1 COMPARISON OF RADIATION DOSE AND IMAGE QUALITY BETWEEN THE SINGLE DETECTOR CT AND MULTI DETECTOR CT ON HEAD EXAMINATION: STUDY PHANTOM Samburi Master of Physics, University of Diponegoro, Indonesia Agency of Testing For Health Facilities (BPFK Jakarta), Ministry Of Health, Indonesia Wahyu Setia Budi Department of Physics, Faculty of Science and Mathematics, University Diponegoro, Indonesia Eko Hidayanto Department of Physics, Faculty of Science and Mathematics, University Diponegoro, Indonesia Manuscript History Number: IJIRAE/RS/Vol.04/Issue04/APAE10105 Received: 19, March 2017 Final Correction: 30, May 2017 Final Accepted: 15, August 2017 Published: August 2017 Citation: Samburi; Wahyu, S. B. & Hidayanto, E. (2017), 'Comparison of Radiation Dose and Image Quality Between the Single Detector CT and Multi Detector CT on Head Examination: Study Phantom, in IJIRAE Journal. Editor: Dr.A.Arul L.S, Chief Editor, IJIRAE, AM Publications, India Copyright: 2017 This is an open access article distributed under the terms of the Creative Commons Attribution License, Which Permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Abstract Research has been done by comparing radiation dose profile and image quality to single detector CT (SDCT) and multi detector CT (MDCT). The method of measuring dose accuracy and linearity of the output dose is done by placing a dose pencil on the iso center position in the middle of the gantry. The scanning parameters use axial mode with the acquisition factor of 120 kv, mas, 1 sec rotation time and 5-10 mm slice thickness. Head Phantom CTDI diameter of 16 cm is used for CT dose index (CTDI) measurement. These measurements use a routine head examination protocol provided by CT Scan. Evaluation of image quality using fantom Gammex ACR 464. The resulting data is normalized to obtain constant image noise between SDCT and MDCT. SDCT resulted in a linearity coefficient of radiation dose output ranging from to higher than MDCT ranging from to The accuracy of DOSIS radiation at 120 kv / 100 mas from SDCT is 23.3 mgy and 31.2 mgy whereas MDCT ranges from mgy. The standard deviation of radiation dose distribution pattern with CTDI head phantom ranged from 1.7 to 5.8. While the results of image quality measurement, linearity response to 5 standard materials indicate that they are within the range of HU reference values. Evaluation of homogeneity of CT number accuracy, uniformity of noise and uniformity of CT,center and CT,edge respectively are in the range -7 and + 7 HU, 0,258 and 1,123 HU and 0,93-3,85 HU. Contrast to Noise Ratio (CNR) measurements range from In addition, the results of spatial resolution measurements range from 6-7 lp / cm. From the overall evaluation shows, that MDCT has a better score than a SDCT. However, when compared with standard reference values, a single CT detector can be retained as an imaging modality to support the patient's diagnostic process. Keywords SDCT, MDCT, Linearitas Dosis, CTDI, Kualitas Citra. IJIRAE , All Rights Reserved Page -46

2 I. INTRODUCTION Total use of CT scans in Indonesia continues to experience a significant increase from year to year. From about 7300-an X-ray machine data are registered in the Nuclear Energy Regulatory Agency (BAPETEN) in 2015, about 5-6% of that amount is the modality of CT Scan with various brands and types. From these data the amount of approximately 10% is a machine with a single detector CT (single slice CT Scan), which is still used in the examination of patients. While the rest are multi-slice CT with a variation of the dual slice to slice 128 that began circulating. Slowly and surely conventional CT displaced population and is replaced with the latest generation in line with technological developments. One problem that still complained of by the physicists is that the use of CT contributing to provide 35-45% of the dose of radiation of all existing diagnostic radiology modalities [1]. Results of research on comparative dose on conventional CT and spiral air by Bouzarjomehri F., et al, 2006, indicating the patient's radiation dose using conventional CT examination is higher than spiral CT [2]. The amount of radiation dose in CT scan parameters determined by parameters such as tube voltage (kv), tube current production - timer (mas), slice thickness (mm), rotation time and pitch. The relationship between the dose of radiation to the current production of tube-timer linear. The higher mas then the higher the output radiation from CT scans. MDCT development to further enhance the ability of CT to improve image quality and reduce radiation dose. [3]. Among the use of CT scans, CT examination was the dominant case in every hospital / clinic. More than 50% CT scan is a head examination [Lisa W]. The study of image quality and dose eyepiece using a head examination protocol with helical and sequential modes was performed by Abdeen et al (2010). The results showed that the relative helix mode resulted in smaller doses than sequential, while the image quality was relatively similar [Abdeen]. Another study in 2013, comparing different CT Scan engines to recognize some types of CT that have the lowest dose of radiation and diagnostic image quality adequate for diagnostic purposes. The study used human observers to evaluate image quality and record radiation doses from selected hospitals [Haytham Ahmad]. In line with Haytham (2013) research, we conducted the same research using a set of measuring instruments, fantom CTDI head and fantom image quality. The goal is to compare the radiation dose and image quality between a single CT detector and a multi-detector to obtain valid data about the advantages and disadvantages of both. II. METHODS A. SURVEY STUDY AND INSTRUMENTATION During the period of March 2016 s.d March 2017, we conducted a research survey of 6 CT Scans consisting of 2 types of CT single detectors and 4 Multi Slices of four different hospital radiology facilities. The six CT Scan aircraft are shown in table 1 below. Table I - Characteristics of CT model Scan survey research. Scanner Type Manufactur Model Slice Class Production A HITACHI PRATICO B TOSHIBA CXXG-010A C PHILIPS MX 4000 DUAL D TOSHIBA AQUILION E HITACHI ROBUSTA F PHILIPS MX-16 SLICE The equipment used for this study consisted of a fantom image quality accreditation, Gammex ACR 464, RTI type piranha 657 and a fantom dose CT head index (figure.1). Figure 1. The equipment used in research IJIRAE , All Rights Reserved Page -47

