M R I Physics Course

Size: px
Start display at page:

Download "M R I Physics Course"

Transcription

1 M R I Physics Course Multichannel Technology & Parallel Imaging Nathan Yanasak, Ph.D. Jerry Allison Ph.D. Tom Lavin, B.S. Department of Radiology Medical College of Georgia

2 References: 1) The Physics of Clinical MR, for Neuroradiology, Taught Through Images 2) The Physics of Clinical MR, Focusing on the Abdomen, Taught Through Images AUTHORS: VAL M. RUNGE1 MD, WOLFGANG R. NITZ2 PHD, STUART H. SCHMEETS2 BS, RT, WILLIAM H. FAULKNER, JR.3 BS, RT, NILESH K. DESAI1 MD 2

3 Multichannel Coil Technology (basics) Radiological Wish List for MR (and perhaps, other modalities as well): Higher spatial resolution Decreased acquisition time Higher signal to noise ratio (SNR) More images per patient for more diagnostic information Minimize SAR (problem with high-field MR) 3

4 Multichannel Coil (cont.) Some factors to help meet these needs: 1) Protocol/Pulse-sequence optimization 2) Faster image reconstruction hardware but also 3) Single-element coil: increasing SNR requires increased acquisition time 4) Receiving coil element size: decrease increased SNR per volume, but smaller tissue volume 5) Tissue proximity: decrease increased SNR But: Single-element coil + one receive channel = slow dataflow. Solution: Number of RF receive channels: increase decreased acquisition time 4

5 Multichannel Coil (cont.) Circularly-polarized (CP) coil led to a ~ 40% increase in SNR (two-element coil). Also called a quadrature coil. Recent development: Multichannel technology Coil uses multiple elements ( loops ) in phased array with overlapping anatomical coverage Each element acquires MR signals from the entire region. Highest signal in closest proximity to the element. Small element size higher received signal / higher overall signal RF hardware uses multiple channels for receiving signal from multiple elements. 5

6 Multichannel Coil (cont.) Putting the coil and RF system together Signal from each element is (ideally) transferred through its own high-bandwidth RF channel. Reconstruction corrects element-to-element signal variations before forming the final image. Advanced reconstruction and storage hardware necessary to process the rapid inflow of information. 6

7 Multichannel Coil (cont.) Figure 1 shows element arrangement for an 8element head coil, and images acquired from each element, together with the final single combined image. Note that images from each surface element show greater sensitivity near the element. 7

8 The scan illustrated is a fat suppressed FLAIR from a patient with brain metastases, with the edema from a metastasis just superior and posterior to the lateral ventricles visualized. Figure 1 (ref #1) 8

9 Multichannel Coil (cont.) Commercially available, eightchannel coil for brain imaging. (MRI Devices Corporation, Waukesha, WI). We use this coil with our GE 3T. Ref. #1 9

10 Multichannel Coil (cont.) Figure 2 shows element arrangement for a 12element body coil, with six of the elements activated, and images acquired from each element together with the final single combined image. Typically, the scanner software routinely reconstructs the signal from all elements and provides only the combined image. Note that images from each surface element show greater sensitivity near the element. 10

11 The scan sequence employed was truefisp with spectral fat saturation. This image depicts two large liver hemangiomas. Figure 2 (Ref #2) 11

12 Multichannel Coil Example #1 Case study: Axial T1- and T2-weighted images of the brain Figures 3A,C (on left) use a standard CP coil Figures 3B,D (on right) use an eight element, phased array coil MR system uses eight, high-bandwidth, RF receive channels All pulse-sequence parameters same between left and right figures. T1 comparison shows increase in SNR, improved gray versus white matter distinction, and improvement in anatomic definition (e.g., cortical gyri). T2 comparison shows increase in SNR, improved definition of gray matter, and improved visualization of the gray matter nuclei. 12

13 Figures 3A,B,C,D (Ref #1) 13

14 Multichannel Coil (cont.) Clinical benefit of multichannel technology: Higher SNR achieved with multichannel technology allows greater flexibility in sequence parameter selection. If SNR is higher than needed, we can afford to lose a little SNR to gain: An increase in spatial resolution A reduction in acquisition time (e.g., minimize motioninduced artifacts, increase number of images per exam). 14

15 Multichannel Coil (cont.) Advancements in multi-element/multichannel technology (to 32 elements and beyond) will continue to play a role in the development of imaging techniques with higher spatial resolution, faster scan times, and increased diagnostic quality. 1) MCG has an 8-element GE 3T scanner (Sept. 2005) 2) UGA has GE 3T research scanner (8-16 channels summer 2006). 15

16 Multichannel Coil (cont.) Advancements in multi-element/multichannel coils: New 96-channel head coil (Wald, MGH) High-field imaging with 8-channel coil 16

17 Parallel Imaging As shown previously, no image from a single surface coil element is optimally sensitive over the whole area. However, an image reconstructed from all coil elements leads to an increased SNR over a standard acquisition, because each region of the image is reasonably sampled by more than one element. If SNR is higher than needed, one can use the technique of parallel imaging to increase acquisition speed. How? We can decrease sampling of data by each element receiver. Also, reduced sampling less RF excitations per unit time lower SAR. Decrease sampling of data = decreased k-space sampling 17

18 Parallel Imaging Rather than fill all of k-space, parallel imaging acquires a fraction of k-space to save time. Because the anatomy is sampled by multiple coil elements, we can reconstruct the missing information (more or less). Less samples, of course, leads to decreased SNR. But, if our multi-channel SNR is better than we diagnostically require, so what? 18

19 Parallel Imaging How fast can we go? If we have M coil elements covering the FOV, we can skip up to M-1 lines for each line in k-space we sample. The number of lines skipped : acceleration factor (R). This can be fractional as well: # of phase-encodes to cover k-space R = # of phase-encodes used in acquisition Names for acceleration factors: ipat factor (Siemens) SENSE factor (Philips) ASSET factor (GE) Parallel systems: ipat (Siemens) SENSE (Philips) ASSET (GE) 19

