Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators
|
|
- Jared Hart
- 5 years ago
- Views:
Transcription
1 Instruction Bulletin MED /2013 Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators Retain for future use. Introduction This bulletin describes the procedures and components necessary to replace a Square D brand, Iso-Gard line isolation monitor (LIM) with an Iso-Gard Series 6 LIM manufactured by Schneider Electric. It also details the components required to connect legacy remote indicators to the Iso-Gard Series 6 LIM. Legacy Iso-Gard LIMs Iso-Gard Series A, B, or C (Series C shown) Iso-Gard Series D Safety Precautions DANGER HAZARD OF ELECTRIC SHOCK, EXPLOSION, OR ARC FLASH Apply appropriate personal protective equipment (PPE) and follow safe electrical work practices. See NFPA 70E or CSA Z462. This equipment must be installed and serviced only by qualified electrical personnel. Turn off all power supplying this equipment before working on or inside equipment. Always use a properly rated voltage sensing device to confirm power is off. Replace all devices, doors, and covers before turning on power to this equipment. Failure to follow these instructions will result in death or serious injury. CAUTION HAZARD OF UNAUTHORIZED MODIFICATIONS Do not make modifications to this equipment. Only use accessories supplied by or recommended by Schneider Electric with this equipment. Failure to follow these instructions can result in personal injury or equipment damage.
2 Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators MED Replacing the LIM 01/2013 Replacing the LIM Use the following instructions to replace an Iso-Gard Series A, B, C, or D LIM with an Iso-Gard Series 6 LIM. Removing a Series A, B, C, or D LIM 1. Review and follow the safety precautions on page Turn off all power supplying this equipment. 3. Unplug the connector from the LIM. 4. Remove the #8-32 hex nuts securing the LIM mounting bracket to the backplate, and remove the LIM and bracket as an assembly. Retain the hex nuts. 5. Disconnect all wires connected to connector plate and label them (L1, L2, ground(s), and remote accessory). 6. Remove the #8-32 hex nuts securing the LIM connector plate to the backplate and remove the connector plate. Retain the hex nuts. Series A, B, or C Series D
3 MED Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators 01/2013 Replacing the LIM Installing a Series 6 LIM Using the LIM Bracket NOTE: The torque range for all mounting hardware is lb-in ( N m). 1. Review and follow the safety precautions on page Turn off all power supplying this equipment. 3. Set the LIM mounting bracket to the correct height (page 4). 4. Attach the Series 6 LIM to the mounting bracket using the provided #4-40 x 0.5 in. hardware. 5. Attach the LIM assembly to the backplate using the retained #8-32 hex nuts. 6. Attach the LIM terminal connector to the backplate using the retained #8-32 hex nuts. 7. Drill and tap two #6-32 holes as required for mounting the auxiliary transformer (4800XMIA); order separately when remote accessories are connected to panel. Collect and remove all metal shavings. 8. Make connections from auxiliary transformer to the IG6C connector plate. 9. Detach the wire marked 12VDC Com from the 15-pin wiring harness. Insulate the uninsulated wire end using shrink tubing and secure it to the harness with a wire tie. 10. Plug in the wiring harnesses from the connector plate to the LIM. NOTE: For proper functionality, both relay settings must remain at the factory setting (N/C-T Mode). See the SETTING/RELAY menu in instruction bulletin MED101101xx, Iso-Gard Line Isolation Monitor
4 Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators MED Replacing the LIM 01/2013 Adjusting the Variable-Height Mounting Bracket The mounting bracket for the Iso-Gard Series 6 LIM (catalog no. IG6BKTVM) can be set to four different heights. (4) #6-32 x 0.25 in. pan head screws (130) 3.75 (95) 3.25 (83) 3.0 (76) Dimensions in Inches (mm) 4
5 MED Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators 01/2013 Replacing the LIM Installing a Series 6 LIM Without Using the LIM Bracket Use this procedure for all installations where the interior-to-trim distance is other than the heights shown on page 4. This procedure can also be used as an optional mounting procedure in existing medical panels. NOTE: Special brackets are available on request; delivery time is 4 6 weeks. NOTE: The torque range for all mounting hardware is lb-in ( N m). 1. Review and follow the safety precautions on page Turn off all power supplying this equipment. 3. Drill four in. (4 mm) diameter holes in the dead front panel. Collect and remove all metal shavings. 4. Attach the LIM to the dead front panel using the four #4-40 screws and hex nuts provided with the LIM. 5. Attach the LIM terminal connector to the backplate using the two #8 self-tapping screws provided with the LIM. 6. Drill and tap two #6-32 holes as required for mounting the auxiliary transformer (4800XMIA); order separately when remote accessories are connected to panel. Collect and remove all metal shavings. 7. Make connections from auxiliary transformer to the connector plate. 8. Detach the wire marked 12VDC Com from the 15-pin wiring harness. Insulate the uninsulated wire end using shrink tubing and secure it to the harness with a wire tie. 9. Plug in the wiring harnesses from the connector plate to the LIM. NOTE: For proper functionality, both relay settings must remain at the factory setting (N/C-T Mode). See the SETTING/RELAY menu in instruction bulletin MED101101xx, Iso-Gard Line Isolation Monitor (83) 6.50 (165) C L Existing LIM opening 2.00 (51) 4.00 C L (102) 3 Dead front panel 4 Dimensions in Inches (mm)
