1 Brief Operating Instructions Smartec S CLD134 Conductivity Measuring System These instructions are Brief Operating Instructions. For detailed information, please read the Operating Instructions. The complete device documentation comprises: these Brief Operating Instructions the Operating Instructions KA401C/07/en/ valid from: Software version 1.13

2 Table of contents Smartec S CLD134 Table of contents 1 Safety instructions Designated use Installation, commissioning and operation Operational safety Installation Installation of the separate version Installation of the compact version or CLS54 sensor for separate version Wiring Cable glands Connection diagram Structure and termination of measuring cable Commissioning Function check Key assignment Quick setup Calibration Safety instructions # Warning! Electric shocks can cause death or severe injury. Proceed with extreme care when touching cables and terminals. If the device/measuring system is installed in a high-voltage environment and malfunctions or installation errors occur, the terminals or the device/measuring system themselves might be high-voltage energized. 1.1 Designated use Smartec S CLD134 is a field-tested and reliable transmitter used to determine the conductivity of liquid media. It is particularly suitable for use in the foodstuffs industry. Any other use than the one described here compromises the safety of persons and the entire measuring system and is, therefore, not permitted. The manufacturer is not liable for damage caused by improper or non-designated use. 1.2 Installation, commissioning and operation The device/measuring system may only be installed, connected, operated and maintained by trained technical personnel (e.g. certified electrician). The technical personnel must strictly adhere to the Operating Instructions, prevailing standards, legal regulations and certificates (depending on application). If the Brief Operating Instructions do not provide sufficient information, you must read the Operating Instructions. There, you can find detailed information on the device. The operator may only perform modifications and repairs of the device/measuring system that are explicitly permitted in the Operating Instructions. Do not operate damaged products and secure them against unintentional commissioning. Mark the damaged product as being defective. If faults can not be rectified, the products must be taken out of service and secured against unintentional commissioning. 2 Endress+Hauser

3 Smartec S CLD134 Installation 1.3 Operational safety The transmitter has been designed and tested according to the state of the art and left the factory in perfect functioning order. Relevant regulations and European standards have been met. As the user, you are responsible for complying with the following safety conditions: Guidelines for explosion protection Installation instructions Local prevailing standards and regulations. " Caution! Pay attention to the technical data on the nameplate! 2 Installation 2.1 Installation of the separate version Wall mounting For wall mounting, attach the mounting plate to the wall by drilling holes as required. Anchors and screws are to be provided by the operator.! Note! Wall mounting is not recommended for hygienically sensitive areas. Post mounting A mounting kit for installing the housing on horizontal or vertical posts or pipes (max. Ø 60 mm / Ø 2.36") is available as an accessory.! Note! For use in hygienically sensitive areas, shorten the threads as much as possible. 1. Remove the mounting plate. 2. Insert the holding bars through the pre-drilled holes of the mounting plate and screw the mounting plate onto the transmitter. 3. Use the brackets to install the Smartec S on the post or pipe. Æ max. 60 / 2.36 Fig. 1: Post mounting of separate version mm / inch a Endress+Hauser 3

4 Wiring Smartec S CLD Installation of the compact version or CLS54 sensor for separate version! Note! Perform an air set and calibrate the sensor before installing the compact version or the sensor. Install the compact version or the CLS54 sensor directly on the pipe or vessel socket via the process connection (depending on ordered version). 1. When installing the Smartec S CLD134 or the sensor, make sure that the flow opening of the sensor is oriented in the flow direction of the medium. An orientation arrow on the sensor facilitates orientation. 2. Tighten the flange. 3 Wiring # Warning! The electrical connection must only be carried out by a certified electrician. Technical personnel must have read and understood the instructions in this manual and must adhere to them. Ensure that there is no voltage at the power cable before beginning the connection work. 3.1 Cable glands A B Fig. 2: Terminal assignment of cable glands A Separate version 1 Cable gland for analog output, binary input 2 Cable gland for alarm contact 3 Cable gland for power supply 4 Housing ground 5 Pressure comp. element PCE (Goretex - filter) 6 Cable gland for sensor connection, M 16x1.5 B Compact version Cable gland for analog output, binary input Cable gland for alarm contact Cable gland for power supply Housing ground Pressure comp. element PCE (Goretex - filter) a Endress+Hauser

5 Smartec S CLD134 Wiring 3.2 Connection diagram A ma Lf F B ma optional S G C V D V H E V ~ ~ I PE Fig. 3: Electrical connection of Smartec S CLD134 A Signal output 1 conductivity B Signal output 2 temperature C Auxiliary power output D Binary input 2 (MRS 1+2) E Binary input 1 (hold / MRS 3+4) F G H I MRS a Conductivity sensor Temperature sensor Alarm (contact position: no current) Power supply Remote parameter set switching (measuring range switching) Endress+Hauser 5

6 Wiring Smartec S CLD134 Connection of binary inputs A D1 B D2 15 V S1 S2 Fig. 4: Connection of binary inputs when using external contacts a A B S1 S2 Auxiliary power output Contact inputs D1 and D2 External contacts, not energized External contacts, not energized 3.3 Structure and termination of measuring cable Semiconductor screen Screen (15) GN (11) WH (12) YE (13) (84) (83) Screen (16) WH RD BN (unused) S GN WH YE BU RD RD BU RD WH CLK5 CLK5 J Fig. 5: Structure of the measuting cable a en a Fig. 6: Electrical connection of the CLS54 sensor for the separate version 6 Endress+Hauser

7 Smartec S CLD134 Commissioning 4 Commissioning 4.1 Function check # Warning! Check all connections for correctness. Make sure that the supply voltage is identical to the voltage indicated on the nameplate! 4.2 Key assignment Key CAL E Assignment Use the CAL key to acknowledge calibration data and to continue through the calibration process. The ENTER key has several functions: It calls up the setup menus from the measuring mode. It is used to store (acknowledge) data entered in setup mode. It is used to move on within function groups. + and - In setup mode, the PLUS and MINUS keys have the following functions: Selection of function groups. Setting of parameters and numeric values. In measuring mode, repeatedly pressing the PLUS key displays the following settings in sequence: 1. Temperature display in F. 2. Hide temperature display. 3. Display of uncompensated conductivity value. 4. Back to basic setting. In measuring mode, repeatedly pressing the MINUS key displays the following settings in sequence: 1. Display of current measuring range. 2. Display of current errors in sequence (max. 10). 3. After all errors are displayed, the standard display is shown again. Press the PLUS and the MINUS key simultaneously to perform the escape function. Endress+Hauser 7