20 Parallel Imaging Example #1: We have an 8-element phased-array head coil. We want an acquisition matrix of 256 x 256. What is the maximum acceleration factor we can achieve? Answer: If we have M elements, we can skip up to M-1 lines in k-space. So, M=8, and M-1=7. In the case of this acceleration, for each 8 lines within k-space, we are acquiring only 1 of these line. # of phase-encodes to cover k-space R = # of phase-encodes used in acquisition = 256 phase encodes / { (1 acquired line/8 lines) * 256 } = 8 20

21 Parallel Imaging Increasing acceleration leads to decreasing SNR. However, the benefits may be greater than saving time as well. For EPI images, which are greatly affected by susceptibility differences, parallel imaging can improve geometric distortion and/or image voids. Because the gradients are switching so quickly for an EPI image, one can accrue errors that lead to distortion. These are alleviated using parallel imaging, where the sequence requires less lines in k-space to be read out. 21

22 Parallel Imaging Example of Parallel Acceleration on the GE 3T: R=1 R=2.0 R=2.8 R=3.2 R=4.0 22

23 SNR vs. Acceleration Short-axis cardiac images 32-channel coil 1.5 T magnet Reeder SB et al. MRM 54:748,

24 Parallel Imaging (cont.) Spatial coil sensitivity = function describing the sensitivity of the coil element at any particular position in the FOV. (Ref. #2) R C1 Total C2 L Both types of parallel imaging techniques rely on this function. 24

25 Parallel Imaging (cont.) How is the spatial sensitivity measured? a b c d Method #1: Acquire quick images from each element (a) and reconstruct the full image using all elements (b). Image (a) divided by (b) gives a noisy sensitivity map (c). Filtering smoothes out the noise, yielding our sensitivity map (d). In short, with this method, one must acquire a map before running a parallel imaging sequence. Takes a minute or so. If one uses the summed image from all elements as a reference, this technique is called Auto-SENSE. 25

26 Parallel Imaging (cont.) Method #2: During the parallel scan, we can acquire extra data in the very center of k-space, using the number of phase encodes in this region that we would have used without parallel imaging. Because the center of k-space is responsible for low spatial resolution, this will also give you spatial sensitivity maps for each coil element. (This is the basis of AUTO-SMASH, VD-AUTOSMASH, and GRAPPA ). Key: White=filled part of k-space Black=unfilled k-space 26

27 Parallel Imaging Two main types of parallel imaging: image based reconstruction SENSE, msense, ASSET k-space based reconstruction SMASH, GRAPPA 27

28 Parallel Imaging (Image Based Reconstruction) Image-based reconstruction is, in principle, easier to understand than k-space-based reconstruction. If data is acquired with less phase encodes than will fill k-space, the reconstructed image will show aliasing. We ve seen this before: less phase encodes in the same region of k-space smaller FOV. If FOV is smaller than the object, we get aliasing. (from Boesiger & Pruessmann, # of encodes

29 Parallel Imaging (Image Based Reconstruction) Each pixel in the aliased image (Ialias) is comprised of overlapping (or summed) data from two or more pixels in the unaliased image (I1, I2, ). Use the spatial sensitivity function for each coil element to reconstruct the image intensity uniquely at each position. (from Boesiger & Pruessmann, 29

30 A Simplistic SENSE Example Ialias,1=s1,A IA + s1,b IB A s1 IA Ialias,1 B IB A s2 B Ialias,2 Ialias,2=s2,A IA + s2,b IB We know s1, s2 (sensitivity maps); we measure Ialias,1, Ialias,2; so we can calculate IA, IB. 30

31 Parallel Imaging (k-space Reconstruction) Let s review some topics quickly again: 1) What does k-space really represent (we know that MRI collects data in k-space before reconstructing an image)? 2) What is the relationship between spatial structure in an image and waves? 31

32 Parallel Imaging (k-space basics) Remember: we can decompose a complicated 1-D wave into a combination of simple waves of a given frequency. 32

33 Parallel Imaging (k-space basics) For each simple component, if we know the amplitude and the phase, we can construct a unique wave. Amplitude change Phase change 33

34 Here s the representation of the waveform as a plot of amplitudes and phases: Complicated Wave representation f amplitude/phase representation x phase Can transform back and forth. Now, let s look at 2D waves 34

35 Fourier Transform Basics y Complicated Wave representation K-space representation x Can transform back and forth. K-space is just a 2D (or 3D) version of the amplitude/phase representation. 35

36 Parallel Imaging (k-space method) Back to 1-D again, for simplicity (i.e., the A-P axis) A Coil 1 Coil 2 Coil 3 C2 C3 P Spatial sensitivities for each element of a multi-element coil are periodically distributed across the FOV. This picture shows three spatial sensitivity functions spanning our imaging FOV. So, each element is sensitive to signal in a particular location. C1 36

37 Parallel Imaging (k-space method) We can use each individual coil sensitivity to our advantage. We can examine the signals across a full field of view by combining the signals in some proportion from each coil. Depending on how we combine the signals (add or subtract) from each coil into the total, we can enhance or suppress our sensitivity to signals of different spatial variation. Each combination of coil signals would result in an effective sensitivity across the full FOV. 37

38 Parallel Imaging (k-space method) The black curves represent two of the effective sensitivities using elements in this example. The upper combination is sensitive to Fundamental: CA (add signal for all three) First Harmonic: CB (subtract middle signal) The upper combination is sensitive to these spatial variations: while the lower combination is sensitive to these spatial variations: 38

39 Parallel Imaging (k-space method) So, each combination is sensitive to spatial variations of different wavelength, or spatial harmonics. This modulation of sensitivity across the FOV mimics modulation of spatial sensitivity by phaseencode gradients. If our acceleration factor is R, we have to reconstruct R of these harmonics, by using R combinations of coil signals. 39