6 Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators MED Connecting Legacy Remote Indicators 01/2013 Connecting Legacy Remote Indicators When a legacy Iso-Gard LIM is upgraded to a Series 6 LIM, additional components may be needed to connect existing remote indicators. Each component can be ordered separately or as part of a kit. See Schneider Electric catalog #8501CT1301 for ordering details. Component Catalog Number Series 6 LIM IG6M Connector plate and cables IG6C Variable-height mounting bracket IG6BKTVM Transformer 4800XMIA Connector box and wiring harness 4800RA1WMAHA NOTE: Wiring diagrams for connecting remote indicators to the Iso-Gard Series 6 LIM are shown on pages 13 and 14. 6
7 MED Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators 01/2013 Connecting Legacy Remote Indicators If you have one of these remote indicators You ll need to add these components ORICA IG6M IG6BKTVM IA1C IG6C 4800XMIA RA1 PM (panel-mounted) IG6M IG6BKTVM RA1 WM (wall-mounted) IG6C 4800RA1WMAHA 7
8 Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators MED Connecting Legacy Remote Indicators 01/2013 If you have one of these remote indicators and your application requires an ammeter per the NEC a You ll need to replace the remote indicator with one of these Plus, add these components IG6M ORICA5C RA1 PM (panel-mounted) IG6C M5IAI IG6BKTVM RA1 WM (wall-mounted) M5IAI RA1WMAHA a NEC Section 517 (B) Line Isolation Monitors (3) Ampmeter Exception: The line isolation monitor shall be permitted to be a composite unit, with sensing section cabled to a separate display panel section on which the alarm or test functions are located. Information Note: It is desirable to locate the ammeter so that it is conspicuously visible to person(s) in the anesthetizing location. 8
9 MED Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators 01/2013 Connecting Legacy Remote Indicators If you have this remote indicator and your application does not require an ammeter per the NEC a You ll need to replace the remote indicator with this one b ORICA5C IG2000PPM (includes IG2000P and panel-mount bracket) M5IAI IG2000PST0614 M5IAI50 IG2000PG4 (designed for mounting in a four-gang electrical box) a NEC Section 517 (B) Line Isolation Monitors (3) Ampmeter Exception: The line isolation monitor shall be permitted to be a composite unit, with sensing section cabled to a separate display panel section on which the alarm or test functions are located. Information Note: It is desirable to locate the ammeter so that it is conspicuously visible to person(s) in the anesthetizing location. b These remote indicators can be connected directly to the Iso-Gard Series 6 LIM. See the instructions included with the device. 9
10 Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators MED Connecting Legacy Remote Indicators 01/2013 Installing a 4800RA1WMA Converter Box Use the 4800RA1WMA converter box to connect an Iso-Gard Series 6 LIM to an RA1 panel- or wall-mounted remote indicator. WARNING HAZARD OF EQUIPMENT SHUTDOWN To prevent an inadvertent shutdown of the Iso-Gard Series 6 LIM, do not energize more than four RA1 remote indicators simultaneously with the RA1WMA converter box. Failure to follow these instructions can result in death or serious injury. Mounting Mount the converter box in a suitable location using the four screws provided. Adding a Voltage Jumper 1. Review and follow the safety precautions on page Turn off all power supplying this equipment. 3. Remove and retain the four screws securing the cover to the converter box. 4. Remove the cover and locate a four-pin, female connector labeled JP1. JP1 120 V jumper 5. If the system voltage is 120 V, connect the voltage jumper labeled 120. or If the system voltage is 208 V, connect the voltage jumper labeled Replace the cover and secure with the four screws removed in step 1. 10
11 MED Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators 01/2013 Connecting Legacy Remote Indicators Connecting to an Iso-Gard IG6C Connector Plate The connector plate for an Iso-Gard Series 6 LIM is connected to the RA1WMA converter box with the RA1HA wiring harness, sold separately. IG6C Connector Plate 1. Remove the converter box cover. 2. Connect the 1 x 15-pin female connector from the RA1HA wiring harness to the 1 x 15-pin male plug inside the converter box. 3. Route the wiring for the female connector through the slot on the converter box. 4. Replace the converter box cover. RA1WMA convertor box RA1HA wiring harness To IG6C connector plate 11
12 Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators MED Connecting Legacy Remote Indicators 01/2013 Connecting Other Devices Annunciator Panels (Nurse Stations) These devices can be fed from multiple LIMs. For each LIM replaced, connect according to the wiring diagrams on pages 13 and 14. IA4CI IA8CI Power Modules To connect these devices, refer to the wiring diagrams on pages 13 and 14. XRIAI Supervisory Stations To connect these devices, refer to the wiring diagrams on pages 13 and 14. 8CIIAI 12
