UV/VIS 8500 DOUBLE-BEAM SPECTROPHOTOMETER OPERATION MANUAL TECHCOMP

Size: px
Start display at page:

Download "UV/VIS 8500 DOUBLE-BEAM SPECTROPHOTOMETER OPERATION MANUAL TECHCOMP"

Transcription

1 UV/VIS 8500 DOUBLE-BEAM SPECTROPHOTOMETER OPERATION MANUAL TECHCOMP

2 Copy Right Techcomp Limited Version: 1.2 All rights reserved worldwide. Printed in HONG KONG. Every effort has been made to avoid errors in text and diagrams, however, TECHCOMP Limited assumes no responsibility for any errors which may appear in this publication. It is the policy of TECHCOMP Limited to improve products as new techniques and components become available. TECHCOMP Limited therefore reserves the right to change specifications at any time. We would appreciate any comments on this publication. TECHCOMP Limited PO Box 72919, Kowloon Central Post Office, Kowloon, Hong Kong. Tel: (852) Fax: (852) Tlx: TECOP HX URL:

3 Chapter 1 System Installation 1.1 Introduction 1.2 Unpacking and Inspection 1.3 Power Requirements 1.4 Environmental Requirements 1.5 Electrical Connection 1.6 Initial System Calibration 1.7 Printer Setup Table of Contents Chapter 2 Functional Description 2.1 Introduction 2.2 Main Functions Fixed Points Measurement Wavelength Scanning Time Scanning Spectrum Processing Function Automatic System Initialization Function 2.3 Instrument Specifications Chapter 3 Routine Operation 3.1 Introduction 3.2 Key-Panel 3.3 Dynamic Keyboard 3.4 Sample Compartment 3.5 Main Menu Chapter 4 Fixed Points Measurement 4.1 Introduction 4.2 Selecting the Fixed Points Function 4.3 Selecting Data Acquisition Mode 4.4 Single Wavelength Measurement 4.5 Multiwavelength Measurement 4.6 Regression Curve Setup 4.7 Sample Test 4.8 Save Files 4.9 Load Files Chapter 5 Wavelength Scanning 5.1 Introduction 5.2 Selecting Wavelength Scan Mode 5.3 Step by Step Operation Selecting Data Acquisition Mode Entering Wavelength Scan Setup Variables Setting the Lamp Switching Wavelength Position 5.4 Collecting a Spectrum in Wavelength Scan

4 5.5 Spectrum Processing Introduction Rescale Zoom Peaks & Valleys Threshold Trace Derivative Adding Spectra Spectra Subtraction Save a Spectrum Load a Spectrum Unload a Spectrum Chapter 6 Time Scanning 6.1 Introduction 6.2 Selecting Time Scan Mode 6.3 Step by Step Operation Selecting Data Acquisition Mode Entering Time Scan Setup Variables Collecting a Spectrum in Time Scan 6.4 Spectrum Processing Introduction Rescale Peaks & Valleys Threshold Trace Derivative Save a Spectrum Load a Spectrum Unload a Spectrum Chapter 7 Maintenance 7.1 Introduction 7.2 Internal Diagnostics and Automatic Adjustments Verifying Wavelength Accuracy and Reproducibility Confirming the Spectrum Bandwidth Confirming Baseline Flatness Chapter 8 Replacement Parts 8.1 Introduction 8.2 Spare Parts 8.3 Lamp Replacement Replacing the Tungsten (WI) Lamp Assembly Replacing the Deuterium (D2) Lamp Assembly 8.4 Main Fuse Replacement

5 Chapter 1 System Installation 1.1 Introduction This chapter contains the necessary information for installing the instrument. The installation procedures involve unpacking, power requirements, environmental requirements and interfacing. Please note this instrument belongs to installation category 2 and pollution degree Unpacking and Inspection Use the following steps to unpack your UV8500 spectrophotometer. 1. Carefully inspect the shipping container for any signs of external damage. If any external damage is noted, contact your agent immediately and do not discard the box. 2. Place the carton in an upright position and open it. 3. Lift the instrument out of the carton and place it on a flat surface, free from dust, vibration, and away from direct sunlight. 4. Open the cover and visually inspect the instrument for loose, bent or broken parts. Report any damage immediately. 5. Check the instrument accessories against the delivery(shipping) note. 6. Please save all packing materials, as it maybe required for later transportation. 1.3 Power Requirements Check the power reference which is attached on the rear panel of the instrument. 8500D is used with power supply 110V/50~60 Hz, 8500E is used with 220V/50~60Hz. Make sure the instrument is grounded in your laboratory. The maximum tolerance of voltage variation is ±10%. Note: Power Rating: 110 or 220VAC/ 50-60Hz, 6A Max. Main Fuse Rating: 250VAC/6A, Quick Blow(F).

6 1.4 Environmental Requirements Place the instrument on a flat, non-vibrating work bench. Avoid direct sunlight and preferable in an air-conditioning environment. Make sure that the back and right side of the instrument have at least 15 cm of clearance from the wall. Insufficient ventilation will cause overheat of the instrument. Note: Working Temperature: 5 C ~ 30 C Relative Humidity: < 85% Altitude: Up to 2000m 1.5 Electrical Connection Use the following steps to complete electrical connection of UV Locate the module power cord in the set of accessories that came with your instrument. Plug its female end into the receptacle of the power entry product, which is located on the back panel of UV8500. The power entry product combines the power socket, main switch, main fuses and power rating display as a single unit. The following diagram is sketch of the back panel of UV MOUSE TECHCOMP LTD. HONGKONG VOLTS FRE.: ***V 50/60Hz SERIAL NO.: H****** PRINTER 2. Plug the male end of the module power cord into an appropriate, grounded wall outlet or power strip. Warning: Do not use a plug adapter to ground the modular cord into a 2- prong(ungrounded) electrical outlet. You may expose yourself to dangerous electrical shock in the event of certain types of instrument failure. 3. Install a Windows 95 compatible mouse on the serial port of UV Install a Windows 95 compatible printer on the parallel port of UV8500.

7 1.6 Initial System Calibration Power on the UV8500 unit after electrical connection, the system will perform an initial calibration automatically. The calibration procedure includes: entering into Windows95 environment, electronic system check, deuterium lamp check, tungsten lamp check, wavelength calibration, and system baseline measurement. UV8500 software will be started automatically after the unit entered into Windows 95 environment.

8 1.7 Printer Setup Canon Bubble-Jet BJ-10e is set as default printer in UV8500. If you want to use another type of printer, first you should install it by using the Add Printer wizard of Windows95, then in UV8500 software, set the newly installed printer as defaulted. The following procedure shows you how to set a newly installed printer as defaulted in Steps: 1. On the File menu, click Print Setup. 2. Click the arrow next to Name, and select the printer type. 3. Click OK.

9 Chapter 2 Functional Description 2.1 Introduction The UV8500 spectrophotometer is designed to perform qualitative and quantitative photometric analyses in the ultraviolet and visible regions. A complete functional description of this instrument is introduced in this chapter. 2.2 Main Functions This section introduces the main functions in UV Fixed Points Measurement Single Wavelength Photometric Measurement UV8500 lets you read the photometric value(%transmittance or Absorbance) at a single wavelength quickly and conveniently. The wavelength slew speed reaches 6,000nm/min. Multiwavelength Photometric Measurement Up to 20 wavelength points can be set up in the multi-wavelength photometric measurement mode. Results will be grouped into a table format automatically. Result can be saved in built-in memory or floppies for later use. Concentration Measurement The UV8500 provides the following capabilities in concentration measurement. You can use up to 3 wavelength measurement to generate the regression curve. You can use up to 20 standards to set up the regression curve. UV8500 will calculate the working curve using a linear equation that fits the data. Abnormal reading for standard and samples can be deleted and modified after measurement. You can enter factor values to generate regression curves. Analytical results can be print out on a printer. Method and results can be saved in built-in memory or floppy for later use.

10 2.2.2 Wavelength Scanning Features You can acquire wavelength spectrum accurately and conveniently with selection of steps of 0.1, 0.2, 0.5, 1.0 and 4.0nm which corresponsing to speed ranges from 10 nm/min to 3,200nm/min. Light source switching position can be user defined from 320nm to 360nm. Light sources and filters are automatically changed during scanning. Peaks and valleys will be automatically detected after scanning. User can define the peak threshold. Powerful spectrum processing functions are provided. ( see section 5.5 ) Spectra can be printed out on a printer. Analytical results can also be saved in the built-in memory or floppy diskettes for later use Time Scanning The UV8500 allows you to record the absorbance or transmittance value of a sample as a function of time at a specified wavelength. On-line display of absorbance or transmittance with time graduations on the abscissa. Spectra can be printed out on a printer. Analytical results can also be saved in built-in memory or floppy diskettes for later use. Photometric values at specified intervals can be printed in table format. Data can also be exported to EXCEL software for statistical calculations.

11 2.2.4 Spectrum Processing Function The UV8500 spectrophotometer can save measured spectra in built-in memory or floppy diskettes for later processing. Spectrum processing functions include: Trace a Spectrum A cursor can be moved to a desired point with mouse in the spectrum displayed on the LCD screen and the photomatric data at this point is displayed. Automatic Peak Detection After a scanning is completed, peaks and valleys can be automatically detected and listed out in a table format. They will also be labelled on the spectrum. Scale Expansion Simultaneous expansion of X and Y axes are provided with Zoom function. Display range can also be changed though Display Setup function. Differentiation You can calculate and display the first through fourth derivative spectra for a given spectrum. Derivative spectra are useful for enhancing spectral data that are not readily apparent in absorbance spectrum. Calculations Between Spectra You can calculate addition and subtraction between two spectra with the resulting data displayed on the LCD Automatic System Initialization Function The UV8500 performs the following calibration and self-diagnosis functions automatically. Memory Check Lamp Ignition Check Wavelength Drive Mechanism Check Automatic Wavelength Calibration

12 2.3 Instrument Specifications Item Description Monochromator Double Beam CT Arrangement, Holographic diffraction grating Spectral Bandwidth < 2.0 nm Wavelength Range 190 to 1100nm Wavelength Setability 0.1nm increment Wavelength Accuracy ±0.5nm (ASTM method #E275-83) Wavelength Reproducibility 0.3nm (ASTM method #E275-83) Photometric Mode Absorbance & % Transmittance Photometric Range to 3.000A; 0 to 300%T Photometric Accuracy ±0.003A; ±0.5%T(0.000 to 1.000A) Photometric 0.002A; 0.3%T(0.000 to 1.000A) Reproducibility Photometric Drift ±0.001A/hr Instrument Noise < 0.2%T Baseline Flatness ±0.005A Stray Light <0.05%T (ASTM method #E387-72) Wavelength Slew Speed 6,000nm/min (full wavelength range) Wavelength Scan Speed 60 to 3,200nm/min Kinetic Sampling Interval 0.5s, 1s, 6s, 30s and 60s Light Source Deuterium Lamp, Tungsten-Halogen Lamp Light Source Switching Automatic light source switching, User selectable within a range from 320 to 360nm Detector Dual Silicon Photodiode Detectors Display UV8500 : 8.4 backlight LCD display (optional color LCD display) UV8500PC: Computer Monitor Sample Compartment 100 mm beam spacing 12.5cm (W) x 21.5cm (L) x 13.0cm (H) Dimensions 61.0cm (W) x 52.0cm (L) x 24.0cm (H) Power Source 110 or 220VAC, 50/60Hz Note: Please note that the instrument can be used at altitude up to 2000 meter. This instrument belongs to installation category 2 and pollution degree 1.

