Collection of Work Projects

Size: px
Start display at page:

Download "Collection of Work Projects"

Transcription

1 EUTEMPE-RX: European Training and Education for Medical Physics Experts in Radiology MPE05: Anthropomorphic phantoms Collection of Work Projects Varna, September

2 2

3 Introduction: The present booklet is created from the report of the trainees in the MPE05 module of the pilot running of the EUTEMPE-RX course. The content of the reports is not changed. The reports have been reformatted to meet a uniform style for the whole collection. One or two slides used by the trainees during their oral presentations are also added to each work. 3

4 4

5 Table of Contents PROJECT ASSIGNMENT No PROJECT ASSIGNMENT No PROJECT ASSIGNMENT No PROJECT ASSIGNMENT No PROJECT ASSIGNMENT No PROJECT ASSIGNMENT No PROJECT ASSIGNMENT No PROJECT ASSIGNMENT No

6 6

7 EUTEMPE-RX: European Training and Education for Medical Physics Experts in Radiology MPE05: Anthropomorphic phantoms Applications of anthropomorphic phantoms for design and evaluation of advanced x-ray imaging techniques PROJECT ASSIGNMENT No. 1 Using computational breast phantom, perform a virtual study to determine the potential of breast tomosynthesis for detectability of breast abnormalities, compared to conventional mammography To be considered: Involved Phantom(s): Breast Phantom Small to Medium size, Glandular to Dense background tissue, Tree min 15 lactiferous branches in total, include skin and Cooper, Output volume matrix size 400 x 400 x 400 voxels (or alternatively 500 x 500 x 500 voxels), voxel size = 0.25 mm. Abnormalities: One spherical mass abnormality up to 10 mm in diameter and one cluster of microcalcifications with a microcalcification of up to 1 mm in diameter. Imaging techniques: Conventional mammography and Breast tomosynthesis. Assumed Incident Dose: 5-12 mgy. Any other parameters are by your discretion and preference. To be completed: Create a phantom, according to the provided specification and add the abnormalities. Obtain synthetic conventional mammographic image of the phantom at an energy in the range kev, selected by your discretion. Obtain 10 synthetic mammography images over an isocentric arc (-9 to 9) degree (every 2 degrees) and 26 synthetic mammography images over an isocentric arc (-25 to 25) degree (every 2 degree) at the same energy as above, keeping the total dose the same. Add a realistic level of noise, based on the required incident dose. Perform image reconstruction at different planes of interest. Compare subjectively (visually) the images. 7

8 Compute objective figure of merits (FOMs) CNR in the region of the abnormalities and compare. To be prepared: Report: a short written summary of the work and the obtained results; containing for example: o appropriate outcome images and profiles o tables with computed FOMs o discussion of the results and conclusions PowerPoint Presentation: to be used for a short (5-10 ) oral presentation. Useful hints: The tool for the phantom creation is the BreastSimulator tool. For introducing abnormalities use the BreastSimulator tool. Projection images are generated using the BreastSimulator tool. Image reconstruction is performed using FDKR tool. References: [1]. BreastSimulator User Guide. [2]. FDKR User Guide. 8

9 PROJECT IMPLEMENTATION REPORT A virtual study to determine the potential of breast tomosynthesis for detectability of breast abnormalities, compared to conventional mammography Şölen Çubukçu(Turkey) ; Agnieszka Kuchcinska (Poland) ; Nicola Pace (Italy) Introduction Virtual anthropomorphic phantoms can be used in order to assess the diagnostic possibilities of different imaging techniques. In particular, breast virtual phantoms containing abnormalities may be employed in a comparative evaluation of mammography and tomosynthesis. Materials and Methods We developed an anthropomorphic phantom to simulate a small dense breast using BreastSimulator software. Two abnormalities were inserted within the simulated breast: 1 spherical abnormal mass simulating breast cancer. Radius = 5mm 1 cluster (of radius = 10 mm) containing 5 microcalcifications each max microcalcification radius = 0.5 mm. Data used for the development of the model are displayed in table 1 and table 2. Simulation was performed considering incident photon energy of 20 kev. Two breast phantoms were simulated with same parameters except attenuation coefficient factor. Breast size Nipple size Total Ducts Cooper ligaments (radius) Matrix size Voxel size Nwalks Nsteps 40x40x40 mm 4x4x5mm 16 (4x4) (2mm) 400x400x Table 1: input parameters for the used breast model Element Attenuation 20keV (mm -1 ) Semiellypsoid, hyperboloid, nipple, ducts, cooper elements, skin microcalcifications Abnormal mass (phantom 1 )/ 0.90 (Phantom 2 ) Table 2: attenuation coefficients assumed 9

10 Correct localization of both abnormalities and microcalcification was based on generated model. Figures below show phantom visualization. Table 3 show example of localization for cluster and abnormality for phantom 2. µcaco3 1 µcaco3 2 µcaco3 3 µcaco3 4 µcaco3 5 mass X [mm] 11,9 14,42 12,62 13,63 12,15-3 Y [mm] 12,59 14,22 10,83 15,03 12,11-11 Z [mm] 9,26 3,64 3,93-2,97 6,21-12 Table 3: localization of abnormalities in phantom 2 Fıg 1: skin and duct system phantom 2 Fıg 2: Phantom 2 yellow color represent abnormality, green color Cooper ligaments Fıg 3: Phantom 2 basic visualisation After the generation of the simulated breast, XrayImagingSimulator was used in order to generate a synthetic X-ray image obtained through conventional mammography (single CC projection), a tomosynthesis reconstructed synthetic image generated from 10 planar acquisition ranging in angles between -9 and 9 degrees, and a tomosynthesis reconstructed synthetic image generated from 26 planar acquisitions ranging in angles between -25 and 25 degrees with step 2 degree. Reconstruction was made using FDKR Software. Images obtained were analyzed with ImageJ software in order to investigate differences in diagnostic capabilities for abnormal masses of the different radiographic techniques considered. Major outcome for comparison was the figure of merit (FOM) CNR in the region of the abnormalities. Other simulations accounting for different abnormality attenuation coefficient, different photon energy and different breast density were made in order to explore the variability of results. Results We obtained 3 images for phantom 1 simulating planar mammography (fig 4) and tomosynthesis acquisition (fig 5 and 6). Abnormal mass was well displayed in all the three images (fig 4 to 6). 10

