TLS Tunable Light Source. Operation Manual

Size: px
Start display at page:

Download "TLS Tunable Light Source. Operation Manual"

Transcription

1 TLS Tunable Light Source Operation Manual January 16, 2007 General Photonics Corp. Ph: (909) Edison Ave. Fax: (909) Chino, CA USA Document #: GP-UM-TLS Page 1 of 40

2 Table of Contents: Section 1. Specifications:... 3 Section 2. Overview:... 4 Section 3. Feature Description: Optical Features: Electrical Features:... 6 Section 4. Operation Instructions: Unpacking Front Panel Operation Front panel operation troubleshooting Testing and Characterization: Remote control and programming: RS-232 operation RS-232 connection Remote operation and commands RS-232 troubleshooting USB control Procedure for Installing USB VCP driver VCP Uninstall Procedure: TLS-101 USB Commands USB Troubleshooting: LabVIEW control programs Ethernet control GPIB control Section 5. Technical Support Document #: GP-UM-TLS Page 2 of 40

3 Section 1. Specifications: Physical Features: Package Dimensions Standard 2U half-19 rack mount size 14 (L) 8.5 (W) 3.5 (H) Output Fiber Type PM or single mode fiber Fiber Connector Type FC/PC, FC/APC, SC/PC or SC/APC 1 Weight 7 lbs System Characteristics: Output power (at pigtail) /- 0.5 dbm Power deviation over band 0.3 to 0.5 db Frequency Range to196.1 THz (C band, 89channels, 50GHz spacing) Wavelength Range to nm (C Band) Number of Channels 89 channels (50GHz spacing) SMSR 40 to 45 db OSNR 50 to 55 db (0.1nm optical bandwidth) Polarization extinction ratio 20 db Return Loss -50 to -40 db Locked frequency accuracy -2.5 to +2.5 GHz Shuttered output suppression 250 to 1000 MHz Tuning Speed 20 ms Operating Temperature 0 to 70º C Storage Temperature -20 to 65ºC Front panel display 2-line LCD display Power Supply VAC, Hz or VAC, Hz PC Interface RS-232, USB, Ethernet, GPIB Note: 1. Per user specification at the time of ordering. Document #: GP-UM-TLS Page 3 of 40

4 Section 2. Overview: General Photonics TLS-101 is a High performance continuous wave (CW) tunable light source for use in the C-band window covering 1528 to 1563 nm. The TLS-101 is a low cost tunable laser source for daily test and measurement applications. The laser can be step-tuned across the C-band communication channels with a tuning resolution of 50 GHz in less than 20 ms. The internal wavelength locker assures that the wavelengths or frequencies of the output light are exactly on the ITU grid of the DWDM channels. Figure 1 Overview of TLS-101 The TLS-101 has four different operation modes: continuous scan mode, manual scan mode, single wavelength mode, and select wavelength mode,. In continuous scan mode, the output wavelength will continue to tune between the start channel and end channel, according to the user s speed setting. In manual scan mode, the wavelength changes when the user presses the ENTER key. In single wavelength mode, the instrument will output the fixed wavelength set by the user. In select wavelength mode, up to 40 different wavelengths, in any combination, may be stored and played back, from any start to any end wavelength. The TLS-101 comes with USB, RS-232, Ethernet, and GPIB communication ports for easy, flexible interface with computers and other instruments. In particular, the Ethernet port allows the instrument to be controlled remotely, from home, office or a hotel room when on travel. The TLS-101 is an ideal source for WDM component testing, sensor interrogation, optical coherence tomography (OCT), PMD and PDL characterization, and for use in general purpose test and measurement applications. Document #: GP-UM-TLS Page 4 of 40

5 Section 3. Feature Description: 3.1 Optical Features: The TLS-101 has one fiber connector on the front panel for the output optical beam. The output can be SM or PM depending on the option chosen when the instrument was ordered. The adapter is a universal connector interface (UCI), which features a male-type adapter top piece that can be removed for direct access to the ferrule end for routine cleaning and maintenance without removing the entire adapter from the panel. This feature helps avoid high insertion loss, high return loss and measurement instability caused by dirty or contaminated connectors. In addition, the TLS-101 s universal interchangeable adapter allows the end user to switch to ST, SC, or FC connectors without opening the instrument panel. Although the instrument is shipped with a customer specified fiber adapter, other interchangeable inserts can be purchased from General Photonics. For additional information on different input fiber adapter inserts, please contact General Photonics. External fiber connectors should be cleaned using industry standard cleaning methods before connection to the TLS-101. If this procedure is followed before each connection, the instrument s internal connector ferrule should not need regular cleaning. However, high insertion loss or measurement instability that does not improve after cleaning the external connectors may indicate that the instrument s internal connector ferrule needs to be cleaned. The connector ferrule is contained in a universal connector interface consisting of a front piece that connects to the external fiber connector, and a base piece that is mounted on the front panel of the instrument. To clean the connector ferrule, first, make sure that no external connector is connected to the universal connector interface. Then, using a Phillips screwdriver, remove the two small screws connecting the front and back parts of the adapter, and carefully pull the front flange straight out. (Note: never remove the adapter base from the front panel). The ferrule end should now be exposed. Clean the ferrule using standard cleaning procedures (compressed air or a fresh lint-free tissue and alcohol). Care must be taken to avoid scratching the ferrule surface. Finally, replace the front flange (position it so that the key notch faces up, and the small alignment pin lines up with the hole in the base piece, before pushing it in) and the screws. It is generally recommended that the user prepare a patch cord fiber to avoid inside connector wear. Document #: GP-UM-TLS Page 5 of 40

6 3.2 Electrical Features: The TLS-101 system uses the standard wall electricity supply. The line voltage is factoryset at 110 VAC or 220 VAC per customer specification. Due to high voltage, the following safety precautions must be exercised during operation. The ground pin on the power supply cord must be connected to earth ground of the wall receptacle. Never touch the boards inside the package without proper insulation. The TLS-101 is not user serviceable and can be serviced by factory-authorized personnel. The front panel of the TLS-101 is shown in Figure 2. The power switch (Power), light source enable key, liquid crystal display (LCD), push button control pad, and output optical connector are mounted on the front panel. The AC power plug, fuse, and RS-232, USB, Ethernet, and GPIB interface connectors are mounted on the back panel, as shown in Figure 3. The TLS-101 includes RS-232, USB, Ethernet, and GPIB interfaces for external computer operation of the system. RS-232 and USB cables are provided for connection to a personal computer. A basic LabVIEW (National Instruments, control program is also provided. The remote control command list and instructions, USB driver installation instructions, and LabVIEW control program descriptions are listed in section 4.5. Document #: GP-UM-TLS Page 6 of 40

7 Figure 2. Front panel layout. Front panel description: LCD display: displays operation mode and parameters POWER: power on/off switch GREEN LED: power LED LIGHT: light source enable/disable key LIGHT OUTPUT: universal connector interface adapter for optical fiber output Keypad: push buttons for operation status control MODE: Select operation mode or parameter setup ENTER: Confirm or select displayed setting : Arrow keys: set parameters or move between setting options. Document #: GP-UM-TLS Page 7 of 40

8 Figure 3. Rear panel Rear panel description: USB: USB interface port RS-232: serial communication port Ethernet: Ethernet interface port GPIB: GPIB interface port Line: external AC supply input connector, 110 V or 220 V BNC: not used in TLS-101 : electrical ground connection Document #: GP-UM-TLS Page 8 of 40

9 Section 4. Operation Instructions: Warning: Never look into the light source fiber connector when the light source is turned on. THE OUTPUT LIGHT FROM A HIGH POWER LASER IS HARMFUL TO HUMAN EYES. Please follow industry standard procedures when operating a high power laser source. The TLS-101 is designed for indoor use only. Avoid water condensation or liquid spills during TLS-101 storage and operation. 4.1 Unpacking Inspect TLS-101 for any physical damage due to shipping and transportation. Contact carrier if any damage is found. Check the packing list to see if any parts or accessories are missing. 4.2 Front Panel Operation Setup procedure is described below: 1. Make sure local AC voltage matches the AC voltage requirement of the TLS-101 system. If not, do not proceed. Contact General Photonics immediately. i. The base model uses VAC, 50/60Hz ii. The TLS-101 can also be configured to use VAC, 50/60Hz by user request at the time it is ordered. 2. Connect the power cord and plug it into the wall receptacle. Make sure the ground pin of the power cord is connected to earth ground. 3. Connect the output fiber to the output connector of the TLS-101. It is important to clean the fiber connector using industry standard procedures and to make sure that the connection is good. Make sure that the optical power source is off during connector cleaning. 4. Turn on the power supply, select operational mode and parameters (see next section) and turn the light enable key. Document #: GP-UM-TLS Page 9 of 40

10 When the instrument is first powered on, it will go through an initialization process. During this time, the LCD will display the following initialization screen: Note that the firmware version may vary. When initialization is complete, the LCD displays: The default operational mode is continuous scan mode. At this point, the user can use the up or down arrow keys to change the work mode, followed by the ENTER key to confirm the setting. The MODE key can be used to return to the menu to select other operational modes. Before the instrument is powered off, all settings that are currently being used, will be saved automatically as the default setting value for the next operation. Mode selection: General Photonics TLS-101, VER CONTINUOUS SCAN 2. MANUAL SCAN Press the MODE key to bring up the operational mode and basic parameter selection screens. The TLS-101 has 5 options available under its mode selection screens, with 2 options listed on each screen. When the MODE key is pressed, one of these screens will appear, with the current operation mode highlighted. The default operation mode is continuous scan mode or the mode used last. Use the UP and DOWN arrows to move between options, and the ENTER key to select the highlighted option. The mode selection screens are shown below. 1. CONTINUOUS SCAN 2. MANUAL SCAN 3. SINGLE WAVELENGTH 4. SELECT WAVELENGTH 5. SET GPIB ADDR Document #: GP-UM-TLS Page 10 of 40

