Policy # MI_SER_GAL Department of Microbiology. Page Quality Manual TABLE OF CONTENTS

Size: px
Start display at page:

Download "Policy # MI_SER_GAL Department of Microbiology. Page Quality Manual TABLE OF CONTENTS"

Transcription

1 Department of Microbiology Version: 1.0 CURRENT 1 of 23 Prepared by QA Committee Issued by: Laboratory Manager Revision Date: 3/26/2018 Approved by Laboratory Director: Annual Review Date: 9/1/2018 Microbiologist-in-Chief Uncontrolled When Printed TABLE OF CONTENTS Introduction... 2 Specimen Collection and Processing... 3 Materials and Reagent... 2 Procedure... 2 System Reagent Preparation... 4 Specimen Treatment... 4 Testing Procedure... 5 Reporting Trouble shooting Maintenance Procedure Weekly Monthly Record of Edited Revisions... 22

2 Version: 1.0 CURRENT 2 of 23 Introduction The Platelia TM Aspergillus EIA is a qualitative enzyme immunoassay for the detection of Aspergillus galactomannan antigen in serum and Bronchoalveolar Lavage (BAL) samples. Reagent And Material 1. EVOLIS Analyzer 2. BioRad Aspergillus Galactomannan kit Store the kit at 2-8ºC. Bring all reagents to room temperature(18-25 ºC)for at least 30 minutes before use. Return all reagents to 2-8 ºC immediately after use. Return unused strips/plates to pouch and reseal. Galactomannan monoclonal antibodies coated microwell plate Concentrated wash solution (20X) Negative control serum (Human) Cut-off control serum (Human) Positive control serum (Human) Conjugate Anti-galactomannan monoclonal antibody/ peroxidase labeled Sample Treatment Solution (EDTA acid solution) Chromogen TMB solution Stopping solution 1N sulphuric acid 3. Distilled or deionized water, for dilution of Concentrated Washing Solution 4. Disposable gloves 5. Pipettes to measure and dispense 50 µl, 100 µl, 300 µl, and 1000 µl ml polypropylene microcentrifuge tubes with airtight stoppers, able to support heating to 120 ºC (heat block) 7. Heat block 8. Votex agitator Reagent Preparation And Storage 1. Microwell Strip Plate (R1) Every frame containing 12 strips I s packaged in a pouch. Open the pouch and take out the frame. Put the frame containing the unused strips back in the original pouch. Carefully reseal the pouch and store at 2-8ºC. After the vacuum-packed pouch has been opened, the strips stored at 2-8ºC in their original pouch that has been carefully resealed are stable for 8 weeks. Check whether the desiccant is still present.

3 Version: 1.0 CURRENT 3 of Washing Solution (R2) Prepare Working Washing Solution as neede by adding one part Concentrated Washing Solution (R2) to 19 parts deionized or distilled water. The Working Washing Solution can be stored for 14 days at 2-30 ºC. Prepare a sufficient amount of Working Washing Solution to complete the run (80ml for one strip: 4 ml R ml distilled water). 3. Negative Control Serum (R3), Cut-off Control Serum (R4) and Positive Control Serum (R5) The controls must be heat-treated with the Sample Treatment Solution (R7) as patient specimens, in order to also be a monitor of the treatment. At least one set of controls is needed for each batch of heating process. After opening, these reagents stored at 2-8ºC, are stable for 8 weeks, in the absence of contamination. 4. Conjugate (R6), Sample Treatment Solution (R7), Chromogen: TMB solution (R9) These reagents are ready to use. After opening, these reagents stored at 2-8ºC are stable for 8 weeks if they are free of contamination. 5. Stopping reaction (R10) This reagent is ready to use. After opening, this reagent stored at 2-8ºC is stable until the validity date shown on the label if there is no contamination. Specimen Collection and Processing The test is performed on serum or BAL fluid 1. Serum Collect blood samples according to standard laboratory procedures. Serum samples must be uncontaminated with fungal spores and/or bacteria. Transport and store samples in sealed tubes, unexposed to air. Unopened samples can be stored at 2-8ºC for up to 5 days prior to testing. After initial opening, samples may be stored at 2-8ºC for 48 hours prior to testing. For longer storage, store the serum at -70 ºC. Serum samples can be subjected to a maximum 4 freezing/thawing cycles. Previously frozen specimens should be thoroughly mixed after thawing prior to testing. The results are not affected by sample containing 20mg/L of bilirubin, lipemic samples containing the equivalent of 2 g/l of triolein (triglyceride) or hemolyzed samples containing 500 mg/dl of hemoglobin. Do not decomplement sera

4 Version: 1.0 CURRENT 4 of BAL Fluid Collect BAL fluid damples according to standard laboratory procedures. BAL fluid samples must be collected in sterile saline and may be tested on neat samples (as is) or supernatants from centrifuged samples (10,000 rpm for 10 min) before proceeding to treat the sample. BAL fluid samples must be uncontaminated with fungal spores and/or bacteria. Transport and store samples in sealed tubes, unexposed air. After initial opening, samples may be stored at 2-8ºC for up to 24 hours. For longer storage, store the BAL samples frozen (-20 ºC or less) up to 5 months. BAL samples can be subjected to a maximum of 4 freezing/thawing cycles. Previously frozen specimens should be thoroughly mixed after thawing prior to testing. Procedure System Reagent Preparation 1. Before beginning EIA assay, fill up Galactomannan BioRad Wash containers when necessary: *Working Galactomannan Wash buffer: Add 50 ml of 20x BioRad wash solution to 950 ml of distilled water. Fill up Galactomannan wash container with black tubing 2. De-ionized Water: Fill up Evolis blue, yellow and red tubing container with de-ionized water 3. The Aspergillus Galactomannan conjugate, TMB and stopping solution will be pipetted from their original containers and can be loaded onto the analyzer as is. 4. Bring all required microwell strips and the reagents mentioned above to room temperature before use. 5. Empty the waste tank and check the system liquid container (refill if necessary). System Liquid : 10L of de-ionized water only Specimen Treatment Separate BAL and serum and according to worklist Serum and BAL samples need to be heat-treated in the presence of EDTA prior to being tested on the analyzer. This pre-analytical procedure helps dissociate immune complexes and causes serum proteins to precipitate which can possibly interfere with the testing procedure. All control sera: negative (R3), cut-off (R4) and positive (R5) must be processed at the same time as serum / BAL fluid samples: Each batch of heating process needs a set of controls to be processed at the same time.

5 Version: 1.0 CURRENT 5 of Label 2 empty 2ml sample tubes. One tube must labelled with the patient barcode label. For the control samples label 2 empty tubes for each: Neg, Cut-Off & Pos, respectively. *The Cut-Off Control will be sampled twice by analyzer but only one sample tube needs to be prepared. 2. Pipette 390ul of each test serum/bal or control into one of the 2 ml sample tubes. 3. Add 130ul of serum treatment solution into the sample tube. 4. Tightly close the caps of the tubes to prevent opening during heating. 5. Mix tubes thoroughly by vortexing each sample. 6. Heat the sample tubes by placing them in a heating block at 120 o C for 6 minutes. Tubes must be placed in the block only when the prescribed temperature is reached. Do not rely on the temperature displayed by the apparatus, please check that the emperature complies with specification by using a calibrated thermometer which will be fitted into a tube containing mineral oil: 120 ºC must be reached inside the tube in a heat block. 7. Carefully remove the heated tubes from the heating block after 6 minutes and place them in the centrifuge. Centrifuge tubes at 10,000 x g for 10 minutes. The supernatant is used for the detection of the galactomannan antigen. 8. Transfer 300ul of supernatant into the second labeled 2ml sample tube which has patient s bar code and will be placed on to the Evolis TM analyzer. 9. After preparation, the supernatant may be removed and stored at 2-8 ºC for up to 48 hours prior to testing. If analysis of the results indicates that retesting of a sample is required, another aliquot of the serum or BAL must be heat treated for retesting. Testing Procedure 1. Bring reagents to room temperature (18-25 ºC) for at least 30 minutes before use. 2. Log onto the system by clicking on OK, no password is required. 3. A self-test of the system is automatically initialized each time the EVOLIS software is run. The self-test is considered satisfactory if the word PASSED appears beside each instrument module. Print a copy of the self-test report and combine it with the work list and result sheets. 4. Down load worklist to EVOLIS Analyzer. 5. Prepare specimens by removing the lids off the heat-treated serum/bal supernatant and load all sample tubes with the barcode facing right on to the sample rack (rack code T). Check the quality of the samples by ensuring all clots, foam and bubbles have been removed.

