Analysis of Combined Probing Measurement Error and Length Measurement Error Test for Acceptance Testing in Dimensional Computed Tomography

Size: px
Start display at page:

Download "Analysis of Combined Probing Measurement Error and Length Measurement Error Test for Acceptance Testing in Dimensional Computed Tomography"

Transcription

1 Digital Industrial Radiology and Computed Tomography (DIR 2015) June 2015, Belgium, Ghent - More Info at Open Access Database Analysis of Combined Probing Measurement Error and Length Measurement Error Test for Acceptance Testing in Dimensional Computed Tomography Fabrício BORGES DE OLIVEIRA, Markus BARTSCHER, Ulrich NEUSCHAEFER-RUBE Coordinate Metrology Department, Physikalisch-Technische Bundesanstalt (PTB) Braunschweig, Germany, Phone: , Fax: , fabricio.borges@ptb.de, markus.bartscher@ptb.de, ulrich.neuschaefer-rube@ptb.de Abstract Industrial X-ray Computed Tomography (CT) has evolved as an alternative measurement method to tactile and optical coordinate measurement systems (CMSs). Acceptance testing e.g. according to ISO standards is a common method to assess locally and globally CMS errors as probing measurement errors P and length measurement errors E, respectively. This paper explores in an experimental approach aspects of performing combined, i.e. simultaneous, P and E acceptance tests in the field of dimensional CT. This combined test approach was proposed in ISO standardization in early Prospective benefits of combined P/E testing are the potential reduction of time and costs. However, the combined P/E test is a completely new concept in the field of coordinate metrology and needs to be analyzed for its metrological characteristics. Thus, this work performed within the framework the European Marie Curie project INTERAQCT discusses this new approach on the basis both of real CT scans and simulations and tries to create input for the international standardization of dimensional CT. Keywords: Standardization, Marie Curie project INTERAQCT, Acceptance testing, combined P/E test, dimensional computed tomography (CT), length measurement error E, probing error P 1. Introduction In the last decade, industrial X-ray Computed Tomography (CT) has evolved as an alternative measurement method to classical (i.e., tactile and optical) Coordinate Measurement Systems (CMSs). Consequently, to further increase the trust in CT measurements, both manufacturer and users of industrial CT are highly interested in standardized tests for these systems. Acceptance testing is part of the main scope of the well-established international ISO series of standards, up to now focused on classical CMSs. A common approach to assess local and global CMS errors is performing probing measurement error P and length measurement error E testing, respectively. This paper explores in an experimental approach advantages and disadvantages of performing combined, i.e. simultaneous, P and E acceptance tests in the field of dimensional CT. This combined test approach was proposed in ISO standardization for CT (CT task force of ISO technical committee (TC) 213 working group (WG) 10 beginning of Two important prospective advantages of combined P/E testing are the potential reduction of time and costs. However, the common P/E test is a completely new concept in the field of dimensional metrology, i.e. in coordinate metrology, and needs to be analyzed for its metrological characteristics. Thus, the paper discusses this new approach and tries to create input for the international standardization of dimensional CT. Therefore, this paper starts with a general description of acceptance testing for CMS, describing its principles. Then, the paper describes the reference standards used for the new proposal of combined P and E test as well as an experimental setup for both CT scans and simulations. In the following section, a description of the evaluation procedure including a new patch-based approach for CT measurements is given. Finally, results and conclusions are discussed closing the paper.

2 2. Acceptance testing of dimensional CT The main principle of acceptance testing, according to ISO series of standards, is to perform an overall test of the entire performance of the CMS. Mainly, the test should create trust upon the technology by means of helping to achieve traceability to the SI-unit metre and enabling comparability with other CMSs [1]. Therefore, the test is to be performed as an integrated system and it should evaluate the system using every process step of the measurement chain (i.e. as a black box). Also, it should be economic and feature satisfactory reliability (satisfactory test value uncertainty [2]). However, the acceptance test by itself does not provide complete traceability for general measurements of industrial parts. For this purpose a statement regarding the measurement uncertainty is required, which is by intent not covered by the acceptance testing procedure. A last important objective of the acceptance testing and the underlying specifications (which are tested using the acceptance test procedure) is to provide comparability between different CMSs as well for CMSs of the same type and sensing principle and as well to CMSs which employ different measurement and sensing technologies. The acceptance test shall reflect the standard use of the CMS and shall cover all dominant error behaviour of the CMS under study. Therefore, real-life effects and e.g. the request for simple geometry reference standards in the test design shall be taken into account. Besides, the difficulty to achieve comparability with other CMS when using real-life specimen and due to its complexity and variety, the use of real-life specimen in the scope of acceptance testing is limited. Another important principle of acceptance testing is to assess both global and local performance i.e. error characteristics of CMSs. The local error behaviour is tested as a probing error test (P-test) assessing the diameter and form deviation of a calibrated sphere (Figure 1). The global system behaviour is tested as a length measurement error (E-test) using reference standards for lengths (e.g. hole plates, step gauges, ball plates, etc.). The hole plate reference standard has shown to be an useful tool to evaluate the E-test of CT systems, due to its high number of possible features and the different radiation absorption lengths along the hole plate. Moreover, the hole plate standard is in accordance with ISO concepts, due to the fact that it features 5 independent lengths in 7 different directions. Besides, it has economic design and satisfactory test value uncertainty can be achieved. However, a recent test study in ISO TC 213 WG10 in 2013 and 2014 has shown that there is material influence present [3]. Length measurements can be evaluated as bi- and/or unidirectional measurements, see Figure 1. Recent resolutions of ISO TC 213 WG 10 state that bi-directional measurements are mandatory for the acceptance testing. On the other hand, uni-directional measurements are optional [3]. Either uni- or bidirectional measurements can be evaluated using patches to create a representative point. This approach which entered into ISO methodology with ISO for optical distances sensors [4] enables more stable results when measuring length by reducing the influence of the sensor noise as well as the patch approach improves comparability between CMSs with different sensor technologies, mainly due to the high density of points obtained by CT and to the morphological filtering being present in tactile probing. On the other hand, the use of patches has intrinsic low pass filtering properties and can hide local effects of the system, which might be relevant for the CMS user/tester. However, in the current international standard for optical distance sensors unsolved issues exist, e.g., the patch geometry is described in a non satisfying way. For CT-based CMS technology, acceptance testing is relatively new. The first national guideline was released in 2011, when the German engineers association VDI/VDE published

3 the first framework for acceptance testing for CT the guideline VDI/VDE [5]. The guideline VDE/VDI tries to map approaches of ISO to CT [6]. The work in the series VDI/VDE 2630 has started in In the international standardization level, the acceptance testing framework for CT started later. In 2010, the International Standards Organization (ISO) working group ISO TC 213 WG10 created the technical task force committee committed to develop acceptance testing for CT creating in future ISO The current status of ISO standardization work for CT can be seen in more details in [1]. In the context of ISO based acceptance testing for CT, a combined probing and length test (i.e. simultaneous, P and E acceptance tests) was proposed in This paper explores in an experimental approach advantages and disadvantages of performing such a combined P and E acceptance test in the field of dimensional CT as well as the impact of patches for CT-based length measurements. Figure 1. global behaviour assessment E-test tested as uni- and bidirectional length measurements (here for the case of cylindrical geometry); Local behaviour assessment P-test tested as probing dispersion using 95% of the points and probing size error using all point, 25 representative points based on 25 patches and all probed points (here for the case of spherical geometry) 3. Combined P/E testing Both experimental CT scans and CT simulations for combined P/E were performed. In the experimental CT scans, a calibrated hole plate and a calibrated sphere were used as reference standards for studying the combined P/E test. Additionally, simulations of CT scans using the BAM software artist [8] were performed where the simulation parameters are comparable to the performed CT scans (e.g. material composition, shape, dimensions and setup parameters). Both CT scans and simulations include the case of independent P and E tests, too. Length measurement- and probing measurement-based analyses were carried out for both CT scans and simulations. 3.1 Reference standards A calibrated hole plate machined by erosion in aluminium (AlSiMgMn/ / EN AW- 6082) and a calibrated silicon nitride sphere were used to investigate whether an impact on P- test and/or E-test when performing combined P/E test exists. The hole plate a has size of 48 mm 48 mm 8 mm and features 28 holes (4 mm diameter and cylindricity less than 5 µm) designed in such way that it features 5 independent lengths in seven main different directions (Figure 2 (a)), in accordance with ISO :2009 concepts [6]. The design of

