Empower Software Release Notes Appendix Changes Implemented through Service Packs

Size: px
Start display at page:

Download "Empower Software Release Notes Appendix Changes Implemented through Service Packs"

Transcription

1 Empower Software Release Notes Appendix Changes Implemented through Service Packs Problems Fixed with Empower Service Pack A PCS #: In certain circumstances, the Server is unable to access data files that have been buffered and the message, unable to write to the server error is returned. The data files are transferred to the server after the buffering but do not add to the total number of files in the appropriate raw data drive. The Empower message center displays data file checksum errors for data files. Any of the data files that showed in the raw data drive of the server but did not add to the total item count were not accessible anymore through the Empower application. The normal recovery procedure of running instrvx, to force buffered files to the server, does not correct the problem in this case. Service Pack A does not address this issue with the Waters ZQ running in Workgroup or Enterprise (client server) configurations. This will be addressed in a separate Service Pack. PCS #: When running a system, consisting of a 717 autosampler running on IEEE, 515 standalone pump, and a 2475 detector on RS-232, the runtime clock continues past the programmed end time for 10 minutes. The message, instrument fails then appears in the status bar. The problem does not occur if the system does not contain a 717 autosampler running on IEEE or if a SAT/in channel is added to the system. PCS #: A problem recovering data can occur when extracting an MS derived channel and using an MS processing method to create a result in Review. The following sequence will cause the error: 1) Extract an MS derived channel and use an MS processing method to create a result in Review. 2) Save the Result in Review. 3) Delete the 3D MS datafile that was used to create the result. 4) Take the result back into review and review crashes. PCS #: MassLynx tune configuration file TUNEPRM.IPR fails Verify Files after a new MS tune is saved. The tunes file (.ipr) is updated/changed each time a user saves any of the values in the tune page. For this reason, the.ipr file should not be part of the verify file routine. Problems Fixed with Empower Service Pack B PCS #: As more methods were created or copied to a project, access to that project would become slower. This would also affect access to software tools such as the Sample Set Editor. This issue has been resolved in Service Pack B. Problems Fixed with Empower Service Pack C PCS #: Page 1 of 26

2 If BCD values from an autosampler were being read into Empower using a SAT/IN module, it was possible for incorrect values to be reported, namely values of 126 or 127. Valid BCD values range from 00 to 99. This behavior has been resolved in Service Pack C. PCS #: If a toolkit application required the user to select data from Empower, it would assume the use of the Pro interface was appropriate for all data selection, This behavior has been corrected in Service Pack C, now whichever interface the user has open (Pro or QuickStart) will be used for data selection. PCS #: When using the AutoArchive option for Empower, sample archive entries are created for any samples acquired into a monitored project. However, if acquisition through an acquisition server was done first to one database, and then to a second database where fewer custom field were being tracked, no sample archives would be created in the second database. This behavior has been corrected in Service Pack C, and sample archive entries will be created correctly when sequentially acquiring to multiple database from a single acquisition server. PCS #: After buffering data on a network installation from a Waters ZQ mass spectrometer, data files could be seen as missing when reviewing the data in Empower. All appropriate information was stored in/with the database but the message Data file checksum error would be seen in the Empower Message Center. This behavior has been corrected in Service Pack C, and checksum errors due to this buffering defect no longer occur. PCS #: It was possible to see very slow performance on a system acquiring data from a Waters ZQ mass spectrometer. This behavior has been corrected in Service Pack C, and checksum errors due to this buffering defect no longer occur. PCS #: It was possible to see very slow performance on a system acquiring data from a Waters ZQ mass spectrometer and displaying a large number of SIR channels in the real time view of QuickStart. This behavior has been corrected in Service Pack C, and up to four SIR channels may be displayed at one time. PCS #: When intersample calculations were used in projects, processing sample sets could take significantly longer than expected. The time taken would increase as the number of vials and/or results in the project increased This behavior has been corrected in Service Pack C, and the time taken for reprocessing when using intersample calculations has been significantly reduced. PCS #: Under the following circumstances it was possible to use the wrong version of a processing method when reviewing/manually reprocessing data: If a result had a version of the processing method associated with it that was not the most current in the project and this result was taken into the Review application, then a different method were loaded, and then the menu option to Reset Processing Method were selected; the most current version of the method was retrieved from the database instead of the original, older method version associated with the result. This behavior has been corrected in Service Pack C, and correct method version will be retrieved when the Reset Processing Method option is selected. Features Implemented with Empower Service Pack ApexTrack Gaussian skimming is available as a timed event. Page 2 of 26

3 Valley-to-Valley integration is available as a timed event. When System Suitability is used in an ApexTrack method to calculate USP Resolution or USP Plate Count, the Tangent field is determined as the width at baseline between two lines drawn tangent to the peak s inflection points. An S29 processing code is added to the peak s results to indicate that this is how the peak s Tangent is calculated. (When using Traditional integration, Tangent is determined as the width at baseline between two tangent lines drawn at the Tangent Percent specified in the processing method.) The Start and End time parameters (located in an ApexTrack processing method) are graphically displayed in the chromatogram plot in Review when the Show Events tool is used. When these markers are dragged, the appropriate time in the method is automatically updated. The ApexTrack Width field has been eliminated. Important ApexTrack Integration Information Because of improvements made to ApexTrack integration, you may find slightly different results with some peaks, as compared to the same peak processed using the same ApexTrack processing method in the released version of Empower without the Service Pack ApexTrack. These changes occur due to a modification in the ApexTrack algorithm that prevents piercing of the data signal with the baseline and results in a one data point shift in the start time or the end time of some peaks, hence changing the peak area. After installation of the Empower Service Pack ApexTrack, requalification is required. Features Implemented with Empower Service Pack Light Scattering Light Scattering (GPCLS) is a new processing option that can be enabled after the GPC and/or GPCV options have been enabled. The Light Scattering option allows a user who has collected light scattering data and concentration data in the same injection to perform GPCLS processing of Narrow or Broad Unknown samples and calculate molecular weights without having to create GPC calibration curves. If the GPCV option is enabled, users can also perform GPCVLS (triple detection) processing with or without GPC Universal calibration curves. The menu choice Process > Calibrate Light Scattering System initiates a LS Calibration wizard to lead the user through the process of calibrating the light scattering detector. Three new menu choices: Process > GPC-LS Data, Process > GPCV Data, and Process > GPCV-LS Data have been added to a GPC processing. You then need to either (1) apply the current method set to the current injection (if the method set is capable of producing the appropriate type of result) or (2) create and apply a new method set based on the injection's instrument method. Important Information for GPC and GPCV Users For GPC and/or GPCV users who do not need the new Light Scattering functionality, it is important to note that several GPC and GPCV problems that are listed below have been fixed. Also listed are several new features and improvements to the GPCV software. GPCV Features Implemented with Empower Service Pack Light Scattering PCS #: When the Fitting Model is set to Calibration, there are 2 new Intrinsic Viscosity Fit choices in the Slicing Table: 1st Order Step and 2nd Order Step. PCS #: Page 3 of 26

4 The Viscometer channels attached to GPCV results are marked visible by default in the 2D channel table of Review (as in V3.20 Millennium32). PCS #: When the alpha LCB in the slicing table is set either too high or too low so that there is a problem calculating g', you now get a V45 processing code (The alpha LCB is too high or too low for the fitted viscosity curve.) instead of a V04 processing code (There is a problem calculating g'.). PCS #: K LCB was added to the slicing table to go with the already existing alpha LCB value. PCS #: In Review, the Autoscale button and plot menu entry now set the Normalize X checkbox to false instead of true in the Plot properties for the 2d chromatogram plot. Setting the Normalize X checkbox to false allows the chromatograms to overlay each other using the same X-axis. PCS #: The processing method wizard in Review now allows you to edit the integration parameters for all the channels in a GPCV result. PCS #: There is a new item in the Process menu in Review, GPCV Data. The Process > GPCV Data menu item creates new Universal GPC processing method using the processing method wizard if needed and then either (1) applies the current method set to the current injection if the method set is capable of producing a GPCV result or (2) creates and applies a new method set based on the injection's instrument method. If a new GPCV processing method is created via the Process > GPCV Data menu item, the flow rate is copied from the instrument method into the new processing method. GPC Problems Fixed with Empower Service Pack Light Scattering PCS #: You can now take Broad GPC results in to Review, Preview, and Reporting and view the distribution plots and/or tables. PCS #: You can now take GPC distributions into Compare and view the distribution plots and/or tables. PCS #: A narrow standard peak quantitated against a GPC Universal calibration curve is now given the correct MP because the K and alpha in the sample are no longer ignored. GPC and GPCV Problem Fixed with Empower Service Pack Light Scattering PCS #: When processing in the Review windows of QuickStart with a GPC processing method containing more than 1 row in the slicing table, the broad peak is now identified correctly. GPCV Problems Fixed with Empower Service Pack Light Scattering PCS #: GPCV Narrow Standard processing will now skip and not assign molecular weights to any peak that does not have a corresponding viscometer peak. PCS #: Page 4 of 26