8 Commissioning Smartec S CLD Quick setup After switching the transmitter on, configure the major functions required for accurate measurement. The following section gives you an example for a basic configuration. Input 1. Press the ENTER key. 2. Enter the code 22 to be able to edit the setup. Press the ENTER key. 3. Press the MINUS key several times until the "Service" function group is displayed. 4. Press the ENTER key to edit this function group. 5. Select your language, e.g. "ENG" for English. Confirm your entry by pressing the ENTER key. 6. Press the PLUS and MINUS keys simultaneously to quit the "Service" function group. 7. Press the MINUS key several times until the "Setup 1" function group is displayed. 8. Press the ENTER key to edit "Setup 1". 9. In A1, select the operating mode, e.g. "cond" = conductivity. Confirm your selection by pressing the ENTER key. 10. In A2, press the ENTER key to confirm the factory setting. 11. In A3, press the ENTER key to confirm the factory setting. 12. In A4, press the ENTER key to confirm the factory settting. 13. In A5, enter the cell constant for the connected sensor. Refer to the sensor s or the compact version s quality certificate for the exact value. Selection or range (factory setting bold) ENG = English GER = German FRA = French ITA = Italian NEL = Dutch ESP = Spanish cond = conductivity conc = concentration % ppm mg/l TDS = Total Dissolved Solids none XX.xx X.xxx XXX.x XXXX auto, μs/cm, ms/cm, S/cm, μs/m, ms/m, S/m Display SERVICE ENG Language SETUP 1 cond Oper.Mode ppm Conc.Unit XX.xx Format auto Unit l/cm Cellconst 14. In A6, press the ENTER key to confirm the factory setting InstFac 8 Endress+Hauser

9 Smartec S CLD134 Commissioning Input 15. If you are working in applications that fluctuate a great deal and you need to stabilize the display, enter the required damping factor in A7. Confirm your entry by pressing ENTER. The display returns to the inital display of "Setup 1". Selection or range (factory setting bold) Display 1 Damping 16. Press the MINUS key to go to the "Setup 2" function group. 17. Press the ENTER key to edit "Setup 2". SETUP In B1, select the temperature sensor of your conductivity sensor. By default, your measuring system is supplied with the CLS54 sensor with Pt 1000 temperature sensor. Confirm your entry by pressing ENTER. 19. In B2, select the appropriate temperature compensation for your process, e.g. "lin" = linear. Confirm your selection by pressing ENTER. 20. In B3, enter the temperature coefficient α. Confirm your entry by pressing ENTER. 21. The real temperature is displayed in B5. If necessary, calibrate the temperature sensor to an external measurement. Confirm your entry by pressing ENTER. 22. The difference between the measured and the entered temperatures is displayed. Press the ENTER key. The display returns to the initial display of the "Setup 2" function group. 23. Press the MINUS key to go to the "Output" function group. 24. Press the ENTER key to edit the output settings. 25. In O1, select your output, e.g. "out1" = output 1. Confirm your selection by pressing ENTER. 26. In O2, select the linear characteristic. Confirm your selection by pressing ENTER. 27. In O211, select the current range for your output, e.g ma. Confirm your selection by pressing ENTER. 28. In O212, enter the conductivity corresponding to the minium current value at the transmitter output, e.g. 0 μs/cm. Confirm your entry by pressing ENTER. Pt100 Pt1k = Pt 1000 NTC30 fixed none lin = linear NaCl = common salt (IEC 60746) Tab %/K %/K Display and entry of real temperature C 0.0 C C out 1 out 2 lin = linear (1) sim = simulation (2) ma ma 0.00 μs/cm 0.00 μs/cm ms/cm Pt1k ProcTemp. lin TempComp %/K Alpha val 0.0 C RealTemp. 0.0 C TempOffs. OUTPUT out1 Sel. Out lin Sel. Type 4-20 Sel.Range 0 μs/cm 0/4 ma 29. In O213, enter the conductivity corresponding to the maximum current value at the transmitter output, e.g. 930 ms/cm. Confirm your entry by pressing ENTER. The display returns to the initial display of the "Output" function group ms/cm 0.0 μs/cm ms/cm 930 ms/cm 20 ma 30. Press the PLUS and MINUS keys simultaneously to return to measuring mode. Endress+Hauser 9

10 Commissioning Smartec S CLD134! Note! You must perform an air set before installing the sensor. To do so, refer to the chapter "Calibration". 4.4 Calibration To access the "Calibration" function group, press the CAL key (code = 22). Code Field Selection or range Display Info C Function group CALIBRATION CALIBRAT Calibration settings C1(1) Compensation of residual coupling Airs=Airset (1) Cellc=cell constant (2) InstF=Installation factor(3) Remove sensor from the medium and dry completely. AirS Calibrat When commissioning inductive sensors, an airset is mandatory. The calibration of the sensor is to be performed in air. The sensor must be dry. C111 Residual coupling start calibration (airset) current measured value 0.0 μs/cm AirSet Start calibration with CAL. C112 Residual coupling is displayed μs/cm 5.3 μs/cm AirSetVal Residual coupling of measuring system (sensor and transmitter). C113 Calibration status is displayed o.k. E xxx o.k. Status If the calibration status is not o.k., the second display line shows an explanation of the error. C114 Store calibration results yes no new yes Store If C113 = E xxx, then only no or new. If new, return to C. If yes/no, return to Measurement. C1(2) Calibration of cell constant Airs=Airset (1) Cellc=cell constant (2) InstF=installation factor(3) Cellc Calibrat Immerse sensor in calibration solution.! Note! This section decribes the calibration for temperature compensated conductivity. For calibration with uncompensated conductivity, set the temperature coefficient α to 0. The sensor should be immersed at a sufficient distance from the vessel wall (installation factor has no influence if a > 15 mm / 0.59"). C121 Enter process temperature (MTC) 25 C C 25.0 C ProcTemp. Only exists if B1 = fixed C122 Enter α value of calibration solution 2.10 %K %K 2.10 %K alpha val This value is specified in the Technical Information of all E+H calibration solutions. You can also use the printed-on table to calculate the value. Set α to 0 for calibration with uncompensated values. C123 Enter correct conductivity value of calibration solution current measured value ms/cm ms/cm Real. val The display is always in ms/cm. 10 Endress+Hauser