4 this plate was developed by NMIJ/AIST and PTB, and is described in more details in [1]. Moreover, a silicon nitride sphere with a carbon fibre shaft (i.e., classical CMS tactile probe) was used, see Figure 2 (b). It features 8 mm in diameter and a sphericity of less than 0.3 µm, suitable for CT probing test analyses. Reference measurements of the hole plate and sphere were performed using tactile CMSs and a special full diamond probe of 2 mm in diameter. The diamond probe was used to prevent pollution effects during the hole plate measurements. These effects easily can impair the quality of tactile CMS measurements, especially when using ruby probes and when performing scanning measurements or the massive multiple single-point particles remain on the probe and disturb tactile probing on aluminium surfaces (aluminium measurements). In the reference measurements of the hole plate, several scanning circumferential lines in different heights were probed within the holes (Figure 3). This calibration approach allows flexibility on the measurands, i.e., it enables the use of single-point 1, multi-point 2, patches 3, etc. Therefore, comparable measurements between the reference measurements and CT measurements can be achieved. (a) (b) Figure 2. (a): Aluminium hole plate 48 mm x 48 mm x 8 mm, designed by NMIJ/AIST and PTB [1], (b): Silicon nitride tactile CMS probe; image shows characterisation of the dark silicon nitride probe using a (red) standard ruby probe Figure 3. Reference measurements scheme of the hole plate using circumferential lines 1 Single-point: use of single point to measure distance 2 Multi-point: use of several points to obtain a element, e.g. a circle to perform centre to centre measurements 3 Patches: use of several points in region of small spatial size to create a representative point

5 3.2 Experimental setup CT scans For the CT scans, the hole plate and the sphere were scanned in the same mounting setup, i.e. combined setup (Figure 4). Also scans of the hole plate and the sphere were performed independently. For the combined setup, the hole plate was place tilted by 45º, slightly off centred to the axis of the rotary table according to the ISO proposal of the combined P/E test. Scans of three different sphere to plate distances were carried out. Distances of 0.5 mm, 2.9 mm and 5.7 mm were used, which corresponds approximately to 11 times, 69 times and 135 times the voxel size of the measurement, respectively. Due to the mechanical mounting, a lateral displacement of the sphere relative to the hole plate occurred, too. Magnification factor of 4.7 times (corresponding to a voxel size of 42 µm) ) and a 2D grid-like scaling correction has been performed to improve the accuracy of CT measurements [7]. The applied scaling correction excludes an additiona error which is otherwise overlaid on the data. Futhermore, soft beam hardening correction were applied for the combined setup as well as for the independent hole plate measurements. However, for the independent sphere measurements, no beam hardening correction was applied, due to the material composition and dimensions of the sphere. The beam hardening effects were verified by plotting the gray value profile of the reconstructed volume of the sphere measurement. The CT scanning parameters can be found in the Table 1. The same CT parameters were used for all measurements. Each CT scan lasts approximately 3 h 40 min. Figure 4. Hole plate and silicon nitride sphere (as part of a tactile CMS probe with a carbon fibre shaft) mounting setup for the combined P/E test Voltage Current Exposure in kv in µa time in s Table 1. CT setup parameters for the combined P/E test Frames/ Filter Cu Number of projection in mm projections Voxel size in µm 42 CT simulation CT simulations of the combined P/E test were performed using the software artist (version 2.4.0) of the Bundesanstalt für Materialforschung und prüfung (BAM) in Berlin [8], see Figure 5 (a). This software has been constantly developed and extended in the last decade and is in its recent versions capable not only to perform simulation of radiographies (Figure 5 (b)), but also to perform simulation of entire CT scans. Most of the physical effects present in CT scanning can be modelled in this software.

6 The simulated measurements were carried out for different relative positions between hole plate and sphere, where the beam path concerning the material penetration length was changed, see Figure 6, i.e. (a) the sphere was hidden by the hole plate in a high number of projections (i.e. the beam had maximum material thickness to penetrate); (b) the sphere was hidden by the hole plate in a high number of projections, however lower material thickness due to the holes in the beam path hitting the sphere; and (c) the sphere was always visible to the beam. Comparable parameters as described in the CT scans section were used for the CT simulation (e.g. material composition, shape, dimensions and setup parameters). However, there are some physical effects which were not simulated in this study, e.g. beam hardening correction and mechanical misalignments; these parameters explain differences between real scans and simulations to some extent. (a) (b) Figure 5. Simulated combined mounting setup (a), graphical interface of simulated radiography using software artist (b) Figure 6. Simulation experimental setups for combined P/E test; (a) sphere beam path with a big amount of material to penetrate; (b) sphere beam path with a less amount of material to penetrate, due to the holes; and (c) sphere beam path hole plate free. 3.3 Evaluation The evaluation of the acceptance test is based on absolute (i.e. compared with reference measurements) length measurement and probing measurement-based analyses. In order to have CT data comparable with tactile CMS measurements, the same data points obtained by tactile CMS reference measurements were fitted in the CT volume, after performing a surface determination step using the commercial CT analysis software VG Studio Max There are two possibilities of performing surface determination in such combined scenario, e.g. using the same threshold starting value for both length and probing evaluations or performing different threshold starting values focused on each material (i.e. aluminium and silicon nitride). Preliminary tests have been carried out and have shown that

7 the surface determination step plays a critical role in the combined P/E test. The test is intended to be mono-material, but it is multi-material, indeed. However, in this paper as a starting point the same surface determination was used for both length measurement- and probing measurement-based analyses. Bidirectional length measurements were carried out based on single- or multi-point (i.e. patches) and unidirectional multi-point measurements (i.e. centre to centre measurements). All the length-based analyses in the hole plate were carried out for the longest distance in each direction, see Figure 7. Figure 7. Longest distances on each direction of the hole plate E-test At the middle plane of the hole plate, least-squares circles with unconstrained radii and positions were fitted for the unidirectional centre to centre measurements. For the bidirectional single-point measurements, least-squares cylinders with unconstrained radii and positions were fitted, combination and intersection between two cylinders leads to the position where the single-point shall be taken. Additionally, bidirectional multi-point measurements (i.e. patches, see Figure 8) were performed. Points within a sphere-shaped region with radii of 35 µm, 70 µm, 100 µm, 200 µm and 1000 µm, respectively, were used to calculate a patch-based Chebyshev best-fitting cylinder (this fitting element is usually applied to create the middle symmetry cylinder for a minimum zone cylinder based form analysis MZC). The centre of gravity of the patch points and the centre of the patch-based cylinder provide interim points for creating the representative point. The two interim points define a line which intersects the Chebyshev best-fitting cylinder. This intersection point defines finally the representative point. Chebyshev best fitting was used in this approach due to its flexibility. Tests were carried out for the usual least-square (LS) fitting, however, for smaller patches, Chebyshev best fitting presented better and more stable results than least-square fitting. Additionally, the patch procedure is in conformance with the new ISO directive issued in early 2015: No information from other patches shall be used when extracting the representative point of a patch. This measurement procedure was performed using scripting mode in the commercial coordinate metrology inpsection software GOM Inspect Professional V7.5 SR2.