5 In the Method Set tree of the method set window of Review (or the method set editor) when the method set lists an instrument method containing a GPCV instrument, you can now view the Properties of one of the Imported data channels via the right mouse menu. PCS #: If alpha LCB is entered in the slicing table, the overall peak g' is used to calculate Slice lambda instead of the Slice g'. PCS #: The overall peak g' is now calculated when the fitting model is Structural, the intrinsic viscosity fit is 1st Order, and there are user-entered K and alpha values for the sample. PCS #: If the fitting model is Structural, the intrinsic viscosity fit is 1st Order, and the Branching Model is Star, the overall Peak g' and g values are now calculated. PCS #: GPCV Broad Standard calibration now adds points to the curve because the calculation of fitted viscosity has been fixed when the intrinsic viscosity fit is 1st Order and the fitting model is Calibration. PCS #: In the default GPCV project for Empower, the instrument method containing the GPCV instrument needs to be recreated using the View Instrument Method menu item in Review after the imported data's injection(s) have been brought in to Review. Known Problems in Service Pack Light Scattering PCS #: The overall value peak g' is not calculated for broad GPC-LS peaks when there is a user-entered epsilon. This cannot be fixed in a service pack because the fix requires a change to the database, but a user can create a custom field to calculate peak g' either from peak g and epsilon or by summing the slice g' values. PCS #: There is a new selection in the report method wizard to create a Light Scattering report method, but the report that is currently generated by this method is a basic GPC report which does not include Light Scattering (LS) fields and plots. PCS #: The peak [n]w value is not calculated for GPC-LS peaks when there is a Universal calibration curve. This cannot be fixed in a service pack because the fix requires a change to the database. PCS #: The default precision of the RI Sensitivity field in Sample Loading or alter Sample needs to be increased from 0 to at least 2. The help states the following "For example, suppose a 2410 RI Detector is calibrated at a sensitivity setting of 4. If a row in the Samples table specifies an instrument method with the 2410 sensitivity set to 16, enter 0.25 in the RI Sensitivity column of the Samples table. If a row in the Samples table specifies an instrument method with the 2410 sensitivity set to 1, enter 4 in the RI Sensitivity column of the Samples table." A user is not currently able to enter a value of This defect was observed in Millennium32 v4.00 and Empower. Problems Fixed with Empower Service Pack D PCS #28812 (User Access To Acquisition Server Properties) A user who has access to a system on an acquisition server but does not have access to the acquisition server properties can access those properties by right clicking on the system itself. While the OK box is grayed out, the owner of the acquisition server is changed to SYSTEM from a non system user. Users who do not have access to an acquisition server will not be allowed to view the acquisition server properties and be presented with a warning dialog box. Page 5 of 26

6 PCS #28777 (Project Access) Via the Project Properties Access tab, a System Administrator sets project access to a specific group in the system. A user logs into the system with custom field access only, accesses the Project Properties Access tab and notes that the Group User Type is changed to "Users Own Group". The user selects OK and resets the Group User Type to Users Own Group from the specific system group previous entered by the System Administrator. PCS #29854 (Raw Data Files) In conditions where network performance is extremely high and the LACE 32 has been up and active for an extended period of time buffering over 1000 files, it is possible to observe a database entry where no D- files have been created. PCS #29984 (RPC Server Errors) During recent internal evaluation for Empower it was discovered that a situation could develop on a LAC/E 32 such that the Microsoft Remote Procedure Call (RPC) server becomes unavailable, placing the LAC/E 32 into a constant state of buffering. The Microsoft RPC programming functions are used to forward application function calls to a remote system across the network. While this Waters Corporation service pack corrects these particular RPC errors, Microsoft Corporation Windows 2000 Service Pack 3 also corrects other RPC errors. If you are experiencing other specific RPC errors, Waters recommends installing Windows 2000 Service Pack 3. PCS #28443 (Needle Was Time) When trying to modify the needle wash time in the 2695 s Default Separation Method, the time remained unaffected at Normal. In Millennium 32 V4.0 the needle wash time is taken from the default Alliance method but in Empower this is not occurring, forcing the Normal needle wash time. PCS #28502 (Margin Issue In Reports) Customers upgrading from Millennium 32 V3.2 to version 4.0/or Empower Software reported their reports were cut off along the left edge. PCS #30142 (Background Processing) When processing Agilent 1100 data you may see slight differences in peak area counts depending on the application used to process the data ( Review or the Process and Report tool). To determine if a difference exists you would need to expand the precision of the area and height columns (in the Project window or in a report) to at least three decimal places. Features Implemented with Instrument Support Service Pack 1 Empower Instrument Support Service Pack 1 lays the groundwork for implementation of laboratory instrument control using a private Ethernet network. The Service Pack provides additional support for COM (Component Object Model) instruments and support for Ethernet control that will be introduced shortly for a number of instruments. Operation of any instrument via Ethernet will require the installation of Instrument Support Service Pack 1 as a prerequisite. Instruments that will be supported on Ethernet include: Waters 2420 Evaporative Light Scattering Detector, and the Waters EMD 1000 Mass Detector. The Service Pack also provides an updated XML schema for COM instruments that use an HTML Instrument Method editor (A1100 modules, for example). Page 6 of 26

7 Supporting instruments in a private Ethernet environment requires that IP addresses be assigned. This task is handled by the Waters DHCP Server, which is greatly enhanced in this Service Pack. The Waters DHCP Server utility operates only on the network card that the user identifies as the card that is attached to a private network of laboratory instruments. The Waters DHCP Server will only assign IP addresses to Waters instruments or to Ethernet instruments that have been specifically identified. Problems Fixed with Instrument Support Service Pack When the Light Scattering Service Pack installed, a derived channel cannot be created for a GPC project. After installation of Instrument Support Service Pack, derived channels can be created in GPC projects An audit trail comment is required after canceling an operation within the users properties page. After installation of the Service Pack, audit trail comments are not required if the action is cancelled Under certain circumstances, audit trail comments for one action could be used for another action without asking the user for another comment. After installation of the Service Pack, the user will be asked for audit trail comments and the comments will be assigned properly The Waters DHCP Server does not have a user interface and remote (client/server) capabilities. After installation of the Service Pack, Waters DHCP Server has a user interface accessible from a Tab in Acquisition Server Properties COM instruments are not uniquely identified within the software. After installation of the Service Pack, COM instruments are uniquely identified within the software Pressing the Scan for Instruments button does not cause the CapLC pump to be recognized. After installation of the Service Pack, the CapLC pump is recognized The Condition Column special control function causes a COM instrument to return a negative value for datastart. After installation of the Service Pack, the correct value is returned Instrument Method editors for COM instruments do not function properly when two or more COM instruments are in a chromatographic system. After installation of the Service Pack, the Instrument Method editor functions properly. Known Issues with Instrument Support Service Pack When more than one type of Ethernet instrument is in use, scanning for instruments may intermittently fail to detect all instruments The Project MFC Application message pops up when the Project Window closes after buffering has occurred. Click OK to exit the application and then click Don t Send After reinstalling Windows XP, Empower, and Instrument Support Service Pack 1, the message Waters DHCP Server Chinese Resource.DLL has encountered a problem and needs to close If this message appears, please continue with the installation and reboot the computer. If the message appears after rebooting, contact Waters customer support. Features Implemented with Instrument Support Service Pack 2 Empower Instrument Support Service Pack 2 fixes several defects in Empower. These defects are listed on page 4 of this document. Additionally, Instrument Support Service Pack 2 provides support for the Agilent A1100 DA detector (DAD models G1315A or G1315B) in 3D mode using Ethernet connectivity. In addition to this service pack, an A1100 Plus License (key disk) is required in order to use an A1100 system containing an A1100 DAD to acquire 3D spectral data. The A1100 Plus License (key disk) is not required in order to install the defect fixes contained in this service pack. Problems Fixed with Empower Instrument Support Service Pack 2 PCS #: 30602: Abort and Change Vial Number not Documented Properly Page 7 of 26

8 When you abort a sample set (using the Abort Now option) and then change a vial number and resume acquisition, the injection occurs from the correct vial, but the original vial is displayed in the software (Project Window, Review, Reports, etc.) Additionally, when you resume acquisition, the row you were running when the abort occurred should be displayed in red but is not. This problem is fixed in Empower Instrument Support Service Pack 2. When this service pack is applied, the vials are documented correctly in the software and the lines in Run Samples highlight in red appropriately. PCS #: 30828: Instrument Method Problems for A1100 Using VWD. When a locked instrument method (containing a variable wavelength detector (VWD)) is specified in Run Samples and the user edits the instrument method when Run Samples is not in an Idle state (i.e. Run Samples is Setting up ), the original instrument method parameters are executed during acquisition, not the modified parameters. This problem is fixed in Empower Instrument Support Service Pack 2. When this service pack is applied, the correct A1100 instrument method parameters are used. PCS #: 28995: Legend does not Print when Printing from Review. When printing a chromatogram from Review, the legend (specified in the Plot Properties) does not print. This problem is fixed in Empower Instrument Support Service Pack 2. When this service pack is applied, chromatogram legends print from Review. PCS #: 31059: Create Project OK to All Copies Previous Comments into Audit Trail. After installing Empower Instrument Support Service Pack 1, which fixes an Audit Trail issue in full audit trail projects (PCS # 30037), one symptom of the original problem still remains. During project creation, two audit trail entries are added to the System Audit Trail. If the user selects OK to All after entering the audit trail comment during project creation, only one of the project creation audit trail entries contains the correct comment. The other project creation audit trail entry contains a comment from a previous, irrelevant audit trail entry. This problem is fixed in Empower Instrument Support Service Pack 2. When this service pack is applied, the audit trail comments are correct for project creation audit trail entries. Known Issues with Empower Instrument Support Service Pack 2 PCS # 31167: File not removed when ISSP2 is Uninstalled. When Empower ISSP2 and all previous Empower service packs are installed, and then only ISSP2 is uninstalled, the Empower\Instruments\HTML\A1100\A1100-PostRunRpt-r100.xsl file is not always removed. The A1100-PostRunRpt-r100.xsl file is a new file for ISSP2 and it is only used by the A1100.dll. If the file is not removed during uninstallation, there is no negative impact on the system. PCS # 30252: Shutdown Method Gives Instrument Failure with A1100 DAD System When a shutdown method is specified with an A1100 DAD system, an instrument failure may occur at the end of the sample set. To avoid this problem, add a zero volume injection line to the end of your sample set. This line should contain a method set with an instrument method to shutdown the system. The method on this line will run for the specified run time. The run time should therefore be set long enough to allow the shutdown changes to occur. Do not incorporate your shutdown parameters directly in the instrument method that you are routinely using on each injection line because when using an A1100 system, the instrument method stops running at the end of the run time and the system reverts back to initial conditions. (Conversely, when using a system containing Waters equipment, the instrument method continues to execute after the specified run time, until another injection occurs.) PCS # 29881: Condition Column Function Not Supported with A1100 Systems When a Condition Column function is present in a sample set using an A1100 system, the sample set will remain on the Condition Column line indefinitely and will not advance to the next line in the sample set. The Condition Column function is not supported with an A1100 system in Millennium32 4.0, Empower or Empower Instrument Support 2 Service Pack. PCS # 30947: Help Does Not Display Page 8 of 26

