Signet 2750 DryLoc ph/orp Sensor Electronics
|
|
- Spencer Cameron
- 5 years ago
- Views:
Transcription
1 Signet 750 DryLoc ph/orp Sensor Electronics * * Rev. G /05 English WARNING! SAFETY INSTRUCTIONS. Depressurize and vent system prior to installation or removal.. Confi rm chemical compatibility before use.. Do not exceed maximum temperature/pressure specifi cations.. Wear safety goggles or faceshield during installation/service. 5. Do not alter product construction. 6. When using chemicals or solvents care should be taken and appropriate eye, face, hand, body, and/or respiratory protection should be used.. Description Signet 750 ph/orp Sensor Electronics provide twowire to 0mA loop output for ph and ORP measurements, without the expense of local display and other luxuries available in fullfeatured transmitters. Preamplifi cation is builtin, reducing system costs while ensuring absolute signal integrity up to 0.8m/,000 feet. These sophisticated fi eldmount devices also provide the digital (S L : Signet Sensor Serial Link) output, the latest development in leadingedge technology from Signet. The 750 Sensor Electronics selfconfi gure for ph or ORP operation via automatic recognition of electrode type, and the DryLoc electrode connector quickly forms a robust assembly for submersible and inline installations. NEMA X Junction Boxes are integral parts of the inline version and are available as accessories for the submersible version. The optional EasyCal feature allows simple pushbutton calibration and includes an LED for visual feedback. English. Description. Specifi cations. Inline Installation. 750 Inline Assembly 5. Submersible Installation 6. Digital (S L) wiring 7. 0 ma Loop wiring 8. Calibration 9. Troubleshooting 0. Ordering Information. Specifications General Compatible Electrodes: Signet DryLoc Electrodes Operational Range: 0.00 to.00 ph, ±000 mv ORP 0 C to 85 C (ºF to 85ºF) Response Time (includes electrode response): ph: <6 s for 95% of change ORP: application dependent ph Temp τ: 0 s (75) 96 s (756) 8 s (756WT) Materials: 750, (inline): PBT (thermal plastic polyester) 750, (submersible): CPVC Cable for, submersible versions: 5 m (5 ft.) conductor shielded, AWG. May be extended up to 8 m (600 ft) with current output. May be extended up to 05 m (000 ft) with S L output. Weight: 750, (inline): 0.75 kg (.75 lb.) 750, (submersible): 0.6 kg (. lb.) Environmental Ambient Temp.: 0 C to 85 C Storage Temp.: 0 C to 85 C Relative Humidity: 95% max, noncondensing Immunity: EN5008 Emissions: EN5008 Enclosure Rating: 750, (inline): NEMA X/IP65 (with electrode connected) 750, (submersible): NEMA 6P/IP68 (with electrode and watertight extension pipe connected) Electrical Input Impedance: >0 Ω Input response time: 500 ms Temperature drift: ±0.00 ph per ºC ±0. mv ORP per ºC Input resolution: 0.0 ph, mv ORP, 0.ºC Current output: Description: ph: Fixed to 0 ma, isolated, 0 to ph (custom scaling available) ORP: Fixed to 0 ma, isolated, 000 to 000 mv (custom scaling available, 000 to 000 mv) Power: Regulated VDC ± 0%, 0 ma max. Max Loop Resistance: 50Ω V 5Ω 8V 600Ω V Accuracy: 5ºC Temperature drift: ±µa per ºC Output resolution: ± 5 µa Error indication:.6 ma Digital (S L) output: Description: Serial ASCII, TTL level 9600 bps Power: 5VDC ± 0% regulated, ma max Accuracy: ph: ± C ORP: ± 5 C Resolution: ph: 0.0 ph ORP: mv Temp. 0. C Error indication: Temp output "999.9" Standards and Approvals CE Manufactured under ISO 900 & ISO 00
2 . Inline Dimensions The 750 inline sensor (without EasyCal) is designed for applications where electrode calibration is facilitated by remote equipment. The 750 InLine sensor includes EasyCal. A Signet installation fi tting ( / in. to in.) or a Signet pipe adapter is required to secure the electrode in the pipe. The Signet Measurement and Instrumentation catalog offers a complete selection of fi ttings. The 750 and 750 submersible sensor electronics are recommended for inline applications using the 76 series differential electrodes and the 77 series Threaded DryLoc electrodes. 76 and 77 series electrodes thread directly into standard in. or / in. NPT fi ttings. 5 mm (.5 in.) 78 mm (7 in.) 89. mm (.5 in.).7 in./ 9 mm 8 mm (9.75 in.) 5.5 in./ 0 mm 5 mm (.75 in.) Mounting position 75 and 76 series electrodes must be mounted upright. Vertical (0 ) position optimum Do not install within 0º of horizontal (Contact factory for horizontal or inverted installation requirements.) Recommended maximum fl ow rate 0 ft/s. Wettap electrodes and 77 series electrodes have pressurized electrolyte chambers, enabling them to be mounted at any angle. Vertical ± 60 YES YES 0 0 NO NO NO Signet 750 DryLoc ph/orp Sensor Electronics
3 . 750 Inline ph Sensor Assembly Note: This procedure applies to systems using standard electrodes. If the 79 ph/orp WetTap is used, refer to the 79 manual for instructions.. Insert electrode into Signet installation fi tting. Seat the electrode tabs into the alignment notches in the fi tting.. Thread the inline retaining cap onto fi tting to secure the electrode into place. Handtighten the cap. DO NOT use any tools to avoid damaging the threads. Retaining cap Note: Keep the electrical interconnection between electrode and sensor electronics dry and clean at all times.. Unlock the ring on base of 750 (The ring is unlocked when the lines on the ring and body of the 750 are not aligned.) Locking Ring Unlocked The molded alignment marks are enhanced for clarity.. Place 750 onto top of electrode and turn until the assembly drops into position. The molded alignment marks are enhanced for clarity. 5. Turn locking ring / turn to secure the 750 assembly. (The ring is locked when the line on the ring and the line on the body of the 750 are aligned.) Locked InLine Sensor and Electrode Removal To remove electronics assembly only: Turn locking ring / turn; lift assembly straight up. To remove electrode from the pipe: Remove electronics assembly, then unthread retaining cap; pull electrode straight up. To remove the complete 750 system: Hold locking ring in place. Unthread retaining cap and pull electrode straight up. Signet 750 DryLoc ph/orp Sensor Electronics CAUTION! Do not remove the electrode from a pressurized pipe. Wear appropriate protective clothing when working with chemicals in pressurized pipe.
