Imaging of a Large Collection of Human Embryo Using a Super-Parallel MR Microscope

Size: px
Start display at page:

Download "Imaging of a Large Collection of Human Embryo Using a Super-Parallel MR Microscope"

Transcription

1 Magn Reson Med Sci, Vol. 6, No. 3, pp , 2007 MAJOR PAPER Imaging of a Large Collection of Human Embryo Using a Super-Parallel MR Microscope Yoshimasa MATSUDA 1,ShinyaONO 1, Yosuke OTAKE 1, Shinya HANDA 1, Katsumi KOSE 1 *, Tomoyuki HAISHI 2,ShigetoYAMADA 3, Chikako UWABE 3, and Kohei SHIOTA 3 1 Institute of Applied Physics, University of Tsukuba Tennodai, Tsukuba, Ibaraki , Japan 2 MR Technology Inc., Tsukuba, Japan 3 Center for Congenital Anatomy Research, Kyoto University, Kyoto, Japan (Received December 6, 2006; Accepted September 5, 2007) Using 4 and 8-channel super-parallel magnetic resonance (MR) microscopes with a horizontal bore 2.34T superconducting magnet developed for 3-dimensional MR microscopy of the large Kyoto Collection of Human Embryos, we acquired T 1 -weighted 3D images of 1204 embryos at a spatial resolution of (40 mm) 3 to (150 mm) 3 in about 2 years. Similarity of image contrast between the T 1 -weighted images and stained anatomical sections indicated that T 1 -weighted 3D images could be used for an anatomical 3D image database for human embryology. Keywords: embryo, ˆxed specimen, MR microscopy, multi-channel receiver, parallel MRI Introduction The 40,000 50,000 human embryos acquired by Kyoto University from 1961 to are famous andbelievedtobethelargestsuchcollectioninthe world. The uniqueness of such a large collection underlies the desire for the non-destructive 3D measurements of these specimens. Several methods for non-destructive 3D measurement may be applied to chemically ˆxed embryos. Although X-ray computed tomography (CT) is widely used, especially for skeletal systems, the poorly established skeletal systems of these human embryos made its use unsuitable in this case. Neither was ultrasound, used clinically for fetuses and embryos in vivo, useful for visualizing the internal ˆne structure of the chemically ˆxed embryos because of the limited wavelength of clinical ultrasound machines. However, MR microscopy is a very powerful tool for 3D measurement of human embryos because the chemically ˆxed human embryos contain large quantities of mobile or NMR visible protons, which are major components of the formalin preservation uid. Three-dimensional *Corresponding author, Phone: , Fax: , kose@bk.tsukuba.ac.jp measurement of the human embryo structures at high spatial resolution replaces the time-consuming histological sectioning, staining, and 3D reconstruction tasks. Consequently, in 1999, Kyoto and Tsukuba Universities began a project to acquire 3D MR microscopic images of thousands of embryos. We report their methods and results. 4,5 Materials and Methods Humanembryospecimens Human embryo specimens were chemically ˆxed in Bouin's uid and stored in 10z formalin solution. 1 Twelve hundred and four undamaged and normal-appearing specimens, ranging from Carnegie stage (CS) 13 to 23 and corresponding to 28 to 56 days post conception, were selected from the collection (Table). The specimens were transferred from preservation tubes to nuclear magnetic resonance (NMR) sample tubes, the diameters of which were minimized to maximize the signal-to-noise ratio (SNR) of the NMR signal. The specimens were transported in NMR sample tubes ˆlled with formalin solution to avoid mechanical shock and biological contamination. 139

2 140 Y. Matsuda et al. Table. Human embryos selected for 3D magnetic resonance microscopy Carnegie Stage Number of Samples Tube inner diameter (mm) Voxel Size (mm) /7.0 40, /9.0 40, 45, 50, /9.0 45, 50, /9.0 50, 55, /18 120, /18 120, 150 Outer and inner diameters (OD/ID; mm) of the NMR test tubes used for the embryos were 5/4; 8/7; 10/9; 12/10.5; 15/13.5; and 20/18. The total number of embryos was Super-parallel MR microscope Four- and eight-channel super-parallel MR microscopes were developed Each consisted of a 2.34T horizontal bore superconducting magnet (40 cm room temperature bore, 16 cm diameter spherical volume [dsv] homogeneous region), 4- and 8- channel linear array gradient probes, 4- and 8- channel parallel NMR transceivers, an 8-channel radiofrequency (RF) transmitter, a 3-channel gradient driver, and industrial personal computers. Figure 1 shows the 8-channel linear-array gradient probe developed for NMR sample tubes measuring 8 mm in outer diameter. The 4-channel linear array gradient probe with replaceable radiofrequency (RF) coil units was used for measuring specimens stored in NMR sample tubes measuring 10-, 12-, 15-, and 20-mm in outer diameter. The RF coils were solenoids, and the numbers of turns were optimized for the diameters of the sample tubes. The 4-channel parallel NMR transceiver system was built in-house utilizing commercial quadrature detection modules. 9 The 8-channel parallel NMR transceiver was a commercial system (DTRX-Octet-array, MRTechnology-DS Technology, Japan). The detected parallel NMR signals were digitized using 4 or 8 ADC boards (PC-414G3, Datel, USA) installed in one or 2 industrial personal computers running on the Windows 98 operating system. Further details of the system were described previously. 9 Imaging pulse sequences Three-dimensional spin echo or gradient echo imaging pulse sequences were used throughout the experiment. Repetition time (TR) varied from 100 to 2400 ms, and echo time (TE) varied from 8 to 28 ms in the spin echo sequences. In the gradient echo Fig. 1. Eight-channel linear array gradient probe. Size: 14 cm (width) 14 cm (height) 18.5 cm (diameter). Outer diameters for the sample tubes were 8 mm. sequences, TR and TE were ˆxed at 100 ms and 6 ms, and ip angles (FA) were varied from 109to 909. The image matrix for most pulse sequences was ; for some sequences, it was Voxel size varied from (40 mm) 3 to (150 mm) 3. Imaging experiments Embryo specimens were imaged in NMR sample tubes ˆlled with formalin solution, which protected the embryos from desiccation and magnetic susceptibility artifacts. To improve the SNR, NMR sig- Magnetic Resonance in Medical Sciences

