Free-Breathing Whole-Heart Coronary MRA: Motion Compensation Integrated into 3D Cartesian Compressed Sensing Reconstruction

Size: px
Start display at page:

Download "Free-Breathing Whole-Heart Coronary MRA: Motion Compensation Integrated into 3D Cartesian Compressed Sensing Reconstruction"

Transcription

1 Free-Breathing Whole-Heart Coronary MRA: Motion Compensation Integrated into 3D Cartesian Compressed Sensing Reconstruction Christoph Forman 1,2, Robert Grimm 1, Jana Hutter 1,2, Andreas Maier 1,2, Joachim Hornegger 1,2, and Michael O. Zenge 3 1 Pattern Recognition Lab, Department of Computer Science, Friedrich-Alexander Universität Erlangen-Nürnberg, Erlangen, Germany 2 Erlangen Graduate School in Advanced Optical Technologies (SAOT), Friedrich-Alexander Universität Erlangen-Nürnberg, Erlangen, Germany 3 MR Product Innovation & Definition, Healthcare Sector, Siemens AG, Erlangen, Germany Abstract. Respiratory motion remains a major challenge for wholeheart coronary magnetic resonance angiography (CMRA). Recently, iterative reconstruction has been augmented with non-rigid motion compensation to correct for the effects of respiratory motion. The major challenge of this approach is the estimation of dense deformation fields. In this work, the application of such a motion-compensated reconstruction is proposed for accelerated 3D Cartesian whole-heart CMRA. Without the need for extra calibration data or user interaction, the nonrigid deformations due to respiratory motion are directly estimated on the acquired image data. In-vivo experiments on 14 healthy volunteers were performed to compare the proposed method with the result of a navigator-gated reference scan. While reducing the acquisition time by one third, the reconstructed images resulted in equivalent vessel sharpness of 0.44 ± 0.06 mm 1 and 0.45 ± 0.05 mm 1, respectively. 1 Introduction The absence of ionizing radiation renders whole-heart coronary magnetic resonance angiography (CMRA) as an appealing alternative to computed tomography angiography. However, uncompensated motion during the data acquisition remains a challenging problem in CMRA which is not yet fully solved. The effects of cardiac motion are sufficiently minimized by segmented acquisitions triggered with an electrocardiogram (ECG) to end-diastole. For each heart beat, only a restricted amount of data can be acquired during this resting phase of the heart. This prolongs the acquisition time and leaves whole-heart imaging susceptible to respiratory motion. The consequences are reconstruction artifacts, such as blurring or aliasing, which significantly impair an accurate depiction of the coronary arteries. Most commonly, respiratory motion is compensated prospectively by gating the acquisition with a navigator [1]. However, acceptance rates less than 50% extend the scan time unfavorably.

2 2 Forman et al. Therefore, retrospective motion correction approaches were proposed to overcome these disadvantages in scan time efficiency. With self-navigation [2, 3], the position of the heart is detected in readouts acquired in the main direction of the respiratory motion of the heart, namely in the superior-inferior (SI) direction. Although this method provides full efficiency, the respiratory motion detection and correction is limited to a translation in one dimension. Recently, iterative reconstruction was augmented with image registration [4, 5], which incorporates the compensation of non-rigid deformations during data acquisition. Assuming that the acquired data can be segmented into subsets with minimum motion, image registration can be used for the compensation of motion between the individual subsets [6]. Alternatively, the estimation of the motion model has been integrated into the image reconstruction [5, 7] without the need for previously reconstructed images. Although promising results have been demonstrated, most of the MR applications utilizing such approaches were limited to 2D cine imaging [5, 8, 9]. This is mainly due to the lack of sufficient data for estimating a dense deformation field. The first application to 3D whole-heart CMRA has been proposed by Schmidt et al. [10]. However, this approach requires the acquisition of extra calibration data to generate the motion model. In the current work, a motion compensated reconstruction is presented for 3D Cartesian whole-heart CMRA that estimates the non-rigid deformations directly from the acquired image data without the need for extra acquisitions. The proposed approach is evaluated in in-vivo experiments on 14 volunteers and the resulting vessel sharpness is compared to a navigator-gated reference scan. 2 Materials and Methods Free-breathing whole-heart CMRA is performed with data acquisition in segments of readouts distributed over multiple heart beats, henceforth denoted as a shot. The Cartesian variable-density spiral phyllotaxis pattern [11] is used for the sampling within the phase-encoding plane. This pattern provides smooth gradient waveforms for the trajectory of each shot to avoid eddy current effects in the balanced steady-state free-precession (bssfp) data acquisition. Furthermore, the incoherent sampling pattern renders this trajectory well-suited for accelerated Cartesian imaging and compressed sensing reconstruction [12]. As illustrated in Fig. 1, the compensation of respiratory motion in image reconstruction [9] requires 1) the detection of respiratory motion in the acquired data, 2) the generation of a motion model, and finally 3) a reconstruction utilizing this model to estimate the motion-free volume. 2.1 Respiratory Motion Detection In order to facilitate a minimum planning effort and the detection of respiratory motion, data is acquired in sagittal slab orientation to capture the main direction of respiration within each readout. An additional readout through k-space center is performed prior to the acquisition of each shot. The Fourier transform of this

3 Free-Breathing Whole-Heart CMRA 3 Fig. 1. Block diagram of the motion-compensated reconstruction: First, respiratory motion is detected and the acquired data are segmented into motion states. After an intermediate reconstruction of the individual motion states, a motion model is estimated using image registration. Finally, the motion-free volume is reconstructed incorporating the motion model. readout provides an orthogonal projection of the imaging volume. In this projection, the effects of respiratory motion are detected as a discrete 1D translational offset relative to a reference using the self-navigation principle [3]. Subsequently acquired data of each shot is assigned to the detected offset. Uncorrected respiratory motion during data scan leads to inconsistencies in the acquired data, which results in motion artifacts in the reconstructed volumes. Hence, a binning of the data according to the detected offsets into a set of finite motion states denoted by D allows to minimize residual motion within each bin. 2.2 Generation of the Motion Model In the next step, a motion model is generated using image registration and an intermediate reconstruction of the acquired image data, which has been segmented in respiratory motion states. Intermediate Reconstruction First, each volume x d assigned to the detected offsets d D is reconstructed. However, since the acquired data is already undersampled, the sampling rate in each respiratory motion state is further decreased by the binning process. As a consequence, none of the individual motion states contains sufficient data to satisfy the Nyquist criterion. But, compressed sensing allows the reconstruction of such undersampled data, while minimizing incoherent artifacts by enforcing a sparse representation. Thereby, registration errors can be avoided in the estimation of the motion model on basis of the intermediately reconstructed volumes. In order to further support the reconstruction of each volume x d, readouts from adjacent motion states are taken into account. Because the effect of motion on the acquired data increases with the respiratory offset relative to d, the readouts are weighted according to this distance using a scaled Gaussian kernel where the maximum is 1. While µ is set to the detected offset of each bin d, the impact of motion on the reconstructed images is adjusted by σ. Although acquisitions from adjacent bins are considered, some motion states have insufficient data for image reconstruction. Thus, only respiratory states D 0.05 with an sampling rate greater than 0.05 compared to