4 5. 750, Submersible Sensor Assembly and Installation. Insert electrode into base of the 750 or 750 and turn until keyed contacts are seated.. Thread retaining cap over electrode and handtighten onto Attach / in. watertight pipe to the top of the 750. Secure the threaded connection to prevent any leakage. For additional defense against possible accumulation of condensation at the back seal area of the sensor, fi ll the lower inches (7500 mm) of conduit or extension pipe with a fl exible sealant such as silicone. 750: / in. NPT threads 750: ISO 7/R / threads A light coat of nonpetroleum based lubricant on the Oring will ease installation. 08 mm/ 8. in. Fill with 7500 mm ( to in.) of sealant Note: Keep the electrical interconnection between electrode and sensor electronics dry and clean at all times. 55 mm/.6 in. Use standard installation hardware to connect the submersible 750 or directly to external equipment. OR The 805 NPT Mount Junction box (or 805 with EasyCal) connects to / in. pipe or conduit and provides convenient wiring termination. OR The 8050 Universal Mount junction box (or 8050 with Easy Cal) mounts fl at onto a wall or can be strapped to a post or pipe. 9 mm/.7 in. 06 mm/. in. 07 mm/. in. / in. NPT threads 9 mm/.7 in. Fill with to in. of sealant Fill with to in. of sealant Cable supplied: 5 m (5 ft.) conductor shld, AWG. May be extended up to 8 m (600 ft) with current output. May be extended up to 05 m (000 ft) with S L output. Signet 750 DryLoc ph/orp Sensor Electronics
5 6. Digital (S L) wiring When the 750 is powered with 5 VDC, the digital (S L) serial data output is automatically selected. S L data is used exclusively by Signet instruments. Remove approximately 0 mm (0. in.) of insulation and tin each conductor before inserting into connectors. S L with no junction box S L with Junction box Black Red White Shield 5 VDC SL data Signal Ground SW Run SW Cal D S 5 VDC S L data Signal Ground Optional Earth ground Connect the 750 cable directly to S L I/O terminals users: If this direct wiring is used, set the CALIBRATE menu to "perform calibration at INSTRUMENT" When the 750 includes a terminal block, connect the 750 terminals as shown to any S L I/O port users: If the 750 includes the EasyCal accessory, set the CALIBRATE menu to "perform Calibration at either SENSOR or INSTRUMENT" If SENSOR, use standard ph buffer values (ph, 7, or 0) to perform periodic calibration. If INSTRUMENT, any ph value can be used ma Loop wiring When the 750 is powered with to VDC, the 0 ma loop output is automatically selected. Remove approximately 0 mm (0. in.) of insulation and tin each conductor before inserting into connectors. Current Loop with no junction box Current loop with Junction box Black Red (no connection) White Shield Optional Earth Ground Power Supply DC V SW Run SW Cal D S Power Supply DC V 0 0 ma loop monitor Optional Earth Ground 0 0 ma loop monitor Connect the 750 cable directly to a Loop device as shown. This confi guration does not provide any calibration capability within the 750 system. Periodic calibration must be performed at the external equipment. When the 750 includes a terminal block, connect the 750 terminals to the Loop device as shown. If the 750 includes the EasyCal accessory, use standard ph buffer values (ph, 7, or 0) to perform periodic calibration. Signet 750 DryLoc ph/orp Sensor Electronics 5
6 8. Calibration All 750 ph/orp Sensor Electronics are factorycalibrated for maximum outofthebox accuracy. Periodic calibration is required to compensate for electrode aging. The optional EasyCal feature allows calibration to be performed local to the sensor. SW EasyCal button SW RUN/CALIBRATE switch for S L system D EasyCal Indicator (green LED) SW Run S SW Cal D S SW Run SW Cal BLK RED WHT SHLD D EasyCal Procedure: The fi rst step (Reset) is recommended each time an electrode is replaced, but is NOT necessary upon initial installation or periodic calibration. In fact, for periodic calibration it is best for the electrode/sensor assembly to remain intact to minimize the possibility of moisture or other contamination entering the electrical interconnection area. The electrode/sensor connection must remain dry and clean at all times.. Reset the 750 ph/orp Sensor to factory calibration: With no electrode connected to the Sensor, press and hold SW until the LED (D) comes on steady then goes off again (approx. 0 seconds). When the LED goes off, release SW; reset is complete.. Connect an electrode to the 750 ph/orp Sensor.. If S L output is being used, place SW in the Cal position. If 0 ma output is being used, SW position is of no consequence.. Place the electrode/sensor assembly into a calibration solution as follows: (If the electrode is healthy, then the 750 will automatically recognize the solution. The order in which the solutions are used during the calibration procedure is of no consequence.) For ph calibration, use any two of these international standards: ph.0, 7.0 or 0.0 buffer solutions. (Signet part number contains one capsule of each value) To produce standards for ORP calibration, mix the chemical Quinhydrone into ph 7.0 and.0 buffers to saturation (/8g per 50ml). Regardless of the size of the container used for calibration, one inch of solution is adequate to completely submerge the tip of the electrode. Allow at least 0 seconds for the electrode response to stabilize before calibration. 