How does the ROI affect the thresholding?

Size: px
Start display at page:

Download "How does the ROI affect the thresholding?"

Transcription

1 How does the ROI affect the thresholding? Micro-computed tomography can be applied for the visualization of the inner structure of a material or biological tissue in a non-destructive manner. Besides visualization, image analysis is an important topic in micro-ct, to obtain quantitative parameters from scanned datasets. SkyScan has developed analysis software CTAn to serve this purpose to a large extend. An important issue to consider in CTAn is the application of a region of interest (ROI). This application note will give an example to illustrate the importance of this ROI and the correct use of it. Intro CTAn performs quantitative measurements both of densitometry (voxel attenuation coefficient or calibrated density) and of morphometry. The morphometry image analysis is based on binary images. Binary images contain only black and white pixels, representing the non-selected or selected pixels respectively. The process of selecting pixels is called binarization and in CTAn this is performed by means of segmentation or thresholding. Simple global or adaptive methods allow for easy segmentation. Acquisition of quantitative data can be greatly influenced by defining the region of interest. CTAn offers a number of tools for drawing ROI s with a wide range of shapes. The importance of this ROI needs to be emphasized. Methodology Let s consider the following illustrative dataset, a sand sample. The dataset has been acquired with a SkyScan1172 with a 10MP Hamamatsu camera. The sample was scanned at 44kV with a 0.5 mm Al filter at 360 and reconstructed with NRecon. This 8-bit reconstructed dataset is loaded in CTAn and has 256 different greyvalues in the range [0,255]. 1

2 Figure 1. Illustrative sand sample in CTAn. The table below (Table 1) shows the different steps in 2 comparative methods. In the first method an ROI is drawn, a new dataset is saved and reloaded from this ROI, and 2 different thresholding values have been applied to select the air in between the sand particles. The second method follows the same steps, but there is an additional step in between. It not only loads the new dataset from the ROI, but also the ROI itself. Understanding the impact of this key step, and how this affects the results (in step 4 and 5) is crucial for analysis. Table 1. Comparison of 2 analysis methods Method 1 Method 2 Step 1: draw an ROI Step 1: draw an ROI 2

3 STEP 2: save new dataset from ROI, load (this) new dataset STEP 2: save new dataset from ROI, load (this) new dataset STEP 3: Load the ROI (again) STEP 3: Apply thresholding [1,25] STEP 4: Thresholding [1,25] STEP 4: thresholding [0,25] STEP 5: thresholding [0,25] 3

4 [0,25] [1,25] Result & discussion CTAn looks at an image as a square collection of pixels. When you are loading an image with a different shape, CTAn will automatically generate a square picture from this shape, by including a number of black pixels with grey value of 0. When reloading the dataset in step 2, the round image has now become square. When thresholding the range [1,25] these pixels are not included, but when the boundary of the thresholding range is changed to [0,25], these pixels are also binarized, showing that these pixels have a greyvalue of 0. In the second method, not only the dataset has been reloaded, but also the ROI. This is a way of defining the image boundary. The image is now round. The pixels in green are not part of the image, which is seen when applying the thresholding ranges [1,25] and [0,25]. The image is now defined within the boundaries of the ROI. This difference has a consequence in analysis. When analysing the percentage of air in this sand sample for example, one can be interested in the object percentage. In the table below, Table 2, this parameter is calculated for each of the four possible outcomes from Table 1. When applying a threshold of [1,25], the object percentage is 39.8%. When neglecting to reload the ROI, an artificially low value of 30.7% is obtained. Although the absolute amount of white pixels remains the same This value is lower because the total ROI area is larger. There are now less white pixels with respect to the ROI. For a thresholding value of [0,25], the object percentage would be 49.8%.. When the reloading of the ROI would have been neglected, this would lead to an artificially high value of 61.3 %. The pixels surrounding the round sand sample are now also included in the image, and considered as part of the object, which is obviously a false statement. Table 2. analysis of different ROI & thresholding Not reloading ROI ROI Area = micron ² Reloading ROI ROI Area = micron ² 30.7% 39.8 % 61.3% 49.8 % 4

5 Conclusion In this application note the importance of understanding the ROI has been pointed out. By means of an example dataset it has been shown that neglecting to reload the ROI may result in an inappropriate amount of pixels being falsely considered as part of the image, which can lead to incorrect analysis results. Obtained values can be artificially high or low, depending on the protocol that is used. In order to prevent this from happening, one needs to carefully address this issue and design an appropriate ROI. SkyScan, Belgium 5

Applying Hounsfield unit density calibration in SkyScan CT-analyser

Applying Hounsfield unit density calibration in SkyScan CT-analyser 1 Bruker-microCT Method note Applying Hounsfield unit density calibration in SkyScan CT-analyser Hounsfield units (HU) are a standard unit of x-ray CT density, in which air and water are ascribed values

