Beginner s Guide to Vevo 2100
|
|
- Sabina Neal
- 6 years ago
- Views:
Transcription
1 PN Beginner s Guide to Vevo 2100 December 2008
2 This Beginner s Guide to Vevo 2100 provides a quick overview of the Vevo 2100 imaging modes and provides key tips to allow users to start a study, acquire images, make measurements and export data. A more detailed step-by-step description of the system is available in the Vevo 2100 Operator Manual. Connect the transducer Attach the transducer that will be used during the imaging session to the connector on the front panel of the Vevo Start the Vevo Turn on the Vevo system using the breaker switch at the back of the system and then the computer by clicking the standby switch on the left side of the system. 2. Wait for the Windows Vista operating system to start-up. The system is configured by VisualSonics to automatically log in. The Vevo 2100 software should automatically launch. Start the Vevo Software 1. When a scanhead is connected the transducer initialization window appears. 2. Select the appropriate application and click Initialize. Initialization sets up the appropriate preset parameters for the selected application and specific transducer. Fig.1 Initialization Dialog 2. Once the transducer is initialized, the application opens in the Study Browser. The Study Browser window is used for sorting acquired data, for review and data export.
3 Click the New button in the Study Browser to open the New Study window. In this window the user can start a brand new study or a new series; a series is a new session that could be part of an old study. Fig.2 New Study dialog 3. Select one of the options. The application is going to open the Study Info window where the user needs to enter required information in order to start the study. The fields marked with yellow have to be filled in before the user can start imaging. Fig.3 Study/Series Information 4. Once all the required information is entered press OK and the software automatically begins imaging in B-Mode. There are various imaging modes available, accessible by the mode buttons located on the keyboard Page 3 of 12
4 B-Mode B-Mode imaging is used to visualize anatomical structures in a 2D greyscale image. 1. Press B-Mode to start acquistion. The acquisition parameters panel is activated by default. Users can change the parameters as needed and can save a custom set by pressing Save Presets. 2. Press <Scan/Freeze> to start and stop data acquisition. 3. Use the Cine Loop Review knob to scroll through the acquired data, once you have frozen the current acquisition, or reloaded an acquired loop. 4. To store a cineloop press <Cine Store>. The system can store a cineloop while actively scanning, or once the current acquisition has been frozen. 5. To store a single frame from a live acquisition or a frozen acquisition press <Frame Store>. 6. Users can simultaneously label and save a cine loop: by pressing Image Label, enter the Image name, and then press enter. Fig.4 B-Mode acquisition window left kidney in transverse view B-Mode is available on all Vevo 2100 systems, the following imaging modes are available as options to the Vevo You will only be able to use the options that have been purchased with your system. Page 4 of 12
5 M-Mode M-Mode imaging is used for visualizing and quantifying tissue movement over time, for example the motion of the myocardium over time. 1. In B-Mode acquisition, optimise the image for the area to be imaged in M-Mode. 2 Press the M-Mode button on the keyboard to display the M- Mode overlay. Adjust the position by moving the trackball, and size by clicking the SV Gate toggle up or down. Press the M- Mode button again or click the right mouse button. 3 Start and stop data acquisition using <Scan/Freeze>. New M-Mode image lines enter the image area from the right side and scroll across the window to the left in real-time. 4 Use the Cine Loop Review knob to scroll through the acquired data. 5 To store a data cineloop press <Cine Store>, this can be done in live scanning or when the current acquisition has been frozen; or use the Auto Image Label function (See B-Mode imaging Step. 6 To store a still frame image select <Frame Store>, again this can be done in live or frozen acquisition. Fig.5 M-Mode acquisition window Left Ventricle short axis view Fig. 5 M-Mode Acquisition Window Abdominal Aorta Page 5 of 12
6 Pulsed Wave (PW) Doppler Mode PW Doppler imaging is used for visualizing and quantifying blood flow through vessels and values in the heart. 1. In B-Mode acquisition, locate the region of interest for PW Doppler acquisition. Ensure that the angle between the direction of blood flow and the transmitted ultrasound beam is 60 o or less. 2. Press the PW button on the keyboard. 3. Position and size the sample volume at the desired location for PW Doppler signal acquisition; use the trackball to change the sample volume position and the SV Gate toggle to change the size of the sample volume. 4. Adjust the PW Doppler angle using the Doppler Angle knob the yellow dotted line should be placed in line with the blood flow. If steering is needed to reduce the angle between blood flow and ultrasound beam use the Beam Angle toggle to steer the angle either to the right or left. 5. Press the right mouse button or the PW button again to switch to the PW Doppler acquisition window. 6. Start and stop data acquisition using <Scan/Freeze>. New Doppler lines enter the image area from the right side and scroll across the window to the left in real-time. The audio representation of the acquired Doppler data is heard through the system speakers, the volume can be adjusted using the rotary dial on the keyboard. 7. Use the Cine Loop Review knob to scroll through the acquired data. 8. To store a cineloop press <Cine Store> or use the Auto-Image Label function. (See B-Mode imaging Step 6) 9. To store a still image press <Frame Store>. Fig.6 PW Doppler acquisition window Mitral Valve apical four chamber view Page 6 of 12