11 Inter-reader variability was assessed, with an agreement on detectability of abnormalities of 100% across the three Readers on all the three images. For phantom 2 detail tomosynthesis analysis was performed and very accurate localization of micro calcification was obtained. For 2D mammography FOM was also calculated for other suspicious part of the image in order to simulate false positive investigation for dense breast. For phantom 2 other region has higher FOM than real abnormal mass. In order to check possible improvement of system performance for dense breast and low difference of attenuation factor between abnormality and background Dual energy image simulation was done. Agreement between simulated Phantoms background was find, for both phantoms mean pixel value was 1, 46. ROI analysis yielded the results listed in tables 3, and figure D mammography -9 to 9 tomosynthesis -25 to 25 tomosynthesis ROI mass mean ROI mass stdev ROI bkgd mean ROI bkgd stdev CNR Table 3: ROI Analysis for Phantom 1 Fıg 4: synthetic X-ray image of phantom 1 obtained through conventional mammography Fıg 5: tomosynthesis reconstructed synthetic image of phantom 1 generated from 10 planar acquisition ranging in angles between -9 and 9 degrees Fıg 6: tomosynthesis reconstructed synthetic image of phantom 1 generated from 26 planar acquisitions ranging in angles between -25 and 25 degrees 11

12 Fıg 7: detaıl of the abnormal mass dısplayed by 2D mammogram (Phantom 1 ) Fıg8: detaıl of the abnormal mass dısplayed by -9 to 9 tomosynthesis (Phantom 1 ) Fıg 9: detaıl of the abnormal mass dısplayed by -25 to 25 tomosynthesis (Phantom 1 ) ABN FP ROImass mean ROImass stdev ROIbkgd mean ROIbkgdstdev CNR Fıg 10: Phantom 1 ; with abnormal mass attenuation coefficient 0,200 [1/mm] ABN FP ROImass mean ROImass stdev ROIbkgd mean ROIbkgdstdev CNR Fıg 11: Phantom 2 ; with abnormal mass attenuation coefficient 0,090 [1/mm] 12

13 Fig 12: Tomo angle 9 reconstuctions planes of phantom 2 contained microcalfification (automatic contrast of image adjustement). Fig 13: Tomo angle 9 reconstuctions planes of phantom 2 contained microcalfification (contrast of image adjusted by operator). Fig 14: Tomo angle 9 reconstuctions planes of phantom 2 contained microcalfification (automatic contrast of image adjustement). 13

14 Fıg 15: Low energy image phantom 1 (Image J) Fıg 16: High energy image phantom 1 (Image J) Fıg 17: Combined image phantom 1 (Image J) Fıg 18: Low energy, high energy, combined images (XraySimulator Conclusions Simualtion of 2D irradiation and obtained images showed that for high density breast 2D mammography can provide false positive results. Tomosynthesis was found better for microcalcification detectability, althought in case of very low difference in attenuation coefficient between abnormal mass and bacground it was found that any method that was under investigation can contribute to correct diagnosis Special care should be taken when choosing between phantom models (solid state vs. Voxels) and Software used for obtaining image simulation Using mathematical antropomorhic phantom can contribute to virtual clinical studies. From the presentation: 14

15 EUTEMPE-RX: European Training and Education for Medical Physics Experts in Radiology MPE05: Anthropomorphic phantoms Applications of anthropomorphic phantoms for design and evaluation of advanced x-ray imaging techniques PROJECT ASSIGNMENT No. 2 Using computational breast phantom, perform a virtual study to determine the potential of the cone-beam CT for detectability of breast abnormalities, compared to conventional mammography To be considered: Involved Phantom(s): Breast Phantom Small to Medium size, Glandular to Dense background tissue, Tree min 15 lactiferous branches in total, Cooper and skin included, Output volume matrix size 400 x 400 x 400 voxels, (or alternatively 500 x 500 x 500 voxels), voxel size = 0.25 mm. Abnormalities: Five single microcalcifications up to 1 mm in diameter, randomly distributed. Imaging techniques: Conventional mammography and Cone-beam CT. Assumed Incident Dose: 5-12 mgy. Any other parameters are by your discretion and preference. To be completed: Create a phantom, according to the provided specification and add the abnormalities. Obtain synthetic conventional mammographic image of the phantom at an energy in the range kev, selected by your discretion. Obtain 180 synthetic projection images over a complete arc (every 2 degrees) at the same energy as above. Add a realistic level of noise, based on the required incident dose. Perform image reconstruction at different planes of interest. Compare subjectively (visually) the images. Compute objective figure of merits (FOMs) CNR in the region of the abnormalities and compare. 15

16 To be prepared: Report: a short written summary of the work and the obtained results; containing for example: o appropriate outcome images and profiles o tables with computed FOMs o discussion of the results and conclusions PowerPoint Presentation: to be used for a short (5-10 ) oral presentation. Useful hints: The tool for the phantom creation is the BreastSimulator tool. For introducing abnormalities use the BreastSimulator tool. Projection images are generated using the BreastSimulator tool. Image reconstruction is performed using FDKR tool. References: [1]. BreastSimulator User Guide. [2]. FDKR User Guide. 16

17 PROJECT IMPLEMENTATION REPORT Using computational breast phantom, perform a virtual study to determine the potential of the cone-beam CT for detectability of breast abnormalities, compared to conventional mammography Yanka Baneva and Anna Radice Our work was to create phantom of breast with Breast Simulator tool: small to medium size glandular to dense background tissue tree - min 15 lactiferous branches in total cooper and skin included five single microcalcifications up to 1 mm in diameter, randomly distributed. For the image techniques we used conventional mammography and Cone-beam CT. We created the phantom with the provided specifications: Energy of the range of 19 kev Semi-ellipsoid: (x0,y0,z0)=(24,0,0); (xr,yr,zr)=(40,40,40) Semi-hyperbolid: (x0,y0,z0,free_offset)=(25,0,55,11), (xr,yr,zr,c_param)=(5,5,5,6) Additional: Cooper ligaments: number =50000; radius=3mm;att_coeff =0.079mm -1 Pectoralis: yes, att_coeff =0.097mm -1 Skin: yes, att_coeff=0.08mm -1 Nipple: (x0,y0,z0)=(-15,0,-2); (xr,yr,zr)=(4,4,5) Ducts: (x0,y0,z0)=(14.5,0,-2);(ϕ,θ)=(-10, -10 ), att_coeff=0.079mm -1 Abnormalities: 5 single microcalcifications up to 1 mm in diameter, randomly distributed. Major ducts Major ducts height(mm) Major duct radius(mm) Lact.ducts Lact.Ducts in.height (mm) Lact.Ducts, lim.height (mm) Lact.Ducts In.radius (mm) Texture: att_coeff= mm-1 Size Voxel(mm) Nwalk N step LPF Dilation Filt.size Increment Hurst. Coeff No No Lobule radius (mm) St_dev. 17

