1 Oxford ICP 2-step DRIE SOP Oxford ICP 2-step DRIE SOP Page 1 of Scope 1.1 This document provides operating procedures for the Oxford 100 ICP 2-step DRIE. 2. Table of Contents 1. Scope Table of Contents Reference Documents Referenced within this Document External Documents Equipment and/or Materials Safety Setup Procedures Record Information in Log Book Turn on Burn Box Hook up Liquid Nitrogen... 3

2 Oxford ICP 2-step DRIE SOP Page 2 of Turn on Chiller Vent Load Lock Load Wafer Pump Down Load Lock step Deep Etching Procedure Set Up and Edit Process Recipe Start Etch Process step Smooth Shallow Etching Procedure Set up and Edit Process Recipe Start the Etch Process Shutdown Procedures Vent Load Lock Shutdown System Maintenance Procedures Chamber Cleaning Check He Leak Rate Check Chamber Leak Rate Troubleshooting Process Notes Typical Etch Characteristics Process Summary Revision History Figure 1, Liquid Nitrogen Tank... 3 Figure 2, Chiller... 3 Figure 3, Pump Control Page... 4 Figure 4, Load Lock... 4 Figure 5, Temperature Controller... 5 Figure 6, Recipe Editor Page... 7 Figure 7, Process Control Page... 9 Figure 8, Etch Rate for 10 um Trenches Figure 9, Etch Rate for 2 um Trenches Figure 10, Etch Rate for 4 um Pillars Figure 11, 2-step Etching SEM Analysis Table 1, Deep Etch Recipe... 6 Table 2, Smooth Etch Recipe Reference Documents 3.1 Referenced within this Document Ecosys Burn Box SOP 3.2 External Documents None

3 Oxford ICP 2-step DRIE SOP Page 3 of Equipment and/or Materials 4.1 Wafer/Sample 4.2 Oxford Liquid Nitrogen 5. Safety 5.1 Follow all Nanofab safety procedures. 6. Setup Procedures 6.1 Record Information in Log Book Record all processing and characterization information in log book The log sheet is an excel sheet on the computer. It should already be open. 6.2 Turn on Burn Box Follow the procedure in the Ecosys Burn Box SOP to turn on the burn box. LN2 Valve 6.3 Hook up Liquid Nitrogen Figure 1, Liquid Nitrogen Tank If the LN2 hose on the back of the Oxford 100 is not connected to the LN2 tank, connect it Remove filler hose, if attached Connect Oxford 100 hose. See Figure 1, Liquid Nitrogen Tank Open the LN2 valve. 6.4 Turn on Chiller Go into the chase behind the Oxford Turn the power switch to start on the chiller. See Figure 2, Chiller. Power Switch Figure 2, Chiller

4 Oxford ICP 2-step DRIE SOP Page 4 of 13 System Button Process Buttton Load Lock Controls Main Chamber Controls Do Not Touch Evacuate Stop Vent 6.5 Vent Load Lock Click on System icon. See Figure 3, Pump Control Page Click on pump page Press stop button corresponding to the load lock mechanical pump. See Figure Press vent button. See Figure Wait for load lock to vent. The pressure will read above 600 Torr and the time will be Open load lock lid Press stop button. See Figure Load Wafer Figure 3, Pump Control Page Place wafer in load lock against the two pins on the transfer arm with the wafer facing the two pins. The wafer should touch the two pins. See Figure 4, Load Lock. Figure 4, Load Lock

5 Oxford ICP 2-step DRIE SOP Page 5 of Pump Down Load Lock Close load lock lid Select the evacuate button. See Figure Enter an ID name or number for your wafer Wait for the load lock to pump down (~1 min) Ensure that the chuck is cooling down. See Figure 5, Temperature Controller. Actual Temperature Set point Figure 5, Temperature Controller 7. 2-step Deep Etching Procedure NOTE: This recipe can be used to etch deep trenches or vias in 100 mm wafers. It is strongly advised that you do a test run on a practice wafer before working with your device wafer. 7.1 Set Up and Edit Process Recipe Click Process button. See Figure Click Recipes page Click on the New button. See Figure 6, Recipe Editor Page Chose Yes to clear the current recipe Click Load to open your recipe Select recipe and click ok If you don t have a recipe, open the Bosch_Template recipe and rename the file The recipe steps will be displayed on the left side of the screen Set the number of etch cycles Click on the Repeat step. See Figure Choose Repeat Step from the menu. See Figure Enter the number of times to repeat the etch and deposition step The etch rate is approx. 1 micron per cycle Click OK.

6 Oxford ICP 2-step DRIE SOP Page 6 of The deep etch recipe is as follows: Table 1, Deep Etch Recipe Step C 4 F 8 SF 6 Pressure ICP Power RF Power Temp. Time (C) Flow 40sccm 0 15mT s C 4 F 8 Strike 40sccm 0 15mT 500W 30W 15 3s Plasma Etch 0 70sccm 1mT 500W 25W 15 11s Deposit 95sccm 0 1mT 500W 10W 15 5s Final Etch 0 70sccm 1mT 500W 25W 15 11s The Final Etch step removes material from the last deposition step In all steps the He backing is To change process time or other parameters, click on the process step and select Edit Step and set the parameter to the desired amount After changing the file name, click Save and Click OK. 7.2 Start Etch Process Click the Run button. See Figure 6. NOTE: The wafer transfer will begin automatically and the etch process will start when the wafer is loaded and the chuck is at the correct temperature. The wafer will automatically transfer to the load lock when the process is complete Listen closely to the machine during the first steps If the machine makes a repetitive clicking sound, restart the process Press Stop Return the wafer to load lock Click the Run button Monitor the parameters during the first steps of the process Go to the Process Control page Monitor the Reflected Power. It should be 0. See Figure 7, Process Control Page step Smooth Shallow Etching Procedure 8.1 Set up and Edit Process Recipe Click Process button. See Figure 3, Pump Control Page Click Recipes page Click on the New button. See Figure 6, Recipe Editor Page Chose Yes to clear the current recipe.