More information

Enhanced material contrast by dual-energy microct imaging

Enhanced material contrast by dual-energy microct imaging Enhanced material contrast by dual-energy microct imaging Method note Page 1 of 12 2 Method note: Dual-energy microct analysis 1. Introduction 1.1. The basis for dual energy imaging Micro-computed tomography

More information

Adaptive thresholding. in CTAn

Adaptive thresholding. in CTAn Adaptive thresholding in CTAn Method note Page 1 of 8 2 Bruker-MicroCT method note:adaptive thresholding in CTAn Introduction 2D or 3D morphometric analysis of micro-ct images always requires binarization

More information

Using X-Ray Micro Tomography to assess the quality of Packaging Foil

Using X-Ray Micro Tomography to assess the quality of Packaging Foil Using X-Ray Micro Tomography to assess the quality of Packaging Foil Martin Koster 1, Gerard van Dalen 2, Robert Hoeve 3 1 Unilever Research 2 Unilever Research 3 Unilever Research, Olivier van Noortlaaan

More information

Computed tomography (Item No.: P )

Computed tomography (Item No.: P ) Computed tomography (Item No.: P2550100) Curricular Relevance Area of Expertise: Biology Education Level: University Topic: Modern Imaging Methods Subtopic: X-ray Imaging Experiment: Computed tomography

More information

Computed tomography of simple objects. Related topics. Principle. Equipment TEP Beam hardening, artefacts, and algorithms

Computed tomography of simple objects. Related topics. Principle. Equipment TEP Beam hardening, artefacts, and algorithms Related topics Beam hardening, artefacts, and algorithms Principle The CT principle is demonstrated with the aid of simple objects. In the case of very simple targets, only a few images need to be taken

More information

Recognition and Measurement of Small Defects in ICT Testing

Recognition and Measurement of Small Defects in ICT Testing 19 th World Conference on Non-Destructive Testing 2016 Recognition and Measurement of Small Defects in ICT Testing Guo ZHIMIN, Ni PEIJUN, Zhang WEIGUO, Qi ZICHENG Inner Mongolia Metallic Materials Research

More information

DUE to beam polychromacity in CT and the energy dependence

DUE to beam polychromacity in CT and the energy dependence 1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact

More information

Development of an automated bond verification system for advanced electronic packages

Development of an automated bond verification system for advanced electronic packages Development of an automated bond verification system for advanced electronic packages Michel Darboux, Jean-Marc Dinten LETI (CEA - Technologies AvancSes) - DSYS/SCSI CENG, 17 rue des Martyrs F38054 Grenoble

More information

The user's guide. CT-Analyser. Contents. (version )

The user's guide. CT-Analyser. Contents. (version ) CT-Analyser (version 1.8.1.3) The user's guide Contents 1. Introduction 2 2. First start-up of the program 2 3. Set-up of preferences 3 4. Opening files 4 4.1. Crossection image window 8 4.2. Dataset bar

More information

Oxygen XRay Step by Step

Oxygen XRay Step by Step 1 Oxygen XRay Step by Step 14 October, 2008 OctoNus Software 2 1 Content OXYGEN XRAY STEP BY STEP... 1 1 CONTENT... 2 2 INSTRUCTION... 3 STEP 1. LOAD PHOTOS... 3 STEP 2. CREATE 3D MODELS MANUALLY OR AUTOMATICALLY...

More information

FINDING THE TRUE EDGE IN CTA

FINDING THE TRUE EDGE IN CTA FINDING THE TRUE EDGE IN CTA by: John A. Rumberger, PhD, MD, FACC Your patient has chest pain. The Cardiac CT Angiography shows plaque in the LAD. You adjust the viewing window trying to evaluate the stenosis

More information

Digital Volume Correlation for Materials Characterization

Digital Volume Correlation for Materials Characterization 19 th World Conference on Non-Destructive Testing 2016 Digital Volume Correlation for Materials Characterization Enrico QUINTANA, Phillip REU, Edward JIMENEZ, Kyle THOMPSON, Sharlotte KRAMER Sandia National

More information

Manual for Bruker-microCT CT-Analyser v

Manual for Bruker-microCT CT-Analyser v Setting a high value for number of projections allows a set of MIP slices to be taken through a selected part or all of the dataset. This can be a useful way of displaying quickly the structure of a scanned

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

Calibrated Image Acquisition for Multi-view 3D Reconstruction

Calibrated Image Acquisition for Multi-view 3D Reconstruction Calibrated Image Acquisition for Multi-view 3D Reconstruction Sriram Kashyap M S Guide: Prof. Sharat Chandran Indian Institute of Technology, Bombay April 2009 Sriram Kashyap 3D Reconstruction 1/ 42 Motivation

