ViewRay System Commissioning
|
|
- Joleen Stanley
- 5 years ago
- Views:
Transcription
1 ViewRay System Commissioning Kyle R. Padgett, PhD, DABR Assistant Professor University of Miami Viewray Team: Matt Studenski, PhD Chet Ford, PhD Yidong Yang, PhD Nesrin Dogan, PhD Medical Physics Residents: Naru Lamichhane, PhD Xing Li, PhD
2 Disclosures None 2
3 Overview Overview of the MRIdian system Commissioning Safety MRI Mechanicals RT System Treatment Planning System End to End Testing
4 Commonly known as Viewray MRIdian Integrated magnetic resonance (MR)-guided radiation therapy (RT) instrument designed to provide simultaneous MR Imaging (MRI) and a range of external beam RT options Consists of three major subsystems MRI System RT System Adaptive RT System Introduction
5 The MRI Gapped horizontal solenoidal superconducting magnet 0.35 T whole body MRI, capable of volumetric and real-time imaging The MRI system consists of two separate magnets abutted with a 28 cm gap. Split gradient coil has an inner diameter of 80 cm. Resolution as low as 0.75 mm Planar real-time imaging at up to 4 frames per second. MR isocenter matched to RT isocenter Body & Surface coils are thin & uniformly attenuating
6 Split Super Conducting Magnet Allows Unobstructed Beam Path Fits in Standard Vaults, Pop-apart design for non-destructive rigging 0.35T Field strength provides 50cm DSV for large FOV imaging Less image distortion and less patient heating Minimal distortion of dose distribution 70cm bore to accommodate large patients Split gradient with a 28cm gap, slew 200mT/m/ms, 18mT/m peak, 30kW heat removal. The MRI
7 Low Field MRI High-field causes a loss of spatial integrity Magnetic susceptibility artifacts due to the patient scales with field strength High field distorts the dose distribution Electron return effects get worse with field strength High field heats the patient -SAR
8 Electron Return Effect The Magnetic Field has an effect upon the secondary electrons The stronger the Magnetic Field the more pronounced the effect Raaijmakers et. al. Phys. Med. Biol. 53 (2008)
9 MRI Sequences Available FISP - Fast Imaging with Steady Precession GE Gradient Echo (Only one small FOV option) FISP is a coherent technique that uses a fully balanced gradient waveform (true FISP) The image contrast is determined by T2*/T1 and mostly depends on the TR The speed and relative motion insensitivity of acquisition help to make the technique reliable. Other MRI sequences may become available on future software releases.
10 Setup Imaging All setup images shown use FISP imaging FOV ranges from 22 54cm Resolution ranges from mm Acquisition time ranges from sec GE setup image has a resolution of 7.5mm, a FOV of ~27cm and a acquisition time of 12min
11 Real Time Imaging / Gating All CINE images shown use FISP imaging FOV ranges from 27cm to 45cm Resolution ranges from 3.5mm to 10mm All are 4 frames per second for a single slice
12 The RT delivery system Equipped with three robotic 60 Co treatment heads mounted with 120 separation on a rotating gantry. Each treatment head operates independently and they can be operated simultaneously. Each head can deliver up to 1.8Gy per minute at the machine isocenter, which is 105 cm from each source, with the nominal source strength of 15kCi. The source is shielded by a depleted-uranium safe in the retracted position
13 The RT delivery system Rotating Gantry Assembly 3 Independent Co60 Headed Design Enabling IMRT, SBRT or 3D- Conformal External Beam Radiation with Asynchronous Delivery Mounted with 120 degree separation 15,000 Ci per source degree Rotation for 2 or 3 Head Operation for increased Reliability. 3 Doubly Focused MLC Systems 180 MLC Leaves. 60 per head 1.05 leaf thickness projected at isocenter
14 Source Operation Beam Indicator Beam Indicator Beam OFF Beam On 14
15 3 Doubly Focused MLC Systems 180 MLC Leaves Per Head 1.05cm leaf thickness projected at isocenter Fully interdigitating Average Leakage less than 0.375% The MLC
16 Adaptive RT System An integrated high-performance radiation planning and delivery software capable of auto contouring, Monte Carlo dose computation, and IMRT or conformal RT planning or both are used to support 3-dimensional conformal RT, IMRT, and on-couch ART. The speed of the TPS (9 field plans with complete optimization, leaf motion calculation, and dose calculations can be accomplished in less than 30 seconds) enables ART treatments based on the volumetric image of the day
17 Commissioning Equipment Water tank with manually driven chamber holder Non-magnetic ion chamber: Exradin Nonmagnetic A28 & standard A12 chambers MRI-compatible IMRT QA device: ArcCheck MR, Sun Nuclear Corporation MRI-compatible beam profiler: ArcCheck IC Profiler 1122-mr, Sun Nuclear Corporation Gafchromic Films and Film Processor 2D Spatial Integrity Phantom Spherical Water Phantoms CIRS Gating Phantom ACR MRI Phantom
18 Radiation shielding survey was performed by the University of Miami radiation safety office. All points outside the treatment vault showed radiation levels in compliance with regulatory limits specified in 10 CFR, part 20 (limit of 2mR/hr) Safety
19 Safety Door Interlocks: Beam was successfully interrupted when door is opened Emergency stop buttons: The 5 emergency stop buttons were enabled and generated safety interlocks on the console Beam on Warning Light: Warning light successfully operated during irradiation Prime Alert Functionality: Prime alert functionality was confirmed with a check source Intercom and AV Monitoring: Confirmed with successful communication between two individuals Backup Timer: Primary timer was disabled and the backup timer successfully terminated the beam Emergency Door operation: Operates when power is off Emergency Couch Retract: Can be initiated manually and when power is off Many other safety interlocks were also tested during ATP
20 Magnetic Field Homogeneity Phased Array Coil Elements Image Homogeneity Spatial Integrity High Contrast Resolution MRI Tests Slice Thickness Slice Position Accuracy Low Contrast Resolution Ghosting Ratio
21 Magnetic Field Homogeneity The homogeneity of the MRI system was measured using a large spherical water phantom A water spectrum was then collected to determine the frequency spread at FWHM Repeated at several different Gantry Angles and with the different RF coils FWHM Gantry 0 Gantry Gantry Gantry Gantry ppm 1.46ppm 1.44ppm 1.60ppm 1.74ppm <= 5ppm
22 Coil Element Tests The signal characteristics for the coil elements for all phased array coils were measured. 12 element Torso Coil 12 element Head and Neck Coil All SNR measurements significantly exceeded specifications Shape of signal profile was also qualitatively evaluated.
