Comparison of Quality of Multiplanar Reconstructions and Direct Coronal Multidetector CT Scans of the Lung

Size: px
Start display at page:

Download "Comparison of Quality of Multiplanar Reconstructions and Direct Coronal Multidetector CT Scans of the Lung"

Transcription

1 Osamu Honda 1 Takeshi Johkoh 1 Shuji Yamamoto 2 Mitsuhiro Koyama 1 Noriyuki Tomiyama 1 Takenori Kozuka 1 Seiki Hamada 1 Naoki Mihara 1 Hironobu Nakamura 1 Nestor L. Müller 3 Received January 2, 22; accepted after revision March 29, Department of Radiology, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita, Osaka, , Japan. Address correspondence to O. Honda. 2 Department of Radiology, Osaka University Hospital, Osaka, , Japan. 3 Department of Radiology, Vancouver Hospital and Health Sciences Centre and University of British Columbia, 855 W. 12th Ave., Vancouver, B. C. V5Z 1M9, Canada. AJR 22;179: X/2/ American Roentgen Ray Society Comparison of Quality of Multiplanar Reconstructions and Direct Coronal Multidetector CT Scans of the Lung OBJECTIVE. The purpose of this study was to compare the quality of coronal multiplanar reconstructions with the quality of direct coronal thin-section multidetector CT (MDCT) scans. MATERIALS AND METHODS. Axial multidetector CT (MDCT) scans were obtained through the entire lung in 1 normal autopsy lung specimens using an MDCT scanner. Four protocols were used:.5-mm collimation with a.5-mm reconstruction interval;.5-mm collimation with a.3-mm reconstruction interval; 1-mm collimation with a.5-mm reconstruction interval; and 2-mm collimation with a 1-mm reconstruction interval. Multiplanar reconstruction images with.5-mm slice thickness were obtained from the four types of data sets. Direct coronal thin-section CT of the same 1 autopsy lung specimens was performed using.5-mm scan collimation, a.3-mm reconstruction interval, a 25.6-cm field of view, and a matrix. Two independent observers compared the image quality of each of the four coronal multiplanar reconstruction sets with that of direct coronal thin-section CT scans. The observers analyzed visualization of anatomic features and artifacts. RESULTS. The total image quality of the multiplanar reconstructions obtained from.5-mm collimation data with or without.3-mm overlapping reconstruction was equal to that of direct coronal thin-section CT scans in all 2 interpretations. The image quality of multiplanar reconstruction images from.5-mm collimation data either with or without overlapping reconstruction was superior to multiplanar reconstruction images obtained from 1- or 2-mm collimation scans ( p <.1, Fisher s exact test). Stairstep artifacts in multiplanar reconstructions using.5-mm collimation without overlapping reconstruction were equal to those with overlapping reconstruction and were fewer than those on 1- or 2-mm collimation (p <.1, Mann-Whitney U test). CONCLUSION. The image quality of coronal multiplanar reconstructions from isotropic voxel data obtained using.5-mm collimation, with or without overlapping reconstruction, is similar to that of direct coronal thin-section CT scans. T o improve the quality of multiplanar reconstruction images, axial multidetector CT (MDCT) images must be reconstructed in an overlapping fashion [1 3]. In 1995, Kalender [1] predicted that isotropic scanning would become possible. MDCT scanners can rapidly scan large longitudinal (z) volumes with high z-axis resolution [4], and MDCT scanners show smaller stairstep artifacts than single-detector CT scanners [5]. MDCT scanners allow contiguous thin-section CT to be performed through the entire lung and isotropic imaging that is, equal spatial resolution in the x y plane and the z-axis. MDCT scanners can potentially allow the depiction of images in any desired plane with a resolution similar to that in the transverse plane. Optimal resolution on thin-section CT is achieved using the highest matrix size ( ) and the lowest field of view that will include the entire region of interest. CT performed using 1-mm collimation and a matrix requires a field of view of 51.2 cm to obtain isotropic voxel data sets. The introduction of MDCT scanners with the ability to obtain.5-mm-thick sections allows the reduction of the field of view to 25.6 cm. To our knowledge, no study has evaluated the image quality of multiplanar reconstruction lung images obtained from isotropic voxel data sets. The purpose of this study was to compare the image quality of coronal multiplanar reconstructions of MDCT scans performed through normal autopsy lung specimens with that of direct coronal thin-section MDCT scans. AJR:179, October