7 Oxford ICP 2-step DRIE SOP Page 7 of Click Load to open your recipe Select recipe and click ok If you don t have a recipe, open the Bosch_Smooth_Template recipe and rename the file The recipe steps will be displayed on the left side of the screen Set the number of etch cycles Click on the Repeat step. See Figure Choose Repeat Step from the popup menu. See Figure Enter the number of times to repeat the etch and deposition step Click OK. Run Button Repeat Step Button The smooth etch recipe is as follows: Figure 6, Recipe Editor Page Table 2, Smooth Etch Recipe Step C 4 F 8 SF 6 Pressure ICP Power RF Power Temp. Time (C) Flow 40sccm 0 15mT s C 4 F 8 Strike 40sccm 0 15mT 400W 30W 15 3s Plasma Etch 0 52sccm 1mT 400W 25W 15 4s Deposit 65sccm 0 1mT 400W 10W 15 3s Final Etch 0 52sccm 1mT 400W 25W 15 4s NOTE: The Final Etch step removes material from the last deposition step.

8 Oxford ICP 2-step DRIE SOP Page 8 of 13 NOTE: In all steps the He backing is To change process time or other parameters, click on the process step and select Edit Step and set the parameter to the desired amount After changing the file name, click Save and Click OK. 8.2 Start the Etch Process Click the Run button. See Figure 6. NOTE: The wafer transfer will begin automatically and the etch process will start when the wafer is loaded and the chuck is at the correct temperature. The wafer will automatically transfer to the load lock when the process is complete. NOTE: The etch rate for aspect ratios less than 5:1 (the depth is 5 times greater than the width) is ~ 1.46 um/min. The average surface roughness is 1.96 nm. The sidewall roughness is approx. 30 nm Listen closely to the machine during the first steps If the machine makes a repetitive clicking sound, restart the process Press Stop Return the wafer to load lock Click the Run button Monitor the parameters during the first steps of the process Go to the Process Control page Monitor the Reflected Power. It should be 0. See Figure 7, Process Control Page.

9 Oxford ICP 2-step DRIE SOP Page 9 of 13 Reflected Power Figure 7, Process Control Page 9. Shutdown Procedures 9.1 Vent Load Lock Click OK when the software says Process completed Vent load lock Press stop button corresponding to the load lock mechanical pump. See Figure Click ok when it says Wafer has finished processing Press vent button. See Figure Wait 3 min. for the load lock to vent Pull up to open load lock lid. Do not force it open, once it is vented it should open easily. If it does not open easily, press stop, then vent, and wait 3 minutes again Press stop button to stop venting Remove wafer. 9.2 Shutdown System Close knob on liquid nitrogen dewar. See Figure Pump down load lock Close load lock lid.

10 Oxford ICP 2-step DRIE SOP Page 10 of Select the evacuate button. See Figure Press cancel when software asks for ID #. 10. Maintenance Procedures 10.1 Chamber Cleaning Chamber cleaning should be done after every 3-5 hours of etching or as needed The cleaning recipe parameters are as follows: SF6 = 10sccm O2 = 40sccm ICP power = 1000W RF power = 100W He backside pressure = 0mT Pressure = 20mT T = Room Temperature 10.2 Check He Leak Rate Check chamber pressure Check icon to make sure wafer is clamped Set He pressure to 10 mt (pressure greater than 20 Torr can damage thin substrates) Check He flow rate. It should be less than?? sccm Check chamber pressure, it should have risen less than 7 mtorr Check Chamber Leak Rate Click on Chamber Click on leak detection Pump down over night Run leak test (step time) for 20 minutes Log interval 5 sec Pumping to 7e-7 Torr Click return to process 10.4 Troubleshooting To remove clamped wafer when the system doesn t recognize there is a wafer in the chamber: Pump down load lock Click on service mode Click on location where the wafer is located in the chamber.

11 Oxford ICP 2-step DRIE SOP Page 11 of Click add wafer Exit service mode Click on green wafer and blue arrow should appear Pump down load lock Click on center of load lock to transfer wafer to load lock. 11. Process Notes 11.1 Typical Etch Characteristics Figure 8, Etch Rate for 10 um Trenches

12 Oxford ICP 2-step DRIE SOP Page 12 of 13 Figure 9, Etch Rate for 2 um Trenches Figure 10, Etch Rate for 4 um Pillars

13 Oxford ICP 2-step DRIE SOP Page 13 of Process Summary Ex. 1-Bosch Template Recipe Ex. 2-Bosch Smooth Template Recipe Figure 11, 2-step Etching SEM Analysis 12. Revision History Rev Date Originator Description of Changes 1 01 Jun 2010 Sam Bell

1. Scope 1.1 This document provides the operating procedure for plasma enhanced chemical vapor deposition with the Oxford 80+.

1. Scope 1.1 This document provides the operating procedure for plasma enhanced chemical vapor deposition with the Oxford 80+. Oxford 80+ PECVD SOP Oxford 80+ PECVD SOP Page 1 of 7 1. Scope 1.1 This document provides the operating procedure for plasma enhanced chemical vapor deposition with the Oxford 80+. 2. Table of Contents

More information

Oxford 80+ RIE SOP. Oxford 80+ RIE SOP Page 1 of 6 Revision

Oxford 80+ RIE SOP. Oxford 80+ RIE SOP Page 1 of 6 Revision Oxford 80+ RIE SOP Oxford 80+ RIE SOP Page 1 of 6 1. Scope 1.1 This document provides the operating procedure for the Oxford 80+ reactive ion etch. 2. Table of Contents 1. Scope... 1 2. Table of Contents...

