FILMETRICS F20 STANDARD OPERATION PROCEDURE
|
|
- Magnus Elliott
- 6 years ago
- Views:
Transcription
1 Arizona State University NanoFab FILMETRICS F20 STANDARD OPERATION PROCEDURE Rev C
2 Table of Contents Contents Table of Contents Purpose / Scope Reference Documents Equipment / Supplies / Material Safety Set Up Procedures Operation Procedures Process Data Revision History...5 Page 1
3 1. Purpose / Scope The Filmetrics F20 uses spectroscopic reflectometry to measure film thickness. Single layer or multilayer films on substrates can be measured and analyzed. Typically, the F20 will measure thickness values between 150A and 100um. The large spot size is about 1.5 mm in diameter. 2. Reference Documents Information about this technique is available on the vendor s web site, 3. Equipment / Supplies / Material Reference Si wafer Reference SiO2/Si wafer Tweezers to handle wafers 4. Safety No unique safety requirements 5. Set Up Procedures 1. Turn on the light switch is on the front of system. Allow the light to warm-up for 5-10 minutes. 2. Log into the computer using your ASURITE or ASUAD user name and password. 3. Open the software FILMeasure by double clicking on the icon. 4. The graph will show the reflectance versus wavelength; Initial set-up and model building are done with buttons in the right column. 5. Baseline Reference Measurement. a. Put a clean Si sample, face up, above the mark on the sample stage. i. Select [Baseline] right side and a dialog box opens entitled Take Reflectance Reference ii. Step 1: Ensure Reflectance Material box displays [Si], Check box [X] next to Autoscale Integration iii. Step 2: Take Reference Reading, click [OK]; data is displayed on graph. b. The next dialog box is entitled Take Reflectance Dark i. Tilt the Si reference wafer to about 45 degrees, so the beam is reflected away from the fiber optic detector. ii. Click [OK] and the results are displayed on the screen; generally it is nearly a straight line with occasional small peaks. c. Next, place a reference SiO2/Si wafer on the sample stage. d. In the [Structure box ], use the pull-down arrow and review the list of recent film/substrate combinations; select SiO2/Si if available; otherwise select [Edit structure]. e. Select the [Edit Structure] box, The screen displays the optical model: i. Air, ii. Layer 1, [SiO2 ], [Thickness estimate], [ x ] d [ ] n, [ ] k iii. Substrate [ Si ] Page 2
4 f. Enter the thickness estimate (18000A for this test wafer), check the box next to d (to fit for thickness). Select [Measure]. g. The calculated (red) and measured (blue) reflectance versus wavelength plots should have the Peaks and Valleys aligned. If the thickness and GOF values (box in right column) are close to the expected thickness and GOF is high ( ), the system is working properly. Otherwise, contact Nanofab staff. 6. Operation Procedures 1. Put your wafer on the sample stage under the fiber optic beam. 2. Select the [Film Structure ] button and find a recent film/substrate combination in the list that is has the same layers as your sample. If there is no match then [cancel] and go to step Otherwise, Open the [Edit Structure Box] and the model window opens. It will show the substrate, Film 1, and Medium Air. Note the insert layer boxes on the left side and pulldown arrows for each layer ID. 4. First, select the substrate; the pull down arrow opens the data base of substrate material; select the substrate material for your sample. 5. Select [Film 1]; use the down arrow to obtain a list of material categories: a. Air b. Photo Resist (includes common resists AZ3312 and AZ1518) c. Dielectrics d. Semiconductors e. Metals (must be semi-transparent metal films) f. Others (CoSi2, Fe2O3, ITO, NiP, PZT, TiN, TiN-Palik, ZnSe) Within each category there is a list of materials; choose the material for the 1 st layer. 6. Enter the approximate thickness for your Film 1 and check the Fit box [X] next to d. 7. Add films and select materials using the insert arrows and pull-down arrows. Enter the approximate thickness for each layer and check the Fit box next to d if the software will fit for the thickness of this layer. 8. Select the [Measure] button and evaluate the thickness value(s) and MSE parameters. Also check the plots to see if the measured (Blue) and calculated (Red) reflectances match. If the match is not good, re-do the optical model and/or the initial thickness estimate. 