Industrial Computer Tomography for Dimensional Metrology: Overview of Influence Factors and Improvement Strategies
|
|
- Naomi Cobb
- 6 years ago
- Views:
Transcription
1 Industrial Computer Tomography for Dimensional Metrology: Overview of Influence Factors and Improvement Strategies F. a, K. Kiekens b, M. Pierlet b, W. Dewulf b, P. Bleys a, J.-P. Kruth a and A. Voet c Abstract While Computer Tomography (CT) has since long been used for medical applications and material inspection, its application field has recently been broadened to include dimensional metrology in industry. In particular, it provides a unique means of measuring assemblies, complex structures as well as the inner geometry of parts made by rapid manufacturing in a non-destructive way. However, the accuracy of CT-based measurements remains yet largely uncertain, due to a number of influencing factors related to the workpiece, the CT equipment as well as the measurement setup. This paper first classifies the factors influencing the CT performance on the basis of the basic components of the CT system. Next, the obtained knowledge is used to suggest improvement strategies to ameliorate the accuracy of CT measurements. Contact information a frank.welkenhuyzen@mech.kuleuven.be philip.bleys@mech.kuleuven.be jean-pierre.kruth@mech.kuleuven.be Department of Mechanical Engineering, K.U.Leuven Celestijnenlaan 300B, B-3001 Leuven, Belgium b kim.kiekens@groept.be mieke.pierlet@groept.be wim.dewulf@groept.be Groep T - International University College Leuven, K.U.Leuven Association A. Vesaliusstraat 13, B-3000 Leuven, Belgium c andre.voet@denayer.wenk.be De Nayer Instituut J. De Nayerlaan 5, B-2860 Sint Katelijne Waver, Belgium 1
2 Introduction Computer Tomography (CT) is a well known technique in the medical world and in the field of material inspection. Its application field has recently been broadened to include dimensional metrology. CT enables measuring the objects outside as well as the inside (e.g. invisible holes) in a non-destructive way. The ability to measure the inside of a part makes industrial Computer Tomography attractive and unique in the world of dimensional metrology. Assemblies, complex structures as well as the inner geometry of parts made by rapid manufacturing can be measured in a non-destructive way Figure 1: Principle of industrial Computer Tomography (illustrated in case of a flat panel detector) [1]. Figure 1 illustrates the principle of Computer Tomography. A source produces X-rays by projecting electrons on a target. As the X-rays penetrate the workpiece (sample), the X-rays are attenuated (made weaker), due to absorption or scattering. The amount of attenuation is determined by the amount of material penetrated, but also by the material composition, After penetrating the workpiece, the attenuated X-rays will incident on the detector, resulting in a 2D gray image. 2D images are taken for many rotation steps of the workpiece. Reconstruction based on these projected images leads to a 3D voxel model (a voxel is the 3D analogue of a pixel), where the gray value of the voxels are a measure for the density of the material. This paper describes the different factors influencing CT performance (Section 2) and considers improvement strategies to ameliorate the accuracy of CT measurements (Section 3). 2. Factors influencing CT performance In order to quantify the accuracy of measurements using Computer Tomography, one has to identify the error sources. Figure 2 classifies the most important factors influencing CT performance using the basic components of the CT system as a starting point. X-ray Source The factors related to the X-ray source are partly determined by the machine, but also partly to be chosen by the operator. The X-ray tube voltage can be chosen by the operator within a machine specific range. The higher the voltage, the more penetrating the X-ray beam becomes. The applied current is a user-defined input as well, which affects the intensity of the beam (i.e. the quantity of radiation energy) [2]. 2
3 Figure 2: Classification of factors influencing the CT performance. Another important quantity is the focal spot size. Figure 3 illustrates the effect induced by the spot size. The smaller the spot size, the sharper the edges will be. In case of large spot sizes unsharpness will occur, known as the penumbra effect. A disadvantage of a smaller spot size is the concentrated heat produced at the spot on the target (transmissive, rotating, ) inside the X- ray tube, requiring cooled targets and limiting the maximum applicable voltage Figure 3: Small spot size (left) compared to large spot size (right). Other influencing factors are the target material and type (reflective or transmission target), the X-ray spectrum, The polychromatic character of conventional X-ray sources causes the well known effect of beam hardening: while the X-ray beam penetrates material, the low-energy X-rays (soft X-rays) are more easily attenuated than the higher-energy X-rays. As a consequence the image on the detector differs from the expected image, resulting in observable errors in the reconstructed volume. The amount of beam hardening depends on the initial X-ray spectrum as well as on the composition, density and the amount of material traversed. One way to reduce beam hardening is to place a thin sheet filter in front of the workpiece in order to absorb the low 3
