Live-cell 3D super-resolution imaging in thick biological samples

Size: px
Start display at page:

Download "Live-cell 3D super-resolution imaging in thick biological samples"

Transcription

1 Nature Methods Live-cell 3D super-resolution imaging in thick biological samples Francesca Cella Zanacchi, Zeno Lavagnino, Michela Perrone Donnorso, Alessio Del Bue, Lauria Furia, Mario Faretta & Alberto Diaspro Supplementary Figure 1 Supplementary Figure 2 Supplementary Figure 3 Supplementary Figure 4 Supplementary Figure 5 Supplementary Figure 6 Supplementary Figure 7 Supplementary Figure 8 Supplementary Figure 9 Supplementary Results 1 Supplementary Results 2 Supplementary Results 3 Supplementary Results 4 Set-up used for IML-SPIM Selective plane photoactivation Background comparison between widefield and SPIM imaging schemes Localization accuracy for IML SPIM images within cellular spheroids expressing Connexin43-PAmCherry 3D super resolution imaging in depth by means of elliptical PSF distortion 3D super resolution imaging in depth of nuclear PAmCherry protein. Scattering effects on deep single molecule detection capability Light sample interaction along the illumination path Point spread function characterization in thick scattering samples Background comparison between widefield SPIM Localization precision and resolution degradation in complex biological samples 3D super resolution by astigmatism-based technique Scattering effects on the single molecule detection and localization Note: Supplementary Video 1 is available on the Nature Methods website.

2 Supplementary Figure 1 Schematic representation of the IML-SPIM set-up.

3 Supplementary Figure 2 Selective plane photoactivation Activation process initiated by violet laser (405 nm) in single plane illumination system: photoactivation experiments have been performed on cells expressing H2B-PAmCherry fusion protein (A). For photoactivation an intensity of 0.2 kw/cm 2 is employed and a readout laser (561 nm) of 0.5 kw/cm 2 is used to excite the sample. The exposure time is 500 ms. Both the activation and the readout lasers have been maintained continuously running after the initiation of photoactivation (t = 13 s). Images related to different total irradiation times are shown in (A). The mean fluorescence intensity (emitted by PAmCherry molecules within the cell nucleus) is reported over time, showing the competition of the photoactivation and the photobleaching processes (B). A 40x detection lens of NA = 0.8 is used and the light sheet thickness is approx. 4 µm.

4 Supplementary Figure 3 Background comparison between widefield and SPIM. Signal to noise ratio improvement provided by SPIM. Imaging of FITC labeled nanocapsules performed using single plane illumination microscopy (A). Comparison between single plane illumination microscopy (B) and widefield imaging (C) of nanocapsules. Images are acquired in SPIM and widefield using the same excitation intensity at 488 nm (0.9 kw/cm 2 ) and same exposure time (100 ms). (B) and (C) clearly show the background reduction, provided by the single plane illumination scheme compared to widefield, due to out of focus contributions. The thickness of the light sheet is approximately 4 µm.

5 Supplementary Figure 4 Localization accuracy for IML SPIM images within cellular spheroids expressing Connexin43- PAmCherry Direct estimation of the effective localization accuracy within the cellular spheroid by repeated measurements on targeted sub-resolved molecular aggregates or on individual molecules. The localization accuracy was determined from repeated localization of point like objects in the spheroid which provide SNR conditions typical of IML SPIM experiments. The typical number of collected photons/molecule in IML SPIM imaging ranges from 100 to 500 photons and the measured background photons within spheroids (which is typically in the range of b = 3 4 photons/pixel) remain constant with depth (up to 100µm). Images are performed under these conditions (histogram of the collected photon number is shown in (B) and the image of a single localized molecule is shown as an example (A)). The spatial distribution of localizations in the radial (C) and axial (D) dimensions are shown. As expected, the localization accuracy in the axial direction was approximately twice the localization accuracy measured in the radial one. The histogram of localizations was generated by aligning different localized clusters by the mean value (each cluster contains more than ten localization events). Fitting with a gaussian function yielded to a standard deviation of 26 nm in the radial direction and 60 nm in the axial one. The corresponding FWHM provided an estimation of the localization accuracy within complex samples such as cellular spheroids (63 nm and 141 nm in the radial and in the axial direction, respectively). Images are performed using the experimental conditions typically used for single molecule detection in IML-SPIM experiments (frame rate of 50 ms and at readout laser power of 5 kw/cm 2 ). The thickness of the light sheet used to perform this experiment is approximately 1.8µm and the detection lens used is a 100x water dipping objective lens (NA = 1.1).

6 Suppplementary Figure 5 3D Super resolution imaging in depth by means of elliptical PSF distortion (A) The intensity point spread function obtained imaging sub-resolved fluorescent beads (Supplementary Results 3) at different depths showed the asymmetry due to astigmatism (induced inserting a 500 mm cylindrical lens in the detection path). IML SPIM is used to image human mammary MCF10A cell spheroids expressing Connexin43 PAmCherry 36 in 2D (B) or in 3Dsuperresolution configuration (C). Total amount of events: 30,000. The average localization precision obtained is 28 nm and the axial resolution obtained using astigmatic PSF is approximately 65 nm (Figure 1g). Photoactivation is achieved by 405 nm employing an activation intensity I = 0.25 kw/cm 2. Images have been acquired using 561 nm excitation at 5 kw/cm 2 and exposure time t = 40 ms. Total acquisition time is approximately 2 min. To increase the optical sectioning capabilities of the system, the thickness of the light sheet used to perform this experiment is approximately 1.8 µm and the detection lens used is a 100x water dipping objective lens (NA 1.1). Scale bar = 1 µm.

7 Supplementary Figure 6 Scattering effects on deep single molecule detection capability (A) Top view schematic diagram of the imaging geometry based on the calculations of the total optical path length within the sample (Supplementary Results 4). The red dot indicates a representative position of the sample within the gel cylinder. SPIM images of MCF10 acini at different depths within the samples (B). Sparse photoactivation has been induced by 405 nm radiation and the signal collected upon continuous 561 nm illumination. Single frame images showing the IML SPIM capability to detect single molecules (C) at different imaging depths (10 μm, 40 μm, 70 μm and 100 μm). Frequency histograms of number of photons collected for each molecule (D) show that single molecules can be detected without any appreciable loss in term of collected photons. For single molecule imaging of PAmCherry nuclear protein the activation intensity used is 0.05 kw/cm 2 and the readout one is 11 kw/cm 2. Scale bar 10 μm.

8 Supplementary Figure 7 Light sample interaction along the illumination path Images of the light sheet thickness for different optical pathways through a phantom sample with scattering coefficient 50 mm -1 (see Online Methods for sample preparation details). Images of the light sheet thickness has been collected rotating (90 ) the cylindrical lens in the conventional SPIM architecture. The position of the light sheet has been centered at 50 µm (A), 100 µm (B), 200 µm (C) deep within the sample. On the right, the intensity profiles along the direction of the light propagation (Y) show that no significant decay occurs over the entire field of view (Intensity fluctuations are within 5%). The excitation wavelength used was λ = 488 nm and laser intensity 0.3 kw/cm 2. Scale bar 10 µm.

