Imaging Guide Pancreatic Imaging

Size: px
Start display at page:

Download "Imaging Guide Pancreatic Imaging"

Transcription

1 Imaging Guide Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Imaging System Ver 1.0

2 Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Imaging System Course Objectives: This guide is designed to assist the user to: - recognize pancreatic structures in typical views - position the correct probe to get the appropriate views of the pancreas - performing the measurements of interest to get anatomical and physiological data Overview of the Research Areas and Anatomy of the Pancreas The pancreas of the mouse is located in the upper left abdominal cavity. It borders the major vessels medially and the spleen laterally. The kidney is inferior to the pancreas while the stomach is superior (Figure 1). In ultrasound, the pancreas has a brick road or cobblestone appearance. Figure 1. Schematic showing the abdominal cavity of the mouse with the pancreas exposed and its relative position to surrounding anatomy. Imaging Guide: Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Page 1

3 Overview of the Imaging Modes to be Used in this Guide B-Mode: B-Mode, or brightness mode, imaging is used to acquire two dimensional images of an area of interest. The grayscale image allows for identification of anatomical structures, and the software allows for various measurements in this mode of imaging. A motor can be attached to the ultrasound probe to translate the probe over a defined range with a defined step-size, the software then recompiles the various 2D slices into a 3D volume. The 3D software allows for various measurements and imaging planes in this type of imaging. Power Doppler Mode: Power Doppler Mode is another tool used to analyze blood flow within vessels. Power Doppler Mode color codes flow according to a frequency Doppler shift and signal intensity. This results in less angle dependent color changes and more sensitivity. However, this mode is often used on smaller vessels having slower blood flow where velocity data isn t important. A Power sample box is placed around the vessels of interest and the system is able to display the Doppler shift in a color (orange to yellow) and is displayed over a B- Mode image. Various Power maps are available to optimize flow data presentation and are user selectable. Please note: Power, Color, and simultaneous B/PW modes display the flow data in the highest resolution/sensitivity as possible in a given physiological state. The B mode resolution will be slightly reduced by the required time sharing found on all ultrasound systems. So high resolution imaging should be done by B-Mode while flow data should be done by PW, Color and Power modes. Color Doppler Mode: Color Doppler Mode is similar to Power Doppler Mode and both utilize many of the same parameters to optimize their sensitivity. Color Doppler Mode displays the mean velocity while power displays a combination of Intensity and Doppler shift. Color mode displays the direction of the Doppler shift so the color is assigned accordingly with blood flowing towards the transducer on a scale from red to white and away from the transducer on a scale from blue to white. BART is an acronym commonly used to remember the colors associated with the direction of flow; Blue Away, Red Towards. Various Power maps are available to optimize flow data presentation and are user selectable. Contrast Mode: Contrast Mode is a mode very similar to B-Mode in that it is used to look at 2D anatomical structures. However, it includes specific analysis tools such as reference subtraction of two images. The result is a copper (or green) contrast overlay used to show the injected contrast agents. Contrast agents are used to enhance the visualization of blood flow in blood vessels including very small vessels such as capillaries. The contrast measurement tools not only produce enhanced flow visualization, but also provides a variety of quantification tools including various hemodynamic parameters. Imaging Guide: Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Page 2

4 Overview of the Parameters to be Optimized Presets All modes come with Presets. These are groups of imaging parameters that can be selected by pressing the Preset key. This reduces the time required to change imaging parameters individually. Users can define and name their own presets to optimize imaging for their own protocols by pressing the Save Preset key. This memorizes the current settings and requires a new name. These presets are probe-specific so each probe must have its own settings. Presets can be imported and exported between users. B-Mode Parameters: There are numerous parameters which can be set to optimize a B-Mode image, outlined below are a few key parameters, for a full list of parameters and how they affect B-Mode imaging please consult our Vevo 2100 Operators Manual. Display Maps Display maps are used to assign grayscale colors to detected ultrasound signal intensities; this allows optimized grayscale differentiation of specific anatomical targets. The Vevo 2100 system comes with numerous factory settings which can be selected from to provide the best quality images. Dynamic Range Dynamic range determines the range of grayscale intensities which can be displayed on an image. Together with the display maps, dynamic range can be adjusted to provide the best quality images with less noise. Focal Zones The Vevo 2100 system can have one, two, or three focal zones, with the spacing of the two and three focal zones being narrow or wide. Using additional focal zones provides an image which is focused over a broader depth, however increasing the focal zones will decrease the maximum frame rate. The number of focal zones and their placement should be selected based on the anatomical structures to be imaged and the amount of image motion. Line Density Line density allows the user to change the space between lines on the ultrasound image. Increasing the line density will increase the resolution, texture, and overall will increase the quality of the image; however it will decrease the maximum frame rate at which the system can image. Pulsed-Wave, Color and Power Doppler Mode Parameters: There are various parameters which can be set to optimize Doppler imaging in PW and Color Doppler imaging modes. Please see our Quick-Start Guide: Imaging in Color Doppler, Power Doppler and Power Doppler 3D Modes for an explanation of these parameters and how they will affect the image quality. Imaging Guide: Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Page 3