More information

Oxford DSiE Revision /25/17 Page 1 of 9. Oxford DSiE SOP

Oxford DSiE Revision /25/17 Page 1 of 9. Oxford DSiE SOP Page 1 of 9 Oxford DSiE SOP The PlasmaPro 100 is a Inductively coupled plasma device (ICP) configured to provide effective high aspect ratio etching of 4 inch silicon wafers, and pieces when a carrier

More information

Xetch SOP Page 1 of 7 Revision

Xetch SOP Page 1 of 7 Revision Xetch SOP Page 1 of 7 Xetch System SOP 1. Scope 1.1 This document provides procedures and process information for etching silicon wafers with the Xetch xenon difluoride etching system. 2. Table of Contents

More information

R I T Title: Lam 4600

R I T Title: Lam 4600 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Lam 4600. All users are expected to have read and understood this document. It

More information

Lam AutoEtch 490 Plasma Etch System Operations and Maintenance Manual

Lam AutoEtch 490 Plasma Etch System Operations and Maintenance Manual Fabrication Laboratory Revision: E - 02/10/2009 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Lam AutoEtch 490 Plasma Etch System

More information

Oerlikon ICP 790 User Instructions. For the Micro-Electronics Laboratory. University of Notre Dame. Department of Electrical Engineering

Oerlikon ICP 790 User Instructions. For the Micro-Electronics Laboratory. University of Notre Dame. Department of Electrical Engineering Oerlikon ICP 790 User Instructions For the Micro-Electronics Laboratory At University of Notre Dame Department of Electrical Engineering General Precautions Contacts For problems, clarifications of procedures,

More information

Heidelberg Pattern Generator SOP

Heidelberg Pattern Generator SOP Heidelberg Pattern Generator SOP Page 1 of 15 Heidelberg Pattern Generator SOP 1. Scope 1.1 This document provides the operating procedures for the Heidelberg Pattern Generator with Version 3.12.5 software.

More information

TECHNICS MACRO-RIE 8800 STANDARD OPERATING PROCEDURES

TECHNICS MACRO-RIE 8800 STANDARD OPERATING PROCEDURES TECHNICS MACRO-RIE 8800 STANDARD OPERATING PROCEDURES Version: 1.0 Sept 2017 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center TABLE OF CONTENTS 1.0 DESCRIPTION.... 3 2.0 SAFETY....... 3

More information

ION MILLING USING PVD 75 SOP. January 2014 START UP

ION MILLING USING PVD 75 SOP. January 2014 START UP ION MILLING USING PVD 75 SOP January 2014 START UP 1. Select the VACUUM tab at the top left of the screen. NOTE: The current vacuum found at the bottom of the screen in Torr. 2. To open the chamber press

More information

Suss Aligner SOP. Suss Aligner SOP Page 1 of 6 Revision Scope 1.1 This document provides the procedure for operating the Suss Aligner.

Suss Aligner SOP. Suss Aligner SOP Page 1 of 6 Revision Scope 1.1 This document provides the procedure for operating the Suss Aligner. Suss Aligner SOP Page 1 of 6 Suss Aligner SOP 1. Scope 1.1 This document provides the procedure for operating the Suss Aligner. 2. Table of Contents 1. Scope... 1 2. Table of Contents... 1 Suss Aligner

More information

AIXTRON (formerly NanoInstruments) Carbon Nanotube Deposition System

AIXTRON (formerly NanoInstruments) Carbon Nanotube Deposition System STANDARD OPERATING PROCEDURE AIXTRON (formerly NanoInstruments) Carbon Nanotube Deposition System CORAL Name: CCNT Model Number: Black Magic Pro Location: 39-428a TRL What it does: Plasma Enhanced Chemical

More information

Introduction to operation

Introduction to operation Introduction to operation C 70 The status of the system is visualized in real time and shows a synoptic of both the transfer chamber and the main components of the machine sheet C 70A A number of different

More information

Plasma Quest RPCVD: Operating Procedure (6/24/02)

Plasma Quest RPCVD: Operating Procedure (6/24/02) Plasma Quest RPCVD: Operating Procedure (6/24/02) Considerations: *There are four basic steps that MUST exist in a recipe: Gas Stabilization, Plasma Tuning, Deposition/Process and End Step. Additional

More information

Surface Technology Systems Advanced Silicon Etching

Surface Technology Systems Advanced Silicon Etching Document nr. : EriHeu-20110410-1V1 Page nr. : 0 Author: Eric van den Heuvel Surface Technology Systems Advanced Silicon Etching 0. Table of contents 0. Table of contents... 0 1. Changes compared to previous

More information

University of Minnesota, MN Nano Center Standard Operating Procedure

University of Minnesota, MN Nano Center Standard Operating Procedure Equipment name: AV Etcher Badger name: av-etch Revision number: 2 Model: Vision 320 Revisionist: Paul Kimani Location: PAN Bay 2 Date: Aug 12, 2015 1. Description The AV Etcher is an RIE system with six

More information

FILMETRICS F20 STANDARD OPERATION PROCEDURE

FILMETRICS F20 STANDARD OPERATION PROCEDURE Arizona State University NanoFab FILMETRICS F20 STANDARD OPERATION PROCEDURE Rev C Table of Contents Contents Table of Contents...1 1. Purpose / Scope...2 2. Reference Documents...2 3. Equipment / Supplies

More information

B4 Nitride Furnace Instructions

B4 Nitride Furnace Instructions B4 Nitride Furnace Instructions The Nitride furnace is a LPCVD furnace capable of depositing standard stoichiometric (Si3N4) silicon nitride, as well as silicon rich/low stress silicon nitride. The Nitride