13 Chapter 3 Basic Operation 3.1 Introduction This chapter provides information for basic operation of UV8500/8500PC spectrophotometer. 3.2 Key-Panel (UV8500 only) The illustration below shows the operation panel of UV8500. Tab A 7 B C D 8 E 9 F Y Z SAVE SHUTS DOWN Caps G 4 H I 5 J K 6 L GOTOλ AU / ZE PRINT Shift M 1 N O P 2 Q 3 R CLEAR EXIT ENTER Ctrl S T U V W 0 X S/P STOP Description of Key Functions The description below gives the key functions that are available during operation. KEY DESCRIPTION Tab Use this key to select items Caps Input Character font Shift Input Characters of the upper position Ctrl Input characters of the lower position 0 ~ 9 Input Values: ( input numeric values ) - Minus?iDecimal Space GO TO λ Use for single wavelength photometric measurement CLEAR Clear error inputs, unwanted data, delete files S/P START/PAUSE the measurement(wavelength scan or time scan) SAVE Save current working files into memory AUTO ZERO Absorbance autozero calibration EXIT Return from sub-menu to main menu STOP Stop the measurement(wavelength scan or time scan) SHUTS DOWN Exit the UV8500 software

14 PRINT connected ENTER Left Arrow Right Arrow Up Arrow Down Arrow Print out current working files(data or spectra) on the printer Confirm entered value or selections Moves the highlight or the cursor left Moves the highlight or the cursor right Moves the highlight or the cursor up Moves the highlight or the cursor down 3.3 Dynamic Keyboard UV8500 provides a dynamic keyboard in the software. Every time the software asks you to input characters, a button representing the dynamic keyboard appears. Clicking this button, the dynamic keyboard will be shown on the LCD as the following. 3.4 Sample Compartment The following figure shows you the structure of sample compartment. REFERENCE LENS DETECTORS SAMPLE

15 3.5 Main Menu After entering the UV8500 software, the Main Menu appears on the display as shown in the illustration below. Three main functions are listed on the dialog box. They are Wave Scan, Fixed Points and Time Scan. The Fixed Points function includes multi-wavelength photometric measurement and concentration measurement. Status Bar Menu bar and Toolbar are both provided in UV8500 software, so you have two ways to select a desired function. On the menu bar, use your keypad or mouse to select the desired function. Almost all the functions listed in the menu bar can be reached by clicking a corresponding button in the toolbar.

16 Many quick-click buttons are provided in the Toolbar. In the following table, we give a brief introduction of these buttons. Button Description Button Description Start a Measurement Pause a Measurement Stop a Measurement Display and Print Setting Load a Spectrum Save a File Remove a Spectrum Print B Wavelength Scan Measurement Time Scan Measurement Background Calibration Fixed Points Measurement Go To λ Zero Calibration Method Setup Set Absorbance as Photometric Mode Activate the ZOOM Function Activate the Trace Cursor Set %Transmittance as Photometric Mode Display Range Setup Restore Zoomed(expanded) Axes to Their Original Scales Add Two Spectra Subtract one Spectrum from Another List out Peaks of a Spectrum Get a Derivative from a Spectrum List out Valleys of a Spectrum Define the Peak/Valley Threshold

17

18 Chapter 4 Fixed Points Measurement 4.1 Introduction This chapter describes three functions in the Fixed Points Measurement. Single Wavelength Photometric Measurement Multi-wavelength Photometric Measurement (up to 20 wavelength points) Concentration Measurement, which include regression curve setup and reading unknown samples. 4.2 Fixed Points Mode Method 1 On the File menu, click New, the following dialog box will be popped out. Either: (a) Use mouse to click Fixed Points from the pop out menu, then click OK, or (b) Use the arrow key on your keypad to select this item and press ENTER. Method 2 Click button on the toolbar. 4.3 Data Acquisition Mode Use the following steps to select the data Acquisition mode(%t, ABS) that you wish to use in the Fixed Points measurement. Step: Click button in the toolbar to select the %transmittance mode or click button to select the absorbance mode. The status bar will be changed accordingly. Abs.: *.**** %T: ***.*

19 4.4 Single Wavelength Measurement The UV8500 provides you a convenient method to measure photometric value at a fixed wavelength. Step: 1. Press key GO TO λ in the keypad or click button in the toolbar. 2. Key in the desired wavelength position and click OK. The wavelength setability is 0.1nm. Please note the wavelength should be within the range from 190nm to 1100nm. 3. Place a reference and a sample in the sample compartment. The wavelength position and photometric value will be displayed in the status bar. For example, if the wavelength position is set to be 555.0nm and Abs. value at this point is 0.226, the status bar will be shown as Ready Pos.: 555.0nm Abs.: Note: The single wavelength photomatric measurement function is also available in other sub-menus.

20 4.5 Multiwavelength Photometric Measurement Mode This section shows you the procedure in the multi-wavelength photometric measurement. Steps: 1. Click the button in the toolbar. The method dialog box will be popped out. 2. Type the number of wavelength points in the Number of points box, or click the up/down arrow next to the box to set the wavelength points. Leave the two boxes of Calculate Concentration and Use Standard Sample blank. 3. Keyin the wavelength positions in the wavelength box. For example, we set six wavelength points, which are 500nm, 510nm, 520nm, 780nm, 770nm and 760nm.

21 4. Click the Sample tab, the LCD will be shown as the following. The control menu contains six buttons, which are Start, Delete, Modify, Recalculate, Title/Format and Print Data. 5. Place a sample and a reference in the sample compartment. 6. Click Start to run a new measurement. The display will be changed as the following. 7. UV8500 will read the photometric value of sample 1 at the fixed wavelength automatically. Keyin the sample name in the Name box with keypad or dynamic keyboard. 8. Click OK after UV8500 complete the measurement, the photometric data for sample 1 will be listed in the sample table. 9. Repeat step 3 to step 8 to test all samples.

22 10. Click Print Data to print out the table displayed. The following table shows the format of the printout. Fixed Points Measurement By Techcomp UV8500 Sample name 500.0nm 510.0nm 520.0nm 780.0nm 770.0nm 760.0nm Sample-1 Sample-2 Sample-3 *Abs unit is used. Note: The Title/Format button enables you to definite your report title and the print format. Click the Title/Format button in the control menu, the following dialog box will appear. You can change the Main Title, Second Title and Foot Note in this dialog box.

23 4.6 Linear Regression Analysis Method Set Up There are two methods to set up the linear regression curve. You can use standards to set up the regression curve or just key in the parameters manually. Use the following steps to select the method that you want to use. Steps 1. Click the Method Tab. 2. Enter the number of wavelength points in the Number of points box, or click the up/down arrow next to the box to set the wavelength points. 3. Key in the wavelength positions in Wavelength boxes. 4. Tick the Calculate Concentration check box to activate concentration calculation. If you leave this check box blank, you will only get results in absorbance or in %transmittance. 5. Tick the Use Standard Samples check box if you want to set up the regression curve with prepared standards. Leave this check box blank if you want to use the existing standard curve parameters. Wavelength Points One to three wavelength points can be used in the concentration measurement. The calculation functions are different according to the number of wavelength points. 1. One Wavelength A=A 2. Two Wavelengths A= A2-A1 3. Three Wavelengths A=A2-(mA1+nA3)/(λ3-λ1) where m=λ3-λ2, n=λ2-λ Key in the curve parameters to set up the regression curve The regression curve is represented by the following function. C = C0 + C1 *Abs C0 represents the intercept and C1 is the slope. Keyin the value of C0 and C1 in their boxes.

24 4.6.3 Using Standards for Calibration Curve Setup This section shows you the procedure of using standard to set up the calibration curve. In the following example, we use two wavelength points, which are 260.0nm and 280.0nm to set up the regression curve. steps 1. Click the Standard tab. The following dialog box will be popped out. 2. Click the B button for background correction. 3. If a blank is prepared, place cuvettes which contain blank solution in both the reference and sample holders. Click for zero correction. 4. Place standard 1 in the sample compartment.

25 5. Click Start to run a measurement. 6. Key in the concentration value of standard 1 in the box of Conc. %. 7. You can define a new name for the standard in the Name box, otherwise, it will be defaulted as Standard After UV8500 completes the measurement of standard 1, click OK. The photometric data, A and concentration will be shown in the standard table. Repeat step 4 to step 7 to measure all the standards prepared. 9. Incorrect results can be deleted. To do this, first click the standard name in the sample column, then click Delete. The following dialog box will be shown on the display. Click Yes to delete the standard reading.

26 10. Click the Fitting tab to display the regression curve. 11. Display range can be modified by clicking the Display Setting tab. After you changed the display parameters in the above dialog box, click Apply, the fitting display will be changed according to your settings. Note: The unit of concentration you set in the above dialog box should be the same as the one you have set for your standards.