5 Selecting the Appropriate Probe for Imaging Transducer (probe) selection involves many factors. Higher frequencies provide the highest resolution but limit both penetration, imaging depth, and image width. Lower frequency probes have the best Doppler sensitivity with the least aliasing. So, using one probe for the best imaging while using another probe for the best Doppler isn t uncommon and highly recommended. To image the pancreas of a mouse the MS550D is the ideal probe to use, but the MS550S and MS400 may also be used. When imaging rat pancreas, B-Mode images should be acquired with the MS400 probe, while the Doppler images are best acquired with the MS250 or MS200 probes. Software Measurements When imaging the pancreas, some of the measurements that can be made are: Overall Pancreatic Blood Flow (Color Doppler, Power Doppler and Contrast Modes) To obtain flow within the pancreas using Color Doppler Mode and/or Power Doppler Mode, consult the Quick-Start Guide: Imaging in Color Doppler, Power Doppler and Power Doppler 3D Modes To obtain micro-vasculature flow within the pancreas, please see the VisualSonics Application Protocol Image Enhancement by Bolus Injection using untargeted MicroMarker contrast agents. Animal Preparation The animal should be anaesthetized to a level appropriate to a surgical procedure. Typically isofluorane is used at a concentration of 1.5-2%. After the animal is anaesthetized in the knock down box, it should be positioned on the platform in an optimal position to inject the target structure. Eye lubricant should be placed on each eye to prevent drying of the area. A small amount of ECG gel should be placed on the copper leads on the platforms and the paws taped to them, this will provide the ECG and respiratory physiology. A rectal probe should be inserted to monitor the temperature of the animal during the imaging session. The hair on the abdomen should be removed using a depilatory cream, such as Nair, and rinsed with water. Imaging Guide: Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Page 4

6 B-Mode Imaging To image the pancreas, the animal platform should be level and the probe should be placed transverse that is, the probe lies across the stomach from left to right as opposed to head-to-toe. Place the probe to the animal below the ribs and slightly to the left side. Using the Y-axis manipulator, move the animal to bring the kidney and spleen into view. When both are visible, the tissue in between the two is the pancreas (Figure 2). It has a very typical brick road or cobblestone appearance in ultrasound, but no defining wall; it is more of a diffuse structure. To ensure the liver is not in view, scan towards the head and the liver will appear it is darker in appearance in ultrasound and has a capsule enclosing it. Figure 2. Mouse pancreas in the transverse view showing relative position to kidney and spleen. Imaging Guide: Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Page 5

7 Color and Power Doppler Mode Imaging Turn on the Color (Figure 3) or Power Doppler (Figure 4) Mode and optimize the setting by the Preset key. Typical presets for pancreas might be: Very Slow, Slow Flow. Find the highest velocity flow in the Color regional area of interest. Figure 3. Color Doppler Mode image of the mouse pancreas in transverse view showing relative position to kidney and spleen. After the vessel is located in Color Doppler Mode, PW Doppler Mode can be used to quantify blood flood in this vessel. Press PW from the mode buttons and this will bring up the sample volume for PW Doppler Mode. Moving the cursor will move the sample volume. Place the sample volume in the vessel of interest. Adjust the Doppler Angle to align with blood flow. Press PW one more time to bring up the PW Doppler window. If the waveform is too big or too small, adjust the display by turning the Velocity dial. If the waveform is positioned too high or too low for the display window, it can be moved up or down with the Baseline button. Imaging Guide: Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Page 6

8 Figure 4. Power Doppler Mode image of the mouse pancreas in transverse view showing relative position to kidney and spleen. Need help? Call us toll-free at (North America) or (other regions) or contact us via at Imaging Guide: Guide to Small Animal Pancreatic Imaging using the Vevo 2100 Page 7

Beginner s Guide to Vevo 2100

Beginner s Guide to Vevo 2100 PN 12090 Beginner s Guide to Vevo 2100 December 2008 This Beginner s Guide to Vevo 2100 provides a quick overview of the Vevo 2100 imaging modes and provides key tips to allow users to start a study, acquire

More information

VEVO 1100 Sole Source Specifications

VEVO 1100 Sole Source Specifications VEVO 1100 Sole Source Specifications Table of Contents 1.SYSTEM... 2 1-1 Licensing for the User... 2 1-2 Safety... 2 1-3 Portability... 2 2.TRANSDUCERS... 2 2-1 MicroScan transducers... 2 3. IMAGING STATION...

More information

VisualSonics Application Protocol 3D Tumor Imaging using the MicroMarker Non-Targeted Contrast Agents

VisualSonics Application Protocol 3D Tumor Imaging using the MicroMarker Non-Targeted Contrast Agents VisualSonics Application Protocol 3D Tumor Imaging using the MicroMarker Non-Targeted Contrast Agents 1 Objective Part Number: 11902 The objective of the 3D Tumor Imaging using the MicroMarker Non-Targeted

More information

Vevo770 Ultrasound. Quick Operation Guide AIPT

Vevo770 Ultrasound. Quick Operation Guide AIPT Vevo770 Ultrasound Quick Operation Guide Vevo770 Probe head (707B/704) Preparation The acoustic membrane is very fragile and sensitive, it must be handled with care. The probe head should be filled with

More information

VEVO 2100 Sole Source Specifications

VEVO 2100 Sole Source Specifications VEVO 2100 Sole Source Specifications Table of Contents 1.SYSTEM... 2 1-1 Licensing for the User... 2 1-2 Safety... 2 1-3 Portability... 2 2.TRANSDUCERS... 2 2-1 MicroScan transducers... 2 3. IMAGING STATION...

More information

US 1.

