1 abc February 2005 Report Wide bore CT scanner comparison report version 12 best choice best practice nww.medical-devices.nhs.uk

2 About MHRA evaluation reports. What you can expect. The Device Evaluation Service (DES) aims to provide independent and objective evaluations of medical devices available on the UK market. Specialist centres, mainly in NHS Trusts, do the evaluations under long term contract to, and in accordance with protocols approved by, the Medicines and Healthcare products Regulatory Agency (MHRA). The evaluations are usually of a unit supplied by the manufacturer. We would expect this unit to be representative of the product on the market but cannot guarantee this. Prospective purchasers should satisfy themselves with respect to any modifications that might be made to the product type after MHRA s evaluation. The reports are intended to supplement, not replace, information already available to prospective purchasers. The MHRA DES does not have access to any information held by the Agency in its capacity as the Competent Authority for the UK, apart from any information already in the public domain. The reports will contain data given by the manufacturer on the regulatory status of their devices but, apart from this, they are not an indicator of the regulatory status of a product. Occasionally, DES refers products to the regulatory arm of the MHRA for considerations of breaches of the legislation governing medical devices. DES plays no further part in any regulatory investigation that ensues and does not have advance notification of any regulatory action that may follow. How to obtain MHRA reports. To order evaluation reports, a copy of the publications catalogue or to sign up for our alert service contact: MHRA Orders Department, Room 1207, Hannibal House, Elephant & Castle, London, SE1 6TQ. T: +44 (0) F: +44 (0) E: des@ mhra.gsi.gov.uk Visit www. mhra.gov.uk for a comprehensive list of publications, details of forthcoming evaluations, services and contacts. Colour reports. Full colour versions of all reports published after 2002 are available from the Internet at

3 Wide bore CT scanner comparison report version 12 David Platten, Nicholas Keat Maria Lewis, Sue Edyvean ImPACT Bence Jones Offices St George s Hospital London SW17 0QT Tel: +44 (0) Fax: +44 (0) impact@impactscan.org Crown Copyright 2005 For more information on ImPACT visit Apart from any fair dealing for the purposes of research or private study, or criticism, or review, as permitted under the Copyright, Designs & Patents Act, 1998, this publication may only be reproduced, stored, or transmitted in any form or by any means with the prior permission, in writing, of the Controller of Her Majesty s Stationery Office (HMSO). Information on reproduction outside these terms can be found on the HMSO website ( or hmsolicensing@cabinet-office.x.gsi.gov.uk. The MHRA is an executive agency of the Department of Health. ISBN

4 Contents Introduction 5 Purpose of this report...5 Comparison method...5 Specification comparison...5 Scanners covered in this report...6 Table 1: Scanners covered in this report...6 Specification comparison 7 Table 2: Scanner gantry...7 Table 3: Couch...7 Table 4: X-ray generator...8 Table 5: X-ray tube...8 Table 6: Detection system...9 Table 7: System start-up and calibration...9 Table 8: Scan parameters...10 Table 9: Helical scanning...11 Table 10: Scan projection radiography (SPR)...11 Table 11: Manufacturers' performance data...12 Table 12: Factors affecting image quality...13 Table 13: Operator's console...14 Table 14: Main computer...14 Table 15: Image storage...15 Table 16: Image reconstruction on main console...16 Table 17: 3D reconstruction display...17 Table 18: Optional facilities...18 Table 19: Installation requirements...19 Table 20: Independent workstation...20 Table 21: Image transfer and connectivity...21 Appendix 1: ImPACT and the MHRA 22 Background...22 ImPACT...22 MHRA support to purchasers and users

5 Introduction Purpose of this report In May 2003 the UK Government announced a 90 million fund to replace all CT and MRI scanners installed before 1997 as part of the NHS Cancer Plan. ImPACT have produced comparison reports for each phase of the purchase program. The primary aim of these reports is to aid the equipment selection process by providing comparisons of CT scanners that are currently on the market. This report is for phase 7 of the Cancer Plan funding. There are four separate reports for six to ten, sixteen, and 32 to 64 slice CT scanners, as well as a report on wide bore systems. The scope of this report is limited to CT scanners that have a gantry bore wider than a standard system. These systems have been developed primarily for applications in radiotherapy planning. Comparison method The data given in this report are representative of the scanners as of February 2005, and are liable to change as the performance of individual scanner models is changed and upgraded. In particular, optional features such as workstations and software packages may be listed as standard for the scanner replacement programme, but may not be included in other, separate scanner purchases. There are two main areas for comparison of the scanners: performance and specification. None of the scanners in this report have been tested by ImPACT at the time of publication, so the scope of this document is restricted to a comparison of system specifications. Specification comparison The specification comparison is presented as a side-by-side summary comparison of the specification of each scanner, workstation and related equipment. It is grouped into a series of sub-sections relating to different aspects of the scanner, such as gantry, tube and detectors etc. Manufacturers supplied the specification data in response to a template issued by ImPACT. The data has not been verified by ImPACT. 5

6 Introduction Scanners covered in this report At the time of writing, there are four manufacturers of medical CT scanners that sell their systems in the UK (in alphabetical order); GE Medical Systems, Philips Medical Systems, Siemens AG and Toshiba Medical Systems. The wide bore models are listed below. Table 1: Scanners covered in this report Manufacturer GE Philips Siemens Scanner model LightSpeed RT AcQsim RT Somatom Sensation The is a four slice system with a gantry aperture of 80 cm. It has a maximum scan field of view of 50 cm, but is able to reconstruct images up to 65 cm using extrapolation methods. The fourth generation AcQsim RT from Philips is a single slice scanner with a patient bore of 85 cm, and a maximum scan field of view of 60 cm. The 20-slice Siemens Somatom Sensation has a gantry bore of 82 cm, with a maximum scan field of view of 50 cm. It is possible to reconstruct images up to 82 cm using extrapolation methods. The detector bank consists of 32 x 0.6 mm and 8 x 1.2 mm rows. The features the Siemens Straton x-ray tube, a new novel design of x-ray tube with fast cooling rates. Toshiba and Philips have recently announced new wide bore systems, but detailed specifications were not available at the time of publication. The Philips Brilliance CT Big Bore will be a 16-slice system with an aperture of 85 cm, a scan field of view of 60 cm, and a maximum image size of 60 cm. The Toshiba Aquilion Large Bore will also be a 16-slice system, with a gantry aperture of 90 cm, a scan field of view of 70 cm, and a maximum image size of 85 cm. 6