2 Honda et al. Materials and Methods Ten lungs of patients who did not have any history of pulmonary disease were autopsied, inflated, and fixed by the method of Markarian and Dailey [6]. The lungs were distended through the main bronchus with a fixative fluid that contained polyethylene glycol 4, 95% ethyl alcohol, 4% formalin, and water in a proportion of 1:5:2:3. The specimens were immersed in the fixative for 2 days. The fixed lungs were then air-dried. Four of these 1 lungs were right lungs and six were left lungs. Axial continuous MDCT scans were obtained on an MDCT scanner (Aquillion-VZ; Toshiba, Tokyo, Japan) using the following protocols:.5-mm collimation with a.5-mm reconstruction interval;.5-mm collimation with a.3-mm reconstruction interval; 1-mm collimation with a.5-mm reconstruction interval; and 2-mm collimation with a 1- mm reconstruction interval. The scanned distances were all 25.6 cm. All scans were obtained using a 6:1 pitch,.5 sec per rotation, 12 kv, 1 mas, a cm field of view, a matrix, and a high-frequency reconstruction algorithm. As the control for image quality, direct coronal thin-section CT scans of the same 1 autopsied lungs were obtained on the same MDCT scanner. Direct coronal thin-section CT was performed using.5-mm scan collimation, a 6:1 pitch, a.3-mm reconstruction interval,.5 sec per rotation, 12 kv, 1 mas, a 25.6-cm field of view, a matrix, and a high-frequency reconstruction algorithm. All images were transferred to a commercially available workstation (Advantage Windows 3.; General Electric Medical Systems, Milwaukee, WI), and coronal multiplanar reconstruction images with.5-mm slice thickness and.5-mm slice intervals were obtained from the four axial MDCT data sets. All coronal multiplanar reconstruction images were reconstructed to fit the direct coronal thin-section CT images on the workstation. All four sets of coronal multiplanar reconstructions and the direct coronal thin-section CT scans were photographed separately on hard copies using a window width of 12 H and a window level of 7 H. After the CT examination, transverse sections were obtained from the same lungs every 1 mm and were stained with H and E and elastica van Gieson stains for histopathologic assessment. Macroscopic and microscopic assessment of the lungs showed no parenchymal abnormalities. Image quality of each coronal multiplanar reconstruction image was evaluated as being equal to or inferior to that of the direct coronal thin-section CT scan by two independent pulmonary radiologists who paid special attention to visualization of the following six normal structures: the interlobular septa, the central bronchovascular bundles coursing vertically, the central bronchovascular bundles coursing horizontally, the peripheral bronchovascular bundles, the interlobar fissures, and the pulmonary parenchyma. In addition, the total image quality of each coronal multiplanar reconstruction image was subjectively evaluated as being equal to or inferior to that of direct coronal thin-section CT scan. Stairstep artifacts of each multiplanar reconstruction image were graded as 1, none; 2, mild; or 3, severe, compared with the direct coronal thin-section CT scan. We required that at least 4 weeks elapse between each set of interpretations. The difference in image quality of each multiplanar reconstruction image was analyzed using Fisher s exact probability test. The scores of stairstep artifacts of each multiplanar reconstruction were evaluated using the Mann-Whitney U test. The interobserver variation of the image quality between two reviewers was evaluated using kappa statistics. The interobserver agreement was classified as follows: poor,.2; fair,.21.4; moderate,.41.6; good,.61.8; and excellent, The interobserver variation for noise and stairstep artifacts was analyzed using Spearman s correlation coefficient. Results The two observers had good agreement (κ =.76) for the evaluation of image quality and stairstep artifacts (Spearman s correlation coefficient, r >.7; p <.1). Therefore, the statistical analysis of the comparison of image quality among the various sets of multiplanar reconstruction images was performed by combining the results of the two observers. TABLE 1 Quality of Multiplanar Reconstructions and Direct Coronal Thin-Section CT Images Multiplanar Reconstruction Protocol Interlobular Septa Coronal Thin-Section CT Images (n = 2) Bronchovascular Bundles The image quality results are summarized in Table 1. The total image quality of multiplanar reconstructions obtained from the.5-mm collimation data sets with or without.3-mm overlapping reconstruction was equal to that of the direct coronal thin-section CT scans in all 2 interpretations (1 lungs, two observers). The image quality of the other multiplanar reconstructions from.5-mm collimation data sets with or without.3-mm overlapping reconstructions was also equal to that of direct coronal thin-section CT scans except in one case of interlobular septa. The total image quality of multiplanar reconstruction images from.5- mm collimation data sets either with or without overlapping reconstruction was superior to that of multiplanar reconstructions obtained from 1- and 2-mm collimation data sets (Fisher s exact probability test, p <.1) (Fig. 1). Stairstep artifacts in multiplanar reconstruction images are summarized in Table 2. The scores of stairstep artifacts in multiplanar reconstructions obtained from.5-mm collimation data sets without overlapping reconstruction were equal to those obtained from.5-mm collimation data sets with overlapping reconstruction. The scores of stairstep artifacts in multiplanar reconstructions obtained from.5- mm collimation data sets with overlapping reconstruction were superior to those from 1- and 2-mm collimation data sets (Mann-Whitney U test, p <.1). Discussion Helical CT has been shown to have a spatial resolution similar to that of conventional CT in the imaging plane and a greater resolution than that of conventional CT in the z- axis [1]. However, until recently multiplanar reconstruction or three-dimensional images from axial helical CT data sets have been accompanied by stairstep artifacts along the z- axis [7]. Resolution in the z-axis can be Major or Minor Fissure Pulmonary Parenchyma Total Image Quality Vertical Central Horizontal Central Peripheral.5-mm collimation,.3-mm overlap mm collimation, no overlap mm collimation,.5-mm overlap mm collimation, 1-mm overlap 1 Note. Total quality of multiplanar reconstructions from.5-mm-collimation data sets either with or without overlapping reconstruction was superior to that of multiplanar reconstructions obtained from 1- and 2-mm-collimation data sets (Fisher s exact probability test, p <.1). Image quality of multiplanar reconstructions from.5-mm-collimation data sets either with or without overlapping reconstruction was superior to that from 1- and 2-mm-collimation data sets. 876 AJR:179, October 22