23 Homogeneity Homogeneity of the Body, Torso, and H&N Coils were measured and exceeded specifications 60% Body Coil 50% Torso Coil 50% H&N Coil The Signal to Noise of the Body, Torso, and H&N Coils were measured and exceeded specifications SNR > 12 Body Coil SNR > 30 Torso Coil SNR > 30 H&N Coil Uniformity % = 100 * 1 [(ROI Signal max ROI Signal min) / (ROI Signal max + ROI Signal min)] SNR = (ROI signal mean) * 0.66 / (ROI Noise SD)
24 Spatial Integrity 2D spatial integrity was measured with a custom made phantom provided by Viewray All measurements within inner circle must have less than 1mm of distortion All measurements within outer circle must have less than 2mm of distortion This test is also performed in our Monthly QA Pass-Rate Pass-Rate Max error 1mm Ring 2mm Ring Axial 100.0% 100.0% 1.764mm Coronal 99.3% 98.7% 2.134mm Sagittal 100.0% 100.0% 1.298mm centered Sagittal 12.5cm N/A 100.0% 1.387mm Patient Left Sagittal 12.5cm Patient Right N/A 100.0% 1.060mm
25 High Contrast Resolution To assess the scanner s ability to resolve small objects when the contrast to noise ratio is sufficiently high. Test utilizes the MRI ACR phantom Should be able to visualize/resolve several of the rows and columns of four. This test is also performed in our Monthly QA Figure 6: from ACR Phantom Test Guidance for Use of the Small MRI Phantom
26 To assess the accuracy with which a slice of specified thickness is achieved Test utilizes the MRI ACR phantom Two signal ramps with a slope of 10 to 1 are utilized Thickness should be 5mm ± 0.7mm This test is also performed in our Monthly QA Slice Thickness Slice thickness = 0.2 x (top x bottom)/(top + bottom)
27 Slice Position Accuracy To assess the accuracy with which slices can be prescribed at specific locations Test utilizes the MRI ACR phantom Two crossed wedged ramps of 45 degrees are utilized The difference in the two bar length determines the slice position accuracy ½ of the measured length is the positional discrepancy This test is also performed in our Monthly QA
28 Low Contrast Resolution To assess the extent to which objects of low contrast are discernible in the images Test utilizes the MRI ACR phantom The low contrast disks are holes drilled in thin sheets of plastic mounted in the phantom Must find 9 spokes on 4 subsequent slices in both the T1 and T2 acquisitions This test is also performed in our Monthly QA
29 Ghosting Ratio To assess the level of ghosting in the images Test utilizes the MRI ACR phantom ROIs are drawn inside the phantom and in 4 surrounding areas to calculate the ghosting. Ratio must be less than This test is also performed in our Monthly QA Ghosting Ratio = [(top + btm) (left + right)] / (2 x Large ROI)
30 Couch Motions Couch Level Couch Sag Couch orthogonal to imaging plane Laser / MRI Coincidence Mechanicals Radiation / MRI Coincidence Radiation Isocentricity Gantry Angle Accuracy
31 Couch Couch Motions Couch Level: A level was placed on the couch oriented in the transverse and longitudinal directions and recorded the values Couch Sag: Solid Water was placed on Head Side and then Foot Side and displacements measured with MRI 0.3mm Head 0.39 Foot Couch motion vs MRI coordinates: Spatial Integrity Phantom was imaged at two known locations then fused to demonstrate proper couch motion Couch orthogonal to MRI coordinates: Spatial Integrity Phantom was imaged in different orientations to confirm orthogonality
32 Radiation/MRI versus Laser Isocenter The radiation versus laser isocenter was determined by using the IC profiler and confirmed with starshots Laser and MRI coincidence was determined with daily QA phantom and MRI visible fiducials. The offset was determined by measuring the fiducials location versus the isocenter location. Sagittal Laser Offset Coronal Laser Offset Axial Laser Offset -0.7 mm 0.0 mm 0.8 mm
33 Gantry Radiation Isocenter 4 Starshots were collected at various gantry angles determine the size of the isocenter Laser / RT coincidence Due to the design of the system films cannot be placed at isocenter manually The Daily QA phantom has a circular film holder for this purpose but requires laser cut film
34 Gantry Angle Accuracy Radiation beams were delivered to the Arccheck device utilizing all three heads at various gantry angles and analyzed in the Arccheck software to determine gantry angle accuracy Gantry angles were also confirmed utilizing starshots collected on the Daily QA phantom Head Set Angle VR Readout AC Measured H H H H H H H
35 MLC Leakage MLC leakage films were collected on all three heads at different gantry angles to ensure leakage is within tolerance and isn t dependent on gantry rotation.