11 Smartec S CLD134 Commissioning Code Field Selection or range Display Info C124 Calculated cell constant is displayed cm /cm Cellconst The calculated cell constant is displayed and entered in A5. C125 Calibration status is displayed o.k. E xxx o.k. Status If the calibration status is not o.k., the second display line shows an explanation of the error. C126 Store calibration results yes no new yes Store If C125 = E xxx, then only no or new. If new, return to C. If yes/no, return to Measurement. C1(3) Calibration with sensor adaptation for inductive sensors The sensor is installed in the process. Airs=airset (1) Cellc= cell constant (2) InstF=installation factor(3) InstF Calibrat Sensor calibration with compensation of wall influence. The distance from the sensor to the pipe wall and the pipe material (conductive or nonconductive) influence the measured value. The installation factor compensates this influence. C131 Enter process temperature 25 C C 25.0 C MTC temp. Only exists if B1 = fixed. C132 Enter α value of calibration solution 2.10 %K %K 2.10 %K alpha value This value is specified in the Technical Information of all E+H calibration solutions. You can also use the printed-on table to calculate the value. Set α to 0 for calibration with uncompensated values. C133 Enter correct conductivity value of calibration solution current measured value ms/cm ms/cm Real val. Determine the correct conductivity value by a reference measurement. C134 Calculated installation factor is displayed InstFact The distance of the sensor to the pipe wall and the pipe material (conductive or nonconductive) influence the measured value. The installation factor compensates this influence. C135 Calibration status is displayed o.k. E xxx o.k. Status If the calibration status is not o.k., the second display line shows an explanation of the error. C136 Store calibration results? yes no new yes Store If C135 = E xxx, then only no or new. If new, return to C. If yes/no, return to Measurement. Endress+Hauser 11

12 KA401C/07/en/07.07 Printed in Germany / FM+SGML

Operating Instructions Smartec CLD134

Operating Instructions Smartec CLD134 BA00401C/07/EN/15.16 71331147 Products Solutions Services Operating Instructions Smartec CLD134 Conductivity measuring system Table of contents Table of contents 1 Document information... 5 1.1 Warnings...

More information

Operating manual. Series 450. Compact thermometer. BA01235O/09/en/

Operating manual. Series 450. Compact thermometer. BA01235O/09/en/ Operating manual Series 450 Compact thermometer BA01235O/09/en/01.13 71220279 2 Table of contents 1 Safety instructions... 3 2 Installation... 4 3 Wiring overview... 7 4 Commissioning... 7 5 Operation...

More information

Operating Instructions Liquisys M CUM223/253

Operating Instructions Liquisys M CUM223/253 BA00200C/07/EN/14.13 71212639 valid as of: software version 2.42 Products Solutions Services Operating Instructions Liquisys M CUM223/253 Transmitter for turbidity and suspended solids About this document

More information

Operating Instructions Smartec CLD18

Operating Instructions Smartec CLD18 BA01149C/07/EN/04.16 71314731 Products Solutions Services Operating Instructions Smartec CLD18 Conductivity measuring system Table of contents Table of contents 1 Document information... 4 1.1 Warnings...

More information

Part 2: Operation. Liquiline M CM42. Operating Instructions. Two-wire transmitter for conductivity (digital sensors)

Part 2: Operation. Liquiline M CM42. Operating Instructions. Two-wire transmitter for conductivity (digital sensors) Operating Instructions Liquiline M CM42 Two-wire transmitter for conductivity (digital sensors) Part 2: Operation BA382C/07/en/POD Valid as of: Software version 13.04.01 Operating concept Fig. 1: Press

More information

DSTHM-2 COMBINED T AND RH DUCT TRANSMITTER. Mounting and operating instructions

DSTHM-2 COMBINED T AND RH DUCT TRANSMITTER. Mounting and operating instructions Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4 STANDARDS 4 OPERATIONAL DIAGRAMS 5 WIRING AND

More information

Technical Information Promonitor NRF560

Technical Information Promonitor NRF560 TI00462G/08/EN/03.14 71264171 Products Solutions Services Technical Information For tank side monitor and control of Proservo NMS5 intelligent tank gauge Application is a monitoring unit for use with Proservo

More information

Instruction Sheet PN 51A-6081C/rev.A June Model 6081-C Wireless Contacting Conductivity Transmitter

Instruction Sheet PN 51A-6081C/rev.A June Model 6081-C Wireless Contacting Conductivity Transmitter Instruction Sheet PN 51A-6081C/rev.A June 2009 Model 6081-C Wireless Contacting Conductivity Transmitter ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures,

More information

Operating instructions. Standstill monitor A / / 2011

Operating instructions. Standstill monitor A / / 2011 Operating instructions Standstill monitor A300 UK 1 2 3 4 5 6 7 8 7390337 / 01 02 / 2011 1 2 3 4 5 6 7 8 switchpoint min max pulse/min power Made in Germany ifm electronic gmbh D 45127 Essen func. I II

More information

Operating instructions AC010S Compact AS-i E-STOP safety module

Operating instructions AC010S Compact AS-i E-STOP safety module Operating instructions AC010S Compact AS-i E-STOP safety module Sachnr. 7390636_/00 05/2007 Contents Safety instructions............................. 3 Installation / Setup............................

More information

English. Operating Manual. Conductive point level switch MLR120. Retain for later reference. Companies / brands of GHM

English. Operating Manual. Conductive point level switch MLR120.  Retain for later reference. Companies / brands of GHM English Operating Manual Conductive point level switch MLR120 Companies / brands of GHM www.ghm-group.de Retain for later reference Table of contents Page 1. Proper use (application areas)... 3 Safety

More information

Technical Information Liquiline To Go CYM290 and Liquiline To Go Ex CYM291

Technical Information Liquiline To Go CYM290 and Liquiline To Go Ex CYM291 TI01198C/07/EN/02.15 71288268 Products Solutions Services Technical Information Liquiline To Go CYM290 and Liquiline To Go Ex CYM291 Portable multiparameter device for Memosens ph, conductivity and oxygen

More information

MANUAL LML-P, LML-S, LML-Ex

MANUAL LML-P, LML-S, LML-Ex PROCESS AUTOMATION MANUAL LML-P, LML-S, LML-Ex Magnetic Immersion Probe for Limit Value Detection ISO9001 With regard to the supply of products, the current issue of the following document is applicable:

More information

PMD70, PMD75, FMD76, FMD77, FMD78

PMD70, PMD75, FMD76, FMD77, FMD78 Brief Operating Instructions Deltabar S PMD70, PMD75, FMD76, FMD77, FMD78 Differential pressure transmitter 8 These are Brief Operating Instructions. For more detailed information, please refer to the

More information

CX3100 Intelligent Conductivity Transmitter

CX3100 Intelligent Conductivity Transmitter CX3100 Intelligent Conductivity Transmitter Operation Manual CX3100 CONTENTS Precautions for Installation Brief Instruction 1 1. Specifications 4 2. Assembly and Installation 2.1 Transmitter Installation

More information

TEMPERIX S Clamp. FAFNIR GmbH Schnackenburgallee 149 c Hamburg Germany

TEMPERIX S Clamp. FAFNIR GmbH Schnackenburgallee 149 c Hamburg Germany Process Automation Edition 2016.09 Technical Changes Reserved 1 TEMPERIX S Clamp 1 General Information These operating instructions contain information necessary for the proper installation and use of

More information

Special Documentation Liquicap M FMI51, FMI52

Special Documentation Liquicap M FMI51, FMI52 SD00198F/00/EN/15.16 71315608 Products Solutions Services Special Documentation Liquicap M FMI51, FMI52 Functional Safety Manual Capacitance level measurement for liquids with a 4-20 ma output signal Table

More information

JUMO ecotrans Lf 01/02

JUMO ecotrans Lf 01/02 JUMO ecotrans Lf 01/02 Transmitter/Switching Device for Conductivity Operating Manual 20273100T90Z002K000 V5.00/EN/00435548 Content 1 Notes... 5 2 Identifying the device version... 6 2.1 Type designation...

More information

Vibration Sensor Type 640

Vibration Sensor Type 640 Vibration Sensor Type 640 Standard Zone-1-21 Zone-2-22 Instruction Manual English Instruction Manual Vibration Sensor Type 640 Standard Zone-1-21 Zone-2-22 Edition: 12.04.10 Attention! Before Start-Up

More information

TMT470L Temperature Transmitter

TMT470L Temperature Transmitter Technical information TMT470L Temperature Transmitter OEM components for compact thermometer/cable transmitter. 2/3 or 4-wire Pt100 can be connected. 4 to 20 ma transmitter, programmable via PC. Application

More information

Operating instructions Safety sensor BNS About this document. Content

Operating instructions Safety sensor BNS About this document. Content afety sensor B 120 1. About this document E Operating instructions.............pages 1 to 6 Original 1.1 Function This operating instructions manual provides all the information you need for the mounting,

More information

English. Operating manual. Universal transmitter UT125. Save for later reference. Companies / brands of GHM

English. Operating manual. Universal transmitter UT125.  Save for later reference. Companies / brands of GHM English Operating manual Universal transmitter UT125 Companies / brands of GHM www.ghm-messtechnik.de Save for later reference. Table of contents page 1. Intended use (areas of application)... 3 1.1 Safety

More information

Manual No Revision MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER

Manual No Revision MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER Manual No. 9835 Revision 0-401 MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER Please read this instruction manual carefully before using the instrument. This manual will provide you with all

More information

How to build a Conductivity Measurement System

How to build a Conductivity Measurement System How to build a Measurement System F B E C D A A Immersion system in open basin B Flow-through system C 1/2 pipe mounted system D 3/4 pipe mounted system E Side-entry system in vessel F Top-entry system

More information

Table of Contents. General Information. Document Sure-Aire Flow Monitoring System. User and Service Manual WARNING CAUTION

Table of Contents. General Information. Document Sure-Aire Flow Monitoring System. User and Service Manual WARNING CAUTION Document 476092 User and Service Manual Installation, Operation and Maintenance Manual Please read and save these instructions for future reference. Read carefully before attempting to assemble, install,

More information

Do not open when energized or when an explosive atmosphere is present. Electrostatic hazard, clean only with damp cloth.

Do not open when energized or when an explosive atmosphere is present. Electrostatic hazard, clean only with damp cloth. D460/D470 Series - ATEX certified position transmitter nnnn II 2 GD ITS 11 ATEX 17431X Ex d IIC T* Gb Tamb -*ºC to +*ºC Ex tb IIIC T* Db Tamb -*ºC to +*ºC IP6X Ambient variation -60 C to +110 C (T4/T130

More information

CDD4 Duct Carbon Dioxide Transmitter

CDD4 Duct Carbon Dioxide Transmitter Drill or punch a 1-1/8 or 1-1/4 hole in the duct at the preferred location and insert the probe into the hole to mark the enclosure mounting holes. Remove the unit and drill the four mounting holes. Clean

More information

CDD Carbon Dioxide Transmitter

CDD Carbon Dioxide Transmitter Introduction The OSA CO2 transmitter uses Infrared Technology to monitor CO2 levels within a range of 0 2000 ppm and outputs a linear 4-20 ma or 0-5/0-10 Vdc signal. The enclosure is designed to operate

More information

BXD17 Series. Features. Analytical instrumentation for your measurement applications.

BXD17 Series. Features. Analytical instrumentation for your measurement applications. BXD17 Series Analytical instrumentation for your measurement applications. Features Single parameter instrument with temperature measurement capabilities Custom IP66 NEMA 4X (144 x 144mm) enclosure Versatile

More information

Electronic Temperature Controller. Instruction Manual Version

Electronic Temperature Controller. Instruction Manual Version Electronic Temperature Controller 701 Instruction Manual Version 1.00.01 Dear Customer, we have made up this operating manual in such a way that all necessary information about the product can be found

More information

Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter. Instruction Sheet

Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter. Instruction Sheet English ph / ISE / COND / RDO / DO Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the

More information

Type Quickstart (from device Version 2)

Type Quickstart (from device Version 2) Type 8022 Flow Transmitter / Pulse divider Durchflusstransmitter / Impulsteiler Transmetteur de débit / Diviseur d impulsions We reserve the right to make technical changes without notice. Technische Änderungen

More information

USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements

USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements www.adwainstruments.com Dear Customer, Thank you for choosing an Adwa product. Please read carefully this manual before starting

More information

Cerabar S. Brief Operating Instructions PMC71, PMP71, PMP75. Process pressure measurement