6 Version: 1.0 CURRENT 6 of Open the door to the sample and reagent unit. Load the sample rack with the bar-codes facing the bar-code reader onto the track marked by the solid red LED light. 7. The Patient Editor dialog box will appear with the specimen numbers from the loaded rack. Double check that the specimen numbers are correct. If there is a blank space under the patient ID column, click on the space and enter the specimen number manually. Select corresponding assay according to the worklist. All specimen numbers entered manually will be flagged with the code ManID on the results report. 8. The Galactomannan assays for BAL and serum should appears along order number Galactomannan assay for BAL= BAL Aspergillus Ag BR V16 Galactomannan assay for serum= Mount Sinai Aspergillus Ag BR V16 9. Change the assay by clicking on the assay box corresponding to the sample number if you want that assay to be run for that sample. A check mark is shown for a sample where the assay has been selected. 10. Once the required assay has been selected for each sample click Close to save. A solid LED light will show up for the next available track in the sample and reagent unit which indicates where the next sample rack can be loaded. Repeat steps 6 to 11 for each new sample rack that is loaded on to the analyzer. 11. A total of four sample racks need to be loaded on to the analyzer even if all the sample racks are not required in order for the reagent template to be used in a later step. When an empty sample rack is loaded on to the analyzer just click Close when the Patient Editor dialog box appears. 12. To check and validate the worklist click on the New Worklist icon located on the upper toolbar and the Set-Up dialog box will appear. 13. Click on + beside the assay file name that is associated with the plate. The plate layout will appear at the right hand side of the dialog box and will show the total number of controls and specimens to be tested, their assigned wells and the number of strips required for the assay. 14. If duplicate samples need to be tested for a given assay in the same run: a) From the Patient Editor dialog box select the assay folder in which the duplicate sample needs to be run and click Add Patient. b) The Select Patient(s) dialog box will appear.

7 Version: 1.0 CURRENT 7 of 23 c) Check the Allow multiple determinations box which appears at the bottom of the Select Patient (s) dialog box. d) The patient sample IDs will appear. Click on the sample ID in which multiple testing is required and click OK. e) In the Set-up Panel dialog box, a (x2) will appear beside the patient ID under the assay folder in which it was selected. 15. If all the information is correct click OK to validate the worklist. 16. A separate dialog box asking for the reagent lot number will appear for each assay that is ordered. Double check the lot number and expiry date and click OK if everything is correct. 17. Click on the + to expand the Work folder and click on Plate Layout, the number of microwell strips required for the assay can be reviewed. Click the Print button located along the top of the toolbar to print the plate layout if desired. 18. Click START (green button) located on the upper tool bar to open the Load dialog box. 19. The Load dialog box appears and shows where to load all the required resources. 20. Load all required resources from left to right: i. Pipette tips: Grey = 1100 ul tips (full rack) Brown= 300 ul tips (full rack) *The pipette tips have to be loaded in the exact position as shown in the layout ii. Load the assay reagents on the reagent racks according to the following templates (Figure 1). Figure 1 is suitable for one batch of heating process Figure 2 is suitable for two batches of heating process The reagent template specific for the Aspergillus Galactomannan assay should be used. iii. Ensure all caps on the reagent/control containers have been removed before loading them onto the analyzer.

8 Version: 1.0 CURRENT 8 of EMPTY 5. Negative Control 6. Cut-Off Control 7. Positive Control Controls for the 1 st batch of heating process Controls for the 2 nd batch of heating process Figure Once all the reagents and controls have been added to the racks, load the reagent racks onto the sample and reagent unit in the following order: 1 st = 0 - Large reagent rack 2 nd = 3 - Small reagent rack 22. In the Load dialog box, click the Open Reagent Layout button. Select the corresponding reagent file and click Open. All reagents and controls will be automatically allocated to their assign position.

9 Version: 1.0 CURRENT 9 of 23 Galactomannan.rea ----one batch of heating process BAL and Blood galactomannan.rea --- two batches of heating process 23. Once all the required resources have been loaded onto the analyzer click OK and the Load dialog box will disappear. The analyzer will begin to check the resources to ensure their amounts are sufficient. 24. The Load Plate dialog box will then appear. Rename the plate ID with the corresponding assay name, for example Galactomannan. The test date will automatically be added to the plate ID by the analyzer. 25. Prepare the required number of microwell strips for the requested assay. The number of strips required is shown in the Plate Layout which is found on the right hand side of the Load Plate dialog box. 26. Insert the microplate into the metal frame microplate holder. Make sure that the A1 position of the plate and holder match. 27. Open the door to the microplate loading compartment and load the microplate and its holder onto the plate transport unit. Once loaded, close the door to the microplate loading compartment. 28. Once the plate is loaded click OK. After the microplate is loaded, the analyzer will start to run automatically. 29. From the Work folder click on Schedule to view the chorological order and completion time of the run. 30. While the samples are being pipetted click on Active Event Log to ensure that all samples were pipetted successfully. Look for any red flags that may appear in the Dispense Sample section which indicates that an error has occurred. The analyzer will pause and flag with an error message if a clot or low sample volume is encountered, allowing for operator intervention. 31. The results report will automatically print out after the assay is completed. 32. Once the assay is finished a dialog box prompting the removable of the test plate and carrier from the analyzer will appear. 33. Open the door to the microplate loading compartment and remove the microplate. After the microplate is unloaded close the door to the microplate loading compartment and click OK.

10 Version: 1.0 CURRENT 10 of A blinking red LED light in the sample and reagent unit indicates that the reagent or sample racks can be unloaded from the analyzer. 35. Do not unload pipette tip racks from the analyzer unless they are completely empty. 36. Close the EVOLIS software. Select File Exit from the menu bar or click the X icon at the top right-hand corner of the Evolis software and shut down the computer. Reporting The presence or absence of galactomannan antigen in the test sample (either serum or BAL) is determined by the calculation of an index for each patient sample. To calculate the Index of a sample, divide the O.D of the patient sample by the mean O.D of the cut-off control. Sample Index = Patient Sample O.D. Mean Cut-off Control O.D Patient serum or BAL with an index < 0.50 is considered to be negative for galactomannan antigen. Patient serum or BAL with an index 0.50 is considered to be positive for galactomannan antigen. All positive specimens should be heat-treated and repeated x2 on the next run. EXCEPTION: A positive 8GAL or 8GALB can be referred, without confirming, for 3weeks after a Confirmed (sample repeated in duplicate) 8GAL or 8GALB. Confirmation must be performed on same sample type, ie. You cannot refer a positive serum result to confirmation results on a BAL. For a sample O.D.= [OD>3.50]: 1. calculate the index for OD=3.50 : 3.50 Mean Cut-off Control O.D 2. Report patient s index > the index of OD=3.50 See Evolis EIA Worksheet for calculations

