Ini$al Performance Test Results on an Elekta MR-Linac
|
|
- Edmund Sherman
- 5 years ago
- Views:
Transcription
1 Ini$al Performance Test Results on an Elekta MR-Linac Eric Paulson, PhD DABR Radia$on Oncology, Radiology, and Biophysics Froedtert and Medical College of Wisconsin Milwaukee, Wisconsin, United States April 7, 2017
2 Disclosures Advancing a Healthier Wisconsin ( / ) Elekta Instruments, AB
3 Why MRI in Radia$on Oncology? Pancreas Planning CT MRI Pillars of success in RT: Accurate delinea$on of disease extent Op$mal alignment of the treatment beam to the target kv CBCT MRI MR Simula$on: Consistent, true GTV More accurate CTV Esophagus MR-guided RT: Reduce inter-, intra-frac$onal varia$ons Biologically adap$ve, dose-escalated RT
4 Challenges Facing MR-guided RT Decoupling the magne$c field from RT components: Minimizing magne$c interference Decoupling the RT components from MRI acquisi$on: Minimizing radiofrequency interference Beam transmission through the MRI Influence of Lorentz force on secondary electrons
5 MR-gRT Implementa$ons MRI on Rails (PMH) Rota$ng Bi-Planar (CCI, Alberta) ViewRay (Wash U, UW Madison, UCLA, etc) Rota$ng Bi-Planar (Australia)
6 Elekta/Philips MR-Linac Philips 1.5T MRI: Ac$ve shielding reduces magne$c interference Faraday cage integrated into cryostat Split solenoid with transmission through the cyrostat Windows designed into gradient, RF coils Elekta Linac: 7MV FFF beam (transverse to B0) Rigid ring mounted gantry with slip rings 160-leaf MLC in S-I orienta$on 7mm projected leaf width at isocenter 142cm SAD 22cm max field size in along S-I EPID for portal imaging, QA
7 MR-Linac Vault at MCW Modified exis$ng 3 rd floor vault Floor built up 3 to provide gantry clearance: Maze with ramp (MCW) Ver$cal lig (MDA)
8 Pre-Acceptance Measurements Magnet ramped up: 3/4/2017 Philips Calibra$ons: 3/5/2017 3/6/2017 AAPM Abstract Deadline: 3/9/2017 Magnet ramped down: 3/11/2017
9 Magne$c Field vs Gantry Rota$on Passive shimmed at gantry = 0 deg: 4.1 Hz RMS Average of all angles: 10 Hz RMS (gantry posi$on can affect magne$c field homogeneity) Paulson, et al, AAPM 2017 (submiked)
10 Magne$c Field vs Gantry Rota$on (+Shimming) Average of all angles: 4.8 Hz RMS Conclusion: Perform shimming at each gantry angle prior to beam-on Paulson, et al, AAPM 2017 (submiked)
11 Gradient Linearity over 50cm FOV Axial Sagikal Coronal Es$mated within 1mm at 20cm radius from isocenter Paulson, et al, AAPM 2017 (submiked)
12 With/Without Radia$on Radia$on Off Radia$on On Difference Paulson, et al, AAPM 2017 (submiked)
13 ACR MRI Phantom QA Meets required tolerances for ACR Accredita$on using protocol from diagnos$c 1.5T Philips Ingenia Paulson, et al, AAPM 2017 (submiked)
14 Mo$on Es$ma$on Accuracy Matched programmed peak-to-peak displacement, trajectory, and frequency Paulson, et al, AAPM 2017 (submiked)
15 On-Board EPID Image Quality All required features on Las Vegas phantom were visible Paulson, et al, AAPM 2017 (submiked)
16 Magne$c Field On Beam Profiles Magne$c Field Off Symmetry within 0.3% (without magne$c field) Paulson, et al, AAPM 2017 (submiked)
17 End-To-End Test (1/2) CT Sim MIM Monaco Mosaiq TSM Integrity
18 CT Sim MIM Monaco Mosaiq TSM End-To-End Test (2/2) Date: 16/03/2017 ArcCHECK QA of Dose Distribution Hospital Name: QA File Parameter ArcCHECK Plan Patient Name : MRLINAC, CIRS Patient ID : MRLINAC06 Plan Date : 3/16/2017 SSD : SDD : Depth : Energy : 0 Angle : Relative Comparison Difference (%) Distance (mm) Threshold (%) Meas Uncertainty : Yes Summary (Gamma Analysis) Total Points Passed Failed % Passed : 3.0 : 3.0 : 5 : 272 : 261 : 11 : 96 *DTA/Gamma is using 2D Mode Dose Values in cgy CAX Norm Sel Max ArcCHECK Plan ArcCHECK Plan % Diff DTA(mm) NA 0.00 NA Coords (y,x) cm Notes 0,0 1, , \ \MRLINA... ArcCHECK-Plan 96% Set1 Set2 CAX Offset X=1 Y=-1 Integrity % X(mm) ArcCheck response affected by concurrent imaging while measuring