9. Position the curser over the curves on the graph and read out the corresponding wavelength and reflectance values in the boxes at the top of the graph. Save the data if desired. 10. Finished? Close the software; Generally, do not save the program unless it s unique to your project. 7. Process Data There are several options to modify the analysis of the reflectance data. 1. The [Edit Structure] menu contains the optical model. In addition, a tab labeled Solver Options has 2 possibilities, a. analyzer data with displayed data ; b. Fixed range From [xxx] nm to [yyy] nm, allowing fitting over limited wavelength ranges; this can be useful for PR films with large k-values at short wavelengths. Try an initial fit at longer wavelengths such as 500nm to 800nm or from 550nm to 800nm. Page 3
5 2. On the right column is a button labeled [Show Statistics]; select this box to open a window containing: a. Mean b. Medium c. Std. Dev. d. Min e. Max f. Range Followed by a table of values with headings, Measure #, Layer 1 d(a), GOF Commands at the bottom are [save], [copy] [clear] or select rows to [delete] If you select [clear], and then take a series of measurements, the statistics for your sample will be displayed in this box. 3. Another button in the right column is labeled [Show Table]; select this box opens a data table with headings of Wavelength and Reflectance. Commands include Select all, Copy selected data, Save data This allows you to save the table of data to your shared drive; text data can be open using Excel. Page 4
6 8. Revision History Effective Date Originator DESCRIPTION OF REVISION Issue 5/5/09 Name Initial Release A 4/21/15 staff Rev B B Wayne Paulson Rev C update & fit new SOP format C D E F G H Page 5
NRF Filmetrics F40 SOP Revision /02/12 Page 1 of 5. Filmetrics F40 SOP
Page 1 of 5 Filmetrics F40 SOP The F40 is a reflectometer used to measure the thickness and optical constants of thin, transparent/semi-transparent films. By fitting it to a microscope the tool can have
More informationSpectroscopic Ellipsometer --- J. A. Woollam alpha-se
Spectroscopic Ellipsometer --- J. A. Woollam alpha-se Introduction Figure 1: J. A. Woollam alpha-se spectroscopic ellipsometer An ellipsometer measures the change in polarization as light reflects or transmits
More informationFilmetrics F40 UV Thin Film Measurement System Standard Operating Procedure
Filmetrics F40 UV Thin Film Measurement System Standard Operating Procedure NYU Tandon School of Engineering Nanofabrication Facility Contents 1 Startup 2 Baseline and reference check 2.1 Baseline setting
More informationOperation Manual of J.A. Woollam Ellipsometer
Operation Manual of J.A. Woollam Ellipsometer 1) Press both the Lamp power and lamp Ignition buttons to turn them on (The buttons are located on the front panel of the bottom control box). 2) Put your
More informationOperating Instructions
Operating Instructions Measurement of Liquid Viscosity The viscosity is measured using the Brookfield Viscometer (see Figure 1a) (a) (b) Figure 1. (a) Top view of the Brookfield Viscometer. The viscosity
More informationNRF Ellipsometer SOP Revision /19/15 Page 1 of 14. Ellipsometer SOP
Page 1 of 14 Ellipsometer SOP The J. A. Woollam M88 is a spectroscopic ellipsometer used to measure film thickness and optical constants of transparent/semi-transparent thin films. It uses a Xenon arc
More information5.2.2 Pins on the alignment detector are fragile. Use caution when inserting and removing the detector. Report any broken pins to staff.
Title: Woollam Vase Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Woollam VASE. All users are expected to have read and understood
More informationNANOSPEC 4150 STANDARD OPERATING PROCEDURES
NANOSPEC 4150 STANDARD OPERATING PROCEDURES Version: 1.0 JAN 2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center TABLE OF CONTENTS 1.0 INTRODUCTION.. 3 2.0 HARDWARE....... 3 3.0 OPERATING
More informationStandard Operating Manual
Standard Operating Manual J.A. Woolam M-2000V Spectroscopic Ellipsometer Version 1.0 Page 1 of 13 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Features 2.3 Sample
More information1. Scope 1.1 This document provides the operating procedure for plasma enhanced chemical vapor deposition with the Oxford 80+.