More information

SmartSoft AES Operator s Guide

SmartSoft AES Operator s Guide SmartSoft AES Operator s Guide Part No. 702064 Rev. A Physical Electronics USA, PHI, SMART-Tool, SmartSoft, MultiPak and Watcher are trademarks of ULVAC-PHI, Inc. All other trademarks are the property

More information

NanoBeam nb4 Standard Operating Procedure

NanoBeam nb4 Standard Operating Procedure NanoBeam nb4 Standard Operating Procedure These instructions are intended for reference only, and will not replace the thorough training required for proper system operation. Contact a clean room staff

More information

Woollam M2000 Operation Manual

Woollam M2000 Operation Manual Woollam M2000 Operation Manual The Woollam M2000 is a spectroscopic ellipsometer used to characterize optically transparent films. The system has the Near IR upgrade that covers 700 wavelengths from 193nm

More information

SUSS SB6E BONDER SOP. October 2013

SUSS SB6E BONDER SOP. October 2013 SUSS SB6E BONDER SOP October 2013 The SUSS SB6 bonder is capable of silicon fusion, anodic, and eutectic bonding. There are two separate heads; one for anodic bonding, and another for fusion bonding. Definitions

More information

Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017

Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017 Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017 Introduction This document is only for operation of the HPLC and ICP-MS instruments to obtain data. It contains no maintenance

More information

SUSS MICROTEC MJB4 MASK ALIGNER DOCUMENTATION

SUSS MICROTEC MJB4 MASK ALIGNER DOCUMENTATION SUSS MICROTEC MJB4 MASK ALIGNER DOCUMENTATION ANITHA & KARTHICK 1 Tool specification: Substrate size :1 up to 4 Mask size : Standard 2 x2 up to 5 x5 Alignment :X,Y-±5mm,theta-5⁰ Alignment Accuracy :Top

More information

Title: Heidelberg DWL66+ Semiconductor & Microsystems Fabrication Laboratory Revision: B Rev Date: 05/03/2017

Title: Heidelberg DWL66+ Semiconductor & Microsystems Fabrication Laboratory Revision: B Rev Date: 05/03/2017 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Heidelberg DWL66+. All users are expected to have read and understood this document.

More information

Equipment Standard Operating Procedure Kimberly Appel

Equipment Standard Operating Procedure Kimberly Appel Date Created: June 26, 2004 Date Modified: The Flexus 2320 Equipment Standard Operating Procedure Kimberly Appel 1. Purpose 1.1. The Flexus Thin Film Stress Measuring Apparatus (TFSMA) measures the changes

More information

Karl Suss SB6 Bonder Operation Tool Id: Rev 01, 09/24/10

Karl Suss SB6 Bonder Operation Tool Id: Rev 01, 09/24/10 Karl Suss SB6 Bonder Operation Tool Id: 119.020 The technical content of this document is the responsibility of UCLA CNSI Integrated Systems Nanofabrication Cleanroom Los Angeles, CA Karl Suss SB6 Bonder

More information

POLOS CPK WAFER SPINNER SOP

POLOS CPK WAFER SPINNER SOP POLOS CPK WAFER SPINNER SOP June 2013 Start-up Procedure: 1. Press the green Start button between the two upper displays to turn on the spinner. 2. If any of the marked orange lights in the upper right

More information

WW Celerity Applications Engineering Mass Flow Controller Troubleshooting Guide

WW Celerity Applications Engineering Mass Flow Controller Troubleshooting Guide WW Celerity Applications Engineering Mass Flow Controller Troubleshooting Guide Troubleshooting Manual for Mass Flow Controllers 1 199-001-0009 REV A TABLE OF CONTENTS Troubleshooting Guide for Digital

More information

Preparing for EBeam Write

Preparing for EBeam Write Preparing for EBeam Write Electron beam lithography is a time-intensive process that takes many steps to properly prepare for writing. In order to make the process as efficient and simple as possible,

More information

MDC CV SYSTEM OPERATION PROCEDURE

MDC CV SYSTEM OPERATION PROCEDURE MDC CV SYSTEM OPERATION PROCEDURE Rev B Table of Contents Contents Table of Contents... 1 1. Purpose / Scope... 2 2. Reference Documents... 2 3. Operation Manual... 2 4. Equipment / Supplies / Material...

More information

Title: Bruce Furnaces Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date: 02/21/06

Title: Bruce Furnaces Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date: 02/21/06 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Bruce Furnaces. All users are expected to have read and understood this document.

More information

University of Minnesota NanoFabrication Center Standard Operating Procedure

University of Minnesota NanoFabrication Center Standard Operating Procedure Equipment Name: JEOL SEM Coral Name: jeol-sem Revision Number: 4 Model: JSM-6610LV Revisionist: K. Roberts Location: Area 3 Date: 9/17/2013 1 Description The JSM-6610LV is a scanning electron microscope

More information

FLAMINGOS Camera Dewar Cooldown On Kitt Peak

FLAMINGOS Camera Dewar Cooldown On Kitt Peak FLAMINGOS Camera Dewar Cooldown On Kitt Peak Dick Joyce Version 1.04, 2012 May 08 FLAMINGOS Camera Cooldown, V 1.04, 2012 May 08 Page 1 of 7 1. Introduction This manual is a guide to cooling down the FLAMINGOS

More information

` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE

` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE ` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE Version: 1.0 April-2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research Center (NRC) TABLE OF CONTENTS 1. Introduction...2 1.1

More information

Dip Coating Machine For Sol-Gel Process Model DM 200. Owner s Manual

Dip Coating Machine For Sol-Gel Process Model DM 200. Owner s Manual Dip Coating Machine For Sol-Gel Process Model DM 200 Owner s Manual LIMITATION OF LIABILITY In no event shall CHEMAT TECHNOLOGY, INC. be held responsible, nor liable for any direct, indirect, incidental,