4 4 Forman et al. the fully sampled k-space are used for further processing. A diagonal matrix W d, containing the estimated weights for each k-space sample and respiratory phase, is directly introduced into the data fidelity term of the cost function of the iterative reconstruction: f( x d ) = argmin W d (A x d y) λ x d TV d D 0.05 (1) x d where the first part of the cost function optimizes the data fidelity of the reconstructed volume, written as a concatenated vector, x d C nxnynz with respect to the measured data y C n kn c with n k acquired k-space samples from n c coil elements. A represents the MR system matrix including the receiver coil sensitivities, the Fourier transform, and the sampling pattern. In the second term, artifacts due to the incoherent sampling are penalized during image reconstruction by enforcing a sparse representation of the estimated volume in the total variation (TV) norm [13]. The described objective function is solved using the limited-memory Broyden-Fletcher-Goldfarb-Shanno (L-BFGS) method [14]. Non-rigid Registration Finally, the reconstructed volumes are used as input for image registration to estimate the motion model. The compensation of nonrigid deformations in the motion-compensated reconstruction requires a dense pixel-wise deformation field and its inverse for every respiratory motion state. Here, both forward and backward deformations are estimated using non-rigid 3D-3D symmetric diffeomorphic registration [15]. The volume reconstructed from the end-expiratory motion state, which contains the most acquisitions, serves as a reference and is consecutively registered to all other volumes. 2.3 Motion Compensated Reconstruction After this, the estimated deformation fields are used to generate the motion matrices U d of size n x n y n z n x n y n z, which are incorporated into the data fidelity term for the reconstruction of the motion-free volume x: f(x) = argmin x d D 0.05 AU d x y d λ x TV, (2) where U d deforms x into the dth motion state to facilitate a comparison to the acquired data in this respiratory phase. 2.4 MR Experiments In-vivo experiments were performed in 14 healthy volunteers (male, age 24.6±5.3) on a 1.5 T clinical MR scanner (MAGNETOM Aera, Siemens AG, Healthcare Sector, Erlangen, Germany), with software release syngo MR D13A. Signal reception was performed using an 18-channel body matrix coil and 8 elements of a spine matrix coil. All measurements were ECG-triggered. The spiral phyllotaxis

5 Free-Breathing Whole-Heart CMRA 5 sampling pattern was generated for the proposed method with 377 shots, and for the navigator-gated reference scan with 233 shots, containing 30 readouts each. Compared to a fully sampled k-space, this results in a sampling ratio of and 0.152, respectively. Data acquisition was performed during the end diastolic phase of the cardiac cycle. At the beginning of each shot, an additional readout for the detection of respiratory motion was acquired. The individual trigger delay to the subject specific cardiac resting period was adjusted using a two-dimensional cine scan with axial slice orientation performed prior to the coronary acquisition. The whole-heart imaging volume was placed in sagittal slab orientation with the blood pool of the heart positioned in the center of the FOV. 3D volume-selective, T2-prepared, fat-saturated bssfp imaging was performed with the following parameters: TR/TE 4.0/2.0 ms, radio frequency (RF) excitation angle 115, FOV mm 3, acquired matrix , reconstructed matrix , slice oversampling 22%, voxel size 1.05 mm 3 and a receiver bandwidth of 849 Hz/Px. For the navigator-gated acquisition an acceptance window with a width of 5 mm was placed in end-expiration. The acquired data of the proposed method was reconstructed with three different methods: M 1 ) using Eq. 1 with W being the identity matrix, M 2 ) with W weighting the input data towards end-expiration with σ = 1.25, and M 3 ) with motion compensation following Eq. 2. The result of all methods were compared to the reconstruction of an additional navigator-gated reference scan (M Ref ). In all cases, the iterative reconstruction was performed with λ = and was terminated after a fixed number of 8 iterations, where no considerable improvement of the fidelity was observed. 2.5 Data Analysis For the assessment of the image quality, vessel sharpness was compared in the resulting volumes obtained from all methods. A centerline was manually segmented for the right coronary artery (RCA) in the 3D isotropic volumes using CoronaViz (Work-in-Progress software, Siemens Corporate Research, Princeton, NJ, USA). Vessel sharpness was evaluated in the proximal and distal part of the RCA. For the proximal evaluation, five points were selected in the volumes obtained from all methods at corresponding positions distributed over the first 40 mm of each centerline. Accordingly, three points were selected in the subsequent 25 mm of the vessel for the distal evaluation. At each measurement point, a cross-section was sampled perpendicular to the centerline. On this cross-section the sharpness was evaluated by the inverse of the average distance between 20% and 80% of the maximum signal intensity on both sides as described in [16]. Finally, vessel sharpness of each part of the coronary vessel was estimated by averaging the results of all cross-sections. 3 Results and Discussion Whole-heart free-breathing acquisitions utilizing the self-navigation principle as well as those with navigator gating were performed successfully in all volunteers.

