` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE

Size: px
Start display at page:

Download "` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE"

Transcription

1 ` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE Version: 1.0 April-2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research Center (NRC)

2 TABLE OF CONTENTS 1. Introduction Scope of Work Description Safety Hardware Requirements Training System Restrictions Operating Procedures System Pre-Checks Hg Lamp Start up..x 4.3 Preparing system for Alignment/Exposure Anodic Bond Alignment Recipes Recipe Guidelines Creating a recipe Editing a recipe Nanofab Recipes

3 1.0 INTRODUCTION 1.1 Scope These procedures apply to the EVG 620 Mask Alignment Wafer system. All maintenance should follow the procedures set forth in the manufacturer s maintenance and operations manuals. This document is for reference only. Users must be trained by Nanofab staff before operating this equipment. 1.2 Description The EVG 620 mask alignment system is a dual-use tool designed for optical double-side lithography and precision alignment for 2 to 4 diameter wafer size and small sample pieces. Top side alignment accuracy of ± 1.0µm 3σ, bottom side alignment ± 1.25µm 3σ, large gap alignment ± 1.5µm 3σ, IR alignment (substrate depending), bond alignment of ± 5.0µm 3σ. An ultra-soft wedge compensation together with a computer controlled contact force between the mask and wafer ensures that both wafer damage is minimized while mask lifetime are dramatically increased. High resolution of 2um is achieved in four exposure modes: proximity, soft, hard, and vacuum contact modes utilizing 350 Watt Mercury arc lamp calibrated to 16 Mw/cm2 light intensity using 365 nm uv probe. 1.3 Safety This machine is connected to HIGH VOLTAGE. Be very careful and remain aware of electrical hazards. If you encounter any electrical malfunctions, contact NanoFAB staff immediately Always keep finger clear of the moving parts such exposure head, mask loader and wafer tray. Failure to do so may result in personal injury This equipment uses a Hg-Arc lamp UV light source that is subjected to 20 to 30 atmospheres of internal pressure. There is always the possibility of a lamp exploding due to internal strains, high lamp temperature or physical abuse. Any vapor released by an exploded lamp would be highly dangerous. In the event of a lamp explosion, clear the area of personnel and contact the NanoFAB staff immediately. Do not return to the area until staff gives the OK to return

4 1.3.4 Be aware of UV light intensity when performing the Lamp start up procedure or when exposing wafers: The light emanating from the light source is a hazard to vision. DO NOT LOOK INTO THE OUTPUT BEAM of the UV Lamp or reflection of the beam. Permanent damage to the retina of the eye and subsequent blindness can result. When checking whether the Lamp is ON, open the Lamp viewport window shutter just a very small amount and for the minimum time to see if the light is ON (the window is equipped with a short UV wavelength filter) Short UV wavelength filter (green-yellow light) Do not operate the system without proper room ventilation as some Deep UV lamps produce significant levels of OZONE This machine has an EMO (Emergency Off) switch/button mounted on the top panel to the right of the keyboard. The EMO switch should be pressed only in an emergency. An emergency would be fire, smoke, electrocution hazards, or an impending injury to anyone using this particular piece of equipment. If the EMO is pressed notify NanoFAB staff immediately Read any posted NanoFab Engineering Change Notices (ECN) for any hardware, process or safety changes before running the tool. 4

5 2.0 HARDWARE 2.1 Vibration isolation table 2.2 Hg-Arc lamp calibrated at 365 nm wavelength 2.3 Wedge compensation: Internal N contact force 2.4 Manual substrate and mask loading 2.5 High resolution top and bottom side split field microscopes 2.6 Alignment stage precision micrometers 3.0 REQUIREMENTS 3.1 Training All users must be trained and authorized on the EVG 620 Mask Aligner to use this system. Training is supplied by a Nanofab staff member please contact the tool owner to schedule training. 3.2 System Restrictions Users must be logged into the EVG computer under operator name/password and log out of the computer when finished with the system Only NanoFab supplied recipes are to be stored on the system computer. Custom user recipes must be stored on a USB drive and should NOT be removed from the system or taken out of the NanoFab building USERS are not allowed to save UV light probe intensities to the hard drive. If you need the intensity uniformity checked notify staff. The UV intensity uniformity is checked on weekly schedule If a system fault is display during an initialization or run sequence notify Nanofab staff. 5

6 to 4 diameter wafer size and small sample pieces can be aligned/exposed. 2, 3 diameters wafers and small sample pieces need to be fixed at a 4 carrier wafer Only 5 x 5 x ( ) Mask can be used Recipe Requirements for all Alignment Modes: For top side, bottom side, Anodic bond and 1 st mask Do not change the Separation parameter of 30um. The alignment camera objectives are 10X so a 30um separation will be within the focal range of the objectives. If you are aligning a deep buried layer or using very thick resist > 50 um and both alignment marks are not in focus call staff to check Ensure your softbake (pre-exposure) bake process is correct so your wafers resist coating does not stick to the mask during wedge compensation Follow the Lamp startup procedure per section: Hg-Lamp Start-up procedure Follow the Lamp Shutdown procedure per section: Hg-Lamp Shut-down procedure 6

7 4. OPERATING PROCEDURES 4.1. System Pre-Checks Check to ensure the hard contact pressure is 2.3 ±.05 barr Check to ensure system N 2 and CDA pressure gauges are 0.6 barr. These gauges are located on the back of the tool as shown Check to ensure the Hg-Arc lamp is ON by moving the Lamp viewport window shutter just a very small amount and for the minimum time to see if the light is ON. (The Lamp viewport window is equipped with a short UV wavelength filter) Short UV wavelength filter (green-yellow light) If the Lamp looks like it is OFF, proceed to section to check the Lamp power supply LCD displays are within specifications. 7

8 Check to ensure Hg-Arc lamp power supply is ON and the power supply knob Constant Intensity setting CI. If these knob CI then move them to these settings. Check to ensure the Lamp power supply LCD displays are within the following spec limits. o The Lamp power at 350 ± 3 Watts. If the Lamp power supplies LCD displays are not within these specifications and tolerances call Nanofab Staff immediately If the Lamp power supply is already ON and within spec limits proceed to section 4.2. Preparing System for Alignment/Exposure to continue your photolithography process. If the lamp looks OFF verify this by the next section Check to ensure Lamp power supply is OFF by looking at the power supply LCD displays will be OFF and the Main power breaker/switch will be in the down position as shown. If the lamp is OFF, proceed to section Hg-Lamp start-up procedure to turn ON the lamp Hg-Lamp start-up procedure : 8