Cerabar S. Brief Operating Instructions PMC71, PMP71, PMP75. Process pressure measurement Brief Operating Instructions Cerabar S PMC71, PMP71, PMP75 Process pressure measurement These Instructions are Brief Operating Instructions; they are not a substitute for the Operating Instructions pertaining

More information

Operating Manual LF1-LF4 i

Operating Manual LF1-LF4 i Operating Manual Oct. 2005 Fillerkampsweg 1 5 31832 Springe OT Eldagsen Tel.: 05044/887-0 (Fax: -99) E-Mail: info@iotronic.de Internet: http://www.iotronic.de Page 3 Contents 1.0 General Guidelines for

More information

BTL6-A_10-M -A1-S115. Data Sheet. english

BTL6-A_10-M -A1-S115. Data Sheet. english BT6-A_10-M -A1-S11 english Data Sheet Balluff GmbH Schurwaldstrasse 9 76 Neuhausen a.d.f. Germany Phone +49 (0) 71 8/1-0 Fax +49 (0) 71 8/0 10 Servicehotline +49 (0) 71 8/1-70 E-Mail: balluff@balluff.de

More information

C100 Handheld Cond./TDS/Sal. Meter. Instruction Manual

C100 Handheld Cond./TDS/Sal. Meter. Instruction Manual C100 Handheld Cond./TDS/Sal. Meter Instruction Manual Table of Contents 1. Introduction ---------------------------------------------------------------------------------------------------------------------------

More information

Operating Instructions HART. Field Communication with Liquisys M CXM223/253. BA208C/07/en/

Operating Instructions HART. Field Communication with Liquisys M CXM223/253. BA208C/07/en/ Operating Instructions HART Field Communication with Liquisys M CXM223/253 6 BA208C/07/en/0.05 557596 Brief operating instructions This explains how to use these Operating Instructions to commission your

More information

Operating instructions. Switching amplifier DN0210 DN / / 2015

Operating instructions. Switching amplifier DN0210 DN / / 2015 Operating instructions Switching amplifier DN0210 DN0220 UK 80011079 / 00 01 / 2015 Contents 1 Preliminary note...4 1.1 Symbols used...4 1.2 Warning signs used...4 2 Safety instructions...5 2.1 General...5

More information

Instruction Manual DULCOMETER 2201 X Cond

Instruction Manual DULCOMETER 2201 X Cond Instruction Manual DULCOMETER 2201 X Cond ProMinent Dosiertechnik GmbH D-69123 Heidelberg Germany ProMinent Dosiertechnik GmbH Im Schuhmachergewann 5 11 D-69123 Heidelberg Postfach 10 17 60 D-69007 Heidelberg

More information

CONDUCELL 4USF ARC SENSORS. Operating Instructions

CONDUCELL 4USF ARC SENSORS. Operating Instructions CONDUCELL 4USF ARC SENSORS Operating Instructions Important note Copyright 2009 HAMILTON Bonaduz AG, Bonaduz Switzerland. All rights reserved. The reproduction of any part of this document in any form

More information

GRUNDFOS ALLDOS DATA BOOKLET. Conex DIS-C. Measuring amplifiers and controllers for conductivity measurement 50/60 Hz

GRUNDFOS ALLDOS DATA BOOKLET. Conex DIS-C. Measuring amplifiers and controllers for conductivity measurement 50/60 Hz GRUNDFOS ALLDOS DATA BOOKLET Measuring amplifiers and controllers for conductivity measurement 50/60 Hz TM042232 if Contents Product features Conex DIS 3 3 Identification Type key,, controller 4 Type key,,

More information

Analyzer/Controller. ph/orp HART. Model 54e ph/orp ESSENTIAL INSTRUCTIONS WARNINGS RISK OF ELECTRICAL SHOCK

Analyzer/Controller. ph/orp HART. Model 54e ph/orp ESSENTIAL INSTRUCTIONS WARNINGS RISK OF ELECTRICAL SHOCK Instruction Sheet PN 51A-54epH/rev J October 2010 ph/orp HART Model 54e ph/orp Analyzer/Controller For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. ESSENTIAL

More information

TFR5. Room or outdoor temperature sensor. Operating instructions for temperature sensor. English page 1...5

TFR5. Room or outdoor temperature sensor. Operating instructions for temperature sensor. English page 1...5 EN/2018-03-14 Design and specifications subject to change without notice Operating instructions for temperature sensor English page 1...5 Field of application CombiTemp is a temperature sensor, based on

More information

CombiLyz AFI4 / AFI5 Inductive conductivity transmitter

CombiLyz AFI4 / AFI5 Inductive conductivity transmitter Main features Range from 500 µs/cm to 1000 ms/cm ll hygienic design Built in graphical display CombiView DFON Very fast temperature compensation Easy and full programmable with FlexProgrammer 9701 FI5

More information

english S-5EC Potentiometer 5-step Operating Instructions Keep for reference! L-BAL-E265-GB 1608 Index 002

english S-5EC Potentiometer 5-step Operating Instructions Keep for reference! L-BAL-E265-GB 1608 Index 002 S-5EC english Potentiometer 5-step Operating Instructions Keep for reference! Content 1 General notes............................................ 3 1.1 Structure of the operating instructions....................

More information

Operating instructions Electronic pressure sensor. PM205x /01 09/2011

Operating instructions Electronic pressure sensor. PM205x /01 09/2011 Operating instructions Electronic pressure sensor PM205x 706071/01 09/2011 Contents 1 Preliminary note...3 1.1 Symbols used...3 2 Safety instructions...3 3 Functions and features...4 3.1 Applications...4

More information

Transmitter Your Consultant: 02/

Transmitter Your Consultant: 02/ Transmitter 7500 Your Consultant: 02/99 52 120 435 Mettler-Toledo GmbH, Process, Postfach, 8902 Urdorf, Switzerland Phone +41 (01) 736 22 11, Fax +41 (01) 736 26 36 Transmitter 7500 Warranty Defects occurring

More information

WK7200. Water Quality Test Kit USER S MANUAL

WK7200. Water Quality Test Kit USER S MANUAL WK7200 USER S MANUAL Water Quality Test Kit INTRODUCTION: Thank you for selecting model WK7200 microprocessor-based waterproof Water Quality Test Kit. It is used to measure a wide range of ph, ORP, Conductivity,