13 MED Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators 01/2013 Wiring Diagrams Wiring Diagrams One Remote Indicator 4800XMIA NOTE: Test is an optional connection that is only required with an IA1CPTT device. Multiple Remote Indicators 4800XMIA NOTE: When there is more than one remote indicator, connect them in parallel. This includes indicators mounted in annunciator panels or power modules. 13
14 Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators MED Wiring Diagrams 01/2013 When Power Source Already Exists in Panel 4800XMIA Power Source... Power Modules and Supervisory Station (Used with Control Panels) 4800XMIA 8-station interlock switch with Off position ( )2 ( )3 A1 A2 A3 B1 B2 B3 C1 C2 C3 D1 D2 D3 E1 E2 E3 F1 F2 F3 G1 G2 G3 H1 H2 H3 14
15 MED Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators 01/2013 Wiring Diagrams 15
16 Upgrading Legacy Iso-Gard Line Isolation Monitors and Remote Indicators MED Instruction Bulletin 01/2013 Schneider Electric USA, Inc S. Roselle Road Palatine, IL USA Electrical equipment should be installed, operated, serviced, and maintained only by qualified personnel. No responsibility is assumed by Schneider Electric for any consequences arising out of the use of this material. All trademarks are owned by Schneider Electric Industries SAS or its affiliated companies. 16
Instruction Bulletin Trim and Rail Extension Kit (6-, 12-, and 18-inch) Installation onto an NF Panelboard Class 1670
Instruction Bulletin Trim and Rail Extension Kit (6-, 12-, and 18-inch) Installation onto an NF Panelboard Class 1670 80043-763-01 06/2008 Peru, IN, USA Retain for future use. Introduction This bulletin
More informationAltivar 61/71 Adjustable Speed Drives Heatsink Fan Kits VZ3V1212 and VZ3V1216
Altivar 61/71 Adjustable Speed Drives Heatsink Fan Kits VZ3V1212 and VZ3V1216 Instruction Bulletin 30072-452-48 Retain for future use. 30072-452-48 Altivar 61/71 Heatsink Fan Kits VZ3V1212 and VZ3V1216
More informationPowerLogic High Density Metering System 1-Meter Enclosure
PowerLogic High Density Metering System 1-Meter Enclosure Installation Guide 63230-508-211A1 Safety information PowerLogic High Density Metering System 1-Meter Enclosure Important information Read these
More informationProcedure to Install an IO Expansion Cage Assembly in a Maxum II Modular Oven Analyzer
Procedure to Install an IO Expansion Cage Assembly in a Maxum II Modular Oven Analyzer Difficulty Level: Medium Estimated time to execute: 1 Hour Revision History Issue Date Reason 001 5/31/2016 Initial
More informationClipsal Bus Couplers. Two Channel (SLC5102BCLEDL) and Four Channel (SLC5104BCL) for Use with C-Bus Wired Systems
Clipsal Bus Couplers Two Channel (SLC5102BCLEDL) and Four Channel (SLC5104BCL) for Use with C-Bus Wired Systems Instruction Bulletin Retain for future use. Clipsal Bus Couplers 63249-420-236A2 Instruction
More informationINSTALLATION INSTRUCTIONS
TM REPLACEMENT PARTS FOR USE WITH CX SERIES PANELS INSTALLATION INSTRUCTIONS Hubbell Control Solutions 9601 Dessau Road Building One Suite 100 Austin, TX 78754 Toll Free: 888-698-3242 Fax: 512-450-1215
More informationInstruction Manual: IM E
Innovative Technology TVSS Products The I.T. Equalizer Installation Instruction Manual: IM01005011E Rev. 2 May 2009 Contents Page 1 Introduction... 1 1.1 Scope... 1 1.2 I.T. Equalizer Catalog Numbering
More informationInstruction Bulletin. Fiber Optic Switch Packs. SLSPSP101 and SLSPSP102 for use with Luminaires INTRODUCTION CONTENTS OF THE BOX
Instruction Bulletin 63249-500-04A3 07/2010 Fiber Optic Switch Packs SLSPSP101 and SLSPSP102 for use with Luminaires INTRODUCTION SLSPSP101 Single Input Switch Pack SLSPSP102 Dual Input Switch Pack The
More informationConext CL-60 Inverter Firmware Upgrade Process
Conext CL-60 Inverter Firmware Upgrade Process http://solar.schneider-electric.com 976-0380-01-01/B August 2017 Application Note EXCLUSION FOR DOCUMENTATION UNLESS SPECIFICALLY AGREED TO IN WRITING, SELLER
More informationMultiTech Router Commissioning for Com X 200/210/510
Knowledge Base 7EN42-0166 05/2017 MultiTech Router Commissioning for Com X 200/210/510 This document describes installation recommendations and how to commission a MultiTech rcell Series router for use
More informationInstruction Bulletin
Instruction Bulletin 63249-401-200/A4 03/2005 POWERLINK TM G3 Controller NF1000G3 for use with POWERLINK G3 Systems Retain for future use. NOTICE Read these instructions carefully and look at the equipment
More informationI-Line Panelboards. Innovative design, smart bus technology, and simple installation
I-Line Panelboards Innovative design, smart bus technology, and simple installation TM Smart: designed for efficient installation Simple: Efficient Installation Process I-Line Breakers Main Lugs: Fixed-mounted
More informationPhase Loss Protection Upgrade. Phase Loss Protection Upgrade. In this bulletin:
Phase Loss Protection Upgrade In this bulletin: Introduction... 2 Purpose... 2 General... 2 Applicability... 2 HD3070 Phase Loss Protection Upgrade Kit Parts... 2 Preparation... 4 Install the Phase Loss
More informationSymmetra PX 160. Maintenance Bypass Enclosure. Installation 05/
Symmetra PX 160 Maintenance Bypass Enclosure Installation 05/2014 www.schneider-electric.com Legal Information The Schneider Electric brand and any registered trademarks of Schneider Electric Industries
More informationWiser EER260LLCR Load Control Control de carga EER260LLCR Contrôle de charge EER260LLCR
Wiser EER260LLCR Load Control Control de carga EER260LLCR Contrôle de charge EER260LLCR Installer s Guide / Guía de instalación / Guide de l'installateur S1A90300 Rev. 03, 08/2012 Retain for future use.