7 3D-Mode 3D-Mode imaging is used to visualize an anatomical structure in three dimensions. Multiple 2D slices are acquired and compiled into a 3D volume. 3D-Mode imaging can be used for B-Mode Power Doppler Mode, and Contrast Mode Data. 1. Connect the 3D motor stage to the Vevo Rail System, attach the motor to the ball joint, where the scanhead clamp can also be attached. 2. Connect the transducer clamp to the 3D motor. 3. Mount the transducer in the clamp; the transducer imaging plane must be perpendicular to the axis along which the motor will move. 4. Press the <3D> button on the keyboard. The 3D motor initialization dialog will come up. Click okay and the motor will initialize. Ensure that the probe is not touching an animal and that nothing will abstruct it s movement. 5. Optimize the image for acquisition, for example in B-Mode and determine the range the motor must move to cover the desired volume using the micromanipulators on the stage for the animal handling platform. 6. When the acquisition ends the application opens the 3D analysis window. If using Contrast Mode, the image is loaded into a 2D window frame which can then be loaded into 3D once all post processing steps are complete; this is done by clicking Load in 3D. 7. To store the data press <3D Store> or Auto-Save on Image Label. Fig.7 3D Mode analysis window Left Ventricle using ECG and Respiration gating Page 7 of 12
8 Power Doppler Mode Power Doppler Mode imaging is used to visualize blood flow; generally Power Doppler is used to visualize flow in structures such as tumors or abdominal organs. 1. In B-Mode imaging optimize the image for the target of interest to be imaged in Power Doppler. 2. Press the Power button on the keyboard to start acquisition 3. Position and size the Power Box at the desired location for Power Doppler signal acquisition. The position (solid box) of the Power Box can be adjusted by moving the Trackball around the image, the size (dotted box) can also be adjusted by using the trackball; to toggle between the two boxes use the right mouse button. 4. Start and stop acquisition using <Scan/Freeze>. 5. To store a cineloop press <Cine Store> or the Auto-Save on Image Label. (see B-Mode imaging Step 6). 6. To store a still image press <Frame Store>. Note: Power Doppler Mode is also supported in 3D-Mode. To acquire Power-3D-Mode data, from Power Doppler Mode click 3D and adjust the 3D acqusition parameters. At the end of the acquisition the application switches to Power 3D analysis window. Fig.8 Power Doppler acquisition window Subcutaneous Tumor on hind limb Page 8 of 12
9 Color Doppler Mode Color Dopper imaging is used to visualize blood flow and give additional information about the direction of flow. Color Doppler is used on a wide range of targets such as vessels, the heart, and even tumors and abdominal organs 1. In B-Mode imaging optimize the image for the target of interest to be imaged in Color Doppler. 2. Press the Color button on the keyboard to start acquisition. 3. Position and size the Color Box at the desired location for Color Doppler signal acquisition. The position (solid box) of the Color Box can be adjusted by moving the Trackball around the image, the size (dotted box) can also be adjusted by using the trackball; to toggle between the two boxes use the right mouse button. 4. Start and stop data acquisition using <Scan/Freeze>. 5. To store a cineloop press <Cine Store> or the Auto-Save on Image Label. (see B-Mode imaging Step 6). 6. To store a still image press <Frame Store>. Fig.9 Color Doppler acquisition window Left Common Carotid Artery transverse view Page 9 of 12
10 Contrast Mode Contrast Mode imaging is used in conjunction with the MicroMarker Contrast Agents to enhance the visualization of blood flow down to the capillary level, and also to detect the expression of various biomarkers on the cell surface of endothelial cells. 1. In B-Mode imaging optimize the image for the target of interest to be imaged in Contrast Mode. 2. Press the Contrast button on the keyboard to start acquisition. 3. Start and stop data acquisition using <Scan/Freeze> 4. Use the Cine Loop Review knob to scroll through the acquired data. 5. To store a cineloop press <Cine Store> or the Auto-Save on Image Label. (see B-Mode imaging Step 6). 6. To store a still image press <Frame Store>. Note: Contrast Mode is also supported in 3D-Mode. To acquire Contrast-3D-Mode data, from Contrast Mode click 3D and set the acquisition parameters. To open the Contrast 3D Mode data in the 3D analysis window, users need to press the Load in 3D button. Fig.10 Loading Contrast 3D data in the 3D window Fig.11 Contrast Mode acquisition window Non-Targeted MicroMarker Contrast Agent in a subcutaneous tumor on the hind limb Page 10 of 12
