Maxum Workstation Update Release Notes
|
|
- Amberlynn George
- 5 years ago
- Views:
Transcription
1 1 August 9, 2017 Maxum Workstation Update Release Notes Requirements Maxum Workstation Software will operate best on computers equipped with a multi-core processor, with at least 4 GB RAM (8+ GB recommended). This installation needs up to 2 GB of free disk space. You must log in with Administrator privileges. This will ensure unrestricted access to the system registry. Otherwise, the installation will fail. Internet Explorer 8.0 or higher must be installed. Maxum Workstation Software has been tested successfully with the following operating systems: Windows 7 Professional SP1 (32 bit and 64 bit) Windows 8.1 Pro (64 bit) Windows 10 Enterprise (64 bit) Windows Server 2008 Standard SP2 (32 bit and 64 bit) Windows Server 2012 R2 Standard (64 bit) Windows Server 2016 Standard (64 bit) Patch Installation Instructions: 1) Download the zip file and Readme files to your workstation 2) Unzip the entire contents of the zip file (.exe file and two {XXX} folders) 3) Run the.exe file to upgrade your Workstation and then upgrade Maxum software to each GC as desired. Note: Make sure in step two that the entire contents of the zip file are unzipped to the same folder. All files are required for the patch upgrade. Description of Changes GCP Workstation: 1) Problem: Incorrect text is displayed for the MPa pressure unit selection in the options window used to select display units. Solution: The displayed text has been corrected.
2 2 2) Problem: When selecting any calibrated group in GCP and then selecting "Show calibration", an exception is created and the calibration is not shown as long as the GCP instrument remains open. Symptom: If viewing a newly added group or existing group that has not been calibrated or is using a surrogate peak, then the "Show Calibration works if it is the first peak selected after opening a connection with GCP. As long as the instrument remains open, all calibrated groups work fine for "Show Calibration". The workaround prior to patch 1 was to add a dummy peak or select one that was not calibrated first after opening an instrument connection. Solution: The logic for drawing the information on the screen was modified to make sure that the blank chart was ready before data was written to it. 3) Problem: GCP crashes if you access the result configuration table after saving the column settings. Solution: The logic was modified to correctly save the column settings in the result configuration table to allow GCP to open it successfully during subsequent sessions with GCP. 4) Problem: Exception when linking appai to a limit in GCP. Symptom: When using GCP to link an appai to a limit and save the change, a database exception for invalid foreign key for the appai table is thrown. Solution: The condition throwing the exception was corrected. 5) Problem: Live Chromatogram Help needs to be updated to clearly state that the feature requires Maxum 5.3 software in the GC to work. Solution: GCP Help files were updated appropriately. 6) Problem: Restore Database does not work to a MicroSam. Solution: The restore database software in GCP was modified to provide a reset command after the files had been restored so that the MicroSam would restart automatically. 7) Problem: Text entry in the Peak table is not working properly if "enter" is pressed on a text field in the last row of the peak table. Solution: A correction was made to allow the last row to function correctly with the enter key. 8) Problem: APC (Advance Personal Computer) entries appear incorrectly in GCN as Gateways to Advance Data Highway. Solution: APC entries will be blocked from appearing in the GCN list of network devices. 9) Enhancement: Manual Response Factor column in the Peak Table should be visible by default. Peak Table default columns changed to include Manual Response Factor column.
3 3 10) Problem: Reload From MBM file option is only available after a method is saved. Solution: The software was modified to allow MBM files to be reloaded. 11) Problem: Fahrenheit calculation of temperature ramp rates is not calculated correctly. Symptom: The display and processing of the ramp is incorrect when Fahrenheit is used for Programmed Temperature Applications. Solution: The temperature ramp calculation was corrected when Fahrenheit units are specified. 12) Enhancement: Simulated Distillation result IDs are limited to 100 for renumbering. The limit was increased to 900 for renumbering SimDis IDs. 13) Problem: Method exports are sometimes much slower using 5.3. Solution: Method exports were optimized to improve export performance. 14) Problem: In Network-Serial Settings of GCP the communication standard setting for Port 1 will not show correctly if "Default" is chosen. Solution: GCP was modified to show the correct communication settings when default was selected. 15) Problem: The GCP User Manual released with the 5.30 DVD is an older version. Solution: The latest GCP User Manual from January 2017 will be installed with this patch. 16) Problem: The numeric entries are not always taken correctly when editing methods. In some cases numbers are very slow to display after pressing a numeric key. In other cases the entries are not entered correctly.. Solution: Software was modified to give better performance and more reliable data entry. 17) Enhancement: Change the default temperature setpoint when a new temperature controller is defined. Current is 100 degc, Desired Change is to 60 degc. The pressure controller default was also changed from 100 kpa to 60 kpa. 18) Problem: SimDis peak properties are not updating on the Calibration and Response Factor Correction page. Solution: Software modified to see properties as they change. 19) Problem: Printed chromatograms do not always show the correct source if a chromatogram is printed from file and then later from an analyzer. Solution: Modification was made to pick up the current source correctly. 20) Enhancement: Modify chromatogram overlay with tool buttons for more overlay features (e.g. batch overlay from different detectors). 21) Enhancement: Add the choice to select a chromatogram overlay directly from an analyzer as well as from a file.