19 Summary MR-gRT extends the advantages of MRI into the RT treatment room, facilita$ng high precision, biologically-adap$ve, dose-escalated radiotherapy The Elekta MR-Linac is one of several implementa$ons pioneering MR-gRT, each uniquely addressing the design challenges of integra$ng MRI and RT devices Based on pre-acceptance measurements: MR image quality and geometric accuracy does not appear to have been compromised by integra$on of a linac Dosimetry effects in the presence of a magne$c field are as expected, and can be calculated and managed in treatment planning
20 Acknowledgements Nikolai Mickevicius X. Allen Li, PhD Ergun Ahunbay, PhD Xingfeng Chen, PhD Slade Klawikowski, PhD Ray Thomas (Elekta) Scok Kotlarz (Philips)
ViewRay System Commissioning
ViewRay System Commissioning Kyle R. Padgett, PhD, DABR Assistant Professor University of Miami Viewray Team: Matt Studenski, PhD Chet Ford, PhD Yidong Yang, PhD Nesrin Dogan, PhD Medical Physics Residents:
More information8/3/2016. MR-guided RT: Commissioning and Quality Control. Topics for MR-guided RT system commissioning & QC
MR-guided RT: Commissioning and Quality Control T. Stanescu, PhD, MCCPM Medical Physicist, Princess Margaret Cancer Centre, UHN Assistant Professor, Radiation Oncology, University of Toronto Affiliated
More informationAcknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF
Page 1 Quality Assurance of IMRT Delivery Systems - Siemens Lynn J. Verhey, Ph.D. Professor and Vice-Chair UCSF Dept. of Radiation Oncology AAPM 22 Annual Meeting, Montreal Acknowledgments Ping Xia, Ph.D.,
More informationReal-time monitoring of linac performance using RT plans and logfiles
Real-time monitoring of linac performance using RT plans and logfiles NCCAAPM Wisconsin Dells 7 April 2017 Mark Wiesmeyer, PhD, DABR Technical Product Manager Technical talk with insights and results based
More informationThank-You Members of TG147 TG 147: QA for nonradiographic
Thank-You Members of TG147 TG 147: QA for nonradiographic localization and positioning systems Twyla Willoughby, M.S. Medical Physicist Clinical AAPM Meeting March 2013 Department of Radiation Oncology
More informationURGENT IMPORTANT FIELD SAFETY NOTIFICATION
Subject: Incorrect Movement of the Treatment Table Product: MOSAIQ Scope: Sites affected will be those: 1. Running MOSAIQ and, 2. Treating on linear accelerators with the RATM license Notification Released:
More information8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques
Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery Lei Ren, Ph.D. Duke University Medical Center 2016 AAPM 58 th annual meeting, Educational Course, Therapy Track Disclosure I have
More informationTomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS
Tomotherapy Physics Machine Twinning and Quality Assurance Emilie Soisson, MS Tomotherapy at UW- Madison Treating for nearly 5 years Up to ~45 patients a day on 2 tomo units Units twinned to facilitate
More informationMR-guided radiotherapy: Vision, status and research at the UMC Utrecht. Dipl. Ing. Dr. Markus Glitzner
MR-guided radiotherapy: Vision, status and research at the UMC Utrecht Dipl. Ing. Dr. Markus Glitzner About myself Training Medizintechnik TU Graz PhD UMC Utrecht Clinical work Software implementation
More informationImage Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD
Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image
More informationRaising the Bar in IMRT QA
MapCHECK 2TM Raising the Bar in IMRT QA The leader in quick and precise measurement of modulated radiotherapy beams Benefits Proven solution for film-less rotational delivery and IMRT QA - More than 1500
More information8/2/2017. PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA
PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA Jennifer R. Clark, M.S., DABR Conflict of Interest Statement: I am an employee of Sun Nuclear Corporation Outline: Platform overview
More informationVolumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT
Volumetric Modulated Arc Therapy - Clinical Implementation Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA Acknowledgement David M. Shepard, Ph.D. Muhammad K. N. Afghan, Ph.D. Fan Chen, Ph.D.