36 Employs a wire jig with 4cm separation between wires Measures MLC positioning accuracy Align the wire jig to the sagittal laser Treatment plan was created for a series of MLC segments centered at 0,- 12,-8,-4,4,8,12cm The MLC positioning accuracy is 0.2cm In house software written to process data, see us at AAPM in Washington DC. MLC Accuracy
37 Field Size Accuracy Square fields were measured with the IC Profiler with a 1 cm depth (the intrinsic buildup of the IC Profiler device) Field Size Field Size Axis Head Gantry Flatness Symmetry Neg_Penumb Pos_Penumb Beam Center Field Size Difference (mm) 4.2x4.2 x-measured x-measured x4.2 Y-measured Y-measured x10.5 X-measured X-measured x10.5 Y-measured Y-measured x21 X-measured X-measured x21 Y-measured Y-measured x27.3 X-measured X-measured x27.3 Y-measured Y-measured x4.2 x-measured x-measured x-measured x4.2 Y-measured Y-measured Y-measured x10.5 X-measured X-measured X-measured x10.5 Y-measured Y-measured Y-measured x21 X-measured X-measured X-measured x21 Y-measured Y-measured Y-measured x27.3 X-measured X-measured X-measured x27.3 Y-measured Y-measured Y-measured
38 Output Calibration A28 MRI compatible chambers employed Calibration conditions: 5 cm depth in solid-water 105 cm SAD, 10.5 x 10.5 cm field. Beams delivered from 90 degrees in solid water phantom For verification, heads 1 and 3 were measured in liquid water at 0 degrees (AP beam), and these measurements matched the measurements in solid water to less than a 0.5% difference.
39 RPC OSLDs Report
40 Percent Depth Doses A28 TPS-measured % Measured vs TPS depth (mm) 4.2x x x % % 0.90% 2.94% 80.00% % 0.00% 0.00% 70.00% % -0.11% -0.12% Percent Depth Dose 60.00% 50.00% 40.00% 30.00% 4.2 x 4.2 TPS 10.5 x 10.5 TPS 27.3 x 27.3 TPS 4.2 x 4.2 Measured % -0.44% -0.02% % -0.46% -0.04% % -0.66% 0.05% 20.00% 10.5 x 10.5 Measured % 0.14% 0.12% 10.00% 27.3 x 27.3 Measured % -0.17% 0.07% 0.00% % -0.67% 0.35% Depth (mm) % -0.87% 0.14%
41 Beam Profiles Profiles with different depths of solid water Three depths (1,5, and 10cm) with detector plane maintained at Iso (105cm SAD) 4.2x4.2 cm, 10.5x10.5 cm, 21.0x21.0 cm, and 27.3x27.3 cm field sizes were measured Same profiles were exported from the TPS and the profiler software was used to calculate the flatness and symmetry All profiles matched within specifications Depth = 10 cm Field Size Axis Head Gantry Flatness %Diff Symmetry %Diff 4.2x4.2 X-measured X-measured X-measured X-calc x4.2 Y-measured Y-measured Y-measured Y-calc x10.5 X-measured X-measured X-measured X-calc x10.5 Y-measured Y-measured Y-measured Y-calc x21.0 X-measured X-measured X-measured X-calc x21.0 Y-measured Y-measured Y-measured Y-calc x27.3 X-measured X-measured X-measured X-calc x27.3 Y-measured Y-measured Y-measured Y-calc x27.3 X-measured X-measured X-calc x27.3 Y-measured Y-measured Y-calc x27.3 X-measured X-measured X-calc x27.3 Y-measured Y-measured Y-calc 6 0.1
42 Shutter Timer Error A28 chamber was irradiated several times for two different durations (30s and 240s)
43 Shutter Dose Compensation 1ss nnnn 200ss nnnn/200 DDDDDDDD CCCCCCCCCCCCCCCCCCCCCCCC = 200ss nnnn/200 An A28 chamber was irradiated several times for two different lengths of time (200seconds and 1second) and the compensation time is calculated using the equation shown above Compensation values from the configuration file was recorded and compared to the calculated value
44 Timer Accuracy A 30-second exposure was delivered for each beam. Exposure time was verified with a stopwatch with NISTtraceable calibration. Head Set Time Measured Time Percent Error % % %
45 Timer Linearity Exposures of 5, 15, 30 and 60 seconds were measured with an ion chamber at 1.5 cm depth in solid water. The measured exposures were fit to a line and the line fit was compared to the measured value for each time. H Rdg Fit % Difference Meas-Fit Time (s) 5 s 15 s 30 s 60 s s 15 s 30 s 60 s Shutter Timer Error (sec) H H H
46 CIRS motion phantom Sensor is placed a known distance inferior to the most superior extent of motion Beam should trigger off when motion phantom reaches the level of this optical sensor Latency is discrepancy between actual beam off and optical sensor triggering. Measured with a four channel oscilloscope. Gating Latency
47 Gating Dosimetry
48 Couch Attenuation Relative Attenuation Gantry Angle
49 TPS Statistics Viewray utilizes a Monte Carlo based treatment planning system The number of particle histories used in the calculation increases the accuracy, but Using a high # of histories in adaptive planning can significantly increase the amount of time the patient is on the table. Max Dose Min Dose Mean Dose Standard Deviation 1-Million Histories Million Histories
50 End to End Tests IMRT Delivery IMRT Two Head Delivery RPC Head and Neck Phantom RPC Lung Phantom TG-119 Tests Head and Neck Prostate Multi Target C-Shape
51 IMRT Delivery Delivery Mode Gamma (3%, 3mm) Full Delivery Measured to Calculated Dose 99.4 % Interrupted Delivery: Measured to Calculated 100 % Interrupted Delivery: Measured to Measured 99.9 %