9 When attempting to launch Help, a message occurs stating that the help file cannot be opened or that it cannot be displayed. If this occurs, locate and delete X:\Empower\Help\Master.GID (where X is the location where Empower is installed.) PCS # 30288: External Contact Closure Event Card is Not Supported in DAD when the LAN Card is Present The external contacts are not available in the DAD when a LAN card is present. The Event card should be placed in another A1100 module; the contacts can be used from there. PCS # 29945: Must Follow Proper Sequence When Deleting A1100 Ethernet Instrument Before deleting A1100 Ethernet instruments from the Instruments tab of the Acquisition Server properties, it is necessary to remove the DHCP entry from the DHCP Configuration Page (using the Configure DHCP tab). Next, remove the A1100 instrument from the Instruments tab. PCS # 30256: Apply Button when no Changes Made During an equilibration, if the user clicks the Change Flow button but makes no changes, if the Apply button is then clicked a message indicating that the flow rate changed is posted to the Message Center. In this scenario, it is not necessary to click Apply. Click Cancel instead. PCS # 30946: Entry of Invalid Bandwidth It is possible to enter an invalid bandwidth (i.e. 0) by entering the value and then saving the method without clicking somewhere else in table. Problems Fixed with Empower Toolkit Service Pack Fix defect PCS #28591 (Toolkit Memory Leak) Fixes a Memory leak surrounding the ResultViewer object. When calling and releasing the ResultViewer and associated objects, system memory will be appropriately released. Fix defect PCS #28480 (Quick Start Interface) Allow full functionality of the Quick Start Interface via Toolkit. When a user logs in to Empower via the Toolkit Project object Login method, Toolkit should check what allowed interfaces the user has and open the appropriate interface based on their Default User Interface setting. This also caused problems when fetching result id information based on the mtkselected option of the ResultViewer. Through further investigation of this defect, changes were also made to the Project Object LaunchMilApp Method. Detail changes on this fix: A fourth optional parameter was added to the Project Object s LaunchMilApp method, mtkuserinterface ui. The Method now is: LaunchMilApp(mtkMillenniumApp milapp, mtkconnectoptions connectoption, VARIANT varargs, mtkuserinterface ui), Where mtkuserinterface is an enumeration of (mtkdefault = 0, mtkquickstart = 1, mtkpro = 2, mtkopenaccess = 3, mtkweb = 4). This allows the ToolKit user to specify which interface they wish to launch, with some caveats. - If the optional argument isn t supplied, Toolkit will look at the supplied User s default interface and use that (and if that selection is set to use the system default, it will use the System Policy Default UI setting). - If the optional argument supplied is mtkdefault, Toolkit will launch via the System Policy Default UI setting. - If the optional argument supplied is mtkpro or mtkquickstart, Toolkit will check to see if the given user has the appropriate privileges for that, if not it will return an error. - In the case where mtkopenaccess or mtkweb is specified or defaulted to, an error will be returned as these are not supported interfaces via ToolKit. Page 9 of 26

10 Note: There are timing issues with launching applications ( i.e. Project, Review) with the QuickStart interface, whereas the behavior is not consistent when launching with a particular view, i.e. Launching Project/QuickStart/Channels may display Find Data/Samplesets view. In these cases the user will have to make an additional selection manually to the appropriate tab view. Fix defect PCS #28474 (Apex Track) When Apex Track is enabled there are different Integration Parameters and Events in Empower. However, the Visual Basic object browser will only show fields corresponding to Traditional Integration. In Empower there are two sets of integration parameters, Apex Track and Traditional. When editing a processing method with Apex Track integration parameters the ToolKit will load the Apex Track fields for the IntegrationParameters object, when editing a processing method with Traditional integration the ToolKit will load those fields. This is seamless to the ToolKit user; you will simply see the appropriate fields based on which kind of processing method you are editing. Fix defect PCS #29096 (LaunchMillApp Method) The Project object LaunchMillApp method has an mtkcreateifnecessary connect option. If this option is used, the application is only started if it is not already running based on the Login parameters. Currently Toolkit will ignore the mtkcreateifnecessary option and create a new application window. Using this mtkcreateifnecessary connect option will now operate correctly and not create a new application window. Fix defect PCS #29563 (Enforce Unique User Account Names) If attempting to create a user through the Toolkit User object and the system policy Enforce Unique User Account Names is set, the Toolkit will not display the appropriate message box stating User name x already in use. The system creates the account by appending a number 1 to the end of the duplicate name attempting to be created. Toolkit will now return an error if that user name is already in use. Toolkit will also check to ensure the account name meet s the minimum length requirements, if not, an error will be returned. Fix defect PCS #29565 (Project Object Login) If attempting to login via the Toolkit Project object, the Toolkit will not display the appropriate errors for the different login conditions. The following list details these conditions and the revised Toolkit action: 1. A new user account is created via Empower or Toolkit and Password Required was enabled. When attempting to log into Empower via Toolkit for the first time using this new account, the Toolkit did not display a message box stating Your Password has Expired. Toolkit will now display an error that the password has expired. The user must log in through the Empower Software to correct this. By design, the Toolkit can not create or modify passwords. This must be done through the Empower Software. 2. If the System Policy Enforce Minimum Password Length of: x characters is enabled and a user logs into Empower via Toolkit with a password not meeting this policy, Toolkit did not display a message box stating, The password for this account no longer meets current system policies for length. Toolkit will now display this message box and direct the user to change their password by logging in through Empower Software. By design, the Toolkit can not create or modify passwords. This must be done through the Empower Software. 3. If the System Policy Passwords Expire every: x days is enabled and a user logs into Empower via Toolkit with a password not meeting this policy, Toolkit did not display a message box stating, Your password will expire in x day(s). Toolkit will now display an error when the password has expired. The user must log in through the Empower Software to correct this. By design, the Toolkit can not create or modify passwords. This must be done through the Empower Software. Fix defect PCS #29594 (Result and Sample Information Changes) Page 10 of 26

11 The New Project Policies tab for Empower System Polices has a "Result Changes" and "Sample Information Changes" check box under 'Default Full Audit Trail Settings' - 'Full Audit Trail Support'. When these are checked and Result or Sample Information changes are made via the Toolkit the user is not prompted to enter a reason for these changes as they are in the base application. If the Sample Information Changes System Policies is enabled, sample information changes made via Toolkit will now prompt the user to enter a reason for these changes. Toolkit programmers may automate this response in code using the Toolkit RevisionComment Property. Setting this Property will automatically enter the comment and not prompt the user for an entry. Note: Results cannot be changed via Toolkit therefore checking this has no effect on Toolkit. Fix defect PCS #29675 (Copy From Non-FAT to FAT Projects) When the System Policy, "Don t allow user to copy from non-fat projects into FAT projects" is checked, any Toolkit object that allows the user to copy between projects should be checked. Currently if this is enabled Toolkit users can still copy between non-fat and FAT projects. When an inappropriate copy is now performed, Toolkit displays the message Can t copy from non-fat projects into FAT projects. Fix defect PCS #28711 (User Object) When creating a user via the Toolkit User object, you are restricted from starting the name with a numeric character. This is allowed within the base Empower application and should function the same within Toolkit. Users now created via the Toolkit User object are not restricted from starting the name with a numeric character. Fix defect PCS #30959 (KPRIME) When the System Suitability and GPC/LS (Light Scattering) options are enabled, both options contain a KPRIME field which causes a conflict in Toolkit. Due to the conflict, Toolkit does not return any KPRIME value. However, Toolkit works fine when only one of the options is enabled. The following list details the Peak field name mappings from Empower Software to Toolkit when the GPC/LS Option is enabled: - Empower Software Toolkit K(LS) KLS Alpha(LS) AlphaLS Linear K(LS) LinearKLS Linear Alpha(LS) LinearAlphaLS Mz(LS) MzLS Mz+1(LS) MzPlus1LS Note: There was no change made to the System Suitability mappings. Fix defect PCS #31170 (Data Acquisition Privileges) User accounts that do not have the Data Acquisition Privilege are able to perform data acquisition via a Toolkit application. With the Data Acquisition Privilege enabled, User accounts will not be able to perform data acquisition (i.e. run a single injection or sample set). Fix defect PCS #31242 (Copy Preferences) When you create a new Project through Empower, You can 'copy preferences' from another project. In Toolkit, there is no property to set for copying preferences. A new Boolean Property was added to the DBProject Object called CopyPreferences, and works the same as other DBProject Copy Methods such as CopyMethods, CopyViewFilters, etc. If CopyPreferences is set to True when creating a new project using the DBProject.Copy method, Toolkit will copy the existing options and preferences from the specified source project. Note: Existing options includes the following options built into the base Empower Software: Photo Diode Array, Mass Spectrometry and CE/CIA. Page 11 of 26