40 Parallel Imaging (k-space method) (from Sodickson, et al., MRM 41: 1009, 1999) 40

41 Parallel Imaging (k-space method) How are the missing k-space lines filled in? 1) Spatial sensitivities of each coil are determined. 2) Given an effective sensitivity that we wish to calculate (i.e., sensitivity across the FOV to a sinusoid of a particular periodicity), the actual sensitivities require summation in particular proportions. So, we use the desired effective sensitivity and the measured sensitivities to determine the necessary proportions. 3) The missing k-space lines are calculated by summing k-space data from each coil element, using the proportions determined in the first step. 41

42 Parallel Imaging (k-space method) Example: The fundamental effective sensitivity covers the original lines in k-space; Data for the first harmonic is shifted up a line in k-space. 42

43 Parallel Imaging and Noise Noise in parallel images is 1) increased, and 2) non-uniform. As shown in this SENSE example, unfolding the alias multiplies the noise within particular regions (non-uniform). A similar effect appears in k-space based methods. 43 Larkman DJ et al. Magn Reson Med 2006; 55:

44 Parallel Imaging (k-space Example) K-space oriented parallel acquisition techniques: SMASH (Simultaneous Acquisition of Spatial Harmonics), AUTO-SMASH PILS (Parallel Imaging with Localized Sensitivities) GRAPPA (Generalized Autocalibrating Partially Parallel Acquisition). Figure 5 shows fast spin echo T2 weighted sagittal scans of the lumbar spine, without (A) and with (B) parallel imaging (the latter using GRAPPA). In (B), every second Fourier line has been skipped (an ipat factor of 2, or acceleration factor). Scan time is thus reduced by a factor of two (comparing B to A). 44

45 Figures 5A,B (Ref #2) 45

46 Parallel Imaging (Image Based Reconstruction) Image-based reconstruction parallel acquisition techniques: SENSE (SENSitivity Encoding). Figure 6 shows fast spin echo T2-weighted sagittal scan of the lumbar spine, without (A) and with (B) parallel imaging (using SENSE) In (B) every second Fourier line (parallel imaging with an IPAT factor of 2). Thus the scan time for (B) is half that of (A). Note that there are residual wrap around artifacts (arrow, B), a major drawback to the use of image-based reconstruction technique when anatomy is larger than FOV. 46

47 Figures 6A,B (Ref. #2) 47

48 Parallel Imaging (Drawbacks) Image-based reconstruction: If an aliasing artifact would be present in the chosen FOV for a non-parallel image sequence, then this aliasing will cause reconstruction problems if parallel imaging is attempted. K-space based reconstruction: The ability to construct effective sensitivities from the spatial sensitivities for each coil element depends on the sensitivity profile. This, in turn, depends on the coil element design; therefore, coil design is more critical with this technique. 48

49 GRAPPA (k-space) SENSE (image) 49

50 When Should You Use Parallel MR Imaging? To reduce total scan time To speed up single-shot MRI methods To reduce TE on long echo-train methods To mitigate susceptibility, chemical shift and other artifacts (may cause others) To decrease RF heating (SAR) by minimizing number of RF pulses

51 USE #1: Reduction of SAR in body imaging Case study: Breath-hold T2-weighted abdominal scans. Figure 4A: 17 second T2-weighted breath-hold acquisition, using 29 echoes Figure 4B: 17 second T2-weighted breath-hold acquisition, using 19 echoes. The missing Fourier lines for B were reconstructed using parallel imaging. Use of parallel imaging can be used to reduce the echo train length while keeping scan time the same (SAR reduction). Use of a shorter echo train more slices can be acquired within the same scan time, or overall time can be reduced. 51 Also, less T2-blurring and motion artifact.

52 Figures 4A,B (Ref #2) 52

53 Use #2: Reduction of T2-blurring Parallel imaging reduces T2-blurring because the readout time is shorter resolution is better. Augustin Me et al. Top Magn Reson Imag 2004; 15:207 53

54 Use #3: Reduction of Susceptibility Artifact (EPI) Parallel imaging reduces number of phase-encoding steps required per imaging time Top normal acquisition, Bottom R=2 acceleration

55 Other Current Uses Contrast enhanced MR (e.g., MRA) Improved spatial resolution for a given scan time. Cardiac MRI R=2 6 heartbeats R= 3 4 heartbeats R=4 3 heartbeats

56 (Not-so-distant) Future uses of parallel imaging: 2D acceleration 2D SENSE reconstruction (2X in L-R and 2X in A-P) from an 8channel head array coil conjugated gradient iterative solver after 10 iterations.

Parallel Magnetic Resonance Imaging (pmri): How Does it Work, and What is it Good For?

Parallel Magnetic Resonance Imaging (pmri): How Does it Work, and What is it Good For? Parallel Magnetic Resonance Imaging (pmri): How Does it Work, and What is it Good For? Nathan Yanasak, Ph.D. Chair, AAPM TG118 Department of Radiology Georgia Regents University Overview Phased-array coils

More information

What is pmri? Overview. The Need for Speed: A Technical and Clinical Primer for Parallel MR Imaging 8/1/2011

What is pmri? Overview. The Need for Speed: A Technical and Clinical Primer for Parallel MR Imaging 8/1/2011 The Need for Speed: A Technical and Clinical Primer for Parallel MR Imaging Nathan Yanasak, Ph.D. Chair, AAPM TG118 Assistant Professor Department of Radiology Director, Core Imaging Facility for Small

More information

Parallel Imaging. Marcin.