6 6 Forman et al. (a) (b) (c) Fig. 2. (a) Sagittal slice of an intermediate reconstruction at end-inspiration, midrange, and end-expiration. A temporal profile is plotted along the dashed line over all respiratory states (b) pre and (c) post non-rigid motion compensation. The arrows in these profiles indicate the position of the right coronary artery. The average navigator acceptance rate was 39.5 ± 10.1%, which prolonged the acquisition time to 10.1 ± 2.3 min. Although the sampling pattern for the proposed method was generated for 60% more shots compared to the reference scan, the acquisition time was reduced to 6.3 ± 0.9 min. This means a reduction of the proposed method by 37%. Table 1 shows the qualitative results of vessel sharpness for all methods and volunteers. On average, in M 1 the effects of motion lead to a vessel sharpness of ± mm 1 (proximal) and ± mm 1 (distal). With M 2 the sharpness is improved to 0.426±0.057 mm 1 (proximal) and 0.427±0.043 mm 1 (distal) and with M 3 to ± mm 1 (proximal) and ± mm 1 (distal). A subsequent reference scan M Ref resulted in an equivalent image quality with ± mm 1 (proximal) and ± mm 1 (distal). The respiratory motion detection resulted in 8 to 18 motion states. Fig. 2(a) shows a sagittal slice of the reconstructed intermediate volume weighted to endinspiration, mid-range, and end-expiration of volunteer 5. In this image, a temporal profile along the dashed line over all intermediate volumes is plotted in Fig. 2(b). The arrow indicates the impact of respiration on the right coronary artery. As seen in Fig. 2(c), the effects of respiratory motion were minimized after application of the inverse deformation from the estimated motion model. For this volunteer, Fig. 3 shows the reformatted images of the coronary arteries obtained from all methods. The M 1 reconstruction exhibits motion artifacts, which affects the proximal part of the coronary vessel right after extension from the aorta as indicated by the top arrow in Fig. 3(a). In this region, the RCA can be recovered by weighting the reconstruction to end-expiration in M 2 as shown in Fig. 3(b). However, this weighting introduces an additional sub-sampling on the acquired data. The reduced available data for reconstruction induced a loss of the overall sharpness in the image. Thereby, the coronary vessel is hardly visible in the distal part of the heart as denoted by the bottom arrow. Especially, the vessel sharpness in this subsection of the coronary benefits from incorporating all data in image reconstruction as performed in M 3, Fig. 3(c). While reducing the

7 Free-Breathing Whole-Heart CMRA 7 (a) (b) (c) (d) Fig. 3. The reformatted images of the RCA of volunteer 5 reconstructed with (a) no motion correction, (b) weighted reconstruction with µ set to end-expiration, and (c) motion-compensated reconstruction incorporating the motion model. (d) An additional scan with navigator gating is performed for reference. Table 1. Numerical results for the comparison between the reconstruction of the data without motion compensation (M 1), weighted to end-expiration (M 2), incorporating the motion model (M 3), and of the additional scan with navigator gating (M Ref ). In cases denoted by - the coronary artery was not fully detectable in the distal part proximal M M M M Ref distal M M M M Ref acquisition time, the vessel sharpness of this image showed comparable results as M Ref illustrated in Fig. 3(d). The presented work uniquely demonstrates the estimation of non-rigid displacement fields for motion-compensated reconstruction directly from the image data. Compared to a navigator-gated reference, the proposed approach resulted in an equivalent vessel sharpness, while reducing the acquisition time by one third. Furthermore, the described approach not only renders additional calibration data unnecessary, but also does not require further user interaction. This improvement to the workflow can hardly be overestimated, since CMRA has not been widely used in clinical practice due to the complexity of the examination. Therefore, a high priority should be given to the integration of the proposed method into the clinical routine. Disclaimer: The concepts and information presented in this paper are based on research and are not commercially available. Acknowledgements: The authors gratefully acknowledge funding of the Erlangen Graduate School in Advanced Optical Technologies (SAOT) by the German Research Foundation (DFG) in the framework of the German excellence initiative.

8 8 Forman et al. References 1. Stuber, M., Botnar, R.M., Danias, P.G., Kissinger, K.V., Manning, W.J.: Submillimeter Three-dimensional Coronary MR Angiography with Real-time Navigator Correction: Comparison of Navigator Locations. Radiology 212 (1999) Stehning, C., Börnert, P., Nehrke, K., Eggers, H., Stuber, M.: Free-Breathing Whole-Heart Coronary MRA With 3D Radial SSFP and Self-Navigated Image Reconstruction. Magn Reson Med 54 (2005) Piccini, D., Littmann, A., Nielles-Vallespin, S., Zenge, M.O.: Respiratory Self- Navigation for Whole-Heart Bright-Blood Coronary MRI: Methods for Robust Isolation and Automatic Segmentation of the Blood Pool. Magn Reson Med 68 (2012) Batchelor, P.G., Atkinson, D., Irarrazaval, P., Hill, D.L., Hajnal, J., Larkman, D.: Matrix Description of General Motion Correction Applied to Multishot Images. Magn Reson Med 54 (2005) Odille, F., Vuissoz, P.A., Marie, P.Y., Felbinger, J.: Generalized Reconstruction by Inversion of Coupled Systems (GRICS) Applied to Free-Breathing MRI. Magn Reson Med 60 (2008) Boldea, V., Sharp, G.C., Jiang, S.B., Sarrut, D.: 4D-CT lung motion estimation with deformable registration: Quantification of motion nonlinearity and hysteresis. Med. Phys. 35(3) (2008) Hinkle, J., Szegedi, M., Wang, B., Salter, B., Joshi, S.: 4D CT image reconstruction with diffeomorphic motion model. Med. Imag. Anal. 16 (2012) Hui, X., Ding, Y., Guetter, C., Jolly, M.P., Guehring, J., Zuehlsdorf, S., Simonetti, O.P.: Motion Compensated Magnetic Resonance Reconstruction Using Inverse- Consistent Deformable Registration: Application to Real-Time Cine Imaging. In: MICCAI. (2011) Usman, M., Atkinson, D., Odille, F., Kolbitsch, C., Vaillant, G., Schaeffter, T., Batchelor, P.G., Prieto, C.: Motion Corrected Compressed Sensing for Free- Breathing Dynamic Cardiac MRI. Magn Reson Med (online) 10. Schmidt, J.F.M., Buehrer, M., Boesiger, P., Kozerke, S.: Nonrigid Retrospective Respiratory Motion Correction in Whole-Heart Coronary MRA. Magn Reson Med 66 (2011) Forman, C., Piccini, D., Hutter, J., Grimm, R., Hornegger, J., Zenge, M.O.: Accelerated Whole-Heart Coronary Imaging using Multiple Breath-holds and Compressed Sensing Monitored by Self-Navigation. In: Proc. ISMRM 20. (2012) Lustig, M., Donoho, D., Pauly, J.M.: Sparse MRI: The Application of Compressed Sensing for Rapid MR Imaging. Magn Reson Med 58 (2007) Rudin, L.I., Osher, S., Fatemi, E.: Nonlinear total variation based noise removal algorithms. Physica D 60 (1992) Nocedal, J.: Updating Quasi-Newton Matrices with Limited Storage. Math. Comp. 35 (1980) Avants, B.B., Epstein, C.L., Grossman, M., Gee, J.C.: Symmetric diffeomorphic image registration with cross-correlation: Evaluating automated labeling of elderly and neurodegenerative brain. Med. Imag. Anal. 12 (2008) Li, D., Carr, J.C., Shea, S.M., Zheng, J., Deshpande, V., Wielopolski, P.A., Finn, J.P.: Coronary Arteries: Magnetization-prepared Contrast-enhanced Threedimensional Volume-targeted Breath-hold MR Angiography. Radiology 219 (2001)