Oxford 80+ PECVD SOP Oxford 80+ PECVD SOP Page 1 of 7 1. Scope 1.1 This document provides the operating procedure for plasma enhanced chemical vapor deposition with the Oxford 80+. 2. Table of Contents
More informationUniversity of Minnesota Nano Fabrication Center Standard Operating Procedure
Equipment Name: Film-Sense Coral Name: film-sense Revision Number: 1 Model: FS-1 Revisionist: T. Whipple Location: Bay 1 Date: 5/16/2017 1 Description The Film Sense FS-1 Multi-Wavelength Ellipsometer
More informationMDC CV SYSTEM OPERATION PROCEDURE
MDC CV SYSTEM OPERATION PROCEDURE Rev B Table of Contents Contents Table of Contents... 1 1. Purpose / Scope... 2 2. Reference Documents... 2 3. Operation Manual... 2 4. Equipment / Supplies / Material...
More informationFlexus 2320 SOP Revision 1 01/25/16 Page 1 of 7. Flexus 2320 SOP
Page 1 of 7 Flexus 2320 SOP The Flexus 2320 determines wafer curvature by measuring the angle of deflection of a laser beam off the surface 6, 4, 3 and 2 diameter substrates. Film stress is determined
More informationSpecification of Thin Film Thickness Measuring Equipment
of Thin Film Thickness Measuring Equipment A. Application: Major application is for solar cell thin film thickness measurement for films such as CIGS, CdS, ITO, ZnO etc. deposited on substrates such as
More informationOCEAN OPTICS Chem 2000 Spectrophotometer
OCEAN OPTICS Chem 2000 Spectrophotometer prepared for: Department of Chemistry and Biochemistry New Mexico State University Las Cruces, NM 88003 last updated May 2003 R.. Cowley TABLE OF CONTENTS I. Introduction...2
More informationXetch SOP Page 1 of 7 Revision
Xetch SOP Page 1 of 7 Xetch System SOP 1. Scope 1.1 This document provides procedures and process information for etching silicon wafers with the Xetch xenon difluoride etching system. 2. Table of Contents
More informationWoollam M2000 Operation Manual
Woollam M2000 Operation Manual The Woollam M2000 is a spectroscopic ellipsometer used to characterize optically transparent films. The system has the Near IR upgrade that covers 700 wavelengths from 193nm
More informationOxford ICP 2-step DRIE SOP
Oxford ICP 2-step DRIE SOP Oxford ICP 2-step DRIE SOP Page 1 of 13 1. Scope 1.1 This document provides operating procedures for the Oxford 100 ICP 2-step DRIE. 2. Table of Contents 1. Scope... 1 2. Table
More information` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE
` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE Version: 1.0 April-2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research Center (NRC) TABLE OF CONTENTS 1. Introduction...2 1.1
More informationOxford 80+ RIE SOP. Oxford 80+ RIE SOP Page 1 of 6 Revision
Oxford 80+ RIE SOP Oxford 80+ RIE SOP Page 1 of 6 1. Scope 1.1 This document provides the operating procedure for the Oxford 80+ reactive ion etch. 2. Table of Contents 1. Scope... 1 2. Table of Contents...
More informationR I T. Title: Rudolf Ellipsometer Auto-EL IV. Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date:12/07/17 1 SCOPE
Title: Rudolf Ellipsometer Auto-EL IV Fabrication Laboratory Revision: C Rev Date:12/07/17 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the
More informationTitle: Heidelberg DWL66+ Semiconductor & Microsystems Fabrication Laboratory Revision: B Rev Date: 05/03/2017
Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Heidelberg DWL66+. All users are expected to have read and understood this document.