More information

HPS-M -2 DIFFERENTIAL PRESSURE TRANSMITTER. Mounting and operating instructions

HPS-M -2 DIFFERENTIAL PRESSURE TRANSMITTER. Mounting and operating instructions DIFFERENTIAL PRESSURE Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA 4 STANDARDS 5 OPERATIONAL

More information

itemp TMT180 Technical information

itemp TMT180 Technical information Technical information itemp TMT180 Temperature head transmitter For resistance thermometers Pt100, settable using a PC, for installation in a sensor head Form B Application PC programmable (PCP) Temperature

More information

Original operating instructions Fail-safe inductive sensor GI711S / / 2010

Original operating instructions Fail-safe inductive sensor GI711S / / 2010 Original operating instructions Fail-safe inductive sensor GI7S 704583 / 0 06 / 200 Contents Preliminary note 3. Explanation of symbols 3 2 Safety instructions 4 2. Safety-related requirements regarding

More information

Operating Manual. English. Digital-Analog Converter DACON8-UMB.

Operating Manual. English. Digital-Analog Converter DACON8-UMB. English Digital-Analog Converter www.lufft.de G. Lufft Mess- und Regeltechnik GmbH, Fellbach, Germany. We reserve the right to amend technical specifications at any time. Contents 1 Please Read Before

More information

Operating instructions Safety sensor BNS About this document. Content. 6 Disassembly and disposal 6.1 Disassembly Disposal...

Operating instructions Safety sensor BNS About this document. Content. 6 Disassembly and disposal 6.1 Disassembly Disposal... Safety sensor BNS 0 Disassembly and disposal. Disassembly..... Disposal... 7 EU Declaration of conformity Operating instructions.............pages to Original x.000 / 09.08 / v.a. - 08- / Q / 08-09- /

More information

KROHNE 12/ GR Installation and Operating Instructions BM 500

KROHNE 12/ GR Installation and Operating Instructions BM 500 KROHNE 12/2003 7.02273.21.00 GR Installation and Operating Instructions BM 500 Subject to change without notice. Variable area flowmeters Vortex flowmeters Flow controllers Electromagnetic flowmeters Ultrasonic

More information

Operating manual GTL 142 GTL 152. Temperature probe. Please keep the manual for future use.

Operating manual GTL 142 GTL 152. Temperature probe. Please keep the manual for future use. Operating manual Temperature probe GTL 142 GTL 152 Please keep the manual for future use. L01.0.1X.6C-02 GREISINGER electronic GmbH Hans-Sachs-Str. 26 93128 Regenstauf Germany Fon +49(0)9402-9383-0 Fax

More information

CDD4 Series Room CO2 Transmitter Installation Instructions

CDD4 Series Room CO2 Transmitter Installation Instructions CDD4 Series Room CO2 Transmitter Installation Instructions Introduction The CO2 transmitter uses Infrared Technology to monitor CO2 levels and outputs a linear 4-20 ma or 0-5/0-10 Vdc signal. Options include

More information

CombiLyz AFI4 / AFI5 Inductive conductivity transmitter

CombiLyz AFI4 / AFI5 Inductive conductivity transmitter Main features Range from 500 µs/cm to 1000 S/cm ll hygienic design Built in graphical display CombiView DFON Very fast temperature compensation Easy and full programmable with FlexProgrammer 9701 FI5 split

More information

Liquipoint T FTW31, FTW32

Liquipoint T FTW31, FTW32 TI00375F/00/EN/14.16 71336942 Products Solutions Services Technical Information Liquipoint T FTW31, FTW32 Conductive Point level switch for multiple point detection in conductive liquids Applications The

More information

Solar Combiner Enclosure

Solar Combiner Enclosure Installation Instructions Solar Combiner Enclosure Catalog Numbers 1000-SB006, 1000-SB012 Topic Page Description 1 Important Safety Instructions 3 Nameplate Data 4 Planning for Installation 4 Install the

More information

Operating Manual / IEC. Explosion-Proofed Junction Boxes, Polyester. Wallücker Bahndamm 7 DE Kirchlengern

Operating Manual / IEC. Explosion-Proofed Junction Boxes, Polyester. Wallücker Bahndamm 7 DE Kirchlengern Operating Manual / IEC Explosion-Proofed Junction Boxes, Polyester Wallücker Bahndamm 7 DE- 32278 Kirchlengern Tel.: +49 5223 49107-0 Fax: +49 5223 49107-28 info@multi-box.com www.multi-box.com Contents

More information

Conductivity meter Manual

Conductivity meter Manual Announcements Conductivity meter Manual 1.When using, please comply with the instruction of the operation procedures and the announcements. 2.If find the instrument working abnormally or damaged, please

More information

Operating and Installation Manual. EASYLAB Expansion modules Type EM-TRF for 230 V AC mains voltage

Operating and Installation Manual. EASYLAB Expansion modules Type EM-TRF for 230 V AC mains voltage Operating and Installation Manual EASYLAB Expansion modules Type EM-TRF for 230 V AC mains voltage Type EM-TRF-USV for 230 V AC mains voltage; provides uninterruptible power supply (UPS) The art of handling

More information

Operating instructions. Speed monitor D / / 2014

Operating instructions. Speed monitor D / / 2014 Operating instructions Speed monitor D200 80005257 / 00 05 / 2014 Contents 1 Preliminary note...4 1.1 Symbols used...4 1.2 Warning signs used...4 2 Safety instructions...5 2.1 General...5 2.2 Target group...5

More information

BA01236F/00/EN/ Products Solutions Services. ToF Adapter FXA291

BA01236F/00/EN/ Products Solutions Services. ToF Adapter FXA291 BA01236F/00/EN/01.14 71240651 Products Solutions Services ToF Adapter FXA291 Contents Contents 1 Safety instructions............................................................. 3 1.1 Designated use......................................................................................