5. Press and hold SW for approximately 8 to 0 seconds. During this time, the LED (D) will come on steady then go back off. Release SW (If the LED blinks several times rapidly, the calibration was not successful. See the troubleshooting section.) 6. Remove the electrode/sensor assembly from the fi rst calibration solution, rinse the electrode with clean water, and place it in a second solution. Allow at least 0 seconds for the electrode response to stabilize before calibration. 7. Press and hold SW for approximately 8 to 0 seconds. During this time, the LED will come on steady then go back off. Release SW. (If the LED blinks several times rapidly, the calibration was not successful. See the troubleshooting section.) 8. For S L systems ONLY: Return SW to the RUN position. Calibration is complete. Return the system to service. 6 Signet 750 DryLoc ph/orp Sensor Electronics
7 9. Troubleshooting LED and Output Condition Current Out: LED off, current output is.6 ma S L: Temp out is During EasyCal, the LED blinks rapidly for seconds and the current output is frozen at a random fixed value. After completing calibration procedure, the output values are inaccurate. Possible Causes No electrode installed Bad/dirty contacts between electrode and 750 The buffer solution is outside of the accepted tolerance for the 750 The electrode is depleted (>. ph or 65 mv offset) Insuffi cient time allowed for electrode stabilization during calibration. Suggested Solutions Install electrode Check interconnection between electrode and 750, clean contacts ph system: Use fresh ph, 7 ph, or 0 ph buffer and restart the calibration. ORP system: Use fresh ph and 7 ph buffer solution saturated with quinhydrone. Replace the electrode. Recalibrate, verify that test solutions are at room temperature and wait at least 0 seconds after placing electrode in solution before pressing S EasyCal button. Signet 750 DryLoc ph/orp Sensor Electronics 7
8 0. Ordering Information Mfr. Part No. Code Description Inline Sensor Electronics w/junction Box Inline Sensor Electronics w/junction Box and EasyCal Submersible Sensor Electronics with 5 ft. cable, / in. NPT threads Submersible Sensor Electronics with 5 ft. cable, ISO 7/R / threads Parts and Accessories Mfr. Part No. Code Description DryLoc ph Electrodes Electrode, ph, DryLoc, fl at, PT Electrode, ph, DryLoc, fl at, KΩ 75HF Electrode, ph, DryLoc, fl at, PT000, HFresist (<%) 75HF Electrode, ph, DryLoc, fl at, KΩ, HFresist (<%) Electrode, ph, DryLoc, bulb, PT DI Electrode, ph, DryLoc, bulb, PT000, DI (<00 µs) 756DI Electrode, ph, DryLoc, bulb, KΩ, DI (<00 µs) 756WT Electrode, ph, DryLoc, bulb, PT000, wettap 756WT Electrode, ph, DryLoc, bulb, KΩ, wettap 756WTP Electrode, ph, DryLoc, bulb, PT000, wettap 756WTP Electrode, ph, DryLoc, plastic bulb, KΩ, wettap Electrode, ph, DryLoc, fl at, KΩ, differential, in. NPT process connection Electrode, ph, DryLoc, fl at, PT000, differential, in. NPT process connection Electrode, ph, DryLoc, fl at, 00Ω, differential, in. NPT process connection Electrode, ph, DryLoc, bulb, KΩ, differential, in. NPT process connection Electrode, ph, DryLoc, bulb, PT000, differential, in. NPT process connection Electrode, ph, DryLoc, bulb, 00Ω, differential, in. NPT process connection Electrode, ph, DryLoc, fl at, KΩ, ¾ in. NPT process connection Electrode, ph, DryLoc, fl at, PT000, ¾ in. NPT process connection Electrode, ph, DryLoc, bulb, KΩ, ¾ in. NPT process connection Electrode, ph, DryLoc, fl at, PT000, ¾ in. NPT process connection DryLoc ORP (Redox) Electrodes Electrode, ORP, DryLoc, fl at, 0KΩ ID Electrode, ORP, DryLoc, bulb, 0KΩ ID 757WT Electrode, ORP, DryLoc, bulb, 0KΩ ID, wettap Electrode, ORP, DryLoc, fl at, 0KΩ ID, differential, in. NPT process connection Electrode, ORP, DryLoc, fl at, PT000, differential, in. NPT process connection Electrode, ORP, DryLoc, fl at, 00Ω, differential, in. NPT process connection Electrode, ORP, DryLoc, bulb, 0KΩ ID, differential, in. NPT process connection Electrode, ORP, DryLoc, bulb, PT000, differential, in. NPT process connection Electrode, ORP, DryLoc, bulb, 00Ω, differential, in. NPT process connection Electrode, ORP, DryLoc, fl at, 0KΩ ID, ¾ in. NPT process connection Electrode, ORP, DryLoc, fl at, no tc, ¾ in. NPT process connection Electrode, ORP, DryLoc, bulb, 0 KΩ ID, ¾ in. NPT process connection Electrode, ORP, DryLoc, bulb, no tc, ¾ in. NPT process connection Miscellaneous Accessories Universal Mount Junction Box Universal Mount Jct. Box w/easycal / in. NPT Mount Junction Box / in. NPT Mount Jct. Box w/easycal ph/orp Simulator/System tester Adapter cable to connect 759 and ph Buffer Kit P550P Universal Pipe Adapter PVC P550C Universal Pipe Adapter CPVC P550V Universal Pipe Adapter PVDF Cable, conductor shield (blk/red/wht/shld) AWG (per ft) George Fischer Signet Inc. 0 Aerojet Avenue, El Monte, CA 9788 U.S.A. Tel. (66) Fax (66) For Worldwide Sales and Service, visit our website: Or call (in the U.S.): (800) Rev. G /05 English George Fischer Signet Inc. 00 Printed in U.S.A. on Recycled Paper
SIGNET ph/orp Transmitter
SIGNET 8750- ph/orp Transmitter ENGLISH 3-8750.090- D-4/03 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.