More informationR I T Title: Lam 4600
Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Lam 4600. All users are expected to have read and understood this document. It
More information2.2. Facilities Requirements
2.2. Facilities Requirements Facilities requirements for the alpha-se system are listed in Table 2-1 and the system dimensions are given in Figure 2-1. As shown in Figure 2-2, the preferred clear work
More informationOperating Procedure for Horiba Raman Microscope
Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch
More informationVEECO FPP Point Probe Operating Manual
VEECO FPP-5000 4-Point Probe Operating Manual Version: 1.0 May 2013 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research and Teaching Facility TABLE OF CONTENTS 1. Introduction....2 1.1 Scope of Work....2
More informationMetricon Model 2010/M Prism Coupler Bulk Material, Thick Film Index/Birefringence/ Thickness Measurement System Standard Operating Procedure
Metricon Model 2010/M Prism Coupler Bulk Material, Thick Film Index/Birefringence/ Thickness Measurement System Standard Operating Procedure Version: 1.0 January 2017 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology
More informationStandard Operating Procedure FS-1 Film Sense Ellipsometer (spectroscopic)
Standard Operating Procedure FS-1 Film Sense Ellipsometer (spectroscopic) The Film Sense FS-1 is a spectroscopic ellipsometer dedicated to advanced and rapid characterization of thin film structures for
More informationLEXT 3D Measuring LASER Microscope
LEXT 3D Measuring LASER Microscope Warning: This instrument may only be operated by those who have been trained by AAF staff and have read and signed the AAF laboratory policies. A) STARTUP 1. Computer
More informationHIGH TEMPERATURE PROBE STATION STANDARD OPERATION PROCEDURE
Arizona State University NanoFab HIGH TEMPERATURE PROBE STATION STANDARD OPERATION PROCEDURE Rev A Table of Contents Contents Table of Contents... 1 1. Purpose / Scope... 2 2. Reference Documents... 2
More informationLaser Applications for Photovoltaics Crystalline and Thin Film Technologies
LASERS & MATERIAL PROCESSING I OPTICAL SYSTEMS I INDUSTRIAL METROLOGY I TRAFFIC SOLUTIONS I DEFENSE & CIVIL SYSTEMS Laser Applications for Photovoltaics Crystalline and Thin Film Technologies Back contact
More informationHeidelberg Pattern Generator SOP
Heidelberg Pattern Generator SOP Page 1 of 15 Heidelberg Pattern Generator SOP 1. Scope 1.1 This document provides the operating procedures for the Heidelberg Pattern Generator with Version 3.12.5 software.
More informationGuideStar II Customer Presentation. February 2012
GuideStar II Customer Presentation February 2012 Outline Product Overview Description Applications and Features Specifications Technical Details Picomotor Mirror Mounts 8784 Cameras Alignment Layout and
More informationChapter 36. Diffraction. Dr. Armen Kocharian
Chapter 36 Diffraction Dr. Armen Kocharian Diffraction Light of wavelength comparable to or larger than the width of a slit spreads out in all forward directions upon passing through the slit This phenomena
More informationSeized Drugs Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division
Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division SHIMADZU ULTRAVIOLET/VISIBLE SPECTROPHOTOMETER (UV/VIS) Instrument Shimadzu UV-2401 UV/VIS Spectrophotometer (Serial Number
More informationABM's High Resolution Mask Aligner Features:
ABM's High Resolution Mask Aligner is a very versatile instrument with interchangeable light sources which allow Near-UV (405-365 nm) as well as Mid- and Deep-UV (254 nm, 220 nm) exposures in proximity
More informationIncreasing laser processing efficiency using multibeam and tailored beam profiles
Increasing laser processing efficiency using multibeam and tailored beam profiles Ulrich Rädel TOPAG Lasertechnik GmbH, Darmstadt Overview Presentation of company Topag Increasing processing efficiency
More informationCharacterization of stratified media using high-resolution thin film measurement techniques
Characterization of stratified media using high-resolution thin film measurement techniques Alberto Aguerri Sensofar-Tech, S.L. Crt. N150 Km14.5 IPCT Mòdul TR-20, 08227 Terrassa (Barcelona), Spain E-mail:
More informationFC Lens series for Cree XLamp 7090 XR and XR-E LEDs
FC Lens series for Cree XLamp 7090 XR and XR-E LEDs High efficiency 4 beams available Easy assembly The FC lens offers low-profile lenses specifically designed for the XLamp 7090XR and XR-E LEDs (1) from
More informationRaith e_line Electron Beam Lithography
Raith e_line Electron Beam Lithography Standard Operating Procedure 1 (For an un-patterned sample) Revision: 7.0 Last Updated: Feb.18/2015, Revised by Mohamad Rezaei Overview This document will provide
More informationReflection and Refraction
Reflection and Refraction INTRODUCTION Geometric optics is one of the oldest branches of physics, dealing with the laws of refraction and reflection. The law of reflection 1 was known to the ancient Greeks
More informationHeidelberg MLA-150 Standard Operating Procedure
Heidelberg MLA-150 Standard Operating Procedure CORAL Name: Model: Location: Purpose: Author: MLA-150 Heidelberg MLA150 Maskless Aligner TRL Photo-Au Room Direct-Write Lithography Heidelberg Instruments
More informationOptical Topography Measurement of Patterned Wafers
Optical Topography Measurement of Patterned Wafers Xavier Colonna de Lega and Peter de Groot Zygo Corporation, Laurel Brook Road, Middlefield CT 6455, USA xcolonna@zygo.com Abstract. We model the measurement
More informationksa MOS Ultra-Scan Performance Test Data
ksa MOS Ultra-Scan Performance Test Data Introduction: ksa MOS Ultra Scan 200mm Patterned Silicon Wafers The ksa MOS Ultra Scan is a flexible, highresolution scanning curvature and tilt-measurement system.
