syngo MR E11 Operator Manual Ortho Answers for life.

Size: px
Start display at page:

Download "syngo MR E11 Operator Manual Ortho Answers for life."

Transcription

1 syngo MR E11 Operator Manual Ortho Answers for life.

2

3 syngo MR E11 Operator Manual Ortho

4 Legend Indicates a hint Is used to provide information on how to avoid operating errors or information emphasizing important details Indicates the solution of a problem Is used to provide troubleshooting information or answers to frequently asked questions Indicates a list item Indicates a prerequisite Is used for a condition that has to be fulfilled before starting a particular operation Indicates a one-step operation Indicates steps within operating sequences Italic Is used for references and for table or figure titles Is used to identify a link to related information as well as previous or next steps Bold Blue Courier Courier Menu > Menu Item <variable> Is used to identify window titles, menu items, function names, buttons, and keys, for example, the Save button Is used to emphasize particularly important sections of the text Is used for on-screen output of the system including code-related elements or commands Is used to identify inputs you need to provide Is used for the navigation to a certain submenu entry Is used to identify variables or parameters, for example, within a string CAUTION Used with the safety alert symbol, indicates a hazardous situation which, if not avoided, could result in minor or moderate injury or material damage. CAUTION consists of the following elements: Information about the nature of a hazardous situation Consequences of not avoiding a hazardous situation Methods of avoiding a hazardous situation 4 Ortho Operator Manual

5 Legend WARNING Indicates a hazardous situation which, if not avoided, could result in death or serious injury. WARNING consists of the following elements: Information about the nature of a hazardous situation Consequences of not avoiding a hazardous situation Methods of avoiding a hazardous situation syngo MR E11 5

6 Legend 6 Ortho Operator Manual

7 Table of contents 1 Introduction Layout of the operator manual The current operator manual Intended use Authorized operating personnel Definitions of different persons 13 2 Measurement Overview of orthopedic imaging concept High throughput high-resolution imaging Advanced imaging techniques Dedicated orthopedic phased array coils Large Joint Dot Engine Knee Dot Engine Planning the examination and measuring the localizer 17 Adapting the examination to the patient 17 Starting the measurement of the scout 18 Changing the examination strategy subsequently Adjusting the slices and performing the measurements D examinations with integrated MPR planning Configuring MPR views (optional) 22 Assigning 3D measurements to the integrated MPR post-processing 23 Adding MPR views 23 syngo MR E11 7

8 Table of contents 2.4 Hip Dot Engine Planning the examination and measuring the localizer 25 Adapting the examination to the patient 25 Starting the measurement of the scout 26 Changing the examination strategy subsequently Adjusting the slices and performing the measurements Bilateral 3D hip examinations with integrated MPR planning Shoulder Dot Engine Planning the examination and measuring the localizer 32 Adapting the examination to the patient 32 Starting the measurement of the scout 33 Changing the examination strategy subsequently Adjusting the slices and performing the measurements D shoulder examinations with integrated MPR planning High-resolution fast 2D imaging Optimization with TSE sequence 37 Advantages 38 Optimizing echo time 38 Optimizing echo train length (ETL) 39 Optimizing fat saturation 40 Optimizing fat suppression 40 BLADE technique 41 Reducing examination noise 42 Reducing SAR for hip imaging syngo WARP: Imaging near orthopedic implants WARP: high bandwidth WARP: View Angle Tilting (VAT) technique WARP: Slice encoding for metal artifact correction (SEMAC) 47 8 Ortho Operator Manual

9 Table of contents 2.8 Isotropic 3D imaging D SPACE D TrueFISP D DESS MEDIC Image examples Parametric mapping with syngo MapIt T2 or R2 mapping with syngo MapIt T2 or R2 mapping: protocol parameters Examples in cartilage repair therapies (microfracture therapy vs MACT therapy) T2* or R2* mapping with syngo MapIt T2* or R2* mapping: protocol parameters Clinical use in cartilage repair therapies (microfracture therapy) Fast T1 mapping with syngo MapIt T1 mapping with B1 correction T1 mapping: protocol parameters Post-processing 3.1 syngo MR E11 Fusing biochemical maps and images Loading the data Loading the maps Loading the original images Optimizing the image display Visualizing the cartilage Saving and filming the images

10 Table of contents 10 Ortho Operator Manual

11 Introduction 1 1 Introduction In order to operate the MR system accurately and safely, the operating personnel must have the necessary expertise as well as knowledge of the complete operator manual. The operator manual must be read carefully prior to using the MR system. 1.1 Layout of the operator manual Your complete operator manual is split up into several volumes to improve readability. Each of these individual operator manuals covers a specific topic: Hardware components (system, coils, etc.) Software (measurement, evaluation, etc.) Another element of the complete operator manual is the information provided for the system owner of the MR system. The extent of the respective operator manual depends on the system configuration used and may vary. All components of the complete operator manual may include safety information that needs to be adhered to. The operator manuals for hardware and software address the authorized user. Basic knowledge in operating PCs and software is a prerequisite. 1.2 The current operator manual This manual may include descriptions covering standard as well as optional hardware and software. Contact your Siemens Sales Organization with respect to the hardware and software available for your system. The description of an option does not infer a legal requirement to provide it. syngo MR E11 11

12 1 Introduction The graphics, figures, and medical images used in this operator manual are examples only. The actual display and design of these may be slightly different on your system. Male and female patients are referred to as the patient for the sake of simplicity. 1.3 Intended use Your MAGNETOM MR system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. Your MAGNETOM MR system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles. The MAGNETOM MR system is not a device with measuring function as defined in the Medical Device Directive (MDD). Quantitative measured values obtained are for informational purposes and cannot be used as the only basis for diagnosis. For the USA only: Federal law restricts this device to sale, distribution and use by or on the order of a physician. Your MR system is a medical device for human use only! 12 Ortho Operator Manual

13 Introduction Authorized operating personnel The MAGNETOM MR system must be operated according to the intended use and only by qualified persons with the necessary knowledge in accordance with country-specific regulations, e.g. physicians, trained radiological technicians or technologists, subsequent to the necessary user training. This user training must include basics in MR technology as well as safe handling of MR systems. The user must be familiar with potential hazard and safety guidelines the same way the user is familiar with emergency and rescue scenarios. In addition, the user has to have read and understood the contents of the operator manual. Please contact Siemens Service for more information on available training options and suggested duration and frequency of such training Definitions of different persons Term used User/Operator/ Operating personnel System owner MR worker Explanation Person who operates the system or software, takes care of the patient or reads images Typically physicians, trained radiological technicians, or technologists Person who is responsible for the MR environment. This includes legal requirements, emergency plans, employee information and qualifications, as well as maintenance/repair. Person who works within the controlled access area or MR environment User/Operator as well as further personnel (for example, cleaning staff, facility manager, service personnel) syngo MR E11 13

14 1 Introduction Term used Siemens Service/service personnel Explanation Group of specially trained persons who are authorized by Siemens to perform certain maintenance activities References to Siemens Service include service personnel authorized by Siemens. 14 Ortho Operator Manual

15 Measurement 2 2 Measurement syngo MR E Overview of orthopedic imaging concept Large Joint Dot Engine Knee Dot Engine Hip Dot Engine Shoulder Dot Engine High-resolution fast 2D imaging syngo WARP: Imaging near orthopedic implants Isotropic 3D imaging Parametric mapping with syngo MapIt 58 15

16 2 Measurement 2.1 Overview of orthopedic imaging concept Orthopedic imaging is a comprehensive clinical applications package focusing on the following areas High throughput high-resolution imaging High-resolution fast 2D imaging: To deliver extremely high in-plane 2D resolution within clinically acceptable examination times. ( Page 37 High-resolution fast 2D imaging) Isotropic 3D imaging: To deliver enhanced workflow and increased diagnostic quality. ( Page 49 Isotropic 3D imaging) Advanced imaging techniques For helping in early diagnosis of osteoarthritis or monitoring of cartilage repair therapy. Parametric mapping (syngo MapIt): For improved diagnostic capabilities and therapy planning. ( Page 58 Parametric mapping with syngo MapIt) Dedicated orthopedic phased array coils The advances in resolution, image quality, workflow, scan speed and new imaging fields (biochemical) are only possible due to the dedicated orthopedic phased Array coils and the flexible coils. Siemens coils: Flex Large 4 (knee, shoulder, hip, ankle, pediatric MSK imaging) Flex Small 4 (wrist, elbow, pediatric MSK imaging) Body 18 (hip) Hand/Wrist 16 (hand, wrist) Foot/Ankle 16 (foot, ankle) Shoulder Large 16 (large shoulders) Shoulder Small 16 (small shoulders) CP Extremity Coil (not capable of parallel imaging) 16 Ortho Operator Manual

17 Measurement 2 QED coil: Tx/Rx 15-Channel Knee Coil 2.2 Large Joint Dot Engine The Large Joint Dot Engine provides a consistent workflow for all large joints. It consists of three Dot Engines: Knee Dot Engine ( Page 17 Knee Dot Engine) Hip Dot Engine ( Page 25 Hip Dot Engine) Shoulder Dot Engine ( Page 31 Shoulder Dot Engine) 2.3 Knee Dot Engine The Knee Dot Engine is intended to simplify and speed up the examination workflow. It provides guidance and easy adaption of the examination strategy. For 3D measurements the creation of MPRs is integrated into the workflow and supported by a guidance step for MPR planning. ( Page 21 3D examinations with integrated MPR planning) The Dot Engine user interfaces shown in this operator manual are examples only. The actual guidance texts and the design may be slightly different on your system. Planning the examination and measuring the localizer Patient has been registered Knee Dot Engine has been selected Adapting the examination to the patient After registration, the Patient View opens automatically. The default examination parameters are loaded. syngo MR E11 17