9 Check to ensure the EVG or Microsoft desktop screen software is not currently running or ON. If the EVG software is running logout and shutdown the tool per section to Shut down then Microsoft /EVG Explorer desktop by clicking on the red shut down icon then choose shutdown option the press OK. (failure to close the software and computer could result is damaging the computer when re-starting the Hg-Arc lamp) Turn OFF the system Power Control by turning the power switch to OFF position as shown. The computer monitor will turn OFF. Power Control by turning the power switch to OFF Check to Hg-Arc lamp power supply knob is set for Constant Intensity setting CI. If these knob are not set for CI then move them to the correct settings Turn on Hg-Arc lamp power supply ON by moving the power switch/breaker to the UP position as shown. The green LED should all be ON. Call staff if these LEDS do not turn ON 9

10 Turn ON (ignite) Hg-Arc lamp by press the Lamp Start button down for 1-2 seconds The Lamp Power LCD display should display 350 ± 3 Watts. If the Lamp Power still displays -000 Watts, then the Lamp didn t ignite. Press the Lamp Start button again per section If the power LCD does not display 350 ± 3 Watts call staff to check. 10

11 Let the Lamp warm up for 15 to 20 minutes then switch the voltage/current switch to voltage. The voltage should read 61+/- 2 volt. If the voltage value is out of spec, please contact Nanofab Personnel Turn ON the system Power Control by turning the power switch to ON position as shown. Power Control by turning the power switch to ON The tool computer will boot up and will start the Microsoft / EVG Explorer software Now the lamp is ON and stable, the EVG Explorer software is running on the desktop you are able to continue your photolithography process per section 4.2 Preparing System for Alignment/Exposure Check to insure the system Main Power and computer are ON.The computer screen should have the EVG 620 desktop displayed as show 11

12 4.3. Preparing System for Alignment/Exposure If you have not completed the System Pre-Checks in steps then you must complete those before proceeding When you first approach the screen the EVG 620 desktop screen is displayed as shown. Double click on EVG6XX tab to run wafer processing software In the grey login screen click on the LOGIN tab at the top left side of the screen. A pop up window will display. Enter user ID and password. User : 2222 Password : Wait for system initialization to complete. Keep fingers clear of the moving parts such exposure head and mask loader. Failure to do so may result in personal injury. During initialization system will prompt you to pull tray out. Gently pull tray out. Follow any initialization prompts and DO NOT do any manual operations unless there is a system prompt. If initialization is OK then continue, if there is any fault code message call staff. 12

13 Select the new folder or open folder icon to create a new or download a previously stored recipe If the new folder icon was chosen a temporary recipe template will display. At this point you create a recipe depending your device requirements ( i.e. 1 st mask, top side alignment, bottom side alignment,bond alignment) There are four process modes you enter by clicking on the arrow down drop down menu : Process, Process Mode, Exposure Mode, Contact Mode.Do not change the Separation parameter( 30um ), Mask thickness or Substrate thickness.the alignment camera objectives are 10X so a 30um separation will be within the focal range of the objectives. If you know the distance from the wafer alignment marks to the surface of the resist coated wafer ( i.e. buried layer+ thick resist) is much greater than 30 um the mask to wafer alignment mark focusing will not be a sharp image. Ok to enter your resist thickness. Here are examples of the four different process mode screens you will need to set up depending on your device requirements and processing step. 1 st mask top side align bottom align bond align 13

14 In each of these General Process Modes you will have to set up the following recipe parameters. Alignment Mode : top side,bottom side,bond,1 st mask. Exposure Mode : constant dose, dose interval, constant time, constant interval time. Contact Mode : hard contact, proximity, soft contact, V+H contact, vacuum contact Note : V+H contact and vacuum contact modes require the use of the vacuum contact oring. Again here are examples of the four different process mode screens you will need to set up depending on your device requirements and processing step. 1 st mask top side align bottom align bond align After you are finished creating a new recipe press the Green arrow at the bottom left to begin the Semi-Automatic alignment / exposure program Each Alignment Mode and Contact Mode recipes will have a different semi-automatic computer controlled sequences. The USER simply needs to follow the software prompt at all times. Note : V+H contact and vacuum contact modes require the use of the vacuum contact oring 14

15 To begin the process press Continue or Exit to stop (abort) the sequence. The first screen configure optics (no options they are10x). Press continue Insert mask holder as press continue (if mask holder is already inserted press continue). To insert the mask holder mate the two brackets 1 st and gently mate the single bracket at the top on the mask as shown Fix mask holder with the three thumbs screws finger tight as press continue Insert chuck and air line press and continue. (if chuck is already installed and air line is connected press continue) Insert load frame and Photo-Mask as shown and press continue. (mask chrome side down, mask edge aligned with white tray bumpers). 15

16 Move Tray in and press continue Move stage to center position by adjusting X,Y and theta micrometers to 5mm on the scaled line and ZERO on the dial. The micrometers zeroing accuracy is critical for the mask vacuum to hold the mask in the subsequent steps. After zeroing the micrometers press continue Adjust optics by using joystick knob for X, Y motion. Rotate knob for Z axis adjustments. Click on [ L ] and [ R ] optic icons to move objectives independently. Adjust left and right camera brightness and contrast as needed Use the pin icon, +, -, 4X, 2X,1X to find and save the recipe s optics step count, ZL and ZR information. 16

17 Adjust theta to correct for mask to optics rotation. After mask adjustments are finished press continue Gently move tray out and remove load frame then press continue. Note : If you do not remove load frame and press continue the auto optics focusing of the wafer marks and mask marks will not focus correctly Insert substrate as shown for WEC ( Wedge Compensation) and press continue. No load frame at this step Move tray in and press continue. Wait for WEC and Auto Position Optics to complete. 17

18 Align substrate to mask by using previously stored alignment mark positions then perform a fine manual alignment at the left and right side of the wafer by using X,Y, theta micrometers to optimize the alignment. Save alignment positions per section ( Note: Refer to section to for camera optics operation and functions ) When alignment is complete press Continue to exposure the wafer to UV light. Wait for exposure to finish. End of process screen will display. End of Process If you are finished processing your wafers press the Exit icon at lower right side of screen. The status screen will display To remove your wafer click on Tray out icon on the upper task bar. Follow the software prompts only. 18

19 To remove your Mask click on Mask out icon on the upper task bar. Follow the software prompts only Log out of the EV6XX program After you have logout of the EV6XX software, the Microsoft /EVG Explorer software will be on the desktop screen. Leave the computer ON with the EV6XX software running Hg-Lamp Shut-down procedure Leave the computer ON with the EV6XX software running Turn OFF Hg-Arc lamp by by moving the power switch/breaker to the down position as shown. 19