7 Specification comparison Table 2: Scanner gantry Generation 3rd 4th 3rd Slipring Low voltage Low voltage Low voltage Aperture (cm) Scan fields of view (cm) 25 and , 82 extended Nominal slice widths for axial scans (mm) 0.625, 1.25, 2.5, 3.75, 5, 7.5, 10 2, 3, 4, 5, 8 1.5, 2.0, 2.4, 3.0, 4.0, 4.8, 5, 6, 7, 8, 10, 12, 24 Tilt range (degrees) ± 30 ± 30 ± 30 Type of positioning lights Laser Laser Laser Accuracy of positioning lights (mm) Table 3: Couch ± 1 at any laser to patient distance ± 0.5 mm at centre of gantry ± 1 Couch top material Carbon fibre Carbon fibre Carbon fibre Couch top length and width (cm) 239 x 62 (42 just for cradle) 168 x 34 (just for cradle) 243 x 40 (just for cradle) Horizontal movement range (cm) Horizontal movement speeds (mm/sec) Accuracy/reproducibility of table positioning (mm) Scannable horizontal range without table top extension (cm) Scannable horizontal range with table top extension(s) (cm) Vertical movement range out of gantry (cm) Vertical movement range in gantry (cm) Minimum couch top height outside gantry (cm) Maximum weight allowed on couch (kg) Maximum weight on couch which still achieves stated performance specifications (kg) up to , 40, 50, ± 0.25 ± 0.25 ± (Axial), 160 (Helical & Scout) 170 (Axial), 160 (Helical & Scout) (±0.25mm) 205 (±1mm)

8 Specification comparison Table 4: X-ray generator Type High frequency High frequency High frequency Location Rotation assembly Rotation assembly Rotation assembly Power rating (kw) kv settings available 80, 100, 120, , 100, 120, 130, , 100, 120, 140 ma Range and Step size (5mA steps) (10mA steps) (1mA steps) Max. ma allowed for each kv 80kV: 400 ma 100kV: 420 ma 120kV: 440 ma 140kV: 380 ma 80kV: 400 ma 100kV: 400 ma 120kV: 400 ma 130kV: 320 ma 140kV: 320 ma 80 kv: 400mA 100kV: 400mA 120kV: 400mA 140kV: 360mA Table 5: X-ray tube Type and make GE Performix Pro Philips Rhino Siemens Straton Focal spot size(s) (mm), quoted to IEC 336/93 standard Settings at which focal spot changes from small to large Total filtration (inherent + beam shaping filter) at central axis (mm Al equivalent) 0.6 x x 0.9 Info. not available 6.8 (70kV, head) 9.5 (70kV, body) 0.4 x x 0.9 Automatically selected dependent on scan protocols 0.8 x 1.1 Not applicable Anode heat capacity (MHU) , equiv to 30 Maximum anode cooling rate (khu/min) Method of cooling Glycol to air Oil to air Oil to air Guaranteed tube life 1 year unlimited guarantee 100,000 revolutions 1 year unlimited guarantee 8

9 Specification comparison Table 6: Detection system Detector type Solid state (HiLight / Lumex) Solid State (Cadmium Tungstate) Solid state (Ultra Fast Ceramic) Number of detectors per row 888 (plus 18 reference elements) (1344 channels) Number of elements along z-axis Effective length of each element at isocentre (mm) Total effective length of detector array at isocentre (mm) Future option for more slices / rotation 16 x x 0.6, 8 x None currently None currently Yes Table 7: System start-up and calibration Power-on to warm-up time from fully off (mins) Tube warm-up time from 'cold' to operating temperature (mins) Time to perform detector calibrations at warm-up (mins) Recommended frequency for any additional calibration by the radiographer Time to perform these additional calibrations (mins) Total time from fully off to scanning in an emergency (mins) Included in tube warmup Part of warm-up procedure Once every 24 hours Not required Not required 13 (inc warm-up) Not required Not required < 3 Info. not available 4 5 9

10 Specification comparison Table 8: Scan parameters Reconstruction fields of view (cm) 9.6 to Extended scan field of view (cm) 65 Entire FOV uses standard reconstruction 82 Nominal slice widths (mm) and number of simultaneous slices (axial scans) Scan times for axial scans (s) * = Partial scans 2 x 0.625, 1 x 1.25, 4 x 1.25, 1 x 2.5, 4 x 3.75, 4 x 5 2, 3, 4, 5, 8 1, 2, 3, 4 0.6*, 1, 1.5, 2, 3, 4, 5 20 x 1.2, 10 x 2.4, 5 x 4.8, 4 x 6, 1 x 5, 2 x 12, 1 x *(option), 0.5 (option), 0.67*, 1.0s kv settings available 80, 100, 120, , 100, 120, 130, , 100, 120, 140 ma Range and Step size (5mA steps) (10mA steps) (1mA steps) Max. ma allowed for each kv 80kV: 400 ma 100kV: 420 ma 120kV: 440 ma 140kV: 380 ma 80kV: 400 ma 100kV: 400 ma 120kV: 400 ma 130kV: 320 ma 140kV: 320 ma 80 kv: 400mA 100kV: 400mA 120kV: 400mA 140kV: 360mA 10

11 Specification comparison Table 9: Helical scanning Rotation times for helical scanning (s) Number of simultaneous slices at each rotation time Pitches available for routine scanning (range and increment) Recommended pitches for optimal image quality 1 1, 1.5, 2, 3, 4, (option), slice : 1 4-slice : 0.75 & slice : 1 4-slice : 0.75 & freeely selectable None freely selectable None Helical interpolation algorithms available GE Proprietary algorithms (SmartHelical & MDMP, including CrossBeamTM & HyperplaneTM) Fan Beam SureView, AMPR conebeam artefact reduction Maximum number of rotations in one helical run at standard abdomen parameters 70sec ( ma) 90sec ( ma) 110sec ( ma) Maximum continuous scan time (s) 120 Limited by disc capacity and tube loading Limited by disc capacity and tube loading 160 (324mA,.5s) 176 (300mA,.5s) 200 (273mA,.5s) 100 Starting with a cold tube, the maximum helical scan distance using a 1 mm imaged slice thickness and a pitch of 1.5 (mm) 1600 Limited by disc capacity and tube loading 1570 Gantry tilt range for helical scanning (degrees) Table 10: Scan projection radiography (SPR) ± 30 ± 30 ± 30 Maximum SPR length (mm) SPR field dimensions (mm x mm) 500 x x x 1536 Angular positions of X-ray tube available for SPR (1 steps) AP, PA, LAT (±1 accuracy) Real time image No No Yes Accuracy of slice prescription from the scanogram (mm) Accuracy of distance measurements from SPR's taken at isocentre (lateral and axial directions) (mm) ± 0.25 ± 0.25 ± 0.5 < 2 x image pixel size ± 0.25 ±