18 Skin Ducts Cooper ligaments Microcalcifications Figure 1. That is the image of the phantom with all parts on it Then we obtained synthetic conventional planar mammographic image at the same energy level, with Breast simulator, with: SID = 600 mm, SDD =900 mm, ray tracing step = 0.1 mm, image size 500x500 pixels; Resolution: 5 pixels/mm. Figure 2. The microcalcifications are visible on the picture. Next step was to obtain 180 synthetic projection images over a complete arc with step 2 degrees, at same energy as above and to reconstruct the images by using FDKR tool at different planes of interest. Figure 3. 18

19 The first image in fig.3 represents the axial plane at z= -9 mm, the second one is lateral, at the same z. On both pictures the calcifications are still visible. Figure 4. At fig.4 we see the axial, lateral and coronal reconstructions at z=30mm. The calcifications in the first two images are still visible while for the third one not. That means that it is very important to put the correct number for z in order to receive good image. The final task was to compute CNR in the region of the abnormalities and to compare. This time we decided to reconstruct coronal images at the same z-level where the 2 of 5 microcalcifications are located in the breast phantom: Image comparison (z1 = - 4 mm), mammographic image versus coronal reconstruction CBCT image. A) Mammo Figure 5. Area Mean StdDev Min Max CNR 1 signal background B) Tomo image Area Mean StdDev Min Max CNR 1 signal background

20 CNR percentage difference = Image comparison (z1 = - 10 mm), mammographic image versus coronal reconstruction CBCT image. z = 10 A) Mammo Figure 6. Area Mean StdDev Min Max CNR 1signal background B)Tomo image ROI Area Mean StdDev Min Max CNR 1signal background CNR percentage difference = As a conclusion we may say that: CBCT is visually better. CBCT allows reconstruction images at different planes. CBCT reconstruction images have higher CNR. From the presentation: 20

21 21

22 22

23 EUTEMPE-RX: European Training and Education for Medical Physics Experts in Radiology MPE05: Anthropomorphic phantoms Applications of anthropomorphic phantoms for design and evaluation of advanced x-ray imaging techniques PROJECT ASSIGNMENT No. 3 Using computational breast phantom, perform a virtual study to determine the potential of cone-beam CT for detectability of breast abnormalities, compared to conventional mammography To be considered: Involved Phantom(s): Voxel-based phantom composed from segmented CT slices: \Phantoms\CT Slices\VoxelBreast2, Voxel size = 0.3 mm. Abnormalities: One Irregular mass up to 10 mm in size and one cluster of microcalcifications with a microcalcification of up to 1 mm in diameter. Imaging techniques: Conventional mammography and Cone-beam CT. Assumed Incident Dose: 5-12 mgy. Any other parameters are by your discretion and preference. To be completed: Create a phantom, according to the provided specification and add the abnormalities. Assign relevant attenuation coefficients to the different phantom regions. Obtain synthetic conventional mammographic image of the phantom at an energy value, selected by your discretion. Obtain 180 synthetic projection images over a complete arc (every 2 degree) at the same energy as above. Add a realistic level of noise, based on the required incident dose. Perform image reconstruction at different planes of interest. Compare subjectively (visually) the images. Compute objective figure of merits (FOMs) CNR in the region of the abnormalities and compare. 23

24 To be prepared: Report: a short written summary of the work and the obtained results; containing for example: o appropriate outcome images and profiles o tables with computed FOMs o discussion of the results and conclusions PowerPoint Presentation: to be used for a short (5-10 ) oral presentation. Useful hints: The tool for the phantom creation is the XRayImageSimulator tool. Microcalcifications may be modelled as ellipsoids or spheres. Exploit the option Voxel-based phantom Monte Carlo to convert the object to voxel based model. For the Irregular abnormality, use the tool AbnormalityMaker in the Tools directory. Use the script file abnormality_maker.m and modify accordingly. For introducing abnormalities use the XRayImageSimulator -> Tools -> Add an abnormality. For assigning linear attenuation coefficients, use the Matlab/Octave script: assign_linear_attenuation_coefficient.m. For inspection of the created volume, ImageJ is used. Tissue Value Skin 255 Air 0 Adipose 85 Gland 170 Projection images are generated using the XRayImageSimulator tool. Image reconstruction is performed using FDKR tool. References: [1]. XRayImageSimulator User Guide. [2]. FDKR User Guide. 24

25 PROJECT IMPLEMENTATION REPORT Using computational breast phantom, perform a virtual study to determine the potential of cone-beam CT for detectability of breast abnormalities, compared to conventional mammography Eric Pace and Lucie Sukupova Three microcalcifications were added into a basic CT breast image. A conventional mammography image and corresponding CT plane were reconstructed, and this is shown in Fig. 1. Fig. 1: The conventional mammography image and the reconstructed CT image Images were compared not only qualitatively but also quantitatively by the use of contrast to noise ratio (CNR). The values are included in Table 1. Mammography Image CT image Mean Background CNR Table 1: Values for an objective assessment The CNR values above show that the CT reconstructed image provides a much higher value of CNR, therefore the microcalcifications ares more visible. While the microcalcifications are visible in the mammography image and may be considered of sufficient quality for the radiologists, the 25

26 aim of this study was to compare the CNR of both imaging methods. In the simulation the microcalcifications were quite large, whereas this would not be so in the clinical situation. Thus it is possible to have microcalcifications which are not visible at all in the mammographic image but would be visible in the CT reconstruction. From the presentation: 26

27 EUTEMPE-RX: European Training and Education for Medical Physics Experts in Radiology MPE05: Anthropomorphic phantoms Applications of anthropomorphic phantoms for design and evaluation of advanced x-ray imaging techniques PROJECT ASSIGNMENT No. 4 Using computational breast phantom, perform a virtual study to determine the potential of breast tomosynthesis for detectability of breast abnormalities, compared to conventional mammography To be considered: Involved Phantom(s): Voxel-based phantom composed from segmented CT slices: \Phantoms\CT Slices\CompressedBreastSlices, Voxel size 0.3 mm. Abnormalities: One Irregular mass up to 10 mm in size and one cluster of microcalcifications with a microcalcification of up to 1 mm in diameter. Imaging techniques: Conventional mammography and Breast tomosynthesis mammography. Assumed Incident Dose: 5-12 mgy. Any other parameters are by your discretion and preference. To be completed: Create a phantom, according to the provided specification and add the abnormalities. Assign relevant attenuation coefficients to the different phantom regions. Obtain synthetic conventional mammographic image of the phantom at an energy in the range kev, selected by your discretion. Obtain 10 synthetic mammography images over an arc (-9 to 9) degree (every 2 degrees) and 26 synthetic mammography images over an arc (-25 to 25) degree (every 2 degrees) at the same energy as above. Add a realistic level of noise, based on the required incident dose. Perform image reconstruction at different planes of interest. Compare subjectively (visually) the images. 27