Parallel Imaging. Marcin. Parallel Imaging Marcin m.jankiewicz@gmail.com Parallel Imaging initial thoughts Over the last 15 years, great progress in the development of pmri methods has taken place, thereby producing a multitude

More information

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street MRI is located in the sub basement of CC wing. From Queen or Victoria, follow the baby blue arrows and ride the CC south

More information

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE MRES 7005 - Fast Imaging Techniques Module 4 K-Space Symmetry Review K-Space Review K-Space Symmetry Partial or Fractional Echo Half or Partial Fourier HASTE Conditions for successful reconstruction Interpolation

More information

Role of Parallel Imaging in High Field Functional MRI

Role of Parallel Imaging in High Field Functional MRI Role of Parallel Imaging in High Field Functional MRI Douglas C. Noll & Bradley P. Sutton Department of Biomedical Engineering, University of Michigan Supported by NIH Grant DA15410 & The Whitaker Foundation

More information

Dynamic Autocalibrated Parallel Imaging Using Temporal GRAPPA (TGRAPPA)

Dynamic Autocalibrated Parallel Imaging Using Temporal GRAPPA (TGRAPPA) Magnetic Resonance in Medicine 53:981 985 (2005) Dynamic Autocalibrated Parallel Imaging Using Temporal GRAPPA (TGRAPPA) Felix A. Breuer, 1 * Peter Kellman, 2 Mark A. Griswold, 1 and Peter M. Jakob 1 Current

More information

Accelerated MRI Techniques: Basics of Parallel Imaging and Compressed Sensing

Accelerated MRI Techniques: Basics of Parallel Imaging and Compressed Sensing Accelerated MRI Techniques: Basics of Parallel Imaging and Compressed Sensing Peng Hu, Ph.D. Associate Professor Department of Radiological Sciences PengHu@mednet.ucla.edu 310-267-6838 MRI... MRI has low

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space White Pixel Artifact Caused by a noise spike during acquisition Spike in K-space sinusoid in image space Susceptibility Artifacts Off-resonance artifacts caused by adjacent regions with different

More information

Scan Acceleration with Rapid Gradient-Echo

Scan Acceleration with Rapid Gradient-Echo Scan Acceleration with Rapid Gradient-Echo Hsiao-Wen Chung ( 鍾孝文 ), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ. Dept. Radiology, Tri-Service General Hospital 1 of 214 The Need

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

Diffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford

Diffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford Diffusion MRI Acquisition Karla Miller FMRIB Centre, University of Oxford karla@fmrib.ox.ac.uk Diffusion Imaging How is diffusion weighting achieved? How is the image acquired? What are the limitations,

More information

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis Clinical Importance Rapid cardiovascular flow quantitation using sliceselective Fourier velocity encoding with spiral readouts Valve disease affects 10% of patients with heart disease in the U.S. Most

More information

Generalized Autocalibrating Partially Parallel Acquisitions (GRAPPA)

Generalized Autocalibrating Partially Parallel Acquisitions (GRAPPA) Magnetic Resonance in Medicine 47:1202 1210 (2002) Generalized Autocalibrating Partially Parallel Acquisitions (GRAPPA) Mark A. Griswold, 1 * Peter M. Jakob, 1 Robin M. Heidemann, 1 Mathias Nittka, 2 Vladimir

More information

(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions

(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions EE225E/BIOE265 Spring 2013 Principles of MRI Miki Lustig Assignment 8 Solutions 1. Nishimura 7.1 P)jc%z 5 ivcr3. 1. I Due Wednesday April 10th, 2013 (a Scrhon5 R2iwd b 0 ZOms Xn,s r cx > qs 4-4 8ni6 4

More information

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005 Field Maps John Pauly October 5, 25 The acquisition and reconstruction of frequency, or field, maps is important for both the acquisition of MRI data, and for its reconstruction. Many of the imaging methods

More information

MRI Physics II: Gradients, Imaging

MRI Physics II: Gradients, Imaging MRI Physics II: Gradients, Imaging Douglas C., Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor Magnetic Fields in MRI B 0 The main magnetic field. Always on (0.5-7 T) Magnetizes

More information

MRI Imaging Options. Frank R. Korosec, Ph.D. Departments of Radiology and Medical Physics University of Wisconsin Madison

MRI Imaging Options. Frank R. Korosec, Ph.D. Departments of Radiology and Medical Physics University of Wisconsin Madison MRI Imaging Options Frank R. Korosec, Ph.D. Departments of Radiolog and Medical Phsics Universit of Wisconsin Madison f.korosec@hosp.wisc.edu As MR imaging becomes more developed, more imaging options

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

COBRE Scan Information

COBRE Scan Information COBRE Scan Information Below is more information on the directory structure for the COBRE imaging data. Also below are the imaging parameters for each series. Directory structure: var/www/html/dropbox/1139_anonymized/human:

More information

Zigzag Sampling for Improved Parallel Imaging

Zigzag Sampling for Improved Parallel Imaging Magnetic Resonance in Medicine 60:474 478 (2008) Zigzag Sampling for Improved Parallel Imaging Felix A. Breuer, 1 * Hisamoto Moriguchi, 2 Nicole Seiberlich, 3 Martin Blaimer, 1 Peter M. Jakob, 1,3 Jeffrey

More information

Basic fmri Design and Analysis. Preprocessing

Basic fmri Design and Analysis. Preprocessing Basic fmri Design and Analysis Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial filtering

More information

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging 1 CS 9 Final Project Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging Feiyu Chen Department of Electrical Engineering ABSTRACT Subject motion is a significant

More information

Motion Artifacts and Suppression in MRI At a Glance

Motion Artifacts and Suppression in MRI At a Glance Motion Artifacts and Suppression in MRI At a Glance Xiaodong Zhong, PhD MR R&D Collaborations Siemens Healthcare MRI Motion Artifacts and Suppression At a Glance Outline Background Physics Common Motion

More information

GE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals.

GE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals. GE Healthcare CLINICAL GALLERY Discovery * MR750w 3.0T This brochure is intended for European healthcare professionals. NEURO PROPELLER delivers high resolution, motion insensitive imaging in all planes.