More informationNovel methods and tools for MR Commissioning and Quality Control
Novel methods and tools for MR Commissioning and Quality Control T. Stanescu, PhD, MCCPM Medical Physicist, RMP, Princess Margaret Cancer Centre Assistant Professor, Radiation Oncology, University of Toronto
More informationImage Guidance and Beam Level Imaging in Digital Linacs
Image Guidance and Beam Level Imaging in Digital Linacs Ruijiang Li, Ph.D. Department of Radiation Oncology Stanford University School of Medicine 2014 AAPM Therapy Educational Course Disclosure Research
More information7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT
CLINICAL COMMISSIONING AND ACCEPTANCE TESTING OF A 3D SURFACE MATCHING SYSTEM Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago Learning Objective Describe acceptance
More informationPhoton beam dose distributions in 2D
Photon beam dose distributions in 2D Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2014 Acknowledgments! Narayan Sahoo PhD! Richard G Lane (Late) PhD 1 Overview! Evaluation
More informationBasic Radiation Oncology Physics
Basic Radiation Oncology Physics T. Ganesh, Ph.D., DABR Chief Medical Physicist Fortis Memorial Research Institute Gurgaon Acknowledgment: I gratefully acknowledge the IAEA resources of teaching slides
More informationVersion 5.6. Quick Start Guide
WWW..COM Version 5.6 Quick Start Guide STANDARD IMAGING, INC. 3120 Deming Way Middleton, WI 53562-1461 Jul / 2018 2018 Standard Imaging, Inc. TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 DOC # 80714-05
More informationDisclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions
LINAC and MLC QA for IMRT Ping Xia, Ph.D., Department of Radiation Oncology Disclosure Received research support from Siemens Medical Solutions University of California San Francisco Therapy Series (SAM)
More informationDelta 4 PT. True Volumetric Pre-treatment Verification. The difference is clear
Delta 4 PT True Volumetric Pre-treatment Verification The difference is clear ScandiDos innovators in advanced volumetric dosimetry solutions ScandiDos ScandiDos is the innovative company that introduced
More informationThe team. Disclosures. Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge.
Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge Dimitre Hristov Radiation Oncology Stanford University The team Renhui Gong 1 Magdalena Bazalova-Carter 1
More informationQuick Reference Datasheet For All RIT113 Packages
Quick Reference Datasheet For All RIT113 Packages For Rotational Therapies, IMRT & TG142 Highlights and selected product information only. A complete TG142 brochure is available. For more information on
More informationState-of-the-Art IGRT
in partnership with State-of-the-Art IGRT Exploring the Potential of High-Precision Dose Delivery and Real-Time Knowledge of the Target Volume Location Antje-Christin Knopf IOP Medical Physics Group Scientific
More informationFinancial disclosure. Onboard imaging modality for IGRT
Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty
More informationInitial Clinical Experience with 3D Surface Image Guidance
Initial Clinical Experience with 3D Surface Image Guidance Amanda Havnen-Smith, Ph.D. Minneapolis Radiation Oncology Ridges Radiation Therapy Center Burnsville, MN April 20 th, 2012 Non-funded research
More informationRITtrend allows you to effectively manage all your physics QA data in one powerful and customizable package.
So much is asked of medical physicists these days We Can Help Automated Radiation Therapy Phantom Analysis in Seconds QC for Therapy OBI s and CT simulators. Radia s Catphan/OBI module performs analysis
More informationTHE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!
THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER! Confidence in complex treatments Modern radiation therapy uses complex plans with techniques such as IMRT, VMAT and Tomotherapy.
More informationTumor motion during liver SBRT
Tumor motion during liver SBRT - projects at Aarhus University Hospital - Per Poulsen, Esben Worm, Walther Fledelius, Morten Høyer Aarhus University Hospital, Denmark SBRT: Stereotactic Body Radiation
More informationS. Guru Prasad, Ph.D., DABR
PURPOSE S. Guru Prasad, Ph.D., DABR Director of Medical Physics IAEA Consultant NorthShore University Health System and University of Chicago, Pritzker School of Medicine Current TPS utilize more information
More informationSlide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.
Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing
More informationUsing a research real-time control interface to go beyond dynamic MLC tracking
in partnership with Using a research real-time control interface to go beyond dynamic MLC tracking Dr. Simeon Nill Joint Department of Physics at The Institute of Cancer Research and the Royal Marsden
More informationThe IORT Treatment Planning System. radiance. GMV, 2012 Property of GMV All rights reserved
The IORT Treatment Planning System radiance Property of GMV All rights reserved WHY RADIANCE? JUSTIFICATION Property of GMV All rights reserved ADVANTAGES OF IORT PRECISION: RT guided by direct vision.
More informationFeasibility study of performing IGRT system daily QA using a commercial QA device
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 3, summer 2011 Feasibility study of performing IGRT system daily QA using a commercial QA device Jean L.Peng, 1 Darren Kahler, 2 Jonathan
More informationClinical implementation of photon beam flatness measurements to verify beam quality
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Clinical implementation of photon beam flatness measurements to verify beam quality Simon Goodall, a Nicholas Harding, Jake Simpson,
More informationFeasibility of 3D Printed Patient specific Phantoms for IMRT QA and Other Dosimetric Special Procedures
Feasibility of 3D Printed Patient specific Phantoms for IMRT QA and Other Dosimetric Special Procedures ehler 046@umn.edu Eric Ehler, PhD Assistant Professor Department of Radiation Oncology What is 3D
More informationTPS COMMISSIONING. Laurence Court University of Texas MD Anderson Cancer Center 4/3/2017 1
TPS COMMISSIONING Laurence Court University of Texas MD Anderson Cancer Center 4/3/2017 1 Conflicts of interest Court receives funding from NIH, CPIRT, Varian and Elekta 4/3/2017 2 Resources your first
More informationIMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays
IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays Sotiri Stathakis Outline Why IMRT/VMAT QA AAPM TG218 UPDATE Tolerance Limits and Methodologies for IMRT Verification QA Common sources
More informationQalitätssicherung an Multileafkollimatoren. Dr. Lutz Müller Würzburg jan 2004
Qalitätssicherung an Multileafkollimatoren Dr. Lutz Müller Würzburg jan 2004 IMRT Verification - present Target Volume Constraints Inverse Planning Algorithm Fluence Map Leaf Sequencer Leaf & Gantry sequence
More informationLucy Phantom MR Grid Evaluation
Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is
More informationAdvanced Radiotherapy
Advanced Radiotherapy IBA Concept for IMRT Verification Dr. Lutz Müller 1 IMRT Verification BIS BIS ADAS Fluence Model Delivered fluence Target Volume Constraints Inverse Backprojection Fluence Map Leaf
More informationUse of Monte Carlo modelling in radiotherapy linac design. David Roberts, PhD Senior Physicist Elekta
Use of Monte Carlo modelling in radiotherapy linac design David Roberts, PhD Senior Physicist Elekta Contents Overview of Elekta What we do Where we use Monte Carlo Codes and resources Example : Agility
More informationAutomated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator
Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4
More informationTG 132: Use of Image Registration and Fusion in RT
TG 132: Use of Image Registration and Fusion in RT Kristy K Brock, PhD, DABR, FAAPM Associate Professor Department of Radiation Oncology, University of Michigan Chair, AAPM TG 132: Image Registration and
More informationMapCHECK 2 & 3DVH The Gold Standard for 2D Arrays
MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.
More information8/3/2016. Image Guidance Technologies. Introduction. Outline
8/3/26 Session: Image Guidance Technologies and Management Strategies Image Guidance Technologies Jenghwa Chang, Ph.D.,2 Department of Radiation Medicine, Northwell Health 2 Hofstra Northwell School of
More informationradiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li NCAAPM Spring Meeting 2010 Madison, WI
GPU-Accelerated autosegmentation for adaptive radiotherapy Andrew Godley, Ergun Ahunbay, Cheng Peng, and X. Allen Li agodley@mcw.edu NCAAPM Spring Meeting 2010 Madison, WI Overview Motivation Adaptive
More informationMonte Carlo methods in proton beam radiation therapy. Harald Paganetti
Monte Carlo methods in proton beam radiation therapy Harald Paganetti Introduction: Proton Physics Electromagnetic energy loss of protons Distal distribution Dose [%] 120 100 80 60 40 p e p Ionization
More informationDesign and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon
Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization
More informationVMAT for dummies: Concepts, Clinical Implementation and Treatment Planning. Rajat Kudchadker, Ph.D. Associate Professor
VMAT for dummies: Concepts, Clinical Implementation and Treatment Planning Rajat Kudchadker, Ph.D Associate Professor 2012 AAPM Spring Clinical Meeting Disclaimer I hold no financial interest nor have
More informationMapCHECK 2 & 3DVH. The Gold Standard for 2D Arrays
MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.