28 Compute objective figure of merits (FOMs) CNR in the region of the abnormalities and compare. To be prepared: Report: a short written summary of the work and the obtained results; containing for example: o appropriate outcome images and profiles o tables with computed FOMs o discussion of the results and conclusions PowerPoint Presentation: to be used for a short (5-10 ) oral presentation. Useful hints: The tool for the phantom creation is the XRayImageSimulator tool. Microcalcifications may be modelled as ellipsoids or spheres. Exploit the option Voxel-based phantom Monte Carlo to convert the object to voxel based model. For the Irregular abnormality, use the tool AbnormalityMaker in the Tools directory. Use the script file abnormality_maker.m and modify accordingly. For introducing abnormalities use the XRayImageSimulator -> Tools -> Add an abnormality. For assigning linear attenuation coefficients, use the Matlab/Octave script: assign_linear_attenuation_coefficient.m. For visualization and inspection of the created volume, ImageJ is used. The initial voxel values in the compressed breast matrix are: Tissue Value Skin 7 Air 119 Adipose 255 Gland 151 Photon noise is added by using add_noise.m script in the mcode directory. Projection images are generated using the XRayImageSimulator tool. Image reconstruction is performed using FDKR tool. References: [1]. XRayImageSimulator User Guide. [2]. FDKR User Guide. 28

29 PROJECT IMPLEMENTATION REPORT Using computational breast phantom, perform a virtual study to determine the potential of breast tomosynthesis for detectability of breast abnormalities, compared to conventional mammography Marius Laurikaitis 1 and Anastasios Konstantinidis 2 1 Oncology Hospital of Kaunas Clinics, Kaunas, Lithuania, marius@medicinosfizika.lt 2 Christie Medical Physics and Engineering, The Christie NHS Foundation Trust, Manchester, UK, taskon25@yahoo.gr Project tasks Create a Voxel-based phantom composed from segmented CT slices. Assign relevant attenuation coefficients to different phantom regions. Obtain conventional Mammographic image of the phantom (energy 20 kev). Obtain Digital Breast Tomosynthesis (DBT) images (energy 20 kev, arc -9 o to 9 o and -25 o to 25 o ). Add quantum noise considering 5 mgy incident dose. Perform image reconstruction at different planes (z = -20, -5, 0, 5, 20 mm). Compare subjectively the images. Compute objective figure of merit, i.e. Contrast-to-Noise Ratio (CNR) in the region of abnormalities and compare. Materials and methods For Voxel-based phantom creation we used XRayImageSimulator [1]. The image size was 641 x 357 x 175 pixels. For adding abnormalities we used XRayImageSimulator [1]. The abnormality objects were masses with size of 200 x 200 x 200 pixels. For assign attenuation coefficients we used the Octave [2] script assign_linnear_attenuation_coefficient.m. For irradiating Voxel-based phantom and getting conventional Mammographic and Digital Breast Tomosynthesis images we used BreastSimulator [3]. For adding quantum noise (incident dose - 5 mgy) to the images we used the Octave [2] script add_noise.m. For reconstruction of Digital Breast Tomosynthesis images we used FDKR [4]. For visualization of images and extraction of objective parameters we used ImageJ [5]. Results and Discussion Figure 1 shows synthetic Mammographic (a) and DBT images using 10 projections (-9 o to 9 o arc) (b) and (c) 26 projections (-25 o to 25 o arc) (c) both for z=-5 mm plane. DBT resulted in better visual detection of abnormality compared to Mammography. In particular, the reconstructed image over 26 projections was better compared to 10 projections. 29

30 (a) (b) (c) Figure 1. Visualization using (a) Mammography and Digital Breast Tomosynthesis modalities with (b) -9 o to 9 o arc and (c) -25 o to 25 o arc (both for z=-5 mm) Figure 2 shows the regions of interest (ROIs) used for the calculation of CNR parameter. The used background ROI was larger in order to take into account background non-uniformities. Figure 2. Computation of CNR: CNR=(mean_object mean_back)/std_back Figure 3 shows projection DBT images for different arcs (-9 o to 9 o (a) and -25 o to 25 o (b)) and planes (z = -20, -5, 0, 5, 20 mm). The abnormality can be visually detected in the range z = -5, 0 and 5 mm but it is hard to define which plane is best. (a) z = -20 mm z = -5 mm z = 0 mm z = 5 mm z = 20 mm (b) Figure 3. Visualization using Digital Breast Tomosynthesis with (a) -9 o to 9 o arc and (b) -25 o to 25 o arc in different planes (z = -20, -5, 0, 5, 20 mm) Table 1 shows the CNR values for different conditions, i.e. using Mammography and DBT for 10 and 26 projection images and 5 planes of interest. Mammography results in the best CNR value. Regarding DBT application, the combination of 26 projections with z = -5 mm resulted in high CNR value. 30

31 Table 1. Comparison of CNR values for different irradiation parameters. Technique CNR Mammography Digital Breast Tomosynthesis z = -20 mm z = -5 mm z = 0 mm z = 5 mm z = 20 mm DBT (-9 to 9 ) DBT (-25 to 25 ) Conclusions 1. DBT resulted in better visual detection of abnormality compared to Mammography. In particular, the reconstructed image over 26 projections was better compared to 10 projections. 2. We were able to visually detect the abnormality in the range z=-5, 0 and 5 mm but we could not say which one was best. 3. The CNR value using Mammography was higher compared to DBT. 4. DBT for 26 projections resulted in better CNR values compared to 10 projections. 5. Reconstructed images for z=-5 mm had the best CNR values for both sets of projections. References 1. XRayImageSimulator User Guide 2. Octave (available online: 3. BreastSimulator User Guide 4. FDKR User Guide 5. ImageJ (available online: From the presentation: 31

32 32

33 EUTEMPE-RX: European Training and Education for Medical Physics Experts in Radiology MPE05: Anthropomorphic phantoms Applications of anthropomorphic phantoms for design and evaluation of advanced x-ray imaging techniques PROJECT ASSIGNMENT No. 5 Using computational breast phantom, perform a virtual study to determine the potential of the dual-energy imaging for detectability of breast abnormalities, compared to conventional mammography To be considered: Involved Phantom(s): Voxel-based phantom composed from segmented CT slices: \Phantoms\CT Slices\CompressedBreastSlices, Voxel size 0.3 mm. Abnormalities: One irregular mass and five individual microcalcifications of up to 1 mm in diameter. Imaging techniques: Conventional mammography and dual-energy mammography. Assumed Incident Dose: 5-12 mgy. Any other parameters are by your discretion and preference. To be completed: Create a phantom, according to the provided specification and add the abnormalities. Assign relevant attenuation coefficients to the different phantom regions. Obtain synthetic conventional mammographic images of the phantom at an energy in the range kev. Obtain synthetic mammography images at three high x-ray energies, for instance: 50 kev, 65 kev and 80 kev or use other ones of your preference. Add a realistic level of noise, based on the required incident dose. Compute dual-energy difference images. Compare subjectively (visually) the images. Compute objective figure of merits (FOMs) CNR in the region of the abnormalities and compare. 33

