ICBM Spinning Disc Microscope System User Notes
|
|
- Wilfred Floyd
- 5 years ago
- Views:
Transcription
1 6/17/09SCAndoriQversion1.9.1DRAFT fastmodeonly ICBM SpinningDiscMicroscopeSystemUserNotes (PerkinElmer)R2 E2 Whenyousignuponthecalendarpleaseindicate: 1)Whichobjectiveyouwilluse. 100xOilisthedefaultobjectiveonthesystem. 60xWisalsoavailable 2)Whichlaserlinesareneeded. 442,488,514,568,647nm 3)Whichdichroiccubeisneeded. CFP/YFP/CY5(labeledCFP/YFP)or FITC/Rhod/CY5(labeledrgb) 3)Ifusinglivecells,isanobjectiveheaterand/orstageheaterneeded? ICBMpersonnelwillstartthesystemupandshutitdown.Donotattemptthisyourself.Askforassistance. Facilitypersonnelwillchangeobjectivesanddichoticcubes.Donotattemptthisyourself.Youdonotwanttobeheld responsiblefordamage.askforassistance. CONTENTS GettingStarted......pg2 ObservationthroughtheOculars......pg3 RevolutionFastLZ (Fast)easier ImageWindow Rev Sutter XYZ (Slow)allowsmultifield Multifield AcquireaSubRegion DATAMANAGEMENT Thisisyourdataandyouareresponsibleforitsstorageandpreservation.TheICBMdoesnotprovidelongterm storageofuserdata. Weprovideacomputer,In Neph ToastforburningofDVDsandCDs.Pleasebringblankmediawithyou.Attheendof yourimagingsession,moveyourfilestoin Neph Toastandburn2copiesofyourdata.BlankCDsandDVDsaremuch cheaperthanthetimeandmaterialsneededtorepeatyourexperiment.afteryouhavecheckedwhetherthefileson yourcdsanddvdsarereadable,atyournextimagingsession,deletethemfromtheperkinelmerandfromin Neph Toast. 1
2 6/17/09 SC Andor iq version DRAFT fast mode only GETTING STARTED Lasers ON? (switch toward case) not on standby? (switch away from case, red arrow ) Remove stage insert. Do not bump the objective. Log in to Windows using your UserID and password. Double click the Andor iq icon to start the acquisition software. Select configuration For Basic Mode (Fast) choose (Revolution Fast LZ) For Complex Mode allows multifield, choose (Rev Sutter XYZ) If you have chosen Rev Sutter XYZ, you will sometimes get a window saying that the stage has not been registered. Check to see that the stage insert has been removed, then click START. The stage will move to its extreme positions then back to center. Replace the stage insert. You will next get a window asking you to select data path. It is best to leave this path as it is. Choose Close. Your data is saved in an internal format as you acquire images. You will save your data to another computer in a format that other programs can read at the end of your imaging session. In the Main Imaging Window Press the Select button, then choose the correct objective. (halfway down left side) Select correct optivar setting this should match the position of the optivar Unchecked = 1.0 Checked = 1.5 This must be done for the software to give you the correct pixel dimensions Optivar position shown is 1.0x. The lever to move the optivar is indicated by the large green arrowhead. To move the optivar into the 1.5 x position, gently move the lever towards the front of the microscope body as indicated by the green arrow. The large focus knob is coarse focus. The small focus knob is fine focus. The stage is moved using the joystick. Gentle motion with the joystick is preferred. Do not run the objective into the edge of the stage opening. This will damage the objective. Please be aware of the position of the objective at all times. 2
3 6/17/09 SC Andor iq version DRAFT fast mode only OBSERVATION THROUGH THE OCCULARS Transillumination Turn transillumination lamp ON and OFF with button on left side of microscope body. Control transillumination intensity with dial below the button. Turret position 5 Port position 1 Epifluorescence Turn epifluorescent illumination lamp ON and OFF using the foot pedal. If the lamp is left on, it will go off by itself after 60 seconds. Green Turret position 3 Red and Green Turret position 2 Port position 3 Set mechanical shutter to open position. Note: Mechanical shutter must be in open position for epifluorescence and for laser scanning. 3
4 6/17/09SCAndoriQversion1.9.1DRAFT fastmodeonly SpinningDiscConfocalAcquisition inrevolutionfastlz (Fast) Forspinningdiscconfocalacquisition Trans illuminationlampoff,epifluorescencelampoff Turretposition5 Portposition5 InAndoriQWindow ChooseFAST Rightclicktheimagenameandchooseedit.Thiswillbetheprefix forthissessionofimages. InAcquisition/AuxiliaryDevicesWindow Undertabs Experiment Sequence LControl checkboxtoactivate EnabledandMove1st(L=wavelength) EnabledL checkboxestoactivatedesirednumberofchannels Emission check Exposure check msisgoodstartingpoint Z controlenabled check ComputedParameter Planes CurrentL chosenchannelisactivechannel channelyouwillsee whenlivescanning Setup excitation/emission Exampleshows Excitation488/Emission525/50 Excitation568/Emission600/45 Excitation647/Emission700LP Toselectexcitationwavelength,checkdesiredbox ToselectEmissionfilter,choosefromdropdownmenu Position Filter 0 OPEN 1 525/ / longpass 4 triple 5 OPEN 6 485/54, / /30 4 Use8,9for CFP/YFPFRET Whendone PressRecord.