11 Data Review 1. To review the data. a. To review the study data press Study Managment b. To select a specific loop for review double click on the thumbnail image, or on the image label 2. To export still images or movies a. Open the desired cineloop b. If exporting a movie adjust the playback controls to the desired frames and playback speed; if exporting a picture select the appropriate frame c. Click on Export d. Select the appropriate Export Type e. In options change the name of the file if desired and choose the appropriate File Type f. The save location can also be chosen on the left side of the window 2. To save the study. a. Select Close from Study Managment to close the study. (Users will have the option of adding a new series to the study at later time points) b. To copy the study to an accessible drive, click Copy To c. To import a study into the software, for example on a workstation, open the software, click on Copy From, and select the study from the accessible drive Data Analysis 1. To add measurements press <Measure>. In Vevo 2100 there are 5 available factory measurement packages: Cardiac Abdominal Embryology Ophtalmology Vascular All these packages are available in all modes except Contrast Mode. Users can define custom measurements packages where they can select: - package content and the display properties of the measurement (font, size, color) There is no limit for the number of measurements of the same type. The measurements report displays all the values, the Page 11 of 12
12 average for all measurements, and the standard deviation. 2. To add annotations right-click the image or press <Annotate>. The appearance of the annotations can be changed in the preferences window, this is opened by pressing Pref on the keyboard, or using the preferences tab on the workstation software. Shutting Down the Vevo At the end of the session click Close Study. 2. Press the computer standby switch on the left side of the system to shutdown the software and turn off the computer. 3. Turn off the Vevo System by turning off the breaker at the back of the system. Need help? Call us toll-free to or via at Page 12 of 12
Quick Start Guide to the Vevo 770
Quick Start Guide to the Vevo 770 PN 11422 This Quick Start Guide provides a quick overview of the Vevo 770 and provides tips to allow users to start using the instrument and its various modes. A detailed
More informationQuick Start Guide to the Vevo 770
Quick Start Guide to the Vevo 770 PN 11422 This Quick Start Guide provides a quick overview of the Vevo 770 and provides tips to allow users to start using the instrument and its various modes. A detailed
More informationVisualSonics Application Protocol 3D Tumor Imaging using the MicroMarker Non-Targeted Contrast Agents
VisualSonics Application Protocol 3D Tumor Imaging using the MicroMarker Non-Targeted Contrast Agents 1 Objective Part Number: 11902 The objective of the 3D Tumor Imaging using the MicroMarker Non-Targeted
More informationImaging Guide Pancreatic Imaging
Imaging Guide Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Imaging System Ver 1.0 Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Imaging System Course Objectives: This guide
More informationVevo770 Ultrasound. Quick Operation Guide AIPT
Vevo770 Ultrasound Quick Operation Guide Vevo770 Probe head (707B/704) Preparation The acoustic membrane is very fragile and sensitive, it must be handled with care. The probe head should be filled with
More informationVEVO 1100 Sole Source Specifications
VEVO 1100 Sole Source Specifications Table of Contents 1.SYSTEM... 2 1-1 Licensing for the User... 2 1-2 Safety... 2 1-3 Portability... 2 2.TRANSDUCERS... 2 2-1 MicroScan transducers... 2 3. IMAGING STATION...
More informationMyLab Alpha Quick Start
MyLab Alpha Quick Start Basic Setup The system has a 2.5 minute start time, but can be put in standby mode to allow for an 8 second boot time. From the patient data screen, click on the Menu button Then
More informationVEVO 2100 Sole Source Specifications
VEVO 2100 Sole Source Specifications Table of Contents 1.SYSTEM... 2 1-1 Licensing for the User... 2 1-2 Safety... 2 1-3 Portability... 2 2.TRANSDUCERS... 2 2-1 MicroScan transducers... 2 3. IMAGING STATION...
More informationLOGIQ. V2 Ultrasound. Part of LOGIQ Vision Series. Imagination at work LOGIQ is a trademark of General Electric Company.
TM LOGIQ V2 Ultrasound Part of LOGIQ Vision Series Imagination at work The brilliance of color. The simplicity of GE. Now you can add the advanced capabilities of color Doppler to patient care with the
More informationVisualSonics Vevo 770 High-Resolution Imaging System. Operator Manual
VisualSonics Vevo 770 High-Resolution Imaging System Operator Manual VisualSonics Vevo 770 High-Resolution Imaging System Operator Manual January 2006 PN 11322 Revision 1.3 V1.3 VisualSonics Vevo 770
More informationQuick reference guide. MyLab50
Quick reference guide MyLab50 Quick reference guide MyLab50 Introduction This Quick Guide describes only the basic operating procedures for the diagnostic ultrasound MyLab50 Family, named in the following
More informationQuick Guide. EPIQ 7 Ultrasound System. English
Quick Guide EPIQ 7 Ultrasound System English Contents For easy reference, each section is color-coded. The color of each section title correlates with the color band on each page in that section. Controls
More information10 Working With Setup Tools
10 Working With Setup Tools The Setup window allows you to set defaults and control various functions of the Terason software. The Setup window tabs display sets of controls for defaults and other settings.
