1 Arizona State University NanoFab HIGH TEMPERATURE PROBE STATION STANDARD OPERATION PROCEDURE Rev A

2 Table of Contents Contents Table of Contents Purpose / Scope Reference Documents Equipment / Supplies / Material Safety Set Up Procedures Operation Procedures Process Data Revision History Page 1

3 1. Purpose / Scope This SOP provides the operating instructions for the High Temperature Probe Station. This includes the mechanical probe station details, the micromanipulators, electrical connections, measurements (IV s with Agilent 4155B or CV s with HP4284), software control of the switch matrix (ICV Lite) and software to automate measurements(ics). 2. Reference Documents Agilent 4155C Users Guide V1.pdf Agilent 5146C V2, Agilent 4155C Screen setup ref Agilent 4284A CV Manual Agilent Impedance Measurement Handbook Dieter K. Schroeder, Semiconductor Material and Device Characterization 3. Equipment / Supplies / Material Micromanipulator Probe Station with Video monitor Micromanipulator, Model 525 Probe holder and tungsten probe tips Thermal Chuck with Temperature controller Agilent 4155C for IV measurements HP 4284A for CV measurements Agilent 41501B Pulse Generator HP E5250A Switch Matrix PC with software to control Measurements PC for data storage 4. Safety The main safety concern is when the probe chuck is hot ( C); do not touch samples or chuck or any surrounding fixtures. 5. Set Up Procedures The setup procedures are organized into the following topics: Mechanical, Micromanipulators, Connections, Software Instrument Control A. Mechanical Probe Station Controls 1. The probe station is inside a light-tight, shielded metal box with lid. 2. X & Y stage controls are on the left and right sides. 3. Z control of the probe ring in on the front side; note that it lifts the ring, not the stage. 4. Micromanipulators are positioned around the probe ring. 5. Microscope with video camera provide sample view on monitor; focusing knobs are behind the microscope. 6. Microscope light & zoom controls are on the black box; front-right side of enclosure. Light source is below the probe box with ON/OFF switch and fiber optics to microscope. 7. Traveling microscope X & Y controls are at the back, right side. Page 2

4 8. Triax connections to the measuring equipment are on the back right side of the probe box and labeled SMU1, SMU2, SMU3, SMU4 (front to back). Connection to the chuck is on the bottom right side. 9. Note the strain relief connectors on both the left and right side of the probe station; the top 2 feed-through connectors are triax and the bottom 3 are co-ax connectors; IMPORTANT: Do not mixup triax and co-ax connectors; All the probe stage wiring uses triax connectors. 10. Triax to coax connectors are required for the direct connection to the cap bridge. 11. Vacuum control is the switch, front, left side; holds sample in place. 12. Emergency shut off is the red button on the front side. B. Micromanipulators 1. The MM micromanipulators have a vacuum hold down; vacuum switch is on the side of the base; push and hold to release and re-position; release to hold. 2. Micromanipulators have X, Y and Z controls on the backside; a coarse Z control is on the front of the manipulators; note the triax wire is connected to the probe holder. 3. The triax connector from the manipulator goes to the triax feedthrough connector on the back-left or back-right side of the stage; NOTE: the top 2 feedthrough connectors on both sides are triax, the bottom 3 are co-ax connectors on both sides; All wiring in the station uses triax wires and connectors; DO NOT MIX! 4. At the tip of the holder is the probe needle. Standard needles are ~20 mil diameter tungsten wires with sharp needles on one end. The probe holder has an angled hole for the needle; a set-screw in the end of the holder clamps the needle in place. Please do not bend the needles! 5. Needle heights are set for most wafer samples, but can be adjusted if needed. To change, contact CSSER staff 6. If the needles are damaged, contact CSSER staff for replacement. 7. The needles often pick up material from probe pads and can generate contact problems for subsequent probing. To clean the needles, use a swab with a little IPA and gently clean the needle tips; inspect using the microscope. 8. In order to position the probe needles, follow this sequence; a. Lower the needle to just above the sample using the coarse Z control. b. Release the vacuum control and slide the manipulator until needle is visible in the microscope. c. Use the standard X, Y and Z controls to position the needle above the probe pad. d. Slowly lower the needle until it makes contact (at contact, needle moves horizontally) e. Follow the above steps to connect other manipulator needles to probe pads. f. If the next pad is outside the microscope view, use the traveling microscope to locate the next probe pad. 9. When you finish probing, lift the needle(s) from your sample using the Z and coarse-z controls; release the vacuum; rotate the manipulator to the back of the stage; Turn off the vacuum; remove your sample. 10. Turn off the microscope light and the monitor; close the lid. Page 3

5 C. Connections 1. The triax connectors from the manipulators goes to the Strain Relief triax feedthrough connectors on the left or right side of the probe station. Triax wires (white) connect the backside of these connectors to the corresponding feedthroughs on the wall of the probe box. 2. Ensure that each micromanipulator is connected to the corresponding SMU feedthrough; NOTE: they are labeled (from front to back) SMU1, SMU2, SMU3, SMU4. 3. If you are using a connection to the back side of the wafer, ensure a connection between the chuck connector on the back, bottom of the probe station and the appropriate SMU. D. Software Instrument Control 1. ICV-Lite software controls the switch matrix. This allows you to choose between the Agilent 4155C and HP 4284A measurement systems. 2. Log into the computer using your ISSAC ID and password; contact nanofab staff for computer access. 3. Open the ICV-Lite software; Use the pull-down menus; Select Development a. Select Device Connections Editor b. Choose 4155C (for IV s) or 4284A (for CV s) c. Select Edit ; Switch matrix schematic appears; Inputs are the first column; outputs are across the top row (connections to the probe box feedthroughs). O U T P U T S I 1 o N 2 o P 3 o U 4 o T 5 S Default connections for the Agilent 4255C. Page 4

6 O U T P U T S I 1 N 2 P 3 U 4 T 5 S o 10 o Default connections for the 4284A Note that Hi Pot & Hi I are connected to #9. Low Pot & Low I to #10 d. Select Make Connections Now e. Select Finish 4. This completed the section on selecting the measurement tool. You can close out the software and continue with your measurements. 6. Operation Procedures The operation procedures are divided into the following sections: IV Measurements, CV Measurements, Temperature Control, Software Measurement Control. A. IV Measurements This probe station uses an Agilent 4155C for IV measurements; Current range is from sub-pa to 100 ma; there are both voltage and power limitations. The 4155C has 4 independent power supplies [SMU1, SMU2, SMU3, SMU4] that will be connected to the corresponding 4 connectors on the probe box (back right side). 1. Place your sample on the probe stage; turn on the video monitor and light; Find correct probe pads and position micromanipulator needles (see section B ); note corresponding SMU 2. Agilent 4255C Front Panel Operation (refer to photo). Note the following blocks: a. Display on the left side with 7 vertical soft keys and 8 horizontal soft keys. Soft key functions change with screen. b. Page control block; just to the right side of the display. The 4 gray keys in a column allow one to set up and measure devices; start with the top key [Chan], then [Meas] then [Display] and then [Graph/List]. c. Knob to read out data points and arrow keys to move around screens. d. Measurement Block (right top); 6 keys; i. top row [Single, Repeat, Standby]; ii. bottom row [stop, Append, Stress(yellow)] e. Integration time keys; [Short, Medium, Long]: short tends to be noisy, medium is usually good, long yields the best resolution, but is slower] Page 5

