CT imaging using energy-sensitive photon-counting detectors. Disclosure. Vision 20/20 paper

Size: px
Start display at page:

Download "CT imaging using energy-sensitive photon-counting detectors. Disclosure. Vision 20/20 paper"

Transcription

1 CT imaging using energy-sensitive photon-counting detectors Katsuyuki Ken Taguchi, Ph.D. Division of Medical Imaging Physics Department of Radiology Johns Hopkins University No financial interest Research grants Siemens Healthcare Disclosure Past relationships/grants Former employee of Toshiba Medical Systems Former Co-I of a project funded by Philips Healthcare AHA, NIH R01 and SBIRs (DxRay) Consultant Life Saving Imaging Technology (LISIT) (Tokyo, Japan) 2 Vision 20/20 paper Detector technology Imaging technology System technology Potential clinical benefits Med Phys 40(10), (19 pages) (2013) 1

2 Number of photons n (E) Outline Current CT detectors to higher dose exams Low-dose CT by integrating 3 technologies Photon counting detectors (PCDs) Iterative reconstruction for PCDs (PIECE) Joint estimation with tissue types (JE-MAP) Whole-body prototype PCD-CT system Other clinical merits of PCD-CT 4 Major problems of current CT Relatively high-dose procedure Insufficient contrast between tissues CT images are not tissue-type specific Pixel values, Hounsfield units or linear atten. coeff., are not quantitative but qualitative 5 Energy weighted integration 1.5E E E E Energy [kev] Incident Transmitted Weighted by E 6 2

3 Energy binning for spectral CT Linear attenuation coefficient [cm -1 ] Transmitted x-ray spectrum [A.U.] 1.5 Rib Bismuth 1.40E-03 Blood only 1.20E-03 Iodine-mixed blood u(e, blood) Energy [kev] 1.00E-03 u(e, Gd) u(e, dry_spine) Gd-mixed blood u(e, dry_rib) 8.00E-04 u(e, iodine) Gd u(e, bismuth) Bi-mixed blood 6.00E-04 Gd-blood (0.26%) I-blood (0.49%) Spine Bi-blood (0.49%) 4.00E-04 Iodine 2.00E-04 Water Four materials would provide an identical pixel value if scanned with current CT 0.00E Energy [kev] Transmitted spectra with 25 cm water and 5 cm blood w/ or w/o contrast agents 7 Strategy for low-dose CT Combine the following 3 technologies: Technologies Relative amount of dose reduction Liver CT (msv) Current CT system with FBP *1 N/A 10 Photon counting detectors (PCDs) % 6 7 Iterative reconstruction for PCDs (PIECE) >35 % Joint estimation with tissue types (JE-MAP) >35 % Note 1. Dose for current CT with iterative reconstruction may be 6 8 msv Note 2. Annual background radiation: 3.1 msv 8 Strategy for low-dose CT Combine the following 3 technologies: Technologies Relative amount of dose reduction Liver CT (msv) Current CT system with FBP *1 N/A 10 Photon counting detectors (PCDs) % 6 7 Iterative reconstruction for PCDs (PIECE) >35 % Joint estimation with tissue types (JE-MAP) >35 % Note 1. Dose for current CT with iterative reconstruction may be 6 8 msv Note 2. Annual background radiation: 3.1 msv 9 3

4 Photon counting detector (PCD) Incident x-ray Amplifier Photon to charge Charge measured, counted, and stored digitally Counting processing Courtesy of W. Barber, Ph.D. and J. Iwanczyk, Ph.D. (DxRay, Inc) 10 Detection mechanism Thresholds to pulse height comparators Comparators DAC Counter 1 Digital output Input analog pulses Shaper DAC Counter 2 Digital output Preamp DAC Counter N Digital output Pulse height = photon Threshold #3 energy N 3 Threshold #2 N 2 n 3 = N 3 n 2 = N 2 N 3 Threshold #1 N 1 n 1 = N 1 N 2 Time 11 Spectral response (SR) Incident x-ray photons (a) (b) (c) (d) Sensor Anode A 90 kev photon 60 kev time 30 kev Charge sharing time 12 4

5 Probability Output count rate [Mcps] Counts Probability [a.u.] Spectral response (SR) Incident x-ray photons (a) (b) (c) (d) Sensor Anode Response to 59.5 kev photons (Am-241 source) Noise floor Continuum: Photo peak - Charge trapping - K-escape - Charge sharing - Compton scatter Incident spectrum Recorded spectrum Energy [kev] Energy [kev] Pulse height (photon energy) Observed pulses True pulses Pulse pileup Count rate loss Non-paralyzable detector ( =20 ns) True 10 Paralyzable detector ( =20 ns) Input count rate [Mcps] Dead Live Time Pulse pileup spectrum True Recorded Time Events on detector Energy [kev] 14 Spectral distortion Spectral response (SR): Charge sharing, K-escape, Compton, etc. Larger, slower PCD Incident x-ray photons (a) (b) (c) (d) Sensor Anode Pulse pileups Smaller, faster PCD Pulse height (photon energy) Observed pulses True pulses 15 5