More informationModel 2460-KIT. Screw Terminal Connector Kit. Description / September 2014 *P * 1
Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.keithley.com Model 2460-KIT Screw Terminal Connector Kit Description The Model 2460-KIT Screw Terminal Connector Kit
More informationInstalling Powerlink NF3500G4 Controllers
Installing Powerlink NF3500G4 Controllers for Use with Powerlink Systems RETAIN FOR FUTURE USE Installing Powerlink NF3500G4 Controllers 63249-420-374A3 Instruction Bulletin 09/2015 2 2015 Schneider Electric.
More informationSeries 3700 Screw Terminal Assemblies Installation Instructions
Keithley Instruments, Inc. 28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY www.keithley.com Series 3700 Screw Terminal Assemblies Installation Instructions Introduction This document contains handling
More informationInstruction Bulletin. Sepam Series 20 & 40 Relays Quick Start Guide
Instruction Bulletin 63230-216-239B1 LaVergne, TN, USA Sepam Series 20 & 40 Relays Quick Start Guide Retain for future use. Introduction Sepam Series 20 Relays Sepam Series 40 Relays Storage Commissioning
More informationGalaxy VM. System Bypass Cabinet 640 kva Installation GVMSBC640KHEL 06/2016.
Galaxy VM System Bypass Cabinet 640 kva Installation GVMSBC640KHEL 06/2016 www.schneider-electric.com Legal Information The Schneider Electric brand and any registered trademarks of Schneider Electric
More informationInstallation Instructions
Installation Instructions Variable Frequency Drive (VFD) 7 ½ - 25 Ton Units with 2 Wire Control WARNING UNINTENDED EQUIPMENT OPERATION Modifying or changing parame eters whose function is not described
More informationPowerLogic ION7550 RTU option
70052-0213-02 PRODUCT OPTION 01/2011 PowerLogic ION7550 RTU option The PowerLogic ION7550 Remote Terminal Unit (RTU) option is designed for data acquisition from WAGES (water, air, gas, electricity, steam)
More informationInstallation Guide SMT 2200/3000 VA Input/Output Hardwire Kit
Installation Guide SMT 2200/3000 VA Input/Output Hardwire Kit Important Safety Messages SAVE THESE INSTRUCTIONS - This section contains important instructions that should be followed during installation
More informationINSTALLATION INSTRUCTIONS STZ-R SERIES (ENGLISH)
Surge-Trap STZ-R series Surge Protective Device (SPD) Danger Hazard of Electrical Shock, Explosion or Arc Flash Only qualified licensed electricians should install or service SPDs Verify that all power
More informationPush Buttons and Operator Interface
Specifier s Guide Type K Rotary Cam Switch Catalog 9003CT1301 2013 Class 9003 CONTENTS DescriptionPage General Information...............................................................3 Type K Rotary
More informationHubbell Surge Protective DeviceS
Hubbell Surge Protective DeviceS Hubbell Modular Panel Products Surge Protective Devices (SPD) The Hubbell Modular Panel Product Surge Protective Devices (SPD) are designed to provide protection at the
More informationPACSystems* RX3i IC695ACC650
September 2012 PACSystems* RX3i IC695ACC650 Series 90-70 to RX3i Conversion Rack The conversion rack provides a solution for converting an existing Series 90-70 system to PACSystems RX3i, minimizing outage
More informationElectrical network protection Sepam series 20 Sepam series 40 Quick start
03146790FE+01+NP00000000 Electrical network protection Sepam series 20 Sepam series 40 Quick start PE50465 Storage Sepam may be stored in its original packaging in a closed sheltered location: b ambient
More informationType VR Electrically Operated Ground and Test Device
Type VR Electrically Operated Ground and Test Device For Use with Masterclad Switchgear 4.76 kv 15 kv, 1200, 2000, 3000, 4000 A Up to 63 ka Short Circuit Rating Class 6055 Instruction Bulletin 6055-37
More informationInstalling a Power over Ethernet injector
Installing a Power over Ethernet injector AlphaEclipse StreetSmart and RoadStar signs The instructions in this document explain how to install/replace a Power over Ethernet (PoE) injector in a StreetSmart
More informationInstallation and Operation Manual
Installation and Operation Manual APC 2X Series Hardwire Products PMP2X-A PMF2X-A PMG2X-A PMH2X-A APC hardwired products are designed to provide specification-grade performance at the service entrance,
More informationPS/IO Circuit Board Retrofit
S&C 6800 Series Automatic Switch Controls PS/IO Circuit Board Retrofit Table of Contents Section Page Introduction Qualified Persons.... 2 Read this Instruction Sheet.... 2 Retain this Instruction Sheet....