11 OPTION 1: Continuous Scan Mode When the ENTER key is first selected for continuous scan mode, the LCD displays: Start: 01 ( THz) λ: nm This is the initial starting channel setting screen. The top line shows the channel setting and the frequency of this channel, and the bottom line shows the wavelength of this channel at the time the MODE key was pressed. Use the arrow keys to set the desired channel number. The LEFT and RIGHT arrows move the cursor between digits, and the UP and DOWN arrows increment and decrement the active digit. Once the value is set, the ENTER key implements the setting. Then the LCD will display the screen for the next setting: End: 89 (196.10THz) λ: nm The first line shows the ending channel number and the frequency of this channel, and the bottom line shows the wavelength of the ending channel. Press the arrow key to set the desired channel number as described above. Pressing ENTER from the operation screen switches the display to the step length and dwell time setting screen: Step: 1x 50 = 50GHz Delay: 200 ms/step The step setting is displayed in the top line, and can be changed with the arrow keys, the bottom line displays the rate setting which is the dwelling time at each step. Pressing ENTER after completing the step setting and the cursor goes to the bottom line. At this time the rate value may be changed by arrow keys. Pressing the ENTER key again from the operation screen switches to the next setting screen: Trace: SAW Cycle: 50 Trace: SAW Cycle: 50 ENTER key to start In this screen there are still two parameters. Trace gives the option to set either the triangle or saw wave scanning mode. When the triangle wave is selected, the wavelengths follow the pattern traced out by the triangle waveform, which is a smooth up and down selection from the start to the end wavelength and back down, never stopping or skipping any wavelength. The saw waveform begins at the start wavelength and moves toward the Document #: GP-UM-TLS Page 11 of 40

12 end wavelength, however when it reaches the end wavelength, it skips back to the start wavelength for another run. In this way a saw tooth waveform is traced out. Use the UP and DOWN keys to select the desired trace. Pressing the ENTER key after completing the selection and the cursor will go to the cycle setting. This is the number of runs that the system will make. User the arrow keys to change this setting. Then press ENTER again and the screen will display the indication information at the bottom line ENTER key to start If some setting needs to be modified at this point, press the MODE key, then repeat the operation as above. If all the settings are correct, then press the ENTER key to run. Then the screen will display the running information: (1) if no error screen, the current channel and wavelength will be shown. (2) the screen displays TLS is not ready, please try again!. It indicates that the tunable laser has not initialized, you should wait for a moment and press ENTER again. If the same information is still displayed, it may be that something is wrong with the tunable laser. Please consult with GENERAL PHOTONICS. (3)the screen displays Setting error!. This indicates that the difference between the start and the end channel is not a multiple of the step. You should press the MODE key to set the parameter again. Current Output CH: xx λ: xxxx.xx nm TLS is not ready, Please try again! Setting error! OPTION 2: Manual Scan Mode Manual scan mode allows the user to control the outputting channel. When Option 2 is first selected, the LCD will display: Start: 01 (191.70THz) λ: nm This is the initial starting channel setting screen. The top line shows the channel setting and the frequency of this channel, and the bottom line shows the wavelength of this channel at the time the MODE key was pressed. Use the arrow keys to set the desired channel number. The LEFT and RIGHT arrows move the cursor between digits, and the UP and DOWN arrows increment and decrement the active digit. Once the value is set, the ENTER key implements the setting. Then the LCD will display the screen for the next setting: Document #: GP-UM-TLS Page 12 of 40

13 End: 89 (196.10THz) λ: nm The first line shows the ending channel number and the frequency of this channel, the bottom line shows the wavelength of ending channel. Press the arrow key to set the desired channel number as described above. Pressing ENTER from the operation screen switches the display to step length and dwell time setting screen: Step: 1x 50 = 50GHz The step setting is displayed in the top line, and can be changed with the arrow keys. Pressing ENTER key from the operation screen switches to the next setting screen: Trace: Triangle ENTER key to start Trace gives the option to set either the triangle or saw wave scanning mode. When the triangle wave is selected, the wavelengths follow the pattern traced out by the triangle waveform, which is a smooth up and down selection from the start to the end wavelength and back down, never stopping or skipping any wavelength. The saw waveform begins at the start wavelength and moves toward the end wavelength, however when it reaches the end wavelength, it skips back to the start wavelength for another run. In this way a saw tooth waveform is traced out. Use the UP and DOWN keys to select the desired trace. If some setting needs to be modified at this point, press the MODE key, then repeat the operation as above. If all the settings are correct, then press the ENTER key to run. Then the screen will display the running information: (1) if no error screen, the current channel and wavelength will be shown. (2) the screen displays TLS is not ready, please try again!. It indicates that the tunable laser has not initialized, you should wait for a moment and press ENTER again. If the same information is still displayed, it may be that something is wrong with the tunable laser. Please consult with GENERAL PHOTONICS. (3)the screen displays Setting error!. This indicates that the difference between the start and the end channel is not a multiple of the step. You should press the MODE key to set the parameter again. Current Output CH: xx λ: xxxx.xx nm TLS is not ready, Please try again! Setting error! Document #: GP-UM-TLS Page 13 of 40

14 OPTION 3: Single Wavelength Mode Under fixed wavelength mode the instrument outputs the fixed channel according to user s setting. When Option 3 is first selected, the LCD displays: Set CH: 01(191.70THz) λ: nm This is the outputting channel setting screen. The top line shows the channel setting and the frequency of this channel, and the bottom line shows the wavelength of this channel at the time the MODE key was pressed. Use the arrow keys to set the desired channel number. Once the value is set, the ENTER key implements the setting. Then the instrument begins to output light, and the LCD will display the running information: (1) if no error the screen displays the current channel and wavelength will be shown. Press the MODE key to return to the mode selection screen. (2) the screen displays TLS is not ready, please try again!. It indicates that the tunable laser has not initialized and you should wait for a moment and press ENTER again. If the same information is still displayed, it may be that something is wrong with the tunable laser. Please consult with GENERAL PHOTONICS. Current Output CH: xx λ: xxxx.xx nm TLS is not ready, Please try again! OPTION 4: Select Wavelength Mode Select Wavelength Mode allows the user to input up to 40 wavelengths in any combination, to be stored and played back, from any start to any end wavelength. Coincidentally, these wavelengths are stored as channels and should not be confused with the channels that are also associated with the wavelengths. For further clarity, CH1 will not be referenced by the word channel, even though it is an abbreviation. Channel will be reserved for the wavelengths. When the ENTER key is first selected for Select Wavelength Mode, the LCD displays: CH1: 01 ( nm) key to continue Document #: GP-UM-TLS Page 14 of 40

15 This is the first wavelength to be entered (CH1). The top line shows the channel setting and the wavelength of this channel. Use the arrow keys to set the desired channel number. The LEFT arrow key moves the cursor between digits, and the UP and DOWN arrows increment and decrement the active digit. The RIGHT arrow key is reserved for moving to CH2, CH3, etc. Then the LCD will display the screen for the next setting: CH2: 02 ( nm) key to continue Any channel can be chosen for CH2. In this manner, the laser can be made to sweep any wavelength pattern including holding for a period at a special wavelength by entering the same wavelength through many CH s. When the user is satisfied with the number of entries, or CH40 has been reached, pressing ENTER will take the user to the Delay screen: Pressing ENTER will start the scan loop which will go on infinitely unless interrupted by pressing MODE. OPTION 5: Set GPIB Address Delay: 200 ms/step Enter key to start Option 5 allows the user to set the instrument s GPIB address. When this option is selected, the LCD displays the current GPIB address information, as shown below: GPIB ADDR: 11 Use the UP and DOWN arrow keys to change the GPIB address (range is 1-30), then press ENTER to confirm the setting. 4.3 Front panel operation troubleshooting 1. No light output: Make sure that the light enable key is in the on position. 2. Unstable light output: Make sure that the TLS-101 has been allowed to warm up for minutes when first powered on, or has been allowed minutes to stabilize if the operating temperature setting has been changed. 3. Optical output power lower than expected: Make sure that the output fiber connector type matches the TLS-101 s output fiber connector, and that all connectors and adapters are clean. Document #: GP-UM-TLS Page 15 of 40

16 4.4 Testing and Characterization: The TLS-101 can be serviced only by manufacturer-authorized personnel. There are no user serviceable components in this system. The performance of the TLS-101 can be evaluated using standard power meters and spectrum analyzers. 4.5 Remote control and programming: RS-232 operation RS-232 connection The RS-232 serial interface port allows the user to remotely control the TLS-101. Any program that supports RS-232 communication protocols can be used to send ASCII commands to the TLS-101 to remotely access the system. The accompanying LabVIEW program shows a programming example. The RS-232 connector on the rear panel of the TLS-101 is a DB9 male connector. Use a straight connection RS-232 cable to connect the TLS-101 to the RS-232 port of a computer. To ensure proper communication, use a serial cable with direct pin-to-pin connected wires (see Fig. 4) at both ends of the cable. GND RXD TXD Figure 4. RS-232 connector pin assignment on TLS-101 back panel Remote operation and commands General Photonics provides a test program for remote control of the TLS-101. The control software is included with the TLS-101 package. Installation and application procedures for the control software are described later in this section. Document #: GP-UM-TLS Page 16 of 40