5 6/17/09SCAndoriQversion1.9.1DRAFT fastmodeonly LaserIntensitySettings inrevolutionfastlz (Fast InAcquisition/AuxiliaryDevicesWindow Undertabs Experiment Wavelength Excitation Adjustlaserlevels.(20%isgoodstartingpoint.) Laserforcurrentchannelwillbechecked.Goaheadand setaninitialintensitylevelforallofthelasersthatyouwill beusing.pressrecord. LeavetheEditLaserintensitywindowopensothatyoucan finetunethelaserintensityeasilylater. 5
6 6/17/09SCAndoriQversion1.9.1DRAFT fastmodeonly CameraControlSettings inrevolutionfastlz (Fast) InAcquisition/AuxiliaryDevicesWindow Undertabs DV897_BV2302Settings Exposure. Window512:512 BinningX=Y,1x1 ShutterClose Framestobeaveraged unchecked FastLambdaZ checked DisplayControl displayeveryframe EMGainenabled Checked Valuesgenerallybetween ok Ifyouhaveasuperbrightsample,you mayneedtosetemgainlower. Pre AmpGain1.00 ExposureTime beginwith ms Timeshowninthiswindowisforactive channel. Pressrecordtokeepchanges. NoteCamerastatusframeatbottomofwindow. Tempmustbe 70forimaging.Thisnormallyjust takesafewminutes.thetemperatureisusually downto 70bythetimeyoureachthispointin setup. 6
7 6/17/09SCAndoriQversion1.9.1DRAFT fastmodeonly OPTIMIZEACQUISITIONSETTINGS inrevolutionfastlz (Fast) Whatintensitylevelshouldyourimages have?itdependsonthesampleandit dependsonthetypeofexperimentthatyou aredoing.ifyouaredoingliveimagingover time,toreducephotodamageand photobleaching,youwantthelowestlaser powerandlowestexposuretimethatwill giveyouanadequateimage.youmay decidethatyoucanworkwithaslowasan intensityvalueof300.withfixedsamples thatyouwillonlyimageoncemorelaser powerandlongerexposurescanbeused. Typicallyintensityvaluesforfixedsamples varywidelybutingeneralintensityvaluesin therangeof willproduceagood image.avoidsaturation. Returntothiswindow. InAcquisition/AuxiliaryDevicesWindow Undertabs Experiment Sequence PresstheLivebuttonatthetopofthewindow.Themicroscope willbeginscanning,usingthechannelchosenincurrentl.the LivebuttonwilltogglewhenpressedtosayIdle.PressIdleto stopscanning. AtthebottomoftheImageWindow,thereisacheckboxlabeled Autoscale.Thisshouldbechecked. Withautoscalechecked,dimimagesappeargrainyandthe backgroundisgreyinsteadofblack.oversaturatedimageslose detaininbrightareasandmayevenappearsmeared(thisis calledblooming)whenapixelisoverfilledwithlight. DimImages Increaselaserintensityand/orincrease exposuretime. SaturatedImages Decreaselaserintensityand/or decreaseexposuretime.ifimageisstilltoo bright, decreaseemgain. AdjustLaserIntensityandExposureTimeforeachchannel.Press Recordtokeepsettings. 7
8 6/17/09SCAndoriQversion1.9.1DRAFT fastmodeonly Z SCAN AcquiringanImageVolume inrevolutionfastlz (Fast) InAcquisition/AuxiliaryDevicesWindow Undertabs Experiment Sequence Z controlenabled checked Z controlmovefirst unchecked ComputedParameter Planes FastScan checked Stepsize typeinastepsize.agoodplacetostartis0.5µm. CurrentZ PressGoCentre Thiscentersthepiezoinitstravel 100µm. Q:HowdoIchooseastepsize? A:Itdependsonthesample. Somestartingpointsare: 1/2theopticalresolutioninz(~0.5µm) or z=thex,ypixelsize But,dependingonwhatyouaretryingto visualize,youmaybeabletotakez steps2xthisormore. Commonlyusedstepsizesonthissystemrange from µm PressLive.Findandfocusoncellofinterestusingfocusknobonmicroscopebody(Donotuseknobonscopeafterthis step) UsingslidernexttoCurrentZorarrowsunderStepthroughplanes,movetothestartpositionforthe3Dvolume,then pressstart.movetotheendposition,thenpressend. InAndoriQWindow PressRun. ThefirsttimethatRunispressed,adialogueboxopensandsaysthattheshutter isnotopen,doyouwanttoopentheshutter.clickyes. Thisdialoguewillnotappearagaininthecurrentsession. ClickOKtostartprotocol. TheimagesareautomaticallysavedintheImageList.Youwillexportallofyourimagesattheendofyoursessionina formatreadablebyothersoftware.ifyouhaveseveraldifferentsamples,youshouldrenameyourimagesasyou acquirethem.seeimagelist,page10. 8
9 6/17/09SCAndoriQversion1.9.1DRAFT fastmodeonly TIMELAPSE inrevolutionfastlz (Fast) SetupparametersformultichannelandZ scan. IntheAndorIQwindow,rightclickonRepeat,then chooseedit.typeinnumberofrepeats. IntheAndorIQwindow,rightclickonInterval.Default Intervalisnopausebetweenscans.Typeindesired interval. 9
10 6/17/09SCAndoriQversion1.9.1DRAFT fastmodeonly IMAGELIST ChoosingImagelistonthemenubar ofanyofthewindowswillopenthe ImageManger,whichdisplaysallof yourimagesalongwiththeir acquisitioninformation. RenameImage AndorIQsoftwareautomatically namesandsavesimages.stacksand timeseriesarenamedbeginningwith theprefixthatyouassignedatthe beginningofyourimagingsession (usuallythedateandanexperiment identifier)andhaveanumber appended.snapsarenamedwiththe activechannelnamewithanumber appended. Ifyouhavemorethanonesample group,eitherchangetheprefixfor eachsamplegroup,orrenameyour filesasyouacquirethemtoavoid confusionlater. WhentheImageListisopened,the mostrecentimageacquiredappears atthetopofthelist.ifadditional imagesareacquiredwiththelist open,theydonotautomaticallyappearonthelist.clickrefreshandthemostrecentimageacquiredwillappear atthetopofthelist. TheRenameImagebuttonallowsyoutospecifyadifferentnameforthecurrentlychosenimage.Selectanimageby clickingonitintheimagemanagerwindow.clickrenameimage.typethenewnameintothetextboxandclick OK.ThenewnamewillnowappearintheImageManagerwindow. DeleteImage TheDeleteImagebuttonwilldeletecurrentlyselectedimagesfromtheImageManager. SaveImage Ifyouareanewuser,logontotheCD/DVDburningmaching,In Neph Toast.Loggingonwillcreateyouruserfolderon In Neph Toast.Nowlogoff.ItisnotnecessarytostayloggedonwhilesavingyourimagestoIn Neph Toast. 10