20 The Lamp Power display will show -000 Watts and the lamp will be Off Enter the required information in the logbook. 4.3 Anodic Bond Alignment If you have not completed the System Pre-Checks in steps then you must complete those before proceeding. For Anodic bonding alignment OK to leave the Hg-Arc lamp OFF When you first approach the screen the EVG 620 desktop screen is displayed as shown. Double click on EVG6XX tab to run wafer processing software. 20

21 4.3.3 In the grey login screen click on the LOGIN tab at the top left side of the screen. A pop up window will display. Enter user ID and password. User : 2222 Password : Wait for system initialization to complete. Keep fingers clear of the moving parts such exposure head and mask loader. Failure to do so may result in personal injury. During initialization system will prompt you to pull tray out. Gently pull tray out. Follow any initialization prompts and DO NOT do any manual operations unless there is a system prompt. If initialization is OK then continue, if there is any fault code message call staff. 21

22 4.3.5 Remove Mask Holder and 4 wafer chuck and carefully place on storage tray below tool Select the new folder or open folder icon to create a new or download a previously stored recipe If the new folder icon was chosen a temporary recipe template will display. There are four process modes you enter by clicking on the arrow down drop down menu : Process, Process Mode, Exposure Mode, Contact Mode.Do not change the Separation parameter( 30um ), Mask thickness or Substrate thickness Click on the Process arrow down drop down menu and click on Man. Anodic Bond and choose Crosshair mode After you chosen Man. Anodic Bond and Crosshair mode press the Green arrow at the bottom left to begin the Semi-Automatic Bond alignment The Man. Anodic Bond and Crosshair recipe will have a semi-automatic computer controlled sequences. The USER simply needs to follow the software prompt at all times. 22

23 To begin the process press Continue or Exit to stop (abort) the sequence. The first screen configure optics (no options ).Press continue The system will prompt for Save changes to Recipe1? Press NO. The process sequence will begin Insert Bonding tool with clamp levers facing up and press continue as shown (if mask holder is already inserted follow prompts to remove.) Remove clamps by rotating left clamp lever CW and right clamp lever CCW as shown then press continue. 23

24 Remove separation flags by gently pushing the arm outwards and press continue. (separation flag 12,4,8 o clock positions) Insert chuck connect vacuum line and press continue Insert bond glass (correct side up) and ruler as shown and press continue Load top substrate as shown and press continue. 24

25 Remove ruler and press continue Gently move tray IN as shown Move to center stage by adjusting the three X,Y and theta micrometers to 5mm on the scaled line and zero on the dial. Then press continue. Wait for wedge compensation to complete( WEC) Adjust microscope and press continue ( refer to sections to ) Adjust crosshair and press continue. 25

26 Gently move tray OUT and press continue Insert bond glass and press continue Insert ruler and press continue Insert bottom substrate and continue Remove ruler and press continue Gently move tray IN and press continue Pre-aligne substrate and press continue. Wait for WEC Aligne substrate and press continue. Wait for WEC Insert clamps by rotating left clamp lever CCW and right clamp lever CW as shown then press continue. Wait for bond alignment to complete Remove bond tool and press continue. 26

27 Gently move tray OUT. End of Process Press Exit icon at lower right side of screen. The status screen will display To move tray IN click on Tray out/in icon on the upper task bar. Follow the software prompts only Log out of system and enter the required information in the logbook. 27

28 5 RECIPES 5.1 Recipe Guidelines Users must store all of their recipes on a USB drive. This drive must not leave the NanoFab building.warning - Recipes left on the main system hard drive will be deleted Editing a recipe must be approved by Nanofab Staff before proceeding. 5.2 Creating a recipe You must be logged into the EVG system software in order to create a recipe To create a recipe to sections to Nanofab recipes 28

ABM's High Resolution Mask Aligner Features:

ABM's High Resolution Mask Aligner Features: ABM's High Resolution Mask Aligner is a very versatile instrument with interchangeable light sources which allow Near-UV (405-365 nm) as well as Mid- and Deep-UV (254 nm, 220 nm) exposures in proximity

More information

Title: Karl Suss MA150 Semiconductor & Microsystems Fabrication Laboratory Revision: E Rev Date: 03/01/06

Title: Karl Suss MA150 Semiconductor & Microsystems Fabrication Laboratory Revision: E Rev Date: 03/01/06 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Karl Suss MA150 Mask Aligner. All users are expected to have read and understood

More information

SUSS MICROTEC MJB4 MASK ALIGNER DOCUMENTATION

SUSS MICROTEC MJB4 MASK ALIGNER DOCUMENTATION SUSS MICROTEC MJB4 MASK ALIGNER DOCUMENTATION ANITHA & KARTHICK 1 Tool specification: Substrate size :1 up to 4 Mask size : Standard 2 x2 up to 5 x5 Alignment :X,Y-±5mm,theta-5⁰ Alignment Accuracy :Top

More information

Suss Aligner SOP. Suss Aligner SOP Page 1 of 6 Revision Scope 1.1 This document provides the procedure for operating the Suss Aligner.

Suss Aligner SOP. Suss Aligner SOP Page 1 of 6 Revision Scope 1.1 This document provides the procedure for operating the Suss Aligner. Suss Aligner SOP Page 1 of 6 Suss Aligner SOP 1. Scope 1.1 This document provides the procedure for operating the Suss Aligner. 2. Table of Contents 1. Scope... 1 2. Table of Contents... 1 Suss Aligner

More information

Karl Suss SB6 Bonder Operation Tool Id: Rev 01, 09/24/10

Karl Suss SB6 Bonder Operation Tool Id: Rev 01, 09/24/10 Karl Suss SB6 Bonder Operation Tool Id: 119.020 The technical content of this document is the responsibility of UCLA CNSI Integrated Systems Nanofabrication Cleanroom Los Angeles, CA Karl Suss SB6 Bonder

More information

NXQ8000 Series Mask Aligner

NXQ8000 Series Mask Aligner NXQ8000 Series Mask Aligner The NXQ8000 Production Mask Aligner and Front to Back Overlay Inspection System integrates the latest in Robotic Automation with state of the art next generation alignment stage

More information

SUSS SB6E BONDER SOP. October 2013

SUSS SB6E BONDER SOP. October 2013 SUSS SB6E BONDER SOP October 2013 The SUSS SB6 bonder is capable of silicon fusion, anodic, and eutectic bonding. There are two separate heads; one for anodic bonding, and another for fusion bonding. Definitions