18 2 Measurement Selecting the examination strategy From the list: Select a suitable Exam Strategy for the patient. standard Speed focus Motion-insensitive (BLADE) High Bandwidth (WARP) For standard procedures. Provides fast protocols for when the patient cannot stay in the scanner for a longer period of time. For uncooperative/moving patients. Provides motion-insensitive protocols. Provides protocols with reduced sensitivity to susceptibility artifacts if the patient has MR Conditional implants. Please adhere to all safety instructions regarding implants. (Refer to Operator Manual MR System.) The pending protocols of the measurement queue are updated upon your selection. Starting the measurement of the scout The AAscout is used to determine anatomical structures. To start the Knee Dot Engine workflow, confirm the settings in the Patient View. 18 Ortho Operator Manual

19 Measurement 2 Results: The AutoAlign Scout is automatically measured and displayed. The localizer is the basis for the AutoAlign functionality, which provides consistent slice positioning of knee protocols without user interaction. (For a detailed description of the AutoAlign feature, please refer to: Operator Manual System and data management.) The next protocol opens. Changing the examination strategy subsequently Accessing the Patient View You can access the Patient View at any time during the examination. 1 To open the view, click the icon. 2 To confirm the settings and close the view, click the icon. Modifying parameters of measured protocols Changes in the Patient View only apply to pending protocols in the measurement queue. 1 To change the status of a protocol from measured to pending, select the measured protocol. 2 Select Rerun from here from the context menu (right-click with the mouse). 3 Open the Patient View. or Select Rerun from here with from the context menu (right-click with the mouse) The Patient View opens automatically. 4 Enter the requested modifications. syngo MR E11 19

20 2 Measurement Adjusting the slices and performing the measurements Localizers are displayed In the GSP segments, the slices for the following protocol are positioned by AutoAlign Knee. (For a detailed description of the AutoAlign feature, please refer to: Operator Manual System and data management.) Sample images with typical slice positioning are displayed in the Guidance View. Example: Guidance View for coronal slice positioning. 1 Check the slice positions for all subsequent measurements and adjust them, if necessary. You can also modify several sequence parameters of the current protocol using the Parameter View. Here you find the most important sequence parameters, e.g. the number of slices. To display the complete sequence parameters of the Routine parameter card, click the icon. 2 Start the measurement. The measurement is performed. The next protocol opens. 20 Ortho Operator Manual

21 Measurement 2 3 Repeat the above steps for all subsequent measurements D examinations with integrated MPR planning The 3D Knee Dot Engine provides an integrated planning step for MPR post-processing. MPRs of one or multiple 3D measurements are calculated immediately after each measurement. The MPR planning opens after the 3D measurements have been started. You are able to plan multiple MPR views with different orientations while the 3D measurements are running. During registration, the 3D Knee Dot Engine was selected 3D measurement has been started and the MPR planning step has opened Example: Guidance View for sagittal MPR view. 1 Select the desired MPR view from the list on the left side of the Guidance View. The corresponding MPR slice positions and orientations are displayed in the GSP segments. 2 Adapt the slice positioning, if necessary. In the Parameter View: You can also modify the view parameters alpha-numerically, e.g. the FoV. syngo MR E11 21

22 2 Measurement 3 Repeat the above steps for all MPR views. 4 Save the MPR settings. As soon as the 3D measurements have been concluded, the reconstruction of the MPRs is started automatically. For each 3D measurement, the defined MPR views are generated. The names of the resulting image series are a combination of the protocol name and the MPR view. For example: First 3D measurement: T1_SPC_FS Planned MPR views: transversal coronal sagittal Reconstructed MPR image series: T1_SPC_FS_MPR_tra T1_SPC_FS_MPR_cor T1_SPC_FS_MPR_sag If you repeat a 3D measurement, a new set of MPRs is calculated. You may use the resulting MPR views for the slice planning of subsequent measurements Configuring MPR views (optional) Dot Engine Step: The Dot Engine Step defines which strategies, decisions and global parameters are valid for the complete Dot Engine workflow examination. (For a detailed description, please refer to: Operator Manual Dot Cockpit.) Dot add-ins are predefined add-ins for Dot Engine Steps and program steps. Depending on the selected Dot add-in, you can configure different parameters of the Dot Engine Step. 22 Ortho Operator Manual

23 Measurement 2 Knee Dot add-ins: Generic Views MPR Assignment MPR Planning Assigning 3D measurements to the integrated MPR postprocessing Using the MPR Assignment Dot add-in you can define, for which 3D protocols the MPR post-processing is performed. 1 In the Dot Cockpit - Explorer: Select a 3D protocol. 2 Select Edit to open the protocol in the Program Editor. 3 Open the Step Properties dialog window by double-clicking the protocol. 4 Select AddIn Configuration. 5 Select MPR Assignment and activate the desired checkboxes. Adding MPR views Using the MPR Planning add-in, you can add new MPR views. 1 In the Queue: Select the MPR planning program step. 2 Open the Step Properties dialog window by double-clicking. syngo MR E11 23

24 2 Measurement 3 Select AddIn Configuration. 4 Select MPR Planning. 5 Add the MPR view by clicking Add MPR range. A new default MPR view is added to the list of available views. Renaming views 1 Right-click the added view with the mouse. 2 From the context menu: Select Rename. 24 Ortho Operator Manual

25 Measurement 2 3 Enter the new name. 4 Set the Guidance and Parameter Views. (Refer to Operator Manual Dot Cockpit.) Removing views 1 Right-click the view with the mouse. 2 From the context menu: Select Remove. 2.4 Hip Dot Engine The Hip Dot Engine is intended to simplify and speed up the examination workflow. It provides guidance and easy adaption of the examination strategy. The Hip Dot Engine offers three workflows: Two standard conventional 2D-workflows (unilateral and bilateral) and One bilateral 3D-workflow For 3D measurements the creation of MPRs is integrated into the workflow and supported by a guidance step for MPR planning. ( Page 29 Bilateral 3D hip examinations with integrated MPR planning) The Dot Engine user interfaces shown in this operator manual are examples only. The actual guidance texts and the design may be slightly different on your system. Planning the examination and measuring the localizer Patient has been registered Hip Dot Engine has been selected Adapting the examination to the patient After registration, the Patient View opens automatically. The default examination parameters are loaded. syngo MR E11 25

26 2 Measurement Selecting the examination strategy From the list: Select a suitable Exam Strategy for the patient. standard Speed focus High Bandwidth (WARP) For standard procedures. Provides fast protocols for when the patient cannot stay in the scanner for a longer period of time. Provides protocols with reduced sensitivity to susceptibility artifacts if the patient has MR Conditional implants. Please adhere to all safety instructions regarding implants. (Refer to Operator Manual MR System.) The pending protocols of the measurement queue are updated upon your selection. Starting the measurement of the scout The AAHip_Scout is used to determine anatomical structures. To start the Hip Dot Engine workflow, confirm the settings in the Patient View. 26 Ortho Operator Manual

27 Measurement 2 Results: The AutoAlign Scout is automatically measured and displayed. The localizer is the basis for the AutoAlign functionality, which provides consistent slice positioning of hip protocols without user interaction. (For a detailed description of the AutoAlign feature, please refer to: Operator Manual System and data management.) The next protocol opens. Changing the examination strategy subsequently Accessing the Patient View You can access the Patient View at any time during the examination. 1 To open the view, click the icon. 2 To confirm the settings and close the view, click the icon. Modifying parameters of measured protocols Changes in the Patient View only apply to pending protocols in the measurement queue. 1 To change the status of a protocol from measured to pending, select the measured protocol. 2 Select Rerun from here from the context menu (right-click with the mouse). 3 Open the Patient View. or Select Rerun from here with from the context menu (right-click with the mouse) The Patient View opens automatically. 4 Enter the requested modifications. syngo MR E11 27

28 2 Measurement Adjusting the slices and performing the measurements Localizers are displayed In the GSP segments, the slices for the following protocol are positioned by AutoAlign Hip. Sample images with typical slice positioning are displayed in the Guidance View. Example: Guidance View for coronal slice positioning. 1 Check the slice positions for all subsequent measurements and adjust them, if necessary. 2 To modify several sequence parameters of the current protocol, open the Parameters View. 28 Ortho Operator Manual

29 Measurement 2 Here you find the most important sequence parameters, e.g., the number of slices. To display the complete sequence parameters of the Routine parameter card, click the icon. 3 Start the measurement. The measurement is performed. The next protocol opens. 4 Repeat the above steps for all subsequent measurements Bilateral 3D hip examinations with integrated MPR planning The 3D workflows of the Hip Dot Engine provide an integrated planning step for MPR post-processing. MPRs of one or multiple 3D measurements are calculated immediately after each measurement. The MPR planning opens after the 3D measurements have been started. You are able to plan multiple MPR views with different orientations while the 3D measurements are running. For a detailed description of configuring MPR views, please refer to: ( Page 22 Configuring MPR views (optional)). syngo MR E11 29