4 4 22) Problem: ChromViewer crashes when it's associated ChromChart is closed if the dialog window is still open. Solution: Close OverlayPropertyDlg when its associated ChromChart is closed to prevent ChromViewer crash. 23) Problem: Cycle time in the overlay selection dialog should be in local time instead of GMT time.. Solution: Time shown changed to local time. 24) Enhancement: Add numerical fill down context menu to numeric columns of a method tables (the menu becomes available after selecting at least two contiguous cells in a column). Database: 1) Problem: Serial port configuration changes in the Serial Settings menu disables the Enable Modbus TCP checkbox. Solution: Software was modified to insure that the state of the Enable Modbus TCP checkbox is retained during serial modifications. 2) Problem: A nuisance debug message is filling up log files: setting module to error. Solution: This debug message is now turned off when not in debug mode. 3) Enhancement: A warning alarm has been added to let customers know that the Maxum calendar clock is invalid. (A typical source for this warning would be loss of power with a low battery on Syscon that needs replaced.). HMI: 1) Problem: Real-Time Chromatogram cannot be viewed on the local HMI. Symptom: After selecting View Chrom, after a few seconds an hour glass is displayed and the display reverts back to the Maintenance home screen. Solution: The problem was corrected by building with the correct version of Chart.dll. 2) Problem: Realtime Chrom on the Workstation and Local HMI will not allow autoscale to be disabled.
5 5 Solution: The code was modified to allow autoscaling to be turned off. Stream Logger : 1) Enhancement: Stream Logger should be hardened to run continuously through a wider set of conditions long-term. Software was modified to continue operation throughout a wider set of conditions.
Item 1: Microsoft.Net Framework generates an exception when viewing Group calibration information. Known Issues: Gas Chromatograph Portal 5.30.
The following information covers issues and known work-arounds discovered that can be helpful in working with Maxum Gas Chromatograph Portal version 5.30.00. Known issues that have been discovered are
More informationDescription: Update for Gas Chromatograph Portal
1 August 20, 2013 Maxum Workstation 5.10.10 Update Release Notes Requirements Maxum System Manager will operate best on computers equipped with a Pentium-166 or faster CPU, using at least 64 MB RAM. For
More informationTechnical Support Information No. 202c September 2013
Customer Focus Page 1 of 14 Release of Maxum Workstation Software Patch Versions 5.10.02 and 5.10.10 Earlier this year as decision was made to reintroduce the Workstation Patch program providing users
More informationThe patch (Maxum Workstation Patch for Windows 7 Operating System) will load operating system libraries that Microsoft did not include with Windows
1 August 20, 2013 Maxum Workstation 5.00.10 Update Release Notes Requirements Maxum System Manager will operate best on computers equipped with a Pentium-166 or faster CPU, using at least 64 MB RAM. For
More informationSummary of 4.10c Patch 3 Changes
MAXUM 4.10c Patch 3 Read the Summary of 4.1c Patch 3 section to determine if you need Patch 3. The Maxum 4.1c Patch 3 Upgrade Procedure section describes how to upgrade your workstation and individual
More information1. Fixed problems with lost communication between the Syscon and the SNE that was causing SNE resets
MAXUM 3.2b Patch 3 Read the Summary of 3.2b Patch 3 section to determine if you need Patch 3. The Maxum 3.2b Patch 3 Upgrade Procedure section describes how to upgrade your workstation and individual Maxum
More informationUsing SIMDIS Expert 9 (Separation Systems) with Clarity
Using SIMDIS Expert 9 (Separation Systems) with Clarity The SIMDIS Expert 9 may be used with Clarity or Clarity Lite software to process the acquired chromatograms according to Simulated Distillation.