More information10 Working With Setup Tools
10 Working With Setup Tools Setup Tools Overview The Setup window allows you to set defaults and control various functions of the Terason software. The Setup window tabs display sets of controls for defaults
More informationHigh Resolution Multi-modal in vivo Imaging Platform
High Resolution Multi-modal in vivo Imaging Platform The world s only customizable imaging platform combining ultra high frequency ultrasound and photoacoustics Experience the next generation of in vivo
More information1 About the Terason Ultrasound System
1 About the Terason Ultrasound System Overview The Terason Ultrasound System is an easy-to-use, portable ultrasound system that produces high resolution images. The microminiaturized ultrasound system
More informationACUSON P300 Ultrasound System Control Panel Diagram
www.siemens.com ACUSON P300 Ultrasound System Control Panel Diagram 1 Power On / Off 1 2 TGC Controls 3 Left Mouse Click 4 Action 4 7 5 Right Mouse Click 6 Trackball 2 3 6 5 8 7 Pointer 8 Freeze 16 Color
More informationAORTA CTA VPMC-12419
AORTA CTA VPMC-12419 Workflow Overview: The Aorta can be post-processed in various ways. Auto Bone Removal and Vessel Pick provide a quick overview of the entire Aorta. Vessel Probe creates a centerline
More informationThis is your imaging dashboard. There is a bevy of valuable information to read from here. Let s break it up.
The Volume Strikes Back Press the (4D) button. This is your imaging dashboard. There is a bevy of valuable information to read from here. Let s break it up. In the Left GUI (Graphic User Interface), you
More informationVisualsonics Vevo2100: Manual for Collecting RF Data. Updated: 10/29/2012 M Kurowski
Visualsonics Vevo2100: Manual for Collecting RF Data Updated: 10/29/2012 M Kurowski 1 : System Start-Up/Shutdown 1.1 The Visualsonics mvevo2100 system must be the only device plugged into an outlet; failure
More informationTELEMED. Echo Wave II Software LB-2. Ultrasound Consolle. User Manual
SmartUs - ClarUs - LS128 - LS64 - Echo Blaster 64 - Echo Blaster 128 Echo Wave II Software LB-2 Ultrasound Consolle User Manual T E L E M E D Ultrasound Medical Systems Telemed Medical Systems WEB: Telemed
More informationMindray. Hand-carried Diagnostic Ultrasound System. Expanding the Envelope of Performance and Flexibility
M7 Mindray M7 Hand-carried Diagnostic Ultrasound System Expanding the Envelope of Performance and Flexibility M7 Diagnostic Ultrasound System Equipped for Quality Radiology Cardiology Outpatient Emergency
More informationSONOLINE G40 ultrasound system
SONOLINE G40 ultrasound system Quick Scan Card: General/OB For system software version 2.0 Press the Power Switch on the upper left side of the keyboard. A blue light will indicate the power is on. Getting
More informationUltrasound. Q-Station software. Streamlined workflow solutions. Philips Q-Station ultrasound workspace software
Ultrasound Q-Station software Streamlined workflow solutions Philips Q-Station ultrasound workspace software Managing your off-cart workf low Everyone is being asked to do more with fewer resources it
More informationSONOS 7500/5500 Using 3-Dimensional and BiPlane Imaging And Other System Changes for Software Revision D.1
SONOS 7500/5500 Using 3-Dimensional and BiPlane Imaging And Other System Changes for Software Revision D.1 User s Guide Using 3-Dimensional and BiPlane Imaging And Other System Changes for Software Revision
More informationDiagnostic Ultrasound System. Operation Manual. Version English MI A
Diagnostic Ultrasound System Operation Manual Version 1.00.00 English MI68-02334A Table of Contents Table of Contents Volume 2 Chapter 6 Starting Diagnosis System Power... 6-3 Turning the Power On...