30 2 Measurement During registration, a 3D workflow was selected 3D measurement has been started and the MPR planning step has opened Example: Parameters View for sagittal left MPR planning. 1 Select the desired MPR view from the list on the left side of the Parameters View. The corresponding MPR slice positions and orientations are displayed in the GSP segments. 2 Adapt the slice positioning, if necessary. In the Parameters View: You can also modify the view parameters alpha-numerically, e.g. the FoV. 3 Repeat the above steps for all MPR views. 4 Save the MPR settings. As soon as the 3D measurements have been concluded, the reconstruction of the MPRs is started automatically. For each 3D measurement, the defined MPR views are generated. 30 Ortho Operator Manual

31 Measurement 2 The names of the resulting image series are a combination of the protocol name and the MPR view. For example: First 3D measurement: PD_SPC_FS Planned MPR views: transversal coronal sagittal Reconstructed MPR image series: PD_SPC_FS_MPR_tra PD_SPC_FS_MPR_cor PD_SPC_FS_MPR_sag If you repeat a 3D measurement, a new set of MPRs is calculated. You may use the resulting MPR views for the slice planning of subsequent measurements. 2.5 Shoulder Dot Engine The Shoulder Dot Engine is intended to simplify and speed up the examination workflow. It provides guidance and easy adaption of the examination strategy. The Shoulder Dot Engine offers two workflows: Standard conventional 2D-workflow and 3D-workflow For 3D measurements the creation of MPRs is integrated into the workflow and supported by a guidance step for MPR planning. ( Page 35 3D shoulder examinations with integrated MPR planning) syngo MR E11 31

32 2 Measurement The Dot Engine user interfaces shown in this operator manual are examples only. The actual guidance texts and the design may be slightly different on your system Planning the examination and measuring the localizer Patient has been registered Shoulder Dot Engine has been selected Adapting the examination to the patient After registration, the Patient View opens automatically. The default examination parameters are loaded. Selecting the examination strategy From the list: Select a suitable Exam Strategy for the patient. standard Speed focus Motion-insensitive (BLADE) For standard procedures. Provides fast protocols for when the patient cannot stay in the scanner for a longer period of time. For uncooperative/moving patients. Provides motion-insensitive protocols. 32 Ortho Operator Manual

33 Measurement 2 Please adhere to all safety instructions regarding implants. (Refer to Operator Manual MR System.) The pending protocols of the measurement queue are updated upon your selection. Starting the measurement of the scout The AAShoulder_Scout is used to determine anatomical structures. To start the Shoulder Dot Engine workflow, confirm the settings in the Patient View. Results: The AutoAlign Scout is automatically measured and displayed. The localizer is the basis for the AutoAlign functionality, which provides consistent slice positioning of shoulder protocols without user interaction. (For a detailed description of the AutoAlign feature, please refer to: Operator Manual System and data management.) The next protocol opens. Changing the examination strategy subsequently Accessing the Patient View You can access the Patient View at any time during the examination. 1 To open the view, click the icon. 2 To confirm the settings and close the view, click the icon. Modifying parameters of measured protocols Changes in the Patient View only apply to pending protocols in the measurement queue. 1 To change the status of a protocol from measured to pending, select the measured protocol. 2 Select Rerun from here from the context menu (right-click with the mouse). syngo MR E11 33

34 2 Measurement 3 Open the Patient View. or Select Rerun from here with from the context menu (right-click with the mouse) The Patient View opens automatically. 4 Enter the requested modifications Adjusting the slices and performing the measurements Localizers are displayed In the GSP segments, the slices for the following protocol are positioned by AutoAlign Shoulder. Sample images with typical slice positioning are displayed in the Guidance View. Example: Guidance View for transversal slice positioning. 1 Check the slice positions for all subsequent measurements and adjust them, if necessary. 2 To modify several sequence parameters of the current protocol, open the Parameters View. 34 Ortho Operator Manual

35 Measurement 2 Here you find the most important sequence parameters, e.g., the number of slices. To display the complete sequence parameters of the Routine parameter card, click the icon. 3 Start the measurement. The measurement is performed. The next protocol opens. 4 Repeat the above steps for all subsequent measurements D shoulder examinations with integrated MPR planning The 3D workflow of the Shoulder Dot Engine provides an integrated planning step for MPR post-processing. MPRs of one or multiple 3D measurements are calculated immediately after each measurement. The MPR planning opens after the 3D measurements have been started. You are able to plan multiple MPR views with different orientations while the 3D measurements are running. For a detailed description of configuring MPR views, please refer to: ( Page 22 Configuring MPR views (optional)). syngo MR E11 35

36 2 Measurement During registration, the 3D workflow was selected 3D measurement has been started and the MPR planning step has opened Example: Parameters View for sagittal MPR planning 1 Select the desired MPR view from the list on the left side of the Parameters View. The corresponding MPR slice positions and orientations are displayed in the GSP segments. 2 Adapt the slice positioning, if necessary. In the Parameters View: You can also modify the view parameters alpha-numerically, e.g. the FoV. 3 Repeat the above steps for all MPR views. 4 Save the MPR settings. As soon as the 3D measurements have been concluded, the reconstruction of the MPRs is started automatically. For each 3D measurement, the defined MPR views are generated. 36 Ortho Operator Manual

37 Measurement 2 The names of the resulting image series are a combination of the protocol name and the MPR view. For example: First 3D measurement: PD_SPC_FS Planned MPR views: transversal coronal sagittal Reconstructed MPR image series: PD_SPC_FS_MPR_tra PD_SPC_FS_MPR_cor PD_SPC_FS_MPR_sag If you repeat a 3D measurement, a new set of MPRs is calculated. You may use the resulting MPR views for the slice planning of subsequent measurements. 2.6 High-resolution fast 2D imaging Imaging of the musculoskeletal (MSK) system requires high resolution with the necessary contrast for precise detection of small and complex structures Optimization with TSE sequence The flexible TSE sequence is optimized for a maximum matrix size and small FoV. This TSE sequence, together with parallel imaging techniques, delivers extremely high in-plane 2D resolution within clinically acceptable examination times, allowing for a more accurate diagnosis. syngo MR E11 37

38 2 Measurement Alternative protocols with faster scan times are also available in the Siemens protocol tree. The concept of high-resolution fast imaging utilizing the TSE sequence. The high-resolution fat-suppressed PD-weighted TSE image is a tool helping in accurate diagnosis. Advantages By using new reordering techniques, the TSE sequence allows for: a more flexible choice of TE for better optimization of contrast a more flexible choice of echo train lengths to maintain image contrast while optimizing protocols a flexible choice of fat suppression techniques (WEAK, STRONG, SPAIR, STIR) Optimizing echo time PD-weighted contrast is the gold standard in differentiating cartilage defects. Contrast is optimized by varying the echo time (the preferred TE varies from physician to physician). 38 Ortho Operator Manual

39 Measurement 2 (1) TE 20 ms (2) TE 40 ms Optimizing echo train length (ETL) Increasing the ETL will reduce the scan time while the contrast is mostly maintained. The flexible reordering scheme allows greater flexibility (even and odd ETL, reduced blurring). (1) TE 24 ms; ETL 5; 4 min (2) TE 24 ms; ETL 7; 3 min (3) TE 24 ms; ETL 10; 2 min (4) TE 24 ms; ETL 15; 1:20 min syngo MR E11 39

40 2 Measurement Optimizing fat saturation Fat saturation is useful when imaging suspected bone trauma or edema. It allows good differentiation of surface cartilage lesions. FatSat: Uses a spectrally selective pulse to saturate the fat spins before the imaging sequence. Has 2 settings: (a) STRONG which delivers a darker fat than (b) WEAK. SPAIR: Uses an adiabatic spectrally selective pulse which is insensitive to dielectric effects. Useful at 3 Tesla in the hip. STIR: Uses a spatial inversion pulse with a short inversion time to null fat. Only recommended where fat suppression is difficult due to B0 issues, i.e., with MR Conditional implants. Example: A cartilage tear can sometimes be better visualized using FatSat techniques (right) as shown above. Optimizing fat suppression For MSK imaging the fat suppression can be optimized for TSE, SE and SPACE sequences. The main advantages are: increased saturation homogeneity increased saturation strength decreased sensitivity to B0 inhomogeneities 40 Ortho Operator Manual

41 Measurement 2 The optimized fat suppression region Joints for MSK imaging can be selected in the Fat Suppression Optimization dialog window. To open the dialog window, click the button on the right side of the Fat suppr. dropdown menu. Prerequisite: Fat sat. is selected. Protocols for joints imaging are provided for 3 Tesla only. The optimized fat suppression should only be used for foot-ankle, knee, hand-wrist and elbow imaging. BLADE technique The BLADE technique is available for all MSK regions and can be configured for T1, T2, and PD contrasts. It is compatible with multichannel coils, and ipat can be employed. Any orientation can be used (sagittal, coronal and axial). BLADE is fully integrated in the TSE sequence. For a detailed description, please refer to: Application Brochure Pulse Sequences. syngo MR E11 41

