Philips SPECT/CT Systems
|
|
- Junior Gilbert
- 6 years ago
- Views:
Transcription
1 Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET for Molecular Imaging
2 Philips SPECT/CT Portfolio Precedence 6 & 16 BrightView XCT (New) 2
3 Philips SPECT/CT Portfolio Product BrightView XCT (New) Precedence 6 Precedence 16 General positioning NM-based Radiology-based Radiology-based Premium CT Customer value Low dose high resolution localization Ultra-low dose AC CoPlanar FP NM tailored workflow Small room size Diagnostic CT apps Ca++ scoring Exceptional CT quality Diagnostic CT apps Ca++ scoring *Coronary CTA Highest coverage/throughput Short breath hold exams Exceptional CT quality 3
4 Precedence System Precedence 6/16 slice Product Names: Precedence 6 slice (NM and Oncology) Precedence 16 slice (Cardiology and Oncology) Key Performances: SPECT (Key advances): Detector intrinsic resolution 3.2 mm > 5 mm SPECT reconstruction resolution (Astonish) Half-time acquisition (Cardiac) Capable of full cardiac CT: Speed: 0.5 second rotation Slice thickness: 0.67 mm slice (spiral), 0.6 mm slice (anode) Resolution: 24 lp/cm Dose (CTDI): mgy/100 mas (head), 6.5 mgy/100 mas (body) Registration error 4mm (one pixel) 4
5 BrightView XCT Technology: Flat Panel Based Volume CT technology Coplanar with SPECT Imaging (cardiac -14 cm) Key Performances: SPECT (Key advances): Detector intrinsic resolution 3.2 mm > 5 mm SPECT reconstruction resolution (Astonish) Half-time acquisition Localization, CT-AC, Bone Imaging Max. Rotation Speed: 12 sec for 360 o (14 cm axial FOV) Slice thickness: mm (isotropic voxel) Resolution: >15 lp/cm Dose (CTDI) - Typical: ~ 6 mgy body localization, ~ 1 mgy Attenuation Correction Registration error 4mm (one pixel) 5
6 BrightView XCT System Components Low profile gamma detector Volumetric CT components X-ray tube and collimator Rotating anode X-ray tube 120 kvp X-ray generator, pulsed or continuous 4030CB flat panel detector 10, 30, 60 fps, dynamic gain X-ray collimator and beam shaper CBCT image reconstruction using GPU Volumetric CT system goals X-ray cone-beam overlaps SPECT FOV 360 o Gantry rotation within a breath-hold Low-dose CT acquisition parameters Integrated hybrid software solution SPECT FOV 54 x 40 cm Patient table X-ray flat panel detector 14 cm axial coverage X-ray flat panel detector 40 x 30 cm SPECT/C T Gantry 6
7 System Geometry to Increase Field-of-View Gantry center of rotation Projection Overlap ~4.5 cm 88 cm 45 cm Detector effectively doubles in size to increase imaging FOV to 47 cm Reduces truncation Flat panel detector is offset ~17.4 cm relative to its center Half projections from 180 o opposite views are weighted and combined 3- XCT spins to cover 40 cm SPECT FOV Unweighted Projection Weighted Projection 7
8 Key SPECT/CT Acquisition Features -Precedence and BrightView XCT SPECT (Precedence & BrightView): True Energy Independence 16 Energy Windows Concurrent Imaging Half-time Acquisition (Astonish 3D Reconstruction) Uniform Performance at all Energy Levels for Multi-Peak Imaging (<300 kev) MI Tracer Batteries & Advanced Corrections Simultaneous MI & Conventional Acquisition Protocols Optimized Filtering & Resolution Compensation XCT (BrightView XCT): NM-driven acquisition user interface/protocols No compromise to SPECT functions X-ray CT and SPECT are in coplanar for cardiac imaging Reconstruction with GPU for faster processing speed 8
9 New Workstation -Bringing Simplicity to Nuclear Medicine Workflow to improve clinical impact and cost of ownership PET, SPECT and CT on the same platform Improved access, connectivity - taking Nuclear Medicine to enterprise Clinically complete Applications Suite Full QC functionality on the workstation Streamlined and fully configurable workflow User-adaptive protocols (change parameters, defaults on the fly) Access to applications and tools where needed when needed Imbedding best in class algorithms into one workflow (AutoSPECT, Astonish AC) 9