17 If users write their own control programs, the following steps and commands are recommended for remote operation of the TLS-101 using the RS-232 communication port. 1. Connect the TLS-101 to the PC using the RS-232 straight cable (DB-9 female to female). 2. Turn the power switch on. 3. Open the correct remote port. 4. Referring to Table 1, the user can send a command string to the TLS-101 through the RS-232 port. There are many programming languages that support serial communications, including Visual Basic, LabVIEW, and C. RS-232 command notes: 1. RS-232 port uses asynchronous framing, 8 data bits, no parity bit, and 1 stop bit. 2. RS-232 baud rate: 9600 bps. 3. Only one command is allowed in each command string. Table 1 Remote control commands Note: The commands and responses listed in tables 1 and 2 are the same for all communication protocols: RS-232, USB, Ethernet, and GPIB. System commands Command Description Response: (Command acknowledgement, error code, or requested data) All responses start with * and end with # *IDN? Queries the TLS-101 identification. *IDN GP-TLS-101# *VER? Queries the TLS-101 Firmware Version *VER 1.0# (may vary) *IND ON/OFF# Enable/disable the indication of commands *E00#: OK when a response is given *E01#: undefined command *ADR nn# Set the GPIB address, nn is from 1 to 30 *E00#: OK *E01#: undefined command *E06#:parameter outside *ADR? Query the GPIB address *ADR 9# *E01#: undefined command *NOTE: Disabling the indication command may make some accompanying Labview programs not work. Basic optical value commands *POW? Return the power value and use the optimal *POW 2.55# unit. *REG hhhhhh? Query the register of TLA,hhhhhh is the register,data1 and data2. for example, if I want know the status of register 0x32,data1 The status of register is the response, hh hh hh hh. Refer to OIF-ITLA-MSA for the meaning of this data. is 0x00,data2 is 0x00, so the command is *TLS:REG ? *CHN nn# Sets the channel of laser source. nn=1~89 *E00#: OK *E01#: undefined command *E06#:parameter outside Document #: GP-UM-TLS Page 17 of 40

18 *CHN? *FRQ? *WAV? *STA n# *STA? *SWP:MOD n# Returns the current channel number. This command is available for sweep mode Returns the current frequency of laser source Returns the current wavelength of laser source Switch the laser source on or off 0: laser off; 1: laser on ; 2: laser not available Queries the laser state. 0: laser off. 1: laser on Sets the sweep mode. 0: manual mode. 1: cycle mode. *CHN 12# *E01#: undefined command *STA 0#: laser off *STA 1#: laser on *STA 2#: laser not available *SWP:MOD? Query the current working mode *MOD 0/1# *SWP:CYC nnn# Sets the number of cycles for sweep mode, Cycle is 1-999, if cycle=0, the cycle will be *TLS:SWP:CYC nnn# *TLS:SWP:CYC INF# infinite (INF), namely continuous sweep *SWP:CYC? Returns the number of cycles *CYC 0#: cycle is infinite *TLS:SWP:DWE nnnn# Sets the dwell time for sweep mode Can only be used when sweep mode is under cycle mode; nnnn is ms *TLS:SWP:DWE 100# *SWP:DWE? *SWP:REP n# Returns the dwell time of the sweep mode Sets the repeat mode. N=0 or1 0: one way. Every stepped or continuous sweep cycle starts at the start wavelength of the sweep and ends at the stop wavelength of the sweep(saw wave) 1: Two way. Every odd stepped sweep cycle starts at the start wavelength of the sweep, and every even stepped sweep cycle starts at the stop wavelength of the sweep.(triangle wave) *SWP:REP? Returns the status of repeat mode. *0#; one way *1#: two way *SWP:BGN nn# Sets the start channel of a sweep. nn: 1~89 *TLS:SWP:BGN 0# *TLS:SWP:BGN 88# *SWP:BGN? Returns the start channel of the sweep *SWP:END nn# Sets the end channel of a sweep. nn 1~89 *SWP:END? *SWP:STE nn# *SWP:STE? *SWP:STA n# Returns the end channel of the sweep Sets the width of the sweep step. nn:1~88 Returns the channel step of the sweep Stops, starts, pauses and continues a channel sweep 0: stop 1:start 2: pause 3: continuous *E00#: OK *E01#: command error *E13#: parameter setting error, the starting and ending channel is not the multiple of the step size. *E14#: commands conflict, stop can be used under start or continuous, start can Document #: GP-UM-TLS Page 18 of 40

19 *SWP:STA? *SWP:ssss# *RST# *WAV nn? *FRQ nn? *CHN:MAX? Returns the state of a sweep. 0: Sweep is not running; 1:Sweep is running ssss: next. Performs the next sweep step in stepped sweep if it is in manual sweep. Resets laser source Returns the wavelength of the channel nn. Returns the frequency of the channel nn Returns the number of the last channel be used under stop or pause, pause can be used under start or continuous; continuous can be used under pause. *E15#: TLS is not ready for output, and try again please. *E00#: OK *E01#: command error *E14#: commands conflict, this command only can be used in manual sweep. Table 2 Command Response Codes E00 E01 E02 E03 E04 E05 E06 E13 E14 E15 No error (Correct command received) Undefined Command Missing parameter Invalid syntax in command string String of characters too long (>buffer limit) Parameter has too many digits after the decimal point Parameter outside the allowed range Parameter setting error, the difference between start and end channel number is not the multiple of the step size. Commands conflict, stop can be used under start or continuous; start can be used under start or continuous; continuous can be used under pause. TLS is not ready for output, and try again please. Document #: GP-UM-TLS Page 19 of 40

20 RS-232 troubleshooting If a problem occurs during RS-232 control, please check the following: 1. Cable should be straight-wired: Select a straight-wired (pin-to-to pin wired) cable; 2. Verify active Com Port (COM1, COM2, etc.); 3. Verify Com Port settings: 8 data bits, 1 stop bit, no parity bits; 4. Check baud rate: must be 9600 bps; 5. Check commands: command should begin with * and end with # or? USB control Connect the instrument to the control computer using the provided USB cable Procedure for Installing USB VCP driver 1. Copy VCP driver from CD-ROM to your computer. 2. Turn on power supply and plug in USB cable; this should bring up the Found New Hardware Wizard : 3. Click Next : Document #: GP-UM-TLS Page 20 of 40

21 4. Click Next : 5. Check Specify a location, then click Next : Document #: GP-UM-TLS Page 21 of 40

22 6. Use Browse to locate the file directory where the VCP driver is. Click OK to proceed to the next screen: 7. Click Next : Document #: GP-UM-TLS Page 22 of 40

23 8. Click Finish. This completes the first part of the installation, during which the USB to serial converter is installed. The second part consists of the installation of the serial port driver. This should follow automatically once the converter installation is completed: Document #: GP-UM-TLS Page 23 of 40

24 9. Click Next : 10. Click Next : Document #: GP-UM-TLS Page 24 of 40

25 11. Use Browse to locate the file directory where the VCP driver is. Click OK to proceed to the next screen: 12. Click Next : Document #: GP-UM-TLS Page 25 of 40

26 13. Click Finish. Depending on the system software and the location of the driver, the sequence of screens between the found new hardware screen and the completion screen may be slightly different, but in all cases, the result will be the installation first of the USB serial converter and then the USB serial port. Once the installation of the serial port driver has been completed, confirm that the installation has been successful by looking under Device Manager on the System Properties screen: select Control Panel System Properties Hardware Device Manager, then Select View Device by type. The resulting screens are shown below. The device should have installed as a USB Serial Port (COMx) attached to USB High Speed Serial Converter. Note the port number that the device is using; it will need to be entered in the control program. The USB driver can drive more than one instrument, but the correct port number must be set in the control programs for each of the instruments. Document #: GP-UM-TLS Page 26 of 40

27 In this example, the USB device is connected to COM3. Document #: GP-UM-TLS Page 27 of 40

28 VCP Uninstall Procedure: Unplug device. Run Add/Remove Programs from Control Panel: Document #: GP-UM-TLS Page 28 of 40

29 TLS-101 USB Commands For users who prefer to write their own control programs, the command list is the same as for RS-232 (see table 1 in section ) USB Troubleshooting: Sometimes, for unknown reasons, the USB Serial Port cannot be found in Device Manager. To troubleshoot this issue, run Add/Remove Programs from the Control Panel, click FTDI USB Serial Converter Drivers, and then click the Change/Remove button. A warning window will tell you that you have to unplug the USB cable if you want to remove the drives. Unplug the USB cable, then press the Continue button. Restore the computer, and then run the VCP installation procedure again. Document #: GP-UM-TLS Page 29 of 40

30 4.5.3 LabVIEW control programs LabView control programs are provided for computer control of the TLS-101 through different control interfaces. Before running these programs, the user should determine whether the LabView developing environment is installed in the control computer. If the LabView environment (version 8.2 or above) is present, either the source code (vi) or the executable versions of the remote control programs can be used without installing further drivers. To control the ER meter through the USB port, the USB drivers need to be installed (see section 4.5.2) Note: For simple control of the instrument, it is often easier to use the executable versions of the program than the source code. However, the executable versions will not run properly while LabView is running. If using the executable versions of the program on a computer with LabView installed, the user should quit LabView before running them. If LabVIEW is not installed on the control computer, the user should set up the LabVIEW Run-Time engine by running the programs under \LVRunTimeEng_8.2 (on the CD-ROM), and install the driver program for VISA (used for RS-232, distributed by National Instruments) by running the program named Visa400runtime.exe (under the root directory). For USB control, the USB drivers also need to be installed (see section 4.5.2). For further details on Ethernet and GPIB control, see sections and 4.5.5, respectively. Once the necessary drivers and other setup parameters are in place, the executable programs in the application directory can be run. After the necessary drivers are installed, the programs (executable and/or source code) should be copied to the hard drive and run from there. Document #: GP-UM-TLS Page 30 of 40