US 1. US 1 Sample image: Normal pancreas seen on sonogram. Looking up from abdomen toward the head of the patient. The liver is in front of the pancreas. A vein draining the spleen is behind the pancreas http://www.radiologyinfo.org/photocat/photos.cfm?image=abdo-us-pancr.jpg&&subcategory=abdomen&&stop=9

More information

LOGIQ. V2 Ultrasound. Part of LOGIQ Vision Series. Imagination at work LOGIQ is a trademark of General Electric Company.

LOGIQ. V2 Ultrasound. Part of LOGIQ Vision Series. Imagination at work LOGIQ is a trademark of General Electric Company. TM LOGIQ V2 Ultrasound Part of LOGIQ Vision Series Imagination at work The brilliance of color. The simplicity of GE. Now you can add the advanced capabilities of color Doppler to patient care with the

More information

Quick Start Guide to the Vevo 770

Quick Start Guide to the Vevo 770 Quick Start Guide to the Vevo 770 PN 11422 This Quick Start Guide provides a quick overview of the Vevo 770 and provides tips to allow users to start using the instrument and its various modes. A detailed

More information

High Resolution Multi-modal in vivo Imaging Platform

High Resolution Multi-modal in vivo Imaging Platform High Resolution Multi-modal in vivo Imaging Platform The world s only customizable imaging platform combining ultra high frequency ultrasound and photoacoustics Experience the next generation of in vivo

More information

Semi-Interactive Console Layout and Instructions

Semi-Interactive Console Layout and Instructions Semi-Interactive Console Layout and Instructions A video of the SIC accompanies these instructions, which you may refer to for further understanding. Click here to open the video link. NOTE: The images

More information

Quick Start Guide to the Vevo 770

Quick Start Guide to the Vevo 770 Quick Start Guide to the Vevo 770 PN 11422 This Quick Start Guide provides a quick overview of the Vevo 770 and provides tips to allow users to start using the instrument and its various modes. A detailed

More information

Image Optimisation. -How to Get the Most out of your Echo Machine. Katarina Winkler, Biomedicinsk analytiker, Barnhjärtcentrum i Lund

Image Optimisation. -How to Get the Most out of your Echo Machine. Katarina Winkler, Biomedicinsk analytiker, Barnhjärtcentrum i Lund Image Optimisation -How to Get the Most out of your Echo Machine analytiker, Barnhjärtcentrum i Lund Image Optimisation Helpful tips to improve the 2D-Image and Doppler registrations. Common adjustments

More information

KNOBOLOGY. Jamie Ternes, RVT William Beaumont Hospital Royal Oak, Michigan. From the Beginning

KNOBOLOGY. Jamie Ternes, RVT William Beaumont Hospital Royal Oak, Michigan. From the Beginning KNOBOLOGY Jamie Ternes, RVT William Beaumont Hospital Royal Oak, Michigan From the Beginning Power On/Off Patient information Presets Probe selection Scan 1 Modalities /Gray scale/2d Pulse Wave Doppler

More information

MyLab Alpha Quick Start

MyLab Alpha Quick Start MyLab Alpha Quick Start Basic Setup The system has a 2.5 minute start time, but can be put in standby mode to allow for an 8 second boot time. From the patient data screen, click on the Menu button Then

More information

Certificate in Clinician Performed Ultrasound (CCPU)

Certificate in Clinician Performed Ultrasound (CCPU) Certificate in Clinician Performed Ultrasound (CCPU) Syllabus Physics Tutorial Physics Tutorial Purpose: Training: Assessments: This unit is designed to cover the theoretical and practical curriculum for

More information

Nuclear Associates and

Nuclear Associates and Nuclear Associates 84-317 and 84-317-7000 Multipurpose Tissue/Cyst Ultrasound Phantoms Users Manual February 2005 Manual No. 84-317-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in

More information

ACUSON P300 Ultrasound System Control Panel Diagram

ACUSON P300 Ultrasound System Control Panel Diagram www.siemens.com ACUSON P300 Ultrasound System Control Panel Diagram 1 Power On / Off 1 2 TGC Controls 3 Left Mouse Click 4 Action 4 7 5 Right Mouse Click 6 Trackball 2 3 6 5 8 7 Pointer 8 Freeze 16 Color

More information

Brilliance CT Big Bore.

Brilliance CT Big Bore. 1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.

More information

GE Healthcare. Agile Ultrasound. The Next Revolution in Ultrasound Imaging

GE Healthcare. Agile Ultrasound. The Next Revolution in Ultrasound Imaging Agile Ultrasound The Next Revolution in Ultrasound Imaging Abstract Diagnostic use of ultrasound has greatly expanded over the past couple of decades because it offers many advantages as an imaging modality.