1 Pattern Recognition Lab, Department of Computer Science, Friedrich-Alexander University of Erlangen-Nuremberg, Erlangen, Germany

1 Pattern Recognition Lab, Department of Computer Science, Friedrich-Alexander University of Erlangen-Nuremberg, Erlangen, Germany High-Resolution 3D Whole-Heart Coronary MRA: A Study on the Combination of Data Acquisition in Multiple Breath-Holds and 1D Residual Respiratory Motion Compensation Christoph Forman 1,2, Davide Piccini

More information

Total Variation Regularization Method for 3D Rotational Coronary Angiography

Total Variation Regularization Method for 3D Rotational Coronary Angiography Total Variation Regularization Method for 3D Rotational Coronary Angiography Haibo Wu 1,2, Christopher Rohkohl 1,3, Joachim Hornegger 1,2 1 Pattern Recognition Lab (LME), Department of Computer Science,

More information

Total Variation Regularization Method for 3-D Rotational Coronary Angiography

Total Variation Regularization Method for 3-D Rotational Coronary Angiography Total Variation Regularization Method for 3-D Rotational Coronary Angiography Haibo Wu 1,2, Christopher Rohkohl 1,3, Joachim Hornegger 1,2 1 Pattern Recognition Lab (LME), Department of Computer Science,

More information

Compressed Sensing for Rapid MR Imaging

Compressed Sensing for Rapid MR Imaging Compressed Sensing for Rapid Imaging Michael Lustig1, Juan Santos1, David Donoho2 and John Pauly1 1 Electrical Engineering Department, Stanford University 2 Statistics Department, Stanford University rapid

More information

Fast Isotropic Volumetric Coronary MR Angiography Using Free-Breathing 3D Radial Balanced FFE Acquisition

Fast Isotropic Volumetric Coronary MR Angiography Using Free-Breathing 3D Radial Balanced FFE Acquisition Fast Isotropic Volumetric Coronary MR Angiography Using Free-Breathing 3D Radial Balanced FFE Acquisition C. Stehning, 1 * P. Börnert, 2 K. Nehrke, 2 H. Eggers, 2 and O. Dössel 1 Magnetic Resonance in

More information

Single Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction

Single Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction Single Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction Felix Lugauer 1,3, Jens Wetzl 1, Christoph Forman 2, Manuel Schneider 1, Berthold Kiefer

More information

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space White Pixel Artifact Caused by a noise spike during acquisition Spike in K-space sinusoid in image space Susceptibility Artifacts Off-resonance artifacts caused by adjacent regions with different

More information

G Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine. Compressed Sensing

G Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine. Compressed Sensing G16.4428 Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine Compressed Sensing Ricardo Otazo, PhD ricardo.otazo@nyumc.org Compressed

More information

Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data

Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data Christopher M Rank 1, Thorsten Heußer 1, Andreas Wetscherek 1, and Marc Kachelrieß 1 1 German Cancer

More information

Highly Efficient Respiratory Motion Compensated Free-Breathing Coronary MRA Using Golden-Step Cartesian Acquisition

Highly Efficient Respiratory Motion Compensated Free-Breathing Coronary MRA Using Golden-Step Cartesian Acquisition JOURNAL OF MAGNETIC RESONANCE IMAGING 41:738 746 (2015) Technical Development Highly Efficient Respiratory Motion Compensated Free-Breathing Coronary MRA Using Golden-Step Cartesian Acquisition Claudia

More information

Respiratory Motion Estimation using a 3D Diaphragm Model

Respiratory Motion Estimation using a 3D Diaphragm Model Respiratory Motion Estimation using a 3D Diaphragm Model Marco Bögel 1,2, Christian Riess 1,2, Andreas Maier 1, Joachim Hornegger 1, Rebecca Fahrig 2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg 2

More information

Redundancy Encoding for Fast Dynamic MR Imaging using Structured Sparsity

Redundancy Encoding for Fast Dynamic MR Imaging using Structured Sparsity Redundancy Encoding for Fast Dynamic MR Imaging using Structured Sparsity Vimal Singh and Ahmed H. Tewfik Electrical and Computer Engineering Dept., The University of Texas at Austin, USA Abstract. For

More information

Iterative CT Reconstruction Using Curvelet-Based Regularization

Iterative CT Reconstruction Using Curvelet-Based Regularization Iterative CT Reconstruction Using Curvelet-Based Regularization Haibo Wu 1,2, Andreas Maier 1, Joachim Hornegger 1,2 1 Pattern Recognition Lab (LME), Department of Computer Science, 2 Graduate School in

More information

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street MRI is located in the sub basement of CC wing. From Queen or Victoria, follow the baby blue arrows and ride the CC south

More information

Spatial-temporal Total Variation Regularization (STTVR) for 4D-CT Reconstruction

Spatial-temporal Total Variation Regularization (STTVR) for 4D-CT Reconstruction Spatial-temporal Total Variation Regularization (STTVR) for 4D-CT Reconstruction Haibo Wu a, b, Andreas Maier a, Rebecca Fahrig c, and Joachim Hornegger a, b a Pattern Recognition Lab (LME), Department

More information

Module 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution.

Module 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution. MRES 7005 - Fast Imaging Techniques Module 5: Dynamic Imaging and Phase Sharing (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review Improving Temporal Resolution True-FISP (I) True-FISP (II) Keyhole

More information

Parallel Imaging. Marcin.