More information

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY David K. Lehmann 1, Kathleen Brockdorf 1 and Dirk Neuber 2 1 phoenix x-ray Systems + Services Inc. St. Petersburg, FL, USA 2 phoenix x-ray Systems + Services

More information

Quality control phantoms and protocol for a tomography system

Quality control phantoms and protocol for a tomography system Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing

More information

Micro-CT Methodology Hasan Alsaid, PhD

Micro-CT Methodology Hasan Alsaid, PhD Micro-CT Methodology Hasan Alsaid, PhD Preclinical & Translational Imaging LAS, PTS, GlaxoSmithKline 20 April 2015 Provide basic understanding of technical aspects of the micro-ct Statement: All procedures

More information

Method note. Page 1 of 7

Method note. Page 1 of 7 Advanced image coregistration in Dataviewer Method note Page 1 of 7 2 Bruker microct method note: Advanced image co-registration Introduction The 2D and 3D image co-registration in Dataviewer provides

More information

Evaluation of specific network properties of paper using computer tomography and the latest Avizo Fire analysing tools

Evaluation of specific network properties of paper using computer tomography and the latest Avizo Fire analysing tools Evaluation of specific network properties of paper using computer tomography and the latest Avizo Fire analysing tools Tobias Wolfinger 1, Peter Westenberger 2 1 WEIDMANN Electrical Technology AG, Neue

More information

CHAPTER 2. Morphometry on rodent brains. A.E.H. Scheenstra J. Dijkstra L. van der Weerd

CHAPTER 2. Morphometry on rodent brains. A.E.H. Scheenstra J. Dijkstra L. van der Weerd CHAPTER 2 Morphometry on rodent brains A.E.H. Scheenstra J. Dijkstra L. van der Weerd This chapter was adapted from: Volumetry and other quantitative measurements to assess the rodent brain, In vivo NMR

More information

Micro-CT in geosciences: Getting the inside story. Wesley De Boever

Micro-CT in geosciences: Getting the inside story. Wesley De Boever Micro-CT in geosciences: Getting the inside story Wesley De Boever November 24, 2016 Bruker microct Micro-CT scanner manufacturer Life science and material science Hardware and software Headquarters in

More information

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information

THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS

THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS A. Campos a*, A. Zorrilla a, F. Lasagni a a Materials & Processes dept., Center for Advanced Aerospace Technologies

More information

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct J. DHAENE* 1, E. PAUWELS 1, T. DE SCHRYVER 1, A. DE MUYNCK 1, M. DIERICK 1, L. VAN HOOREBEKE 1 1 UGCT Dept. Physics

More information

Surface and volume rendering of large datasets

Surface and volume rendering of large datasets Surface and volume rendering of large datasets Kevin Mackenzie Institute of Medical Sciences, University of Aberdeen, Aberdeen, A25 2ZD Aims As part of a project I was involved with looking at the development

More information

Fluorescence Tomography Source Reconstruction and Analysis

Fluorescence Tomography Source Reconstruction and Analysis TECHNICAL NOTE Pre-clinical in vivo imaging Fluorescence Tomography Source Reconstruction and Analysis Note: This Technical Note is part of a series for Fluorescence Imaging Tomography (FLIT). The user

More information

Creating a Mask in ENVI

Creating a Mask in ENVI Creating a Mask in ENVI When mapping particles or materials of interest you may only want to see them and find their distribution within a certain area, like a cell or area of interest. This can be accomplished

More information

Japan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field. Kouichi Inagaki ICC / IHI Corporation

Japan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field. Kouichi Inagaki ICC / IHI Corporation Japan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field Kouichi Inagaki ICC / IHI Corporation 13 th WORLD CONFERENCE ON INVESTMENT CASTING Paper: T3 Copyright

More information

μct for the Pharmaceutical Industries Innovation with Integrity From R&D to Production, Inspection and Failure Analysis Microtomography

μct for the Pharmaceutical Industries Innovation with Integrity From R&D to Production, Inspection and Failure Analysis Microtomography μct for the Pharmaceutical Industries From R&D to Production, Inspection and Failure Analysis Innovation with Integrity Microtomography Find out What's Inside Porosity and Pore Size Distribution Micro-cracks

More information

3D X-ray Microscopy Characterization. Metal Additively Manufactured Parts. Application Note

3D X-ray Microscopy Characterization. Metal Additively Manufactured Parts. Application Note 3D X-ray Microscopy Characterization Metal Additively Manufactured Parts Application Note 3D X-ray Microscopy Characterization Metal Additively Manufactured Parts Author: Luke Hunter ZEISS X-ray Microscopy