More information

DOT MATRIX PRINTER SP6000 SERIES

DOT MATRIX PRINTER SP6000 SERIES DOT MATRIX PRINTER SP6000 SERIES Hardware Manual < Approval: CEL > Trademark acknowledgments SP6000 : Star Micronics Co., Ltd. Notice All rights reserved. Reproduction of any part of this manual in any

More information

Denton Vacuum Phoenix 400 Continuous In-Line Sputtering Deposition System General Specifications

Denton Vacuum Phoenix 400 Continuous In-Line Sputtering Deposition System General Specifications Phoenix 400 Continuous In-Line Sputtering Deposition System s Phoenix 400X System Overview Denton Vacuum s Phoenix 400X Continuous In-Line Sputtering system provides users optimal uniform coating thicknesses

More information

Gas Cooling Accessory

Gas Cooling Accessory TA Instruments 109 Lukens Drive New Castle, DE 19720 Thermal Analysis & Rheology A SUBSIDIARY OF WATERS CORPORATION Gas Cooling Accessory Operator's Manual PN 991426.001 Rev. D (Text and Binder) PN 991426.002

More information

Tutorial: set up of measurement recipe with database

Tutorial: set up of measurement recipe with database Tutorial: set up of measurement recipe with database This tutorial explains 1. how to create a recipe to perform repeated measurements at a given place on the wafer 2. how to make a data analysis routine

More information

Cyrostat Design -Camera Section Justin Holwell

Cyrostat Design -Camera Section Justin Holwell Cyrostat Design -Camera Section Justin Holwell July 6 th, 2004 JLH 1 Camera Dewar 3D Layout JLH 2 Camera Dewar X-Section JLH 3 Camera Components Lens Mounts 12 Lens Cells and 2 Mounts Grism Assembly 2

More information

Short instructions. Olympus FV1000

Short instructions. Olympus FV1000 Olympus FV1000-1- Short instructions Short instructions. Olympus FV1000 Version 140416 This short instruction will guide you through the turn on and off sequence of the microscope system. For a detailed

More information

Simulator System SUN ABET

Simulator System SUN ABET SoP of I V Simulator System Part 2 SUN 3000 SOLAR SIMULATOR ABET Technologies Standard operating procedure: Flowchart of sequence to be followed: System switch ON sequence Stage and device probing Measuring

More information

Sygonix Home Starter Kit

Sygonix Home Starter Kit App Instruction Manual Sygonix Home Starter Kit Item No. 1597114 www.conradconnect.de Contents Conrad Connect... 4 Symbols... 4 Core functions... 5 App to device communication... 6 App Installation...

More information

Manual. LC-16 system. LC-16 Inkjet Printer 1

Manual. LC-16 system. LC-16 Inkjet Printer 1 Manual LC-16 system LC-16 Inkjet Printer 1 Index ENVIRONMENT. 3 OPERATOR S SAFETY 3 OPERATION SAFETY 3 PART 1 INSTALLATION AND PARAMETER SETTING 4 1) Preparing 4 2) Installation 4 3) Priming 4 4) Parameter

More information

Optimizer Collect User Manual. for MFC/MFM calibration with Integrator 110 and Drycal

Optimizer Collect User Manual. for MFC/MFM calibration with Integrator 110 and Drycal Optimizer Collect User Manual for MFC/MFM calibration with Integrator 110 and Drycal Table of Contents Overview of Optimizer Collect...2 PC Requirements...2 Equipment Required...2 Installing Collect on

More information

ZEPHYR Compact Haze Machine

ZEPHYR Compact Haze Machine ZEPHYR Compact Haze Machine Item ref: 160.464UK User Manual Version 1.0 Caution: Please read this manual carefully before operating Damage caused by misuse is not covered by the warranty Introduction Thank

More information

USING EMULATION TO VALIDATE A CLUSTER TOOL SIMULATION MODEL

USING EMULATION TO VALIDATE A CLUSTER TOOL SIMULATION MODEL Proceedings of the 2000 Winter Simulation Conference J. A. Joines, R. R. Barton, K. Kang, and P. A. Fishwick, eds. USING EMULATION TO VALIDATE A CLUSTER TOOL SIMULATION MODEL H. Todd LeBaron Brooks Automation,

More information

FC 1800 #1 Operation Manual University of Notre Dame

FC 1800 #1 Operation Manual University of Notre Dame \ FC 1800 #1 Operation Manual University of Notre Dame For Training Please contact: Dave Heemstra 8 June 2015 Page 1 of 12 General Precautions Contacts For problems, clarification of procedures, or general

More information

Shutdown & Startup Procedure for JEOL JSM-7600F FESEM

Shutdown & Startup Procedure for JEOL JSM-7600F FESEM Shutdown & Startup Procedure for JEOL JSM-7600F FESEM Shutdown Procedure 1. Ensure the Liquid Nitrogen Trap (LNT) as shown in Fig. 1 has no liquid nitrogen (LN2) & is at room temperature (not cold). If

More information

Operating Procedure for Horiba Raman Microscope

Operating Procedure for Horiba Raman Microscope Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch

More information

Title: Karl Suss MA150 Semiconductor & Microsystems Fabrication Laboratory Revision: E Rev Date: 03/01/06

Title: Karl Suss MA150 Semiconductor & Microsystems Fabrication Laboratory Revision: E Rev Date: 03/01/06 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Karl Suss MA150 Mask Aligner. All users are expected to have read and understood

More information

Organist Palette Guide

Organist Palette Guide Organist Palette Guide December 2016 North American Office: 5600 General Washington Drive Suite B211, Alexandria, VA 22312 USA Toll Free: (800) 272-4775 T: (703) 933-0024 F: (703) 933-0025 European Office:

More information

FORWARDING ALL CALLS TO ANOTHER PERSON

FORWARDING ALL CALLS TO ANOTHER PERSON UNDERSTANDING POLYCOM VVX 40 TABLE OF CONTENTS: COVER : UNDERSTANDING POLYCOM VVX 40... FORWARDING CALLS... LISTENING TO VOICEMAILS... MANAGING CONFERENCE CALLS... 4 TRANSFERRING CALLS... 6 UNDERSTANDING

More information

SPA III Quick Reference Guide for Custom Panels

SPA III Quick Reference Guide for Custom Panels SPA III Quick Reference Guide for Custom Panels This guide contains instructions for using the BD FACS Sample Prep Assistant III (SPA III) with BD FACS SPA software version.0 and later. Initial Setup Custom

More information

Shanghai Xixun-----professional supplier of Led sign control system. Quick Guide Book of M-series control card take M10 controller for example

Shanghai Xixun-----professional supplier of Led sign control system. Quick Guide Book of M-series control card take M10 controller for example Quick Guide Book of M-series control card ------take M10 controller for example 1 1. Hardware connection as show in image 1-1 in below: a. Connect M10 controller with HUB via 50pin cable, make sure the

More information

Bent Pins Inside the Sample Chamber

Bent Pins Inside the Sample Chamber PPMS Service Note 1070-311 Bent Pins Inside the Sample Chamber It has come to the attention of Quantum Design that there has been an increase in the instances of sample chambers with bent pins. We would

More information

Information Technology Help Desk George Washington Hall B College Avenue Fredericksburg, VA 22401

Information Technology Help Desk George Washington Hall B College Avenue Fredericksburg, VA 22401 Crestron TO START THE SYSTEM: Touch the screen on the small black Crestron box located on the lectern s desktop, this will wake up the system. On the system s start up screen, touch the Touch Here To Start

More information

DIMATIX Printer Tutorial and Operating Instructions

DIMATIX Printer Tutorial and Operating Instructions DIMATIX Printer Tutorial and Operating Instructions The Operating Principle: The DIMATIX printer is suited to use unusual fluids such as protein solutions or conductive polymers as ink materials. These

More information

The Dektak XT is a 2D contact profilometer used for step height, pitch and surface roughness

The Dektak XT is a 2D contact profilometer used for step height, pitch and surface roughness Dektak XT 2D Profilometer Operation Manual The Dektak XT is a 2D contact profilometer used for step height, pitch and surface roughness measurements. Vision 64 application software controls the system

More information

Visitron Spinning Disk Unit 1

Visitron Spinning Disk Unit 1 Startup Procedure and Notes (Last update: Sept. 2014) 6 4 2 5 1 3 7 8 9 1. Switch on the fluorescence illumination for widefield viewing (HXP lamp) 2. Switch on power-bar on the left side of the shelf

More information

YELLOW JACKET P TITAN User Manual UPC# 40860

YELLOW JACKET P TITAN User Manual UPC# 40860 YELLOW JACKET P51-860 TITAN User Manual UPC# 40860 Note: These instructions do not cover the manifold attached to the instrument. For instructions on use of the TITAN 4-Valve manifold, please visit: http://yellowjacket.com/wpcontent/uploads/2015/01/500716_rev.f-1.pdf

More information

Operating Instructions

Operating Instructions Operating Instructions Measurement of Liquid Viscosity The viscosity is measured using the Brookfield Viscometer (see Figure 1a) (a) (b) Figure 1. (a) Top view of the Brookfield Viscometer. The viscosity

More information

Equipment Overview: Safety:

Equipment Overview: Safety: PerkinElmer Series 200 HPLC: Operating Procedure 12/2/11 R.C Equipment Overview: This document outlines the basic procedure for using the Sedex light scattering detector and the PE series 200 HPLC system

More information

GSC300 CONFIGURATOR SOFTWARE INTERFACE. Installation and User Manual for the GSC300 Configurator PC Software Interface

GSC300 CONFIGURATOR SOFTWARE INTERFACE. Installation and User Manual for the GSC300 Configurator PC Software Interface GSC300 CONFIGURATOR SOFTWARE INTERFACE Installation and User Manual for the GSC300 Configurator PC Software Interface File: ConfiguratorRev1.5.2.doc Feb.21, 2006 2 READ MANUAL BEFORE INSTALLING UNIT END-USER

More information

REV 1. Quick Start Guide. MBE Control Solutions

REV 1. Quick Start Guide. MBE Control Solutions Quick Start Guide MBE Control Solutions AMBER Setup AMBER is usually sold preinstalled on a Windows 7-64 Workstation. A Self installation guide and installation software is available from MBE Control Solutions:

More information

Influence of Geometrical Configuration of Cantilever Structure on Sensitivity of MEMS Resonant Sensors

Influence of Geometrical Configuration of Cantilever Structure on Sensitivity of MEMS Resonant Sensors Influence of Geometrical Configuration of Cantilever Structure on Sensitivity of MEMS Resonant Sensors Georgeta Ionascu 1, Adriana Sandu 2, Elena Manea 3, Lucian Bogatu 4 1 Professor, Mechatronics & Precision

More information

SI-150-GIST. DART Thermo Quantiva, Endura, and Fusion VAPUR Interface User Manual. IonSense Inc. 999 Broadway Suite 404 Saugus, MA 01906

SI-150-GIST. DART Thermo Quantiva, Endura, and Fusion VAPUR Interface User Manual. IonSense Inc. 999 Broadway Suite 404 Saugus, MA 01906 SI-150-GIST DART Thermo Quantiva, Endura, and Fusion VAPUR Interface User Manual Manual Revision 1 Document # 7.5.122 IonSense Inc. 999 Broadway Suite 404 Saugus, MA 01906 Copyright 2005-2015 by IonSense