12 New Features Added with Empower Toolkit Service Pack PCS #28590 (Summarize Function) The Empower feature Summarize Custom Fields was not implemented via Toolkit. The Constant information is as follows: Identifier Value Used By mtksummarizecustom 39 Static Fields Type Custom Menu Item Argument Toolkit Custom Menu Items were created to provide users access to any application from within the Empower Software. The last parameter is a space-delimited list of command line arguments that are passed to the executable upon startup. These arguments are placed in the registry key for the custom menu item and would expand during program runtime. As an example %p expands to the current project name, expand meaning the current project name is added to the command line during runtime in place of the %p parameter. A new Custom Menu command line parameter of %a has been created. This would expand to the current acquisition server node when launched from Run Samples (Pro Interface) or Quick Start Interface. Com Instrument XML Interface A new object InstSetupCom will be added to InstSetups family. This object is unlike other InstSetup objects in that it will handle multiple instrument setups defined as COM instruments. COM instruments that save their instrument setup data as XML will be handled by this new object. The InstSetupCOM object will be a general purpose object designed to expose Empower instrument setups that are stored as XML (i.e. new COM instruments) to Toolkit users. Unlike existing InstSetup objects that are tailored to the individual instrument, (InstSetup600, InstSetup2790, etc.) this object will handle multiple types of instruments via a simple set of commands. The InstSetupCOM object will allow a Toolkit user to fetch and store the XML blob that contains the instrument method parameters for a particular COM instrument. The XML data will be fetched and stored as a single entity. The Toolkit Programmer is responsible for validating any and all modifications made to the XML prior to setting the XML property. There will be little or NO error checking and/or range validation done on the XML being sent down to the database. The Toolkit Programmer should parse through the XML schema capturing each field s minimum and maximum parameter, which can then be validated in your code prior to modification. The XML schema (whose extension is.xsd) for each COM instrument is supplied by Waters and can be found in the appropriate directory for that instrument under the Empower/Instruments/HTML subdirectory. The InstSetupCOM object derives from the base class InstSetup in order to allow polymorphic handling of all types of InstSetups in any Instrument Method. The InstSetupCom object has all the properties and methods of the InstSetup object and the following additional property. NOTE: The Type property will always return mtkcom for any and all COM based instruments. XML() as String Set: Sets the XML method data for the instrument Get: Returns the XML method data for the instrument Default: N/A IEEE Instrument Setup for Waters 486, 410, 2410, 2414, 2790/5 and Agilent 5890/7673, 5890S/7673S, 6890 GC s - InstSetup486 The InstSetup486 encapsulates the setup information for the Waters Corporation 486 instrument. The InstSetup486 object has all the properties and methods as described by InstSetup. Its dynamic fields are specific to the 486 instrument setup. It also has the following additional properties. EventLines as EventLine486S Page 12 of 26

13 Set: Sets the value for this event lines collection. Get: Returns the current collection of event lines. Default: An EventLine486S collection containing the event line information for this InstSetup InstSetup410 The InstSetup410 encapsulates the setup information for the 410 instrument. The InstSetup410 object has all the properties and methods as described by InstSetup. Its dynamic fields are specific to the 410 instrument setup. Note: The 2410 and 2414 show up as a 410 instrument in Empower. - InstSetup2790 The InstSetup2790 encapsulates the setup information for the 2790/5 instrument. The InstSetup2790 object has all the properties and methods as described by InstSetup. Its dynamic fields are specific to the 2790 instrument setup. It also has the following additional properties. EventLines as EventLine2790S Set: Sets the value for this event lines collection. Get: Returns the current collection of event lines. Default: An EventLine2790S collection containing the event line information for this InstSetup2790. GradientLines as GradientLines Set: Sets the value for this gradient lines collection. Get: Returns the current collection of gradient lines. Default: A GradientLines collection containing the gradient line information for this InstSetup InstSetup5890 The InstSetup5890 encapsulates the setup information for the HP 5890 and 5890S series instruments. The InstSetup5890 object has all the properties and methods as described by InstSetup. Its dynamic fields are specific to the 5890 and 5890S instrument setups. It also has the following additional properties and sub objects. RampLines as RampLine5890S Set: Sets the value for this ramp lines collection. Get: Returns the current collection of ramp lines. Default: An RampLine5890S collection containing the ramp line information for this InstSetup5890. Channels as Channel5890S Get: Returns the current collection of channels. Default: A Channel5890S collection containing the channels information for this InstSetup5890. RampLine5890S Dynamic collection of RampLine5890 objects. Used by InstSetup5890 object to manipulate the ramp lines of a HP 5890 series instrument setup. There are no additional properties or methods. RampLine5890 The RampLine5890 object encapsulates a HP 5890 series ramp line object. It contains the fields specific to a HP 5890 series ramp line. Fields() as Fields Get: Returns the current collection of Fields. Default: A Fields collection containing all the RampLine5890 fields, FieldType is mtkeventline. XXX([FormatData as Variant]) as Variant Dynamic property where XXX stands for the name of the field whose value is to be returned. Only the names of the fields in the Fields property can be used. All RampLine5890 dynamic properties can use both Set() and Get() functions. After a property is set, its get will reflect the current setting; yet, it will only be officially stored in the database at the InstrumentMethod Store(). Page 13 of 26

14 If the optional FormatData parameter is set to non-zero, then the returned data will be a formatted string using the Empower width and precision specifications. Methods: Set(FieldName, Value) FieldName as string: Contains name of the dynamic field whose value is to be assigned. Action: This provides another way to set a dynamic property. Using object.xxx = val in Visual Basic is equivalent to doing object.set( xxx, val). The Set gives greater programmatic flexibility. Get(FieldName, [FormatData]) as Variant FieldName as string: Contains name of the dynamic field whose value is to be retrieved. FormatData as BOOLEAN: Set True if data is to be returned as a Empower formatted string. Default value - False. Action: Returns the value of the desired dynamic field name. This provides another way to retrieve a dynamic property. Using val = object.xxx in Visual Basic is equivalent to doing val = object.get( xxx ). The Get gives greater programmatic flexibility. Channel5890S Dynamic collection of Channel5890 objects. Used by InstSetup5890 object to manipulate the channels of a HP 5890 series instrument setup. There are no additional properties or methods. Channel5890 The Channel5890 object encapsulates a HP 5890 series channel object. It contains the fields specific to a HP 5890 series channel. Fields() as Fields Get: Returns the current collection of Fields. Default: A Fields collection containing all the Channel5890 fields, FieldType is mtkischannel. XXX([FormatData as Variant]) as Variant Dynamic property where XXX stands for the name of the field whose value is to be returned. Only the names of the fields in the Fields property can be used. All Channel5890 dynamic properties can use both Set() and Get() functions. After a property is set, its get will reflect the current setting; yet, it will only be officially stored in the database at the InstrumentMethod Store(). If the optional FormatData parameter is set to non-zero, then the returned data will be a formatted string using the Empower width and precision specifications. Methods: Set(FieldName, Value) FieldName as string: Contains name of the dynamic field whose value is to be assigned. Action: This provides another way to set a dynamic property. Using object.xxx = val in Visual Basic is equivalent to doing object.set( xxx, val). The Set gives greater programmatic flexibility. Get(FieldName, [FormatData]) as Variant FieldName as string: Contains name of the dynamic field whose value is to be retrieved. FormatData as BOOLEAN: Set True if data is to be returned as a Empower formatted string. Default - False. Action: Returns the value of the desired dynamic field name. This provides another way to retrieve a dynamic property. Using val = object.xxx in Visual Basic is equivalent to doing val = object.get( xxx ). The Get gives greater programmatic flexibility. - InstSetup6890 The InstSetup6890 encapsulates the setup information for the HP 6890 series instrument. The InstSetup6890 object has all the properties and methods as described by InstSetup. Its dynamic fields are specific to the 6890 instrument setup. It also has the following additional properties. FrontDetector as HP6890Detector Page 14 of 26