6 Strategy for low-dose CT Combine the following 3 technologies: Technologies Relative amount of dose reduction Liver CT (msv) Current CT system with FBP *1 N/A 10 Photon counting detectors (PCDs) % 6 7 Iterative reconstruction for PCDs (PIECE) >35 % Joint estimation with tissue types (JE-MAP) >35 % Note 1. Dose for current CT with iterative reconstruction may be 6 8 msv Note 2. Annual background radiation: 3.1 msv 16 Image artifacts due to SR and PP X-ray focus Bowtie filters n 0 (E) Object, x(e) Energy, E PCD n t (E) Ray PCD pixels Observed spectrum n R (E) n t (E) Energy, E Recorded counts y={y 1, y 2, y 3, y 4 } Reconstruct images Energy, E x={x 1, x 2, x 3, x 4 } 17 Image artifacts due to SR and PP x 2 X-ray focus Bowtie filters n 0 (E) x E = Object, x(e) m Energy, E PCD w m Ψ m E PCD pixels Ray [cm -1 ] n t (E) Observed spectrum n R (E) n t (E) Photoelectric Energy, E Compton Scattering Recorded counts K-edge(Iodine) y={y 1, y 2, y 3, y 4 } Reconstruct images Energy, E x={x 1, x 2, x 3, x 4 } Photon energy [kev] 18 6

7 COV W (%) Counts/keV Counts/keV SRE+PPE, P, unipolar, no Gd SRE x DL, P, unipolar, no Gd SRE, P, unipolar, no Gd trans x DL, P, unipolar, no Gd trans, P, unipolar, no Gd Counts/keV X-ray focus PCD compensation PIECE = Physics-modeled Iterative reconstruction for Energy-sensitive photon Counting detector PCD Line integrals, Bowtie v W j filters n 0 (E) n t (E) Ray PCD pixels Observed spectrum n R (E) n t (E) Energy, E Recorded counts y={y 1, y 2, y 3, y 4 } Energy, E p y v Energy, E Estimate object v by maximizing the likelihood 19 Energy Spectral response Pulse pileups Recorded spectrum Energy 1: J. Cammin, Med. Phys. 41: (2014) 2: K. Taguchi, Med. Phys., 37: 3957 (2010) 3: K. Taguchi, Med. Phys., 38: 1089 (2011) PCD model Transmitted spectrum Spectral response 1 n SRE E = a SRE E, E k n t E k k or n SRE = A SRE n t Pulse pileups 2,3 n R E = n SRE E Pr "rec" = 1 Pr l "rec" = 1 Pr E l l l = # of photons piled up for one recorded count 20 Monte Carlo simulation study for Siemens PCD measured SRE+PPE( (E)) 16,000 SRE, DL transmitted, DL Energy (kev) Energy (kev) 200 ma 800 ma 25 ma 800 ma DE onlyh 2 O, Al P, unipolar I t = 25 ma SRE+DE DLR = 0.9% COV = 2.4% SRE+pileups W PCD data measured SRE+PPE( (E)) DE onlysre, DL transmitted, DL SRE+DE H 2 O, Al P, unipolar I t = 800 ma SRE+pileups DLR = 25.3% COV = 1.5% W PCD data Energy (kev) Energy (kev) 100 mm water, 3.5 mm Al, 0.9 mm Ti; 120 kvp; pulse height analyzer; unipolar pulse deadtime loss ratio: 1-n recorded/n t (%) 100-DE (%) DE = Detection efficiency (%) COVw = Weighted coefficient of variation (%) J. Cammin, presented at IEEE NSS/MIC

8 Evaluation of PIECE-1 CNR 6.0 CNR 6.0 CNR 4.1 CNR Strategy for low-dose CT Combine the following 3 technologies: Technologies Relative amount of dose reduction Liver CT (msv) Current CT system with FBP *1 N/A 10 Photon counting detectors (PCDs) % 6 7 Iterative reconstruction for PCDs (PIECE) >35 % Joint estimation with tissue types (JE-MAP) >35 % Note 1. Dose for current CT with iterative reconstruction may be 6 8 msv Note 2. Annual background radiation: 3.1 msv 23 Joint estimation with tissue types Sinogram y Characteristic coefficients w Tissue type z E Projection noise K. Nakada, et al., IEEE MIC (2013); Med Phys (2015) 24 8