10 New Workstation: NM Application Suite -Bringing Simplicity to Nuclear Medicine Comprehensive clinical applications and methods. Comprehensive Planar Suite Planar Gated Analysis Whole Body Analysis Pulmonary Analysis Renal Analysis Endocrine Analysis Hepatobiliary Analysis Gastric Analysis Esophageal Analysis Salivary Analysis Comprehensive SPECT Suite Integrated SPECT/CT AC, Reconstruction, Review Comprehensive QA Suite Daily QC NEMA Suite 10
11 Automated Workflow Example: - CT-AC Reconstruction Automate your Workflow with Auto-Proceed or pause at each Workstep. Set Up AC Map Recon Reorient Review 11
12 Image Quality & Acquisition Speed 3D Astonish Reconstruction 3D OSEM with resolution, attenuation and scatter correction Multiple peak isotope Better image quality Toward to better quantitation Half-time Acquisition Astonish provide equal or better with half-time acquisition FB P Astonish Typical SPECT/CT acquisition (typical US Protocols.) Total body scan 15 min SPECT + 2 min CT 7 min SPECT + 2 min CT Cardiac scan 15 min SPECT + 2 min CT 7 min SPECT + 2 min CT 12
13 Image Quality and Acquisition Speed Astonish provides constant image quality with wide range of count statistics 5 Observers: A relative image quality (RIQ) score with respect to a standard image (64x64x64 matrix, clinical counts, FBP reconstruction) was assigned to each image by each observer (3: much better, 2: better, 1: slightly better, 0: equivalent, -1: slightly worse, -2: worse and -3: much worse). Relative Image quality Score Comparison IQ Score ASTONISH FBP Count Statistics ( % with respect to reference image) Astonish projection FBP 64 projection Reference image (conventional acquisition time) 13
14 Clinical Images of BrightView XCT Separate control room is not required for the CT acquisition Plan the SPECT/CT off the nuclear medicine p-scope Flexible breathing protocols to minimize mis-registration Images from Radiological Associates of Sacramento, Courtesy of Dr. Richard Myers, MD 14
15 Patient 1: Cardiac Attenuation Correction 194 lbs (88 kg), Male, 49 yrs Dx: Pre-op clearance, Abnormal EKG Single 60-second CT, tidal breathing Single 12-second CT, end expiration breath hold CT parameters: SPECT parameters: 60 second 12 second Persantine stress, 2.5 hrs p.i. 120 kvp 120 kvp 35 mci Tc-99m MIBI 5 ma 2.5 ma 64 x msec pulse width continuous 64 azimuths 1.2 mgy CTDI VOL 0.79 mgy CTDI VOL 20 sec/azimuth 1.46 zoom 15
16 Comparison of breath hold and tidal breathing 12 second Breath hold CT, 0.8mGy 60 second BrightView XCT launch to sales Tidal breathing CT, 1.2mGy Images courtesy of Radiological Associates of Sacramento
17 Inferior wall attenuation 12 sec breath hold AC No AC AC No AC AC No AC Images courtesy of Radiological Associates of Sacramento 17
18 Inferior wall attenuation 60 sec tidal breathing AC No AC AC No AC AC No AC Images courtesy of Radiological Associates of Sacramento 18
19 Patient 2: Localization 143 lbs (65 kg), Female, 54 yrs Dx: Lung CA Three 12 second CT (breath hold) CT parameters: SPECT parameters: 120 kvp 22.9 mci Tc-99m MDP, 2.5 hrs p.i. 20 ma 128 x 128 continuous pulse width 128 azimuth 6.8 mgy CTDIVOL 20 sec/azimuth 1.0 zoom 19
20 Lung Cancer bilateral facet disease BrightView XCT launch to sales Images courtesy of Radiological Associates of Sacramento
21 Patient 3: Bone Localization 202 lbs (92 kg), Male, 25 yrs Dx: Right knee sarcoma One 24 second CT CT parameters: SPECT parameters: 120 kvp 25.3 mci Tc-99m MDP, 3 hrs p.i. 80 ma 128 x msec pulse width 128 azimuths 14.9 mgy CTDI VOL 20 sec/azimuth 1.4 zoom 21
22 Osteosarcoma, CT resampled to 0.64 mm voxels BrightView XCT launch to sales Images courtesy of Radiological Associates of Sacramento
23 Thank you Thank you 23
3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH
3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk
More informationFits you like no other
Fits you like no other BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase in room
More informationIf it matters to you, it matters to us
If it matters to you, it matters to us Philips clinical innovations in nuclear medicine Innovation with insight We understand that clinical innovations are only as valuable as the day-to-day difference