More information

Panalytical XPert Powder X-Ray Diffraction SOP

Panalytical XPert Powder X-Ray Diffraction SOP Revision 1 2/06/2018 Panalytical XPert Powder SOP Page 1 of 11 Panalytical XPert Powder X-Ray Diffraction SOP NOTE: Latest revisions are in Blue. Table of Contents 1.0 Safety 2.0 Training 3.0 Sample Preparation

More information

Olympus cellsens Live/TIRF Unit

Olympus cellsens Live/TIRF Unit Version 4 (July 2017) Startup (Widefield Mode) Switch on power strips #1-4 (behind the monitor/lasers) in the order as they are labeled. Switch on #2 (TIRF lasers) only if you need one or more lasers for

More information

VLSIresearch - Critical Subsystems and Components Service Overview

VLSIresearch - Critical Subsystems and Components Service Overview VLSIresearch - Critical Subsystems and Components Service Overview The market for Critical Subsystems and Components for use in the semiconductor and related manufacturing equipment has emerged to become

More information

PROGRAMMING INSTRUCTIONS

PROGRAMMING INSTRUCTIONS PROGRAMMING INSTRUCTIONS SCROLL UP ARROW DOWN ARROW Display Numbers and letters on display in quotation marks are flashing characters. Owners Instructions: The Display should look as follows before you

More information

Raith e_line Electron Beam Lithography

Raith e_line Electron Beam Lithography Raith e_line Electron Beam Lithography Standard Operating Procedure 1 (For an un-patterned sample) Revision: 7.0 Last Updated: Feb.18/2015, Revised by Mohamad Rezaei Overview This document will provide

More information

KINETICS CALCS AND GRAPHS INSTRUCTIONS

KINETICS CALCS AND GRAPHS INSTRUCTIONS KINETICS CALCS AND GRAPHS INSTRUCTIONS 1. Open a new Excel or Google Sheets document. I will be using Google Sheets for this tutorial, but Excel is nearly the same. 2. Enter headings across the top as

More information

NANOSPEC 4150 STANDARD OPERATING PROCEDURES

NANOSPEC 4150 STANDARD OPERATING PROCEDURES NANOSPEC 4150 STANDARD OPERATING PROCEDURES Version: 1.0 JAN 2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center TABLE OF CONTENTS 1.0 INTRODUCTION.. 3 2.0 HARDWARE....... 3 3.0 OPERATING

More information

Mesa Met Lab Series Bi-Directional Communications Protocol

Mesa Met Lab Series Bi-Directional Communications Protocol Mesa Met Lab Series Bi-Directional Communications Protocol To enable Mesa customers to integrate their Met Lab Series primary piston provers with their unique applications, our Met Lab systems come standard

More information

Flexus 2320 SOP Revision 1 01/25/16 Page 1 of 7. Flexus 2320 SOP

Flexus 2320 SOP Revision 1 01/25/16 Page 1 of 7. Flexus 2320 SOP Page 1 of 7 Flexus 2320 SOP The Flexus 2320 determines wafer curvature by measuring the angle of deflection of a laser beam off the surface 6, 4, 3 and 2 diameter substrates. Film stress is determined

More information

CPC. Composite Processing Control For Windows. Software Operations Manual. Copyright 2009, ASC Process Systems Documentation revision: 2/16/2009

CPC. Composite Processing Control For Windows. Software Operations Manual. Copyright 2009, ASC Process Systems Documentation revision: 2/16/2009 CPC Composite Processing Control For Windows Software Operations Manual Copyright 2009, ASC Process Systems Documentation revision: 2/16/2009 CPC Operations Manual Index Page #1-1 1. STARTING CPC...1-1

More information

Quick Reference Guide

Quick Reference Guide Quick Reference Guide Envizio Pro & Envizio Pro II Overview The Envizio Pro and Envizio Pro II are all-in-one field computers capable of providing product control for up to four products (with required

More information

Label Printer QL-810W/820NWB. Quick Setup Guide (English)

Label Printer QL-810W/820NWB. Quick Setup Guide (English) Label Printer QL-810W/820NWB Quick Setup Guide (English) Label Printer QL-810W/820NWB Printed in China D00RPG001 Quick Setup Guide (English) Available Manuals Product Safety Guide This guide provides safety

More information

Rack Installation Instructions

Rack Installation Instructions Rack Installation Instructions Review the documentation that comes with your rack cabinet for safety and cabling information. Before installing your server in a rack cabinet, review the following guidelines:

More information

Toucan LT board printer

Toucan LT board printer Toucan LT board printer Setup and Operating instructions Unpack the Toucan LT board printer as you would any Toucan LT. Follow all cautions associated with installing a standard Toucan LT. Change in wash

More information

NR-TEC S2S SURFACE TO SANDFACE DATA LOGGER

NR-TEC S2S SURFACE TO SANDFACE DATA LOGGER NR-TEC S2S SURFACE TO SANDFACE DATA LOGGER INTRODUCTION RIGGING IN / OUT CONNECTING TO DEVICE TAKING A SINGLE SHOT BASIC OPERATIONS GUIDE SCHEDULING VIEW A SHOT SCHEDULE UPLOAD A NEW SHOT SCHEDULE ERASING

More information

Operation Manual. for the. Data Logging Software. Version 7.1. (Isoft.xls)

Operation Manual. for the. Data Logging Software. Version 7.1. (Isoft.xls) for the Data Logging Software Version 7.1 (Isoft.xls) TetraTec Instruments GmbH 1 GENERAL HINTS 1.1 Typographical Conventions Displayment Means marks a work procedure, which you must implement references

