PHI VersaProbe III User Guide

Size: px
Start display at page:

Download "PHI VersaProbe III User Guide"

Transcription

1 PHI VersaProbe III User Guide Created for Stanford NanoShared Facilities (SNSF) Juliet Jamtgaard October, 2016 This document was created for training purposes only. Screenshots are from SmartSoft- VersaProbe TM by Physical Electronics, a Division of ULVAC- PHI. 2

2 PHI VersaProbe 3 User Guide Features: Greater count collection with better energy resolution. Hot/cold stage. Intro Camera for easy navigation. SmartSoft software with Queue options. FIG5 Ar+ Ion gun. GCIB Ar+ cluster gun. Collection- limiting aperture for AR- XPS. Spot sizes: 9um 200um plus HP (1400x100um) Auto Neutralize, Auto Z- align options. 3

3 PHI VersaProbe Loading Sample A. C. B. D. 1. Choose Holder A. XPS Tab B. Platen Manager tab (RHS). C. Platen tab. D. Choose holder size from dropdown list. 1 inch 2 inch AR Hot Cold E. Press Load. F. If asked to save changes to previous holder, say No. **Note** The Max Z Height of the 1 and 2 holders should be 18.5mm; 16.5mm for AR holder. If your sample is thicker than 0.5mm, subtract sample thickness from Max Z height. 4

4 PHI VersaProbe Loading Sample F. C. D. B. 2. Set up file directory A. Platen Manager A. tab (RHS). B. Directory tab. C. Choose the desired directory. D. Press Update directory button. E. Press Properties button. F. Choose Lab Book tab. G. Choose desired file name settings. E. 5

5 PHI VersaProbe Loading Sample A. Ion gauge must read 3.7x10-4 Pa before gate valve will open. 5. B. 3. Backfill Intro A. System Tab B. Right- click on Intro chamber. C. Select Backfill Intro from drop- down menu. 4. Load Sample in Intro Chamber. Line up Notch in holder with screw on transfer fork. 5. Drag- and- drop Image in Intro chamber into Main chamber. 6. Follow all prompts to take photo and load sample into main chamber. 7. Never lean on transfer arm! 6

6 PHI VersaProbe 3 Choose Position(s) Sample Tab 2. Press Define Point button A. Nearby buttons for Define Area B. Define Line 3. Add Name, Comment and update type. Choose HP setting in Type. 4. Highlight desired point/area (green) and press Move to position button. 7 To use HP setting, must choose HP in Type for each position and also in X- ray Setting on Spectrum or Profile tab before starting scan.

7 PHI VersaProbe 3 Z Align Sample Tab 2. Z- Align Tab (RHS) 3. Move to desired position on Stage (see previous page.) 4. Choose X- ray Setting for Z- Align. 5. **Check Auto settings (see pg 7 step 6). 6. Press Z- Align. 7. Z- align status box will open. Click Properties button and choose Graph tab to see plot. 8. If needed, update position to new Z- height. 8

8 PHI VersaProbe 3 Spectrum Setup XPS Tab 2. Spectrum Tab (RHS) 3. Choose Xray Setting 4. Press Periodic Table button to choose elements or SU for survey. Adjust parameters for each Region. 5. Choose # of Cycles. 6. Press XPS Setup Select/De- select Auto Neutralize settings as needed. 6. 9

9 PHI VersaProbe 3 Notes before Acquisition Before starting acquisition: Make sure only desired positions are checked on Sample tab. Make sure only desired Regions are checked (XPS tab/ Spectrum Tab (RHS)). Make sure XPS Properties dialogue has been examined Auto Neutralizer is on/off as needed. If using queue feature and moving to multiple positions, can use Auto Z- Align. Ideally, Neutralizer and Ion Gun should warm up in standby for a little while before use. Can use Queue feature to automatically perform multiple tasks. If desired, take SXI image and Define points on image. Must push camera icon to save SXI before defining points. ** Press Add to Q and then delete all previous items in Q before pressing Start.*** 10

10 PHI VersaProbe 3 Spectrum Setup (cont d) 1. Two options for start: 1. If Queue is empty, press Spe to start acquisition. 2. Use the Q: Press Add to Q and delete all unwanted lines in Queue. Press Start in Queue menu. 11

11 PHI VersaProbe 3 Hardware Setup for Sputtering Before Ar+ Ion Gun sputtering: 1. Hardware tab 2. Ion gun tab (RHS) 3. Choose desired Gun Ion Gun GCIB 4. Load sputter file (never press Save ) 5. Press Standby. Ion gun will turn on auto- matically during process. 6. Return to off when done. For GCIB sputtering: Manually open valve on GCIB. Put GCIB in blank mode, not standby. Return to off when done. Manually close valve on GCIB. 12

12 PHI VersaProbe 3 Hardware Setup for Neutralizing Before sputtering: 1. Hardware tab 2. E- neut tab (RHS) 3. Load Neutralize file (never press Save ) 4. Press Standby. Neu- tralizer will operate auto- matically during process. 5. Return to off when done. 6. Neutralizer should warm up in standby for ~15 before use. 13