7 f. Entry Block; Edit keys on left, Key pad with numbers or alphabet, Far right [Enter key], [blue key] for alphabet, [green key] for special characters. Note right column next to key board with suffix keys (m=milli, u=micro, n=nano, p=pico, f=femto). g. Ibasic programming keys (generally not used any more) h. File keys for saved programs. (generally not used any more) 3. Short Version of front panel measurement setup The following is a very brief description to set up and measure a 2-terminal device. a. Find the page control block with 4 gray keys in a column. b. Start with the top key [CHAN]; use soft keys (right side); Mode box [sweep]; choose best device representation for your sample [MOSFET], [Bipolar], [Diode]. c. Select [MEAS]; enter Start, Stop, Step voltages and Compliance current. d. Select [Display] key; enter graph X-axis Min, Max & Y-axis Min, Max, Linear or log scales. e. Select [Graph]; the blank graph is displayed. f. Select [Medium] integration time in the measurement bock. g. Select [Single] in the measurement block; measurement starts; green light is on; results show in graph; save your data. Page 6

8 Agilent 4155C photo of the front panel. For this 2-terminal example, connect your device to micromanipulators wired to SMU1 and SMU3. 4. Detailed Front panel measurement setup: Details steps and corresponding screen views a. Select [Chan] key in the Page control block; in the display move curser to Mode box; Soft keys (right side) include SWEEP, SAMPLING, QSCV plus Defaults, Bipolar, FET Vd-Id, (more ½), FET Vgs-Id, and Diode VF-IF, (more 2/2); Choose best fit for your device. For example, Select [Sweep] and then [Diode] for any 2 terminal measurement. Display: Unit Vname Iname Mode Func SMU1 VF IF V VAR1 SMU2 SMU3 V I Common Const SMU4 Page 7

9 b. Select [MEAS], the 2nd gray key, in the Page control Block; Use arrows to move around screen; Use soft key selections and/or key-in values and hit [Enter] key (on right side of entry block). Display: VAR1 VAR2 UNIT SMU1 NAME VF Sweep Mode Single Lin/Log Linear Start 0 Stop 10V Step 0.1 No of steps 101 Compliance 10mA Power Comp off c. Arrow to a box labeled Sweep*, located in the lower right of the screen; Soft key options for compliance: 1. Cont at any 2. Stop at any abnormality 3. Stop at compliance Choose #3, stop at Compliance. (Turns power off if current exceeds value [10mA for this example]). d. Select [Display], the 3 rd gray key, in the page control block. This page sets the plot limits; Xmin, Xmax, Ymin, Ymax as well as liner or log scales Display: X axis Y axis Y2 axis Name VF IF Scale Linear Linear Min 0 0 Max 10 10m Arrow around the screen, pick the names and scale (linear or log) using the softkeys. Key in values for Min and Max, use suffix keys if needed, & hit [enter] key. For example, in the Max for Y axis, key in 10, hit the m key, and [enter] to get 10mA. Note: Log works for all positive or all negative values, but not for mixed positive and negative values. e. Select [Graph/LIST], the 4 th gray key, in the page control block. The graph will be displayed according to the above limits; you can always re-scale or autoscale later. Page 8

10 f. Measure Now, find the Measurement Block and its 6 keys (Top right): Single Repeat Standby Stop Append Stress Below these keys are the integration time keys: Short Medium Long Centered at the top of the display graph is the integration time; default is short ; However, [Medium] is a generally a better starting choice. Now, Select [Single] in the measurement block; Measurement starts; Voltage steps and current is measured; Hit [Stop] if any problems! Green light is on. Results are plotted as data is taken; Green light goes out when finished. Note: select Soft key below the graph: [Marker] Note: select soft key on Right side, [Marker on]; Turn dial to read measurement values; they are displayed above the graph. Note: Repeat erases the 1 st data set and re-measures the device, and continues until [stop]. Note: [Append] keeps the 1 st data set and re-measures the device; both data sets show on graph. g. Save the data! You can save the data to either a 3.5 floppy drive or to a Windows XP computer. Upon 4255C power up, the default drive is Floppy. Ensure that the windows XP computer is turned on 1. To change the drive: 2. Select [System] Key; Select [Filer] softkey; Arrow to field at the top *DISK [PC] and select PC from soft keys on the right (or else select the default FLOPPY). 3. After completing measurement, toggle the [Graph/List] key to LIST. 4. Select softkey [SPREAD SHEET]; Displays ASCII SAVE setup screen. a. The box at the top *DISK [ ] should read *DISK [PC] or *[FLOPPY] 5. Arrow to Name box; Key in file name [FileName]; hit ENTER key (on far right side). a. <= 8 characters for floppy disc files b. <= 36 characters for network files 6. The rest of the fields should be automatically filled in with default values: a. Output Data Boxes from [ 1 ] to [max] b. Unit [OFF] c. Delimiter [Space] d. String Mark [NONE] 7. Select softkey [EXECUTE], bottom left. 8. After files are written, select softkey on the right [EXIT] 9. Select [Graph/List] to return to graph and/or more measurements. Excel can read the text files to import data into a spreadsheet (use comma & space delimiters). Find a computer with a 3.5 drive for the file transfer; save on a USB memory stick. The PC for data storage has a 3.5 drive (see above) Page 9

11 10. Find your files on the PC: Go to the Windows XP computer; turn on if needed; Allow ~5 min for computer to boot up. Go to My Computer ; Open Drive labeled D:\Agilent 4155C (shortcut on Desktop) Saved files should be in this folder as filename.txt files. Move files onto your thumb drive and onto your own computer. EXCEL will open the text files; use TAB and SPACE for delimiters. Test conditions are at the top of the file followed by the data; Appended data follows the initial values. IMPORTANT: Delete your files from the common area D:\Agilent 4155C when finished. 5. Re-assigning SMU s One more useful and helpful option is the following: This shortcut saves moving connectors or micromanipulators. Assume a 2-terminal measurement with micromanipulators connected to SMU2 and SMU4. Select [Chan] key; in the display move curser to Mode box; select [Sweep]. Soft key choices (right side) include MOSFET, Bipolar, Diode, ; Select [diode] for any 2 terminal measurement. Display is as below. Display: Unit Vname Iname Mode Func SMU1 VF IF V VAR1 SMU2 SMU3 V I Common Const SMU4 Use arrow keys to go to first Vname entry [VF]. Now, select soft key on the right column labeled [channel assign]; Curser moves to SMU1 box; soft keys are now labeled [SMU1], [SMU2], [SMU3] and [SMU4]. For this example, re-assign the VF row to SMU2 using the softkey; Curser moves down one row; re-assign SMU2 to SMU1, since it is not used for this example. Curser moves to SMU3 row; re-assign SMU3 to SMU4 using the softkey; Again, since SMU4 is not in use, re-assign to SMU3. Important! use softkey [exit channel assignment]; display now reads: Display: Unit Vname Iname Mode Func SMU1 SMU2 VF IF V VAR1 SMU3 SMU4 V I Common Const Now, continue through the next 3 page control keys and set up the measurement range & display. Page 10