31 Note: Always close the control program (and LabVIEW as well if using the VI version of the control program) before turning off the instrument. The control programs TLS_REMOTE.vi (source code) or TLS_REMOTE.exe (executable) are compatible with RS-232, USB and GPIB interfaces. The program will auto-sense which interface is being used and display the appropriate port configuration option. The control program interface is otherwise identical for RS-232, USB and GPIB control. Figure 5 Continuous scan mode Port number: Displays the selected port for the RS-232, USB or GPIB connection. If the selected port and the actual connection do not match, the program will not function correctly. Set the correct port number here. The port number can be determined from the Device Manager (see section 4.5.2) or from the Measurement & Automation program from National Instruments. Note that the port number cannot be set while the program is executing (trying to make a measurement). Stop the program first and choose the correct port if necessary. Notes: (a) ASRL1:INSTR is the same as COM1 for RS-232. (b) GPIB:XX:INSTR means the GPIB port with an address of XX. Once the port configuration is set, the program can be started by clicking the white arrow in the top left corner of the screen. The continuous scan mode screen allows the user to set the six parameter values. The default value may different from the data saved in the instrument, therefore make sure the desired data has been set successfully, and then click OK. Document #: GP-UM-TLS Page 31 of 40

32 The STOP button stops continuous scan mode operation. It may be necessary to press STOP twice if the first one fails to properly return the system to the MODE menu. Manual Scan Mode The manual scan mode control screen is shown below: Figure 6 Manual Scan Mode The manual scan mode screen allows the user to set the four parameter values to be maintained by typing in the boxes. Note that the default value may be different with the values saved in the instrument, therefore make sure the values have been set successfully, and then click OK. The STOP button stops this process. Document #: GP-UM-TLS Page 32 of 40

33 Single Wavelength Mode Figure 7 Fixed Wavelength Mode This screen is used to set the TLS-101 to fixed wavelength mode. The user selects the desired channel to be maintained by typing it in the box. Note that the default value may be different with the values saved in the instrument, therefore make sure the channel number has been set successfully, and then click OK. The STOP button disables the fixed wavelength output. Document #: GP-UM-TLS Page 33 of 40

34 Select Wavelength Mode Figure 8 Selected wavelength mode Under Select Wavelength Mode, the user inputs the desired channel in the box and presses OK. This will increment the number (previously identified as CH1, CH2 etc.). A correctly stored number will be given as an E00 response in the associated box. A sufficient time should be given to allow the computer to successfully communicate and store the value to the memory of the TLS. This time will differ with connection speed and interface type. A reasonably safe time frame may be about 1 sec. A value that is not stored correctly may show an error code of E06 which literally means value outside allowed range, but this may result if the information is sent too quickly to the device. If the TLS responds with a code other than E00, the Labview control program must be stopped and restarted. This is accomplished by pressing STOP, and pressing the right arrow key near the upper left corner of the window. The numbers will only increment, so if a change is desired to a previous entry, the program must be restarted. This will clear the program and begin again from the first number. The dwell time can also be changed by pressing SET as well. Once the user is satisfied with the programmed wavelengths, pressing OK will start the scan. Pressing STOP will terminate the scan and return to the mode selection menu. Document #: GP-UM-TLS Page 34 of 40

35 4.5.4 Ethernet control The TLS-101 includes a small server. The following steps describe how to use it. 1. Connect Instrument There are two connection configurations for Ethernet control: a) Connect the instrument directly to the PC, as shown in figure 9. b) Connect the instrument with a DHCP server (router or exchange server), as shown in figure 10. DHCP server PC TLS-101 PC TLS-101 Figure 9 Direct connection: TLS-101 to PC Figure 9 Connect TLS-101 with DHCP server 2. There are two ways of assigning the instrument an IP address: a) Static IP: the administrator assigns the instrument a fixed IP address. b) Dynamic IP: the DHCP server assigns the instrument an available address when requested. 3. Set Ethernet configuration and get IP address: Open the LabVIEW program (GP_SET_ETHERNET.vi or GP_SET_ETHERNET.exe) to set the Ethernet configuration. NOTE: This program was designed for an RS-232 interface, so the instrument should be connected to the computer via RS-232 before running it. Please see the instructions at the beginning of section for preparations to run an RS-232 LabView program. The program interface is shown below. a) To use a static IP configuration, click the fixed IP mode tab, as shown in figure 11. Set the 5 parameters shown in figure 11. They can be obtained from your network administrator. The default values are those shown in Fig. 11. Once the parameters are set, click the Set button to confirm. Document #: GP-UM-TLS Page 35 of 40

36 Figure 11 Fixed IP mode b) To use dynamic IP address assignment, click the DHCP MODE tab, as shown in figure 12. Click the OK button to confirm. Note: this is the default mode. Figure 12 Dynamic IP mode Document #: GP-UM-TLS Page 36 of 40

37 After IP address configuration, select the GET IP ADDRESS tab to determine the IP address of the instrument, as shown in figure 13. Write down the IP address; it will be needed to access the TLS-101 s server. Figure 10 Get IP Address 4. Ethernet control of the TLS-101: Open the LabVIEW program (GP_TLS_ETHERNET.vi or GP_TLS_ETHERNET.exe) to control the instrument. The program interface is shown in figure 14. Input the IP address obtained in the previous step (see Fig. 13). If the GP_SET_ETHERNET program s static IP configuration was used, the Port number is the one set from that screen (see Fig. 11). The default value is 23. Document #: GP-UM-TLS Page 37 of 40

38 Figure 14 TLS-101 Ethernet Control Program From this point, the program interface is very similar to the RS-232, USB, and GPIB versions of the control program, with the same CONTINUOUS SCAN MODE, MANUAL SCAN MODE, SINGLE WAVELENGTH MODE, and SELECT WAVELENGTH tabs. See section for a more detailed description of the control program interface GPIB control Setting the GPIB address Each device on the GPIB must have a unique address, including the computer card. There are three methods to find the TLS-101 s current GPIB address: 1) Front panel: Press the MODE button and select option 5. The LCD will display the current GPIB address. 2) Other remote control interface: Send the command *ADR? over an RS-232 or USB connection. The TLS-101 will respond with the active GPIB address. 3) Use National Instruments Measurement & Automation Explorer software to query the address of the TLS-101. This will display all devices on the GPIB. Note: The TLS-101 s GPIB address is stored in nonvolatile memory, so when the instrument is powered down and then powered up again, the active GPIB address will be the last one set. The accompanying program entitled GP_GPIB_ADDRESS.vi or GP_GPIB_ADDRESS.exe may also be used to identify the GPIB port by querying the Document #: GP-UM-TLS Page 38 of 40

39 instrument with a serial, or USB connection. The GPIB cable must also be plugged in. A screenshot of the program in action can be seen below. Figure 15 Querying the TLS-101 with the GP_GPIB_ADDRESS program using a serial connection GPIB Control Connect the instrument to the computer with a GPIB cable. After determining/setting the TLS-101 s GPIB address, select the corresponding port number in the corresponding box of the LabView control program (see section 4.5.3) and run the control program. In addition to using the provided control program, individual commands can be sent through the GPIB port. Enter the GPIB address in the communication program being used. The instrument will then be ready to receive commands. The command list is given in Table 1 at the beginning of section To ensure proper communication, use a GPIB cable that is fully compatible with the IEEE standard. All GPIB/IEEE 488 interface connections must be made before turning on the instruments. Document #: GP-UM-TLS Page 39 of 40

40 Section 5. Technical Support General Photonics is committed to high quality standards and customer satisfaction. For any questions regarding the quality and use of the TLS-101, or future suggestions, please contact General Photonics Corporation at (909) (telephone) or (909) (fax), or by at General Photonics will respond to all customer questions within 24 hours during regular business hours. General Photonics can also be contacted by mail at: General Photonics 5228 Edison Avenue Chino, California USA Document #: GP-UM-TLS Page 40 of 40

PCD 005 PolaMIX. Operation Manual

PCD 005 PolaMIX. Operation Manual PCD 005 PolaMIX All-Fiber Polarization Scrambler Module with Microprocessor Controller Operation Manual July 29, 2011 General Photonics Corp. Ph: (909) 590-5473 5228 Edison Ave. Fax: (909) 902-5536 Chino,

More information

PCD 104. Operation Manual

PCD 104. Operation Manual PCD 104 Wide Bandwidth Fiber-Optic Polarization Scrambler Operation Manual April 22, 2011 General Photonics Corp. Ph: (909) 590-5473 5228 Edison Ave. Fax: (909) 902-5536 Chino, CA 91710 USA www.generalphotonics.com

More information

PDLE 101. Operation Manual

PDLE 101. Operation Manual PDLE 101 PDL Emulator Operation Manual Version 1.1 June 6, 2011 General Photonics Corp. Ph: (909) 590-5473 5228 Edison Ave. Fax: (909) 902-5536 Chino, CA 91710 USA www.generalphotonics.com Document #:

More information

MDL 002. Operation Manual

MDL 002. Operation Manual MDL 002 Miniature Motorized Variable Optical Delay Line Operation Manual Version: 1.3 May 15, 2015 General Photonics Corp. Tel: (909) 590-5473 5228 Edison Ave. Fax: (909) 902-5536 Chino, CA 91710 USA www.generalphotonics.com

More information

VDL 002 VariDelay I. Operation Manual

VDL 002 VariDelay I. Operation Manual VDL 002 VariDelay I Miniature Manual Variable Fiber Optic Delay Line Operation Manual May 6, 2009 General Photonics Corp. Ph: (909) 590-5473 5228 Edison Ave. Fax: (909) 902-5536 Chino, CA 91710 USA www.generalphotonics.com