More information

Visualsonics Vevo2100: Manual for Collecting RF Data. Updated: 10/29/2012 M Kurowski

Visualsonics Vevo2100: Manual for Collecting RF Data. Updated: 10/29/2012 M Kurowski Visualsonics Vevo2100: Manual for Collecting RF Data Updated: 10/29/2012 M Kurowski 1 : System Start-Up/Shutdown 1.1 The Visualsonics mvevo2100 system must be the only device plugged into an outlet; failure

More information

Echo Blaster 64/128 and LogicScan 64/128 Series Ultrasound Systems Echo Wave II Software User Manual

Echo Blaster 64/128 and LogicScan 64/128 Series Ultrasound Systems Echo Wave II Software User Manual Echo Blaster 64/128 and LogicScan 64/128 Series Ultrasound Systems Echo Wave II Software User Manual Internet page: http://www.telemed.lt Information E-mail: info@telemed.lt Ultrasound Medical Systems

More information

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI Figure 1. In the brain, the gray matter has substantially more blood vessels and capillaries than white matter. The magnified image on the right displays the rich vasculature in gray matter forming porous,

More information

n o r d i c B r a i n E x Tutorial DSC Module

n o r d i c B r a i n E x Tutorial DSC Module m a k i n g f u n c t i o n a l M R I e a s y n o r d i c B r a i n E x Tutorial DSC Module Please note that this tutorial is for the latest released nordicbrainex. If you are using an older version please

More information

How to Adjust the 16-Bit CLUT (Color Look Up Table) Editor in OsiriX

How to Adjust the 16-Bit CLUT (Color Look Up Table) Editor in OsiriX How to Adjust the 16-Bit CLUT (Color Look Up Table) Editor in OsiriX 1. Import the series you are interested in into the 2D/3D Viewer. I m using the OBELIX data set, slices 1-395, available for download

More information

Lecture 6: Medical imaging and image-guided interventions

Lecture 6: Medical imaging and image-guided interventions ME 328: Medical Robotics Winter 2019 Lecture 6: Medical imaging and image-guided interventions Allison Okamura Stanford University Updates Assignment 3 Due this Thursday, Jan. 31 Note that this assignment

More information

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space White Pixel Artifact Caused by a noise spike during acquisition Spike in K-space sinusoid in image space Susceptibility Artifacts Off-resonance artifacts caused by adjacent regions with different

More information

Zebrascope user manual

Zebrascope user manual Zebrascope user manual Date: 17/Mar/2016 Deliberate misuse of the Zebrascope to image tissues not approved on this protocol may introduce additional risks. These additional risks may include ventricular

More information

TELEMED. Echo Wave II Software LB-2. Ultrasound Consolle. User Manual

TELEMED. Echo Wave II Software LB-2. Ultrasound Consolle. User Manual SmartUs - ClarUs - LS128 - LS64 - Echo Blaster 64 - Echo Blaster 128 Echo Wave II Software LB-2 Ultrasound Consolle User Manual T E L E M E D Ultrasound Medical Systems Telemed Medical Systems WEB: Telemed

More information

Getting Started. What is SAS/SPECTRAVIEW Software? CHAPTER 1

Getting Started. What is SAS/SPECTRAVIEW Software? CHAPTER 1 3 CHAPTER 1 Getting Started What is SAS/SPECTRAVIEW Software? 3 Using SAS/SPECTRAVIEW Software 5 Data Set Requirements 5 How the Software Displays Data 6 Spatial Data 6 Non-Spatial Data 7 Summary of Software

More information

45 µm polystyrene bead embedded in scattering tissue phantom. (a,b) raw images under oblique

45 µm polystyrene bead embedded in scattering tissue phantom. (a,b) raw images under oblique Phase gradient microscopy in thick tissue with oblique back-illumination Tim N Ford, Kengyeh K Chu & Jerome Mertz Supplementary Figure 1: Comparison of added versus subtracted raw OBM images 45 µm polystyrene

More information

Quick reference guide. MyLab50

Quick reference guide. MyLab50 Quick reference guide MyLab50 Quick reference guide MyLab50 Introduction This Quick Guide describes only the basic operating procedures for the diagnostic ultrasound MyLab50 Family, named in the following

More information

Red Orange the reflected ray. Yellow Green and the normal. Blue Indigo line. Colours of visible reflection

Red Orange the reflected ray. Yellow Green and the normal. Blue Indigo line. Colours of visible reflection distance the carrying the moves away from rest position Brightness Loudness The angle between the incident ray and the normal line Amplitude Amplitude of a light Amplitude of a sound incidence Angle between

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Quick Guide. EPIQ 7 Ultrasound System. English

Quick Guide. EPIQ 7 Ultrasound System. English Quick Guide EPIQ 7 Ultrasound System English Contents For easy reference, each section is color-coded. The color of each section title correlates with the color band on each page in that section. Controls

More information

This is your imaging dashboard. There is a bevy of valuable information to read from here. Let s break it up.

This is your imaging dashboard. There is a bevy of valuable information to read from here. Let s break it up. The Volume Strikes Back Press the (4D) button. This is your imaging dashboard. There is a bevy of valuable information to read from here. Let s break it up. In the Left GUI (Graphic User Interface), you

More information

An image analysis method for quantification of hepatic perfusion based on contrast-enhanced ultrasound

An image analysis method for quantification of hepatic perfusion based on contrast-enhanced ultrasound J. Biomedical Science and Engineering, 2008, 1, 116-120 An image analysis method for quantification of hepatic perfusion based on contrast-enhanced ultrasound imaging Yi-xin Li, Fang Yang & Ning Gu* State

More information

MR Advance Techniques. Vascular Imaging. Class III

MR Advance Techniques. Vascular Imaging. Class III MR Advance Techniques Vascular Imaging Class III 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology

More information

Gamepad Controls. Figure 1: A diagram of an Xbox controller. Figure 2: A screenshot of the BodyViz Controller Panel. BodyViz 3 User Manual 1

Gamepad Controls. Figure 1: A diagram of an Xbox controller. Figure 2: A screenshot of the BodyViz Controller Panel. BodyViz 3 User Manual 1 BodyViz User Manual Gamepad Controls The first step in becoming an expert BodyViz user is to get acquainted with the Xbox gamepad, also known as a controller, and the BodyViz Controller Panel. These can