More informationProduct Specifications
Product Specifications The ksa RateRat Pro is a turnkey, real-time, in-situ optical reflectance probe designed for deposition monitoring of semi-absorbent thin films. The RateRat Pro measures deposition
More informationUsing Graphical Analysis This exercise will familiarize you with using Graphical Analysis in Beer s Law and kinetics.
Using Graphical Analysis 3.1.1 This exercise will familiarize you with using Graphical Analysis in Beer s Law and kinetics. Beer s Law In this exercise you will plot a calibration curve of absorbance versus
More informationEquipment Standard Operating Procedure Kimberly Appel
Date Created: June 26, 2004 Date Modified: The Flexus 2320 Equipment Standard Operating Procedure Kimberly Appel 1. Purpose 1.1. The Flexus Thin Film Stress Measuring Apparatus (TFSMA) measures the changes
More informationLaser Applications for Photovoltaics Crystalline and Thin Film Technologies
LASERS & MATERIAL PROCESSING I OPTICAL SYSTEMS I INDUSTRIAL METROLOGY I TRAFFIC SOLUTIONS I DEFENSE & CIVIL SYSTEMS Laser Applications for Photovoltaics Crystalline and Thin Film Technologies Back contact
More informationPY106 Class31. Index of refraction. Refraction. Index of refraction. Sample values of n. Rays and wavefronts. index of refraction: n v.
Refraction Index of refraction When an EM wave travels in a vacuum, its speed is: c = 3.00 x 10 8 m/s. In any other medium, light generally travels at a slower speed. The speed of light v in a material
More informationSemiconSoft, Inc. Optical Metrology Company. MProbe: Measurement Guide Thin-films measurement using Reflectance or/and Transmittance spectroscopy
MProbe: Measurement Guide Thin-films measurement using Reflectance or/and Transmittance spectroscopy It is easy to be a measurement expert with MProbe The Purpose This document describes the measurement
More informationPCAMS for Windows NT. Trend Program Documentation
PCAMS for Windows NT Trend Program Documentation 1 Table of Contents TREND OVERVIEW... 3 STARTING TREND... 4 TIME ENTRY... 5 LINE COLOR... 6 PRINTING... 8 GRAPH FUNCTIONS... 9 PLOTTING DATA... 10 Precipitator
More informationFORMATION AND CHARACTERIZATION OF POLYIMIDE TENTS FOR NIF-TARGET CAPSULES. Jordan Alexander, Forbes Powell, Mark Hagen, Jeff Guckian, Bobby Warren
FORMATION AND CHARACTERIZATION OF POLYIMIDE TENTS FOR NIF-TARGET CAPSULES Jordan Alexander, Forbes Powell, Mark Hagen, Jeff Guckian, Bobby Warren Luxel Corporation, PO Box 515, Friday Harbor, WA 98250
More informationStandard Operating Procedure of Triboindenter (Hysitron TI 950)
Standard Operating Procedure of Triboindenter (Hysitron TI 950) I Sample Loading and Preparation DO NOT TOUCH the bottom of transducer and optical microscope. Always place the tall samples on the most
More informationUniversity of Minnesota Nano Fabrication Center Standard Operating Procedure
Equipment Name: University of Minnesota Nano Fabrication Center Coral Name: hs-scope Revision Number: 1.5 Model: HS200A Revisionist: M. Fisher Location: Bay 1 Date: 9/12/2013 1 Description The Hyphenated
More informationT-Solar Overview. * Patent-pending
T-Solar T-Solar Overview The T-Solar system combines our best photovoltaic measurement technology into a system designed specifically for measuring textured samples. Based on the established M-2000 rotating
More informationChapter 8: Physical Optics
Chapter 8: Physical Optics Whether light is a particle or a wave had puzzled physicists for centuries. In this chapter, we only analyze light as a wave using basic optical concepts such as interference
More informationTHICKNESS MEASUREMENTS USING PRISM COUPLING. Daniel J. Hahn 5th Year Microelectronic Engineering Student Rochester Institute of Technology ABSTRACT
THICKNESS MEASUREMENTS USING PRISM COUPLING Daniel J. Hahn 5th Year Microelectronic Engineering Student Rochester Institute of Technology ABSTRACT A HeNe laser and a 45-90 degree prism with an index of
More informationSuss Aligner SOP. Suss Aligner SOP Page 1 of 6 Revision Scope 1.1 This document provides the procedure for operating the Suss Aligner.
