SIEMENS HEALTHCARE INTEGRATED ELECTRICAL CABINET

Size: px
Start display at page:

Download "SIEMENS HEALTHCARE INTEGRATED ELECTRICAL CABINET"

Transcription

1 Product Data Sheet Technical Publications SIEMENS HEALTHCARE INTEGRATED ELECTRICAL CABINET for ARTIS Q / Q.ZEN / ZEE BIPLANE IECAX480V225A Main Disconnect Panel

2 Bevco Engineering Company, Inc. CONFIDENTIAL AND PROPRIETARY This Bevco Engineering Company document contains confidential and proprietary information of Bevco Engineering Company. The intellectual properties contained herein may not be used or copied to the detriment of Bevco Engineering Company. The information contained in this document may not be copied without the express written approval of Bevco Engineering Company. CAUTION! Troubleshooting and servicing should be performed by a qualified electrician. WARNING! De-energize and lockout the main disconnect before servicing this panel. Table of Contents 1.0 DESCRIPTION OF OPERATION... 3 SYSTEM CONFIGURATION SPECIFICATIONS... 6 Electrical Specifications... 6 Seismic Specifications:... 6 Electrical Wiring Diagram... 6 Environmental Specifications: PANEL DIMENSIONS DOCUMENT REVISION HISTORY... 8 COMPONENTS SUPPLIED WITH EACH PANEL: The IEC Main Disconnect Panel This Installation, Operations & Service Manual (3) sets of Emergency Power Off pushbuttons with 3NC contacts on each EPO Drawings and Electrical Schematics IECAX480V225A Product Data Sheet_R51 Siemens IEC Manual For AX Systems Page 2 of 8

3 1.0 DESCRIPTION OF OPERATION INTEGRATED ELECTRICAL CABINET This panel incorporates several features desirable for system installations to minimize down time, protect the generator electronics, and to reduce operational delays after a power outage. This product replaces the power panel, the lighting control relay box, and the Emergency Power Off controls. The panel has a main circuit breaker, Q1 with individual branch breakers for each generator and various cooling options. When the main circuit breaker is turned off, all power circuits within the panel will be de-energized except the power circuits for the Room Light control and the X-Ray ON warning lights. The conductors that are fed from other sources are wired using yellow conductors. The generators power is protected by electronic circuit breakers, Q2 and Q3, and is controlled by contactors, K2 and K6. The contactors will open with any loss of power or by pressing any Emergency Power Off (EPO) pushbutton. The K2 and K6 contactor control circuits are factory configured to automatically re-energize the generators upon restoration of facilities power. The control circuit may be re-configured to require the operator to manually restart the generators once the incoming power has been restored. This protects the sensitive electronic circuits of the generators from sags and surges that immediately follow power loss from blackouts, storms, utility reclosure operations, and out of phase automatic transfer switch operations. The SC1 cabinet power is protected by an electronic circuit breaker, Q4, and is controlled by contactor, K7. The contactor will open with any loss of power or by pressing any Emergency Power Off (EPO) pushbutton. The K7 contactor control circuit are factory configured to automatically re-energize the SC1 cabinet upon restoration of facilities power. The control circuit may be re-configured to require the operator to manually restart the SC1 cabinet once the incoming power has been restored. This protects the sensitive electronic circuits of the generators from sags and surges that immediately follow power loss from blackouts, storms, utility reclosure operations, and out of phase automatic transfer switch operations. The control circuits are low voltage 24 VDC and are fully powered from within the panel. The restart functionality and EPO circuitry is controlled with a safety relay, K10. See page 11 for detail. The white SAFETIES OK indicator light on the front of the panel is illuminated when none of the EPOs are pressed and the safety relay K10 is reset. When the white light is active, pressing the green START pushbutton will cause the Angio system to be energized. The green START button will illuminate, and the white SAFETIES OK light will go off. Pressing the STOP button will de-energize the imaging system. Any EPO pressed while the generator systems are energized will result in the immediate de-energizing of the Angio system. If an EPO is pressed at any time, the EPO must be reset which will cause the SAFETIES OK light to activate. Then the START button will activate the imaging system. IMPORTANT: If building power is removed from the panel while the system is energized, the system will re-energize when building power is restored without any human interaction. Additional provisions are made to integrate the X-Ray ON warning lights and room lighting with the X- Ray Equipment. The facility lighting panel provides 120- or 277-volt power that is controlled by contacts relays of K4 and K5 mounted in the IEC. The signal controlling the relays comes directly from the Siemens Generator/Power Cabinet. The IEC will accept signals from the generator at either 24 Volts AC or DC. IECAX480V225A Product Data Sheet_R51 Siemens IEC Manual For AX Systems Page 3 of 8

4 IECAX480V225A Product Data Sheet_R51 Siemens IEC Manual For AX Systems Page 4 of 8

5 UTILITY SUPPLY 480V 3-PHASE 180A MAX Q1 225A WIRESIZE: 4AWG- 350MCM SYSTEM CONFIGURATION KEY TO COMPONENTS: Q1 MAIN CIRCUIT BREAKER Q2 GENERATOR 2 CIRCUIT BREAKER Q3 GENERATOR 1 CIRCUIT BREAKER F2 POWER SUPPLY FUSES F4 DC CIRCUIT BREAKER G01 DC POWER SUPPLY K1 SAFETY CONTACTOR K2 GENERATOR 2 CONTACTOR K4 ROOM LIGHT RELAY K4A ROOM LIGHT RELAY 24VDC K5 X-RAY ON WARNING LIGHT RELAY K5A X-RAY ON WARNING LIGHT RELAY 24VDC K6 - GENERATOR 1 CONTACTOR K7 SYSTEM CABINET RELAY K10 SAFETY RELAY Q2 100A F2 4A Q4 30A Q3 100A G01 DC POWER SUPPLY K A WIRE SIZE: 10-2/0AWG F4 5A SC1 DOOR SWITCH ROOM LIGHT X-RAY ON 30A MAX K A MAX WIRE SIZE: 10-2/0AWG DC CONTROL CIRCUITS K1 SAFETY CONTACTOR K4 ROOM LIGHT RELAY K4A ROOM LIGHT RELAY 24VDC K5 X-RAY ON WARNING LIGHT RELAY K5A X-RAY ON WARNING LIGHT RELAY 24VDC K10 SAFETY RELAY EMERGENCY POWER OFF CIRCUITS X-RAY ON WARNING LIGHT 20A MAX LIGHTING POWER PANEL 120/277V K A WIRE SIZE: 10-2/0AWG GENERATOR 2 80A MAX EPO EPO EPO ROOM LIGHTS 20A MAX GENERATOR 2 80A MAX IECAX480V225A Product Data Sheet_R51 Siemens IEC Manual For AX Systems Page 5 of 8