11 Version: 1.0 CURRENT 11 of 23 All repeats have to be recorder inevolis REPEATS WORKLIST, and give the filled-up worklist to a head LIS officer. An absorbance value of less than may indicate a procedure or instrument error and the result is considered invalid and the specimen should be re-run. Reporting Rules for reporting Index 0.5 Index Initial Reporting After repeat in duplicate 0.5 Index 1.0 Index> Order 8COM 2. Result 8COM:type initial index value 3. Order 8GAL1 and verify 4. Order 8GAL2 and verify 1. 8GAL:Positive Report index in the comment: (Index Values: #.##) *Preliminary testing POSITIVE, to be repeated. 2. Verify 8GAL result 3. Order 8GAL1 and verify 3. Order 8GAL2 and verify 2 of 3 results are positive 8GAL:Positive Report 3 index in the comment: (Index Values: #.##/#.##/#.##) Round off to 2 decimal places 2 of 3 results are Negative 8GAL:Negative Report 3 index in the comment: (Index Values: #.##/#.##/#.##) Round off to 2 decimal places 2 of 3 results are positive 8GAL:Positive Report 3 index in the comment: (Index Values: #.##/#.##/#.##) Round off to 2 decimal places 2 of 3 results are Negative 8GAL:Negative Report 3 index in the comment: (Index Values: #.##/#.##/#.##) Round off to 2 decimal places Trouble shooting No index, but has the following error: P_max_high 1.Order 8COM 2.Result 8COM:type the error Only needs to be repeat once Report according the

12 Version: 1.0 CURRENT 12 of 23 P_min_low P_stop_high P_static_high P_static_high P_mean_low 3.Order 8GAL1 and verify result!

13 Version: 1.0 CURRENT 13 of 23 Maintenance Procedure General Maintenance Tasks Always turn off the instrument before cleaning Daily 1. Perform start-up maintenance (check liquid levels) Make sure the system liquid container is full or almost full and correctly connected to the system. If necessary, refill the liquid-system container. Make sure the liquid waste container is ok. 2. Run the washer maintenance: (Procedure takes approximately 20 min) Fill all wash buffer containers with de-ionized water Click New Worklist located in the upper toolbar The Set-up panel dialog box will appear In the Set-up Panel dialog box click Add Plate In the Set-up Panel dialog box click Add Assay and then select the file WasherClean BR.asy, click Open File. Click OK to validate the various dialog boxes until the worklist appears. Click Start (green button) Load an empty washer microplate with the metal plate holder onto the analyzer when the Load Plate dialog box appears. Click OK after the plate is loaded onto the analyzer. When the run is complete, unload the washer plate and replace the de-ionized water with the appropriate wash buffers. *The daily washer maintenance does not have to be completed on the day the monthly washer maintenance is being done. 3. Perform end-of-day maintenance assay: Check the bag of the tip ejection waste container. If full or nearly full, replace as described. Check the liquid waste level in the liquid waste container, if full or nearly full, decontaminate and empty as described as below: i. Remove cap and pour bleach into the container (without emptying the container first). With ordinary domestic-use bleach (12 chlorine), the

14 Version: 1.0 CURRENT 14 of 23 ii. iii. volume of bleach should equal 10% of the volume of liquid waste in the container. Let it stand for a minimum of 30 minutes (or overnight). Empty the container and rinse it thoroughly with tap water. When adding bleach, it is recommended that you operate under a fume hood as chlorine fumes may occur After adding bleach, do not place the waste container screw cap back on as this may damage the sensor. iv. The screw cap and sensor part may be wiped with a cloth moistened with Microcide (diluted in water at 1.6% 15 ml per 950mL ) Close the EVOLIS software. Select File Exit from the menu bar or click the X icon in the top right-hand corner of the EVOLIS software window. Exit Windows Wait until Windows displays the following message: It is now safe to turn off your computer. Switch off the system (switch on back panel of the instrument). Always exit the EVOLIS software and the Windows software before switching off the system! Open the instrument cover and wipe the tip adapter (pipettor head) using a soft linfree cloth moistened with 70% Ethanol Inspect the instrument (inner and outer surfaces) for stains and spills. If necessary, clean as described as below:

15 Version: 1.0 CURRENT 15 of 23 Outer surfaces Cover and handle Outer panels, including washer window Inner surfaces Dilution and tip rack area Sample and reagent unit Test plate pipetting area Pipettor guide rail Bottom drawer Plate transport unit and compartment Room-temperature incubators Heating incubators Under wash solution containers Computer Computer and monitor Clean with any bactericide, virucide and fungicide hospital-grade disinfectant. To prepare and apply the disinfectant (dilution, spray, wait, wipe ) refer to the instructions prescribed by the product manufacture Same as above Same as above Clean once a month with the same disinfectant as above. Spray the disinfectant on a lint-free paper wipe and wipe the inside of each incubator

16 Version: 1.0 CURRENT 16 of 23 Weekly 1. Run the weekly washer maintenance: (Procedure takes approximately 20 min) Fill all wash buffer containers with de-ionized water Click New Worklist located in the upper toolbar The Set-up panel dialog box will appear In the Set-up Panel dialog box click Add Plate In the Set-up Panel dialog box click Add Assay and then select the file WasherClean BR.asy, click Open File. Click OK to validate the various dialog boxes until the worklist appears. Click Start (green button) Load an empty washer microplate with the metal plate holder onto the analyzer when the Load Plate dialog box appears. Click OK after the plate is loaded onto the analyzer. When the run is complete, unload the washer plate and replace the de-ionized water with the appropriate wash buffers. *The weekly washer maintenance does not have to be completed during the week the monthly washer maintenance is being done. 2. Decontaminate the pipettor wash station: Pour 5 ml of decontamination solution consisting of 1.6% Microcide (15ml of Microcide into 950mL of water) into the pipettor wash station and let it soak for a minimum of 15 minutes. Do not empty, the liquid will drain automatically when the system is initialized. 3. Decontaminate and wipe the tip ejection slide. 4. Clean the instrument surfaces and work area. Monthly 2. Run the monthly washer maintenance: (Procedure takes approximately 1 hour) Fill all the wash buffer containers with de-ionized water Click New Worklist located in the upper toolbar The Set-up panel dialog box will appear In the Set-up Panel dialog box click Add Plate

17 Version: 1.0 CURRENT 17 of 23 In the Set-up Panel dialog box click Add Assay and then select the file WasherManifoldDisinfect BR.asy, click Open File Click OK to validate the various dialog boxes until the worklist appears. Click Start (green button) The Load dialog box will appear. Pour 50ml of 1.6% Microcide into a 60ml of container and load the reagent onto a 0 reagent rack. Then load the rack onto the analyzer. In the Load dialog box, allocate the bottle to the corresponding rack position in which the container was loaded then click OK. Load an empty washer microplate with the metal holder onto the analyzer when the Load Plate dialog box appears. Click OK after the plate is loaded onto the analyzer. A few minutes later a message saying to open the drawer, unscrew waste bottle 1 and close the drawer will appear. Unscrew the cap to waste bottle 1, which is located behind the wash buffer bottles. Close the drawer when done and click OK. After 15 minutes a message prompts you to re-open the drawer and re-screw the cap to waste bottle 1, close the drawer when completed and click OK. When the run is complete, unload the plate and the reagent rack and replace the de-ionized water with the appropriate wash buffers. 3. Decontaminate the sample and reagent racks, plate carrier and tip ejection slide. 4. Decontaminate the system liquid container Empty the system liquid container. Empty the container and rinse thoroughly, twice with tap water and once with de-ionized water. Inspect the filter (attached to the cap) to see if damaged or in need of replacing. Refill the container with freshly prepared System Liquid. Preparation of System Liquid: 10L of de-ionized water. 5. Clean the washer buffer bottles only and NOT the caps & sensor devices. 6. Backup System Files Click Backup button found at the upper tool bar. The System Backup dialog box will open up. Click Backup System Files Once completed, click Close from the System Backup dialog box.