More information

TECHNICAL SPECIFICATIONS

TECHNICAL SPECIFICATIONS TECHNICAL SPECIFICATIONS FOR THE SUPPLY OF A MASK ALIGNER FOR SCUOLA SUPERIORE SANT ANNA ALLEGATO A LOTTO 4 PROCEDURA APERTA IN LOTTI PER LA FORNITURA DI APPARECCHIATURE SCIENTIFICHE PER IL PROGETTO PIC

More information

ANNEALSYS RAPID THERMAL ANNEALER STANDARD OPERATING PROCEDURE

ANNEALSYS RAPID THERMAL ANNEALER STANDARD OPERATING PROCEDURE ANNEALSYS RAPID THERMAL ANNEALER STANDARD OPERATING PROCEDURE Purpose of this Instrument: Rapid thermal annealing, compound semiconductor annealing, and crystallization and densification. Location: Engineering

More information

for R&D and Production Advanced Technology OAI is a Silicon Valley-based manufacturer of advanced precision equipment

for R&D and Production Advanced Technology OAI is a Silicon Valley-based manufacturer of advanced precision equipment Advanced Technology for R&D and Production OAI is a Silicon Valley-based manufacturer of advanced precision equipment for the MEMS, Semiconductor, Nanotechnology, Microfluidics, MEMS Micro TAS and Flat

More information

HEIDELBERG MLA150 MASKLESS ALLIGNER

HEIDELBERG MLA150 MASKLESS ALLIGNER HEIDELBERG MLA150 MASKLESS ALLIGNER MLA150 Maskless Aligner 1. Introduction Heidelberg MLA150 Maskless Aligner is a full-scale production level laser writer which can be used to perform lithography directly

More information

5.2.2 Pins on the alignment detector are fragile. Use caution when inserting and removing the detector. Report any broken pins to staff.

5.2.2 Pins on the alignment detector are fragile. Use caution when inserting and removing the detector. Report any broken pins to staff. Title: Woollam Vase Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Woollam VASE. All users are expected to have read and understood

More information

NANOSPEC 4150 STANDARD OPERATING PROCEDURES

NANOSPEC 4150 STANDARD OPERATING PROCEDURES NANOSPEC 4150 STANDARD OPERATING PROCEDURES Version: 1.0 JAN 2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center TABLE OF CONTENTS 1.0 INTRODUCTION.. 3 2.0 HARDWARE....... 3 3.0 OPERATING

More information

SUSS MJB4. Manual Aligner For Research, Development and Operator Assisted Production October, 2009

SUSS MJB4. Manual Aligner For Research, Development and Operator Assisted Production October, 2009 SUSS MJB4 Manual Aligner For Research, Development and Operator Assisted Production October, 2009 Overview Product Portfolio Aligner MA/BA 8 MA200Compact LithoFab200 MJB4 MA300Plus MA/BA 6 MA150e LithoPack300

More information

University of Minnesota Nano Fabrication Center Standard Operating Procedure

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name: University of Minnesota Nano Fabrication Center Coral Name: hs-scope Revision Number: 1.5 Model: HS200A Revisionist: M. Fisher Location: Bay 1 Date: 9/12/2013 1 Description The Hyphenated

More information

Title: Heidelberg DWL66+ Semiconductor & Microsystems Fabrication Laboratory Revision: B Rev Date: 05/03/2017

Title: Heidelberg DWL66+ Semiconductor & Microsystems Fabrication Laboratory Revision: B Rev Date: 05/03/2017 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Heidelberg DWL66+. All users are expected to have read and understood this document.

More information

Heidelberg MLA-150 Standard Operating Procedure

Heidelberg MLA-150 Standard Operating Procedure Heidelberg MLA-150 Standard Operating Procedure CORAL Name: Model: Location: Purpose: Author: MLA-150 Heidelberg MLA150 Maskless Aligner TRL Photo-Au Room Direct-Write Lithography Heidelberg Instruments

More information

Raith e_line Electron Beam Lithography

Raith e_line Electron Beam Lithography Raith e_line Electron Beam Lithography Standard Operating Procedure 1 (For an un-patterned sample) Revision: 7.0 Last Updated: Feb.18/2015, Revised by Mohamad Rezaei Overview This document will provide

More information

DWL-66 Laser Lithography System Operating Procedure. Effective Date: 09/24/2012 Author(s): Jiong Hua Phone:

DWL-66 Laser Lithography System Operating Procedure. Effective Date: 09/24/2012 Author(s): Jiong Hua Phone: DWL-66 Laser Lithography System Operating Procedure Effective Date: 09/24/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Laser Lithography System, Optical

More information

VEECO FPP Point Probe Operating Manual

VEECO FPP Point Probe Operating Manual VEECO FPP-5000 4-Point Probe Operating Manual Version: 1.0 May 2013 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research and Teaching Facility TABLE OF CONTENTS 1. Introduction....2 1.1 Scope of Work....2

More information

Equipment Standard Operating Procedure Kimberly Appel

Equipment Standard Operating Procedure Kimberly Appel Date Created: June 26, 2004 Date Modified: The Flexus 2320 Equipment Standard Operating Procedure Kimberly Appel 1. Purpose 1.1. The Flexus Thin Film Stress Measuring Apparatus (TFSMA) measures the changes

More information

PANALYTICAL X PERT PRO XRD

PANALYTICAL X PERT PRO XRD STANDARD OPERATING PROCEDURE: PANALYTICAL X PERT PRO XRD Purpose of this Instrument: Identification of signal-phase materials and multi-phase mixture; Quantitative determination of amounts of different

More information

JEOL CarryScope SEM Revision /07/17 Page 1 of 7. JEOL CarryScope SEM

JEOL CarryScope SEM Revision /07/17 Page 1 of 7. JEOL CarryScope SEM Page 1 of 7 JEOL CarryScope SEM The JEOL CarryScope is a compact and portable SEM that utilizes a standard tungsten filament. It is to be used for inspecting and measuring samples processed in the NRF

More information

Operating Procedure for Horiba Raman Microscope

Operating Procedure for Horiba Raman Microscope Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch

More information

NRF Ellipsometer SOP Revision /19/15 Page 1 of 14. Ellipsometer SOP

NRF Ellipsometer SOP Revision /19/15 Page 1 of 14. Ellipsometer SOP Page 1 of 14 Ellipsometer SOP The J. A. Woollam M88 is a spectroscopic ellipsometer used to measure film thickness and optical constants of transparent/semi-transparent thin films. It uses a Xenon arc

More information

TECHNICS MACRO-RIE 8800 STANDARD OPERATING PROCEDURES

TECHNICS MACRO-RIE 8800 STANDARD OPERATING PROCEDURES TECHNICS MACRO-RIE 8800 STANDARD OPERATING PROCEDURES Version: 1.0 Sept 2017 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center TABLE OF CONTENTS 1.0 DESCRIPTION.... 3 2.0 SAFETY....... 3

More information

Prism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis

Prism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis Start Up: Upon entering the laser room turn on the wall mounted Laser Power Button by pulling it away from the wall. Turn on Shutter controllers (toggle switch on back of unit). There should be a U in

More information

Contents. - i - Ver.:2

Contents. - i - Ver.:2 Contents 1 Accessories of D-Point 2... 2 2 Connecting Projector to PC... 4 3 Pen Action... 4 4 Driver Installation... 5 5 Configuration Setting... 8 6 D-Point 2 Pull-up Menu... 12 7 DT02 Multi-touch Interactive

More information

Table of Contents. Chapter 1. Safety Precautions Notice Chapter 2. Components and accessories Chapter 3. Camera components...