More information

PSW-002. Fiber Optic Polarization Switch. User Guide

PSW-002. Fiber Optic Polarization Switch. User Guide PSW-002 Fiber Optic Polarization Switch User Guide Version: 1.1 Date: June 30, 2015 General Photonics Corporation is located in Chino California. For more information visit the company's website at: www.generalphotonics.com

More information

POL-001 Fiber-Optic In-Line Polarizer. Operation Manual

POL-001 Fiber-Optic In-Line Polarizer. Operation Manual POL-001 Fiber-Optic In-Line Polarizer Operation Manual Feb. 18, 2002 General Photonics Corp. Ph: (909) 590-5473 5228 Edison Ave. Fax: (909) 902-5536 Chino, CA 91710 USA www.generalphotonics.com Document

More information

VDL-004. Miniature Manual Variable Optical Delay Line. User Guide

VDL-004. Miniature Manual Variable Optical Delay Line. User Guide VDL-004 Miniature Manual Variable Optical Delay Line User Guide Version: 1.1 Date: July 24, 2018 General Photonics Corporation is located in Chino, California. For more information visit the company's

More information

3700 SERIES USER MANUAL

3700 SERIES USER MANUAL SAFETY GUIDE This manual contains the precautions necessary to ensure your personal safety as well as for protection for the products and the connected equipment. These precautions are highlighted with

More information

PDL-201. Polarization Dependent Loss Multimeter. User Guide

PDL-201. Polarization Dependent Loss Multimeter. User Guide PDL-201 Polarization Dependent Loss Multimeter User Guide Version: 1.0 Date: May 21, 2014 General Photonics, Incorporated is located in Chino California. For more information visit the company's website

More information

FTE-6000 Hand Held Tuneable Laser Source

FTE-6000 Hand Held Tuneable Laser Source FTE-6000 Hand Held Tuneable Laser Source USER S MANUAL Avoid Exposure Laser Radiation Emitted From The Aperture This product conforms to CDRH standards for laser product Per 21 CFR 1040.10 & 1040.11 Table

More information

RS-232/422/485 to Copper or Fiber. Ethernet Converter. User s Manual

RS-232/422/485 to Copper or Fiber. Ethernet Converter. User s Manual RS-232/422/485 to Copper or Fiber Ethernet Converter User s Manual Table Of Contents TABLE OF CONTENTS... 1 INTRODUCTION... 3 PRODUCT OVERVIEW... 3 PRODUCT FEATURES... 3 PACKING LIST... 4 LED INDICATORS...

More information

Precise Fiber Couple Laser Light Source

Precise Fiber Couple Laser Light Source Precise Fiber Couple Laser Light Source User Guide User Guide Document No. AEO-LD1480-0001 1. SAFTEY REQUIREMENT... 2 Precise Laser Light Source Page 1 / 15 1.1 User esponsibility...3 2. Description..

More information

OTEB-CO-P PRE-AMP EDFA INSTRUCTION MANUAL

OTEB-CO-P PRE-AMP EDFA INSTRUCTION MANUAL OTEB-CO-P PRE-AMP EDFA INSTRUCTION MANUAL Phone: (209) 586-1022 (800) 545-1022 Fax: (209) 586-1026 E-Mail: salessupport@olsontech.com www.olsontech.com 025-000521 Rev. X1 10/01/2012 SAFETY WARNINGS LASER

More information

HCC-CW-201A. CWDM Analyzer. Operation Guide

HCC-CW-201A. CWDM Analyzer. Operation Guide CWDM Analyzer Operation Guide Table of Contents Description... 3 Features... 3 Specifications... 4 Safety Information... 5 Preparing for Operation... 6 Unpacking the instrument... 6 Front keys definition...

More information

PCD M02/MPD 001. Operation Manual

PCD M02/MPD 001. Operation Manual PCD M02/MPD 001 Miniature Multi-Channel Piezoelectric Actuator Driver Card With or without Integrated PolaRITE III Polarization Controller Operation Manual Version 1.3 October 30, 2015 General Photonics

More information

RC-SV Configuration Guide Revision 3

RC-SV Configuration Guide Revision 3 Kramer Electronics, Ltd. RC-SV Configuration Guide Revision 3 Software Version 2.1.2.32 Intended for Kramer Technical Personnel or external System Integrators. To check that you have the latest version,

More information

HDMI MATRIX SWITCHER B-120-HDMATRIX-4x4/8x8 CONFIGURATION UTILITY MANUAL

HDMI MATRIX SWITCHER B-120-HDMATRIX-4x4/8x8 CONFIGURATION UTILITY MANUAL HDMI MATRIX SWITCHER B-120-HDMATRIX-4x4/8x8 CONFIGURATION UTILITY MANUAL IMPORTANT SAFETY INSTRUCTIONS To reduce the risk of fire or electric shock, read and follow all instructions and warnings in this

More information

LT900SERIES. 1xN Multi-Channel Switch Operation Manual

LT900SERIES. 1xN Multi-Channel Switch Operation Manual LT900SERIES 1xN Multi-Channel Switch Operation Manual Table of Contents General Information.. 1 General Specifications...... 2 Interface Connectors..... 3 Connector Assignment 16 pin........ 4 26 pin....

More information

Fiber Optic. Foreword. Special Tips:

Fiber Optic. Foreword. Special Tips: Model : WE1550HD-20 Foreword This manual applies to 1550nm EDFA. It mainly describes the performance characteristics, technical parameters, installation and debugging, common troubleshooting, and other

More information

Master Oscillator Power Amplifier SLD System MOPA-SLD-850

Master Oscillator Power Amplifier SLD System MOPA-SLD-850 Master Oscillator Power Amplifier SLD System MOPA-SLD-850 1. Product Description The Superlum MOPA-SLD-850 is an ultra-high power SLD-based light source that features an extremely weak sensitivity to optical

More information

PM-212-MPO Pocket Optical Power Meter INSTRUCTION MANUAL

PM-212-MPO Pocket Optical Power Meter INSTRUCTION MANUAL PM-212-MPO Pocket Optical Power Meter INSTRUCTION MANUAL Revision 1.2 is the registered trademark of OPTOKON, a.s. Other names and trademarks mentioned herein may be the trademarks of their respective

More information

FIBER PIGTAILED ULTRA STABLE LASER MODULE OZ-1000, OZ-2000 & OZ-3000 SERIES. Preliminary

FIBER PIGTAILED ULTRA STABLE LASER MODULE OZ-1000, OZ-2000 & OZ-3000 SERIES. Preliminary 219 Westbrook Rd, Ottawa, ON, Canada, K0A 1L0 Toll Free: 1-800-361-41 Tel:(613) 831-0981 Fax:(613) 836-089 E-mail: sales@ozoptics.com FIBER PIGTAILED ULTRA STABLE LASER MODULE OZ-00, OZ-2000 & OZ-3000

More information

OTEB-CO-B Series. Booster EDFA OPERATING MANUAL

OTEB-CO-B Series. Booster EDFA OPERATING MANUAL OTEB-CO-B Series Booster EDFA OPERATING MANUAL 24926 Highway 108 Sierra Village, CA 95346 Phone: (800) 545-1022 Fax: (209 586-1022 025-000533 Rev X6 E-Mail: sales@olsontech.com 27 Apr 2010 TABLE OF CONTENTS

More information

RC-SV Configuration Guide (Rev 4)

RC-SV Configuration Guide (Rev 4) Kramer Electronics, Ltd. RC-SV Configuration Guide (Rev 4) Software Version 2.1.2.69 Intended for Kramer Technical Personnel or external System Integrators. To check that you have the latest version, go

More information

8 Port USB to RS- 232/422/485 Octal Adapter. Product Manual. Coolgear, Inc. Version 1.1 April 2018 Model Number: USB-8COMi-RM.

8 Port USB to RS- 232/422/485 Octal Adapter. Product Manual. Coolgear, Inc. Version 1.1 April 2018 Model Number: USB-8COMi-RM. 8 Port USB to RS- 232/422/485 Octal Adapter Product Manual Coolgear, Inc. Version 1.1 April 2018 Model Number: USB-8COMi-RM 2 USB-8COMi-RM Product Manual Revision History Revision Date Author Comments

More information

99 Washington Street Melrose, MA Phone Toll Free Visit us at

99 Washington Street Melrose, MA Phone Toll Free Visit us at 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Table of Contents 1. General Safety Requirements... 1 2. Safety Terms and Symbols...

More information

LegacyPlus 7-OR SERIES RETURN TRANSMITTER INSTRUCTION MANUAL 7-OR-RT/xxx

LegacyPlus 7-OR SERIES RETURN TRANSMITTER INSTRUCTION MANUAL 7-OR-RT/xxx LegacyPlus 7-OR SERIES RETURN TRANSMITTER INSTRUCTION MANUAL 7-OR-RT/xxx INSTRUCTION MANUAL 025-370546 Rev X1 11-20-06 TABLE OF CONTENTS SAFETY WARNINGS Page 3 INTRODUCTION Page 4 Figure 1-7-OR-RT Module

More information

USB-COMi-TB USB to Industrial Single RS-422 / 485 Adapter Manual. Specifications and Features

USB-COMi-TB USB to Industrial Single RS-422 / 485 Adapter Manual. Specifications and Features USB-COMi-TB USB to Industrial Single RS-422 / 485 Adapter Manual The USB-COMi-TB USB-to-Industrial Single RS-422/485 Adapter is designed to make industrial communication port expansion quick and simple.