18 Version: 1.0 CURRENT 18 of 23 Quality Control 1. Calculate the mean absorbance of the Cut-Off Control (Cut-Off X): Sum of the O.D values for the two Cut-Off controls divided by 2. The O.D of each Cut-Off control must be and to be considered valid. 2. Calculate the Negative Control Index by dividing the O.D of the Negative Control by the mean absorbance of the Cut-Off Control. Negative Control Index = O.D Negative Control Mean Cut-off Control O.D The index of the Negative Control Serum must be less than 0.40 to be considered valid. 3. Calculate the Positive Control Index by dividing the O.D of the Positive Control by the mean absorbance of the Cut-Off Control. Positive Control Index= O.D Positive Control Mean Cut-off Control O.D The index of the Positive Control Serum must be greater than 2.00 to be considered valid. All QC is checked by senior technologist. Download Worklist To EVOLIS Analyzer 1. Login to the Softlab, double click Instrument Menu under Interface

19 Version: 1.0 CURRENT 19 of High light EVOLIS and click OK button 3. Select Download All in Loadlist drop down menu

20 Version: 1.0 CURRENT 20 of Downloading Procedure window pops up, choose By Order for How to Download 5. For Starting From option, keep the letter and change the first three digits to all 0, then click OK button 6. The Instrument Menu window pops up and shows how many orders will be downloaded. Click OK button

21 Version: 1.0 CURRENT 21 of 23 Related Documents Serology Specimen Management Manual GAL Calculations for Reports Evolis Repeat Worklist Serology External QC Current Serology Inventory Current T:\microbiology\Virology\QC statistics\external QC and INVENTORY Logs\Serology External QC & Inventory CURRENT xls QPEMI03001 Reference: Package insert from Platelia Aspergillus EIA Revised July 2007 EVOLIS Short User Manual software version 1.90, revised June 2008 EVOLIS User Manual software version 1.90, revised June 2000

22 Version: 1.0 CURRENT 22 of 23 Record of Edited Revisions Manual Section Name: Aspergillus Galactomannan Antigen Number / Item Date of Revision Signature of Approval New Heating procedure with template added. QC set done May 25, 2016 Dr. T. Mazzulli with each eating processed. Evolis Repeat Worklist Added to Related documents table June 16, 2016 Dr. T. Mazzulli Reporting table modified: All sites now reporting as UHN. 1 November 22, 2016 Dr. T. Mazzulli General table created. Added note to all reporting round off to 2 decimal places. Annual Review August 8, 2017 Dr. T. Mazzulli Updated rivascope to Microcide cleaning solution and modified quantities. Updated System liquid: removed tween, only 10L of deionized water required. March 21, 2018 Dr. T. Mazzulli

23 Version: 1.0 CURRENT 23 of 23 Number / Item Date of Revision Signature of Approval

Issued by: LABORATORY MANAGER Original Date: March 14, 2001 Approved by: Laboratory Director Revision Date: June 6, 2003.

Issued by: LABORATORY MANAGER Original Date: March 14, 2001 Approved by: Laboratory Director Revision Date: June 6, 2003. Policy # MI/SER/06/v01 Page 1 of 5 Section: Subject Title: Epstein Barr Virus Serology Issued by: LABORATORY MANAGER Original Date: March 14, 2001 Approved by: Laboratory Director Revision Date: June 6,

More information

GEMINI. Compact Microplate Processor. System Guide

GEMINI. Compact Microplate Processor. System Guide GEMINI Compact Microplate Processor System Guide 1 Gemini System Guide Table of Contents 1. Introduction 3 2. User Levels 4 3. Instrument Components & Accessories 6 4. Basic Operation - Quick Guide: 1-Lane

More information

BioRobot 9604 Operating Procedure

BioRobot 9604 Operating Procedure BioRobot 9604 Operating Procedure Note: An import table (directions given in number 20) must be made prior to a robot run. 1. Prior to operating the BioRobot 9604, perform all daily and weekly maintenance

More information

4 plate fully automated ELISA analyzer Reagent system. Open (Human tests predefined) Throughput (max.) Up to 7 plates / run Reaction / reading system

4 plate fully automated ELISA analyzer Reagent system. Open (Human tests predefined) Throughput (max.) Up to 7 plates / run Reaction / reading system Specifications & more > > System Specifications > > Consumables > > Accessories System Overview Cat. No. 16300 Analyzer type 4 plate fully automated ELISA analyzer Reagent system Open (Human tests predefined)

More information

Protein Quantification Kit (BCA Assay)

Protein Quantification Kit (BCA Assay) Protein Quantification Kit (BCA Assay) Booklet Item NO. KTD3001 Product Name Protein Quantification Kit (BCA Assay) ATTENTION For laboratory research use only. Not for clinical or diagnostic use Version

More information

1 plate fully automated ELISA analyzer Reagent system. Open (Human tests predefined) Throughput (Max.) 1 plate /run Barcode reader

1 plate fully automated ELISA analyzer Reagent system. Open (Human tests predefined) Throughput (Max.) 1 plate /run Barcode reader Specifications & more Elisys Uno > > System Specifications > > Consumables > > Accessories System Overview Cat.. 17350 Analyzer type 1 plate fully automated ELISA analyzer Reagent system Open (Human tests

More information

STANDARD OPERATING PROCEDURE

STANDARD OPERATING PROCEDURE Document No: LAB/023/2 Page 1 of 7 Uncontrolled copy if not printed on yellow paper. Title: STANDARD OPERATING PROCEDURE Use of the Magtration System 8Lx Document No: LAB/023 Version No: 2 CopyNo/Holder:

More information

icolumn 12 Operation Manual

icolumn 12 Operation Manual icolumn 12 Operation Manual AccuBioMed Co., Ltd. 8F.-8, No.5, Wuquan 1st Rd., Xinzhuang Dist., New Taipei City 24892, Taiwan (R.O.C.) Tel: +886-2-2299-5989 Fax: +886-2-2299-2678 www.accubiomed.com AccuBioMed

More information

Policy # MI_VTK Department of Microbiology. Page Quality Manual VITEK MANUAL TABLE OF CONTENTS

Policy # MI_VTK Department of Microbiology. Page Quality Manual VITEK MANUAL TABLE OF CONTENTS Department of Microbiology Version: 1.0 CURRENT 1 of 28 Prepared by QA Committee Issued by: Laboratory Manager Revision Date: 10/2/2017 Approved by Laboratory Director: Annual Review Date: 5/1/2018 Microbiologist-in-Chief

More information

QUICKStart Guide APT Versions 3.0 through 3.3

QUICKStart Guide APT Versions 3.0 through 3.3 QUICKStart Guide APT Versions 3.0 through 3.3 PCS QUICKStart Guide Powering On & Off...2-3 Seiko Printer...4-5 Consumable Components...6-7 Pipette Calibration Setup...8-11 Pipette Calibration...12-13 Completing

More information

Microlab F.A.M.E. Version 2.0

Microlab F.A.M.E. Version 2.0 Microlab F.A.M.E. U S E R M A N U A L Version 2.0 Important Notice This manual may not be used or reproduced in any way whatsoever without the express written consent of Hamilton Bonaduz AG. Copyright

More information

BCA protein assay kit for low concentrations. For measuring total protein concentration of pure proteins, extracts or lysates.