Suss Aligner SOP Page 1 of 6 Suss Aligner SOP 1. Scope 1.1 This document provides the procedure for operating the Suss Aligner. 2. Table of Contents 1. Scope... 1 2. Table of Contents... 1 Suss Aligner
More informationSurface Plasmon Resonance Simulate your reflectivity curve with WinSpall
Introduction tutorial #2 WinSpall is a software for the simulation of surface plasmon resonance curves based on the Fresnel formalism. Winspall is easy to use and lets you model reflection curves pretty
More informationTABLE OF CONTENTS INTRODUCTION...2
WINGARD PE 6.0 August 2010 TABLE OF CONTENTS TABLE OF CONTENTS...1 1. INTRODUCTION...2 2. PROGRAM OVERVIEW...3 2.1. PROGRAM MENUS... 4 2.2. MAIN TOOLBAR... 5 2.3. ANALYSIS... 7 2.4. OUTPUT... 10 2.5. DISPLAY
More informationD&S Technical Note 09-2 D&S A Proposed Correction to Reflectance Measurements of Profiled Surfaces. Introduction
Devices & Services Company 10290 Monroe Drive, Suite 202 - Dallas, Texas 75229 USA - Tel. 214-902-8337 - Fax 214-902-8303 Web: www.devicesandservices.com Email: sales@devicesandservices.com D&S Technical
More informationUniversity of Minnesota Nano Center Standard Operating Procedure
University of Minnesota Nano Center Standard Operating Procedure Equipment Name: Optical Particle Analyzer Model: Microtrac SIA Location: PAN 185 Badger Name: Not on Badger Revision Number: 1 Revisionist:
More informationTutorial: Package, analyze, and validate a sample lens system. December
Tutorial: Package, analyze, and validate a sample lens system December 2017 www.zemax.com/lmx/ LMxSales@Zemax.com LMxSupport@Zemax.com Introduction In this tutorial, you learn how to package, analyze,
More informationOxford DSiE Revision /25/17 Page 1 of 9. Oxford DSiE SOP
Page 1 of 9 Oxford DSiE SOP The PlasmaPro 100 is a Inductively coupled plasma device (ICP) configured to provide effective high aspect ratio etching of 4 inch silicon wafers, and pieces when a carrier
More informationFCP Lens Series for Cree XLamp 1 LEDs
FCP Lens Series for Cree XLamp 1 LEDs High efficiency design 5 beams patterns available Available as a lens alone for maximum design flexibility or in a holder for easy assembly The FCP lens offers low
More informationksa ICE - Integrated Control for Epitaxy
Introduction The k-space Integrated Control for Epitaxy system (ksa ICE) is a modular in-situ metrology tool designed for today s MOCVD reactors. It combines proven ksa MOS, ksa BandiT, and ksa RateRat
More informationSession 1B Transparent Materials
Session 1B Transparent Materials Andrew Martin UPenn, February 2014 2014 J.A. Woollam Co., Inc. www.jawoollam.com 1 Overview Transparent substrates & films Cauchy equation Common complexities Evaluating,
More informationFitting NMR peaks for N,N DMA
Fitting NMR peaks for N,N DMA Importing the FID file to your local system Any ftp program may be used to transfer the FID file from the NMR computer. The description below will take you through the process
More informationSample study by 3D optical profiler Contour Elite K for KTH university.