More information+GF+ SIGNET ph/orp Transmitter Instructions
GF SIGNET 8750 ph/orp Transmitter Instructions ENGLISH 8750.090 A9/99 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury.
More informationSignet ph/orp Transmitter
Signet 87501 ph/orp *38750.0901* 38750.0901 Rev. H 06/10 English English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury.
More informationContents. 1. Installation
ENGLISH 8750.090 C/0 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents. Installation. Specifications. Electrical
More informationSignet Pressure Transmitter
Signet 850 Pressure English *850.090* 850.090 Rev. H /06 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.
More informationSignet 2552 Metal Magmeter Flow Sensors
Signet 2552 Metal Magmeter Flow s Features Test certificate included Award winning hot-tap magnetic flow sensor up to DN2550 (102 in.) Patented Magmeter technology* Operating range 0.05 to 10 m/s (0.15
More informationSignet Temperature Transmitter
Loop Output Output 7 Sensr IN (RED) Sensr V (BLACK) Signet 80 Temperature 80.090 English 80.090 Rev. J 0/ English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions
More informationSignet 5700 ph/orp Monitor - ph Instructions
Signet 5700 ph/orp Monitor ph Instructions English *35700.0901* 35700.0901 Rev. K 3/06 English/pH CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output
More informationSignet Conductivity Electrodes
Signet 2839-2842 Conductivity Electrodes Features ± 1% accuracy - Custom calibration certificate provided Dual-threaded Compact electrode length for easy in-line installation in small pipe sizes Triple
More informationSignet 5700 ph/orp Monitor - ORP Instructions
Signet 5700 ph/ Monitor Instructions English *5700.0901* 5700.0901 Rev. K /06 English/ Contents 1. Power Connections. Compatible Electrode/Preamp Wiring. 4 0 ma Current Output Connections 4. Relay Connections
More informationSIGNET 5700 ph/orp Monitor - ORP Instructions
SIGNET 5700 ph/orp Monitor ORP Instructions ENGLISH 35700.0901 H (6/01) English CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections. Follow
More information+GF+ SIGNET Conductivity/Resistivity Transmitter Instructions
GF SIGNET 8850 Conductivity/Resistivity Transmitter Instructions ENGLISH 8850.090 A9/99 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to
More informationSignet Flow Transmitter
Signet 8550 Flow *8550.090* English 8550.090 Rev. L 0/ English WARNING! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.
More informationSignet 9900 Transmitter
Signet 9900 Transmitter Member of the SmartPro Family of Instruments Features Panel Mount Field Mount NEW! Modbus Module supports RS485 Serial Modbus Communications Multiple sensor types supported with
More informationSignet 5900 Salinity Monitor
Signet 5900 Salinity Monitor *35900.0901* 35900.0901 Rev. E 3/06 English Contents 1. Power Connections 2. Compatible Sensor Wiring 3. 4 20 ma Current Output Connections 4. Relay Connections 5. Relay Operation
More informationSignet ph/orp Transmitter * *
Loop 7 6 5 Signet 8750- ph/orp *-8750.090-* CAUTION! Contents 0.0 ph 5.0C Signet ph/orp. Installation. Panel Installation gasket panel terminals mounting bracket 96 mm (.8 in.) 8050 8 mm (. in.) Optional
More informationFLS M9.06 PH/ORP MONITOR. safety instructions. packing List
FLS M9.06 PH/ORP MONITOR safety instructions General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other instruments
More informationGeneral Purpose ph/orp Sensor
Instruction Sheet PN 51A-399VP/rev.J January 2011 Model 399VP General Purpose ph/orp Sensor SPECIFICATIONS Sensor Type: General purpose 399VP Measured Range: ph: ACCUGLASS 0-14 ORP: -1500 to +1500 mv Percent
More informationFLOW MONITOR & TRANSMITTER
FLS M9.02 FLOW MONITOR & TRANSMITTER SAFETY INSTRUCTIONS General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other
More informationINSTRUCTION MANUAL HBPH DIFFERENTIAL 2-WIRE (4-20 ma) ph Sensor
INSTRUCTION MANUAL HBPH DIFFERENTIAL 2-WIRE (4-20 ma) ph Sensor TABLE OF CONTENTS 1. GENERAL INFORMATION 2 2. SPECIFICATIONS 2 3. INSTALLATION 3 3.1. General Instructions 3 3.2. HBPH 4-20 ma Sensor 3 4.
More informationFLS M9.06. ph/orp MONITOR
FLS M9.06 /ORP MONITOR Safety Instructions General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other instruments
More informationOPERATING INSTRUCTION MANUAL MODEL P/R60 & P/R65 DIFFERENTIAL PH & ORP PROBES
OPERATING INSTRUCTION MANUAL MODEL P/R60 & P/R65 DIFFERENTIAL PH & ORP PROBES N116-19 REV. 10 Water Analytics, Inc. 100 School Street Andover, MA 01810 Tel: (978) 749-9949 (855) 747-7623 Fax: (978) 749-9961
More informationSignet Conductivity/Resistivity Electrodes
Signet 2819-2823 Conductivity/Resistivity Electrodes 2819 2820 2821 2822 2823 Sanitary Features Standard process connections - ¾ in. NPT Polypro - Tri-clamp 1-1½ in., 2 - Opt. ½ in. NPT 316 SS 316 SS or
More informationSIGNET Level Transmitter
ENGLISH 850.090 A8/0 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents. Description. Specifications. Panel
More informationSubmersion/Insertion ph/orp Sensors
Instruction Sheet PN 51A-396VP/rev.G January 2011 Models 396VP and 398VP Submersion/Insertion ph/orp Sensors For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/.