8 Figure 8. Description of the applied patch analyses for bidirectional multi-point length measurements Probing error analyses were performed for form and size of the test sphere. Form analysis is evaluated as probing dispersion value (P Form.Sph.D95%::CT ) and probing form error (P Form.Sph.1x25::CT ). For size analyses, probing size error all (P Size.Sph.All::CT ); and probing size error (P Size.Sph.1x25::CT ) [9],[1] are evaluated. For an explanation of notation see e.g. [10] and [1]. The probing dispersion value is determined as the smallest width of a spherical shell that includes 95% of all data points. Probing size error all is measured as the difference of the diameter of a unconstrained least-squared fit of all points measured on a sphere and its calibrated diameter. Probing form error is the error of indication within the range of the Gaussian radial distance can be determined by a least-square fit of 25 representative points on a test sphere. Probing size error is the error of indication of the difference between the diameter of a least-square fit of 25 representative points on a test sphere and its calibrated diameter. These 25 representative points shall provide comparability between CT and classical tactile CMS, due to the noise present in CT and to the morphological filtering present in tactile probing. However, a well defined procedure on how to assess these characteristics in CT is not finally established yet and is required to create an accepted testing scheme. The following 8 steps method was adopted in this contribution for probing form error and probing size error evaluation (Figure 9): 1. Fit unconstrained least-squares (LS) sphere to all data points 2. Select 25 patch centres on this LS sphere (e.g. ISO pattern [11]) 3. Construct 25 cones symmetric to the selected patch centres starting the cone apex from main LS sphere centre 4. Intersect all data points point cloud with each cone resulting in 25 sub-point clouds 5. Construct 25 LS spheres to the 25 sub-point clouds created before 6. Construct 25 lines through sphere centres created in step 5. and respective patch points centre of gravity 7. Intersect the 25 lines with the respective 25 sub LS spheres resulting in 25 points 8. Analyse the 25 points for form and diameter using unconstrained LS fitting

9 Figure 9. Description of the applied analyses of probing form error and probing size error All sphere fitted points were exported as ASCII files by the measurement software (VGStudio MAX 2.2.6) and the following P-test analyses were carried out with a MATLAB dedicated application, developed at PTB [12]. 4. Results CT scans were obtained with different sphere to hole plate distances and lateral positions (top part of Figure 10), for multi-point unidirectional centre to centre length, bidirectional external single-point and multi-point (i.e., patches) length measurements. The results of dimensional and geometrical evaluations allows to investigate whether there is an influence on the combined P/E acceptance test presented in Figure 10 (a), (b) and (c). Figure 10. CT Scan results: sphere to plate distance 0.5 mm; 2.9 mm; 5.7 mm; (a) unidirectional multipoint centre to centre length measurement error results; (b) bidirectional single points based length measurement error results; and (c) results of probing error analyses for the four probing error characteristics (cf. main text)

10 In general, no significant variations were observed for the unidirectional centre to centre measurements (Figure 10 (a)) when changing the distance of sphere relative to the plate. However, significant measurement instabilities can be observed for bidirectional single-point based length measurements (Figure 10 (b)). Indeed, a significant influence was observed on the length measurements depending on the lateral position of the sphere relative to the hole plate, mainly concerning to the X-rays path, see Figure 10 (sphere to plate 0.5 mm, 2.9 mm and 5.7 mm). However, it was expected to obtain smaller measurement errors for the larger sphere to plate distances. This effect did not occurred; see distance 1-28 in Figure 10 (b). Furthermore, distance 1-28 shows the biggest influence for the presence of the sphere as the sphere is next to hole 28. But in general there is no significant influence in the length measurements regarding the sphere to plate distance for those distances used in this study, as the distances used were not sufficient large to have a much lower number of projections with the beam hole plate free. The same is true for the probing error analysis (Figure 10 (c)). This effect was confirmed by simulation data. The simulation setup, Position 1, Position 2 and Position 3, can be seen in Figure 6 and the simulation results are presented in Figure 11. However, for probing analyses in Position 3 (Figure 6 (c)) where the sphere is far away from the central plane a strong pole artefact due to the cone beam geometry was observed (Figure 11 (c) and (d)). This effect can frequently be observed with many cone beam CT systems when a spherical reference standard is measured far away from the central plane of the detector [1]. Differences between the results obtained by CT scans and simulation are explained by the missing effects (e.g., beam hardening correction, mechanical misalignments) on CT simulations. Figure 11. Simulation results of the setups described in Figure 6: (a) unidirectional multi-point centre to centre measurements results; (b) single-point based bidirectional measurements results; (c) results of probing analyses; and (d) observed pole artefact

11 Figure 12 presents the patch analyses of CT scans. Figure 12 (a) shows the mean value of three repeated measurements, without repositioning the reference standard, for each distance. In general, an improvement was achieved concerning the stability of the results for most of the bidirectional external measured lengths. The improvement can be seen by the decrease of the variation (i.e., maximum value minimum value of three repeated measurements) while increasing the patch size (Figure 12 (b)). Figure 12. CT scans Mean value of the external bidirectional length measurement error - patch analyses for three repeated measurements without repositioning the reference standards (a); and variation (Max - Min) of 3 repeated measurements for each distances (b) 5. Conclusions and Summary (a) CT scans and CT simulations were carried out for the new proposal for common P/E acceptance testing for CT and it was analyzed for its metrological characteristics. Also, for the length measurement error analysis a new concept of a patch operator for CT data was presented and analyzed in this paper. Probing- and length-based evaluations were carried out for different distances and lateral positions of the sphere relative to the hole plate. In general, measurement errors could be observed for both length measurement and probing error. For the length measurements a high dependence on the lateral position of the sphere relative to the hole plate was observed, mainly concerning to the X-rays path. However, no significant influence was observed in the length measurements regarding the sphere to plate distance. The same is true for the probing error analysis. This effect was confirmed by simulation data. Regarding the patch analyses, considerable improvement was achieved concerning the stability of the results for most of the bidirectional external measured lengths while applying patches instead of single-point based length measurements. A critical point of the common setup is that the test is intended to be mono-material but it is in reality multi-material. Thus, surface determination plays an important role when performing such a combined P/E testing. Another critical point observed in the tests is the mechanical mounting/stability. The sphere should not be placed in a position where a lot of material in the beam path exists. Finally, common P/E testing appears to be a nice idea from the first look. But it features several intrinsic problems. First of all it is a new concept which changes (increases) the characteristic numbers to be reported. Thus, it is unfair to one (new) measurement technology to urge only this technology (e.g.) CT to state numbers according to this new concept as measurement technologies are in competition to each other and need to be comparable. Secondly, performing this combined setup appears to be delicate for CT. Testing conditions (b)