12 Specification comparison Table 11: Manufacturers' performance data Spatial resolution (lp/cm) for sharpest clinical algorithm. Acquisition parameters in brackets. MTF0: 15.4 MTF10: 13.9 (10cm DFOV, Edge alg, Small Focus) MTF0 15 lp/cm MTF0: 15.4 ±10% MTF2: 14.5 ±10% (1s, 160 ma, 120 kv, 1.2 mm) Contrast resolution: smallest rod size (mm) discernable at given parameters in 20 cm CATPHAN mgy with 95%CL: 120kVp, 135mAs, 10mm, Std alg % CT number accuracy (HU) Water : ± 3 ± 4 CTDI settings for std head CTDI (mgy/100mas), centre of head phantom CTDI (mgy/100mas), periphery of head phantom CTDI settings for std body CTDI (mgy/100mas), centre of body phantom CTDI (mgy/100mas), periphery of body phantom 120 kvp, 260 mas, 2 x 10 mm Spiral: mgy: 120 kv, 140 mas 10 mm Air: ± 10 Water: ± kv 120 kv, 20 x 1.2 mm kvp, 260 mas, 2 x 10 mm 130 kv 20 x 1.2 mm Dose profile FWHM (mm) (focal spot size in brackets) 20: 20.6(l) 15: 16.5 (s) 10: 11.5 (s) 5: 6.7 (s) 1.25: 3.5 (s) 2 x 0.63:1.9(s) ± 10% for all collimations Info. not available 12

13 Specification comparison Table 12: Factors affecting image quality Post-patient collimation for narrow slices Automatic ma control (AEC / ma modulation) software No No No 3D Dose Modulation No Yes (CAREDose 4D) - ma adjustment for patient size Yes No Yes - ma adjustment along the z-axis Yes No Yes - ma modulation during rotation Yes No Yes Adaptive filtration for noise reduction Low signal correction No Yes (automatic) Quarter detector shift Yes No Yes Moving (dynamic/flying) focal spot No No Yes Number of imaging detectors per row Sampling frequency (Hz) s scan (hardware capable to 1760) 672 (1344 channels) 2400 Max Artefact reduction algorithms Cone beam correction Iterative Bone Option (IBO), Motion correction, Recon of thick slices from thinner ones GE Proprietary algorithms (SmartHelical & MDMP, including CrossBeamTM & HyperplaneTM) No Not required Modified beam hardening (abdomen, pelvis, shoulder), Motion correction (sequential modes), Posterior Fossa optimisation Yes, SureView and AMPR cone-beam artefact reduction 13

14 Table 13: Operator's console Specification comparison Diagonal dimension of image screen (inches) CRT 18.1 Number of monitors at console (functions of each if > 1) 2 (patient info and technique selection / image display) 1 1 acquisition and optional 1 x review and processing shared database console Image area matrix dimensions 512 x 512, 768 x 768, 1024 x x x 1024 Usual range of CT Number displayed (HU) to to to (-10,240 to 30,710 if metal implants) Accuracy of distance measurements in x-y plane (mm) < 2 times image pixel size ± 0.25 depends on pixel size Weighted CTDI (CTDIw or CTDIvol) displayed on console Dose Length Product (DLP) displayed on console Geometric Efficiency displayed on console when <70% Control methods Yes No Yes Yes No Yes Yes No No Mouse, trackball, keyboard Mouse, keyboard Mouse, keyboard Table 14: Main computer Make and model HP XW8000 Sun Ultrasparc Siemens PC compatible with array processors Operating system Linux RedHat 7.3 Unix Windows XP Type and speed of CPU 2 x 2.66 GHz 64 bit RISC Amount of computer RAM supplied as standard (Gbytes) Maximum amount of computer RAM (Gbytes) 10 x Pentium Xeon 3.2 GHz Info. not available 18 14

15 Specification comparison Table 15: Image storage Total standard hard disk capacity (Gbytes) Maximum hard disk capacity (Gbytes) Hard disk capacity for image storage (Gbytes and no. of uncompressed 512x512 images) 146 (250,000 images) 26 (11,000 data files) 146 (260,000 images) Hard disk capacity for storage of raw data files (Gbytes and no. of data files) 72 ( slice axial raw data files) 26 (1,000 raw data files) 300 Archive options MOD [images] & DVD [scan data, protocols] (standard) Tape (standard) MOD Option MOD and CD writer (standard) Capacity of a single archive disk (Gbytes and no. of images) 4.6 (9400 losslessly compressed 512x512 images or 700 raw data files) Tape 6500 Images per cartridge 2.3 Gb MOD 1400 Images MOD: 4.1GB (26,000 lossless images), CD- R: 0.65GB (4100 lossless images) 512 x 512 Time to mount an archive disk or tape (s) 5-6 in background operation Tape - depends upon images requested Immediate (disk continually accessible) Approx 30 for full disk Archive data transfer rate (images / s) 1 (read) 0.7 (write) >

16 Specification comparison Table 16: Image reconstruction on main console Reconstruction matrix Minimum reconstruction interval in helical scanning (mm) Reconstruction time from the start of data acquisition to the appearance of the 30th image of a series for a standard axial brain scan (s) (with IBO) 90 3 (up to 20 f/s) Reconstruction time from the start of data acquisition to the appearance of the 30th image of a series for an axial spine scan (s) (up to 20 f/s) Reconstruction time from the start of data acquisition to the appearance of the 30th image of a series for a helical abdomen scan (s) (up to 20 f/s) Simultaneous scanning and reconstruction Any delay in either scanning or reconstruction when performed concurrently Simultaneous scanning and routine analysis Yes No Yes No Info. not available No Yes Yes Yes Simultaneous scanning and archiving and/or hard copying Simultaneous scanning and transfer to second console/workstation Yes Yes Yes Yes Yes Yes 16

17 Table 17: 3D reconstruction display Specification comparison MIPs and MinIPs (maximum and minimum intensity projections) MC-standard WS-standard (MIP & MinIP) MC-standard WS-standard (Angio MIP) MC-standard SSD (3D Shaded Surface Display) WS-standard (3D) MC-standard WS-standard (3D SSD) MC-standard 3D Volume rendering software WS-standard (Volume Rendering) MC-Option WS-standard (Vol Rend 4D Angio) 3D Virtual endoscopy WS-standard (Navigator) Option MPR (Multi-planar reconstruction) MC-standard WS-standard (MPR & MPVR) MC-standard WS-standard (MPR) MC-standard Planes available in MPR Axial, para-axial, sagittal, coronal, oblique, curvilinear All planes, any oblique Axial, sagittal, coronal, oblique, curvilinear 17