12 B. CV measurements We will review the front panel of the 4284A Impedance Bridge and the many options and then illustrate a simple CV measurement. This bridge measures impedance (Z) and phase angle (θ) and then calculates the equivalent parameters such as parallel Cp with parallel G (conductance). 1. Use the ICV Lite software to connect the High and Low bridge terminals to 2 of the SMU connectors (see section D. Software Instrument Control). 2. The 4284A has a display with 5 soft keys on the right. Choose fields using the Arrow keys; key in values with the number pad and the Enter key. At the top center are 3 keys: Display Format, Meas Setup, Catalog/system. 3. Select Measurement Setup key; Arrow to Function; soft key choices are below; See the Agilent Impedance Measurement Handbook for definitions of terms Function Cp-D Cs-D Lp-D Ls-D R-X G-B Cp-Q Cs-Q Lp-Q Ls-Q Z-O(deg) Y-O(deg) Cp-G Cs-G Lp-G Ls-G Z-O(rad) Y-O(rad) Cp-Rp Lp-Rp More 1/6 More 2/6 More 3/6 More 4/6 More 5/6 More 6/6 Similarly, arrow to FREQ and select frequency with soft keys; note the top and bottom keys move in decade steps, the smaller arrows move in smaller frequency steps. For example, if you start at 10KHz, the top key will shift to 100KHz and the bottom key to 1KHz; the smaller arrows allow the following frequencies:[10k, 12K, 15K, 20K, 25K, 30K, 40K, 50K, 60K, 80K, 100K]; similar intermediate values are available starting at 1K, 10K, 100K. Next arrow to LEVEL and set the ac voltage; typical values are 25 mv to 50 mv. Arrow to Bias field and enter a DC voltage on the number pad, followed by the enter key; turn on DC voltage with the key (bottom center); red light should turn on. Arrow around the screen and fill in other fields as needed. 4. Connect your sample using the correct micromanipular(s) and/or substrate connection. Do you get a good capacitance reading? Do you see the expected change in C with ±DC voltages? See software section if you want to automate the CV measurements. Page 11

13 C. Temperature controls The chuck temperature is controlled by the large unit to the left of the probe station. For high temperatures (25C<T<300C) above room temperature, there is a heater in the chuck. For low temperatures, a refrigerant is circulated through the chuck. 1. The control panel is on the left side of the temperature control unit. 2. The main power switch is at the bottom center of the panel; to operate, turn on. 3. The top temperature control box is for temperatures above RT. 4. Set the desired temperature using the dials and red readout screen. 5. The current temperature is the green reading. 6. Turn on the power switch for the heater. Wait! System checks conditions before turning on the heater. 7. Chuck temperature will slowly increase to the setpoint. 8. Temperature will arrive at the setpoint and generally oscillates a couple of degrees around the setpoint. Now, run your measurement and collect data. Set a new temperature and repeat. When finished turn the heater off; chuck begins to cool. Cooling rates are slow so you can often collect measurements within a very short time, relative to the cooling rates. 9. For temperatures below RT, there are two set points that need to be controlled. 10. First, set the refrigerant temperature using the setpoint arrow keys; turn on the refrigeration unit. IMPORTANT; this temperature must be below the desired target temperature. 11. Second, set the target temperature on the second temperature control box; turn on cool. 12. If you plan to go very far below room temperature, you will need to place the extra shields above your sample. This setup will prevent moisture from condensing on your sample. Put the shields in place; note the side openings for the micromanipulators. Position the probe needles on your sample. 13. Next, turn on the N2 gas on the back wall and adjust the flow meter. Purging the chamber of water vapor is a slow process. When finished, turn off the N2 gas on the back wall. 14. Measure your sample; the chuck temperature changes are slow. 15. Remove shields etc when finished. Turn off the heater or cooling control unit; turn off the main power switch. Page 12

14 D. Software control of measurements ICS software allows the programming of measurement conditions and data collection and saving both the program and data. A short overview of the software follows 1. Log in to the computer and open the software ICS 2. On the screen are icons and at the top are the typical windows pull down menus. 3. Use the icons and start on the left and continue to the right. 4. Instrument: select either the 4155 or the 4284 and select connect. 5. New: enter a name for your program; select OK; this will become the file name. 6. Device: select the schematic for your device from the menu; schematic shows up with small adjacent boxes for each connection point. 7. Source units: Power supplies are listed, SMU1, SMU2, SMU3, SMU4; highlight the desired power supply, click on the corresponding square on the schematic; continue with other SMU s. 8. Double click on SMU; resulting screen is similar to the front panel; you will need to fill in all the measurement conditions and select the data (V, I) to be measured. 9. Options: select this key and set up the compliance and integration time. 10. MEAS key; opens the measurement controls: single, append, stop, auto seq. Selecting the single key starts the measurement. 11. AUTO Sequence: you can generate several programs and chain them together in an auto sequence. 12. Measure you sample: select single, results are placed in a spreadsheet ; select the spreadsheet key and select the data for each column. 13. Select Graph: This generates a graph of the data in the spreadsheet; auto scales the data. The graph button allows changes to all graph paramters. 14. Save to EXCEL: open Excel in the background, select the columns of data (shift, select); Go to Edit menu and select COPY (note that Ctrl C does not work!); open Excel, paste data into Excel (note that Ctrl V does work!), save your data! 7. Process Data Frequently, the measured data is exported to Excel. Calculations and graphs can be generated within Excel. For example, if you measure both forward and reverse currents for a diode, you can plot the log[abs(i)] versus V. This allows you to plot a wide range of currents with both positive & negative values. Page 13

15 8. Revision History Effective Date Originator DESCRIPTION OF REVISION Issue 7/13/2016 Wayne Paulson Initial Release A B C D E F G H Page 14

MDC CV SYSTEM OPERATION PROCEDURE

MDC CV SYSTEM OPERATION PROCEDURE MDC CV SYSTEM OPERATION PROCEDURE Rev B Table of Contents Contents Table of Contents... 1 1. Purpose / Scope... 2 2. Reference Documents... 2 3. Operation Manual... 2 4. Equipment / Supplies / Material...