More informationFits you like no other
Fits you like no other Philips BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase
More informationConflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program
James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives 1. Be familiar with recommendations
More informationJames R Halama, PhD Loyola University Medical Center
James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives Be familiar with the tests recommended
More informationGPU implementation for rapid iterative image reconstruction algorithm
GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial
More informationOptimisation of Toshiba Aquilion ONE Volume Imaging
Optimisation of Toshiba Aquilion ONE Volume Imaging Jane Edwards, RPRSG Royal Free London NHS Foundation Trust Dr Mufudzi Maviki, Plymouth Hospitals NHS Trust Background In 2011/12 Radiology at RFH was
More informationOptimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago
Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic
More informationSymbia T Series System Specifications Answers for life.
www.siemens.com/mi System Specifications Answers for life. 2 Diagnostic Excellence Today s healthcare institutions have increasingly diverse patient needs, greater budget limitations and the need for higher
More informationCT Basics Principles of Spiral CT Dose. Always Thinking Ahead.
1 CT Basics Principles of Spiral CT Dose 2 Who invented CT? 1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections Sir Godfrey Hounsfield worked for the Central
More informationPURE. ViSION Edition PET/CT. Patient Comfort Put First.
PURE ViSION Edition PET/CT Patient Comfort Put First. 2 System features that put patient comfort and safety first. Oncology patients deserve the highest levels of safety and comfort during scans. Our Celesteion
More informationRADIOLOGY AND DIAGNOSTIC IMAGING
Day 2 part 2 RADIOLOGY AND DIAGNOSTIC IMAGING Dr hab. Zbigniew Serafin, MD, PhD serafin@cm.umk.pl 2 3 4 5 CT technique CT technique 6 CT system Kanal K: RSNA/AAPM web module: CT Systems & CT Image Quality
More informationSPECT/CT Basics, Technology Updates, Quality Assurance, and Applications
SPECT/CT Basics, Technology Updates, Quality Assurance, and Applications S. Cheenu Kappadath, PhD Department of Imaging Physics University of Texas M D Anderson Cancer Center, Houston, Texas skappadath@mdanderson.org
More informationMulti-slice CT Image Reconstruction Jiang Hsieh, Ph.D.
Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced
More informationSymbia E and S System Specifications Answers for life.
www.siemens.com/mi Symbia E and S System Specifications Answers for life. Symbia E: Work with confidence. What does it mean to work with confidence? It is clarity not only image clarity but the clarity
More informationSpiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA
Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization
More informationIntroduction to Positron Emission Tomography
Planar and SPECT Cameras Summary Introduction to Positron Emission Tomography, Ph.D. Nuclear Medicine Basic Science Lectures srbowen@uw.edu System components: Collimator Detector Electronics Collimator
More informationDesign and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon
Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization
More informationComputer-Tomography II: Image reconstruction and applications
Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de
More informationCT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION
CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology
More informationCorridor4DM Feature List
Software Services Features Corridor4DM Feature List 4DM v2013 FDG PET Standard Uptake Value (SUV) Calculation Auto-Selection of Report Template Motion Correction of SPECT datasets New Normals Databases
More informationThe Near Future in Cardiac CT Image Reconstruction
SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de
More information4DM Packages. 4DM Packages & License Types. Information to help you order the appropriate licenses for your site.