3 B. MEASUREMENT PARAMETER. Linearity dose calculated with linearity coefficient (CL), aims to determine the consistency of performance response of certain significant components of the x-ray plane when given different inputs. The linearity coefficient is formulated as [7]: CTDI describe the sum of all contributions dose along a line parallel to the axis of rotation of the scanner (ie the z- axis) is formulated as[8]: D (z) is the value of the dose at a given location along the z axis, N is the number of images and T is the nominal slice thickness. While the present CTDI 100 dose accumulation of multiple scans in the middle of 100 mm along the detector and does not calculate the accumulated dose of scan length exceeding 100 mm [8]. The relationship between CTDI 100 with CTDI w formulated as : CTDI 100,p showed the average doses edge and CTDI 100, c is the result of the measurement at the center. CTDI 100 relationship with CTDI V: the quantity p is called the pitch factor and is defined as the ratio of the total width of the beam (N.T) divided by increments of movement of the patient table per rotation of the tube [8]. In CT Scan some determinants of image quality characteristics are CT Number accuracy, homogeneity and contrast image resolution. According to the reference of the imagery imagery used, for CT number accuracy evaluation, CT HU reference values for polyethylene (- 95 HU), equivalent water material (0 HU), acrylic (+120 HU) and air (-1000 HU) [ACR CT ]. (Figure 2). Figure 2: CT number linearity response image. To evaluate the homogeneity of the image, the acceptance criteria mean CT number difference between the center and the edge is less than 5 HU fourth edge position. Value CT number in the middle between -7 and +7 HU HU HU selected with ± 5 [9]. IJIRAE , All Rights Reserved Page -48

4 Figure 3: Image homogeneity of CT HU. To quantify low contrast resolution is to calculate the value of contrast-to-noise ratio (CNR) is formulated with Figure 4: Low Contrast Image Resolution A Rod Where is the ROI on nodule diameter of 50 cm, and Rod B is the ROI on the left of Rod A. SD is the standard deviation value of Rod B. criterion noise level contrast ratio allowable absolute value of contrast noise ratio in adult abdominal examination protocol and adult head 1.0. Figure 5: High Contrast Image Resolusi. To assess the spatial resolution of the CT image of the scan, window width and window level (contrast and brightness) is set at 100 WW and 1100 WL. Be visually seen eight bar pattern composed resolution 4, 5, 6, 7, 8, 9, 10 and 12 lp / cm, each measuring 15 mm x 15 mm square. The acceptance criteria for the spatial resolution is a scanner capable of detecting a minimum of 6 lp / cm. IJIRAE , All Rights Reserved Page -49

5 C. Measurement Technique. Multimeter x-ray RTI piranha type 657 and pencil dose detector used for measurement accuracy and linearity of the output radiation and a phantom head CT dose index of 16 cm diameter is used for the measurement of CTDI head. For measurement of dose linearity and accuracy, the position detector is placed on the holder in the free state in the air on a gantry isocenter. First do a scan of the entire detector to obtain a projection image detector positions. Furthermore, the scanning fixed by setting 8-10 mm thick slices of the image against a predetermined area of the detector. Used tube voltage of 120 kv was fixed for the fourth variation CT with tube current of 25 ma to 300 ma. Measurements are converted to dose per rotation and CTDI in air. All measurements are normalized to 100 mas CTDIair and performed with axial mode. From this measurement process resulting output value of radiation dose at 120 kv per 100 mas and the coefficient of linearity of the output radiation. For the measurement of CT dose index used multimeter x-ray detector and a pencil and CTDI phantom head. Parameter measurements refer to the user guide of the manufacturer or using measurement information from the monitor console by using the mode of examination of the head. Measurements performed 5 times from middle position, top, right, bottom and left. The measurement results in the form of exposure dose (CTDI) is converted into a number of CT dose descriptors. Phantom image quality accreditation, Gammex ACR 464, used for evaluation of image quality assessment. Parameters assessed restricted to the measurement results CT number, low contrast resolution, homogeneity and high contrast spatial resolution. III. RESULTS AND DISCUSSION A. The output accuracy and linearity of the output radiation Point dose measurements at the free in air position detector are expressed in dosage form in units of mgy / 100mAs and linearity coefficient of radiation output. The used tube voltage is 120 kv. The measurement results are shown by Table 2. The measurement data shows that the average of the dosage in the air varies greatly, whereas for the linearity coefficient dose in air type A and type B which is the representation of CT single detector is between 0,017-0,0085 and type C, D, E and F are repren- sations of CT Multi Slices ranging from to However, the overall linearity coefficient scanner value is still below the threshold (CL Threshold 0.1). Scanner Type Table II. LINEARITY DOSE MEASUREMENT DETECTOR WITH FLOATING IN THE AIR. Tube Voltage (kv) acquisition parameter Range of Tube Current (ma) Rotation Time (sec) Slice Thicknes (mm) Average (mgy/100mas) Coeffisien of Linearity A , ,77 0,085 B , ,60 0,017 C ,0 5 11,17 0,024 D , ,57 0,018 E , ,82 0,055 F ,0 5 13,31 0,007 B. CTDI and dose distribution pattern. Measurements of CT dose index (CTDI) with the head diameter of fantom are shown in table 3 and histogram 6. The radiation dose pattern of CT Scan at 5 detector positions is shown in 6-histogram. The magnitude of the dose is influenced by various acquisition parameters, such as tube voltage (kv), tube current (ma), rotation time (sec), pitch, slice thickness and collimation options with reference to manufacturer-specified acquisition parameters, each shown on the monitor Console except for single-slice CT scanners. In general, the multi-slice CT plane on the console has provided an estimate of the main CTDI values for CTDI vol and DLP descriptors. This value is at least the physicist's reference to the radiation dose received by the patient. IJIRAE , All Rights Reserved Page -50