4 energy radiation from the spectrum, hence approximating a more monochromatic energy distribution. Disadvantages are the loss of intensity with corresponding reduction of the signal-to-noise ratio due to a limited dynamic range of the detector [3] and inadequate results in case of multi-material objects [4]. Rotation Table and Workpiece Since CT implies a reconstruction of X-ray images taken at different orientations, the workpiece is mounted on a rotation table. Influencing factors of this rotation table are the geometrical errors of the mechanical axes and the number of rotation steps chosen by the operator. Also workpiece characteristics including e.g. material composition and dimensions influence the accuracy of the CT results. Detector Two types of detectors can be distinguished: 2D flat panel detectors and 1D line detectors. When using a flat panel detector, a single rotation of the object normally suffices, provided the detector is sufficiently large (in combination with restricted magnification). For line detectors, the rotation of the object should be complemented with an additional vertical displacement for every slice to be measured (e.g. 10,000 shifts of 10 µm for a 100 mm workpiece with 10 µm resolution). Consequently, the use of a line detector is more time consuming. Line detectors may however accommodate higher beam power (i.e. thicker objects) and better accuracy. A recent possibility (yet, since long applied in the medical world) is the use of a helix CT [5], where the workpiece makes a helical movement. Other important detector characteristics include pixel size, number of pixels, exposure time and signal-to-noise ratio. Data Processing Processing the detector output consists of two steps: reconstruction of the 2D images into a 3D voxel model, followed by the actual measurements (including determination of the edges, thresholding). While discussing the drawbacks of the polychromatic character of the X-ray source above, the problem of beam hardening was already mentioned: without appropriate corrections, beam hardening results in observable errors in the reconstructed volume. Another problem related to the reconstruction is X-ray scattering: as the X-ray beam passes through material some of the original energy in the beam can be deflected onto a new path. The effect changes with each projection resulting in artifacts in the reconstructed image [6]. Positional Relationship The distances between the detector, X-ray source and workpiece have a great influence on the accuracy of the CT measurements. A geometrical magnification is achieved by positioning the object close to the source. As a result more pixels of the detector are used, theoretically improving the resolution. However, at the same time the image becomes more blurred annihilating part of the benefit: A smaller distance between the source and the workpiece results in less parallel X-ray beams and a larger penumbra effect (i.e. less sharp edges). 4
5 Improvement strategies Section 2 provided a short overview of the factors influencing CT performance. Hilpert et al. [7] provided a list of strategies improving this performance; an extended list is provided and discussed in this section. Improvement of dedicated CT components This improvement strategy is more devoted to CT equipment manufacturers; for users it is primarily a concern at the time of purchase. Particularly the X-ray source and the detector are concerned in this improvement strategy, e.g. a detector with smaller pixel size and better signal-to-noise ratio, or an X-ray source with a smaller spot size. Improvement of the hardware stability Using CT as a measuring device raises the question whether the machine construction techniques used for Coordinate Measuring Machines (CMMs) should be adopted: granite tables, air bearings, Referring to Figure 2, this concerns the geometrical errors of the mechanical axes, precision of and minimal rotation steps of the rotational table, and the positional relationship between rotation table, source and detector. Wenig et al. [8] investigate the dependency of the measurement error and the measurement quality on the adjustment accuracy of the mechanical positioning system by simulation of various alignment accuracies. 0,4 0,35 0, Error (mm) 0,25 0,2 positive Z-direction: measurement 1 0,15 positive Z-direction measurement 2 0,1 negative Z-direction measurement 1 0,05 negative Z-direction measurement , Position (mm) Figure 4: Positioning errors of the Z-axis (Magnification axis). To illustrate the occurrence of geometrical errors of the mechanical axes, Figure 4 represents the positioning errors (error between inputted and actual position) of the Z-axis (magnification axis, axis between source and detector) of a CT machine available to KULeuven. The errors on the position are measured over a range of 700 mm. The values are represented both when moving in positive Z-direction and when moving in negative Z-direction. All errors are measured twice, using a laser interferometer, similar to the international standard ISO 230, Part 2: Determination of accuracy and repeatability of positioning numerically controlled axes [9]. The results show a high difference in repeatability of the both directions. Furthermore there is a large direction dependency. It can be concluded that positioning of the object at the same Z-position can lead to high differences in the actual position. For instance if a reference standard 5
6 is used, the Z-axis by preference shouldn t be moved when interchanging the object and the reference standard. If movement of the Z-position is nevertheless necessary the final movement direction to obtain the required position should be the same (for this machine the positive Z-direction is preferred because of the higher repeatability). Optimization of user defined settings Different settings are to be chosen by the operator: voltage and current of the X-ray source, detector exposure time, threshold value, use of appropriate filters (material and thickness) to avoid beam hardening,. There is no simple textbook describing the optimal settings. Instead the user has to determine the values based on his own knowledge and experience, or by prior testing or computer simulation [8,10]. Furthermore the operator is responsible for an optimal orientation of the workpiece. This is illustrated by an experiment, measuring a stepped workpiece with several length dimensions in two directions. Figure 5 demonstrates 2 possible orientations to measure the workpiece. The lengths in Y direction and in Z-direction were measured in both orientations. The results demonstrated that the length measurements in the rotational plane (XZ plane) were much more accurate than those in the Y-direction. Figure 6 also shows that there is much more noise on the planes parallel to the XZ plane (lower accuracy in Y) Figure 5: Two possible orientations for the workpiece of Figure Figure 6: Effects of orientation (left: set-up 1, right: set-up 2) on the workpiece measurements: noise on planes parallel to XZ-plane. 6