3 Imaging of a Large Collection of Human Embryo nals were usually accumulated 16 or 24 times. Several spin echo and gradient echo imaging sequences were used to determine an optimum pulse sequence for visualization of embryo anatomy. The optimum pulse sequence described in the results was used for 3D imaging of the 1204 human embryos. Post processing of the image dataset Acquired 3D image datasets were interpolated from to voxels using a zero-ˆlled Fourier transform technique. Midsagittal cross-sectional images of the embryos were selected after 3D rotation of the interpolated 3D image datasets. Results and Discussion 141 Optimum pulse sequence for visualization of embryo anatomy Figure 2 shows mid-sagittal planes selected from 3D spin echo images of a CS22 human embryo, of which maximum intensity projection (MIP) images are shown in Fig. 3. TE was ˆxed to 8 ms, and TR was varied from 100 to 2400 ms (Fig. 2a-e). To keep the total imaging time constant (11.5 hours), the number of accumulated signals was varied from 24 to one (Fig. 2a-e). In Fig. 2f-h, TR was ˆxed to 800 ms, and TE was varied from 8 to 28 ms. T 1-weighted (T 1W) images are shown in Fig. 2a-b, and T 2-weighted (T 2W) images are shown in Fig. 2g-h. The proton density weighted (PDW) image (Fig. 2e) shows that the density of mobile protons is nearly uniform throughout the embryo Fig. 2. Mid-sagittal planes selected from 3D spin echo images of a CS22 human embryo. Repetition time/echo time were: a: 100/8 ms. b: 200/8 ms. c: 800/8 ms. d: 1200/8 ms. e: 2400/8 ms. f: 1200/8 ms. g: 1200/18 ms. h: 1200/28 ms. Field of view, mm mm mm; image matrix, ; voxel size, (120 mm) 3. Image acquisition times (about 11.5 hours) were identical for all images. Vol. 6 No. 3, 2007

4 142 Y. Matsuda et al. Fig. 3. Maximum intensity projection images of the CS22 human embryo used to image the crosssections shown in Fig. 2. These images were made by the VOLUME-ONE software provided by the VOLUME-ONE developer group and available on the Internet ( specimen, except in the liver. The low intensity of the liver results from its short T 2 ( 10 ms). The results described above demonstrate that the image contrast in the T 1WandT 2Wimagesisdetermined predominantly by T 1 and T 2, respectively. Because the development of the neural system is most important in human embryology, 11 image contrast of the neural system should be optimized. Figure 2 shows that T1W imagesmeetthisrequirement. The contrast-to-noise ratio (CNR) for the vertebrae shown in Fig. 2a was better than that shown in Fig. 2b, so we concluded that Fig. 2a was the best spin echo image for visualization of embryo anatomy. Figure 4 shows mid-sagittal planes selected from 3D gradient echo images of the CS22 human embryo used for the 3D spin echo sequences. The contrastofimageswithsmall ipangles(fig.4a-b)is similartothatofthepdwimageinfig.2e,and the contrast of images with large ip angles (Fig. 4e-f)issimilartothatoftheT1W imageshownin Fig. 2a. To compare the T 1W images obtained by spin echo and gradient echo sequences, the CNR for the vertebrae was calculated by dividing the signal intensity dišerence between the bright and dark regions of the vertebrae by the background noise. The CNR for the gradient echo image (Fig. 4f) was 80z higher than that for the spin echo image (Fig. 2a). However, as shown in Figs. 4 and 5, gradient echo images were quite sensitive to the magnetic susceptibility ešect produced by air bubbles on the surface of the embryo. Air bubbles were inevitable because all the embryo specimens were Fig. 4. Mid-sagittal planes selected from 3D gradient echo images of a CS22 human embryo. Repetition time/echo time: 100 ms/6 ms; ip angles: a: 159. b: 209. c: 309. d: 459. e: 609. f: 909; ˆeldof vision: mm mm mm; image matrix: ; voxel size: (120 mm) 3 ; and number of excitations: 16 for a-f. transferred from the preservation sample tubes to the NMR sample tubes for MR microscopy measurements. We therefore concluded that the 3D spin echo sequence with a TR/TE=100 ms/8 ms should be used to acquire images of the 1204 embryos. MR microscopy of the 1204 embryos Figure 6 shows typical mid-sagittal images selected from 3D image datasets of CS13 to CS23 embryosacquiredwitht 1W sequences. To accelerate the imaging speed, 4 (CS17 23) or 8 (CS13 16) embryos were imaged simultaneously. For the T 1Wsequences, TR was kept at 100 ms and TE was minimized: TE was varied from 8 ms to 12 ms according to the voxel size, which was determined by the achievable gradient ˆeld strength. While the image matrix was ˆxed to , voxel size was varied from (40 mm) 3 to (120 mm) 3 in Fig. 6. The 3D images of the 1204 embryos were acquired from 2003 to Magnetic Resonance in Medical Sciences

5 Imaging of a Large Collection of Human Embryo 143 Fig. 5. Enlarged mid-sagittal head images acquired with a: the spin echo sequence (repetition time/echo time [TR/TE]: 100 ms/8 ms) and b: gradient echo sequence (TR/TE/ ip angle: 100 ms/6 ms/909). Signal loss areas caused by air bubbles in the gradient echo image are indicated by arrows. Fig. 6. Typical mid-sagittal cross sections acquired with T 1 -weighted spin echo sequences and selected from 3D image datasets of CS13 23 embryos. Repetition time was set at 100 ms for all sequences. Echo time was minimized depending on the voxel size. Image matrix was Carnegie stage and voxel size (mm) 3 were: a:13and40;b:14and45;c:15and50;d:16and55;e:17 and 70; f: 18and80;g: 19 and 100; h: 20 and 100; i: 21 and 120; j: 22 and 120; and k: 23 and 120. Figure 7 shows typical mid-sagittal images of CS17 embryos. Although the embryos have various dimensions, the internal structure seems identical. Comparison of image contrast with anatomical sections Figure 8 shows sagittal cross sections of CS23 embryos: Fig. 8a-c depict a T 1-weighted SE image (TR/TE=100 ms/8 ms), a Nissl-stained anatomi- Vol. 6 No. 3, 2007