BCA protein assay kit for low concentrations. For measuring total protein concentration of pure proteins, extracts or lysates. ab207002 BCA protein assay kit for low concentrations Instructions for use: For measuring total protein concentration of pure proteins, extracts or lysates. This product is for research use only and is

More information

(11-nor-9-carboxy- 9-THC) PCP Phencyclidine (Phencyclidine)

(11-nor-9-carboxy- 9-THC) PCP Phencyclidine (Phencyclidine) Profile-V Reader System Quick Reference Instructions Before performing the test, refer to the PROFILE -V MEDTOXScan Drugs of Abuse Test System Insert and MEDTOXScan User Manual for complete operating instructions

More information

NCSBI Forensic Biology Section DNA SOP Effective Date: January 30, 2007 Title: ABI PRISM 7000: DNA Quantitation Revision 02

NCSBI Forensic Biology Section DNA SOP Effective Date: January 30, 2007 Title: ABI PRISM 7000: DNA Quantitation Revision 02 1. Preparing the DNA Standards 1.1. Label eight microcentrifuge tubes to be used for dilution of standards (i.e. Std. 1, Std. 2, Std. 3, etc.) 1.2. Add required amount of TE/glycogen to each tube (See

More information

TG Autoantibody Assays

TG Autoantibody Assays TG Autoantibody Assays 24 May 2011 Page 1 of 14 Contents Making Label.....3 Plating.....6 Diluting Label........7 Washing and adding PAS/IgA-Agarose...9 Washing Plates and Transferring 10 Getting Counts

More information

US IVD Bacterial Test Standard

US IVD Bacterial Test Standard Instructions for Use US IVD Bacterial Test Standard Mass calibration standard containing a typical Escherichia coli DH5 alpha peptide and protein profile plus additional proteins. For quality control and

More information

Objectives 6/15/2015. Review of Customer Communications. Benefits of Immucor s Website

Objectives 6/15/2015. Review of Customer Communications. Benefits of Immucor s Website Review of Customer Communications Presented by: Chris Beritela, MS, BSMT(ASCP)SBB Area Technical Consultant- East Immucor, Inc. The content of this slide presentation is for information and illustration

More information

MAKE THE CHANGE FOR BETTER NEONATAL SCREENING. Introducing the new AutoDELFIA immunoassay system for newborn screening

MAKE THE CHANGE FOR BETTER NEONATAL SCREENING. Introducing the new AutoDELFIA immunoassay system for newborn screening Introducing the new AutoDELFIA immunoassay system for newborn screening MAKE THE CHANGE FOR BETTER NEONATAL SCREENING THE WORLD S LEADING SCREENING PLATFORM PerkinElmer s AutoDELFIA immunoassay system

More information

CHLORINE DIOXIDE (0 to 5.00 mg/l) Method 10126

CHLORINE DIOXIDE (0 to 5.00 mg/l) Method 10126 CHLORINE DIOXIDE (0 to 5.00 mg/l) Method 10126 DPD Method* Scope and Application: for measuring chlorine dioxide in clean water For water 530 nm 1. Enter the stored program for chlorine dioxide (ClO 2

More information

intellipath Research Software Manual

intellipath Research Software Manual intellipath Research Software Manual www.biocare.net Biocare Medical Technical Support: 800-799-9499 Option 3 Customer Service: 800-799-9499 Fax: 925-603-8080 Corporate Office: 4040 Pike Lane Concord,

More information

Magnetic Immunoassay Reader Operator Manual

Magnetic Immunoassay Reader Operator Manual Page 1 of 8 Magnetic Immunoassay Reader Operator Manual Table of contents 1 Introduction...1 2 Magnetic Immunoassay Reader...1 2.1 Contact information and spare parts...1 3 Contents of delivery...1 4 Technical

More information

Preprocessing Methods for the Maxprep Liquid Handler Technical Manual

Preprocessing Methods for the Maxprep Liquid Handler Technical Manual Preprocessing Methods for the Maxprep Liquid Handler Technical Manual Instructions for use of AS9100, AS9101, AS9200 and AS9201 TM529 11/17 Table of Contents 1 Description...1 2...2 2.1. Maxprep Liquid

More information

FULLY AUTOMATED 2 MICROPLATE ANALYZER

FULLY AUTOMATED 2 MICROPLATE ANALYZER FULLY AUTOMATED 2 MICROPLATE ANALYZER The new Personal Lab,(Plab) is an integration between our unique extensive expertise and information gathered from thousands of installations. The new Plab is innovative

More information

NEW IMPROVED SCREENING PLATFORM MATERNAL FETAL HEALTH

NEW IMPROVED SCREENING PLATFORM MATERNAL FETAL HEALTH Introducing the new AutoDELFIA immunoassay system for prenatal screening Brochure not for distribution in the USA NEW IMPROVED MATERNAL FETAL HEALTH SCREENING PLATFORM ALREADY 60 MILLION PREGNANCIES SCREENED

More information

STAINING PROTOCOL FOR CYTOMETRY (Réf LabEx #1)

STAINING PROTOCOL FOR CYTOMETRY (Réf LabEx #1) STAINING PROTOCOL FOR CYTOMETRY (Réf LabEx #1) Aim To provide detailed description of all steps, material and methods used for fluorescent surface staining and preparation of whole blood cells for analysis

More information

ab BCA protein assay kit reducing agent compatible (test tube)

ab BCA protein assay kit reducing agent compatible (test tube) ab207004 BCA protein assay kit reducing agent compatible (test tube) Instructions for use: For measuring total protein concentration of pure proteins, extracts or lysates in the presence of reducing agents.

More information

Microplate Systems. Absorbance Microplate Readers 366. Performance Validation and Software Tools 368. Microplate Washer 370

Microplate Systems. Absorbance Microplate Readers 366. Performance Validation and Software Tools 368. Microplate Washer 370 /microplate/ Absorbance Microplate Readers 366 Performance Validation and Software Tools 368 Microplate Washer 370 Singleplex Assays and Reagents 371 Microplate Systems Absorbance Microplate Readers /microplate/

More information

Insulin Autoantibody Assays

Insulin Autoantibody Assays Insulin Autoantibody Assays 24 May 2011 Page 1 of 13 Contents Initial dilution of Insulin Label.. 3 IAA Plating.....4 ciaa Plating... 5 Diluting IAA Label........6 Diluting ciaa Label...... 7 Washing PAS,

More information

MAKE THE CHANGE FOR BETTER NEONATAL SCREENING. Introducing the new AutoDELFIA immunoassay system for newborn screening

MAKE THE CHANGE FOR BETTER NEONATAL SCREENING. Introducing the new AutoDELFIA immunoassay system for newborn screening MAKE THE CHANGE FOR BETTER NEONATAL SCREENING Introducing the new AutoDELFIA immunoassay system for newborn screening THE WORLD S LEADING SCREENING PLATFORM PerkinElmer s AutoDELFIA immunoassay system

More information

TECHNICAL MANUAL VERSION RoboSep The Fully Automated Cell Separator

TECHNICAL MANUAL VERSION RoboSep The Fully Automated Cell Separator TECHNICAL MANUAL VERSION 2.3.1 RoboSep The Fully Automated Cell Separator i ii SAFETY INFORMATION AND WARNINGS This instrument is intended for research use only and must be operated by trained professionals.

More information

DSX Automated ELISA System

DSX Automated ELISA System DSX Automated ELISA System Operator s Manual For Revelation 6.0 and above IMPORTANT Please read carefully before using the DSX ii Revision History Revision Date: September 19, 2005 12-16-2005 Part No.

More information

Vial block. Power switch

Vial block. Power switch Powering On & Off The color-coded bars within the text throughout this Guide indicate the sections where specific instructions are given. Vial block Power switch Powering On Connect system according to

More information

MINISCAN INSTRUCTIONS

MINISCAN INSTRUCTIONS MINISCAN INSTRUCTIONS DR VIRGINIA GORDON Version 3, July 2011 General Hints on Running Mini-Scan Tests... 1 Total Glycerin Test (TG)... 3 GENERAL HINTS ON RUNNING MINI-SCAN TESTS PIPETTES To use the pipettes

More information

SymBiot I Sample Workstation

SymBiot I Sample Workstation SymBiot I Sample Workstation Getting Started Guide DRAFT August 28, 2001 8:42 pm Symbfmgs.fm5 Copyright 1997, 2001, Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic

More information

FACSAria Fusion SORP QUICK REFERENCE GUIDE

FACSAria Fusion SORP QUICK REFERENCE GUIDE FACSAria Fusion SORP QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Admin Password: BDIS 2. Ensure the flow cell access door is open. You will want to

More information

Manual UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY

Manual UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY Quality Version: 1.0 CURRENT 1 of 74 INTRODUCTION... 3 WASPLAB COMPONENTS... 3 WASPLab WebApp Workflow... 5 PROCEDURE... 6 Software Access... 6 Assigning a password to new users... 6 Changing passwords...