More informationGE Healthcare. Vivid 7 Dimension 08 Cardiovascular ultrasound system
GE Healthcare Vivid 7 Dimension 08 Cardiovascular ultrasound system ltra Definition. Technology. Performance. Start with a system that s proven its worth in LV quantification and 4D imaging. Then add even
More informationWhat you should know about the new release
T.A8.4401-01 TomTec-Arena 1.1 What you should know about the new release Congratulations to the purchase of your new TomTec Software TomTec-Arena and thank you for choosing TomTec Imaging Systems GmbH
More informationEchoPAC Software Only v201
EchoPAC Software Only v201 Product Description The EchoPAC Application is a comprehensive clinical software package for viewing, analyzing, and reporting of multi-dimensional echo, vascular and abdominal
More informationClinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis
Clinical Importance Rapid cardiovascular flow quantitation using sliceselective Fourier velocity encoding with spiral readouts Valve disease affects 10% of patients with heart disease in the U.S. Most
More informationstep into the future Ultrasound Diagnostic Imaging System
step into the future Ultrasound Diagnostic Imaging System Moving into a new era for ultrasound diagnostics. With a proven track record in medical imaging, FUJIFILM has partnered with SonoSite, an innovator
More informationXRADIA microxct Manual
XRADIA microxct Manual Multiscale CT Lab Table of Contents 1. Introduction and Basics 1.1 Instrument Parts 1.2 Powering up the system 1.3 Preparing your sample 2. TXM Controller 2.1 Starting up 2.2 Finding
More informationGet to Know the VideoStudio Workspace
Get to Know the VideoStudio Workspace This written tutorial is a handy guide to understanding the various components of the VideoStudio interface. We will review the Capture, Edit and Share workspaces
More informationSemi-Interactive Console Layout and Instructions
Semi-Interactive Console Layout and Instructions A video of the SIC accompanies these instructions, which you may refer to for further understanding. Click here to open the video link. NOTE: The images
More informationSimplicity, quality, and flexibility
Ultrasound Lumify Simplicity, quality, and flexibility Philips Lumify ultrasound system specifications Contents 1 Introduction 3 1.1 Applications 3 2 System overview 4 2.1 System architecture 4 2.2 Imaging
More informationEcho Blaster 64/128 and LogicScan 64/128 Series Ultrasound Systems Echo Wave II Software User Manual
Echo Blaster 64/128 and LogicScan 64/128 Series Ultrasound Systems Echo Wave II Software User Manual Internet page: http://www.telemed.lt Information E-mail: info@telemed.lt Ultrasound Medical Systems
More informationImage Optimisation. -How to Get the Most out of your Echo Machine. Katarina Winkler, Biomedicinsk analytiker, Barnhjärtcentrum i Lund
Image Optimisation -How to Get the Most out of your Echo Machine analytiker, Barnhjärtcentrum i Lund Image Optimisation Helpful tips to improve the 2D-Image and Doppler registrations. Common adjustments
More informationUS 1.
US 1 Sample image: Normal pancreas seen on sonogram. Looking up from abdomen toward the head of the patient. The liver is in front of the pancreas. A vein draining the spleen is behind the pancreas http://www.radiologyinfo.org/photocat/photos.cfm?image=abdo-us-pancr.jpg&&subcategory=abdomen&&stop=9
More informationSONOSITE X-PORTE TRANSDUCERS
SONOSITE X-PORTE TRANSDUCERS C11xp 8-5 MHz Curved Abdominal, Arterial, Neonatal, Nerve, Venous C35xp 8-3 MHz Curved Abdominal, MSK, Nerve C60xp 5-2 MHz Curved Abdominal, Gynecology, MSK, Nerve, Obstetrics
More informationAmbra User Guide. If you need help. Ambra Support (any time)
If you need help Ambra Support 888 315 0790 (any time) support@ambrahealth.com Ambra User Guide Envision Radiology, a Health Images Organization, has provided a list of your site s personnel that need
More informationModule 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution.
MRES 7005 - Fast Imaging Techniques Module 5: Dynamic Imaging and Phase Sharing (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review Improving Temporal Resolution True-FISP (I) True-FISP (II) Keyhole
More informationMedical Image Viewer Guide
Cloud Medical Image Management Medical Image Viewer Guide March 2016 Table of Contents Indications for Use 3 Browser Requirements 3 User Interface Overview 4-5 Study Page 6 Main Toolbar 7 Sub-Toolbars
More informationLogger Pro 3. Quick Reference
Logger Pro 3 Quick Reference Getting Started Logger Pro Requirements To use Logger Pro, you must have the following equipment: Windows 98, 2000, ME, NT, or XP on a Pentium processor or equivalent, 133
More informationQUANTAX EDS SYSTEM SOP
QUANTAX EDS SYSTEM SOP December 2017 Energy-Dispersive X-Ray Spectroscopy (EDS, EDX, EDXS or XEDS), is an analytical technique used for the elemental analysis or chemical characterization of a sample.
More informationImage Acquisition Systems
Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS
More informationA Slim and Ultra Compact System, With Advanced Performance
Samsung Medison is a global leading medical devices company. Founded in 1985, the company now sells cutting-edge medical devices including diagnostic ultrasound, digital X-ray and blood analyzer, in 110
More informationTechnical Principles of Transthoracic Three-Dimensional Echocardiography
Technical Principles of Transthoracic Three-Dimensional Echocardiography 2 Stein Inge Rabben 2.1 Introduction The last years we have experienced a rapid development in three-dimensional echocardiography
More informationUltrasound To Go. The MySono U5 -
Ultrasound To Go The MySono U5 - Ultrasound To Go With the introduction of the MySono U5, MEDISON brings you a fully featured ultrasound imaging system to go. Delivering exceptional image quality and featuring
More informationMedDream WEB DICOM Viewer User's Manual
MedDream WEB DICOM Viewer User's Manual Version 4.01 2013, UAB Softneta, info@softneta.com Table of Contents Change log:...3 1. Logging on to MedDream...4 2. Search of studies...5 2.1 Opening multiple
More informationOperator's Guide FMD-3200/FMD-3200-BB/FMD-3300
Electronic Chart Display and Information System (ECDIS) English Model Operator's Guide FMD-3200/FMD-3200-BB/FMD-3300 The purpose of this guide is to provide the basic operation for this equipment. For
More informationVisage 7 Clinical Training Basic Features
Visage 7 Clinical Training Basic Features Contents Overview... 4 Usage... 4 Client Server Architecture... 5 Client Login... 6 Study Browser... 7 Query Section... 8 Study Labels... 10 Query Labeled Studies...