More informationINSTALLATION INSTRUCTIONS
Wired Remote Controller 7 Day Programmable Ductless Systems KSACN0401AAA (High Wall Models) KSACN0501AAA (Ducted/Cassette Models) INSTALLATION INSTRUCTIONS NOTE: Read the entire instruction manual before
More informationQuick Start Guide TS A
Quick Start Guide TS 930 125-630A DANGER HAZARD OF ELECTRICAL SHOCK, EXPLOSION, OR ARC FLASH Read and understand this quick start guide before installing and operating the transfer switch The installer
More informationMLHG6-AU. Iso-Gard Line Isolation Monitor (LIM) Setup Guide DANGER. Components
Iso-Gard Line Isolation Monitor (LIM) MLHG6-AU Setup Guide This document is intended as a guide for the basic physical setup of the Line Isolation Monitor (LIM). This document includes wiring diagrams
More informationInstruction Sheet WPC 2000 Option 1
Instruction Sheet WPC 2000 Option 1 This document shows you how to install, wire, program, monitor, and troubleshoot WPC 2000 Option 1. Option 1 is compatible with WPC 2000 systems running WPC 2000 version
More informationValidator Update Instructions for Rowe BC1200 $1 - $20
Validator Update Instructions for Rowe BC1200 $1 - $20 Kit Overview The purpose of the kit is to replace the Rowe BA50 transport and stacker with a 120 volt Mars validator with a compact mask. The kit
More informationKBMG MULTI-SPEED BOARD
TM INSTALLATION AND OPERATION MANUAL KBMG MULTI-SPEED BOARD KB Part No. 8833 Multi-Speed Board for KBMG Series Regenerative Drive Pending! See Safety Warning on Page 1 The information contained in this
More information4100/ VDC Converter Installation Instructions
4100/4120-0156 8 VDC Converter Installation Instructions Introduction This publication describes the installation procedure for the 8 VDC Converter. Related Documentation Field Wiring Diagram for 4100
More informationModel 7705 Control Module
www.keithley.com Model 7705 Control Module User s Guide PA-696 Rev. D / October 2006 A G R E A T E R M E A S U R E O F C O N F I D E N C E Safety Precautions The following safety precautions should be
More informationSquare D Clipsal Pascal Automation Controller
Square D Clipsal Pascal Automation Controller SLC5500PACA For Use with Wired C-Bus Networks Instruction Bulletin Retain for future use. Square D Clipsal Pascal Automation Controller 63249-420-258A2 Instruction
More informationInstruction Sheet Board Style Low Water Cutoff
Instruction Sheet Board Style Low Water Cutoff 102-305 SUPERSEDES: REVISION E DATED December 12, 2007 #5401173-REV F PLANT ID 001-3902 US Patents 6,904,800, 7,243,540, and 7,317,993 Other Patents Pending
More informationInstallation Instructions
CRDMNDMD001A00 CRDMNDMD002A00 Energy Demand Load Controller Rooftop Unit Accessory Installation Instructions NOTE: C13331 Read the entire instruction manual before starting the installation, and follow
More informationSAVE THESE INSTRUCTIONS
READ AND FOLLOW ALL SAFETY INSTRUCTIONS! SAVE THESE INSTRUCTIONS AND DELIVER TO OWNER AFTER INSTALLATION IMPORTANT SAFEGUARDS! When using electrical equipment, basic safety precautions should always be
More informationInstallation Instructions
CRDMNDGW001A00 CRDMNDGW002A00 Energy Demand Gateway Rooftop Unit Accessory Installation Instructions NOTE: C13331 Read the entire instruction manual before starting the installation, and follow instructions
More information2260B-RMK-Series Rack Mount Kit
Keithley Instruments, Inc. 28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY http://www.keithley.com Assembly and Mounting Instructions Introduction The 2260B-RMK-Series Rack Mount Kit is suited for
More informationINSTALL INSTRUCTIONS RSXMR-9886-P-48
INSTALL INSTRUCTIONS RSXMR-9886-P-48 Installation Instructions RSXMR-9886-P-48 (CEQ. 20194) INSTALL INSTRUCTIONS RSXMR-9886-P-48 Contents Section Description Page 1 Copyright 2 1.1 Disclaimer 2 1.2 Warnings
More informationHydroGuard 4 to 20 ma Output Module. For use with HydroGuard Water Quality Monitor and Controller
HydroGuard 4 to 20 ma Output Module For use with HydroGuard Water Quality Monitor and Controller Technician's Manual Installation, Operation, and Maintenance Guide Chapter 1: Preface... 3 1.1 Intended
More informationCOOPER POWER SERIES. RS-485 digital communications accessory board installation and operation instructions. Voltage Regulators MN225074EN
Voltage Regulators MN225074EN Effective March 2017 Supersedes January 2012 (S225-40-7) RS-485 digital communications accessory board installation and operation instructions COOPER POWER SERIES DISCLAIMER
More informationINSTRUCTION MANUAL RA 19-5B RACK ADAPTER HOT SWAP RACK ADAPTER FOR HSF SERIES POWER SUPPLIES
INSTRUCTION MANUAL RA 19-5B RACK ADAPTER HOT SWAP RACK ADAPTER FOR HSF SERIES POWER SUPPLIES KEPCO INC. An ISO 9001 Company. MODEL RA 19-5B RACK ADAPTER ORDER NO. REV. NO. IMPORTANT NOTES: 1) This manual
More informationResolver to Digital Expansion Board
Resolver to Digital Expansion Board Catalog No. EXB009A01 Installation and Operating Manual 6/98 MN1313 Table of Contents Section 1 General Information............................. 1-1 Introduction....................................
More informationHUBBELL SPIKESHIELD Wired-In Branch Panels Surge Protective Device (SPD)
HUBBELL SPIKESHIELD Wired-In Branch Panels Surge Protective Device (SPD) Instruction Bulletin 8222-0513B Retain for future use. SPIKESHIELD Wired-In Branch Panels 8222-0513B 05/2012 Table of Contents Precautions...
More informationThis 4200-RM Rack Mount Kit is for installation in 4200-CAB series cabinets only.