9 Joint estimation with tissue types Sinogram y Characteristic coefficients w Tissue type z E Projection noise Geometry Statistical relationship 0.20 w pe 0.10 Iodine-mixed blood z i = z j = blood 2 α w i w j rib Smooth lung fat z i z spine j β blood 0.00 muscle Discontinuous liver muscle location w K. Nakada, et al., IEEE MIC (2013); Med Phys (2015) cs 25 Joint estimation with tissue types Sinogram y Characteristic coefficients w Tissue type z E Projection noise Geometry Statistical relationship w z i = z j = blood α w i w j 2 Smooth z i z j z β blood muscle Discontinuous location K. Nakada, et al., IEEE MIC (2013); Med Phys (2015) 26 Truth FBP PML K. Nakada, CERN wrokshop (2015) WW 400HU, WL -50HU 27 9

10 Truth FBP PML K. Nakada, CERN wrokshop (2015) WW 400HU, WL -50HU 28 Joint estimation with tissue types PML FBP Noise reduction at FWHM=3 pixels by 30%: FBP PML by 33%: PML JE-MAP by 53%: FBP JE-MAP JE-MAP K. Nakada, et al., IEEE MIC (2013) 29 Photon counting low-dose CT Current CT detectors lead to higher dose exams Low-dose CT by integrating 3 technologies Photon counting detectors (PCDs) Iterative reconstruction for PCDs (PIECE) Joint estimation with tissue types (JE-MAP) Technologies Relative amount of dose reduction Liver CT (msv) Current CT system with FBP *1 N/A 10 Photon counting detectors (PCDs) % 6 7 Iterative reconstruction for PCDs (PIECE) >35 % Joint estimation with tissue types (JE-MAP) >35 %

11 Outline Current CT detectors to higher dose exams Low-dose CT by integrating 3 technologies Photon counting detectors (PCDs) Iterative reconstruction for PCDs (PIECE) Joint estimation with tissue types (JE-MAP) Whole-body prototype PCD-CT system Other clinical merits of PCD-CT 31 Whole-body prototype PCD-CT system Developed by Siemens, installed at Mayo Clinic Dual-source CT, EID for 50 cm, PCD for 27.5 cm (225 m) 2 pixel, 2 thresholds (staggered 4 thresholds) 128x10 6 counts/s/mm 2 with 13.5% loss, 256x10 6 with 25.2% loss Shading artifacts due to energetic cross-talks (not shown) Cadavers In-vivo swines Courtesy of Cynthia H. McCollough, Ph.D. (Mayo Clinic) 32 Clinical benefits of spectral CT Improvement of current CT images 1) Contrast-to-noise ratio (CNR) and contrast of CT images, or doses of radiation & contrast agents 2) Quantitative CT imaging 3) Material- or tissue type-specific imaging 4) Accurate K-edge CT imaging 5) Simultaneous multi-agent imaging 6) Molecular CT with nanoparticles 7) Personalized medicine New class of CT imaging K Taguchi, Med Phys 40(10), (19 pages) (2013) 33 11

12 Molecular CT with nanoparticle contrast agents or probes Bi D. Pan, Angew Chem Int Ed 49: (2010) 34 Imaging macrophages in atherosclerotic plaques Nano agent Nano agent Conventional Injection F. Hyafil, Nat Med 13(5): (2007) Nano agent Conventional agent 35 The use of PCDs and coded-aperture for simultaneous absorption, phase, and differential phase contrast imaging Low-dose single scan followed by a novel retrieval algorithm 5cm average breast tissue 1.5 mgy dose 6 energy bins, CdTe detector M. Das, AAPM 2015, TH-AB (8:30 am) Absorption Phase Diff. phase 36 12

13 Summary Current CT detectors cause the major limitations of CT images PCDs address them and enable new applications Low-dose CT by integrating 3 technologies Photon counting detectors (PCDs) Iterative reconstruction for PCDs (PIECE) Joint estimation with tissue types (JE-MAP) A few whole-body prototype PCD-CT systems being evaluated at hospitals 37 Acknowledgment Kenji Amaya, Ph.D. (TITech) Kento Nakada, B.Sc. (TITech) Cynthia H. McCollough, Ph.D. (Mayo Clinic) Jan S. Iwanczyk, Ph.D. (DxRay, Inc) William Barber, Ph.D. (DxRay, Inc) Neal E. Harsough, Ph.D. (DxRay, Inc) Steffen Kappler, Ph.D. (Siemens Healthcare) Thomas G. Flohr, Ph.D. (Siemens Healthcare) Karl Stierstorfer, Ph.D. (Siemens Healthcare) Eric C. Frey, Ph.D. (JHU) Jochen Cammin, Ph.D. (JHU) Somesh Srivastava, Ph.D. (JHU) Benjamin M.W. Tsui, Ph.D. (JHU) Current and former students and colleagues in DMIP and JHU 38 13

14 Thresholds to pulse height comparators DAC Comparators Counter 1 Digital output Input analog pulses Preamp Shaper DAC Counter 2 Digital output DAC Counter N Digital output Clinical benefits of spectral CT 1) Contrast-to-noise ratio (CNR) and contrast of CT images, or doses of radiation & contrast agents 2) Quantitative CT imaging 3) Material- or tissue type-specific imaging 4) Accurate K-edge CT imaging by >2 energy windows 5) Simultaneous multi-agent imaging 41 Joint estimation with tissue types Sinogram y Characteristic coefficients w Tissue type z E Projection noise w z blood muscle location K. Nakada, et al., IEEE MIC (2013); Med Phys (2015) 42 14

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology.