13 PHI VersaProbe 3 Depth Profile Setup XPS Tab 2. Profile Tab (RHS). 3. Depth Profile. 4. Choose X- ray spot size. 5. Choose Regions, sweeps and pass energies. 6. Press Sputter Rate Table to set up depth profile

14 PHI VersaProbe 3 Depth Profile Setup Check XPS Setup for Auto Neutralizer and Auto Z- align 8. Press Profile to begin or use Queue

15 PHI VersaProbe 3 Angle Profile Setup XPS Tab 2. Profile Tab (RHS). 3. Angle Profile. 4. Choose X- ray spot size. 5. Choose Regions, sweeps and pass energies. 6. Press Angle Table to set up tilt profile

16 PHI VersaProbe 3 Angle Profile Setup Check XPS Setup for Auto Neutralizer and Auto Z- align 8. Choose Acceptance Angle = Narrow. 9. Turn beam- limited aperture into place on instrument. 10. Press Angle to begin or use Queue

17 PHI VersaProbe Unloading Sample 1. Before unloading, deselect any Auto Sputter/Auto Neutralize commands from XPS Acquisition Setup. XPS tab, bottom left corner of either Spectrum or Profile tabs. See pages 13, Turn off power to Neutralizer and Ion Gun. Hardware tab, Neutralizer or Ion Gun tabs. See pages If used Ar Ion Gun (not GCIB) for sputtering or neutralizing, Pump Intro. See next page, part D. 4. If used GCIB, close valve on gun. 5. If using AR, return Beam Acceptance Aperture to Standard Position. 18

18 PHI VersaProbe Unloading Sample 1. A. C., D. 1. System tab A. Drag- and- drop image from Main chamber into Intro chamber. B. Follow all prompts to insert transfer arm and remove sample. C. Backfill Intro Right- click on Intro. Choose Backfill. Remove sample. D. Pump Intro Right- click on Intro. Choose Pump Intro. 19

SmartSoft AES Operator s Guide

SmartSoft AES Operator s Guide SmartSoft AES Operator s Guide Part No. 702064 Rev. A Physical Electronics USA, PHI, SMART-Tool, SmartSoft, MultiPak and Watcher are trademarks of ULVAC-PHI, Inc. All other trademarks are the property

More information

Quick User Guide for XPS Analysis

Quick User Guide for XPS Analysis Quick User Guide for XPS Analysis Please refer to General User Guide and its table of contents to find details on how to execute the following tasks. When in doubt, discuss your procedure with Dr. Lozovyy

More information

SmartSoft-XPS Software Release Notes

SmartSoft-XPS Software Release Notes SmartSoft-XPS Software Release Notes Release 3.10.0 October, 2018 (Win7: 32-Bit) Release 4.1.0 October, 2018 (Win10: 64-Bit) SEE Diagnostics, Dual Source GCIB/Ar Ion Gun, SXI Image Registration Enhancements,

More information

SmartSoft-VersaProbe Software Release Notes

SmartSoft-VersaProbe Software Release Notes SmartSoft-VersaProbe Software Release Notes Release 2.10.0 October, 2018 (Win7: 32-Bit) Release 3.1.0 October, 2018 (Win10: 64-Bit) SEE Diagnostics, REELS, LEIPS, PHI-UPS Controller, SXI Image Registration

More information

Operation Procedure for the DLD Axis Ultra XPS

Operation Procedure for the DLD Axis Ultra XPS Operation Procedure for the DLD Axis Ultra XPS Version 2.0* Note This user handbook was a modification of the Kratos technician offered version Users are suggested to mount samples that can be analyzed

More information

XPS1 Automated Multi-Sample Run Procedure

XPS1 Automated Multi-Sample Run Procedure XPS1 Automated Multi-Sample Run Procedure Follow the XPS Operating Procedure to load samples into the SAC chamber. Once the samples are in the SAC chamber, the following procedure can be used to automate

More information

Lab 4: Pass the Data Streams to a Match Processor and Define a Match Rule

Lab 4: Pass the Data Streams to a Match Processor and Define a Match Rule Lab 4: Pass the Data Streams to a Match Processor and Define a Match Rule In this lab you will feed both the data records and the error records to a match processor and define a match rule. At the end

More information

JEOL CarryScope SEM Revision /07/17 Page 1 of 7. JEOL CarryScope SEM

JEOL CarryScope SEM Revision /07/17 Page 1 of 7. JEOL CarryScope SEM Page 1 of 7 JEOL CarryScope SEM The JEOL CarryScope is a compact and portable SEM that utilizes a standard tungsten filament. It is to be used for inspecting and measuring samples processed in the NRF

More information

University of Minnesota NanoFabrication Center Standard Operating Procedure

University of Minnesota NanoFabrication Center Standard Operating Procedure Equipment Name: JEOL SEM Coral Name: jeol-sem Revision Number: 4 Model: JSM-6610LV Revisionist: K. Roberts Location: Area 3 Date: 9/17/2013 1 Description The JSM-6610LV is a scanning electron microscope