52 IMRT Two-Head Mode Delivery Mode Gamma (3%, 3mm) Full Delivery Measured to Calculated Dose 99.4 % Two Head Mode: (Two Head Mode) Measured to (Three Head Mode) Calculated (Two Head Mode) Measured to (Three Head Mode) Measured 99% 100%
53 RPC Head and Neck Primary PTV Rx Secondary PTV Rx OAR Max Normal Tissue 6.6Gy 5.4Gy 4.5Gy 7.26Gy
54 RPC Head and Neck Constraints Primary PTV Rx Secondary PTV Rx OAR Max Normal Tissue 6.6Gy 5.4Gy 4.5Gy 7.26Gy
55 RPC Head and Neck
56 RPC Lung Primary PTV Rx Cord Max Heart <33% Heart <66% Heart <100% Both Lungs <37% 6.0Gy 5.0Gy 6.0Gy 4.5Gy 4.0Gy 2.0Gy
57 RPC Lung Constraints Primary PTV Rx Cord Max Heart <33% Heart <66% Heart <100% Both Lungs <37% 6.0Gy 5.0Gy 6.0Gy 4.5Gy 4.0Gy 2.0Gy
58 RPC Lung
59 TG-119 Summary Test Gamma (2.5%/2.5mm) Head and Neck 95% Prostate 97.4% Multitarget 98.2% C-Shape 96.3%
60 Acknowledgements Alan Pollack, MD PhD Radka Stoyanova, PhD Viewray Inc: David Holloway, Maria Bellon, Rebecca Sandbrook, James Victoria, Michael Saracen, John Ryan UCLA: Minsong Cao and James Lamb WashU: Olga Green
Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.
Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing
More informationS. Guru Prasad, Ph.D., DABR
PURPOSE S. Guru Prasad, Ph.D., DABR Director of Medical Physics IAEA Consultant NorthShore University Health System and University of Chicago, Pritzker School of Medicine Current TPS utilize more information
More informationVolumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT
Volumetric Modulated Arc Therapy - Clinical Implementation Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA Acknowledgement David M. Shepard, Ph.D. Muhammad K. N. Afghan, Ph.D. Fan Chen, Ph.D.
More informationFAST, precise. qa software
qa software FAST, precise Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, rotational or brachytherapy plans no film, no phantoms, no linac
More informationTHE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!
THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER! Confidence in complex treatments Modern radiation therapy uses complex plans with techniques such as IMRT, VMAT and Tomotherapy.
More informationLucy Phantom MR Grid Evaluation
Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is
More informationIMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays
IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays Sotiri Stathakis Outline Why IMRT/VMAT QA AAPM TG218 UPDATE Tolerance Limits and Methodologies for IMRT Verification QA Common sources
More informationTomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS
Tomotherapy Physics Machine Twinning and Quality Assurance Emilie Soisson, MS Tomotherapy at UW- Madison Treating for nearly 5 years Up to ~45 patients a day on 2 tomo units Units twinned to facilitate
More informationAcknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF
Page 1 Quality Assurance of IMRT Delivery Systems - Siemens Lynn J. Verhey, Ph.D. Professor and Vice-Chair UCSF Dept. of Radiation Oncology AAPM 22 Annual Meeting, Montreal Acknowledgments Ping Xia, Ph.D.,
More informationIMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE
QA SOFTWARE FAST, PRECISE Software for accurate and independent verification of monitor units, dose, and overall validity of standard, IMRT, VMAT, SRS and brachytherapy plans no film, no phantoms, no linac
More informationC a t p h a n / T h e P h a n t o m L a b o r a t o r y
C a t p h a n 5 0 0 / 6 0 0 T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 Internationally recognized for measuring the maximum obtainable performance of axial, spiral and multi-slice
More informationIni$al Performance Test Results on an Elekta MR-Linac
Ini$al Performance Test Results on an Elekta MR-Linac Eric Paulson, PhD DABR Radia$on Oncology, Radiology, and Biophysics Froedtert and Medical College of Wisconsin Milwaukee, Wisconsin, United States
More information7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT
CLINICAL COMMISSIONING AND ACCEPTANCE TESTING OF A 3D SURFACE MATCHING SYSTEM Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago Learning Objective Describe acceptance
More informationRaising the Bar in IMRT QA
MapCHECK 2TM Raising the Bar in IMRT QA The leader in quick and precise measurement of modulated radiotherapy beams Benefits Proven solution for film-less rotational delivery and IMRT QA - More than 1500
More information8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques
Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery Lei Ren, Ph.D. Duke University Medical Center 2016 AAPM 58 th annual meeting, Educational Course, Therapy Track Disclosure I have
More informationBasic Radiation Oncology Physics
Basic Radiation Oncology Physics T. Ganesh, Ph.D., DABR Chief Medical Physicist Fortis Memorial Research Institute Gurgaon Acknowledgment: I gratefully acknowledge the IAEA resources of teaching slides
More informationQalitätssicherung an Multileafkollimatoren. Dr. Lutz Müller Würzburg jan 2004