More informationGE Healthcare. Vivid E9 XDclear. with
GE Healthcare Vivid E9 XDclear with capabilities for every day Whether in TTE or TEE, capture the entire heart in a single beat. Image a valve, or the entire ventricle with excellent image quality. And
More informationGE Medical Systems Training In Partnership LOGIQ 400. Quick Start Guide A Training in Partnership Program
GE Medical Systems Training In Partnership LOGIQ 400 Quick Start Guide 2237879 100 A Training in Partnership Program TiPTM Regulatory Requirement This product complies with regulatory requirements of
More informationLearn Image Segmentation Basics with Hands-on Introduction to ITK-SNAP. RSNA 2016 Courses RCB22 and RCB54
Learn Image Segmentation Basics with Hands-on Introduction to ITK-SNAP RSNA 2016 Courses RCB22 and RCB54 RCB22 Mon, Nov 28 10:30-12:00 PM, Room S401CD RCB54 Thu, Dec 1 2:30-4:30 PM, Room S401CD Presenters:
More informationGamepad Controls. Figure 1: A diagram of an Xbox controller. Figure 2: A screenshot of the BodyViz Controller Panel. BodyViz 3 User Manual 1
BodyViz User Manual Gamepad Controls The first step in becoming an expert BodyViz user is to get acquainted with the Xbox gamepad, also known as a controller, and the BodyViz Controller Panel. These can
More informationCompactly SAMSUNG powerful ULTRASOUND H60
Samsung Medison is a global leading medical devices company. Founded in 1985, the company now sells cutting-edge medical devices including diagnostic ultrasound, digital X-ray and blood analyzer around
More informationUser s Guide. Creative Desktop Wireless 8000
User s Guide Creative Desktop Wireless 8000 Information in this document is subject to change without notice and does not represent a commitment on the part of Creative Technology Ltd. No part of this
More informationUltrasound To Go. MySono U5
Ultrasound To Go MySono U5 Ultrasound To Go With the introduction of the MySono U5, Samsung Medison brings you a fully featured ultrasound imaging system to go. Delivering exceptional image quality and
More informationLiving Image Software
Living Image Software User s Manual Version 3.1 2002-2008 Xenogen Corporation. All rights reserved. PN 124908 Xenogen Corporation 68 Elm Street Hopkinton, MA 01748 USA 1.877.522.2447 (US) 1.508.435.9500
More informationSection 9. Human Anatomy and Physiology
Section 9. Human Anatomy and Physiology 9.1 MR Neuroimaging 9.2 Electroencephalography Overview As stated throughout, electrophysiology is the key tool in current systems neuroscience. However, single-
More informationSL-6397 DECUS GAMING MOUSE
Profile selection and configuration, lighting effects You can switch between 5 different profiles at the top of the left-hand column. A default profile is configured when the driver is installed for the
More informationTutorial BOLD Module
m a k i n g f u n c t i o n a l M R I e a s y n o r d i c B r a i n E x Tutorial BOLD Module Please note that this tutorial is for the latest released nordicbrainex. If you are using an older version please
More informationRev: August 29, CIRS Motion Control User Manual
CIRS Motion Control User Manual 1 Table of Contents Introduction... 3 Installation... 3 Note... 3 Interface... 4 Using the Motion Library... 7 Running a Motion... 9 Adjusting Motion Parameters... 10 Motion
More informationSegment CT Instructions For Use - English
Segment CT Instructions For Use - English February 28, 2018 Software platform v2.2 R6239 MEDVISO AB http://www.medviso.com Griffelvägen 3 SE-224 67 Lund Sweden Tel: +46-76-183 6442 ii Contents 1 Terms
More informationComputed tomography of simple objects. Related topics. Principle. Equipment TEP Beam hardening, artefacts, and algorithms