9 Supplementary Figure 8 Point Spread Function characterization in thick scattering samples Lateral and axial experimental point spread function in non scattering samples (A) and in a phantom -1 sample (B) with scattering coefficient 50 mm (as described in Online Methods). PSF are measured imaging sub- diffraction sized beads (40 nm). Images are acquired 100 µm deep into the sample. The FWHM of the lateral PSF in non scattering samples (485 ± 4 nm) and in samples with scattering properties (489 ± 7 nm) remained constant. The axial extent of the PSF is (2.69 ± 0.25 µm) in case of non scattering sample and is (2.71 ± 0.33 µm) in the scattering one. Excitation wavelength used was λ = 488 nm and the laser intensity 0.15 kw/cm 2. Scale bar = 1 µm.

10 Supplementary Figure 9 3D Super resolution imaging in depth of nuclear PAmCherry protein. IML SPIM is used to create 3D super resolution images of a nucleus within a human mammary MCF10A cell spheroids expressing H2B-PAmCherry. IML SPIM imaging (A) of cell nuclei deep within a spheroid. The final image has been reconstructed after the collection and localization of 31,000 molecules. As a comparison the SPIM image obtained by adding the total signal over the frames is shown in (B). A z color coded map of the cell represented in the right region of (A) and (B) is shown in (C) to visualize z localizations corresponding to µm. XY section of the very same cell nucleus (D) and the 3D volume corresponding to the ROI in (D) is represented in (E) showing the XZ intensity projection. Supplementary Video 1 steps through XY slices within the sample. Imaging depth is 50 µm. Images are acquired with a frame rate of 50 ms and at a laser power of 5 kw/cm 2. Activation laser intensity I = 0.04 kw/cm 2. The activation and excitation wavelength used are λ = 405 nm and λ = 561 nm respectively. The thickness of the light sheet used to perform this experiment is approximately 1.8 µm and the detection lens used is a 100x water dipping objective lens (NA = 1.1). Total acquisition time for (A) is 3 minutes.

11 Supplementary Results 1 BACKGROUND COMPARISON BETWEEN WIDEFIELD SPIM Supplementary Figure 3A shows the sample as imaged by conventional SPIM. Supplementary Figure 3B, obtained using SPIM configuration, demonstrates the improvement in terms of background signal when compared to the image produced by conventional widefield illumination (Supplementary Figure 3C). The average number of photons/pixel in the background, calculated in regions inside the nanocapsules (Supplementary Figure 3B, C), is reduced in case of single plane illumination configuration (3.3 ± 0.4 photons/pixel) compared to the widefield illumination scheme (6.4 ± 0.6 photons/pixel). The measurement has been obtained by repeating the measurements statistically over 20 samples. Measurements are obtained by calculating the background in the center of the nanocapsules, and the obtained results are in agreement with measurements previously performed in the single molecule regime and presented in the literature 27. Supplementary Results 2 LOCALIZATION PRECISION AND RESOLUTION DEGRADATION IN COMPLEX BIOLOGICAL SAMPLES In order to provide a demonstration of the IML SPIM imaging capabilities we reported in the manuscript the value of the localization precision loc which was found to be less than 35 nm. This value represents the error due to the localization process and was calculated according to the analytical model earlier presented in literature 28, 29 : 2 loc 2 2 a s sb 2 N 9 a N where the term due to the background noise can be neglected thanks to the background suppression provided by single plane illumination. a represents the pixel size, N is the number of collected photons, s is the radial point spread function of the system and b is the background noise. For practical imaging of large scattering biological samples several limiting factors, mainly related to scattering and aberration effects, can contribute to a decreased effective localization precision. To consider additional errors induced in the localization process, the precision can be redefined by considering also the standard deviation inst of the instabilities of the system loc inst where the factor of the EMCCD. 2 takes into account for the excess noise introduced by electron multiplying process Concerning the axial accuracy of IML SPIM, tested on model samples by repeated localizations of subresolved fluorescent beads (Supplementary Results 3), the measured value was found to be approximately 65 nm. However, localization of a bead, composed of bright and randomly oriented fluorophores, does not provide the same localization performance of single molecules. Again, in case of

12 localization of single emitters within large scattering samples the effective localization precision will be limited by the low photon regime. In fact, even if the axial accuracy provided by the system was found to be 65 nm in model samples, the real localization accuracy could be affected by background signal and low light conditions when imaging biological complex samples. In order to experimentally evaluate the effective accuracy of the axial coordinate determination and to estimate the overall localization precision, point-like fluorescent objects have been repeatedly localized directly inside the cellular spheroids (Supplementary Figure 4). The measurement has been performed in the same experimental conditions used for IML SPIM imaging and led to an effective accuracy of about 63 nm in the radial direction and 141 nm in the axial one. This measurement provides the closest estimate of the effective resolution 31 obtained by IML-SPIM in large samples, since it takes into account for aberrations due to light-sample interactions and for additional errors in the localization precision due to instabilities of the system. Therefore, the experimental measure of the localization accuracy deep within cellular spheroids proves the super-resolution imaging capabilities of IML SPIM within large scattering samples. In this case, even if the localization precision itself strongly differed from the localization accuracy, a radial resolution of 63nm and axial resolution of 140 nm has been demonstrated (Supplementary Figure 4). Supplementary Results 3 3D SUPER RESOLUTION BY ASTIGMATISM BASED TECHNIQUE 3D super-resolution techniques, based on multiplane detection 32 and on astigmatic distortions, have been presented recently in literature 33, 34. In order to extend the imaging capabilities of the system to 3D super-resolution we implemented the method based on astigmatic distortions 33,35. The introduction of weak astigmatism, by inserting a cylindrical lens (f = 500 mm), into the detection path provided an elliptically stretched point spread function along a lateral axis depending on the axial position (Supplementary Figure 5A). The experimental 3D point spread function can be obtained by imaging subresolved fluorescent beads by scanning the sample through the focus of the detection lens. Calibration has been performed by acquiring a z stack of 40nm fluorescent beads (Invitrogen) with z step of 50 nm. Model images of 40 nm fluorescent markers at sampled z position (the 0 coordinate is referred to the light sheet center) are shown in Supplementary Figure 5A. Images are acquired in a scattering phantom sample (50 mm -1 scattering coefficient) at an imaging depth of 60 µm. Although the PSF was found not to be depth dependent in case of homogeneously scattering samples (Supplementary Figure 8), sample-dependent variations in the astigmatic PSF, due to irregular shape of the sample and scattering variations, can occur while imaging biological samples. This can induce small distortions in the elliptical shape of the PSF which cannot be easily taken into account by a theoretical PSF model. In order to remove errors induced in the assignment of the axial position of single molecules, the calibration was always performed by acquiring beads in the surroundings of the imaged region. To this end fiducial fluorescent markers are embedded together with the sample in the agarose gel. The 3D intensity distribution obtained is fitted to an elliptical Gaussian function and used to extract calibration curves. In such a way the relation between the x and y dimensions of the radial PSF and the z position of the particle is made explicit. Once the calibration curves are available it is possible to parameterize the x and y dimensions in terms of z. Non linear regression is then used to fit the intensity distributions in order to extract the z coordinate. The spatial resolution along the z direction provided by the system can be calculated by localizing sub-resolved fluorescent beads several times.