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology. Photon Counting Detectors and Their Applications in Medical Imaging Principles, Pitfalls and Progress in Photon-Counting-Detector Technology Taly Gilat Schmidt, PhD Associate Professor Department of Biomedical

More information

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Joonki Noh, Jeffrey A. Fessler EECS Department, The University of Michigan Paul E. Kinahan Radiology Department,

More information

Digital Scatter Removal in Mammography to enable Patient Dose Reduction

Digital Scatter Removal in Mammography to enable Patient Dose Reduction Digital Scatter Removal in Mammography to enable Patient Dose Reduction Mary Cocker Radiation Physics and Protection Oxford University Hospitals NHS Trust Chris Tromans, Mike Brady University of Oxford

More information

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland Metal Artifact Reduction CT Techniques R S S S Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland N. 1 v o 4 1 0 2. Postoperative CT Metal Implants CT is accurate for assessment

More information

Cardiac Dual Energy CT: Technique

Cardiac Dual Energy CT: Technique RSNA 2013, VSCA51-01, Chicago, Dec. 5, 2013 Cardiac Radiology Series Cardiac Dual Energy CT: Technique Willi A. Kalender, Ph.D. Institute of Medical Physics University of Erlangen www.imp.uni-erlangen.de

More information

The Near Future in Cardiac CT Image Reconstruction

The Near Future in Cardiac CT Image Reconstruction SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de

More information

Acknowledgments and financial disclosure

Acknowledgments and financial disclosure AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories

More information

ML reconstruction for CT

ML reconstruction for CT ML reconstruction for CT derivation of MLTR rigid motion correction resolution modeling polychromatic ML model dual energy ML model Bruno De Man, Katrien Van Slambrouck, Maarten Depypere, Frederik Maes,

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application

Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application Physics in Medicine & Biology Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application To cite this article: Polad M Shikhaliev and Shannon G Fritz 2011

More information

Financial disclosure. Onboard imaging modality for IGRT

Financial disclosure. Onboard imaging modality for IGRT Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty

More information

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel Proc. Of SPIE Vol 0626, Application of Optical Instrumentation in Medicine XIV and Picture Archiving and Communication Systems(PACS IV) for Medical Applications IV, ed. S J Dwyer/R H Schneider (Jan 1986)

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

Machine Learning-based Techniques to Address Spectral Distortions in Photon Counting X-ray Computed Tomography

Machine Learning-based Techniques to Address Spectral Distortions in Photon Counting X-ray Computed Tomography Machine Learning-based Techniques to Address Spectral Distortions in Photon Counting X-ray Computed Tomography by MengHeng Touch Graduate Program in Medical Physics Duke University Date: Approved: Cristian

More information

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial

8/2/2017. Disclosure. Philips Healthcare (Cleveland, OH) provided the precommercial 8//0 AAPM0 Scientific Symposium: Emerging and New Generation PET: Instrumentation, Technology, Characteristics and Clinical Practice Aug Wednesday 0:4am :pm Solid State Digital Photon Counting PET/CT Instrumentation

More information

Photon counting spectroscopic CT with dynamic beam attenuator

Photon counting spectroscopic CT with dynamic beam attenuator Photon counting spectroscopic CT with dynamic beam attenuator Haluk Atak 1) and Polad M. Shikhaliev a) 1) Department of Nuclear Engineering, Hacettepe University, Ankara, Turkey Abstract Purpose: Photon

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Acknowledgments. Nesterov s Method for Accelerated Penalized-Likelihood Statistical Reconstruction for C-arm Cone-Beam CT.