Related topics Beam hardening, artefacts, and algorithms Principle The CT principle is demonstrated with the aid of simple objects. In the case of very simple targets, only a few images need to be taken
More informationIntroduction to Microsoft Office PowerPoint 2010
Introduction to Microsoft Office PowerPoint 2010 TABLE OF CONTENTS Open PowerPoint 2010... 1 About the Editing Screen... 1 Create a Title Slide... 6 Save Your Presentation... 6 Create a New Slide... 7
More informationThe Supreme 3D/4D Ultrasound. The Supreme 3D/4D Ultrasound. CT-V20-ICM
The Supreme 3D/4D Ultrasound The Supreme 3D/4D Ultrasound www.medison.com info@medison.com CT-V20-ICM-06.20.2008 The Supreme 3D/4D Ultrasound Since launching the first commercially available 3D ultrasound
More informationAnsur ProSim 6/8. Users Manual. Software Plug-In
Ansur ProSim 6/8 Software Plug-In Users Manual FBC-0002 April 2011, Rev. 2, 8/11 2011 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies. Table of Contents
More informationKNOBOLOGY. Jamie Ternes, RVT William Beaumont Hospital Royal Oak, Michigan. From the Beginning
KNOBOLOGY Jamie Ternes, RVT William Beaumont Hospital Royal Oak, Michigan From the Beginning Power On/Off Patient information Presets Probe selection Scan 1 Modalities /Gray scale/2d Pulse Wave Doppler
More informationRemote Support. User Guide 7.23
Remote Support User Guide 7.23 Copyright 1997 2011 Cisco and/or its affiliates. All rights reserved. WEBEX, CISCO, Cisco WebEx, the CISCO logo, and the Cisco WebEx logo are trademarks or registered trademarks
More informationn o r d i c B r a i n E x Tutorial DTI Module
m a k i n g f u n c t i o n a l M R I e a s y n o r d i c B r a i n E x Tutorial DTI Module Please note that this tutorial is for the latest released nordicbrainex. If you are using an older version please
More informationUNIVERSITY OF SOUTHAMPTON
UNIVERSITY OF SOUTHAMPTON PHYS2007W1 SEMESTER 2 EXAMINATION 2014-2015 MEDICAL PHYSICS Duration: 120 MINS (2 hours) This paper contains 10 questions. Answer all questions in Section A and only two questions
More informationPOWERPOINT 2003 OVERVIEW DISCLAIMER:
DISCLAIMER: POWERPOINT 2003 This reference guide is meant for experienced Microsoft Office users. It provides a list of quick tips and shortcuts for familiar features. This guide does NOT replace training
More informationCase Study 1: Piezoelectric Rectangular Plate
Case Study 1: Piezoelectric Rectangular Plate PROBLEM - 3D Rectangular Plate, k31 Mode, PZT4, 40mm x 6mm x 1mm GOAL Evaluate the operation of a piezoelectric rectangular plate having electrodes in the
More informationQUADRA-CHEK 2000 Demo User's Manual. Evaluation Unit
QUADRA-CHEK 2000 Demo User's Manual Evaluation Unit English (en) 06/2018 Contents Contents 1 Fundamentals...7 2 Software Installation...11 3 Basic Operation... 17 4 Software Configuration...41 5 Quick
More informationn o r d i c B r a i n E x Tutorial DSC Module
m a k i n g f u n c t i o n a l M R I e a s y n o r d i c B r a i n E x Tutorial DSC Module Please note that this tutorial is for the latest released nordicbrainex. If you are using an older version please
More informationRetrospective Spiral Respiratory Correlated Imaging with Varian RPM
Retrospective Spiral Respiratory Correlated Imaging with Varian RPM This is a Quick Step Guide for Retrospective Spiral Respiratory Correlated Imaging (4D CT) using the Varian RPM device v1.7 with the
More information8 Setting Preferences
PRELIMINARY Setting Preferences 8-1 8 Setting Preferences About this Section This section explains the Preferences panel. CableEye is a powerful software that includes several options that can be customized.
More informationTerason s Light-Weight, Power-Packed Ultrasound System. Specifications. The New Look of Ultrasound.