15 Get: Returns the current FrontDetector object. Default: A HP6890Detector object containing the front detector information for this InstSetup6890. BackDetector as HP6890Detector Get: Returns the current BackDetector object. Default: An HP6890Detector object containing the back detector information for this InstSetup6890. FrontInjector as HPInjector Get: Returns the current FrontInjector object. Default: An HPInjector object containing the front Injector information for this InstSetup6890. BackInjector as HPInjector Get: Returns the current BackInjector object. Default: An HPInjector object containing the back Injector information for this InstSetup6890. FrontInlet as HP6890Inlet Get: Returns the current FrontInlet object. Default: An HP6890Inlet object containing the front Inlet information for this InstSetup6890. BackInlet as HP6890Inlet Get: Returns the current BackInlet object. Default: An HP6890Inlet object containing the back Inlet information for this InstSetup6890. Column1 as HP6890Column Get: Returns the current Column1 object. Default: An HP6890Column object containing the column information for this InstSetup6890. Column2 as HP6890Column Get: Returns the current Column2 object. Default: An HP6890Column object containing the column information for this InstSetup6890. RampLines as RampLine6890S Set: Sets the value for this ramp lines collection. Get: Returns the current collection of ramp lines. Default: An RampLine6890S collection containing the ramp line information for this InstSetup6890. Channels as Channel6890S Get: Returns the current collection of channels. Default: A Channel6890S collection containing the channels information for this InstSetup6890. RampLine6890S Dynamic collection of RampLine6890 objects. Used by InstSetup6890 object to manipulate the ramp lines of a HP 6890 series instrument setup. There are no additional properties or methods. RampLine6890 The RampLine6890 object encapsulates a HP 6890 series ramp line object. It contains the fields specific to a HP 6890 series ramp line. Fields() as Fields Get: Returns the current collection of Fields. Default: A Fields collection containing all the RampLine6890 fields, FieldType is mtkeventline. XXX([FormatData as Variant]) as Variant Dynamic property where XXX stands for the name of the field whose value is to be returned. Only the names of the fields in the Fields property can be used. All RampLine6890 dynamic properties can use both Set() and Get() functions. Page 15 of 26

16 After a property is set, its get will reflect the current setting; yet, it will only be officially stored in the database at the InstrumentMethod Store(). If the optional FormatData parameter is set to non-zero, then the returned data will be a formatted string using the Empower width and precision specifications. Methods: Set(FieldName, Value) FieldName as string: Contains name of the dynamic field whose value is to be assigned. Action: This provides another way to set a dynamic property. Using object.xxx = val in Visual Basic is equivalent to doing object.set( xxx, val). The Set gives greater programmatic flexibility. Get(FieldName, [FormatData]) as Variant FieldName as string: Contains name of the dynamic field whose value is to be retrieved.. FormatData as BOOLEAN: Set True if data is to be returned as a Empower formatted string. Default value - False. Action: Returns the value of the desired dynamic field name. This provides another way to retrieve a dynamic property. Using val = object.xxx in Visual Basic is equivalent to doing val = object.get( xxx ). The Get gives greater programmatic flexibility. Channel6890S Dynamic collection of Channel6890 objects. Used by InstSetup6890 object to manipulate the channels of a HP 6890 series instrument setup. There are no additional properties or methods. Channel 6890 The Channel6890 object encapsulates a HP 6890 series channel object. It contains the fields specific to a HP 6890 series channel. Fields() as Fields Get: Returns the current collection of Fields. Default: A Fields collection containing all the Channel6890 fields, FieldType is mtkischannel. XXX([FormatData as Variant]) as Variant Dynamic property where XXX stands for the name of the field whose value is to be returned. Only the names of the fields in the Fields property can be used. All Channel6890 dynamic properties can use both Set() and Get() functions. After a property is set, its get will reflect the current setting; yet, it will only be officially stored in the database at the InstrumentMethod Store(). If the optional FormatData parameter is set to non-zero, then the returned data will be formatted string using the Empower width and precision specifications. Methods: Set(FieldName, Value) FieldName as string: Contains name of the dynamic field whose value is to be assigned. Action: This provides another way to set a dynamic property. Using object.xxx = val in Visual Basic is equivalent to doing object.set( xxx, val). The Set gives greater programmatic flexibility. Get(FieldName, [FormatData]) as Variant FieldName as string: Contains name of the dynamic field whose value is to be retrieved. FormatData as BOOLEAN: Set True if data is to be returned as a Empower formatted string. Default value - False. Action: Returns the value of the desired dynamic field name. This provides another way to retrieve a dynamic property. Using val = object.xxx in Visual Basic is equivalent to doing val = object.get( xxx ). The Get gives greater programmatic flexibility. HP6890Inlet The HP6890Inlet object encapsulates a HP 6890 series inlet object. It contains the fields specific to a HP 6890 series inlet. Fields() as Fields Get: Returns the current collection of Fields. Page 16 of 26

Waters Empower Software Build 1154 Toolkit Service Pack Release Notes

Waters Empower Software Build 1154 Toolkit Service Pack Release Notes Waters Empower Software Build 1154 Toolkit Service Pack Release Notes These release notes are intended as supplemental information and should be used in conjunction with the Empower Software Release Notes.

More information

Empower Toolkit Whitepaper March 2009

Empower Toolkit Whitepaper March 2009 Empower Toolkit Whitepaper March 2009 Empower 2 Introduction... 6 Project Object & User Object Changes... 6 System Audit Trail Policy Changes... 7 Project Audit Trail Policy Changes... 10 Project Hierarchies...

More information

Empower Light Scattering Software

Empower Light Scattering Software Empower Light Scattering Software Getting Started Guide 34 Maple Street Milford, MA 01757 71500043903, Revision A NOTICE The information in this document is subject to change without notice and should

More information

For additional information, see these documents on the Breeze 2 software DVD:

For additional information, see these documents on the Breeze 2 software DVD: R E L E A S E N O T E S Waters Breeze 2 Read these release notes before you install and operate Breeze 2. They contain information necessary to ensure successful liquid chromatography results, including:

More information

Waters Empower Software A1100 System Configuration Instructions for Ethernet Control

Waters Empower Software A1100 System Configuration Instructions for Ethernet Control Waters Empower Software A1100 System Configuration Instructions for Ethernet Control In order to configure an A1100 system for Ethernet control or to acquire A1100 DAD 3D data in Empower Software, it is

More information

What s New in Empower 3

What s New in Empower 3 What s New in Empower 3 Revision A Copyright Waters Corporation 2010 All rights reserved Copyright notice 2010 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND IN IRELAND. ALL RIGHTS RESERVED.

More information

SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.31: PSS

SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.31: PSS SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.31: PSS0051-05 Introduction This document details the installation and known issues with OMNISEC v10.31. This release is a maintenance release to

More information

New features and Workflow Enhancements Empower 3 Software

New features and Workflow Enhancements Empower 3 Software New features and Workflow Enhancements Empower 3 Software Mogens Hallas Nordic User Training 2011 Långvik, September 6th to 9th, 2011 2011 Waters Corporation 1 Empower 3 Architecture and operating systems

More information

Chromatography Software Training Materials. Contents

Chromatography Software Training Materials. Contents Chromatography Software Training Materials This document contains information on how to build a method, start the instrument to acquire data, and then process the data using the Galaxie Program. You will

More information

Agilent Instrument Control Framework (ICF) Update A Driver Update 1 (DU1) Hot Fix 2 (HF2)

Agilent Instrument Control Framework (ICF) Update A Driver Update 1 (DU1) Hot Fix 2 (HF2) R E L E A S E N O T E S Agilent Instrument Control Framework (ICF) Update A.02.03 Driver Update 1 (DU1) Hot Fix 2 (HF2) These release notes explain the installation of the updated Agilent Instrument Control

More information

Waters Headspace Control Option, Version 3.0

Waters Headspace Control Option, Version 3.0 R E L E A S E N O T E S Waters Headspace Control Option, Version 3.0 Version 3.0 of the Headspace Control Option (HCO) contains instrument drivers for Agilent Headspace Samplers and a toolkit program that

More information

Data Acquisition with CP-2002/2003 Micro-GC Control

Data Acquisition with CP-2002/2003 Micro-GC Control Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA 94598 Star Chromatography Workstation Version 6 Data Acquisition with CP-2002/2003 Micro-GC Control Operation Manual Varian, Inc. 2002

More information

Operating Manual. High Performance Liquid Chromatograph. Scientific Equipment Center, Prince Of Songkla University

Operating Manual. High Performance Liquid Chromatograph. Scientific Equipment Center, Prince Of Songkla University Operating Manual High Performance Liquid Chromatograph Agilent 1100 ; VWD, DAD, FLD and RID Scientific Equipment Center, Prince Of Songkla University Operating Manual High Performance Liquid Chromatograph

More information

Quick Guide to the Star Bar

Quick Guide to the Star Bar Saturn View Saturn Writer Quick Guide to the Star Bar Last active Chromatogram in SaturnView Last active method in the Method Editor Click with the right mouse button on the Star Bar to get this menu Sample,

More information

Validation Sample Set Generator

Validation Sample Set Generator Validation Sample Set Generator General Information This tool is an example program which demonstrates some capabilities of the Empower Toolkit. It was not created using Waters Software Development Lifecycle.

More information

ChromQuest 5.0 Quick Reference Guide

ChromQuest 5.0 Quick Reference Guide ChromQuest 5.0 Quick Reference Guide This guide contains an overview of the ChromQuest chromatography data system, with topics organized by workflow. For more information, refer to the ChromQuest User

More information

SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.30: PSS

SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.30: PSS SOFTWARE UPDATE NOTIFICATION (SUN) OMNISEC SOFTWARE v10.30: PSS0051-04 Introduction This document details the installation and known issues with OMNISEC v10.30. This release is a major release to fix a

More information

Waters Empower 2 Service Pack G

Waters Empower 2 Service Pack G R E L E A S E N O T E S Waters Empower 2 Service Pack G Empower 2 Service Pack G enhances LDAP capabilities. These release notes also address the following information and instructions: System and software

More information

MassLynx 4.2 SCN 989 Release Notes

MassLynx 4.2 SCN 989 Release Notes MassLynx 4.2 SCN 989 Release Notes Page 1 of 20 Waters, THE SCIENCE OF WHAT S POSSIBLE., ACQUITY, Alliance, MassLynx, nanoacquity, and Xevo are registered trademarks and OpenLynx, Quanpedia, and TargetLynx

More information

Agilent 6400 Series Triple Quadrupole LC/MS System

Agilent 6400 Series Triple Quadrupole LC/MS System Agilent 6400 Series Triple Quadrupole LC/MS System Quick Start Guide Where to find information 4 Getting Started 6 Step 1. Start the Data Acquisition software 7 Step 2. Prepare the LC modules 13 Step 3.