Sample study by 3D optical profiler Contour Elite K for KTH university Samuel.lesko@bruker.com Objectives Objectives Main goals for the visit consist of evaluating 3D optical profiler: Confirm capability
More informationLam AutoEtch 490 Plasma Etch System Operations and Maintenance Manual
Fabrication Laboratory Revision: E - 02/10/2009 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Lam AutoEtch 490 Plasma Etch System
More informationP6 Profilometer Operating Procedure
P6 Profilometer Operating Procedure I. Introduction The KLA-Tencor P-6 profiler is a highly sensitive surface profiler that measures step height. The key features of P-6 include: A vertical range of 326
More informationPerkinElmer Series 200 HPLC: Operating Procedure
PerkinElmer Series 200 HPLC: Operating Procedure Chem Eng 5503 Version 1.01 Last Update 8/9/07 This document outlines the basic operating procedure for the PE series 200 HPLC, UV/Vis and SEDEX Light Scattering
More informationDWL-66 Laser Lithography System Operating Procedure. Effective Date: 09/24/2012 Author(s): Jiong Hua Phone:
DWL-66 Laser Lithography System Operating Procedure Effective Date: 09/24/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Laser Lithography System, Optical
More informationTextbook Reference: Physics (Wilson, Buffa, Lou): Chapter 24
AP Physics-B Physical Optics Introduction: We have seen that the reflection and refraction of light can be understood in terms of both rays and wave fronts of light. Light rays are quite compatible with
More informationGrazing Angle 2 Theta Phase Analysis
Page 1 of 7 Grazing Angle 2 Theta Phase Analysis 1. Log into the User Log System on the SMIF web site Hardware Setup X-Ray Tube The line focus configuration of the x-ray tube is used. This is the default
More informationManufacturing Challenges for Lithography in the Textured Disc Paradigm. September 18 th, 2008 Babak Heidari
Manufacturing Challenges for Lithography in the Textured Disc Paradigm September 18 th, 2008 Babak Heidari Longitudinal Perpendicular Pattern media + HAMR 6,25 T/in 2 TDK: DTR 602 Gb/in 2 1 T/in 2 150
More informationIndustrial Example I Semiconductor Manufacturing Photolithography Can you tell me anything about this data!
Can you tell me anything about this data! 1 In Semiconductor Manufacturing the Photolithography process steps are very critical to ensure proper circuit and device performance. Without good CD (critical
More informationSPECTRAL IMAGING VIEWER SOFTWARE MANUAL
103 Quality Circle, Suite 215 Huntsville, AL 35806 Phone: (256) 704-3332 Fax: (256) 971-2073 E-Mail: info@axometrics.com Website: http://www.axometrics.com SPECTRAL IMAGING VIEWER SOFTWARE MANUAL 2012
More informationChemistry Instrumental Analysis Lecture 6. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 6 UV to IR Components of Optical Basic components of spectroscopic instruments: stable source of radiant energy transparent container to hold sample device
More informationMaintenance Package Measurement
Maintenance Package Measurement Contents Contents 1. Package measurement flow...1 2. Measurement procedures...3 2.1 Startup... 3 2.2 Hardware setup... 4 2.3 Setting Package measurement conditions... 7
More informationPowerMax-Pro kw Sensors
PowerMax-Pro kw Sensors 1W to 3 kw PowerMax-Pro laser detectors have been enhanced to enable multi-kw continuous power measurement of laser beams. PowerMax-Pro detectors, first introduced in 2014, are
More informationSwitchboard. Creating and Running a Navigation Form
Switchboard A Switchboard is a type of form that displays a menu of items that a user can click on to launch data entry forms, reports, queries and other actions in the database. A switchboard is typically
More informationStandard Operating Procedure for the Horiba FluroMax-4
Standard Operating Procedure for the Horiba FluroMax-4 Adapted from Horiba Operations Manual Created by Michael Delcau, Modified by Brian Lamp The Fluoromax is capable of making a variety of measurements.
More informationC. FEATURES OF SPA-4000
C. FEATURES OF SPA-4000 1. Key Features 2. Specification 3. Thermo-Optic Coefficient Analysis SPEC 4. Thermo-Optic Coefficient Analysis GRAPH 5. Configuration 11 C. 1. Key Features Propagation Loss measurement
More informationBasic configuration. 1 Basic configuration
Basic configuration 1 Basic configuration 1-1 Series Displacement Sensors.................... 22 1-2 Basic Operation Flow....................................... 23 1-3........... 25 1-1 Series Displacement
More informationACCURATE TEXTURE MEASUREMENTS ON THIN FILMS USING A POWDER X-RAY DIFFRACTOMETER
ACCURATE TEXTURE MEASUREMENTS ON THIN FILMS USING A POWDER X-RAY DIFFRACTOMETER MARK D. VAUDIN NIST, Gaithersburg, MD, USA. Abstract A fast and accurate method that uses a conventional powder x-ray diffractometer
More informationCrystal Quality Analysis Group
Crystal Quality Analysis Group Contents Contents 1. Overview...1 2. Measurement principles...3 2.1 Considerations related to orientation and diffraction conditions... 3 2.2 Rocking curve measurement...