34 To be prepared: Report: a short written summary of the work and the obtained results; containing for example: o appropriate outcome images and profiles o tables with computed FOMs o discussion of the results and conclusions PowerPoint Presentation: to be used for a short (5-10 ) oral presentation. Useful hints: The tool for the phantom creation is the XRayImageSimulator tool. Microcalcifications may be modelled as ellipsoids or spheres. Exploit the option Voxel based phantom Monte Carlo to convert the object to voxel based model. For the Irregular abnormality, use the tool AbnormalityMaker in the Tools directory. Use the script file abnormality_maker.m and modify accordingly. For introducing abnormalities use the XRayImageSimulator -> Tools -> Add an abnormality. For assigning linear attenuation coefficients, use the Matlab/Octave script: assign_linear_attenuation_coefficient.m. For inspection of the created volume, ImageJ is used. The initial voxel values in the compressed breast matrix are: Tissue Value Skin 7 Air 119 Adipose 255 Gland 151 Projection images are generated using the XRayImageSimulator tool. Photon noise is added by using add_noise.m script in the mcode directory. Dual-energy images are computed using the script calculate_dual_energy_image.m in Matlab/Octave or it may be also performed using ImageJ. References: [1]. XRayImageSimulator User Guide. [2]. FDKR User Guide. 34

35 PROJECT IMPLEMENTATION REPORT Using computational breast phantom, perform a virtual study to determine the potential of the dual-energy imaging for detectability of breast abnormalities, compared to conventional mammography. Rita Demou and Simona Avramova-Cholakova The aim of this report was to investigate whether dual-energy imaging improves detectability of breast abnormalities compared to conventional mammography. Both a visual assessment and a comparison of the CNR were employed in order to evaluate the two techniques. METHODOLOGY Using XRaySimlator a 3D matrix was constructed from slices recorded in a CT modality. A 641x357x175 matrix of a 0.3mm voxel size was created. The model represented a compressed breast phantom. Then, a solid geometry phantom was constructed to represent five microcalcifications and an irregular mass. Table 1 records the spatial and content information of the abnormalities. The phantom was converted to a vocalized matrix of 84x84x84 pixels of a size of 0.3mm. Table 1. Specifications of abnormalities Type Center (mm) Dimensions (mm) Compound φ, θ, ψ degrees MCal Ellipsoid (0, 0, 0) (0.7, 1, 0.6) CaCO3 (0, 0, 0) MCal Ellipsoid (2, -2, 3) (0.9, 1.2, 1.0) CaCO3 (0, 0, 0) MCal Ellipsoid (0, 3, -3) (0.9, 0.3, 0.5) CaCO3 (0, 0, 0) MCal Ellipsoid (-2, 4, 5) (0.5, 0.3, 0.4) CaCO3 (0, 0, 0) MCal Ellipsoid (2, 2, 2) (0.9, 1.2, 0.8) CaCO3 (0, 0, 0) IrMass Ellipsoid (8, 8, 8) (10, 5, 7) Water (45, -75, 0) The two matrices were combined utilizing XRaySimulator. The matrix representing the abnormalities was positioned at position (180, 200, 70) inside the CT breast model. Using Image J, the combined model was visualized both as a stack of slices and as a 3D volume. Figure 1 represents the combined phantom. 35

36 Octave was used to transform modeled coefficients of the materials included in the combined phantom to linear attenuation coefficients (LAC). Phantoms were created taking into account the liner attenuation coefficients at 20, 50, 65 and 80 kev. Table 2 records modeled and linear attenuation coefficients. Figure 1. Slices and 3D Volume of the combined phantom Table 2. Modeled and Linear attenuation coefficients Object Modeled Coefficient LAC 20keV LAC 50keV LAC 65keV LAC 80keV Skin Air Adipose Gland Water CaCO Planar projections were simulated for the four phantoms using analytical image simulator in XRaySimulator. Set irradiation properties were identical for the four phantoms and are recorded in table 3. Table 3. Irratiation properties of simulated projections Source to Isocenter Distance Source to Detector Distance Ray Step Resolution 600 mm 650 mm 0.1 mm 5 pixels/mm Image Size 1024 x 1024 Geometry Output Fan Beam Line Integral 36

37 Using Octave the 20keV projection was subtracted from 50, 65and 80keV image, consecutively to obtain the dual energy images. Finally, using ImageJ, the Contrast-to-Noise Ratio (CNR) was determined for 20keV planar projection and for the three dual energy images, applying the formulae:, where MPVobj is the mean pixel value of the regions of interest (ROI) in the object, and MPVbckgr and STDEVbckgr are the mean pixel value and the standard deviation of the ROI in the background, respectively. RESULTS Figure 2 shows the visual assessment of the two different techniques. Figure 2. The 20keV planar projection and the dual energy images. The ROI, chosen for the CNR calculation, are depicted in Figures 3-6 for both the biggest microcalcification and the irregular mass. 37

38 Figure 3. ROI in the calcification. Figure 4. ROI in the background of calcification. Figure 5. ROI in the mass. Figure 6. ROI in the background of mass. The results from the CNR calculations are presented in table 4. Mean is for mean pixel value and StdDev is the standard deviation. Planar Mammo_20 is for the planar mammography simulated image and DE describes the dual energy images for the different energy combinations. Table 4. Input data and results from calculation of CNR. Planar Mammo_20 Planar Mammo_20 DE_20_80 DE_20_65 DE_20_50 DE_20_80 DE_20_65 DE_20_50 Area Mean StdDev CNR Mass Background Calcification Background Mass Background Mass Background Mass Background Calcification Background Calcification Background Calcification Background

39 Figures 7 and 8 present a summary of the results in table 4. CONCLUSIONS Figure 7. Figure 8. Both visual assessment and comparison of the CNR clearly shows that dual energy technique significantly improves the detectability of breast abnormalities compared to conventional mammography. The improvement of the contrast of the micro-calcifications is more significant while increasing energy difference of the dual energy images. In the case of the low contrast mass, the CNR becomes worst when a low energy difference is employed. However, when energies above 65keV are used to get the high energy projections, the CNR improves in comparison with conventional mammography. From the presentation: 39