More information

OPERATOR MANUAL OSD361 FIBER OPTIC CCTV TRANSMITTER MODULE

OPERATOR MANUAL OSD361 FIBER OPTIC CCTV TRANSMITTER MODULE OPERATOR MANUAL OSD361 FIBER OPTIC CCTV TRANSMITTER MODULE OSD361 FIBER OPTIC CCTV TRANSMITTER MODULE Document No. 10102303 PAGE 1 INDEX 1 1 TECHNICAL SUMMARY... 3 1.1 BRIEF DESCRIPTION...3 1.1.1 OVERVIEW...

More information

OPERATOR MANUAL OSD461A/OSD463A AUDIO FIBER OPTIC TRANSMISSION SYSTEM

OPERATOR MANUAL OSD461A/OSD463A AUDIO FIBER OPTIC TRANSMISSION SYSTEM OPERATOR MANUAL OSD461A/OSD463A AUDIO FIBER OPTIC TRANSMISSION SYSTEM OSD461A/OSD463A AUDIO FIBER OPTIC TRANSMISSION SYSTEM Document No. 101052 Rev. 01 PAGE 2 INDEX 1 1 TECHNICAL SUMMARY... 4 1.1 BRIEF

More information

USB-16COMi-M 16-Port RS-422/485 USB Serial Adapter User Manual. Features and Specifications. Power Supply

USB-16COMi-M 16-Port RS-422/485 USB Serial Adapter User Manual. Features and Specifications. Power Supply USB-16COMi-M 16-Port RS-422/485 USB Serial Adapter User Manual The USB to industrial 16-Port RS-422/485 Adapter is designed to make serial port expansion quick and simple. Connecting to a USB port on your

More information

LASER PLUS OPERATING INSTRUCTIONS

LASER PLUS OPERATING INSTRUCTIONS LASER PLUS OPERATING INSTRUCTIONS Phone: (209) 586-1022 (800) 545-1022 Fax: (209) 586-1026 E-Mail: salessupport@olsontech.com 025-370403 REV X9 www.olsontech.com 12/16/03 TABLE OF CONTENTS SYSTEM OVERVIEW

More information

2.2. Facilities Requirements

2.2. Facilities Requirements 2.2. Facilities Requirements Facilities requirements for the alpha-se system are listed in Table 2-1 and the system dimensions are given in Figure 2-1. As shown in Figure 2-2, the preferred clear work

More information

Product Manual. 8-Port RS-232 USB to Serial Adapter Data Control Box. Coolgear, Inc. Version 1.1 September 2017 Model Number: USB-8COM

Product Manual. 8-Port RS-232 USB to Serial Adapter Data Control Box. Coolgear, Inc. Version 1.1 September 2017 Model Number: USB-8COM 8-Port RS-232 USB to Serial Adapter Data Control Box Product Manual Coolgear, Inc. Version 1.1 September 2017 Model Number: USB-8COM 2 USB-8COM Product Manual Revision History Revision Date Author Comments

More information

Table of Contents. 1 Welcome. 2 Install your Access Point-I. 3 Using the Access Point-I

Table of Contents. 1 Welcome. 2 Install your Access Point-I. 3 Using the Access Point-I Table of Contents 1 Welcome Introducing the Avaya Wireless LAN 1-1 About the Access Point-I 1-2 Finding Information 1-3 2 Install your Access Point-I Overview 2-1 Verify Kit Contents 2-2 Write Down Product

More information

Models 2601B, 2602B, and 2604B System SourceMeter Instruments Quick Start Guide

Models 2601B, 2602B, and 2604B System SourceMeter Instruments Quick Start Guide Models 2601B, 2602B, and 2604B System SourceMeter Instruments Quick Start Guide Safety precautions Observe the following safety precautions before using this product and any associated instrumentation.

More information

SMK585 1U rackmount. With 8 Ports KVM Switch

SMK585 1U rackmount. With 8 Ports KVM Switch SMK585 1U rackmount Monitor Keyboard Drawer With 8 Ports KVM Switch TABLE OF CONTENTS Content FEATURES...1 BASIC SPECIFICATION...2 DISPLAY...2 PACKAGE CONTENTS...2 TECHNICAL SPECIFICATIONS...3 SYSTEM REQUIREMENT...3

More information

OPERATOR MANUAL OSD381 FIBER OPTIC CCTV TRANSMITTER MODULE

OPERATOR MANUAL OSD381 FIBER OPTIC CCTV TRANSMITTER MODULE OPERATOR MANUAL OSD381 FIBER OPTIC CCTV TRANSMITTER MODULE OSD381 FIBER OPTIC CCTV TRANSMITTER MODULE Document No. 10100902 PAGE 1 INDEX 1 1 TECHNICAL SUMMARY... 3 1.1 BRIEF DESCRIPTION...3 1.1.1 OVERVIEW...

More information

User s Manual. ASE-1019 ASE Light Source

User s Manual. ASE-1019 ASE Light Source ASE-1019 ASE Light Source ASE-1019 ASE Light Source User s Manual Triple Play Communications Document 3013800-701 250 East Drive, Suite F Rev 1.1 Melbourne, FL 32904 July 2015 1 Revision History Document

More information

TigerStop Ethernet-to-Serial Converter Installation & User s Guide v4.2

TigerStop Ethernet-to-Serial Converter Installation & User s Guide v4.2 E1 / 2010 ESC 4.2 TigerStop Ethernet-to-Serial Converter Installation & User s Guide v4.2 September 2010 TigerStop Version 5.03+ CONTACT: TigerStop LLC, Assembly Plant, 12909 NE 95 th St., Vancouver, WA

More information

DPS INC ASCII MUX. Operation Guide. Table Of Contents. Overview. T/kda

DPS INC ASCII MUX. Operation Guide. Table Of Contents. Overview. T/kda DPS INC Operation Guide "Your Partners in Telemetry Monitoring Systems" ASCII MUX Fig. - ASCII MUX Expands T/MonXM ASCII s up to Eight Times Table Of Contents Overview... Functional Schematic... Typical

More information

Installing the Cisco Unified Videoconferencing 3545 MCU

Installing the Cisco Unified Videoconferencing 3545 MCU CHAPTER 2 Installing the Cisco Unified Videoconferencing 3545 MCU The Cisco Unified Videoconferencing 3545 MCU works together with a Cisco Unified Videoconferencing 3545 EMP Enhanced Media Processor (EMP)

More information

TT230SM THERMAL TRANSFER PRINTER USER S MANUAL

TT230SM THERMAL TRANSFER PRINTER USER S MANUAL TT230SM THERMAL TRANSFER PRINTER USER S MANUAL Operations Overview Unpacking and Inspection This printer has been specially packaged to withstand damage during shipping. Please carefully inspect the packaging

More information

L300 user manual. Programmable DC Electronic Load CONTENTS. Electronic Load Software CAUTION...2 SAFETY NOTES...2

L300 user manual. Programmable DC Electronic Load CONTENTS. Electronic Load Software CAUTION...2 SAFETY NOTES...2 Programmable DC Electronic Load L300 user manual CONTENTS CAUTION......2 SAFETY NOTES......2 Chapter 1 General Introduction...3 1.1 General Introduction......3 1.2 Specification......3 1.3 Features......3

More information

Model 1931/2931 Series Model 1936/2936 Series

Model 1931/2931 Series Model 1936/2936 Series Model 1931/2931 Series Model 1936/2936 Series Single and Dual-Channel Optical Meters Start Up Guide User s Manual and Future Updates Dear Customer, This Start Up Guide contains essential information,

More information

Training Training Manual: Manual: KI9800A Series Pocket Fibre Sources Level 1

Training Training Manual: Manual: KI9800A Series Pocket Fibre Sources Level 1 June 2011 Training Manual: KI9800A Series Pocket Fibre s Level 1 Course Contents 1. Typical Applications 2. General Features 3. Models 4. Keypad 5. Instrument care 6. Operation 7. Operating Modes 8. Additional

More information

8 Button RS232/IR. Control Panel. MuxLab Inc A / SE A

8 Button RS232/IR. Control Panel. MuxLab Inc A / SE A 8 Button RS232/IR Control Panel 500816 MuxLab Inc. 2016 94-000833-A / SE-000833-A SAFETY PRECAUTIONS To insure the best use from the product, please read all instructions carefully before using the device.

More information

Product Manual. 2 Port USB to RS-422 /485 Optical Isolated Adapter. Coolgear, Inc. Version 1.1 March 2018 Model Number: USB-2COMi-Si-M

Product Manual. 2 Port USB to RS-422 /485 Optical Isolated Adapter. Coolgear, Inc. Version 1.1 March 2018 Model Number: USB-2COMi-Si-M 2 Port USB to RS-422 /485 Optical Isolated Adapter Product Manual Coolgear, Inc. Version 1.1 March 2018 Model Number: USB-2COMi-Si-M 2 USB-2COMi-Si-M Product Manual Revision History Revision Date Author

More information

PM-212-SI3 Pocket Optical Power Meter INSTRUCTION MANUAL

PM-212-SI3 Pocket Optical Power Meter INSTRUCTION MANUAL PM-212-SI3 Pocket Optical Power Meter INSTRUCTION MANUAL Revision 2.4 is the registered trademark of OPTOKON, a.s. Other names and trademarks mentioned herein may be the trademarks of their respective

More information

User Manual. AN A GPON Optical Network Unit. Version: A. Code: MN Date: December 2011

User Manual. AN A GPON Optical Network Unit. Version: A. Code: MN Date: December 2011 An Expert in Optical Communications User Manual AN5506-01-A GPON Optical Network Unit Version: A Code: MN000000943 Date: December 2011 FiberHome Telecommunication Technologies Co., Ltd. Version Version

More information

POWER Shiloh Road Alpharetta, Georgia (770) FAX (770) Toll Free

POWER Shiloh Road Alpharetta, Georgia (770) FAX (770) Toll Free Instruction Manual Model 1582-10M Protection Switch January 2009 Rev O ALARMS MENU OUTPUT = CH1 AUTO POWER 1 2 MODEL 1582 SWITCH CROSS TECHNOLOGIES INC. CH1 CH2 REMOTE EXECUTE Data, drawings, and other

More information

Co-Lase Users Guide. DM00001 Rev A (DCN 031) These instructions are protected by copyright. All rights reserved.