13 The histogram of the distribution of z coordinates (Figure 1g) is obtained localizing at least six beads 300 times each 33. The z coordinate is calculated and each distribution is aligned by the mean value (the obtained distribution is shown in Figure 1g). The standard deviation of the obtained gaussian distribution is 27 nm, which corresponds to an effective resolution of about 65 nm. IML-SPIM is used to create 3D super-resolution images in depth of overexpressed Connexin43-PAmCherry (Supplementary Figure 5C) and H2B- PAmCherry in infected human mammary MCF10A cell spheroids. Supplementary Figure 9A, 9B show the comparison between SPIM and 3D superresolution images. In addition, 3D images are shown using both a z-color coded map (Supplementary Figure 9C) and orthogonal projections (Supplementary Figure 9D, 9E). Supplementary Video 1 shows XY slices through the sample with 116 nm z step. Supplementary Results 4 SCATTERING EFFECTS ON THE SINGLE MOLECULE DETECTION AND LOCALIZATION The single molecule imaging capability can be affected by scattering phenomena and light-sample interactions while imaging large biological samples. Since image quality degradation with the optical path length through the sample is expected, we investigated the scattering effects both in the illumination and the detection pathways. We considered that the total optical path length is composed by the contribution due to the illumination and to the distance covered by the emitted photons to the imaging lens. To schematically show the geometry of the SPIM imaging architecture in relation to the length of the optical pathway (Supplementary Figure 6A) we calculated the total optical path length within an object having cylindrical symmetry (i.e. the agar cylinder experimentally used). The optical path length (OPL) is related to the image quality and is defined as: C Optical Path Length n s ds (1) where n(s) represents the local refractive index of the sample and C is the pathway through the sample. The optical path length has been calculated (Wolfram Mathematica 7), according to eq.1, for a homogeneous sample. The imaging geometry is sketched in Supplementary Figure 6A. A density map schematically visualizes the relationship between sample position and path length. Dark regions correspond to a short OPL and a better image quality while bright regions represent a long OPL and correspond to a degraded imaging capability. Since the imaging performance degrades with the increased total travel of the light within the sample, the contributions due to scattering processes along the detection path and the illumination path needed to be investigated separately. To this end, the number of photons collected for each single molecule has been verified at different depths within a large biological sample. Supplementary Figure 6B showed a series of selected planes of an early-stage mammary acinus formed by MCF10A infected cells, 30 μm spaced through its entire extension and positioned 50 μm apart from the agar edge. In Supplementary Figure 6B SPIM optical sections show the growing cells with the formation of a cavity leading to generation of the internal duct. The ability to perform individual molecule detection (Supplementary Figure 6C) is shown at different depths (10 µm, 40 µm,70 µm and 100 µm) within the whole 3D spatial organization of the sample. In order to evaluate the effects due to scattering through the detection path, the effective number of photons for each single event is calculated at different depths within the

14 entire spheroid. Despite a small fraction of emitted photons lost due to scattering effects, histograms in Supplementary Figure 6D showed that no appreciable loss occurs in the depth range µm within the spheroid. Nevertheless, is worth noting that scattering effects could play a relevant role in single molecule detection at greater imaging depths (> 100 µm), affecting the number of collected photons/molecule and leading to a depth-dependent localization precision. Here we showed that no appreciable loss of collected photons/molecule through the entire volume of the spheroid occurs (Supplementary Figure 6D). This implies a negligible effect of the imaging depth on the localization precision which strictly depends on the number of collected photons, according to analytical models presented in the literature 28,29. To investigate the effects due to the scattering processes through the illumination path, the light sheet intensity distribution at different positions is acquired (Supplementary Figure 7). Distortions in the shape of the illumination intensity distribution due to light-matter interaction may occur while imaging scattering and non transparent samples. To investigate this phenomenon, direct imaging of the emitted fluorescent intensity is performed using fluorescent phantom samples. The cylindrical lens is rotated by 90 in order to image the thickness of the light sheet directly onto the CCD camera (Supplementary Figure 7). Images of the light sheet intensity profile are acquired at different light sheet positions (50 µm, 100 µm and 200 µm). The intensity distribution along the direction of the light propagation (y) at different illumination path lengths clearly shows that no significant decay occurs over the entire field of view (intensity fluctuations are within 5%). The obtained result demonstrated that, under these experimental conditions, the illumination intensity distribution can be assumed as uniformly distributed along the entire field of view (80 µm x 80 µm). Furthermore it also confirms a negligible distortion in the light sheet shape. Since the localization precision depends on the number of collected photons/event as well as on the characteristics of the imaging system used 28, the experimental measurement of the PSF in scattering phantom sample is also provided (Supplementary Figure 8). Experimental point spread functions are measured by imaging sub-resolved 40 nm fluorescent beads (Invitrogen). Images are acquired at 100 µm imaging depth. Lateral and axial experimental point spread functions have been experimentally measured in non scattering samples (Supplementary Figure 8A) and in a phantom scattering sample (Supplementary Figure 8B) having a scattering coefficient of 50 mm -1 (prepared as described in Online Methods). The x z section of the intensity point spread functions showed weak aberration effects in both the cases. The PSF FWHM, which influences the localization precision value 28, is calculated by gaussian fitting. The lateral PSF in non-scattering samples (485 ± 4 nm) and in samples with scattering properties (489 ± 7 nm) remained constant. The invariance of the radial PSF, coupled with the neglectable loss of collected photons/molecule along the detection path, assured an invariant localization precision (within an imaging depth range of µm). The experiments described in this section have been performed using a light sheet thickness of approximately 4 µm and a 40x imaging lens (NA = 0.8). REFERENCES 27. Ritter et al. Opt. Express 16, (2008). 28. Thompson, R.E., Larson, D.R. & Webb, W.W. Biophys. J. 82, (2002) 29. Mortensen, K.I. et al. Nat Methods 7, (2010)

15 30. Aquino, D. et al. Nat Methods 8, (2011) 31. Hess, S.T. et al. Methods Mol Biol 544, (2009) 32. Juette, M.F. et al. Nat Methods 5, (2008) 33. Huang, B. et al. Science 319, (2008) 34. York, A.G. et al. Nat Methods 8, (2011) 35. Kao, H.P. & Verkman, A.S. Biophys. J. 67, (1994) 36. McLachlan, E. et al. Cancer Res 66, (2006)

Super-resolution 3D Reconstruction of thick biological samples: a computer vision perspective

Super-resolution 3D Reconstruction of thick biological samples: a computer vision perspective 2013 IEEE International Conference on Computer Vision Workshops Super-resolution 3D Reconstruction of thick biological samples: a computer vision perspective Alessio Del Bue, Francesca Cella Zanacchi and

More information

Supplementary Figure 1

Supplementary Figure 1 Supplementary Figure 1 Experimental unmodified 2D and astigmatic 3D PSFs. (a) The averaged experimental unmodified 2D PSF used in this study. PSFs at axial positions from -800 nm to 800 nm are shown. The

More information

Independent Resolution Test of

Independent Resolution Test of Independent Resolution Test of as conducted and published by Dr. Adam Puche, PhD University of Maryland June 2005 as presented by (formerly Thales Optem Inc.) www.qioptiqimaging.com Independent Resolution

More information

Lightsheet Z.1. Light Sheet Fluorescence Microscopy by Carl Zeiss. Fabrice Schmitt, Sales Manager Carl ZEISS France

Lightsheet Z.1. Light Sheet Fluorescence Microscopy by Carl Zeiss. Fabrice Schmitt, Sales Manager Carl ZEISS France Lightsheet Z.1 Light Sheet Fluorescence Microscopy by Carl Zeiss Fabrice Schmitt, Sales Manager Carl ZEISS France 12.12.2012 Light Sheet Fluorescence Microscopy (LSFM) Principle The Principle of Light

