method on image uniformity and binding potential estimates on micropet Marie-Laure Camborde, Arman Rhamim, Danny F. Newport, Stefan Siegel, Ken R. Buckley, Eric Vandervoort, Thomas
More informationAcknowledgments and financial disclosure
AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories
More informationCorso di laurea in Fisica A.A Fisica Medica 5 SPECT, PET
Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 5 SPECT, PET Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes
More informationIs deformable image registration a solved problem?
Is deformable image registration a solved problem? Marcel van Herk On behalf of the imaging group of the RT department of NKI/AVL Amsterdam, the Netherlands DIR 1 Image registration Find translation.deformation
More information438 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 31, NO. 2, FEBRUARY 2012
438 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 31, NO. 2, FEBRUARY 2012 Super-Resolution in Respiratory Synchronized Positron Emission Tomography Daphné Wallach*, Frédéric Lamare, Giorgos Kontaxakis, and
More informationSemi-Quantitative Metrics in Positron Emission Tomography. Michael Adams. Department of Biomedical Engineering Duke University.
Semi-Quantitative Metrics in Positron Emission Tomography by Michael Adams Department of Biomedical Engineering Duke University Date: Approved: Timothy G. Turkington, Supervisor Adam P. Wax Terence Z.
More informationA publicly accessible Monte Carlo database for validation purposes in emission tomography
Short communication A publicly accessible Monte Carlo database for validation purposes in emission tomography I. Castiglioni 1, I. Buvat 2, G. Rizzo 1, M. C. Gilardi 1, J. Feuardent 2, F. Fazio 1 1 IBFM-CNR,
More informationDUAL energy X-ray radiography [1] can be used to separate
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy
More informationCHAPTER 3. Preprocessing and Feature Extraction. Techniques
CHAPTER 3 Preprocessing and Feature Extraction Techniques CHAPTER 3 Preprocessing and Feature Extraction Techniques 3.1 Need for Preprocessing and Feature Extraction schemes for Pattern Recognition and
More informationAuto-Segmentation Using Deformable Image Registration. Disclosure. Objectives 8/4/2011
Auto-Segmentation Using Deformable Image Registration Lei Dong, Ph.D. Dept. of Radiation Physics University of Texas MD Anderson Cancer Center, Houston, Texas AAPM Therapy Educational Course Aug. 4th 2011
More informationFunctional MRI in Clinical Research and Practice Preprocessing
Functional MRI in Clinical Research and Practice Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization
More informationAn introduction to plotting data
An introduction to plotting data Eric D. Black California Institute of Technology February 25, 2014 1 Introduction Plotting data is one of the essential skills every scientist must have. We use it on a
More informationFast Timing and TOF in PET Medical Imaging
Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part
More informationA Projection Access Scheme for Iterative Reconstruction Based on the Golden Section
A Projection Access Scheme for Iterative Reconstruction Based on the Golden Section Thomas Köhler Philips Research Laboratories Roentgenstrasse - Hamburg Germany Abstract A new access scheme for projections
More informationWalk Through of CERR Capabilities. CERR: Introduction. Outline. Getting CERR: Control Panel. Documentation Support Community
Walk Through of CERR Capabilities Aditya P. Apte, Ph.D. Department of Medical Physics Memorial Sloan Kettering Cancer Center New York aptea@mskcc.org AAPM 2015, July 15, 2015 CERR: Computational Environment
More informationBackground. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /
Radiographic Tomosynthesis: Image Quality and Artifacts Reduction Baojun Li, Ph.D Department of Radiology Boston University Medical Center 2012 AAPM Annual Meeting Background Linear Trajectory Tomosynthesis
More information