42 2 Measurement Advantages of oversampling: Wrap-around artifacts (streaking) avoided Good results with respect to movement artifacts Increased SNR (1) Knee coronal: FoV , 0.4 mm 0.4 mm, TE = 47 ms (2) Wrist: FoV , 0.3 mm 0.3 mm, TE = 47 ms Reducing examination noise The quiet MR sequences increase patient comfort and enable MR examinations for noise-sensitive patients. The noise reduction mainly depends on the protocol settings. Compared to conventional imaging, a reduction of approximately 10 db(a) can be achieved. Quiet protocols are available in the Siemens protocol tree. The protocols for quiet MSK imaging include: localizer based on GRE sequence TSE sequence (proton density weighted with/without fat saturation) For a detailed description of the quiet MR sequences, please refer to Operator Manual Neuro. Reducing SAR for hip imaging In particular unilateral hip examinations are limited by SAR. For this reason, B1 shimming provides an option to reduce the SAR level of a protocol by optimizing the transmit parameters of the ptx channels. 42 Ortho Operator Manual

43 Measurement 2 In addition, protocols with reduced SAR are provided for unilateral hip imaging. Select Low SAR patient specific from the B1 Shim mode list in the System ptx Volumes parameter card. The SAR reduction depends on the size and position of the FOV and requires an additional adjustment scan. 2.7 syngo WARP: Imaging near orthopedic implants MR imaging in the vicinity of metal implants is challenging, firstly because of safety concerns and secondly because of image distortions. In general, an MR examination is contraindicated for patients with electronic or electronically conductive implants or metals, especially those containing ferromagnetic foreign matter. Particularly with regard to magnetic forces and heating of implants and surrounding tissue, please adhere to all safety instructions. (Refer to Operator Manual MR System.) The main source of artifacts seen when imaging patients with MR Conditional implants is related to distortions of the local magnetic field caused by the large difference in magnetic susceptibility between metal and tissue. The degree of field distortions depends on the shape, the location and the material properties of the metal implant. syngo WARP provides dedicated imaging techniques based on the TSE sequence in order to reduce susceptibility-related artifacts: syngo MR E11 43

44 2 Measurement high bandwidth optimizations to reduce the level of artifacts, for example geometric distortions and contrast changes are minimized VAT (View Angle Tilting) technique to correct for in-plane distortions (geometric distortion of the frequency encoding, for example image pixels are displaced along the frequency-encoding direction) SEMAC (Slice Encoding for Metal Artifact Correction) technique to correct for through-plane distortions (geometric distortions of the excited slice profile, for example the excited slices are curved instead of being plane) VAT and SEMAC are available if the WARP checkbox is activated in the Sequence Part 2 parameter card WARP: high bandwidth By activating the WARP checkbox, the bandwidth of the RF pulses will be increased. In combination with a high readout bandwidth selected, the TSE sequence becomes less sensitive to field distortions. High readout bandwidth: 44 Ortho Operator Manual

45 Measurement 2 reduces in-plane distortions avoids image blurring when using the VAT technique reduces the SNR of the image High RF pulse bandwidth (combined with small slice thickness): reduces through-plane distortions depends on the RF pulse type: Fast has a higher bandwidth, but also a higher SAR increases the specific absorption rate (SAR) Practical hints for imaging near MR Conditional implants with the TSE sequence: always select the WARP checkbox select a high readout bandwidth, for most cases Hz/ pixel is an advisable value select RF pulse type Fast unless the SAR gets too high SAR can be reduced by selecting a lower refocusing flip angle replace conventional fat suppression by STIR imaging (Short Tau Inversion Recovery): select magnetic preparation Slice-sel. IR with TI approx. 160ms (1.5T) or 200ms (3T) If WARP is selected, a W is added to the sequence string displayed in the image (only if VAT and SEMAC are not selected). syngo MR E11 45

46 2 Measurement (1) without syngo WARP (2) with syngo WARP WARP: View Angle Tilting (VAT) technique VAT compensates in-plane distortions by adding an additional readout gradient in the slice direction. The corresponding value in the Sequence Part 2 parameter card specifies the amplitude of the view angle tilting gradient: 0%: VAT is deactivated (no distortion correction) 100%: VAT has the same amplitude as the slice-selective gradient (full readout distortion correction) Performing the measurement with VAT may cause blurring of the image. Blurring can be reduced by: using thin slices using a high readout bandwidth The result of VAT strongly depends on the type and orientation of the implant, as well as other imaging parameters. In general, VAT can only reduce in-plane distortions. If VAT does not improve the image quality, this means that other effects are dominating, for example through-plane distortions. In that case, try to use thinner slices, RF pulse type Fast or SEMAC. 46 Ortho Operator Manual

47 Measurement 2 If VAT is selected, a V is added to the sequence string displayed in the image (only if SEMAC is not selected). (1) without VAT (2) with VAT WARP: Slice encoding for metal artifact correction (SEMAC) SEMAC provides a through-plane distortion correction by performing additional phase encoding in the slice direction. SEMAC is most effective for severe field distortions, for example near large metal structures such as full joint replacement of the hip or knee. SEMAC can only be applied in combination with: VAT set to 100% slice distance of 0% A SEMAC value of 0 means that no SEMAC is applied. syngo MR E11 47

48 2 Measurement The measurement time will be prolonged by the number of SEMAC steps. Since SEMAC increases the SNR, try to reduce the scan time by reducing averages, apply parallel imaging or partial fourier. The required number of SEMAC phase-encoding steps depends on the size, shape and material of the implant and may vary from patient to patient. For joint arthroplasty 8-12 SEMAC steps is usually a reasonable value. Metals causing stronger artifacts, such as stainless steel or cobalt chromium alloys, require higher SEMAC steps, whereas titanium implants usually require lower values or even no SEMAC at all. SEMAC cannot be used in combination with 3D TSE, BLADE, TimCT, multiple slice groups, and DIXON. If SEMAC is selected, a S is added to the sequence string displayed in the image text. (1) without SEMAC (2) with SEMAC 48 Ortho Operator Manual

49 Measurement Isotropic 3D imaging High-resolution fast isotropic 3D imaging is becoming increasingly more important as a means of enhancing workflow and providing more accurate diagnosis. Isotropic 3D imaging allows for fully flexible examination, i.e. depending on the suspected diagnosis, images can be reformatted in any plane. Isotropic sequences: Using an isotropic 3D sequence, the images can be reformatted with high in-plane and through-plane resolution. Utilization of such a sequence introduces further improvements in the workflow by acquiring one 3D series that can be reformatted in the different planes needed for precise diagnosis. Siemens provides five different sequences with specified contrasts: 3D SPACE ( Page 49 3D SPACE) 3D TrueFISP ( Page 50 3D TrueFISP) 3D DESS (Dual Echo Steady State) ( Page 51 3D DESS) 3D MEDIC ( Page 53 MEDIC) 3D FISP For a detailed description of the sequences, please refer to: Application Brochure Pulse Sequences D SPACE Concept: Different planes can be reconstructed from the isotropic data set for high-resolution diagnosis of cartilage, ligaments, and meniscus in the knee joint. syngo MR E11 49

50 2 Measurement (1) PD-weighted SPACE, isotropic resolution 0.5 mm (2) PD-weighted SPACE with FatSat, isotropic resolution 0.5 mm D TrueFISP The TrueFISP sequence is used for balanced steady state imaging. On each of the gradient axes the net gradient moment is zero. It has a good SNR, but is prone to banding artifacts in regions of compromised homogeneity, i.e., MR Conditional implants. (1) RF (2) Slice selection 50 Ortho Operator Manual

51 Measurement 2 (3) Phase encoding (4) Frequency encoding (5) Signal (SS-FID/SS-echo) Minimize TR by selecting a high readout bandwidth to avoid interference streaks in the image. Ankle, isotropic resolution 0.3 mm, reformatted below in sagittal, coronal, and axial direction D DESS In the steady state, two signals are produced an FID signal and an echo. With DESS (Dual Echo Steady State) these two signals are measured and combined to produce a single image. syngo MR E11 51

52 2 Measurement (1) RF (2) Slice selection (3) Phase encoding (4) Frequency encoding (5) Signal (SS-FID/SS-echo) Use primarily in orthopedic imaging with good contrast between synovial fluid and cartilage. A non-selective excitation pulse is activated beforehand for fat suppression at a short TR. 52 Ortho Operator Manual

53 Measurement 2 (1) Shoulder, isotropic resolution 0.7 mm (2) Knee, isotropic resolution 0.6 mm MEDIC The MEDIC sequence is a multi-echo GRE sequence where up to 6 echoes are combined to produce a single image. (1) RF (2) Slice selection (3) Phase encoding syngo MR E11 53

54 2 Measurement (4) Frequency encoding (5) Signal (dashed line: T2* decay) Gives very good contrast for meniscus evaluation. (1) Shoulder coronal (2) Knee sagittal Image examples Example: 10 minute knee exam 54 Ortho Operator Manual

55 Measurement 2 Isotropic data set 0.5 mm. syngo MR E11 55

56 2 Measurement Reformat along x diagnosis-based planes: (1) Meniscus reformat (2) ACL reformat (3) Coronal reformat (4) Patella reformat Example: Precise localization of anatomy (meniscus) 56 Ortho Operator Manual

57 Measurement 2 Example: Complex geometry simple reformatting (radial reformats for hip imaging) syngo MR E11 57

58 2 Measurement 2.9 Parametric mapping with syngo MapIt Parametric mapping with syngo MapIt may improve both the accuracy of diagnosis and planning and also may monitor the effectiveness of therapy. At present, the main focus is on: T2 and T2* mapping, T1 mapping 58 Ortho Operator Manual