Keithley Instruments, Inc. 28775 Aurora Road Cleveland, Ohio 44139 (440) 248-0400 Fax: (440) 248-6168 www.keithley.com Model 4200-RM Rack Mount Kit Packing List Introduction NOTE This 4200-RM Rack Mount
More informationInstruction Manual. Multilin TM 8 Series Retrofit Kit * A1* GE Grid Solutions LISTED
GE Grid Solutions Multilin TM 8 Series Retrofit Kit Instruction Manual 8 Series version: 1.6x GE publication code: 1601-0701-A1 (GEK-130906) LISTED *1601-0701-A1* Copyright 2016 GE Multilin Inc. All rights
More informationTorque Series LCD Remote Panel Installation/Operation Manual Model: TQ-DSP-12/24
Torque Series LCD Remote Panel Installation/Operation Manual Model: TQ-DSP-12/24 Section Page Introduction 1 Materials Provided 1 I) Safety Instructions 1 A) Inverter Safety Instructions 1 B) Battery Safety
More informationISOLATION RELAY RACK ISOLATION KIT 19 RACK MOUNT RELAY RACK ISOLATION KIT, SLOTTED UNEQUAL FLANGE RACK ISOLATION KIT 23 RACK MOUNT
ISOLATION Info: 9070 = Isolation RELAY RACK ISOLATION KIT 19 RACK MOUNT 9070-1019 19" Rack Mount 23 RACK MOUNT 9070-1023 23" Rack Mount RELAY RACK ISOLATION (KIT) RELAY RACK ISOLATION KIT, SLOTTED 19 RACK
More informationQuick Start Guide TS 870 ATS
Quick Start Guide TS 870 ATS 100A - 800A (with TSC 900 Controller) DANGER HAZARD OF ELECTRICAL SHOCK, EXPLOSION, OR ARC FLASH Read and understand this quick start guide before installing and operating
More informationModel 2657A-LIM-3 LO Interconnect Module
Keithley Instruments, Inc. 28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY http://www.keithley.com Model 2657A-LIM-3 LO Interconnect Module User's Guide Description The Model 2657A-LIM-3 LO Interconnect
More informationE1135C PDU and Pod Upgrade Procedure
E4030-90010 Rev. B 12/2003 In this Document... Tools Needed, 2 Contents of the Upgrade Kits, 2 Installation Procedures, 4 Verifying the Power Option of the New PDU, 4 Removing the PDU from the Support
More informationhp uninterruptible power system r12000 xr models installation instructions
hp uninterruptible power system r000 xr models installation instructions Overview These instructions show how to install an uninterruptible power system (UPS). For detailed information about the UPS, refer
More informationSERVICE BULLETIN DEPARTMENT OF COMPLIANCE VEHICLE SAFETY AND RECALL MANAGEMENT BUILDING N MAIN ST MIDDLEBURY, INDIANA
SERVICE BULLETIN DEPARTMENT OF COMPLIANCE VEHICLE SAFETY AND RECALL MANAGEMENT BUILDING 11 423 N MAIN ST MIDDLEBURY, INDIANA 46540-9218 Technical Service Bulletin: 225-0754 O Integrity O Safety O Quality
More informationACCESS 9340/9360 Meter Input/Output Module
Installation Manual PMIM-IOMOD-0208 ACCESS 9340/9360 Meter Input/Output Module 9340-60-I/O2222 and 9340-60-I/O26 HAZARD CATEGORIES AND SPECIAL SYMBOLS Read these instructions carefully and look at the
More informationMODEL SST3-MV INSTALLATION INSTRUCTIONS FOR FEDERAL MODEL SST3-MV SELECTABLE SIRENTONE-MULTI VOLT
MODEL INSTALLATION INSTRUCTIONS FOR FEDERAL MODEL SELECTABLE SIRENTONE-MULTI VOLT INSTALLATION INSTRUCTIONS FOR FEDERAL MODEL SELECTABLE SIRENTONE-MULTI VOLT A. GENERAL. The Federal Model is a solid-state
More informationInstallation and Operation Manual
Installation and Operation Manual APC Panel Mount Modular SPDs PMP3S-A PMP3DS-A PMP4S-A PMP4DS-A PMF3S-A PMF3DS-A PMF4S-A PMF4DS-A PMG3S-A PMG3DS-A PMG4S-A PMG4DS-A PML3S-A PML3DS-A PML4S-A PML4DS-A The
More informationOPERATING AND SERVICE MANUAL. Universal Interface Device 47
OPERATING AND SERVICE MANUAL Universal Interface Device 47 MAGNA-POWER ELECTRONICS, INC. 39 ROYAL ROAD, FLEMINGTON, NJ 08822 May 24, 2012 SAFETY NOTICE Universal Interface Device 47 (UID46) connects
More informationModel 2380 Rack-Mount Kit
Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.tek.com/keithley Model 2380 Rack-Mount Kit Installation Instructions Introduction The Model 2380 Fixed Rack-Mount
More informationVM 6.1. Capacitive Air Gap Measuring Chains User s Manual P/N: M7A-110
VM 6.1 Capacitive Air Gap Measuring Chains User s Manual P/N: 9428-25M7A-110 Safety Information The following manual contains information and warnings. They must be followed in order to keep the instrument
More informationPD PA AP Procedure to Install a SSSI I.S. Power Barrier in a Maxum II Analyzer
Overview 1 Detailed Description 2 General Safety Considerations 3 PD PA AP Procedure to Install a SSSI I.S. Power Barrier in a Maxum II Analyzer Installation Instructions 4 Commissioning and Hardware Installation
More informationRMB Peripheral Units Installation Guide
RMB Peripheral Units Installation Guide Part Number 65-000101 2011 by Kentrox, Inc. All rights reserved. Copyright 2011 by Kentrox, Inc. All Rights Reserved. The material discussed in this publication
More informationOTPC 150/225/260 Amp Transfer Switch
OTPC 50/5/60 Amp Transfer Switch Printed in U.S.A. 6 08B (Spec A) 5-00 This catalog covers models produced under the Cummins /Onan and Cummins Power Generation brand names. To avoid errors or delay in