4DM Packages 4DM Packages & License Types. Information to help you order the appropriate licenses for your site. Nuclear Cardiac Quantification, Review, and Reporting Select Your 4DM Package and corresponding
More informationTechnical aspects of SPECT and SPECT-CT. John Buscombe
Technical aspects of SPECT and SPECT-CT John Buscombe What does the clinician need to know? For SPECT What factors affect SPECT How those factors should be sought Looking for artefacts For SPECT-CT Issues
More informationDX-D 100 with wireless detector
M O B I L E D R S O L U T I O N DX-D 100 with wireless detector Patients who most need imaging exams may lack the mobility necessary to move to the X-ray room or to position themselves properly for optimum
More information8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial
8//0 AAPM0 Scientific Symposium: Emerging and New Generation PET: Instrumentation, Technology, Characteristics and Clinical Practice Aug Wednesday 0:4am :pm Solid State Digital Photon Counting PET/CT Instrumentation
More informationMediso AnyScan. Single-Head and Dual-Head SPECT
Mediso AnyScan S Single-Head and Dual-Head SPECT ANYSCAN S (SINGLE-HEAD & DUAL-HEAD SPECT) Low Cost of Ownership Associated Imaging Services is the exclusive U.S. source in the Midwest for Mediso Medical
More informationSPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.
SPECT QA and QC Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC What is needed? Why? How often? Who says? QA and QC in Nuclear Medicine QA - collective term for all the efforts made to produce
More informationMediso AnyScan S Single-Head and Dual-Head SPECT
Mediso AnyScan S Single-Head and Dual-Head SPECT ANYSCAN S (SINGLE-HEAD & DUAL-HEAD SPECT) Low Cost of Ownership Small Footprint Absolute Imaging Solutions tm is the exclusive U.S. source for Mediso Medical
More informationAnnexure XII SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER
Annexure XII SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER A) Scanner Design X-Ray generator and tube: 1. Scanner: Whole body spiral CT scanner (16 slices) of latest technology.
More informationFinancial disclosure. Onboard imaging modality for IGRT
Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty
More informationCh. 4 Physical Principles of CT
Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between
More informationDX-D 100 with WITH ITS EXCELLENT IMAGE QUALITY AND
M O B I L E D R S O L U T I O N DX-D 100 with wireless detector Patients who most need imaging exams may lack the mobility necessary to move to the X-ray room or to position themselves properly for optimum
More informationAIDR 3D Iterative Reconstruction:
Iterative Reconstruction: Integrated, Automated and Adaptive Dose Reduction Erin Angel, PhD Manager, Clinical Sciences, CT Canon Medical Systems USA Iterative Reconstruction 1 Since the introduction of
More informationDeviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies
g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,
More informationDX-D 100 WITH WIRELESS DETECTOR
DX-D 100 WITH WIRELESS DETECTOR MOBILE DR SOLUTION WITH ITS EXCELLENT IMAGE QUALITY AND FLEXIBLE HANDLING, THE MOBILE DX-D 100 WITH WIRELESS DETECTOR OFFERS FAST IMAGING THAT CAN BE VALIDATED IMMEDIATELY.