18 Specification comparison Table 18: Optional facilities Contrast injector Option Option Option Contrast media bolus tracking Standard (SmartPrep) Not available CT fluoroscopy software and hardware Option (SmartStep) Not available Standard (CAREBolus) Option (CAREVision with HandCARE) Hard-copy imaging device Option Option Option Radiotherapy planning table top Option (RT flat pad and Exact couch top) Standard Option Carbon fibre breast board Option Option Option Means for attaching patient immobilisation devices and a stereotactic frame to the end of the couch Option (Exact couch) Option Option Bone Mineral Densitometry MC-Not available (BMD) (Osteo CT) CT Angiography MC-standard WS-standard AVA (Vessel Assessment) option on WS WS option MC-standard basic package Dental Radiotherapy CT simulation software Prospective ECG-triggered cardiac software Retrospective ECG-gated cardiac software CT Perfusion software (Dentascan) MC-Not available (CT sim) (SmartScore) MC-Not available MC & (Cardiac Imaging) (CT Perfusion) WS option Not available Not available WS option (Dental CT) Coherence Dosimetrist (separate workstation) Not available Not available (Perfusion CT) 18

19 Table 19: Installation requirements Specification comparison Environmental requirements (max/min temperature, humidity) in scanner room ºC, 30-60% relative humidity C, 30-60% relative humidity ºC, 15-75% relative humidity Environmental requirements (max/min temperature, humidity) in scanner control room ºC, 30-60% relative humidity C, 30-60% relative humidity ºC, 15-75% relative humidity Peak heat output from system during scanning (kw) (add 0.07 for CT fluoro) System cooling method Output to air Water - air Water - water Air conditioning requirements for scanner room of minimum floor area Minimum floor area required for the system (m²) Recommended To meet environmental requirements None Gantry dimensions (H x W x D (mm)) and weight (kg) 1992 x 2439 x 1007, 1701kg 1960 x 2240 x 863, 1532 kg 1990 x 2280 x 940, 2000 kg Couch dimensions (H x W x L (mm)) and weight (kg) 1121 x 610 x 2387, 330kg 560 x 340 x 1680, 550kg 1022 x 690 x 2430, 500 kg Supplementary unit dimensions (H x W x D (mm)) and weight (kg) Power supply requirements Power Unit: 1067 x 711 x 559, 350kg 3 phase V, 150kVA None 3 phase V, 75 kva Power unit: 1815 x 905 x 800, 550kg Cooling unit: 1815 x 905 x 860, 200kg 3 phase V, 66-80kVA 19

20 Table 20: Independent workstation Specification comparison Is a workstation provided? Yes No, option (Brilliance Extended Workspace) No, option (LEONARDO) Computer make and model HP XP8000 Dell Xeon Siemens Fujitsu Xeon Operating system Linux Red Hat 7.3 Windows XP Windows XP Type and speed (GHz) of CPU 2 x x x Xeon 3.2 Amount of computer RAM supplied as standard (Gbytes) Maximum amount of computer RAM (Gbytes) Total hard disk storage capacity supplied as standard (Gbytes) Maximum total hard disk storage capacity (Gbytes) Archive options Capacity of a single archive disk (Gbytes) Environmental requirements (max/min temperature, humidity) for workstation CD-R standard MOD optional CD-R standard EOD option ºC, % relative humidity 0-40 C CD-R standard MOD option MOD: 4.1 CD-R: ºC, 20-80% relative humidity 20

21 Specification comparison Table 21: Image transfer and connectivity Speed of scanner / workstation connections to local area networks (Mbits/s) Remote PC access to images on workstation DICOM services on Main Console or Option Option (Easyweb) Option Storage SCU, SCP SCU, SCP SCU, SCP Print SCU SCU SCU Query / retrieve SCU, SCP SCU, SCP SCU, SCP Modality worklist Option Option SCU Performed procedure step Option Option SCU Storage commitment Not available Option SCU DICOM services on Workstation Storage SCU, SCP SCU, SCP SCU, SCP Print SCU SCU SCU Query / retrieve SCU, SCP SCU, SCP SCU, SCP Modality worklist management Not available Not available Not available Performed procedure step Not available Not available Not available Storage commitment Not available Not available SCU. 21

22 Appendix 1: ImPACT and the MHRA Background One of the roles of the Medicines and Healthcare products Regulatory Agency (MHRA) is to fund evaluation programmes for medical devices and equipment. The programme includes evaluation of x-ray Computed Tomography Equipment currently available on the UK market. MHRA aims to ensure that evaluation techniques keep abreast of improvements in CT imaging performance and that MHRA reports present evaluation information that is timely, useful and readily understood. ImPACT ImPACT (Imaging Performance Assessment of Computed Tomography) is the MHRA's CT evaluation facility. It is based at St George's Hospital, London, part of St George's Healthcare NHS Trust. ImPACT have developed test objects and measurement procedures suitable for inter-comparing CT scanner performance. For each CT evaluation hundreds of images are obtained from the system under test and subsequently analysed using custom written software. Dose measurements are made using ion chambers, and x- ray film is used to obtain additional x-ray dose information. MHRA support to purchasers and users The ImPACT team is available to answer any queries with regard to the details of this report, and also to offer general technical and user advice on CT purchasing, acceptance testing and quality assurance. ImPACT Bence Jones Offices St George s Hospital London SW17 0QT T: +44 (0) F: +44 (0) E: impact@impactscan.org W: MHRA contact point for general information on the CT evaluation programme: Device Evaluation Service MHRA Elephant and Castle London SE1 6TQ T: +44 (0) F: +44 (0) E: des@ mhra.gsi.gov.uk W: www. mhra.gov.uk 22

Annexure XII SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER

Annexure XII SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER Annexure XII SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER A) Scanner Design X-Ray generator and tube: 1. Scanner: Whole body spiral CT scanner (16 slices) of latest technology.