Qalitätssicherung an Multileafkollimatoren Dr. Lutz Müller Würzburg jan 2004 IMRT Verification - present Target Volume Constraints Inverse Planning Algorithm Fluence Map Leaf Sequencer Leaf & Gantry sequence
More informationInitial Clinical Experience with 3D Surface Image Guidance
Initial Clinical Experience with 3D Surface Image Guidance Amanda Havnen-Smith, Ph.D. Minneapolis Radiation Oncology Ridges Radiation Therapy Center Burnsville, MN April 20 th, 2012 Non-funded research
More informationTG-148 overview. Introduction. System Overview. System Overview. QA for Helical Tomotherapy: Report of the AAPM Task Group 148. Conflict of Interest:
QA or Helical Tomotherapy: Report o the AAPM Tas Group 148 Members: Conlict o Interest: r. John Balog owns TomoTherapy stoc. Katja Langen (Co-chair) Nio Papaniolaou (Co-chair) Walter Grant Richard Crilly
More informationGafChromic Protocol Multi-Channel Film Dosimetry + Gamma Map Analysis
GafChromic Protocol Multi-Channel Film Dosimetry + Gamma Map Analysis Micke A. Ashland Inc. Advanced Materials Ashland proprietary patented technology GafChromic Film for Dose Measurement Radiotherapy
More informationMRI-LINAC Dynamic Phantom
MRI-LINAC Dynamic Phantom Model 008M IMAGE ACQUISITION TREATMENT PLANNING DOSE DELIVERY Strict QA procedures for the imaging, planning and delivery of radiotherapy using respiratory management devices
More informationMapCHECK 2 & 3DVH. The Gold Standard for 2D Arrays
MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.
More informationImage Guidance and Beam Level Imaging in Digital Linacs
Image Guidance and Beam Level Imaging in Digital Linacs Ruijiang Li, Ph.D. Department of Radiation Oncology Stanford University School of Medicine 2014 AAPM Therapy Educational Course Disclosure Research
More informationUNCOMPROMISING QUALITY
ION CHAMBERS UNCOMPROMISING QUALITY Designed with over 30 years of scientific integrity for a broad range of dosimetry measurements in diverse radiation beams Farmer-type Chambers For absolute dosimetry
More informationChapter 9 Field Shaping: Scanning Beam
Chapter 9 Field Shaping: Scanning Beam X. Ronald Zhu, Ph.D. Department of Radiation Physics M. D. Anderson Cancer Center Houston, TX June 14-18, 2015 AAPM - Summer School 2015, Colorado Spring Acknowledgement
More informationComparison of Two Phantoms for End-to-End SRS/SBRT Testing. Vikren Sarkar, Ph.D. Associate Professor University of Utah Huntsman Cancer Center
Comparison of Two Phantoms for End-to-End SRS/SBRT Testing Vikren Sarkar, Ph.D. Associate Professor University of Utah Huntsman Cancer Center Disclosure Travel and Speaker Honorarium from Sun Nuclear Corp.
More informationMapCHECK 2 & 3DVH The Gold Standard for 2D Arrays
MapCHECK 2 & 3DVH The Gold Standard for 2D Arrays Your Most Valuable QA and Dosimetry Tools THE GOLD STANDARD FOR 2D ARRAYS The MapCHECK 2 is the world s most selected independent 2D measurement array.
More informationMR-guided radiotherapy: Vision, status and research at the UMC Utrecht. Dipl. Ing. Dr. Markus Glitzner
MR-guided radiotherapy: Vision, status and research at the UMC Utrecht Dipl. Ing. Dr. Markus Glitzner About myself Training Medizintechnik TU Graz PhD UMC Utrecht Clinical work Software implementation
More informationImage Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD
Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image
More informationThe MSKCC Approach to IMRT. Outline
The MSKCC Approach to IMRT Spiridon V. Spirou, PhD Department of Medical Physics Memorial Sloan-Kettering Cancer Center New York, NY Outline Optimization Field splitting Delivery Independent verification
More informationImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing
ImPACT Information Leaflet No. 1: CT Scanner Acceptance Testing Version 1.02, 18/05/01 CONTENTS: 1. SCOPE OF LEAFLET 2. GENERAL PRINCIPLES OF ACCEPTANCE AND COMMISSIONING 2.1 PHANTOMS 2.2 EXPOSURE AND
More informationClinical implementation of photon beam flatness measurements to verify beam quality
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Clinical implementation of photon beam flatness measurements to verify beam quality Simon Goodall, a Nicholas Harding, Jake Simpson,
More informationph fax Deming Way Middleton WI USA
www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 3120 Deming Way Middleton WI 53562-1461 USA 2007 Standard Imaging, Inc. 1239-21 (03.07) beam qa LINAC AND TOMOTHERAPY QA The NEW
More informationDisclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions
LINAC and MLC QA for IMRT Ping Xia, Ph.D., Department of Radiation Oncology Disclosure Received research support from Siemens Medical Solutions University of California San Francisco Therapy Series (SAM)
More information3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending
3DVH : On The Accuracy Of The Planned Dose Perturbation Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools *Patent Pending introduction State-of-the-art IMRT QA of static gantry
More informationThe team. Disclosures. Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge.
Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge Dimitre Hristov Radiation Oncology Stanford University The team Renhui Gong 1 Magdalena Bazalova-Carter 1
More informationDynalog data tool for IMRT plan verification
Dynalog data tool for IMRT plan verification Poster No.: R-0051 Congress: 2014 CSM Type: Scientific Exhibit Authors: V. Sashin; FOOTSCRAY/AU Keywords: Computer applications, Radiation physics, Experimental,
More informationDelta 4 PT. True Volumetric Pre-treatment Verification. The difference is clear
Delta 4 PT True Volumetric Pre-treatment Verification The difference is clear ScandiDos innovators in advanced volumetric dosimetry solutions ScandiDos ScandiDos is the innovative company that introduced
More informationNew Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC
New Technology in Radiation Oncology James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC Technology s s everywhere From the imaging chain To the planning system To the linac To QA..it..it
More informationA secondary monitor unit calculation algorithm using superposition of symmetric, open fields for IMRT plans
Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2013 A secondary monitor unit calculation algorithm using superposition of symmetric, open fields for IMRT plans Adam
More informationMonte Carlo methods in proton beam radiation therapy. Harald Paganetti
Monte Carlo methods in proton beam radiation therapy Harald Paganetti Introduction: Proton Physics Electromagnetic energy loss of protons Distal distribution Dose [%] 120 100 80 60 40 p e p Ionization
More informationMachine and Physics Data Guide
WWW..COM Machine and Physics Data Guide STANDARD IMAGING, INC. 3120 Deming Way Middleton, WI 53562-1461 May / 2008 2008 Standard Imaging, Inc. TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 www.standardimaging.com
More informationICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588
ICARO Vienna April 27-29 2009 Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588 M. Saiful Huq, Ph.D., Professor and Director, Dept. of Radiation
More information8/3/2016. MR-guided RT: Commissioning and Quality Control. Topics for MR-guided RT system commissioning & QC
MR-guided RT: Commissioning and Quality Control T. Stanescu, PhD, MCCPM Medical Physicist, Princess Margaret Cancer Centre, UHN Assistant Professor, Radiation Oncology, University of Toronto Affiliated
More informationAdvanced Radiotherapy
Advanced Radiotherapy IBA Concept for IMRT Verification Dr. Lutz Müller 1 IMRT Verification BIS BIS ADAS Fluence Model Delivered fluence Target Volume Constraints Inverse Backprojection Fluence Map Leaf
More informationPreface. Med. Phys. 35(9), , Mechanical QA. Radiation Survey Mechanical tests Light radiation Table, Collimator, Gantry Jaws.
AAPM-SAM-2012-Das (1) Beam Data Collection and Commissioning for Linear Accelerators: Technical Considerations and Recommendations Preface Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation
More informationSPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.
SPECT QA and QC Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC What is needed? Why? How often? Who says? QA and QC in Nuclear Medicine QA - collective term for all the efforts made to produce
More informationTomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram
TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization
More informationNovel methods and tools for MR Commissioning and Quality Control
Novel methods and tools for MR Commissioning and Quality Control T. Stanescu, PhD, MCCPM Medical Physicist, RMP, Princess Margaret Cancer Centre Assistant Professor, Radiation Oncology, University of Toronto
More informationVersion 5.6. Quick Start Guide
WWW..COM Version 5.6 Quick Start Guide STANDARD IMAGING, INC. 3120 Deming Way Middleton, WI 53562-1461 Jul / 2018 2018 Standard Imaging, Inc. TEL 800.261.4446 TEL 608.831.0025 FAX 608.831.2202 DOC # 80714-05
More informationUse of MRI in Radiotherapy: Technical Consideration
Use of MRI in Radiotherapy: Technical Consideration Yanle Hu, PhD Department of Radiation Oncology, Mayo Clinic Arizona 04/07/2018 2015 MFMER slide-1 Conflict of Interest: None 2015 MFMER slide-2 Objectives
More informationIAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques
IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques Report of the Coordinated Research Project (CRP) on Development of Procedures
More informationPhoton beam dose distributions in 2D
Photon beam dose distributions in 2D Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2014 Acknowledgments! Narayan Sahoo PhD! Richard G Lane (Late) PhD 1 Overview! Evaluation
More informationQuality assurance of a helical tomotherapy machine
INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 49 (2004) 2933 2953 PHYSICS IN MEDICINE AND BIOLOGY PII: S0031-9155(04)71892-8 Quality assurance of a helical tomotherapy machine J D Fenwick 1,2,WATomé
More informationAcceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems. Bruce Curran University of Michigan Medical Center Ann Arbor, MI