More information

IO-PT4. Component identification. User safety and equipment protection guidelines. Unitronics Industrial Automation Systems 1

IO-PT4. Component identification. User safety and equipment protection guidelines. Unitronics Industrial Automation Systems 1 IO-PT4 I/O Expansion Module 4 PT100 Inputs (-50 to 460 C) The IO-PT4 is an I/O expansion module that can be used in conjunction with specific Unitronics OPLC controllers. The module offers 4 PT100 inputs

More information

Differential Liquid/Gas Pressure Transmitter

Differential Liquid/Gas Pressure Transmitter Installation Instruction Differential Liquid/Gas Pressure Transmitter Catalog Number(s) 1414-CPZ10FWFAA, 1414-IPZ10FWFAA Explosion Hazard WARNING Do not use in an explosive or hazardous environment, with

More information

PumpDrive. Self -cooling, motor-independent frequency inverter. Brief Instructions

PumpDrive. Self -cooling, motor-independent frequency inverter. Brief Instructions Brief Instructions PumpDrive 4070.801/2--11 Self -cooling, motor-independent frequency inverter Mounting variants: Motor mounting (MM) Wall mounting (WM) Cabinet mounting (CM) Brief Instructions 1 About

More information

Intrinsically safe batch controller Batching Master 110i

Intrinsically safe batch controller Batching Master 110i Intrinsically safe batch controller Batching Master 110i Installation Guide BVS 04 AT E 172 Revision 12.2 IBS BatchControl GmbH Im Sträßchen 2-4 Tel.: ++49 2441 9199 801 53925 Kall Fax.: ++49 2441 9199

More information

Operating instructions Evaluation system for flow sensors VS / / 2013

Operating instructions Evaluation system for flow sensors VS / / 2013 Operating instructions Evaluation system for flow sensors VS3000 7097 / 0 07 / 203 Contents Preliminary note...2 2 Safety instructions...3 3 Function and features... Mounting.... Mounting of the sensors...

More information

Operating Instructions (Translation) 3. Safety Information. 1. Description. 2. Explosion Protection. Supply module Type 17-21BB-170x

Operating Instructions (Translation) 3. Safety Information. 1. Description. 2. Explosion Protection. Supply module Type 17-21BB-170x 1. Description The supply module was developed specially for direct mounting in hazardous areas in Zone 1 and 21 and is ATEX-certified. The supply module is a permanently installed piece of electrical

More information

HDM97BP HDM97BO HDM97BN

HDM97BP HDM97BO HDM97BN HDM97Pocket Reference Meter Series HDM97BP HDM97BO HDM97BN Safety Instructions and Quick Start Guide This document describes the major functions of the HDM97Pocket instrument series. The complete manual

More information

USER MANUAL AD 3000 EC/TDS/Temperature Bench Meter

USER MANUAL AD 3000 EC/TDS/Temperature Bench Meter USER MANUAL AD 3000 EC/TDS/Temperature Bench Meter www.adwainstruments.com 1 Dear Customer, Thank you for choosing an Adwa product. Please read carefully this manual before starting operations. This instrument

More information

SolConeX extra-low voltage flange socket

SolConeX extra-low voltage flange socket SolConeX extra-low voltage flange socket Operating instructions Additional languages www.stahl-ex.com Contents 1 General Information...3 1.1 Manufacturer...3 1.2 Information regarding the operating instructions...3

More information

Line reactors SINAMICS. SINAMICS G130 Line reactors. Safety information 1. General. Mechanical installation 3. Electrical installation

Line reactors SINAMICS. SINAMICS G130 Line reactors. Safety information 1. General. Mechanical installation 3. Electrical installation Safety information 1 General 2 SINAMICS SINAMICS G130 Mechanical installation 3 Electrical installation 4 Technical specifications 5 Operating Instructions Control version V4.7 04/2014 A5E00331462A Legal

More information

+GF+ SIGNET Conductivity/Resistivity Transmitter Instructions

+GF+ SIGNET Conductivity/Resistivity Transmitter Instructions GF SIGNET 8850 Conductivity/Resistivity Transmitter Instructions ENGLISH 8850.090 A9/99 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to

More information

Signet Conductivity Electrodes

Signet Conductivity Electrodes Signet 2839-2842 Conductivity Electrodes Features ± 1% accuracy - Custom calibration certificate provided Dual-threaded Compact electrode length for easy in-line installation in small pipe sizes Triple

More information

Air-Saver. Version: 0403 INDEX

Air-Saver. Version: 0403 INDEX Air-Saver Version: 0403 INDEX Page 1 Page 2 Page 3 Page 4 Page 5 Page 6 Page 7 General safety instructions Installation instructions Programming instructions Programming instructions and specifications

More information

Operating instructions Safe AS-i input module ASIM-C-M About this document. Content

Operating instructions Safe AS-i input module ASIM-C-M About this document. Content 7 Set-up and maintenance 7.1 Functional testing....10 7.2 Maintenance...10 EN Operating instructions.............pages 1 to 6 Original 8 Disassembly and disposal 8.1 Disassembly....10 8.2 Disposal...10

More information

Operation and Installation Manual

Operation and Installation Manual Operation and Installation Manual SmartLINK M 13 x Content 1 Document Content...1 1.1 Function...1 1.2 Target group: Authorised personnel...1 1.3 Symbols and abbreviations used...1 1.4 Proper use...2 1.5

More information

GRUNDFOS DATA BOOKLET. Conex DIS-C. Measuring amplifiers and controllers. For conductivity measurement (conductive, inductive)

GRUNDFOS DATA BOOKLET. Conex DIS-C. Measuring amplifiers and controllers. For conductivity measurement (conductive, inductive) GRUNDFOS DATA BOOKLET Measuring amplifiers and controllers For conductivity measurement (conductive, inductive) Table of contents 1. Product features 3 Conex DIS 3 Conex DIS-C 3 2. Identification 4 Type

More information

EX-RC1 Remote I/O Adapter

EX-RC1 Remote I/O Adapter EX-RC1 Remote I/O Adapter The EX-RC1 interfaces between Unitronics Vision OPLCs and remote I/O Expansion Modules distributed throughout your system. The adapter is connected to a PLC via CANbus. Each adapter

More information

GRUNDFOS INSTRUCTIONS. Control MPC. Installation and operating instructions

GRUNDFOS INSTRUCTIONS. Control MPC. Installation and operating instructions GRUNDFOS INSTRUCTIONS Control MPC Installation and operating instructions English (GB) English (GB) Installation and operating instructions Original installation and operating instructions CONTENTS Page

More information

Smart Wireless THUM Adapter

Smart Wireless THUM Adapter Reference Manual Smart Wireless THUM Adapter www.rosemount.com Reference Manual Smart Wireless THUM Adapter Smart Wireless THUM Adapter Smart Wireless THUM Adapter Hardware Revision 1 HART Device Revision

More information

Please enter the identity code for your device here!