11 6/17/09SCAndoriQversion1.9.1DRAFT fastmodeonly ImagesaresavedwithintheAndorsoftwareduringacquisition,buttheyareinaformthatcanonlybeaccessedfrom withintheandorsoftware.attheendofyourimagingsessionyoumustsaveyourimagestoanothermachinein TIFFformat.TIFFformatisreadablebyalargenumberofotherimagevisualizationandprocessingprograms. Filescanbesavedoneatatimeorasagroup.AndorIQsoftwarehasdifferentdialogueboxesforindividualand groupsaves. SaveImage MultipleImages Highlightthefilesthatyouwishtosavebyclickingonthenamesin theimagemanagewindow.tohighlightmorethanonefile,hold downctrlandcontinueselectingfiles. ClickSaveImage.TheMutiSaveOptionsandSaveMultiplefiles dialoqueboxeswillopen. ChooseTIFF(singleimagefiles)*.tif Uncheckthedeletetheimagesbox. (Deletethissessionsimagesthenexttimethatyouusethespinningdisc microscope,afteryouhavemadeacopyoncd/dvdoronyourownlab machine.) ClickOK TheBrowseforFolderdialogueboxwillopen.BrowsetoyourfolderonIn Neph ToastandclickOK. GotoIn Neph ToastandburntwocopiesofyourdataontoCDorDVD. Spotcheckthatyoucanopenyourimageswhenyougetbacktoyourown lab.deleteyourdatafromtheimagemanagerandfromin Neph Toast duringyournextimagingsession. SaveImage singleimage Highlightthesinglefilethatyouwishtosavebyclickingonthename intheimagemanagerwindow. ClickSaveImage. Browsetowhereyouwanttosavethefiles In Neph Toast ChooseTIFF(singleimagefiles)*.tif Givethefileaname(filename.tif) ClickSave 11
12 6/17/09SCAndoriQversion1.9.1DRAFT fastmodeonly IMAGEWINDOW(windowislabeled AndorCameraWithFastZandWavelength) IntensityScaling AtthebottomoftheImageWindow,thereisacheckboxlabeledAutoscale.Thisshouldbechecked.TheAndorcamera acquires14bitimages(2 14 levelsofgrey).mostmonitorscanonlydisplay8bitimages(2 8 levelsofgrey).whenyouclick autoscale,thesoftwarewillscalewhatevernumberofgreylevelsofsignalthatyouhave,upordown,tofitwithinthe8 bitdisplay.thisdoesnotchangeyourdata,itjustmakesvisualizationeasier.movethemouseovertheimagetoseethe intensityvalueforthatareaoftheimage.thex,ycoordinatesandtheintensityvaluewillbedisplayed atthebottomoftheimagewindow.todisplayahistogramoftheimageintensityvalues,press: toopentheimagecontrastcontroldialoguebox. 12
13 6/17/09SCAndoriQversion1.9.1DRAFT fastmodeonly ColorandColoroverlay Whenyouhaveacquiredamultichannelimage,thesecontrolswillbeavailableinthelowerleft oftheimagewindow.whenmultiisunchecked,onechannelatatimeisdisplayedandyoucan togglebetweenchannelsbypressingthebuttons.youcanchangethedisplaycolorforthe channelsbyrightclickingonthebutton,thenchoosingacolorfromthelutmenu.tooverlay twochannels,checkthemultiboxandpressbothbuttons.keepinmindthatthisisdisplay coloronlyanddoesnotaffectthedata.dataisacquiredasgreyscale,intensityvalues. ScrollingthroughZ,TimeandFields InthelowerleftoftheImagewindow,therearetabsthatcorrespondtothedimensions acquiredinanimage. ToscrollthroughZ,clickonthetablabeledZandmovetheslider. Toscrollthroughtime,clickonthetablabeledTimeandmovetheslider. Tomovebetweenfieldsinamultifielddataset,clickonthetablabeledXYandmovethe slider. 13
14 6/17/09SCAndoriQversion1.9.1DRAFT fastmodeonly AcquireaSubRegion Snapimageofspecimenwithchannelthatletsyoudefinedesiredarea. IntheImageWindowToolbar(AndorCameraWithFastZandWavelength) UsingImage ToolDefineRectangle drawarectangularroi(regionofinterest)aroundthesubregionthatyou wanttocapture PressDeviceSetupButtonandopen: Acquisition/AuxiliaryDevicesWindow DV897_BV2302Settings Exposuretabs. ClickontheWindowbutton(ItwillchangetosayFullChip) Snapimagetoconfirmthatcorrectregionissetup Acquire3D,multichannel,timelapseasusual.SubRegionwillbeappliedtofutureimages.Thisisusefulwhenyouwant tospeedupacquisition. Toreturntothenormal512x512,clickontheFullChipbutton.ItwillchangebacktoWindowandthesubregionisno longerselected.toclearthedefinedregion,rightclickontheimageandchoosedeleteall. 14
Olympus IX-70 Imaging Protocol
Olympus IX-70 Imaging Protocol 1) System Startup Please note our sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused
More informationLSM 5 MP, LSM 510 and LSM 510 META Laser Scanning Microscopes
LSM 5 MP, LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4.2 January 2007 Contents Page Starting the System...3 Setting the microscope...6 Configuring the beam path
More informationOlympus IX-70 Imaging Protocol
Olympus IX-70 Imaging Protocol 1) System Startup F Please note our sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for
More informationIndiana Center for Biological Microscopy. BioRad MRC 1024 MP Confocal & Multi-Photon Microscope