More informationMachine and Physics Data Guide
WWW..COM Machine and Physics Data Guide STANDARD IMAGING, INC. 3120 Deming Way Middleton, WI 53562-1461 May / 2008 2008 Standard Imaging, Inc. TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 www.standardimaging.com
More information1. Learn to incorporate QA for surface imaging
Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago ***No disclosures*** 1. Learn to incorporate QA for surface imaging into current QA procedures for IGRT. 2. Understand
More informationDAILY LINAC QA BEAM QA
BEAM QA DAILY LINAC QA The QA BeamChecker Plus allows for fast, reliable, and uncomplicated daily QA of Varian, Elekta, Siemens, and Accuray Treatment Machines. The QA BeamChecker Plus is specifically
More informationImage Co-Registration II: TG132 Quality Assurance for Image Registration. Image Co-Registration II: TG132 Quality Assurance for Image Registration
Image Co-Registration II: TG132 Quality Assurance for Image Registration Preliminary Recommendations from TG 132* Kristy Brock, Sasa Mutic, Todd McNutt, Hua Li, and Marc Kessler *Recommendations are NOT
More informationOverview of Proposed TG-132 Recommendations
Overview of Proposed TG-132 Recommendations Kristy K Brock, Ph.D., DABR Associate Professor Department of Radiation Oncology, University of Michigan Chair, AAPM TG 132: Image Registration and Fusion Conflict
More informationICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588
ICARO Vienna April 27-29 2009 Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588 M. Saiful Huq, Ph.D., Professor and Director, Dept. of Radiation
More informationFacility Questionnaire PART I (General Information for 3DCRT and IMRT)
Facility Questionnaire PART I (General Information for 3DCRT and IMRT) The following items are required before you can enter cases on any RTOG protocol that requires data submission to the Image-Guided
More informationA comparative analysis for verification of IMRT and VMAT treatment plans using a 2-D and 3-D diode array
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2014 A comparative analysis for verification of IMRT and VMAT treatment plans using a 2-D and 3-D diode array
More informationHugues Mailleux Medical Physics Department Institut Paoli-Calmettes Marseille France. Sunday 17 July 2016
Hugues Mailleux Medical Physics Department Institut Paoli-Calmettes Marseille France Sunday 17 July 2016 AGENDA 1. Introduction 2. Material 3. Optimization process 4. Results 5. Comments 6. Conclusion
More informationOp#miza#on of an On- Board Imaging System for Rapid Radiotherapy. Erica Kemmerling, Meng Wu, He Yang, and Rebecca Fahrig
Op#miza#on of an On- Board Imaging System for Rapid Radiotherapy Erica Kemmerling, Meng Wu, He Yang, and Rebecca Fahrig Problem: mo#on during treatment Two overlapping thorax views during normal breathing
More informationIntroduction. Quality Assurance for Image- Guided Radiation Therapy. Justification for IGRT. Image-Guided Radiation Therapy
Introduction Quality Assurance for Image- Guided Radiation Therapy Jean-Pierre Bissonnette, Ph.D., MCCPM Princess Margaret Hospital, Toronto, Canada IGRT What is it? Rationale Equipment Quality Assurance
More informationThe MSKCC Approach to IMRT. Outline
The MSKCC Approach to IMRT Spiridon V. Spirou, PhD Department of Medical Physics Memorial Sloan-Kettering Cancer Center New York, NY Outline Optimization Field splitting Delivery Independent verification
More informationFAST, precise. qa software
qa software FAST, precise Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, rotational or brachytherapy plans no film, no phantoms, no linac
More informationDynalog data tool for IMRT plan verification
Dynalog data tool for IMRT plan verification Poster No.: R-0051 Congress: 2014 CSM Type: Scientific Exhibit Authors: V. Sashin; FOOTSCRAY/AU Keywords: Computer applications, Radiation physics, Experimental,
More information3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending
3DVH : On The Accuracy Of The Planned Dose Perturbation Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools *Patent Pending introduction State-of-the-art IMRT QA of static gantry
More informationATC Conformance Statement Overview
Advanced-Technology QA Consortium DICOM Workshop: ATC Conformance Statement Overview Walter R. Bosch, D.Sc. ATC DICOM Conformance Statement Overview DICOM Part 10 File Set Reader Application (ITC) Defines
More information3D Scanning Comes Full Circle
3D SCANNER TM 3D Scanning Comes Full Circle Relative 3D Dosimetry offering the easiest setup, most objectivity, and best consistency available Cross-Plane Advanced Design Ring and diameter drive mechanism
More information7/31/ D Cone-Beam CT: Developments and Applications. Disclosure. Outline. I have received research funding from NIH and Varian Medical System.