59 Measurement T2 or R2 mapping with syngo MapIt Clinical use to aid in: Cartilage repair therapies (microfracture therapy vs MACT therapy). (MACT = Matrix-associated Autologous Chondrocyte Transplantation). Cartilage repair therapies (MACT therapy follow up) Early OA detection (femoral acetabular impingement). (OA = Osteoarthritis). T2 provides information on structural changes in the collagen within the cartilage. T2 depends on the orientation as a function of the magic angle effect. T2 provides information on changes in the water content of the cartilage. Compressed areas show less water content lower T2. Less compressed areas show greater water content higher T2. T2 has been used to study cartilage repair therapies and to provide information on cartilage softening. T2 is measured using a multi-echo spin echo with up to 32 echoes. Each echo produces an image. The map is produced by a pixel-bypixel analysis. Instead of T2, R2 = 1/T2 can optionally be calculated T2 or R2 mapping: protocol parameters For T2 or R2 mapping use the se_mc sequence. Use the Inline MapIt parameter card. syngo MR E11 59

60 2 Measurement MapIt Select T2 or R2 to turn on the Inline technology. Measurements Are set to 1. Contrasts TE TR Noise threshold Save original images Up to 32 echoes possible. More echoes provide a better fit. Set the echo time for each echo. Keep maximum TE to less than or similar to expected T2 values. Use TR values > 1000 ms. Echoes with signals less than this value will be ignored in the fit. If in doubt, leave at default value. Keeps the morphological base images in the database Examples in cartilage repair therapies (microfracture therapy vs MACT therapy) MFX therapy creates microfractures in the bone. Cartilage regeneration is promoted by the released blood and marrow. 60 Ortho Operator Manual

61 Measurement 2 These are the characteristics of this method: Reduced T2 in the region of MFX therapy. Matrix-associated autologous chondrocyte transplantation (MACT). An operative procedure using a cell seeded collagen matrix. Used for the treatment of localized full thickness cartilage defects. Courtesy of S. Trattnig, Dept. Radiology, University Vienna T2* or R2* mapping with syngo MapIt T2* is the apparent transverse relaxation rate and has a T2 component. It can be used as a substitute for T2 mapping. Instead of T2*, R2* = 1/T2* can be optionally calculated. syngo MR E11 61

62 2 Measurement T2* also has a component which depends on the field change within the voxel, resulting from: Macroscopic field changes, i.e. main field homogeneity, large susceptibility fields from implants. Macroscopic field changes due to susceptibility variations, i.e. bone-cartilage interface. Microscopic field changes from the microstructure within the voxel. T2* is measured using a multi-echo gradient-echo sequence. Each TE gives an image. A pixel-by-pixel analysis produces the T2* map. (1) RF (2) Slice selection (3) Phase encoding (4) Frequency encoding (5) Signal (dashed line: T2* decay) 62 Ortho Operator Manual

63 Measurement T2* or R2* mapping: protocol parameters For T2* or R2* mapping use the gre sequence. Use the Inline MapIt parameter card. MapIt Select T2* or R2* to turn on the Inline technology. Measurements Are set to 1. Contrasts TE Noise threshold Save original images Up to 12 echoes possible. Set the echo time for each echo. Keep maximum TE to less than or similar to expected T2 values. Echoes with signals less than this value will be ignored in the fit. If in doubt, leave at default value. Keeps the morphological base images in the database. In the Sequence Part 1 parameter card: syngo MR E11 63

64 2 Measurement Contrasts Bandwidth Flow comp. Readout mode Up to 12 echoes possible. Must be entered for each echo. Use the same bandwidth for all echoes. Useful in the abdomen Select mono or bi-polar Clinical use in cartilage repair therapies (microfracture therapy) MFX therapy creates microfractures in the bone. Cartilage regeneration is promoted by the released blood and marrow. T2* is reduced in the region of MFX therapy. A small patient study suggests that variability is better with T2*. 64 Ortho Operator Manual

65 Measurement Fast T1 mapping with syngo MapIt Clinical use to aid in: dgemric Pre-operative staging Prediction of therapy outcome Therapy follow up T1 mapping is used in cartilage to track proteoglycans. The gold standard was multiple IR spin echo measurements. However, the very long acquisition time of up to 30 minutes makes it clinically difficult. Fast T1 mapping with syngo uses a 2 angle VIBE measurement. The acquisition time is significantly reduced to 3 minutes. To calculate T1, this technique utilizes 2 spoiled GRE measurements, each identical except for different flip angles. syngo MapIt calculates T1 on a pixel-by-pixel basis. T1 mapping with B1 correction Variable flip angle techniques used by syngo MapIt are intrinsically sensitive to inhomogeneities of the transmit RF field (B1). syngo MapIt can perform B1 corrections to improve the spatial homogeneity and the reproducibility of the acquired T1 maps. syngo MR E11 65

66 2 Measurement To enable this feature, run the Siemens B1mapForT1mapping protocol prior to the T1 mapping protocol. The three-dimensional FoV of the B1mapForT1mapping protocol should be greater or equal to the FoV of the T1 mapping protocol. B1 corrections will be performed automatically. The image series with the corrected T1 maps is marked as T1_Images_B1corr instead of T1_Images. Please note: The resulting T1 map values may overestimate the absolute T1 values! This behavior is expected due to the technique used the variable flip angle method. Thus a relative comparison of T1 values within tissues should be made T1 mapping: protocol parameters For fast T1 mapping use the vibe sequence. Use the Inline MapIt parameter card. MapIt Auto angle calculation T1 estimate Select T1 map. Select, if required. Two optimum angles will be automatically calculated, based on the estimated T1. 66 Ortho Operator Manual

67 Measurement 2 Flip angle Alternatively, select two flip angles. syngo MR E11 67

68 2 Measurement 68 Ortho Operator Manual

69 Post-processing 3 3 Post-processing 3.1 Fusing biochemical maps and images 70 syngo MR E11 69

70 3 Post-processing 3.1 Fusing biochemical maps and images The following example describes how to overlay T2/T2* maps generated with syngo MapIt with their corresponding anatomical images. Subsequently, manual cartilage segmentation is performed. The same procedure applies when using T1 maps. In this case, two series containing the morphological base images for the two flip angles are stored together with the T1 map series Loading the data T2/T2* maps have been generated with syngo MapIt Anatomical base images are available (Save original images activated during mapping) Loading the maps 1 Select the mapping series in the Patient Browser. 2 Click the 3D MPR icon to start image processing as MPR. 3 Window the mapping images to optimize their contrast and brightness. Loading the original images The series preceding the T2/T2* maps contains the morphological base images for all echoes. 1 Select the morphological series in the Patient Browser. 2 Load the data to 3D Fusion with the icon. 3 In the 3D Series List, select the echo images you want to use. 70 Ortho Operator Manual

71 Post-processing 3 The Fusion Registration dialog window is displayed. Registration is not needed in this case. 4 Skip Fusion Registration with OK Optimizing the image display 1 Open the Fusion Definition MPR dialog window with Fusion > Fusion Definition. 2 In order to get access to the mask menu, click Advanced. 3 Hide the background noise in the mapping images by increasing the L value for Masking to 1 (left spin box). 4 Select suitable Color Lookup Tables for the map and the morphological images to optimize their view. Recommended settings: Rainbow for mapping images (left selection list), Gray Scale for morphological images (right selection list). 5 Set the T2/T2* baseline map Window Level to a default value by changing the left C and W values (e.g. 45 each). syngo MR E11 71

72 3 Post-processing Visualizing the cartilage It is easier to segment the cartilage if you work on the morphological images. 1 In the Fusion Definition MPR dialog window, move the Mixing Ratio slider to the right to display the morphological image only. 2 Activate VOI Punch Mode with the icon in the Settings subtask card. The VOI Punch Mode dialog window opens. VOI drawing is activated automatically. 3 Trace the cartilage in the morphological image. Double-click to finish. 4 Remove the non-cartilage part of the mapping image with the Keep Inside icon. 5 In order to provide for good fusion, set a Mixing Ratio of 50% in the Fusion Definition MPR dialog window. 72 Ortho Operator Manual

73 Post-processing Saving and filming the images 1 Select the respective segment for saving or filming. 2 Save the images as a new series with the icon. 3 To film the images selected, click the icon. syngo MR E11 73

Orthopedic MRI Protocols. Philips Panorama HFO

Orthopedic MRI Protocols. Philips Panorama HFO Orthopedic MRI Protocols Philips Panorama HFO 1 2 Prepared in collaboration with Dr. John F. Feller, Medical Director of Desert Medical Imaging, Palm Springs, CA. Desert Medical Imaging will provide the

More information

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street MRI is located in the sub basement of CC wing. From Queen or Victoria, follow the baby blue arrows and ride the CC south

More information

COBRE Scan Information

COBRE Scan Information COBRE Scan Information Below is more information on the directory structure for the COBRE imaging data. Also below are the imaging parameters for each series. Directory structure: var/www/html/dropbox/1139_anonymized/human:

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

MRI. When to use What sequences. Outline 2012/09/19. Sequence: Definition. Basic Principles: Step 2. Basic Principles: Step 1. Govind Chavhan, MD