Parallel Imaging. Marcin. Parallel Imaging Marcin m.jankiewicz@gmail.com Parallel Imaging initial thoughts Over the last 15 years, great progress in the development of pmri methods has taken place, thereby producing a multitude

More information

Respiratory Motion Compensation for C-arm CT Liver Imaging

Respiratory Motion Compensation for C-arm CT Liver Imaging Respiratory Motion Compensation for C-arm CT Liver Imaging Aline Sindel 1, Marco Bögel 1,2, Andreas Maier 1,2, Rebecca Fahrig 3, Joachim Hornegger 1,2, Arnd Dörfler 4 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg

More information

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis Clinical Importance Rapid cardiovascular flow quantitation using sliceselective Fourier velocity encoding with spiral readouts Valve disease affects 10% of patients with heart disease in the U.S. Most

More information

Dynamic Autocalibrated Parallel Imaging Using Temporal GRAPPA (TGRAPPA)

Dynamic Autocalibrated Parallel Imaging Using Temporal GRAPPA (TGRAPPA) Magnetic Resonance in Medicine 53:981 985 (2005) Dynamic Autocalibrated Parallel Imaging Using Temporal GRAPPA (TGRAPPA) Felix A. Breuer, 1 * Peter Kellman, 2 Mark A. Griswold, 1 and Peter M. Jakob 1 Current

More information

Motion Artifacts and Suppression in MRI At a Glance

Motion Artifacts and Suppression in MRI At a Glance Motion Artifacts and Suppression in MRI At a Glance Xiaodong Zhong, PhD MR R&D Collaborations Siemens Healthcare MRI Motion Artifacts and Suppression At a Glance Outline Background Physics Common Motion

More information

Basic fmri Design and Analysis. Preprocessing

Basic fmri Design and Analysis. Preprocessing Basic fmri Design and Analysis Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial filtering

More information

Functional MRI in Clinical Research and Practice Preprocessing

Functional MRI in Clinical Research and Practice Preprocessing Functional MRI in Clinical Research and Practice Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization

More information

Role of Parallel Imaging in High Field Functional MRI

Role of Parallel Imaging in High Field Functional MRI Role of Parallel Imaging in High Field Functional MRI Douglas C. Noll & Bradley P. Sutton Department of Biomedical Engineering, University of Michigan Supported by NIH Grant DA15410 & The Whitaker Foundation

More information

SPM8 for Basic and Clinical Investigators. Preprocessing

SPM8 for Basic and Clinical Investigators. Preprocessing SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial

More information

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial

More information

Depth-Layer-Based Patient Motion Compensation for the Overlay of 3D Volumes onto X-Ray Sequences

Depth-Layer-Based Patient Motion Compensation for the Overlay of 3D Volumes onto X-Ray Sequences Depth-Layer-Based Patient Motion Compensation for the Overlay of 3D Volumes onto X-Ray Sequences Jian Wang 1,2, Anja Borsdorf 2, Joachim Hornegger 1,3 1 Pattern Recognition Lab, Friedrich-Alexander-Universität

More information

The Impact of Navigator Timing Parameters and Navigator Spatial Resolution on 3D Coronary Magnetic Resonance Angiography

The Impact of Navigator Timing Parameters and Navigator Spatial Resolution on 3D Coronary Magnetic Resonance Angiography JOURNAL OF MAGNETIC RESONANCE IMAGING 14:311 318 (2001) Technical Note The Impact of Navigator Timing Parameters and Navigator Spatial Resolution on 3D Coronary Magnetic Resonance Angiography Elmar Spuentrup,

More information

Make the Most of Time

Make the Most of Time Make the Most of Time Temporal Extension of the itv Algorithm for 4D Cardiac C-Arm CT Viktor Haase 1,5, Oliver Taubmann 2,3, Yixing Huang 2, Gregor Krings 4, Günter Lauritsch 5, Andreas Maier 2,3, Alfred

More information

Spiral keyhole imaging for MR fingerprinting

Spiral keyhole imaging for MR fingerprinting Spiral keyhole imaging for MR fingerprinting Guido Buonincontri 1, Laura Biagi 1,2, Pedro A Gómez 3,4, Rolf F Schulte 4, Michela Tosetti 1,2 1 IMAGO7 Research Center, Pisa, Italy 2 IRCCS Stella Maris,

More information

Self-Navigation with Compressed Sensing for 2D Translational Motion Correction in Free-Breathing Coronary MRI: A Feasibility Study

Self-Navigation with Compressed Sensing for 2D Translational Motion Correction in Free-Breathing Coronary MRI: A Feasibility Study Self-Navigation with Compressed Sensing for 2D Translational Motion Correction in Free-Breathing Coronary MRI: A Feasibility Study Gabriele Bonanno 1,2, Gilles Puy 3, Yves Wiaux 1,2,3,4,5, Ruud B. van

More information

Zigzag Sampling for Improved Parallel Imaging

Zigzag Sampling for Improved Parallel Imaging Magnetic Resonance in Medicine 60:474 478 (2008) Zigzag Sampling for Improved Parallel Imaging Felix A. Breuer, 1 * Hisamoto Moriguchi, 2 Nicole Seiberlich, 3 Martin Blaimer, 1 Peter M. Jakob, 1,3 Jeffrey

More information

Compressed-Sensing Motion Compensation (CosMo): A Joint Prospective Retrospective Respiratory Navigator for Coronary MRI

Compressed-Sensing Motion Compensation (CosMo): A Joint Prospective Retrospective Respiratory Navigator for Coronary MRI Magnetic Resonance in Medicine 66:1674 1681 (2011) Compressed-Sensing Motion Compensation (CosMo): A Joint Prospective Retrospective Respiratory Navigator for Coronary MRI Mehdi H. Moghari, 1 Mehmet Akçakaya,

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

EPI Data Are Acquired Serially. EPI Data Are Acquired Serially 10/23/2011. Functional Connectivity Preprocessing. fmri Preprocessing

EPI Data Are Acquired Serially. EPI Data Are Acquired Serially 10/23/2011. Functional Connectivity Preprocessing. fmri Preprocessing Functional Connectivity Preprocessing Geometric distortion Head motion Geometric distortion Head motion EPI Data Are Acquired Serially EPI Data Are Acquired Serially descending 1 EPI Data Are Acquired

More information

King s Research Portal

King s Research Portal King s Research Portal DOI: 10.1016/j.mri.2016.12.021 Document Version Peer reviewed version Link to publication record in King's Research Portal Citation for published version (APA): Usman, M., Ruijsink,

More information

K-Space Trajectories and Spiral Scan

K-Space Trajectories and Spiral Scan K-Space and Spiral Scan Presented by: Novena Rangwala nrangw2@uic.edu 1 Outline K-space Gridding Reconstruction Features of Spiral Sampling Pulse Sequences Mathematical Basis of Spiral Scanning Variations

More information

Accelerated MRI Techniques: Basics of Parallel Imaging and Compressed Sensing

Accelerated MRI Techniques: Basics of Parallel Imaging and Compressed Sensing Accelerated MRI Techniques: Basics of Parallel Imaging and Compressed Sensing Peng Hu, Ph.D. Associate Professor Department of Radiological Sciences PengHu@mednet.ucla.edu 310-267-6838 MRI... MRI has low

More information

Retrospective Respiratory Motion Correction for Navigated Cine Velocity Mapping

Retrospective Respiratory Motion Correction for Navigated Cine Velocity Mapping JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE 1 Vol. 6, No. 4, pp. 785 792, 2004 VELOCITY MAPPING Retrospective Respiratory Motion Correction for Navigated Cine Velocity Mapping Christof Baltes, 1 Sebastian

More information

Development of fast imaging techniques in MRI From the principle to the recent development

Development of fast imaging techniques in MRI From the principle to the recent development 980-8575 2-1 2012 10 13 Development of fast imaging techniques in MRI From the principle to the recent development Yoshio MACHIDA and Issei MORI Health Sciences, Tohoku University Graduate School of Medicine

More information

NIH Public Access Author Manuscript Med Phys. Author manuscript; available in PMC 2009 March 13.