3 Multidetector CT and Multiplanar Reconstructions of the Lung A B Fig. 1. Lung of 68-year-old male cadaver. A, Direct coronal thin-section multidetector CT image of lung with.5-mm collimation and pitch of 6:1 clearly shows interlobular septum (arrowhead ), central and peripheral bronchovascular bundles, and major fissure (arrow ). B, Coronal multiplanar reconstruction image from.5-mm-collimation data sets with.3-mm overlapping reconstruction shows image quality similar to that of A. C, Coronal multiplanar reconstruction image from.5-mm-collimation data set without overlapping reconstruction shows image quality similar to that of B. (Fig. 1 continues on next page) C AJR:179, October

4 Honda et al. D E Fig. 1. (continued) Lung of 68-year-old male cadaver. D, On coronal multiplanar reconstruction image from 1-mm-collimation data set with.5-mm overlapping reconstruction, interlobular septum (arrowhead ) is ambiguous compared with direct coronal thin-section CT image (A) and coronal multiplanar reconstruction images from.5-mm-collimation data set with (B) and without (C).3-mm overlapping reconstruction. E, Coronal multiplanar reconstruction image from 2-mm-collimation data set with 1-mm overlapping reconstruction shows considerable noise. Interlobular septum, central and peripheral bronchovascular bundles, and major fissure (arrow ) are ambiguous. improved by using overlapping reconstructions [1, 2]. However, depiction of anatomic structures in arbitrarily chosen planes with the same image quality as in the transverse plane requires isotropic imaging. In this study, we used.5-mm collimation, a 25.6-cm field of view, and a matrix. Pixel size in the x y plane was calculated as follows: 256-mm field of view / 512 pixels =.5 mm. Therefore, isotopic imaging has been theoretically achieved. The results of the study showed similar image quality for coronal multiplanar images reconstructed from axial scans obtained using.5-mm collimation and for direct coronal CT scans. Furthermore, these multiplanar reconstructions had no stairstep artifacts. These findings confirm the hypothesis that isotropic imaging was achieved. Caldemeyer et al. [8] reported that the diagnostic efficacy of coronal multiplanar reconstruction images with.5-mm collimation on dual-detector helical CT was comparable to or superior to that of direct coronal CT scans with mm collimation on temporal bone. Venema et al. [9] also reported that coronal multiplanar reconstructions from axial dual-detector helical CT obtained with.5-mm collimation could replace direct coronal sequential CT scans of petrosal bone. In their study, overlapping reconstruction was used. However, our results suggest that overlapping reconstruction is not necessary for coronal multiplanar recon- TABLE 2 structions from isotropic voxel data sets for lung parenchyma. Our study has several limitations. First, because we used autopsy lung specimens, it was not possible to evaluate the influence of motion artifacts. However, ethical concerns exist about performing repeated CT in any given patient. Furthermore, it is not possible to obtain direct coronal images of human lungs using Noise and Stairstep Artifacts in Multiplanar Reconstructions Stairstep Artifacts (n = 2) Multiplanar Reconstruction Protocol.5-mm collimation,.3-mm overlap None Mild Severe mm collimation, no overlap mm collimation,.5-mm overlap mm collimation, 1-mm overlap Note. Data are numbers of reconstructions in each scoring category. Multiplanar reconstructions obtained from.5-mmcollimation data sets without overlapping reconstruction were equal to those from.5-mm-collimation data sets with overlapping reconstruction (Mann-Whitney U test, p <.5). Multiplanar reconstructions obtained from.5-mm-collimation data sets with overlapping reconstruction were superior to those from 1- and 2-mm-collimation data sets (Mann-Whitney U test, p <.1). AJR:179, October 22