More information

Coordinate Measuring Machines with Computed Tomography

Coordinate Measuring Machines with Computed Tomography Always a Step Ahead with Quality Coordinate Measuring Machines with Computed Tomography Multisensor Coordinate Measuring Machines with Computed Tomography Computed Tomography in Coordinate Measuring Machines

More information

Algorithm User Guide:

Algorithm User Guide: Algorithm User Guide: Membrane Quantification Use the Aperio algorithms to adjust (tune) the parameters until the quantitative results are sufficiently accurate for the purpose for which you intend to

More information

TEP Hounsfield units. Related topics Attenuation coefficient, Hounsfield units

TEP Hounsfield units. Related topics Attenuation coefficient, Hounsfield units Hounsfield units TEP Related topics Attenuation coefficient, Hounsfield units Principle Depending on the type of CT scanner and the settings, the result of a CT scan of the same material can be different

More information

Physical bases of X-ray diagnostics

Physical bases of X-ray diagnostics Physical bases of X-ray diagnostics Dr. István Voszka Possibilities of X-ray production (X-ray is produced, when charged particles of high velocity are stopped) X-ray tube: Relatively low accelerating

More information

Latest developments in 3D analysis of geomaterials by Morpho+

Latest developments in 3D analysis of geomaterials by Morpho+ Latest developments in 3D analysis of geomaterials by Morpho+ Cnudde, Veerle *,**, Vlassenbroeck, Jelle **, De Witte, Yoni **, Brabant, Loes **, Boone, Matthieu **, Dewanckele, Jan *,**, Van Hoorebeke,

More information

MULTI-MODALITY 3D VISUALIZATION OF HARD-ALPHA IN TITANIUM

MULTI-MODALITY 3D VISUALIZATION OF HARD-ALPHA IN TITANIUM MULTI-MODALITY 3D VISUALIZATION OF HARD-ALPHA IN TITANIUM BILLETS UTILIZING CT X-RAY AND ULTRASONIC IMAGING Mano Bashyam and Vi Sub GE Aircraft Engines, Mail Drop Q45 1, Neumann Way Cincinnati, OH 45215

More information

Layout Segmentation of Scanned Newspaper Documents

Layout Segmentation of Scanned Newspaper Documents , pp-05-10 Layout Segmentation of Scanned Newspaper Documents A.Bandyopadhyay, A. Ganguly and U.Pal CVPR Unit, Indian Statistical Institute 203 B T Road, Kolkata, India. Abstract: Layout segmentation algorithms

More information

CT Reconstruction with Good-Orientation and Layer Separation for Multilayer Objects

CT Reconstruction with Good-Orientation and Layer Separation for Multilayer Objects 17th World Conference on Nondestructive Testing, 25-28 Oct 2008, Shanghai, China CT Reconstruction with Good-Orientation and Layer Separation for Multilayer Objects Tong LIU 1, Brian Stephan WONG 2, Tai

More information

1. Editorial. N 10 February Content

1. Editorial. N 10 February Content N 10 February 2012 Content 1. Editorial 2. In vivo micro CT Imaging Technology Available at CIF Epalinges 3. Zeiss LSM 700 Confocal Microscope at CIF Dorigny 4. New Powerful Image Processing Setups 5.

More information

Adapted acquisition trajectory and iterative reconstruction for few-views CT inspection

Adapted acquisition trajectory and iterative reconstruction for few-views CT inspection Adapted acquisition trajectory and iterative reconstruction for few-views CT inspection Caroline Vienne 1, Marius Costin 1 More info about this article: http://www.ndt.net/?id=21917 1 CEA, LIST, Département

More information

CHARACTERIZATION OF THE POWDER PARTICLES SHAPES AND PROPERTIES USING THE FRACTAL THEORY

CHARACTERIZATION OF THE POWDER PARTICLES SHAPES AND PROPERTIES USING THE FRACTAL THEORY CHARACTERIZATION OF THE POWDER PARTICLES SHAPES AND PROPERTIES USING THE FRACTAL THEORY *M.Bodea, *N.Jumate, **M.Danca * Technical University of Cluj, ** Department of Mathematics Tehnofrig Technical College

More information

The user's guide. CT-Volume (version 2.0) Contents

The user's guide. CT-Volume (version 2.0) Contents CT-Volume (version 2.0) The user's guide Contents 1. Introduction 2 2. Starting CTVol 2 3. Creating 3d models 4 4. Loading a model into CTVol 6 5. Movement and rotation of a model 7 6. CTVol preferences

More information

A Study of Medical Image Analysis System

A Study of Medical Image Analysis System Indian Journal of Science and Technology, Vol 8(25), DOI: 10.17485/ijst/2015/v8i25/80492, October 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 A Study of Medical Image Analysis System Kim Tae-Eun