NIH Public Access Author Manuscript Med Phys. Author manuscript; available in PMC 2009 March 13. NIH Public Access Author Manuscript Published in final edited form as: Med Phys. 2008 February ; 35(2): 660 663. Prior image constrained compressed sensing (PICCS): A method to accurately reconstruct dynamic

More information

Scaling Calibration in the ATRACT Algorithm

Scaling Calibration in the ATRACT Algorithm Scaling Calibration in the ATRACT Algorithm Yan Xia 1, Andreas Maier 1, Frank Dennerlein 2, Hannes G. Hofmann 1, Joachim Hornegger 1,3 1 Pattern Recognition Lab (LME), Friedrich-Alexander-University Erlangen-Nuremberg,

More information

Fast Imaging Trajectories: Non-Cartesian Sampling (1)

Fast Imaging Trajectories: Non-Cartesian Sampling (1) Fast Imaging Trajectories: Non-Cartesian Sampling (1) M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.05.03 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business

More information

Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT

Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT Marcus Brehm 1,2, Thorsten Heußer 1, Pascal Paysan 3, Markus Oehlhafen 3, and Marc Kachelrieß 1,2 1 German Cancer Research Center

More information

Sparse sampling in MRI: From basic theory to clinical application. R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology

Sparse sampling in MRI: From basic theory to clinical application. R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology Sparse sampling in MRI: From basic theory to clinical application R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology Objective Provide an intuitive overview of compressed sensing

More information

COBRE Scan Information

COBRE Scan Information COBRE Scan Information Below is more information on the directory structure for the COBRE imaging data. Also below are the imaging parameters for each series. Directory structure: var/www/html/dropbox/1139_anonymized/human:

More information

The Near Future in Cardiac CT Image Reconstruction

The Near Future in Cardiac CT Image Reconstruction SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de

More information

An Accurate, Robust, and Computationally Efficient Navigator Algorithm for Measuring Diaphragm Positions #,z

An Accurate, Robust, and Computationally Efficient Navigator Algorithm for Measuring Diaphragm Positions #,z JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE 1 Vol. 6, No. 2, pp. 483 490, 2004 TECHNIQUE An Accurate, Robust, and Computationally Efficient Navigator Algorithm for Measuring Diaphragm Positions #,z Yiping

More information

Constrained Reconstruction of Sparse Cardiac MR DTI Data

Constrained Reconstruction of Sparse Cardiac MR DTI Data Constrained Reconstruction of Sparse Cardiac MR DTI Data Ganesh Adluru 1,3, Edward Hsu, and Edward V.R. DiBella,3 1 Electrical and Computer Engineering department, 50 S. Central Campus Dr., MEB, University

More information

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Benedikt Lorch 1, Martin Berger 1,2, Joachim Hornegger 1,2, Andreas Maier 1,2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg

More information

Fast Implementation of Iterative Image Reconstruction

Fast Implementation of Iterative Image Reconstruction Fast Implementation of Iterative Image Reconstruction Tobias Schaeffter 1, Michael Schacht Hansen 2, Thomas Sangild Sørensen 3 1 Division of Imaging Sciences and Biomedical Engineering, Kings College London

More information

8/11/2009. Common Areas of Motion Problem. Motion Compensation Techniques and Applications. Type of Motion. What s your problem

8/11/2009. Common Areas of Motion Problem. Motion Compensation Techniques and Applications. Type of Motion. What s your problem Common Areas of Motion Problem Motion Compensation Techniques and Applications Abdominal and cardiac imaging. Uncooperative patient, such as pediatric. Dynamic imaging and time series. Chen Lin, PhD Indiana

More information

Gradient-Based Differential Approach for Patient Motion Compensation in 2D/3D Overlay

Gradient-Based Differential Approach for Patient Motion Compensation in 2D/3D Overlay Gradient-Based Differential Approach for Patient Motion Compensation in 2D/3D Overlay Jian Wang, Anja Borsdorf, Benno Heigl, Thomas Köhler, Joachim Hornegger Pattern Recognition Lab, Friedrich-Alexander-University

More information

Model-Based Respiratory Motion Compensation for Image-Guided Cardiac Interventions

Model-Based Respiratory Motion Compensation for Image-Guided Cardiac Interventions Model-Based Respiratory Motion Compensation for Image-Guided Cardiac Interventions February 8 Matthias Schneider Pattern Recognition Lab Friedrich-Alexander-University Erlangen-Nuremberg Imaging and Visualization

More information

MODEL-BASED FREE-BREATHING CARDIAC MRI RECONSTRUCTION USING DEEP LEARNED & STORM PRIORS: MODL-STORM

MODEL-BASED FREE-BREATHING CARDIAC MRI RECONSTRUCTION USING DEEP LEARNED & STORM PRIORS: MODL-STORM MODEL-BASED FREE-BREATHING CARDIAC MRI RECONSTRUCTION USING DEEP LEARNED & STORM PRIORS: MODL-STORM Sampurna Biswas, Hemant K. Aggarwal, Sunrita Poddar, and Mathews Jacob Department of Electrical and Computer

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

High spatial and temporal resolution retrospective cine cardiovascular magnetic resonance from shortened free breathing real-time acquisitions

High spatial and temporal resolution retrospective cine cardiovascular magnetic resonance from shortened free breathing real-time acquisitions Xue et al. Journal of Cardiovascular Magnetic Resonance 2013, 15:102 RESEARCH Open Access High spatial and temporal resolution retrospective cine cardiovascular magnetic resonance from shortened free breathing