More information8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit
Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant
More informationImage Acquisition Systems
Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS
More informationShadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS
Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic
More informationCorso di laurea in Fisica A.A Fisica Medica 4 TC
Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure
More informationMEDICAL IMAGING 2nd Part Computed Tomography
MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional
More informationQIBA PET Amyloid BC March 11, Agenda
QIBA PET Amyloid BC March 11, 2016 - Agenda 1. QIBA Round 6 Funding a. Deadlines b. What projects can be funded, what cannot c. Discussion of projects Mechanical phantom and DRO Paul & John? Any Profile
More informationComputed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA
Computed Tomography Principles, Design, Artifacts, and Recent Advances THIRD EDITION Jiang Hsieh SPIE PRESS Bellingham, Washington USA Table of Contents Preface Nomenclature and Abbreviations xi xv 1 Introduction
More informationInveon User Notes. Inveon Scanners and Inveon Acquisition Workplace 2.0
Inveon User Notes Inveon Scanners and Inveon Acquisition Workplace 2.0 Legal Notice 2014 by Siemens Medical Solutions USA, Inc. All rights reserved. Siemens reserves the right to modify the design and
More informationCLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0
Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed
More informationHitachi Medical Systems America, Inc. Hitachi Medical Corporation
Advanced 128 Advanced 128 128 Slice Capability Enlarged 75cm Aperture Thin Gantry Profile 88cm Intelli IP Advanced, Iterative Reconstruction IntelliCenter Auto-Lateral Shifting Table, up to 16cm Range
More informationPRODUCT DATA. Advanced 128
PRODUCT DATA Advanced 128 Supria 16 Slice CT Puts You On The Path of High Quality, Cost-Effective CT Scanning Solid Capabilities Are Built Into the Supria Addressing the challenges of controlling healthcare
More informationSolid Capabilities Are Built Into the Supria Plus. Putting You On The Path of High Quality, Cost-Effective CT Scanning
Specification Data Putting You On The Path of High Quality, Cost-Effective CT Scanning Solid Capabilities Are Built Into the Supria Plus Addressing the challenges of controlling healthcare organization
More informationIntroduction to Emission Tomography
Introduction to Emission Tomography Gamma Camera Planar Imaging Robert Miyaoka, PhD University of Washington Department of Radiology rmiyaoka@u.washington.edu Gamma Camera: - collimator - detector (crystal
More informationDX-D WITH DX-D 40/45 DETECTOR FAMILY MOBILE DR SOLUTION
DX-D 100 + WITH DX-D 40/45 DETECTOR FAMILY MOBILE DR SOLUTION WITH ITS EXCELLENT IMAGE QUALITY AND FLEXIBLE HANDLING, THE MOBILE DX-D 100 + WITH WIRELESS DETECTOR OFFERS FAST IMAGING THAT CAN BE VALIDATED
More informationLower Dose with Confidence
Lower Dose with Confidence The Next Generation SCENARIA CT is Here. 64 or 128 Slice Capability The SCENARIA SE delivers lower dose technologies*, enhanced patient access, fast workflow speed and added
More informationEmission Computed Tomography Notes
Noll (24) ECT Notes: Page 1 Emission Computed Tomography Notes Introduction Emission computed tomography (ECT) is the CT applied to nuclear medicine. There are two varieties of ECT: 1. SPECT single-photon
More informationEnhancement Image Quality of CT Using Single Slice Spiral Technique
Enhancement Image Quality of CT Using Single Slice Spiral Technique Doaa. N. Al Sheack 1 and Dr.Mohammed H. Ali Al Hayani 2 1 2 Electronic and Communications Engineering Department College of Engineering,
More informationAAPM Standard of Practice: CT Protocol Review Physicist
AAPM Standard of Practice: CT Protocol Review Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center September 11, 2014 2014 Texas Radiation Regulatory Conference Goals Understand purpose
More informationAcknowledgments and financial disclosure
AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories
More informationSCENARIA 64-Slice CT Putting Advanced CT Within Reach.
SCENARIA 64-Slice CT Putting Advanced CT Within Reach. Advancing the technology of Computed Tomography for over 30 years, Hitachi is a recognized innovator of lower dose*, high diagnostic value CT solutions.
More informationIntroduction to Biomedical Imaging
Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray
More informationUSING cone-beam geometry with pinhole collimation,
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE 2009 687 A Backprojection-Based Parameter Estimation Technique for Skew-Slit Collimation Jacob A. Piatt, Student Member, IEEE, and Gengsheng L.
More informationRadiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada
Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Tecniques of imaging in Overview Introduction Tecniques of imaging
More informationMotion Compensation from Short-Scan Data in Cardiac CT
Motion Compensation from Short-Scan Data in Cardiac CT Juliane Hahn 1,2, Thomas Allmendinger 1, Herbert Bruder 1, and Marc Kachelrieß 2 1 Siemens Healthcare GmbH, Forchheim, Germany 2 German Cancer Research
More informationA closer look at CT scanning
Vet Times The website for the veterinary profession https://www.vettimes.co.uk A closer look at CT scanning Author : Charissa Lee, Natalie Webster Categories : General, Vets Date : April 3, 2017 A basic
More informationAutomated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator
Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4
More informationValidation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition
Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.