6 2.0 SPECIFICATIONS Electrical Specifications UL design and ratings of this panel require the application of wire at the 75 C ampacity ratings of NEC table Wire rated at higher temperature ratings, such as 90 C, may be used but ONLY applied at the 75 C ampacity ratings of NEC Table Catalog Number Incoming Voltage & Frequency Incoming Power Configuration Main Circuit Breaker Rating Regulatory Markings IECAX480V225A 480VAC, 50/60Hz 3Ø + Ground + Neutral 225 Amps culus Short Circuit Current Rating is 35,000 Amperes RMS 480V Seismic Specifications: This Integrated Electrical Cabinet has been certified by an independent California structural engineer in conformance with the shake testing requirements of ICC- AC 156. The California OSHPD number is OSP The seismic performance characteristics are as follows: SDS(g) 2.56; z/h 1.0 ; Ip 1.5 Electrical Wiring Diagram The Electrical Wiring Diagram is adhered to the inside of the enclosure door for convenient reference to the circuitry. A complete set of the panel drawings are also attached to the back of this manual and shipped with the disconnect panel. Environmental Specifications: FOR INDOOR USE ONLY Temperature F (15-32 C) Humidity 30-75%, NON-CONDENSING PARTS, TECHNICAL SUPPORT AND SERVICE FOR INQUIRIES DURING INSTALLATION CONTACT BEVCO ENGINEERING, SUSSEX WI (262) FOR ISSUES FOLLOWING INSTALLATION, CONTACT SIEMENS UPTIME CENTER, (800) IECAX480V225A Product Data Sheet_R51 Siemens IEC Manual For AX Systems Page 6 of 8

7 4.0 PANEL DIMENSIONS Refer to the Table and Illustrations below for the physical dimensions of the panel. The panel may be recessed into the wall for semi-flush installations. See Illustrations below for typical front and side views with the mounting hole pattern. Refer to the drawing set provided with the panel for total weight and center of gravity dimensions. PANEL EXTERNAL PHYSICAL DIMENSIONS Parameter Dimensions Height 36.0 inches 914 mm Width 30.0 inches 762 mm Total Depth 9.1 inches 230 mm Enclosure Depth 8.0 inches 203 mm Mounting Hole Diameter 7/16 inches 11.1 mm Mounting Hole Width 28.5 inches 724 mm Mounting Hole Height 34.5 inches 876 mm Weight 157 pounds 71.2 kg IECAX480V225A Product Data Sheet_R51 Siemens IEC Manual For AX Systems Page 7 of 8

8 6.0 DOCUMENT REVISION HISTORY REV DATE AUTHOR DESCRIPTION OF CHANGE 1 04/19/2014 RJB Initial Release 2 02/28/2015 RJB Changed Artis name on first page 3 03/09/2015 XYUE Update manual based on current schematic 4 03/19/2015 JRB Added additional Provision on page /01/2016 DRO IEC Redesign 51 07/10/18 JT Updated system configuration data IECAX480V225A Product Data Sheet_R51 Siemens IEC Manual For AX Systems Page 8 of 8

Installation, Operation, and Service Manual SIEMENS HEALTHCARE INTEGRATED ELECTRICAL CABINET

Installation, Operation, and Service Manual SIEMENS HEALTHCARE INTEGRATED ELECTRICAL CABINET Technical Publications SIEMENS HEALTHCARE INTEGRATED ELECTRICAL CABINET for ARTIS Q / Q.ZEN / ONE ARTIS ZEEGO / ZEEGO.ZEN ARTIS ZEE MULTIPURPOSE SINGLE PLANE IECAX480V125A Main Disconnect Panel Bevco Engineering

More information

ASCO 920 Remote Control (RC) Switches

ASCO 920 Remote Control (RC) Switches Maximum reliability and excellent value ASCO 920 Remote Control (RC) Switches ASCO 920 Remote Control Switch in Type Enclosure The ASCO 920 is designed as a feeder disconnect switch for load capacities

More information

920 Remote Control (RC) Switches

920 Remote Control (RC) Switches 920 Remote Control (RC) Switches ASCO is the global leader in power switching and controls ASCO 920 PRODUCT SPECIFICATIONS SWITCH ONLY (PANELBOARD MOUNTING) 1,2,3 () DIMENSIONS, IN.(MM) WIDTH HEIGHT DEPTH

More information

Specifications, Installation, and Operating Instructions

Specifications, Installation, and Operating Instructions Specifications, Installation, and Operating Instructions Model: Power Vantage AC Panel Protector CAUTION: The installation of a surge protection device () must be done by qualified electrical personnel.

More information

DANGER DANGER: is used in this manual to warn of high voltages capable of causing shock, burns, or death.

DANGER DANGER: is used in this manual to warn of high voltages capable of causing shock, burns, or death. Installation/Operation Manual TABLE OF CONTENTS Page UNPACKING AND INSTALLATION Unpacking and Preliminary Inspection...2, 3 Model Number...9 LOCATION CONSIDERATIONS Environment, Equipment Performance,

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS LIGHTING CONTROL PANELS 16 AND 24 RELAYS INSTALLATION INSTRUCTIONS INSTALLATION OVERVIEW The installation instructions contained in this document are provided as a guide for proper and reliable installation.