More informationProcedure to Replace a TCD DPM with an IS TCD DPM in a Maxum or Maxum II Analyzer
Procedure to Replace a TCD DPM with an IS TCD DPM in a Maxum or Maxum II Analyzer Difficulty Level: Medium Estimated time to execute: 1 Hour Revision History Issue Date Reason 001 1/31/2017 Initial Issue
More informationMP8000 SOFTWARE Instruction Manual
Tel: +1-800-832-3873 E-mail: techline@littelfuse.com www.littelfuse.com/mp8000 MP8000 SOFTWARE Instruction Manual REVISION 0-B-030818 Copyright 2018 Littelfuse. All rights reserved. Page A All brand or
More informationMaxum OPC Server /Client
Maxum OPC Server /Client New Features / Changes document 3.00.02 February 4, 2019 Requirements Maxum Workstation Software will operate best on computers equipped with a multi-core processor, with at least
More informationFor Research Use Only. Not for use in diagnostic procedures.
Release Notes Agilent Feature Extraction 12.1 (12.1.0.3) Product Numbers # G2566AA - Feature Extraction provided with the Agilent Scanner # G4460AA - G4462AA - Feature Extraction Commercial License # G4463AA
More informationSOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.30: PSS
SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.30: PSS0051-04 Introduction This document details the installation and known issues with OMNISEC v10.30. This release is a major release to fix a
More information9936A LogWare III. User s Guide. Revision
9936A LogWare III User s Guide Revision 680301 Table of Contents 1 Introduction... 1 1.1 Symbols Used...1 1.2 Conventions...1 1.3 What is LogWare?...1 1.4 License Agreement...2 1.5 Requirements...2 1.6
More informationCurrent Version: Ver.Type Ver No. Date Firmware September 2016 Software Web Hardware September 2016 (Ver.2) Reference: 11272
Product Name: VIA Campus Current Version: Ver.Type Ver No. Date Firmware 2.0.0916.788 September 2016 Software Web Hardware RN Date: September 2016 (Ver.2) Reference: 11272 Kramer Product Affected: Product
More informationFor additional information, see these documents on the Breeze 2 software DVD:
R E L E A S E N O T E S Waters Breeze 2 Read these release notes before you install and operate Breeze 2. They contain information necessary to ensure successful liquid chromatography results, including:
More informationxponent Version 4.2 RELEASE NOTES
xponent Version 4.2 RELEASE NOTES Copyright Luminex Corporation, 1996-2013. All rights reserved. Part Number: 89-30000-00-462 Disclaimer Subject to change without notice. Although prepared to ensure accuracy,
More informationTechnical Support Information No. 108c January 2013
Customer Focus Page 1 of 17 Maxum/MicroSAM Modbus Transmission Summary In this documentation is a step by step procedure on how to successfully setup the Modbus transmission for the Maxum and/or MicroSAM
More informationThis is a known issue (SVA-700) that will be resolved in a future release IMPORTANT NOTE CONCERNING A VBASE RESTORE ISSUE
SureView Analytics 6.1.1 Release Notes ================================= --------- IMPORTANT NOTE REGARDING DOCUMENTATION --------- The Installation guides, Quick Start Guide, and Help for this release
More informationAvigilon Control Center 5.10 Release Notes
Version 5.10.2.2 Released Tuesday October 12 th, 2016 Release Summary This is a service release of Avigilon Control Center 5.10. Users experiencing issues listed in the Issues Fixed list should consider
More informationPulse SCADA/HMI. Version 4.11 Release Notes
Pulse SCADA/HMI Version 4.11 Release Notes October 2017 Copyright The information in this document is subject to change without prior notice and does not represent a commitment on the part of AFCON Control
More informationWaters Empower Software A1100 System Configuration Instructions for Ethernet Control
Waters Empower Software A1100 System Configuration Instructions for Ethernet Control In order to configure an A1100 system for Ethernet control or to acquire A1100 DAD 3D data in Empower Software, it is
More informationProcedure to Install a CIM Door Retrofit in an Existing Airbath/Airless Maxum (SYSCON1/SYSCON+ with 5.10 or 5.20 Software)
to Install a CIM Door Retrofit in an Existing Airbath/Airless Maxum (SYSCON1/SYSCON+ with 5.10 or 5.20 Software) Difficulty Level: Medium Estimated time to execute: 2 Hours Revision History Issue Date
More information9936A LogWare III. User s Guide. Revision
9936A LogWare III User s Guide Revision 850701 Table of Contents 1 Introduction...1 1.1 Symbols Used... 1 1.2 Conventions... 1 1.3 What is LogWare?... 1 1.4 License Agreement... 2 1.5 Requirements...
More informationSOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.31: PSS
SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.31: PSS0051-05 Introduction This document details the installation and known issues with OMNISEC v10.31. This release is a maintenance release to
More informationHigh Motor Current Shutdown High Motor Current Warning Low Motor Current Shutdown Compressor Auxiliary Shutdown
Important: To upgrade from Quantum HD version 10.21 or earlier to Quantum HD Unity version 10.30 or later requires the flashcard to be changed. These are available through the Baltimore Parts Center, part
More informationProcedure to Install a CIM Door Retrofit in an Existing Airbath/Airless Maxum II (with SYSCON1/SYSCON+)
to Install a CIM Door Retrofit in an Existing Airbath/Airless Maxum II (with SYSCON1/SYSCON+) Difficulty Level: Medium Estimated time to execute: 2 Hours Revision History Issue Date Reason 00A 8/18/12
More informationCopyright ANRITSU. All rights reserved.
General Caution --------------- Please use a USB Memory Stick for firmware updates. If there is a firmware update question, please contact Anritsu service support at https://techsupport.anritsu.com, and
More information3M Molecular Detection System Software Upgrade/Installation Instructions
User Manual Supplement Number: TB.342837.03 Effective Date: March 2018 Supersedes: TB.342837.02 Technology Platform: 3M Molecular Detection System Originating Location: St. Paul, MN 3M Molecular Detection
More informationProcedure to Upgrade from B&W Maintenance Panel to TIB Color Touchscreen Door in a Working Maxum or Maxum II Analyzer Equipped With a SYSCON2
Procedure to Upgrade from B&W Maintenance Panel to TIB Color Touchscreen Door in a Working Maxum or Maxum II Analyzer Equipped With a SYSCON2 Difficulty Level: High Estimated time to execute: 3 Hours Revision
More informationRelease Notes Life Technologies Attune NxT Software v2.3
Release Notes Life Technologies Attune NxT Software v2.3 In the following pages you will find instructions describing: New software features Known software/system issues with troubleshooting guidance Software
More informationVirtual Appliance Installation Guide
Virtual Appliance Installation Guide 2017 RapidFire Tools, Inc. All rights reserved. V20171027 Ver 2M Contents Purpose of this Guide...3 System Requirements...3 Hyper-V installation System Requirements...3
More informationNOTICE: This document contains references to Varian. Please note that Varian, Inc. is now part of Agilent Technologies. For more information, go to
NOTICE: This document contains references to Varian. Please note that Varian, Inc. is now part of Agilent Technologies. For more information, go to www.agilent.com/chem. Varian MS Workstation 6.9 Service
More informationCA ERwin Data Modeler
CA ERwin Data Modeler Installation Guide Release 9.6.0 This Documentation, which includes embedded help systems and electronically distributed materials (hereinafter referred to as the Documentation ),
More informationRelease Notes P/N REV A01 August 11, 2010
EMC Unisphere Analyzer Version 1.0.0 Release Notes P/N 300-011-113 REV A01 August 11, 2010 These release notes contain supplemental information about EMC Unisphere Analyzer, version 1.0.0. Topics include:
More informationAvigilon Control Center 5.10 Release Notes
Version 5.10.0.16 Released Tuesday September 27 th, 2016 Release Summary This is the initial release of Avigilon Control Center 5.10. This release introduces support for Avigilon fisheye cameras and adds
More informationHansaton LabMaster Installation Guide
Hansaton LabMaster Installation Guide LabMaster Manufacturing Service Software Installation Code: R2j99x5 Contents Welcome to LabMaster... 3 Installation... 4 Install LabMaster... 4 Network Database Installation...