MRI. When to use What sequences. Outline 2012/09/19. Sequence: Definition. Basic Principles: Step 2. Basic Principles: Step 1. Govind Chavhan, MD MRI When to use What sequences Govind Chavhan, MD Assistant Professor and Staff Radiologist The Hospital For Sick Children, Toronto Planning Acquisition Post processing Interpretation Patient history and

More information

syngo MR E11 Operator Manual System and data management Answers for life.

syngo MR E11 Operator Manual System and data management Answers for life. www.siemens.com/healthcare syngo MR E11 Operator Manual System and data management Answers for life. syngo MR E11 Operator Manual System and data management Legend Indicates a hint Is used to provide

More information

SIEMENS MAGNETOM Avanto syngo MR B15

SIEMENS MAGNETOM Avanto syngo MR B15 \\USER\INVESTIGATORS\Ravi\ADNI-Subject\Localizer TA: 0:10 PAT: Voxel size: 1.9 1.5 8.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer Inline movie

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

A Virtual MR Scanner for Education

A Virtual MR Scanner for Education A Virtual MR Scanner for Education Hackländer T, Schalla C, Trümper A, Mertens H, Hiltner J, Cramer BM Hospitals of the University Witten/Herdecke, Department of Radiology Wuppertal, Germany Purpose A

More information

Siemens AG, Healthcare Sector. syngo MR D Operator Manual - System and data management

Siemens AG, Healthcare Sector. syngo MR D Operator Manual - System and data management Siemens AG, Healthcare Sector Cs2 syngo System Operator 2010-2012 MR-05011 630 02 English 06/2012 n.a. Informatik and Manual D13 data management syngo MR D13 Operator Manual - System and data management

More information

New Technology Allows Multiple Image Contrasts in a Single Scan

New Technology Allows Multiple Image Contrasts in a Single Scan These images were acquired with an investigational device. PD T2 T2 FLAIR T1 MAP T1 FLAIR PSIR T1 New Technology Allows Multiple Image Contrasts in a Single Scan MR exams can be time consuming. A typical

More information

SIEMENS MAGNETOM Skyra syngo MR D13

SIEMENS MAGNETOM Skyra syngo MR D13 Page 1 of 8 SIEMENS MAGNETOM Skyra syngo MR D13 \\USER\CIND\StudyProtocols\PTSA\*dm_ep2d_mono70_b0_p2_iso2.0 TA:1:05 PAT:2 Voxel size:2.0 2.0 2.0 mm Rel. SNR:1.00 :epse Properties Routine Prio Recon Load

More information

Breast MRI Accreditation Program Clinical Image Quality Guide

Breast MRI Accreditation Program Clinical Image Quality Guide Breast MRI Accreditation Program Clinical Image Quality Guide Introduction This document provides guidance on breast MRI clinical image quality and describes the criteria used by the ACR Breast MRI Accreditation

More information

SIEMENS MAGNETOM Avanto syngo MR B15

SIEMENS MAGNETOM Avanto syngo MR B15 \\USER\INVESTIGATORS\Ravi\ADNI-phantom\QC Phantom-Localizer TA: 0:10 PAT: Voxel size: 1.9 1.5 8.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer

More information

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE MRES 7005 - Fast Imaging Techniques Module 4 K-Space Symmetry Review K-Space Review K-Space Symmetry Partial or Fractional Echo Half or Partial Fourier HASTE Conditions for successful reconstruction Interpolation

More information

GE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals.

GE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals. GE Healthcare CLINICAL GALLERY Discovery * MR750w 3.0T This brochure is intended for European healthcare professionals. NEURO PROPELLER delivers high resolution, motion insensitive imaging in all planes.

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

SIEMENS MAGNETOM Verio syngo MR B15V

SIEMENS MAGNETOM Verio syngo MR B15V \\USER\ZAHID_RESEARCH\MS\No Name\3D SWI TA: 6:39 PAT: 2 Voxel size: 1.0 0.5 2.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer Inline movie Auto

More information

SIEMENS MAGNETOM Verio syngo MR B17

SIEMENS MAGNETOM Verio syngo MR B17 \\USER\Dr. Behrmann\routine\Ilan\ep2d_bold_PMU_resting TA: 8:06 PAT: Voxel size: 3.03.03.0 mm Rel. SNR: 1.00 USER: ep2d_bold_pmu Properties Special sat. Prio Recon System Before measurement Body After

More information

Siemens AG, Healthcare Sector. syngo MR D13 0. Supplement - Parameters and image text 0.

Siemens AG, Healthcare Sector. syngo MR D13 0. Supplement - Parameters and image text 0. Siemens AG, Healthcare Sector 0 0 n.a. English Cs2 syngo Neuro Operator MR-05014 630 05/2010 01 02 Informatik, Manual D11 Cape syngo MR D13 0.0 Supplement - Parameters and image text 0. syngo MR D13 0.

More information

MRI Physics II: Gradients, Imaging

MRI Physics II: Gradients, Imaging MRI Physics II: Gradients, Imaging Douglas C., Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor Magnetic Fields in MRI B 0 The main magnetic field. Always on (0.5-7 T) Magnetizes

More information

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Outline High Spectral and Spatial Resolution MR Imaging (HiSS) What it is How to do it Ways to use it HiSS for Radiation

More information

syngo MR E11 Operator Manual Neuro Answers for life.

syngo MR E11 Operator Manual Neuro Answers for life. www.siemens.com/healthcare syngo MR E11 Operator Manual Neuro Answers for life. syngo MR E11 Operator Manual Neuro Legend Indicates a hint Is used to provide information on how to avoid operating errors

More information

Data. MAGNETOM Symphony, A Tim System Matrix Coils

Data. MAGNETOM Symphony, A Tim System Matrix Coils Data MAGNETOM Symphony, A Tim System Matrix Coils Head Matrix Coil The Head Matrix coil is part of the standard system configuration. 12-element design with 12 integrated preamplifiers, two rings of 6

More information

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space White Pixel Artifact Caused by a noise spike during acquisition Spike in K-space sinusoid in image space Susceptibility Artifacts Off-resonance artifacts caused by adjacent regions with different

More information

M R I Physics Course

M R I Physics Course M R I Physics Course Multichannel Technology & Parallel Imaging Nathan Yanasak, Ph.D. Jerry Allison Ph.D. Tom Lavin, B.S. Department of Radiology Medical College of Georgia References: 1) The Physics of

More information

(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions

(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions EE225E/BIOE265 Spring 2013 Principles of MRI Miki Lustig Assignment 8 Solutions 1. Nishimura 7.1 P)jc%z 5 ivcr3. 1. I Due Wednesday April 10th, 2013 (a Scrhon5 R2iwd b 0 ZOms Xn,s r cx > qs 4-4 8ni6 4

More information

ADNI, ADNI_QH, SURVEY. Geometry. connection

ADNI, ADNI_QH, SURVEY. Geometry. connection ADNI, ADNI_QH, SURVEY Geometry Coil selection = Head connection = d Multi coil Homogeneity correction ne FOV (mm) = 250.00 RFOV (%) = 100.00 Foldover suppression Matrix scan = 256 reconstruction = 256

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

MSK EXTREME. V-SPEC Technology HIGH FIELD 1.0 TESLA TRULY OPEN, DEDICATED MRI. The Power of High Field. Rapid ROI. Optimized Performance

MSK EXTREME. V-SPEC Technology HIGH FIELD 1.0 TESLA TRULY OPEN, DEDICATED MRI. The Power of High Field. Rapid ROI. Optimized Performance MSK EXTREME HIGH FIELD 1.0 TESLA TRULY OPEN, DEDICATED MRI The Power of High Field Rapid ROI Optimized Performance High Throughput Ease of Use Increased Patient Comfort V-SPEC Technology THE POWER OF HIGH

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

MR Advance Techniques. Vascular Imaging. Class III

MR Advance Techniques. Vascular Imaging. Class III MR Advance Techniques Vascular Imaging Class III 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology

More information

surface Image reconstruction: 2D Fourier Transform

surface Image reconstruction: 2D Fourier Transform 2/1/217 Chapter 2-3 K-space Intro to k-space sampling (chap 3) Frequenc encoding and Discrete sampling (chap 2) Point Spread Function K-space properties K-space sampling principles (chap 3) Basic Contrast

More information

A novel noise removal using homomorphic normalization for multi-echo knee MRI

A novel noise removal using homomorphic normalization for multi-echo knee MRI A novel noise removal using homomorphic normalization for multi-echo knee MRI Xuenan Cui 1a),HakilKim 1b), Seongwook Hong 1c), and Kyu-Sung Kwack 2d) 1 School of Information and Communication Engineering,

More information

XI Signal-to-Noise (SNR)

XI Signal-to-Noise (SNR) XI Signal-to-Noise (SNR) Lecture notes by Assaf Tal n(t) t. Noise. Characterizing Noise Noise is a random signal that gets added to all of our measurements. In D it looks like this: while in D

More information

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI Figure 1. In the brain, the gray matter has substantially more blood vessels and capillaries than white matter. The magnified image on the right displays the rich vasculature in gray matter forming porous,

More information

MRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health

MRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health MRI image formation Indiana University School of Medicine and Indiana University Health Disclosure No conflict of interest for this presentation 2 Outlines Data acquisition Spatial (Slice/Slab) selection

More information

SIEMENS MAGNETOM TrioTim syngo MR B17

SIEMENS MAGNETOM TrioTim syngo MR B17 \\USER\KNARRGROUP\MultiBand\LavretskyMultiBand\trufi localizer 3-plane TA: 5.1 s PAT: Voxel size: 1.2 1.2 5. Rel. SNR: 1.00 SIEMENS: trufi Load to stamp Slice group 1 Slices 1 Dist. factor 20 % Phase enc.