More information

Spectroscopic Ellipsometer --- J. A. Woollam alpha-se

Spectroscopic Ellipsometer --- J. A. Woollam alpha-se Spectroscopic Ellipsometer --- J. A. Woollam alpha-se Introduction Figure 1: J. A. Woollam alpha-se spectroscopic ellipsometer An ellipsometer measures the change in polarization as light reflects or transmits

More information

Raith e_line Electron Beam Lithography

Raith e_line Electron Beam Lithography Raith e_line Electron Beam Lithography Standard Operating Procedure 1 (For an un-patterned sample) Revision: 7.0 Last Updated: Feb.18/2015, Revised by Mohamad Rezaei Overview This document will provide

More information

SUSS SB6E BONDER SOP. October 2013

SUSS SB6E BONDER SOP. October 2013 SUSS SB6E BONDER SOP October 2013 The SUSS SB6 bonder is capable of silicon fusion, anodic, and eutectic bonding. There are two separate heads; one for anodic bonding, and another for fusion bonding. Definitions

More information

3D Scanning Comes Full Circle

3D Scanning Comes Full Circle 3D SCANNER TM 3D Scanning Comes Full Circle Relative 3D Dosimetry offering the easiest setup, most objectivity, and best consistency available Cross-Plane Advanced Design Ring and diameter drive mechanism

More information

Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER

Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER DISCLAIMER Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 This document is intended to be a brief, bare-bones user s

More information

TraceFinder Analysis Quick Reference Guide

TraceFinder Analysis Quick Reference Guide TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in the Thermo TraceFinder 3.0 analytical software. For detailed descriptions

More information

PANalytical Epsilon 5 XRF spectral acquisition, processing and reporting

PANalytical Epsilon 5 XRF spectral acquisition, processing and reporting QSM Approval: PANalytical Epsilon 5 XRF spectral acquisition, processing and reporting 1. Introduction and Scope 1.1. This Standard Operating Procedure provides instruction for the operation, acquisition

More information

Unscrew and remove the cap from the waste tank

Unscrew and remove the cap from the waste tank Stanford Cytoflex HTS User Guide 11232015 Starting up 1. Check the sheath fluid level every time you use the cytometer. This ensures that you do not run out of sheath fluid during an experiment. Replenish

More information

2.2. Facilities Requirements

2.2. Facilities Requirements 2.2. Facilities Requirements Facilities requirements for the alpha-se system are listed in Table 2-1 and the system dimensions are given in Figure 2-1. As shown in Figure 2-2, the preferred clear work

More information

Jasco 4700 FTIR. Laser Spectroscopy Labs. Jasco 4700 FTIR. Operation instructions. Laser Spectroscopy Labs, UCI

Jasco 4700 FTIR. Laser Spectroscopy Labs. Jasco 4700 FTIR. Operation instructions. Laser Spectroscopy Labs, UCI Laser Spectroscopy Labs Jasco 4700 FTIR Operation instructions Turn the system on 1. Turn on the computer in case it has been switched off. Normally computer stays on. 2. The instrument should be always

More information

Calendar: Scheduling, invitations, attachments, and printing

Calendar: Scheduling, invitations, attachments, and printing Does your Calendar look different than what s shown here? To fix this, switch to the new look! Calendar: Scheduling, invitations, attachments, and printing Your calendar view Sign in to Google Calendar.

More information

DATA LOGGING SOFTWARE FOR 4-WAY BALL VALVE DIGITAL MANIFOLD

DATA LOGGING SOFTWARE FOR 4-WAY BALL VALVE DIGITAL MANIFOLD DATA LOGGING SOFTWARE FOR 4-WAY BALL VALVE DIGITAL MANIFOLD The Mastercool Data Logger Application provides all the necessary Recorder and Display Functions to record, display and save new data or playback

More information

for High-Performance Routine Analysis

for High-Performance Routine Analysis ESCA + 105-V01/11.04 ESCA+ for High-Performance Routine Analysis - Reliable and Fast ESCA Analysis - Ease of Use - Impressive XPS Horsepower Performance - High Energy Resolution - Quantification with Industry-

More information

Bruker. SDD X-ray Detector

Bruker. SDD X-ray Detector Operating Manual Part 2 Bruker SDD X-ray Detector Contacts Cornell Police: 255-5111 Emergency: 911 Phil Carubia: 255-6757 8.00 Step 1 - setup: Select Objects and click on MultiPoint MultiPoint Mode Set

More information

PANALYTICAL X PERT PRO XRD

PANALYTICAL X PERT PRO XRD STANDARD OPERATING PROCEDURE: PANALYTICAL X PERT PRO XRD Purpose of this Instrument: Identification of signal-phase materials and multi-phase mixture; Quantitative determination of amounts of different

More information

HIGHLIGHTS OF NEW FEATURES TOPICAL DESCRIPTION OF SOFTWARE

HIGHLIGHTS OF NEW FEATURES TOPICAL DESCRIPTION OF SOFTWARE HIGHLIGHTS OF NEW FEATURES 1. New Capture Desktop - Experiment Mode 4 2. New Capture Desk Top - Recipe Mode 5 3. Export of Raw Data 6 4. ESCA Control Panel Replacement for A Quick 8 5. Motor Control Panel