More information

Optical Sectioning. Bo Huang. Pharmaceutical Chemistry

Optical Sectioning. Bo Huang. Pharmaceutical Chemistry Optical Sectioning Bo Huang Pharmaceutical Chemistry Approaches to 3D imaging Physical cutting Technical difficulty Highest resolution Highest sensitivity Optical sectioning Simple sample prep. No physical

More information

EPFL SV PTBIOP BIOP COURSE 2015 OPTICAL SLICING METHODS

EPFL SV PTBIOP BIOP COURSE 2015 OPTICAL SLICING METHODS BIOP COURSE 2015 OPTICAL SLICING METHODS OPTICAL SLICING METHODS Scanning Methods Wide field Methods Point Scanning Deconvolution Line Scanning Multiple Beam Scanning Single Photon Multiple Photon Total

More information

2, the coefficient of variation R 2, and properties of the photon counts traces

2, the coefficient of variation R 2, and properties of the photon counts traces Supplementary Figure 1 Quality control of FCS traces. (a) Typical trace that passes the quality control (QC) according to the parameters shown in f. The QC is based on thresholds applied to fitting parameters

More information

DeltaVision OMX SR super-resolution microscope. Super-resolution doesn t need to be complicated

DeltaVision OMX SR super-resolution microscope. Super-resolution doesn t need to be complicated DeltaVision OMX SR super-resolution microscope Super-resolution doesn t need to be complicated Live-cell super-resolution microscopy DeltaVision OMX SR is a compact super-resolution microscope system optimized

More information

Supplementary Figure 1 Guide stars of progressively longer wavelengths can be used for direct wavefront sensing at increasingly large depth in the

Supplementary Figure 1 Guide stars of progressively longer wavelengths can be used for direct wavefront sensing at increasingly large depth in the Supplementary Figure 1 Guide stars of progressively longer wavelengths can be used for direct wavefront sensing at increasingly large depth in the cortex of the living mouse. Typical SH images of guide

More information

Supporting Information for Azimuthal Polarization Filtering for Accurate, Precise, and Robust Single-Molecule Localization Microscopy

Supporting Information for Azimuthal Polarization Filtering for Accurate, Precise, and Robust Single-Molecule Localization Microscopy Nano Letters Supporting Information for Azimuthal Polarization Filtering for Accurate, Precise, and Robust Single-Molecule Localization Microscopy Matthew D. Lew, and W. E. Moerner *, Departments of Chemistry

More information

Vutara 350. Innovation with Integrity. The Fastest, Super-Resolution Microscope Deep 3D Imaging on Live Cells, Quickly and Easily

Vutara 350. Innovation with Integrity. The Fastest, Super-Resolution Microscope Deep 3D Imaging on Live Cells, Quickly and Easily Vutara 350 The Fastest, Super-Resolution Microscope Deep 3D Imaging on Live Cells, Quickly and Easily Innovation with Integrity Fluorescence Microscopy Vutara 350 Don t Get Left Behind Bruker s Vutara

More information

Back Illuminated Scientific CMOS

Back Illuminated Scientific CMOS Prime 95B Scientific CMOS Camera Datasheet CMOS, EMCCD AND CCD CAMERAS FOR LIFE SCIENCES Back Illuminated Scientific CMOS Discovery depends on every photon Primary applications: Super-Resolution Microscopy

More information

MetaMorph Super-Resolution Software. Localization-based super-resolution microscopy

MetaMorph Super-Resolution Software. Localization-based super-resolution microscopy MetaMorph Super-Resolution Software Localization-based super-resolution microscopy KEY BENEFITS Lateral Resolution up to 20 nm & Axial Resolution up to 40nm Real-time and offline superresolution image

More information

ALBERT-LUDWIG UNIVERSITY FREIBURG Institute of Computer Science Chair for Pattern Recognition and Image Processing

ALBERT-LUDWIG UNIVERSITY FREIBURG Institute of Computer Science Chair for Pattern Recognition and Image Processing ALBERT-LUDWIG UNIVERSITY FREIBURG Institute of Computer Science Chair for Pattern Recognition and Image Processing Multiview Deblurring for 3-D Images from Light Sheet based Fluorescence Microscopy - Supporting

More information

COLOCALISATION. Alexia Ferrand. Imaging Core Facility Biozentrum Basel

COLOCALISATION. Alexia Ferrand. Imaging Core Facility Biozentrum Basel COLOCALISATION Alexia Ferrand Imaging Core Facility Biozentrum Basel OUTLINE Introduction How to best prepare your samples for colocalisation How to acquire the images for colocalisation How to analyse

More information

Supplementary Information

Supplementary Information Supplementary Information Interferometric scattering microscopy with polarization-selective dual detection scheme: Capturing the orientational information of anisotropic nanometric objects Il-Buem Lee

More information

Confocal Microscope Imaging of Single-Emitter Fluorescence and Hanbury Brown & Twiss Setup for Photon Antibunching. Edward Pei

Confocal Microscope Imaging of Single-Emitter Fluorescence and Hanbury Brown & Twiss Setup for Photon Antibunching. Edward Pei Confocal Microscope Imaging of Single-Emitter Fluorescence and Hanbury Brown & Twiss Setup for Photon Antibunching Edward Pei Abstract The purpose of these labs was to study single photon sources and measure

More information

High Resolution BSI Scientific CMOS

High Resolution BSI Scientific CMOS CMOS, EMCCD AND CCD CAMERAS FOR LIFE SCIENCES High Resolution BSI Scientific CMOS Prime BSI delivers the perfect balance between high resolution imaging and sensitivity with an optimized pixel design and

More information

Virtual Frap User Guide

Virtual Frap User Guide Virtual Frap User Guide http://wiki.vcell.uchc.edu/twiki/bin/view/vcell/vfrap Center for Cell Analysis and Modeling University of Connecticut Health Center 2010-1 - 1 Introduction Flourescence Photobleaching

More information

Mode-Field Diameter and Spot Size Measurements of Lensed and Tapered Specialty Fibers

Mode-Field Diameter and Spot Size Measurements of Lensed and Tapered Specialty Fibers Mode-Field Diameter and Spot Size Measurements of Lensed and Tapered Specialty Fibers By Jeffrey L. Guttman, Ph.D., Director of Engineering, Ophir-Spiricon Abstract: The Mode-Field Diameter (MFD) and spot

More information

1. ABOUT INSTALLATION COMPATIBILITY SURESIM WORKFLOWS a. Workflow b. Workflow SURESIM TUTORIAL...

1. ABOUT INSTALLATION COMPATIBILITY SURESIM WORKFLOWS a. Workflow b. Workflow SURESIM TUTORIAL... SuReSim manual 1. ABOUT... 2 2. INSTALLATION... 2 3. COMPATIBILITY... 2 4. SURESIM WORKFLOWS... 2 a. Workflow 1... 3 b. Workflow 2... 4 5. SURESIM TUTORIAL... 5 a. Import Data... 5 b. Parameter Selection...