27 4.7 Sample Test The following procedure shows you the step to read samples. 1. Click the Sample tab. The interface will display as the following. 2. Place sample 1 in the sample cuvette holder. 3. Click START to run a measurement. 4. UV8500 will read out the photometric value of sample 1 at the fixed wavelength positions automatically. Type the sample name in the Name box, otherwise, it will be defaulted as Sample Click OK, the photomatric readout for Sample-1 will be listed in the sample table. DeltaAbs and concentration value of Sample-1 will also be displayed in column 3 and column 4.

28 6. Repeat the step 2 to 5 to test all the unknown samples prepared. 7. Click Print Data to print out the table displayed. The following table shows the format of the printout. Fixed Points Measurement By UV8500 Sample name 260.0nm 280.0nm DeltaAbs % Sample-1 Sample-2 Sample-3 *Abs unit is used. Note: The Title/Format button enables you to definite your report title and the print format. Click the Title/Format button in the control menu, the following dialog box will be popped out. You can change the Main Title, Second Title and Foot Note in this dialog box.

29 4.8 Save Files Step: 1. On the File menu, click Save, or click in the toolbar. The interface will be changed as the following. 2. Select a folder and key in a file name in the File name box. The file type for fixed points measurement is defaulted as *.fix. There are two ways to key in a filename. One is to use the membrane keypad. The other is to use the dynamic keyboard, which will be popped out by clicking the button located beside the File name box. 3. Click Save.

30 4.9 Load Files Steps : 1. On the File menu, click Open, or click in toolbar. The interface will be changed as the following. 2. Select a folder and a filename. 3. Click OK to open the file selected.

31 Chapter 5 Wavelength Scanning 5.1 Introduction This chapter describes how to collect a spectrum while using Wavelength Scan function. 5.2 Selecting Wavelength Scan Mode Method 1 On the File menu, click New, the following dialog box will be popped out. Either: (a) Use mouse to click Wave Scan from the pop out menu, then click OK, or (b) Use the arrow key on your keypad to select this item and press ENTER. Method 2 Click button on the toolbar. The LCD will display the wavelength scan interface as below.

32 5.3 Setting the Lamp Switching Wavelength Position During system initialization, UV8500 will automatically compare the energy of D2 lamp and tungsten lamp within the range from 320nm to 360nm to set the lamp switching point. You can set a new switching point to replace the current one through the following steps. Step: 1. On the UV8500 menu, click D2/W Switch Point, the interface will be changed as the following. 2. Key in the lamp switching wavelength position in the New point box. It should be within the range from 320nm to 360nm. 3. Click Setup button or press ENTER on the keypad. The interface will return to the wavelength scan sub-menu.

33 5.4 Step by Step Operation This section describes how to operate the UV8500 spectrophotometer in the Wavelength Scan Mode Selecting Data Acquisition Mode Use the following steps to select the data Acquisition mode(%t, ABS) that you wish to use in the wavelength scanning measurement. Click button in the toolbar to select the %transmittance mode or click button to select the absorbance mode Entering Wavelength Scan Setup Variables You can use the following steps to set up the variables for wavelength scan. 1. Click button in the toolbar. The interface will be changed as the following. 2. Key in the lower limit of scan range in the From box. Acceptable entries range from 180 to Key in the upper limit of scan range in the To box. Acceptable entries range from 180 to Click the arrow next to Step, and select a scan interval. Six scan intervals can be selected from 0.1nm, 0.2nm, 0.5nm, 1.0nm, 2.0nm and 4.0nm. 5. Click OK, the interface will return to the wavelength scan sub-menu.

34 5.5 Collecting a Spectrum in Wavelength Scan Once you have set up the operation conditions for wavelength scan, you are ready to collect a spectrum. The following procedure shows you how to collect a spectrum. 1. Place a sample in the sample cuvette holder. Close the cover of the sample compartment. 2. Click button in the toolbar. The instrument will start scanning automatically. 3. During scanning, if you want to stop it, click in the toolbar. The scan procedure can also be paused by clicking the button. Click this button again, the instrument will continue scanning. 4. The real time spectrum will be displayed on the LCD during scanning.

35 Baseline During scanning, the unit will use the system baseline which was checked during system initilization. For particular experiments where the background matrix is a strong absorber, we suggest you run the baseline again with the following steps. 1. Place sample cuvettes which contain a reference or blank solution in both the reference and sample cuvette holders. 2. On the UV8500 menu, click Autozero, or click button in the toolbar. The unit will begin the collection of baseline automatically. The new baseline will be stored in the baseline memory and will be used in all subsequent experiments until you run a new baseline to replace the current one.

36 5.5 Spectrum Processing Introduction Once you have acquired and displayed a spectrum, the following options are available. 1. RESCALE 2. ZOOM 3. PEAKS & VALLEYS 4. TRACE 5. ARITHMETIC PROCESS 6. FILES The following table shows you the general functions of spectrum processing. Option Variable Range Function RESCALE X-Axis 190 to 1100nm Reset the display scale of a spectrum Y-Axis -1 to 3 ZOOM X-Axis 190 to 1100nm Expands either or both axes for more detailed viewing. Y-Axis -1 to 3 PEAKS & VALLEYS TABLE List out the peaks and valleys of a spectrum TRACE ARITHMETIC PROCESS FILES THRESHOLD ABS=0.001 to 1.000, in increments, %T=0.1 to in 0.1 increments A + B A - B DERIV SAVE LOAD UNLOAD List the Y-axis values over which the instrument detects a peak or valley Permits reading of values from the on-screen spectrum using the cursor Sums together two spectra Subtracts one spectrum from another Displays a 1st through 4th order derivative spectrum save the current processing spectrum load a spectrum saved before Remove a spectrum from display

37

38 Select a Current Spectrum As UV8500 can displayed several spectra overlaid on the screen, you should specify the spectrum before you want to process it. Step: Down Arrow 1. Click the down arrow. All spectra displayed will be listed in the pull-down menu. 2. Click the spectrum you want to select. Its name will be listed in the Name Box and we call it as Current Spectrum.

39 5.5.2 RESCALE The following steps shows you how to change the display range of a spectrum. 1. On the Settings menu, click Display Range, or click the button in the toolbar. The interface will be changed as follows. 2. Key in the display range variables for x-axis and y-axis. The maximum range for X- axis is from 180 to 1100 and the maximum range for Y-axis is from -1 to You can also set a display intervals on both X and Y axes. To do this, first tick the box of Manual Settings and then key in the intervals. 4. Click OK. Note: The button state. in the toolbar lets you restore the display settings to the defaulted

40 5.5.3 ZOOM By using a mouse, you can zoom a part of the spectrum more conveniently. Steps. 1. Click the button in the toolbar. 2. Position the cursor in the upper-left corner of the area you want to select. 3. Hold the left mouse button to drag the cursor to outline the spectrum area you want to enlarge. 4. Release the mouse button. The spectrum which displayed within the outlined area will be enlarged. The following is an example of this function. Zoomed Area Note: The button state. in the top tools bar lets you restore the display settings to the defaulted

41 5.5.4 PEAKS & VALLEYS List the Peaks and Valleys of a spectrum Once a spectrum is displayed on the screen and it is appointed as Current Spectrum. (refer to section 5.5.1) On the View menu, click Peaks, Or clicking button in the toolbar. All peaks detected will be listed in a table format beside the spectrum. On the View menu, click Valleys, Or clicking button in the toolbar. Valleys of the spectrum displayed will be listed. Following is an example of this function. Peaks are listed above the spectrum, while valleys are listed below it. In the table, peaks are noted with,while valleys are noted with.

42 5.5.5 THRESHOLD The threshold value measures the absorbance from a valley to peak. If the value is greater than the one you choose for Threshold, the instrument will detect it and report it as a peak. Therefore, smaller Threshold values will detect smaller peaks and even background noise, if you lower the value far enough. UV8500 let you change the threshold value which will be used in peak and valley detection. On the Settings menu, click Peak Height, or click the button, the following dialog box will be popped out. Type the threshold value in the box and click OK. UV8500 will then use the new threshold to detect peaks and valleys.

43 5.5.6 TRACE Step: 1. Once a spectrum is displayed on the screen and it is appointed as Current Spectrum. (refer to section ) on the View menu, click Search, or click the button in the toolbar. The interface will be changed as the following. 2. A crosshair cursor will appear with x and y axes values displayed. 3. Move the crosshair cursor left or right on the spectrum. 4. The data in the cursor window indicate the X-axis and Y-axis values for the current cursor location. 5. Double click the left mouse button to release the crosshair cursor.

44 5.5.7 DERIVATIVE Steps: 1. Once a spectrum is displayed on the screen and it is appointed as Current Spectrum, (refer to section 5.5.1) on the Compute menu, click Derivative, or click the button in the toolbar. The following dialogue box will be popped out. 2. Key in the class of derivative and type a name for the result spectrum, then click OK. The result spectrum will be displayed overlaid with the original one.

45 5.5.8 Adding Spectra Adding spectra can assist in the development of artificial structure of spectra in multicomponent mixtures. UV8500 will only adding two spectra that are already displayed on the screen, so before arithmetic processing, load two spectra from the memory. In the following example, we have two spectra saved in directory D:\UV8500\Samples, which are 1.wav and 2.wav. Step 1. On the Compute menu, click Add, or click the button in the toolbar. The following dialogue box will be popped out. 2. Click the down arrow next to Source 1 to select a spectrum and define it as source 1. Select a spectrum for Source 2 in the same way. UV8500 will not allow you to select a spectrum not displayed on the screen or select a same spectrum twice. 3. Key in a name for the Result spectrum and click OK. The UV8500 will start processing with the result displayed on the screen.

46 5.5.9 Spectra Subtraction Subtracting one spectrum from another has been a classical technique to offset spectral interference from the spectrum of interest. UV8500 will only process subtraction of two spectra that are already displayed on the screen, so before arithmetic processing, load two spectra from the memory. In the following example, we have two spectra saved in directory D:\UV8500\Samples, they are 1.wav and 2.wav. Steps 1. On the Compute menu, click Sub, or click the button in the toolbar. The following dialogue box will be popped out. 2. Click the down arrow next to Source 1 to select a spectrum and define it as source 1. Select a spectrum for Source 2 in the same way. UV8500 will not allow you to select a spectrum not displayed on the screen or select a same spectrum twice. 3. Key in a name for the Result spectrum and click OK. The UV8500 will start processing with the result displayed on the screen.