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

Functional MRI in Clinical Research and Practice Preprocessing

Functional MRI in Clinical Research and Practice Preprocessing Functional MRI in Clinical Research and Practice Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization

More information

Fast Imaging UCLA. Class Business. Class Business. Daniel B. Ennis, Ph.D. Magnetic Resonance Research Labs. Tuesday (3/7) from 6-9pm HW #1 HW #2

Fast Imaging UCLA. Class Business. Class Business. Daniel B. Ennis, Ph.D. Magnetic Resonance Research Labs. Tuesday (3/7) from 6-9pm HW #1 HW #2 Fast Imaging Daniel B. Ennis, Ph.D. Magnetic Resonance Research Labs Class Business Tuesday (3/7) from 6-9pm 6:00-7:30pm Groups Avanto Sara Said, Yara Azar, April Pan Skyra Timothy Marcum, Diana Lopez,

More information

Controlled Aliasing in Parallel Imaging Results in Higher Acceleration (CAIPIRINHA)

Controlled Aliasing in Parallel Imaging Results in Higher Acceleration (CAIPIRINHA) www.siemens.com/magnetom-world Controlled Aliasing in Parallel Imaging Results in Higher Acceleration (CAIPIRINHA) Felix Breuer; Martin Blaimer; Mark Griswold; Peter Jakob Answers for life. Controlled

More information

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial

More information

XI Signal-to-Noise (SNR)

XI Signal-to-Noise (SNR) XI Signal-to-Noise (SNR) Lecture notes by Assaf Tal n(t) t. Noise. Characterizing Noise Noise is a random signal that gets added to all of our measurements. In D it looks like this: while in D

More information

Module 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution.

Module 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution. MRES 7005 - Fast Imaging Techniques Module 5: Dynamic Imaging and Phase Sharing (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review Improving Temporal Resolution True-FISP (I) True-FISP (II) Keyhole

More information

EPI Data Are Acquired Serially. EPI Data Are Acquired Serially 10/23/2011. Functional Connectivity Preprocessing. fmri Preprocessing

EPI Data Are Acquired Serially. EPI Data Are Acquired Serially 10/23/2011. Functional Connectivity Preprocessing. fmri Preprocessing Functional Connectivity Preprocessing Geometric distortion Head motion Geometric distortion Head motion EPI Data Are Acquired Serially EPI Data Are Acquired Serially descending 1 EPI Data Are Acquired

More information

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI Figure 1. In the brain, the gray matter has substantially more blood vessels and capillaries than white matter. The magnified image on the right displays the rich vasculature in gray matter forming porous,

More information

SPM8 for Basic and Clinical Investigators. Preprocessing

SPM8 for Basic and Clinical Investigators. Preprocessing SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial

More information

Partially Parallel Imaging With Localized Sensitivities (PILS)

Partially Parallel Imaging With Localized Sensitivities (PILS) Partially Parallel Imaging With Localized Sensitivities (PILS) Magnetic Resonance in Medicine 44:602 609 (2000) Mark A. Griswold, 1 * Peter M. Jakob, 1 Mathias Nittka, 1 James W. Goldfarb, 2 and Axel Haase

More information

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 16 Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 3.1 MRI (magnetic resonance imaging) MRI is a technique of measuring physical structure within the human anatomy. Our proposed research focuses

More information

VD-AUTO-SMASH Imaging

VD-AUTO-SMASH Imaging Magnetic Resonance in Medicine 45:1066 1074 (2001) VD-AUTO-SMASH Imaging Robin M. Heidemann, Mark A. Griswold, Axel Haase, and Peter M. Jakob* Recently a self-calibrating SMASH technique, AUTO-SMASH, was

More information

Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm

Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm L.P. Panych 1,3, B. Madore 1,3, W.S. Hoge 1,3, R.V. Mulkern 2,3 1 Brigham and

More information

Partial k-space Reconstruction

Partial k-space Reconstruction Chapter 2 Partial k-space Reconstruction 2.1 Motivation for Partial k- Space Reconstruction a) Magnitude b) Phase In theory, most MRI images depict the spin density as a function of position, and hence

More information

MRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health

MRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health MRI image formation Indiana University School of Medicine and Indiana University Health Disclosure No conflict of interest for this presentation 2 Outlines Data acquisition Spatial (Slice/Slab) selection

More information

Breast MRI Accreditation Program Clinical Image Quality Guide

Breast MRI Accreditation Program Clinical Image Quality Guide Breast MRI Accreditation Program Clinical Image Quality Guide Introduction This document provides guidance on breast MRI clinical image quality and describes the criteria used by the ACR Breast MRI Accreditation

More information

Partial k-space Recconstruction

Partial k-space Recconstruction Partial k-space Recconstruction John Pauly September 29, 2005 1 Motivation for Partial k-space Reconstruction a) Magnitude b) Phase In theory, most MRI images depict the spin density as a function of position,

More information

A Virtual MR Scanner for Education

A Virtual MR Scanner for Education A Virtual MR Scanner for Education Hackländer T, Schalla C, Trümper A, Mertens H, Hiltner J, Cramer BM Hospitals of the University Witten/Herdecke, Department of Radiology Wuppertal, Germany Purpose A

More information

divided into structured and unstructured artifacts. Hence unstructured artifact is defined as an random noise, and

divided into structured and unstructured artifacts. Hence unstructured artifact is defined as an random noise, and Improving Quality of MR Images Caused by Ghosting and Noise S. Jayaprakash 1, C. Madhubala 2 1 Head of the Department, CSE Dept., Idhaya Engineering College for Women 2 PG Scholar, CSE Department, Idhaya

More information

Sparse sampling in MRI: From basic theory to clinical application. R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology

Sparse sampling in MRI: From basic theory to clinical application. R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology Sparse sampling in MRI: From basic theory to clinical application R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology Objective Provide an intuitive overview of compressed sensing

More information

MR Advance Techniques. Vascular Imaging. Class III

MR Advance Techniques. Vascular Imaging. Class III MR Advance Techniques Vascular Imaging Class III 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology

More information

MRI. When to use What sequences. Outline 2012/09/19. Sequence: Definition. Basic Principles: Step 2. Basic Principles: Step 1. Govind Chavhan, MD

MRI. When to use What sequences. Outline 2012/09/19. Sequence: Definition. Basic Principles: Step 2. Basic Principles: Step 1. Govind Chavhan, MD MRI When to use What sequences Govind Chavhan, MD Assistant Professor and Staff Radiologist The Hospital For Sick Children, Toronto Planning Acquisition Post processing Interpretation Patient history and