More information

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing ImPACT Information Leaflet No. 1: CT Scanner Acceptance Testing Version 1.02, 18/05/01 CONTENTS: 1. SCOPE OF LEAFLET 2. GENERAL PRINCIPLES OF ACCEPTANCE AND COMMISSIONING 2.1 PHANTOMS 2.2 EXPOSURE AND

More information

Comparison of the Imaging Performance of CT Scanners, Issue 11

Comparison of the Imaging Performance of CT Scanners, Issue 11 Imaging Performance Assessment of CT Scanners Comparison of the Imaging Performance of CT Scanners, Issue 11 Comparison data is presented on the following scanners : Elscint Twin Flash IGE HiSpeed Advantage

More information

Equipment Specification

Equipment Specification MULTISLICE CT SCANNER ( 16 Slices ) Merk : Hitachi Japan Model : SUPRIA 5 MHU Price : Rp 6.512.360.215,27 No Equipment Specification 1 2 3 Scanner Gantry Object for scanning : Whole body including head

More information

Fits you like no other

Fits you like no other Fits you like no other BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase in room

More information

PRODUCT DATA. Advanced 128

PRODUCT DATA. Advanced 128 PRODUCT DATA Advanced 128 Supria 16 Slice CT Puts You On The Path of High Quality, Cost-Effective CT Scanning Solid Capabilities Are Built Into the Supria Addressing the challenges of controlling healthcare

More information

Solid Capabilities Are Built Into the Supria Plus. Putting You On The Path of High Quality, Cost-Effective CT Scanning

Solid Capabilities Are Built Into the Supria Plus. Putting You On The Path of High Quality, Cost-Effective CT Scanning Specification Data Putting You On The Path of High Quality, Cost-Effective CT Scanning Solid Capabilities Are Built Into the Supria Plus Addressing the challenges of controlling healthcare organization

More information

Fits you like no other

Fits you like no other Fits you like no other Philips BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase

More information

Optimisation of Toshiba Aquilion ONE Volume Imaging

Optimisation of Toshiba Aquilion ONE Volume Imaging Optimisation of Toshiba Aquilion ONE Volume Imaging Jane Edwards, RPRSG Royal Free London NHS Foundation Trust Dr Mufudzi Maviki, Plymouth Hospitals NHS Trust Background In 2011/12 Radiology at RFH was

More information

SCENARIA 64-Slice CT Putting Advanced CT Within Reach.

SCENARIA 64-Slice CT Putting Advanced CT Within Reach. SCENARIA 64-Slice CT Putting Advanced CT Within Reach. Advancing the technology of Computed Tomography for over 30 years, Hitachi is a recognized innovator of lower dose*, high diagnostic value CT solutions.

More information

O-ARM IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE

O-ARM IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE SYSTEM FEATURES CATEGORY PHYSICAL DIMENSIONS IMAGING MODALITY PERFORMANCE SPECIFICATION Length Width Height Weight Gantry Opening Bore Diameter Single Plane

More information

Computer-Tomography II: Image reconstruction and applications

Computer-Tomography II: Image reconstruction and applications Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de

More information

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead.

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead. 1 CT Basics Principles of Spiral CT Dose 2 Who invented CT? 1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections Sir Godfrey Hounsfield worked for the Central

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

ImPACT. CT dosimetry and a data base for CTDI values. EFOMP Workshop at ECR Imaging Performance Assessment of CT Scanners

ImPACT. CT dosimetry and a data base for CTDI values. EFOMP Workshop at ECR Imaging Performance Assessment of CT Scanners ImPACT Imaging Performance Assessment of CT Scanners A Medical Devices Agency Evaluation Group CT dosimetry and a data base for CTDI values EFOMP Workshop at ECR 2002 S. Edyvean, ImPACT St George s Hospital,

More information

Evaluation of AutoQA Lite TM Image Quality Measurement Software

Evaluation of AutoQA Lite TM Image Quality Measurement Software Evaluation of AutoQA Lite TM Image Quality Measurement Software Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4 2XU Phone: 0131 537 1161

More information

RADIOLOGY AND DIAGNOSTIC IMAGING

RADIOLOGY AND DIAGNOSTIC IMAGING Day 2 part 2 RADIOLOGY AND DIAGNOSTIC IMAGING Dr hab. Zbigniew Serafin, MD, PhD serafin@cm.umk.pl 2 3 4 5 CT technique CT technique 6 CT system Kanal K: RSNA/AAPM web module: CT Systems & CT Image Quality

More information

SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER

SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER A) Scanner Design X-Ray generator and tube: 1. Scanner: Whole body spiral CT scanner (16 slices) of latest technology. 2. X-Ray

More information

Medical Image Processing: Image Reconstruction and 3D Renderings

Medical Image Processing: Image Reconstruction and 3D Renderings Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,

More information

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic

More information

SCENARIA Hitachi Innovation at Work for You.

SCENARIA Hitachi Innovation at Work for You. PRODUCT DATA SCENARIA Hitachi Innovation at Work for You. Continuing its commitment to the SCENARIA scalable CT platform, Hitachi introduces cutting-edge 64/128-slice technology with the SCENARIA SE. The

More information

SRI GURU RAMDAS INSTITUTE OF MEDICAL SCIENCES & RESEARCH MEHTA ROAD, VALLAH, SRI AMRITSAR

SRI GURU RAMDAS INSTITUTE OF MEDICAL SCIENCES & RESEARCH MEHTA ROAD, VALLAH, SRI AMRITSAR SRI GURU RAMDAS INSTITUTE OF MEDICAL SCIENCES & RESEARCH MEHTA ROAD, VALLAH, SRI AMRITSAR TECHNICAL SPECIFICATIONS FOR PURCHASE OF CT SIMULATOR FOR DEPARTMENT OF RADIOTHERAPY 1. MANDATORY REQUIREMENTS

More information

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography

More information

DX-D 300. Flexible Direct Radiography System

DX-D 300. Flexible Direct Radiography System DX-D 300 Flexible Direct Radiography System MUSICA 2 processing provides superior contrast detail and exam-independent, consistent image quality Cesium Iodide DR detector technology offers potential for

More information

Z-MOTION. Universal Digital Radiographic System Z-MOTION. Control-X Medical CONTROL-X MEDICAL

Z-MOTION. Universal Digital Radiographic System Z-MOTION. Control-X Medical CONTROL-X MEDICAL Control-X Medical Z-MOTION Compact design, low ceiling height requirement Motorized and manual movement capability Wide motion / SID range Best-in-class image quality Flexible connectivity to PACS systems

More information

DX-D 300. Flexible Direct Radiography System

DX-D 300. Flexible Direct Radiography System DX-D 300 Flexible Direct Radiography System MUSICA² processing provides outstanding contrast detail and consistently excellent image quality Universal, flexible and affordable modality combines a single

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4

More information

Fujifilm DR Solution. FDR AcSelerate. The new pinnacle in diagnostic imaging from Fujifilm ISS. CsI. Dynamic Visualization. Technology.