6 144 Y. Matsuda et al. Fig. 7. Mid-sagittal cross sections selected from 3D image datasets of CS17 embryos acquired with T 1 -weighted spin echo sequences. Repetition time/echo time: 100 ms/12 ms; ˆeld of view: 8.96 mm 8.96 mm mm; image matrix, ; voxel size, (70 mm) 3 ; an number of excitations: 16. Fig. 8. Mid-sagittal cross sections of CS23 embryos. a: T 1 -weighted spin echo image (repetition time/echo time:100 ms/8 ms); b: anatomical section with Nissl staining; 11 c: inverted contrast image of b. cal section as presented in a textbook, 11 andaninverted contrast image of Fig. 8b, respectively. The ˆgure shows that the image contrast of the T 1Wimage is quite similar to that of the stained anatomical section. The image intensities of several organs of the embryos shown in Fig. 8a and 8c are plotted in Fig. 9. The organs selected were tongue, liver, neural tube, Magnetic Resonance in Medical Sciences

7 Imaging of a Large Collection of Human Embryo heart, foot, vertebra (high and low intensity areas), and midbrain. The graph shows that there was a positive correlation (R 2 =0.61) between the set of images, which we interpreted as follows. It is well-known that the image intensity of the stained sections correlates positively with cell density and that longitudinal relaxation rate (1/T 1)correlates positively with intensity of the T1W image. This, and the graph in Fig. 9, suggests that cell density correlates positively with longitudinal relaxation rate. This is reasonable because when cell density increases, motion of the mobile protons among Fig. 9. Correlation between the intensity of the T 1 - weighted image and intensity of the Nissl stain. The correlation is statistically signiˆcant in Pº or in the cells tends to be slow, the spectral components of proton motion at the Larmor frequency (100 MHz) tends to increase, and longitudinal relaxation is prompted if the Larmor frequency is smaller than the inverse of the correlation time of the mobile protons. 12 The last assumption will be valid for most ˆxed specimens because the T 1 of chemically ˆxed specimens is shorter than that of fresh tissues in usual NMR frequencies In summary, the T1-weighted 3D images can be used for the 3D anatomical database of human embryos. Achievable spatial resolution The major purpose of this study was to acquire 3D microscopic images of a large number of human embryos within a limited timeframe. Thus, spatial resolution was not optimized for each embryo. Figure 10 shows a high-resolution mid-sagittal section of a CS22 embryo acquired with a gradient echo sequence (TR/TE=100 ms/6 ms; ip angle [FA]=909; number of excitations [NEX]=12) in 24 hours. The image matrix was , and the voxel size was 50 mm 3. Such a high-resolution image could be acquired for late-stage (CS21 23) embryos if measurement time was considerably extended. As shown by the gradient echo images in Fig. 4, T 2 * for most organs seems longer than 10 ms. This suggests that the spatial resolution is limited only by the SNR of the NMR signal per voxel. Higher magnetic ˆelds or improved RF coils will thus yield higher spatial resolution of the embryo specimens. Conclusions A super-parallel MR microscope with a horizontal bore 2.34T superconducting magnet developed for 3D MR microscopy of a large number of human embryos permitted acquisition of T 1-weighted 3D images of 1204 embryos at the spatial resolution of (40 mm) 3 (150 mm) 3 in about 2 years. The T 1 W image contrast was similar to that of stained anatomical sections, indicating that the T 1 W3Dimages can be used for an anatomical 3D image database for human embryology. In the future, higher spatial resolution will be attained using a higher magnetic ˆeld and improved RF coils. Fig. 10. Mid-sagittal cross-section of a CS22 embryo acquired with a gradient echo sequence (repetition time/echo time: 100 ms/6 ms; ip angle: 909; number of excitations: 8). The image matrix was and the voxel size was (50 mm) 3. Acknowledgments We gratefully acknowledge Professor Yoshiteru Seo at Dokkyo University School of Medicine and Dr. Masae Yaguchi at Keio University for their valuable advice throughout this project. Vol. 6 No. 3, 2007

8 146 Y. Matsuda et al. References 1. Nishimura H, Takano K, Tanimura T, Yasuda M. Normal and abnormal development of human embryos: ˆrst report of the analysis of 1,213 intact embryos. Teratology 1968; 1: Matsunaga E, Shiota K. Holoprosencephaly in human embryos: epidemiologic studies of 150 cases. Teratology 1977; 16: Shiota K. Neural tube defects and maternal hyperthermia in early pregnancy: epidemiology in a human embryo population. Am J Med Genet 1982; 12: Matsuda Y, Ono S, Handa S, Haishi T, Kose K. 3D MR microscopy of a large human embryo collection (Kyoto Collection) to create a 3D image database for human embryology, In: Proceedings of the 12th Annual Meeting of ISMRM, Kyoto, 2004; Matsuda Y, Ono S, Handa S, et al. 3D MR microscopy of a large human embryo collection: ˆnal report for 1,200 human embryos, In: Proceedings of the 13th Annual Meeting of ISMRM, Miami, 2005; Kose K, Haishi T, Matsuda Y, Anno I. Superparallel MR microscope, In: Proceedings of the 9th Annual Meeting of ISMRM, Glasgow, 2001; Matsuda Y, Haishi T, Yamazaki Y, Kose K, Anno I. Development of multi-channel gradient probe for super-parallel MR microscope, In: Proceedings of the 9th Annual Meeting of ISMRM, Glasgow, 2001; Matsuda Y, Utsuzawa S, Kurimoto T, et al. Development of a compact 8-channel parallel MRI console for super-parallel MR microscope, In: Proceedings of the 10th Annual Meeting of ISMRM, Hawaii, 2002; Matsuda Y, Utsuzawa S, Kurimoto T, et al. Superparallel MR microscope. Magn Reson Med 2003; 50: Otake Y, Matsuda Y, Kose K. 8-channel linear array gradient probe for the super-parallel MR microscope, In: Proceedings of the 13th Annual Meeting of ISMRM, Miami, 2005; Drews U. Taschenatlas der embryologie. 1st ed. Stuttgart: Georg Thieme Verlag, Bloembergen N, Purcell EN, Pound RV. Relaxation ešects in nuclear magnetic resonance absorption. Phys Rev 1948; 73: Thickman DI, Kundel HL, Wolf G. Nuclear magnetic resonance characteristics of fresh and ˆxed tissue: the ešect of elapsed time. Radiology 1983; 148: Kamman RL, Go KG, Stomp GP, Hulstaert CE, Berendsen HJ. Changes of relaxation times T 1 and T 2 in rat tissues after biopsy and ˆxation. Magn Reson Imaging 1985; 3: Johnson GA, Benveniste H, Black RD, Hedlund LW, Maronpot RR, Smith BR. Histology by magnetic resonance microscopy. Magn Reson Q 1993; 9:1 30. Magnetic Resonance in Medical Sciences