More information

A Study of Nonlinear Approaches to Parallel Magnetic Resonance Imaging

A Study of Nonlinear Approaches to Parallel Magnetic Resonance Imaging University of Wisconsin Milwaukee UWM Digital Commons Theses and Dissertations December 2012 A Study of Nonlinear Approaches to Parallel Magnetic Resonance Imaging Yuchou Chang University of Wisconsin-Milwaukee

More information

SIEMENS MAGNETOM Avanto syngo MR B15

SIEMENS MAGNETOM Avanto syngo MR B15 \\USER\INVESTIGATORS\Ravi\ADNI-Subject\Localizer TA: 0:10 PAT: Voxel size: 1.9 1.5 8.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer Inline movie

More information

Fmri Spatial Processing

Fmri Spatial Processing Educational Course: Fmri Spatial Processing Ray Razlighi Jun. 8, 2014 Spatial Processing Spatial Re-alignment Geometric distortion correction Spatial Normalization Smoothing Why, When, How, Which Why is

More information

MR Imaging Artifacts and Parallel Imaging Techniques with Calibration Scanning: A New Twist on Old Problems 1

MR Imaging Artifacts and Parallel Imaging Techniques with Calibration Scanning: A New Twist on Old Problems 1 IMAGING PHYSICS 532 Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights.

More information

New Technology Allows Multiple Image Contrasts in a Single Scan

New Technology Allows Multiple Image Contrasts in a Single Scan These images were acquired with an investigational device. PD T2 T2 FLAIR T1 MAP T1 FLAIR PSIR T1 New Technology Allows Multiple Image Contrasts in a Single Scan MR exams can be time consuming. A typical

More information

SIEMENS MAGNETOM Skyra syngo MR D13

SIEMENS MAGNETOM Skyra syngo MR D13 Page 1 of 8 SIEMENS MAGNETOM Skyra syngo MR D13 \\USER\CIND\StudyProtocols\PTSA\*dm_ep2d_mono70_b0_p2_iso2.0 TA:1:05 PAT:2 Voxel size:2.0 2.0 2.0 mm Rel. SNR:1.00 :epse Properties Routine Prio Recon Load

More information

Controlled Aliasing in Volumetric Parallel Imaging (2D CAIPIRINHA)

Controlled Aliasing in Volumetric Parallel Imaging (2D CAIPIRINHA) Magnetic Resonance in Medicine 55:549 556 (2006) Controlled Aliasing in Volumetric Parallel Imaging (2D CAIPIRINHA) Felix A. Breuer,* Martin Blaimer, Matthias F. Mueller, Nicole Seiberlich, Robin M. Heidemann,

More information

SIEMENS MAGNETOM Avanto syngo MR B15

SIEMENS MAGNETOM Avanto syngo MR B15 \\USER\INVESTIGATORS\Ravi\ADNI-phantom\QC Phantom-Localizer TA: 0:10 PAT: Voxel size: 1.9 1.5 8.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer

More information

Use of Multicoil Arrays for Separation of Signal from Multiple Slices Simultaneously Excited

Use of Multicoil Arrays for Separation of Signal from Multiple Slices Simultaneously Excited JOURNAL OF MAGNETIC RESONANCE IMAGING 13:313 317 (2001) Technical Note Use of Multicoil Arrays for Separation of Signal from Multiple Slices Simultaneously Excited David J. Larkman, PhD, 1 * Joseph V.

More information

TOF-MRA Using Multi-Oblique-Stack Acquisition (MOSA)

TOF-MRA Using Multi-Oblique-Stack Acquisition (MOSA) JOURNAL OF MAGNETIC RESONANCE IMAGING 26:432 436 (2007) Technical Note TOF-MRA Using Multi-Oblique-Stack Acquisition (MOSA) Ed X. Wu, PhD, 1,2 * Edward S. Hui, BEng, 1,2 and Jerry S. Cheung, BEng 1,2 Purpose:

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

Orthopedic MRI Protocols. Philips Panorama HFO

Orthopedic MRI Protocols. Philips Panorama HFO Orthopedic MRI Protocols Philips Panorama HFO 1 2 Prepared in collaboration with Dr. John F. Feller, Medical Director of Desert Medical Imaging, Palm Springs, CA. Desert Medical Imaging will provide the

More information

G Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine. Compressed Sensing

G Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine. Compressed Sensing G16.4428 Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine Compressed Sensing Ricardo Otazo, PhD ricardo.otazo@nyumc.org Compressed

More information

Fast Imaging Trajectories: Non-Cartesian Sampling (1)

Fast Imaging Trajectories: Non-Cartesian Sampling (1) Fast Imaging Trajectories: Non-Cartesian Sampling (1) M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.05.03 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business

More information

MRI Hot Topics Motion Correction for MR Imaging. medical

MRI Hot Topics Motion Correction for MR Imaging. medical MRI Hot Topics Motion Correction for MR Imaging s medical Motion Correction for MR Imaging Kyle A. Salem, PhD Motion Artifacts in a Clinical Setting Patient motion is probably the most common cause of

More information

3-D Gradient Shimming

3-D Gradient Shimming 3-D Gradient Shimming 3-D Gradient Shimming on imaging and micro-imaging systems Technical Overview Introduction Advantage statement The GE3DSHIM method is a 3-D gradient shimming procedure developed for

More information

Functional MRI. Jerry Allison, Ph. D. Medical College of Georgia

Functional MRI. Jerry Allison, Ph. D. Medical College of Georgia Functional MRI Jerry Allison, Ph. D. Medical College of Georgia BOLD Imaging Technique Blood Oxygen Level Dependent contrast can be used to map brain function Right Hand Motor Task Outline fmri BOLD Contrast

More information

Functional MRI data preprocessing. Cyril Pernet, PhD

Functional MRI data preprocessing. Cyril Pernet, PhD Functional MRI data preprocessing Cyril Pernet, PhD Data have been acquired, what s s next? time No matter the design, multiple volumes (made from multiple slices) have been acquired in time. Before getting

More information

MultiSlice CAIPIRINHA Using View Angle Tilting Technique (CAIPIVAT)

MultiSlice CAIPIRINHA Using View Angle Tilting Technique (CAIPIVAT) RESEARCH ARTICLE MultiSlice CAIPIRINHA Using View Angle Tilting Technique (CAIPIVAT) Min-Oh Kim, Taehwa Hong, and Dong-Hyun Kim Department of Electrical and Electronic Engineering, Yonsei University, Korea.