More informationThe following links are to self-extracting Zip files on our web site which contain all the installation files:
AssetGen Professional Download and Installation Instructions 1.0 Platform Requirements The platform requirements are: - Windows 7 or greater. 2GB RAM (4GB recommended for 64 bit systems). - Microsoft.Net
More informationJNIOR Support Tool Release Notes
JNIOR Support Tool Release Notes Version 7.5.518.1209 18 May 2018 7.4.417.1208 17 April 2018 7.4.410.1512 10 April 2018 RELEASED to the website May 25, 2018 NOTE: This version or later IS REQUIRED with
More informationStarFinder. Operation Manual. Star Chromatography Workstation Version 6
Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA 94598-1675/USA Star Chromatography Workstation Version 6 StarFinder Operation Manual Varian, Inc. 2002 03-914723-00:Rev. 2 Trademark Acknowledgment
More informationPlatform Compatibility...1 New Features...2 Known Issues...4 Related Technical Documentation...5
ViewPoint SonicWALL ViewPoint 6.0 Contents Platform Compatibility...1 New Features...2 Known Issues...4 Related Technical Documentation...5 Platform Compatibility Operating System Requirements SonicWALL
More informationData Acquisition with CP-2002/2003 Micro-GC Control
Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA 94598 Star Chromatography Workstation Version 6 Data Acquisition with CP-2002/2003 Micro-GC Control Operation Manual Varian, Inc. 2002
More informationSoftware Release Notes
Software Release Notes 24-February-2017 The Version and Build Number can be found in the Help menu, About ViewPoint window. Note: Build Release Date is encoded in the Build # in yyyymmdd format. For details
More informationdctrack Version 6.1.1
dctrack Version 6.1.1 These Release Notes contain important information regarding the release of this product, so please read them carefully. We strongly recommend that you read the entire document. Applicability
More informationINSTRUCTION MANUAL V1. 2x A
INSTRUCTION MANUAL V1. 2x A INTRODUCTION... 3 CONNECTIONS AND INSTALLATION... 4 MECHANICAL INSTALLATION... 4 ATTACHING AND DETACHING THE FRONT COVER... 8 ATTACHING AND DETACHING THE HMI... 9 ELECTRICAL
More informationQlikView NPrinting release notes
QlikView NPrinting 16.4.0.0 release notes What s new in this release? QlikView NPrinting is the reporting system for QlikView. Version 16.4.0.0 contains a lot of bug fixes. Support for Office 2016 Starting
More informationStonesoft Management Center. Release Notes for Version 5.6.1
Stonesoft Management Center Release Notes for Version 5.6.1 Updated: January 9, 2014 Table of Contents What s New... 3 Fixes... 3 System Requirements... 6 Basic Management System Hardware Requirements...
More informationQuest Desktop Authority Full Build Update Release Notes
Quest Desktop Authority Full Build Update 10.1.0.1006 Release Notes January 5, 2018 Version 10.1.0.1006 Importance: Mandatory Contents Readme Resolved Issues Compatibility of This Update Installing This
More informationQuest Desktop Authority Full Build Update Release Notes
Quest Desktop Authority Full Build Update 10.2.0.256 Release Notes July 13, 2018 Version 10.2.0.256 Importance: Mandatory Contents Readme Resolved Issues Compatibility of This Update Installing This Update
More informationUpgrade Guide. BCM Business Communications Manager
Upgrade Guide BCM50 2.0 Business Communications Manager Document Status: Standard Document Number: NN40020-401 Document Version: 01.01 Part Code: N0107119 Date: October 2006 Copyright 2006 Nortel Networks,
More informationLabChip EZ Reader V3.0 SP2 Software Release Notes
LabChip EZ Reader V3.0 SP2 Software Release Notes Installation Notes... 2 Starting the Installation... 2 Upgrading From V2.0, V2.1, or V2.2 to V3.0 SP2... 2 Uninstalling the Software... 3 Documentation...
More informationOperating guide. OPC server for ECL Comfort 310. Table of Contents
Operating guide OPC server for ECL Comfort 310 Table of Contents 1. Introduction to Danfoss ECL OPC Server... 2 2. Configuring databases... 6 3. Exporting configuration data... 7 4. Importing data from
More informationUnity Loader V14.0 Readme
Unity Loader V14.0 Readme Table of Contents 1 MAIN FEATURES... 2 2 SYSTEM REQUIREMENTS... 2 3 SETUP PROCEDURE... 3 3.1 Recommendations before installation... 3 3.2 Unity Loader Setup procedure... 3 3.3
More informationPowerCenter Repository Maintenance
PowerCenter Repository Maintenance 2012 Informatica Corporation. No part of this document may be reproduced or transmitted in any form, by any means (electronic, photocopying, recording or otherwise) without
More informationThis document contains release notes for MobilePrint 2.0 General Release.