6 Table III CTDI DISTRIBUTION PATTERNS IN 5 POSITION DETECTORS IN MEASURING HEAD WITH FANTOM CTDI. Scanner Type The results of measurements dose (mgy) Center Top edge Right edge Bottom edge Left edge Standard Deviation A 29,94 34,16 31,81 23,43 28,85 4,61 B 44,70 49,34 45,85 41,92 43,93 3,07 C 33,12 30,37 34,39 31,35 28,93 1,79 D 25,07 39,2 34,03 32,3 36,26 5,84 E 25,92 27,49 24,64 23,59 24,87 1,68 F 40,29 40,56 40,10 35,86 38,23 2,24 Figure 1. Graph CTDI distribution pattern of four types of scanners at the 5 position of the detector with CTDI phantom head. C. Comparison of linearity response accuracy The measurement data of the CT number of the six scanners is shown in table 4, the slope curve is shown in table 5. The value of CT number is obtained from the measurement of the region of interest (ROI) to five standard materials that have certain CT HU reference value (table 4). The electron density of the five reference materials is shown in Table 4. Table IV - MEASUREMENT OF THE VALUE OF CT HU 4 TYPE SCANNER FOR RESPONSE ASSESSMENT LINEARITY. Material Electron Measured Value (CT HU) HU Density A B C D E F Reference Polyethelene 4-94, ,57-94,34-98,8-96,54-95 Bone , ,1 958, ,1 967, Air , ,96-971,26-990,2-976, Acrylic , ,96 120,17 120,1 119, Water ,26 0-2,69 1,26 2,5-0,46 0 Tabel V - LINEAR REGRESSION OF CT-HU MEASUREMENT RESULTS Tipe Scanner Nilai CT-HU Scanner Regresi Linear Polyethelene Bone Air Acrylic Water (R) A -94,34 958,53-971,26 120,17 1,26 0,996 B -98,19 962,12-996,09 122,23 2,77 0,997 C -93,57 818,1-1009,96 127,96-2,69 0,999 D -94,34 958,53-971,26 120,17 1,26 0,996 E -98,8 1110,1-990,2 120,1 2,5 0,991 F -96,54 967,16-976,75 119,59-0,46 0,996 IJIRAE , All Rights Reserved Page -51

7 Phantom scanning uses axial mode or sequential routine head examination, with a 120 kv, mas acquisition factor and 5-10 cm thick wedges. The CT-HU values generated by each scanner were compared with the reference HU and the linear regression value (R) was calculated. Limit of linearity response tolerance is expressed with value R 0,99. D. Comparison of Homogeneity CT-HU. The comparison of the images homogeneity is shown in Table 6. The data were extracted from a scan of the fantom that has a homogeneous material equivalent to water. The value of CT number is obtained from the measurement of the region of interest (ROI) on the five position of the scanned image. Phantom scanning also uses axial mode or sequential routine head checking, with a 120 kv, mas acquisition factor and 5-10 cm thick wedges. To draw closer to the same measurement parameters all scanners of homogeneity measurements were converted to acquisition parameters of 120 kv, 300 mas and 8 mm thickness of the slice. The results of the evaluation of the measurements of CT summation are shown in Table 5. From the table, it can be concluded that the CT scores of the four types of scanners are included in the range (- 7 HU and + 7 HU). Similarly, for the uniformity parameters of noise and maximum deviation of CT HU centers and edges entered in the acceptance criteria of the test (-5 CT HU and +5 CT HU). Scanner Type A B C D E F Table VI- ACCURACY AND UNIFORMITY CT NUMBER OF WATER EQUIVALENT MATERIAL. Parameter ROI Potition (CT HU) Top Edge Right Edge Left Edge Bottom Edge Center Mean ROI 0,46 0,78-0,10 0,25-0,15 Noise 3,63 3,23 3,34 3,61 3,20 Deviation Mean ROI 0,46 0,78 0,10 0,25 Noise converted 0,45 0,76-0,10 0,24 Mean ROI 3,26 3,82 3,28 2,92 3,61 Noise 4,08 4,02 3,95 3,95 4,02 Deviation Mean ROI 3,26 3,82 3,28 2,92 Noise converted 3,16 3,70 3,18 2,83 Mean ROI -0,49-4,12-4,29-2,65-4,14 Noise 3,70 3,27 3,07 3,07 3,13 Deviation Mean ROI 0,49 4,12 4,29 2,65 Noise converted -0,45-3,80-3,96-2,44 Mean ROI -0,27-1,70-1,64-1,34-1,57 Noise 5,44 3,68 3,78 3,76 3,50 Deviation Mean ROI 0,27 1,70 1,64 1,34 Noise converted -0,26-1,65-1,59-1,30 Mean ROI 1,60 2,50 2,30 1,60 2,20 Noise 3,55 2,79 2,91 2,81 2,83 Deviation Mean ROI 1,60 2,50 2,30 1,60 Noise converted 1,55 2,42 2,23 1,55 Mean ROI -3,40-3,20-2,90-2,70-3,00 Noise 2,90 2,60 2,40 2,40 2,30 Deviation Mean ROI 3,40 3,20 2,90 2,70 Noise converted -3,04-2,86-2,59-2,41 Table VII- EVALUATION OF MEASUREMENT ACCURACY CT NUMBER OF MATERIAL AND WATER EQUIVALENT NUMBER UNIFORMITY CT. Parameter Evaluation Measured Values A B C D E F Acceptance Criteria Accuracy of CT Number (water) 0,76-4,29 0,45-1,65 2,42 3,04-7 s.d +7 Noise Uniformity 0,85 0,35 3,50 1,38 0,87 0,63-5 s.d +5 Maximum Deviation (CT HU center CT HU edge) 0,78 3,28 4,29 1,70 2,50 3,40-5 s.d +5 IJIRAE , All Rights Reserved Page -52