40 40

41 EUTEMPE-RX: European Training and Education for Medical Physics Experts in Radiology MPE05: Anthropomorphic phantoms Applications of anthropomorphic phantoms for design and evaluation of advanced x-ray imaging techniques PROJECT ASSIGNMENT No. 6 Using computational breast phantom, perform a virtual study to determine the potential of dual-energy imaging for detectability of breast abnormalities, compared to conventional mammography To be considered: Involved Phantom(s): LUCMR Small to Medium size, Average compressed breast thickness, Spheres sized mm in diameter, bulk material. Abnormalities: Two spherical mass abnormalities of different materials and five individual microcalcifications of up to 1 mm in diameter. Imaging techniques: Conventional mammography and dual-energy mammography. Assumed Incident Dose: 5-12 mgy. Any other parameters are by your discretion and preference. To be completed: Create a phantom, according to the provided specification and add the abnormalities. Choose appropriate materials from table 1. Obtain synthetic conventional mammographic images of the phantom at two energies in the range kev. Obtain synthetic mammography images at three higher x-ray energies, for instance: 45 kev, 60 kev and 75 kev. Add a realistic level of noise, based on the required incident dose. Compute dual-energy difference images. Compare subjectively (visually) the images. Compute objective figure of merits (FOMs) CNR in the region of the abnormalities and compare. To be prepared: Report: a short written summary of the work and the obtained results; containing for example: 41

42 o appropriate outcome images and profiles o tables with computed FOMs o discussion of the results and conclusions PowerPoint Presentation: to be used for a short (5-10 ) oral presentation. Useful hints: The tool for the phantom creation is the LUCMRGen tool. Take into account table 1 that contains list of modelled in XRayImageSimulator materials. For introducing abnormalities use the XRayImageSimulator tool. Take into account table 1. Projection images are generated using the XRayImageSimulator tool. Photon noise is added by using add_noise.m script file in the mcode directory. Modify the script file accordingly. The image subtraction may be performed in ImageJ but more flexibility is provided by Matlab/Octave. Dual-energy images are computed using the script calculate_dual_energy_image.m in Matlab/Octave or it may be also performed using ImageJ. Table 1: Modelled Materials C Al Si Cu Pb water bone compact ICRU-44 lung tissue - inflated brain, grey/white matter breast mass 75% gland 25% fat plexiglas (lucite, PMMA) ideal detector adipose ICRU-44 muscle ICRU-44 Gd2O2S (phosphor) air CsI (phosphor) bone cortical ICRU-44 polystyrene BaFBr gland CaCO3 soft tissue ICRU- 44 Breast 50% gland 50% fat ideal air SrFBr paraffin wax, C25H52 skin gray matter white matter blood SiO2 silicon gel epoxy resin nylon-6,6 water vapour ethanol ethanol vapour polyethylene povidoneiodine References: [1]. Bliznakova, K., I. Buliev, et al. (2013). Development and Evaluation of a Software Phantom Dedicated for Breast Imaging Experimentation: Preliminary (First) Results. 5th Pan-Hellenic Conference on Biomedical Technology (ELEBIT 2013). 4-6 April 2013, Athens, Greece. [2]. Cockmartin, L., N. Marshall, et al. (2012). Design and evaluation of a phantom with structured background for digital mammography and breast tomosynthesis. Philadelphia, PA LNCS:

43 PROJECT IMPLEMENTATION REPORT Potential of dual energy imaging for detectability of breast abnormalities Ciprian Cindea and Oliver Diaz Purpose : To explore the potential of computer simulation using phantom models and study the detection performance of several breast lesions in dual energy (DE) mammography and conventional mammography. Methods: L UCMR phantom was created using the LUCMRGen software. The phantom contains a PMMA case of semicircular shape (radius 40 and height 35mm). It was filled with 2246 water spheres of different radii to produce breast texture. Five spheres (1 mm 3 radius) of CaCO3 were randomly inserted within the LUCMR phantom to simulate microcalcifications. Moreover, two masses were modelled as spheres of radius 3 and 6 mm 3 containing breast mass tissue (75% glandular and 25% adipose tissue) and 100% glandular tissue respectively. XRAYImagingSimulator was used to generate the projection images for several monoenergetic beams: 17, 23, 45, 60, 75 kev. Poisson noise was added to the idealised projected images assuming 6 mgy entrance dose. DE images were generated from the subtraction of the noisy projections and contrast to noise ratio (CNR) was used to analyse the detection performance of the simulated lesions. Results: From all the 6 DE combinations of low high energy, the best configuration observed was 1760 kev. This setup illustrates an improvement of 191% and 196% in CNR for microcalcifications and masses respectively when compared with conventional energy projections (17 and 23 kev). Conclusions: Results have demonstrated that, according to CNR, DE is superior to mammography for lesion detection as background tissue could be eliminated from the image. All microcalcifications could be visually differentiated in both DE and conventional mammography, however, breast masses, which have similar attenuation coefficient to background tissue, were only visible in DE images. Computer simulations using phantom models have demonstrated great potential to optimize imaging parameters. LUCMR phantom model Conventional mammography Dual Energy projection 43

44 From the presentation: 44

45 EUTEMPE-RX: European Training and Education for Medical Physics Experts in Radiology MPE05: Anthropomorphic phantoms Applications of anthropomorphic phantoms for design and evaluation of advanced x-ray imaging techniques PROJECT ASSIGNMENT No. 7 Using computational breast phantom, perform a virtual study to determine the potential of breast tomosynthesis for detectability of breast abnormalities, compared to conventional mammography To be considered: Involved Phantom(s): LUCMR Small to Medium size, Average compressed breast thickness, Spheres sized mm in diameter, bulk material. Abnormalities: Two spherical mass abnormalities of different materials and five individual microcalcifications of up to 1 mm in diameter. Imaging techniques: Conventional mammography and breast tomosynthesis mammography. Assumed Incident Dose: 5-12 mgy. Any other parameters are by your discretion and preference. To be completed: Create a phantom, according to the provided specification and add the abnormalities. Choose appropriate materials from table 1. Obtain synthetic conventional mammographic images of the phantom at an energy selected by your discretion. Obtain 26 synthetic mammography images over an arc of your discretion. Add a realistic level of noise, based on the required incident dose. Perform image reconstruction at the planes of interest. Compare subjectively (visually) the images. Compute objective figure of merits (FOMs) CNR in the region of the abnormalities and compare. 45