More information

UniPoint System Software User s Guide

UniPoint System Software User s Guide UniPoint System Software User s Guide LT2137 1998 Gilson, Inc. All rights reserved Exercise 3-Creating operations list... 4-22 Create and name operations list... 4-22 Set up steps... 4-23 Identify sample

More information

Clarity. Clarity Demo. Code/Rev.: M003/60A Date: Fax: Petrzilkova 2583/ Prague 5

Clarity. Clarity Demo. Code/Rev.: M003/60A Date: Fax: Petrzilkova 2583/ Prague 5 Clarity Clarity Demo Clarity ENG Code/Rev.: M003/60A Date: 21.4.2015 Phone: +420 251 013 400 DataApex Ltd. Fax: +420 251 013 401 Petrzilkova 2583/13 clarity@dataapex.com 158 00 Prague 5 www.dataapex.com

More information

EZ IQ. Chromatography Software PN P1A

EZ IQ. Chromatography Software PN P1A O P E R A T I N G M A N U A L EZ IQ Chromatography Software Trademarks The trademarks of the products mentioned in this manual are held by the companies that produce them. Windows is a registered trademark

More information

Agilent 7890 GC Instrument Control Software Version 2.01

Agilent 7890 GC Instrument Control Software Version 2.01 R E L E A S E N O T E S Agilent 7890 GC Instrument Control Software Version 2.01 These release notes explain how to install Waters Instrument Control Software (ICS), version 2.01, for the Agilent 7890

More information

Agilent 7890 GC Instrument Control Software Version 2.5

Agilent 7890 GC Instrument Control Software Version 2.5 R E L E A S E N O T E S Agilent 7890 GC Instrument Control Software Version 2.5 The Agilent 7890 GC Instrument Control Software version 2.5 enables dual-tower injection capability and provides support

More information

[ WHITE PAPER ] System Policies and Privileges for Processing Data in Empower Software with Data Integrity in Mind INTRODUCTION

[ WHITE PAPER ] System Policies and Privileges for Processing Data in Empower Software with Data Integrity in Mind INTRODUCTION System Policies and Privileges for Processing Data in Empower Software with Data Integrity in Mind Explore how Empower Chromatography Data Software System Policies and privileges can be used to control

More information

Agilent 7890 GC Instrument Control Software Version 2.6

Agilent 7890 GC Instrument Control Software Version 2.6 R E L E A S E N O T E S Agilent 7890 GC Instrument Control Software Version 2.6 Version 2.6 of the instrument control software (ICS) for the Agilent 7890 GC addresses several issues, provides support for

More information

irc Pro (Not for diagnostic use)

irc Pro (Not for diagnostic use) irc Pro (Not for diagnostic use) User guide 2016 2 Table of Contents ircpro... 3 Overview... 3 Release notes... 4 Requirements... 5 Setup... 6 Setup by DVD... 7 Setup by Internet... 8 step 1 - Install

More information

NOTICE: This document contains references to Varian. Please note that Varian, Inc. is now part of Agilent Technologies. For more information, go to

NOTICE: This document contains references to Varian. Please note that Varian, Inc. is now part of Agilent Technologies. For more information, go to NOTICE: This document contains references to Varian. Please note that Varian, Inc. is now part of Agilent Technologies. For more information, go to www.agilent.com/chem. Varian MS Workstation 6.9 Service

More information

Galaxie Workstation Configuration Manager

Galaxie Workstation Configuration Manager Varian, Inc. 2700 Mitchell Drive Walnut Creek, CA 94598-1675/USA Galaxie Workstation Configuration Manager User s Guide Varian, Inc. 2002-2005 Printed in U.S.A. 03-914973-00:Rev. 3 Table of Contents Using

More information

Agilent ChemStation Plus

Agilent ChemStation Plus Agilent ChemStation Plus Getting Started Guide Agilent Technologies Notices Agilent Technologies, Inc. 2004, 2006-2008 No part of this manual may be reproduced in any form or by any means (including electronic

More information

Manual HPLC-4 Alliance-UV

Manual HPLC-4 Alliance-UV Manual HPLC-4 Alliance-UV MAKE SURE NEVER TO: 1. pump particles or air through the column. 2. inject when the injector is dry. 3. inject solutions containing particles. 4. apply large (sudden) pressure

More information

Thermo Xcalibur Getting Started (Quantitative Analysis)

Thermo Xcalibur Getting Started (Quantitative Analysis) Thermo Xcalibur Getting Started (Quantitative Analysis) XCALI-97207 Revision B September 2010 2010 Thermo Fisher Scientific Inc. All rights reserved. Xcalibur, Surveyor, and Accela are registered trademarks

More information

Guidelines for Setting Auto-Peak Width in ApexTrack Processing Methods and its Effect on USP Resolution and USP Plate Count

Guidelines for Setting Auto-Peak Width in ApexTrack Processing Methods and its Effect on USP Resolution and USP Plate Count TECN10008147 New Doc. No. Rev. 01 Page 1 of 6 Guidelines for Setting Auto-Peak Width in ApexTrack Processing Methods and its Effect on USP Resolution and USP Plate Count The information in this article

More information

Procedure to set up an HPIB 82350B card on an M57. M58 or C20x PC. Table 1: Parts required. Part Number Description Qty

Procedure to set up an HPIB 82350B card on an M57. M58 or C20x PC. Table 1: Parts required. Part Number Description Qty 1 OF 21 Procedure to set up an HPIB 82350B card on an M57. M58 or C20x PC Parts required Table 1: Parts required Part Number Description Qty 289000764 Kit, HPIB Card, Inter, PCI 1 Procedure The procedure

More information

Agilent OpenLAB CDS Waters Alliance HPLC Instrument Control Add-On. Installation, Upgrade and Migration Guide

Agilent OpenLAB CDS Waters Alliance HPLC Instrument Control Add-On. Installation, Upgrade and Migration Guide Agilent OpenLAB CDS Waters Alliance HPLC Instrument Control Add-On Installation, Upgrade and Migration Guide Notices Agilent Technologies, Inc. 2013, 2014 No part of this manual may be reproduced in any

More information

Analyst QS. Administrator s Guide. Part Number: B July 2004

Analyst QS. Administrator s Guide.   Part Number: B July 2004 Analyst QS Administrator s Guide Part Number: 1001930 B July 2004 http://www.appliedbiosystems.com This document is provided to customers who have purchased MDS Sciex equipment to use in the operation

More information

Data Handling and Reports

Data Handling and Reports Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA 94598-1675/USA Star Chromatography Workstation Version 6 Data Handling and Reports Tutorials Varian, Inc. 2002 03-914736-00:Rev. 4 Trademark

More information

ChromQuest 4.2 Chromatography Data System

ChromQuest 4.2 Chromatography Data System ChromQuest 4.2 Chromatography Data System Installation Guide CHROM-97200 Revision A April 2005 2006 Thermo Electron Corporation. All rights reserved. Surveyor is a registered trademark and ChromQuest is

More information

Agilent Instrument Control Framework A for Agilent LC/CE and GC Instruments in Non-Agilent Chromatography Data Systems Release Notes

Agilent Instrument Control Framework A for Agilent LC/CE and GC Instruments in Non-Agilent Chromatography Data Systems Release Notes Contents Agilent Instrument Control Framework A.02.01 for Agilent LC/CE and GC Instruments in Non-Agilent Chromatography Data Systems Release Notes Instrument Control Framework (ICF) Revision A.02.01 [085]

More information

Impact Analysis for Software changes in OpenLAB CDS A.01.04

Impact Analysis for Software changes in OpenLAB CDS A.01.04 Impact Analysis for Software changes in OpenLAB CDS A.01.04 Document Information: Filename Product Identifier Product Revision Project Identifier Document Revision Impact_Analysis_OpenLAB_CDS_A0104 OpenLAB

More information

67 Series Spectrophotometer PC Software

67 Series Spectrophotometer PC Software 67 Series Spectrophotometer PC Software Instruction Manual 670 004/Rev D/04-17 IMPORTANT Please ensure the SD Card is fitted into the socket as detailed in the image below. -------------------------------------------------------------------------------------------------------------------------

More information

Getting low-resolution GC-MS data on the JEOL GCMate Introduction

Getting low-resolution GC-MS data on the JEOL GCMate Introduction Getting low-resolution GC-MS data on the JEOL GCMate Introduction GC-MS is a technique that combines gas phase separation and mass-based detection/characterization of analytes. It s a very sensitive method

More information

Soprane II Documentation

Soprane II Documentation Soprane II Documentation p. 1 / 198 Table of contents 1. INTRODUCTION 6 2. INSTALLATION 6 3. ANALYZER CONFIGURATION 8 3.1. Creation 9 3.1.1. Analyzer creation with network connection 11 3.1.2. Analyzer

More information

Release Notes Life Technologies Attune NxT Software v2.3

Release Notes Life Technologies Attune NxT Software v2.3 Release Notes Life Technologies Attune NxT Software v2.3 In the following pages you will find instructions describing: New software features Known software/system issues with troubleshooting guidance Software

More information

. Waters Corporation. Functionality ... April, 2004

. Waters Corporation. Functionality ... April, 2004 .. Waters Corporation Empower Software ality.......... April, 2004 Table Of Contents SECTION 1 - INTRODUCTION........ 1 Basic Principles........ 1 What Makes Empower Unique?... 1 Multiple User Interfaces........