Indiana Center for Biological Microscopy BioRad MRC 1024 MP Confocal & Multi-Photon Microscope Microscope and the Attached Accessories A: B: C: D: E: F: G: H: Mercury Lamp Transmission Light Kr/Ar Laser
More informationSpinning Disk Protocol
Spinning Disk Protocol 1) System Startup F Please note our sign-up policy You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time
More informationNikon A1-B Confocal Operating Manual. Start-up. Microscope
Nikon A1-B Confocal Operating Manual Start-up 1. Turn on Excite 120 LED power supply. a. No need for cool down as for mercury bulb b. Open shutter by pushing down contol knob. c. Adjust intensity by turning
More informationDSU Start-Up instructions
DSU Start-Up instructions Always: - start with the 10x objective - properly center the stage around the current objective before changing to another objective - when done, leave the 10x objective in standby
More informationLeica TCS SP8 X Confocal Microscope and Leica Application Suite X software DRAFT VERSION Room B123b
Leica TCS SP8 X Confocal Microscope and Leica Application Suite X software DRAFT VERSION Room B123b User Guide Biomedicum Imaging Unit (BIU) University of Helsinki www.biu.helsinki.fi 11.11.2016 1 GENERAL...1
More informationQuick Start Guide - cellsens
Quick Start Guide - cellsens IQS002- Rev 02 Issue Date: 25/05/2017 1 Integrated CoolLED Products: pe-2 pe-4000 N.B. The pe-300 Series are not officially integrated by Olympus, but it is possible to gain
More informationQuick Start Guide for LAS X
Quick Start Guide for LAS X IQS004- Rev 02 Issue Date: 25/05/2017 1 Integrated CoolLED Products: pe-300 white, pe-300 ultra & pe-340 fura pe-4000 2 Setting up your CoolLED Illumination System COM Port
More informationWide Guy: Inverted Widefield Microscope
Wide Guy: Inverted Widefield Microscope Kyle Marchuk Adam Fries Jordan Briscoe August 2017 Contents 1 Introduction 2 2 Initial Setup 3 2.1 Hardware Startup........................................... 3
More informationLSM510 Confocal Microscope Standard Operation Protocol Basic Operation
LSM510 Confocal Microscope Standard Operation Protocol Basic Operation Please make sure that the COMPRESSED AIR has been TURNED ON prior to the use of the equipment. Kindly inform the administrator if
More informationZeiss AxioImager.Z2 Fluorescence Protocol
Zeiss AxioImager.Z2 Fluorescence Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge
More informationLEXT 3D Measuring LASER Microscope
LEXT 3D Measuring LASER Microscope Warning: This instrument may only be operated by those who have been trained by AAF staff and have read and signed the AAF laboratory policies. A) STARTUP 1. Computer
More informationCell Imaging Unit UIC Bárbara Fekete ZEISS AXIOIMAGER MICROSCOPE MANUAL
Cell Imaging Unit UIC Bárbara Fekete ZEISS AXIOIMAGER MICROSCOPE MANUAL 1 TABLE OF CONTENTS ANATOMY OF THE ZEISS AXIOIMAGE MICROSCOPE 3 INITIALISATION PROCEDURE 4 LEFT SIDE OF MICROSCOPE (DETAILED VIEW)
More informationStart-up system. Nikon Wide Field Microscope Ee1078. Complete manual
Start-up system Switch on the power socket located on the floor Switch devices on in the order of: o Mercury lamp control-unit switch on and ignite o Halogen lamp o Incubator control-box o Microscope stand
More informationManual ANGSTROM grid confocal microscope
Manual ANGSTROM grid confocal microscope Switching on the system 1) Start by switching on the different components of the microscope system. When using live cell imaging also start the temperature and
More informationBasic Users Guide for the LSM 510 Confocal Microscope
Basic Users Guide for the LSM 510 Confocal Microscope Ian Jones and Adam Westmacott Department of Biology & Biochemistry, University of Bath, Bath, UK Updated 01 October 2003 This guide describes the basic
More informationPCIC Plant Cell Imaging Center (BTI, room 104B) CONFOCAL LEICA SP5 user manual
PCIC Plant Cell Imaging Center (BTI, room 104B) CONFOCAL LEICA SP5 user manual -Turn on the mercury lamp if you need to see the fluorescence through the eye pieces. Check that its shutter is on the open
More informationConfocal Application Notes April 2010
April 2010 Motorized Stage Applications Mark & Find Prepared by Louise Bertrand Applications and Technical Support Group, Leica Microsystems, Inc In this issue of Leica Confocal Application Notes, we describe
More informationRenishaw invia Raman Microscope (April 2006)
Renishaw invia Raman Microscope (April 2006) I. Starting the System 1. The main system unit is ON all the time. 2. Switch on the Leica microscope and light source for reflective bright field (BF) imaging.
More informationZeiss Elyra Microscope Standard Operation Protocol Basic Operation
Machine Startup Zeiss Elyra Microscope Standard Operation Protocol Basic Operation 1. Turn on the Incubator power to warm up the incubator for a couple of hours prior to sample acquisition as image could
More informationOlympus DeltaVision Microscope Start-Up and Shut-Down Instructions.