46 To be prepared: Report: a short written summary of the work and the obtained results; containing for example: o appropriate outcome images and profiles o tables with computed FOMs o discussion of the results and conclusions PowerPoint Presentation: to be used for a short (5-10 ) oral presentation. Useful hints: The tool for the phantom creation is the LUCMRGen tool. Take into account table 1 that contains list of modelled in XRayImageSimulator materials. For introducing abnormalities use the XRayImageSimulator tool. Take into account table 1. Projection images are generated using the XRayImageSimulator tool. Photon noise is added by using add_noise.m script file in the mcode directory. Modify the script file accordingly. Image reconstruction is performed using FDKR tool. Table 1: Modelled Materials C Al Si Cu Pb water bone compact ICRU-44 lung tissue - inflated brain, grey/white matter breast mass 75% gland 25% fat plexiglas (lucite, PMMA) ideal detector adipose ICRU-44 muscle ICRU-44 Gd2O2S (phosphor) air CsI (phosphor) bone cortical ICRU-44 polystyrene BaFBr gland CaCO3 soft tissue ICRU-44 Breast 50% gland 50% fat ideal air SrFBr paraffin wax, C25H52 skin gray matter white matter blood SiO2 silicon gel epoxy resin nylon-6,6 water vapour ethanol ethanol vapour polyethylene povidoneiodine References: [1]. Bliznakova, K., I. Buliev, et al. (2013). Development and Evaluation of a Software Phantom Dedicated for Breast Imaging Experimentation: Preliminary (First) Results. 5th Pan-Hellenic Conference on Biomedical Technology (ELEBIT 2013). 4-6 April 2013, Athens, Greece. [2]. Cockmartin, L., N. Marshall, et al. (2012). Design and evaluation of a phantom with structured background for digital mammography and breast tomosynthesis. Philadelphia, PA LNCS:

47 PROJECT IMPLEMENTATION REPORT A virtual phantom study to evaluate the detectability of breast abnormalities in tomosynthesis compared to conventional mammography Lesley COCKMARTIN and Loredana BOGDAN PURPOSE To perform a virtual study to determine the potential of breast tomosynthesis compared to conventional mammography for detectability of breast abnormalities using a computational structured phantom. MATERIALS AND METHODS A virtual structured phantom was designed using the LUCMRGen tool that constructs solid geometrical objects. The compressed breast-shaped container was created based on semicylindrical objects and randomly filled with 1430 differently sized spheres (radii of 7.94 mm; 6.35 mm; 4.76 mm; 3.18 mm and 1.50 mm). The container dimensions are 120 x 60 x 35 mm and the wall thickness was chosen at 3 mm. Afterwards different attenuation coefficients representing different materials were assigned to the objects: skin tissue for the container, glandular tissue as bulk material and adipose tissue for the background spheres. Breast abnormalities i.e. masses and microcalcifications were afterwards added as spherical objects. For the representation of two masses, two spheres with diameter of 8 mm were created and consisted of two different materials (75% glandular/25% adipose and 50% glandular/50% adipose tissue). The masses were positioned at two different heights in the phantom. Four individual microcalcifications were also inserted as small spheres of radii 125, 150, 175 and 200 µm made of Calcium Carbonate, at the same height in the phantom as the masses. Next, projection images were generated using the XRayImageSimulator. An iso-centric geometry with source-to-isocenter distance of 600 mm and source-to-detector distance of 650 mm was selected together with a beam energy of 20 kev. The detector image size was assigned as 150 x 150 mm with a pixel size of 0.1 mm. For 2D mammography a cranio-caudal projection with the x-ray tube positioned at 0 was acquired, while for tomosynthesis 26 projections at different angles ranging from -25 to +25 with 2 increment were acquired. A Poison noise distribution was simulated for 8 mgy entrance dose and added to the projection images. Finally, the 26 tomosynthesis projections were reconstructed into a 3D volume using the FDKR tool and applying the filtered back-projection algorithm. Images of both modalities were visually compared and contrast-to-noise ratio (CNR) was calculated in the images with pixel values consisting of line integrals. RESULTS AND DISCUSSION Microcalcifications were equally visible in 2D mammography and tomosynthesis whereas masses were better visualized and easier detected in tomosynthesis compared to 2D mammography (Figure 1). CNR values for microcalcifications ranged from 2.15 to 5.25 in 2D mammography and from to in tomosynthesis (Table 1). While CNR values were higher in tomosynthesis images compared to mammograms, their visibility and detectability was rated equally. For masses, CNR values were equal to 3.19 and 0.75 for 2D mammography and 1.51 and 0.56 in tomosynthesis for the 50% glandular and 75% glandular mass respectively (Table 1). Higher CNR 47

48 values were found in 2D mammography although mass visibility was better in tomosynthesis (Figure 1). These lower CNR values may be due to the limited dynamic range of reconstructed tomosynthesis images. The better visibility of masses in tomosynthesis can be assigned to the reduction of structure overlap and therefore increasing the visibility of the mass in the reconstructed in-focus plane. Note that higher CNR values were found for the 50% glandular sphere compared to the 75% glandular sphere. We assume this is due to the random generation of the phantom with unknown percent division of background materials: skin (wall), 100% glandular (bulk) and adipose tissue (spheres). CONCLUSIONS Microcalcifications were equally visible in both modalities while masses were better represented in tomosynthesis images. In the future, CNR may be replaced by other quantitative metrics for the assessment of the detectability of breast abnormalities. Table 1. Contrast-to-noise ratio values for masses and microcalcifications embedded in a structured phantom measured in 2D mammographic projection and tomosynthesis reconstructed images. Material Dimension (radius in mm) CNR 2D mammography CNR Tomosynthesis Mass 1 50% gland Mass 2 75% gland µcalc 1 CaCO µcalc 2 CaCO µcalc 3 CaCO µcalc 4 CaCO

49 Figure 1. 2D projection (A and B) and tomosynthesis reconstructed in-focus plane (C and D) images of two masses and four individual microcalcifications. 49

50 From the presentation: 50

51 EUTEMPE-RX: European Training and Education for Medical Physics Experts in Radiology MPE05: Anthropomorphic phantoms Applications of anthropomorphic phantoms for design and evaluation of advanced x-ray imaging techniques PROJECT ASSIGNMENT No. 8 Using computational breast phantom, perform a virtual study to determine the influence of breast thickness on the detectability of breast abnormalities in conventional mammography and breast tomosynthesis To be considered: Involved Phantom(s): LUCMR Small to Medium size, Two different heights 20 and 60 mm Spheres sized mm in diameter, bulk material. Abnormalities: One spherical mass abnormality and one individual microcalcification of up to 1 mm in diameter, placed at different z-level. Imaging techniques: Conventional mammography and breast tomosynthesis mammography. Assumed Incident Dose: 5-12 mgy. Any other parameters are by your discretion and preference. To be completed: Create the phantoms, according to the provided specification and add the abnormalities. Choose appropriate materials from table 1. Obtain synthetic conventional mammographic images of the two phantoms at an energy selected by your discretion. Obtain 21 synthetic mammography images of the three phantoms over an arc -20 to 20 degrees (every 2 degrees) using the same energy. Add a realistic level of noise, based on the required incident dose. Perform image reconstruction at the plane of interest, where abnormalities are located. Compare subjectively (visually) the images. Compute objective figure of merits (FOMs) CNR in the region of the abnormalities and compare. To be prepared: Report: a short written summary of the work and the obtained results; containing for example: 51