More information

M R I Physics Course

M R I Physics Course M R I Physics Course Multichannel Technology & Parallel Imaging Nathan Yanasak, Ph.D. Jerry Allison Ph.D. Tom Lavin, B.S. Department of Radiology Medical College of Georgia References: 1) The Physics of

More information

Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation

Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation Bastian Bier 1, Chris Schwemmer 1,2, Andreas Maier 1,3, Hannes G. Hofmann 1, Yan Xia 1, Joachim Hornegger 1,2, Tobias Struffert

More information

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging 1 CS 9 Final Project Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging Feiyu Chen Department of Electrical Engineering ABSTRACT Subject motion is a significant

More information

Multi-Dimensional Flow-Preserving Compressed Sensing (MuFloCoS) for Time-Resolved Velocity-Encoded Phase Contrast MRI

Multi-Dimensional Flow-Preserving Compressed Sensing (MuFloCoS) for Time-Resolved Velocity-Encoded Phase Contrast MRI 400 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 34, NO. 2, FEBRUARY 2015 Multi-Dimensional Flow-Preserving Compressed Sensing (MuFloCoS) for Time-Resolved Velocity-Encoded Phase Contrast MRI Jana Hutter*,

More information

k-t BLAST Reconstruction From Non-Cartesian k-t Space Sampling

k-t BLAST Reconstruction From Non-Cartesian k-t Space Sampling Magnetic Resonance in Medicine 55:85 91 (006) k-t BLAST Reconstruction From Non-Cartesian k-t Space Sampling Michael S. Hansen, 1, * Christof Baltes, 1 Jeffrey Tsao, 3 Sebastian Kozerke, 1 Klaas P. Pruessmann,

More information

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Qiao Yang 1,4, Meng Wu 2, Andreas Maier 1,3,4, Joachim Hornegger 1,3,4, Rebecca Fahrig

More information

Deep Learning for Fast and Spatially- Constrained Tissue Quantification from Highly-Undersampled Data in Magnetic Resonance Fingerprinting (MRF)

Deep Learning for Fast and Spatially- Constrained Tissue Quantification from Highly-Undersampled Data in Magnetic Resonance Fingerprinting (MRF) Deep Learning for Fast and Spatially- Constrained Tissue Quantification from Highly-Undersampled Data in Magnetic Resonance Fingerprinting (MRF) Zhenghan Fang 1, Yong Chen 1, Mingxia Liu 1, Yiqiang Zhan

More information

29 th NATIONAL RADIO SCIENCE CONFERENCE (NRSC 2012) April 10 12, 2012, Faculty of Engineering/Cairo University, Egypt

29 th NATIONAL RADIO SCIENCE CONFERENCE (NRSC 2012) April 10 12, 2012, Faculty of Engineering/Cairo University, Egypt K1. High Performance Compressed Sensing MRI Image Reconstruction Ahmed Abdel Salam, Fadwa Fawzy, Norhan Shaker, Yasser M.Kadah Biomedical Engineering, Cairo University, Cairo, Egypt, ymk@k-space.org Computer

More information

Fast Dynamic MRI for Radiotherapy

Fast Dynamic MRI for Radiotherapy 1 Fast Dynamic MRI for Radiotherapy KE SHENG, PH.D., FAAPM DEPARTMENT OF RADIATION ONCOLOGY UNIVERSITY OF CALIFORNIA, LOS ANGELES Disclosures I receive research grants from Varian Medical Systems I am

More information

Accelerated 4D flow MRI

Accelerated 4D flow MRI Accelerated 4D flow MRI Comparing SENSE, k-t PCA and Compressed Sensing Tomas Kaandorp 10356169 Patient comfort can be increased by accelerating an MRI scanner to reduce scan time. 4D flow MRI scans can

More information

Blind Sparse Motion MRI with Linear Subpixel Interpolation

Blind Sparse Motion MRI with Linear Subpixel Interpolation Blind Sparse Motion MRI with Linear Subpixel Interpolation Anita Möller, Marco Maaß, Alfred Mertins Institute for Signal Processing, University of Lübeck moeller@isip.uni-luebeck.de Abstract. Vital and

More information

Optimal Sampling Geometries for TV-Norm Reconstruction of fmri Data

Optimal Sampling Geometries for TV-Norm Reconstruction of fmri Data Optimal Sampling Geometries for TV-Norm Reconstruction of fmri Data Oliver M. Jeromin, Student Member, IEEE, Vince D. Calhoun, Senior Member, IEEE, and Marios S. Pattichis, Senior Member, IEEE Abstract

More information

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,

More information

Motion compensated reconstruction

Motion compensated reconstruction Motion compensated reconstruction Educational course: Image acquisition and Reconstruction 25th Annual meeting of the ISMRM, Honolulu, 2017 Sajan Goud Lingala Siemens Healthineers, Princeton, USA Declaration

More information

Spread Spectrum Using Chirp Modulated RF Pulses for Incoherent Sampling Compressive Sensing MRI

Spread Spectrum Using Chirp Modulated RF Pulses for Incoherent Sampling Compressive Sensing MRI Spread Spectrum Using Chirp Modulated RF Pulses for Incoherent Sampling Compressive Sensing MRI Sulaiman A. AL Hasani Department of ECSE, Monash University, Melbourne, Australia Email: sulaiman.alhasani@monash.edu

More information

Motion Compensation from Short-Scan Data in Cardiac CT

Motion Compensation from Short-Scan Data in Cardiac CT Motion Compensation from Short-Scan Data in Cardiac CT Juliane Hahn 1,2, Thomas Allmendinger 1, Herbert Bruder 1, and Marc Kachelrieß 2 1 Siemens Healthcare GmbH, Forchheim, Germany 2 German Cancer Research

More information

FOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION

FOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION Philips J. Res. 51 (1998) 197-201 FOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION This special issue of Philips Journalof Research includes a number of papers presented at a Philips

More information

A Non-Linear Image Registration Scheme for Real-Time Liver Ultrasound Tracking using Normalized Gradient Fields

A Non-Linear Image Registration Scheme for Real-Time Liver Ultrasound Tracking using Normalized Gradient Fields A Non-Linear Image Registration Scheme for Real-Time Liver Ultrasound Tracking using Normalized Gradient Fields Lars König, Till Kipshagen and Jan Rühaak Fraunhofer MEVIS Project Group Image Registration,

More information

Compressed sensing and undersampling k-space

Compressed sensing and undersampling k-space Boston University OpenBU Theses & Dissertations http://open.bu.edu Boston University Theses & Dissertations 2015 Compressed sensing and undersampling k-space Rahimpour, Yashar https://hdl.handle.net/2144/16257