More information

SystemVue 2011 Fundamentals (version ) LAB EXERCISE 1. SystemVue Basics

SystemVue 2011 Fundamentals (version ) LAB EXERCISE 1. SystemVue Basics SystemVue 2011 Fundamentals (version 1-2012) LAB EXERCISE 1 SystemVue Basics This lab exercise introduces the basic use model of SystemVue, including using the built-in templates, running analyses and

More information

Nicolet FT-IR Procedure

Nicolet FT-IR Procedure Nicolet FT-IR Procedure Getting Started Jerry Hu (Lab Manager), ext. 7914 or 7940 jghu@mrl.ucsb.edu Jaya Nolt (Lab Tech), ext. 4997 jaya@mrl.ucsb.edu Materials Research Laboratory University of California,

More information

Forms/Distribution Acrobat X Professional. Using the Forms Wizard

Forms/Distribution Acrobat X Professional. Using the Forms Wizard Forms/Distribution Acrobat X Professional Acrobat is becoming a standard tool for people and businesses to use in order to replicate forms and have them available electronically. If a form is converted

More information

Panalytical MRD X-Ray Diffraction SOP

Panalytical MRD X-Ray Diffraction SOP Panalytical MRD X-Ray Diffraction SOP Table of Contents 1.0 Safety 2.0 Training 3.0 Sample Preparation 4.0 Pre-Operation 5.0 Sample Height Adjustment 6.0 Running Programs 7.0 Sample Unloading 8.0 Post-Operation

More information

Bruker OminiFlex MALDI-TOF Mass Spectrometer Operation Quick Start

Bruker OminiFlex MALDI-TOF Mass Spectrometer Operation Quick Start Bruker OminiFlex MALDI-TOF Mass Spectrometer Operation Quick Start Sample preparation for the Bruker OminiFlex MALDI-TOF Mass Spectrometer is of crucial importance. Please take care when preparing your

More information

GETTING STARTED TABLE OF CONTENTS

GETTING STARTED TABLE OF CONTENTS Sketchup Tutorial GETTING STARTED Sketchup is a 3D modeling program that can be used to create 3D objects in a 2D environment. Whether you plan to model for 3D printing or for other purposes, Sketchup

More information

R I T. Title: Rudolf Ellipsometer Auto-EL IV. Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date:12/07/17 1 SCOPE

R I T. Title: Rudolf Ellipsometer Auto-EL IV. Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date:12/07/17 1 SCOPE Title: Rudolf Ellipsometer Auto-EL IV Fabrication Laboratory Revision: C Rev Date:12/07/17 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the

More information

Designing Horn Antenna utilizing FEM Symmetry Boundary Conditions

Designing Horn Antenna utilizing FEM Symmetry Boundary Conditions Designing Horn Antenna utilizing FEM Symmetry Boundary Conditions If a structure has any symmetry (E or M i.e. Electric or Magnetic), the structure s physical size can be reduced symmetric plane boundary

More information

Nicolet RAMAN Procedure

Nicolet RAMAN Procedure Nicolet RAMAN Procedure Jerry Hu (Lab Manager), ext. 7914 or 7940 jghu@mrl.ucsb.edu Jaya Nolt (Lab Tech), ext. 4997 jaya@mrl.ucsb.edu Materials Research Laboratory University of California, Santa Barbara

More information

ULTRAVISION 3.7R21. Product Bulletin. UltraVision, a complete UT and Phased Array inspection package!

ULTRAVISION 3.7R21. Product Bulletin. UltraVision, a complete UT and Phased Array inspection package! ULTRAVISION 3.7R21 Product Bulletin UltraVision, a complete UT and Phased Array inspection package! www.zetec.com Table of Content Table of Content... 2 Purpose of UltraVision 3.7R21... 3 Corrosion...

More information

Preparing for EBeam Write

Preparing for EBeam Write Preparing for EBeam Write Electron beam lithography is a time-intensive process that takes many steps to properly prepare for writing. In order to make the process as efficient and simple as possible,

More information

SE800 Short Manual SENTECH Instruments GmbH

SE800 Short Manual SENTECH Instruments GmbH SE800 Spectroscopic Ellipsometry Documentation SE800 Page 2 of 16 Documentation SE800 SE 800 Page 3 of 16 Documentation SE800 Document Information: Authors: Date: File: Version: 2/5/2004 3:24:34 PM SE800_ShortManual.doc

More information

MALDI Software Users Guide (simplified version)

MALDI Software Users Guide (simplified version) MALDI Software Users Guide (simplified version) Central Analytical Lab Ekeley E266 The Acquisition program is called Voyager Control Panel. Once you shoot your sample and determine it looks good, you transfer

More information

LSRFortessa Guide. Cell Evaluation and Therapy Core HO324

LSRFortessa Guide. Cell Evaluation and Therapy Core HO324 LSRFortessa Guide Cell Evaluation and Therapy Core HO324 Before Use: Prime the instrument twice with no tube and arm to the side Ensure that sheath is full and waste is empty To Begin: Log into ilabs tracker