8 E. Comparison of Low Contrast Resolution The contrast resolution can be defined as the ability of a CT-scanner to distinguish relatively large objects that differ only slightly in the density of the background. To avoid subjectivity in the assessment of low-contrast images, the quantity of contrast to noise ratio (CNR) is used. The CNR value limit for adult head examination is 1.0. The result of measurement CNR value is shown in table 8. From the table it can be concluded all scanners meet the acceptance criteria. Table VIII - CONTRAST TO NOISE RATIO VALUE. Scanner System ROI Target background ROI background noise CNR measured A 93,5 87,4 3,7 1,7 B 93,5 87,2 4,1 1,5 C 96,4 88,7 3,7 2,1 D 85,6 78,6 3,7 1,9 E 86,3 82,5 3,7 1,0 F 93,1 86,9 3,1 2,0 Acceptance Criteria 1,0 F. Comparison of Hight Contrast Resolution In this test verified the ability of scanners to detect small lesions that have very different attenuation values than the surrounding tissue. This characterizes the ability of the imaging system to distinguish between two very small objects in adjacent positions. Spatial resolution measurements are performed with objects that have high contrast from a uniform background. The criteria for acceptance of high-contrast spatial resolution of CT Scan according to AAPM with the image imagery of Gammex 464 ACR is 6 lp / cm. From the results of the test table 9, it can be concluded that all CT scans meet the acceptance of the same or above the acceptability limit. Table VIII - MEASUREMENT OF SPATIAL RESOLUTION Acquisition factors Scanner Type Spatial Resolution (lp/cm) kv mas slice A B C D ,5 5 7 E ,5 6 F Acceptance Criteria 6 lp/cm Scanner Type Spatial Resolution (lp/cm) Acceptance Criteria CT Scan A 6 CT Scan B 7 CT Scan C 6 6 (lp /cm) CT Scan D 6 IV. CONCLUSIONS The result of comparison of radiation dose and image quality of the six CT types shows that all CT Scan performance is in accordance with standard criteria. Quantitatively there is no significant difference between SDCT and MDCT. A number of other variables that affect the reliability level of CT Scanner is maintenance program, production year, frequency of use and brand of tool manufacturer. If performance grading of all six CT Scan types is seen, it is relatively superior in terms of radiation dose and image quality compared to SDCT. However, if SDCT is compared to the threshold value of passing the test of internationally accepted standards, SDCT can still be maintained as imaging modalities to support the patient's diagnostic process. In the measurement of radiation doses at 120 kvp / 100mAs, 2 types of SDCT scanners had mgy and mgy, 2 MDCT scanners scored mgy and mgy and the other two MDCT types were mgy and 13, 31 mgy. IJIRAE , All Rights Reserved Page -53

9 The CT Scan test method from BAPETEN provides dose guidance at 120 kvp / 100mAs ranging from mgy [7]. The dose range is intended to ensure adequate quality in obtaining diagnostic image quality. If you look at the measurement results, MDCT with 120 kv / 100mAs resulting in dose below that range with adequate image quality, it is necessary to propose amendment of the guidelines. The linearity coefficient of this measurement ranges from , below the guideline level threshold (CL 0.1). The standard deviation of the radiation radiation pattern at 5 detector positions shows varying values. To some extent, the higher standard deviation of the radiation distribution pattern will have an impact on the homogeneity of the image. In the image quality measurement shows all types of CT Scan meet the acceptance criteria test. The linearity response ratio shows the slope curve (R) of all CT scans at the threshold ( 0.99). Evaluation of measurement accuracy of CT number, noise uniformity and maximum deviation between edge and center HU CT is in tolerance range (± 5 HU). Low-contrast resolution is shown in CNR quantification. CNR measurement results are in the acceptance range ( 1.0). Similarly, the results of high contrast resolution measurements of all CT scans are within tolerable limits (6 lp / cm) REFERENCES 1. H. Imhof, et al., 2002, Spiral CT and radiation dose, European Journal of Radiology 47 (2003) 29_/ F.Bouzarjomehri1, et. al., 2006, Conventional and spiral CT dose indices in Yazd general hospitals, Iran, Iran. J. Radiat. Res., 2006; 3 (4): 3. S.T. Schindera, et. al, 2012 : Effect of automatic tube voltage selection on image quality and radiation dose in abdominal CT angiography of various body sizes: A phantom study, Clinical Radiology xxx (2013) e79ee Lisa Williams et al, 2005, Computed tomography of the head: An experimental study to investigate the effectiveness of lead shielding during three scanning protocols, The College of Radiographers, doi: /j.radi Abdeen et.al, 2010, Comparison of image quality and lens dose in helical and sequentially acquired head CT, Clinical Radiology 65 (2010) 868e873N. 6. Haytham Ahmad AL Ewaidat, Studies of CT Dose and Image Quality Using Clinical and Phantom Images, A thesis presented to Charles Sturt University in fulfillment of the requirements for the degree of Doctor of Philosophy, Metode Uji Pesawat Sinar-X CT Scan, KU/PD/DKKN/07, BAPETEN, AAPM REPORT No. 96, The Measurement, Reporting, and Management of Radiation Dose in CT Report of AAPM Task Group 23: CT Dosimetry Diagnostic Imaging Council CT Committee, American College of Radiology CT Accreditation Program Testing Instructions - Revised, January 6, Robert L. Dixon & John Boone, 2010 The CTDI Paradigm: A Practical Explanation for Medical Physicists, American College of Radiology Bushberg JT, Siebert JA, Leidholdt EM, et al: The Essential Physics of Medical Imaging. Baltimore, Lippincott Williams & Wilkins, Huda, W., Ogden, K., & Khorasani, M. (2008). Effect of dose metrics and radiation risk models when optimizing CT x-ray tube voltage. Physics in Medicine and Biology, 53, ImPACT. (2002). Eight and Sixteen Slice CT Scanner Comparison Report (Vol. 7). 14. ImPACT. (2003). Single Slice CT Scanner Comparison Report (Vol. 8). European Commission s Study Group. (1998). 15. ImPACT. (2005). Sixteen slice CT scanner comparison report (Vol. 12). 16. Jaengsri Nuttawan, 2004, CT Protocol, Radiology Departement Of Takshin Hospital, Bangkok Neseth R., 2000, Procedurs and Documentation for CT and MRI, CIC Edizioni Internazionali. 17. Johns H.E and Cunningham J.R The physics of radiology 4th edition. Charles C. Thomas publisher. USA 18. Kalra MK, Maher MM, Toth TL, et al. Strategies for CT radiation dose optimization. Radiology 2004; 230: Lewis, M. (2005). Radiation dose issues in multi-slice CT scanning. ImPACT technology 20. Maria Lewis, 2005, : Radiation Dose Issue in Multi-Slice CT Scanning, ImPACT technology update no Michael, G. (2001). X-ray computed tomography. Physics Education, 36, Papp, Jeffrey, PhD,RT (R) (QM), 2006, Quality Management in The Imajing Sciences, third edition. 23. Seeram, E. (2009). Computed Tomography,Physical Principles, Clinical Application and Quality Control (third ed.). USA. IJIRAE , All Rights Reserved Page -54