5 Multidetector CT and Multiplanar Reconstructions of the Lung commercially available CT scanners. Second, the study was limited to normal lungs. It was reported that MDCT high-resolution CT images of various modes were compared with single-detector high-resolution images in the axial mode, and high-resolution MDCT images of a 6:1 pitch were unsuitable for detecting subtle lung abnormalities on axial MDCT images of and 2.5-mm collimation [1]. In our study, we used a pitch of 6:1 and.5-mm collimation. We chose.5-mm collimation to obtain the multiplanar reconstruction images from isotropic voxels, and a 6:1 pitch to cover a large area in one breath-hold. If axial high-resolution CT is used for direct coronal high-resolution CT instead of a high-resolution MDCT pitch of 6:1, multiplanar reconstruction images from MDCT data of a 6:1 pitch would be inferior to direct coronal high-resolution axial CT scans. Further study is needed with scanning of abnormal lungs, especially lungs with subtle abnormalities. If more rapid scanning could be achieved, it would be possible to obtain superior multiplanar reconstruction images of the whole lung from isotropic voxel data. The third potential limitation is a possible bias because the observers knew which of the images were obtained using the direct coronal technique, which may have biased them to make a superior evaluation of the multiplanar reconstruction images with.5-mm collimation. However, comparing the images with direct coronal images has merit to test for the inferiority of multiplanar reconstruction images as well. As for the multiplanar reconstructions, the observers did not know how the images were scanned and reconstructed; therefore, their findings that the image quality of coronal multiplanar reconstruction images from isotropic voxel data sets without overlapping reconstruction is almost identical to that with overlapping reconstruction are reliable. In conclusion, the results of our study indicate that the image quality of coronal multiplanar reconstructions obtained from isotropic voxel data sets is similar to the image quality of direct coronal thin-section CT scans. The image quality of coronal multiplanar reconstructions from isotropic voxel data sets without overlapping reconstruction is almost identical to that of sets with overlapping reconstruction. References 1. Kalender WA. Thin-section three-dimensional spiral CT: is isotropic imaging possible? Radiology 1995;197: Kasales CJ, Hopper KD, Ariola DN, et al. Reconstructed helical CT scans: improvement in z-axis resolution compared with overlapped and nonoverlapped conventional CT scans. AJR 1995; 164: Wang G, Vannier MW. Stair-step artifacts in three-dimensional helical CT: an experimental study. Radiology 1994;191: Hu H. Multi-slice helical CT: scan and reconstruction. Med Phys 1999;26: Fleischmann D, Rubin GD, Paik DS. Stair-step artifacts with single versus multiple detector-row helical CT. Radiology 2;216: Markarian B, Dailey ET. Preparation of inflated lung specimens. In: Heitzman ER, ed. The lung: radiologic pathologic correlations, 3rd ed. St. Louis: Mosby, 1993: Brink JA. Technical aspects of helical (spiral) CT. Radiol Clin North Am 1995;33: Caldemeyer KS, Sandrasegaran K, Shinaver CN, Mathews VP, Smith RR, Kopecky KK. Temporal bone: comparison of isotropic helical CT and conventional direct axial and coronal CT. AJR 1999;172: Venema HW, Phoa SS, Mirck PG, Hulsmans FJ, Majoie CB, Verbeeten B Jr. Petrosal bone: coronal reconstructions from axial spiral CT data obtained with.5-mm collimation can replace direct coronal sequential CT scans. Radiology 1999;213: Honda O, Johkoh T, Tomiyama N, et al. High-resolution CT using multidetector CT equipment: evaluation of image quality in 11 cadaveric lungs and a phantom. AJR 21;177: AJR:179, October

Medical Image Processing: Image Reconstruction and 3D Renderings

Medical Image Processing: Image Reconstruction and 3D Renderings Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead.

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead. 1 CT Basics Principles of Spiral CT Dose 2 Who invented CT? 1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections Sir Godfrey Hounsfield worked for the Central

More information

Optimisation of Toshiba Aquilion ONE Volume Imaging

Optimisation of Toshiba Aquilion ONE Volume Imaging Optimisation of Toshiba Aquilion ONE Volume Imaging Jane Edwards, RPRSG Royal Free London NHS Foundation Trust Dr Mufudzi Maviki, Plymouth Hospitals NHS Trust Background In 2011/12 Radiology at RFH was

More information

Motion artifact detection in four-dimensional computed tomography images

Motion artifact detection in four-dimensional computed tomography images Motion artifact detection in four-dimensional computed tomography images G Bouilhol 1,, M Ayadi, R Pinho, S Rit 1, and D Sarrut 1, 1 University of Lyon, CREATIS; CNRS UMR 5; Inserm U144; INSA-Lyon; University

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Fundamentals of CT imaging

Fundamentals of CT imaging SECTION 1 Fundamentals of CT imaging I History In the early 1970s Sir Godfrey Hounsfield s research produced the first clinically useful CT scans. Original scanners took approximately 6 minutes to perform

More information

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

Improvement of Efficiency and Flexibility in Multi-slice Helical CT J. Shanghai Jiaotong Univ. (Sci.), 2008, 13(4): 408 412 DOI: 10.1007/s12204-008-0408-x Improvement of Efficiency and Flexibility in Multi-slice Helical CT SUN Wen-wu 1 ( ), CHEN Si-ping 2 ( ), ZHUANG Tian-ge

More information

Philips SPECT/CT Systems

Philips SPECT/CT Systems Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET

More information

Denoising of Coronary artery X-ray angiography image using UDWT

Denoising of Coronary artery X-ray angiography image using UDWT Denoising of Coronary artery X-ray angiography image using UDWT Mrs.A.Umarani 1, Dr.A.Asha 2, R.Jayavalanarasu 3, T.J.Ramkumar 4 Department of EIE 1,3,4, Department of Mechanical Engineering 2 duraiumarani@gmail.com

More information

Extraction and recognition of the thoracic organs based on 3D CT images and its application

Extraction and recognition of the thoracic organs based on 3D CT images and its application 1 Extraction and recognition of the thoracic organs based on 3D CT images and its application Xiangrong Zhou, PhD a, Takeshi Hara, PhD b, Hiroshi Fujita, PhD b, Yoshihiro Ida, RT c, Kazuhiro Katada, MD

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

A closer look at CT scanning

A closer look at CT scanning Vet Times The website for the veterinary profession https://www.vettimes.co.uk A closer look at CT scanning Author : Charissa Lee, Natalie Webster Categories : General, Vets Date : April 3, 2017 A basic