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 16 Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 3.1 MRI (magnetic resonance imaging) MRI is a technique of measuring physical structure within the human anatomy. Our proposed research focuses

More information

Constrained Reconstruction of Sparse Cardiac MR DTI Data

Constrained Reconstruction of Sparse Cardiac MR DTI Data Constrained Reconstruction of Sparse Cardiac MR DTI Data Ganesh Adluru 1,3, Edward Hsu, and Edward V.R. DiBella,3 1 Electrical and Computer Engineering department, 50 S. Central Campus Dr., MEB, University

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

NMR IMAGING OF SNOW. KEYWORD: NMR, three-dimensional imaging, snow structure, ice sphere, dodecane 1. INTRODUCTUION

NMR IMAGING OF SNOW. KEYWORD: NMR, three-dimensional imaging, snow structure, ice sphere, dodecane 1. INTRODUCTUION NMR IMAGING OF SNOW Toshihiro Ozeki* Ship Research Institute, Mitaka, Japan Akihiro Hachikubo Kitami Institute of Technology, Kitami, Japan Katsumi Kose Institute of Applied Physics, University of Tsukuba,

More information

k-space Interpretation of the Rose Model: Noise Limitation on the Detectable Resolution in MRI

k-space Interpretation of the Rose Model: Noise Limitation on the Detectable Resolution in MRI k-space Interpretation of the Rose Model: Noise Limitation on the Detectable Resolution in MRI Richard Watts and Yi Wang* Magnetic Resonance in Medicine 48:550 554 (2002) Noise limitation on the detected

More information

Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm

Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm L.P. Panych 1,3, B. Madore 1,3, W.S. Hoge 1,3, R.V. Mulkern 2,3 1 Brigham and

More information

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI Figure 1. In the brain, the gray matter has substantially more blood vessels and capillaries than white matter. The magnified image on the right displays the rich vasculature in gray matter forming porous,

More information

3D NMR Imaging of Foam Structures

3D NMR Imaging of Foam Structures JOURNAL OF MAGNETIC RESONANCE, Series A 118, 195 201 (1996) ARTICLE NO. 0027 3D NMR Imaging of Foam Structures KATSUMI KOSE Institute of Applied Physics, University of Tsukuba, Tsukuba, Ibaraki 305, Japan

More information

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space White Pixel Artifact Caused by a noise spike during acquisition Spike in K-space sinusoid in image space Susceptibility Artifacts Off-resonance artifacts caused by adjacent regions with different

More information

Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs

Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Doug Dean Final Project Presentation ENGN 2500: Medical Image Analysis May 16, 2011 Outline Review of the

More information

MRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health

MRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health MRI image formation Indiana University School of Medicine and Indiana University Health Disclosure No conflict of interest for this presentation 2 Outlines Data acquisition Spatial (Slice/Slab) selection

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

New Technology Allows Multiple Image Contrasts in a Single Scan

New Technology Allows Multiple Image Contrasts in a Single Scan These images were acquired with an investigational device. PD T2 T2 FLAIR T1 MAP T1 FLAIR PSIR T1 New Technology Allows Multiple Image Contrasts in a Single Scan MR exams can be time consuming. A typical

More information

Role of Parallel Imaging in High Field Functional MRI

Role of Parallel Imaging in High Field Functional MRI Role of Parallel Imaging in High Field Functional MRI Douglas C. Noll & Bradley P. Sutton Department of Biomedical Engineering, University of Michigan Supported by NIH Grant DA15410 & The Whitaker Foundation

More information

Orthopedic MRI Protocols. Philips Panorama HFO

Orthopedic MRI Protocols. Philips Panorama HFO Orthopedic MRI Protocols Philips Panorama HFO 1 2 Prepared in collaboration with Dr. John F. Feller, Medical Director of Desert Medical Imaging, Palm Springs, CA. Desert Medical Imaging will provide the

More information

MRI Physics II: Gradients, Imaging

MRI Physics II: Gradients, Imaging MRI Physics II: Gradients, Imaging Douglas C., Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor Magnetic Fields in MRI B 0 The main magnetic field. Always on (0.5-7 T) Magnetizes

More information

A Model-Independent, Multi-Image Approach to MR Inhomogeneity Correction

A Model-Independent, Multi-Image Approach to MR Inhomogeneity Correction Tina Memo No. 2007-003 Published in Proc. MIUA 2007 A Model-Independent, Multi-Image Approach to MR Inhomogeneity Correction P. A. Bromiley and N.A. Thacker Last updated 13 / 4 / 2007 Imaging Science and

More information

MR Advance Techniques. Vascular Imaging. Class III

MR Advance Techniques. Vascular Imaging. Class III MR Advance Techniques Vascular Imaging Class III 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

UNIVERSITY OF SOUTHAMPTON

UNIVERSITY OF SOUTHAMPTON UNIVERSITY OF SOUTHAMPTON PHYS2007W1 SEMESTER 2 EXAMINATION 2014-2015 MEDICAL PHYSICS Duration: 120 MINS (2 hours) This paper contains 10 questions. Answer all questions in Section A and only two questions

More information

Computational Medical Imaging Analysis

Computational Medical Imaging Analysis Computational Medical Imaging Analysis Chapter 2: Image Acquisition Systems Jun Zhang Laboratory for Computational Medical Imaging & Data Analysis Department of Computer Science University of Kentucky