More information

Computer. Test Fixture SPA. Key board

Computer. Test Fixture SPA. Key board Semiconductor Parameter Analyzer (Operating instructions) A PDF copy of this document should be available on the Windows Desktop of the SPA computer Read this document fully before attempting to operate

More information

Basics of Keithley Interactive Test Environment

Basics of Keithley Interactive Test Environment Basics of Keithley Interactive Test Environment 0 Launching KITE Power Up and Log On ACTION 1. From Power-up: disconnect DUTs, stay clear of SMU output connectors/probes Log-on: KIUSER (no password) or

More information

Probe station system

Probe station system Probe station system Manufacturer: Micromanipulator Mode: 450PM-HR Descriptions: The Model 450PM 8 probe station offers stable and reliable probing performance. It is based on a design of Micromanipulator

More information

Condensed AFM operating instructions:

Condensed AFM operating instructions: Condensed AFM operating instructions: 1. Log onto system at access controller 2. Take the parts you need to mount a probe out of the drawers. You need the appropriate probe holder, tweezers (these are

More information

FILMETRICS F20 STANDARD OPERATION PROCEDURE

FILMETRICS F20 STANDARD OPERATION PROCEDURE Arizona State University NanoFab FILMETRICS F20 STANDARD OPERATION PROCEDURE Rev C Table of Contents Contents Table of Contents...1 1. Purpose / Scope...2 2. Reference Documents...2 3. Equipment / Supplies

More information

METRICS WIN4145. How Win4145 Stores Information. Project Files. The Win4145 Initialization File. Data and Plot Windows.

METRICS WIN4145. How Win4145 Stores Information. Project Files. The Win4145 Initialization File. Data and Plot Windows. METRICS WIN4145 REFERENCE GUIDE CONTENTS CHAPTER 1: A QUICK TOUR OF WIN4145 How Win4145 Stores Information Project Files The Win4145 Initialization File Data and Plot Windows The Menu Bars The Toolbar

More information

Installation and Getting Started Guide

Installation and Getting Started Guide Installation and Getting Started Guide Metrics ICV Version 4.1.0 Copyright 1999-2015, Metrics Technology, Inc. All rights reserved. Table of Contents Installing the ICV Software... 4 Obtain a Codeword...

More information

TOSHIBA DesignAIRS. Help & Guide Manual

TOSHIBA DesignAIRS. Help & Guide Manual TOSHIBA DesignAIRS Help & Guide Manual Contents 1. MAIN TOOLBAR... 3 1.1. Tools / Options... 3 1.1.1. Application Settings... 3 1.1.2. Email Settings... 4 1.1.3. Internet Connection Settings... 4 1.1.4.

More information

MT9085 Series. ACCESS Master. Quick User s Guide

MT9085 Series. ACCESS Master. Quick User s Guide Quick User s Guide MT9085 Series ACCESS Master For safety and warning information, please read this manual before attempting to use the equipment. Keep this manual with the equipment. MT9085 Series ACCESS

More information

Prism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis

Prism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis Start Up: Upon entering the laser room turn on the wall mounted Laser Power Button by pulling it away from the wall. Turn on Shutter controllers (toggle switch on back of unit). There should be a U in

More information

High-Temperature Susceptometer (Kappabridge)

High-Temperature Susceptometer (Kappabridge) High-Temperature Susceptometer (Kappabridge) [Kappa] Room 262 (Kappa) Sensitivity 4 10 8 SI; Frequency: 920 Hz Temperatures: ambient to 1000 K Several hours/sample ( 200 mg chip or powder) for complete

More information

Suss Aligner SOP. Suss Aligner SOP Page 1 of 6 Revision Scope 1.1 This document provides the procedure for operating the Suss Aligner.

Suss Aligner SOP. Suss Aligner SOP Page 1 of 6 Revision Scope 1.1 This document provides the procedure for operating the Suss Aligner. Suss Aligner SOP Page 1 of 6 Suss Aligner SOP 1. Scope 1.1 This document provides the procedure for operating the Suss Aligner. 2. Table of Contents 1. Scope... 1 2. Table of Contents... 1 Suss Aligner

More information

OPERATING INSTRUCTION

OPERATING INSTRUCTION OPERATING INSTRUCTION AUTORANGING MULTIMETER MAX Ω F C 10A MAX every 15 min. COM V SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation

More information

Standard Operating Procedure Inhouse Solar Simulator

Standard Operating Procedure Inhouse Solar Simulator Standard Operating Procedure Inhouse Solar Simulator By Sanchit Khatavkar and Guru Burkul National Centre for Photovoltaic Research and Education (NCPRE), Indian Institute of Technology-Bombay, Powai,

More information

MPI TS150-HP 150 mm High Power Manual Probe System For accurate High Power measurements up to 10 kv, 600 A

MPI TS150-HP 150 mm High Power Manual Probe System For accurate High Power measurements up to 10 kv, 600 A MPI TS15-HP 15 mm High Power Manual Probe System For accurate High Power measurements up to 1 kv, 6 A FEATURES / BENEFITS Universal Use Designed specifically for high power device measurement and wide

More information

Standard Operating Procedure For The Kaye Validator 2000 Thermal Mapping System

Standard Operating Procedure For The Kaye Validator 2000 Thermal Mapping System Standard Operating For The Kaye Validator 2000 Thermal Mapping System SOP No: V01/SOP/505/15 Version No: 01 No of Copies issued: 01 Copy Number: 01 PharmaTherm Limited Circulation: 1. Mr Andrew Varley,

More information

Metrics Tutorial. Table of Contents. 1. Introduction

Metrics Tutorial. Table of Contents. 1. Introduction Metrics Tutorial Table of Contents 1. Introduction 2. The Set-Up Editor 3. The Project File Manager 4. Data Windows 5. Plot Windows 6. Exporting and Importing Data 7. Linking Metrics and Excel 1. Introduction

More information

MPI TS2500-RF 200 mm Fully Automated Probe System For RF Production Test Measurements

MPI TS2500-RF 200 mm Fully Automated Probe System For RF Production Test Measurements MPI TS2500-RF 200 mm Fully Automated Probe System For RF Production Test Measurements FEATURES / BENEFITS Designed for Wide Variety of RF On-Wafer Production Applications RF applications up to 67 GHz &

More information

P6 Profilometer Operating Procedure

P6 Profilometer Operating Procedure P6 Profilometer Operating Procedure I. Introduction The KLA-Tencor P-6 profiler is a highly sensitive surface profiler that measures step height. The key features of P-6 include: A vertical range of 326

More information

User Guide. Video Borescope. Models BR200 and BR250

User Guide. Video Borescope. Models BR200 and BR250 User Guide Video Borescope Models BR200 and BR250 Introduction Congratulations on your purchase of this Extech BR200 (17mm camera head) or BR250 (9mm camera head) Video Borescope. This instrument was designed

More information

MPI TS150-AIT 150 mm Manual Probe System

MPI TS150-AIT 150 mm Manual Probe System MPI TS150-AIT 150 mm Manual Probe System Industry s first explicitly designed 150 mm probe system providing accurate tests for mm-wave, THz, and automated impedance tuner applications FEATURES / BENEFITS

More information

Equipment Standard Operating Procedure Kimberly Appel

Equipment Standard Operating Procedure Kimberly Appel Date Created: June 26, 2004 Date Modified: The Flexus 2320 Equipment Standard Operating Procedure Kimberly Appel 1. Purpose 1.1. The Flexus Thin Film Stress Measuring Apparatus (TFSMA) measures the changes

More information

` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE

` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE ` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE Version: 1.0 April-2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research Center (NRC) TABLE OF CONTENTS 1. Introduction...2 1.1

More information

Autoranging True RMS Multimeter User Manual

Autoranging True RMS Multimeter User Manual Autoranging True RMS Multimeter User Manual Please read this manual before switching the unit on. Important safety information inside. Contents Page 1. Safety Information... 4 2. Safety Symbols... 5 3.