More information

Normalization for clinical data

Normalization for clinical data Normalization for clinical data Christopher Rorden, Leonardo Bonilha, Julius Fridriksson, Benjamin Bender, Hans-Otto Karnath (2012) Agespecific CT and MRI templates for spatial normalization. NeuroImage

More information

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS 23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS P. Helberg 1,

More information

Non-Destructive Failure Analysis and Measurement for Molded Devices and Complex Assemblies with X-ray CT and 3D Image Processing Techniques

Non-Destructive Failure Analysis and Measurement for Molded Devices and Complex Assemblies with X-ray CT and 3D Image Processing Techniques SINCE2013 Singapore International NDT Conference & Exhibition 2013, 19-20 July 2013 Non-Destructive Failure Analysis and Measurement for Molded Devices and Complex Assemblies with X-ray CT and 3D Image

More information

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4

More information

Optimization of scanner parameters for dual energy micro-ct

Optimization of scanner parameters for dual energy micro-ct Optimization of scanner parameters for dual energy micro-ct E. PAUWELS* 1, J. DHAENE 1, A. DE MUYNCK 1 E., M. DIERICK 1, L. VAN HOOREBEKE 1 1 UGCT Dept. Physics and Astronomy, Ghent University, Proeftuinstraat

More information

SEM Drift Correction. Procedure Guide

SEM Drift Correction. Procedure Guide SEM Drift Correction Procedure Guide Introduction Vic-2D includes experimental functionality to correct for both drift and geometric distortions that occur in images taken using SEMs. The correction requires

More information

Quantitative imaging for clinical dosimetry

Quantitative imaging for clinical dosimetry Quantitative imaging for clinical dosimetry Irène Buvat Laboratoire d Imagerie Fonctionnelle U678 INSERM - UPMC CHU Pitié-Salpêtrière, Paris buvat@imed.jussieu.fr http://www.guillemet.org/irene Methodology

More information

The Canty SolidSizer TS system was selected for the analysis of the Sugar crystals. Below are the details regarding the testing.

The Canty SolidSizer TS system was selected for the analysis of the Sugar crystals. Below are the details regarding the testing. 1. Introduction 1.1 Background Information and Research The Canty SolidSizer TS system was selected for the analysis of the Sugar crystals. Below are the details regarding the testing. 1.2 Objective The

More information

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Beam Attenuation Grid Based Scatter Correction Algorithm for More Info at Open Access Database

More information

Limited view X-ray CT for dimensional analysis

Limited view X-ray CT for dimensional analysis Limited view X-ray CT for dimensional analysis G. A. JONES ( GLENN.JONES@IMPERIAL.AC.UK ) P. HUTHWAITE ( P.HUTHWAITE@IMPERIAL.AC.UK ) NON-DESTRUCTIVE EVALUATION GROUP 1 Outline of talk Industrial X-ray

More information

microct Systems for Life Science Innovation with Integrity High Resolution Microtomography for in vivo and ex vivo Applications Microtomography

microct Systems for Life Science Innovation with Integrity High Resolution Microtomography for in vivo and ex vivo Applications Microtomography microct Systems for Life Science High Resolution Microtomography for in vivo and ex vivo Applications Innovation with Integrity Microtomography SkyScan 1278 High Throughput Large Field of View in vivo

More information

Advanced Computed Tomography System for the Inspection of Large Aluminium Car Bodies

Advanced Computed Tomography System for the Inspection of Large Aluminium Car Bodies ECNDT 2006 - Th.3.4.2 Advanced Computed Tomography System for the Inspection of Large Aluminium Car Bodies M. SIMON, I. TISEANU, C. SAUERWEIN Hans WÄLISCHMILLER, Meersburg, Germany M. SINDEL, M. BRODMANN,

More information

AnalySIS Tutorial part 2

AnalySIS Tutorial part 2 AnalySIS Tutorial part 2 Sveinung Lillehaug Neural Systems and Graphics Computing Laboratory Department of Anatomy University of Oslo N-0317 Oslo Norway www.nesys.uio.no Using AnalySIS to automatically

More information

Image Analysis and Morphometry

Image Analysis and Morphometry Image Analysis and Morphometry Lukas Schärer Evolutionary Biology Zoological Institute University of Basel 1 13. /15.3.2013 Zoology & Evolution Block Course Summary Quantifying morphology why do we need

More information

Effect of an initial solution in iterative reconstruction of dynamically changing objects

Effect of an initial solution in iterative reconstruction of dynamically changing objects Effect of an initial solution in iterative reconstruction of dynamically changing objects M. HEYNDRICKX 1, T. DE SCHRYVER 1, M. DIERICK 1, M. N. BOONE* 1, T. BULTREYS 2, V. CNUDDE 2, L. VAN HOOREBEKE 1