More information

Magnetic Resonance Elastography (MRE) of Liver Disease

Magnetic Resonance Elastography (MRE) of Liver Disease Magnetic Resonance Elastography (MRE) of Liver Disease Authored by: Jennifer Dolan Fox, PhD VirtualScopics Inc. jennifer_fox@virtualscopics.com 1-585-249-6231 1. Overview of MRE Imaging MRE is a magnetic

More information

The University of Chicago. Center for EPR Imaging in Vivo Physiology. Image Registration. Boris Epel

The University of Chicago. Center for EPR Imaging in Vivo Physiology. Image Registration. Boris Epel The University of Chicago Center for EPR Imaging in Vivo Physiology Image Registration Boris Epel Imaging Methods are Complimentary CT MRI EPRI High resolution anatomic images Quantitative Poor soft tissue

More information

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street MRI is located in the sub basement of CC wing. From Queen or Victoria, follow the baby blue arrows and ride the CC south

More information

A Transmission Line Matrix Model for Shielding Effects in Stents

A Transmission Line Matrix Model for Shielding Effects in Stents A Transmission Line Matrix Model for hielding Effects in tents Razvan Ciocan (1), Nathan Ida (2) (1) Clemson University Physics and Astronomy Department Clemson University, C 29634-0978 ciocan@clemon.edu

More information

A novel noise removal using homomorphic normalization for multi-echo knee MRI

A novel noise removal using homomorphic normalization for multi-echo knee MRI A novel noise removal using homomorphic normalization for multi-echo knee MRI Xuenan Cui 1a),HakilKim 1b), Seongwook Hong 1c), and Kyu-Sung Kwack 2d) 1 School of Information and Communication Engineering,

More information

Supplementary Information

Supplementary Information Supplementary Information Magnetic resonance imaging reveals functional anatomy and biomechanics of a living dragon tree Linnea Hesse 1,2,*, Tom Masselter 1,2,3, Jochen Leupold 4, Nils Spengler 5, Thomas

More information

Partially Parallel Imaging With Localized Sensitivities (PILS)

Partially Parallel Imaging With Localized Sensitivities (PILS) Partially Parallel Imaging With Localized Sensitivities (PILS) Magnetic Resonance in Medicine 44:602 609 (2000) Mark A. Griswold, 1 * Peter M. Jakob, 1 Mathias Nittka, 1 James W. Goldfarb, 2 and Axel Haase

More information

COBRE Scan Information

COBRE Scan Information COBRE Scan Information Below is more information on the directory structure for the COBRE imaging data. Also below are the imaging parameters for each series. Directory structure: var/www/html/dropbox/1139_anonymized/human:

More information

Super-resolution Reconstruction of Fetal Brain MRI

Super-resolution Reconstruction of Fetal Brain MRI Super-resolution Reconstruction of Fetal Brain MRI Ali Gholipour and Simon K. Warfield Computational Radiology Laboratory Children s Hospital Boston, Harvard Medical School Worshop on Image Analysis for

More information

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Outline High Spectral and Spatial Resolution MR Imaging (HiSS) What it is How to do it Ways to use it HiSS for Radiation

More information

3-D Gradient Shimming

3-D Gradient Shimming 3-D Gradient Shimming 3-D Gradient Shimming on imaging and micro-imaging systems Technical Overview Introduction Advantage statement The GE3DSHIM method is a 3-D gradient shimming procedure developed for

More information

2.1 Signal Production. RF_Coil. Scanner. Phantom. Image. Image Production

2.1 Signal Production. RF_Coil. Scanner. Phantom. Image. Image Production An Extensible MRI Simulator for Post-Processing Evaluation Remi K.-S. Kwan?, Alan C. Evans, and G. Bruce Pike McConnell Brain Imaging Centre, Montreal Neurological Institute, McGill University, Montreal,

More information

surface Image reconstruction: 2D Fourier Transform

surface Image reconstruction: 2D Fourier Transform 2/1/217 Chapter 2-3 K-space Intro to k-space sampling (chap 3) Frequenc encoding and Discrete sampling (chap 2) Point Spread Function K-space properties K-space sampling principles (chap 3) Basic Contrast

More information

We have proposed a new method of image reconstruction in EIT (Electrical Impedance

We have proposed a new method of image reconstruction in EIT (Electrical Impedance Impedance tomography using internal current density distribution measured by nuclear magnetic resonance Eung Je Woo, Soo Yeol Lee, Chi Woong Mun Kon Kuk University, Department of Biomedical Engineering

More information

Lecture 6: Medical imaging and image-guided interventions

Lecture 6: Medical imaging and image-guided interventions ME 328: Medical Robotics Winter 2019 Lecture 6: Medical imaging and image-guided interventions Allison Okamura Stanford University Updates Assignment 3 Due this Thursday, Jan. 31 Note that this assignment

More information

RADIOMICS: potential role in the clinics and challenges

RADIOMICS: potential role in the clinics and challenges 27 giugno 2018 Dipartimento di Fisica Università degli Studi di Milano RADIOMICS: potential role in the clinics and challenges Dr. Francesca Botta Medical Physicist Istituto Europeo di Oncologia (Milano)

More information

Breast MRI Accreditation Program Clinical Image Quality Guide

Breast MRI Accreditation Program Clinical Image Quality Guide Breast MRI Accreditation Program Clinical Image Quality Guide Introduction This document provides guidance on breast MRI clinical image quality and describes the criteria used by the ACR Breast MRI Accreditation

More information

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005 Field Maps John Pauly October 5, 25 The acquisition and reconstruction of frequency, or field, maps is important for both the acquisition of MRI data, and for its reconstruction. Many of the imaging methods

More information

Use of Multicoil Arrays for Separation of Signal from Multiple Slices Simultaneously Excited

Use of Multicoil Arrays for Separation of Signal from Multiple Slices Simultaneously Excited JOURNAL OF MAGNETIC RESONANCE IMAGING 13:313 317 (2001) Technical Note Use of Multicoil Arrays for Separation of Signal from Multiple Slices Simultaneously Excited David J. Larkman, PhD, 1 * Joseph V.