Co-Lase Users Guide. DM00001 Rev A (DCN 031) These instructions are protected by copyright. All rights reserved. Co-Lase Users Guide These instructions are protected by copyright. All rights reserved., Inc. http://www.propel-labs.com 131 East Lincoln Avenue, Suite 200 Fort Collins, CO 80524 1 Table of Contents Chapter

More information

USER MANUAL. RC-74DL Master Room Controller MODEL: P/N: Rev 4

USER MANUAL. RC-74DL Master Room Controller MODEL: P/N: Rev 4 KRAMER ELECTRONICS LTD. USER MANUAL MODEL: RC-74DL Master Room Controller P/N: 2900-000691 Rev 4 Contents 1 Introduction 1 2 Getting Started 2 2.1 Achieving the Best Performance 2 2.2 Safety Instructions

More information

PC Software R&S FS300-K1

PC Software R&S FS300-K1 Copyright 0 Copyright Copyright Licence Agreement Rohde & Schwarz grants you the right to install the R&S FS300-K1 software package on one or more PCs of your choice. The licence included in the software

More information

AD-8923-BCD. Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD

AD-8923-BCD. Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD AD-8923-BCD Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD4002137 2010 A&D Company, Limited. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated

More information

OPTICAL HEADEND PLATFORM OTOHP-CH 3RU CHASSIS OTOHP-PS POWER SUPPLY MODULE OTOHP-BP BLANK PANEL INSTRUCTION MANUAL

OPTICAL HEADEND PLATFORM OTOHP-CH 3RU CHASSIS OTOHP-PS POWER SUPPLY MODULE OTOHP-BP BLANK PANEL INSTRUCTION MANUAL OPTICAL HEADEND PLATFORM OTOHP-CH 3RU CHASSIS OTOHP-PS POWER SUPPLY MODULE OTOHP-BP BLANK PANEL INSTRUCTION MANUAL Phone: (209) 586-1022 (800) 545-1022 Fax: (209) 586-1026 OTOHP-CH Rev. X1 E-Mail: salessupport@olsontech.com

More information

Operational Manual Spectrophotometer Model: SP-830 PLUS Metertech Inc.

Operational Manual Spectrophotometer Model: SP-830 PLUS Metertech Inc. Operational Manual Spectrophotometer Model: SP-830 PLUS Metertech Inc. Version 1.06 Metertech Inc. provides this publication as is without warranty of any kind, either express or implied, including, but

More information

TIME SERVER NETSILON. Quick start.

TIME SERVER NETSILON. Quick start. TIME SERVER NETSILON Quick start This document refers to the following products: 907,900 NETSILON 7 (100-240 VAC) 907,901 NETSILON 7 (18-36 VDC) 907,902 NETSILON 7 (100-240 VAC + 18-36 VDC) www.bodet-time.com

More information

Installation Manual. Mounting Instructions Mechanical Mounting. Luminato. Teleste Corporation

Installation Manual. Mounting Instructions Mechanical Mounting. Luminato. Teleste Corporation Luminato Installation Manual Teleste Corporation Mounting Instructions Mechanical Mounting Luminato Mechanical Installation, agile_59300316, rev0044 Introduction 1 Contents Introduction 4 General... 4

More information

This 4200-RM Rack Mount Kit is for installation in 4200-CAB series cabinets only.

This 4200-RM Rack Mount Kit is for installation in 4200-CAB series cabinets only. Keithley Instruments, Inc. 28775 Aurora Road Cleveland, Ohio 44139 (440) 248-0400 Fax: (440) 248-6168 www.keithley.com Model 4200-RM Rack Mount Kit Packing List Introduction NOTE This 4200-RM Rack Mount

More information

3205 Evaluation Kit: Getting Started

3205 Evaluation Kit: Getting Started Application Note 3205 Evaluation Kit: Getting Started Introduction The JDSU 3205 is a high-performance continuous-wave tunable laser assembly (TLA). This product is designed to be integrated with an external

More information

Nearus USB2.0 Camera Manual NU-350-USB2PTZ-B

Nearus USB2.0 Camera Manual NU-350-USB2PTZ-B Nearus USB2.0 Camera Manual NU-350-USB2PTZ-B Safety Tips Please read this manual carefully before installing the camera. Keep the camera away from violent vibration, physical stress, moisture, extreme

More information

TRC-190 User s Manual

TRC-190 User s Manual User s Manual Edition 3.2, May 2017 www.moxa.com/product 2017 Moxa Inc. All rights reserved. User s Manual The software described in this manual is furnished under a license agreement and may be used only

More information

Installation Manual. 65 Interactive LED/LCD. Model: HILF65101 (64.56 )

Installation Manual. 65 Interactive LED/LCD. Model: HILF65101 (64.56 ) Installation Manual 65 (64.56 ) Model: HILF65101 65 Interactive LED/LCD QUICK SETUP GUIDE For further information, see the user manual. Please contact directly if you have questions on the use of the touch

More information

Serial Data DIN Fiber Link System

Serial Data DIN Fiber Link System USER GUIDE RLH Industries, Inc. The leader in rugged fiber optic technology. U-120 2017A-0420 DIN Fiber Link System COMPACT, RUGGED & TEMPERATURE HARDENED Introduction The DIN Fiber Link system transports

More information

BBA-Series Precision Balances

BBA-Series Precision Balances Schematic BBA-Series Precision Balances BBA-600/BBA-1200 Service Manual 11 Schematic 1 10 Troubleshooting You have purchased a quality precision weighing instrument that requires handling with care. Trouble

More information

SYNERGY GLOBAL INC. Toll Free : Fax :

SYNERGY GLOBAL INC. Toll Free : Fax : SYNERGY GLOBAL INC Toll Free : 1-888-865-6888 Fax : 510-226-8968 Email : info@rackmountmart.com LCD1U15-03 series & LCD1U17-10 series User s manual 4-in-1 (KVM switch, Keyboard, LCD display, touch pad)

More information

EtherSeries. EtherSeries CR-2. CR-2-Opto. User s Guide. Revised October 7, 2013 Firmware Version 1.X

EtherSeries. EtherSeries CR-2. CR-2-Opto. User s Guide. Revised October 7, 2013 Firmware Version 1.X EtherSeries EtherSeries CR-2 & CR-2-Opto User s Guide Revised October 7, 2013 Firmware Version 1.X TABLE OF CONTENTS SECTION 1 - DESCRIPTION... 2 SECTION 2 - SPECIFICATIONS... 4 SECTION 3 - INSTALLATION...

More information

AKELA Vector Network Analyzer (VNA) Quick Start Guide

AKELA Vector Network Analyzer (VNA) Quick Start Guide AKELA Vector Network Analyzer (VNA) Quick Start Guide Copyright AKELA, Inc. 2012, all rights reserved http:\\akelainc.com LabVIEW and LabWindows are registered trademarks of National Instruments Incorporated

More information

USER MANUAL. RC-43SL 6-Button Room Controller MODEL: P/N: Rev 3.

USER MANUAL. RC-43SL 6-Button Room Controller MODEL: P/N: Rev 3. USER MANUAL MODEL: RC-43SL 6-Button Room Controller P/N: 2900-300450 Rev 3 www.kramerav.com Contents 1 Introduction 1 2 Getting Started 2 2.1 Achieving the Best Performance 2 2.2 Safety Instructions

More information

Model 7705 Control Module

Model 7705 Control Module www.keithley.com Model 7705 Control Module User s Guide PA-696 Rev. D / October 2006 A G R E A T E R M E A S U R E O F C O N F I D E N C E Safety Precautions The following safety precautions should be

More information

RoHS COMPLIANT 2002/95/EC

RoHS COMPLIANT 2002/95/EC Superlum Broadband Light Sources cblmd-series (2nd Generation) Compact Broadband Light Technical Product Specification Document Number SL.3328.00.000D3 June 2017 Revision 001 ATTENTION ELECTROSTATIC SENSITIVE

More information

OLS Series Light Sources, OPM Series Optical Power Meters, and Related Test Kits User s Guide

OLS Series Light Sources, OPM Series Optical Power Meters, and Related Test Kits User s Guide OLS Series Light Sources, OPM Series Optical Power Meters, and Related Test Kits User s Guide 2004-2009, AFL Telecommunications, all rights reserved. COM4-00-1001 Revision E, 2009-06-16 Specifications

More information

Stepper Motor Driver Board. Instruction Note

Stepper Motor Driver Board. Instruction Note Stepper Motor Driver Board Instruction Note August 31, 2012 General Photonics Corp. Ph: (909) 590-5473 5228 Edison Ave. Fax: (909) 902-5536 Chino, CA 91710 USA www.generalphotonics.com Document #: GP-IN-MDL-003-CONTROLB-10

More information

USER MANUAL. RC-43SL 6-Button Room Controller MODEL: P/N: Rev 1.