7 Improvement of data processing The captured images obtained by the detector have to be processed. Two steps can be distinguished: Tomographic data reconstruction Reconstructed voxel data post-processing Tomographic data reconstruction combines the 2D projection images to a 3D voxel model. Section 2 (X-ray Source and Data Processing) described two unwanted effects occurring when performing CT: beam hardening and scattering. One way to reduce beam hardening is to use a filter (Section 2, X-ray Source). Another way to decrease these unwanted effect is by algorithmic adaptations, as described by several authors (e.g. [3,4,8,11]). Krumm et al. [4] provide a short overview of common methods to correct beam hardening: pre-filtering (described before), linearization, the dual energy approach and the post-reconstruction method. In addition, Krumm et al. [4] propose a new method to perform beam hardening correction on CT images of multi-material objects. The method combines the advantages of the linearization technique and the post-reconstruction method. Another model to correct beam hardening is presented by Graham et al. [3]. A 3D voxel model is obtained after reconstruction. The next step before being able to measure dimensions is the detection of the edges. The voxel model doesn t simply consist of voxels of 2 (or more in case of multi-material) possible gray values: one representing material and the other one representing air (background). Instead there is a wide range of gray values. A gradual transition, not a simple sharp edge, can be observed at the borders of the workpiece. It is the operator s (or software s) responsibility to determine the edges (Figure 7). Different strategies are possible: one threshold value can be calculated for the complete voxel model (all voxels having a gray value less than the threshold value represent air and the voxels having a higher gray value represent material). More sophisticated methods can be used that evaluate the local gray values near the edge, for instance by taking the average of the local gray values as threshold value or by evaluating the gradients Figure 7: Edge detection [12]. Reduction of systematic measurement errors via calibrated reference standards Reference standards can be used to reduce measurement errors and to ensure traceability of the measurements to the unit of length (the meter). Reference standards can for instance be used for the correction of scale errors or for the determination of threshold values [13], but also for acceptance and accuracy verification of the CT system. PTB (Physikalisch-Technische Bundesanstalt) developed reference standards for applications in the field of CT metrology. Figure 8 provides some examples [13]: Aluminum hollow cylinders for determining the CT threshold; An Aluminum step-cylinder for CT system analysis and parameterization for cast parts; 7
8 A Ball-bar made from ceramic balls on a carbon fibre rod for assessment of scale errors and traceability Figure 8: Reference standards for CT, developed by PTB [13] Figure 9: Sphere calotte plate made from Zerodur [7]. In [7] and [13] a sphere calotte plate (figure 9), manufactured by PTB is used as reference standard. The results in these publications proof the accuracy improvement. Additional measurement of the workpiece Making use of the data retrieved by an additional (more accurate) measurement of the workpiece by a conventional Coordinate Measuring Machine or by an optical measurement system (e.g. laser line scanner or fringe projection system) improves the CT-data. The extra information can for instance be used to rescale the CT-data. The major advantage of industrial Computer Tomography is the possibility to measure the inside (e.g. holes) of a workpiece. Rescaling the CT data by measurement data obtained from external features of the workpiece by a more accurate machine (which is not capable of measuring the inside) improves the accuracy of the overall CT measurement results (including the inside measurements). Experiments using a flat panel detector illustrate that the use of one scale factor is not sufficient, and at least two distinct scale factors have to be used. The experiment consisted of measuring the lengths in X, Y and Z direction of the workpiece in Figure 5 and Figure 6. It was clear that one scale factor was needed for the Y direction, and another scale factor was necessary for the X and Z direction. Some manufacturers combine the CT principle and additional measurement systems in one machine. Another possibility is accomplishing the additional measurement with a stand-alone machine. Neumayer et al. [14] use an additional optical measurement system to enhance the overall accuracy of CT-data by merging the dataset obtained by CT measurement and a dataset obtained by optical measurement. Franz and Kasperl [11] explain a method that makes use of an additional optical sensor for beam hardening correction, which results in a quality and speed improvement for the industrial CT. This way, beam hardening correction can be achieved during the data collection, whereas this is normally performed afterwards in a very time-consuming step. 8
9 Conclusion Computer Tomography is an interesting technique, well-known in the medical world and in the field of material inspection. The possibility to examine the outside as well as the (invisible) inside of objects makes the technique very attractive for dimensional metrology applications. Nevertheless, significant research efforts remain necessary to enhance the quality of the CT measurements. This paper first classified the factors influencing CT performance. This allows spotting the weaknesses and detecting ways to improve the CT measurements. Next, a list of improvement strategies was listed and discussed. Improvements regarding hardware, software, input parameters as well as the possibility to do an additional measurement of the workpiece or of a reference standard were considered. 