12 strongly depend from the way the test is performed. This is in general not acceptable for a standard test recommended for CT in future ISO Acknowledgements We would like to thank the EU Marie Curie Initial Training Networks (ITN) INTERAQCT, Grant agreement no.: , beneficiary no.: 7 for the project funding, more information can be seen on http//: Also we want to thank our colleagues Jakob Schlie (for the probing analyses application), Jens Illemann (for the discussion and scaling correction application and procedure), Konrad Hierse (for tactile measurement of the Si 3 N 4 sphere), Michael Neugebauer (for the probe selection discussion and for the assistance with the mechanical mounting) and Uwe Langner (for tactile CMS measurement of the hole plate). References 1. Bartscher, M., Sato, O., Härtig, F., & Neuschaefer-Rube, U. (2014). Current state of standardization in the field of dimensional computed tomography. Measurement Science and Technology, 25(6), ISO/TS 23165:2006. Geometrical product specifications (GPS) - Guidelines for the evaluation of coordinate measuring machine (CMM) test uncertainty. International Organization for Standardization, Geneva. 3. ISO 2014 Resolutions adopted at the 37 th Meeting of ISO TC 213 WG10, /13/14 (Stockholm, Sweden) Document ISO TC 213 WG 10 N ISO :2013 Geometrical product specifications (GPS) - Acceptance and reverification tests for coordinate measuring systems (CMS) Part 8: CMMs with optical distance sensors. International Organization for Standardization, Geneva. 5. VDI/VDE 2617/ Part 13/Part 1.3 (2011). Guideline for the application of DIN EN ISO for Coordinate Measurement Machines with CT sensors. VDI, Duesseldorf, ISO : Geometrical Product Specifications (GPS) - Acceptance and reverification tests for coordinate measuring machines (CMM) - Part 2: CMMs used for measuring linear dimensions. International Organization for Standardization, Geneva. 7. J Illemann, M Bartscher, U Neuschaefer-Rube, An efficient procedure for traceable dimensional measurements and the characterization of industrial CT systems, Digital Industrial Radiology and Computed Tomography (DIR 2015), Ghent, Belgium, June, C Bellon und G.-R Jaenisch, artist analytical RT inspection simulation tool, Proceedings of DIR, S , 2007 and http//: 9. Draft document ISO TC 213 WG 10 N1051, Geometrical Product Specifications (GPS) Acceptance and reverification tests for coordinate measuring machines (CMM) Part CT: CT, ISO 2012 Symbol language for the ISO series Generation III (G3), Rev. 4.3, Document ISO TC 2013 WG 10 N ISO :2010 Geometrical product specifications (GPS) - Acceptance and reverification tests for coordinate measuring machines (CMM) Part 5: CMMs using single and multiple stylus contacting probing. International Organization for Standardization, Geneva. 12. ISO 2014, DIN Deutsches Institut für Normung, Presentation: Welcome to TF CT meeting (Stockholm, Sweden), Document ISO TC 2013 WG 10 N1026.

Creating a Multi-material Probing Error Test for the Acceptance Testing of Dimensional Computed Tomography Systems

Creating a Multi-material Probing Error Test for the Acceptance Testing of Dimensional Computed Tomography Systems Creating a Multi-material Probing Error Test for the Acceptance Testing of Dimensional Computed Tomography Systems More info about this article: http://www.ndt.net/?id=20848 Fabricio Borges de Oliveira

More information

Creating a Multi-Material Length Measurement Error Test for the Acceptance Testing of Dimensional Computed Tomography Systems

Creating a Multi-Material Length Measurement Error Test for the Acceptance Testing of Dimensional Computed Tomography Systems Creating a Multi-Material Length Measurement Error Test for the Acceptance Testing of Dimensional Computed Tomography Systems More info about this article: http://www.ndt.net/?id=20847 Fabrício Borges

More information

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Wenjuan Sun and Stephen Brown 10 th May 2016 1 Possible factors influencing XCT measurements Components Influencing variables Possible

More information

How beam hardening influences the dimensional geometry in X-ray computed tomography

How beam hardening influences the dimensional geometry in X-ray computed tomography How beam hardening influences the dimensional geometry in X-ray computed tomography More info about this article: http://www.ndt.net/?id=20809 Hannah Corcoran 1, 2, Stephen Brown 2, Stuart Robson 1, Robert

More information

Case Studies in Nondestructive Testing and Evaluation

Case Studies in Nondestructive Testing and Evaluation Case Studies in Nondestructive Testing and Evaluation 6 (2016) 79 92 More info about this article: http://www.ndt.net/?id=20771 Contents lists available at ScienceDirect Case Studies in Nondestructive

More information

Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system

Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system ECNDT 2014 Prague October 6-10, 2014 Dr. Eberhard Neuser Dr. Alexander Suppes Imagination

More information

Experimental investigation of new multi-material gap reference standard for testing computed tomography systems

Experimental investigation of new multi-material gap reference standard for testing computed tomography systems More info about this article: http://www.ndt.net/?id=20837 Experimental investigation of new multi-material gap reference standard for testing computed tomography systems Petr Hermanek 1, Fabrício Borges

More information

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling nd International Symposium on NDT in Aerospace - We..A. Comparison of Probing Error in Dimensional Measurement by Means of D Computed Tomography with Circular and Helical Sampling Jochen HILLER, Stefan

More information

Coordinate Measuring Machines with Computed Tomography

Coordinate Measuring Machines with Computed Tomography Always a Step Ahead with Quality Coordinate Measuring Machines with Computed Tomography Multisensor Coordinate Measuring Machines with Computed Tomography Computed Tomography in Coordinate Measuring Machines

More information

THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS

THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS A. Campos a*, A. Zorrilla a, F. Lasagni a a Materials & Processes dept., Center for Advanced Aerospace Technologies

More information

Wieblinger Weg 92a, Heidelberg, Germany, Phone: , Fax: ;

Wieblinger Weg 92a, Heidelberg, Germany, Phone: , Fax: ; HOW INDUSTRIAL COMPUTER TOMOGRAPHY ACCELERATES PRODUCT DEVELOPMENT IN THE LIGHT METAL CASTING AND INJECTION MOULDING INDUSTRY. Christof REINHART 1, Christoph POLIWODA 1, Thomas GÜNTHER 1 1 Volume Graphics

More information

Parameter Dependent Thresholding for Dimensional X-ray Computed Tomography

Parameter Dependent Thresholding for Dimensional X-ray Computed Tomography International Symposium on Digital Industrial Radiology and Computed Tomography - Poster 12 Parameter Dependent Thresholding for Dimensional X-ray Computed Tomography Kim KIEKENS *, Ye TAN *, Jean-Pierre

More information

Experimental accuracy assessment of different measuring sensors on workpieces with varying properties

Experimental accuracy assessment of different measuring sensors on workpieces with varying properties Experimental accuracy assessment of different measuring sensors on workpieces with varying properties Rauf Oezden 1,*, Metin Aclan 1 and Saliba Danho 2 1 Technical University of Cluj-Napoca, Ph.D. Stud.,

More information

Investigation on the influence of image quality in X-ray CT metrology

Investigation on the influence of image quality in X-ray CT metrology Investigation on the influence of image quality in X-ray CT metrology Pavel Müller 1, Jochen Hiller 1, Angela Cantatore 1, Markus Bartscher 2, Leonardo De Chiffre 1 1 Technical University of Denmark, Department

More information

Frequency-based method to optimize the number of projections for industrial computed tomography

Frequency-based method to optimize the number of projections for industrial computed tomography More Info at Open Access Database www.ndt.net/?id=18739 Frequency-based method to optimize the number of projections for industrial computed tomography Andrea Buratti 1, Soufian Ben Achour 1, Christopher

More information

Investigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography

Investigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography Investigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography Natalia Grozmani 1, Andrea Buratti 1, Robert H. Schmitt 1 More info about this article:

More information

On the impact of probing form error on form measurement in Computed Tomography

On the impact of probing form error on form measurement in Computed Tomography On the impact of probing form error on form measurement in Computed Tomography Jan Lasson Andreasen 1, Filippo Montanari 2, Leonardo De Chiffre 2 More info about this article: http://www.ndt.net/?id=23729

More information

More Info at Open Access Database

More Info at Open Access Database More Info at Open Access Database www.ndt.net/?id=18750 Abstract Enhancing the Accuracy of Computed Tomography Measurements using Data Filtering Alessandro Stolfi a, Maarja-Helena Kallasse b, Lorenzo Carli

More information

Simulation Based POD Estimation for Radiographic Testing of Turbine Blades

Simulation Based POD Estimation for Radiographic Testing of Turbine Blades 5 th European-American Workshop on Reliability of NDE Lecture 12 Simulation Based POD Estimation for Radiographic Testing of Turbine Blades Hans-Uwe BARON *, Benjamin HENKEL *, Carsten BELLON **, Andreas