More information

Computer-Tomography II: Image reconstruction and applications

Computer-Tomography II: Image reconstruction and applications Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de

More information

RADIOLOGY AND DIAGNOSTIC IMAGING

RADIOLOGY AND DIAGNOSTIC IMAGING Day 2 part 2 RADIOLOGY AND DIAGNOSTIC IMAGING Dr hab. Zbigniew Serafin, MD, PhD serafin@cm.umk.pl 2 3 4 5 CT technique CT technique 6 CT system Kanal K: RSNA/AAPM web module: CT Systems & CT Image Quality

More information

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed

More information

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist CT Protocol Review: Practical Tips for the Imaging Physicist Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center August 8, 2013 AAPM Annual Meeting Goals Understand purpose and importance

More information

Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice

Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice Guido Hilgers, a Tonnis Nuver, André Minken Department

More information

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare

More information

MDCT and 3D Workstations

MDCT and 3D Workstations MDCT and 3D Workstations Scott A. Lipson, MD Associate Director of Imaging, Long Beach Memorial Medical Center, Long Beach, California MDCT and 3D Workstations A Practical How-To Guide and Teaching File

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

Agenda : Lung Density Breakout Session

Agenda : Lung Density Breakout Session Agenda : Lung Density Breakout Session 1. : Mathew Fuld and Bernice Hoppel 2. Automatic Exposure Control (AEC) Evaluation : Sean Fain Round 4 Project 3. Dose reduction effects on emphysema metrics : Philip

More information

Computational Medical Imaging Analysis Chapter 4: Image Visualization

Computational Medical Imaging Analysis Chapter 4: Image Visualization Computational Medical Imaging Analysis Chapter 4: Image Visualization Jun Zhang Laboratory for Computational Medical Imaging & Data Analysis Department of Computer Science University of Kentucky Lexington,

More information

An approximate cone beam reconstruction algorithm for gantry-tilted CT

An approximate cone beam reconstruction algorithm for gantry-tilted CT An approximate cone beam reconstruction algorithm for gantry-tilted CT Ming Yan a, Cishen Zhang ab, Hongzhu Liang a a School of Electrical & Electronic Engineering, Nanyang Technological University, Singapore;

More information

Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners

Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners Poster No.: C-1748 Congress: ECR 2012 Type: Educational Exhibit Authors: S. Miyashita

More information

CT: Physics Principles & Equipment Design

CT: Physics Principles & Equipment Design CT: Physics Principles & Equipment Design James Kofler, Ph.D Radiology Mayo Clinic Rochester, MN June 27, 2012 Disclosures Nothing to disclose Learning Objectives Understand fundamental concepts of - CT

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Cardiac Dual Energy CT: Technique

Cardiac Dual Energy CT: Technique RSNA 2013, VSCA51-01, Chicago, Dec. 5, 2013 Cardiac Radiology Series Cardiac Dual Energy CT: Technique Willi A. Kalender, Ph.D. Institute of Medical Physics University of Erlangen www.imp.uni-erlangen.de

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

Feldkamp-type image reconstruction from equiangular data

Feldkamp-type image reconstruction from equiangular data Journal of X-Ray Science and Technology 9 (2001) 113 120 113 IOS Press Feldkamp-type image reconstruction from equiangular data Ben Wang a, Hong Liu b, Shiying Zhao c and Ge Wang d a Department of Elec.

More information

Image Post-Processing, Workflow, & Interpretation

Image Post-Processing, Workflow, & Interpretation Image Post-Processing, Workflow, & Interpretation Richard L. Hallett, MD Chief, Cardiovascular Imaging Northwest Radiology Network Indianapolis, IN Adjunct Assistant Professor Stanford University Stanford,

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

,RSNA, Helical CT: Principles and Technical Considerations1

,RSNA, Helical CT: Principles and Technical Considerations1 Helical T: Principles and Technical onsiderations1 James A. Brink, MD #{149} Jay P. Heiken, MD #{149} Ge Wang, PhD #{149} Kevin W. McEnery, MD Francisj Schlueter, MD #{149} Michael W. Vannier, MD. INTRODUTION

More information

AAPM Standard of Practice: CT Protocol Review Physicist

AAPM Standard of Practice: CT Protocol Review Physicist AAPM Standard of Practice: CT Protocol Review Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center September 11, 2014 2014 Texas Radiation Regulatory Conference Goals Understand purpose

More information

Automated Lesion Detection Methods for 2D and 3D Chest X-Ray Images

Automated Lesion Detection Methods for 2D and 3D Chest X-Ray Images Automated Lesion Detection Methods for 2D and 3D Chest X-Ray Images Takeshi Hara, Hiroshi Fujita,Yongbum Lee, Hitoshi Yoshimura* and Shoji Kido** Department of Information Science, Gifu University Yanagido

More information

CT vs. VolumeScope: image quality and dose comparison

CT vs. VolumeScope: image quality and dose comparison CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,

More information

Enhancement Image Quality of CT Using Single Slice Spiral Technique

Enhancement Image Quality of CT Using Single Slice Spiral Technique Enhancement Image Quality of CT Using Single Slice Spiral Technique Doaa. N. Al Sheack 1 and Dr.Mohammed H. Ali Al Hayani 2 1 2 Electronic and Communications Engineering Department College of Engineering,