MobilePrint 2.0 Release Notes This document contains release notes for MobilePrint 2.0 General Release. Contents MobilePrint 2.0 Release Notes... 1 Install and Upgrade Notes... 1 Supported upgrade path...
More informationLabMax-Pro PC data collection
LabMax-Pro PC data collection Table of Contents 1. The Question 2. Setting up data logging ahead of time 3. Exporting data after it s already been collected 4. Importing and exporting data from the LabMax-Pro
More informationColligo Briefcase for Mac. Release Notes
Colligo Briefcase for Mac Release Notes Contents Technical Requirements... 3 Release 7.5 06 Oct0ber 2017... 4 New in this Release... 4 Release 7.5 18 May 2017... 4 New in 7.5... 4 Issues 7.5... 5 Known
More informationCorpSystem Workpaper Manager
CorpSystem Workpaper Manager Version 6.8 August 2011 This bulletin provides important information relating to CorpSystem Workpaper Manager version 6.8. Please review this bulletin carefully prior to installation,
More informationIntegration Guide: innovi XProtect
Integration Guide: innovi XProtect Version 1.4, October 2018 Table of Contents 1 Overview... 3 1.1. About this Guide... 3 1.2. Enabling Procedure... 3 1.3. Mapping Cameras... 5 1.3.1. Professional+...
More informationGraphic Generation Tool Installation Instructions Code No. LIT Software Release Issued February 2017
Code No. LIT-12011685 Software Release 1.3.1 Issued February 2017 Refer to the QuickLIT website for the most up-to-date version of this document. Introduction...2 Summary of Changes...2 Related Documentation...2
More informationCurrent Firmware Release Notes
Current Firmware Release Notes CPI recommends that all econnect PDU customers upgrade to the current firmware version. Release Versions 1.23.126 & 2.23.126: Upgrade Procedure: Download the zip file from
More informationHow to update OS image from V1421 to V1501 by Exe file
Author Sean Hsu Version 1.0.0 Date 2010/4/20 Page 1/13 How to update OS image from V1421 to V1501 by Exe file Applies to: Platform OS version WinPAC/ViewPAC utility version WP-8x3x N/A N/A WP-8x4x All
More informationInstalling ValveLink Software. ValveLink Solo ValveLink SNAP ON ValveLink DTM ValveLink PLUG IN for PRM. Installation Guide D102800X012
Installation Guide ValveLink Software Installing ValveLink Solo ValveLink SNAP ON ValveLink DTM ValveLink PLUG IN for PRM www.fisher.com This page intentionally left blank. Installation Guide Table of
More informationPulse 4.11 SCADA/HMI. Installation Guide
Pulse 4.11 SCADA/HMI Installation Guide June 2017 Copyright The information in this document is subject to change without prior notice and does not represent a commitment on the part of AFCON Control and
More informationCTECS Connect 2.2 Release Notes December 10, 2009
(Formerly VTECS) CTECS Connect 2.2 Release Notes December 10, 2009 This document contains information that supplements the CTECS Connect 2.2 documentation. Please visit the CTECS Connect Support area of
More informationSpectrometer Visible Light Spectrometer V4.4
Visible Light Spectrometer V4.4 Table of Contents Package Contents...3 Trademarks...4 Manual Driver and Application installation...5 Manual Application Installation...6 First Start of the Application...8
More informationApplied Biosystems RapidFinder Software Version 1.0 and SDS Software Version Beta Release 10/05 Release Notes (PN Rev A)
Applied Biosystems RapidFinder Software Version 1.0 and SDS Software Version 1.3.2 Beta Release 10/05 Release Notes (PN 4370225 Rev A) Software Release notes for: Applied Biosystems RapidFinder Software
More informationWRTU Client User Manual. Date: 29 May, 2014 Document Revision: 1.05
WRTU Client User Manual Date: 29 May, 2014 Document Revision: 1.05 2014 by BiPOM Electronics, Inc. All rights reserved. WRTU Client User Manual. No part of this work may be reproduced in any manner without
More informationThe Unified CVP upgrade process has been described in this chapter in the following sections:
Cisco voice solution components upgrade is a multistage process; solution components are grouped in several stages for upgrading. Users must follow the solution level upgrade order mentioned in the Upgrade