More information

Oxford ICP 2-step DRIE SOP

Oxford ICP 2-step DRIE SOP Oxford ICP 2-step DRIE SOP Oxford ICP 2-step DRIE SOP Page 1 of 13 1. Scope 1.1 This document provides operating procedures for the Oxford 100 ICP 2-step DRIE. 2. Table of Contents 1. Scope... 1 2. Table

More information

iq DIGITAL PICTURE FRAME iq Digital Picture Frame DPF701SB USER GUIDE

iq DIGITAL PICTURE FRAME iq Digital Picture Frame DPF701SB USER GUIDE iq Digital Picture Frame DPF701SB USER GUIDE 1 Table of Contents Important Safety Precautions... 3 Cleaning the LCD Screen... 3 Cleaning the Digital Picture Frame... 3 Introduction... 4 What s in the Box...

More information

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter S-14 Compact Digital Multimeter S-14 Compact Digital Multimeter SAFETY INFORMATION The following safety information must be observed to insure maximum personal safety during the operation at this meter

More information

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER SAFETY INFORMATION The following safety information must be observed to ensure maximum personal safety during the operation of this meter: This meter is

More information

LHe/LN2 Cryogenic Probe Station

LHe/LN2 Cryogenic Probe Station This model of the ARS Probe Station is designed for the ultimate in flexibility for non-destructive device testing. The ability to upgrade and modify this system for further device testing has been in

More information

Heidelberg Pattern Generator SOP

Heidelberg Pattern Generator SOP Heidelberg Pattern Generator SOP Page 1 of 15 Heidelberg Pattern Generator SOP 1. Scope 1.1 This document provides the operating procedures for the Heidelberg Pattern Generator with Version 3.12.5 software.

More information

MPI TS150-THZ 150 mm Manual Probe System

MPI TS150-THZ 150 mm Manual Probe System MPI TS150-THZ 150 mm Manual Probe System Industry s first explicitly designed probe system for accurate measurements at mm-wave and sub-mm wave (THz) frequency range FEATURES / BENEFITS Variety of Applications

More information

Chapter 1: A Quick Tour of ICS Chapter The Setup Editor 26

Chapter 1: A Quick Tour of ICS Chapter The Setup Editor 26 Chapter 1: A Quick Tour of ICS 5 What is ICS SOFTWARE? How ICS Stores Information Test Setups The Test Setup/Project File Relationship Test Setup Applications The Setup Editor Project Files Creating Project

More information

User's Guide. Video Borescope. Models BR200 and BR250

User's Guide. Video Borescope. Models BR200 and BR250 User's Guide Video Borescope Models BR200 and BR250 Introduction Congratulations on your purchase of this Extech BR200 (17mm camera head) or BR250 (9mm camera head) Video Borescope. This instrument was

More information

Introduction. Chapter 2

Introduction. Chapter 2 The Vantage PRO unit (Figure 2-1) combines a digital and graphing multimeter, lab scope and ignition scope with a powerful diagnostic database. Figure 2-1 Vantage PRO This diagnostic database gives you

More information

Olympus DeltaVision Microscope Start-Up and Shut-Down Instructions.

Olympus DeltaVision Microscope Start-Up and Shut-Down Instructions. DeltaVision Instructions: 1. Start-up (Olympus DV) 2. Basic operation (Olympus DV) 2.1 set-up 2.2 designing and running an experiment 3. Transferring files via FTP 4. Deconvolving files 5. File conversion

More information

Navigator Software User s Manual. User Manual. Navigator Software. Monarch Instrument Rev 0.98 May Page 1 of 17

Navigator Software User s Manual. User Manual. Navigator Software. Monarch Instrument Rev 0.98 May Page 1 of 17 User Manual Navigator Software Monarch Instrument Rev 0.98 May 2006 Page 1 of 17 Contents 1. NAVIGATOR SOFTWARE 2. INSTALLATION 3. USING NAVIGATOR SOFTWARE 3.1 STARTING THE PROGRAM 3.2 SYSTEM SET UP 3.3

More information

4.2 NORMAL OPERATION NORMAL START UP PROCEDURE

4.2 NORMAL OPERATION NORMAL START UP PROCEDURE 4.2 NORMAL OPERATION 4.2.1 NORMAL START UP PROCEDURE Kimball Physics recommends that the gun be run only in a vacuum of 10-5 torr or better for an alkali metal ion source. Poorer vacuum may result in source

More information

1. Scope 1.1 This document provides the operating procedure for plasma enhanced chemical vapor deposition with the Oxford 80+.

1. Scope 1.1 This document provides the operating procedure for plasma enhanced chemical vapor deposition with the Oxford 80+. Oxford 80+ PECVD SOP Oxford 80+ PECVD SOP Page 1 of 7 1. Scope 1.1 This document provides the operating procedure for plasma enhanced chemical vapor deposition with the Oxford 80+. 2. Table of Contents

More information

Inspection Camera. User Manual

Inspection Camera. User Manual ic TM Inspection Camera User Manual Overview Contents Product overview...4 Accessories...5 Driver installation...6 Operation Live mode...9 Browse mode...11 Setting mode...13...14 Cautions for Use...15

More information

Installation and Getting Started Guide

Installation and Getting Started Guide Installation and Getting Started Guide Metrics Win4156 Version 4.1.0 Copyright 1999-2015, Metrics Technology, Inc. All rights reserved. Table of Contents Installing the Win4156 Software... 4 Obtain a Codeword...

More information

UNIVERSITY OF CENTRAL FLORIDA DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE HEC 406 NANO/MEMS LABORATORY

UNIVERSITY OF CENTRAL FLORIDA DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE HEC 406 NANO/MEMS LABORATORY UNIVERSITY OF CENTRAL FLORIDA DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE HEC 406 NANO/MEMS LABORATORY HIGH TEMPERATURE & HIGH FREQUENCY TESTING SYSTEM LABORATORY USER MANUAL 4/29/2011 Overview

More information

Installation and Getting Started Guide

Installation and Getting Started Guide Installation and Getting Started Guide Metrics Win4145 Version 4.2 Copyright 1999-2017, Metrics Technology, Inc. All rights reserved. Table of Contents Installing the Win4145 Software... 4 Obtain a Codeword...