More information

A study on the section sensitivity profile in multi-row-detector spiral CT

A study on the section sensitivity profile in multi-row-detector spiral CT Journal of X-Ray Science and Technology (3) IOS Press A study on the section sensitivity profile in multi-row-detector spiral CT Ge Wang a, Mark Madsen a, Katherine Redford a, Shiying Zhao b and Michael

More information

Effects of Mid-Sagittal Plane Perturbation and Image Interpolation on Corpus Callosum Area Calculation

Effects of Mid-Sagittal Plane Perturbation and Image Interpolation on Corpus Callosum Area Calculation Effects of Mid-Sagittal Plane Perturbation and Image Interpolation on Corpus Callosum Area Calculation Omer Ishaq 1, Ghassan Hamarneh 1, Roger Tam and Anthony Traboulsee 1 Medical Image Analysis Lab, School

More information

Computer-Tomography I: Principles, History, Technology

Computer-Tomography I: Principles, History, Technology Computer-Tomography I: Principles, History, Technology Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de History

More information

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT Assessment of 3D performance metrics D and 3D Metrics of Performance Towards Quality Index: Volumetric imaging systems X-ray based Volumetric imaging systems: CBCT/CT Tomosynthesis Samuel Richard and Ehsan

More information

Anatomic measurement accuracy: CT parameters and 3D rendering effects

Anatomic measurement accuracy: CT parameters and 3D rendering effects Anatomic measurement accuracy: CT parameters and 3D rendering effects Brian J Whyms a, E Michael Schimek a, Houri K Vorperian a, Lindell R Gentry b, and Edward T Bersu c University of Wisconsin-Madison

More information

S. Guru Prasad, Ph.D., DABR

S. Guru Prasad, Ph.D., DABR PURPOSE S. Guru Prasad, Ph.D., DABR Director of Medical Physics IAEA Consultant NorthShore University Health System and University of Chicago, Pritzker School of Medicine Current TPS utilize more information

More information

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1 Modifications for P551 Fall 2013 Medical Physics Laboratory Introduction Following the introductory lab 0, this lab exercise the student through

More information

doi: /

doi: / Yiting Xie ; Anthony P. Reeves; Single 3D cell segmentation from optical CT microscope images. Proc. SPIE 934, Medical Imaging 214: Image Processing, 9343B (March 21, 214); doi:1.1117/12.243852. (214)

More information

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing ImPACT Information Leaflet No. 1: CT Scanner Acceptance Testing Version 1.02, 18/05/01 CONTENTS: 1. SCOPE OF LEAFLET 2. GENERAL PRINCIPLES OF ACCEPTANCE AND COMMISSIONING 2.1 PHANTOMS 2.2 EXPOSURE AND

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Benedikt Lorch 1, Martin Berger 1,2, Joachim Hornegger 1,2, Andreas Maier 1,2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg

More information

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE / Radiographic Tomosynthesis: Image Quality and Artifacts Reduction Baojun Li, Ph.D Department of Radiology Boston University Medical Center 2012 AAPM Annual Meeting Background Linear Trajectory Tomosynthesis

More information

Evaluation of AutoQA Lite TM Image Quality Measurement Software

Evaluation of AutoQA Lite TM Image Quality Measurement Software Evaluation of AutoQA Lite TM Image Quality Measurement Software Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4 2XU Phone: 0131 537 1161

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

ImPACT. CT dosimetry and a data base for CTDI values. EFOMP Workshop at ECR Imaging Performance Assessment of CT Scanners

ImPACT. CT dosimetry and a data base for CTDI values. EFOMP Workshop at ECR Imaging Performance Assessment of CT Scanners ImPACT Imaging Performance Assessment of CT Scanners A Medical Devices Agency Evaluation Group CT dosimetry and a data base for CTDI values EFOMP Workshop at ECR 2002 S. Edyvean, ImPACT St George s Hospital,

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

DUE to beam polychromacity in CT and the energy dependence

DUE to beam polychromacity in CT and the energy dependence 1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact

More information

Automatic Lung Surface Registration Using Selective Distance Measure in Temporal CT Scans

Automatic Lung Surface Registration Using Selective Distance Measure in Temporal CT Scans Automatic Lung Surface Registration Using Selective Distance Measure in Temporal CT Scans Helen Hong 1, Jeongjin Lee 2, Kyung Won Lee 3, and Yeong Gil Shin 2 1 School of Electrical Engineering and Computer

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60601-2-44 2001 AMENDMENT 1 2002-09 Amendment 1 Medical electrical equipment Part 2-44: Particular requirements for the safety of X-ray equipment for computed tomography Amendement

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

Recognition and Measurement of Small Defects in ICT Testing

Recognition and Measurement of Small Defects in ICT Testing 19 th World Conference on Non-Destructive Testing 2016 Recognition and Measurement of Small Defects in ICT Testing Guo ZHIMIN, Ni PEIJUN, Zhang WEIGUO, Qi ZICHENG Inner Mongolia Metallic Materials Research