More information

Agilent 1290 Infinity II Evaporating Light Scattering Detector (ELSD) in Empower Environment

Agilent 1290 Infinity II Evaporating Light Scattering Detector (ELSD) in Empower Environment Agilent 1290 Infinity II Evaporating Light Scattering Detector (ELSD) in Empower Environment Technical Note Technical Guide for the configuration and use of the Agilent Evaporating Light Scattering Detector

More information

Chromeleon-PA Analyzer User s Guide Dionex Corporation

Chromeleon-PA Analyzer User s Guide Dionex Corporation Chromeleon-PA Analyzer User s Guide 2003 Dionex Corporation Document No. 031964 Revision 01 October 2003 2003 by Dionex Corporation All rights reserved worldwide. Printed in the United States of America.

More information

OpenLynx User's Guide

OpenLynx User's Guide OpenLynx User s Guide OpenLynx User's Guide Version 4.0 Waters part No 715000405 Micromass Part No - 6666670 5 February, 2002 i OpenLynx User s Guide OpenLynx User s Guide The software described in this

More information

MassLynx 4.1 SCN 804. Release Notes. SQ Detector 2

MassLynx 4.1 SCN 804. Release Notes. SQ Detector 2 R E L E A S E N O T E S MassLynx 4.1 SCN 804 Release Notes SQ Detector 2 43464 Rev 01 Copyright Waters Corporation 2011. All rights reserved. Copyright 2011 Waters Corporation 43464_1. Waters is a registered

More information

Agilent EZChrom Elite. User s Guide

Agilent EZChrom Elite. User s Guide Agilent EZChrom Elite User s Guide Notices Copyright Scientific Software, Inc 1997-2005 Agilent Technologies, Inc. 2006. No part of this manual may be reproduced in any form or by any means (including

More information

C h r o m S t a r 7. M a n u a l

C h r o m S t a r 7. M a n u a l C h r o m S t a r 7 M a n u a l ÿþýü Software für Chromatographie und Prozess-Analytik GmbH Am Weidufer 32, D-28844 Weyhe Tel.: 0421-802806, Fax: 0421-890346 E-mail: info@scpa.de, Internet: www.scpa.de

More information

Chromatography Systems News Flash

Chromatography Systems News Flash Number: Prepared By: Date: DHF-2005-12 Gary Burce September 20, 2005 Chromatography Systems News Flash Worldwide GC and LC Distribution PE TotalChrom vs. Varian Galaxie Recently, we were fortunate to have

More information

Access Gateway Client User's Guide

Access Gateway Client User's Guide Sysgem Access Gateway Access Gateway Client User's Guide Sysgem AG Sysgem is a trademark of Sysgem AG. Other brands and products are registered trademarks of their respective holders. 2013-2015 Sysgem

More information

DESCRIPTION. DataApex Clarity 32bit multi instrument chromatography station for Windows

DESCRIPTION. DataApex Clarity 32bit multi instrument chromatography station for Windows wwwdataapexcom e-mail: clarity@dataapexcom DESCRIPTION DataApex Clarity 32bit multi instrument chromatography station for Windows Multi-detector instruments Easy to install and easy to use Graphical user

More information

LTQ Velos SP3 Release Notes

LTQ Velos SP3 Release Notes LTQ Velos 2.6.0 SP3 Release Notes These release notes briefly list changes in the LTQ Velos 2.6.0 SP3 experiment setup, tuning, and calibration application; fixes to previously reported issues; and known

More information

Breeze 2 Installation and Configuration Guide

Breeze 2 Installation and Configuration Guide Breeze 2 Installation and Configuration Guide 715001951/Revision B Copyright Waters Corporation 2008 2009 All rights reserved Copyright notice 2008 2009 WATERS CORPORATION. PRINTED IN THE UNITED STATES

More information

R e p o r t E d i t o r M a n u a l

R e p o r t E d i t o r M a n u a l R e p o r t E d i t o r 2. 1 M a n u a l Report Editor - Contents Page 1-1 Contents Page 1. Introduction...1-2 2. Description of the Menu Procedures...2-1 2.1 File...2-2 2.2 Edit...2-3 2.3 Format...2-4

More information

Agilent MassHunter Qualitative Data Analysis

Agilent MassHunter Qualitative Data Analysis Agilent MassHunter Qualitative Data Analysis Qualitative Navigator B.08.00 Presenters: Kevin Costalunga Stephen Harnos With Matt Leyden & Kevin Costalunga 1 MassHunter Webinar Series MassHunter Qualitative

More information

Getting Started. Environmental Analysis Software

Getting Started. Environmental Analysis Software Getting Started Environmental Analysis Software Environmental Software Overview The Environmental software is a data acquisition and data analysis package designed to assist you in complying with United

More information

Galaxie Photodiode Array Software

Galaxie Photodiode Array Software Varian, Inc. 2700 Mitchell Drive Walnut Creek, CA 94598-1675/USA Galaxie Photodiode Array Software User s Guide Varian, Inc. 2002-2006 03-914950-00:6 Table of Contents Introduction... 3 General... 3 PDA

More information

IBM Tivoli Storage Manager HSM for Windows Version 7.1. Messages

IBM Tivoli Storage Manager HSM for Windows Version 7.1. Messages IBM Tivoli Storage Manager HSM for Windows Version 7.1 Messages IBM Tivoli Storage Manager HSM for Windows Version 7.1 Messages Note: Before using this information and the product it supports, read the

More information

Silk Performance Manager Installation and Setup Help

Silk Performance Manager Installation and Setup Help Silk Performance Manager 18.5 Installation and Setup Help Micro Focus The Lawn 22-30 Old Bath Road Newbury, Berkshire RG14 1QN UK http://www.microfocus.com Copyright 2004-2017 Micro Focus. All rights reserved.

More information

N2KExtractor. Maretron Data Extraction Software User s Manual

N2KExtractor. Maretron Data Extraction Software User s Manual N2KExtractor Maretron Data Extraction Software User s Manual Revision 3.1.6 Copyright 2017 Maretron, LLP All Rights Reserved Maretron, LLP 9014 N. 23rd Ave #10 Phoenix, AZ 85021-7850 http://www.maretron.com

More information

Thermo Scientific. GRAMS Envision. Version 2.1. User Guide

Thermo Scientific. GRAMS Envision. Version 2.1. User Guide Thermo Scientific GRAMS Envision Version 2.1 User Guide 2013 Thermo Fisher Scientific Inc. All rights reserved. Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase

More information

CTECS Connect 2.2 Release Notes December 10, 2009

CTECS Connect 2.2 Release Notes December 10, 2009 (Formerly VTECS) CTECS Connect 2.2 Release Notes December 10, 2009 This document contains information that supplements the CTECS Connect 2.2 documentation. Please visit the CTECS Connect Support area of

More information

Contact: Systems Alliance, Inc. Executive Plaza III McCormick Road, Suite 1203 Hunt Valley, Maryland Phone: / 877.

Contact: Systems Alliance, Inc. Executive Plaza III McCormick Road, Suite 1203 Hunt Valley, Maryland Phone: / 877. Contact: Systems Alliance, Inc. Executive Plaza III 11350 McCormick Road, Suite 1203 Hunt Valley, Maryland 21031 Phone: 410.584.0595 / 877.SYSALLI Fax: 410.584.0594 http://www.systemsalliance.com http://www.siteexecutive.com

More information

CLARITY DEMO. Clarity Demo Software. Code/Rev.: M003/70E Date:

CLARITY DEMO. Clarity Demo Software. Code/Rev.: M003/70E Date: CLARITY DEMO Clarity Demo Software ENG Code/Rev.: M003/70E Date: 24.10.2017 Phone: +420 251 013 400 DataApex Ltd. Fax: +420 251 013 401 Petrzilkova 2583/13 clarity@dataapex.com 158 00 Prague 5 www.dataapex.com

More information

Waters 2707 Autosampler ICS v1.20 Installer

Waters 2707 Autosampler ICS v1.20 Installer R E L E A S E N O T E S Waters 2707 Autosampler ICS v1.20 Installer Read these release notes before you install and operate the Waters 2707 autosampler instrument control software (ICS). They contain information

More information

Waters ACQUITY QDa Instrument Component Software version 1.54

Waters ACQUITY QDa Instrument Component Software version 1.54 R E L E A S E N O T E S Waters ACQUITY QDa Instrument Component Software version 1.54 The ACQUITY QDa Detector is a single-quadrupole mass detector. The instrument component software (ICS) adds the capability

More information

Agilent EZChrom Elite. PDA Analysis

Agilent EZChrom Elite. PDA Analysis Agilent EZChrom Elite PDA Analysis Notices Copyright Scientific Software, Inc 1997-2003 Agilent Technologies, Inc. 2006. No part of this manual may be reproduced in any form or by any means (including

More information

Agilent ChemStation. ECM Interface Guide. Agilent Technologies

Agilent ChemStation. ECM Interface Guide. Agilent Technologies Agilent ChemStation ECM Interface Guide Agilent Technologies Notices Agilent Technologies, Inc. 2004, 2005-2007 No part of this manual may be reproduced in any form or by any means (including electronic