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed

More information

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist CT Protocol Review: Practical Tips for the Imaging Physicist Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center August 8, 2013 AAPM Annual Meeting Goals Understand purpose and importance

More information

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography

More information

S. Guru Prasad, Ph.D., DABR

S. Guru Prasad, Ph.D., DABR PURPOSE S. Guru Prasad, Ph.D., DABR Director of Medical Physics IAEA Consultant NorthShore University Health System and University of Chicago, Pritzker School of Medicine Current TPS utilize more information

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Optimisation of Toshiba Aquilion ONE Volume Imaging

Optimisation of Toshiba Aquilion ONE Volume Imaging Optimisation of Toshiba Aquilion ONE Volume Imaging Jane Edwards, RPRSG Royal Free London NHS Foundation Trust Dr Mufudzi Maviki, Plymouth Hospitals NHS Trust Background In 2011/12 Radiology at RFH was

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

ImPACT. CT dosimetry and a data base for CTDI values. EFOMP Workshop at ECR Imaging Performance Assessment of CT Scanners

ImPACT. CT dosimetry and a data base for CTDI values. EFOMP Workshop at ECR Imaging Performance Assessment of CT Scanners ImPACT Imaging Performance Assessment of CT Scanners A Medical Devices Agency Evaluation Group CT dosimetry and a data base for CTDI values EFOMP Workshop at ECR 2002 S. Edyvean, ImPACT St George s Hospital,

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing ImPACT Information Leaflet No. 1: CT Scanner Acceptance Testing Version 1.02, 18/05/01 CONTENTS: 1. SCOPE OF LEAFLET 2. GENERAL PRINCIPLES OF ACCEPTANCE AND COMMISSIONING 2.1 PHANTOMS 2.2 EXPOSURE AND

More information

AAPM Standard of Practice: CT Protocol Review Physicist

AAPM Standard of Practice: CT Protocol Review Physicist AAPM Standard of Practice: CT Protocol Review Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center September 11, 2014 2014 Texas Radiation Regulatory Conference Goals Understand purpose

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60601-2-44 2001 AMENDMENT 1 2002-09 Amendment 1 Medical electrical equipment Part 2-44: Particular requirements for the safety of X-ray equipment for computed tomography Amendement

More information

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT.

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT. Progress report: Development of assessment and predictive metrics for quantitative imaging in chest CT Subaward No: HHSN6801000050C (4a) PI: Ehsan Samei Reporting Period: month 1-18 Deliverables: 1. Deployment

More information

A closer look at CT scanning

A closer look at CT scanning Vet Times The website for the veterinary profession https://www.vettimes.co.uk A closer look at CT scanning Author : Charissa Lee, Natalie Webster Categories : General, Vets Date : April 3, 2017 A basic

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

Computer-Tomography II: Image reconstruction and applications

Computer-Tomography II: Image reconstruction and applications Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de

More information

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare

More information

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead.

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead. 1 CT Basics Principles of Spiral CT Dose 2 Who invented CT? 1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections Sir Godfrey Hounsfield worked for the Central

More information

Limits of Dose Reduction in CT: How low is too low? Acknowledgement. Disclosures 8/2/2012

Limits of Dose Reduction in CT: How low is too low? Acknowledgement. Disclosures 8/2/2012 8/2/22 Limits of Dose Reduction in CT: How low is too low? Ehsan Samei Duke University Acknowledgement Disclosures Research grant: NIH Research grant: DHS Research grant: RSNA, QIBA Research grant: General

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Tecniques of imaging in Overview Introduction Tecniques of imaging

More information

AIDR 3D Iterative Reconstruction:

AIDR 3D Iterative Reconstruction: Iterative Reconstruction: Integrated, Automated and Adaptive Dose Reduction Erin Angel, PhD Manager, Clinical Sciences, CT Canon Medical Systems USA Iterative Reconstruction 1 Since the introduction of

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

CT vs. VolumeScope: image quality and dose comparison

CT vs. VolumeScope: image quality and dose comparison CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,

More information

GPU implementation for rapid iterative image reconstruction algorithm

GPU implementation for rapid iterative image reconstruction algorithm GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial

More information

Equipment Specification

Equipment Specification MULTISLICE CT SCANNER ( 16 Slices ) Merk : Hitachi Japan Model : SUPRIA 5 MHU Price : Rp 6.512.360.215,27 No Equipment Specification 1 2 3 Scanner Gantry Object for scanning : Whole body including head

More information

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

C a t p h a n / T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 Internationally recognized for measuring the maximum obtainable performance of axial, spiral and multi-slice