More information

User s Manual. ACS550-PD Stock 3R Irrigation Packaged Drive Supplement to ACS550-U1 User s Manual

User s Manual. ACS550-PD Stock 3R Irrigation Packaged Drive Supplement to ACS550-U1 User s Manual User s Manual ACS550-PD Stock 3R Irrigation Packaged Drive Supplement to ACS550-U1 User s Manual 2 ACS550-PD 3R Irrigation Packaged Drive ACS550 Drive Manuals GENERAL MANUALS ACS550-U1 User s Manual (1

More information

Surge Protective Devices. STV100K Series. Instruction Manual

Surge Protective Devices. STV100K Series. Instruction Manual Surge Protective Devices STV100K Series Instruction Manual Contents 1.0 Installation... 4 1.1 Environment... 4 1.2 Mounting... 4 1.3 Proper Connection of SPDs... 4 1.4 Mechanical Dimensions... 5 1.5 Voltage

More information

Surge Protective Devices. STV100K Series. Instruction Manual

Surge Protective Devices. STV100K Series. Instruction Manual Surge Protective Devices STV100K Series Instruction Manual Contents 1.0 Installation... 4 1.1 Environment... 4 1.2 Mounting... 4 1.3 Proper Connection of SPDs... 4 1.4 Mechanical Dimensions... 5 1.5 Voltage

More information

(1986; Rev. 1989) Molded Case Circuit Breakers and Molded Case Switches. (1985; Rev. 1988) Enclosures for Electrical Equipment (1000 Volts Maximum)

(1986; Rev. 1989) Molded Case Circuit Breakers and Molded Case Switches. (1985; Rev. 1988) Enclosures for Electrical Equipment (1000 Volts Maximum) Section 16306 50/60-HERTZ (HZ) SOLID STATE FREQUENCY CONVERTER Index PART 1 GENERAL PART 2 PRODUCTS 1.1 References 2.1 Frequency Converters 1.2 Factory Tests 2.2 Conduit and Fittings 1.3 Submittals 2.3

More information

Quick Start Guide TS A

Quick Start Guide TS A Quick Start Guide TS 930 125-630A DANGER HAZARD OF ELECTRICAL SHOCK, EXPLOSION, OR ARC FLASH Read and understand this quick start guide before installing and operating the transfer switch The installer

More information

Maximum Reliability & Excellent Value. Features. UL Listed Withstand & Close-On Ratings. Series 300 Power Transfer Switches

Maximum Reliability & Excellent Value. Features. UL Listed Withstand & Close-On Ratings. Series 300 Power Transfer Switches Series 300 Power Transfer Switches Maximum Reliability & Excellent Value With a Series 300 Transfer Switch, you get a product backed by ASCO Power Technologies, the industry leader responsible for virtually

More information

Power Supply (48Vdc, 2.5A)

Power Supply (48Vdc, 2.5A) Telecommunications Group Section 855-180-201 Equipment Issue 1 Fourth Printing, September 2006 8551-80 Power Supply (48Vdc, 2.5A) Compliant with UL Standard 60950, Second Edition* CONTENTS PAGE Part 1.

More information

PLT. Breaker-based transfer switch open, closed or soft transition amps. Specification sheet

PLT. Breaker-based transfer switch open, closed or soft transition amps. Specification sheet Specification sheet PLT Breaker-based transfer switch open, closed or soft transition 800 3000 amps Description PLT breaker-based transfer switches are designed for operation and switching of electrical

More information

Always "On" UPS Systems Inc. USER MANUAL

Always On UPS Systems Inc. USER MANUAL Always "On" UPS Systems Inc. SMART HOME PROTECTOR USER MANUAL Bldg 1-150 Campion Road, Kelowna, BC, Canada, V1X 7S8 Phone: (250) 491-9777 Ext. 451, Fax: (250) 491-9775, Email: sales@alwaysonups.com Web

More information

Line Impedance Stabilization Network (LISN)

Line Impedance Stabilization Network (LISN) Model 3816/2 Line Impedance Stabilization Network (LISN) User Manual ETS-Lindgren Inc. reserves the right to make changes to any product described herein in order to improve function, design, or for any

More information

PowerFlex 400 AC Drive Guide Specification

PowerFlex 400 AC Drive Guide Specification PowerFlex 400 AC Drive Guide Specification Adjustable Frequency Drives with Bypass 3.0 50HP @ 208V AC 3.0 350HP @ 480V AC PART 1 GENERAL 1.01 Quality Assurance A. The manufacturer shall have minimum 5

More information

BT300 HVAC Drives Electronic Bypass (E-Bypass) Options

BT300 HVAC Drives Electronic Bypass (E-Bypass) Options BT300 HVAC Drives Electronic Bypass (E-Bypass) Options Description The BT300 Electronic Bypass (E-Bypass) Options are companion packages for the family of BT300 Variable Frequency Drives (VFDs). For information

More information

HUBBELL SPIKESHIELD Wired-In Branch Panels Surge Protective Device (SPD)

HUBBELL SPIKESHIELD Wired-In Branch Panels Surge Protective Device (SPD) HUBBELL SPIKESHIELD Wired-In Branch Panels Surge Protective Device (SPD) Instruction Bulletin 8222-0513B Retain for future use. SPIKESHIELD Wired-In Branch Panels 8222-0513B 05/2012 Table of Contents Precautions...

More information

APPLICATION CONTROL GUIDELINES. IntelliROL Power Supply PN Revision Date: March 15, 2017

APPLICATION CONTROL GUIDELINES. IntelliROL Power Supply PN Revision Date: March 15, 2017 APPLICATION CONTROL GUIDELINES IntelliROL Power Supply PN 1176718 Revision Date: March 15, 2017 Table of Contents List of Tables...3 TGW Safety Recommendation...4 Warnings and Safety Instructions...5 Introduction...6

More information

Models 120LS, 200LS, 300LS, 400LS, 560LS Installation Instructions

Models 120LS, 200LS, 300LS, 400LS, 560LS Installation Instructions 299-600-96 Rev. F MCG Surge Protection Models 120LS, 200LS, 300LS, 400LS, 560LS Installation Instructions Important Warranty Information: MCG surge protectors are designed to work at specific voltages

More information

Liebert AccuVar (ACV Series) Surge Protective Device (SPD) (With Noise Filtering) GUIDE SPECIFICATIONS for a Parallel Surge Suppression System

Liebert AccuVar (ACV Series) Surge Protective Device (SPD) (With Noise Filtering) GUIDE SPECIFICATIONS for a Parallel Surge Suppression System Liebert AccuVar (ACV Series) Surge Protective Device (SPD) (With Noise Filtering) GUIDE SPECIFICATIONS for a Parallel Surge Suppression System Part 1 General 1.01 Summary A. These specifications describe

More information

MAINTENANCE MANUAL. EDACS REDUNDANT POWER SUPPLY SYSTEM 350A1441P1 and P2 POWER MODULE CHASSIS 350A1441P3, P4, AND P5 POWER MODULES TABLE OF CONTENTS

MAINTENANCE MANUAL. EDACS REDUNDANT POWER SUPPLY SYSTEM 350A1441P1 and P2 POWER MODULE CHASSIS 350A1441P3, P4, AND P5 POWER MODULES TABLE OF CONTENTS MAINTENANCE MANUAL EDACS REDUNDANT POWER SUPPLY SYSTEM 350A1441P1 and P2 POWER MODULE CHASSIS 350A1441P3, P4, AND P5 POWER MODULES TABLE OF CONTENTS SPECIFICATIONS*... 2 INTRODUCTION... 3 DESCRIPTION...