More information

Removal of EPI Nyquist Ghost Artifacts With Two- Dimensional Phase Correction

Removal of EPI Nyquist Ghost Artifacts With Two- Dimensional Phase Correction Removal of EPI Nyquist Ghost Artifacts With Two- Dimensional Phase Correction Nan-kuei Chen 1,5 and Alice M. Wyrwicz 4 * Magnetic Resonance in Medicine 51:147 153 (004) Odd even echo inconsistencies result

More information

MB-EPI PCASL. Release Notes for Version February 2015

MB-EPI PCASL. Release Notes for Version February 2015 MB-EPI PCASL Release Notes for Version 1.0 20 February 2015 1 Background High-resolution arterial spin labeling (ASL) imaging is highly desirable in both neuroscience research and clinical applications

More information

SIEMENS MAGNETOM Verio syngo MR B15V

SIEMENS MAGNETOM Verio syngo MR B15V \\USER\ZAHID_RESEARCH\MS\No Name\3D SWI TA: 6:39 PAT: 2 Voxel size: 1.0 0.5 2.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer Inline movie Auto

More information

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Outline High Spectral and Spatial Resolution MR Imaging (HiSS) What it is How to do it Ways to use it HiSS for Radiation

More information

Use of MRI in Radiotherapy: Technical Consideration

Use of MRI in Radiotherapy: Technical Consideration Use of MRI in Radiotherapy: Technical Consideration Yanle Hu, PhD Department of Radiation Oncology, Mayo Clinic Arizona 04/07/2018 2015 MFMER slide-1 Conflict of Interest: None 2015 MFMER slide-2 Objectives

More information

8/11/2009. Common Areas of Motion Problem. Motion Compensation Techniques and Applications. Type of Motion. What s your problem

8/11/2009. Common Areas of Motion Problem. Motion Compensation Techniques and Applications. Type of Motion. What s your problem Common Areas of Motion Problem Motion Compensation Techniques and Applications Abdominal and cardiac imaging. Uncooperative patient, such as pediatric. Dynamic imaging and time series. Chen Lin, PhD Indiana

More information

Non-Cartesian Parallel Magnetic Resonance Imaging

Non-Cartesian Parallel Magnetic Resonance Imaging Non-Cartesian Parallel Magnetic Resonance Imaging Dissertation zur Erlangung des naturwissenschaftlichen Doktorgrades der Bayerischen Julius-Maximilians-Universität Würzburg vorgelegt von Robin Heidemann

More information

Supplementary methods

Supplementary methods Supplementary methods This section provides additional technical details on the sample, the applied imaging and analysis steps and methods. Structural imaging Trained radiographers placed all participants

More information

Magnetic Resonance Elastography (MRE) of Liver Disease

Magnetic Resonance Elastography (MRE) of Liver Disease Magnetic Resonance Elastography (MRE) of Liver Disease Authored by: Jennifer Dolan Fox, PhD VirtualScopics Inc. jennifer_fox@virtualscopics.com 1-585-249-6231 1. Overview of MRE Imaging MRE is a magnetic

More information

Correction for EPI Distortions Using Multi-Echo Gradient-Echo Imaging

Correction for EPI Distortions Using Multi-Echo Gradient-Echo Imaging Correction for EPI Distortions Using Multi-Echo Gradient-Echo Imaging Nan-kuei Chen and Alice M. Wyrwicz* Magnetic Resonance in Medicine 41:1206 1213 (1999) A novel and effective technique is described

More information

Combination of Parallel Imaging and Compressed Sensing for high acceleration factor at 7T

Combination of Parallel Imaging and Compressed Sensing for high acceleration factor at 7T Combination of Parallel Imaging and Compressed Sensing for high acceleration factor at 7T DEDALE Workshop Nice Loubna EL GUEDDARI (NeuroSPin) Joint work with: Carole LAZARUS, Alexandre VIGNAUD and Philippe

More information

design as a constrained maximization problem. In principle, CODE seeks to maximize the b-value, defined as, where

design as a constrained maximization problem. In principle, CODE seeks to maximize the b-value, defined as, where Optimal design of motion-compensated diffusion gradient waveforms Óscar Peña-Nogales 1, Rodrigo de Luis-Garcia 1, Santiago Aja-Fernández 1,Yuxin Zhang 2,3, James H. Holmes 2,Diego Hernando 2,3 1 Laboratorio

More information

Unaliasing by Fourier-Encoding the Overlaps Using the Temporal Dimension (UNFOLD), Applied to Cardiac Imaging and fmri

Unaliasing by Fourier-Encoding the Overlaps Using the Temporal Dimension (UNFOLD), Applied to Cardiac Imaging and fmri 1999 ISMRM YOUNG INVESTIGATORS MOORE AWARD PAPERS Magnetic Resonance in Medicine 42:813 828 (1999) Unaliasing by Fourier-Encoding the Overlaps Using the Temporal Dimension (UNFOLD), Applied to Cardiac

More information

SIEMENS MAGNETOM Verio syngo MR B17

SIEMENS MAGNETOM Verio syngo MR B17 \\USER\Dr. Behrmann\routine\Ilan\ep2d_bold_PMU_resting TA: 8:06 PAT: Voxel size: 3.03.03.0 mm Rel. SNR: 1.00 USER: ep2d_bold_pmu Properties Special sat. Prio Recon System Before measurement Body After