More information

PHASEQUANT (User Manual)

PHASEQUANT (User Manual) PHASEQUANT (User Manual) ROI Pane Toolbar Threshold Bar Canvas Threshold Display Training Pane Buttons Fig 1. PhaseQuant GUI 1. INSTALLATION AND LOADING 1.1 Installing PhaseQuant Download the zip file

More information

A new concept for high-speed atline and inline CT for up to 100% mass production process control allowing both 3D metrology and failure analysis

A new concept for high-speed atline and inline CT for up to 100% mass production process control allowing both 3D metrology and failure analysis A new concept for high-speed atline and inline CT for up to 100% mass production process control allowing both 3D metrology and failure analysis Oliver Brunke 1, Ferdinand Hansen 2, Ingo Stuke 3, Friedrich

More information

MULTI-PURPOSE 3D COMPUTED TOMOGRAPHY SYSTEM

MULTI-PURPOSE 3D COMPUTED TOMOGRAPHY SYSTEM MULTI-PURPOSE 3D COMPUTED TOMOGRAPHY SYSTEM M. Simon, C. Sauerwein, I. Tiseanu, S. Burdairon Hans Wälischmiller GmbH Klingleweg 8, D-88709 Meersburg, Germany e-mail: ms@hwm.com ABSTRACT A new flexible

More information

LSM 5 MP, LSM 510 and LSM 510 META Laser Scanning Microscopes

LSM 5 MP, LSM 510 and LSM 510 META Laser Scanning Microscopes LSM 5 MP, LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4.2 January 2007 Contents Page Starting the System...3 Setting the microscope...6 Configuring the beam path

More information

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory Introduction In this lab exercise, you will investigate the linearity of the DeskCAT scanner by making measurements

More information

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Qiao Yang 1,4, Meng Wu 2, Andreas Maier 1,3,4, Joachim Hornegger 1,3,4, Rebecca Fahrig

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!

More information

OCR For Handwritten Marathi Script

OCR For Handwritten Marathi Script International Journal of Scientific & Engineering Research Volume 3, Issue 8, August-2012 1 OCR For Handwritten Marathi Script Mrs.Vinaya. S. Tapkir 1, Mrs.Sushma.D.Shelke 2 1 Maharashtra Academy Of Engineering,

More information

8/3/2017. Contour Assessment for Quality Assurance and Data Mining. Objective. Outline. Tom Purdie, PhD, MCCPM

8/3/2017. Contour Assessment for Quality Assurance and Data Mining. Objective. Outline. Tom Purdie, PhD, MCCPM Contour Assessment for Quality Assurance and Data Mining Tom Purdie, PhD, MCCPM Objective Understand the state-of-the-art in contour assessment for quality assurance including data mining-based techniques

More information

Fundamentals of CT imaging

Fundamentals of CT imaging SECTION 1 Fundamentals of CT imaging I History In the early 1970s Sir Godfrey Hounsfield s research produced the first clinically useful CT scans. Original scanners took approximately 6 minutes to perform

More information

Design and Execution of Model Experiments to Validate Numerical Modelling of 2D Ship Operations in Pack Ice

Design and Execution of Model Experiments to Validate Numerical Modelling of 2D Ship Operations in Pack Ice Design and Execution of Model Experiments to Validate Numerical Modelling of 2D Ship Operations in Pack Ice Roelof C. Dragt Offshore Engineering Faculty of Mechanical, Maritime and Material Engineering

More information

X-ray tomographic study of wetting bentonite

X-ray tomographic study of wetting bentonite UNIVERSITY OF JYVÄSKYLÄ X-ray tomographic study of wetting bentonite Tero Harjupatana Department of Physics, University of Jyväskylä BOA seminar 19.8.2014 Goal To develop and apply experimental techniques

More information

Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system

Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system ECNDT 2014 Prague October 6-10, 2014 Dr. Eberhard Neuser Dr. Alexander Suppes Imagination

More information

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT More Info at Open Access Database www.ndt.net/?id=18754 Novel evaluation method of low contrast resolution performance of dimensional X-ray CT Makoto Abe 1, Hiroyuki Fujimoto 1, Osamu Sato 1, Kazuya Matsuzaki

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

A Review on Plant Disease Detection using Image Processing

A Review on Plant Disease Detection using Image Processing A Review on Plant Disease Detection using Image Processing Tejashri jadhav 1, Neha Chavan 2, Shital jadhav 3, Vishakha Dubhele 4 1,2,3,4BE Student, Dept. of Electronic & Telecommunication Engineering,