4D Cone-Beam CT: Developments and Applications Lei Ren, PhD, DABR Department of Radiation Oncology Duke University Medical Center Disclosure I have received research funding from NIH and Varian Medical
More informationIMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE
QA SOFTWARE FAST, PRECISE Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, VMAT, SRS and brachytherapy plans no film, no phantoms, no linac
More informationISO ISO ISO OHSAS ISO
ISO 9001 ISO 13485 ISO 14001 OHSAS 18001 ISO 27001 Pro-NM Performance 08-101 - standard version 08-103 - version with the PET Lid Phantom for NM and PET systems performance evaluation (collimator, artifacts,
More informationBrilliance CT Big Bore.
1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.
More informationProtocol. Technical evaluation of X-ray tomographic image-guided radiotherapy devices CEP10070
Protocol Technical evaluation of X-ray tomographic image-guided radiotherapy devices CEP10070 March 2010 Contents 2 Introduction... 3 Protocol design and validation... 4 General information... 7 Technical
More informationph fax Deming Way Middleton WI USA
www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 3120 Deming Way Middleton WI 53562-1461 USA 2007 Standard Imaging, Inc. 1239-21 (03.07) beam qa LINAC AND TOMOTHERAPY QA The NEW
More informationTable of Contents. About AQUA
User Manual About AQUA Table of Contents 1 About AQUA... 6 1.1 Radiation Therapy... 6 1.1.1 Workflow manager... 6 1.1.2 Reports... 7 1.1.3 Tests... 7 1.1.4 Many Ways to Work... 7 1.2 Technology... 7 1.3
More informationTomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram
TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization
More informationAutomated Quality Assurance for Image-Guided Radiation Therapy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 1, WINTER 2009 Automated Quality Assurance for Image-Guided Radiation Therapy Eduard Schreibmann, a Eric Elder, Tim Fox Department of Radiation
More informationOptical Guidance. Sanford L. Meeks. July 22, 2010
Optical Guidance Sanford L. Meeks July 22, 2010 Optical Tracking Optical tracking is a means of determining in real-time the position of a patient relative to the treatment unit. Markerbased systems track
More informationCommissioning, beam matching and small field dosimetry experience with Elekta Versa
Commissioning, beam matching and small field dosimetry experience with Elekta Versa Raj Varadhan, PhD Director of Physics, Technology Minneapolis Radiation Oncology Outline Elekta Versa/Infinity platform
More informationAbbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology
Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Outline High Spectral and Spatial Resolution MR Imaging (HiSS) What it is How to do it Ways to use it HiSS for Radiation
More informationDICOM CONFORMANCE STATEMENT. BrainLAB PatientBrowser 1.0
Technical Publication DICOM CONFORMANCE STATEMENT BrainLAB PatientBrowser 1.0 Revision 1.0 BrainLAB AG last change: 21.02.01 11:57 Page 1 of 12 Table of contents DICOM Conformance Statement 0. INTRODUCTION...