More information

cief Data Analysis Chapter Overview Chapter 12:

cief Data Analysis Chapter Overview Chapter 12: page 285 Chapter 12: cief Data Analysis Chapter Overview Analysis Screen Overview Opening Run Files How Run Data is Displayed Viewing Run Data Data Notifications and Warnings Checking Your Results Group

More information

Agilent MassHunter Easy Access Software

Agilent MassHunter Easy Access Software Agilent MassHunter Easy Access Software Quick Start Guide Where to Find More Information 3 Installation 4 Sharing Configurations with Multiple Instruments 10 Installed Files 11 Using Easy Access Software

More information

Release Notes Life Technologies Attune NxT Software v2.2

Release Notes Life Technologies Attune NxT Software v2.2 Release Notes Life Technologies Attune NxT Software v2.2 In the following pages you will find instructions describing: New software features Known software/system issues with troubleshooting guidance Software

More information

Agilent G2727AA LC/MS Data Browser Quick Start

Agilent G2727AA LC/MS Data Browser Quick Start What is Data Browser? Agilent G2727AA LC/MS Data Browser Quick Start A guide to get started with Data Browser Use this quick reference guide as a road map for your first steps with the Data Browser software,

More information

Agilent MassHunter Workstation Software 7200 Accurate-Mass Quadrupole Time of Flight GC/MS

Agilent MassHunter Workstation Software 7200 Accurate-Mass Quadrupole Time of Flight GC/MS Agilent MassHunter Workstation Software 7200 Accurate-Mass Quadrupole Time of Flight GC/MS Familiarization Guide Before you begin 3 Prepare your system 3 Prepare the samples required for data acquisition

More information

EASY-nLC Series User Guide for the Xcalibur Data System (Xcalibur version 2.1 or later)

EASY-nLC Series User Guide for the Xcalibur Data System (Xcalibur version 2.1 or later) Thermo EASY-nLC Series User Guide for the Xcalibur Data System (Xcalibur version 2.1 or later) P/N 60053-97230 Revision B January 2013 2013 Thermo Fisher Scientific Inc. All rights reserved. EASY-nLC,

More information

SCN 848 Release Notes

SCN 848 Release Notes R E L E A S E N O T E S SCN 848 Release Notes 43198 Rev 1 Copyright Waters Corporation 2011. All rights reserved. 43198 Rev 1. Waters is a registered trademark, Waters the Science of What s Possible and

More information

Navigator Software User s Manual. User Manual. Navigator Software. Monarch Instrument Rev 0.98 May Page 1 of 17

Navigator Software User s Manual. User Manual. Navigator Software. Monarch Instrument Rev 0.98 May Page 1 of 17 User Manual Navigator Software Monarch Instrument Rev 0.98 May 2006 Page 1 of 17 Contents 1. NAVIGATOR SOFTWARE 2. INSTALLATION 3. USING NAVIGATOR SOFTWARE 3.1 STARTING THE PROGRAM 3.2 SYSTEM SET UP 3.3

More information

EZChrom Elite Compact. What s New in EZChrom Elite Compact Service Pack 2

EZChrom Elite Compact. What s New in EZChrom Elite Compact Service Pack 2 EZChrom Elite Compact What s New in EZChrom Elite Compact 3.3.2 Service Pack 2 Notices Agilent Technologies, Inc. 2010 No part of this manual may be reproduced in any form or by any means (including electronic

More information

A. About this Manual

A. About this Manual Section A. About this Manual A. About this Manual 1. Table of Contents 1. Table of Contents... 1 2. How to Use this User Manual... 2 3. General Introduction to the PAL3 System... 3 4. PAL3 System Specifications...

More information

EASY-nLC Virtual Instrument for Xcalibur 2.1.x Quick Reference Guide

EASY-nLC Virtual Instrument for Xcalibur 2.1.x Quick Reference Guide EASY-nLC Virtual Instrument for Xcalibur 2.1.x Quick Reference Guide This guide describes how to install, configure, and use the Thermo EASY-nLC Virtual Instrument (VI) software with the Thermo Xcalibur

More information

Equipment Overview: Safety:

Equipment Overview: Safety: PerkinElmer Series 200 HPLC: Operating Procedure 12/2/11 R.C Equipment Overview: This document outlines the basic procedure for using the Sedex light scattering detector and the PE series 200 HPLC system

More information

Maxum Workstation Update Release Notes

Maxum Workstation Update Release Notes 1 August 9, 2017 Maxum Workstation 5.30.01 Update Release Notes Requirements Maxum Workstation Software will operate best on computers equipped with a multi-core processor, with at least 4 GB RAM (8+ GB

More information

Agilent GC/MSD Instructions

Agilent GC/MSD Instructions Agilent GC/MSD Instructions GC/MS SAMPLE PREPARATION: Sample Components to Avoid Completely: The following should never be injected: metals, strong acids or bases, salts, oligomeric and polymeric material.

More information

Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017

Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017 Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017 Introduction This document is only for operation of the HPLC and ICP-MS instruments to obtain data. It contains no maintenance

More information

Agilent ChemStation Plus

Agilent ChemStation Plus Agilent ChemStation Plus Getting Started Guide Agilent Technologies Notices Agilent Technologies, Inc. 2004 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Instruction Manual XLComm Data Acquisition Software For XL Series Meters

Instruction Manual XLComm Data Acquisition Software For XL Series Meters Data Acquisition Software For XL Series Meters (XL 15, 20, 25, 30, 40, 50 and 60 meters) 68X090832 Rev. 0 JAN 07 Preface This manual serves to explain the use of the Acquisition Software. The manual functions

More information

OpenLAB CDS Report Template Editor

OpenLAB CDS Report Template Editor OpenLAB CDS Report Template Editor Concepts Guide Agilent Technologies Notices Agilent Technologies, Inc. 2009-2016 No part of this manual may be reproduced in any form or by any means (including electronic

More information

StarFinder. Operation Manual. Star Chromatography Workstation Version 6

StarFinder. Operation Manual. Star Chromatography Workstation Version 6 Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA 94598-1675/USA Star Chromatography Workstation Version 6 StarFinder Operation Manual Varian, Inc. 2002 03-914723-00:Rev. 2 Trademark Acknowledgment

More information

FractionLynx Software

FractionLynx Software FractionLynx Software User s Guide 34 Maple Street Milford, MA 01757 71500054302, Revision A NOTICE The information in this document is subject to change without notice and should not be construed as a

More information

TraceFinder Analysis Quick Reference Guide

TraceFinder Analysis Quick Reference Guide TraceFinder Analysis Quick Reference Guide This quick reference guide describes the Analysis mode tasks assigned to the Technician role in Thermo TraceFinder analytical software. For detailed descriptions

More information

Retention Time Locking with the MSD Productivity ChemStation. Technical Overview. Introduction. When Should I Lock My Methods?

Retention Time Locking with the MSD Productivity ChemStation. Technical Overview. Introduction. When Should I Lock My Methods? Retention Time Locking with the MSD Productivity ChemStation Technical Overview Introduction A retention time is the fundamental qualitative measurement of chromatography. Most peak identification is performed

More information

Release Notes for Agilent Drivers for Thermo Scientific Chromeleon 7 (Rev. 1.1)

Release Notes for Agilent Drivers for Thermo Scientific Chromeleon 7 (Rev. 1.1) Release Notes for Agilent Drivers for Thermo Scientific Chromeleon 7 (Rev. 1.1) Release Notes for Agilent Drivers for Thermo Scientific Chromeleon 7 (Rev. 1.1) This document provides information on the

More information

PDA EXTENSION. Clarity Extension. Code/Rev.: M054/80A Date: 6/8/2018. Fax: Petrzilkova 2583/ Prague 5

PDA EXTENSION. Clarity Extension. Code/Rev.: M054/80A Date: 6/8/2018. Fax: Petrzilkova 2583/ Prague 5 PDA EXTENSION Clarity Extension ENG Code/Rev.: M054/80A Date: 6/8/2018 Phone: +420 251 013 400 DataApex Ltd. Fax: +420 251 013 401 Petrzilkova 2583/13 clarity@dataapex.com 158 00 Prague 5 www.dataapex.com

More information

Cliquid Software for Routine Analysis

Cliquid Software for Routine Analysis Cliquid 3.0.1 Software for Routine Analysis Release Notes Table of Contents Known Issues... 2 LC Systems...2 Initial Log On... 2 Continuous Usage Limitation... 2 Acquisition Fails due to Locked Windows

More information

Spectrometer Visible Light Spectrometer V4.4

Spectrometer Visible Light Spectrometer V4.4 Visible Light Spectrometer V4.4 Table of Contents Package Contents...3 Trademarks...4 Manual Driver and Application installation...5 Manual Application Installation...6 First Start of the Application...8

More information

Thermo Scientific. GRAMS Envision. Version 2.0. User Guide. Revision A

Thermo Scientific. GRAMS Envision. Version 2.0. User Guide. Revision A Thermo Scientific GRAMS Envision Version 2.0 User Guide Revision A 2010 Thermo Fisher Scientific Inc. All rights reserved. Thermo Fisher Scientific Inc. provides this document to its customers with a product

More information

FlowAccess TM GETTING STARTED. Windows software for Skalar SAN plus systems. Version and up

FlowAccess TM GETTING STARTED. Windows software for Skalar SAN plus systems. Version and up FlowAccess TM V3 Windows software for Skalar SAN plus systems Version 3.1.0.4 and up GETTING STARTED FlowAccess TM V3 Windows software for Skalar SAN plus systems Version 3.1.0.4 and up Getting started

More information