DeltaVision Instructions: 1. Start-up (Olympus DV) 2. Basic operation (Olympus DV) 2.1 set-up 2.2 designing and running an experiment 3. Transferring files via FTP 4. Deconvolving files 5. File conversion
More information1. Getting Started: Brief Step-By-Step Guide (PDF File)
Page 1 of 5 1. Getting Started: Brief Step-By-Step Guide (PDF File) Leica SP2 confocal microscope is controlled via software LCS (Leica Confocal Software). The latest version is V2.5. Bellowing is a brief
More informationConfoCor 2 Fluorescence Correlation Microscope
ConfoCor 2 Fluorescence Correlation Microscope Operating Manual Release 3.2 INTRODUCTION Carl Zeiss ConfoCor 2 Knowledge of this manual is required for the operation of the instrument. Would you therefore
More informationMulti-Photon Training
Multi-Photon Training Overview This training will take approximately 4 hours. The first 2 hours will be spent learning one-on-one with your trainer, while the following 2 hours will be your opportunity
More informationCCD Image Acquisition Tutorial
CCD Image Acquisition Tutorial Gatan, Inc. 5933 Coronado Lane, Pleasanton, CA 94588 Tel: (925) 463-0200 Fax: (925) 463-0204 May 2001 1. Introduction This document will guide users of Gatan CCD camera through
More informationOperating Procedure for Horiba Raman Microscope
Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch
More informationPrism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis
Start Up: Upon entering the laser room turn on the wall mounted Laser Power Button by pulling it away from the wall. Turn on Shutter controllers (toggle switch on back of unit). There should be a U in
More informationSamples Carolina sample slides (pollen, algae, ). Clean off oil with lens paper then OpticPad around lens (metal not glass) when done.
Bi/BE 227 Winter 2018 Assignment #1 Widefield and confocal laser scanning microscopy Schedule: Jan 3: Lecture Jan 3-12: Students get trained on how to use scopes, start on assignment Jan 3-17: Carrying
More informationFluorimeter User s Booklet
Fluorimeter User s Booklet 1) Verify that the instrument power and the computer are both off. Then turn on the lamp. The red indicator light will glow when the lamp is on. Turn on lamp switch here. This
More information` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE
` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE Version: 1.0 April-2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research Center (NRC) TABLE OF CONTENTS 1. Introduction...2 1.1
More informationABM's High Resolution Mask Aligner Features:
ABM's High Resolution Mask Aligner is a very versatile instrument with interchangeable light sources which allow Near-UV (405-365 nm) as well as Mid- and Deep-UV (254 nm, 220 nm) exposures in proximity
More informationConfocal 6 Guide. Switching On Components of the system should be switched on in the following order: 1. Lasers Argon laser (488nm and 514nm)
Confocal 6 Guide The system is kept at 37⁰C. If you require the system at a lower temperature book at least one hour before the beginning of your session to allow the system to cool down. It is your responsibility
More informationCCD Acquisition Function
File Edit View Props Region Open Flatfield Corr. Optics Image Stats Setup Open Background Corr. Reset Camera Image Profile Temperature Close Autoexpose Autofocus Correction Output Dual Shutter From the
More informationBiology Shared Instrumentation Facility Olympus IX70 Inverted Microscope and Simple PCI Imaging System
Biology Shared Instrumentation Facility Olympus IX70 Inverted Microscope and Simple PCI Imaging System Use Protocols: A. Reservation and Starting Up B. Sample Observation (No Image Capture) Transmitted
More informationThe Paper Project Guide to Confocal Imaging at Home & in the Classroom
The Paper Project Guide to Confocal Imaging at Home & in the Classroom http://lifesciences.asu.edu/paperproject The Paper Project Guide to Confocal Imaging at Home & in the Classroom CJ Kazilek & Dennis
More informationUse of the Binocular Microscope
Use of the Binocular Microscope Before you begin this learning module be sure that you have the following materials in front of you: A microscope A packet of lens paper and Kimwipes A glass slide of a
More informationXRADIA microxct Manual
XRADIA microxct Manual Multiscale CT Lab Table of Contents 1. Introduction and Basics 1.1 Instrument Parts 1.2 Powering up the system 1.3 Preparing your sample 2. TXM Controller 2.1 Starting up 2.2 Finding
More informationUser Operation Manual
User Operation Manual Manual Version: 1.3 For Single Cell Bioanalyzer (SCB-402) Doc# 102-SCB-402-1.3 Table of Contents 1. General Information... 1 1.1 How to Use This Manual... 1 1.2 Operation Safety...
More informationSUSS MICROTEC MJB4 MASK ALIGNER DOCUMENTATION
SUSS MICROTEC MJB4 MASK ALIGNER DOCUMENTATION ANITHA & KARTHICK 1 Tool specification: Substrate size :1 up to 4 Mask size : Standard 2 x2 up to 5 x5 Alignment :X,Y-±5mm,theta-5⁰ Alignment Accuracy :Top
More informationSIMPLE USER GUIDE RAMAN INSTRUMENT
SIMPLE USER GUIDE RAMAN INSTRUMENT 27 March 2009 version Reviced 14 April 2011 Arne Karlsson TYPICAL START-UP PROCEDURE 1. Software 2. Hardware To log on: click this icon. When you are asked for a password
More informationGOODJOB. ICAN-FluorUV Easy Guide. Contents. Before Installation (Make Sure You Have) 1. Preparing the Camera Installing Camera and Filter...
GOODJOB ICAN-FluorUV Easy Guide Contents Before Installation (Make Sure You Have) 1 Preparing the Camera... 2 Installing Camera and Filter... 2 Reading This First... 2 Getting Started 2 Before installation
More informationANDOR/OLYMPUS SPINNING DISK USER GUIDE
ANDOR/OLYMPUS SPINNING DISK USER GUIDE CONTENTS Overview... 2 Turn off sequence:... 2 Turn on sequence:... 2 Getting started... 2 Multi Dimensional Acquisition (MDA)... 3 FRAP... 3 Focus motors and long-term
More informationZeiss Z.1 Lightsheet. Quick-Start Guide
Zeiss Z.1 Lightsheet Quick-Start Guide Contents Start-up Preparing for Imaging Part 1 Chamber and Sample preparation Part 2 Image Acquisition Shut-down Start-up: Turn on the Microscope / Computers Turn
More information3-D. Here red spheres show the location of gold nanoparticles inside/around a cell nucleus.