More informationConductivity Transmitters
Conductivity Transmitters CDTX-90 Series Starts at $ 576 Electronics Only Display in µs, ms, kω, MΩ, ppm (TDS) Simulate Function Programmable Temperature Compensation Relay Options Dual Output Option Allows
More information+GF+ SIGNET 5800CR Conductivity/Resistivity Monitor Instructions
GF SIGNET 5800CR Conductivity/Resistivity Monitor Instructions ENGLISH 35800CR.0901 D1/98 CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections.
More informationTUpH ph Sensors. Specifications. Storage. Sensor Installation. Electrode Preparation. 397 Performance and Physical Specifications
Instruction Manual 397 LIQ_MAN_ABR_397 November 2013 TUpH ph Sensors For additional information, please refer visit our website at www.rosemountanalytical.com. Specifications 397 Performance and Physical
More informationHMT330 Series Humidity and Temperature Transmitters for Industrial Applications
HMT330 Series Humidity and Temperature Transmitters for Industrial Applications The HMT330 transmitter family has the solution for demanding industrial humidity measurements. Six models for demanding industrial
More informationSignet Conductivity/Resistivity Electrodes
Signet 2819-2823 Conductivity/Resistivity Electrodes 2819 2820 2821 2822 2823 Sanitary Features Standard process connections - ¾ in. NPT Polypro - Tri-clamp 1-1½ in., 2 - Opt. ½ in. NPT 316 SS 316 SS or
More informationAnalyzer/Controller. ph/orp HART. Model 54e ph/orp ESSENTIAL INSTRUCTIONS WARNINGS RISK OF ELECTRICAL SHOCK
Instruction Sheet PN 51A-54epH/rev J October 2010 ph/orp HART Model 54e ph/orp Analyzer/Controller For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. ESSENTIAL
More informationTable of Contents. General Information. Document Sure-Aire Flow Monitoring System. User and Service Manual WARNING CAUTION
Document 476092 User and Service Manual Installation, Operation and Maintenance Manual Please read and save these instructions for future reference. Read carefully before attempting to assemble, install,
More informationOPERATION MANUAL MODEL TT-LP
TURTLE TOUGH OPERATION MANUAL MODEL TT-LP 2-Wire Isolated ph/orp Transmitter turtletough.com.au CONTENT INITIAL INSPECTION 3 INTRODUCTION 3 ASSEMBLY 4 PREPARATION 5 Connection Diagram of Turtle Tough Sensors
More informationSono-Trak Transit Time Clamp-on Ultrasonic Flow Meter
A trademark of Spirax Sarco, Inc. Sono-Trak Transit Time Clamp-on Ultrasonic Flow Meter Description Engineered for performance excellence, the Sono-Trak Transit Time ultrasonic meter combines non-invasive,
More information200CRS Single-Channel Conductivity/Resistivity Measurement System THORNTON. Leading Pure Water Analytics
THORNTON Leading Pure Water Analytics - NIST Traceable Calibration - Two Analog Outputs Standard - Three-Wire Temperature Measurement - Wide Range of Sensor and Cable Selections 200CRS Single-Channel Conductivity/Resistivity
More informationSignet 5800CR Conductivity/Resistivity Monitor
Signet 5800CR Conductivity/Resistivity Monitor *35800CR.0901* 35800CR.0901 Rev. H 3/06 English Contents 1. Power Connections 2. Compatible Sensor Wiring 3. 4 20 ma Current Output Connections 4. Relay Connections
More informationOperation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter
Operation Manual MODEL 2TX 2-wire Isolated ph/orp Transmitter 0 2TX CONTENTS INITIAL INSPECTION.....2 INTRODUCTION......2 ASSEMBLY...3 PREPARATION....4 CONNECTING THE ELECTRODE...4 CONNECTING THE TEMPERATURE
More informationDescription, continued
DL424/425 DirectLine Sensor Module for DL5000 Dissolved Oxygen Probes Overview 70-82-03-48 4/02 Page of 8 Specification The DirectLine DL424/425 Sensor Modules are part of a family of products developed
More informationOperation Manual TX10 1/8 DIN Microcomputer Based ph/orp Controller
Operation Manual TX10 1/8 DIN Microcomputer Based /ORP Controller 11751 Markon Drive Garden Grove, CA. 92841 U.S.A. pg. 1 CONTENTS GENERAL INTRODUCTION.....3 INITIAL INSPECTION......3 USING THE Sensorex
More informationPDW30 Point-To-Point Wireless Bridge
PDW30 Point-To-Point Wireless Bridge WIRELESS WIreless System FEATURES Wire Replacement in a Rugged, Industrial Housing Inputs and Outputs on Both Field Units Simple to Order, Configure, & Install 1 Mile
More informationJUMO ecotrans ph 03 Microprocessor transmitter/ switching device for ph/redox voltage and temperature
Page 1/7 JUMO ecotrans 03 Microprocessor transmitter/ switching device for /Redox voltage and temperature with a 2-line LCD for mounting on a 35 mm DIN rail Brief description Depending on the configuration,
More informationP2-04DAL-2 Analog Output. Removable Terminal Block Specifications
WARNING: To minimize the risk of potential safety problems, you should follow all applicable local and national codes that regulate the installation and operation of your equipment. These codes vary from
More informationProcess Analysis Systems. Stratos Pro A4 MSPH/MSPH. Chem Energy Pharm Food Water. Connection
Process Analysis Systems Chem Energy Pharm Food Water Stratos Pro A4 MSPH/MSPH Connection Connection of Memosens interface of 4-wire device with 2 Memosens sensors Model used: Stratos Pro A401N MSPH/MSPH
More informationCole-Parmer Instrument Company
Operating Instructions Cole-Parmer ph/orp 400 ph/orp Controller Cole-Parmer Instrument Company 625 East Bunker Court, Vernon Hills, Illinois 60061 Tel: 1-847-549-7600 or Toll-free: 1-800-323-4340 Fax:
More informationKOBOLD TDA Series Digital Temperature Transmitter
KOBOLD TDA Series Digital Temperature Transmitter TDA-15 Series User Instructions KOBOLD Instruments Inc. 1801 Parkway View Drive Pittsburgh, PA 15205 Phone (412) 788-2830 Fax (412)-788-4890 www.koboldusa.com
More informationph/orp 300 (Cat. No ) MICROCOMPUTER BASED OPERATION MANUAL ph /ORP/ Temperature CONTROLLER Phone: , Fax:
OPERATION MANUAL /ORP 300 (Cat. No. 0580212) MICROCOMPUTER BASED /ORP/ Temperature CONTROLLER 625 East Bunker CT, Vernon Hills, IL 60061-1844 Phone: 800-32-4340, Fax: 847-247-2929 E-mail: info@coleparmer.com,
More information1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882
1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882 This manual is intended for use in support of installation, commissioning and configuration of the 1/32-DIN Temperature
More information1/2 NPTM mm Nose cone fitting Ideal for clean liquids
Submersible Liquid Level Transmitter Series: PT-500 1.0 Hex 25.4 mm 7.0 177.8 mm Re-Usable cage fitting Ideal for dirty liquids 5.63 143 mm 5.33 136 mm 1/2 NPTM 3.5 88.9 mm 1.15 29.21 mm Nose cone fitting
More informationKOBOLD PDA Series Digital Pressure Transmitter
KOBOLD PDA Series Digital Pressure Transmitter PDA-153 Series User Instructions KOBOLD Instruments Inc. 1801 Parkway View Drive Pittsburgh, PA 15205 Phone (412) 788-2830 Fax (412)-788-4890 www.koboldusa.com
More informationType Electrically Actuated 3-way Ball Valves
GF Piping Systems Electrically Actuated 3-way Ball Valves Type 167-17 General Size: ⅜ 2 aterial: PVC, CPVC, PROGEF Standard PP, ABS, SYGEF Standard PVDF Seat: PTFE Seals: EPD, FP End Connection: Solvent
More informationHI 9828 Multiparameter Water Quality Meter...
HI 9828 Multiparameter Water Quality Meter... Backlit LCD can display up to 12 Measurements at a time Feature Highlights Display up to 12 parameters Graphic LCD with backlight Meter features IP67 and probe
More informationSeries ULT Ultrasonic Level Transmitter
Bulletin L-45 Series ULT Ultrasonic Level Transmitter Specifications - Installation and Operating Instructions 3 29/32 [99.21] 4 9/32 [108.74] 2X 1/2 NPT ø3 21/64 [84.53] 8 1/16 [204.79] Series ULT Ultrasonic
More informationINSTRUCTION MANUAL MODEL RTI-EC RESISTANCE TEMPERATURE INDICATOR TABLE OF CONTENTS 1. GENERAL DESCRIPTION PAGE 1
DEVAR Inc. 706 Bostwick Avenue, Bridgeport, CT 06605 TEL: 203-368-6751 FAX: 203-368-3747 1-800-566-6822 www.devarinc.com INSTRUCTION MANUAL MODEL RTI-EC RESISTANCE TEMPERATURE INDICATOR TABLE OF CONTENTS
More informationMultiparameter Waterproof Meters. Multiparameter measurement in compact, IP67 rated waterproof enclosures
Multiparameter Waterproof Meters Multiparameter measurement in compact, IP67 rated waterproof enclosures HI98195 Features HI98195 Multiparameter Waterproof Meter ph, ORP, EC, TDS, Resistivity, Salinity,
More informationCole-Parmer Instrument Company
Operating Instructions Cole-Parmer ph/orp 350 ph/orp Transmitter Cole-Parmer Instrument Company 625 East Bunker Court, Vernon Hills, Illinois 60061 Tel: 1-847-549-7600 or Toll-free: 1-800-323-4340 Fax:
More informationMODEL 8000MP LEVEL SENSOR
1 MODEL 8000MP LEVEL SENSOR INSTRUCTIONS FOR INSTALLATION, OPERATION & MAINTENANCE VISIT OUR WEBSITE SIGMACONTROLS.COM 2 SERIES 8000MP LEVEL SENSOR 1. DESCRIPTION The Model 8000MP Submersible Level Sensor
More informationMODEL: M5DY SEN TRONIC AG CURRENT LOOP SUPPLY. [2] POWER INPUT AC Power M: V AC (Operational voltage range V, Hz)
Super-mini Terminal Block Signal Conditioners M5-UNIT CURRENT LOOP SUPPLY Functions & Features Powering a 0 ma DC current loop Applicable to smart transmitters Isolation between the input and output Fast
More information955 ebrik INSTALLATION MANUAL. Series ebrik ABSOLUTE PROCESS CONTROL KNOW WHERE YOU ARE... REGARDLESS LINEAR DISPLACEMENT TRANSDUCERS
Series ebrik INSTALLATION MANUAL LINEAR DISPLACEMENT TRANSDUCERS ABSOLUTE PROCESS CONTROL KNOW WHERE YOU ARE... REGARDLESS Introduction The is an accurate programmable, auto-tuning, noncontact, linear
More information+GF+ SIGNET 5900 Salinity Monitor Instructions
+GF+ SIGNET 5900 Salinity Monitor Instructions ENGLISH 35900.0901 (B8/99), English CAUTION! Refer to instruction manual for more details. Remove power to unit before wiring input and output connections.
More information1. Installation. 2. Configuration - Operation. 3. Specifications. Safety
Safety 1. Installation 1.1 OP_ext: Services 1.2 Driver Card Installation 1.3 Sensor Types 1.4 Sensor Wiring 1.5 Controller Wiring 2. Configuration - Operation 2.1 Replaces ph Sensor 2.2 AS -Flex Series
More informationWall-/Pipe Mounted Universal Process Indicator
Data Sheet SS/_8 Wall-/Pipe Mounted Universal Process Indicator High visibility LED display the clearest view of your process status 0.1% measurement accuracy precise indication of process measurement
More informationLinear Actuators (235N)
Features: Force limiting adaptor Digital or analog models Patented Enerdrive Fail-Safe System* (optional) Auto stroke on power up (analog models) Simple manual override Direct or reverse acting (configurable)
More informationDWYER INSTRUMENTS, INC. Series AVUL Air Velocity Transmitter. Specifications - Installation and Operating Instructions.