More informationEstimating 3D Respiratory Motion from Orbiting Views
Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng, Jeffrey A. Fessler, James M. Balter The University of Michigan Oct. 2005 Funding provided by NIH Grant P01 CA59827 Motivation Free-breathing
More informationSCENARIA Hitachi Innovation at Work for You.
PRODUCT DATA SCENARIA Hitachi Innovation at Work for You. Continuing its commitment to the SCENARIA scalable CT platform, Hitachi introduces cutting-edge 64/128-slice technology with the SCENARIA SE. The
More informationTomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram
TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization
More informationImage Post-Processing, Workflow, & Interpretation
Image Post-Processing, Workflow, & Interpretation Richard L. Hallett, MD Chief, Cardiovascular Imaging Northwest Radiology Network Indianapolis, IN Adjunct Assistant Professor Stanford University Stanford,
More informationCT Protocol Review: Practical Tips for the Imaging Physicist Physicist
CT Protocol Review: Practical Tips for the Imaging Physicist Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center August 8, 2013 AAPM Annual Meeting Goals Understand purpose and importance
More informationImage Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD
Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image
More informationVenus Explorer Processing Technical specifications
24, rue du champ de l Alouette F 75013 Paris, France Venus Explorer Processing Technical specifications Venus Explorer Processing Server software Web base application sever; tomcat Apache technology. Include
More informationTechnological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI. Floris Jansen, GE Healthcare July, 2015
Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI Floris Jansen, GE Healthcare July, 2015 PET/MR 101 : challenges Thermal Workflow & Apps RF interactions?!!
More informationAcknowledgments. Nesterov s Method for Accelerated Penalized-Likelihood Statistical Reconstruction for C-arm Cone-Beam CT.
June 5, Nesterov s Method for Accelerated Penalized-Likelihood Statistical Reconstruction for C-arm Cone-Beam CT Adam S. Wang, J. Webster Stayman, Yoshito Otake, Gerhard Kleinszig, Sebastian Vogt, Jeffrey
More informationZ-MOTION. Universal Digital Radiographic System Z-MOTION. Control-X Medical CONTROL-X MEDICAL
Control-X Medical Z-MOTION Compact design, low ceiling height requirement Motorized and manual movement capability Wide motion / SID range Best-in-class image quality Flexible connectivity to PACS systems
More informationO-ARM IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE
IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE SYSTEM FEATURES CATEGORY PHYSICAL DIMENSIONS IMAGING MODALITY PERFORMANCE SPECIFICATION Length Width Height Weight Gantry Opening Bore Diameter Single Plane
More informationFLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300
FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300 The DX-D 300 DR system unites excellent image quality with complete convenience. It offers top-of-the-line technology, a single detector and a fully-motorized
More informationSimulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images
Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),
More informationMEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah
MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography
More informationImplementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system
Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation
More informationBrilliance CT Big Bore.
1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.
More informationDX-D 300. Flexible Direct Radiography System
DX-D 300 Flexible Direct Radiography System MUSICA 2 processing provides superior contrast detail and exam-independent, consistent image quality Cesium Iodide DR detector technology offers potential for
More informationReview of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA
Review of PET Physics Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Chart of Nuclides Z (protons) N (number of neutrons) Nuclear Data Evaluation Lab.
More informationDX-D 300. Flexible Direct Radiography System
DX-D 300 Flexible Direct Radiography System MUSICA² processing provides outstanding contrast detail and consistently excellent image quality Universal, flexible and affordable modality combines a single
More informationRoutine excellence. Philips Ingenuity Core specifications
Routine excellence Philips Ingenuity Core specifications Table of contents 1 Introduction 3 2 User interface 4 2.1 ipatient key benefits 4 2.2 ExamCards 4 2.3 ScanRuler 4 3 DoseWise 5 3.1 DoseRight Index
More informationV/Q Imaging Protocols
Elizabeth ailey ppsc(mrs) M Department of Nuclear Medicine Royal North Shore Hospital, Sydney, ustralia email: ebailey@nsccahs.health.nsw.gov.au The V/Q scan has traditionally used planar imaging techniques.