USER MANUAL. RC-43SL 6-Button Room Controller MODEL: P/N: Rev 1. USER MANUAL MODEL: RC-43SL 6-Button Room Controller P/N: 2900-300450 Rev 1 www.kramerav.com Contents 1 Introduction 1 2 Getting Started 2 2.1 Achieving the Best Performance 2 2.2 Safety Instructions

More information

Installing and Managing the Switch

Installing and Managing the Switch CHAPTER 2 This chapter describes how to install and manage the Cisco SFS 7008 system hardware and contains these sections: Safety, page 2-2 Preparing the Site, page 2-3 Rack-Mounting the Switch, page 2-4

More information

1:N REDUNDANT SWITCHING UNIT REMOTE CONTROL TERMINAL

1:N REDUNDANT SWITCHING UNIT REMOTE CONTROL TERMINAL MITEQ TECHNICAL NOTE 25T051 OCTOBER 2001 REV B 1.1 GENERAL DESCRIPTION 1.1.1 PHYSICAL 1:N REDUNDANT SWITCHING UNIT REMOTE CONTROL TERMINAL SECTION 1 INTRODUCTION Figure 1-1. Front Panel, 1:N Redundant

More information

RS-232 Fiber Link Card System

RS-232 Fiber Link Card System USER GUIDE The leader in rugged fiber optic technology. U-152 2018B-0406 Fiber Link Card System SYSTEM INSTALLATION INFORMATION Introduction The Fiber Link Card system transports a full 9-Pin copper signal

More information

Product Manual. USB BAY-4 Port Serial DB- 9 RS-232 Adapter with FTDI Chipset. Coolgear, Inc. Version 1.1 September 2017 Model Number: USBG-BAY4

Product Manual. USB BAY-4 Port Serial DB- 9 RS-232 Adapter with FTDI Chipset. Coolgear, Inc. Version 1.1 September 2017 Model Number: USBG-BAY4 USB BAY-4 Port Serial DB- 9 RS-232 Adapter with FTDI Chipset Product Manual Coolgear, Inc. Version 1.1 September 2017 Model Number: USBG-BAY4 2 USBG-BAY4 Product Manual Revision History Revision Date Author

More information

KI Channel Optical CWDM Power Meter. User Manual. KI6503 series 48-Channel Optical DWDM Power Meter

KI Channel Optical CWDM Power Meter. User Manual. KI6503 series 48-Channel Optical DWDM Power Meter User Manual KI6503 series 48-Channel Optical DWDM Power Meter Warranty: Information in this manual is given in good faith for the benefit of the user. It cannot be used as the basis for claims against

More information

Model 2460-KIT. Screw Terminal Connector Kit. Description / September 2014 *P * 1

Model 2460-KIT. Screw Terminal Connector Kit. Description / September 2014 *P * 1 Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.keithley.com Model 2460-KIT Screw Terminal Connector Kit Description The Model 2460-KIT Screw Terminal Connector Kit

More information

Fiber Optic Distributed Temperature Sensors (B-DTS)

Fiber Optic Distributed Temperature Sensors (B-DTS) Fiber Optic Distributed Temperature Sensors (B-DTS) Low-cost Brillouin BOTDA scattering version For more information about our strain and temperature sensor system and related products, please visit www.ozoptics.com

More information

USB to Serial Converter User s Guide

USB to Serial Converter User s Guide USB to Serial Converter User s Guide Important Note! In order to minimize possible installation problems and/or resource conflicts: Read Me First! About This User s Guide This User s Guide is designed

More information

BIPOLAR OPERATIONAL POWER SUPPLY

BIPOLAR OPERATIONAL POWER SUPPLY QUICK START GUIDE I INTRODUCTION KEPCO An ISO 9001 Company. BIPOLAR OPERATIONAL POWER SUPPLY 1.1. SCOPE OF MANUAL. This Quick Start Guide covers simple installation and local operation of the Kepco 100W,

More information

INSTALLATION INSTRUCTIONS FOR THE BV10-100/1000

INSTALLATION INSTRUCTIONS FOR THE BV10-100/1000 INSTALLATION INSTRUCTIONS FOR THE BV10-100/1000 This document describes the basic steps for installing your BV10-100 or BV10-1000. For detailed information about the BV10-100/1000, see the Ethernet Performance

More information

OPERATOR MANUAL OSD157 FIBER OPTIC RS422/RS232 MODEM/MULTIPLEXER

OPERATOR MANUAL OSD157 FIBER OPTIC RS422/RS232 MODEM/MULTIPLEXER / OPERATOR MANUAL OSD157 FIBER OPTIC RS422/RS232 MODEM/MULTIPLEXER PAGE 2 OF 17 INDEX 1 1. TECHNICAL SUMMARY... 4 1.1 BRIEF DESCRIPTION... 4 1.1.1 OVERVIEW... 4 1.1.2 APPLICATIONS... 4 1.1.3 FEATURES AND

More information

CMU 100 RS485 to 7x, 6x, 5x, 4x Multimode Fiber Optic Star Coupler User Manual

CMU 100 RS485 to 7x, 6x, 5x, 4x Multimode Fiber Optic Star Coupler User Manual CMU 100 RS485 to 7x, 6x, 5x, 4x Multimode Fiber Optic Star Coupler User Manual CMU 100 / 2.5.6.6.6.6.6.6.6 12, CMU 100 / 8.5.6.6.6.6.6.6.6 12, CMU 100 / 2.5.6.6.6.6.6.6 12, CMU 100 / 8.5.6.6.6.6.6.6 12,

More information

Model 2380 Rack-Mount Kit

Model 2380 Rack-Mount Kit Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.tek.com/keithley Model 2380 Rack-Mount Kit Installation Instructions Introduction The Model 2380 Fixed Rack-Mount

More information

RS-232 Industrial Media Converter

RS-232 Industrial Media Converter USER GUIDE The leader in rugged fiber optic technology. U-151 2018C-0327 Industrial Media Converter COMPACT, RUGGED & TEMPERATURE HARDENED Introduction This Industrial Serial Data media converter transports

More information

CARE AND MAINTENANCE

CARE AND MAINTENANCE CARE AND MAINTENANCE Cleaning the painted or plastic surfaces. Avoid the use of any organic solvents (such as thinner, alcohol, ether, etc.) to clean the painted or plastic surfaces of the accessory. Instead,

More information

GV3000/SE Operator Interface Module (OIM) User Guide Version 2.0 M/N 2RK3000

GV3000/SE Operator Interface Module (OIM) User Guide Version 2.0 M/N 2RK3000 GV3000/SE Operator Interface Module (OIM) User Guide Version 2.0 M/N 2RK3000 Instruction Manual D2-3342-2 The information in this manual is subject to change without notice. Throughout this manual, the

More information

Economy Single Channel Output DC Power Supply

Economy Single Channel Output DC Power Supply Economy Single Channel Output DC Power Supply User Manual www.owon.com.cn Feb. 2019 edition V1.1.0 Copyright LILLIPUT Company. All rights reserved. The LILLIPUT's products are under the protection of the

More information

User s manual 19 1U RACKMOUNT CONTROL CENTER. Rack-KVM in-1 TABLE OF CONTENTS INTRODUCTION... (KVM switch, Keyboard, LCD display, touch pad)

User s manual 19 1U RACKMOUNT CONTROL CENTER. Rack-KVM in-1 TABLE OF CONTENTS INTRODUCTION... (KVM switch, Keyboard, LCD display, touch pad) Rack-KVM9000 User s manual 4-in-1 (KVM switch, Keyboard, LCD display, touch pad) 19 1U RACKMOUNT CONTROL CENTER Rev 1.1 TABLE OF CONTENTS INTRODUCTION... FEATURES.... PACKAGE CONTENTS..... TECHNICAL SPECIFICATIONS...

More information

Edition:2A ODM- 88. DVI Matrix Router USER MANUAL

Edition:2A ODM- 88. DVI Matrix Router USER MANUAL Edition:2A ODM- 88 DVI Matrix Router USER MANUAL ODM-88 Manual Page 2 ODM-88 Manual Page 3 Table of Contents 1 Introduction and Installation. 6 1.1 Key features... 6 1.2 ODM-88 Shipping Content... 6 1.3

More information

Operation Manual for OLS-1315 Optical Light Source

Operation Manual for OLS-1315 Optical Light Source Operation Manual for OLS-1315 Optical Light Source www.tonercable.com Table of Contents 1. Introduction... 1,2 1.1 Summary... 1 1.2 Product Features... 2 1.3 Technical Parameters... 2 1.4 Main Applications...

More information

POWER AVAILABILITY 2U POD USER MANUAL. Power Output Distribution 120 Volt

POWER AVAILABILITY 2U POD USER MANUAL. Power Output Distribution 120 Volt POWER AVAILABILITY 2U POD USER MANUAL Power Output Distribution 120 Volt TABLE OF CONTENTS IMPORTANT SAFETY INSTRUCTIONS.......................................... 1 GLOSSARY OF SYMBOLS..................................................

More information

Series 370 S SoftwSystem are Qui Switch/Mult ck Start G imeter uide Quick Start Guide

Series 370 S SoftwSystem are Qui Switch/Mult ck Start G imeter uide Quick Start Guide Series ACS 3700A Software System Quick Switch/Multimeter Start Guide Quick Start Guide Safety precautions Observe the following safety precautions before using this product and any associated instrumentation.

More information

Model: VA/230W UPS Backup System

Model: VA/230W UPS Backup System IMPORTANT SAFETY INSTRUCTIONS (SAVE THESE INSTRUCTIONS) This manual contains important safety instructions. Please read and follow all instructions carefully during installation and operation of unit.

More information

Handheld Optical Power Meter. Model SSF-TKITP-400 OPTICAL POWER METER. Manual

Handheld Optical Power Meter. Model SSF-TKITP-400 OPTICAL POWER METER. Manual Model SSF-TKITP-400 OPTICAL POWER METER Manual August 2016 Contents Chapter 1. Standard Configuration.....2 Chapter 2. Overview...........2 Chapter 3. Data Sheet.....3 Chapter 4. Function.........4 4.1

More information