3-D The CytoViva 3-D System allows the user can locate objects of interest in a 3-D space. It does this by acquiring multiple Z planes and performing our custom software routines to locate and observe
More informationUser s Guide to the LMD Laser Micro-dissection. System
User s Guide to the LMD-6000 Laser Micro-dissection System Page 1 Glen MacDonald October 31, 2018 Start-up Procedure for Leica LMD-6000. 1. Turn on mercury lamp by pressing the rocker switch; a. beige
More informationBio-Pyramid / STYLUS 1s MBP01-S1SB User s Manual
Bio-Pyramid / STYLUS 1s MBP01-S1SB User s Manual Standard Accessories 1) Olympus STYLUS 1s 2) Lithium ion battery (BLS-50) 3) Lithium ion charger (BCS-5) 4) Lithium ion charger cable 5) USB Cable (CB-USB6)
More informationCoSMoS Starter Guide 1.4 Hoskins Lab Last Modified 03/03/2017 Chris DeCiantis
Contents Start Up:... 2 Turn on Shutter controllers... 2 Turn on Lasers... 2 Turn on other equipment:... 2 Load Slide onto Sample Holder:... 2 Turn on Computer/ Start Labview/Glimpse:... 3 Shutters Activate.vi
More informationOlympus BX51 on Aura. Introduction to the NRI-MCDB Microscopy Facility BX51 Upright Microscope
Olympus BX51 on Aura Introduction to the NRI-MCDB Microscopy Facility BX51 Upright Microscope Contents Start-up Preparing for Imaging Part I General Part II Transmitted Brightfield Part III Transmitted
More informationHeit/Kerfoot/Dikeakos. Leica DMI6000 B. User Manual
Heit/Kerfoot/Dikeakos Leica DMI6000 B User Manual WARNING This microscope is a multi lab resource that many individuals are dependent on. As such the following rules must be followed in order to minimize
More informationCoSMoS Starter Guide 1.6 Hoskins Lab Last Modified 03/30/2017 Chris DeCiantis
Contents Start Up:... 2 Turn on Shutter controllers... 2 Turn on Lasers... 2 Turn on other equipment:... 2 Load Slide onto Sample Holder:... 2 Turn on Computer/ Start Labview/Glimpse:... 3 Shutters Activate.vi
More informationQUANTAX EDS SYSTEM SOP
QUANTAX EDS SYSTEM SOP December 2017 Energy-Dispersive X-Ray Spectroscopy (EDS, EDX, EDXS or XEDS), is an analytical technique used for the elemental analysis or chemical characterization of a sample.
More informationNANOSPEC 4150 STANDARD OPERATING PROCEDURES
NANOSPEC 4150 STANDARD OPERATING PROCEDURES Version: 1.0 JAN 2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center TABLE OF CONTENTS 1.0 INTRODUCTION.. 3 2.0 HARDWARE....... 3 3.0 OPERATING
More informationPresentation and analysis of multidimensional data sets
Presentation and analysis of multidimensional data sets Overview 1. 3D data visualisation Multidimensional images Data pre-processing Visualisation methods Multidimensional images wavelength time 3D image
More informationUniversity of Minnesota Nano Fabrication Center Standard Operating Procedure
Equipment Name: University of Minnesota Nano Fabrication Center Coral Name: hs-scope Revision Number: 1.5 Model: HS200A Revisionist: M. Fisher Location: Bay 1 Date: 9/12/2013 1 Description The Hyphenated
More informationFLUOSTAR Rhodamine B- encapsulating microspheres are seeding particles optimized for Particle Image Velocimetry. Technical handbook ver.
www.ebm.vc FLUOSTAR Rhodamine B- encapsulating microspheres are seeding particles optimized for Particle Image Velocimetry Technical handbook ver.1, Feb, 2010 CONTENTS 1) Introduction of fluorescent PIV
More informationIndependent Resolution Test of
Independent Resolution Test of as conducted and published by Dr. Adam Puche, PhD University of Maryland June 2005 as presented by (formerly Thales Optem Inc.) www.qioptiqimaging.com Independent Resolution
More informationJasco FP-6500 Spectrofluorimeter Updated November 14, 2017
Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017 Instrument instructions can be found at: http://academic.bowdoin.edu/chemistry/resources/instructions.shtml If you have any problems with the
More informationAUTOFOCUS SENSORS & MICROSCOPY AUTOMATION IR LASER SCANNING CONFOCAL MICROSCOPE IRLC DEEP SEE. Now See Deeper than ever before
AUTOFOCUS SENSORS & MICROSCOPY AUTOMATION IR LASER SCANNING CONFOCAL MICROSCOPE IRLC DEEP SEE Now See Deeper than ever before Review and inspection of non visible subsurface defects Non visible and subsurface
More informationTransmitted light Illuminator VIS-LED, color temperature. Epifluorescence lamp module X-Cite 120PC Q (Excelitas
ELYRA PS.1 Multi-functional fluorescence inverted widefield microscope enabling live-cell imaging, TIRF or HILO illuminion and two super-resolution techniques: structured illuminion microscopy (SIM) and
More informationDV2. Alignment Procedure. Install DV2 on Microscope NOTE: PLEASE READ THE ENTIRE PROCEDURE BEFORE YOU BEGIN ALIGNMENT OF THE DV2. Alignment Procedure
H I G H - P E R F O R M A N C E E M C C D & C C D C A M E R A S F O R L I F E S C I E N C E S DV2 This document provides a straightforward, step-by-step outline of the alignment procedure for the Photometrics
More informationSeized Drugs Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division
Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division SHIMADZU ULTRAVIOLET/VISIBLE SPECTROPHOTOMETER (UV/VIS) Instrument Shimadzu UV-2401 UV/VIS Spectrophotometer (Serial Number