Fujifilm DR Solution. FDR AcSelerate. The new pinnacle in diagnostic imaging from Fujifilm ISS. CsI. Dynamic Visualization. Technology. Fujifilm DR Solution FDR AcSelerate The new pinnacle in diagnostic imaging from Fujifilm CsI Scintillator ISS Technology Dynamic Visualization Welcome to the X-ray room of the future! A streamlined solution

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

(restyle 2011) Anatomic programs

(restyle 2011) Anatomic programs Product Data Panoramic X-Ray System with CCD sensor (restyle 2011) Examination programs Panoramic - adult - child - child with deceleration ramp for spine compensation TMJ - open/close mouth (4 views on

More information

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA Computed Tomography Principles, Design, Artifacts, and Recent Advances THIRD EDITION Jiang Hsieh SPIE PRESS Bellingham, Washington USA Table of Contents Preface Nomenclature and Abbreviations xi xv 1 Introduction

More information

Symbia T Series System Specifications Answers for life.

Symbia T Series System Specifications Answers for life. www.siemens.com/mi System Specifications Answers for life. 2 Diagnostic Excellence Today s healthcare institutions have increasingly diverse patient needs, greater budget limitations and the need for higher

More information

DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300

DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300 DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM MUSICA processing provides excellent contrast detail and exam-independent, consistent image quality Cesium Iodide DR detector technology offers potential for

More information

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed

More information

Whole Body X-ray CT System Supria. DICOM Conformance Statement

Whole Body X-ray CT System Supria. DICOM Conformance Statement Whole Body X-ray CT System Supria DICOM Conformance Statement Tokyo, Japan E1E-HC0001-02 Hitachi, Ltd. 2016. All rights reserved. ii E1E-HC0001 History Revision Description Date Rev. 1.0 Initial. 2016/4/1

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

Practice-oriented robust solution for mobile X-ray. The portable DR system for mobile X-ray. Leonardo DR 121OP with dicom PACS DX-R Software

Practice-oriented robust solution for mobile X-ray. The portable DR system for mobile X-ray. Leonardo DR 121OP with dicom PACS DX-R Software Leonardo DR 121OP The portable DR system for mobile X-ray total weight only ca. 26kg The portable DR system for mobile X-ray Practice-oriented robust solution for mobile X-ray O and R Technology Digital

More information

DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300 DR SYSTEM

DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300 DR SYSTEM DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM The DX-D 300 DR system unites excellent image quality with the ultimate convenience. It offers top-of-the-line technology: a fullymotorized positioner enabling

More information

Philips SPECT/CT Systems

Philips SPECT/CT Systems Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET

More information

PURE. ViSION Edition PET/CT. Patient Comfort Put First.

PURE. ViSION Edition PET/CT. Patient Comfort Put First. PURE ViSION Edition PET/CT Patient Comfort Put First. 2 System features that put patient comfort and safety first. Oncology patients deserve the highest levels of safety and comfort during scans. Our Celesteion

More information

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist CT Protocol Review: Practical Tips for the Imaging Physicist Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center August 8, 2013 AAPM Annual Meeting Goals Understand purpose and importance

More information

Routine excellence. Philips Ingenuity Core specifications

Routine excellence. Philips Ingenuity Core specifications Routine excellence Philips Ingenuity Core specifications Table of contents 1 Introduction 3 2 User interface 4 2.1 ipatient key benefits 4 2.2 ExamCards 4 2.3 ScanRuler 4 3 DoseWise 5 3.1 DoseRight Index

More information

AAPM Standard of Practice: CT Protocol Review Physicist

AAPM Standard of Practice: CT Protocol Review Physicist AAPM Standard of Practice: CT Protocol Review Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center September 11, 2014 2014 Texas Radiation Regulatory Conference Goals Understand purpose

More information

Planmeca ProMax 3D Max CBCT unit

Planmeca ProMax 3D Max CBCT unit D00107623 1(6) Planmeca ProMax 3D Max CBCT unit Introduction The Planmeca ProMax 3D Max X-ray unit uses cone beam computerized tomography (CBCT) to produce threedimensional X-ray images. Panoramic and

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300

FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300 The DX-D 300 DR system unites excellent image quality with complete convenience. It offers top-of-the-line technology, a single detector and a fully-motorized

More information

PLANMECA PROMAX 3D MID CBCT UNIT

PLANMECA PROMAX 3D MID CBCT UNIT 1(5) PLANMECA PROMAX 3D MID CBCT UNIT Introduction The Planmeca ProMax 3D Mid X-ray unit uses cone beam computerized tomography (CBCT) to produce three-dimensional X-ray images. Panoramic and cephalometric

More information

DX-D 100 with wireless detector

DX-D 100 with wireless detector M O B I L E D R S O L U T I O N DX-D 100 with wireless detector Patients who most need imaging exams may lack the mobility necessary to move to the X-ray room or to position themselves properly for optimum

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

Computer-Tomography I: Principles, History, Technology

Computer-Tomography I: Principles, History, Technology Computer-Tomography I: Principles, History, Technology Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de History

More information

Fundamentals of CT imaging

Fundamentals of CT imaging SECTION 1 Fundamentals of CT imaging I History In the early 1970s Sir Godfrey Hounsfield s research produced the first clinically useful CT scans. Original scanners took approximately 6 minutes to perform

More information

Planmeca ProMax 3D s CBCT unit

Planmeca ProMax 3D s CBCT unit D0010759 1(5) Planmeca ProMax 3D s CBCT unit Introduction The Planmeca ProMax 3D s X-ray unit uses cone beam computerized tomography (CBCT) to produce threedimensional X-ray images. Panoramic and cephalometric

More information

A closer look at CT scanning

A closer look at CT scanning Vet Times The website for the veterinary profession https://www.vettimes.co.uk A closer look at CT scanning Author : Charissa Lee, Natalie Webster Categories : General, Vets Date : April 3, 2017 A basic

More information

Financial disclosure. Onboard imaging modality for IGRT

Financial disclosure. Onboard imaging modality for IGRT Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty

More information

Protocol. Technical evaluation of X-ray tomographic image-guided radiotherapy devices CEP10070

Protocol. Technical evaluation of X-ray tomographic image-guided radiotherapy devices CEP10070 Protocol Technical evaluation of X-ray tomographic image-guided radiotherapy devices CEP10070 March 2010 Contents 2 Introduction... 3 Protocol design and validation... 4 General information... 7 Technical

More information

CT: Physics Principles & Equipment Design

CT: Physics Principles & Equipment Design CT: Physics Principles & Equipment Design James Kofler, Ph.D Radiology Mayo Clinic Rochester, MN June 27, 2012 Disclosures Nothing to disclose Learning Objectives Understand fundamental concepts of - CT

More information

DX-D 800 DX-D 800 DYNAMIC, 3-IN-1 DIRECT RADIOGRAPHY SYSTEM OFFERING REAL TIME IMAGES FOR FLUOROSCOPY, GENERAL RADIOGRAPHY AND DIRECT EXPOSURES.