More information

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar Doctoral Thesis Time-of-Flight Time-of-Flight Positron Positron Emission Emission Tomography Tomography Using Using Cherenkov Cherenkov Radiation Radiation Rok Dolenec Supervisor: Prof. Dr. Samo Korpar

More information

Introductory Guide to Light Microscopy - Biomedical Confocal Microscopy

Introductory Guide to Light Microscopy - Biomedical Confocal Microscopy Introductory Guide to Light Microscopy - Biomedical Confocal Microscopy 7 May 2007 Michael Hooker Microscopy Facility Michael Chua microscopy@unc.edu 843-3268 6007 Thurston Bowles Wendy Salmon wendy_salmon@med.unc.edu

More information

Last class. This class. Single molecule imaging Deconvolution. FLIM Confocal

Last class. This class. Single molecule imaging Deconvolution. FLIM Confocal FLIM, Confocal Last class Single molecule imaging Deconvolution This class FLIM Confocal FLIM Fluorescence Lifetime IMaging FLIM Fluorescence lifetime imaging Easier and more accurate quantitation Much

More information

Faster STORM using compressed sensing. Supplementary Software. Nature Methods. Nature Methods: doi: /nmeth.1978

Faster STORM using compressed sensing. Supplementary Software. Nature Methods. Nature Methods: doi: /nmeth.1978 Nature Methods Faster STORM using compressed sensing Lei Zhu, Wei Zhang, Daniel Elnatan & Bo Huang Supplementary File Supplementary Figure 1 Supplementary Figure 2 Supplementary Figure 3 Supplementary

More information

TN425: A study of fluorescence standards confirms that OptiGrid confocal images are suitable for quantitative microscopy

TN425: A study of fluorescence standards confirms that OptiGrid confocal images are suitable for quantitative microscopy TN425: A study of fluorescence standards confirms that OptiGrid confocal images are suitable for quantitative microscopy Introduction The OptiGrid converts the illumination system of a conventional wide

More information

A fast method for estimation of light flux in fluorescence image guided surgery

A fast method for estimation of light flux in fluorescence image guided surgery A fast method for estimation of light flux in fluorescence image guided surgery 1. Introduction In this document, we present a theoretical method to estimate the light flux in near-infrared fluorescence

More information

Presentation and analysis of multidimensional data sets

Presentation and analysis of multidimensional data sets Presentation and analysis of multidimensional data sets Overview 1. 3D data visualisation Multidimensional images Data pre-processing Visualisation methods Multidimensional images wavelength time 3D image

More information

NDD FLIM Systems for Leica SP2 MP and SP5 MP Multiphoton Microscopes

NDD FLIM Systems for Leica SP2 MP and SP5 MP Multiphoton Microscopes NDD FLIM Systems for Leica SP2 MP and SP5 MP Multiphoton Microscopes bh FLIM systems for the confocal and the multiphoton versions of the Leica SP2 and SP5 microscopes are available since 2002 [4]. These

More information

Ray Optics I. Last time, finished EM theory Looked at complex boundary problems TIR: Snell s law complex Metal mirrors: index complex

Ray Optics I. Last time, finished EM theory Looked at complex boundary problems TIR: Snell s law complex Metal mirrors: index complex Phys 531 Lecture 8 20 September 2005 Ray Optics I Last time, finished EM theory Looked at complex boundary problems TIR: Snell s law complex Metal mirrors: index complex Today shift gears, start applying

More information

High spatial resolution measurement of volume holographic gratings

High spatial resolution measurement of volume holographic gratings High spatial resolution measurement of volume holographic gratings Gregory J. Steckman, Frank Havermeyer Ondax, Inc., 8 E. Duarte Rd., Monrovia, CA, USA 9116 ABSTRACT The conventional approach for measuring

More information

4.2 Megapixel Scientific CMOS Camera

4.2 Megapixel Scientific CMOS Camera CMOS, EMCCD AND CCD CAMERAS FOR LIFE SCIENCES 4.2 Megapixel Scientific CMOS Camera Photometrics Prime is the first intelligent scientific CMOS (scmos) camera to incorporate a powerful FPGA-based Embedded

More information

3-D. Here red spheres show the location of gold nanoparticles inside/around a cell nucleus.

3-D. Here red spheres show the location of gold nanoparticles inside/around a cell nucleus. 3-D The CytoViva 3-D System allows the user can locate objects of interest in a 3-D space. It does this by acquiring multiple Z planes and performing our custom software routines to locate and observe

More information

Quantifying Three-Dimensional Deformations of Migrating Fibroblasts

Quantifying Three-Dimensional Deformations of Migrating Fibroblasts 45 Chapter 4 Quantifying Three-Dimensional Deformations of Migrating Fibroblasts This chapter presents the full-field displacements and tractions of 3T3 fibroblast cells during migration on polyacrylamide

More information

45 µm polystyrene bead embedded in scattering tissue phantom. (a,b) raw images under oblique

45 µm polystyrene bead embedded in scattering tissue phantom. (a,b) raw images under oblique Phase gradient microscopy in thick tissue with oblique back-illumination Tim N Ford, Kengyeh K Chu & Jerome Mertz Supplementary Figure 1: Comparison of added versus subtracted raw OBM images 45 µm polystyrene

More information

Detectors for Future Light Sources. Gerhard Grübel Deutsches Elektronen Synchrotron (DESY) Notke-Str. 85, Hamburg

Detectors for Future Light Sources. Gerhard Grübel Deutsches Elektronen Synchrotron (DESY) Notke-Str. 85, Hamburg Detectors for Future Light Sources Gerhard Grübel Deutsches Elektronen Synchrotron (DESY) Notke-Str. 85, 22607 Hamburg Overview Radiation from X-Ray Free Electron lasers (XFEL, LCLS) Ultrafast detectors

More information

Raster Image Correlation Spectroscopy and Number and Brightness Analysis

Raster Image Correlation Spectroscopy and Number and Brightness Analysis Raster Image Correlation Spectroscopy and Number and Brightness Analysis 15 Principles of Fluorescence Course Paolo Annibale, PhD On behalf of Dr. Enrico Gratton lfd Raster Image Correlation Spectroscopy

More information

COLOCALISATION. Alexia Loynton-Ferrand. Imaging Core Facility Biozentrum Basel

COLOCALISATION. Alexia Loynton-Ferrand. Imaging Core Facility Biozentrum Basel COLOCALISATION Alexia Loynton-Ferrand Imaging Core Facility Biozentrum Basel OUTLINE Introduction How to best prepare your samples for colocalisation How to acquire the images for colocalisation How to

More information

FLUOSTAR Rhodamine B- encapsulating microspheres are seeding particles optimized for Particle Image Velocimetry. Technical handbook ver.