More information

Range Imaging Through Triangulation. Range Imaging Through Triangulation. Range Imaging Through Triangulation. Range Imaging Through Triangulation

Range Imaging Through Triangulation. Range Imaging Through Triangulation. Range Imaging Through Triangulation. Range Imaging Through Triangulation Obviously, this is a very slow process and not suitable for dynamic scenes. To speed things up, we can use a laser that projects a vertical line of light onto the scene. This laser rotates around its vertical

More information

Advanced Image Reconstruction Methods for Photoacoustic Tomography

Advanced Image Reconstruction Methods for Photoacoustic Tomography Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic

More information

A guide to animation effects for the MAC 2000 Performance

A guide to animation effects for the MAC 2000 Performance A guide to animation effects for the MAC 2000 Performance A guide to animation effects for the MAC 2000 Performance General Introduction The art of gobo animation has been around for many years. The adaptation

More information

1 About the Terason Ultrasound System

1 About the Terason Ultrasound System 1 About the Terason Ultrasound System Overview The Terason Ultrasound System is an easy-to-use, portable ultrasound system that produces high resolution images. The microminiaturized ultrasound system

More information

John R. Mandeville Senior Consultant NDICS, Norwich, CT Jesse A. Skramstad President - NDT Solutions Inc., New Richmond, WI

John R. Mandeville Senior Consultant NDICS, Norwich, CT Jesse A. Skramstad President - NDT Solutions Inc., New Richmond, WI Enhanced Defect Detection on Aircraft Structures Automatic Flaw Classification Software (AFCS) John R. Mandeville Senior Consultant NDICS, Norwich, CT Jesse A. Skramstad President - NDT Solutions Inc.,

More information

Beyond Performance and Value

Beyond Performance and Value Putting the back in ultrasound Beyond Performance and Value Esaote s new ultra-performance MyLab 9 exp ultrasound system is designed to support a full range of shared service diagnostic imaging environments.

More information

Module 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution.

Module 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution. MRES 7005 - Fast Imaging Techniques Module 5: Dynamic Imaging and Phase Sharing (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review Improving Temporal Resolution True-FISP (I) True-FISP (II) Keyhole

More information

VisualSonics Vevo 770 High-Resolution Imaging System. Operator Manual

VisualSonics Vevo 770 High-Resolution Imaging System. Operator Manual VisualSonics Vevo 770 High-Resolution Imaging System Operator Manual VisualSonics Vevo 770 High-Resolution Imaging System Operator Manual January 2006 PN 11322 Revision 1.3 V1.3 VisualSonics Vevo 770

More information

연구용유방초음파질관리 원광대학병원김혜원

연구용유방초음파질관리 원광대학병원김혜원 연구용유방초음파질관리 원광대학병원김혜원 Why US QC? Quality control (QC) testing of ultrasound scanners is important to verify the proper and consistent operation of these devices. main goal ; quality improvement Guidelines

More information

Display. Introduction page 67 2D Images page 68. All Orientations page 69 Single Image page 70 3D Images page 71

Display. Introduction page 67 2D Images page 68. All Orientations page 69 Single Image page 70 3D Images page 71 Display Introduction page 67 2D Images page 68 All Orientations page 69 Single Image page 70 3D Images page 71 Intersecting Sections page 71 Cube Sections page 72 Render page 73 1. Tissue Maps page 77

More information

Discontinued Products

Discontinued Products Discontinued Products January 2019 CIRS was founded in 1983 to improve upon existing tissue simulation methodology and provide quantitative reference standards for Computed Tomography. Today CIRS is recognized

More information

Motion Analysis. Motion analysis. Now we will talk about. Differential Motion Analysis. Motion analysis. Difference Pictures

Motion Analysis. Motion analysis. Now we will talk about. Differential Motion Analysis. Motion analysis. Difference Pictures Now we will talk about Motion Analysis Motion analysis Motion analysis is dealing with three main groups of motionrelated problems: Motion detection Moving object detection and location. Derivation of

More information

Learn Image Segmentation Basics with Hands-on Introduction to ITK-SNAP. RSNA 2016 Courses RCB22 and RCB54

Learn Image Segmentation Basics with Hands-on Introduction to ITK-SNAP. RSNA 2016 Courses RCB22 and RCB54 Learn Image Segmentation Basics with Hands-on Introduction to ITK-SNAP RSNA 2016 Courses RCB22 and RCB54 RCB22 Mon, Nov 28 10:30-12:00 PM, Room S401CD RCB54 Thu, Dec 1 2:30-4:30 PM, Room S401CD Presenters:

More information

UGEO G60 Performance in style

UGEO G60 Performance in style CT-UGEO-G60-CWW-MCI-0731-EN Performance in style 2012 Samsung Electronics All Rights Reserved. Samsung Electronics Reserves The Right To Modify The Design, Packaging, Specifications And Features Shown

More information

Medical Images Analysis and Processing

Medical Images Analysis and Processing Medical Images Analysis and Processing - 25642 Emad Course Introduction Course Information: Type: Graduated Credits: 3 Prerequisites: Digital Image Processing Course Introduction Reference(s): Insight

More information

What is it? How does it work? How do we use it?