Acknowledgments. Nesterov s Method for Accelerated Penalized-Likelihood Statistical Reconstruction for C-arm Cone-Beam CT. June 5, Nesterov s Method for Accelerated Penalized-Likelihood Statistical Reconstruction for C-arm Cone-Beam CT Adam S. Wang, J. Webster Stayman, Yoshito Otake, Gerhard Kleinszig, Sebastian Vogt, Jeffrey

More information

Computer-Tomography II: Image reconstruction and applications

Computer-Tomography II: Image reconstruction and applications Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de

More information

Quantitative Attenuation Correction for PET/CT Using Iterative Reconstruction of Low-Dose Dual-Energy CT

Quantitative Attenuation Correction for PET/CT Using Iterative Reconstruction of Low-Dose Dual-Energy CT Quantitative Attenuation Correction for PET/CT Using Iterative Reconstruction of Low-Dose Dual-Energy CT Paul E. Kinahan, Senior Member, IEEE, Jeffrey A. Fessler, Senior Member, IEEE, Adam M. Alessio,

More information

Acknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury

Acknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury A. Sisniega et al. (presented at RSNA 214) High Performance Cone-Beam CT of Acute Traumatic Brain Injury A. Sisniega 1 W. Zbijewski 1, H. Dang 1, J. Xu 1 J. W. Stayman 1, J. Yorkston 2, N. Aygun 3 V. Koliatsos

More information

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT Assessment of 3D performance metrics D and 3D Metrics of Performance Towards Quality Index: Volumetric imaging systems X-ray based Volumetric imaging systems: CBCT/CT Tomosynthesis Samuel Richard and Ehsan

More information

Spectral CT reconstruction with an explicit photon-counting detector model: a one-step approach

Spectral CT reconstruction with an explicit photon-counting detector model: a one-step approach Spectral CT reconstruction with an explicit photon-counting detector model: a one-step approach Pierre-Antoine Rodesch, V. Rebuffel, C. Fournier, Florence Forbes, L. Verger To cite this version: Pierre-Antoine

More information

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation 26 IEEE Nuclear Science Symposium Conference Record M6-349 Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation Habib Zaidi, Senior Member, IEEE and Mohammad

More information

Background 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J.

Background 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging Stephen J. Glick* J. Michael O Connor**, Clay Didier**, Mini Das*, * University of Massachusetts Medical

More information

Limits of Dose Reduction in CT: How low is too low? Acknowledgement. Disclosures 8/2/2012

Limits of Dose Reduction in CT: How low is too low? Acknowledgement. Disclosures 8/2/2012 8/2/22 Limits of Dose Reduction in CT: How low is too low? Ehsan Samei Duke University Acknowledgement Disclosures Research grant: NIH Research grant: DHS Research grant: RSNA, QIBA Research grant: General

More information

Philips SPECT/CT Systems

Philips SPECT/CT Systems Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET

More information

Prospects for spectral CT with Medipix detectors

Prospects for spectral CT with Medipix detectors Nikhef, Amsterdam, The Netherlands E-mail: e.schioppa@nikhef.nl Jan Visser Nikhef, Amsterdam, The Netherlands E-mail: janvs@nikhef.nl Els Koffeman Nikhef, Amsterdam, The Netherlands E-mail: e.koffeman@nikhef.nl

More information

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Stefan C. Saur 1, Hatem Alkadhi 2, Luca Regazzoni 1, Simon Eugster 1, Gábor Székely 1, Philippe Cattin 1,3 1 Computer Vision

More information

GPU Based Convolution/Superposition Dose Calculation

GPU Based Convolution/Superposition Dose Calculation GPU Based Convolution/Superposition Dose Calculation Todd McNutt 1, Robert Jacques 1,2, Stefan Pencea 2, Sharon Dye 2, Michel Moreau 2 1) Johns Hopkins University 2) Elekta Research Funded by and Licensed

More information

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti Monte Carlo methods in proton beam radiation therapy Harald Paganetti Introduction: Proton Physics Electromagnetic energy loss of protons Distal distribution Dose [%] 120 100 80 60 40 p e p Ionization

More information

Fast Timing and TOF in PET Medical Imaging

Fast Timing and TOF in PET Medical Imaging Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

Hot Topics in Imaging Physics A Key Interdisciplinary Science for 21 st Century Medicine

Hot Topics in Imaging Physics A Key Interdisciplinary Science for 21 st Century Medicine Hot Topics in Imaging Physics A Key Interdisciplinary Science for 21 st Century Medicine Jeff Siewerdsen, PhD FAAPM Department of Biomedical Engineering Department of Computer Science Russell H. Morgan

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

Fits you like no other

Fits you like no other Fits you like no other BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase in room

More information

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004 Spiral ASSR Std p = 1.0 Spiral EPBP Std p = 1.0 Kachelrieß et al., Med. Phys. 31(6): 1623-1641, 2004 256 slices (0/300) Advantages of Cone-Beam Spiral CT Image quality nearly independent of pitch Increase

More information

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation

More information

Performance Evaluation of CT Systems

Performance Evaluation of CT Systems 8/2/206 Performance Evaluation of CT Systems AAPM Task Group 233 Foundations, Methods, and Prospects Ehsan Samei Duke University Medical Center Disclosures Research grant: NIH R0 EB00838 Research grant:

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION 2009 IEEE Nuclear Science Symposium Conference Record M03-7 FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION Wonseok Huh, Jeffrey A. Fessler, Adam M. Alessio, and Paul E. Kinahan Department

More information

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO Abstract C.S. Levin, Y-C Tai, E.J. Hoffman, M. Dahlbom, T.H. Farquhar UCLA School of Medicine Division

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

in PET Medical Imaging

in PET Medical Imaging Fast Timing and TOF in PET Medical Imaging William W. Moses Lawrence Berkeley National Laboratory October 15, 2008 Outline: Time-of-Flight PET History Present Status Future This work was supported in part

More information

DUAL-ENERGY CT IN PROTON THERAPY

DUAL-ENERGY CT IN PROTON THERAPY 10/31/17 DUAL-ENERGY CT IN PROTON THERAPY Isabel Almeida, MAASTRO Clinic 7th NCS Lustrum Symposium 1 10/31/17 http://zonptc.bouwwebcam.nl https://www.youtube.com/watch?v=3vvvf5bqn7g Range uncertainties

More information

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Qiao Yang 1,4, Meng Wu 2, Andreas Maier 1,3,4, Joachim Hornegger 1,3,4, Rebecca Fahrig

More information

CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona

CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona Introduction CBCT has become a routine procedure for

More information

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Tecniques of imaging in Overview Introduction Tecniques of imaging

More information

CT vs. VolumeScope: image quality and dose comparison

CT vs. VolumeScope: image quality and dose comparison CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,

More information

Computer-Tomography I: Principles, History, Technology

Computer-Tomography I: Principles, History, Technology Computer-Tomography I: Principles, History, Technology Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de History

More information

Michael Speiser, Ph.D.

Michael Speiser, Ph.D. IMPROVED CT-BASED VOXEL PHANTOM GENERATION FOR MCNP MONTE CARLO Michael Speiser, Ph.D. Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX September 1 st, 2012 CMPWG Workshop Medical

More information

Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI. Floris Jansen, GE Healthcare July, 2015

Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI. Floris Jansen, GE Healthcare July, 2015 Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI Floris Jansen, GE Healthcare July, 2015 PET/MR 101 : challenges Thermal Workflow & Apps RF interactions?!!

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

Enhanced material contrast by dual-energy microct imaging

Enhanced material contrast by dual-energy microct imaging Enhanced material contrast by dual-energy microct imaging Method note Page 1 of 12 2 Method note: Dual-energy microct analysis 1. Introduction 1.1. The basis for dual energy imaging Micro-computed tomography

More information

Michal E. Kulon, MD 1,2 1. Peter Komlosi, MD, PhD 3 3. Radiology Universe Institute, 2

Michal E. Kulon, MD 1,2 1. Peter Komlosi, MD, PhD 3 3. Radiology Universe Institute, 2 Blend+Proximity, a novel algorithm achieves high suppression of metallic streak artifacts and maximal preservation of contrast between soft tissues and iodinated contrast material on dual-energy CT scans

More information

Gamma spectroscopic measurements using the PID350 pixelated CdTe radiation detector

Gamma spectroscopic measurements using the PID350 pixelated CdTe radiation detector Gamma spectroscopic measurements using the PID350 pixelated CdTe radiation detector K. Karafasoulis, K. Zachariadou, S. Seferlis, I. Papadakis, D. Loukas, C. Lambropoulos, C. Potiriadis Abstract Spectroscopic

More information

Advanced Multi Material Decomposition of Dual Energy in Computed Tomography Image

Advanced Multi Material Decomposition of Dual Energy in Computed Tomography Image Advanced Multi Material Decomposition of Dual Energy in Computed Tomography Image A.Prema 1, M.Priyadharshini 2, S.Renuga 3, K.Radha 4 UG Students, Department of CSE, Muthayammal Engineering College, Rasipuram,

More information

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead.

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead. 1 CT Basics Principles of Spiral CT Dose 2 Who invented CT? 1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections Sir Godfrey Hounsfield worked for the Central

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

Introduction to Positron Emission Tomography

Introduction to Positron Emission Tomography Planar and SPECT Cameras Summary Introduction to Positron Emission Tomography, Ph.D. Nuclear Medicine Basic Science Lectures srbowen@uw.edu System components: Collimator Detector Electronics Collimator

More information

arxiv: v1 [physics.med-ph] 18 Apr 2016

arxiv: v1 [physics.med-ph] 18 Apr 2016 Segmentation-free x-ray energy spectrum estimation for computed tomography arxiv:164.4986v1 [physics.med-ph] 18 Apr 216 Wei Zhao a, Qiude Zhang a, Tianye Niu b a Department of Biomedical Engineering, Huazhong