More information

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY David K. Lehmann 1, Kathleen Brockdorf 1 and Dirk Neuber 2 1 phoenix x-ray Systems + Services Inc. St. Petersburg, FL, USA 2 phoenix x-ray Systems + Services

More information

The Virtual CMM a software tool for uncertainty evaluation practical application in an accredited calibration lab

The Virtual CMM a software tool for uncertainty evaluation practical application in an accredited calibration lab The Virtual CMM a software tool for uncertainty evaluation practical application in an accredited calibration lab Michael Trenk *, Matthias Franke **, Dr. Heinrich Schwenke ** * FEINMESS GmbH&Co. KG **

More information

Use of reference objects for correction of measuring errors in X-ray computed tomography

Use of reference objects for correction of measuring errors in X-ray computed tomography Downloaded from orbit.dtu.dk on: Nov 19, 2018 Use of reference objects for correction of measuring errors in X-ray computed tomography Müller, Pavel Publication date: 2010 Document Version Publisher's

More information

Industrial Computer Tomography for Dimensional Metrology: Overview of Influence Factors and Improvement Strategies

Industrial Computer Tomography for Dimensional Metrology: Overview of Influence Factors and Improvement Strategies 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 Industrial Computer Tomography for Dimensional Metrology: Overview of Influence Factors and

More information

Registration concepts for the just-in-time artefact correction by means of virtual computed tomography

Registration concepts for the just-in-time artefact correction by means of virtual computed tomography DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Registration concepts for the just-in-time artefact correction by means of virtual

More information

Application of an Industrial CT Reference Standard for Cast Freeform Shaped Workpieces

Application of an Industrial CT Reference Standard for Cast Freeform Shaped Workpieces International Symposium on Digital Industrial Radiology and Computed Tomography - We.4.3 Application of an Industrial CT Reference Standard for Cast Freeform Shaped Workpieces Markus BARTSCHER *, Karsten

More information

ASSESSMENT OF MEASUREMENT UNCERTAINTY CAUSED IN THE PREPARATION OF MEASUREMENTS USING COMPUTED TOMOGRAPHY

ASSESSMENT OF MEASUREMENT UNCERTAINTY CAUSED IN THE PREPARATION OF MEASUREMENTS USING COMPUTED TOMOGRAPHY XIX IMEKO World Congress Fundamental and Applied Metrology September 6 11, 2009, Lisbon, Portugal ASSESSMENT OF MEASUREMENT UNCERTAINTY CAUSED IN THE PREPARATION OF MEASUREMENTS USING COMPUTED TOMOGRAPHY

More information

Investigation of the kinematic system of a 450 kv CT scanner and its influence on dimensional CT metrology applications

Investigation of the kinematic system of a 450 kv CT scanner and its influence on dimensional CT metrology applications Investigation of the kinematic system of a 450 kv CT scanner and its influence on dimensional CT metrology applications Frank Welkenhuyzen 1, Bart Boeckmans 1, Ye Tan 1,2, Kim Kiekens 1,2, Wim Dewulf 1,2,

More information

Experimental study on performance verification tests for coordinate measuring systems with optical distance sensors

Experimental study on performance verification tests for coordinate measuring systems with optical distance sensors Experimental study on performance verification tests for coordinate measuring systems with optical distance sensors Simone Carmignato University of Padova, DTG, Stradella San Nicola 3, 36100 Vicenza, Italy

More information

Calibration of the depth-measuring device of Rockwell hardness testing machines

Calibration of the depth-measuring device of Rockwell hardness testing machines Calibration of the depth-measuring device of Rockwell hardness testing machines F. Löffler, A. Sawla, P. Strobel* Physikalisch-Technische Bundesanstalt, Braunschweig * Heinrich Bareiss Prüfgerätebau GmbH,

More information

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT More Info at Open Access Database www.ndt.net/?id=18754 Novel evaluation method of low contrast resolution performance of dimensional X-ray CT Makoto Abe 1, Hiroyuki Fujimoto 1, Osamu Sato 1, Kazuya Matsuzaki

More information

PERFORMANCE ASSESSMENT OF GEOMETRY MEASUREMENTS WITH MICRO-CT USING A DISMOUNTABLE WORK-PIECE-NEAR REFERENCE STANDARD

PERFORMANCE ASSESSMENT OF GEOMETRY MEASUREMENTS WITH MICRO-CT USING A DISMOUNTABLE WORK-PIECE-NEAR REFERENCE STANDARD PERFORMANCE ASSESSMENT OF GEOMETRY MEASUREMENTS WITH MICRO-CT USING A DISMOUNTABLE WORK-PIECE-NEAR REFERENCE STANDARD Markus BARTSCHER, Marko NEUKAMM, Marc KOCH, Ulrich NEUSCHAEFER-RUBE Physikalisch-Technische

More information

INFLUENCE OF GEOMETRICAL MAGNIFICATION ON COMPUTED TOMOGRAPHY DIMENSIONAL MEASUREMENTS

INFLUENCE OF GEOMETRICAL MAGNIFICATION ON COMPUTED TOMOGRAPHY DIMENSIONAL MEASUREMENTS DOI: 10.2507/27th.daaam.proceedings.090 INFLUENCE OF GEOMETRICAL MAGNIFICATION ON COMPUTED TOMOGRAPHY DIMENSIONAL MEASUREMENTS Amalija Horvatić Novak, Biserka Runje, Danijel Butković This Publication has

More information

Combining Analytical and Monte Carlo Modelling for Industrial Radiology

Combining Analytical and Monte Carlo Modelling for Industrial Radiology 19 th World Conference on Non-Destructive Testing 2016 Combining Analytical and Monte Carlo Modelling for Industrial Radiology Carsten BELLON, Gerd-Rüdiger JAENISCH, Andreas DERESCH BAM Bundesanstalt für

More information

Distance measurements on inaccessible wires in plastic parts with X-ray computed tomography

Distance measurements on inaccessible wires in plastic parts with X-ray computed tomography Distance measurements on inaccessible wires in plastic parts with X-ray computed tomography More info about this article: http://www.ndt.net/?id=20814 Michael Reiter 1, Christian Gusenbauer 1, Johann Kastner

More information

INVESTIGATION OF MEASURING STRATEGIES IN COMPUTED TOMOGRAPHY

INVESTIGATION OF MEASURING STRATEGIES IN COMPUTED TOMOGRAPHY NEWTECH 2011 NEW TECHNOLOGIES IN MANUFACTURING ISBN 978-80-214-4267-2 INVESTIGATION OF MEASURING STRATEGIES IN COMPUTED TOMOGRAPHY Pavel Müller, Jochen Hiller, Angela Cantatore, Leonardo De Chiffre Technical

More information

Multi-spectral (W, Mo, Cu, and Ag) XCT measurements

Multi-spectral (W, Mo, Cu, and Ag) XCT measurements Multi-spectral (W, Mo, Cu, and Ag) XCT measurements Marko Katić 1, Damir Markučič 1, Gorana Baršić 1 Faculty of mechanical engineering and naval architecture, Ivana Lucica 5, Zagreb, Croatia, marko.katic@fsb.hr

More information

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS 23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS P. Helberg 1,

More information

CT for Industrial Metrology Accuracy and Structural Resolution of CT Dimensional Measurements

CT for Industrial Metrology Accuracy and Structural Resolution of CT Dimensional Measurements CT for Industrial Metrology Accuracy and Structural Resolution of CT Dimensional Measurements Simone Carmignato 1, Anna Pierobon 2, Paola Rampazzo 1, Matteo Parisatto 3, Enrico Savio 2 1 University of

More information

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China 17th World Conference on Nondestructive Testing, 25-28 Oct 28, Shanghai, China Structural Index of Metal Matrix Composites Evaluated With Computed Tomography Juergen GOEBBELS Federal Institute for Materials