More information

Equipment Specification

Equipment Specification MULTISLICE CT SCANNER ( 16 Slices ) Merk : Hitachi Japan Model : SUPRIA 5 MHU Price : Rp 6.512.360.215,27 No Equipment Specification 1 2 3 Scanner Gantry Object for scanning : Whole body including head

More information

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology Diagnostic imaging techniques Krasznai Zoltán University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology 1. Computer tomography (CT) 2. Gamma camera 3. Single Photon

More information

ECG-Correlated Imaging of the Heart with Subsecond Multislice Spiral CT

ECG-Correlated Imaging of the Heart with Subsecond Multislice Spiral CT 888 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 19, NO. 9, SEPTEMBER 2000 ECG-Correlated Imaging of the Heart with Subsecond Multislice Spiral CT Marc Kachelrieß*, Stefan Ulzheimer, and Willi A. Kalender

More information

An Efficient Technique For Multi-Phase Model Based Iterative Reconstruction

An Efficient Technique For Multi-Phase Model Based Iterative Reconstruction 1 An Efficient Technique For Multi-Phase Model Based Iterative Reconstruction Shiyu Xu, Debashish Pal and Jean-Baptiste Thibault Abstract Multi-phase scan is a fundamental CT acquisition technology used

More information

A prototype table-top inverse-geometry volumetric CT system

A prototype table-top inverse-geometry volumetric CT system A prototype table-top inverse-geometry volumetric CT system Taly Gilat Schmidt a Department of Radiology, Stanford University, Stanford, California 94305 Josh Star-Lack NexRay, Inc., Los Gatos, California

More information

Intraoperative Prostate Tracking with Slice-to-Volume Registration in MR

Intraoperative Prostate Tracking with Slice-to-Volume Registration in MR Intraoperative Prostate Tracking with Slice-to-Volume Registration in MR Sean Gill a, Purang Abolmaesumi a,b, Siddharth Vikal a, Parvin Mousavi a and Gabor Fichtinger a,b,* (a) School of Computing, Queen

More information

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography

More information

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT.

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT. Progress report: Development of assessment and predictive metrics for quantitative imaging in chest CT Subaward No: HHSN6801000050C (4a) PI: Ehsan Samei Reporting Period: month 1-18 Deliverables: 1. Deployment

More information

DICOM Correction Item

DICOM Correction Item DICOM Correction Item Correction Number CP-668 Log Summary: Type of Modification Addition Name of Standard PS 3.3, 3.17 2006 Rationale for Correction The term axial is common in practice, but is incorrectly

More information

Rule-Based Ventral Cavity Multi-organ Automatic Segmentation in CT Scans

Rule-Based Ventral Cavity Multi-organ Automatic Segmentation in CT Scans Rule-Based Ventral Cavity Multi-organ Automatic Segmentation in CT Scans Assaf B. Spanier (B) and Leo Joskowicz The Rachel and Selim Benin School of Computer Science and Engineering, The Hebrew University

More information

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Marc Kachelrieß, a Katia Sourbelle, and Willi A. Kalender Institute of Medical Physics, University of

More information

The Near Future in Cardiac CT Image Reconstruction

The Near Future in Cardiac CT Image Reconstruction SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de

More information

3D temporal subtraction on chest MDCT images using nonlinear image registration technique

3D temporal subtraction on chest MDCT images using nonlinear image registration technique 3D temporal subtraction on chest MDCT images using nonlinear image registration technique Poster No.: C-1072 Congress: ECR 2010 Type: Topic: Authors: Keywords: DOI: Scientific Exhibit Computer Applications

More information

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

C a t p h a n / T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 Internationally recognized for measuring the maximum obtainable performance of axial, spiral and multi-slice

More information

ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma

ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma ORAL DEFENSE 8 th of September 2017 Charlotte MARTIN Supervisor: Pr. MP REVEL M2 Bio Medical

More information

An Automated Initialization System for Robust Model-Based Segmentation of Lungs in CT Data

An Automated Initialization System for Robust Model-Based Segmentation of Lungs in CT Data Fifth International Workshop on Pulmonary Image Analysis -111- An Automated Initialization System for Robust Model-Based Segmentation of Lungs in CT Data Gurman Gill 1,3, Matthew Toews 4, and Reinhard

More information

GPU implementation for rapid iterative image reconstruction algorithm

GPU implementation for rapid iterative image reconstruction algorithm GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

The VesselGlyph: Focus & Context Visualization in CT-Angiography

The VesselGlyph: Focus & Context Visualization in CT-Angiography The VesselGlyph: Focus & Context Visualization in CT-Angiography Matúš Straka M. Šrámek, A. La Cruz E. Gröller, D. Fleischmann Contents Motivation:» Why again a new visualization method for vessel data?