More informationEnFocus Your Upgrade Path to High Performance Intrasurgical OCT
Your Upgrade Path to High Performance Intrasurgical OCT is FDA 510(k) Cleared > Ultra HD OCT EXTEND YOUR MICROSCOPE S POTENTIAL WITH INTRASURGICAL OCT Brilliant Images, Sub-Surface Knowledge is an intrasurgical
More informationLynx Instruction Manual
Lynx Instruction Manual 1.0 INTRODUCTION 2.0 HARDWARE - 2.1 Slider 3.0 USER INTERFACE - 3.1 Main Menu - 3.2 Fire Test Shot - 3.3 Backlight - 3.4 Bluetooth - 3.5 Reset - 3.6 Motor Sleep - 3.7 Torque 4.0
More informationThe Microscope. Lab Exercise. Contents. Light Microscope. Objectives. Scanning Electron Microscope. Introduction. Transmission Electron Microscope
Lab Exercise The Microscope Contents Objectives 1 Introduction 1 Activity.1 Care of Microscopes 2 Activity.2 Parts of the Microscope 3 Activity.3 Viewing Images 4 Activity.4 Magnification of Images 4 Resutls
More informationMulti Time Series Rev
Multi Time Series Rev. 4.0.12 The Multi Time Series program is designed to provide flexible programming of automated time dependent experiments. The basic programming unit is a single Time Series Block
More informationFrom Eye to Insight. Leica DMC2900. Digital microscope camera for easy, efficient documentation and presentation in industry and research
From Eye to Insight Leica DMC2900 Digital microscope camera for easy, efficient documentation and presentation in industry and research 3 High Speed Imaging Leica DMC2900 with USB 3.0 interface for highest
More informationApplication for Inverted Research Microscope ECLIPSE Ti2 Series Ti2Control Ver Instruction Manual (for Android)
M705E 17.3.Nx.4 Application for Inverted Research Microscope ECLIPSE Ti2 Series Ti2Control Ver.1.0.2 Instruction Manual (for Android) Introduction Introduction Thank you for purchasing a Nikon product.
More informationOlympus cellsens TIRF Guide
Startup Power up the system as described in the "Startup & Environmental Control" guide, including power strip #2 ("TIRF"). ON For each laser line needed, turn the key on the control unit right and press
More informationVibroFlex Compact Compactness meets versatility Preliminary datasheet. VibroFlex Compact
VibroFlex Compact The Polytec VibroFlex laser Doppler vibrometer is a modular high-performance solution for non-contact vibration mea sure ment. It offers unrivalled measurement performance and versatility
More informationLiving Image Software
Living Image Software User s Manual Version 3.1 2002-2008 Xenogen Corporation. All rights reserved. PN 124908 Xenogen Corporation 68 Elm Street Hopkinton, MA 01748 USA 1.877.522.2447 (US) 1.508.435.9500
More informationMICHELSON S INTERFEROMETER
MICHELSON S INTERFEROMETER Objectives: 1. Alignment of Michelson s Interferometer using He-Ne laser to observe concentric circular fringes 2. Measurement of the wavelength of He-Ne Laser and Na lamp using
More informationAtomic Force Microscope
Atomic Force Microscope Preparation Before our first appointment to use the microscope, please watch the video tutorials at https://www.afmworkshop.com/atomic-force-microscope-animated-tutorials/. The
More informationApplication for Inverted Research Microscope ECLIPSE Ti2 Series Ti2 Control Ver Instruction Manual (for Android)
M705E 17.6.Nx.5 Application for Inverted Research Microscope ECLIPSE Ti2 Series Ti2 Control Ver.1.1.0 Instruction Manual (for Android) Introduction Introduction Thank you for purchasing a Nikon product.
More informationSuss Aligner SOP. Suss Aligner SOP Page 1 of 6 Revision Scope 1.1 This document provides the procedure for operating the Suss Aligner.
Suss Aligner SOP Page 1 of 6 Suss Aligner SOP 1. Scope 1.1 This document provides the procedure for operating the Suss Aligner. 2. Table of Contents 1. Scope... 1 2. Table of Contents... 1 Suss Aligner
More informationAll-In-One. digital inverted microscope. High-quality Imaging Has Never Been Easier. evos-ca.com
All-In-One digital inverted microscope High-quality Imaging Has Never Been Easier evos-ca.com Fluorescence Imaging Is Now Easier Than Ever ALL-IN-ONE, digital inverted fluorescence microscope The easiest
More informationLS-785 High Throughput Lens Spectrograph User Manual
15 Discovery Way, Acton, MA 01720 Phone: (978)263-3584, Fax: (978)263-5086 Web Site: www.piacton.com LS-785 High Throughput Lens Spectrograph User Manual Ver. 1.2 Table of Contents Chapter 1 -- General...5
More informationStandard Operating Procedure for the Horiba FluroMax-4
Standard Operating Procedure for the Horiba FluroMax-4 Adapted from Horiba Operations Manual Created by Michael Delcau, Modified by Brian Lamp The Fluoromax is capable of making a variety of measurements.
More informationDigital Microscope. Model: AY Read the precautions and important tips in the instructions before use. Chapter One Overview...
Digital Microscope Model: AY12214 Read the precautions and important tips in the instructions before use. Chapter One Overview... 1 1-1 Appearance... 1 Included Accessories... 2 Specifications... 2 Magnification
More informationWorking Copy For Review Purposes Only!