More informationLine reactors SINAMICS. SINAMICS G130 Line reactors. Safety information 1. General. Mechanical installation 3. Electrical installation
Safety information 1 General 2 SINAMICS SINAMICS G130 Mechanical installation 3 Electrical installation 4 Technical specifications 5 Operating Instructions Control version V4.7 04/2014 A5E00331462A Legal
More informationApplication Note on Modifying Disconnect Settings for Xantrex GT Series Single Phase Inverters and Conext TX Series Inverters
Application Note on Modifying Disconnect Settings for Xantrex GT Series Single Phase Inverters and Conext TX Series Inverters 976-0312-01-01 Revision A DANGER RISK OF FIRE, ELECTRIC SHOCK, EXPLOSION, AND
More informationPM Series Power Meter
PM Series Power Meter Quick Setup Guide - PMC-1000, PMC- 1001, PMM-1000, PMB-1960 Safety Information DANGER! HAZARD OF ELECTRIC SHOCK, EXPLOSION, OR ARC FLASH Follow safe electrical work practices. See
More informationINSTRUCTION MANUAL RA 19-4B RACK ADAPTER HOT SWAP RACK ADAPTER FOR HSF SERIES POWER SUPPLIES
INSTRUCTION MANUAL RA 19-4B RACK ADAPTER HOT SWAP RACK ADAPTER FOR HSF SERIES POWER SUPPLIES KEPCO INC. An ISO 9001 Company. MODEL RA 19-4B RACK ADAPTER ORDER NO. REV. NO. IMPORTANT NOTES: 1) This manual
More informationDCM Digital Control Modules
DCM Digital Control Modules TECHNICAL MANUAL Version 1.2 November 2011 Safety Precautions Caution Read Instructions: Read and understand all safety and operating instructions before using the equipment.
More informationCVU-3K-KIT. 3 kv Bias Tee Kit. Description. Parts list / October 2014 *P * 1
Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.keithley.com CVU-3K-KIT 3 kv Bias Tee Kit Description The CVU-3K-KIT Bias Tee Kit consists of three bias tees for
More informationQUICK SETUP GUIDE PMC-1000, PMC-1001, PMM-1000, PMB PM Series Power Meter. Safety Information. Equipment Maintenance and Service.
PM Series Power Meter QUICK SETUP GUIDE PMC-1000, PMC-1001, PMM-1000, PMB-1960 Safety Information DANGER! HAZARD OF ELECTRIC SHOCK, EXPLOSION, OR ARC FLASH Follow safe electrical work practices. See NFPA
More informationEAP 300 REMOTE ANNUNCIATOR INSTALLATION MANUAL. r.0454a
EAP 300 REMOTE ANNUNCIATOR INSTALLATION MANUAL r.0454a PM084 Rev 0 08/06/20 9087A 198 th Street, Langley, BC Canada V1M 3B1 Telephone (604) 888-0110 Telefax (604) 888-3381 E-Mail: info@thomsontechnology.com
More informationSINAMICS G130. Terminal Module 150 (TM150) Operating Instructions 03/2013 SINAMICS
SINAMICS G130 Operating Instructions 03/2013 SINAMICS s Safety information 1 General information 2 SINAMICS SINAMICS G130 Mechanical installation 3 Electrical installation 4 Technical specifications 5
More informationInstallation Guide. Retrofit Kit for USB Ready Intraoral Systems
Installation Guide Retrofit Kit for USB Ready Intraoral Systems Table of Contents Wall-Mount Retrofit Kit... 2 Introduction... 2 Connecting the Articulating and Horizontal Arm Cables... 2 Installing the
More informationAcquiSuite Ally 12 & 48 Advanced Multi-Circuit Meter
AcquiSuite Ally 12 & 48 Advanced Multi-Circuit Meter Install Guide Revision C (12/18) DANGER HAZARD OF ELECTRIC SHOCK, EXPLOSION, OR ARC FLASH Revision C (12/18) Turn off all power supplying equipment
More informationDigital Keypad Introduction
K2 Digital Keypad Introduction The K02 uses the latest microprocessor technology to operate door strikes and security systems that require a momentary (timed) or latching dry contact closure. All programming
More information170 ADM to 60 VDC Module Base / 170 ADO Pt. Relay Out Module Base. Version 1.0
170 ADM 850 10 10 to 60 VDC Module Base / 170 ADO 830 30 6 Pt. Relay Out Module Base Version 1.0 31005197 00 Safety Information NOTICE Read these instructions carefully, and look at the equipment to become
More informationInstruction Bulletin Boletín de instrucciones Directives d'utilisation , Rev. 02, 03/2016
Surgelogic EMA Series Surge Protective Devices (SPDs) Dispositivos de protección contra sobretensiones transitorias (SPD) Dispositifs de protection contre les surtensions transitoires (SPD) Instruction
More informationAutoScan to MultiScan Replacement Kit
Instruction Sheet P/N AutoScan to MultiScan Replacement Kit WARNING: Allow only qualified personnel to perform the following tasks. Observe and follow the safety instructions in this document and all other
More informationCOOPER POWER SERIES. Input/Output (I/O) module installation instructions. Voltage Regulators MN225067EN
Voltage Regulators MN225067EN Effective November 2016 Supersedes June 2014 (S225-70-13) COOPER POWER Input/Output (I/O) module installation instructions SERIES DISCLAIMER OF WARRANTIES AND LIMITATION OF
More informationMDE-4815B INSTALLATION MANUAL. This document is based on Orpak s WGT outdoor unit installation manual P/N
MDE-4815B WGT Outdoor Unit INSTALLATION MANUAL This document is based on Orpak s WGT outdoor unit installation manual P/N 817439300 SAFETY CONSIDERATIONS Read all warning notes and instructions carefully.