More information

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005 Field Maps John Pauly October 5, 25 The acquisition and reconstruction of frequency, or field, maps is important for both the acquisition of MRI data, and for its reconstruction. Many of the imaging methods

More information

Compressed Sensing Reconstructions for Dynamic Contrast Enhanced MRI

Compressed Sensing Reconstructions for Dynamic Contrast Enhanced MRI 1 Compressed Sensing Reconstructions for Dynamic Contrast Enhanced MRI Kevin T. Looby klooby@stanford.edu ABSTRACT The temporal resolution necessary for dynamic contrast enhanced (DCE) magnetic resonance

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

VD-AUTO-SMASH Imaging

VD-AUTO-SMASH Imaging Magnetic Resonance in Medicine 45:1066 1074 (2001) VD-AUTO-SMASH Imaging Robin M. Heidemann, Mark A. Griswold, Axel Haase, and Peter M. Jakob* Recently a self-calibrating SMASH technique, AUTO-SMASH, was

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

Combination of Parallel Imaging and Compressed Sensing for high acceleration factor at 7T

Combination of Parallel Imaging and Compressed Sensing for high acceleration factor at 7T Combination of Parallel Imaging and Compressed Sensing for high acceleration factor at 7T DEDALE Workshop Nice Loubna EL GUEDDARI (NeuroSPin) Joint work with: Carole LAZARUS, Alexandre VIGNAUD and Philippe

More information

GE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals.

GE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals. GE Healthcare CLINICAL GALLERY Discovery * MR750w 3.0T This brochure is intended for European healthcare professionals. NEURO PROPELLER delivers high resolution, motion insensitive imaging in all planes.

More information

3D Guide Wire Navigation from Single Plane Fluoroscopic Images in Abdominal Catheterizations

3D Guide Wire Navigation from Single Plane Fluoroscopic Images in Abdominal Catheterizations 3D Guide Wire Navigation from Single Plane Fluoroscopic Images in Abdominal Catheterizations Martin Groher 2, Frederik Bender 1, Ali Khamene 3, Wolfgang Wein 3, Tim Hauke Heibel 2, Nassir Navab 2 1 Siemens

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Separate CT-Reconstruction for Orientation and Position Adaptive Wavelet Denoising

Separate CT-Reconstruction for Orientation and Position Adaptive Wavelet Denoising Separate CT-Reconstruction for Orientation and Position Adaptive Wavelet Denoising Anja Borsdorf 1,, Rainer Raupach, Joachim Hornegger 1 1 Chair for Pattern Recognition, Friedrich-Alexander-University

More information

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI Figure 1. In the brain, the gray matter has substantially more blood vessels and capillaries than white matter. The magnified image on the right displays the rich vasculature in gray matter forming porous,

More information

4D Computed Tomography Reconstruction from Few-Projection Data via Temporal Non-local Regularization

4D Computed Tomography Reconstruction from Few-Projection Data via Temporal Non-local Regularization 4D Computed Tomography Reconstruction from Few-Projection Data via Temporal Non-local Regularization Xun Jia 1, Yifei Lou 2, Bin Dong 3, Zhen Tian 1,4, and Steve Jiang 1 1 Department of Radiation Oncology

More information

Brilliance CT Big Bore.

Brilliance CT Big Bore. 1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.

More information

Computational Aspects of MRI

Computational Aspects of MRI David Atkinson Philip Batchelor David Larkman Programme 09:30 11:00 Fourier, sampling, gridding, interpolation. Matrices and Linear Algebra 11:30 13:00 MRI Lunch (not provided) 14:00 15:30 SVD, eigenvalues.

More information

Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset

Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset Andreas Maier 1, Patrick Kugler 2, Günter Lauritsch 2, Joachim Hornegger 1 1 Pattern Recognition Lab and SAOT Erlangen,

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE MRES 7005 - Fast Imaging Techniques Module 4 K-Space Symmetry Review K-Space Review K-Space Symmetry Partial or Fractional Echo Half or Partial Fourier HASTE Conditions for successful reconstruction Interpolation

More information

ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma

ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma ORAL DEFENSE 8 th of September 2017 Charlotte MARTIN Supervisor: Pr. MP REVEL M2 Bio Medical

More information

Evaluations of k-space Trajectories for Fast MR Imaging for project of the course EE591, Fall 2004

Evaluations of k-space Trajectories for Fast MR Imaging for project of the course EE591, Fall 2004 Evaluations of k-space Trajectories for Fast MR Imaging for project of the course EE591, Fall 24 1 Alec Chi-Wah Wong Department of Electrical Engineering University of Southern California 374 McClintock

More information

Nonrigid Motion Compensation of Free Breathing Acquired Myocardial Perfusion Data

Nonrigid Motion Compensation of Free Breathing Acquired Myocardial Perfusion Data Nonrigid Motion Compensation of Free Breathing Acquired Myocardial Perfusion Data Gert Wollny 1, Peter Kellman 2, Andrés Santos 1,3, María-Jesus Ledesma 1,3 1 Biomedical Imaging Technologies, Department

More information

Supplemental Material for Efficient MR Image Reconstruction for Compressed MR Imaging

Supplemental Material for Efficient MR Image Reconstruction for Compressed MR Imaging Supplemental Material for Efficient MR Image Reconstruction for Compressed MR Imaging Paper ID: 1195 No Institute Given 1 More Experiment Results 1.1 Visual Comparisons We apply all methods on four 2D

More information

SIEMENS MAGNETOM Avanto syngo MR B15

SIEMENS MAGNETOM Avanto syngo MR B15 \\USER\INVESTIGATORS\Ravi\ADNI-Subject\Localizer TA: 0:10 PAT: Voxel size: 1.9 1.5 8.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer Inline movie

More information

Sampling, Ordering, Interleaving

Sampling, Ordering, Interleaving Sampling, Ordering, Interleaving Sampling patterns and PSFs View ordering Modulation due to transients Temporal modulations Timing: cine, gating, triggering Slice interleaving Sequential, Odd/even, bit-reversed

More information

Whole Body MRI Intensity Standardization

Whole Body MRI Intensity Standardization Whole Body MRI Intensity Standardization Florian Jäger 1, László Nyúl 1, Bernd Frericks 2, Frank Wacker 2 and Joachim Hornegger 1 1 Institute of Pattern Recognition, University of Erlangen, {jaeger,nyul,hornegger}@informatik.uni-erlangen.de

More information