More information

SPA III Quick Reference Guide for Custom Panels

SPA III Quick Reference Guide for Custom Panels SPA III Quick Reference Guide for Custom Panels This guide contains instructions for using the BD FACS Sample Prep Assistant III (SPA III) with BD FACS SPA software version.0 and later. Initial Setup Custom

More information

Total WBC count and differential WBC count HemoCue WBC DIFF system

Total WBC count and differential WBC count HemoCue WBC DIFF system TLM_PRO_029 and Version No. 001 Page 1 of 6 Department: Clinical Sciences Unit: Tropical Laboratory Medicine Total WBC count and differential WBC count HemoCue WBC DIFF system Project/study: Not applicable

More information

ARRAYMATE. Operating Manual

ARRAYMATE. Operating Manual ARRAYMATE Operating Manual Version 2.1 Alere Technologies 2013 Table of Contents: 1. 1.1 Introduction... 3 Instrument Overview... 3 2. 2.1 Safety Information... 4 Important Operation Notes... 4 2.2 Sign

More information

ONCORE Manual Automated Slide Staining System

ONCORE Manual Automated Slide Staining System ONCORE Manual Automated Slide Staining System www.biocare.net Biocare Medical Technical Support: 800-799-9499 Option 3 Customer Service: 800-799-9499 Option 2 Fax: 925-603-8080 Corporate Office: 60 Berry

More information

Unscrew and remove the cap from the waste tank

Unscrew and remove the cap from the waste tank Stanford Cytoflex HTS User Guide 11232015 Starting up 1. Check the sheath fluid level every time you use the cytometer. This ensures that you do not run out of sheath fluid during an experiment. Replenish

More information

QIAsymphony SP/AS Operating the QIAsymphony SP

QIAsymphony SP/AS Operating the QIAsymphony SP December 2017 QIAsymphony SP/AS Operating the QIAsymphony SP For use with software version 5.0 QIAGEN GmbH, QIAGEN Strasse 1, 40724 Hilden GERMANY R1 1107354 Sample to Insight Contents 1 Introduction...8

More information

FACSLSRFortessa SORP QUICK REFERENCE GUIDE

FACSLSRFortessa SORP QUICK REFERENCE GUIDE FACSLSRFortessa SORP QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Administrator Password: BDIS 2. Unlock the screen with your PPMS account (UTSW username

More information

ABI PRISM 7700 Sequence Detection System: Routine Maintenance

ABI PRISM 7700 Sequence Detection System: Routine Maintenance ABI PRISM 7700 Sequence Detection System: Routine Maintenance Overview The ABI PRISM 7700 Sequence Detection System (SDS) and the Macintosh computer that runs the SDS need to be properly and routinely

More information

Thermo Scientific Samco Transfer Pipettes. Precisely what you need

Thermo Scientific Samco Transfer Pipettes. Precisely what you need Thermo Scientific Samco Transfer Pipettes Precisely what you need Yesterday, today Thermo Scientific Samco transfer pipettes have been the industry leaders for over 35 years Design Grips easily with latex

More information

Quick Solutions Packet

Quick Solutions Packet Pub. No. J33039_EN 2008-11-15 Quick Solutions Packet ORTHO AutoVue Innova ORTHO AutoVue Ultra Topic No. Title 1 Interrupted Liquid Flow Error (ASPM255) 2 Replace the Air Filter 3 Defragment the Hard Drive

More information

User manual Code: ORG Name: Alegria

User manual Code: ORG Name: Alegria User manual Code: ORG 300-00 Name: Alegria ORG 300-00_user manual_en_qm113123_2017-07-11_3 page 1 of 51 Content 1. MANUFACTURER... 4 2. IDENTIFICATION... 4 3. INTENDED USE AND PURPOSE... 4 4. CONDITIONS

More information

The NGAL Test Application Note for Abbott Architect c8000

The NGAL Test Application Note for Abbott Architect c8000 The NGAL Test Application Note for Abbott Architect c8 General parameters Name: NGAL Assay number: *1 Assay version: 1 Assay type: Photometric Assay availability: Enabled Cal version: 1 Reaction definition

More information

STANDARD OPERATING PROCEDURES DIVISON OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

STANDARD OPERATING PROCEDURES DIVISON OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA STANDARD OPERATING PROCEDURES DIVISON OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA SOP#: 1139.5 Date Issued: 12/03 Date Revised: 8/16 Page 1 of 5 TITLE: SCOPE: RESPONSIBILITY: PURPOSE: Animal Care

More information

AnaPrep 12. Instruction Manual. BioChain Institute Inc. Automated Nucleic Acid Preparation System

AnaPrep 12. Instruction Manual. BioChain Institute Inc. Automated Nucleic Acid Preparation System BioChain Institute Inc. Instruction Manual Automated Nucleic Acid Preparation System AnaPrep 12 Catalog No.: Z1321001 Rev. Date: April 30, 2015 Manual No. Z1001 Version: Z1321001UA (MV1.14) For General

More information

Fast Facts. for Operating the scil Vet abc Plus. Hematology Chemistry Digital Radiography Ultrasound Orthopedics Education

Fast Facts. for Operating the scil Vet abc Plus. Hematology Chemistry Digital Radiography Ultrasound Orthopedics Education Fast Facts for Operating the scil Vet abc Plus Hematology Chemistry Digital Radiography Ultrasound Orthopedics Education SCIL Vet ABC Plus Fast Facts Table of Contents Start Up Cycle pg. 2 Reagents Status

More information

Standard IA2A, GADA and ZnT8A Assays

Standard IA2A, GADA and ZnT8A Assays Standard IA2A, GADA and ZnT8A Assays 24 May 2011 Page 1 of 14 Contents Making Standard Labels.....3 Plating.....6 Diluting Label........8 Washing PAS, Adding PAS....10 Washing Plates and Transferring 11

More information

Ion 540 Kit Chef QUICK REFERENCE. Dilute the libraries. Prepare the libraries and consumables. Create a Planned Run. Catalog Numbers A27759, A30011

Ion 540 Kit Chef QUICK REFERENCE. Dilute the libraries. Prepare the libraries and consumables. Create a Planned Run. Catalog Numbers A27759, A30011 QUICK REFERENCE Ion 540 Kit Chef Catalog Numbers A27759, A30011 Pub. No. MAN0014119 Rev. C.0 Note: For safety and biohazard guidelines, see the Safety appendix in the Ion 540 Kit Chef User Guide (Pub.