More informationEZChrom Elite Compact. What s New in EZChrom Elite Compact Service Pack 2
EZChrom Elite Compact What s New in EZChrom Elite Compact 3.3.2 Service Pack 2 Notices Agilent Technologies, Inc. 2010 No part of this manual may be reproduced in any form or by any means (including electronic
More informationCP30/G30/MC31 Firmware Version 3100 Known Issues
CP30/G30/MC31 Firmware Version 3100 Known Issues Introduction This document lists issues that have been identified with firmware version 3100 for the Kingfisher CP30/G30/MC31 modules. Where possible, workarounds
More informationSystem Administration Guide
System Administration Guide THERAPY MANAGEMENT SOFTWARE FOR DIABETES Accessing data files CareLink Pro stores user and device data in a central data file (carelink.cldb) and an associated license file
More informationUNIVERSAL SOFTWARE. Universal Software. Data Sheet
Universal Software Data Sheet System Requirements: The minimum requirements for using the Software are: 1). Windows XP/Vista/7 2). A minimum of 512 MB RAM 3). 1 GB of hard disk space 4). Microsoft Office
More informationSecurity Management System - Managing Infinova 307X NVR channels. Video display in Security Management System server software
Security Management System - Managing Infinova 307X NVR channels This document describes steps to configure Infinova 307X NVR cameras in Security Management System software. Note Microsoft Visual C++ 2015
More informationTechnical Support Information No January Complementary Service Information for MicroSAM Part 4
Process Analytics USA Customer Focus Page 1 of 7 Complementary Service Information for MicroSAM Part 4 According to the increasing number of MicroSAM applications and the experience resulting from it,
More informationInfraStruxure Central 6.0 Release Notes
InfraStruxure Central 6.0 Release Notes Table of Contents Page # Part Numbers Affected.......1 Minimum System Requirements...1 New Features........1 Issues Fixed....3 Known Issues......4 Upgrade Procedure......6
More informationPerceptive Intelligent Capture
Perceptive Intelligent Capture Release Notes Hot Patch Build 5553 Version: 5.5 SP2 Written by: Product Documentation, R&D Date: September 2016 2014 Perceptive Software. All rights reserved. Perceptive
More informationWaters Headspace Control Option, Version 3.0
R E L E A S E N O T E S Waters Headspace Control Option, Version 3.0 Version 3.0 of the Headspace Control Option (HCO) contains instrument drivers for Agilent Headspace Samplers and a toolkit program that
More informationAvigilon Control Center 5.0 Release Notes
Version 5.0.2.28 Released March 7 th, 2014 V2.6.0.6 for HD Micro Dome cameras V2.6.0.14 for HD PTZ cameras V2.6.0.14 for HD Bullet cameras V2.6.0.18 for HD H.264 H3 cameras V2.2.0.30 for H.264 encoders
More informationDell Storage Compellent Integration Tools for VMware
Dell Storage Compellent Integration Tools for VMware Administrator s Guide Version 3.1 Notes, Cautions, and Warnings NOTE: A NOTE indicates important information that helps you make better use of your
More informationPHRED Installation Guide
PHRED Installation Guide ColdFusion Version 10, SQL Server Database January 27, 2014 PHRED Installation Guide Page 1 Table of Contents Application Environment... 3 Application Tailoring... 4 Web Server
More informationSoftware Release Notes
Software Release Notes January 16, 2014 The Version and Build Number can be found in the Help menu, About DataTrace Pro screen. Note: Build Release Date is encoded in the Build # in yyyymmdd format. For
More informationEZ IQ. Chromatography Software PN P1A
O P E R A T I N G M A N U A L EZ IQ Chromatography Software Trademarks The trademarks of the products mentioned in this manual are held by the companies that produce them. Windows is a registered trademark
More informationAllegro CX Field PC Release 1.07a
Allegro CX Field PC Release 1.07a Installation Instructions June 2007 Enhancements Enhancements included in the Allegro CX Field PC operating system release 1.07a are as follows: Integration of the OS
More informationIMC Network Traffic Analyzer 7.1 (E0301P04) Copyright (c) 2015 Hewlett-Packard Development Company, L.P. All Rights Reserved.