Acceptance Testing and Commissioning of Monte Carlo Dose Calculation Systems Bruce Curran University of Michigan Medical Center Ann Arbor, MI Disclosure Some of the work discussed in this talk was supported
More informationRadiotherapy Plan Competition TomoTherapy Planning System. Dmytro Synchuk. Ukrainian Center of TomoTherapy, Kirovograd, Ukraine
Radiotherapy Plan Competition 2016 TomoTherapy Planning System Dmytro Synchuk Ukrainian Center of TomoTherapy, Kirovograd, Ukraine Beam Geometry 6MV fan beam 3 jaw options 1.0, 2.5 and 5 cm 64 leaves binary
More informationFeasibility of 3D Printed Patient specific Phantoms for IMRT QA and Other Dosimetric Special Procedures
Feasibility of 3D Printed Patient specific Phantoms for IMRT QA and Other Dosimetric Special Procedures ehler 046@umn.edu Eric Ehler, PhD Assistant Professor Department of Radiation Oncology What is 3D
More information4 Measurement. and Analysis. 4.1 Overview and Underlying Principles 4-1
Measurement and Analysis.1 Overview and Underlying Principles.1.1 Introductory Remarks The physics and setup for film dosimetry have been described in the previous chapters. The measurement setup for IMRT
More informationAnthropomorphic Verification of the Eclipse AAA Algorithm for Spine SBRT Treatments Involving Titanium Implants
Anthropomorphic Verification of the Eclipse AAA Algorithm for Spine SBRT Treatments Involving Titanium Implants Mike Grams Clinical Medical Physicist Mayo Clinic, Rochester, MN NCCAAPM Meeting, April 17,
More informationIntegrated proton-photon treatment planning
Pinnacle 3 Proton Planning Integrated proton-photon treatment planning Philips Pinnacle 3 Proton Planning specifications Pinnacle 3 Proton Planning is designed to simplify treatment planning for proton
More informationADVANCING CANCER TREATMENT
The RayPlan treatment planning system makes proven, innovative RayStation technology accessible to clinics that need a cost-effective and streamlined solution. Fast, efficient and straightforward to use,
More informationTPS COMMISSIONING. Laurence Court University of Texas MD Anderson Cancer Center 4/3/2017 1
TPS COMMISSIONING Laurence Court University of Texas MD Anderson Cancer Center 4/3/2017 1 Conflicts of interest Court receives funding from NIH, CPIRT, Varian and Elekta 4/3/2017 2 Resources your first
More informationPatient Set-ups and Tumor Localizations
Patient Set-ups and Tumor Localizations Amy S. Harrison Patient Positioning Prior to starting any localization or simulation procedure patients need to be positioned and immobilized Patients disease location
More informationThe IORT Treatment Planning System. radiance. GMV, 2012 Property of GMV All rights reserved
The IORT Treatment Planning System radiance Property of GMV All rights reserved WHY RADIANCE? JUSTIFICATION Property of GMV All rights reserved ADVANTAGES OF IORT PRECISION: RT guided by direct vision.
More informationADVANCING CANCER TREATMENT
3 ADVANCING CANCER TREATMENT SUPPORTING CLINICS WORLDWIDE RaySearch is advancing cancer treatment through pioneering software. We believe software has un limited potential, and that it is now the driving
More information8/2/2017. PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA
PerFRACTION from Sun Nuclear Automated Transit Dosimetry for In-Vivo QA Jennifer R. Clark, M.S., DABR Conflict of Interest Statement: I am an employee of Sun Nuclear Corporation Outline: Platform overview
More informationCLINICAL ASPECTS OF COMPACT GANTRY DESIGNS
CLINICAL ASPECTS OF COMPACT GANTRY DESIGNS J. Heese, J. Wulff, A. Winnebeck, A. Huggins, M. Schillo VARIAN PARTICLE THERAPY JUERGEN HEESE New gantry developments Viewpoint from user and vendor perspective
More informationOutline. Outline 7/24/2014. Fast, near real-time, Monte Carlo dose calculations using GPU. Xun Jia Ph.D. GPU Monte Carlo. Clinical Applications
Fast, near real-time, Monte Carlo dose calculations using GPU Xun Jia Ph.D. xun.jia@utsouthwestern.edu Outline GPU Monte Carlo Clinical Applications Conclusions 2 Outline GPU Monte Carlo Clinical Applications
More informationISO ISO ISO OHSAS ISO
ISO 9001 ISO 13485 ISO 14001 OHSAS 18001 ISO 27001 Pro-NM Performance 08-101 - standard version 08-103 - version with the PET Lid Phantom for NM and PET systems performance evaluation (collimator, artifacts,
More informationThe Dose Junction Issue Associated with Photon Beams for Large Volume Radiation Therapy and the Sensitivity to Set-up Error
Research Article imedpub Journals http://www.imedpub.com/ Journal of Medical Physics and Applied Sciences The Dose Junction Issue Associated with Photon Beams for Large Volume Radiation Therapy and the
More informationCode of practice: Create a verification plan for OCTAVIUS Detector 729 in Philips Pinnacle³
Technical Note D655.208.03/00 Code of practice: Create a verification plan for OCTAVIUS Detector 729 in There are basically three ways to carry out verification of IMRT fluences. For all options, it is
More informationMichael Speiser, Ph.D.