DX-D 800 DX-D 800 DYNAMIC, 3-IN-1 DIRECT RADIOGRAPHY SYSTEM OFFERING REAL TIME IMAGES FOR FLUOROSCOPY, GENERAL RADIOGRAPHY AND DIRECT EXPOSURES. D I R E C T R A D I O G R A P H Y S Y S T E M DX-D 800 The DX-D 800, part of Agfa HealthCare s family of Direct Radiography (DR) solutions, provides enhanced flexibility in diagnostics through its direct

More information

Toshiba Aquilion 4 3d Manual READ ONLINE

Toshiba Aquilion 4 3d Manual READ ONLINE Toshiba Aquilion 4 3d Manual READ ONLINE TOSHIBA AQUILION VISION: Head. EP No: 4. Boost 3D: OFF: OSR *** Filter: OFF. SureStart - Manual trigger as soon as contrast is seen: Scan mode: TOSHIBA AQUILION

More information

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Tecniques of imaging in Overview Introduction Tecniques of imaging

More information

VieW 3D. 3D Post-Processing WorKstation THE THIRD DIMENSION. Version 3.1

VieW 3D. 3D Post-Processing WorKstation THE THIRD DIMENSION. Version 3.1 VieW 3D 3D Post-Processing WorKstation THE THIRD DIMENSION Version 3.1 iq-view 3D THE FULLY-FEATURED 3D MEDICAL IMAGING SOLUTION FOR RADIOLOGISTS iq-view 3D contains all components of iq-view with the

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60601-2-44 2001 AMENDMENT 1 2002-09 Amendment 1 Medical electrical equipment Part 2-44: Particular requirements for the safety of X-ray equipment for computed tomography Amendement

More information

DX-D 100 with WITH ITS EXCELLENT IMAGE QUALITY AND

DX-D 100 with WITH ITS EXCELLENT IMAGE QUALITY AND M O B I L E D R S O L U T I O N DX-D 100 with wireless detector Patients who most need imaging exams may lack the mobility necessary to move to the X-ray room or to position themselves properly for optimum

More information

product data Raybow dr mobile digital radiographic system with wireless detector product data

product data Raybow dr mobile digital radiographic system with wireless detector product data Raybow dr mobile digital radiographic system with wireless detector Integrating all the advanced technologies, the ergonomic and high quality system Raybow dr lets you perform digital radiographies anywhere:

More information

Biomedical Imaging. Computed Tomography. Patrícia Figueiredo IST

Biomedical Imaging. Computed Tomography. Patrícia Figueiredo IST Biomedical Imaging Computed Tomography Patrícia Figueiredo IST 2013-2014 Overview Basic principles X ray attenuation projection Slice selection and line projections Projection reconstruction Instrumentation

More information

Lower Dose with Confidence

Lower Dose with Confidence Lower Dose with Confidence The Next Generation SCENARIA CT is Here. 64 or 128 Slice Capability The SCENARIA SE delivers lower dose technologies*, enhanced patient access, fast workflow speed and added

More information

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT CLINICAL COMMISSIONING AND ACCEPTANCE TESTING OF A 3D SURFACE MATCHING SYSTEM Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago Learning Objective Describe acceptance

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

Whole Body Submillimeter Scan

Whole Body Submillimeter Scan 128 Whole Body Submillimeter Scan The real value of 64ch/128slice CT systems does not come from the ability to perfom cardiac scans but the capability to scan all parts of the body in high definition submillimeter

More information

Overview. Digital X-ray imaging

Overview. Digital X-ray imaging Overview Digital X-ray imaging OR Technology Headquarters The OR Technology competence group: Your competent partner for digital X-ray imaging Producing digital X-ray technology and developing image management

More information

Evaluation report. X-ray tomographic image guided radiotherapy systems CEP10071

Evaluation report. X-ray tomographic image guided radiotherapy systems CEP10071 Evaluation report X-ray tomographic image guided radiotherapy systems CEP10071 March 2010 Contents 2 Summary... 3 Introduction... 6 Product description... 9 Methods... 14 Technical performance... 23 Purchasing...

More information

DX-D 100 WITH WIRELESS DETECTOR

DX-D 100 WITH WIRELESS DETECTOR DX-D 100 WITH WIRELESS DETECTOR MOBILE DR SOLUTION WITH ITS EXCELLENT IMAGE QUALITY AND FLEXIBLE HANDLING, THE MOBILE DX-D 100 WITH WIRELESS DETECTOR OFFERS FAST IMAGING THAT CAN BE VALIDATED IMMEDIATELY.

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

Future Topics. Projection Imaging Dose Reporting, XA 3D Volume Objects. for DICOM WG-02. Presented by Heinz Blendinger, Siemens Medical Solutions

Future Topics. Projection Imaging Dose Reporting, XA 3D Volume Objects. for DICOM WG-02. Presented by Heinz Blendinger, Siemens Medical Solutions Future Topics for Projection Imaging Dose Reporting, XA 3D Volume Objects DICOM WG-02 Presented by Heinz Blendinger, Siemens Medical Solutions 1 Presentation outline Dose Reporting Why Dose Reporting?