More information

Applications Guide for Interleaved

Applications Guide for Interleaved Applications Guide for Interleaved rephase/dephase MRAV Authors: Yongquan Ye, Ph.D. Dongmei Wu, MS. Tested MAGNETOM Systems : 7TZ, TRIO a Tim System, Verio MR B15A (N4_VB15A_LATEST_20070519) MR B17A (N4_VB17A_LATEST_20090307_P8)

More information

Supplementary Information

Supplementary Information Supplementary Information Magnetic resonance imaging reveals functional anatomy and biomechanics of a living dragon tree Linnea Hesse 1,2,*, Tom Masselter 1,2,3, Jochen Leupold 4, Nils Spengler 5, Thomas

More information

Module 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution.

Module 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution. MRES 7005 - Fast Imaging Techniques Module 5: Dynamic Imaging and Phase Sharing (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review Improving Temporal Resolution True-FISP (I) True-FISP (II) Keyhole

More information

Diffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford

Diffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford Diffusion MRI Acquisition Karla Miller FMRIB Centre, University of Oxford karla@fmrib.ox.ac.uk Diffusion Imaging How is diffusion weighting achieved? How is the image acquired? What are the limitations,

More information

Supplementary methods

Supplementary methods Supplementary methods This section provides additional technical details on the sample, the applied imaging and analysis steps and methods. Structural imaging Trained radiographers placed all participants

More information

Role of Parallel Imaging in High Field Functional MRI

Role of Parallel Imaging in High Field Functional MRI Role of Parallel Imaging in High Field Functional MRI Douglas C. Noll & Bradley P. Sutton Department of Biomedical Engineering, University of Michigan Supported by NIH Grant DA15410 & The Whitaker Foundation

More information

Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs

Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Doug Dean Final Project Presentation ENGN 2500: Medical Image Analysis May 16, 2011 Outline Review of the

More information

UNIVERSITY OF SOUTHAMPTON

UNIVERSITY OF SOUTHAMPTON UNIVERSITY OF SOUTHAMPTON PHYS2007W1 SEMESTER 2 EXAMINATION 2014-2015 MEDICAL PHYSICS Duration: 120 MINS (2 hours) This paper contains 10 questions. Answer all questions in Section A and only two questions

More information

8/11/2009. Common Areas of Motion Problem. Motion Compensation Techniques and Applications. Type of Motion. What s your problem

8/11/2009. Common Areas of Motion Problem. Motion Compensation Techniques and Applications. Type of Motion. What s your problem Common Areas of Motion Problem Motion Compensation Techniques and Applications Abdominal and cardiac imaging. Uncooperative patient, such as pediatric. Dynamic imaging and time series. Chen Lin, PhD Indiana

More information

MRI Imaging Options. Frank R. Korosec, Ph.D. Departments of Radiology and Medical Physics University of Wisconsin Madison

MRI Imaging Options. Frank R. Korosec, Ph.D. Departments of Radiology and Medical Physics University of Wisconsin Madison MRI Imaging Options Frank R. Korosec, Ph.D. Departments of Radiolog and Medical Phsics Universit of Wisconsin Madison f.korosec@hosp.wisc.edu As MR imaging becomes more developed, more imaging options

More information

ThE ultimate, INTuITIVE Mr INTErFAcE

ThE ultimate, INTuITIVE Mr INTErFAcE ThE ultimate, INTuITIVE Mr INTErFAcE Empowering you to do more The revolutionary Toshiba M-power user interface takes Mr performance and flexibility to levels higher than ever before. M-power is able to

More information

Magnetic Resonance Elastography (MRE) of Liver Disease

Magnetic Resonance Elastography (MRE) of Liver Disease Magnetic Resonance Elastography (MRE) of Liver Disease Authored by: Jennifer Dolan Fox, PhD VirtualScopics Inc. jennifer_fox@virtualscopics.com 1-585-249-6231 1. Overview of MRE Imaging MRE is a magnetic

More information

Use of MRI in Radiotherapy: Technical Consideration

Use of MRI in Radiotherapy: Technical Consideration Use of MRI in Radiotherapy: Technical Consideration Yanle Hu, PhD Department of Radiation Oncology, Mayo Clinic Arizona 04/07/2018 2015 MFMER slide-1 Conflict of Interest: None 2015 MFMER slide-2 Objectives

More information

Midterm Review

Midterm Review Midterm Review - 2017 EE369B Concepts Noise Simulations with Bloch Matrices, EPG Gradient Echo Imaging 1 About the Midterm Monday Oct 30, 2017. CCSR 4107 Up to end of C2 1. Write your name legibly on this

More information

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead.

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead. 1 CT Basics Principles of Spiral CT Dose 2 Who invented CT? 1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections Sir Godfrey Hounsfield worked for the Central

More information

Computational Medical Imaging Analysis

Computational Medical Imaging Analysis Computational Medical Imaging Analysis Chapter 2: Image Acquisition Systems Jun Zhang Laboratory for Computational Medical Imaging & Data Analysis Department of Computer Science University of Kentucky

More information

LEAD LOCALIZATION. Version 1.0. Software User Guide Revision 1.1. Copyright 2018, Brainlab AG Germany. All rights reserved.

LEAD LOCALIZATION. Version 1.0. Software User Guide Revision 1.1. Copyright 2018, Brainlab AG Germany. All rights reserved. LEAD LOCALIZATION Version 1.0 Software User Guide Revision 1.1 Copyright 2018, Brainlab AG Germany. All rights reserved. TABLE OF CONTENTS TABLE OF CONTENTS 1 GENERAL INFORMATION...5 1.1 Contact Data...5

More information

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 16 Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 3.1 MRI (magnetic resonance imaging) MRI is a technique of measuring physical structure within the human anatomy. Our proposed research focuses

More information

Exam 8N080 - Introduction MRI

Exam 8N080 - Introduction MRI Exam 8N080 - Introduction MRI Friday January 23 rd 2015, 13.30-16.30h For this exam you may use an ordinary calculator (not a graphical one). In total there are 6 assignments and a total of 65 points can

More information

OBJECT MANAGEMENT OBJECT MANIPULATION

OBJECT MANAGEMENT OBJECT MANIPULATION OBJECT MANAGEMENT OBJECT MANIPULATION Version 1.1 Software User Guide Revision 1.0 Copyright 2018, Brainlab AG Germany. All rights reserved. TABLE OF CONTENTS TABLE OF CONTENTS 1 GENERAL INFORMATION...5

More information

Scan Acceleration with Rapid Gradient-Echo

Scan Acceleration with Rapid Gradient-Echo Scan Acceleration with Rapid Gradient-Echo Hsiao-Wen Chung ( 鍾孝文 ), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ. Dept. Radiology, Tri-Service General Hospital 1 of 214 The Need

More information

WE PROVIDE KNOWLEDGE, SERVICES AND PRODUCTS FOR ANY KIND OF PROCESS PLANTS

WE PROVIDE KNOWLEDGE, SERVICES AND PRODUCTS FOR ANY KIND OF PROCESS PLANTS MAGNETOM ESSENZA A Tim+Dot System Tim [25x8] www.siemens.com/magnetom-essenza 2 MAGNETOM ESSENZA Established 1.5 T performance. With Tim+Dot Lomisa Distribuciones y Proyectos S.L., socidad inscrita en

More information

Fast Imaging Trajectories: Non-Cartesian Sampling (1)

Fast Imaging Trajectories: Non-Cartesian Sampling (1) Fast Imaging Trajectories: Non-Cartesian Sampling (1) M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.05.03 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business

More information

Functional MRI. Jerry Allison, Ph. D. Medical College of Georgia

Functional MRI. Jerry Allison, Ph. D. Medical College of Georgia Functional MRI Jerry Allison, Ph. D. Medical College of Georgia BOLD Imaging Technique Blood Oxygen Level Dependent contrast can be used to map brain function Right Hand Motor Task Outline fmri BOLD Contrast

More information

Fmri Spatial Processing

Fmri Spatial Processing Educational Course: Fmri Spatial Processing Ray Razlighi Jun. 8, 2014 Spatial Processing Spatial Re-alignment Geometric distortion correction Spatial Normalization Smoothing Why, When, How, Which Why is

More information

A Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data

A Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data A Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data Alexey Samsonov, Julia Velikina Departments of Radiology and Medical

More information

SISCOM (Subtraction Ictal SPECT CO-registered to MRI)

SISCOM (Subtraction Ictal SPECT CO-registered to MRI) SISCOM (Subtraction Ictal SPECT CO-registered to MRI) Introduction A method for advanced imaging of epilepsy patients has been developed with Analyze at the Mayo Foundation which uses a combination of

More information

Release Notes. Multi-Band EPI C2P. Release HCP_v1 10 February 2014

Release Notes. Multi-Band EPI C2P. Release HCP_v1 10 February 2014 Release Notes Multi-Band EPI C2P Release HCP_v1 10 February 2014 Installation 1. Restart the system (reboot host and MRIR) 2. Extract the.zip file to a temporary directory 3. Run the installer.bat file