More information

Characterization of microshells experimented on Laser Megajoule using X-Ray tomography

Characterization of microshells experimented on Laser Megajoule using X-Ray tomography Characterization of microshells experimented on Laser Megajoule using X-Ray tomography More info about this article: http://www.ndt.net/?id=20881 Alexandre Choux, Lise Barnouin, Ludovic Reverdy, Marc Theobald

More information

The Influence of Data Filtering on Dimensional Measurements with CT

The Influence of Data Filtering on Dimensional Measurements with CT 18th World Conference on Nondestructive Testing, 16-20 April 2012, Durban, South Africa The Influence of Data Filtering on Dimensional Measurements with CT Markus BARTSCHER 1, Andreas STAUDE 2, Karsten

More information

Selection of items for InteraqCT Comparison on Assemblies

Selection of items for InteraqCT Comparison on Assemblies More Info at Open Access Database www.ndt.net/?id=18776 Selection for InteraqCT Comparison on Assemblies Alessandro Stolfi, Leonardo De Chiffre Technical University of Denmark, Produktionstorvet Building

More information

Simulation of Radiographic Testing for POD Assessment

Simulation of Radiographic Testing for POD Assessment 4th European-American Workshop on Reliability of NDE - Th.5.B.2 Simulation of Radiographic Testing for POD Assessment Gerd-Rüdiger JAENISCH, Carsten BELLON, Uwe EWERT, BAM, Berlin, Germany Abstract. Efficient

More information

Uncertainty Evaluation of Pore Analysis for Additively Manufactured Parts using Cross Sections

Uncertainty Evaluation of Pore Analysis for Additively Manufactured Parts using Cross Sections Uncertainty Evaluation of Pore Analysis for Additively Manufactured Parts using Cross Sections More info about this article: http://www.ndt.net/?id=23659 Abstract Leonard Schild 1, Manuel Fülling 1, Benjamin

More information

An efficient procedure for traceable dimensional measurements and the characterization of industrial CT systems

An efficient procedure for traceable dimensional measurements and the characterization of industrial CT systems Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18050 An efficient procedure

More information

Quality control phantoms and protocol for a tomography system

Quality control phantoms and protocol for a tomography system Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing

More information

Traceable 3D X-Ray Measurements ZEISS METROTOM

Traceable 3D X-Ray Measurements ZEISS METROTOM 20 Traceable 3D X-Ray Measurements With a computer tomograph, you can successfully perform measuring and inspection jobs yourself with only one X-ray scan. The standard acceptance test, the precision engineering

More information

AN EVALUATION OF NON-CONTACT LASER SCANNER PERFORMANCE USING CALIBRATED ARTIFACTS

AN EVALUATION OF NON-CONTACT LASER SCANNER PERFORMANCE USING CALIBRATED ARTIFACTS AN EVALUATION OF NON-CONTACT LASER SCANNER PERFORMANCE USING CALIBRATED ARTIFACTS Michael Gomez, Colleen Vaccaro, Chase Maness, Connor Lawrence, Carson Murray, Luis Vargas, Greg Caskey, Chris Evans, and

More information

CT Systems and their standards

CT Systems and their standards CT Systems and their standards Stephen Brown Engineering Measurement 11 th April 2012 Industrial X-ray computed tomography: The future of co-ordinate metrology? Burleigh Court, Loughborough University

More information

Length, Germany, PTB (Physikalisch-Technische Bundesanstalt)

Length, Germany, PTB (Physikalisch-Technische Bundesanstalt) Laser radiations Laser radiations Laser radiations Laser radiations Length Length Frequency stabilized laser (He- Ne): vacuum wavelength Frequency stabilized laser (He- Ne): absolute frequency Frequency

More information

Investigation on Comparability of Surface- and Material Dependent Measurements in Multi-Sensor Coordinate Measuring Machines

Investigation on Comparability of Surface- and Material Dependent Measurements in Multi-Sensor Coordinate Measuring Machines Investigation on Comparability of Surface- and Material Dependent Measurements in Multi-Sensor Coordinate Measuring Machines M. Vogel, M. Rosenberger, G. Linß, S.C.N. Töpfer Technische Universität Ilmenau,

More information

Comparision of contact and contactless measuring methods for form evaluation

Comparision of contact and contactless measuring methods for form evaluation Available online at www.sciencedirect.com Procedia Engineering 48 (2012 ) 273 279 MMaMS 2012 Comparision of contact and contactless measuring methods for form evaluation Peter Ka uch a, Miroslav Dovica

More information

Evaluation strategies in CT scanning

Evaluation strategies in CT scanning Downloaded from orbit.dtu.dk on: Dec 20, 2017 Evaluation strategies in CT scanning Hiller, Jochen Publication date: 2012 Document Version Publisher's PDF, also known as Version of record Link back to DTU

More information

Reconstruction Methods for Coplanar Translational Laminography Applications

Reconstruction Methods for Coplanar Translational Laminography Applications Reconstruction Methods for Coplanar Translational Laminography Applications U. EWERT, K.-U. THIESSENHUSEN, A. DERESCH, C. BELLON, S. HOHENDORF, S. KOLKOORI, N. WROBEL, B. REDMER, M. TSCHAIKNER, BAM, Berlin

More information

MICROTOPOGRAPHY THE PHOTOGRAMMETRIC DETERMINATION OF FRICTION SURFACES

MICROTOPOGRAPHY THE PHOTOGRAMMETRIC DETERMINATION OF FRICTION SURFACES MICROTOPOGRAPHY THE PHOTOGRAMMETRIC DETERMINATION OF FRICTION SURFACES Matthias HEMMLEB, Jörg ALBERTZ Technical University Berlin, Photogrammetry and Cartography Phone: +49-(0)30-314-23991, Fax: +49-(0)30-314-21104

More information

Form evaluation algorithms in coordinate metrology

Form evaluation algorithms in coordinate metrology Form evaluation algorithms in coordinate metrology D. Zhang, M. Hill, J. McBride & J. Loh Mechanical Engineering, University of Southampton, United Kingdom Abstract Form evaluation algorithms that characterise

More information

Three-Dimensional Laser Scanner. Field Evaluation Specifications

Three-Dimensional Laser Scanner. Field Evaluation Specifications Stanford University June 27, 2004 Stanford Linear Accelerator Center P.O. Box 20450 Stanford, California 94309, USA Three-Dimensional Laser Scanner Field Evaluation Specifications Metrology Department

More information

EMRP JRP IND62 TIM: Use of on-board metrology systems for area-scanning on machine tools

EMRP JRP IND62 TIM: Use of on-board metrology systems for area-scanning on machine tools EMRP JRP IND62 TIM: Use of on-board metrology systems for area-scanning on machine tools Author: Co-authors: doc. Ing. Vít Zelený, CSc. doc. Ing. Ivana Linkeová, Ph.D. Ing. Jakub Sýkora Pavel Skalník Jaromír

More information

Radiographic Simulator artist: Version 2

Radiographic Simulator artist: Version 2 18th World Conference on Nondestructive Testing, 16-20 April 2012, Durban, South Africa Radiographic Simulator artist: Version 2 Carsten Bellon 1, Andreas Deresch 1, Christian Gollwitzer 1, Gerd-Rüdiger

More information

Comparison of Reconstruction Methods for Computed Tomography with Industrial Robots using Automatic Object Position Recognition

Comparison of Reconstruction Methods for Computed Tomography with Industrial Robots using Automatic Object Position Recognition 19 th World Conference on Non-Destructive Testing 2016 Comparison of Reconstruction Methods for Computed Tomography with Industrial Robots using Automatic Object Position Recognition Philipp KLEIN 1, Frank

More information

Computer aided error analysis for a threedimensional precision surface mapping system