What is it? How does it work? How do we use it? What is it? How does it work? How do we use it? Dual Nature http://www.youtube.com/watch?v=dfpeprq7ogc o Electromagnetic Waves display wave behavior o Created by oscillating electric and magnetic fields

More information

icatvision Quick Reference

icatvision Quick Reference icatvision Quick Reference Navigating the i-cat Interface This guide shows how to: View reconstructed images Use main features and tools to optimize an image. REMINDER Images are displayed as if you are

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

KULLEGG MARIA REGINA BOYS SECONDARY MOSTA HALF-YEARLY EXAMINATIONS 2012/2013. SUBJECT: PHYSICS Form 4 TIME: 1 HR 30 MIN NAME :

KULLEGG MARIA REGINA BOYS SECONDARY MOSTA HALF-YEARLY EXAMINATIONS 2012/2013. SUBJECT: PHYSICS Form 4 TIME: 1 HR 30 MIN NAME : KULLEGG MARIA REGINA BOYS SECONDARY MOSTA HALF-YEARLY EXAMINATIONS 2012/2013 SUBJECT: PHYSICS Form 4 TIME: 1 HR 30 MIN NAME : CLASS : INDEX NO : Track 2 Answer ALL questions in the spaces provided on the

More information

Acoustic Field Comparison of High Intensity Focused Ultrasound by using Experimental Characterization and Finite Element Simulation

Acoustic Field Comparison of High Intensity Focused Ultrasound by using Experimental Characterization and Finite Element Simulation Acoustic Field Comparison of High Intensity Focused Ultrasound by using Experimental Characterization and Finite Element Simulation J. L. Teja, A. Vera, and L. Leija Department of Electrical Engineering,

More information

SONOS 7500/5500 Using 3-Dimensional and BiPlane Imaging And Other System Changes for Software Revision D.1

SONOS 7500/5500 Using 3-Dimensional and BiPlane Imaging And Other System Changes for Software Revision D.1 SONOS 7500/5500 Using 3-Dimensional and BiPlane Imaging And Other System Changes for Software Revision D.1 User s Guide Using 3-Dimensional and BiPlane Imaging And Other System Changes for Software Revision

More information

Magnetic Resonance Elastography (MRE) of Liver Disease

Magnetic Resonance Elastography (MRE) of Liver Disease Magnetic Resonance Elastography (MRE) of Liver Disease Authored by: Jennifer Dolan Fox, PhD VirtualScopics Inc. jennifer_fox@virtualscopics.com 1-585-249-6231 1. Overview of MRE Imaging MRE is a magnetic

More information

Object Identification in Ultrasound Scans

Object Identification in Ultrasound Scans Object Identification in Ultrasound Scans Wits University Dec 05, 2012 Roadmap Introduction to the problem Motivation Related Work Our approach Expected Results Introduction Nowadays, imaging devices like

More information

EnSite NavX Navigation and Visualization Technology EnSite Fusion Registration Module Procedure Guide

EnSite NavX Navigation and Visualization Technology EnSite Fusion Registration Module Procedure Guide EnSite NavX Navigation and Visualization Technology EnSite Fusion Registration Module Procedure Guide The EnSite Fusion Module enables Dynamic Registration, allowing a DIF model (a surface rendering of

More information

UNIVERSITY OF SOUTHAMPTON

UNIVERSITY OF SOUTHAMPTON UNIVERSITY OF SOUTHAMPTON PHYS2007W1 SEMESTER 2 EXAMINATION 2014-2015 MEDICAL PHYSICS Duration: 120 MINS (2 hours) This paper contains 10 questions. Answer all questions in Section A and only two questions

More information

RSNA, /rg

RSNA, /rg RSNA, 2015 10.1148/rg.2015140320 Appendix As noted in the main article, DICOM image files cannot be used directly for 3D printing; further steps are necessary to make them readable by 3D printers. The

More information

Phased Array Assisted Manual Nozzle Inspection Solution with Data Archiving Capability

Phased Array Assisted Manual Nozzle Inspection Solution with Data Archiving Capability 19 th World Conference on Non-Destructive Testing 2016 Phased Array Assisted Manual Nozzle Inspection Solution with Data Archiving Capability Jason HABERMEHL 1, Nicolas BADEAU 1, Martin ST-LAURENT 1, Guy

More information

VJ Technologies Inspection Services Division. By Bob Maziuk, February 2015

VJ Technologies Inspection Services Division. By Bob Maziuk, February 2015 White Paper on the Inspection of Underground Electrical Conductors, Insulation, and Splices in Oil-Filled Piping Using High Energy Computed Tomography (CT) VJ Technologies Inspection Services Division

More information

The Supreme 3D/4D Ultrasound. The Supreme 3D/4D Ultrasound. CT-V20-ICM

The Supreme 3D/4D Ultrasound. The Supreme 3D/4D Ultrasound.   CT-V20-ICM The Supreme 3D/4D Ultrasound The Supreme 3D/4D Ultrasound www.medison.com info@medison.com CT-V20-ICM-06.20.2008 The Supreme 3D/4D Ultrasound Since launching the first commercially available 3D ultrasound

More information

Chapter 24. Creating Surfaces for Displaying and Reporting Data

Chapter 24. Creating Surfaces for Displaying and Reporting Data Chapter 24. Creating Surfaces for Displaying and Reporting Data FLUENT allows you to select portions of the domain to be used for visualizing the flow field. The domain portions are called surfaces, and

More information

CP Generalize Concepts in Abstract Multi-dimensional Image Model Component Semantics. David Clunie.