FLUOSTAR Rhodamine B- encapsulating microspheres are seeding particles optimized for Particle Image Velocimetry. Technical handbook ver. www.ebm.vc FLUOSTAR Rhodamine B- encapsulating microspheres are seeding particles optimized for Particle Image Velocimetry Technical handbook ver.1, Feb, 2010 CONTENTS 1) Introduction of fluorescent PIV

More information

SUPPLEMENTARY INFORMATION DOI: /NPHOTON

SUPPLEMENTARY INFORMATION DOI: /NPHOTON DOI:.38/NPHOTON.2.85 This supplement has two parts. In part A, we provide the rigorous details of the wavefront correction algorithm and show numerical simulations and experimental data for the cases of

More information

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information

OPTI-521 Graduate Report 2 Matthew Risi Tutorial: Introduction to imaging, and estimate of image quality degradation from optical surfaces

OPTI-521 Graduate Report 2 Matthew Risi Tutorial: Introduction to imaging, and estimate of image quality degradation from optical surfaces OPTI-521 Graduate Report 2 Matthew Risi Tutorial: Introduction to imaging, and estimate of image quality degradation from optical surfaces Abstract The purpose of this tutorial is to introduce the concept

More information

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation

More information

Biological Image Information I: A Description of the Modalities

Biological Image Information I: A Description of the Modalities 2.771J BEH.453J HST.958J Spring 2005 Lecture 21 April 2005 Biological Image I: A Description of the Modalities The Modalities Direct photography SEM and TEM Cryo-EM Two-photon microscopy Confocal microscopy

More information

On the uncertainty of Astigmatic Particle Tracking Velocimetry in the depth direction

On the uncertainty of Astigmatic Particle Tracking Velocimetry in the depth direction On the uncertainty of Astigmatic Particle Tracking Velocimetry in the depth direction Massimiliano Rossi 1,*, Christian J. Kähler 1 1: Institute for Fluid Mechanics and Aerodynamics, Bundeswehr University

More information

Supporting Information: Highly tunable elastic dielectric metasurface lenses

Supporting Information: Highly tunable elastic dielectric metasurface lenses Supporting Information: Highly tunable elastic dielectric metasurface lenses Seyedeh Mahsa Kamali, Ehsan Arbabi, Amir Arbabi, u Horie, and Andrei Faraon SUPPLEMENTAR NOTE : SAMPLING FREQUENC OF THE PHASE

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

Depth-variant blind restoration with pupil-phase constraints for 3D confocal microscopy

Depth-variant blind restoration with pupil-phase constraints for 3D confocal microscopy Journal of Physics: Conference Series OPEN ACCESS Depth-variant blind restoration with pupil-phase constraints for 3D confocal microscopy To cite this article: Saima Ben Hadj et al 2013 J. Phys.: Conf.

More information

Chapter 24. Wave Optics. Wave Optics. The wave nature of light is needed to explain various phenomena

Chapter 24. Wave Optics. Wave Optics. The wave nature of light is needed to explain various phenomena Chapter 24 Wave Optics Wave Optics The wave nature of light is needed to explain various phenomena Interference Diffraction Polarization The particle nature of light was the basis for ray (geometric) optics

More information

Supplementary Figure 1 Optimum transmissive mask design for shaping an incident light to a desired

Supplementary Figure 1 Optimum transmissive mask design for shaping an incident light to a desired Supplementary Figure 1 Optimum transmissive mask design for shaping an incident light to a desired tangential form. (a) The light from the sources and scatterers in the half space (1) passes through the

More information

FLUOVIEW FV1000/FV1200

FLUOVIEW FV1000/FV1200 FLUOVIEW FV1000/FV1200 UPGRADE TO 3D NANOIMAGING AND SINGLE MOLECULE TRACKING FOR OLYMPUS FLUOVIEW FV1000/FV1200 Within the past few years, several methods have been devised in order to obtain images with

More information

QUANTITATIVE PHASE IMAGING OF BIOLOGICAL CELLS USING OFF-AXIS METHOD OF WIDE FIELD DIGITAL INTERFERENCE MICROSCOPY (WFDIM)

QUANTITATIVE PHASE IMAGING OF BIOLOGICAL CELLS USING OFF-AXIS METHOD OF WIDE FIELD DIGITAL INTERFERENCE MICROSCOPY (WFDIM) http:// QUANTITATIVE PHASE IMAGING OF BIOLOGICAL CELLS USING OFF-AXIS METHOD OF WIDE FIELD DIGITAL INTERFERENCE MICROSCOPY (WFDIM) Pradeep Kumar Behera 1, Dalip Singh Mehta 2 1,2 Physics,Indian Institute

More information

PH880 Topics in Physics

PH880 Topics in Physics PH880 Topics in Physics Modern Optical Imaging (Fall 2010) Overview of week 8 Monday Nonlinear Microscopy Wednesday No class (Mid term week) Quantum Optics Electro- magnetic Optics Wave Optics Ray Optics

More information

Chapter 24. Wave Optics. Wave Optics. The wave nature of light is needed to explain various phenomena

Chapter 24. Wave Optics. Wave Optics. The wave nature of light is needed to explain various phenomena Chapter 24 Wave Optics Wave Optics The wave nature of light is needed to explain various phenomena Interference Diffraction Polarization The particle nature of light was the basis for ray (geometric) optics

More information

Diffraction. Single-slit diffraction. Diffraction by a circular aperture. Chapter 38. In the forward direction, the intensity is maximal.

Diffraction. Single-slit diffraction. Diffraction by a circular aperture. Chapter 38. In the forward direction, the intensity is maximal. Diffraction Chapter 38 Huygens construction may be used to find the wave observed on the downstream side of an aperture of any shape. Diffraction The interference pattern encodes the shape as a Fourier

More information

Nonlinear optics and two photon microscopy. Table of contents. Sam Whiteley and Seth Parker PHYS 173/BGGN 266 July 13, 2014

Nonlinear optics and two photon microscopy. Table of contents. Sam Whiteley and Seth Parker PHYS 173/BGGN 266 July 13, 2014 Nonlinear optics and two photon microscopy Sam Whiteley and Seth Parker PHYS 173/BGGN 266 July 13, 2014 Table of contents 1. Introduction 2. Optical setup 3. Initial images and troubleshooting 4. Determining

More information

LIFA KEY FEATURES APPLICATIONS. Fluorescence Lifetime Attachment LIFA 15001A02 16/03/2015

LIFA KEY FEATURES APPLICATIONS. Fluorescence Lifetime Attachment LIFA 15001A02 16/03/2015 LIFA Fluorescence Lifetime Attachment LIFA 151A2 16/3/215 The LIFA is a dedicated system for Fluorescence Lifetime Imaging Microscopy (FLIM). It allows the generation of lifetime images on any widefield

More information

Range Imaging Through Triangulation. Range Imaging Through Triangulation. Range Imaging Through Triangulation. Range Imaging Through Triangulation

Range Imaging Through Triangulation. Range Imaging Through Triangulation. Range Imaging Through Triangulation. Range Imaging Through Triangulation Obviously, this is a very slow process and not suitable for dynamic scenes. To speed things up, we can use a laser that projects a vertical line of light onto the scene. This laser rotates around its vertical

More information

Continuation Format Page

Continuation Format Page C.1 PET with submillimeter spatial resolution Figure 2 shows two views of the high resolution PET experimental setup used to acquire preliminary data [92]. The mechanics of the proposed system are similar

More information

ratio of the volume under the 2D MTF of a lens to the volume under the 2D MTF of a diffraction limited

ratio of the volume under the 2D MTF of a lens to the volume under the 2D MTF of a diffraction limited SUPPLEMENTARY FIGURES.9 Strehl ratio (a.u.).5 Singlet Doublet 2 Incident angle (degree) 3 Supplementary Figure. Strehl ratio of the singlet and doublet metasurface lenses. Strehl ratio is the ratio of

More information

Transmitted light Illuminator VIS-LED, color temperature. Epifluorescence lamp module X-Cite 120PC Q (Excelitas

Transmitted light Illuminator VIS-LED, color temperature. Epifluorescence lamp module X-Cite 120PC Q (Excelitas ELYRA PS.1 Multi-functional fluorescence inverted widefield microscope enabling live-cell imaging, TIRF or HILO illuminion and two super-resolution techniques: structured illuminion microscopy (SIM) and

More information

Supplementary Figure 1: Schematic of the nanorod-scattered wave along the +z. direction.