47 SAVE A SPECTRUM Step: 1. On the File menu, click Save, or click button in the toolbar. The interface will be changed as the following. 2. Select a folder and key in a file name in the File name box. The file type for wavelength scanning is defaulted as *.wav. There are two ways to key in a filename. One is to use the membrane keypad. The other is to use the dynamic keyboard, which will be popped out by clicking the icon located beside the File name box. The following dialog box will be popped out after you clicking this icon. 3. Click Save.

48 LOAD A SPECTRUM Step: 1. Click the button in the toolbar. The interface will be changed as the following. 2. Select the directory and select the file name. 3. Click Open.

49 Unload a spectrum You can unload a spectrum if you do not want it displayed on the screen. For example, we suppose that three spectra are displaying on the LCD. They are demo2.wav, 1.wav and 2.wav and we want to remove 2.wav from the screen. Step: 1. Click the down arrow. All spectra displayed will be listed in the pull-down menu. 2. Click 2.wav and appoint it as Current Spectrum(refer to section 5.5.1). 3. Click the button in the toolbar, 2.wav will be removed from the display. This will not removed the spectrum from the memory, 2.wav is still saved in directory D:\UV8500\samples\.

50

51 Chapter 6 Time Scanning (Kinetic Analysis) 6.1 Introduction This chapter tells you how to obtain the absorbance or transmittance value for a sample as a function of time at a given wavelength. 6.2 Selecting Time Scan Mode Method 1 On the File menu, click New, the following dialog box will appear. Either: (a) Use mouse to click Time Scan from the pop out menu, then click OK, or (b) Use the arrow key on your keypad to select this item and press ENTER. Method 2 Click icon following. in the toolbar. The LCD will display the time scan interface as the

52 6.3 Step by Step Operation This section describe how to operate the UV8500 spectrophotometer in the Time Scan Mode Selecting Data Acquisition Mode Click button in the toolbar to select the %transmittance mode or click button to select the absorbance mode Entering Time Scan Setup Variables You can use the following steps to set up the variables for time scan. Step : 1. Click button in the toolbar. A dialogue box will be popped out as follows. 2. Key in the wavelength, total time(in seconds) and scan step in the above dialog box. The wavelength range should be within 190 to 1100nm. The upper limit for total time is 10 5 second, around 27 hours and 42 minutes. Four scan interval can be selected from 0.5s, 1s, 60s, and 120s. 3. Click OK, the interface will be returned to the Time Scan sub-menu.

53 6.3.3 Collecting a Spectrum in Time Scan Once you have set up the operation conditions, you are ready to collect a spectrum. To collect a spectrum in time scan, follow this procedure: 1. Close the cover of the sample compartment, on the UV8500 menu, click Background, B or click button on the toolbar. UV8500 will run a background correction at this wavelength point automatically. Then the status bar will show OK. 2. Place sample cuvettes which contain a reference or blank solution in both the reference and sample cuvette holders. On the UV8500 menu, click Autozero, or click button in the toolbar. 3. Take out the reference or blank in the sample holder, place a sample in it and close the cover. 4. Click button in the toolbar. The instrument will start scanning automatically. 5. You can stop scanning by clicking the button. The scan procedure can also be paused by clicking the button. Click this button again, the instrument will continue scanning. 6. The spectrum will be displayed on the LCD during scanning. For rate measurement, please refer to section 6.4.6, use the Derivative Function to calculate slope.

54 6.4 Spectrum Processing Introduction After you have acquired and displayed a spectrum, the following options are available: 1. RESCALE 2. PEAKS & VALLEYS 3. TRACE 4. ARITHMETIC PROCESS 5. FILES The following table shows you the general functions of spectrum processing. Option Variable Range Function RESCALE X-Axis 0 to (second) Expands either or both axes for more detailed viewing Y-Axis -1 to 3 PEAKS & VALLEYS TABLE List out the peaks and valleys of a spectrum TRACE THRESHOLD ABS=0.001 to 1.000, in increments, %T=0.1 to in 0.1 increments List the Y-axis values over which the instrument detects a peak or valley Permits reading of values from the on-screen spectrum using the cursor PROCESS DERIV Displays a 1st through 4th order derivative spectrum FILES SAVE LOAD UNLOAD save the current processing spectrum load a spectrum saved before Remove a spectrum from display

55 Appoint a Current Spectrum As UV8500 can displayed several spectra overlaid on the screen, you should specify the spectrum before you want to process it. Step: Down Arrow 1. Click the down arrow. All spectra displayed will be listed in the pull-down menu. 2. Click the spectrum you want to select. Its name will be listed in the Name Box and we call it as Current Spectrum.

56 6.4.2 RESCALE The following steps shows you how to change the display range of a spectrum. 1. On the Settings menu, click Display Range, or click the button in the toolbar. The interface will be changed as the following. 2. Key in the display range variables. The maximum range for X-axis is from 0 to and the maximum range for Y-axis is from -1 to 3 if the photometric mode is absorbance. 3. Click OK. Note: The button state. in the top tools bar lets you restore the display settings to the defaulted

57 6.4.3 PEAKS & VALLEYS List the Peaks and Valleys of a spectrum Once a spectrum is displayed on the screen and it is appointed as Current Spectrum,(refer to section 6.4.1) on the View menu, click Peaks, or click button in the toolbar. All peaks detected will be listed in a table format beside the spectrum. On the View menu, click Valleys, Or clicking button in the toolbar. Valleys of the spectrum displayed will be listed. Following is an example of this function.

58 6.4.4 THRESHOLD The threshold value measures the absorbance from a valley to peak. If the value is greater than the one you choose for Threshold, the instrument will detect it and report it as a peak. Therefore, smaller Threshold values will detect lower peaks and even background noise, if you lower the value far enough. UV8500 let you change the threshold value which will be used in peak and valley detection. Click the button on the toolbar, the following dialog box will be popped out. Type the threshold value in the box and click OK. UV8500 will then use the new threshold to detect peaks and valleys.

59 6.4.5 TRACE Step: 1. Once a spectrum is displayed on the screen and it is appointed as Current Spectrum, (refer to section 6.4.1) click the button in the top tools bar. The interface will be changed as the following. 2. A crosshair cursor will appear. Move the crosshair cursor left or right on the spectrum. 3. The data in the cursor window indicate the X-axis and Y-axis values for the current cursor location. 4. Double click the left mouse button to will release the crosshair cursor.

60 6.4.6 DERIVATIVE Steps: 1. Once a spectrum is displayed on the screen and it is appointed as Current Spectrum, (refer to section ) on the Compute menu, click Derivative, or click the button on the toolbar. The following dialogue box will be popped out. 2. Key in the class of derivative and select a name for the result spectrum, then click OK. The result spectrum will be displayed overlaid with the original one.

61 6.4.7 SAVE A SPECTRUM Step: 1. On the File menu, click Save, or click button on the toolbar. The interface will be changed as the following. 2. Select a folder and key in a file name in the File name box. The file type for time scan measurement is defaulted as *.tim. There are two ways to key in a filename. One is to use the membrane keypad. The other is to use the dynamic keyboard, which will be popped out by clicking the button next to the File name box. The following diagram is a draft of the dynamic keyboard. 3. Click Save.

62 6.4.8 LOAD A SPECTRUM Step: 1. Click the button on the toolbar. The interface will be changed as the following. 2. Select the file directory and select the file name. 3. Click Open.

63 6.4.9 Unload a spectrum You can unload a spectrum if you do not want it to be displayed on the screen. For example, we suppose there are two spectra displayed on the LCD. They are inktest.tim and test1.tim. Now we want to remove test1.tim from the screen. Step: 1. Click the down arrow. All spectra displayed will be listed in the pull-down menu. 2. Click test1.tim and appoint it as Current Spectrum(refer to section 6.4.1). 3. Click the button in the toolbar, test1.tim will be removed from the display. This will not removed the spectrum from the memory, test1.tim is still saved in directory D:\UV8500\samples\.

64

65 Chapter 7 Maintenance 7.1 Introduction The UV8500 spectrophotometer is designed to require minimal maintenance. This section describes the routine maintenance tasks you should perform and the diagnostic functions that are used to characterize the instrument. 7.2 Internal Diagnostics and Automatic Adjustments The UV8500 spectrophotometer executes a self-diagnostic each time it is power on. If you want to conduct certain performance verifications manually, follow the procedures listed below Verifying Wavelength Accuracy and Reproducibility Definition Wavelength Accuracy: the deviation of the average wavelength reading at an absorption band or emission band from the known wavelength of the band. Wavelength Reproducibility: a measure of the ability of a spectrophotometer to return to the same spectral position as measured by an absorption band or emission band of known wavelength when the instrument is reset or read at a given wavelength. Standard Most spectrophotometric methods employ compounds or unknown mixtures for the purpose of calibrating instruments photometrically at specified analytical wavelengths. We can provide you with a calibration standard ( Cat no. CS-100, holmium oxide glass ) to verify your spectrophotometer performance. The absorption spectrum of CS100 is obtained by scanning it in the whole wavelength range. Absorption peak wavelength values of holmium oxide glass are listed below. Band No. Wav. (nm) Band No. Wav. (nm) Band No. Wav. (nm) Absorption peaks of holmium oxide glass Method Use Wave Scan function to scan the CS-100 standard from 250 to 650nm. We suggest you to use 0.1nm as scanning step. After you have acquired and displayed the scanned spectrum, list out the peaks.