More information

Counting Particles or Cells Using IMAQ Vision

Counting Particles or Cells Using IMAQ Vision Application Note 107 Counting Particles or Cells Using IMAQ Vision John Hanks Introduction To count objects, you use a common image processing technique called particle analysis, often referred to as blob

More information

Measurements using three-dimensional product imaging

Measurements using three-dimensional product imaging ARCHIVES of FOUNDRY ENGINEERING Published quarterly as the organ of the Foundry Commission of the Polish Academy of Sciences ISSN (1897-3310) Volume 10 Special Issue 3/2010 41 46 7/3 Measurements using

More information

Image Compression With Haar Discrete Wavelet Transform

Image Compression With Haar Discrete Wavelet Transform Image Compression With Haar Discrete Wavelet Transform Cory Cox ME 535: Computational Techniques in Mech. Eng. Figure 1 : An example of the 2D discrete wavelet transform that is used in JPEG2000. Source:

More information

VARIABILITY IN MICRO-COMPUTED TOMOGRAPHY IMAGES FOR NON-DESTRUCTIVE EVALUATION OF FRICTION-STIR WELD QUALITY

VARIABILITY IN MICRO-COMPUTED TOMOGRAPHY IMAGES FOR NON-DESTRUCTIVE EVALUATION OF FRICTION-STIR WELD QUALITY Proceedings of the South Dakota Academy of Science, Vol. 93 (2014) 61 VARIABILITY IN MICRO-COMPUTED TOMOGRAPHY IMAGES FOR NON-DESTRUCTIVE EVALUATION OF FRICTION-STIR WELD QUALITY Alex Wulff*, William Cross,

More information

ECE 172A: Introduction to Intelligent Systems: Machine Vision, Fall Midterm Examination

ECE 172A: Introduction to Intelligent Systems: Machine Vision, Fall Midterm Examination ECE 172A: Introduction to Intelligent Systems: Machine Vision, Fall 2008 October 29, 2008 Notes: Midterm Examination This is a closed book and closed notes examination. Please be precise and to the point.

More information

Chapter 3. Image Processing Methods. (c) 2008 Prof. Dr. Michael M. Richter, Universität Kaiserslautern

Chapter 3. Image Processing Methods. (c) 2008 Prof. Dr. Michael M. Richter, Universität Kaiserslautern Chapter 3 Image Processing Methods The Role of Image Processing Methods (1) An image is an nxn matrix of gray or color values An image processing method is algorithm transforming such matrices or assigning

More information

Tumor Detection and classification of Medical MRI UsingAdvance ROIPropANN Algorithm

Tumor Detection and classification of Medical MRI UsingAdvance ROIPropANN Algorithm International Journal of Engineering Research and Advanced Technology (IJERAT) DOI:http://dx.doi.org/10.31695/IJERAT.2018.3273 E-ISSN : 2454-6135 Volume.4, Issue 6 June -2018 Tumor Detection and classification

More information

Mirror Based Framework for Human Body Measurement

Mirror Based Framework for Human Body Measurement 362 Mirror Based Framework for Human Body Measurement 1 Takeshi Hashimoto, 2 Takayuki Suzuki, 3 András Rövid 1 Dept. of Electrical and Electronics Engineering, Shizuoka University 5-1, 3-chome Johoku,

More information

A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY

A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY A PRACTICAL ALGORITHM FOR RECONSTRUCTION FROM X-RAY BACKSCATTER DATA Young S. Ham and C.F. Poranski Materials Chemistry Branch Naval Research Laboratory Washington DC 20375 E.C. Greenawald Geo-Centers,

More information

Flying Start AS Computer Science. September 2015

Flying Start AS Computer Science. September 2015 Flying Start AS Computer Science September 2015 Name: To your first AS Computing lesson, you will need to bring: 1. A folder with dividers An A4 ring binder with labelled A4 dividers would be ideal. The

More information

SUPPLEMENTARY FILE S1: 3D AIRWAY TUBE RECONSTRUCTION AND CELL-BASED MECHANICAL MODEL. RELATED TO FIGURE 1, FIGURE 7, AND STAR METHODS.