More information

M R I Physics Course

M R I Physics Course M R I Physics Course Multichannel Technology & Parallel Imaging Nathan Yanasak, Ph.D. Jerry Allison Ph.D. Tom Lavin, B.S. Department of Radiology Medical College of Georgia References: 1) The Physics of

More information

divided into structured and unstructured artifacts. Hence unstructured artifact is defined as an random noise, and

divided into structured and unstructured artifacts. Hence unstructured artifact is defined as an random noise, and Improving Quality of MR Images Caused by Ghosting and Noise S. Jayaprakash 1, C. Madhubala 2 1 Head of the Department, CSE Dept., Idhaya Engineering College for Women 2 PG Scholar, CSE Department, Idhaya

More information

A Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data

A Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data A Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data Alexey Samsonov, Julia Velikina Departments of Radiology and Medical

More information

Outline: Contrast-enhanced MRA

Outline: Contrast-enhanced MRA Outline: Contrast-enhanced MRA Background Technique Clinical Indications Future Directions Disclosures: GE Health Care: Research support Consultant: Bracco, Bayer The Basics During rapid IV infusion, Gadolinium

More information

Enhao Gong, PhD Candidate, Electrical Engineering, Stanford University Dr. John Pauly, Professor in Electrical Engineering, Stanford University Dr.

Enhao Gong, PhD Candidate, Electrical Engineering, Stanford University Dr. John Pauly, Professor in Electrical Engineering, Stanford University Dr. Enhao Gong, PhD Candidate, Electrical Engineering, Stanford University Dr. John Pauly, Professor in Electrical Engineering, Stanford University Dr. Greg Zaharchuk, Associate Professor in Radiology, Stanford

More information

Imaging Notes, Part IV

Imaging Notes, Part IV BME 483 MRI Notes 34 page 1 Imaging Notes, Part IV Slice Selective Excitation The most common approach for dealing with the 3 rd (z) dimension is to use slice selective excitation. This is done by applying

More information

A Virtual MR Scanner for Education

A Virtual MR Scanner for Education A Virtual MR Scanner for Education Hackländer T, Schalla C, Trümper A, Mertens H, Hiltner J, Cramer BM Hospitals of the University Witten/Herdecke, Department of Radiology Wuppertal, Germany Purpose A

More information

MRI. When to use What sequences. Outline 2012/09/19. Sequence: Definition. Basic Principles: Step 2. Basic Principles: Step 1. Govind Chavhan, MD

MRI. When to use What sequences. Outline 2012/09/19. Sequence: Definition. Basic Principles: Step 2. Basic Principles: Step 1. Govind Chavhan, MD MRI When to use What sequences Govind Chavhan, MD Assistant Professor and Staff Radiologist The Hospital For Sick Children, Toronto Planning Acquisition Post processing Interpretation Patient history and

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

Denoising the Spectral Information of Non Stationary Image using DWT

Denoising the Spectral Information of Non Stationary Image using DWT Denoising the Spectral Information of Non Stationary Image using DWT Dr.DolaSanjayS 1, P. Geetha Lavanya 2, P.Jagapathi Raju 3, M.Sai Kishore 4, T.N.V.Krishna Priya 5 1 Principal, Ramachandra College of

More information

3-D Compounding of B-Scan Ultrasound Images

3-D Compounding of B-Scan Ultrasound Images 3-D Compounding of B-Scan Ultrasound Images Jochen F. Krücker, Charles R. Meyer, Theresa A. Tuthill, Gerald L. LeCarpentier, J. Brian Fowlkes, Paul L. Carson University of Michigan, Dept. of Radiology,

More information

Biomedical Image Processing for Human Elbow

Biomedical Image Processing for Human Elbow Biomedical Image Processing for Human Elbow Akshay Vishnoi, Sharad Mehta, Arpan Gupta Department of Mechanical Engineering Graphic Era University Dehradun, India akshaygeu001@gmail.com, sharadm158@gmail.com

More information

Accelerated MRI Techniques: Basics of Parallel Imaging and Compressed Sensing

Accelerated MRI Techniques: Basics of Parallel Imaging and Compressed Sensing Accelerated MRI Techniques: Basics of Parallel Imaging and Compressed Sensing Peng Hu, Ph.D. Associate Professor Department of Radiological Sciences PengHu@mednet.ucla.edu 310-267-6838 MRI... MRI has low

More information

Development and Evaluation of a New Method for Measuring of Signal-to-Noise Ratio in Magnetic Resonance Images

Development and Evaluation of a New Method for Measuring of Signal-to-Noise Ratio in Magnetic Resonance Images Development and Evaluation of a New Method for Measuring of Signal-to-Noise Ratio in Magnetic Resonance Images Poster No.: C-0707 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Fukuyama, K. Imai,

More information

A Study of Medical Image Analysis System

A Study of Medical Image Analysis System Indian Journal of Science and Technology, Vol 8(25), DOI: 10.17485/ijst/2015/v8i25/80492, October 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 A Study of Medical Image Analysis System Kim Tae-Eun

More information

Compressed Sensing Reconstructions for Dynamic Contrast Enhanced MRI

Compressed Sensing Reconstructions for Dynamic Contrast Enhanced MRI 1 Compressed Sensing Reconstructions for Dynamic Contrast Enhanced MRI Kevin T. Looby klooby@stanford.edu ABSTRACT The temporal resolution necessary for dynamic contrast enhanced (DCE) magnetic resonance

More information

GE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals.

GE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals. GE Healthcare CLINICAL GALLERY Discovery * MR750w 3.0T This brochure is intended for European healthcare professionals. NEURO PROPELLER delivers high resolution, motion insensitive imaging in all planes.