Terason s Light-Weight, Power-Packed Ultrasound System Specifications The New Look of Ultrasound. usmart 3200T Ultrasound System Specifications Overview Clinical Applications Abdominal Anesthesia Arterial
More informationESS I-Cap ver 1.0x. ESS I-cap Program Users Manual. clear-cut. Image Capture, Catalog, and Report Generation
ESS I-Cap ver 1.0x clear-cut Image Capture, Catalog, and Report Generation - 1 - Table of Contents System Requirements...3 Hardware Installation:...3 Software installation:...3 Video Source...5 The Video
More informationIMAGE STUDIO LITE. Tutorial Guide Featuring Image Studio Analysis Software Version 3.1
IMAGE STUDIO LITE Tutorial Guide Featuring Image Studio Analysis Software Version 3.1 Notice The information contained in this document is subject to change without notice. LI-COR MAKES NO WARRANTY OF
More informationQuick Start for DATS on Windows 7
Skip directly to the section that covers your version of Windows (7, Vista or XP) Quick Start for DATS on Windows 7 *** VERY IMPORTANT SAFETY NOTE *** NEVER connect the test leads of the DATS to a speaker
More informationPalm Reader Handbook
Palm Reader Handbook Copyright 2000-2002 Palm, Inc. All rights reserved. Graffiti, HotSync, the Palm logo, and Palm OS are registered trademarks of Palm, Inc. The HotSync logo and Palm are trademarks of
More informationCertificate in Clinician Performed Ultrasound (CCPU)
Certificate in Clinician Performed Ultrasound (CCPU) Syllabus Physics Tutorial Physics Tutorial Purpose: Training: Assessments: This unit is designed to cover the theoretical and practical curriculum for
More informationChapter 3 Operating instructions
Chapter 3 Operating instructions Summary This chapter describes the how to control and navigate through the TVR 30 s menus and options. Content Control interfaces 10 Controlling the TVR 30 10 Front panel
More informationQuick Start for WT3 (v1.5) on Windows XP (and earlier)
Skip directly to the section that covers your version of Windows (XP and earlier, Vista or Windows 7) Quick Start for WT3 (v1.5) on Windows XP (and earlier) *** VERY IMPORTANT SAFETY NOTE *** NEVER connect
More informationGE Healthcare. Vivid E9 XDclear. with. healthymagination. GE imagination at work
GE Healthcare with Vivid E9 XDclear healthymagination GE imagination at work Extraordinary capabilities Whether in TTE or TEE, capture the entire heart in a single beat. Image a valve, or the entire ventricle
More information5 Working With Studies
5 Working With Studies The Terason software provides many tools for you to work with studies. These tools allow: Storing Images and Loops; see page 90 Reviewing Patient Studies; see page 92 Exporting Studies;
More informationVascular Research Tools 5
Vascular Research Tools 5 Documentation Medical Imaging Applications, LLC c Medical Imaging Applications, LLC July 15, 2006 Contents 1 The Family of Vascular Research Tools 7 1.1 Image Format Compatibility
More informationClinicalConnect TM eunity TM Training Guide
ClinicalConnect TM eunity TM Training Guide October, 2013 Launch eunity TM from ClinicalConnect TM Search and select the patient whose record you wish to view. Navigate to the Radiology module in ClinicalConnect
More informationStandard Plus Player. User Guide. i-tech Company LLC TOLL FREE: (888) WEB:
Standard Plus Player User Guide i-tech Company LLC TOLL FREE: (888) 483-2418 EMAIL: info@itechlcd.com WEB: www.itechlcd.com 1. INTRODUCTION OF THE Standard Plus PLAYER... 3 2. MAIN MENU... 4 2.1 START
More informationM7Vet Diagnostic Ultrasound System
2011.06 V1.0 M7Vet Diagnostic Ultrasound System System Description The diagnostic ultrasound system is intended for use in abdomen, cardiac, small parts, peripheral vascular, transrectal, intraoperative
More informationUsing TeraRecon intuition Viewer with STAT Quick Reference Guide
Using TeraRecon intuition Viewer with STAT Quick Reference Guide 1. Launching intuition 2. 3D Assessment of Coronary Arteries for Stenosis 3. Time Volume Analysis (TVA) for Determination of Left Ventricular
More informationGE Healthcare. Venue 40. an innovation of healthymagination
GE Healthcare Venue 40 an innovation of healthymagination Product Description Venue 40 gives physicians precision and exceptional image quality in an intuitive and affordable ultrasound system. Venue is
More informationBeyond Performance and Value
Putting the back in ultrasound Beyond Performance and Value Esaote s new ultra-performance MyLab 9 exp ultrasound system is designed to support a full range of shared service diagnostic imaging environments.
More informationNSS evo3 Quick Start Guide
NSS evo3 Quick Start Guide EN Front panel 2 3 4 5 6 7 8 9 10 11 12 12 1 No. Description 1 Touch screen. Pages/Home key - press to open the Home page for page selection and setup 2 options. WheelKey - user
More informationComputed tomography (Item No.: P )
Computed tomography (Item No.: P2550100) Curricular Relevance Area of Expertise: Biology Education Level: University Topic: Modern Imaging Methods Subtopic: X-ray Imaging Experiment: Computed tomography
More informationContents. Introduction... iii. CHAPTER 1 - Getting Started... 5
Transducer Characterization User s Guide Version 5 5 1 build 12 or later UTEX Scientific Instruments Inc support@utex com Winspect Transducer Characterization User s Guide Contents Table of Contents and
More informationSpecifications. The New Look of Ultrasound.
Specifications The New Look of Ultrasound. usmart 3300 Ultrasound System Specifications Overview Clinical Applications Abdominal Arterial Breast Carotid Critical Care FAST-Emergency Medicine MSK Neonatal
More informationMedis Suite 3.0. Quick Start Manual. October 6, v3.0
Medis Suite 3.0 Quick Start Manual October 6, 2017 9.45.250.30.2 v3.0 Medis medical imaging systems bv Schuttersveld 9, 2316 XG Leiden, the Netherlands http://www.medis.nl Medis medical imaging systems
More information