Series AVUL Air Velocity Transmitter Specifications - Installation and Operating Instructions Bulletin P-AVUL 3-49/64 [95.71] 2-43/64 [67.92] 1/2 NPS 3-3/16 [80.81] 1-19/32 [40.59] 31/32 24.58 3-33/64
More informationContinuous level measurement Ultrasonic transducers
Siemens G 2013 Overview transducers use ultrasonic technology to measure level in a wide range of liquids and solids. Benefits Integral temperature compensation Low ringing effect reduces blanking distance
More informationContinuous level measurement Ultrasonic transducers
Siemens G 2012 Level Measurement Overview Echomax XPS/XCT transducers use ultrasonic technology to measure level in a wide range of liquids and solids. Benefits Integral temperature compensation Low ringing
More informationSubmersible Liquid Level Transmitter Series: PT-500
Submersible Liquid Level Transmitter Series: PT-500 The PT-500 is a rugged submersible pressure transmitter designed for tank liquid level measurement, water and wastewater applications, and well depth
More informationP2-04DA Analog Output. Removable Terminal Block Specifications
WARNING: To minimize the risk of potential safety problems, you should follow all applicable local and national codes that regulate the installation and operation of your equipment. These codes vary from
More informationC160 Wall-/Pipe Mounted Universal Process Indicator
Data sheet DS/ EN Rev. I Wall-/Pipe Mounted Universal Process Indicator reliable process indicator, wherever it s needed High visibility LED display the clearest view of your process status 0.1% measurement
More informationPLEASE READ INSTRUCTIONS CAREFULLY BEFORE INSTALLATION!
TUC2 / TUCH2 Model Series Installation Instructions PLEASE READ INSTRUCTIONS CAREFULLY BEFORE INSTALLATION! APPLICATION The TUC2 provides temperature space monitoring with a backlit LCD. The TUCH2 provides
More informationInform IT Single Pen Circular Chart Recorder
Data Sheet SS/_ Inform IT Single Pen Circular Chart Recorder Universal process input ma, mv, V, thermocouples and resistance Signal linearization full range of linearizers included as standard -digit indicator
More informationOperations Manual EcoSense ph100a
Operations Manual EcoSense ph100a Portable ph, mv and Temperature Instrument English CONTENTS GENERAL INTRODUCTION... 3 INITIAL INSPECTION... 3 THE INSTRUMENT... 3 BATTERY INSTALLATION... 3 Battery Disposal...
More informationMultiparameter Waterproof Meters. Multiparameter measurement in compact, IP67 rated waterproof enclosures
Multiparameter Waterproof Meters Multiparameter measurement in compact, IP67 rated waterproof enclosures HI98194 Features HI98194 Multiparameter Waterproof Meter ph, ORP, EC, TDS, Resistivity, Salinity,
More informationTRANSMISSOR DE PRESSÃO PIEZO RESISTIVO INTELIGENTE MICRO PROGRAMÁVEL SÉRIE PAS
TRANSMISSOR DE PRESSÃO PIEZO RESISTIVO INTELIGENTE MICRO PROGRAMÁVEL SÉRIE PAS Flexibility Data configuration with HART configurator Measuring of gauge and absolute pressure Reliability Continuous self-diagnostic
More informationECHOMAX XPS-10 ULTRASONIC TRANSDUCER
Datasheet ECHOMX XPS-10 ULTRSONIC TRNSDUCER Supplied by.com Call us on + (0)118 916 920 Email info@27able.com Overview transducers use ultrasonic technology to measure level in a wide range of liquids
More informationHumidity goes Digital
Humidity goes Digital Networkable Industrial Transmitter Humidity, Temperature and Dew Point or other Computed Parameter Optional metal enclosure Unique HygroClip technology for superior performance Durable,
More informationph600 2-wire ph/orp Transmitter
600 2-wire /ORP Transmitter Installation and Operating Instructions 900582-D - Sensolution 600 Isolated 2-wire transmitter instruction manualdoc 1 of 20 1 Contents 1 Contents 3 2 Introduction 5 3 Installation
More informationOPERATING INSTRUCTION MANUAL
OPERATING INSTRUCTION MANUAL MODEL P/R60 DIFFERENTIAL 6-WIRE ph & ORP PROBES N116-140 REV. 0 Aquametrix, by Water Analytics Inc. 100 School Street Andover, MA 01810 P/R60 Page 1 Tel: (978) 749-9949 Toll
More informationSafety. 2. Configuration - Operation 2.1 RTD Selection 2.2 Driver Test Header. 3. Diagnostics 3.1 ph Input 3.2 Temperature Input
Safety 1. Installation 1.1 PT:Services 1.2 erature Compensation of 1.3 Controller Services 1.4 Driver Card Installation 1.5 Sensor Part Numbers 1.6 Sensor Wiring 2. Configuration - Operation 2.1 Selection
More informationP2-04AD Analog Input. Removable Terminal Block Specifications
WARNING: To minimize the risk of potential safety problems, you should follow all applicable local and national codes that regulate the installation and operation of your equipment. These codes vary from
More informationX Year mfg
4to20 ma Output S L Loop PWR Loop2 PWR N/C S L DATA 7 8 9 6,*(7 S L Output 8 S L SIGNET 8900 MultiParameter Controller Supplemental Manual *8900.090SUP* 8900.090SUP Rev. A 2/04 (English). Description This
More informationHygroClip M22 Series. 2-wire loop powered humidity temperature transmitters INSTRUCTION MANUAL
HygroClip M22 Series 2-wire loop powered humidity temperature transmitters INSTRUCTION MANUAL 20060329 CONTENTS Overview... 3 Operation... 5 Power supply... 5 HygroClip S digital probe... 5 Temperature
More informationInstruction Sheet PN 51A-6081C/rev.A June Model 6081-C Wireless Contacting Conductivity Transmitter
Instruction Sheet PN 51A-6081C/rev.A June 2009 Model 6081-C Wireless Contacting Conductivity Transmitter ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures,
More informationBA/BS4S Temperature Sensor Installation & Operating Instructions
Product Overview and Identification The BAPI-Stat 4 room temperature sensor features a large format LCD and slider setpoint adjustment. Additional options include button override and communication jack.