More information

GPU implementation for rapid iterative image reconstruction algorithm

GPU implementation for rapid iterative image reconstruction algorithm GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial

More information

Fits you like no other

Fits you like no other Fits you like no other Philips BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase

More information

DUAL energy CT (DECT) is a modality where one and. Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography

DUAL energy CT (DECT) is a modality where one and. Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography Marc Kachelrieß, Member, IEEE, Timo Berkus, Philip Stenner, Willi A. Kalender Abstract Material selective imaging using dual energy

More information

Some reference material

Some reference material Some reference material Physics reference book on medical imaging: A good one is The Essential Physics of Medical Imaging, 3 rd Ed. by Bushberg et al. ($170! new). However, there are several similar books

More information

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,

More information

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Workshop on Quantitative SPECT and PET Brain Studies 14-16 January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Físico João Alfredo Borges, Me. Corrections in SPECT and PET SPECT and

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

Correlation between Model and Human Observer Performance on a Lesion Shape Discrimination Task in CT

Correlation between Model and Human Observer Performance on a Lesion Shape Discrimination Task in CT Correlation between Model and Human Observer Performance on a Lesion Shape Discrimination Task in CT Yi Zhang, Shuai Leng, Lifeng Yu and Cynthia McCollough Department of Radiology Mayo Clinic, Rochester

More information

Motion Compensation from Short-Scan Data in Cardiac CT

Motion Compensation from Short-Scan Data in Cardiac CT Motion Compensation from Short-Scan Data in Cardiac CT Juliane Hahn 1,2, Thomas Allmendinger 1, Herbert Bruder 1, and Marc Kachelrieß 2 1 Siemens Healthcare GmbH, Forchheim, Germany 2 German Cancer Research

More information

Improved activity estimation with MC-JOSEM versus TEW-JOSEM in 111 In SPECT

Improved activity estimation with MC-JOSEM versus TEW-JOSEM in 111 In SPECT Improved activity estimation with MC-JOSEM versus TEW-JOSEM in 111 In SPECT Jinsong Ouyang, a Georges El Fakhri, and Stephen C. Moore Department of Radiology, Harvard Medical School and Brigham and Women

More information

Tomographic Reconstruction

Tomographic Reconstruction Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)

More information

CT: Physics Principles & Equipment Design

CT: Physics Principles & Equipment Design CT: Physics Principles & Equipment Design James Kofler, Ph.D Radiology Mayo Clinic Rochester, MN June 27, 2012 Disclosures Nothing to disclose Learning Objectives Understand fundamental concepts of - CT

More information

Outline. Outline 7/24/2014. Fast, near real-time, Monte Carlo dose calculations using GPU. Xun Jia Ph.D. GPU Monte Carlo. Clinical Applications

Outline. Outline 7/24/2014. Fast, near real-time, Monte Carlo dose calculations using GPU. Xun Jia Ph.D. GPU Monte Carlo. Clinical Applications Fast, near real-time, Monte Carlo dose calculations using GPU Xun Jia Ph.D. xun.jia@utsouthwestern.edu Outline GPU Monte Carlo Clinical Applications Conclusions 2 Outline GPU Monte Carlo Clinical Applications

More information

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives 1. Be familiar with recommendations

More information

A closer look at CT scanning

A closer look at CT scanning Vet Times The website for the veterinary profession https://www.vettimes.co.uk A closer look at CT scanning Author : Charissa Lee, Natalie Webster Categories : General, Vets Date : April 3, 2017 A basic

More information

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed

More information

arxiv: v2 [cond-mat.mtrl-sci] 5 Jan 2010

arxiv: v2 [cond-mat.mtrl-sci] 5 Jan 2010 Gamma scattering scanning of concrete block for detection of voids. Shivaramu 1, Arijit Bose 2 and M. Margret 1 1 Radiological Safety Division, Safety Group, IGCAR, Kalpakaam - 63 12 (India) 2 Chennai

More information

7/13/2015 EVALUATION OF NONLINEAR RECONSTRUCTION METHODS. Outline. This is a decades-old challenge

7/13/2015 EVALUATION OF NONLINEAR RECONSTRUCTION METHODS. Outline. This is a decades-old challenge EVALUATION OF NONLINEAR RECONSTRUCTION METHODS Kyle J. Myers, Ph.D. Director, Division of Imaging, Diagnostics, and Software Reliability Office of Science and Engineering Laboratories, CDRH, FDA 2 Outline

More information

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Review of PET Physics Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Chart of Nuclides Z (protons) N (number of neutrons) Nuclear Data Evaluation Lab.