More information

Curve Tracing Systems

Curve Tracing Systems Curve Tracing Systems Models Available MultiTrace: The most flexible solution for devices up to 625 pins, capable of any of the applications described here. Comes with a PGA-625 fixture MegaTrace: A larger

More information

#61-844SW ThermalVision Software Instruction Manual

#61-844SW ThermalVision Software Instruction Manual ThermalVision Software Instruction Manual ND-7035-1 Page 1 of 23 Contents 1. Installing the ThermalVision software onto a PC... 3 2. Transferring saved images from the camera to the PC.... 3 2.1. Direct

More information

Title: Karl Suss MA150 Semiconductor & Microsystems Fabrication Laboratory Revision: E Rev Date: 03/01/06

Title: Karl Suss MA150 Semiconductor & Microsystems Fabrication Laboratory Revision: E Rev Date: 03/01/06 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Karl Suss MA150 Mask Aligner. All users are expected to have read and understood

More information

Tutorial 3: Using the Waveform Viewer Introduces the basics of using the waveform viewer. Read Tutorial SIMPLIS Tutorials SIMPLIS provide a range of t

Tutorial 3: Using the Waveform Viewer Introduces the basics of using the waveform viewer. Read Tutorial SIMPLIS Tutorials SIMPLIS provide a range of t Tutorials Introductory Tutorials These tutorials are designed to give new users a basic understanding of how to use SIMetrix and SIMetrix/SIMPLIS. Tutorial 1: Getting Started Guides you through getting

More information

University of Minnesota Nano Fabrication Center Standard Operating Procedure

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name: University of Minnesota Nano Fabrication Center Coral Name: hs-scope Revision Number: 1.5 Model: HS200A Revisionist: M. Fisher Location: Bay 1 Date: 9/12/2013 1 Description The Hyphenated

More information

COMPACT PRECISION LINEAR MOTORIZED ACTUATORS LSMA Series LSMA-173

COMPACT PRECISION LINEAR MOTORIZED ACTUATORS LSMA Series LSMA-173 COMPACT PRECISION LINEAR MOTORIZED ACTUATORS LSMA Series Narrow Width ized Translation Stages Compact (30 mm) design Precision bearing system Resolution 0.1 microns Standard stroke 20 and 50 mm Optional

More information

Simulator System SUN ABET

Simulator System SUN ABET SoP of I V Simulator System Part 2 SUN 3000 SOLAR SIMULATOR ABET Technologies Standard operating procedure: Flowchart of sequence to be followed: System switch ON sequence Stage and device probing Measuring

More information

Primary Use. Operating Principle

Primary Use. Operating Principle Primary Use The Leica DVM6 is an optical microscope that has the ability observe samples at a high magnification at a high resolution. The microscope allows users to view their sample with up to a 2350x

More information

First Steps with ADS and Coax Modeling

First Steps with ADS and Coax Modeling First Steps with ADS and Coax Modeling ADS Session 1 Start ADS and Create an Empty Project Look for a desktop icon or start menu item entitled Advanced Design System 2009 ADS will start up and you will

More information

Thermal Transient Test Installation and Operating Manual

Thermal Transient Test Installation and Operating Manual Thermal Transient Test Installation and Operating Manual 2705A De La Vina Street Santa Barbara, California 93105 Telephone (805) 682-0900 descon@silcom.com www. santabarbaraautomation.com Installation

More information

FLS M9.06. ph/orp MONITOR

FLS M9.06. ph/orp MONITOR FLS M9.06 /ORP MONITOR Safety Instructions General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other instruments

More information

Visual Physics - Introductory Lab Lab 0

Visual Physics - Introductory Lab Lab 0 Your Introductory Lab will guide you through the steps necessary to utilize state-of-the-art technology to acquire and graph data of mechanics experiments. Throughout Visual Physics, you will be using

More information

DPL4000 Portable Low Range Dew Point Analyzer

DPL4000 Portable Low Range Dew Point Analyzer OPERATIONS MANUAL DPL4000 Portable Low Range Dew Point Analyzer 7205 Edington Drive / Cincinnati, OH 45249 / Tel (513) 772-0060 / Fax (513) 772-9466 Page #1 of 16 M4581 DPL4000 Product Description: This

More information

MASTERLINE Ultimate Hybrid High Voltage 4/12 (MLUH HV 4/12) blunt cut at BTS side

MASTERLINE Ultimate Hybrid High Voltage 4/12 (MLUH HV 4/12) blunt cut at BTS side INSTALLATION MANUAL HUBER+SUHNER AG Fiber Optics MASTERLINE Ultimate Hybrid High Voltage 4/12 (MLU HV 4/12) blunt cut at BTS side DOC-0000803885 Rev A June 18, 2018 Page 1 of 12 (MLUH HV 4/12) blunt cut

More information

N2KExtractor. Maretron Data Extraction Software User s Manual

N2KExtractor. Maretron Data Extraction Software User s Manual N2KExtractor Maretron Data Extraction Software User s Manual Revision 3.1.6 Copyright 2017 Maretron, LLP All Rights Reserved Maretron, LLP 9014 N. 23rd Ave #10 Phoenix, AZ 85021-7850 http://www.maretron.com

More information

Keithley Instruments Customer Documentation Series 2600A UL Safety Supplement

Keithley Instruments Customer Documentation Series 2600A UL Safety Supplement Keithley Instruments, Inc. 28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY www.keithley.com Keithley Instruments Customer Documentation Series 2600A UL Safety Supplement 1. Introduction The Keithley

More information

Operating Procedure for Horiba Raman Microscope

Operating Procedure for Horiba Raman Microscope Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch

More information

FLS M9.06 PH/ORP MONITOR. safety instructions. packing List

FLS M9.06 PH/ORP MONITOR. safety instructions. packing List FLS M9.06 PH/ORP MONITOR safety instructions General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other instruments

More information

INSTALLATION AND OPERATIONAL INSTRUCTIONS FOR THE OFITE MODEL 1000 VERSION 1.0

INSTALLATION AND OPERATIONAL INSTRUCTIONS FOR THE OFITE MODEL 1000 VERSION 1.0 OFI Testing Equipment Model 1000 HPHT Viscometer (130-75) Instructions Page 1 of 7 INSTALLATION AND OPERATIONAL INSTRUCTIONS FOR THE OFITE MODEL 1000 VERSION 1.0 INSTALL DIRECTIONS FOR OFITE SOFTWARE.