More information

DDFile Catcher

DDFile Catcher DDFile Catcher http://www.ddfilecatcher.com/ DDFile Catcher is a program that documents your directory structure as well as selected properties of folder and files in your directory structure. These properties

More information

Quick Start: GeoDAS-LT with Single Beam Echo Sounder

Quick Start: GeoDAS-LT with Single Beam Echo Sounder Quick Start: GeoDAS-LT with Single Beam Echo Sounder I. Setting up GeoDAS-LT for Single Beam Echo Sounder Survey 1. Connect your single beam echo sounder, GPS receiver and other sensors (e.g. gyro and

More information

Equipment Standard Operating Procedure Kimberly Appel

Equipment Standard Operating Procedure Kimberly Appel Date Created: June 26, 2004 Date Modified: The Flexus 2320 Equipment Standard Operating Procedure Kimberly Appel 1. Purpose 1.1. The Flexus Thin Film Stress Measuring Apparatus (TFSMA) measures the changes

More information

Lab 8-1 Lab Multiple Form Types - Setup

Lab 8-1 Lab Multiple Form Types - Setup In the following lab instructions, you are going to add a second form type and learn to set up automatic form identification. In addition, you ll be introduced to registration zones and index group zones.

More information

Steps for LCMS-8040 Triple Quadrupole

Steps for LCMS-8040 Triple Quadrupole Steps for LCMS-8040 Triple Quadrupole June 2017 Version 2 A. Data Acquisition 1. Check lights on MS. There should be a total of three lights lit by LED. One of them is blinking. Power Blinking Gas Status

More information

Collect and Reduce Intensity Data -- APEX

Collect and Reduce Intensity Data -- APEX Collect and Reduce Intensity Data -- APEX General Steps in Collecting Intensity Data Note that the steps outlined below are generally followed when using all modern automated diffractometers, regardless

More information

Welcome To The Paragon TM. 3.4 CMA Guide

Welcome To The Paragon TM. 3.4 CMA Guide Show Welcome To The Paragon TM 3.4 CMA Guide The following tutorial will assist you in performing the steps necessary in creating a CMA Packet in Paragon. For Additional information, login to Paragon and

More information

ClonePix TM 2 QUICK SET-UP INTRUCTIONS MANUAL SOFTWARE RELEASE

ClonePix TM 2 QUICK SET-UP INTRUCTIONS MANUAL SOFTWARE RELEASE ClonePix TM 2 QUICK SET-UP INTRUCTIONS MANUAL SOFTWARE RELEASE 1.2.80.1071 07MAN1181.A1 Effective Date: 30-Jan-11 ECO #: 3093 Contents What are Quick Set-Up Instructions?... 3 Starting up ClonePix 2...

More information

Perkin-Elmer Lambda 3B Spectrophotometer

Perkin-Elmer Lambda 3B Spectrophotometer Written by Mark Csele Written by Jeff Brown Operation Turn on the Lambda 3B and the Computer Remove all samples from the sample chamber. Turn on the lamp for the range being investigated using the switches

More information

Quick reference guide

Quick reference guide EN Quick reference guide You can find legal and warranty information on the CD. Further information is available at HP s support Web site (http://www.hp.com/go/knowledge_center/). Legal notices The information

More information

The Dektak XT is a 2D contact profilometer used for step height, pitch and surface roughness

The Dektak XT is a 2D contact profilometer used for step height, pitch and surface roughness Dektak XT 2D Profilometer Operation Manual The Dektak XT is a 2D contact profilometer used for step height, pitch and surface roughness measurements. Vision 64 application software controls the system

More information

Primary Use. Operating Principle

Primary Use. Operating Principle Primary Use The Leica DVM6 is an optical microscope that has the ability observe samples at a high magnification at a high resolution. The microscope allows users to view their sample with up to a 2350x

More information

NYU A&S AEM Implementation. Exercise Sheets

NYU A&S AEM Implementation. Exercise Sheets NYU A&S AEM Implementation Exercise Sheets Table of Contents Table of Contents... 2 Exercise 1: Blank Page with Title... 1 A. Create page using template Blank Page with Title... 1 B. Author component Generic

More information

Prism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis

Prism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis Start Up: Upon entering the laser room turn on the wall mounted Laser Power Button by pulling it away from the wall. Turn on Shutter controllers (toggle switch on back of unit). There should be a U in

More information

1

1 In the following tutorial we will determine by fitting the standard instrumental broadening supposing that the LaB 6 NIST powder sample broadening is negligible. This can be achieved in the MAUD program

More information

Physics 101, Lab 1: LINEAR KINEMATICS PREDICTION SHEET

Physics 101, Lab 1: LINEAR KINEMATICS PREDICTION SHEET Physics 101, Lab 1: LINEAR KINEMATICS PREDICTION SHEET After reading through the Introduction, Purpose and Principles sections of the lab manual (and skimming through the procedures), answer the following

More information

University of Minnesota Nano Fabrication Center Standard Operating Procedure

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name: University of Minnesota Nano Fabrication Center Coral Name: hs-scope Revision Number: 1.5 Model: HS200A Revisionist: M. Fisher Location: Bay 1 Date: 9/12/2013 1 Description The Hyphenated