Supplementary Figure 1: Schematic of the nanorod-scattered wave along the +z. direction. Supplementary Figure 1: Schematic of the nanorod-scattered wave along the +z direction. Supplementary Figure 2: The nanorod functions as a half-wave plate. The fast axis of the waveplate is parallel to

More information

Fundamental Optics for DVD Pickups. The theory of the geometrical aberration and diffraction limits are introduced for

Fundamental Optics for DVD Pickups. The theory of the geometrical aberration and diffraction limits are introduced for Chapter Fundamental Optics for DVD Pickups.1 Introduction to basic optics The theory of the geometrical aberration and diffraction limits are introduced for estimating the focused laser beam spot of a

More information

LAGRANGIAN PARTICLE TRACKING IN ISOTROPIC TURBULENT FLOW VIA HOLOGRAPHIC AND INTENSITY BASED STEREOSCOPY. By Kamran Arjomand

LAGRANGIAN PARTICLE TRACKING IN ISOTROPIC TURBULENT FLOW VIA HOLOGRAPHIC AND INTENSITY BASED STEREOSCOPY. By Kamran Arjomand LAGRANGIAN PARTICLE TRACKING IN ISOTROPIC TURBULENT FLOW VIA HOLOGRAPHIC AND INTENSITY BASED STEREOSCOPY By Kamran Arjomand I. Background A. Holographic Imaging 1. Acquire Hologram 3. Numerical Reconstruction

More information

Lens Implementation on GATE for Optical Imaging Simulation

Lens Implementation on GATE for Optical Imaging Simulation 2017 IEEE NSS/MIC USA, Atlanta Lens Implementation on GATE for Optical Imaging Simulation Han Gyu Kang 1, Seong Hyun Song 1, Young Been Hang 1, Kyeong Min Kim 2, and Seong Jong Hong 1,3* 1.Dept. of Senior

More information

Supporting Information. Super Resolution Imaging of Nanoparticles Cellular Uptake and Trafficking

Supporting Information. Super Resolution Imaging of Nanoparticles Cellular Uptake and Trafficking Supporting Information Super Resolution Imaging of Nanoparticles Cellular Uptake and Trafficking Daan van der Zwaag 1,2, Nane Vanparijs 3, Sjors Wijnands 1,4, Riet De Rycke 5, Bruno G. De Geest 2* and

More information

Beam Profilier - Beamage 3.0

Beam Profilier - Beamage 3.0 Profilier - age 3.0 KEY FEATURES High resolution (160x120 points) 2.2 MPixels resolution gives accurate profile measurements on very small beams Large Area The 11.3 x 6.0 mm sensor allows to measure very

More information

Supporting information for: A highly directional room-temperature single. photon device

Supporting information for: A highly directional room-temperature single. photon device Supporting information for: A highly directional room-temperature single photon device Nitzan Livneh,, Moshe G. Harats,, Daniel Istrati, Hagai S. Eisenberg, and Ronen Rapaport,, Applied Physics Department,

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION DOI: 1.138/NPHOTON.234 Detection and tracking of moving objects hidden from view Genevieve Gariepy, 1 Francesco Tonolini, 1 Robert Henderson, 2 Jonathan Leach 1 and Daniele Faccio 1 1 Institute of Photonics

More information

12/7/2012. Biomolecular structure. Diffraction, X-ray crystallography, light- and electron microscopy. CD spectroscopy, mass spectrometry

12/7/2012. Biomolecular structure. Diffraction, X-ray crystallography, light- and electron microscopy. CD spectroscopy, mass spectrometry phase difference at a given distance constructive/destructive interference Biomolecular structure. Diffraction, X-ray crystallography, light- and electron microscopy. CD spectroscopy, mass spectrometry

More information

Confocal Microscopy Imaging of Single Emitter Fluorescence and Hanbury Brown, and Twiss Setup for Photon Antibunching. Abstract

Confocal Microscopy Imaging of Single Emitter Fluorescence and Hanbury Brown, and Twiss Setup for Photon Antibunching. Abstract James Maslek 10/26/12 Confocal Microscopy Imaging of Single Emitter Fluorescence and Hanbury Brown, and Twiss Setup for Photon Antibunching Abstract The purpose of this experiment was to observe fluorescence

More information

User Operation Manual

User Operation Manual User Operation Manual Manual Version: 1.3 For Single Cell Bioanalyzer (SCB-402) Doc# 102-SCB-402-1.3 Table of Contents 1. General Information... 1 1.1 How to Use This Manual... 1 1.2 Operation Safety...

More information

Cover Page. The handle holds various files of this Leiden University dissertation

Cover Page. The handle   holds various files of this Leiden University dissertation Cover Page The handle http://hdl.handle.net/1887/48877 holds various files of this Leiden University dissertation Author: Li, Y. Title: A new method to reconstruct the structure from crystal images Issue

More information

Coupling of surface roughness to the performance of computer-generated holograms

Coupling of surface roughness to the performance of computer-generated holograms Coupling of surface roughness to the performance of computer-generated holograms Ping Zhou* and Jim Burge College of Optical Sciences, University of Arizona, Tucson, Arizona 85721, USA *Corresponding author:

More information

Chapter 38. Diffraction Patterns and Polarization

Chapter 38. Diffraction Patterns and Polarization Chapter 38 Diffraction Patterns and Polarization Diffraction Light of wavelength comparable to or larger than the width of a slit spreads out in all forward directions upon passing through the slit This

More information

3. Image formation, Fourier analysis and CTF theory. Paula da Fonseca

3. Image formation, Fourier analysis and CTF theory. Paula da Fonseca 3. Image formation, Fourier analysis and CTF theory Paula da Fonseca EM course 2017 - Agenda - Overview of: Introduction to Fourier analysis o o o o Sine waves Fourier transform (simple examples of 1D

More information

E x Direction of Propagation. y B y

E x Direction of Propagation. y B y x E x Direction of Propagation k z z y B y An electromagnetic wave is a travelling wave which has time varying electric and magnetic fields which are perpendicular to each other and the direction of propagation,

More information

AP* Optics Free Response Questions

AP* Optics Free Response Questions AP* Optics Free Response Questions 1978 Q5 MIRRORS An object 6 centimeters high is placed 30 centimeters from a concave mirror of focal length 10 centimeters as shown above. (a) On the diagram above, locate

More information

A picture can say more than 1000 words comparing camera images. Jürgen Bretschneider, 2015

A picture can say more than 1000 words comparing camera images. Jürgen Bretschneider, 2015 A picture can say more than 1000 words comparing camera images Jürgen Bretschneider, 2015 Allied Vision Profile Foundation: 1989, Headquarters: Stadtroda (Germany) Employees: > 300 (2015) Know how: Development

More information

dq dt I = Irradiance or Light Intensity is Flux Φ per area A (W/m 2 ) Φ =

dq dt I = Irradiance or Light Intensity is Flux Φ per area A (W/m 2 ) Φ = Radiometry (From Intro to Optics, Pedrotti -4) Radiometry is measurement of Emag radiation (light) Consider a small spherical source Total energy radiating from the body over some time is Q total Radiant

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!