5. Acknowledgement The authors acknowledge the support of the Institute for the Promotion of Innovation by Science and Technology in Flanders (IWT-Vlaanderen) in the framework of the TETRA programme (project IWT080137). References [1] Phoenix X-ray, [2] BALL, J., MOORE, A. D., Essential Physics for Radiographers. Oxford: Blackwell Science, [3] GRAHAM, D., NITIN, J., JAMES, E., A modelling approach to beam hardening correction. In: Proceedings of the SPIE, September 2008, San Diego. S.R. STOCK, ed., 2008, [4] KRUMM, M., KASPERL, S., FRANZ, M., Reducing non-linear artifacts of multi-material objects in industrial 3D computed tomography. NDT and E International, 41 (4), , [5] Helix-CT, Fraunhofer-Institut für Integrierte Schaltungen ISS, Produktbeschreibung, [6] KAK, A., SLANEY, M., Principles of Computerized Tomographic Imaging, Philadelphia: SIAM,2001 [7] HILPERT, U., BARTSCHER, M., NEUGEBAUER, M., GOEBBELS, J., WEIDEMANN, G., BELLON, C., Simulation-aided computed tomography (CT) for dimensional measurements, International Symposium on Digital industrial Radiology and Computed Tomography, June , Lyon, France. [8] WENIG, P., KASPERL, S., Examination of the Measurement Uncertainty on Dimensional Measurements by X-ray Computed Tomography. In: ECNDT Proceedings, September 2006, Berlin. [9] ISO 230-2:2006, Test Code for Machine Tools, Part 2: Determination of Accuracy and Repeatability of Positioning Numerically Controlled Axes. [10] KERKHOFS, G., MOESEN, M., VAN BAEL, S., ELICEGUI, L., MAES, F., LOECKX, D., SCHROOTEN, J., WEVERS, M., Mechanical characterization of porous structures by combining micro- CT imaging and 3D image analysis with in-situ loading, FEA and local strain mapping, submitted to Advanced engineering materials. [11] FRANZ, M., KASPERL, S., Quality and Speed Improvements in Industrial CT by the Use of an additional Optical Sensor, In: ECNDT Proceedings, September 2006, Berlin. [12] Metris, [13] BARTSCHER, M., HILPERT, U., GOEBBELS, J., WEIDEMANN, G., Enhancement and Proof of Accuracy of Industrial Computed Tomography (CT) Measurements, Annals of the CIRP, 56 (1), , [14] NEUMAYER, D., MODRICH, K., MAISL, M., KASPERL, S., Computed Tomography as a tool for industrial measurement. In: ECNDT Proceedings, September 2006, Berlin. 9
Parameter Dependent Thresholding for Dimensional X-ray Computed Tomography
International Symposium on Digital Industrial Radiology and Computed Tomography - Poster 12 Parameter Dependent Thresholding for Dimensional X-ray Computed Tomography Kim KIEKENS *, Ye TAN *, Jean-Pierre
More informationInvestigation of the kinematic system of a 450 kv CT scanner and its influence on dimensional CT metrology applications
Investigation of the kinematic system of a 450 kv CT scanner and its influence on dimensional CT metrology applications Frank Welkenhuyzen 1, Bart Boeckmans 1, Ye Tan 1,2, Kim Kiekens 1,2, Wim Dewulf 1,2,
More informationHIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY
HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY David K. Lehmann 1, Kathleen Brockdorf 1 and Dirk Neuber 2 1 phoenix x-ray Systems + Services Inc. St. Petersburg, FL, USA 2 phoenix x-ray Systems + Services
More informationCoordinate Measuring Machines with Computed Tomography
Always a Step Ahead with Quality Coordinate Measuring Machines with Computed Tomography Multisensor Coordinate Measuring Machines with Computed Tomography Computed Tomography in Coordinate Measuring Machines
More informationInvestigation on the influence of image quality in X-ray CT metrology
Investigation on the influence of image quality in X-ray CT metrology Pavel Müller 1, Jochen Hiller 1, Angela Cantatore 1, Markus Bartscher 2, Leonardo De Chiffre 1 1 Technical University of Denmark, Department
More informationDevelopments in Dimensional Metrology in X-ray Computed Tomography at NPL
Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Wenjuan Sun and Stephen Brown 10 th May 2016 1 Possible factors influencing XCT measurements Components Influencing variables Possible
More informationMulti-spectral (W, Mo, Cu, and Ag) XCT measurements
Multi-spectral (W, Mo, Cu, and Ag) XCT measurements Marko Katić 1, Damir Markučič 1, Gorana Baršić 1 Faculty of mechanical engineering and naval architecture, Ivana Lucica 5, Zagreb, Croatia, marko.katic@fsb.hr
More informationTHREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS
THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS A. Campos a*, A. Zorrilla a, F. Lasagni a a Materials & Processes dept., Center for Advanced Aerospace Technologies
More informationRegistration concepts for the just-in-time artefact correction by means of virtual computed tomography
DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Registration concepts for the just-in-time artefact correction by means of virtual
More informationINVESTIGATION OF MEASURING STRATEGIES IN COMPUTED TOMOGRAPHY
NEWTECH 2011 NEW TECHNOLOGIES IN MANUFACTURING ISBN 978-80-214-4267-2 INVESTIGATION OF MEASURING STRATEGIES IN COMPUTED TOMOGRAPHY Pavel Müller, Jochen Hiller, Angela Cantatore, Leonardo De Chiffre Technical
More informationUse of reference objects for correction of measuring errors in X-ray computed tomography
Downloaded from orbit.dtu.dk on: Nov 19, 2018 Use of reference objects for correction of measuring errors in X-ray computed tomography Müller, Pavel Publication date: 2010 Document Version Publisher's
More informationIntroduction to XCT and its growth
Introduction to XCT and its growth Professor Richard Leach FInstP FIoN Engineering Measurement Division Measurement Network Meeting, Loughborough April 2012 X-ray computed tomography History X-ray were
More informationHow beam hardening influences the dimensional geometry in X-ray computed tomography
How beam hardening influences the dimensional geometry in X-ray computed tomography More info about this article: http://www.ndt.net/?id=20809 Hannah Corcoran 1, 2, Stephen Brown 2, Stuart Robson 1, Robert
More informationASSESSMENT OF MEASUREMENT UNCERTAINTY CAUSED IN THE PREPARATION OF MEASUREMENTS USING COMPUTED TOMOGRAPHY
XIX IMEKO World Congress Fundamental and Applied Metrology September 6 11, 2009, Lisbon, Portugal ASSESSMENT OF MEASUREMENT UNCERTAINTY CAUSED IN THE PREPARATION OF MEASUREMENTS USING COMPUTED TOMOGRAPHY
More informationComputed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system
Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system ECNDT 2014 Prague October 6-10, 2014 Dr. Eberhard Neuser Dr. Alexander Suppes Imagination
More informationAdvanced Non-Destructive Testing by High Resolution Computed Tomography for 3D analysis of Automotive Components
Advanced Non-Destructive Testing by High Resolution Computed Tomography for 3D analysis of Automotive Components J. Luebbehuesen GE Sensing & Inspection Technologies GmbH, phoenix x-ray, Niels-Bohr-Str.
More informationHIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS
23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS P. Helberg 1,
More informationComparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling
nd International Symposium on NDT in Aerospace - We..A. Comparison of Probing Error in Dimensional Measurement by Means of D Computed Tomography with Circular and Helical Sampling Jochen HILLER, Stefan
More informationAdvanced Reconstruction Techniques Applied to an On-Site CT System
2nd International Symposium on NDT in Aerospace 2010 - We.1.A.4 Advanced Reconstruction Techniques Applied to an On-Site CT System Jonathan HESS, Markus EBERHORN, Markus HOFMANN, Maik LUXA Fraunhofer Development
More informationAn initial experimental study on the influence of beam hardening in X-ray CT for dimensional metrology
An initial experimental study on the influence of beam hardening in X-ray CT for dimensional metrology Joseph J. Lifton 1,2, Andrew A. Malcolm 2, John W. McBride 3 1 The University of Southampton, United
More informationDigital Laminography and Computed Tomography with 600 kv for Aerospace Applications
4th International Symposium on NDT in Aerospace 2012 - Tu.3.A.1 Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications Malte KURFISS 1, Gerd STRECKENBACH 2 1 YXLON International
More informationCase Studies in the Use of Computed Tomography for Non-Destructive Testing, Inspection and Measurement
SINCE2013 Singapore International NDT Conference & Exhibition 2013, 19-20 July 2013 Case Studies in the Use of Computed Tomography for Non-Destructive Testing, Inspection and Measurement Andrew A. MALCOLM,
More informationA new Concept for High-Speed atline and inlinect for up to 100% Mass Production Process Control
18th World Conference on Nondestructive Testing, 16-20 April 2012, Durban, South Africa A new Concept for High-Speed atline and inlinect for up to 100% Mass Production Process Control Oliver BRUNKE 1,
More informationRadiographic Simulator artist: Version 2
18th World Conference on Nondestructive Testing, 16-20 April 2012, Durban, South Africa Radiographic Simulator artist: Version 2 Carsten Bellon 1, Andreas Deresch 1, Christian Gollwitzer 1, Gerd-Rüdiger
More informationA new Concept for High-Speed atline and inlinect for up to 100% Mass Production Process Control
6 th International Congress of Metrology, 06003 (203) DOI: 0.05/ metrology/20306003 C Owned by the authors, published by EDP Sciences, 203 A new Concept for High-Speed atline and inlinect for up to 00%
More informationCreating a Multi-material Probing Error Test for the Acceptance Testing of Dimensional Computed Tomography Systems
Creating a Multi-material Probing Error Test for the Acceptance Testing of Dimensional Computed Tomography Systems More info about this article: http://www.ndt.net/?id=20848 Fabricio Borges de Oliveira
More informationOptiv CT160 Accurate measurement for every detail. Computed Tomography System
Optiv CT160 Accurate measurement for every detail Computed Tomography System Computed Tomography CT Technology. New prospects for metrology. Computed tomography (CT), a technology with unique advantages,
More informationDr. Javier Santillan, San Carlos, CA
X-ray diffraction as a tool for automated residual st ress analysis & a non-synchrotron based nanofocus x-ray computed tomography technique for materials characterization and metrology Dr. Javier Santillan,
More informationWieblinger Weg 92a, Heidelberg, Germany, Phone: , Fax: ;
HOW INDUSTRIAL COMPUTER TOMOGRAPHY ACCELERATES PRODUCT DEVELOPMENT IN THE LIGHT METAL CASTING AND INJECTION MOULDING INDUSTRY. Christof REINHART 1, Christoph POLIWODA 1, Thomas GÜNTHER 1 1 Volume Graphics
More informationA new concept for high-speed atline and inline CT for up to 100% mass production process control allowing both 3D metrology and failure analysis
A new concept for high-speed atline and inline CT for up to 100% mass production process control allowing both 3D metrology and failure analysis Oliver Brunke 1, Ferdinand Hansen 2, Ingo Stuke 3, Friedrich
More informationQuality control phantoms and protocol for a tomography system
Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing
More informationJapan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field. Kouichi Inagaki ICC / IHI Corporation
Japan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field Kouichi Inagaki ICC / IHI Corporation 13 th WORLD CONFERENCE ON INVESTMENT CASTING Paper: T3 Copyright
More informationDEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE
Rajesh et al. : Proceedings of the National Seminar & Exhibition on Non-Destructive Evaluation DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh V Acharya, Umesh Kumar, Gursharan
More informationFrequency-based method to optimize the number of projections for industrial computed tomography
More Info at Open Access Database www.ndt.net/?id=18739 Frequency-based method to optimize the number of projections for industrial computed tomography Andrea Buratti 1, Soufian Ben Achour 1, Christopher
More informationExperimental investigation of new multi-material gap reference standard for testing computed tomography systems
More info about this article: http://www.ndt.net/?id=20837 Experimental investigation of new multi-material gap reference standard for testing computed tomography systems Petr Hermanek 1, Fabrício Borges
More informationIndustrial Computed Tomography Innovations
GE Inspection Technologies Industrial Computed Tomography Innovations Premium performance for premium quality and speed. gemeasurement.com/ct Premium CT technologies. Faster than ever before. Every industrial
More informationDigital Image Processing
Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture
More informationCombining Analytical and Monte Carlo Modelling for Industrial Radiology
19 th World Conference on Non-Destructive Testing 2016 Combining Analytical and Monte Carlo Modelling for Industrial Radiology Carsten BELLON, Gerd-Rüdiger JAENISCH, Andreas DERESCH BAM Bundesanstalt für
More informationMore Info at Open Access Database
More Info at Open Access Database www.ndt.net/?id=18750 Abstract Enhancing the Accuracy of Computed Tomography Measurements using Data Filtering Alessandro Stolfi a, Maarja-Helena Kallasse b, Lorenzo Carli
More informationCT in Dimensional Metrology, Engineering and Manufacture