IMPROVED CT-BASED VOXEL PHANTOM GENERATION FOR MCNP MONTE CARLO Michael Speiser, Ph.D. Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX September 1 st, 2012 CMPWG Workshop Medical
More informationDosimetric Analysis Report
RT-safe 48, Artotinis str 116 33, Athens Greece +30 2107563691 info@rt-safe.com Dosimetric Analysis Report SAMPLE, for demonstration purposes only Date of report: ----------- Date of irradiation: -----------
More informationNEW METHOD OF COLLECTING OUTPUT FACTORS FOR COMMISSIONING LINEAR ACCELERATORS WITH SPECIAL EMPHASIS
NEW METHOD OF COLLECTING OUTPUT FACTORS FOR COMMISSIONING LINEAR ACCELERATORS WITH SPECIAL EMPHASIS ON SMALL FIELDS AND INTENSITY MODULATED RADIATION THERAPY by Cindy D. Smith A Thesis Submitted to the
More information3D Scanning Comes Full Circle
3D SCANNER TM 3D Scanning Comes Full Circle Relative 3D Dosimetry offering the easiest setup, most objectivity, and best consistency available Cross-Plane Advanced Design Ring and diameter drive mechanism
More informationMEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah
MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography
More informationQuick Reference Datasheet For All RIT113 Packages
Quick Reference Datasheet For All RIT113 Packages For Rotational Therapies, IMRT & TG142 Highlights and selected product information only. A complete TG142 brochure is available. For more information on
More informationAn Investigation of a Model of Percentage Depth Dose for Irregularly Shaped Fields
Int. J. Cancer (Radiat. Oncol. Invest): 96, 140 145 (2001) 2001 Wiley-Liss, Inc. Publication of the International Union Against Cancer An Investigation of a Model of Percentage Depth Dose for Irregularly
More informationCentralite CT Moving Laser Patient Positioning System (MRR-1)
Centralite CT Moving Laser Patient Positioning System (MRR-1) Installation and Setup Manual DIACOR, INC. 2550 DECKER LAKE BLVD., SUITE 26, WEST VALLEY CITY, UTAH 84119 800 342-2679 / 801 467-0050 / 801
More informationDAILY LINAC QA BEAM QA
BEAM QA DAILY LINAC QA The QA BeamChecker Plus allows for fast, reliable, and uncomplicated daily QA of Varian, Elekta, Siemens, and Accuray Treatment Machines. The QA BeamChecker Plus is specifically
More informationAn Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy
An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy Chenyang Xu 1, Siemens Corporate Research, Inc., Princeton, NJ, USA Xiaolei Huang,
More informationCreating a Knowledge Based Model using RapidPlan TM : The Henry Ford Experience
DVH Estimates Creating a Knowledge Based Model using RapidPlan TM : The Henry Ford Experience Karen Chin Snyder, MS, DABR AAMD Region V Meeting October 4, 2014 Disclosures The Department of Radiation Oncology
More informationChapter 7 Feasibility Study for 3D Online Motion Compensation
Chapter 7 Feasibility Study for 3D Online Motion Compensation The current section presents the results of the experimental study of the feasibility of three-dimensional motion online compensation (3D-OMC)
More informationEvaluation of AutoQA Lite TM Image Quality Measurement Software
Evaluation of AutoQA Lite TM Image Quality Measurement Software Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4 2XU Phone: 0131 537 1161
More informationAutomated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator
Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4
More informationDiscontinued Products
Discontinued Products January 2019 CIRS was founded in 1983 to improve upon existing tissue simulation methodology and provide quantitative reference standards for Computed Tomography. Today CIRS is recognized
More informationTable of Contents. About AQUA
User Manual About AQUA Table of Contents 1 About AQUA... 6 1.1 Radiation Therapy... 6 1.1.1 Workflow manager... 6 1.1.2 Reports... 7 1.1.3 Tests... 7 1.1.4 Many Ways to Work... 7 1.2 Technology... 7 1.3
More informationINSTRUCTIONS RPD INFORMATION RPD PRODUCT INFORMATION. Address 5218 Barthel Industrial Drive Albertville, MN Website
Expect Service INSTRUCTIONS Radiation Products Design Inc RPD INFORMATION Address 5218 Barthel Industrial Drive Albertville, MN 55301 Website www.rpdinc.com Email sales@rpdinc.com Phone 763-497-2071 or
More information3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools
3D SCANNER 3D Scanning Comes Full Circle Your Most Valuable QA and Dosimetry Tools BETTER OUTCOMES THROUGH TECHNOLOGY The 3D SCANNER is different by design. 3D SCANNER has been developed from the ground
More informationData. ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy
Data ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy The ModuLeaf Mini Multileaf
More informationContinuation Format Page
C.1 PET with submillimeter spatial resolution Figure 2 shows two views of the high resolution PET experimental setup used to acquire preliminary data [92]. The mechanics of the proposed system are similar
More information1. Learn to incorporate QA for surface imaging
Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago ***No disclosures*** 1. Learn to incorporate QA for surface imaging into current QA procedures for IGRT. 2. Understand
More informationThank-You Members of TG147 TG 147: QA for nonradiographic
Thank-You Members of TG147 TG 147: QA for nonradiographic localization and positioning systems Twyla Willoughby, M.S. Medical Physicist Clinical AAPM Meeting March 2013 Department of Radiation Oncology
More informationDesign and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon
Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization
More informationThree-dimensional dosimetry of small fields
Three-dimensional of small fields Travis J. McCaw Medical Radiation Research Center Department of Medical Physics University of Wisconsin, Madison, WI NCCAAPM 2013 Fall Meeting 10 October 2013 Medical
More informationSpiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA
Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization
More informationA SYSTEM OF DOSIMETRIC CALCULATIONS
A SYSTEM OF DOSIMETRIC CALCULATIONS INTRODUCTION Dose calculation based on PDD and TAR have Limitations The dependence of PDD on SSD Not suitable for isocentric techniques TAR and SAR does not depend on
More informationOverview of Proposed TG-132 Recommendations
Overview of Proposed TG-132 Recommendations Kristy K Brock, Ph.D., DABR Associate Professor Department of Radiation Oncology, University of Michigan Chair, AAPM TG 132: Image Registration and Fusion Conflict
More informationIMRT site-specific procedure: Prostate (CHHiP)
IMRT site-specific procedure: Prostate (CHHiP) Scope: To provide site specific instructions for the planning of CHHIP IMRT patients Responsibilities: Radiotherapy Physicists, HPC Registered Therapy Radiographers
More informationPURE. ViSION Edition PET/CT. Patient Comfort Put First.
PURE ViSION Edition PET/CT Patient Comfort Put First. 2 System features that put patient comfort and safety first. Oncology patients deserve the highest levels of safety and comfort during scans. Our Celesteion
More information