More information

SIEMENS MAGNETOM Symphony syngo MR A30

SIEMENS MAGNETOM Symphony syngo MR A30 \\USER\ADNI STUDY\MAIN PROTOCOL\HUMAN PROTOCOL\localizer Scan Time: 9.2 [s] Voxel size: 2.2 1.1 10.0 [mm] Rel. SNR: 1.00 SIEMENS: gre Slice group 1 Slice group 2 Slice group 3 280 [mm] 10 [mm] 20 [ms]

More information

Motion Artifacts and Suppression in MRI At a Glance

Motion Artifacts and Suppression in MRI At a Glance Motion Artifacts and Suppression in MRI At a Glance Xiaodong Zhong, PhD MR R&D Collaborations Siemens Healthcare MRI Motion Artifacts and Suppression At a Glance Outline Background Physics Common Motion

More information

Technical Publications

Technical Publications GE Medical Systems Technical Publications Direction 2188003-100 Revision 0 Tissue Volume Analysis DICOM for DICOM V3.0 Copyright 1997 By General Electric Co. Do not duplicate REVISION HISTORY REV DATE

More information

MB-EPI PCASL. Release Notes for Version February 2015

MB-EPI PCASL. Release Notes for Version February 2015 MB-EPI PCASL Release Notes for Version 1.0 20 February 2015 1 Background High-resolution arterial spin labeling (ASL) imaging is highly desirable in both neuroscience research and clinical applications

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

3T/7T MRI FACILITY. 3T MRI System Start-Up and Shutdown

3T/7T MRI FACILITY. 3T MRI System Start-Up and Shutdown 3T/7T MRI FACILITY SOP Number: 300.02 Title 3T MRI System Start-Up and Shutdown Revision Chronology Version Number Date Changes 300.01 01 March 2010 New 300.02 21 January 2013 Updated procedures Facility

More information

T 1 MAPPING FOR DCE-MRI

T 1 MAPPING FOR DCE-MRI T 1 MAPPING FOR DCE-MRI A dissertation submitted to the Faculty of Medicine, University of Malaya in partial fulfillment of the requirements for the degree of Master of Medical Physics By NURUN NAJWA BINTI

More information

Imaging Notes, Part IV

Imaging Notes, Part IV BME 483 MRI Notes 34 page 1 Imaging Notes, Part IV Slice Selective Excitation The most common approach for dealing with the 3 rd (z) dimension is to use slice selective excitation. This is done by applying

More information

LOGIQ. V2 Ultrasound. Part of LOGIQ Vision Series. Imagination at work LOGIQ is a trademark of General Electric Company.

LOGIQ. V2 Ultrasound. Part of LOGIQ Vision Series. Imagination at work LOGIQ is a trademark of General Electric Company. TM LOGIQ V2 Ultrasound Part of LOGIQ Vision Series Imagination at work The brilliance of color. The simplicity of GE. Now you can add the advanced capabilities of color Doppler to patient care with the

More information

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1 Modifications for P551 Fall 2013 Medical Physics Laboratory Introduction Following the introductory lab 0, this lab exercise the student through

More information

Single Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction

Single Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction Single Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction Felix Lugauer 1,3, Jens Wetzl 1, Christoph Forman 2, Manuel Schneider 1, Berthold Kiefer

More information

Whole Body MRI Intensity Standardization

Whole Body MRI Intensity Standardization Whole Body MRI Intensity Standardization Florian Jäger 1, László Nyúl 1, Bernd Frericks 2, Frank Wacker 2 and Joachim Hornegger 1 1 Institute of Pattern Recognition, University of Erlangen, {jaeger,nyul,hornegger}@informatik.uni-erlangen.de

More information

Fast Imaging UCLA. Class Business. Class Business. Daniel B. Ennis, Ph.D. Magnetic Resonance Research Labs. Tuesday (3/7) from 6-9pm HW #1 HW #2

Fast Imaging UCLA. Class Business. Class Business. Daniel B. Ennis, Ph.D. Magnetic Resonance Research Labs. Tuesday (3/7) from 6-9pm HW #1 HW #2 Fast Imaging Daniel B. Ennis, Ph.D. Magnetic Resonance Research Labs Class Business Tuesday (3/7) from 6-9pm 6:00-7:30pm Groups Avanto Sara Said, Yara Azar, April Pan Skyra Timothy Marcum, Diana Lopez,

More information

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis Clinical Importance Rapid cardiovascular flow quantitation using sliceselective Fourier velocity encoding with spiral readouts Valve disease affects 10% of patients with heart disease in the U.S. Most

More information

AAPM Standard of Practice: CT Protocol Review Physicist

AAPM Standard of Practice: CT Protocol Review Physicist AAPM Standard of Practice: CT Protocol Review Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center September 11, 2014 2014 Texas Radiation Regulatory Conference Goals Understand purpose

More information

Segmentation of Bony Structures with Ligament Attachment Sites

Segmentation of Bony Structures with Ligament Attachment Sites Segmentation of Bony Structures with Ligament Attachment Sites Heiko Seim 1, Hans Lamecker 1, Markus Heller 2, Stefan Zachow 1 1 Visualisierung und Datenanalyse, Zuse-Institut Berlin (ZIB), 14195 Berlin

More information

Philips MRI Protocol Dump Created on Comment Software Stream

Philips MRI Protocol Dump Created on Comment Software Stream Page 1 of 5 Philips MRI Protocol Dump Created on 2/17/2011 4:11:01 PM Comment Created by ExamCard_to_XML with inputs: "J:\ADNI GO - ADNI 2 Phantom5.ExamCard" on system (BU SCHOOL OF MEDICINE :: 192.168.71.10)

More information

SICAT SUITE VERSION 1.0. Instructions for use SIDEXIS XG English

SICAT SUITE VERSION 1.0. Instructions for use SIDEXIS XG English SICAT SUITE VERSION 1.0 Instructions for use SIDEXIS XG English Table of contents Table of contents 1 System requirements... 3 2 Overview of SICAT Suite... 5 3 Differences between full featured mode and

More information

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005 Field Maps John Pauly October 5, 25 The acquisition and reconstruction of frequency, or field, maps is important for both the acquisition of MRI data, and for its reconstruction. Many of the imaging methods

More information

Basic fmri Design and Analysis. Preprocessing

Basic fmri Design and Analysis. Preprocessing Basic fmri Design and Analysis Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial filtering

More information

3D imaging with SPACE vs 2D TSE in MR guided prostate biopsy

3D imaging with SPACE vs 2D TSE in MR guided prostate biopsy 3D imaging with SPACE vs 2D TSE in MR guided prostate biopsy Poster No.: C-2004 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper M. Garmer, S. Mateiescu, M. Busch, D. Groenemeyer; Bochum/

More information

FOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION

FOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION Philips J. Res. 51 (1998) 197-201 FOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION This special issue of Philips Journalof Research includes a number of papers presented at a Philips

More information

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist CT Protocol Review: Practical Tips for the Imaging Physicist Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center August 8, 2013 AAPM Annual Meeting Goals Understand purpose and importance

More information

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland Metal Artifact Reduction CT Techniques R S S S Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland N. 1 v o 4 1 0 2. Postoperative CT Metal Implants CT is accurate for assessment

More information

Medical Image Processing: Image Reconstruction and 3D Renderings

Medical Image Processing: Image Reconstruction and 3D Renderings Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,

More information

Accelerated MRI Techniques: Basics of Parallel Imaging and Compressed Sensing

Accelerated MRI Techniques: Basics of Parallel Imaging and Compressed Sensing Accelerated MRI Techniques: Basics of Parallel Imaging and Compressed Sensing Peng Hu, Ph.D. Associate Professor Department of Radiological Sciences PengHu@mednet.ucla.edu 310-267-6838 MRI... MRI has low

More information

The SIMRI project A versatile and interactive MRI simulator *

The SIMRI project A versatile and interactive MRI simulator * COST B21 Meeting, Lodz, 6-9 Oct. 2005 The SIMRI project A versatile and interactive MRI simulator * H. Benoit-Cattin 1, G. Collewet 2, B. Belaroussi 1, H. Saint-Jalmes 3, C. Odet 1 1 CREATIS, UMR CNRS

More information

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial

More information

Outline: Contrast-enhanced MRA

Outline: Contrast-enhanced MRA Outline: Contrast-enhanced MRA Background Technique Clinical Indications Future Directions Disclosures: GE Health Care: Research support Consultant: Bracco, Bayer The Basics During rapid IV infusion, Gadolinium

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

ADNI GO - ADNI 2 Human7 (9) 38:30.1

ADNI GO - ADNI 2 Human7 (9) 38:30.1 Philips RI Protocol Dump Create on 11/25/2013 11:40:24 A Comment Create by ExamCar_to_XL with inputs: "K:\ADNI GO - ADNI 2 Human7.ExamCar" on system (BU SCHOOL OF EDICINE :: 192.168.71.10) Software Stream

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Parallel Magnetic Resonance Imaging (pmri): How Does it Work, and What is it Good For?

Parallel Magnetic Resonance Imaging (pmri): How Does it Work, and What is it Good For? Parallel Magnetic Resonance Imaging (pmri): How Does it Work, and What is it Good For? Nathan Yanasak, Ph.D. Chair, AAPM TG118 Department of Radiology Georgia Regents University Overview Phased-array coils

More information