Computer aided error analysis for a threedimensional precision surface mapping system Computer aided error analysis for a threedimensional precision surface mapping system M. Hill, J.W. McBride, D. Zhang & J. Loh Mechanical Engineering, U~riversity of Southampton, UK Abstract This paper

More information

A NON-IMAGING OPTICAL SYSTEM FOR CHARACTERISATION OF BALL- SHAPED MICRO-INDENTERS

A NON-IMAGING OPTICAL SYSTEM FOR CHARACTERISATION OF BALL- SHAPED MICRO-INDENTERS IMEKO 22 nd TC3, 12 th TC5 and 3 rd TC22 International Conferences 3 to 5 February, 2014, Cape Town, Republic of South Africa A NON-IMAGING OPTICAL SYSTEM FOR CHARACTERISATION OF BALL- SHAPED MICRO-INDENTERS

More information

CT Reconstruction with Good-Orientation and Layer Separation for Multilayer Objects

CT Reconstruction with Good-Orientation and Layer Separation for Multilayer Objects 17th World Conference on Nondestructive Testing, 25-28 Oct 2008, Shanghai, China CT Reconstruction with Good-Orientation and Layer Separation for Multilayer Objects Tong LIU 1, Brian Stephan WONG 2, Tai

More information

Simulation-based correction of systematic errors for CT measurements

Simulation-based correction of systematic errors for CT measurements Simulation-based correction of systematic errors for CT measurements Andreas Michael Müller 1, Florian Wohlgemuth 1, Tino Hausotte 1 1 Institute of Manufacturing Metrology, Friedrich-Alexander-Universität

More information

COMPARISON OF ERROR MAPPING TECHNIQUES FOR COORDINATE MEASURING MACHINES USING THE PLATE METHOD AND LASER TRACER TECHNIQUE

COMPARISON OF ERROR MAPPING TECHNIQUES FOR COORDINATE MEASURING MACHINES USING THE PLATE METHOD AND LASER TRACER TECHNIQUE I IMEKO World Congress Fundamental and Applied Metrology September 6, 9, Lisbon, Portugal COMPARISON OF ERROR MAPPING TECHNIQUES FOR COORDINATE MEASURING MACHINES USING THE PLATE METHOD AND LASER TRACER

More information

X-rays see all X-ray computed microtomography (µct) a novel technique available at CERN for material and metrological inspection

X-rays see all X-ray computed microtomography (µct) a novel technique available at CERN for material and metrological inspection X-rays see all X-ray computed microtomography (µct) a novel technique available at CERN for material and metrological inspection Mariusz Jedrychowski EN/MME-MM Introduction of new technique at CERN 12/12/2017

More information

Predicting the mechanical behaviour of large composite rocket motor cases

Predicting the mechanical behaviour of large composite rocket motor cases High Performance Structures and Materials III 73 Predicting the mechanical behaviour of large composite rocket motor cases N. Couroneau DGA/CAEPE, St Médard en Jalles, France Abstract A method to develop

More information

Case Studies in the Use of Computed Tomography for Non-Destructive Testing, Inspection and Measurement

Case Studies in the Use of Computed Tomography for Non-Destructive Testing, Inspection and Measurement SINCE2013 Singapore International NDT Conference & Exhibition 2013, 19-20 July 2013 Case Studies in the Use of Computed Tomography for Non-Destructive Testing, Inspection and Measurement Andrew A. MALCOLM,

More information

Enhanced material contrast by dual-energy microct imaging

Enhanced material contrast by dual-energy microct imaging Enhanced material contrast by dual-energy microct imaging Method note Page 1 of 12 2 Method note: Dual-energy microct analysis 1. Introduction 1.1. The basis for dual energy imaging Micro-computed tomography

More information

Improvements Metrosoft QUARTIS R12

Improvements Metrosoft QUARTIS R12 Improvements Metrosoft QUARTIS R12 Improvements Metrosoft QUARTIS R12 At a glance Metrosoft QUARTIS R12 offers a wide range of improvements for all users and significantly contributes to optimize daily

More information

OUT OF MACHINE CALIBRATION TECHNIQUE FOR ANALOG PROBES

OUT OF MACHINE CALIBRATION TECHNIQUE FOR ANALOG PROBES XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil OUT OF MACHINE CALIBRATION TECHNIQUE FOR ANALOG PROBES J.J. Aguilar 1, J.A. Yagüe 1, J.A.

More information

Traceability for computationally-intensive metrology User manual for Chebyshev algorithm testing

Traceability for computationally-intensive metrology User manual for Chebyshev algorithm testing Traceability for computationally-intensive metrology User manual for Chebyshev algorithm testing 1 Copyright: TraCIM e. V. Certification number: 008 Authors: Klaus Wendt 1 ; Ulrich Brandt 1, Ulrich Lunze

More information

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications 4th International Symposium on NDT in Aerospace 2012 - Tu.3.A.1 Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications Malte KURFISS 1, Gerd STRECKENBACH 2 1 YXLON International

More information

MULTIPLE-SENSOR INTEGRATION FOR EFFICIENT REVERSE ENGINEERING OF GEOMETRY

MULTIPLE-SENSOR INTEGRATION FOR EFFICIENT REVERSE ENGINEERING OF GEOMETRY Proceedings of the 11 th International Conference on Manufacturing Research (ICMR2013) MULTIPLE-SENSOR INTEGRATION FOR EFFICIENT REVERSE ENGINEERING OF GEOMETRY Feng Li, Andrew Longstaff, Simon Fletcher,

More information

Introduction to XCT and its growth

Introduction to XCT and its growth Introduction to XCT and its growth Professor Richard Leach FInstP FIoN Engineering Measurement Division Measurement Network Meeting, Loughborough April 2012 X-ray computed tomography History X-ray were

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 10360-5 Second edition 2010-09-15 Geometrical product specifications (GPS) Acceptance and reverification tests for coordinate measuring machines (CMM) Part 5: CMMs using single

More information

Available online at ScienceDirect. Procedia Engineering 161 (2016 ) Bohdan Stawiski a, *, Tomasz Kania a

Available online at   ScienceDirect. Procedia Engineering 161 (2016 ) Bohdan Stawiski a, *, Tomasz Kania a Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 161 (16 ) 937 943 World Multidisciplinary Civil Engineering-Architecture-Urban Planning Symposium 16, WMCAUS 16 Testing Quality

More information

CIVA Computed Tomography Modeling

CIVA Computed Tomography Modeling CIVA Computed Tomography Modeling R. FERNANDEZ, EXTENDE, France S. LEGOUPIL, M. COSTIN, D. TISSEUR, A. LEVEQUE, CEA-LIST, France page 1 Summary Context From CIVA RT to CIVA CT Reconstruction Methods Applications

More information

Optiv CT160 Accurate measurement for every detail. Computed Tomography System

Optiv CT160 Accurate measurement for every detail. Computed Tomography System Optiv CT160 Accurate measurement for every detail Computed Tomography System Computed Tomography CT Technology. New prospects for metrology. Computed tomography (CT), a technology with unique advantages,

More information

Advanced Reconstruction Techniques Applied to an On-Site CT System

Advanced Reconstruction Techniques Applied to an On-Site CT System 2nd International Symposium on NDT in Aerospace 2010 - We.1.A.4 Advanced Reconstruction Techniques Applied to an On-Site CT System Jonathan HESS, Markus EBERHORN, Markus HOFMANN, Maik LUXA Fraunhofer Development

More information

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China 7th World Conference on Nondestructive Testing, 25-28 Oct 2008, Shanghai, China Image Registration Combining Digital Radiography and Computer-Tomography Image Data Frank HEROLD YXLON International X-Ray

More information

1.5µm lidar for helicopter blade tip vortex detection

1.5µm lidar for helicopter blade tip vortex detection 1.5µm lidar for helicopter blade tip vortex detection June 24 th 2009 A. Dolfi, B. Augere, J. Bailly, C. Besson, D.Goular, D. Fleury, M. Valla objective of the study Context : European project AIM (Advanced