More informationF-1000 DIGITAL PADDLEWHEEL FLOWMETER
F-1000 DIGITAL PADDLEWHEEL FLOWMETER F1000 with Machined In-line Fitting High accuracy digital paddlewheel technology with very low pressure drop. 3/8, 1/2, 3/4, 1, 1-1/2, and 2 female pipe threads. Flow
More informationph, ORP, and TDS ORP Continuous Monitor OPERATING INSTRUCTIONS 8. Warranty 9. Return of Goods
ENTER 8. Warranty Oakton Instruments warrants this product to be free from significant deviations in material and workmanship for a period of three year from the date of purchase (excludes electrode).
More informationABB MEASUREMENT & ANALYTICS DATA SHEET. TB84PH ph /ORP / pion transmitter
ABB MEASUREMENT & ANALYTICS DATA SHEET TB84PH ph /ORP / pion transmitter 2 TB 8 4PH P H /ORP / P I ON TR AN S MIT TE R DS/ TB8 4PH-E N RE V. D Measurement made easy Hazardous area rated transmitter with
More informationSX90 Process Controller
Local regulations may restrict the use of this product to below the conditions quoted. In the interests of development and improvement of the product, we reserve the right to change the specification without
More informationDINALOG A 144 x 36 A1400 Light-Strip Indicator
DINALOG A x 6 A00 Light-Strip Indicator -9-0-0 /.00 Front panel dimensions: x 6 mm Light-strip indicator with 7 high-contrast LEDs Red LED display color Digital display range for portrait format: 999 to
More informationModel HU-224/225. Technical Information TI.224/ Humidity Transduce r FOR ADDITIONAL INFORMATION SEE HU-224/225 DATA SHEET
Model /225 FOR ADDITIONAL INFORMATION SEE /225 DATA SHEET SPECIFICATIONS Accuracy*: ± 2% / ± 3% RH Range: 0-100% RH Hysteresis: ± 1% Supply Voltage:12-40 VDC 12-35 VAC (VDC output units only) Supply Current:
More informationThreaded DryLoc ph/orp Electrode
27742777 Threaded DryLoc ph/orp Electrode Features Durable DryLoc connector with gold plated contacts Description The 2774 2777 ph and ORP Electrodes feature a unique foulproof DryLoc connector with goldplated
More informationPortable pressure calibrator Model CPH7600
Calibration technology Portable pressure calibrator Model CPH7600 WIKA data sheet CT 17.01 Wally Box III Applications Calibration service companies and service industry Measurement and control laboratories
More informationLFC Liquid Flow Controller
LFC Liquid Flow Controller Type 879 can be combined with High dynamic control through fast flow measurement Applicable for liquid dosing up to 600 ml/min (6 l/h) No moving parts in medium Fieldbus optional
More information100mm Process Indicator Recorder
Data Sheet SS/_3 100mm Process Indicator Recorder One or two pen continuous line 100mm recorder universal input for thermocouple, RTD, mv, ma and V Clear, 5-digit LED display high visibility, wide viewing
More information60W Power over Ethernet Waterproof Adapter PoE IEEE BT Single Port Injector for Outdoor Application
WWW.PHIHONG.COM 60W Power over Ethernet Waterproof Adapter PoE IEEE BT Single Port Injector for Outdoor Application Features Compliant with the IEEE802.3bt Standard Non-Vented Case with Mounting Bracket
More informationGeneral Specifications. P2-08THM Analog Input
General Specifications Operating Temperature Storage Temperature Humidity Environmental Air Vibration Shock Field to Logic Side Isolation Heat Dissipation Enclosure Type Agency Approvals Module Keying
More informationph/orp 3/4-inch Combination ph and ORP Sensor Kits Features and Benefits DATA SHEET Versatile Mounting Styles Low Price High Performance
DATA SHEET 3/4-inch Combination ph and ORP Sensor Kits ph/orp Use the Digital Gateway to make any Hach analog combination ph or ORP sensor compatible with the Hach sc1000 Controller. Features and Benefits
More informationGeneral Specifications. P2-8AD4DA-2 Analog Input/Output
General Specifications Operating Temperature Storage Temperature Humidity Environmental Air Vibration Shock Field to Logic Side Isolation Insulation Resistance Heat Dissipation 1.95W º to 6ºC (32º to 14ºF)
More informationportable waterproof multiparameter meter
water resistant battery powered interchangeable electrode DIN connection replaceable electrode portable waterproof multiparameter meter Professional testing and analysis out in the field is supported by
More informationDigital O.R.P. Transmitter Type 8206
Advantages / Benefits Design Compact O.R.P. Transmitter Type 8206 The O.R.P. transmitter compactly combines an O.R.P. sensor and an electronic with display in splash-proof plastic NEMA 4 (IP 65) enclosure.
More informationAdapter Kit - QSI Electronics to G-Series Meter Installation Instructions
Adapter Kit - QSI Electronics to G-Series Meter Installation Instructions These installation instructions cover the adapter kit installation of the QSI1, QSI2 and QSI3 versions of communications electronics
More information