More information

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid by Yingxuan Chen Graduate Program in Medical Physics Duke University Date: Approved: Lei Ren, Supervisor Fang Fang Yin, Chair

More information

CT: Physics Principles & Equipment Design

CT: Physics Principles & Equipment Design CT: Physics Principles & Equipment Design James Kofler, Ph.D Radiology Mayo Clinic Rochester, MN June 27, 2012 Disclosures Nothing to disclose Learning Objectives Understand fundamental concepts of - CT

More information

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT More Info at Open Access Database www.ndt.net/?id=18754 Novel evaluation method of low contrast resolution performance of dimensional X-ray CT Makoto Abe 1, Hiroyuki Fujimoto 1, Osamu Sato 1, Kazuya Matsuzaki

More information

Tomographic Reconstruction

Tomographic Reconstruction Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,

More information

1970 Projection Radiography 2D projection of 3D anatomy

1970 Projection Radiography 2D projection of 3D anatomy Speakers: L. N. Rothenberg, Ph.D. Computed Tomography G. D. Clarke, Ph.D. Magnetic Resonance Imaging J. A. Zagzebski, Ph.D. Ultrasonic Imaging August 1, 2012 Lawrence N. Rothenberg, Ph.D. Keith S. Pentlow,

More information

Micro-CT Methodology Hasan Alsaid, PhD

Micro-CT Methodology Hasan Alsaid, PhD Micro-CT Methodology Hasan Alsaid, PhD Preclinical & Translational Imaging LAS, PTS, GlaxoSmithKline 20 April 2015 Provide basic understanding of technical aspects of the micro-ct Statement: All procedures

More information

RADIOLOGY AND DIAGNOSTIC IMAGING

RADIOLOGY AND DIAGNOSTIC IMAGING Day 2 part 2 RADIOLOGY AND DIAGNOSTIC IMAGING Dr hab. Zbigniew Serafin, MD, PhD serafin@cm.umk.pl 2 3 4 5 CT technique CT technique 6 CT system Kanal K: RSNA/AAPM web module: CT Systems & CT Image Quality

More information

Lorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc.

Lorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc. Lorad FFDM QC Procedures for Medical Physicist Tao Wu, Ph.D. Hologic, Inc. Lorad Selenia FFDM Tests Following Lorad QC Manual Collimation Assessment Artifact Evaluation System Resolution Phantom Image

More information

Kevin O Donnell Senior R&D Manager Toshiba Medical Research Institute Co-Chair, DICOM Standards Committee Past Chair, IHE Radiology Planning

Kevin O Donnell Senior R&D Manager Toshiba Medical Research Institute Co-Chair, DICOM Standards Committee Past Chair, IHE Radiology Planning 1 Kevin O Donnell Senior R&D Manager Toshiba Medical Research Institute Co-Chair, DICOM Standards Committee Past Chair, IHE Radiology Planning Committee Standards & Tools 2 Learning Objectives 1) DICOM

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

DUE to beam polychromacity in CT and the energy dependence

DUE to beam polychromacity in CT and the energy dependence 1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact

More information

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic

More information

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT Assessment of 3D performance metrics D and 3D Metrics of Performance Towards Quality Index: Volumetric imaging systems X-ray based Volumetric imaging systems: CBCT/CT Tomosynthesis Samuel Richard and Ehsan

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

James R Halama, PhD Loyola University Medical Center

James R Halama, PhD Loyola University Medical Center James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives Be familiar with the tests recommended

More information

Comparison of the Imaging Performance of CT Scanners, Issue 11

Comparison of the Imaging Performance of CT Scanners, Issue 11 Imaging Performance Assessment of CT Scanners Comparison of the Imaging Performance of CT Scanners, Issue 11 Comparison data is presented on the following scanners : Elscint Twin Flash IGE HiSpeed Advantage

More information

Preparing for Part 2 of the ABR Diagnostic Physics Exam

Preparing for Part 2 of the ABR Diagnostic Physics Exam Preparing for Part 2 of the ABR Diagnostic Physics Exam Joseph Zambelli, Ph.D. Spectrum Health, Grand Rapids, MI Outline Exam Background Exam Preparation Suggested Resources Location, Time, and Format

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

Annexure XII SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER

Annexure XII SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER Annexure XII SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER A) Scanner Design X-Ray generator and tube: 1. Scanner: Whole body spiral CT scanner (16 slices) of latest technology.

More information

CURRICULUM COMMITTEE MEETING Friday, March 18, :00 p.m. Student Life Center, Faculty Dining Room (Building 23, First Floor) AGENDA

CURRICULUM COMMITTEE MEETING Friday, March 18, :00 p.m. Student Life Center, Faculty Dining Room (Building 23, First Floor) AGENDA CURRICULUM COMMITTEE MEETING Friday, March 18, 2016-2:00 p.m. Student Life Center, Faculty Dining Room (Building 23, First Floor) I. Call to Order AGENDA II. Roll Call III. Minutes of meeting of January

More information

Dose-volume product (DVP) as descriptor for estimating total energy imparted to patient undergoing CT examination

Dose-volume product (DVP) as descriptor for estimating total energy imparted to patient undergoing CT examination DIAGNOSTIC AND INTERVENTIONAL IMAGING Journal of Medical Physics and Biophysics, Vol. 3, No. 1, December 2016 Dose-volume product (DVP) as descriptor for estimating total energy imparted to patient undergoing

More information

Dose-Volume Product (DVP) as descriptor for estimating total energy imparted to patient undergoing CT examination

Dose-Volume Product (DVP) as descriptor for estimating total energy imparted to patient undergoing CT examination DIAGNOSTIC AND INTERVENTIONAL IMAGING Journal of Medical Physics and Biophysics, Vol. 2, No. 1, September 2016 Dose-Volume Product (DVP) as descriptor for estimating total energy imparted to patient undergoing