66 Calculation Calculate the difference of each observed wavelength from the average and sum the differences. The wavelength accuracy is the difference between the true wavelength and the averaged observed wavelength. Wavelength accuracy can be calculated as follows: n λ = 1/n λi - λr i=1 where: λi = observed wavelength value, λr = true or reference wavelength value, n = number of measurements The average value of the differences is the wavelength reproducibility which can be expressed as: n δλ = max λi - 1/n λi i=1 where: λi = observed wavelength value, n = number of measurements

67 7.2.2 Confirming the Spectrum Bandwidth Definition The bandwidth, in nanometers of the segment of the spectrum that will be isolated by the monochromator. It is the range of wavelength between the points at which the transmittance is equal to one-half the peak. Steps 1. Enter the interface of Wave Scan. 2. On the UV8500 menu, click D2 Spectral Slitwidth. The instrument will automatically scanned the deuterium lamp energy around its emission peak at 656.1nm. 3. Print out the scanned spectrum. Calculations Confirm the bandwidth as instructed and illustrated below: 1. Measure the peak width in nm at half-height. 2. If the value is below 2nm, then the bandwidth is normal. x 5 A x 1/2A Bandwidth Bandwidth T t

68 7.2.3 Confirming Baseline Flatness The baseline flatness is defined by the amount of baseline drift, excluding electronic noise. Be sure to warm up the instrument sufficiently (approx. 2 hours), and then proceed as instructed below: 1. From the main menu, select Wave Scan function. 2. Set up the wavelength scan variables as shown in the following table. DATA MODE ABS START WAVE nm STOP WAVE nm UP SCALE LOW SCALE SCAN STEP 0.2nm LAMP SWITCH 340nm 3. Click OK to acquire the spectrum. The baseline is normal if it is within ±0.005 Abs in the whole wavelength range.

69 Chapter 8 Replacement Parts 8.1 Introduction The UV8500 contains only a few parts that are considered to be spare parts. It is good preventive maintenance to keep these spare parts on hand to minimize instrument downtime in case of need. 8.2 Spare Parts The following parts are recommended to be kept on hand as spares: 1. Tungsten (WI) Lamp Assembly 500 hours expected service life (P/N: A) 2. Deuterium (D2) Lamp Assembly 300 hours expected service life (P/N: A) 3. Main Power Fuse 250VAC/6A, 50~60Hz, quick blow(f) Ver.: 1.2

70 8.3 Lamp Replacement Please observe the following safety precaution when using the lamp installation procedures to install a new lamp in your UV8500 spectrophotometer. WARNING: Do not attempt to replace either lamp with the power switch on. Both lamps operate at high temperatures that can cause skin burns on contact. Also, potentially injurious voltages will remain in the lamp connections for a short period of time after the instrument is turned off. For these reasons, turn the instrument off and allow a minimum of 20 minutes to allow lamp cool down before making a replacement. Removing the lamp compartment cover will expedite the lamp cooling process. Caution: Do not touch the reflective surface of the focusing reflector and the bulb. Any finger prints on these surfaces must be removed with ethanol Ver.: 1.2

71 8.3.1 Replacing the Tungsten (WI) Lamp Assembly Follow these steps and precautions to replace a tungsten (WI) lamp: 1. Turn off the instrument and disconnect its power cord from the power source. Allow a minimum of 20 minutes to pass so that the lamp cools down and the lighting circuitry completely discharges before proceeding to step Remove the slide door which is located on the top of the lamp compartment. 3. Unscrew the two locking nuts which fasten the lamp compartment cover. Then lift off the cover and set it aside. The illustration below shows the location of the main components inside the lamp compartment. Focussing Reflector Connection Terminal D2 L WI Lamp 4. Unscrew the two locking screws in the connection terminal for WI lamp. The connection terminal is illustrated in the following diagram. The two blue wires are connected to the WI lamp. D2 WI 5. Unscrew the two screws which fasten the WI lamp assembly on the chassis. 6. Take out the lamp holder and replace it with a new one. Note: You need not to take the lamp off the lamp holder as these two parts should only be replaced as a whole assembly. 7. Lock the new lamp assembly on the chassis with two fastening screws. 8. Fasten the two blue wires in the connection terminal for WI Lamp. 9. Close the lamp compartment cover and re-tighten the lamp compartment screws. 10. Install the slide door on the instrument cover Replacing the Deuterium (D2) Lamp Assembly Blue Wires Follow these steps and precautions to replace a tungsten (WI) lamp: Ver.: 1.2

72 1. Turn off the instrument and disconnect its power cord from the power source. Allow a minimum of 20 minutes to pass so that the lamp cools down and the lighting circuitry completely discharges before proceeding to step Remove the slide door which is located on the top of the lamp compartment. 3. Unscrew the two locking nuts which fasten the lamp compartment cover. Then lift off the cover and set it aside. The illustration below shows the location of the main components inside the lamp compartment. Focussing Reflector Connection Terminal D2 L WI Lamp 4. To replace the D2 lamp, you should unscrew the three locking screws in the connection terminal for deuterium lamp first. The connection terminal is illustrated in the following diagram. The color of the high voltage one is different from the low voltage ones. Please note that the high voltage should be connected to a specified position while the positions of the low voltage ones are interchangeable. D2 W High Voltage 5. Unscrew the two screws which fasten the D2 lamp holder on the chassis. Low Voltage 6. Take out the lamp holder and replace it with a new one. Note: You need not to take the lamp off the lamp holder as these two parts should only be replaced as a whole assembly. 7. Lock the new lamp assembly on the chassis with two fastening screws. 8. Fasten the three color wires in the connection terminal for D2 Lamp. 9. Close the lamp compartment cover and re-tighten the lamp compartment screws Ver.: 1.2

Contents Safety.. 1 General. 1 Electrical.. 1 Warning 1 Working Principle 2 Unpacking Instructions 2 Specifications 3 Installation.. 3 Introduction..

Contents Safety.. 1 General. 1 Electrical.. 1 Warning 1 Working Principle 2 Unpacking Instructions 2 Specifications 3 Installation.. 3 Introduction.. Contents Safety.. General. Electrical.. Warning Working Principle 2 Unpacking Instructions 2 Specifications 3 Installation.. 3 Introduction.. 4 Description of keys and operating method.. 5 Operating Procedures.

More information

UV / VIS SPECTROPHOTOMETER

UV / VIS SPECTROPHOTOMETER www.mapada.com.cn A Professional Manufacturer Shanghai Mapada Instruments Co., Ltd. Add: Building D-10, 261# Sanbang Rd., Songjiang Export Processing Zone,Shanghai, 201611 Tel: +86-21-5488 1172 +86-21-5488-0273

More information

Operational Manual Spectrophotometer Model: SP-830 PLUS Metertech Inc.

Operational Manual Spectrophotometer Model: SP-830 PLUS Metertech Inc. Operational Manual Spectrophotometer Model: SP-830 PLUS Metertech Inc. Version 1.06 Metertech Inc. provides this publication as is without warranty of any kind, either express or implied, including, but

More information

Model : LUV-300 Please read this manual carefully before using the instrument Labnics Equipment

Model : LUV-300 Please read this manual carefully before using the instrument Labnics Equipment UV/VIS SPECTROPHOTOMETER Instruction Manual Model : LUV-300 Please read this manual carefully before using the instrument Labnics Equipment Table of Content Sr.No. Contents Pg. No. 1 Principles of Operation

More information

Model FP-6500 Spectrofluorometer Instruction Manual. FP-6500 for Windows

Model FP-6500 Spectrofluorometer Instruction Manual. FP-6500 for Windows Model FP-6500 Spectrofluorometer Instruction Manual FP-6500 for Windows P/N: 0302-9999 April 2000 Contents Safety Considerations...i Regulatory Statements... iii Preface... iv Installation Conditions...v

More information

UV-VIS Spectrophotometer

UV-VIS Spectrophotometer UV-VIS Spectrophotometer LOW STRAY LIGHT EXCELLENT STABILITY EASILY UPDATED MANY APPLICATIONS LOW COST HIGH QUALITY SMALL FOOTPRINT USER FRIENDLY SOFTWARE UV-VIS Spectrophotometer 1/2 PG INSTRUMENTS LIMITED

More information

UV-6 Series Double Beam UV/Vis Spectrophotometers

UV-6 Series Double Beam UV/Vis Spectrophotometers UV-6 Series Double Beam UV/Vis Spectrophotometers Model UV-6100 UV-6100PC UV-6300 UV-6300PC avelength Range 190-1100nm 190-1100nm Spectral Bandwidth 1.8nm 1.8nm 1.0nm 1.0nm Optical System avelength Accuracy

More information

Table of Contents 1. PRINCIPLE, APPLICATION AND FEATURES PRINCIPLE APPLICATION FEATURES SPECIFICATIONS AND ACCESSORIE

Table of Contents 1. PRINCIPLE, APPLICATION AND FEATURES PRINCIPLE APPLICATION FEATURES SPECIFICATIONS AND ACCESSORIE 752N UV-Visible Spectrophotometer Operating Instructions This Operating Instructions provides details on the operation, maintenance and troubleshooting of 752N UV-visible spectrophotometer. Please read

More information

UV-VIS Spectrophotometer

UV-VIS Spectrophotometer T60 UV-VIS Spectrophotometer LOW STRAY LIGHT EXCELLENT STABILITY EASILY UPDATED MANY APPLICATIONS LOW COST HIGH QUALITY SMALL FOOTPRINT USER FRIENDLY SOFTWARE T60 UV-VIS Spectrophotometer 1/2 PG INSTRUMENTS

More information

Affordable Lab Technology

Affordable Lab Technology T6 T6 UV/ VIS Spectrophotometer LOW STRAY LIGHT EXCELLENT STABILITY EASILY UPDATED MANY APPLICATIONS LOW COST HIGH QUALITY SMALL FOOTPRINT USER FRIENDLY SOFTWARE 1. 2 Low stray light Stray light below

More information

67 Series Spectrophotometer PC Software

67 Series Spectrophotometer PC Software 67 Series Spectrophotometer PC Software Instruction Manual 670 004/Rev D/04-17 IMPORTANT Please ensure the SD Card is fitted into the socket as detailed in the image below. -------------------------------------------------------------------------------------------------------------------------

More information

USER S MANUAL. SP-8001 UV-VIS Spectrophotometer. Metertech Inc. Version 1.08

USER S MANUAL. SP-8001 UV-VIS Spectrophotometer. Metertech Inc. Version 1.08 USER S MANUAL SP-8001 UV-VIS Spectrophotometer Metertech Inc. Version 1.08 provides this publication as is without warranty of any kind, either express or implied, including, but not limited to the implied

More information

SP-VIS 200. Operating Manual. SP-VIS 200 Series UV-Vis Spectrophotometer. Instruments. User s Manual SP VIS Spectrum Instruments