More information

Quick Reference Guide

Quick Reference Guide Quick Reference Guide Viper Pro Software Version 3.10 Viper Pro Do s and Don ts Do have the Viper Pro serial number and firmware revision available when calling for technical assistance. It is best if

More information

3 Trial of Pro-Server EX

3 Trial of Pro-Server EX 3 Trial of Pro-Server EX 3.1 Starting 'Pro-Studio EX'...3-2 3.2 Trial of New Form...3-10 3.3 Trial of Recipe Function...3-40 3.4 Trial of Logging Funtion...3-53 3.5 Trial of Send Mail Function...3-67 3-1

More information

ION BEAM MILLING SYSTEM FOR TEM, SEM AND LM PREPARATION. Leica EM RES102

ION BEAM MILLING SYSTEM FOR TEM, SEM AND LM PREPARATION. Leica EM RES102 ION BEAM MILLING SYSTEM FOR TEM, SEM AND LM PREPARATION Leica EM RES102 ION BEAM MILLING In recent years, ion milling has been developed into the most applicable method of sample preparation for the analysis

More information

Thermal Diffusivity Instruments DLF-2 with EM-1600

Thermal Diffusivity Instruments DLF-2 with EM-1600 Thermal Diffusivity Instruments DLF-2 with EM-1600 Site Preparation Guide Revision A Issued February 2017 Table of Contents Table of Contents... 1 Ideal Setup... 2-3 System Components... 3 Instrument Measurements...

More information

SILENCING AN ALARM. When the alarm bell or siren is sounding, enter your user code or present your keyfob to your keypad.

SILENCING AN ALARM. When the alarm bell or siren is sounding, enter your user code or present your keyfob to your keypad. S Y S T E M U S E R G U I D E SILENCING AN ALARM When the alarm bell or siren is sounding, enter your user code or present your keyfob to your keypad. IS THIS A FALSE ALARM? YES NO displays. REAL ALARM

More information

2016 SFS Chemical Safety, Inc. All Rights Reserved Page 1

2016 SFS Chemical Safety, Inc. All Rights Reserved Page 1 New Users, consult the Navigation section first MANIFEST The Manifest form contains all open and closed Uniform Hazardous Waste Manifests of the system. A manifest is the shipping document required for

More information

Integrated System for PAGE and Blotting

Integrated System for PAGE and Blotting Mini-V 8 10 Integrated System for PAGE and Blotting Mini-V 8 10 Gel Electrophoresis System Gel size 8.7 cm x 7.2 cm (W x L) Patented drop-in wedge automatically positions gels without the use of gaskets

More information

Twickenham. Scientific Instruments. The Transfer Control Module (TCM) Instruction manual, release 1.1

Twickenham. Scientific Instruments. The Transfer Control Module (TCM) Instruction manual, release 1.1 .. Twickenham TSI Scientific Instruments The Transfer Control Module (TCM) Instruction manual, release 1.1 Twickenham Scientific Instruments Ltd. Registered in England No 2991094 Registered Office: 61d

More information

Raman Lab - Laser 1 operation

Raman Lab - Laser 1 operation Raman Lab - Laser 1 operation Cover limit switches cooling water supply optimate panel light control laser remote control sample stage Remote Shutter Control laser power supply WARNING! Never ever try

More information

Wagner Smart Screen Service/Troubleshooting Screens

Wagner Smart Screen Service/Troubleshooting Screens Section 7-2 Wagner Smart Screen Service/Troubleshooting Screens Menu Button PIN code login When you fi rst click on the Menu button, you will be prompted for a PIN code. Enter 0070 and click the check

More information

Operating Manual. Version 7.3.4

Operating Manual. Version 7.3.4 Operating Manual Version 7.3.4 March 2011 Index Page 1. Installation instructions 4 1.1 System requirements 4 1.2 Installation procedure 4 1.3 Meitav-tec USB adaptor installation 7 2. Open Maxinet and

More information

LECO (TruSpec CN) - Quick Reference Guide

LECO (TruSpec CN) - Quick Reference Guide LECO (TruSpec CN) - Quick Reference Guide Logging On (if necessary) Coming soon Items to check prior to beginning analysis Gases Compressed Air should read 40psi. Helium and Oxygen should read 35psi on

More information

SLEE Tissue Stainer MAS. Instruction Manual. IMPORTANT: Read this manual before starting the instrument. Rev. 01 ( )

SLEE Tissue Stainer MAS. Instruction Manual. IMPORTANT: Read this manual before starting the instrument. Rev. 01 ( ) SLEE Tissue Stainer MAS Instruction Manual IMPORTANT: Read this manual before starting the instrument Rev. 01 (05.2007) Copyright: Duplication of this manual by any means for any purpose without written

More information

Eye-Pal Vision Quick Start Rev. A

Eye-Pal Vision Quick Start Rev. A Eye-Pal Vision Quick Start 440793-001 Rev. A Setting Up the Unit 1. Place the Eye-Pal Vision on a flat surface with the front panel and buttons, facing toward you. 2. Unfold the camera. Pull the camera

More information

MagicSoft Version 3.02

MagicSoft Version 3.02 MagicSoft Version 3.02 Chapter 1 Introduction MagicSoft is designed to be intuitive and easy to navigate. You ll never find yourself buried deep in a menu without quick access to the screens and tools

More information

Owner s Manual. Digital Player Addendum. For Heat Siphon Swimming Pool Heat Pumps

Owner s Manual. Digital Player Addendum. For Heat Siphon Swimming Pool Heat Pumps Made in Latrobe Since 1983 Pennsylvania U.S.A. Owner s Manual Digital Player Addendum For Heat Siphon Swimming Pool Heat Pumps Heating Only Models: Z250HP, Z375HP, Z575HP & Z700HP Z250HP50, Z375HP50, Z575HP50

More information