More information

User Manual. Trinocular Stereo Microscope

User Manual. Trinocular Stereo Microscope User Manual Trinocular Stereo Microscope Model V434 XV434 Series MicroscopeNet.com Table of Contents i. Caution.. 1 ii. Care and Maintenance... 2 1. Component Illustration... 3 2. Installation 4 3. Operation....6

More information

Microscopic Imaging Research Station (MIRS) Assembly Guide. Version 1.0.0

Microscopic Imaging Research Station (MIRS) Assembly Guide. Version 1.0.0 Microscopic Imaging Research Station (MIRS) Assembly Guide www.adsyscontrols.com Adsys Controls, Inc.2012 Version 1.0.0 I. Assembly of the Adsys Controls MIRS system This document explains the assembly

More information

BS 287 DUAL CHANNEL POWER SUPPLY. User Manual. January 2017 V1.0

BS 287 DUAL CHANNEL POWER SUPPLY. User Manual. January 2017 V1.0 BS 287 DUAL CHANNEL POWER SUPPLY User Manual January 2017 V1.0 Table of contents 1.0 SAFETY INSTRUCTIONS... 3 2.0 GENERAL DESCRIPTION PS 289... 4 3.0 MECHANICAL INSTALLATION... 5 4.0 MAINS POWER & SAFETY

More information

minipcr Mobile App operation (Android devices) Appendix: maintenance and operating conditions

minipcr Mobile App operation (Android devices) Appendix: maintenance and operating conditions minipcr TM Thermal Cycler User s Guide Contents I. Loading your samples II. Setting up your minipcr TM thermal cycler III. Creating, storing, and editing protocols IV. Running and monitoring protocols

More information

Equipment Operating Procedure. Model Number: Dage-Series-4000PA Machine Name: Shear Tester Date: 8/17/2013 Serial Number:

Equipment Operating Procedure. Model Number: Dage-Series-4000PA Machine Name: Shear Tester Date: 8/17/2013 Serial Number: Equipment Operating Procedure Manufacturer: Dage Precision Location: PRC Assembly Laboratory Industries Model Number: Dage-Series-4000PA Machine Name: Shear Tester Date: 8/17/2013 Serial Number: 1922708

More information

Getting Started: Creating and Executing a Test Setup. Step 1: Cable the Hardware Connections. Step 2: Connect the HP4145 Instrument Driver

Getting Started: Creating and Executing a Test Setup. Step 1: Cable the Hardware Connections. Step 2: Connect the HP4145 Instrument Driver METRICS TECHNOLOGY WIN4145 GETTING STARTED GUIDE CONTENTS Getting Started: Creating and Executing a Test Setup Step 1: Cable the Hardware Connections Step 2: Connect the HP4145 Instrument Driver Step 3:

More information

BPL SERIES INSTALLATION INSTRUCTIONS THIS SHEET CONTAINS IMPORTANT SAFETY INSTRUCTIONS. SAVE THESE INSTRUCTIONS.

BPL SERIES INSTALLATION INSTRUCTIONS THIS SHEET CONTAINS IMPORTANT SAFETY INSTRUCTIONS. SAVE THESE INSTRUCTIONS. BPL SERIES INSTALLATION INSTRUCTIONS Important Warning THIS SHEET CONTAINS IMPORTANT SAFETY INSTRUCTIONS. SAVE THESE INSTRUCTIONS. This product must be installed in accordance with National Electrical

More information

3700 SERIES USER MANUAL

3700 SERIES USER MANUAL SAFETY GUIDE This manual contains the precautions necessary to ensure your personal safety as well as for protection for the products and the connected equipment. These precautions are highlighted with

More information

Removal and Installation8

Removal and Installation8 8 Screw Types 8-4 Top Cover Assembly 8-5 Left Hand Cover 8-6 Right Hand Cover 8-10 Front Panel Assembly 8-14 Left Rear Cover 8-15 Right Rear Cover 8-16 Extension Cover (60" Model only) 8-17 Media Lever

More information

User Manual. Binocular Zoom Stereo Microscope with Boom Stand. MicroscopeNet.com

User Manual. Binocular Zoom Stereo Microscope with Boom Stand. MicroscopeNet.com User Manual Binocular Stereo Microscope with Boom Stand Model W42C1 Series MicroscopeNet.com Table of Contents i. Caution.. 1 ii. Care and Maintenance... 1 1. Component Illustration... 2 2. Installation...3

More information

VEECO FPP Point Probe Operating Manual

VEECO FPP Point Probe Operating Manual VEECO FPP-5000 4-Point Probe Operating Manual Version: 1.0 May 2013 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research and Teaching Facility TABLE OF CONTENTS 1. Introduction....2 1.1 Scope of Work....2

More information

Configuring a Precision System for On-Wafer Capacitance

Configuring a Precision System for On-Wafer Capacitance Application Note Innovating Test Technologies Configuring a Precision System for On-Wafer Capacitance Development of new aggressively scaled MOS processes and devices depends on high accuracy CV measurements.

More information

FluoroSELECT Fluorometer User s Manual

FluoroSELECT Fluorometer User s Manual FluoroSELECT Fluorometer User s Manual Version 4.4.A Sigma-Aldrich Chemie GmbH Industriestrasse 25 CH-9470 Buchs Switzerland EurTechServ@sial.com techservice@sial.com (US and Canada) sigma-aldrich.com

More information

LAB EXERCISE 3 Harmonic Balance Techniques

LAB EXERCISE 3 Harmonic Balance Techniques ADS 2012 Circuit Simulation Techniques (v.2 - April 2013) LAB EXERCISE 3 Harmonic Balance Techniques Topics: Harmonic Balance simulations are covered including swept variables, 2-tone simulations, equations

More information

FLOW MONITOR & TRANSMITTER

FLOW MONITOR & TRANSMITTER FLS M9.02 FLOW MONITOR & TRANSMITTER SAFETY INSTRUCTIONS General Statements Do not install and service the product without following the Instruction Manual. This item is designed to be connected to other

More information

ABM's High Resolution Mask Aligner Features:

ABM's High Resolution Mask Aligner Features: ABM's High Resolution Mask Aligner is a very versatile instrument with interchangeable light sources which allow Near-UV (405-365 nm) as well as Mid- and Deep-UV (254 nm, 220 nm) exposures in proximity

More information

HP Pavilion dv7-6c90us Cooling fan Replacement

HP Pavilion dv7-6c90us Cooling fan Replacement HP Pavilion dv7-6c90us Cooling fan Replacement This guide will walk you through the process of replacing the cooling fan in an HP Pavilion dv7 laptop. Written By: Angelina Clayton ifixit CC BY-NC-SA www.ifixit.com

More information

HP 4145A & HP 4284A with Micromanipulator Probe Station model 6000 Electrical Test Station Standard Operating Procedure (SOP)

HP 4145A & HP 4284A with Micromanipulator Probe Station model 6000 Electrical Test Station Standard Operating Procedure (SOP) DOCUMENT: Teaching Fab Electrical Test Station Standard Operating Procedure (SOP) Version: 1 HP 4145A & HP 4284A with Micromanipulator Probe Station model 6000 Electrical Test Station Standard Operating

More information

Wireless Circuits and Systems Laboratory

Wireless Circuits and Systems Laboratory Wireless Circuits and Systems Laboratory Procedure #9 8714 VNA Data Acquisition and Manipulation (using Microsoft Excel) Contained in this procedure: 1) Acquiring 1- and 2-port data from the HP8714 and

More information

XPS 15 2-in-1. Service Manual. Computer Model: XPS Regulatory Model: P73F Regulatory Type: P73F001

XPS 15 2-in-1. Service Manual. Computer Model: XPS Regulatory Model: P73F Regulatory Type: P73F001 XPS 15 2-in-1 Service Manual Computer Model: XPS 15-9575 Regulatory Model: P73F Regulatory Type: P73F001 Notes, cautions, and warnings NOTE: A NOTE indicates important information that helps you make better

More information

XRADIA microxct Manual

XRADIA microxct Manual XRADIA microxct Manual Multiscale CT Lab Table of Contents 1. Introduction and Basics 1.1 Instrument Parts 1.2 Powering up the system 1.3 Preparing your sample 2. TXM Controller 2.1 Starting up 2.2 Finding

More information

Pulsed Frequency TM. EM272/EM272A Software Download, Installation and User Guide

Pulsed Frequency TM. EM272/EM272A Software Download, Installation and User Guide Pulsed Frequency TM EM272/EM272A Software Download, Installation and User Guide Account Registration... 3 Create an Account... 4 Download the Software... 6 Install the Software... 7 Connect EM272 / EM272A

More information

Co-Lase Users Guide. DM00001 Rev A (DCN 031) These instructions are protected by copyright. All rights reserved.