More informationOPERATING AND SERVICE MANUAL. Universal Interface Device 47
OPERATING AND SERVICE MANUAL Universal Interface Device 47 MAGNA-POWER ELECTRONICS, INC. 39 ROYAL ROAD, FLEMINGTON, NJ 08822 May 24, 202 SAFETY NOTICE Universal Interface Device 47 (UID47) connects two
More informationInstallation Instructions
50ES, 50EZ, 50GL, 50GS, 50GX, 50JS, 50JX, 50JZ, 50SD, 50SZ, 50VL, 50VT, 50XP, 50XZ 601A, 602A, 602B, 604A, 604B, 604D, 607C, 701A, 702A, 702B, 704A, 704B, 704D, 707C PA1P, PA2P, PA3P, PH1P, PH2P, PH3P
More informationWARNING ELECTRICAL SHOCK HAZARD
1 Product Overview These instructions provide information about installation of finger guard kits. The instructions are targeted at users already familiar with the VLT Soft Starters MCD 500, or VLT Compact
More informationWARNING ELECTRICAL SHOCK HAZARD
Description Safety Instructions The top-entry sub D9 connector kit is designed for enclosure sizes D1h D8h, E1 E2, and E1h E4h with VLT PROFIBUS DP MCA 101 option. It fits the following drives: VLT HVAC
More informationPowerFlex 400 Frame G and H Replacement Procedure for Gate Power Board
Service Bulletin PowerFlex 400 Frame G and H Replacement Procedure for Gate Power Board Contents This publication provides instructions for replacing the gate power board for PowerFlex 400 Frame G and
More informationConnecting a Cisco Output Module
CHAPTER 5 Overview The optional Cisco Output Module (Figure 5-1) is attached to a Cisco Physical Access Gateway or Cisco Reader Module to provide additional connections for up to 8 outputs, each of which
More informationIMPORTANT SAFETY INSTRUCTIONS
READ AND FOLLOW ALL SAFETY INSTRUCTIONS SAVE THESE INSTRUCTIONS AND DELIVER TO OWNER AFTER INSTALLATION IMPORTANT SAFETY INSTRUCTIONS WARNING RISK OF SEVERE INJURY OR DEATH To reduce the risk of death,
More informationXDBKITS5 Bus / DB Terminal Kit (Size 5) Installation Manual DPD00116
XDBKITS5 Bus / DB Terminal Kit (Size 5) Installation Manual DPD00116 XDBKITS5 Option Kit Installation Manual vacon 3 Installing the Bus / DB Terminal Option Kit Introduction The XDBKITS5 option kit is
More informationS66 Series Electronic Fan Speed Control
FANs 121, 125, 1628.3 Product/Technical Bulletin S66 Issue Date 0918 S66 Series Electronic Fan Speed Control The S66 Series Electronic Fan Speed Control is designed to modulate the speed of single-phase,
More informationINSTRUCTION MANUAL RA 19-7B RACK ADAPTER HOT SWAP RACK ADAPTER FOR HSF SERIES POWER SUPPLIES
INSTRUCTION MANUAL RA 19-7B RACK ADAPTER HOT SWAP RACK ADAPTER FOR HSF SERIES POWER SUPPLIES KEPCO INC. An ISO 9001 Company. MODEL RA 19-7B RACK ADAPTER ORDER NO. REV. NO. IMPORTANT NOTES: 1) This manual
More informationModel 2380 Rack-Mount Kit
Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.tek.com/keithley Model 2380 Rack-Mount Kit Installation Instructions Introduction The Model 2380 Fixed Rack-Mount
More informationINSTRUCTION MANUAL RA 19-6B RACK ADAPTER HOT SWAP RACK ADAPTER FOR HSF SERIES POWER SUPPLIES
INSTRUCTION MANUAL RA 19-6B RACK ADAPTER HOT SWAP RACK ADAPTER FOR HSF SERIES POWER SUPPLIES KEPCO INC. An ISO 9001 Company. MODEL RA 19-6B RACK ADAPTER ORDER NO. REV. NO. IMPORTANT NOTES: 1) This manual
More informationCisco CRS 3-Phase AC Power Distribution Unit Installation Guide 2. Cisco CRS 3-Phase AC Power Distribution Unit 2
Cisco CRS 3-Phase AC Power Distribution Unit Installation Guide Cisco CRS 3-Phase AC Power Distribution Unit Installation Guide 2 Cisco CRS 3-Phase AC Power Distribution Unit 2 Revised: November 18, 2016,
More information