CP Generalize Concepts in Abstract Multi-dimensional Image Model Component Semantics. David Clunie. CP-1390 - Generalize Concepts in Abstract Multi-dimensional Image Model Semantics Page 1 STATUS Date of Last Update Person Assigned Submitter Name Submission Date Assigned 2014/06/09 David Clunie mailto:dclunie@dclunie.com

More information

Tomographic Reconstruction

Tomographic Reconstruction Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)

More information

Beyond Performance and Value

Beyond Performance and Value Putting the back in ultrasound Beyond Performance and Value Esaote s new ultra-performance MyLab 9 exp ultrasound system is designed to support a full range of shared service diagnostic imaging environments.

More information

Exploration and Study of MultiVolume Image Data using 3D Slicer

Exploration and Study of MultiVolume Image Data using 3D Slicer Exploration and Study of MultiVolume Image Data using 3D Slicer Meysam Torabi and Andriy Fedorov torabi@bwh.harvard.edu, fedorov@bwh.harvard.edu Surgical Navigation and Robotics Lab and Surgical Planning

More information

Resona 7. New Waves in Ultrasound Innovation. Premium Ultrasound System

Resona 7. New Waves in Ultrasound Innovation. Premium Ultrasound System Resona 7 Premium Ultrasound System New Waves in Ultrasound Innovation With over 20 years of experience, Mindray hosts a wide range of ultrasound imaging solutions including cart-based and portable systems.

More information

SONOLINE G40 ultrasound system

SONOLINE G40 ultrasound system SONOLINE G40 ultrasound system Quick Scan Card: General/OB For system software version 2.0 Press the Power Switch on the upper left side of the keyboard. A blue light will indicate the power is on. Getting

More information

Linescan System Design for Robust Web Inspection

Linescan System Design for Robust Web Inspection Linescan System Design for Robust Web Inspection Vision Systems Design Webinar, December 2011 Engineered Excellence 1 Introduction to PVI Systems Automated Test & Measurement Equipment PC and Real-Time

More information

Product information. Hi-Tech Electronics Pte Ltd

Product information. Hi-Tech Electronics Pte Ltd Product information Introduction TEMA Motion is the world leading software for advanced motion analysis. Starting with digital image sequences the operator uses TEMA Motion to track objects in images,

More information

HOOK-X Series. Operation manual ENGLISH. HOOK-4x HOOK-5x HOOK-7x

HOOK-X Series. Operation manual ENGLISH. HOOK-4x HOOK-5x HOOK-7x HOOK-X Series Operation manual ENGLISH HOOK-4x HOOK-5x HOOK-7x Lowrance and Navico are registered trademarks of Navico. Navico may find it necessary to change or end our policies, regulations and special

More information

BCC Multi Stretch Wipe

BCC Multi Stretch Wipe BCC Multi Stretch Wipe The BCC Multi Stretch Wipe is a radial wipe with three additional stretch controls named Taffy Stretch. The Taffy Stretch parameters do not significantly impact render times. The

More information

Unit 21 - Creating a Navigation Bar in Macromedia Fireworks

Unit 21 - Creating a Navigation Bar in Macromedia Fireworks Unit 21 - Creating a Navigation Bar in Macromedia Fireworks Items needed to complete the Navigation Bar: Unit 21 - House Style Unit 21 - Graphics Sketch Diagrams Document ------------------------------------------------------------------------------------------------

More information

GE Healthcare. Vivid 7 Dimension 08 Cardiovascular ultrasound system

GE Healthcare. Vivid 7 Dimension 08 Cardiovascular ultrasound system GE Healthcare Vivid 7 Dimension 08 Cardiovascular ultrasound system ltra Definition. Technology. Performance. Start with a system that s proven its worth in LV quantification and 4D imaging. Then add even

More information

SIVIC GUI Overview. SIVIC GUI Layout Overview

SIVIC GUI Overview. SIVIC GUI Layout Overview SIVIC GUI Overview SIVIC GUI Layout Overview At the top of the SIVIC GUI is a row of buttons called the Toolbar. It is a quick interface for loading datasets, controlling how the mouse manipulates the

More information

Data Loading & 3D Visualization

Data Loading & 3D Visualization Neuroimage Analysis Center Data Loading & 3D Visualization Sonia Pujol, Ph.D. Surgical Planning Laboratory Harvard Medical School Leonardo da Vinci (1452-1519), Virgin and Child Alte Pinakothek, München

More information

MEASURING SURFACE CURRENTS USING IR CAMERAS. Background. Optical Current Meter 06/10/2010. J.Paul Rinehimer ESS522

MEASURING SURFACE CURRENTS USING IR CAMERAS. Background. Optical Current Meter 06/10/2010. J.Paul Rinehimer ESS522 J.Paul Rinehimer ESS5 Optical Current Meter 6/1/1 MEASURING SURFACE CURRENTS USING IR CAMERAS Background Most in-situ current measurement techniques are based on sending acoustic pulses and measuring the

More information

AN ULTRA-SOUND AND DOPPLER MEDICAL IMAGE ANALYSIS FRAMEWORK

AN ULTRA-SOUND AND DOPPLER MEDICAL IMAGE ANALYSIS FRAMEWORK AN ULTRA-SOUND AND DOPPLER MEDICAL IMAGE ANALYSIS FRAMEWORK Ioan Stoian*, Ovidiu Dancea*, Mihaela Gordan**, Adina Stoian***, Gratian Botis**** * S.C. IPA S.A. Cluj-Napoca Subsidiary 109 Republicii Str.,400489