More information

Apple idvd 11 Tutorial

Apple idvd 11 Tutorial Apple idvd 11 Tutorial GETTING STARTED idvd is a program that allows you to create a DVD with menus and graphics of a professionally made commercial disc to play on your home DVD player. To Begin your

More information

Epilog Laser Galvo Model G2 Operating Instructions

Epilog Laser Galvo Model G2 Operating Instructions Epilog Laser Galvo Model G2 Operating Instructions The Epilog G2 galvo system is a three-axis galvanometer comprised of two galvanometers (motors), a translating lens, and a fixed focus length lens. This

More information

P6 Profilometer Operating Procedure

P6 Profilometer Operating Procedure P6 Profilometer Operating Procedure I. Introduction The KLA-Tencor P-6 profiler is a highly sensitive surface profiler that measures step height. The key features of P-6 include: A vertical range of 326

More information

Quality Control (QC) Online in SobekCM

Quality Control (QC) Online in SobekCM Page 1 of 10 Quality Control (QC) Online in SobekCM 1) How to open the Quality Control tool Once the item is opened on the webpage, click on the Quality Control icon in the grey bar which shows up. (Note

More information

Quick User Guide SURVEYTIME. User Manual Page 1 INTRODUCTION TO SURVEYTIME

Quick User Guide SURVEYTIME. User Manual Page 1 INTRODUCTION TO SURVEYTIME User Manual Page 1 SURVEYTIME Quick User Guide 2015 User Manual Page 2 SURVEY SELECTION... 3 CREATING A CROSSTAB REPORT... 4 Sorting your columns:... 10 SINGLE-SURVEY CHART... 11 CREATING A TREND REPORT...

More information

SolidWorks Intro Part 1b

SolidWorks Intro Part 1b SolidWorks Intro Part 1b Dave Touretzky and Susan Finger 1. Create a new part We ll create a CAD model of the 2 ½ D key fob below to make on the laser cutter. Select File New Templates IPSpart If the SolidWorks

More information

ION BEAM MILLING SYSTEM FOR TEM, SEM AND LM PREPARATION. Leica EM RES102

ION BEAM MILLING SYSTEM FOR TEM, SEM AND LM PREPARATION. Leica EM RES102 ION BEAM MILLING SYSTEM FOR TEM, SEM AND LM PREPARATION Leica EM RES102 ION BEAM MILLING In recent years, ion milling has been developed into the most applicable method of sample preparation for the analysis

More information

Operating Procedure for Horiba Raman Microscope

Operating Procedure for Horiba Raman Microscope Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch

More information

GAZIANTEP UNIVERSITY INFORMATICS SECTION SEMETER

GAZIANTEP UNIVERSITY INFORMATICS SECTION SEMETER GAZIANTEP UNIVERSITY INFORMATICS SECTION 2010-2011-2 SEMETER Microsoft Excel is located in the Microsoft Office paket. in brief Excel is spreadsheet, accounting and graphics program. WHAT CAN WE DO WITH

More information

Warrior Control Panel

Warrior Control Panel Warrior Control Panel The Control Panel sets certain global characteristics of the Warrior system. The Warrior Control Panel is started from its icon in the Warrior program group or via the Windows Program

More information

Version Beta, pre-release. zspace Studio Demonstration Script

Version Beta, pre-release. zspace Studio Demonstration Script zspace Studio Demonstration Script Version 1.0-2014 Beta, pre-release zspace.com zspace Studio Demonstration Script zspace is a registered trademark of zspace, Inc. All other trademarks are the property

More information

Vanderbilt Small Molecule NMR Center

Vanderbilt Small Molecule NMR Center Vanderbilt Small Molecule NMR Center Instructions for Using TOPSPIN version 2.0 DRX 400: Setting up 1H, 13C (1H decoupled), and 31P NMR Experiments Getting started: Sample Preparation: Make sure your sample

More information

4.2 NORMAL OPERATION NORMAL START UP PROCEDURE

4.2 NORMAL OPERATION NORMAL START UP PROCEDURE 4.2 NORMAL OPERATION 4.2.1 NORMAL START UP PROCEDURE Kimball Physics recommends that the gun be run only in a vacuum of 10-5 torr or better for an alkali metal ion source. Poorer vacuum may result in source

More information

3G/WiFi In-Vehicle Surveillance & GPS Tracking System

3G/WiFi In-Vehicle Surveillance & GPS Tracking System Model: MDVR-J200 3G/WiFi In-Vehicle Surveillance & GPS Tracking System Quick Start Guide Thank you for purchasing a Securview Vehicle Surveillance & Tracking System This Quick Start Guide covers basic

More information

Chapter 2 Surfer Tutorial

Chapter 2 Surfer Tutorial Chapter 2 Surfer Tutorial Overview This tutorial introduces you to some of Surfer s features and shows you the steps to take to produce maps. In addition, the tutorial will help previous Surfer users learn

More information

AUTO ATTENDANT. Spectrum Business Voice. Auto Attendant. Quick Start Guide SpectrumBusiness.net

AUTO ATTENDANT. Spectrum Business Voice. Auto Attendant. Quick Start Guide SpectrumBusiness.net Spectrum Business Voice Auto Attendant 1 1-800-314-7195 SpectrumBusiness.net Quick Start Guide CONTENTS Overview 3 Accessing the Easy Attendant 4 Managing Your Auto Attendant 5 Setting Up an Auto Attendant

More information

Multiviewer Stream Media Player User Manual

Multiviewer Stream Media Player User Manual Multiviewer Stream Media Player User Manual Contents Chapter One: Installation of Multiviewer... 2 Chapter Two: Multiviewer stream media player... 3 Chapter Three: Instruction of MultiVewer interface...