More information

Medical X-ray Imaging

Medical X-ray Imaging Medical X-ray Imaging HYBRID PORTABLE X-RAY 20BT lite / 20BT / 40BT The world s only hybrid powered portable X-ray Light weight, compact size and durable cover Light weight portable X-ray with high power

More information

Medical X-ray Imaging

Medical X-ray Imaging Medical X-ray Imaging HYBRID PORTABLE X-RAY 20BT lite / 20BT / 40BT The world s only hybrid powered portable X-ray Light weight, compact size and durable cover Light weight portable X-ray with high power

More information

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

C a t p h a n / T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 Internationally recognized for measuring the maximum obtainable performance of axial, spiral and multi-slice

More information

Computed tomography - outline

Computed tomography - outline Computed tomography - outline Computed Tomography Systems Jørgen Arendt Jensen and Mikael Jensen (DTU Nutech) October 6, 216 Center for Fast Ultrasound Imaging, Build 349 Department of Electrical Engineering

More information

As fl exible as your care requires

As fl exible as your care requires As fl exible as your care requires Philips Ingenuity Flex 32 CT Built on Ingenuity The Philips Ingenuity Flex 32 helps you provide excellent care with outstanding flexibility, now and in the future. Built

More information

Micro-CT Methodology Hasan Alsaid, PhD

Micro-CT Methodology Hasan Alsaid, PhD Micro-CT Methodology Hasan Alsaid, PhD Preclinical & Translational Imaging LAS, PTS, GlaxoSmithKline 20 April 2015 Provide basic understanding of technical aspects of the micro-ct Statement: All procedures

More information

S. Guru Prasad, Ph.D., DABR

S. Guru Prasad, Ph.D., DABR PURPOSE S. Guru Prasad, Ph.D., DABR Director of Medical Physics IAEA Consultant NorthShore University Health System and University of Chicago, Pritzker School of Medicine Current TPS utilize more information

More information

Hitachi Medical Systems America, Inc. Hitachi Medical Corporation

Hitachi Medical Systems America, Inc. Hitachi Medical Corporation Advanced 128 Advanced 128 128 Slice Capability Enlarged 75cm Aperture Thin Gantry Profile 88cm Intelli IP Advanced, Iterative Reconstruction IntelliCenter Auto-Lateral Shifting Table, up to 16cm Range

More information

Symbia E and S System Specifications Answers for life.

Symbia E and S System Specifications Answers for life. www.siemens.com/mi Symbia E and S System Specifications Answers for life. Symbia E: Work with confidence. What does it mean to work with confidence? It is clarity not only image clarity but the clarity

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

CT Dose Profiler. Probe for evaluation of CT systems. CT Dose Profiler User's Manual - English - Version 6.4A. RTI article number:

CT Dose Profiler. Probe for evaluation of CT systems. CT Dose Profiler User's Manual - English - Version 6.4A. RTI article number: CT Dose Profiler Probe for evaluation of CT systems - English - Version 6.4A RTI article number: 9630512-00 CT Dose Profiler The CT Dose Profiler probe makes it possible to evaluate the performance of

More information

The protocols used to scan the musculoskeletal system are tailored to each patient and

The protocols used to scan the musculoskeletal system are tailored to each patient and Chapter 22. Musculoskeletal Protocols The protocols used to scan the musculoskeletal system are tailored to each patient and region being examined. The clinical indication for the examination will also

More information

OPERA Swing. Revolutionary R/F multifunctional remote controlled system all in one

OPERA Swing. Revolutionary R/F multifunctional remote controlled system all in one Revolutionary R/F multifunctional remote controlled system all in one Applications Musculoskeletal Chest Genitourinary Gastrointestinal Vascular Stitching Tomography Mylography Swallows ERCP Patient positioning

More information

DX-D 600 DX-D 600 HIGH-PRODUCTIVITY, EXCELLENT IMAGE QUALITY, DIRECT RADIOGRAPHY SYSTEM WITH STATE-OF-THE-ART DESIGN IN THREE CONFIGURATION OPTIONS.

DX-D 600 DX-D 600 HIGH-PRODUCTIVITY, EXCELLENT IMAGE QUALITY, DIRECT RADIOGRAPHY SYSTEM WITH STATE-OF-THE-ART DESIGN IN THREE CONFIGURATION OPTIONS. D i r e c t r a d i o g r a p h y s y s t e m DX-D 600 HIGH-PRODUCTIVITY, EXCELLENT IMAGE QUALITY, DIRECT RADIOGRAPHY SYSTEM WITH STATE-OF-THE-ART DESIGN IN THREE CONFIGURATION OPTIONS. Two-detector, high-productivity,

More information

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon WHITE PAPER Introduction An image guidance system based on Cone Beam CT (CBCT) is included in Leksell

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

One Machine... Mac OS Compatible! Multiple Modalities. ProMax 3D & 3D s

One Machine... Mac OS Compatible! Multiple Modalities. ProMax 3D & 3D s One Machine... Multiple Modalities Mac OS Compatible! 3D Bitewing Panoramic Cephalometric Features Availability of multiple imaging modalities in one machine (3D, bitewing, panoramic, and cephalometric)

More information

International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN:

International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: COMPARISON OF RADIATION DOSE AND IMAGE QUALITY BETWEEN THE SINGLE DETECTOR CT AND MULTI DETECTOR CT ON HEAD EXAMINATION: STUDY PHANTOM Samburi Master of Physics, University of Diponegoro, Indonesia Agency

More information

Cardiac Dual Energy CT: Technique

Cardiac Dual Energy CT: Technique RSNA 2013, VSCA51-01, Chicago, Dec. 5, 2013 Cardiac Radiology Series Cardiac Dual Energy CT: Technique Willi A. Kalender, Ph.D. Institute of Medical Physics University of Erlangen www.imp.uni-erlangen.de

More information

The Near Future in Cardiac CT Image Reconstruction

The Near Future in Cardiac CT Image Reconstruction SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de

More information

Profound understanding of anatomy

Profound understanding of anatomy ENGLISH Profound understanding of anatomy The unique Planmeca ProMax 3D product family offers equipment for all maxillofacial imaging. All volume sizes from the smallest special cases to whole head images

More information

DICOM Conformance Statement. CT AV Series Product Line. Document Number February 1997

DICOM Conformance Statement. CT AV Series Product Line. Document Number February 1997 Philips Medical Systems DICOM Conformance Statement CT AV Series Product Line Document Number 4522 983 65911 28 February 1997 Copyright Philips Medical Systems Nederland B.V. 1997 Philips Medical Systems

More information

DX-D WITH DX-D 40/45 DETECTOR FAMILY MOBILE DR SOLUTION

DX-D WITH DX-D 40/45 DETECTOR FAMILY MOBILE DR SOLUTION DX-D 100 + WITH DX-D 40/45 DETECTOR FAMILY MOBILE DR SOLUTION WITH ITS EXCELLENT IMAGE QUALITY AND FLEXIBLE HANDLING, THE MOBILE DX-D 100 + WITH WIRELESS DETECTOR OFFERS FAST IMAGING THAT CAN BE VALIDATED

More information

Integrating Modalities to Enterprise PACS through Thinking Systems

Integrating Modalities to Enterprise PACS through Thinking Systems Integrating Modalities to Enterprise PACS through 1 The Need For most enterprise PACS implementations, traditional CT, MR, CR, DR and general Ultrasound are of primary concerns. Not much attention has

More information