More information

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Stefan C. Saur 1, Hatem Alkadhi 2, Luca Regazzoni 1, Simon Eugster 1, Gábor Székely 1, Philippe Cattin 1,3 1 Computer Vision

More information

Enhancement Image Quality of CT Using Single Slice Spiral Technique

Enhancement Image Quality of CT Using Single Slice Spiral Technique Enhancement Image Quality of CT Using Single Slice Spiral Technique Doaa. N. Al Sheack 1 and Dr.Mohammed H. Ali Al Hayani 2 1 2 Electronic and Communications Engineering Department College of Engineering,

More information

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives 1. Be familiar with recommendations

More information

ACMP 25th Annual Meeting

ACMP 25th Annual Meeting Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation ACMP 25th Annual Meeting Seattle, WA May 3, 2008 Melissa C. Martin, M.S., FACR Therapy Physics, Inc. 879 West 190th St., Ste 419,

More information

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA Computed Tomography Principles, Design, Artifacts, and Recent Advances THIRD EDITION Jiang Hsieh SPIE PRESS Bellingham, Washington USA Table of Contents Preface Nomenclature and Abbreviations xi xv 1 Introduction

More information

Quality control phantoms and protocol for a tomography system

Quality control phantoms and protocol for a tomography system Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

CT Dose Profiler. Probe for evaluation of CT systems. CT Dose Profiler User's Manual - English - Version 6.4A. RTI article number:

CT Dose Profiler. Probe for evaluation of CT systems. CT Dose Profiler User's Manual - English - Version 6.4A. RTI article number: CT Dose Profiler Probe for evaluation of CT systems - English - Version 6.4A RTI article number: 9630512-00 CT Dose Profiler The CT Dose Profiler probe makes it possible to evaluate the performance of

More information

Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc.

Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc. Unique Features of the GE Senoclaire Tomosynthesis System Tyler Fisher, M.S., DABR Therapy Physics, Inc. Conflict of Interest Disclosure I have no conflicts to disclose. Learning Objectives Overview of

More information

Application of MCNP Code in Shielding Design for Radioactive Sources

Application of MCNP Code in Shielding Design for Radioactive Sources Application of MCNP Code in Shielding Design for Radioactive Sources Ibrahim A. Alrammah Abstract This paper presents three tasks: Task 1 explores: the detected number of as a function of polythene moderator

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

A study of densitometry comparison among three radiographic processing solutions

A study of densitometry comparison among three radiographic processing solutions Iran. J. Radiat. Res., 2006; 4 (2): 81-86 A study of densitometry comparison among three radiographic processing solutions V. Changizi 1*, E. Jazayeri 1,A.Talaeepour 2 1 Department of Radiology Technology,

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

RITtrend allows you to effectively manage all your physics QA data in one powerful and customizable package.

RITtrend allows you to effectively manage all your physics QA data in one powerful and customizable package. So much is asked of medical physicists these days We Can Help Automated Radiation Therapy Phantom Analysis in Seconds QC for Therapy OBI s and CT simulators. Radia s Catphan/OBI module performs analysis

More information

DETERMINISTIC 3D RADIATION TRANSPORT SIMULATION FOR DOSE DISTRIBUTION AND ORGAN DOSE EVALUATION IN DIAGNOSTIC CT

DETERMINISTIC 3D RADIATION TRANSPORT SIMULATION FOR DOSE DISTRIBUTION AND ORGAN DOSE EVALUATION IN DIAGNOSTIC CT DETERMINISTIC 3D RADIATION TRANSPORT SIMULATION FOR DOSE DISTRIBUTION AND ORGAN DOSE EVALUATION IN DIAGNOSTIC CT Monica Ghita,, Glenn Sjoden, Manuel Arreola, Ahmad Al-Basheer Basheer, Choonsik Lee, Wesley

More information

Philips SPECT/CT Systems

Philips SPECT/CT Systems Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET

More information

Agenda : Lung Density Breakout Session

Agenda : Lung Density Breakout Session Agenda : Lung Density Breakout Session 1. : Mathew Fuld and Bernice Hoppel 2. Automatic Exposure Control (AEC) Evaluation : Sean Fain Round 4 Project 3. Dose reduction effects on emphysema metrics : Philip

More information

UNCOMPROMISING QUALITY

UNCOMPROMISING QUALITY ION CHAMBERS UNCOMPROMISING QUALITY Designed with over 30 years of scientific integrity for a broad range of dosimetry measurements in diverse radiation beams Farmer-type Chambers For absolute dosimetry

More information

Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners

Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners Poster No.: C-1748 Congress: ECR 2012 Type: Educational Exhibit Authors: S. Miyashita

More information

PURE. ViSION Edition PET/CT. Patient Comfort Put First.

PURE. ViSION Edition PET/CT. Patient Comfort Put First. PURE ViSION Edition PET/CT Patient Comfort Put First. 2 System features that put patient comfort and safety first. Oncology patients deserve the highest levels of safety and comfort during scans. Our Celesteion

More information

Fundamentals of CT imaging

Fundamentals of CT imaging SECTION 1 Fundamentals of CT imaging I History In the early 1970s Sir Godfrey Hounsfield s research produced the first clinically useful CT scans. Original scanners took approximately 6 minutes to perform

More information

Basic Radiation Oncology Physics

Basic Radiation Oncology Physics Basic Radiation Oncology Physics T. Ganesh, Ph.D., DABR Chief Medical Physicist Fortis Memorial Research Institute Gurgaon Acknowledgment: I gratefully acknowledge the IAEA resources of teaching slides

More information

Measurement of Skin Dose

Measurement of Skin Dose Measurement of Skin Dose Sources of Uncertainty Kenneth A. Fetterly, Ph.D. William Pavlicek, Ph.D. Dan Bednarek, PhD 2014 AAPM Annual Meeting, Austin Texas 2013 MFMER slide-1 Purpose 1. Present a framework