SP-VIS 200. Operating Manual. SP-VIS 200 Series UV-Vis Spectrophotometer. Instruments. User s Manual SP VIS Spectrum Instruments Instruments SP-VIS 200 SP-VIS 200 Series UV-Vis Spectrophotometer Operating Manual User s Manual SP VIS 200 1 SPECTRUM INSTRUMENTS AUST PTY LTD. Add: 1211/ 7, Riverside Quay, South Bank, Victoria 3006,

More information

UNICO 2100 SERIES SPECTROPHOTOMETER Model 2100 Spectrophotometer Model UV 2100 Spectrophotometer

UNICO 2100 SERIES SPECTROPHOTOMETER Model 2100 Spectrophotometer Model UV 2100 Spectrophotometer UNICO 2100 SERIES SPECTROPHOTOMETER Model 2100 Spectrophotometer Model UV 2100 Spectrophotometer SERVICE MANUAL UNITED PRODUCTS & INSTRUMENTS INC. MANUFACTURER OF QUALITY OPTICAL SYSTEMS 1-732-274-1155

More information

Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017

Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017 Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017 Instrument instructions can be found at: http://academic.bowdoin.edu/chemistry/resources/instructions.shtml If you have any problems with the

More information

V-600 Operation. JASCO V-600 for Windows Software Manual

V-600 Operation. JASCO V-600 for Windows Software Manual V-600 Operation JASCO V-600 for Windows Software Manual P/N:0302-7420A September 2006 Preface This instruction manual serves as a guide for using this instrument. It is intended to instruct first-time

More information

Thermo Scientific GENESYS 30 Visible Spectrophotometer Legendary Reliability Unrivaled Usability

Thermo Scientific GENESYS 30 Visible Spectrophotometer Legendary Reliability Unrivaled Usability Thermo Scientific GENESYS 30 Visible Spectrophotometer Legendary Reliability Unrivaled Usability Colorimetry Workhorse, Lab Ready, Simple Controls Designed for Users, Designed for Results The Thermo Scientific

More information

Provide Accurate Results. Operational Manual. Visible Spectrophotometer. SP-880 Mate. PC Software. Metertech Inc. Version 1.2.3

Provide Accurate Results. Operational Manual. Visible Spectrophotometer. SP-880 Mate. PC Software. Metertech Inc. Version 1.2.3 Provide Accurate Results Operational Manual Visible Spectrophotometer SP-880 Mate PC Software Metertech Inc. Version 1.2.3 Metertech Inc. provides this publication as is without warranty of any kind, either

More information

Thermo Scientific Spectronic 200 Visible Spectrophotometer. The perfect tool. for routine measurements

Thermo Scientific Spectronic 200 Visible Spectrophotometer. The perfect tool. for routine measurements Thermo Scientific Spectronic 200 Visible Spectrophotometer The perfect tool for routine measurements The Standard for Routine Measurements Thermo Scientific SPECTRONIC spectrophotometers have served as

More information

Precise & Affordable

Precise & Affordable Water Analysis Instruments Thermo Scientific Orion AquaMate and AQUAfast Product Brochure Reliable water analysis Precise & Affordable A complete line of Spectrophotometers, Colorimeters and Turbidity

More information

Provide Accurate Results. Operational Manual. Visible Spectrophotometer SP-880. Metertech Inc. Version 1.10

Provide Accurate Results. Operational Manual. Visible Spectrophotometer SP-880. Metertech Inc. Version 1.10 Provide Accurate Results Operational Manual Visible Spectrophotometer SP-880 Metertech Inc. Version 1.10 Metertech Inc. provides this publication as is without warranty of any kind, either express or implied,

More information

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer Operating instructions - Cary 100 Bio UV-visible Spectrophotometer This document describes how to power up the spectrophotometer, set its measurement parameters, insert one or more of samples for analysis

More information

SoftWare. M.Wave Professional PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

SoftWare. M.Wave Professional PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Operation Manual SoftWare M.Wave Professional PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC.011.14 Contents 1.

More information

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software Spectrophotometers with Vernier Data Acquisition Software The absorbance of a sample is given as A = log I o I, where I o is the intensity without sample present and I is the intensity with the sample

More information

Hitachi U-2001 UV/Vis Spectrometer Updated November 14, 2017

Hitachi U-2001 UV/Vis Spectrometer Updated November 14, 2017 Hitachi U-2001 UV/Vis Spectrometer Updated November 14, 2017 Instrument instructions can be found at: http://academic.bowdoin.edu/chemistry/resources/instructions.shtml If you have any problems with the

More information

MODELS 6500/6505 SPECTROPHOTOMETERS OPERATING MANUAL

MODELS 6500/6505 SPECTROPHOTOMETERS OPERATING MANUAL MODELS 6500/6505 SPECTROPHOTOMETERS OPERATING MANUAL 650 050/REV A/09-99 CONTENTS SAFETY 1 Page SECTION 1 SECTION 2 SECTION 3 SECTION 4 SECTION 5 SECTION 6 INTRODUCTION 1.1 Instrument Description 2 1.2

More information

Seized Drugs Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division

Seized Drugs Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division SHIMADZU ULTRAVIOLET/VISIBLE SPECTROPHOTOMETER (UV/VIS) Instrument Shimadzu UV-2401 UV/VIS Spectrophotometer (Serial Number

More information

3700 SERIES USER MANUAL

3700 SERIES USER MANUAL SAFETY GUIDE This manual contains the precautions necessary to ensure your personal safety as well as for protection for the products and the connected equipment. These precautions are highlighted with

More information

Fluorimeter User s Booklet

Fluorimeter User s Booklet Fluorimeter User s Booklet 1) Verify that the instrument power and the computer are both off. Then turn on the lamp. The red indicator light will glow when the lamp is on. Turn on lamp switch here. This

More information

OCEAN OPTICS Chem 2000 Spectrophotometer

OCEAN OPTICS Chem 2000 Spectrophotometer OCEAN OPTICS Chem 2000 Spectrophotometer prepared for: Department of Chemistry and Biochemistry New Mexico State University Las Cruces, NM 88003 last updated May 2003 R.. Cowley TABLE OF CONTENTS I. Introduction...2

More information

AnalytPro. Analysis software. User Manual

AnalytPro. Analysis software. User Manual AnalytPro Analysis software User Manual Contents 1. Introduction... 1 1.1 Main function... 1 1.2 Spectra treatment... 2 2. Installation... 3 2.1 System configuration... 3 2.2 Installation AnalytPro...

More information

Diving into a new dimension: Quality meets efficiency

Diving into a new dimension: Quality meets efficiency Diving into a new dimension: Quality meets efficiency DR 6000 UV-VIS spectrophotometer Combining quality and cost-effectiveness The new DR 6000 UV-VIS spectrophotometer delivers top performance for both

More information

Agilent EZChrom Elite. PDA Analysis

Agilent EZChrom Elite. PDA Analysis Agilent EZChrom Elite PDA Analysis Notices Copyright Scientific Software, Inc 1997-2003 Agilent Technologies, Inc. 2006. No part of this manual may be reproduced in any form or by any means (including

More information

Standard Operating Procedure for the Horiba FluroMax-4

Standard Operating Procedure for the Horiba FluroMax-4 Standard Operating Procedure for the Horiba FluroMax-4 Adapted from Horiba Operations Manual Created by Michael Delcau, Modified by Brian Lamp The Fluoromax is capable of making a variety of measurements.

More information

SPEKOL UV VIS Spectrophotometer

SPEKOL UV VIS Spectrophotometer SPEKOL 1300 UV VIS Spectrophotometer Kundendienst Konrad-Zuse-Str. 1 07745 Jena Phone: Hotline: + 49 (3641) 77-7407 Fax: + 49 (3641) 77-7449 e-mail: service@analytik-jena.de General information about on

More information

Perkin-Elmer Lambda 3B Spectrophotometer

Perkin-Elmer Lambda 3B Spectrophotometer Written by Mark Csele Written by Jeff Brown Operation Turn on the Lambda 3B and the Computer Remove all samples from the sample chamber. Turn on the lamp for the range being investigated using the switches

More information

Turner Spectrophotometer

Turner Spectrophotometer BARNSTEAD THERMOLYNE CORPORATION Turner Spectrophotometer SP-870 OPERATION MANUAL LT1102X6 Model # SM110235 SM110230-33 Voltage 115V 230V with European cord set 6/97 1 2 Table of Contents Introduction...

More information

CHAPTER 2: USING MANAGER PREFERENCE. 2.1 Using Manager Mode Using CSAFE Console (back side) Access Layout...

CHAPTER 2: USING MANAGER PREFERENCE. 2.1 Using Manager Mode Using CSAFE Console (back side) Access Layout... S3x_OM_090507.indd 1 S3x_OM_090507.indd 2 CHAPTER 1: IMPORTANT SAFETY INSTRUCTIONS PAGES 1.1 Read and Save These Instructions... 01 1.2 Setting up the Stepper... 01 1.3 Installation Requirements... 01

More information

Quick Setup & Getting Started

Quick Setup & Getting Started Quick Setup & Getting Started HP Compaq Business PC Copyright 2007 Hewlett-Packard Development Company, L.P. The information contained herein is subject to change without notice. Microsoft, Windows, and

More information

Galaxie Photodiode Array Software

Galaxie Photodiode Array Software Varian, Inc. 2700 Mitchell Drive Walnut Creek, CA 94598-1675/USA Galaxie Photodiode Array Software User s Guide Varian, Inc. 2002-2006 03-914950-00:6 Table of Contents Introduction... 3 General... 3 PDA

More information

6 x 6 Peltier-Thermostatted Multicell Holder Series II Accessory

6 x 6 Peltier-Thermostatted Multicell Holder Series II Accessory 6 x 6 Peltier-Thermostatted Multicell Holder Series II Accessory Overvoltage Category II Pollution Degree 2 Equipment Class III Safety information Introduction Before using this accessory, you must read

More information

7310 and 7315 Glossary of Icons. Operating Manual

7310 and 7315 Glossary of Icons. Operating Manual 7310 and 7315 Glossary of Icons Operating Manual Glossary of Icons Mode ICON Description Back key Tick icon - Done/yes Cross icon Cancel/no Printer icon - Print/open printer settings No results to send

More information

Datalogging RH / Temperature Pen Instruction Manual

Datalogging RH / Temperature Pen Instruction Manual Datalogging RH / Temperature Pen 800013 Instruction Manual 1 Datalogging RH / Temperature Pen 800013 Copyright 2014 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents of this manual

More information

CF3000 Dealer Diagnostic Tool Instruction Manual

CF3000 Dealer Diagnostic Tool Instruction Manual CF3000 Dealer Diagnostic Tool Instruction Manual Table of Contents: About the CF3000......3 Important Precautions......4 Components....5 Charging the CF3000......7 Licensing the CF3000.......8 Updating

More information

Thermo Scientific Multiskan GO User Manual Rev. 1.0

Thermo Scientific Multiskan GO User Manual Rev. 1.0 Thermo Scientific Multiskan GO User Manual Rev. 1.0 Thermo Scientific Multiskan GO User Manual Rev. 1.0, Cat. no. N10588 Copyright Copyright 2010 Thermo Fisher Scientific Corporation. All rights reserved.