More information

ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma

ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma ORAL DEFENSE 8 th of September 2017 Charlotte MARTIN Supervisor: Pr. MP REVEL M2 Bio Medical

More information

Diffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford

Diffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford Diffusion MRI Acquisition Karla Miller FMRIB Centre, University of Oxford karla@fmrib.ox.ac.uk Diffusion Imaging How is diffusion weighting achieved? How is the image acquired? What are the limitations,

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

Supplementary methods

Supplementary methods Supplementary methods This section provides additional technical details on the sample, the applied imaging and analysis steps and methods. Structural imaging Trained radiographers placed all participants

More information

design as a constrained maximization problem. In principle, CODE seeks to maximize the b-value, defined as, where

design as a constrained maximization problem. In principle, CODE seeks to maximize the b-value, defined as, where Optimal design of motion-compensated diffusion gradient waveforms Óscar Peña-Nogales 1, Rodrigo de Luis-Garcia 1, Santiago Aja-Fernández 1,Yuxin Zhang 2,3, James H. Holmes 2,Diego Hernando 2,3 1 Laboratorio

More information

Supplementary Figure 1

Supplementary Figure 1 Supplementary Figure 1 BOLD and CBV functional maps showing EPI versus line-scanning FLASH fmri. A. Colored BOLD and CBV functional maps are shown in the highlighted window (green frame) of the raw EPI

More information

Monday, Tuesday, Wednesday, and Thursday, 1 pm to 3 or 4 pm. (See Course Schedule for details)

Monday, Tuesday, Wednesday, and Thursday, 1 pm to 3 or 4 pm. (See Course Schedule for details) Anatomy 6201 Course Director: Dr. Ernesto Phone: (303) 724-3430 Office: RC1 South Rm 11124 Office Hours: by appointment Email: ernesto.salcedo@ucdenver Location ED 2 South Room 2206.! Course Hours Monday,

More information

Fast Imaging Trajectories: Non-Cartesian Sampling (1)

Fast Imaging Trajectories: Non-Cartesian Sampling (1) Fast Imaging Trajectories: Non-Cartesian Sampling (1) M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.05.03 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business

More information

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis Clinical Importance Rapid cardiovascular flow quantitation using sliceselective Fourier velocity encoding with spiral readouts Valve disease affects 10% of patients with heart disease in the U.S. Most

More information

Plane Wave Imaging Using Phased Array Arno Volker 1

Plane Wave Imaging Using Phased Array Arno Volker 1 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic More Info at Open Access Database www.ndt.net/?id=16409 Plane Wave Imaging Using Phased Array

More information

From Simple to Complex: An Incremental Approach to Help Teach Students. the Complex Anatomy of Midgut Rotation in the Human Embryo

From Simple to Complex: An Incremental Approach to Help Teach Students. the Complex Anatomy of Midgut Rotation in the Human Embryo 1 From Simple to Complex: An Incremental Approach to Help Teach Students the Complex Anatomy of Midgut Rotation in the Human Embryo Using Animations and an Interactive Website Takami Iijima June 24, 2008

More information

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE MRES 7005 - Fast Imaging Techniques Module 4 K-Space Symmetry Review K-Space Review K-Space Symmetry Partial or Fractional Echo Half or Partial Fourier HASTE Conditions for successful reconstruction Interpolation

More information

Development of fast imaging techniques in MRI From the principle to the recent development

Development of fast imaging techniques in MRI From the principle to the recent development 980-8575 2-1 2012 10 13 Development of fast imaging techniques in MRI From the principle to the recent development Yoshio MACHIDA and Issei MORI Health Sciences, Tohoku University Graduate School of Medicine

More information

MEDICAL IMAGE ANALYSIS

MEDICAL IMAGE ANALYSIS SECOND EDITION MEDICAL IMAGE ANALYSIS ATAM P. DHAWAN g, A B IEEE Engineering in Medicine and Biology Society, Sponsor IEEE Press Series in Biomedical Engineering Metin Akay, Series Editor +IEEE IEEE PRESS

More information

Exam 8N080 - Introduction MRI

Exam 8N080 - Introduction MRI Exam 8N080 - Introduction MRI Friday January 23 rd 2015, 13.30-16.30h For this exam you may use an ordinary calculator (not a graphical one). In total there are 6 assignments and a total of 65 points can

More information

Computer-Tomography I: Principles, History, Technology

Computer-Tomography I: Principles, History, Technology Computer-Tomography I: Principles, History, Technology Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de History

More information

DEVELOPMENT OF PREVENTIVE MAINTENANCE SYSTEM ARISING IN ADVANCED EDDY CURRENT TESTING USING NETWORK TOMOGRAPHY

DEVELOPMENT OF PREVENTIVE MAINTENANCE SYSTEM ARISING IN ADVANCED EDDY CURRENT TESTING USING NETWORK TOMOGRAPHY DEVELOPMENT OF PREVENTIVE MAINTENANCE SYSTEM ARISING IN ADVANCED EDDY CURRENT TESTING USING NETWORK TOMOGRAPHY F. Kojima 1, F.Kobayashi 1, A. Nishimizu 2, M. Koike 2, T. Matsui 2, H. Endo 3, T. Uchimoto

More information

Single-sided Magnet Diffusion Measurements

Single-sided Magnet Diffusion Measurements Single-sided Magnet Diffusion Measurements Kuldeep S. Panesar April 2013 (kuldeep.panesar@gmail.com) Contents 1. Initial configuration... 2 2. Preparation for diffusion measurement... 5 3. Starting a diffusion

More information

NUS: non-uniform sampling

NUS: non-uniform sampling FMP, 10.10.2012 2/48 NMR-spectroscopy uses the nuclear spin that can be thought of as a mixture between gyroscope and magnet 3/48 The frequency of the rotation of the spin in a magnetic field is what we

More information

Bio-Medical RF Simulations with CST Microwave Studio

Bio-Medical RF Simulations with CST Microwave Studio Bio-Medical RF Simulations with CST Microwave Studio Biological Models Specific Absorption Rate (SAR) Bio-Medical Examples Biological Models The right choice of the biological model is essential for the

More information

The Near Future in Cardiac CT Image Reconstruction

The Near Future in Cardiac CT Image Reconstruction SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de

More information

High dynamic range magnetic resonance flow imaging in the abdomen

High dynamic range magnetic resonance flow imaging in the abdomen High dynamic range magnetic resonance flow imaging in the abdomen Christopher M. Sandino EE 367 Project Proposal 1 Motivation Time-resolved, volumetric phase-contrast magnetic resonance imaging (also known

More information

Following on from the two previous chapters, which considered the model of the

Following on from the two previous chapters, which considered the model of the Chapter 5 Simulator validation Following on from the two previous chapters, which considered the model of the simulation process and how this model was implemented in software, this chapter is concerned