More information

Selective Optical Assembly of Highly Uniform. Nanoparticles by Doughnut-Shaped Beams

Selective Optical Assembly of Highly Uniform. Nanoparticles by Doughnut-Shaped Beams SUPPLEMENTARY INFORMATION Selective Optical Assembly of Highly Uniform Nanoparticles by Doughnut-Shaped Beams Syoji Ito 1,2,3*, Hiroaki Yamauchi 1,2, Mamoru Tamura 4,5, Shimpei Hidaka 4,5, Hironori Hattori

More information

NEW OPTICAL MEASUREMENT TECHNIQUE FOR SI WAFER SURFACE DEFECTS USING ANNULAR ILLUMINATION WITH CROSSED NICOLS

NEW OPTICAL MEASUREMENT TECHNIQUE FOR SI WAFER SURFACE DEFECTS USING ANNULAR ILLUMINATION WITH CROSSED NICOLS NEW OPTICAL MEASUREMENT TECHNIQUE FOR SI WAFER SURFACE DEFECTS USING ANNULAR ILLUMINATION WITH CROSSED NICOLS Satoru Takahashi 1, Takashi Miyoshi 1, Yasuhiro Takaya 1, and Takahiro Abe 2 1 Department of

More information

Condenser Optics for Dark Field X-Ray Microscopy

Condenser Optics for Dark Field X-Ray Microscopy Condenser Optics for Dark Field X-Ray Microscopy S. J. Pfauntsch, A. G. Michette, C. J. Buckley Centre for X-Ray Science, Department of Physics, King s College London, Strand, London WC2R 2LS, UK Abstract.

More information

specular diffuse reflection.

specular diffuse reflection. Lesson 8 Light and Optics The Nature of Light Properties of Light: Reflection Refraction Interference Diffraction Polarization Dispersion and Prisms Total Internal Reflection Huygens s Principle The Nature

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

A SUPER-RESOLUTION MICROSCOPY WITH STANDING EVANESCENT LIGHT AND IMAGE RECONSTRUCTION METHOD

A SUPER-RESOLUTION MICROSCOPY WITH STANDING EVANESCENT LIGHT AND IMAGE RECONSTRUCTION METHOD A SUPER-RESOLUTION MICROSCOPY WITH STANDING EVANESCENT LIGHT AND IMAGE RECONSTRUCTION METHOD Hiroaki Nishioka, Satoru Takahashi Kiyoshi Takamasu Department of Precision Engineering, The University of Tokyo,

More information

doi: /

doi: / Yiting Xie ; Anthony P. Reeves; Single 3D cell segmentation from optical CT microscope images. Proc. SPIE 934, Medical Imaging 214: Image Processing, 9343B (March 21, 214); doi:1.1117/12.243852. (214)

More information

MAT 17C - DISCUSSION #4, Counting Proteins

MAT 17C - DISCUSSION #4, Counting Proteins MAT 17C - DISCUSSION #4, Counting Proteins Visualization of molecules inside a living cell is one of the most important tools of molecular biology in the 21 st century. In fact, it was the subject of the

More information

Module 3: Velocity Measurement Lecture 14: Analysis of PIV data. The Lecture Contains: Flow Visualization. Test Cell Flow Quality

Module 3: Velocity Measurement Lecture 14: Analysis of PIV data. The Lecture Contains: Flow Visualization. Test Cell Flow Quality The Lecture Contains: Flow Visualization Test Cell Flow Quality Influence of End-Plates Introduction To Data Analysis Principle of Operation of PIV Various Aspects of PIV Measurements Recording of the

More information

Supplemental information. Appendix to Wavelength-scale light concentrator made by direct 3D laser writing of polymer metamaterials

Supplemental information. Appendix to Wavelength-scale light concentrator made by direct 3D laser writing of polymer metamaterials Supplemental information Appendix to Wavelength-scale light concentrator made by direct 3D laser writing of polymer metamaterials J. Moughames 1,2, S. Jradi 1, T.M. Chan 3, S. Akil 4, Y. Battie 4, A. En

More information

dq dt I = Irradiance or Light Intensity is Flux Φ per area A (W/m 2 ) Φ =

dq dt I = Irradiance or Light Intensity is Flux Φ per area A (W/m 2 ) Φ = Radiometry (From Intro to Optics, Pedrotti -4) Radiometry is measurement of Emag radiation (light) Consider a small spherical source Total energy radiating from the body over some time is Q total Radiant

More information

LIGHT SHEET MICROSCOPE. Alphα NEW GENERATION. Ground Breaking Technology For Selective Plane Illumination Microscopy

LIGHT SHEET MICROSCOPE. Alphα NEW GENERATION. Ground Breaking Technology For Selective Plane Illumination Microscopy LIGHT SHEET MICROSCOPE NEW GENERATION 3 Alphα Ground Breaking Technology For Selective Plane Illumination Microscopy Alphα 3 GET THE MOST OUT OF LIGHT SHEET IMAGING Optimal SPIM Architecture SHARP OPTICAL

More information

Measurement of droplets temperature by using a global rainbow technique with a pulse laser

Measurement of droplets temperature by using a global rainbow technique with a pulse laser , 23rd Annual Conference on Liquid Atomization and Spray Systems, Brno, Czech Republic, September 2010 Measurement of droplets temperature by using a global rainbow technique with a pulse laser S. Saengkaew,

More information

Effect of OTF attenuation on single-slice SR-SIM reconstruction

Effect of OTF attenuation on single-slice SR-SIM reconstruction Effect of OTF attenuation on single-slice SR-SIM reconstruction Supplementary Figure 1: Actin filaments in U2OS cells, labelled with Phalloidin-Atto488, measured on a DeltaVision OMX, excited at 488 nm

More information

LECTURE 37: Ray model of light and Snell's law

LECTURE 37: Ray model of light and Snell's law Lectures Page 1 Select LEARNING OBJECTIVES: LECTURE 37: Ray model of light and Snell's law Understand when the ray model of light is applicable. Be able to apply Snell's Law of Refraction to any system.

More information

STEREO VISION AND LASER STRIPERS FOR THREE-DIMENSIONAL SURFACE MEASUREMENTS

STEREO VISION AND LASER STRIPERS FOR THREE-DIMENSIONAL SURFACE MEASUREMENTS XVI CONGRESO INTERNACIONAL DE INGENIERÍA GRÁFICA STEREO VISION AND LASER STRIPERS FOR THREE-DIMENSIONAL SURFACE MEASUREMENTS BARONE, Sandro; BRUNO, Andrea University of Pisa Dipartimento di Ingegneria

More information

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube T. Donath* a, F. Pfeiffer a,b, O. Bunk a, W. Groot a, M. Bednarzik a, C. Grünzweig a, E. Hempel c, S. Popescu c, M. Hoheisel

More information

AP Physics Problems -- Waves and Light

AP Physics Problems -- Waves and Light AP Physics Problems -- Waves and Light 1. 1975-4 (Physical Optics) a. Light of a single wavelength is incident on a single slit of width w. (w is a few wavelengths.) Sketch a graph of the intensity as

More information

(Refer Slide Time: 00:10)

(Refer Slide Time: 00:10) Fundamentals of optical and scanning electron microscopy Dr. S. Sankaran Department of Metallurgical and Materials Engineering Indian Institute of Technology, Madras Module 02 Unit-4 Phase contrast, Polarized

More information