More information

2 1.1 Safety using methods Definition of warning symbols Standard accessories Installation...

2 1.1 Safety using methods Definition of warning symbols Standard accessories Installation... Table of Contents 1 Satety precautions... 2 1.1 Safety using methods...2 1.2 Definition of warning symbols...2 2 Standard accessories......6 3 Installation... 7 3.2 Leg frame... 7 3.3 Installation and

More information

KM-4800w. Installation Guide

KM-4800w. Installation Guide KM-4800w Installation Guide TABLE OF CONTENTS page 1 Installation Requirements 2 2 Unpacking 3 2. 1 Unpacking 3 2. 2 Confirmation of Accessories 5 3 Leveling the Machine 7 4 Setup of the Roll Deck 9 5

More information

Operating Procedure PerkinElmer LAMBDA 950 (Integrating sphere Mode)

Operating Procedure PerkinElmer LAMBDA 950 (Integrating sphere Mode) Operating Procedure PerkinElmer LAMBDA 950 (Integrating sphere Mode) National Centre for Photovoltaic Research and Education (NCPRE) Indian Institute of Technology- Bombay Powai, Mumbai-400076 INDIA Operating

More information

ChromQuest 5.0 Quick Reference Guide

ChromQuest 5.0 Quick Reference Guide ChromQuest 5.0 Quick Reference Guide This guide contains an overview of the ChromQuest chromatography data system, with topics organized by workflow. For more information, refer to the ChromQuest User

More information

Contents Contents 2 Preface 3 Introduction 4 Components 5 Operation 6 Timer 7 Lamp Replacement 9 Specification 10 Ordering Information 11 2

Contents Contents 2 Preface 3 Introduction 4 Components 5 Operation 6 Timer 7 Lamp Replacement 9 Specification 10 Ordering Information 11 2 SPOT-LITE CURING SYSTEM USER S GUIDE http://www uvp com UVP, Inc Ultra-Violet Products Ltd 2066 W 11 th Street Unit 1, Trinity Hall Estate, Nuffield Road Upland, CA 91768 USA Cambridge, CB4 1TG, UK (800)

More information

UV WinLab Software The Lambda Series

UV WinLab Software The Lambda Series UV WinLab Software The Lambda Series Key Features Quality software for dependable results Complete instrument and accessory control Powerful spectral arithmetic function Comprehensive Kinetics package

More information

NIR Moisture Analyzer KB-30

NIR Moisture Analyzer KB-30 NIR Moisture Analyzer KB-30 Operation Manual Safety Precautions FPL950701 Improper use of the NIR moisture analyzer in violation of the following safety notes may result in death, injury or damage to property

More information

Manual. User Reference Guide. Analysis Application (EMG) Electromyography Analysis

Manual. User Reference Guide. Analysis Application (EMG) Electromyography Analysis Phone: (888) 765-9735 WWW.MINDWARETECH.COM User Reference Guide Manual Analysis Application Electromyography Analysis (EMG) Copyright 2014 by MindWare Technologies LTD. All Rights Reserved. 1 Phone: (614)

More information

Benchmark Plus Microplate Reader User Manual. This Instrument for Laboratory Use Only Rev C

Benchmark Plus Microplate Reader User Manual. This Instrument for Laboratory Use Only Rev C Benchmark Plus Microplate Reader User Manual This Instrument for Laboratory Use Only 4100172 Rev C 4100172 Rev C Benchmark Plus System with Incubator Catalog Number 170-6930 Benchmark Plus Reader with

More information

MindWare Electromyography (EMG) Analysis User Reference Guide Version Copyright 2011 by MindWare Technologies LTD. All Rights Reserved.

MindWare Electromyography (EMG) Analysis User Reference Guide Version Copyright 2011 by MindWare Technologies LTD. All Rights Reserved. MindWare Electromyography (EMG) Analysis User Reference Guide Version 3.0.12 Copyright 2011 by MindWare Technologies LTD. All Rights Reserved. MindWare EMG 3.0.12 User Guide Internet Support E-mail: sales@mindwaretech.com

More information

TEL: / FAX:

TEL: / FAX: Self-leveling Rotary Laser FRE-211 Operating Manual P.R.Engineering Ltd UK www.laser-level.co.uk TEL: 01246 269 777 / FAX: 01246 260 007 SAFETY PRECAUTIONS: During instrument operation, be careful not

More information

Student Quick Reference Guide

Student Quick Reference Guide Student Quick Reference Guide How to use this guide The Chart Student Quick Reference Guide is a resource for PowerLab systems in the classroom laboratory. The topics in this guide are arranged to help

More information

icore Kiosk system Installation Guide

icore Kiosk system Installation Guide icore Kiosk system Installation Guide The reproduction, transmission or use of this document or its contents is not permitted without express authority. Offenders will be liable for damages. All rights,

More information

cief Data Analysis Chapter Overview Chapter 12:

cief Data Analysis Chapter Overview Chapter 12: page 285 Chapter 12: cief Data Analysis Chapter Overview Analysis Screen Overview Opening Run Files How Run Data is Displayed Viewing Run Data Data Notifications and Warnings Checking Your Results Group

More information

New software features of the 7010 Particle Size. Analyzer Danielle Chamberlin May May 5, Page 1

New software features of the 7010 Particle Size. Analyzer Danielle Chamberlin May May 5, Page 1 New software features of the 7010 Particle Size May 5, 2009 Analyzer 2.0.0 Danielle Chamberlin May 5 2009 Page 1 Introduction to the new 7010 software Instrument operation New look and feel easy to operate

More information

Part No. Z , IA Mar OPERATION MANUAL. High Voltage Digitalmeter A

Part No. Z , IA Mar OPERATION MANUAL. High Voltage Digitalmeter A Part No. Z1-109-820, IA001704 Mar. 2011 OPERATION MANUAL High Voltage Digitalmeter 149-10A Use of Operation Manual Please read through and understand this Operation Manual before operating the product.

More information

CVI SPECTRAL PRODUCTS. 111 Highland Drive Putnam CT, (860) SM32Pro

CVI SPECTRAL PRODUCTS. 111 Highland Drive Putnam CT, (860) SM32Pro CVI SPECTRAL PRODUCTS 111 Highland Drive Putnam CT, 06260 (860) 928-5834 SM32Pro 2.8.28 Table of Contents Warranty and Liability 1 Quick Start Installation Guide 2 System Requirements 3 Requirements for

More information

Quick Start Guide. January EAZ0081L02A Rev. A

Quick Start Guide. January EAZ0081L02A Rev. A Quick Start Guide January 2014 EAZ0081L02A Rev. A Trademarks Snap-on and Vantage Ultra are trademarks of Snap-on Incorporated. All other marks are trademarks or registered trademarks of their respective

More information

Series 400 Colorimetry Spectrophotometer User's Manual P/N 1658 Rev. C

Series 400 Colorimetry Spectrophotometer User's Manual P/N 1658 Rev. C Photonics Series 400 Colorimetry Spectrophotometer User's Manual P/N 1658 Rev. C S.I. Photonics, Inc. Tucson, Arizona Table of Contents Table of Contents 1. Introduction to UV -Vis Spectroscopy...1 1.1

More information

Operation Manual. Concorde 600 Power Supply. *This instrument is intended for laboratory use only.

Operation Manual. Concorde 600 Power Supply. *This instrument is intended for laboratory use only. Concorde 600 Power Supply Operation Manual Cat.no. R10-1001011 *This instrument is intended for laboratory use only http://www.recenttec.com E-mail : support@recenttec.com Version 1.1 Packing List x 1

More information

TRIAX Series Spectrometers

TRIAX Series Spectrometers Getting Started TRIAX Series Spectrometers TRIAX 180/190 TRIAX 320 TRIAX 550 Part Number 81034 Issued February 1999 This manual is intended to provide the first-time user with basic information and simple

More information

Deckblatt von Marketing / Foto fehlt

Deckblatt von Marketing / Foto fehlt R-Biopharm AG RIDA QUICK SCAN II (V2.5) Deckblatt von Marketing / Foto fehlt Bedienungsanleitung 2017-11-04 2 All rights reserved. This manual may be used only for its intended purpose. It may not be reproduced

More information

Part No. Z , IA Nov OPERATION MANUAL. High Voltage Digitalmeter A

Part No. Z , IA Nov OPERATION MANUAL. High Voltage Digitalmeter A Part No. Z1-109-920, IA001723 Nov. 2005 OPERATION MANUAL High Voltage Digitalmeter 149-30A Use of Operation Manual Please read through and understand this Operation Manual before operating the product.

More information

TL6 Ultra Micro-volume Spectrophotometer

TL6 Ultra Micro-volume Spectrophotometer TL6 Ultra Micro-volume Spectrophotometer TL6 is the latest Ultra Micro-volume Spectrophotometer of PERSEE ANALYTICS. It is a small portable instrument for fast analysis of nucleic acid and protein concentrations

More information

FluxGage. FluxGage. LED Luminaire Measurement System User Manual

FluxGage. FluxGage. LED Luminaire Measurement System User Manual FluxGage FluxGage LED Luminaire Measurement System User Manual 1 Acronyms... 3 2 Introduction... 4 2.1 Operation principle... 4 3 Specifications... 5 4 Mechanical and Electrical Installation... 9 4.1 Unpacking...