More informationQuantitative imaging for clinical dosimetry
Quantitative imaging for clinical dosimetry Irène Buvat Laboratoire d Imagerie Fonctionnelle U678 INSERM - UPMC CHU Pitié-Salpêtrière, Paris buvat@imed.jussieu.fr http://www.guillemet.org/irene Methodology
More informationRetrospective Spiral Respiratory Correlated Imaging with Varian RPM
Retrospective Spiral Respiratory Correlated Imaging with Varian RPM This is a Quick Step Guide for Retrospective Spiral Respiratory Correlated Imaging (4D CT) using the Varian RPM device v1.7 with the
More informationS. Guru Prasad, Ph.D., DABR
PURPOSE S. Guru Prasad, Ph.D., DABR Director of Medical Physics IAEA Consultant NorthShore University Health System and University of Chicago, Pritzker School of Medicine Current TPS utilize more information
More informationISO ISO ISO OHSAS ISO
ISO 9001 ISO 13485 ISO 14001 OHSAS 18001 ISO 27001 Pro-NM Performance 08-101 - standard version 08-103 - version with the PET Lid Phantom for NM and PET systems performance evaluation (collimator, artifacts,
More informationAssessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT
Assessment of 3D performance metrics D and 3D Metrics of Performance Towards Quality Index: Volumetric imaging systems X-ray based Volumetric imaging systems: CBCT/CT Tomosynthesis Samuel Richard and Ehsan
More informationThe Emory Reconstruction Toolbox Version 1.0
The Emory Reconstruction Toolbox Version 1.0 Operating Instructions Revision 02 (April, 2008) Operating Instructions The Emory Reconstruction Toolbox Application Copyrights, Trademarks, Restrictions
More informationEducational Objectives SPECT/CT: Basics, Quality Assurance, and Clinical Applications
Educational Objectives SPECT/CT: Basics, Quality Assurance, and Clinical Applications 1. To understand the physics principles underlying SPECT/CT image acquisition and reconstruction 2. To understand the
More informationAs fl exible as your care requires
As fl exible as your care requires Philips Ingenuity Flex 32 CT Built on Ingenuity The Philips Ingenuity Flex 32 helps you provide excellent care with outstanding flexibility, now and in the future. Built
More informationEquipment Specification
MULTISLICE CT SCANNER ( 16 Slices ) Merk : Hitachi Japan Model : SUPRIA 5 MHU Price : Rp 6.512.360.215,27 No Equipment Specification 1 2 3 Scanner Gantry Object for scanning : Whole body including head
More informationDX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300 DR SYSTEM
DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM The DX-D 300 DR system unites excellent image quality with the ultimate convenience. It offers top-of-the-line technology: a fullymotorized positioner enabling
More informationMedical Image Processing: Image Reconstruction and 3D Renderings
Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,
More informationAn Efficient Technique For Multi-Phase Model Based Iterative Reconstruction
1 An Efficient Technique For Multi-Phase Model Based Iterative Reconstruction Shiyu Xu, Debashish Pal and Jean-Baptiste Thibault Abstract Multi-phase scan is a fundamental CT acquisition technology used
More informationBME I5000: Biomedical Imaging
1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.
More informationAgenda : Lung Density Breakout Session
Agenda : Lung Density Breakout Session 1. : Mathew Fuld and Bernice Hoppel 2. Automatic Exposure Control (AEC) Evaluation : Sean Fain Round 4 Project 3. Dose reduction effects on emphysema metrics : Philip
More informationVolumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT
Volumetric Modulated Arc Therapy - Clinical Implementation Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA Acknowledgement David M. Shepard, Ph.D. Muhammad K. N. Afghan, Ph.D. Fan Chen, Ph.D.
More information