More informationPRODUCT OVERVIEW. Rupert Perera President, EUV Tech
PRODUCT OVERVIEW Rupert Perera President, EUV Tech EUV TECH OVERVIEW Started in 1997, EUV Tech has pioneered the development of EUV metrology tools: EUV Reflectometer o Measures the reflectivity and uniformity
More informationratio of the volume under the 2D MTF of a lens to the volume under the 2D MTF of a diffraction limited
SUPPLEMENTARY FIGURES.9 Strehl ratio (a.u.).5 Singlet Doublet 2 Incident angle (degree) 3 Supplementary Figure. Strehl ratio of the singlet and doublet metasurface lenses. Strehl ratio is the ratio of
More informationOlympus cellsens TIRF Guide
Startup Power up the system as described in the "Startup & Environmental Control" guide, including power strip #2 ("TIRF"). ON For each laser line needed, turn the key on the control unit right and press
More informationRgBLase LLC P/N: FB 671 XXX Diode Pumped Solid State Laser Module
Key Features: 671nm output, TEM00 Fiber Coupled (optional) Remote control TTL Modulation (optional) ESD protection Plug & Play Low Noise (optional) Applications: Bio Technology Photo Finishing Semiconductor
More informationAgilent Cary Universal Measurement Spectrophotometer (UMS)
Agilent Cary Universal Measurement Spectrophotometer (UMS) See what you ve been missing Date: 13 th May 2013 TRAVIS BURT UV-VIS-NIR PRODUCT MANAGER AGILENT TECHNOLOGIES 1 Agenda Introducing the Cary 7000
More informationTutorial: Package, analyze, and validate a sample lens system. by Isis Peguero, Optomechanical Engineer. August 2016
Tutorial: Package, analyze, and validate a sample lens system by Isis Peguero, Optomechanical Engineer August 2016 www.zemax.com/lmx/ LMxSales@zemax.com LMxSupport@zemax.com In this tutorial, you will
More informationE3S-A. Built-in Amplifier Photoelectric Sensor (Medium Size) Ordering Information. Built-in Amplifier Photoelectric Sensors. Horizontal. 7 m.
Built-in Amplifier (Medium Size) ES-A CSM_ES-A_DS_E Be sure to read Safety Precautions on page 0. Ordering Information Built-in Amplifier s Red light Infrared light Sensing method Appearance Connection
More informationFour-Point Probe System
Manual version: 2.0.E Product code: T2001A2 Product version: 2.0 Software version: 1.1 enabling materials science ossila.com Contents 1. Safety... 3 1.1 Warning... 3 1.2 Caution... 3 2. Introduction...
More informationLecture 41: WED 29 APR
Physics 2102 Jonathan Dowling Lecture 41: WED 29 APR Ch. 36: Diffraction PHYS 2102-2 FINAL 5:30-7:30PM FRI 08 MAY COATES 143 1/2 ON NEW MATERIAL 1/2 ON OLD MATERIAL Old Formula Sheet: http://www.phys.lsu.edu/classes/
More informationPanalytical XPert Powder X-Ray Diffraction SOP
Revision 1 2/06/2018 Panalytical XPert Powder SOP Page 1 of 11 Panalytical XPert Powder X-Ray Diffraction SOP NOTE: Latest revisions are in Blue. Table of Contents 1.0 Safety 2.0 Training 3.0 Sample Preparation
More information3. Using TFCompanion. 3.1 Filmstack. Introduction Layer types and properties
3. Using TFCompanion. 3.1 Filmstack. Introduction. Filmstack is an optical model of the sample that is measured - it consists of a substrate, collection of layers and an ambient. There is no limitation
More informationLab 12 - Interference-Diffraction of Light Waves
Lab 12 - Interference-Diffraction of Light Waves Equipment and Safety: No special safety equipment is required for this lab. Do not look directly into the laser. Do not point the laser at other people.
More informationChromQuest 5.0 Quick Reference Guide
ChromQuest 5.0 Quick Reference Guide This guide contains an overview of the ChromQuest chromatography data system, with topics organized by workflow. For more information, refer to the ChromQuest User
More information