Table of Contents. Chapter 1. Safety Precautions Notice Chapter 2. Components and accessories Chapter 3. Camera components... P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046 P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046 Table of Contents Chapter 1. Safety Precautions Notice...

More information

Heidelberg Pattern Generator SOP

Heidelberg Pattern Generator SOP Heidelberg Pattern Generator SOP Page 1 of 15 Heidelberg Pattern Generator SOP 1. Scope 1.1 This document provides the operating procedures for the Heidelberg Pattern Generator with Version 3.12.5 software.

More information

User Operation Manual

User Operation Manual User Operation Manual Manual Version: 1.3 For Single Cell Bioanalyzer (SCB-402) Doc# 102-SCB-402-1.3 Table of Contents 1. General Information... 1 1.1 How to Use This Manual... 1 1.2 Operation Safety...

More information

Working Copy For Review Purposes Only!

Working Copy For Review Purposes Only! Working Copy For Review Purposes Only! Version #1.5 08/29/06 Aaron Holland Up-dated D.Turnbull 10-18-2006 (MarkerSOP3.doc) Up-dated D. Turnbull 12-11-06 (MarkerSOP4.doc) Up-dated D.Turnbull 05-12-2008

More information

FILMETRICS F20 STANDARD OPERATION PROCEDURE

FILMETRICS F20 STANDARD OPERATION PROCEDURE Arizona State University NanoFab FILMETRICS F20 STANDARD OPERATION PROCEDURE Rev C Table of Contents Contents Table of Contents...1 1. Purpose / Scope...2 2. Reference Documents...2 3. Equipment / Supplies

More information

UHP-T-LED User Manual

UHP-T-LED User Manual UHP-T-LED User Manual Ver 03 July 2014 UHP-T-LED System Description The UHP-T-LED is an Ultra High Power LED light source for fluorescence microscopy, Optogenetics, High power illumination and other applications.

More information

Mercury Manual No. IM-M1000 Rev G

Mercury Manual No. IM-M1000 Rev G Mercury 1000 Analog Output Encoder System Installation Manual and Reference Guide Manual No. IM-M1000 Rev G Introduction MicroE Systems was founded to advance encoder technology to a level never before

More information

Spectroscopic Ellipsometer --- J. A. Woollam alpha-se

Spectroscopic Ellipsometer --- J. A. Woollam alpha-se Spectroscopic Ellipsometer --- J. A. Woollam alpha-se Introduction Figure 1: J. A. Woollam alpha-se spectroscopic ellipsometer An ellipsometer measures the change in polarization as light reflects or transmits

More information

Flip-Chip-Aligner-Bonder Model M-8HP

Flip-Chip-Aligner-Bonder Model M-8HP Flip-Chip-Aligner-Bonder Model M-8HP 1. Introduction DO NOT CHANGE THE SETTINGS OF THE VIDEO CAMERAS UNLESS YOU ARE AUTHORIZED TO CALIBRATE THE MACHINE AND KNOW WHAT YOU ARE DOING! The laminar flow hood

More information

Probe station system

Probe station system Probe station system Manufacturer: Micromanipulator Mode: 450PM-HR Descriptions: The Model 450PM 8 probe station offers stable and reliable probing performance. It is based on a design of Micromanipulator

More information

Polarization Microscope. Omano OM349P Series. User Guide

Polarization Microscope. Omano OM349P Series. User Guide Polarization Microscope Omano OM349P Series User Guide Table of Contents Getting Started... 3 Components... 3 Technical Specifications... 5 Setup Instructions... 6 Basic Operation... 7 Changing the Light

More information

Avonic AV-CON300. PTZ Camera Controller

Avonic AV-CON300. PTZ Camera Controller Avonic AV-CON300 PTZ Camera Controller User Manual Version 1.0 Update notes: Join Avonic linkedin.com/company/avonic twitter.com/avonic1 facebook.com/avonic www.avonic.eu 1 Contents Inhoud Contents...

More information

R I T Title: Lam 4600

R I T Title: Lam 4600 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Lam 4600. All users are expected to have read and understood this document. It

More information

OPERATING INSTRUCTIONS:

OPERATING INSTRUCTIONS: List OPERATING INSTRUCTIONS: Preparation For Installation, Product Identification Charging The Batteries ing And Adjusting The Lift Mounting Head s Brackets Mounting Clamps And Quick-Locking Arms Installation

More information

OMM200 and OMM200T. Metallurgical Microscope

OMM200 and OMM200T. Metallurgical Microscope OMM200 and OMM200T Metallurgical Microscope Instruction Manual (Please read the manual carefully before using the microscope) Contents Safety ----------------------------------------------------------------------------

More information

Preparing for EBeam Write

Preparing for EBeam Write Preparing for EBeam Write Electron beam lithography is a time-intensive process that takes many steps to properly prepare for writing. In order to make the process as efficient and simple as possible,

More information

MT8050iE series. Installation Instruction (1) (2)

MT8050iE series. Installation Instruction (1) (2) MT8050iE series 3 Installation Instructions Installation Instruction Secure the operator panel in position, using all the fastening holes and the provided brackets and screws: (A) 1 Installation and Startup

More information

VDL-004. Miniature Manual Variable Optical Delay Line. User Guide

VDL-004. Miniature Manual Variable Optical Delay Line. User Guide VDL-004 Miniature Manual Variable Optical Delay Line User Guide Version: 1.1 Date: July 24, 2018 General Photonics Corporation is located in Chino, California. For more information visit the company's

More information

Xetch SOP Page 1 of 7 Revision

Xetch SOP Page 1 of 7 Revision Xetch SOP Page 1 of 7 Xetch System SOP 1. Scope 1.1 This document provides procedures and process information for etching silicon wafers with the Xetch xenon difluoride etching system. 2. Table of Contents

More information

INSTRUCTION MANUAL POLARIZING & ORE MICROSCOPES. Polarizing Microscopes. Ore Microscopes. Model: NYMC62B00