More information

Surge Protective Devices Installation, Operation and Maintenance Manual. LoadTrack LTL

Surge Protective Devices Installation, Operation and Maintenance Manual. LoadTrack LTL LoadTrack LTL Surge Protective Devices Installation, Operation and Maintenance Manual P.O. Box 3760 Winter Park, FL 32790 USA TEL: 800-647-1911 www.tpssurge.com LOADTRACK LTL INSTALLATION, OPERATION AND

More information

Installation Manual. APEX Series. Silicon and MOV Panel Protection. APEX Series

Installation Manual. APEX Series. Silicon and MOV Panel Protection. APEX Series Installation Manual APEX Series Silicon and MOV Panel Protection APEX Series APEX Series Installation Guide 2 Table of Contents Mechanical Installation 3 Table One - Application Rating 3 Figure 1 - Mounting

More information

Catalog Number Identification with Elements Explained V V V V X 347 V Hz

Catalog Number Identification with Elements Explained V V V V X 347 V Hz Owner s Manual 917 & 918 Remote Control Switches TABLE OF CONTENTS page Ratings & Installation... 1 2 Optional Accessories... 3 4 Troubleshooting & Parts Kits... 5 Wiring & Outline Diagrams... back of

More information

Installation Instructions

Installation Instructions 50ES, 50EZ, 50GL, 50GS, 50GX, 50JS, 50JX, 50JZ, 50SD, 50SZ, 50VL, 50VT, 50XP, 50XZ 601A, 602A, 602B, 604A, 604B, 604D, 607C, 701A, 702A, 702B, 704A, 704B, 704D, 707C PA1P, PA2P, PA3P, PH1P, PH2P, PH3P

More information

User s Manual. ACS550-PD 3R Irrigation Packaged Drive Supplement to ACS550-U1 User s Manual

User s Manual. ACS550-PD 3R Irrigation Packaged Drive Supplement to ACS550-U1 User s Manual User s Manual ACS550-PD 3R Irrigation Packaged Drive Supplement to ACS550-U1 User s Manual 2 ACS550-PD 3R Irrigation Packaged Drive ACS550 Drive Manuals GENERAL MANUALS ACS550-U1 User s Manual (1 200 HP)

More information

IIM Warwick Court. Williamsburg VA REV 5

IIM Warwick Court. Williamsburg VA REV 5 1 1-757-258-3939 100 Warwick Court Williamsburg VA 23185 REV 5 2 Technical Specification >Accuracy Better than + 1% over a range of 1.5 to 200 amperes with power factor ranging from 1 to 0.5 Approved by

More information

Overview These instructions are presented as a guideline for installing and setting up the LX Series Breaker Control Panel.

Overview These instructions are presented as a guideline for installing and setting up the LX Series Breaker Control Panel. LX Breaker Control Panels Installation and Setup Procedure Hubbell Building Automation, Inc. 9601 Dessau Road Building One Suite 100 Austin, Texas 78754 512-450-1100 512-450-1215 Fax www.hubbell-automation.com

More information

ASCO SERIES 300 Manual Transfer Switch Solutions

ASCO SERIES 300 Manual Transfer Switch Solutions ASCO SERIES 300 Manual Transfer Switch Solutions When power fails, businesses suffer. For some organizations, a permanent generator is too costly. Instead, permanent means for quickly connecting a portable

More information

DTS 310. DIN Rail Mounted, Indoor Rated Revenue Grade Electrical Sub-meter. Page 1. Measurlogic MQ A

DTS 310. DIN Rail Mounted, Indoor Rated Revenue Grade Electrical Sub-meter. Page 1. Measurlogic MQ A DTS 310 DIN Rail Mounted, Indoor Rated Revenue Grade Electrical Sub-meter Page 1 1 PRODUCT OVERVIEW... 3 1.1 SUPPLIED ITEMS... 3 1.2 DOCUMENT CONVENTIONS... 4 1.3 PRODUCT SPECIFICATION... 4 1.3.1 Current

More information

READ AND FOLLOW ALL SAFETY INSTRUCTIONS! SAVE THESE INSTRUCTIONS AND DELIVER TO OWNER AFTER INSTALLATION IMPORTANT SAFETY INSTRUCTIONS

READ AND FOLLOW ALL SAFETY INSTRUCTIONS! SAVE THESE INSTRUCTIONS AND DELIVER TO OWNER AFTER INSTALLATION IMPORTANT SAFETY INSTRUCTIONS LRC0 8X0A Relay Control Panel BACnet MS/TP Enabled INSTALLATI INSTRUCTIS READ AND FOLLOW ALL SAFETY INSTRUCTIS! SAVE THESE INSTRUCTIS AND DELIVER TO OWNER AFTER INSTALLATI IMPORTANT SAFETY INSTRUCTIS!

More information

DeltaV Bulk Power Supplies

DeltaV Bulk Power Supplies DeltaV Distributed Control System Product Data Sheet DeltaV Bulk Power Supplies DeltaV 5 A, 10A, 20 A, 40 A Bulk Power Supplies. DeltaV 20 A, 40 A and 80 A Bulk Power Supply Redundancy Modules. Easy to

More information

BSPD Surge Protective Devices

BSPD Surge Protective Devices BUSSMANN Supersedes December 2016 SERIES BSPD Surge Protective Devices High capacity Type 1 and Type 2 SPDs for overvoltage protection on Contents Description Page Introduction 2 Applications 2 Features

More information

60W Power over Ethernet Waterproof Adapter PoE IEEE BT Single Port Injector for Outdoor Application

60W Power over Ethernet Waterproof Adapter PoE IEEE BT Single Port Injector for Outdoor Application WWW.PHIHONG.COM 60W Power over Ethernet Waterproof Adapter PoE IEEE BT Single Port Injector for Outdoor Application Features Compliant with the IEEE802.3bt Standard Non-Vented Case with Mounting Bracket