More information

A RADIAL WHITE LIGHT INTERFEROMETER FOR MEASUREMENT OF CYLINDRICAL GEOMETRIES

A RADIAL WHITE LIGHT INTERFEROMETER FOR MEASUREMENT OF CYLINDRICAL GEOMETRIES A RADIAL WHITE LIGHT INTERFEROMETER FOR MEASUREMENT OF CYLINDRICAL GEOMETRIES Andre R. Sousa 1 ; Armando Albertazzi 2 ; Alex Dal Pont 3 CEFET/SC Federal Center for Technological Education of Sta. Catarina

More information

REPRESENTATION REQUIREMENTS OF AS-IS BUILDING INFORMATION MODELS GENERATED FROM LASER SCANNED POINT CLOUD DATA

REPRESENTATION REQUIREMENTS OF AS-IS BUILDING INFORMATION MODELS GENERATED FROM LASER SCANNED POINT CLOUD DATA REPRESENTATION REQUIREMENTS OF AS-IS BUILDING INFORMATION MODELS GENERATED FROM LASER SCANNED POINT CLOUD DATA Engin Burak Anil 1 *, Burcu Akinci 1, and Daniel Huber 2 1 Department of Civil and Environmental

More information

A cone-beam CT geometry correction method based on intentional misalignments to render the projection images correctable

A cone-beam CT geometry correction method based on intentional misalignments to render the projection images correctable A cone-beam CT geometry correction method based on intentional misalignments to render the projection images correctable Felix Meli, Benjamin A. Bircher, Sarah Blankenberger, Alain Küng, Rudolf Thalmann

More information

Mr. Chaobin Li China Lucky Film Corp. No. 6 Lekai South Street 6, Baoding, Hebei Province, PRC Tel.:

Mr. Chaobin Li China Lucky Film Corp. No. 6 Lekai South Street 6, Baoding, Hebei Province, PRC Tel.: Evaluation of the image quality of X-ray exposed radiographic film samples of Lucky Film L4 in standard AGFA NDT S machine processing (8 min cycle time) BAM reference Copy Customer BAM 8.3/7728a 1 st copy

More information

INSPECTION OF THE TURBINE BLADES USING SCANNING TECHNIQUES

INSPECTION OF THE TURBINE BLADES USING SCANNING TECHNIQUES INSPECTION OF THE TURBINE BLADES USING SCANNING TECHNIQUES H. Nieciag, M. Traczynski and Z. Chuchro Department of Geometrical Quanities Metrology The Institute of Metal Cutting, 30-011 Cracow, Poland Abstract:

More information

Advanced Computed Tomography System for the Inspection of Large Aluminium Car Bodies

Advanced Computed Tomography System for the Inspection of Large Aluminium Car Bodies ECNDT 2006 - Th.3.4.2 Advanced Computed Tomography System for the Inspection of Large Aluminium Car Bodies M. SIMON, I. TISEANU, C. SAUERWEIN Hans WÄLISCHMILLER, Meersburg, Germany M. SINDEL, M. BRODMANN,

More information

A universal method to optimise measurement uncertainty, time and cost for CMM scanning technology

A universal method to optimise measurement uncertainty, time and cost for CMM scanning technology universal method to optimise measurement uncertainty, time and cost for CMM scanning technology Prof. Dr.-Ing Robert Schmitt, Dipl.-Ing. Susanne Nisch Laboratory for Machine Tools and Production Engineering

More information

18th World Conference on Nondestructive Testing, April 2012, Durban, South Africa

18th World Conference on Nondestructive Testing, April 2012, Durban, South Africa 18th World Conference on Nondestructive Testing, 16-20 April 2012, Durban, South Africa TOWARDS ESTABLISHMENT OF STANDARDIZED PRACTICE FOR ASSESSMENT OF SPATIAL RESOLUTION AND CONTRAST OF INTERNATIONAL

More information

Micro Cutting Tool Measurement by Focus-Variation

Micro Cutting Tool Measurement by Focus-Variation Micro Cutting Tool Measurement by Focus-Variation Stefan Scherer 1, Reinhard Danzl 2, and Franz Helmli 3 1 CEO Alicona*; e-mail: stefan.scherer@alicona.com 2 Alicona Research*; e-mail: reinhard.danzl@alicona.com

More information

SWING ARM OPTICAL CMM

SWING ARM OPTICAL CMM SWING ARM OPTICAL CMM Peng Su, Chang Jin Oh, Robert E. Parks, James H. Burge College of Optical Sciences University of Arizona, Tucson, AZ 85721 OVERVIEW The swing arm profilometer described in reference

More information

arxiv: v1 [physics.ins-det] 21 Aug 2009

arxiv: v1 [physics.ins-det] 21 Aug 2009 NOTE arxiv:0908.3209v1 [physics.ins-det] 21 Aug 2009 Practical Way to Measure Large-Scale 2D Surfaces Using Repositioning on Coordinate-Measuring Machines Sergey Kosarevsky Saint-Petersburg Institute of

More information

Advanced Non-Destructive Testing by High Resolution Computed Tomography for 3D analysis of Automotive Components

Advanced Non-Destructive Testing by High Resolution Computed Tomography for 3D analysis of Automotive Components Advanced Non-Destructive Testing by High Resolution Computed Tomography for 3D analysis of Automotive Components J. Luebbehuesen GE Sensing & Inspection Technologies GmbH, phoenix x-ray, Niels-Bohr-Str.

More information

Process Monitoring using three dimensional Computed Tomography and Automatic Image Processing

Process Monitoring using three dimensional Computed Tomography and Automatic Image Processing DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Process Monitoring using three dimensional Computed Tomography and Automatic Image

More information

Good Practice guide to measure roundness on roller machines and to estimate their uncertainty

Good Practice guide to measure roundness on roller machines and to estimate their uncertainty Good Practice guide to measure roundness on roller machines and to estimate their uncertainty Björn Hemming, VTT Technical Research Centre of Finland Ltd, Finland Thomas Widmaier, Aalto University School

More information

GEOMETRIC CALIBRATION OF CMMs USING 3D LENGTH MEASUREMENTS

GEOMETRIC CALIBRATION OF CMMs USING 3D LENGTH MEASUREMENTS GEOMETRIC CALIBRATION OF CMMs USING 3D LENGTH MEASUREMENTS F.P. d Hooghe 1, P. Schellekens 2 and L. Levasier 2 1 Metrology Laboratory in Europe, Mitutoyo Nederland B.V. 2 Section Precision Engineering,

More information

Industrial Computed Tomography Innovations

Industrial Computed Tomography Innovations GE Inspection Technologies Industrial Computed Tomography Innovations Premium performance for premium quality and speed. gemeasurement.com/ct Premium CT technologies. Faster than ever before. Every industrial

More information

Multi-sensor measuring technology. O-INSPECT The best of optical and contact measuring technology for true 3D measurements.

Multi-sensor measuring technology. O-INSPECT The best of optical and contact measuring technology for true 3D measurements. Multi-sensor measuring technology O-INSPECT The best of optical and contact measuring technology for true 3D measurements. 2 // multifunctionality made BY CarL Zeiss The moment you realize that new requirements

More information

QUALITY CONTROL OF PARTS AND ASSEMBLIES USING SOFTWARE GOM INSPECT BASED ON PORTABLE CMM DATA

QUALITY CONTROL OF PARTS AND ASSEMBLIES USING SOFTWARE GOM INSPECT BASED ON PORTABLE CMM DATA 8 th International Quality Conference May 23 rd 2014 Center for Quality, Faculty of Engineering, University of Kragujevac Milan Blagojevic 1) Miroslav Zivkovic 1) 1) Faculty of Engineering, University

More information