INSTRUCTION MANUAL POLARIZING & ORE MICROSCOPES. Polarizing Microscopes. Ore Microscopes. Model: NYMC62B00 POLARISING & ORE MICROSCOPES INSTRUCTION MANUAL Monocular Microscope Binocular Microscope Trinocular Microscope Polarizing Microscopes POLARIZING & ORE MICROSCOPES Model: NYMC62B00 Monocular Microscope

More information

University of Minnesota NanoFabrication Center Standard Operating Procedure

University of Minnesota NanoFabrication Center Standard Operating Procedure Equipment Name: JEOL SEM Coral Name: jeol-sem Revision Number: 4 Model: JSM-6610LV Revisionist: K. Roberts Location: Area 3 Date: 9/17/2013 1 Description The JSM-6610LV is a scanning electron microscope

More information

R I T. Title: Rudolf Ellipsometer Auto-EL IV. Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date:12/07/17 1 SCOPE

R I T. Title: Rudolf Ellipsometer Auto-EL IV. Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date:12/07/17 1 SCOPE Title: Rudolf Ellipsometer Auto-EL IV Fabrication Laboratory Revision: C Rev Date:12/07/17 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the

More information

Astrel Instruments. AST-8300-B GETTING STARTED rev A. Getting Started 1/21 rev A

Astrel Instruments. AST-8300-B GETTING STARTED rev A. Getting Started 1/21 rev A Astrel Instruments GETTING STARTED rev A Getting Started 1/21 rev A Getting started with the Astrel Instruments AST8300 camera Thank you for purchasing an Astrel Instruments AST8300 CCD camera. Your new

More information

Oxford DSiE Revision /25/17 Page 1 of 9. Oxford DSiE SOP

Oxford DSiE Revision /25/17 Page 1 of 9. Oxford DSiE SOP Page 1 of 9 Oxford DSiE SOP The PlasmaPro 100 is a Inductively coupled plasma device (ICP) configured to provide effective high aspect ratio etching of 4 inch silicon wafers, and pieces when a carrier

More information

Updated GCA Stepper Training Guide & References

Updated GCA Stepper Training Guide & References Updated 06-04-03 GCA Stepper Training Guide & References General Information about system... 1 Aligning reticle (mask) to system... 2 Logging into computer/checking the system... 3 Setting up / editing

More information

CoSMoS Starter Guide 1.4 Hoskins Lab Last Modified 03/03/2017 Chris DeCiantis

CoSMoS Starter Guide 1.4 Hoskins Lab Last Modified 03/03/2017 Chris DeCiantis Contents Start Up:... 2 Turn on Shutter controllers... 2 Turn on Lasers... 2 Turn on other equipment:... 2 Load Slide onto Sample Holder:... 2 Turn on Computer/ Start Labview/Glimpse:... 3 Shutters Activate.vi

More information

STANDARD OPERATING PROCEDURE: HORIBA SPECTROFLUOROMETER_STEADY STATE

STANDARD OPERATING PROCEDURE: HORIBA SPECTROFLUOROMETER_STEADY STATE STANDARD OPERATING PROCEDURE: HORIBA SPECTROFLUOROMETER_STEADY STATE Purpose of this Instrument: Essential tool for characterizing the relationship between absorbed and emitted photos at specified wavelengths.

More information

Flexus 2320 SOP Revision 1 01/25/16 Page 1 of 7. Flexus 2320 SOP

Flexus 2320 SOP Revision 1 01/25/16 Page 1 of 7. Flexus 2320 SOP Page 1 of 7 Flexus 2320 SOP The Flexus 2320 determines wafer curvature by measuring the angle of deflection of a laser beam off the surface 6, 4, 3 and 2 diameter substrates. Film stress is determined

More information

DV5-2. SC-DV5_Main.jpg. SC-DV5_Main.jpg. SC-DV5_Main.jpg

DV5-2. SC-DV5_Main.jpg. SC-DV5_Main.jpg. SC-DV5_Main.jpg DV5-2 SC-DV5_Main.jpg SC-DV5_Main.jpg SC-DV5_Main.jpg Overview Camera: Overview -screen Display: The DV5-2 body camera is an advanced recording system with many options and features available. Please review

More information

MPI TS150-THZ 150 mm Manual Probe System

MPI TS150-THZ 150 mm Manual Probe System MPI TS150-THZ 150 mm Manual Probe System Industry s first explicitly designed probe system for accurate measurements at mm-wave and sub-mm wave (THz) frequency range FEATURES / BENEFITS Variety of Applications

More information

CoSMoS Starter Guide 1.6 Hoskins Lab Last Modified 03/30/2017 Chris DeCiantis

CoSMoS Starter Guide 1.6 Hoskins Lab Last Modified 03/30/2017 Chris DeCiantis Contents Start Up:... 2 Turn on Shutter controllers... 2 Turn on Lasers... 2 Turn on other equipment:... 2 Load Slide onto Sample Holder:... 2 Turn on Computer/ Start Labview/Glimpse:... 3 Shutters Activate.vi

More information

2.2. Facilities Requirements

2.2. Facilities Requirements 2.2. Facilities Requirements Facilities requirements for the alpha-se system are listed in Table 2-1 and the system dimensions are given in Figure 2-1. As shown in Figure 2-2, the preferred clear work

More information

Reference Guide for the KODAK Picture Kiosk GS Compact. Software Version 2.0

Reference Guide for the KODAK Picture Kiosk GS Compact. Software Version 2.0 Reference Guide for the KODAK Picture Kiosk GS Compact Software Version 2.0 Eastman Kodak Company, 2008 Contents Product Overview... 4 Console... 4 605 Photo Printer... 5 Product Choices... 6 Startup and

More information

Complete Barrel Measuring and Inspection System. PS Series. User s manual

Complete Barrel Measuring and Inspection System. PS Series. User s manual Complete Barrel Measuring and Inspection System PS Series User s manual SAFETY PRECAUTIONS... 3 ELECTROMAGNETIC COMPATIBILITY... 3 LASER SAFETY... 3 GENERAL INFORMATION... 3 BASIC DATA AND PERFORMANCE

More information

Product Parts Front view... 2 Rear view... 2 Inside the product... 3 Scanner parts... 3 Control panel... 4

Product Parts Front view... 2 Rear view... 2 Inside the product... 3 Scanner parts... 3 Control panel... 4 Table of Contents Product Parts Front view................................................................. 2 Rear view.................................................................. 2 Inside the product...........................................................