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image

More information

Small Angle Gamma Ray Scatter: What Is The Impact On Image Quality

Small Angle Gamma Ray Scatter: What Is The Impact On Image Quality ISPUB.COM The Internet Journal of Medical Technology Volume 4 Number 2 Small Angle Gamma Ray Scatter: What Is The Impact On Image Quality G CUrrie, J Wheat Citation G CUrrie, J Wheat.. The Internet Journal

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

Using the ACR CT accreditation phantom for routine image quality assurance on both CT and CBCT imaging systems in a radiotherapy environment

Using the ACR CT accreditation phantom for routine image quality assurance on both CT and CBCT imaging systems in a radiotherapy environment JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 4, 2014 Using the ACR CT accreditation phantom for routine image quality assurance on both CT and CBCT imaging systems in a radiotherapy environment

More information

Applying Hounsfield unit density calibration in SkyScan CT-analyser

Applying Hounsfield unit density calibration in SkyScan CT-analyser 1 Bruker-microCT Method note Applying Hounsfield unit density calibration in SkyScan CT-analyser Hounsfield units (HU) are a standard unit of x-ray CT density, in which air and water are ascribed values

More information

Acknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury

Acknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury A. Sisniega et al. (presented at RSNA 214) High Performance Cone-Beam CT of Acute Traumatic Brain Injury A. Sisniega 1 W. Zbijewski 1, H. Dang 1, J. Xu 1 J. W. Stayman 1, J. Yorkston 2, N. Aygun 3 V. Koliatsos

More information

연구용유방초음파질관리 원광대학병원김혜원

연구용유방초음파질관리 원광대학병원김혜원 연구용유방초음파질관리 원광대학병원김혜원 Why US QC? Quality control (QC) testing of ultrasound scanners is important to verify the proper and consistent operation of these devices. main goal ; quality improvement Guidelines

More information

Medical Image Processing: Image Reconstruction and 3D Renderings

Medical Image Processing: Image Reconstruction and 3D Renderings Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,

More information

Biomedical Imaging. Computed Tomography. Patrícia Figueiredo IST

Biomedical Imaging. Computed Tomography. Patrícia Figueiredo IST Biomedical Imaging Computed Tomography Patrícia Figueiredo IST 2013-2014 Overview Basic principles X ray attenuation projection Slice selection and line projections Projection reconstruction Instrumentation

More information

Fits you like no other

Fits you like no other Fits you like no other BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase in room

More information

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

Improvement of Efficiency and Flexibility in Multi-slice Helical CT J. Shanghai Jiaotong Univ. (Sci.), 2008, 13(4): 408 412 DOI: 10.1007/s12204-008-0408-x Improvement of Efficiency and Flexibility in Multi-slice Helical CT SUN Wen-wu 1 ( ), CHEN Si-ping 2 ( ), ZHUANG Tian-ge

More information

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti Monte Carlo methods in proton beam radiation therapy Harald Paganetti Introduction: Proton Physics Electromagnetic energy loss of protons Distal distribution Dose [%] 120 100 80 60 40 p e p Ionization

More information

The effects of metal on size specific dose estimation (SSDE) in CT : a phantom study

The effects of metal on size specific dose estimation (SSDE) in CT : a phantom study The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2016 The effects of metal on size specific dose estimation (SSDE) in CT : a phantom study Maram M. Alsanea

More information

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4

More information

Recognition and Measurement of Small Defects in ICT Testing

Recognition and Measurement of Small Defects in ICT Testing 19 th World Conference on Non-Destructive Testing 2016 Recognition and Measurement of Small Defects in ICT Testing Guo ZHIMIN, Ni PEIJUN, Zhang WEIGUO, Qi ZICHENG Inner Mongolia Metallic Materials Research

More information

Computer-Tomography I: Principles, History, Technology

Computer-Tomography I: Principles, History, Technology Computer-Tomography I: Principles, History, Technology Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de History

More information

Z-MOTION. Universal Digital Radiographic System Z-MOTION. Control-X Medical CONTROL-X MEDICAL

Z-MOTION. Universal Digital Radiographic System Z-MOTION. Control-X Medical CONTROL-X MEDICAL Control-X Medical Z-MOTION Compact design, low ceiling height requirement Motorized and manual movement capability Wide motion / SID range Best-in-class image quality Flexible connectivity to PACS systems

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

Diagnostic quality of time-averaged ECG-gated CT data

Diagnostic quality of time-averaged ECG-gated CT data Diagnostic quality of time-averaged ECG-gated CT data Almar Klein a, Luuk J. Oostveen b, Marcel J.W. Greuter c, Yvonne Hoogeveen b, Leo J. Schultze Kool b, Cornelis H. Slump a and W. Klaas Jan Renema b

More information

A study on image quality provided by a kilovoltage cone-beam computed tomography

A study on image quality provided by a kilovoltage cone-beam computed tomography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 1, 2013 A study on image quality provided by a kilovoltage cone-beam computed tomography Julia Garayoa a and Pablo Castro Servicio de Radiofísica,

More information

Computed Tomography Imaging: CT Protocol Management. Caveat 8/3/2017

Computed Tomography Imaging: CT Protocol Management. Caveat 8/3/2017 Computed Tomography Imaging: CT Protocol Management Mark P. Supanich, Ph.D., DABR AAPM Annual Meeting 3 rd August, 2017 Slides at: goo.gl/k8n8jf Rush is an academic health system comprising Rush University

More information

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications 4th International Symposium on NDT in Aerospace 2012 - Tu.3.A.1 Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications Malte KURFISS 1, Gerd STRECKENBACH 2 1 YXLON International

More information

Automated Calculation of Water-equivalent Diameter (D W. ) Based on AAPM Task Group 220

Automated Calculation of Water-equivalent Diameter (D W. ) Based on AAPM Task Group 220 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 Automated Calculation of Water-equivalent Diameter (D W ) Based on AAPM Task Group 220 Choirul Anam, 1,2a Freddy Haryanto, 2 Rena

More information