More information

Installation Instructions for Eaton Surge Protective Device XXCFXXX10-DIN and XXCFXXX10-DIN2

Installation Instructions for Eaton Surge Protective Device XXCFXXX10-DIN and XXCFXXX10-DIN2 Supersedes 6/2015 Surge Protective Device XXCFXXX10-DIN and XXCFXXX10-DIN2 XXCFXXX10-DIN Contents Description Page 1.0 Setup...2 1.1 Before Installation...2 1.2 Installation...3 For AC Applications...3

More information

S66 Series Electronic Fan Speed Control

S66 Series Electronic Fan Speed Control FANs 121, 125, 1628.3 Product/Technical Bulletin S66 Issue Date 0918 S66 Series Electronic Fan Speed Control The S66 Series Electronic Fan Speed Control is designed to modulate the speed of single-phase,

More information

Total Home Surge Protection for Point-of-Entry. FirstSurge TM. User Manual - Canada (English)

Total Home Surge Protection for Point-of-Entry. FirstSurge TM. User Manual - Canada (English) Total Home Surge Protection for Point-of-Entry FirstSurge TM User Manual - Canada (English) www.siemens.ca/surge Introduction Thank you for choosing the Siemens FirstSurge Residential Surge Protective

More information

DeltaV Bulk Power Supplies

DeltaV Bulk Power Supplies DeltaV Distributed Control System Product Data Sheet DeltaV Bulk Power Supplies DeltaV 5 A, 10A, 20 A, 40 A Bulk Power Supplies. DeltaV 20 A, 40 A and 80 A Bulk Power Supply Redundancy Modules. Easy to

More information

Thermometer / Barometer System

Thermometer / Barometer System TBX USER S MANUAL Thermometer / Barometer System 616TA Pg.2of12 TABLE OF CONTENTS QUICK START GUIDE... 3 INTRODUCTION... 6 SPECIFICATIONS... 6 TERMINAL COMPARTMENT LAYOUT... 7 FEATURES... 8 OPERATION...

More information

AcquiSuite Ally 12 & 48 Advanced Multi-Circuit Meter

AcquiSuite Ally 12 & 48 Advanced Multi-Circuit Meter AcquiSuite Ally 12 & 48 Advanced Multi-Circuit Meter Install Guide Revision C (12/18) DANGER HAZARD OF ELECTRIC SHOCK, EXPLOSION, OR ARC FLASH Revision C (12/18) Turn off all power supplying equipment

More information

to 48V DC Power Supply

to 48V DC Power Supply Telecommunications Group Section 855-080-201 Equipment Issue 1 Second Printing, July 2007 8550-80 24 to 48V DC Power Supply Compliant with UL Standard 60950, Second Edition* CONTENTS PAGE Part 1. GENERAL.............................................................................

More information

SLI 15 Inverter Series Data Sheet 1500 Watt, 1U Compact Inverters

SLI 15 Inverter Series Data Sheet 1500 Watt, 1U Compact Inverters Applications Telecom IT Industrial Safety in compliance with IEC60950:1999, 3 rd edition; EN60950:2000; UL60950, 3 rd edition; CSA Standard C22.2 No.60950-00, 3 rd edition Features New, compact design:

More information

Surge Protective Devices. STV25K Series. Instruction Manual

Surge Protective Devices. STV25K Series. Instruction Manual Surge Protective Devices STV25K Series Instruction Manual Contents 1.0 Installation & Operation... 4 1.1 Environment... 4 1.2 Mechanical Dimensions... 4 1.3 DIN Rail Mounting... 5 1.4 Supplemental Enclosures...

More information

SECTION 16496E ATS MOLDED CASE WITH ATC 900 CONTROLLER

SECTION 16496E ATS MOLDED CASE WITH ATC 900 CONTROLLER ATS MOLDED CASE WITH ATC 900 CONTROLLER PART 1 GENERAL 01 02 03 04 SCOPE Furnish and install the low voltage automatic transfer switches having the ratings, features/accessories and enclosures as specified

More information

Expansion Unit Catalog Nos , - 152, - 153, - 154, - 156, -E157

Expansion Unit Catalog Nos , - 152, - 153, - 154, - 156, -E157 PRODUCT DA TA SLC 150 110 Expansion Unit Catalog Nos. 1745-151, - 152, - 153, - 154, - 156, -E157 7 : The EXpdnSiQn Unit The SLC 150 expansion unit can be used with either the SLC 150 processor unit or

More information

CONTROL PANEL WIRING SPECIFICATIONS

CONTROL PANEL WIRING SPECIFICATIONS CONTROL PANEL WIRING ISSUE STATUS FIRST REV DESCRIPTION DATE 001 1 ST ISSUE 22/04/2014 002 1 ST AMMENDMENT 10/03/2017 003 2 ND AMMENDMENT 08/09/2017 THIS DOCUMENT IS DESIGNED TO COVER ALL APPLICATIONS,

More information

PGR-4300 MANUAL GENERATOR GROUND-FAULT RELAY

PGR-4300 MANUAL GENERATOR GROUND-FAULT RELAY Tel: +1-800-832-3873 E-mail: techline@littelfuse.com www.littelfuse.com PGR-4300 MANUAL GENERATOR GROUND-FAULT RELAY REVISION 3-B-041318 Copyright 2018 by Littelfuse, Inc. All rights reserved. Document

More information

Specifications 1 Installation Procedures 2 Wiring Diagrams 3 Installation Checklist 5 Maintenance 5 Warranty Information 6 Material Return Policy 6

Specifications 1 Installation Procedures 2 Wiring Diagrams 3 Installation Checklist 5 Maintenance 5 Warranty Information 6 Material Return Policy 6 CFS Family TITLE PAGE Specifications 1 Installation Procedures 2 Wiring Diagrams 3 Installation Checklist 5 Maintenance 5 Warranty Information 6 Material Return Policy 6 Part Number Model Description A70-00-5006

More information

Dual-Feed 80A Auto Low Voltage Disconnect Panel - 80LVD02

Dual-Feed 80A Auto Low Voltage Disconnect Panel - 80LVD02 Dual-Feed 80A Auto Low Voltage Disconnect Panel - 80LVD02 User Manual 1.1 Overview Telect s 80A, -48 Vdc Low-Voltage Disconnect (LVD) panel is a rack-mounted component with coil-operated contactors. These