CT in Production Quality Control - Leuven, June 4, 2013 CT in Dimensional Metrology, Engineering and Manufacture Prof. Dr. Ir. Jean-Pierre KRUTH Katholieke Universiteit Leuven (K.U.Leuven) Mechanical Engineering
More informationInvestigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography
Investigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography Natalia Grozmani 1, Andrea Buratti 1, Robert H. Schmitt 1 More info about this article:
More informationCreating a Multi-Material Length Measurement Error Test for the Acceptance Testing of Dimensional Computed Tomography Systems
Creating a Multi-Material Length Measurement Error Test for the Acceptance Testing of Dimensional Computed Tomography Systems More info about this article: http://www.ndt.net/?id=20847 Fabrício Borges
More informationMetal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland
Metal Artifact Reduction CT Techniques R S S S Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland N. 1 v o 4 1 0 2. Postoperative CT Metal Implants CT is accurate for assessment
More informationVoxel dataset noise reduction algorithm based on CMM reference measurements for geometrical tolerance analysis of prismatic test pieces
Voxel dataset noise reduction algorithm based on CMM reference measurements for geometrical tolerance analysis of prismatic test pieces More info about this article: http://www.ndt.net/?id=21929 Crhistian
More informationAUTOMATED HIGH SPEED VOLUME COMPUTED TOMOGRAPHY FOR INLINE QUALITY CONTROL
AUTOMATED HIGH SPEED VOLUME COMPUTED TOMOGRAPHY FOR INLINE QUALITY CONTROL R. Hanke 1, A. Kugel 2, and P. Troup 2 1 Fraunhofer-Entwicklungszentrum für Röntgentechnik, Fürth, Germany; 2 Austria Aluguss
More informationCIVA Computed Tomography Modeling
CIVA Computed Tomography Modeling R. FERNANDEZ, EXTENDE, France S. LEGOUPIL, M. COSTIN, D. TISSEUR, A. LEVEQUE, CEA-LIST, France page 1 Summary Context From CIVA RT to CIVA CT Reconstruction Methods Applications
More informationProcess Monitoring using three dimensional Computed Tomography and Automatic Image Processing
DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Process Monitoring using three dimensional Computed Tomography and Automatic Image
More informationProcess Monitoring using Three Dimensional Computed Tomography and Automatic Image Processing
ECNDT 2006 - We.3.7.1 Process Monitoring using Three Dimensional Computed Tomography and Automatic Image Processing Michael MAISL, Fraunhofer EZRT, Saarbrücken, Germany Stefan KASPERL, STEVEN OEKL, Fraunhofer
More informationCharacterization of microshells experimented on Laser Megajoule using X-Ray tomography
Characterization of microshells experimented on Laser Megajoule using X-Ray tomography More info about this article: http://www.ndt.net/?id=20881 Alexandre Choux, Lise Barnouin, Ludovic Reverdy, Marc Theobald
More informationCT Measurement Realization Process: The Need for Advanced Correction Methods
CT Measurement Realization Process: The Need for Advanced Correction Methods Crhistian R Baldo 1, Thiago L Fernandes 2, Gustavo D Donatelli 2, Gergely Szikszay-Molnar 3, Kostadin Doytchinov 3 1 Federal
More informationArtifact Mitigation in High Energy CT via Monte Carlo Simulation
PIERS ONLINE, VOL. 7, NO. 8, 11 791 Artifact Mitigation in High Energy CT via Monte Carlo Simulation Xuemin Jin and Robert Y. Levine Spectral Sciences, Inc., USA Abstract The high energy (< 15 MeV) incident
More informationX-ray Industrial Computed Laminography (ICL) Simulation Study of Planar Objects: Optimization of Laminographic Angle
More info about this article: http://www.ndt.net/?id=21086 X-ray Industrial Computed Laminography (ICL) Simulation Study of Planar Objects: Optimization of Laminographic Angle Lakshminarayana Yenumula
More informationAnalysis of Combined Probing Measurement Error and Length Measurement Error Test for Acceptance Testing in Dimensional Computed Tomography
Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18038 Analysis of Combined Probing
More informationArion: a realistic projection simulator for optimizing laboratory and industrial micro-ct
Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct J. DHAENE* 1, E. PAUWELS 1, T. DE SCHRYVER 1, A. DE MUYNCK 1, M. DIERICK 1, L. VAN HOOREBEKE 1 1 UGCT Dept. Physics
More informationThe new generation of industrial computed tomography. The Desktop CT exact S
The new generation of industrial computed tomography The Desktop CT S Volume scanning technology Innovation with a family tradition Founded in 1968, the family-owned, WEN- ZEL Metrology Group is one of
More informationINFLUENCE OF GEOMETRICAL MAGNIFICATION ON COMPUTED TOMOGRAPHY DIMENSIONAL MEASUREMENTS
DOI: 10.2507/27th.daaam.proceedings.090 INFLUENCE OF GEOMETRICAL MAGNIFICATION ON COMPUTED TOMOGRAPHY DIMENSIONAL MEASUREMENTS Amalija Horvatić Novak, Biserka Runje, Danijel Butković This Publication has
More informationINDUSTRIAL SYSTEM DEVELOPMENT FOR VOLUMETRIC INTEGRITY
INDUSTRIAL SYSTEM DEVELOPMENT FOR VOLUMETRIC INTEGRITY VERIFICATION AND ANALYSIS M. L. Hsiao and J. W. Eberhard CR&D General Electric Company Schenectady, NY 12301 J. B. Ross Aircraft Engine - QTC General
More informationTranslational Computed Tomography: A New Data Acquisition Scheme
2nd International Symposium on NDT in Aerospace 2010 - We.1.A.3 Translational Computed Tomography: A New Data Acquisition Scheme Theobald FUCHS 1, Tobias SCHÖN 2, Randolf HANKE 3 1 Fraunhofer Development
More informationFast iterative beam hardening correction based on frequency splitting in computed tomography
Fast iterative beam hardening correction based on frequency splitting in computed tomography Qiao Yang a,b, Matthias Elter b, Ingo Schasiepen b, Nicole Maass b and Joachim Hornegger a,c a Pattern Recognition
More informationCOMPARISON OF ERROR MAPPING TECHNIQUES FOR COORDINATE MEASURING MACHINES USING THE PLATE METHOD AND LASER TRACER TECHNIQUE
I IMEKO World Congress Fundamental and Applied Metrology September 6, 9, Lisbon, Portugal COMPARISON OF ERROR MAPPING TECHNIQUES FOR COORDINATE MEASURING MACHINES USING THE PLATE METHOD AND LASER TRACER
More informationCT Systems and their standards
CT Systems and their standards Stephen Brown Engineering Measurement 11 th April 2012 Industrial X-ray computed tomography: The future of co-ordinate metrology? Burleigh Court, Loughborough University