More information

Skyla VB1 Quick Reference Guide

Skyla VB1 Quick Reference Guide Skyla VB1 Quick Reference Guide 1 VB1 DM/JAC 2014 v1.0 Quick Reference Guide Contents External features of the analyser 3 VB1 Function Buttons 4 Testing a Sample 5 How to reconstitute a biochemistry control

More information

Instructions for How to Setup the Robot to Print

Instructions for How to Setup the Robot to Print Instructions for How to Setup the Robot to Print Introduction This document provides guidelines for setting up a Pat O. Brown arrayer * to print antigens onto glass slides. Pat Brown s website, http://cmgm.stanford.edu/pbrown/mguide/index.html,

More information

JMJ-Group Microbiology BMB SDU

JMJ-Group Microbiology BMB SDU JMJ-Group Microbiology BMB SDU Title: Daily use of FACS Aria II SOP number: SOPFACS0042_v01 Version number: 01 Date issued: 2013.01.16 Review date: 2014.12.01 Written by: Tina Kronborg and Michelle Madelung

More information

Random Access Analyzer For Immuno-Assay, Clinical Chemistry & Plasma Proteins

Random Access Analyzer For Immuno-Assay, Clinical Chemistry & Plasma Proteins Random Access Analyzer For Immuno-Assay, Clinical Chemistry & Plasma Proteins Technical Features Batch and random assess. Eclectica is the first compact random access system for Immuno-Assay and Clinical

More information

QIAsymphony SP/AS User Manual Operating the QIAsymphony SP

QIAsymphony SP/AS User Manual Operating the QIAsymphony SP April 2012 QIAsymphony SP/AS User Manual Operating the QIAsymphony SP For use with software version 4.0 Sample & Assay Technologies Trademarks QIAGEN, QIAsymphony (QIAGEN Group); Falcon (Becton, Dickinson

More information

Ultrospec 10 User Manual. English. Deutsch. Français. Español. Italiano

Ultrospec 10 User Manual. English. Deutsch. Français. Español. Italiano English Deutsch Ultrospec 10 User Manual Français Español Italiano Biochrom US Telephone: 1-508-893-8999 84 October Hill Road Toll Free: 1-800-272-2775 Holliston, MA Fax: 1-508-429-5732 01746-1388 support@hbiosci.com

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS Automatic Tissue Processor MTP INS5500GB 2011-07-08 Instructions MTP 1 Instructions MTP 2 CONTENTS 1. INTENDED USE... 4 2. SYMBOLS... 4 3. SAFETY NOTES... 4 4. COMPONENTS... 5 5.

More information

VIAFLO 96 VIAFLO 384. Microplate Pipetting Easy and Affordable

VIAFLO 96 VIAFLO 384. Microplate Pipetting Easy and Affordable VIAFLO 96 VIAFLO 384 Microplate Pipetting Easy and Affordable VITARIS AG Blegistrasse 9 CH-6340 Baar Tel.: 041 769 00 00 Fax: 041 769 00 01 E-mail: info@vitaris.com www.vitaris.com VIAFLO 96 VIAFLO 384

More information

Simple, intuitive, and compact. Liquid Handling Station

Simple, intuitive, and compact. Liquid Handling Station Simple, intuitive, and compact. Liquid Handling Station F I R S T C L A S S B R A N D The name BRAND is synonymous with high precision volumetrics and high-performance life science products. Many years

More information

The Olis Titrator Series

The Olis Titrator Series The Olis Titrator Series Perfect Titrations into any type of cuvette With considerable effort, one can do precise titration studies by hand. Or, one can let the OLIS titrators do the technician s work

More information

Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette

Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette Pipettes allow transfer of accurately known volume of solution from one container to another. Volumetric or transfer pipettes deliver a

More information

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Section 1: Starting an experiment 1.1 Opening a file 1. Open the File menu and select Open Activity. 2. In the Open dialog box, navigate to

More information

ClonePix TM 2 QUICK SET-UP INTRUCTIONS MANUAL SOFTWARE RELEASE

ClonePix TM 2 QUICK SET-UP INTRUCTIONS MANUAL SOFTWARE RELEASE ClonePix TM 2 QUICK SET-UP INTRUCTIONS MANUAL SOFTWARE RELEASE 1.2.80.1071 07MAN1181.A1 Effective Date: 30-Jan-11 ECO #: 3093 Contents What are Quick Set-Up Instructions?... 3 Starting up ClonePix 2...

More information

Thermo Scientific Samco Transfer Pipettes. Precisely what you need

Thermo Scientific Samco Transfer Pipettes. Precisely what you need Thermo Scientific Samco Transfer Pipettes Precisely what you need Application Guide LIFE SCIENCE APPLICATIONS INDUSTRIAL, RESEARCH AND TESTING CLINICAL APPLICATIONS Top 9 Pipettes: 6, 3, 5, 20, 23, 24,

More information

FACA-1240 Auto Chemistry Analyzer User s Manual

FACA-1240 Auto Chemistry Analyzer User s Manual FACA-1240 Auto Chemistry Analyzer User s Manual PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC.VER.01-4.12 I How

More information

BD FACS Sample Prep Assistant III Instructions For Use

BD FACS Sample Prep Assistant III Instructions For Use BD FACS Sample Prep Assistant III Instructions For Use IVD For In Vitro Diagnostic Use bdbiosciences.com 23-13406-02 4/2016 Becton, Dickinson and Company BD Biosciences 2350 Qume Drive, San Jose, CA 95131

More information

Quick Start Guide Axxin T8-ISO. Instrument. In association with: E: W: D Version 1.5

Quick Start Guide Axxin T8-ISO. Instrument. In association with: E: W:   D Version 1.5 Quick Start Guide Axxin T8-ISO Instrument In association with: E: support@axxin.com W: D006093 Version 1.5 Office DOCUMENT CONTROL DETAILS TITLE DOC No. CUSTOMER CONTACT QSG, Quick Start Guide, Axxin T8-ISO

More information

other sizes available on request

other sizes available on request ratiolab Cryo Boxes, Cardboard, Format 133 x 133 mm ratiolab Cryo Boxes, Cardboard, Format 133 x 133 mm Ultra low temperature-resistant cardboard. temperature-resistant down to 80 C available with a water-repellent

More information

INSTRUCTION MANUAL HBPH DIFFERENTIAL 2-WIRE (4-20 ma) ph Sensor

INSTRUCTION MANUAL HBPH DIFFERENTIAL 2-WIRE (4-20 ma) ph Sensor INSTRUCTION MANUAL HBPH DIFFERENTIAL 2-WIRE (4-20 ma) ph Sensor TABLE OF CONTENTS 1. GENERAL INFORMATION 2 2. SPECIFICATIONS 2 3. INSTALLATION 3 3.1. General Instructions 3 3.2. HBPH 4-20 ma Sensor 3 4.

More information

DS2 Automated ELISA System

DS2 Automated ELISA System DS2 Automated ELISA System Operator s Manual For Use with DS-Matrix version 1.24 and above IMPORTANT Please read carefully before using the DS2 Revision History Manual Version Revision Date: Revision F

More information

FACSLyric TM General User instructions

FACSLyric TM General User instructions flowkon Flow Cytometry User Instructions at the University of Konstanz FACSLyric TM General User instructions In BD FACSuite software, you can measure and analyze samples using either assays or experiments.

More information

VIAFLO 96 VIAFLO 384. Microplate Pipetting Easy and Affordable

VIAFLO 96 VIAFLO 384. Microplate Pipetting Easy and Affordable VIAFLO 96 VIAFLO 384 Microplate Pipetting Easy and Affordable VIAFLO 96 VIAFLO 384 Handheld microplate processing Have you reached the limits of productivity with traditional handheld pipettes? Would you

More information

DPP Micro Reader. The DPP Micro Reader is maintenance-free, not configurable by the user and is operated with a single, multi-function button.

DPP Micro Reader. The DPP Micro Reader is maintenance-free, not configurable by the user and is operated with a single, multi-function button. Read this User Guide completely before using the product. Storage conditions: Store between -20 to 80 C (-4 to 176 F) DPP Micro Reader For use under Emergency Use Authorization (EUA) only. For In Vitro

More information

STANDARD OPERATING PROCEDURE TONOPORT V

STANDARD OPERATING PROCEDURE TONOPORT V Page 1 of 24 1. Scope This Standard Operating Procedure (SOP) applies to the staff and students using the Tonoport V in the Pharmacy Practice Resource Unit (PPRU) at the Pharmacy Department, University

More information

RIDASCREEN Aflatoxin M1

RIDASCREEN Aflatoxin M1 R-Biopharm AG Product information Reliable Extensively tested, excellent performance Sensitive Low limit of detection (5 ppt for milk) Easy All reagents ready to use Automatable Options available R-Biopharm

More information

Universal Fit. Pipette Tips.