Working Copy For Review Purposes Only! Version #1.5 08/29/06 Aaron Holland Up-dated D.Turnbull 10-18-2006 (MarkerSOP3.doc) Up-dated D. Turnbull 12-11-06 (MarkerSOP4.doc) Up-dated D.Turnbull 05-12-2008
More informationHCImage. Image Acquisition and Analysis Software. Light... Camera... Acquisition...
HCImage Image Acquisition and Analysis Software Light... Camera... Acquisition... HCImage Analysis HCImage Analysis provides comprehensive control of Hamamatsu cameras, microscopes, stages and other peripheral
More information4 OPTIMIS FUSION LASER DESCRIPTION
Introduction 4 OPTIMIS FUSION LASER DESCRIPTION 4.1 GENERAL DESCRIPTION Bionocular head Slit lamp Magnification Controle module LCD screen Aiming beam Headrest Fixation LED Chinrest Slit width control
More informationFluorescence Microscope with Spinning Disk Confocal and Total Internal Reflection Fluorescence Modules
Open Tender Notification for the procurement of Fluorescence Microscope with Spinning Disk Confocal and Total Internal Reflection Fluorescence Modules at the Indian Institute of Science, Bangalore (Last
More information»»»LASER GENERATION IV The Innovation for your Success
»»»LASER GENERATION IV The Innovation for your Success VISION LWI IV FIBER FLE X X FLEXABILITY FOR NEW POSSIBILITYS Based on the well-established and reliable LWI IV Series the Fibre Fle x x is the choice
More informationOptical Ptychography Imaging
Optical Ptychography Imaging Summer Project Annafee Azad Supervisors: Dr Fucai Zhang Prof Ian Robinson Summer 2014 23 October 2014 Optical Ptychography Imaging P a g e 2 Abstract This report details a
More informationHigh Speed Digital Microscope
High Speed Digital Microscope Product datasheet Page Description 2 Benefits 3 System Specifications 3 IP License 4 High Speed Digital Microscope datasheet Page 1 of 5 Part No. Part Description # 3200531
More informationCOMPACT MANUAL USE OF SPARK M10 PLATE READER Room HG General Instrumentation
0 COMPACT MANUAL USE OF SPARK M10 PLATE READER Room HG01.228 General Instrumentation SPECIFICATIONS ASSISTANCE BOOKINGS SWITCH ON CREATE/EDIT METHODS (IN MAGELLAN) MEASUREMENT STORAGE DATA USERS AND METHODS
More informationQuick Start. MS-9 Laser Scanner
Quick Start MS-9 Laser Scanner This guide is designed to get your scanner up and running quickly. When connected to a host computer with Windows operating system (2000 or above), you can use ESP (Easy
More informationRaman Spectrometer Installation Manual
RI Raman Spectrometer Instruction Manual Application software (included in standard configuration) Connection RI RAMAN is recognized by Windows operational system as standard HID device. That is why there
More informationConfocal Microscope - Radiance. Basic Instructions. (Stefanie Reichelt) The Radiance. Confocal Scanning Microscope. designed and developed by
Confocal Microscope - Radiance Basic Instructions (Stefanie Reichelt) The Radiance Confocal Scanning Microscope designed and developed by Brad Amos (MRC-LMB) Commercialised by Bio-Rad Inc. (1998 2004)
More informationFLEX 2 NEW. Key points. 3D Confocal Raman, 2 lasers, fiber based, AFM combined
FLEX 2 3D Confocal Raman, 2 lasers, fiber based, AFM combined NEW Key points Compact size 2 lasers, easily switchable 2 confocal operation modes, easily switchable : High Spatial Resolution 35 nm High
More informationInterference. Electric fields from two different sources at a single location add together. The same is true for magnetic fields at a single location.
Interference Electric fields from two different sources at a single location add together. The same is true for magnetic fields at a single location. Thus, interacting electromagnetic waves also add together.
More informationOptical Sectioning. Bo Huang. Pharmaceutical Chemistry
Optical Sectioning Bo Huang Pharmaceutical Chemistry Approaches to 3D imaging Physical cutting Technical difficulty Highest resolution Highest sensitivity Optical sectioning Simple sample prep. No physical
More informationDESIGNED & MANUFACTURED by FLOATCAM
DESIGNED & MANUFACTURED by FLOATCAM DC-Slider Motion Control PN377719 Package Includes: Motor Microprocessor Motor to Microprocess or Remote Controller Connection Cable AC Power Cable w/north American
More informationBOTTOM SIDE ALIGNMENTSOP
BOTTOM SIDE ALIGNMENTSOP October 2013 The wafer is aligned to the mask using the bottom side alignment microscope (BSA). 1. Load mask.please refer to Step 1-3 in the previous SOP for loading a mask. 2.
More informationZeiss LSM 510 Meta inverted confocal manual
Power Up Zeiss LSM 510 Meta inverted confocal manual *adapted from the Duke University Light Microscopy Core Facility manual 1. Turn on the mercury lamp (if required) and sign in the log book, recording
More informationPhysical Optics. 1 st year physics laboratories. University of Ottawa.
Physical Optics 1 st year physics laboratories University of Ottawa https://uottawa.brightspace.com/d2l/home INTRODUCTION Physical optics deals with light as a wave which can bend around obstacles (diffraction)
More informationLightsheet Z.1. Light Sheet Fluorescence Microscopy by Carl Zeiss. Fabrice Schmitt, Sales Manager Carl ZEISS France
Lightsheet Z.1 Light Sheet Fluorescence Microscopy by Carl Zeiss Fabrice Schmitt, Sales Manager Carl ZEISS France 12.12.2012 Light Sheet Fluorescence Microscopy (LSFM) Principle The Principle of Light
More informationQuickGuide LightSheet
This quickguide will guide you through most of the important steps in using the LightSheet Z.1 from ZEISS. For quick guidance the number of the step will appear below the ZEISS logo on the upper right
More information