Network Traffic Analyzer 7.1 (E0301P04) Copyright (c) 2015 Hewlett-Packard Development Company, L.P. All Rights Reserved. Table of Contents 1. What's New in this Release 2. Problems Fixed in this Release
More informationAvigilon Control Center 5.6 Release Notes
Notes Version 5.6.0.28 Released Monday August 31 st, 2015 Release Summary This is a service release for Avigilon Control Center 5.6. Users experiencing issues listed in the Issues Fixed list should consider
More informationDell Storage Compellent Integration Tools for VMware
Dell Storage Compellent Integration Tools for VMware Version 4.0 Administrator s Guide Notes, Cautions, and Warnings NOTE: A NOTE indicates important information that helps you make better use of your
More informationOrchestrator. Version Release Information. Version 4.0 of the Orchestrator Software has now been released.
Orchestrator Version 4.0.0 Release Information Version 4.0 of the Orchestrator Software has now been released. Version 4.0 contains the following fixes/changes from V3.3 Operating System: Supports Windows
More informationRelease Notes. Contents. About this release. MYOB AccountRight Standard 2011 Beta MYOB AccountRight Plus 2011 Beta
Release Notes MYOB AccountRight Standard 2011 Beta MYOB AccountRight Plus 2011 Beta Contents Section See About this release below Getting started page 2 New features page 6 Features unavailable in this
More informationInstall and Known Issues
Autodesk A360 Collaboration for Revit version 2015.5 Install and Known Issues Introduction: Autodesk A360 Collaboration for Revit version 2015.5 Autodesk A360 Collaboration for Revit version 2015.5 can
More informationWaters Empower 2 Service Pack G
R E L E A S E N O T E S Waters Empower 2 Service Pack G Empower 2 Service Pack G enhances LDAP capabilities. These release notes also address the following information and instructions: System and software
More informationINSTALLATION GUIDE. ID DESIGNER PC-Based Software Installation Guide. Version 4.0
INSTALLATION GUIDE ID DESIGNER PC-Based Software Installation Guide Version 4.0 122 West State Street Traverse City, MI 49684 www.salamanderlive.com 877.430.5171 TABLE OF CONTENTS 1 ABOUT THIS GUIDE...
More informationGRAMS Suite Version 9.1
Thermo Scientific GRAMS Suite Version 9.1 Welcome Guide Revision A 2011 Thermo Fisher Scientific Inc. All rights reserved. Thermo Fisher Scientific Inc. provides this document to its customers with a product
More informationDATA LOGGER MODELS DL-1080/1081
DATA LOGGER MODELS DL-1080/1081 USER MANUAL V1.3x INTRODUCTION... 4 CONNECTIONS AND INSTALLATION... 5 MECHANICAL INSTALLATION... 5 ATTACHING AND DETACHING THE FRONT COVER... 8 ATTACHING AND DETACHING THE
More informationpclamp Electrophysiology Data Acquisition and Analysis Software
pclamp Electrophysiology Data Acquisition and Analysis Software Version 10.4.1 Software Release Notes May 2014 www.moleculardevices.com This document is provided to customers who have purchased Molecular
More informationdocalpha Installation Guide
ARTSYL DOCALPHA INSTALLATION GUIDE 1. docalpha Architecture Overview... 2 1.1. docalpha Server Components... 4 1.2. docalpha Production Environment Stations Overview... 4 1.3. docalpha Setup & Administration
More information25 Saving Setting Guide Import/Export Nodes and Symbols
25 Saving 25.1...25-2 25.2 Import/Export Nodes and Symbols...25-7 25-1 25.1 When you finish the settings of the entry node(s), symbol(s) and function(s), save these set data as a "Network Project File".
More informationOnline Backup Client User Manual
Online Backup Client User Manual Software version 3.21 For Linux distributions October 2010 Version 2.0 Disclaimer This document is compiled with the greatest possible care. However, errors might have
More informationDell Storage Integration Tools for VMware
Dell Storage Integration Tools for VMware Version 4.1 Administrator s Guide Notes, cautions, and warnings NOTE: A NOTE indicates important information that helps you make better use of your product. CAUTION:
More informationCustomer Release Notes
This document contains important information about this release. Be sure to provide this information to all users before proceeding with the installation. Note: The term Color Profiler Suite refers to
More informationXLmanage Version 2.4. Installation Guide. ClearCube Technology, Inc.
XLmanage Version 2.4 Installation Guide ClearCube Technology, Inc. www.clearcube.com Copyright and Trademark Notices Copyright 2009 ClearCube Technology, Inc. All Rights Reserved. Information in this document
More information