More information

How to work with text

How to work with text How to work with text Adobe Flash Professional lets you add text to a Flash application in two formats: You can add Text Layout Framework (TLF) text. You can add Classic text. Using the Text Layout Framework

More information

AORTA CTA VPMC-12419

AORTA CTA VPMC-12419 AORTA CTA VPMC-12419 Workflow Overview: The Aorta can be post-processed in various ways. Auto Bone Removal and Vessel Pick provide a quick overview of the entire Aorta. Vessel Probe creates a centerline

More information

Contents. Introduction... iii. CHAPTER 1 - Getting Started... 5

Contents. Introduction... iii. CHAPTER 1 - Getting Started... 5 Transducer Characterization User s Guide Version 5 5 1 build 12 or later UTEX Scientific Instruments Inc support@utex com Winspect Transducer Characterization User s Guide Contents Table of Contents and

More information

IMPAX Volume Viewing 3D Visualization & Segmentation

IMPAX Volume Viewing 3D Visualization & Segmentation Getting started guide IMPAX Volume Viewing 3D Visualization & Segmentation This guide outlines the basic steps to perform and manipulate a 3D reconstruction of volumetric image data using IMPAX Volume

More information

Sizing and evaluation of planar defects based on Surface Diffracted Signal Loss technique by ultrasonic phased array

Sizing and evaluation of planar defects based on Surface Diffracted Signal Loss technique by ultrasonic phased array Sizing and evaluation of planar defects based on Surface Diffracted Signal Loss technique by ultrasonic phased array A. Golshani ekhlas¹, E. Ginzel², M. Sorouri³ ¹Pars Leading Inspection Co, Tehran, Iran,

More information

Outline. Introduction to photoacoustic computed tomography (PACT) Imaging models and iterative image reconstruction. Success with small animal imaging

Outline. Introduction to photoacoustic computed tomography (PACT) Imaging models and iterative image reconstruction. Success with small animal imaging Outline Advantages of PACT Photoacoustic Computed Tomography with Applications to Breast Imaging Mark A. Anastasio Department of Biomedical Engineering Washington University in St. Louis St. Louis, MO

More information

Direct Image Viewer User Tutorial

Direct Image Viewer User Tutorial Direct Image Viewer User Tutorial A guide to understanding the functions and features within the Direct Image Viewer to best present the data and to get the most understanding from the Direct Image logs.

More information

step into the future Ultrasound Diagnostic Imaging System

step into the future Ultrasound Diagnostic Imaging System step into the future Ultrasound Diagnostic Imaging System Moving into a new era for ultrasound diagnostics. With a proven track record in medical imaging, FUJIFILM has partnered with SonoSite, an innovator

More information

In order to get the G.C.S.E. grade you are capable of, you must make your own revision notes using your Physics notebook.

In order to get the G.C.S.E. grade you are capable of, you must make your own revision notes using your Physics notebook. In order to get the G.C.S.E. grade you are capable of, you must make your own revision notes using your Physics notebook. When summarising notes, use different colours and draw diagrams/pictures. If you

More information

10/16/14. Ultrasound Physics & Instrumentation 5 th Edition. License Agreement. Chapter Outline. All Copyright Laws Apply.

10/16/14. Ultrasound Physics & Instrumentation 5 th Edition. License Agreement. Chapter Outline. All Copyright Laws Apply. Ultrasound Physics & Instrumentation 5 th Edition Companion Presentation Frank R. Miele Pegasus Lectures, Inc. License Agreement This presentation is the sole property of Pegasus Lectures, Inc. No part

More information

BCC Rays Ripply Filter

BCC Rays Ripply Filter BCC Rays Ripply Filter The BCC Rays Ripply filter combines a light rays effect with a rippled light effect. The resulting light is generated from a selected channel in the source image and spreads from

More information

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,

More information

Quantitative Three-Dimensional Imaging of the Posterior Segment with the Heidelberg Retina Tomograph

Quantitative Three-Dimensional Imaging of the Posterior Segment with the Heidelberg Retina Tomograph Quantitative Three-Dimensional Imaging of the Posterior Segment with the Heidelberg Retina Tomograph Heidelberg Engineering GmbH, Heidelberg, Germany Contents 1 Introduction... 1 2 Confocal laser scanning

More information

USER MANUAL VP PLANNING

USER MANUAL VP PLANNING USER MANUAL VP PLANNING 1.0 Mobile application ios Visible Patient SAS RCS Strasbourg TI 794 458 125 1 place de l'ho pital 67000 Strasbourg, FRANCE Share capital: 58330 Tel.: 33 (0)3 88 11 90 00 Fax: 33

More information

Supplementary Figure 1

Supplementary Figure 1 Supplementary Figure 1 BOLD and CBV functional maps showing EPI versus line-scanning FLASH fmri. A. Colored BOLD and CBV functional maps are shown in the highlighted window (green frame) of the raw EPI

More information

Advanced Copy Pro 9.0 Plugin User Guide

Advanced Copy Pro 9.0 Plugin User Guide Advanced Copy Pro 9.0 Plugin User Guide UPDATED ON 6/26/2018 PlanSwift Authored by: Dave Hansen 1 Table of Contents Overview... 3 Purchasing and Installation... 4 Purchasing Plugins... 4 Installation and

More information