More information

Occupancy and Vacancy Sensors ROOM CONTROLLER* DAYLIGHT SENSOR WPP-INT INSTALLATION GUIDE

Occupancy and Vacancy Sensors ROOM CONTROLLER* DAYLIGHT SENSOR WPP-INT INSTALLATION GUIDE Occupancy and Vacancy Sensors ROOM CONTROLLER* DAYLIGHT SENSOR WPP-INT INSTALLATION GUIDE 2 Part Number: WIR-3110 WARNING! SYSTEM MUST BE INSTALLED IN ACCORDANCE WITH LOCAL AND NATIONAL ELECTRICAL CODES

More information

SED2 Variable Frequency Drives with Electronic (E) Bypass Options

SED2 Variable Frequency Drives with Electronic (E) Bypass Options SED2 Variable Frequency Drives with Electronic (E) Options Description The E- Options are companion packages for the family of SED2 Variable Frequency Drives (s). For information on the family of SED2

More information

C. New or replacement products installed in existing facilities shall be consistent with existing equipment.

C. New or replacement products installed in existing facilities shall be consistent with existing equipment. September 2012, rev 00 26 2923 Variable Frequency Controllers PART 1. GENERAL 1.01 Description A. This specification is to cover a complete Variable Frequency motor Drive (VFD) consisting of a pulse width

More information

MG24434 Supplementary Protector SUPPLEMENTARY PROTECTOR, 240VAC, 16A, CIRCUIT BREAKER TYPE: STANDARD

MG24434 Supplementary Protector SUPPLEMENTARY PROTECTOR, 240VAC, 16A, CIRCUIT BREAKER TYPE: STANDARD United States MG24434 Supplementary Protector SUPPLEMENTARY PROTECTOR, 240VAC, 16A, CIRCUIT BREAKER TYPE: STANDARD Product Information Features and Specifications For Use With: OEM Panels Approvals: UL

More information

INSTALLATION INSTRUCTIONS STZ-R SERIES (ENGLISH)

INSTALLATION INSTRUCTIONS STZ-R SERIES (ENGLISH) Surge-Trap STZ-R series Surge Protective Device (SPD) Danger Hazard of Electrical Shock, Explosion or Arc Flash Only qualified licensed electricians should install or service SPDs Verify that all power

More information

HomeWorks Remote Power Feed-Through Panels

HomeWorks Remote Power Feed-Through Panels power feed-through panels are available in two different sizes, each of which may be either surface-mounted or recess-mounted, in an electrical closet or other equipment room. The number of remote power

More information

PGR-4700 MANUAL SENSITIVE GROUND-FAULT RELAY

PGR-4700 MANUAL SENSITIVE GROUND-FAULT RELAY Tel: +1-800-832-3873 E-mail: techline@littelfuse.com www.littelfuse.com/pgr-4700 PGR-4700 MANUAL SENSITIVE GROUND-FAULT RELAY REVISION 2-B-032218 Copyright 2018 by Littelfuse, Inc. All rights reserved.

More information

QuantumR Light Management Hub

QuantumR Light Management Hub Quantum 369375e 1 04.02.2015 QuantumR Light Management Hub The Quantum light management hub provides a centralized connection point for Lutron EcoSystem digital ballast modules, Lutron power panels, GRAFIK

More information

Sync Control. Installation/Operation Manual Rev. E

Sync Control. Installation/Operation Manual Rev. E 9 315 Sync Control Installation/Operation Manual 164201140 Rev. E ------------------------------------------------------------------------ ------------------------------------------------------------------------

More information

QuantumR Light Management Hub (QP2)

QuantumR Light Management Hub (QP2) Quantum 369375f 1 11.05.15 QuantumR Hub (QP2) The Quantum light management hub provides a centralized connection point for Lutron EcoSystem digital ballast modules, Lutron power panels, and QS devices

More information

60103 Circuit Breaker

60103 Circuit Breaker 6003 Circuit Breaker United States MINIATURE CIRCUIT BREAKER, 20/240VAC, 2A, CIRCUIT BREAKER TYPE: STANDARD Image not available Product Information Features and Specifications Circuit Breaker Type: Standard

More information

PGR-3200 MANUAL INSULATION MONITOR

PGR-3200 MANUAL INSULATION MONITOR Tel: +1-800-832-3873 E-mail: techline@littelfuse.com www.littelfuse.com PGR-3200 MANUAL INSULATION MONITOR REVISION 3-E-040918 Copyright 2018 by Littelfuse, Inc. All rights reserved. Document Number: PM-1025-EN

More information

AMP Rack User Manual. Publication Rev G

AMP Rack User Manual. Publication Rev G Publication Introduction This manual contains information for the AMP Rack from Forney Corporation, 16479 Dallas Parkway, Suite 600, Addison, Texas. www.forneycorp.com All personnel should become thoroughly

More information

Installation, Testing, and Operating Procedures 30 AMP PORTABLE AND PERMANENT SERIES GFCI SINGLE and MULTIPHASE

Installation, Testing, and Operating Procedures 30 AMP PORTABLE AND PERMANENT SERIES GFCI SINGLE and MULTIPHASE IMPORTANT! Please read all the information on this sheet. SAVE THESE INSTRUCTIONS! NOTICE BEFORE USING READ INSTRUCTIONS COMPLETELY. TO BE INSTALLED BY A QUALIFIED ELECTRICIAN IN ACCORDANCE WITH NATIONAL

More information

Single Phase Systems-North American PC125/PC420 Series

Single Phase Systems-North American PC125/PC420 Series Single Phase Systems-North American PC125/PC420 Series Enclosure Power Distribution Units (epdu ) PC125-C Front RACK MOUNTED 19 x 3.4 (2U) x 8.5. Approximate shipping weight: 19 lbs. (12) NEMA OUTLETS

More information

PanelView Plus/VersaView CE Terminals and Display Modules

PanelView Plus/VersaView CE Terminals and Display Modules Installation Instructions PanelView Plus/VersaView CE Terminals and Display Modules (Catalog Numbers 2711P-xxxxxx, 6182H-xxxxxx) English Inside: Overview...2 For More Information...2 Modular Components...3

More information

CurrentWatch Current Sensors. EGF Series Ground Fault Sensors. Ground Fault Sensors with Solid-State or Mechanical Relay Outputs.