Co-Lase Users Guide. DM00001 Rev A (DCN 031) These instructions are protected by copyright. All rights reserved. Co-Lase Users Guide These instructions are protected by copyright. All rights reserved., Inc. http://www.propel-labs.com 131 East Lincoln Avenue, Suite 200 Fort Collins, CO 80524 1 Table of Contents Chapter

More information

Elecraft K3 KPA3 Power Connector Replacement Revision B, June 30, 2017 Copyright 2017, Elecraft, Inc. All Rights Reserved

Elecraft K3 KPA3 Power Connector Replacement Revision B, June 30, 2017 Copyright 2017, Elecraft, Inc. All Rights Reserved Introduction Elecraft K3 KPA3 Power Connector Replacement Revision B, June 30, 2017 Copyright 2017, Elecraft, Inc. All Rights Reserved The connectors furnishing high current to the KPA3 module have failed

More information

Inspiron Service Manual. 2-in-1. Computer Model: Inspiron Regulatory Model: P69G Regulatory Type: P69G001

Inspiron Service Manual. 2-in-1. Computer Model: Inspiron Regulatory Model: P69G Regulatory Type: P69G001 Inspiron 13 5000 2-in-1 Service Manual Computer Model: Inspiron 13-5378 Regulatory Model: P69G Regulatory Type: P69G001 Notes, cautions, and warnings NOTE: A NOTE indicates important information that helps

More information

FC 1800 #1 Operation Manual University of Notre Dame

FC 1800 #1 Operation Manual University of Notre Dame \ FC 1800 #1 Operation Manual University of Notre Dame For Training Please contact: Dave Heemstra 8 June 2015 Page 1 of 12 General Precautions Contacts For problems, clarification of procedures, or general

More information

TMA Standard Operating Procedure [Updated April 30, 2015]

TMA Standard Operating Procedure [Updated April 30, 2015] Thermomechanical Analysis Q400 Location of Machine: Composites Lab, RFM 1218 Location of SOP and Machine Operating & Safety Manual: Composites Lab website under resources; Composites Lab TRACS site; and

More information

Digital Proximity System Software Installation Manual

Digital Proximity System Software Installation Manual Digital Proximity System Software Installation Manual Doc# 100961 REV A (February 2018) Table of Contents Purpose...2 How to Download DPS 1.35 Software from the Internet...3 The difference between Simulate

More information

Karl Suss SB6 Bonder Operation Tool Id: Rev 01, 09/24/10

Karl Suss SB6 Bonder Operation Tool Id: Rev 01, 09/24/10 Karl Suss SB6 Bonder Operation Tool Id: 119.020 The technical content of this document is the responsibility of UCLA CNSI Integrated Systems Nanofabrication Cleanroom Los Angeles, CA Karl Suss SB6 Bonder

More information

Amphenol PROTIMETER. Advanced Sensors. Digital Mini. Protimeter Moisture Meter. Instruction Manual

Amphenol PROTIMETER. Advanced Sensors. Digital Mini. Protimeter Moisture Meter. Instruction Manual PROTIMETER 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Digital Mini Protimeter Moisture Meter Digital Mini %WME DRY Pro meter

More information

Optical encoding film disc mini board with integrated infrared photo sensor and LED operating indicator round, 5-pin output socket (180, DIN 41524)

Optical encoding film disc mini board with integrated infrared photo sensor and LED operating indicator round, 5-pin output socket (180, DIN 41524) Quick reference 200 pulses per revolution of measuring drum For use with TG05 to TG50 and BG4 to BG100 Uni-directional Not applicable for ex-proof areas Application The Pulse Generator for RITTER Drum-type

More information

How Computer Mice Work

How Computer Mice Work How Computer Mice Work Inside this Article 1. Introduction to How Computer Mice Work 2. Evolution of the Computer Mouse 3. Inside a Mouse 4. Connecting Computer Mice 5. Optical Mice 6. Optical Mouse Accuracy

More information

1. Review the printer safety information in the Regulatory Compliance document provided with your printer.

1. Review the printer safety information in the Regulatory Compliance document provided with your printer. QUICK REFERENCE This Quick Reference contains supply loading information and general care and maintenance procedures for the Monarch Pathfinder Ultra Platinum 6039 printer. For more detailed information,

More information

DIO-0808RN-USB. N Series for USB Isolated Digital I/O Unit(8ch Relay Output, 8ch DI) DIO-0808RN-USB 1. Ver.1.02

DIO-0808RN-USB. N Series for USB Isolated Digital I/O Unit(8ch Relay Output, 8ch DI) DIO-0808RN-USB 1. Ver.1.02 N Series for USB Isolated Digital I/O Unit(8ch Relay Output, 8ch DI) DIO-0808RN-USB *The above photograph is DIO-1616HN-USB. * Specifications, color and design of the products are subject to change without

More information

Handout Objectives: a. b. c. d. 3. a. b. c. d. e a. b. 6. a. b. c. d. Overview:

Handout Objectives: a. b. c. d. 3. a. b. c. d. e a. b. 6. a. b. c. d. Overview: Computer Basics I Handout Objectives: 1. Control program windows and menus. 2. Graphical user interface (GUI) a. Desktop b. Manage Windows c. Recycle Bin d. Creating a New Folder 3. Control Panel. a. Appearance

More information

ViewConn Pro VC-8200

ViewConn Pro VC-8200 ViewConn Pro VC-8200 1. Introducing ViewConn 2. Using ViewConn Pro 3. Optional Integrated Optical Power Meter 4. Specifications Safety Fiber optic lasers use light wavelengths not visible to the human

More information

Inspiron 22. Service Manual Series. Regulatory Model: W17B Regulatory Type: W17B001

Inspiron 22. Service Manual Series. Regulatory Model: W17B Regulatory Type: W17B001 Inspiron 22 3000 Series Service Manual Regulatory Model: W17B Regulatory Type: W17B001 Notes, cautions, and warnings NOTE: A NOTE indicates important information that helps you make better use of your

More information

DeskCNC setup and operation manual

DeskCNC setup and operation manual DeskCNC setup and operation manual This document explains how to install, setup and cut foam shapes using DeskCNC 4 axis foam cutting software. The document will go through a step by step process of how

More information