More information

Exam 8N080 - Introduction MRI

Exam 8N080 - Introduction MRI Exam 8N080 - Introduction MRI Friday January 23 rd 2015, 13.30-16.30h For this exam you may use an ordinary calculator (not a graphical one). In total there are 6 assignments and a total of 65 points can

More information

Joint SENSE Reconstruction for Faster Multi-Contrast Wave Encoding

Joint SENSE Reconstruction for Faster Multi-Contrast Wave Encoding Joint SENSE Reconstruction for Faster Multi-Contrast Wave Encoding Berkin Bilgic1, Stephen F Cauley1, Lawrence L Wald1, and Kawin Setsompop1 1Martinos Center for Biomedical Imaging, Charlestown, MA, United

More information

This Time. fmri Data analysis

This Time. fmri Data analysis This Time Reslice example Spatial Normalization Noise in fmri Methods for estimating and correcting for physiologic noise SPM Example Spatial Normalization: Remind ourselves what a typical functional image

More information

Advanced Imaging Trajectories

Advanced Imaging Trajectories Advanced Imaging Trajectories Cartesian EPI Spiral Radial Projection 1 Radial and Projection Imaging Sample spokes Radial out : from k=0 to kmax Projection: from -kmax to kmax Trajectory design considerations

More information

Release Notes. Multi-Band EPI C2P. Release HCP_v1 10 February 2014

Release Notes. Multi-Band EPI C2P. Release HCP_v1 10 February 2014 Release Notes Multi-Band EPI C2P Release HCP_v1 10 February 2014 Installation 1. Restart the system (reboot host and MRIR) 2. Extract the.zip file to a temporary directory 3. Run the installer.bat file

More information

Imaging Notes, Part IV

Imaging Notes, Part IV BME 483 MRI Notes 34 page 1 Imaging Notes, Part IV Slice Selective Excitation The most common approach for dealing with the 3 rd (z) dimension is to use slice selective excitation. This is done by applying

More information

SIEMENS MAGNETOM TrioTim syngo MR B17

SIEMENS MAGNETOM TrioTim syngo MR B17 \\USER\KNARRGROUP\MultiBand\LavretskyMultiBand\trufi localizer 3-plane TA: 5.1 s PAT: Voxel size: 1.2 1.2 5. Rel. SNR: 1.00 SIEMENS: trufi Load to stamp Slice group 1 Slices 1 Dist. factor 20 % Phase enc.

More information

LOGIQ. V2 Ultrasound. Part of LOGIQ Vision Series. Imagination at work LOGIQ is a trademark of General Electric Company.

LOGIQ. V2 Ultrasound. Part of LOGIQ Vision Series. Imagination at work LOGIQ is a trademark of General Electric Company. TM LOGIQ V2 Ultrasound Part of LOGIQ Vision Series Imagination at work The brilliance of color. The simplicity of GE. Now you can add the advanced capabilities of color Doppler to patient care with the

More information

Dynamic Contrast enhanced MRA

Dynamic Contrast enhanced MRA Dynamic Contrast enhanced MRA Speaker: Yung-Chieh Chang Date : 106.07.22 Department of Radiology, Taichung Veterans General Hospital, Taichung, Taiwan 1 Outline Basic and advanced principles of Diffusion

More information

Siemens AG, Healthcare Sector. syngo MR D13 0. Supplement - Parameters and image text 0.

Siemens AG, Healthcare Sector. syngo MR D13 0. Supplement - Parameters and image text 0. Siemens AG, Healthcare Sector 0 0 n.a. English Cs2 syngo Neuro Operator MR-05014 630 05/2010 01 02 Informatik, Manual D11 Cape syngo MR D13 0.0 Supplement - Parameters and image text 0. syngo MR D13 0.

More information

MR-Encephalography (MREG)

MR-Encephalography (MREG) MR-Encephalography (MREG) C2P package Version 2.0 For SIEMENS Magnetom Verio/TRIO/Skyra/Prisma Installation and User s Guide NUMARIS/4 VB17, VD11, VD13 Jakob Assländer Pierre LeVan, Ph.D. 10.09.2014 Magnetic

More information

Redundancy Encoding for Fast Dynamic MR Imaging using Structured Sparsity

Redundancy Encoding for Fast Dynamic MR Imaging using Structured Sparsity Redundancy Encoding for Fast Dynamic MR Imaging using Structured Sparsity Vimal Singh and Ahmed H. Tewfik Electrical and Computer Engineering Dept., The University of Texas at Austin, USA Abstract. For

More information

Compressed Sensing for Rapid MR Imaging

Compressed Sensing for Rapid MR Imaging Compressed Sensing for Rapid Imaging Michael Lustig1, Juan Santos1, David Donoho2 and John Pauly1 1 Electrical Engineering Department, Stanford University 2 Statistics Department, Stanford University rapid

More information

Super-resolution Reconstruction of Fetal Brain MRI

Super-resolution Reconstruction of Fetal Brain MRI Super-resolution Reconstruction of Fetal Brain MRI Ali Gholipour and Simon K. Warfield Computational Radiology Laboratory Children s Hospital Boston, Harvard Medical School Worshop on Image Analysis for

More information

Applications Guide for Interleaved

Applications Guide for Interleaved Applications Guide for Interleaved rephase/dephase MRAV Authors: Yongquan Ye, Ph.D. Dongmei Wu, MS. Tested MAGNETOM Systems : 7TZ, TRIO a Tim System, Verio MR B15A (N4_VB15A_LATEST_20070519) MR B17A (N4_VB17A_LATEST_20090307_P8)

More information

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,

More information

Head motion in diffusion MRI

Head motion in diffusion MRI Head motion in diffusion MRI Anastasia Yendiki HMS/MGH/MIT Athinoula A. Martinos Center for Biomedical Imaging 11/06/13 Head motion in diffusion MRI 0/33 Diffusion contrast Basic principle of diffusion

More information

Information about presenter

Information about presenter Information about presenter 2013-now Engineer R&D ithera Medical GmbH 2011-2013 M.Sc. in Biomedical Computing (TU München) Thesis title: A General Reconstruction Framework for Constrained Optimisation

More information