CurrentWatch Current Sensors. EGF Series Ground Fault Sensors. Ground Fault Sensors with Solid-State or Mechanical Relay Outputs. CurrentWatch Current Sensors Ground Fault Sensors with Solid-State or Mechanical Relay s 1 Contents Overview.................... 1 Model Selection, Switches...... 2 Model Selection, Accessories... 3 Wiring

More information

INSTALLATION DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER. Before installation:

INSTALLATION DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER. Before installation: DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER The DKM-409 is a precision instrument designed for displaying various AC parameters in 3-phase distribution panels. Thanks to its isolated

More information

24973 Circuit Breaker MINIATURE CIRCUIT BREAKER, 440VAC, 10A, CIRCUIT BREAKER TYPE: STANDARD

24973 Circuit Breaker MINIATURE CIRCUIT BREAKER, 440VAC, 10A, CIRCUIT BREAKER TYPE: STANDARD United States 24973 Circuit Breaker MINIATURE CIRCUIT BREAKER, 440VAC, 10A, CIRCUIT BREAKER TYPE: STANDARD Image not available Product Information Features and Specifications Approvals: IEC 60947-2 Rated

More information

1. Section Lightning Protection for lightning protection system. D. The installation shall be in conformance with NEC Article 285.

1. Section Lightning Protection for lightning protection system. D. The installation shall be in conformance with NEC Article 285. SECTION 26 04 75 - SURGE PROTECTION DEVICES PART 1 - GENERAL 1.1 SUMMARY A. Applicable provisions of the General Conditions, Supplementary General Conditions and Special Conditions shall govern work performed

More information

PGR-6101 MANUAL GROUND-FAULT & INSULATION MONITOR

PGR-6101 MANUAL GROUND-FAULT & INSULATION MONITOR Tel: +1-800-832-3873 E-mail: techline@littelfuse.com www.littelfuse.com PGR-6101 MANUAL GROUND-FAULT & INSULATION MONITOR Revision 0-C-041918 Copyright 2018 by Littelfuse, Inc. All rights reserved. Document

More information

Table of Contents. Safety. Attention. Warnings

Table of Contents. Safety. Attention. Warnings Solstart Miniature Soft Starter 8-58A, 220-600V Instruction Manual Ver. 21.2. 2002 Table of Contents Page Subject 3 Starter Selection 4 Installation Notes 5 Wiring 6 Starter Settings & Start-up Procedure

More information

ABL7RM24025 regulated SMPS with auto reset - 1 or 2-phase V AC - 24 V A

ABL7RM24025 regulated SMPS with auto reset - 1 or 2-phase V AC - 24 V A Characteristics regulated SMPS with auto reset - 1 or 2-phase - 100...240 V AC - 24 V - 2.5 A Product availability : Stock - Normally stocked in distribution facility Price* : 184.00 USD Main Range of

More information

Line reactors SINAMICS. SINAMICS G130 Line reactors. Safety information 1. General. Mechanical installation 3. Electrical installation

Line reactors SINAMICS. SINAMICS G130 Line reactors. Safety information 1. General. Mechanical installation 3. Electrical installation Safety information 1 General 2 SINAMICS SINAMICS G130 Mechanical installation 3 Electrical installation 4 Technical specifications 5 Operating Instructions Control version V4.7 04/2014 A5E00331462A Legal

More information

B60 Domestic Voltage Optimisation Unit

B60 Domestic Voltage Optimisation Unit B60 Domestic Voltage Optimisation Unit Installation Manual V1.2 Nov 2013 1 P a g e PRODUCT ADVISORY NOTICE This product must be installed by a competent person in accordance with the IEE Wiring Regulations,

More information

OVERLOAD RELAYS POWER & ACTUATION PROVEN. series

OVERLOAD RELAYS POWER & ACTUATION PROVEN. series POWER & ACTUATION OVERLOAD RELAYS series Need reliable, accurate overload and phase-loss protection for your motors? c3controls Series Bimetallic Overload Relays provide unmatched protection and can be

More information

SCS AUTOMATION & CONTROL

SCS AUTOMATION & CONTROL - 1 - SCS AUTOMATION & CONTROL PROJECT XXXX Issue A Winder Customer: XXXXXXX Ltd.. Automation Center 156 Stanley Green Road Poole Dorset England BH15 3AH - 2 - Contents 1. Introduction 2. Safety 3. Specification

More information

Premium Online Series

Premium Online Series Premium Online Series UPS backup power systems backup protection with double conversion that isolates and conditions output power for clean, consistent, reliable power features: Permanent connection to

More information

Page Housing Plastic Metal Mounting DIN Rail or surface mount Panel or bracket mount Wattage Range 10W to 240W 7.5W to 240W 10W to 300W

Page Housing Plastic Metal Mounting DIN Rail or surface mount Panel or bracket mount Wattage Range 10W to 240W 7.5W to 240W 10W to 300W Selection Guide Selection Guide PS5R Slim Line PS5R PS3L Operator Interfaces Communication & Networking PLCs NEW Models Appearance Automation Software Page 107 113 118 Housing Plastic Metal Mounting DIN

More information

PV Rapid Shutdown device

PV Rapid Shutdown device PV Rapid Shutdown device Installation and Operation Manual Solis-RSD-1G(1:1) Solis-RSD-1G(2:2) Manufacturer: Ginlong (Ningbo) Technologies Co.,Ltd., Ningbo, Zhejiang, P.R.China US Office: 565 Metro Pl.

More information

Model KEP. Transfer Switch Standard Features. Automatic Transfer Switches Service Entrance Rated. Enclosed Contact Power Switching Units.

Model KEP. Transfer Switch Standard Features. Automatic Transfer Switches Service Entrance Rated. Enclosed Contact Power Switching Units. Model KEP Automatic Transfer Switches Service Entrance Rated Controller D Decision-Makerr MPAC 1500 Ratings Power Switching Device Molded case (MCCB) Insulated Case (ICCB) Current Voltage, Frequency 200

More information

PDI Modular Compact Remote Power Panel

PDI Modular Compact Remote Power Panel Guide Specifications (Revision 000, 6/17/2016) 1 GENERAL 1.1 Summary These specifications describe requirements for a remote power panel (RPP) distributing power to sensitive loads. These specifications

More information

Three Phase Systems-International PC302-I/MTD

Three Phase Systems-International PC302-I/MTD Three Phase Systems-International PC302-I/MTD Enclosure Power Distribution Units (epdu ) sales@pulizzi.com 800-870-2248 Fax: 605-334-4999 Front RACK MOUNTED 19 x 3.4 (2U) x 8.5 with recess mounting Approximate

More information

Expandable Power Systems

Expandable Power Systems Expandable Power Systems Installation Guide Models Include: Maximal11EV 12VDC @ 4A or 24VDC @ 3A. 12VDC @ 4A or 24VDC @ 3A. Maximal33EV 12VDC or 24VDC @ 6A. 12VDC or 24VDC @ 6A. Maximal37EV 24VDC @ 10A.