More informationCase Studies in Nondestructive Testing and Evaluation
Case Studies in Nondestructive Testing and Evaluation 6 (2016) 79 92 More info about this article: http://www.ndt.net/?id=20771 Contents lists available at ScienceDirect Case Studies in Nondestructive
More informationX-rays see all X-ray computed microtomography (µct) a novel technique available at CERN for material and metrological inspection
X-rays see all X-ray computed microtomography (µct) a novel technique available at CERN for material and metrological inspection Mariusz Jedrychowski EN/MME-MM Introduction of new technique at CERN 12/12/2017
More informationRadiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada
Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Tecniques of imaging in Overview Introduction Tecniques of imaging
More informationAdvanced Computed Tomography System for the Inspection of Large Aluminium Car Bodies
ECNDT 2006 - Th.3.4.2 Advanced Computed Tomography System for the Inspection of Large Aluminium Car Bodies M. SIMON, I. TISEANU, C. SAUERWEIN Hans WÄLISCHMILLER, Meersburg, Germany M. SINDEL, M. BRODMANN,
More informationAn Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography
4th International Symposium on NDT in Aerospace 2012 - Tu.3.A.3 An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography Jonathan HESS *, Patrick KUEHNLEIN *, Steven OECKL
More informationMethods for Quantitative Characterization of Large-Scale High Energy Computed Tomography Systems
More Info at Open Access Database www.ndt.net/?id=16704 Methods for Quantitative Characterization of Large-Scale High Energy Computed Tomography Systems Michael BÖHNEL 1, Andreas GRUBER 2, Nils REIMS 1,
More informationRobot-Based Solutions for NDT Inspections: Integration of Laser Ultrasonics and Air Coupled Ultrasounds for Aeronautical Components
19 th World Conference on Non-Destructive Testing 2016 Robot-Based Solutions for NDT Inspections: Integration of Laser Ultrasonics and Air Coupled Ultrasounds for Aeronautical Components Esmeralda CUEVAS
More informationScienceDirect. Comparison of Different Method of Measurement Geometry using CMM, Optical Scanner and Computed Tomography 3D
Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 69 (2014 ) 255 262 24th DAAAM International Symposium on Intelligent Manufacturing and Automation, 2013 Comparison of Different
More informationDistance measurements on inaccessible wires in plastic parts with X-ray computed tomography
Distance measurements on inaccessible wires in plastic parts with X-ray computed tomography More info about this article: http://www.ndt.net/?id=20814 Michael Reiter 1, Christian Gusenbauer 1, Johann Kastner
More informationNON-COLLIMATED SCATTERED RADIATION TOMOGRAPHY
NON-COLLIMATED SCATTERED RADIATION TOMOGRAPHY Gorshkov V.A., Space Research Institute, Moscow, Russia Yumashev V.M., State corporation "Rosatom", Centre "Atom-innovation", Moscow, Russia Kirilenko K.V.,
More informationOptics Vac Work MT 2008
Optics Vac Work MT 2008 1. Explain what is meant by the Fraunhofer condition for diffraction. [4] An aperture lies in the plane z = 0 and has amplitude transmission function T(y) independent of x. It is
More informationDraft SPOTS Standard Part III (7)
SPOTS Good Practice Guide to Electronic Speckle Pattern Interferometry for Displacement / Strain Analysis Draft SPOTS Standard Part III (7) CALIBRATION AND ASSESSMENT OF OPTICAL STRAIN MEASUREMENTS Good
More informationPhase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube
Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube T. Donath* a, F. Pfeiffer a,b, O. Bunk a, W. Groot a, M. Bednarzik a, C. Grünzweig a, E. Hempel c, S. Popescu c, M. Hoheisel
More informationNovel evaluation method of low contrast resolution performance of dimensional X-ray CT
More Info at Open Access Database www.ndt.net/?id=18754 Novel evaluation method of low contrast resolution performance of dimensional X-ray CT Makoto Abe 1, Hiroyuki Fujimoto 1, Osamu Sato 1, Kazuya Matsuzaki
More informationNon-Destructive Failure Analysis and Measurement for Molded Devices and Complex Assemblies with X-ray CT and 3D Image Processing Techniques
SINCE2013 Singapore International NDT Conference & Exhibition 2013, 19-20 July 2013 Non-Destructive Failure Analysis and Measurement for Molded Devices and Complex Assemblies with X-ray CT and 3D Image
More informationA geometry measurement system for a dimensional cone-beam CT
A geometry measurement system for a dimensional cone-beam CT More info about this article: http://www.ndt.net/?id=21904 Benjamin A. Bircher, Felix Meli, Alain Küng, Rudolf Thalmann Federal Institute of
More informationCOMPUTER SIMULATION OF X-RA Y NDE PROCESS COUPLED
COMPUTER SIMULATION OF X-RA Y NDE PROCESS COUPLED WITH CAD INTERFACE Carsten Bellon, Gerd-Rüdiger Tillack, Christina Nockemann, and Lutz Stenze! Laboratory VIII.33 "Reliability ofnon-destructive Evaluation"
More informationXRADIA microxct Manual
XRADIA microxct Manual Multiscale CT Lab Table of Contents 1. Introduction and Basics 1.1 Instrument Parts 1.2 Powering up the system 1.3 Preparing your sample 2. TXM Controller 2.1 Starting up 2.2 Finding
More information17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China
17th World Conference on Nondestructive Testing, 25-28 Oct 28, Shanghai, China Structural Index of Metal Matrix Composites Evaluated With Computed Tomography Juergen GOEBBELS Federal Institute for Materials
More informationTraceable 3D X-Ray Measurements ZEISS METROTOM
20 Traceable 3D X-Ray Measurements With a computer tomograph, you can successfully perform measuring and inspection jobs yourself with only one X-ray scan. The standard acceptance test, the precision engineering
More informationEnhanced material contrast by dual-energy microct imaging
Enhanced material contrast by dual-energy microct imaging Method note Page 1 of 12 2 Method note: Dual-energy microct analysis 1. Introduction 1.1. The basis for dual energy imaging Micro-computed tomography
More informationAvailable online at ScienceDirect
Available online at www.sciencedirect.com ScienceDirect Procedia CIRP 13 ( 2014 ) 43 48 2 nd CIRP 2nd CIRP Conference on Surface Integrity (CSI) Computed Tomography as a Tool for Examining Surface Integrity
More informationMulti-slice CT Image Reconstruction Jiang Hsieh, Ph.D.
Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced
More informationAdapted acquisition trajectory and iterative reconstruction for few-views CT inspection
Adapted acquisition trajectory and iterative reconstruction for few-views CT inspection Caroline Vienne 1, Marius Costin 1 More info about this article: http://www.ndt.net/?id=21917 1 CEA, LIST, Département
More informationOn the impact of probing form error on form measurement in Computed Tomography
On the impact of probing form error on form measurement in Computed Tomography Jan Lasson Andreasen 1, Filippo Montanari 2, Leonardo De Chiffre 2 More info about this article: http://www.ndt.net/?id=23729
More information17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China
7th World Conference on Nondestructive Testing, 25-28 Oct 2008, Shanghai, China Image Registration Combining Digital Radiography and Computer-Tomography Image Data Frank HEROLD YXLON International X-Ray
More informationThe PolyCT Increasing the sample throughput using multiple rotation axes
The PolyCT Increasing the sample throughput using multiple rotation axes More info about this article: http://www.ndt.net/?id=21946 1 Dr. Nicolas Thiemeyer, Johannes Leisner 2, Michael Salamon 2, Carina
More informationDesign and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon
Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization
More informationKlingleweg 8, Meersburg, Germany, Industriestr. 4, Donauwörth, Germany,
Capabilities and Application of Specialized Computed Tomography Methods for the Determination of Characteristic Material Properties of Fiber Composite Components Michael Krumm 1, Christoph Sauerwein 1,
More informationUsing X-Ray Micro Tomography to assess the quality of Packaging Foil
Using X-Ray Micro Tomography to assess the quality of Packaging Foil Martin Koster 1, Gerard van Dalen 2, Robert Hoeve 3 1 Unilever Research 2 Unilever Research 3 Unilever Research, Olivier van Noortlaaan
More informationMicro-CT in situ study of carbonate rock microstructural evolution for geologic CO2 storage
Downloaded from orbit.dtu.dk on: Apr 10, 2018 Micro-CT in situ study of carbonate rock microstructural evolution for geologic CO2 storage Zheng, Yi; Yang, Yan; Rogowska, M.; Gundlach, Carsten Published
More informationComparison of Reconstruction Methods for Computed Tomography with Industrial Robots using Automatic Object Position Recognition
19 th World Conference on Non-Destructive Testing 2016 Comparison of Reconstruction Methods for Computed Tomography with Industrial Robots using Automatic Object Position Recognition Philipp KLEIN 1, Frank
More informationWAVELENGTH MANAGEMENT
Camera Accessories WAVELENGTH MANAGEMENT UV CONVERTERS UV Converters take advantage of a phenomenon called fluorescence to extend the performance range of the Beamage beam profiling camera to ultraviolet
More informationMEDICAL IMAGING 2nd Part Computed Tomography
MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional
More informationAM INSPECTION / NDT: IS XCT THE SOLUTION?
AM INSPECTION / NDT: IS XCT THE SOLUTION? Dr. Nick Brierley Principal Research Engineer Metrology & NDT Group nick.brierley@the-mtc.org April 2018 OUTLINE Challenges of AM post-build inspection XCT for
More informationApplications of Temporal Phase Shift Shearography
Chapter 5 Applications of Temporal Phase Shift Shearography The main applications of temporal phase shift shearography are in NDT and strain measurement. 5.1. Temporal Phase Shift Shearography for NDT
More informationPERFORMANCE ASSESSMENT OF GEOMETRY MEASUREMENTS WITH MICRO-CT USING A DISMOUNTABLE WORK-PIECE-NEAR REFERENCE STANDARD
PERFORMANCE ASSESSMENT OF GEOMETRY MEASUREMENTS WITH MICRO-CT USING A DISMOUNTABLE WORK-PIECE-NEAR REFERENCE STANDARD Markus BARTSCHER, Marko NEUKAMM, Marc KOCH, Ulrich NEUSCHAEFER-RUBE Physikalisch-Technische
More informationDigital Volume Correlation for Materials Characterization
19 th World Conference on Non-Destructive Testing 2016 Digital Volume Correlation for Materials Characterization Enrico QUINTANA, Phillip REU, Edward JIMENEZ, Kyle THOMPSON, Sharlotte KRAMER Sandia National
More informationIdentification of Shielding Material Configurations Using NMIS Imaging
Identification of Shielding Material Configurations Using NMIS Imaging B. R. Grogan, J. T. Mihalczo, S. M. McConchie, and J. A. Mullens Oak Ridge National Laboratory, P.O. Box 2008, MS-6010, Oak Ridge,
More information