More information

Medical Photon Counting in Israel

Medical Photon Counting in Israel Clinical Use of Photon Counting Detectors in CT Medical Photon Counting in srael D spect NM Alcyone Ventri MB Module manufacturer CT Swift Reuven Levinson CT Engineering GE Healthcare Haifa, srael Jerry

More information

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar

Cherenkov Radiation. Doctoral Thesis. Rok Dolenec. Supervisor: Prof. Dr. Samo Korpar Doctoral Thesis Time-of-Flight Time-of-Flight Positron Positron Emission Emission Tomography Tomography Using Using Cherenkov Cherenkov Radiation Radiation Rok Dolenec Supervisor: Prof. Dr. Samo Korpar

More information

3-D PET Scatter Correction

3-D PET Scatter Correction Investigation of Accelerated Monte Carlo Techniques for PET Simulation and 3-D PET Scatter Correction C.H. Holdsworth, Student Member, IEEE, C.S. Levin", Member, IEEE, T.H. Farquhar, Student Member, IEEE,

More information

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid by Yingxuan Chen Graduate Program in Medical Physics Duke University Date: Approved: Lei Ren, Supervisor Fang Fang Yin, Chair

More information

AIDR 3D Iterative Reconstruction:

AIDR 3D Iterative Reconstruction: Iterative Reconstruction: Integrated, Automated and Adaptive Dose Reduction Erin Angel, PhD Manager, Clinical Sciences, CT Canon Medical Systems USA Iterative Reconstruction 1 Since the introduction of

More information

Proton dose calculation algorithms and configuration data

Proton dose calculation algorithms and configuration data Proton dose calculation algorithms and configuration data Barbara Schaffner PTCOG 46 Educational workshop in Wanjie, 20. May 2007 VARIAN Medical Systems Agenda Broad beam algorithms Concept of pencil beam

More information

Continuation Format Page

Continuation Format Page C.1 PET with submillimeter spatial resolution Figure 2 shows two views of the high resolution PET experimental setup used to acquire preliminary data [92]. The mechanics of the proposed system are similar

More information

CT Iterative Reconstruction Techniques

CT Iterative Reconstruction Techniques RECENT ADVANCES IN CT RADIATION DOSE REDUCTION TECHNIQUES CT Iterative Reconstruction Techniques Kalpana Kanal, PhD, FSCBTMR, FACR, FAAPM University of Washington Seattle, WA SCBT-MR 2017 Nashville, Tennessee

More information

8/3/2016. Outline. The EPID Strikes Back: Future EPID Technology and Applications. Active Matrix Flat-Panel Imagers (AMFPIs)

8/3/2016. Outline. The EPID Strikes Back: Future EPID Technology and Applications. Active Matrix Flat-Panel Imagers (AMFPIs) 8//6 The EPID Strikes Back: Future EPID Technology and Applications Larry E. Antonuk Department of Radiation Oncology University of Michigan, Ann Arbor Acknowledgements: Youcef El-Mohri, Qihua Zhao (U.

More information

Micro-CT Methodology Hasan Alsaid, PhD

Micro-CT Methodology Hasan Alsaid, PhD Micro-CT Methodology Hasan Alsaid, PhD Preclinical & Translational Imaging LAS, PTS, GlaxoSmithKline 20 April 2015 Provide basic understanding of technical aspects of the micro-ct Statement: All procedures

More information

NIH Public Access Author Manuscript Med Phys. Author manuscript; available in PMC 2009 March 13.

NIH Public Access Author Manuscript Med Phys. Author manuscript; available in PMC 2009 March 13. NIH Public Access Author Manuscript Published in final edited form as: Med Phys. 2008 February ; 35(2): 660 663. Prior image constrained compressed sensing (PICCS): A method to accurately reconstruct dynamic

More information

James R Halama, PhD Loyola University Medical Center

James R Halama, PhD Loyola University Medical Center James R Halama, PhD Loyola University Medical Center Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program Learning Objectives Be familiar with the tests recommended

More information

A New Approach for the Enhancement. of Dual-energy Computed Tomography Images. Kyung Kook Park

A New Approach for the Enhancement. of Dual-energy Computed Tomography Images. Kyung Kook Park A New Approach for the Enhancement of Dual-energy Computed Tomography Images by Kyung Kook Park A Dissertation Presented in Partial Fulfillment of the Requirements for the Degree Doctor of Philosophy Approved

More information

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT.

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT. Progress report: Development of assessment and predictive metrics for quantitative imaging in chest CT Subaward No: HHSN6801000050C (4a) PI: Ehsan Samei Reporting Period: month 1-18 Deliverables: 1. Deployment

More information

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization

More information

Limitations in the PHOTON Monte Carlo gamma transport code

Limitations in the PHOTON Monte Carlo gamma transport code Nuclear Instruments and Methods in Physics Research A 480 (2002) 729 733 Limitations in the PHOTON Monte Carlo gamma transport code I. Orion a, L. Wielopolski b, * a St. Luke s/roosevelt Hospital, Columbia

More information