More information

Construction of Voxel-type Phantom Based on Computed Tomographic Data of RANDO Phantom for the Monte Carlo Simulations

Construction of Voxel-type Phantom Based on Computed Tomographic Data of RANDO Phantom for the Monte Carlo Simulations Construction of Voxel-type Phantom Based on Computed Tomographic Data of RANDO Phantom for the Monte Carlo Simulations K. Minami 1, K. Ejiri 1, M. Shimo 1, M. Kato, Y. Takeuchi, K. Yonemochi, H. Toyama

More information

QIBA PET Amyloid BC March 11, Agenda

QIBA PET Amyloid BC March 11, Agenda QIBA PET Amyloid BC March 11, 2016 - Agenda 1. QIBA Round 6 Funding a. Deadlines b. What projects can be funded, what cannot c. Discussion of projects Mechanical phantom and DRO Paul & John? Any Profile

More information

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Qiao Yang 1,4, Meng Wu 2, Andreas Maier 1,3,4, Joachim Hornegger 1,3,4, Rebecca Fahrig

More information

Using Movie Clips in PowerPoint Presentations: Part 1, Compatibility Issues

Using Movie Clips in PowerPoint Presentations: Part 1, Compatibility Issues Movie Clips in Presentation s Computers in Radiology Chun-Shan Yam 1 Jonathan Kruskal Michael Larson Yam C-S, Kruskal J, Larson M DOI:10.2214/AJR.05.0238 Received February 11, 2005; accepted after revision

More information

TEP Hounsfield units. Related topics Attenuation coefficient, Hounsfield units

TEP Hounsfield units. Related topics Attenuation coefficient, Hounsfield units Hounsfield units TEP Related topics Attenuation coefficient, Hounsfield units Principle Depending on the type of CT scanner and the settings, the result of a CT scan of the same material can be different

More information

Continuation Format Page

Continuation Format Page C.1 PET with submillimeter spatial resolution Figure 2 shows two views of the high resolution PET experimental setup used to acquire preliminary data [92]. The mechanics of the proposed system are similar

More information

Data Collection Worksheet

Data Collection Worksheet Data Collection Worksheet Part I: Computed Tomography Liver Fat/Hepatic Steatosis Multidetector Computed Tomography (MDCT) Scan Protocol Individuals were scanned using an eight-slice MDCT (LightSpeed Ultra,

More information

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),

More information

3D Echocardiography Development Timeline

3D Echocardiography Development Timeline 3D Acquisition Strategies and Display Lissa Sugeng, MD, MPH Associate Professor of Medicine Director of the Yale Echo Lab and Echo Core Lab (YCRG) And Peter Flueckiger, MD Advanced Imaging Cardiology Fellow

More information

Chapter 4. Image Reconstruction and Review. Image Reconstruction

Chapter 4. Image Reconstruction and Review. Image Reconstruction Chapter 4 Image Reconstruction and Review Data acquisition with MDCT scanners is relatively straightforward. Scanners offer a fixed number of slice thickness options, and other variables are controlled

More information

Automated segmentations of skin, soft-tissue, and skeleton from torso CT images

Automated segmentations of skin, soft-tissue, and skeleton from torso CT images Automated segmentations of skin, soft-tissue, and skeleton from torso CT images Xiangrong Zhou* a, Takeshi Hara a, Hiroshi Fujita a, Ryujiro Yokoyama b, Takuji Kiryu b, and Hiroaki Hoshi b a Department

More information

Scaling Calibration in the ATRACT Algorithm

Scaling Calibration in the ATRACT Algorithm Scaling Calibration in the ATRACT Algorithm Yan Xia 1, Andreas Maier 1, Frank Dennerlein 2, Hannes G. Hofmann 1, Joachim Hornegger 1,3 1 Pattern Recognition Lab (LME), Friedrich-Alexander-University Erlangen-Nuremberg,

More information

USING cone-beam geometry with pinhole collimation,

USING cone-beam geometry with pinhole collimation, IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE 2009 687 A Backprojection-Based Parameter Estimation Technique for Skew-Slit Collimation Jacob A. Piatt, Student Member, IEEE, and Gengsheng L.

More information

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS 23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS P. Helberg 1,

More information

Motion Compensation from Short-Scan Data in Cardiac CT

Motion Compensation from Short-Scan Data in Cardiac CT Motion Compensation from Short-Scan Data in Cardiac CT Juliane Hahn 1,2, Thomas Allmendinger 1, Herbert Bruder 1, and Marc Kachelrieß 2 1 Siemens Healthcare GmbH, Forchheim, Germany 2 German Cancer Research

More information

Integrated System for Planning Peripheral Bronchoscopic Procedures

Integrated System for Planning Peripheral Bronchoscopic Procedures Integrated System for Planning Peripheral Bronchoscopic Procedures Jason D. Gibbs, Michael W. Graham, Kun-Chang Yu, and William E. Higgins Penn State University Dept. of Electrical Engineering University

More information

Tomographic Reconstruction

Tomographic Reconstruction Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)

More information

Respiratory Motion Estimation using a 3D Diaphragm Model

Respiratory Motion Estimation using a 3D Diaphragm Model Respiratory Motion Estimation using a 3D Diaphragm Model Marco Bögel 1,2, Christian Riess 1,2, Andreas Maier 1, Joachim Hornegger 1, Rebecca Fahrig 2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg 2

More information

AIDR 3D Iterative Reconstruction:

AIDR 3D Iterative Reconstruction: Iterative Reconstruction: Integrated, Automated and Adaptive Dose Reduction Erin Angel, PhD Manager, Clinical Sciences, CT Canon Medical Systems USA Iterative Reconstruction 1 Since the introduction of

More information