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

Whole Body MRI Intensity Standardization

Whole Body MRI Intensity Standardization Whole Body MRI Intensity Standardization Florian Jäger 1, László Nyúl 1, Bernd Frericks 2, Frank Wacker 2 and Joachim Hornegger 1 1 Institute of Pattern Recognition, University of Erlangen, {jaeger,nyul,hornegger}@informatik.uni-erlangen.de

More information

Information about presenter

Information about presenter Information about presenter 2013-now Engineer R&D ithera Medical GmbH 2011-2013 M.Sc. in Biomedical Computing (TU München) Thesis title: A General Reconstruction Framework for Constrained Optimisation

More information

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT More Info at Open Access Database www.ndt.net/?id=18754 Novel evaluation method of low contrast resolution performance of dimensional X-ray CT Makoto Abe 1, Hiroyuki Fujimoto 1, Osamu Sato 1, Kazuya Matsuzaki

More information

Magnetic Resonance Imaging (MRI)

Magnetic Resonance Imaging (MRI) C. A. Bouman: Digital Image Processing - January 12, 215 1 Magnetic Resonance Imaging (MRI) Can be very high resolution No radiation exposure Very flexible and programable Tends to be expensive, noisy,

More information

Analysis of Functional MRI Timeseries Data Using Signal Processing Techniques

Analysis of Functional MRI Timeseries Data Using Signal Processing Techniques Analysis of Functional MRI Timeseries Data Using Signal Processing Techniques Sea Chen Department of Biomedical Engineering Advisors: Dr. Charles A. Bouman and Dr. Mark J. Lowe S. Chen Final Exam October

More information

General and Efficient Super-Resolution Method for Multi-slice MRI

General and Efficient Super-Resolution Method for Multi-slice MRI General and Efficient Super-Resolution Method for Multi-slice MRI D.H.J. Poot 1,2,V.VanMeir 2, and J. Sijbers 2 1 BIGR, Erasmus Medical Center, Rotterdam 2 Visionlab, University of Antwerp, Antwerp Abstract.

More information

Chapter 4. Clustering Core Atoms by Location

Chapter 4. Clustering Core Atoms by Location Chapter 4. Clustering Core Atoms by Location In this chapter, a process for sampling core atoms in space is developed, so that the analytic techniques in section 3C can be applied to local collections

More information

Computed tomography (Item No.: P )

Computed tomography (Item No.: P ) Computed tomography (Item No.: P2550100) Curricular Relevance Area of Expertise: Biology Education Level: University Topic: Modern Imaging Methods Subtopic: X-ray Imaging Experiment: Computed tomography

More information

Classification of Hyperspectral Breast Images for Cancer Detection. Sander Parawira December 4, 2009

Classification of Hyperspectral Breast Images for Cancer Detection. Sander Parawira December 4, 2009 1 Introduction Classification of Hyperspectral Breast Images for Cancer Detection Sander Parawira December 4, 2009 parawira@stanford.edu In 2009 approximately one out of eight women has breast cancer.

More information

DUE to beam polychromacity in CT and the energy dependence

DUE to beam polychromacity in CT and the energy dependence 1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact

More information

FOCUS. Image Based Automatic Shimming Using B 0 Gradients. Installation and Users Guide Version 0.9

FOCUS. Image Based Automatic Shimming Using B 0 Gradients. Installation and Users Guide Version 0.9 FOCUS Image Based Automatic Shimming Using B 0 Gradients Installation and Users Guide Version 0.9 FOCUS - Field Optimization by Computed Update of Shims Joost A. B. Lohman, Bruker Spectrospin ltd, U.K.

More information

ThE ultimate, INTuITIVE Mr INTErFAcE

ThE ultimate, INTuITIVE Mr INTErFAcE ThE ultimate, INTuITIVE Mr INTErFAcE Empowering you to do more The revolutionary Toshiba M-power user interface takes Mr performance and flexibility to levels higher than ever before. M-power is able to

More information

Single Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction

Single Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction Single Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction Felix Lugauer 1,3, Jens Wetzl 1, Christoph Forman 2, Manuel Schneider 1, Berthold Kiefer

More information

Combination of Parallel Imaging and Compressed Sensing for high acceleration factor at 7T

Combination of Parallel Imaging and Compressed Sensing for high acceleration factor at 7T Combination of Parallel Imaging and Compressed Sensing for high acceleration factor at 7T DEDALE Workshop Nice Loubna EL GUEDDARI (NeuroSPin) Joint work with: Carole LAZARUS, Alexandre VIGNAUD and Philippe

More information

Medical Image Analysis

Medical Image Analysis Computer assisted Image Analysis VT04 29 april 2004 Medical Image Analysis Lecture 10 (part 1) Xavier Tizon Medical Image Processing Medical imaging modalities XRay,, CT Ultrasound MRI PET, SPECT Generic

More information

CHAPTER-1 INTRODUCTION

CHAPTER-1 INTRODUCTION CHAPTER-1 INTRODUCTION 1.1 Fuzzy concept, digital image processing and application in medicine With the advancement of digital computers, it has become easy to store large amount of data and carry out

More information

Non-Cartesian Parallel Magnetic Resonance Imaging

Non-Cartesian Parallel Magnetic Resonance Imaging Non-Cartesian Parallel Magnetic Resonance Imaging Dissertation zur Erlangung des naturwissenschaftlichen Doktorgrades der Bayerischen Julius-Maximilians-Universität Würzburg vorgelegt von Robin Heidemann

More information

TEPZZ A_T EP A1 (19) (11) EP A1 (12) EUROPEAN PATENT APPLICATION. (43) Date of publication: Bulletin 2012/36

TEPZZ A_T EP A1 (19) (11) EP A1 (12) EUROPEAN PATENT APPLICATION. (43) Date of publication: Bulletin 2012/36 (19) (12) EUROPEAN PATENT APPLICATION TEPZZ 4979A_T (11) (43) Date of publication: 0.09.2012 Bulletin 2012/36 (1) Int Cl.: G01R 33/0 (2006.01) G01R 33/44 (2006.01) (21) Application number: 121744.3 (22)

More information