More information

NE400 - LED MOTION LIGHT WITH VIDEO CAMERA

NE400 - LED MOTION LIGHT WITH VIDEO CAMERA USER MANUAL NE400 - LED MOTION LIGHT WITH VIDEO CAMERA Table of contents 1. Contents of package 2. NightWatcher NE400 Motion Light with Video Camera 3. Assembly 4. Installation 5. Programming your NightWatcher

More information

Primary Use. Operating Principle

Primary Use. Operating Principle Primary Use The Leica DVM6 is an optical microscope that has the ability observe samples at a high magnification at a high resolution. The microscope allows users to view their sample with up to a 2350x

More information

MPI TS150-AIT 150 mm Manual Probe System

MPI TS150-AIT 150 mm Manual Probe System MPI TS150-AIT 150 mm Manual Probe System Industry s first explicitly designed 150 mm probe system providing accurate tests for mm-wave, THz, and automated impedance tuner applications FEATURES / BENEFITS

More information

User Operation Manual: Motorized Valve Controller for Valved Sources

User Operation Manual: Motorized Valve Controller for Valved Sources User Operation Manual: Motorized Valve Controller for Valved Sources Table of Contents BEFORE YOU BEGIN...2 Safety Precautions...2 Product Concerns...2 Specifications...3 Support...3 INTRODUCTION...4 Product

More information

Operating instructions for DWL66 laser writer

Operating instructions for DWL66 laser writer Operating instructions for DWL66 laser writer Things to remember: 1. The DWL66 runs a 405-nm semiconductor laser. The laser beam path is covered and you must not remove any of the covers. To assure that

More information

Start-up system. Nikon Wide Field Microscope Ee1078. Complete manual

Start-up system. Nikon Wide Field Microscope Ee1078. Complete manual Start-up system Switch on the power socket located on the floor Switch devices on in the order of: o Mercury lamp control-unit switch on and ignite o Halogen lamp o Incubator control-box o Microscope stand

More information

Rigaku Mercury CCD on Ultima III PXRD in CSL 1011

Rigaku Mercury CCD on Ultima III PXRD in CSL 1011 Rigaku Mercury CCD@ FSU for Ultima III PXRD [CSL 1011] Quick Shut-down and start-up procedure for Rigaku MercuryCCD on Ultima III XRD [Updated by Thayumanasamy Somasundaram, V.: Apr 18, 2014; Original

More information

Operational Safety Procedure (OSP) For The. Laser Bay. At the. Texas Center for High Intensity Laser Science. Document Number TPW-D-0011-A

Operational Safety Procedure (OSP) For The. Laser Bay. At the. Texas Center for High Intensity Laser Science. Document Number TPW-D-0011-A Operational Safety Procedure (OSP) For The Laser Bay At the Texas Center for High Intensity Laser Science Document Number TPW-D-0011-A Texas Center for High Intensity Laser Science Dept. of Physics, The

More information

Integrated video camera ImageCam USER S INFORMATION. Camera control panel

Integrated video camera ImageCam USER S INFORMATION. Camera control panel Camera control panel FRREZE image storage BLACK & WHITE changing colour / blackwhite DISPLAY status sign ZOOM TELE digital zoom high ZOOM WIDE digital zoom low POS PRESET set function memory POS RESET

More information

Oxford ICP 2-step DRIE SOP

Oxford ICP 2-step DRIE SOP Oxford ICP 2-step DRIE SOP Oxford ICP 2-step DRIE SOP Page 1 of 13 1. Scope 1.1 This document provides operating procedures for the Oxford 100 ICP 2-step DRIE. 2. Table of Contents 1. Scope... 1 2. Table

More information

National Optical & Scientific Instrument Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR

National Optical & Scientific Instrument Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR National Optical & Scientific Instrument Inc. 6508 Tri-County Parkway Schertz, Texas 78154 Phone (210) 590-9010 Fax (210) 590-1104 INSTRUCTIONS FOR MODEL DC4-156-S COMPOUND BIOLOGICAL MICROSCOPE WITH DIGITAL

More information

MEIJI METALLURGICAL MICROSCOPE STANDARD OPERATING PROCEDURE

MEIJI METALLURGICAL MICROSCOPE STANDARD OPERATING PROCEDURE MEIJI METALLURGICAL MICROSCOPE STANDARD OPERATING PROCEDURE Purpose of this Instrument: This instrument is for illuminating solid specimens to identify, inspect and measure features of interest. Location:

More information

ELIONIX ELECTRON BEAM LITHOGRAPHY SYSTEM

ELIONIX ELECTRON BEAM LITHOGRAPHY SYSTEM Wisconsin Center for Applied Microelectronics 1550 Engineering Drive Phone: 608/262-6877 Madison, WI 53706 Fax: 608/265-2614 ELIONIX ELECTRON BEAM LITHOGRAPHY SYSTEM ELS-G100 Sample Loading and Direct

More information

DV2. Alignment Procedure. Install DV2 on Microscope NOTE: PLEASE READ THE ENTIRE PROCEDURE BEFORE YOU BEGIN ALIGNMENT OF THE DV2. Alignment Procedure

DV2. Alignment Procedure. Install DV2 on Microscope NOTE: PLEASE READ THE ENTIRE PROCEDURE BEFORE YOU BEGIN ALIGNMENT OF THE DV2. Alignment Procedure H I G H - P E R F O R M A N C E E M C C D & C C D C A M E R A S F O R L I F E S C I E N C E S DV2 This document provides a straightforward, step-by-step outline of the alignment procedure for the Photometrics

More information

Co-Lase Users Guide. DM00001 Rev A (DCN 031) These instructions are protected by copyright. All rights reserved.

Co-Lase Users Guide. DM00001 Rev A (DCN 031) These instructions are protected by copyright. All rights reserved. Co-Lase Users Guide These instructions are protected by copyright. All rights reserved., Inc. http://www.propel-labs.com 131 East Lincoln Avenue, Suite 200 Fort Collins, CO 80524 1 Table of Contents Chapter

More information

Vividia HM-250 HDMI/LCD/USB/TV Portable. Digital Measurement Microscope

Vividia HM-250 HDMI/LCD/USB/TV Portable. Digital Measurement Microscope Vividia HM-250 HDMI/LCD/USB/TV Portable Digital Measurement Microscope Table of Contents 1. Precaution. 3 2. Preparation..5 Package Contents.. 6 Part Names & Functions..6 3. Specifications.... 7 4. Use

More information

Colour Set-up Version 7.1

Colour Set-up Version 7.1 1 of 35 Colour Set-up Version 7.1 REVISION HISTORY Date Rev Description Author Approved 4/07/14 G Various updates including Win8. CW/FMC/DMG 26/05/15 G Updated to include ICC Profiles NM 2 of 35 Table

More information

iq DIGITAL PICTURE FRAME iq Digital Picture Frame DPF701SB USER GUIDE

iq DIGITAL PICTURE FRAME iq Digital Picture Frame DPF701SB USER GUIDE iq Digital Picture Frame DPF701SB USER GUIDE 1 Table of Contents Important Safety Precautions... 3 Cleaning the LCD Screen... 3 Cleaning the Digital Picture Frame... 3 Introduction... 4 What s in the Box...