More information

Model NM512. TS1 Conflict Monitor Standard 12 Channel Unit. Naztec Operations Manual. For. Published by

Model NM512. TS1 Conflict Monitor Standard 12 Channel Unit. Naztec Operations Manual. For. Published by Naztec Operations Manual For Model NM512 TS1 Conflict Monitor Standard 12 Channel Unit Published by Naztec, Inc. 820 Park Two Dr. Sugar Land, Texas 77478 Phone: (281) 2407233 Fax: (281) 240-7238 Copyright

More information

SED2 VFD NEMA Type 3R Bypass/Air Conditioner

SED2 VFD NEMA Type 3R Bypass/Air Conditioner SED2 VFD NEMA Type 3R Bypass/Air Conditioner Description The NEMA Type 3R Bypass with Air Conditioner allows SED2 to be employed in a harsh environment. The SED2 is designed specifically for HVAC applications

More information

SAVE THESE INSTRUCTIONS

SAVE THESE INSTRUCTIONS READ AND FOLLOW ALL SAFETY INSTRUCTIONS! SAVE THESE INSTRUCTIONS AND DELIVER TO OWNER AFTER INSTALLATION IMPORTANT SAFEGUARDS! When using electrical equipment, basic safety precautions should always be

More information

A0/A1 Controller Assembly Operation/Installation Manual READ AND FOLLOW ALL SAFETY INSTRUCTIONS

A0/A1 Controller Assembly Operation/Installation Manual READ AND FOLLOW ALL SAFETY INSTRUCTIONS Document Number: 9100-127-2301-99 Release Date: 2/2/15 A0/A1 Controller Assembly Operation/Installation Manual READ AND FOLLOW ALL SAFETY INSTRUCTIONS DO NOT let any supply cords touch hot surfaces higher

More information

Model HM-535 Power Supply Installation and Service Instructions

Model HM-535 Power Supply Installation and Service Instructions Model HM-535 Power Supply Installation and Service Instructions 430-535 0104 2004 Heritage MedCall, Inc SENTRY INSTALLATION & SERVICE INSTRUCTIONS POWER SUPPLY UNIT Model HM-535 IMPORTANT SAFETY INSTRUCTIONS

More information

ICS Regent. Relay Output Modules. Low Power and High Power (T3446L and T3446H) PD-6017

ICS Regent. Relay Output Modules. Low Power and High Power (T3446L and T3446H) PD-6017 ICS Regent PD-6017 Relay Output Modules Low Power and High Power (T3446L and T3446H) Issue 1, March, 06 Relay output modules provide control of eight user output loads. Two types of relay output modules

More information

Symmetra PX 160. Maintenance Bypass Enclosure. Installation 05/

Symmetra PX 160. Maintenance Bypass Enclosure. Installation 05/ Symmetra PX 160 Maintenance Bypass Enclosure Installation 05/2014 www.schneider-electric.com Legal Information The Schneider Electric brand and any registered trademarks of Schneider Electric Industries

More information

SAVE THESE INSTRUCTIONS

SAVE THESE INSTRUCTIONS READ AND FOLLOW ALL SAFETY INSTRUCTIS! SAVE THESE INSTRUCTIS AND DELIVER TO OWNER AFTER INSTALLATI IMPORTANT SAFETY INSTRUCTIS WARNING To reduce the risk of death, injury or property damage from fire,

More information

SPECIAL INSTRUCTIONS FOR CAPACITORS COMPACT GENERATORS

SPECIAL INSTRUCTIONS FOR CAPACITORS COMPACT GENERATORS SPECIAL INSTRUCTIONS FOR CAPACITORS COMPACT GENERATORS (WITH CAPACITOR CHARGER BOARD A3517-02) The process depends on Generator and System configuration. This document applies to installation of Capacitors

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS TM REPLACEMENT PARTS FOR USE WITH CX SERIES PANELS INSTALLATION INSTRUCTIONS Hubbell Control Solutions 9601 Dessau Road Building One Suite 100 Austin, TX 78754 Toll Free: 888-698-3242 Fax: 512-450-1215

More information

AL168 Series Multi-Output Power Supplies

AL168 Series Multi-Output Power Supplies AL168 Series Multi- Power Supplies Installation Guide Models Include: AL168CB - 16VAC @ 6A (100VA) or 18VAC @ 5.5A (100VA). AL168175CB - 16VAC @ 10A (175VA). AL168300CB - 16VAC @ 18A (350VA) or 18VAC @

More information

ASCO SERIES 300 Service Entrance Power Transfer Switch

ASCO SERIES 300 Service Entrance Power Transfer Switch ASCO SERIES 300 Service Entrance Power Transfer Switch The ASCO Service Entrance Power Transfer Switch combines automatic power switching with the necessary disconnecting, grounding, and bonding required

More information

SAVE THESE INSTRUCTIONS

SAVE THESE INSTRUCTIONS OUTDOOR HARDWIRE INSTALLATION INSTRUCTIONS Please read and save these instructions. Read carefully before using product. Protect yourself and others by observing all safety information, warnings and cautions.

More information

SAVE THESE INSTRUCTIONS

SAVE THESE INSTRUCTIONS READ AND FOLLOW ALL SAFETY INSTRUCTIS! SAVE THESE INSTRUCTIS AND DELIVER TO OWNER AFTER INSTALLATI IMPORTANT SAFETY INSTRUCTIS WARNING To reduce the risk of death, injury or property damage from fire,

More information

ZC Series Zone Monitoring Controllers

ZC Series Zone Monitoring Controllers ZC Series Zone Monitoring Controllers Installation Instructions MANUAL Reset Controllers Model Description Part Number ZC-1 1 Zone Controller 0421 ZC-2 2 Zone Controller 0422 ZC-3 3 Zone Controller 0423

More information