Please enter the identity code for your device here! Operating Instructions Two-wire measurement transmitter DULCOMETER DMT Measured variable Conductivity DMTa Please enter the identity code for your device here! Please carefully read through these operating

More information

MAXIMA + Series ROTARY LEVEL CONTROL

MAXIMA + Series ROTARY LEVEL CONTROL Price $5.00 MAXIMA + Series ROTARY LEVEL CONTROL OPERATING INSTRUCTIONS PLEASE READ CAREFULLY Division of Garner Industries 7201 North 98th Street Lincoln, NE 68507-9741 (402) 434-9102 925-0268 Rev. A

More information

Conductivity-Panelmeter LF 9648 Conductivity measurement with 2-and 4-electrode cells

Conductivity-Panelmeter LF 9648 Conductivity measurement with 2-and 4-electrode cells Conductivity-Panelmeter LF 9648 Conductivity measurement with 2-and 4-electrode cells Features M LED-Display 14.2 mm red M Indicating range 2000(0) Digit M Measuring ranges programmable from 0... 2.000

More information

JUMO MIDAS C08. OEM Pressure Transmitter Basic. Applications. Brief description. Special features. Customer benefits

JUMO MIDAS C08. OEM Pressure Transmitter Basic. Applications. Brief description. Special features. Customer benefits Data Sheet 4000 Page /0 JUMO MIDAS C08 OEM Pressure Transmitter Basic Applications compressors machinery and plant engineering industrial pneumatics commercial vehicles building technology Brief description

More information

Original operating instructions Fail-safe inductive sensor GF711S / / 2013

Original operating instructions Fail-safe inductive sensor GF711S / / 2013 Original operating instructions Fail-safe inductive sensor GF7S 8528 / 5 / 23 Contents Preliminary note...3. Explanation of symbols...3 2 Safety instructions...4 2. Safety-related requirements regarding

More information

JUMO dtrans CR 02. Transmitter/controller for conductivity, TDS, resistance, temperature and standard signals Type

JUMO dtrans CR 02. Transmitter/controller for conductivity, TDS, resistance, temperature and standard signals Type JUMO dtrans CR 02 Transmitter/controller for conductivity, TDS, resistance, temperature and standard signals Type 202552 B 202552.0 Operating instructions 2011-07-29/00541516 WARNING: A sudden malfunction

More information

EX- Documentation VA 450 / VA 452. thermal mass flow sensor

EX- Documentation VA 450 / VA 452. thermal mass flow sensor EX- Documentation VA 450 / VA 452 thermal mass flow sensor This document contains important information for the installation and operation of the flow sensor in explosive environments. Please read the

More information

Operating Instructions

Operating Instructions Introduction To protect the signal lines of field devices and systems in the cabinet against lightning. Pepperl+Fuchs covers the complete range of Surge Protection Barriers. Housing types Depending on

More information

ADC7520 SERIES. 1600W Battery Chargers and Power Supplies

ADC7520 SERIES. 1600W Battery Chargers and Power Supplies ADC7520 SERIES 1600W Battery Chargers and Power Supplies Wide output adjustment range 0 72VDC Analog control by external 0-5VDC voltage Temp.comp charging, sense as on option Power fail relay alarm Master-Slave

More information

STR-3 STANDARD 3-PHASE 230 VAC TRANSFORMER CONTROLLER. Mounting and operating instructions

STR-3 STANDARD 3-PHASE 230 VAC TRANSFORMER CONTROLLER. Mounting and operating instructions STANDARD 3-PHASE 230 VAC TRANSFORMER CONTROLLER Mounting and operating instructions Table of contents SAFETY AND PRECAUTIONS 3 PRODUCT DESCRIPTION 4 ARTICLE CODES 4 INTENDED AREA OF USE 4 TECHNICAL DATA

More information

Liquisys M CUM 223 / 253 Transmitter for Turbidity and Solids Content

Liquisys M CUM 223 / 253 Transmitter for Turbidity and Solids Content BA 200C/07/en/07.02 51500274 Valid from software version: 2.40 Liquisys M CUM 223 / 253 Transmitter for Turbidity and Solids Content Operating Instructions ENDRESS+HAUSER LIQUISYS S ENDRESS+HAUSER LIQUISYS

More information

Technical Information Memobase Plus CYZ71D

Technical Information Memobase Plus CYZ71D TI00502C/07/EN/02.13 71228167 Products Solutions Services Technical Information Memobase Plus CYZ71D Multichannel multiparameter software for liquid analysis with digital Memosens sensors Measure - Calibrate

More information

MDM 011-Z1 Regen Resistor

MDM 011-Z1 Regen Resistor MDM 011-Z1 Regen Resistor Date of creation: 10.04.2017 Version date: 10.04.2017 Article number: 09-402-011-Z1-E Publisher: SIGMATEK GmbH & Co KG A-5112 Lamprechtshausen Tel.: 06274/4321 Fax: 06274/4321-18

More information

L Series Humidity/-temperature sensors for use in heating, ventilation and air conditioning Technical data Humidity

L Series Humidity/-temperature sensors for use in heating, ventilation and air conditioning Technical data Humidity Galltec Mess- und Regeltechnik GmbH MELA Sensortechnik GmbH Boschstr. 4 Raasdorfer Str. 18 71145 Bondorf. Germany 07987 Mohlsdorf-Teichwolframsdorf. Germany Tel. +49 (0)7457-9453-0 Tel. +49(0)3661-62704-0

More information

Operating instructions Vortex flow meter. SVxxx / / 2015

Operating instructions Vortex flow meter. SVxxx / / 2015 Operating instructions Vortex flow meter SVxxx4 80226711 / 00 06 / 2015 Contents 1 Preliminary note...3 1.1 Symbols used...3 1.2 Warning signs used...3 2 Safety instructions...3 3 Functions and features...4

More information

Operating instructions Door handle switch BNS-B About this document. Content. 6 Disassembly and disposal 6.1 Disassembly Disposal...

Operating instructions Door handle switch BNS-B About this document. Content. 6 Disassembly and disposal 6.1 Disassembly Disposal... 6 Disassembly and disposal 6.1 Disassembly....6 6.2 Disposal...6 7 EU Declaration of conformity Operating instructions.............pages 1 to 8 Original x.000 / 09.2017 / v.a. - 101178533- / I / 2017-08-31

More information