More information

Installation. Installation 1

Installation. Installation 1 Installation 1 Installation The SP35 printer is easy to install and use. It is also flexible; you can connect and use it in many different ways. Most often, one printer is connected to one PC. However,

More information

Installation Manual. Model: HILU Ultra HD Interactive Flat Panel Display

Installation Manual. Model: HILU Ultra HD Interactive Flat Panel Display Installation Manual Model: HILU750 '' Ultra HD Interactive Flat Panel Display QUICK SETUP GUIDE For further information, see the User Manual. Please contact HITACHI directly if you have questions on the

More information

CDD Carbon Dioxide Transmitter

CDD Carbon Dioxide Transmitter Introduction The OSA CO2 transmitter uses Infrared Technology to monitor CO2 levels within a range of 0 2000 ppm and outputs a linear 4-20 ma or 0-5/0-10 Vdc signal. The enclosure is designed to operate

More information

Thanks for shopping with Improvements! Lighted Canterbury Christmas Greenery Doorway Arch Item #548443

Thanks for shopping with Improvements! Lighted Canterbury Christmas Greenery Doorway Arch Item #548443 Thanks for shopping with Improvements! Lighted Canterbury Christmas Greenery Doorway Arch Item #548443 IMPORTANT, RETAIN FOR FUTURE REFERENCE: READ CAREFULLY. PARTS LIST: 2 Metal Plate Stands 5 Greenery

More information

Operating Procedure for Horiba Raman Microscope

Operating Procedure for Horiba Raman Microscope Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch

More information

1690-Series Logic Analyzers

1690-Series Logic Analyzers 1690-Series Logic Analyzers Quick Start/Installation Installation 2 Making a measurement 4 Snap to Edge Markers 6 Left-Click Menus 7 Tree Structure Labels 7 Offline Analysis 8 Marker Overview 8 Tool Tips

More information

USER S MANUAL Madell Technology Corp Walnut Ave. Suite V Buena Park, CA USA Phone:

USER S MANUAL Madell Technology Corp Walnut Ave. Suite V Buena Park, CA USA Phone: DNM-9602A Microplate Reader USER S MANUAL Madell Technology Corp. 7372 Walnut Ave. Suite V Buena Park, CA 90620 USA Phone: 1-877-670-9023 http://www.madelltech.com 1 CONTENTS 1.Introduction 1.1 Characteristics

More information

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter S-14 Compact Digital Multimeter S-14 Compact Digital Multimeter SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation at this meter

More information

Installation Manual. 65 Interactive LED/LCD. Model: HILF65101 (64.56 )

Installation Manual. 65 Interactive LED/LCD. Model: HILF65101 (64.56 ) Installation Manual 65 (64.56 ) Model: HILF65101 65 Interactive LED/LCD QUICK SETUP GUIDE For further information, see the user manual. Please contact directly if you have questions on the use of the touch

More information

Compatible Visible Spectrophotometer T3/T3M. Operation Instruction Manual

Compatible Visible Spectrophotometer T3/T3M. Operation Instruction Manual Compatible Visible Spectrophotometer T3/T3M Operation Instruction Manual Introduction Thanks for your purchase of the Persee Analytics Inc. T3 spectrophotometer. UV-Visible Spectrophotometer is a well-accepted,

More information

EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM

EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM GAIN 0 Instruction Manual and Experiment Guide for the PASCO scientific Model OS-8537 and OS-8539 02-06575A 3/98 EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM CI-6604A

More information

L300 user manual. Programmable DC Electronic Load CONTENTS. Electronic Load Software CAUTION...2 SAFETY NOTES...2

L300 user manual. Programmable DC Electronic Load CONTENTS. Electronic Load Software CAUTION...2 SAFETY NOTES...2 Programmable DC Electronic Load L300 user manual CONTENTS CAUTION......2 SAFETY NOTES......2 Chapter 1 General Introduction...3 1.1 General Introduction......3 1.2 Specification......3 1.3 Features......3

More information

OPERATING INSTRUCTION

OPERATING INSTRUCTION OPERATING INSTRUCTION AUTORANGING MULTIMETER MAX Ω F C 10A MAX every 15 min. COM V SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation

More information

E1135C PDU and Pod Upgrade Procedure

E1135C PDU and Pod Upgrade Procedure E4030-90010 Rev. B 12/2003 In this Document... Tools Needed, 2 Contents of the Upgrade Kits, 2 Installation Procedures, 4 Verifying the Power Option of the New PDU, 4 Removing the PDU from the Support

More information

TD-700 FLUOROMETER SERVICE MANUAL

TD-700 FLUOROMETER SERVICE MANUAL TD-700 FLUOROMETER SERVICE MANUAL July 1996 CONTENTS Page Section 1 INTRODUCTION 2 Section 2 PRELIMINARY CHECKS 3 Section 3 TROUBLESHOOTING GUIDE 5 A. Lamp (Fluorescent) 5 B. Lamp Heater 7 C. Fan 8 D.

More information

9-pin dot matrix printer

9-pin dot matrix printer 9-pin dot matrix printer All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means, electronic, mechanical, photocopying,

More information

SPECTROMETERS USER MANUAL. Ver. 1.2_09/16. Making spectroscopy brighter

SPECTROMETERS USER MANUAL. Ver. 1.2_09/16. Making spectroscopy brighter SPECTROMETERS USER MANUAL Making spectroscopy brighter 1 Table of contents 1. General Safety Rules... 2 2. Items Supplied... 2 3. Introduction... 3 4. Installation and Operation... 4 4.1. Notes... 7 5.

More information

NIR Moisture Analyzer KB-230

NIR Moisture Analyzer KB-230 NIR Moisture Analyzer KB-230 Operating Manual Contents 1. For safety measurement... 4 2. Features and part names... 6 2-1. Features... 6 2-2. Parts of main unit... 6 2-3. Accessories... 7 3. Specifications...

More information

UWE-1707 SERIES CONTENTS OPERATION MANUAL ELECTRONIC WEIGHING INDICATOR 1. SPECIFICATIONS 2. INSTALLATION

UWE-1707 SERIES CONTENTS OPERATION MANUAL ELECTRONIC WEIGHING INDICATOR 1. SPECIFICATIONS 2. INSTALLATION UWE-1707 SERIES CONTENTS ELECTRONIC WEIGHING INDICATOR OPERATION MANUAL 1. SPECIFICATIONS PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE SCALE 2. INSTALLATION 3. ROUTINE OPERATION

More information

Removal and Installation8

Removal and Installation8 8 Screw Types 8-4 Top Cover Assembly 8-5 Left Hand Cover 8-6 Right Hand Cover 8-10 Front Panel Assembly 8-14 Left Rear Cover 8-15 Right Rear Cover 8-16 Extension Cover (60" Model only) 8-17 Media Lever

More information

FTIR Spectroscopy. (Brucker Vector 22)

FTIR Spectroscopy. (Brucker Vector 22) FTIR Spectroscopy (Brucker Vector 22) Warning: This instrument may only be operated by those who have been trained by AAF staff and have read and signed the AAF laboratory policies. 1.0 PURPOSE: In infrared

More information

CyberComm Pro Data Acquisition Software Installation & User Guide

CyberComm Pro Data Acquisition Software Installation & User Guide CyberComm Pro 2.2.3 Data Acquisition Software Installation & User Guide ph 1100 and ph 2100 Bench ph and Bench ph/ion Meter Technology Made Easy... 68X090822 rev 1 Aug 2002 2 PREFACE Thank you for selecting

More information

FlowAccess TM GETTING STARTED. Windows software for Skalar SAN plus systems. Version and up

FlowAccess TM GETTING STARTED. Windows software for Skalar SAN plus systems. Version and up FlowAccess TM V3 Windows software for Skalar SAN plus systems Version 3.1.0.4 and up GETTING STARTED FlowAccess TM V3 Windows software for Skalar SAN plus systems Version 3.1.0.4 and up Getting started

More information

Agilent ChemStation for UV-visible Spectroscopy

Agilent ChemStation for UV-visible Spectroscopy Agilent ChemStation for UV-visible Spectroscopy Understanding Your Biochemical Analysis Software Agilent Technologies Notices Agilent Technologies, Inc. 2000, 2003-2008 No part of this manual may be reproduced

More information

Thermo Scientific Evolution 60S UV-Visible Spectrophotometer

Thermo Scientific Evolution 60S UV-Visible Spectrophotometer m o l e c u l a r s p e c t r o s c o p y Thermo Scientific Evolution 60S UV-Visible Spectrophotometer Research Performance, Routine Ready Precise Dependable Economical Intuitive Part of Thermo Fisher

More information

CDD4 Series Room CO2 Transmitter Installation Instructions

CDD4 Series Room CO2 Transmitter Installation Instructions CDD4 Series Room CO2 Transmitter Installation Instructions Introduction The CO2 transmitter uses Infrared Technology to monitor CO2 levels and outputs a linear 4-20 ma or 0-5/0-10 Vdc signal. Options include

More information

FROG-4000TM Quick Start Guide

FROG-4000TM Quick Start Guide FROG-4000TM Quick Start Guide 1 Precautions DO NOT invert the instrument with the sparge bottle attached. DO NOT transport or store the instrument with liquid in the sparge bottle. DO NOT handle or carry

More information

Hardware Reference Guide For Thin Clients

Hardware Reference Guide For Thin Clients b Hardware Reference Guide For Thin Clients Document Part Number: 254241-001 July 2002 This book provides basic hardware information for this series of Compaq Thin Clients. Å WARNING: Ä CAUTION: 2002 Compaq

More information

180 DEGREE PEEL TESTER

180 DEGREE PEEL TESTER 180 DEGREE PEEL TESTER MODEL PA-2000-180 OPERATING INSTRUCTIONS CHEMINSTRUMENTS 510 COMMERCIAL DRIVE FAIRFIELD, OHIO 45014 (513) 860-1598 www.cheminstruments.com Revision 1.0 November 5, 2018 CONTENTS

More information

Mercury Tracker-3000 XS

Mercury Tracker-3000 XS Member of the envea Group Mercury Tracker-3000 XS Portable Instrument for Measuring Mercury in Air and Other Gases The new Mercury Tracker-3000 XS lighter - smaller - more comfortable We updated the well-proven

More information