SUPPLEMENTARY FILE S1: 3D AIRWAY TUBE RECONSTRUCTION AND CELL-BASED MECHANICAL MODEL. RELATED TO FIGURE 1, FIGURE 7, AND STAR METHODS. SUPPLEMENTARY FILE S1: 3D AIRWAY TUBE RECONSTRUCTION AND CELL-BASED MECHANICAL MODEL. RELATED TO FIGURE 1, FIGURE 7, AND STAR METHODS. 1. 3D AIRWAY TUBE RECONSTRUCTION. RELATED TO FIGURE 1 AND STAR METHODS

More information

DLIT and FLIT Reconstruction of Sources

DLIT and FLIT Reconstruction of Sources Concept Tech Note 5 DLIT and FLIT Reconstruction of Sources Diffuse Luminescence Imaging Tomography (DLIT) is a technique that analyzes images of the surface light emission from a living subject to generate

More information

Occlusion Detection of Real Objects using Contour Based Stereo Matching

Occlusion Detection of Real Objects using Contour Based Stereo Matching Occlusion Detection of Real Objects using Contour Based Stereo Matching Kenichi Hayashi, Hirokazu Kato, Shogo Nishida Graduate School of Engineering Science, Osaka University,1-3 Machikaneyama-cho, Toyonaka,

More information

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 16 Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 3.1 MRI (magnetic resonance imaging) MRI is a technique of measuring physical structure within the human anatomy. Our proposed research focuses

More information

Ensemble registration: Combining groupwise registration and segmentation

Ensemble registration: Combining groupwise registration and segmentation PURWANI, COOTES, TWINING: ENSEMBLE REGISTRATION 1 Ensemble registration: Combining groupwise registration and segmentation Sri Purwani 1,2 sri.purwani@postgrad.manchester.ac.uk Tim Cootes 1 t.cootes@manchester.ac.uk

More information

LABORATORY SYSTEM FOR X-RAY NANOTOMOGRAPHY

LABORATORY SYSTEM FOR X-RAY NANOTOMOGRAPHY 79 LABORATORY SYSTEM FOR X-RAY NANOTOMOGRAPHY Alexander Sasov, SkyScan, Vluchtenburgstraat 3, Aartselaar B2630, Belgium, www.skyscan.be. ABSTRACT Using advanced X-ray technologies and X-ray scattering

More information

DEVELOPMENT OF PREVENTIVE MAINTENANCE SYSTEM ARISING IN ADVANCED EDDY CURRENT TESTING USING NETWORK TOMOGRAPHY

DEVELOPMENT OF PREVENTIVE MAINTENANCE SYSTEM ARISING IN ADVANCED EDDY CURRENT TESTING USING NETWORK TOMOGRAPHY DEVELOPMENT OF PREVENTIVE MAINTENANCE SYSTEM ARISING IN ADVANCED EDDY CURRENT TESTING USING NETWORK TOMOGRAPHY F. Kojima 1, F.Kobayashi 1, A. Nishimizu 2, M. Koike 2, T. Matsui 2, H. Endo 3, T. Uchimoto

More information

Multi-pass approach to adaptive thresholding based image segmentation

Multi-pass approach to adaptive thresholding based image segmentation 1 Multi-pass approach to adaptive thresholding based image segmentation Abstract - Thresholding is still one of the most common approaches to monochrome image segmentation. It often provides sufficient

More information

THREE DIMENSIONAL EXAMINATION OF DIRECTIVITY PATTERN IN IMMERSION TANK TESTING

THREE DIMENSIONAL EXAMINATION OF DIRECTIVITY PATTERN IN IMMERSION TANK TESTING 7 th International Symposium on NDT in Aerospace We.3.A.7 THREE DIMENSIONAL EXAMINATION OF DIRECTIVITY PATTERN IN IMMERSION TANK TESTING Daniel KOTSCHATE 1, Thomas HECKEL 1 and Dirk GOHLKE 1 1 Federal

More information

Towards a fully automated image analysis system applicable to ecological studies Tested on water fleas, birds and pollen

Towards a fully automated image analysis system applicable to ecological studies Tested on water fleas, birds and pollen Towards a fully automated image analysis system applicable to ecological studies Tested on water fleas, birds and pollen Selwyn Hoeks, RU Nijmegen Supervisors: Eelke Jongejans, RU Nijmegen Marjolein Bruijning,

More information

Multi-spectral (W, Mo, Cu, and Ag) XCT measurements

Multi-spectral (W, Mo, Cu, and Ag) XCT measurements Multi-spectral (W, Mo, Cu, and Ag) XCT measurements Marko Katić 1, Damir Markučič 1, Gorana Baršić 1 Faculty of mechanical engineering and naval architecture, Ivana Lucica 5, Zagreb, Croatia, marko.katic@fsb.hr

More information

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Marc Kachelrieß, a Katia Sourbelle, and Willi A. Kalender Institute of Medical Physics, University of

More information

PART 1. Answers module 6: 'Transformations'

PART 1. Answers module 6: 'Transformations' Answers module 6: 'Transformations' PART 1 1 a A nominal measure scale refers to data that are in named categories. There is no order among these categories. That is, no category is better or more than

More information

UGviewer: a medical image viewer

UGviewer: a medical image viewer Appendix A UGviewer: a medical image viewer As a complement to this master s thesis, an own medical image viewer was programmed. This piece of software lets the user visualize and compare images. Designing

More information