More information

Agilent VnmrJ 3.2. Quick Start Guide

Agilent VnmrJ 3.2. Quick Start Guide Agilent VnmrJ 3.2 Quick Start Guide VnmrJ 3.2 Interface 3 Overview 4 Prepare Sample 5 Load Sample 6 Enter Sample Information 7 Build Study Queue 8 Run Study Queue 8 Process Data 9 Plot Data 10 This Quick

More information

Model FP-6500 Spectrofluorometer Instruction Manual. FP-6500 for Windows

Model FP-6500 Spectrofluorometer Instruction Manual. FP-6500 for Windows Model FP-6500 Spectrofluorometer Instruction Manual FP-6500 for Windows P/N: 0302-9999 April 2000 Contents Safety Considerations...i Regulatory Statements... iii Preface... iv Installation Conditions...v

More information

Android Mobile Client. User Guide. Version 2.9.1

Android Mobile Client. User Guide. Version 2.9.1 Android Mobile Client. User Guide Version 2.9.1 1. General information about the Android mobile client..................... 3 2. Connecting to the server and working with servers in the Android client........

More information

MIOPS Splash User Manual

MIOPS Splash User Manual MIOPS Splash User Manual Firmware Version: 1.0 App Version: 2.6 Document Version: 1.0.0 Release Date: January 21st, 2018 1. Introduction... 4 2. Specifications... 5 3. Smartphone App... 6 3.1. Using the

More information

Creating an Outlook Auto-Response Rule for Departmental Accounts

Creating an Outlook Auto-Response Rule for Departmental Accounts Ready Reference 120408 Creating an Outlook Auto-Response Rule for Departmental Accounts Introduction This document contains instructions on how to create a rule from a departmental account in order to

More information

ClonePix TM FL Quick Set-Up Instructions

ClonePix TM FL Quick Set-Up Instructions ClonePix TM FL Quick Set-Up Instructions Software Release: 1.2.13.971 07MAN1055.A2 Effective Date: 01-Jun-10 QUICK INSTRUCTIONS > CONTENTS Contents What are Quick Set Up Instructions?...3 Starting up ClonePix

More information

ME 365 EXPERIMENT 3 INTRODUCTION TO LABVIEW

ME 365 EXPERIMENT 3 INTRODUCTION TO LABVIEW ME 365 EXPERIMENT 3 INTRODUCTION TO LABVIEW Objectives: The goal of this exercise is to introduce the Laboratory Virtual Instrument Engineering Workbench, or LabVIEW software. LabVIEW is the primary software

More information

MLA100 Maskless Aligner

MLA100 Maskless Aligner Quick Guide MLA100 Maskless Aligner Doc. No.: DWL-HI-060 Revision: 5 (August 2017) Wizard version: 1.9 Copyright 2017 by Heidelberg Instruments Job Setup As mentioned before (Wizard Description), the MLA100

More information

Fig. A. Fig. B. Fig. 1. Fig. 2. Fig. 3 Fig. 4

Fig. A. Fig. B. Fig. 1. Fig. 2. Fig. 3 Fig. 4 Create A Spinning Logo Tutorial. Bob Taylor 2009 To do this you will need two programs from Xara: Xara Xtreme (or Xtreme Pro) and Xara 3D They are available from: http://www.xara.com. Xtreme is available

More information

Vividia HM-250 HDMI/LCD/USB/TV Portable. Digital Measurement Microscope

Vividia HM-250 HDMI/LCD/USB/TV Portable. Digital Measurement Microscope Vividia HM-250 HDMI/LCD/USB/TV Portable Digital Measurement Microscope Table of Contents 1. Precaution. 3 2. Preparation..5 Package Contents.. 6 Part Names & Functions..6 3. Specifications.... 7 4. Use

More information

SMPK4230 User Manual

SMPK4230 User Manual SMPK4230 User Manual Sylvania is a registered trademark of OSRAM Sylvania Inc. Used under License What s in the box? Operating Unit User Manual Quick Start Guide USB Cable Earphone Battery Charger 2 Content

More information

TLS Data Processing and Exploration Manual

TLS Data Processing and Exploration Manual Katherine Shervais (UNAVCO), Marin Clark (University of Michigan) Collecting data in the field is only the first step in the complete TLS workflow. This manual will take you through the skills needed to

More information

Chapter 1 General Information

Chapter 1 General Information Chapter 1 General Information System Description Chapter 1 General Information 1.0 System Description The Affirm attaches to the Selenia Dimensions. A biopsy device attaches to the Affirm. X- and Y-axes