Universal Fit. Pipette Tips. Universal Fit Pipette Tips www.globescientific.com Universal Fit Pipette Tips Manufactured in a cleanroom using fully automated equipment No silicone additives, lubricants or heavy metals are used during

More information

User's Guide. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500

User's Guide. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500 User's Guide ExStik EC500 ph/conductivity / TDS / Salinity / Temperature Meter EC500 Introduction Congratulations on your purchase of the ExStik EC500 ph/conductivity/total Dissolved Solids (TDS) / Salinity

More information

Your Test Your Workflow Your Freedom. Automated Staining Instrument. Freedom

Your Test Your Workflow Your Freedom. Automated Staining Instrument. Freedom Flexibility Your Test Your Workflow Your Freedom Automated Staining Instrument Freedom www.biocare.net www.slidestainer.com Enjoy the Freedom of Maximum Flexibility & Productivity Reagent Rack Cold Station

More information

USER S MANUAL TULIP DIAGNOSTICS (P) LTD. Gitanjali, Tulip Block, Dr. Antonion De Rego Bagh, Alto Santacruz Bambolim P.O GOA.

USER S MANUAL TULIP DIAGNOSTICS (P) LTD. Gitanjali, Tulip Block, Dr. Antonion De Rego Bagh, Alto Santacruz Bambolim P.O GOA. USER S MANUAL TULIP DIAGNOSTICS (P) LTD. Gitanjali, Tulip Block, Dr. Antonion De Rego Bagh, Alto Santacruz Bambolim P.O GOA. Pin: 403202 Phone: +91 832-2458546/51; Fax: +91 832-2458544; Email: sales@tulipgroup.com

More information

VIAFLO 96 VIAFLO 384 Microplate Pipetting Easy and Affordable

VIAFLO 96 VIAFLO 384 Microplate Pipetting Easy and Affordable VIAFLO 96 VIAFLO 384 Microplate Pipetting Easy and Affordable VIAFLO 96 VIAFLO 384 Handheld microplate processing Have you reached the limits of productivity with traditional handheld pipettes? Would you

More information

Manual HPLC-4 Alliance-UV

Manual HPLC-4 Alliance-UV Manual HPLC-4 Alliance-UV MAKE SURE NEVER TO: 1. pump particles or air through the column. 2. inject when the injector is dry. 3. inject solutions containing particles. 4. apply large (sudden) pressure

More information

CryoClear. Cryogenic Vials. Innovation. for the long-term storage of your precious samples.

CryoClear. Cryogenic Vials. Innovation. for the long-term storage of your precious samples. CryoClear Cryogenic Vials Innovation for the long-term storage of your precious samples. CryoClear Cryogenic Vials Innovation to protect your precious samples. Globe Scientific s CryoClear vials are designed

More information

ClonePix TM FL Quick Set-Up Instructions

ClonePix TM FL Quick Set-Up Instructions ClonePix TM FL Quick Set-Up Instructions Software Release: 1.2.13.971 07MAN1055.A2 Effective Date: 01-Jun-10 QUICK INSTRUCTIONS > CONTENTS Contents What are Quick Set Up Instructions?...3 Starting up ClonePix

More information

FRAGMENT ANALYSIS USING ABI 3730/3730XL HANDBOOK. Fragment Analysis using ABI 3730/3730xL, 7-B406, version Devyser AB, 2018 Page 1 of 11

FRAGMENT ANALYSIS USING ABI 3730/3730XL HANDBOOK. Fragment Analysis using ABI 3730/3730xL, 7-B406, version Devyser AB, 2018 Page 1 of 11 FRAGMENT ANALYSIS USING ABI 3730/3730XL HANDBOOK Devyser AB, 2018 Page 1 of 11 Contents 1. DEV-5 SPECTRAL CALIBRATION... 3 1.1. Preparing DEV-5 Spectral Calibration standards... 3 1.2. Creating a Spectral

More information

IFA AUTOMATION BEYOND HUMAN BOUNDARIES HELMED IFA PROCESSOR

IFA AUTOMATION BEYOND HUMAN BOUNDARIES HELMED IFA PROCESSOR IFA AUTOMATION BEYOND HUMAN BOUNDARIES HELMED IFA PROCESSOR WHAT IS THE HELMED? The HELMED is a fully automated and open immunofluorescence processor capable of performing all IFA steps. It is suitable

More information

Canto II Operation Protocol

Canto II Operation Protocol Canto II Operation Protocol Make sure the following actions have been taken before running your samples. Make sure the waste tank is empty and refilled with 500 ml clean (see page 15 for details) Make

More information

Product Overview. your solutions

Product Overview. your solutions Product Overview your solutions 2 CyBio - Product Overview your solutions Pipettors CyBi -Well Provides precise pipetting in 96-, 384- and 1536-well formats. Available with a variety of stackers and accessories

More information

DNA Shearing with E220 Focused-ultrasonicator.

DNA Shearing with E220 Focused-ultrasonicator. Quick Guide: DNA Shearing with Focused-ultrasonicator. This Quick Guide provides DNA Shearing protocols when using -130, -50, -15, -500, or minitube and a Covaris Focused-ultrasonicator. Revision History

More information

DS2 Automated ELISA System

DS2 Automated ELISA System DS2 Automated ELISA System Operator s Manual Part No. 91000200 For Use with DS-Matrix version 1.24 and above IMPORTANT Please read carefully before using the DS2 Revision History Manual Version Revision

More information

COBAS INTEGRA 400 plus

COBAS INTEGRA 400 plus COBAS INTEGRA 400 plus COBAS INTEGRA 400 plus is designed to consolidate testing and increase efficiency while reducing total running costs of the laboratory. The COBAS INTEGRA systems extensive assay

More information

Introduction: Features:

Introduction: Features: Index Features:... 3 Specifications:... 4 Accessories:... 5 Functions of Keyboard:... 7 Preparation:... 8 ... 8 ... 8 Calibration:... 8 ... 8 ... 9 ...

More information

Sartorius Biohit Picus Electronic Pipettes

Sartorius Biohit Picus Electronic Pipettes Sartorius Biohit Picus Electronic Pipettes Enjoy pipetting Achieve better results Learn to use it in a minute Description Picus electronic pipette is intended to be used in liquid handling applications

More information

Palintest PHOTOMETER 7500 INSTRUMENTS. INST.7500.C Version 3-01/09

Palintest PHOTOMETER 7500 INSTRUMENTS. INST.7500.C Version 3-01/09 Palintest INSTRUMENTS PHOTOMETER 7500 INST.7500.C Version 3-01/09 Your Palintest direct-reading photometer is designed to give long and trouble-free operation. To ensure you get the best out of this photometer,

More information

GloMax 96 Microplate Luminometer

GloMax 96 Microplate Luminometer TECHNICAL MANUAL GloMax 96 Microplate Luminometer Instructions for Use of Products E6501, E6511 and E6521 Revised 1/16 TM278 GloMax 96 Microplate Luminometer All technical literature is available at: www.promega.com/protocols/

More information

Aqueous GPC Instructions - Detailed

Aqueous GPC Instructions - Detailed Aqueous GPC Instructions - Detailed Waste Bottle Columns PL Datastream Injection Port GPC Computer UV Detector GPC Pump Eluent Bottle The Aqueous GPC Set Up RI Detector Important The points below must

More information

TABLE OF CONTENTS. Table of Contents. Table of Figures

TABLE OF CONTENTS. Table of Contents. Table of Figures TABLE OF CONTENTS Table of Contents Table of Figures 1 2 1 Introduction 3 2 Installation 4 2.1 Installation Precautions 4 2.2 Environmental Requirements 4 3 Specifications 4 4 Precautions 4 5 Materials

More information