More informationEvaluation of 3D Gamma index calculation implemented in two commercial dosimetry systems
University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2015 Evaluation of 3D Gamma index calculation implemented
More informationCLINICAL ASPECTS OF COMPACT GANTRY DESIGNS
CLINICAL ASPECTS OF COMPACT GANTRY DESIGNS J. Heese, J. Wulff, A. Winnebeck, A. Huggins, M. Schillo VARIAN PARTICLE THERAPY JUERGEN HEESE New gantry developments Viewpoint from user and vendor perspective
More informationDose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them
Dose Distributions George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Purpose To familiarize the resident with dose distributions and the factors that affect them
More informationPhilips SPECT/CT Systems
Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET
More informationINTEGRATED DATA MIGRATION:
Technical Perspectives Integrated data migration INTEGRATED DATA MIGRATION: Patient record transfer with discrete RT data elements ARIA Database DATA MIGRATION Third-party Database The purpose of this
More informationMonaco VMAT. The Next Generation in IMRT/VMAT Planning. Paulo Mathias Customer Support TPS Application
Monaco VMAT The Next Generation in IMRT/VMAT Planning Paulo Mathias Customer Support TPS Application 11.05.2011 Background What is Monaco? Advanced IMRT/VMAT treatment planning system from Elekta Software
More informationUse of MRI in Radiotherapy: Technical Consideration
Use of MRI in Radiotherapy: Technical Consideration Yanle Hu, PhD Department of Radiation Oncology, Mayo Clinic Arizona 04/07/2018 2015 MFMER slide-1 Conflict of Interest: None 2015 MFMER slide-2 Objectives
More informationA NEW LASER FOCUS ON PATIENT SAFETY
A NEW LASER FOCUS ON PATIENT SAFETY PATIENT SAFETY STARTS HERE During CT Simulation, clinics rely on moveable lasers to mark where radiation will enter a patient s body to target the tumor. Fixed lasers
More informationComprehensive treatment planning for brachytherapy. Advanced planning made easy
Comprehensive treatment planning for brachytherapy Advanced planning made easy Oncentra Brachy offers a variety of smart tools that facilitate many of the repetitive tasks for you. In contemporary brachytherapy,
More informationComputational modelling and analysis: IOP Medical Physics Group
Computational modelling and analysis: IOP Medical Physics Group Using Bespoke Matlab Software for the Collection and Processing of Dynalog Files from the Varian Millennium 120 MLC system following delivery
More informationCT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION
CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology
More informationMichael Speiser, Ph.D.
IMPROVED CT-BASED VOXEL PHANTOM GENERATION FOR MCNP MONTE CARLO Michael Speiser, Ph.D. Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX September 1 st, 2012 CMPWG Workshop Medical
More informationDosimetric Analysis Report
RT-safe 48, Artotinis str 116 33, Athens Greece +30 2107563691 info@rt-safe.com Dosimetric Analysis Report SAMPLE, for demonstration purposes only Date of report: ----------- Date of irradiation: -----------
More informationProton dose calculation algorithms and configuration data
Proton dose calculation algorithms and configuration data Barbara Schaffner PTCOG 46 Educational workshop in Wanjie, 20. May 2007 VARIAN Medical Systems Agenda Broad beam algorithms Concept of pencil beam
More informationSimple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Simple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field Hideharu Miura, 1a Shuichi
More informationQuality assurance of a helical tomotherapy machine
INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 49 (2004) 2933 2953 PHYSICS IN MEDICINE AND BIOLOGY PII: S0031-9155(04)71892-8 Quality assurance of a helical tomotherapy machine J D Fenwick 1,2,WATomé
More informationLASER SOLUTIONS. Ensure Accurate Setup from Simulation to Treatment. MOVEABLE: CT SIM+ // FIXED: MICRO and MICRO+
LASER SOLUTIONS Ensure Accurate Setup from Simulation to Treatment MOVEABLE: CT SIM+ // FIXED: MICRO and MICRO+ THE LEADER IN LASER ALIGNMENT Gammex was the first company to replace incandescent lightbulbs
More informationI. INTRODUCTION. Figure 1. Radiation room model at Dongnai General Hospital
International Journal of Computational Engineering Research Vol, 04 Issue, 4 Simulation of Photon and Electron dose distributions 5 code for the treatment area using the linear electron accelerator (LINAC)
More informationMeasurement of inter and intra fraction organ motion in radiotherapy using cone-beam CT projection images
Measurement of inter and intra fraction organ motion in radiotherapy using cone-beam CT projection images T E Marchant, A M Amer 1 and C J Moore North Western Medical Physics, Christie Hospital NHS Foundation
More information