More information

Improvement of the correlative AFM and ToF-SIMS approach using an empirical sputter model for 3D chemical characterization

Improvement of the correlative AFM and ToF-SIMS approach using an empirical sputter model for 3D chemical characterization Improvement of the correlative AFM and ToF-SIMS approach using an empirical sputter model for 3D chemical characterization T. Terlier 1, J. Lee 1, K. Lee 2, and Y. Lee 1 * 1 Advanced Analysis Center, Korea

More information

Scheduling Book Set-Up & Check-In Preferences

Scheduling Book Set-Up & Check-In Preferences Clinical Informatics Scheduling Book Set-Up & Check-In Preferences 10/23/2018 Clinical Informatics 1 Scheduling Book Settings It is important that ALL settings in the Scheduling Book are set up correctly,

More information

home page b c d e f g h

home page b c d e f g h a b c d e f g h home page i a Home Clicking globaldesign anywhere in Lobo will take you back to the main Lobo landing page b View Projects Clicking here gives you access to projects c Search Bar Quickly

More information

Determination of Drag Coefficient

Determination of Drag Coefficient DEPARTMENT OF MECHANICAL, INDUSTRIAL AND MANUFACTURING ENGINEERING MIMU 505 - MEASUREMENT AND ANALYSIS Determination of Drag Coefficient You will need to bring a zip disk or USB storage device to the lab

More information

Editing Drupal 7 Sites. Basic page editing. June To begin editing, log in by clicking Login at the bottom of the homepage. Username.

Editing Drupal 7 Sites. Basic page editing. June To begin editing, log in by clicking Login at the bottom of the homepage. Username. Editing Drupal 7 Sites June 2013 Basic page editing To begin editing, log in by clicking Login at the bottom of the homepage Username Password Navigate to the page you want to edit Click the Edit tab at

More information

Using Graphics. Digital Camera. Auto Shapes

Using Graphics. Digital Camera. Auto Shapes AutoShape Using Graphics Internet The following graphic elements are available to enhance your presentation Clip Art AutoShapes Fill effects Shadow effects 3D effects Digital Camera WordArt Digital Camera

More information

ETT Olfactometer C. Direct Control Software Manual

ETT Olfactometer C. Direct Control Software Manual 1 ETT Olfactometer C Direct Control Software Manual 2 Table of Contents Part 1: Installation 3 Download 3 Installation Process 5 Part 2: Operation 10 Startup 10 Manual Mode Paradigm Designer 13 Hierarchy

More information

5 Axis Cutting Using Delcam Powermill Written by: John Eberhart DM Lab Tutorial

5 Axis Cutting Using Delcam Powermill Written by: John Eberhart DM Lab Tutorial 5 Axis Cutting Using Delcam Powermill Written by: John Eberhart DM Lab Tutorial This tutorial covers how to setup a job for a Multi-Axis Toolpath specifi cally using the Robot. Note: You need to follow

More information

IGSS 13 Configuration Workshop - Exercises

IGSS 13 Configuration Workshop - Exercises IGSS 13 Configuration Workshop - Exercises Contents IGSS 13 Configuration Workshop - Exercises... 1 Exercise 1: Working as an Operator in IGSS... 2 Exercise 2: Creating a new IGSS Project... 28 Exercise

More information

*413323* CTI Toolbar 5.4a. Quick Reference Guide. Accessing Training Videos. Opening CTI Toolbar

*413323* CTI Toolbar 5.4a. Quick Reference Guide. Accessing Training Videos. Opening CTI Toolbar Accessing Training Videos Opening CTI Toolbar Tip: If you share a computer, a CTI Toolbar icon can be set up for each user so that individual preferences and phone lines can be customized. CTI Toolbar

More information

Findmyshift - Getting started with Findmyshift

Findmyshift - Getting started with Findmyshift Findmyshift - Getting started with Findmyshift Managers Creating your first roster Adding staff to your team Inviting your staff to log in Choosing your settings Entering shifts Publishing shifts Handling

More information

Quick Start Guide to using Light Converse along with Pangolin LD2000 and BEYOND

Quick Start Guide to using Light Converse along with Pangolin LD2000 and BEYOND Quick Start Guide to using Light Converse along with Pangolin LD2000 and BEYOND First Steps Regardless of when or from whom you purchased Light Converse, we recommend you do the following steps before

More information

Visual Physics - Introductory Lab Lab 0

Visual Physics - Introductory Lab Lab 0 Your Introductory Lab will guide you through the steps necessary to utilize state-of-the-art technology to acquire and graph data of mechanics experiments. Throughout Visual Physics, you will be using

More information

Typical System Construction Schedule; Balzers Chambers: BAK600 BAK760 QLC800

Typical System Construction Schedule; Balzers Chambers: BAK600 BAK760 QLC800 Typical System Construction Schedule; Balzers Chambers: BAK600 BAK760 QLC800 Chamber / Framework System fully dismantled. Old controls and instruments removed. Chamber and framework stripped and prepared

More information