More information

3 Indexer Installation For PRSalpha Tools

3 Indexer Installation For PRSalpha Tools 888-680-4466 ShopBotTools.com 3 Indexer Installation For PRSalpha Tools Copyright 2016 ShopBot Tools, Inc. page 1 Copyright 2016 ShopBot Tools, Inc. page 2 Table of Contents General Safety and Precautions...5

More information

XRADIA microxct Manual

XRADIA microxct Manual XRADIA microxct Manual Multiscale CT Lab Table of Contents 1. Introduction and Basics 1.1 Instrument Parts 1.2 Powering up the system 1.3 Preparing your sample 2. TXM Controller 2.1 Starting up 2.2 Finding

More information

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210) INSTRUCTIONS FOR National Optical & Scientific Instruments Inc. 6508 Tri-County Parkway Schertz, Texas 78154 Phone (210) 590-9010 Fax (210) 590-1104 INSTRUCTIONS FOR MODEL DC5-163 COMPOUND BIOLOGICAL MICROSCOPE WITH DIGITAL

More information

The Dektak XT is a 2D contact profilometer used for step height, pitch and surface roughness

The Dektak XT is a 2D contact profilometer used for step height, pitch and surface roughness Dektak XT 2D Profilometer Operation Manual The Dektak XT is a 2D contact profilometer used for step height, pitch and surface roughness measurements. Vision 64 application software controls the system

More information

Lam AutoEtch 490 Plasma Etch System Operations and Maintenance Manual

Lam AutoEtch 490 Plasma Etch System Operations and Maintenance Manual Fabrication Laboratory Revision: E - 02/10/2009 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Lam AutoEtch 490 Plasma Etch System

More information

User Manual. Binocular Zoom Stereo Microscope with Boom Stand. MicroscopeNet.com

User Manual. Binocular Zoom Stereo Microscope with Boom Stand. MicroscopeNet.com User Manual Binocular Stereo Microscope with Boom Stand Model W42C1 Series MicroscopeNet.com Table of Contents i. Caution.. 1 ii. Care and Maintenance... 1 1. Component Illustration... 2 2. Installation...3

More information

MLA100 Maskless Aligner

MLA100 Maskless Aligner Quick Guide MLA100 Maskless Aligner Doc. No.: DWL-HI-060 Revision: 5 (August 2017) Wizard version: 1.9 Copyright 2017 by Heidelberg Instruments Job Setup As mentioned before (Wizard Description), the MLA100

More information

SCHOTT FLM 4 Fiber Lighting Modul. Operating Instructions

SCHOTT FLM 4 Fiber Lighting Modul. Operating Instructions SCHOTT FLM 4 Fiber Lighting Modul Operating Instructions Contents 1. Important information 3 2. Scope of Delivery 3 3. Intended use 3 4. Safety information 4 5. Operation 5 5.1. LED Optics Module 5 5.2.

More information

3025 MICROSCOPE SERIES INSTRUCTIONS

3025 MICROSCOPE SERIES INSTRUCTIONS 3025 MICROSCOPE SERIES INSTRUCTIONS Polysciences, Inc. 400 Valley Road Warrington, PA 18976 (800) 523-2575 / (215) 343-6484 (800) 343-3291 / (215) 343-0214 fax info@polysciences.com www.polysciences.com

More information

Multi-Photon Training

Multi-Photon Training Multi-Photon Training Overview This training will take approximately 4 hours. The first 2 hours will be spent learning one-on-one with your trainer, while the following 2 hours will be your opportunity

More information

Manual ANGSTROM grid confocal microscope

Manual ANGSTROM grid confocal microscope Manual ANGSTROM grid confocal microscope Switching on the system 1) Start by switching on the different components of the microscope system. When using live cell imaging also start the temperature and

More information

200 mm Semi-automated Probe System with BlueRay Technology

200 mm Semi-automated Probe System with BlueRay Technology PA200BlueRay 200 mm Semi-automated Probe System with BlueRay Technology DATA SHEET The sets a new standard for high-speed accuracy. Its precision ensures smooth probe landing with safe, repeatable electrical

More information

Laser Alignment. For the Epilog Fusion CO2 10/10/2013

Laser Alignment. For the Epilog Fusion CO2 10/10/2013 EPILOG LASER 16371 Table Mountain Parkway Golden, Colorado 80403 Phone 303-215-9171 FAX 303-277-9669 www.epiloglaser.com For the Epilog 10/10/2013 Requirements: Tools: 5/32 Allan Wrench 3/32 Allan Wrench

More information

TrueChrome Camera Quick Start Guide

TrueChrome Camera Quick Start Guide TrueChrome Camera Quick Start Guide This Quick Start Guide is for TrueChrome series cameras. To ensure security and best use experience, we recommend that you read this guide thoroughly and carefully before

More information

Ebeam Advantest F7000+VD02 user manual

Ebeam Advantest F7000+VD02 user manual Ebeam Advantest F7000+VD02 user manual Eric (Mita Lab) 2016/02/24 Please, leave this manual here. You can find an electronic version in the TakedaCR Wiki: http://takeda-cr.t.u-tokyo.ac.jp (accessible only

More information

Condensed AFM operating instructions:

Condensed AFM operating instructions: Condensed AFM operating instructions: 1. Log onto system at access controller 2. Take the parts you need to mount a probe out of the drawers. You need the appropriate probe holder, tweezers (these are

More information

Product Specifications

Product Specifications Product Specifications The ksa RateRat Pro is a turnkey, real-time, in-situ optical reflectance probe designed for deposition monitoring of semi-absorbent thin films. The RateRat Pro measures deposition

More information

Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017

Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017 Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017 Instrument instructions can be found at: http://academic.bowdoin.edu/chemistry/resources/instructions.shtml If you have any problems with the

More information

Firearms & Tool Marks Comparison Microscope LEEDS FORENSIC SYSTEMS. Discovery

Firearms & Tool Marks Comparison Microscope LEEDS FORENSIC SYSTEMS. Discovery Firearms & Tool Marks Comparison Microscope LEEDS FORENSIC SYSTEMS Discovery The Discovery, the world s premier firearms and tool marks comparison microscope, features Zeiss optics and provides the largest

More information