52 o appropriate outcome images and profiles o tables with computed FOMs o discussion of the results and conclusions PowerPoint Presentation: to be used for a short (5-10 ) oral presentation. Useful hints: The tool for the phantom creation is the LUCMRGen tool. Take into account table 1 that contains list of modelled in XRayImageSimulator materials. For introducing abnormalities use the XRayImageSimulator tool. Take into account table 1. Projection images are generated using the XRayImageSimulator tool. Photon noise is added by using add_noise.m script file in the mcode directory. Modify the script file accordingly. Image reconstruction is performed using FDKR tool. Table 1: Modelled Materials C Al Si Cu Pb water bone compact ICRU-44 lung tissue - inflated brain, grey/white matter breast mass 75% gland 25% fat plexiglas (lucite, PMMA) ideal detector adipose ICRU-44 muscle ICRU-44 Gd2O2S (phosphor) air CsI (phosphor) bone cortical ICRU-44 polystyrene BaFBr gland CaCO3 soft tissue ICRU-44 Breast 50% gland 50% fat ideal air SrFBr paraffin wax, C25H52 skin gray matter white matter blood SiO2 silicon gel epoxy resin nylon-6,6 water vapour ethanol ethanol vapour polyethylene povidoneiodine References: [1]. Bliznakova, K., I. Buliev, et al. (2013). Development and Evaluation of a Software Phantom Dedicated for Breast Imaging Experimentation: Preliminary (First) Results. 5th Pan-Hellenic Conference on Biomedical Technology (ELEBIT 2013). 4-6 April 2013, Athens, Greece. [2]. Cockmartin, L., N. Marshall, et al. (2012). Design and evaluation of a phantom with structured background for digital mammography and breast tomosynthesis. Philadelphia, PA LNCS:

53 PROJECT IMPLEMENTATION REPORT Influence of breast thickness on the delectability of breast abnormalities on conventional mammography and breast tomosynthesis D.Petrov 1, A.Hustuc 2 1 dimitarbp@gmail.com 2 ahustuc@gmail.com Abstract In order to study the difference in detectability for different phantom thicknesses, for two imaging modalities Digital breast tomosynthesis (DBT) and digital mammography. Two different type lesions were placed inside the phantom case. Images were modified for specific dose level, adding blurriness and noise according to the chosen dose level. Detectability measurements has been done in CNR figure of merit. The results show better detectability in digital mammography for thinner phantom, and not significant difference for DBT modality. Method Using LUCMRGen home-made program, two anthropomorphic sphere phantoms has created with the following parameters: Container radius, mm Height, mm Spheres radius [mm] 1,59; 3,18; 5,0; 6,0; 7,0 1,59; 3,18; 5,0; 6,0; 7,0 No. of ellipsoids Material container Container filling material 453 objects skin gland 1383 objects skin gland Sphere material Adipose IRCU- 44 Adipose IRCU- 44 In order to study detectability we simulated two lesion types in each phantom. We chose breast calcification and non-spiculated mass lesions, with the following parameters: Lesions Abnormality material Object type Dimensions Calcification CaCO 3 Ellipsoid 1x1x1 Non-spiculated mass Breast mass Ellipsoid 6x6x6 We simulated a projection images using XrayImagingSimulator software to obtain a single digital mammography image and a set of DBT images with initial beam energy 27keV. System SID [mm] SDD [mm] Angles [deg] Image resolution Digital mammography px/mm Image sizes, [px] 512x x

54 DBT From -20 to +20, with step 1 2px/mm 512x x1024 To acquire real images, we used a home-made tool for adding noise and blurring in order to simulate 10mGy dose. For the DBT projection images a FBP reconstruction has been made, to acquire a stack of coronar slice images. Results For measurement we used ImageJ, with circular ROIs matching the size of the lesions. We calculated Contrast to Noise Ratio(CNR): System Phantom height CNR masses CNR Calcifications Digital Mammography Digital Breast Tomosynthesis 20mm 2,06 3,78 60mm 1,66 2,67 20mm 0,73 13,88 60mm 0,97 12,30 Discussion and conclusions In general using CNR as a figure of merit, we cannot compare the detectability of lesions between DM and DBT, because of the difference of the dynamic range. This causes a lower CNR for DBT, but lesions are easily visible. Human observer reading in 4AFC paradigm should be made and an evaluation of the detectability in percent correct figure of merit should be taken into account. Compare the imaging method In our particular case the CNR of microcalcification in DBT is greater for both phantom heights compared to digital mammography, and opposite for CNR comparison for masses. Compare the lesions For both imaging methods the CNR for masses is lower than the CNR for calcifications, which follows the human detectability of the images. Although the ratio for DBT is larger than the same ratio for digital mammography(15,82 and 1,73 respectively), this could be caused by the dynamic range difference between the two modalities. Compare the phantom 54

55 The phantoms are similar and the only difference is the height. This difference permitted to fit more objects (ellipsoids) in 60 mm phantom. For mammography this causes lower detectability for the thicker phantom, for both of the abnormalities, because of the more overlapping objects. For DBT since the CNR was evaluated in the central plane of the lesions and no overlapping was observed. In this case CNR20mm is lower, than CNR60mm for the mass lesions, and opposite for the microcalcifications, so we cannot conclude that there is a specific detectability difference depending of the phantom thickness. Two phantoms were simulated with different lesion types and different height. Projection images were acquired, then noise was added and comparison of the detectability in the two modalities has been made. From the presentation: 55

56 56

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),

More information

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT Assessment of 3D performance metrics D and 3D Metrics of Performance Towards Quality Index: Volumetric imaging systems X-ray based Volumetric imaging systems: CBCT/CT Tomosynthesis Samuel Richard and Ehsan

More information

Power Spectrum Analysis of an Anthropomorphic Breast Phantom Compared to Patient Data in 2D Digital Mammography and Breast Tomosynthesis

Power Spectrum Analysis of an Anthropomorphic Breast Phantom Compared to Patient Data in 2D Digital Mammography and Breast Tomosynthesis Power Spectrum Analysis of an Anthropomorphic Breast Phantom Compared to Patient Data in 2D Digital Mammography and Breast Tomosynthesis Lesley Cockmartin 1,*, Predrag R. Bakic 2, Hilde Bosmans 1, Andrew

More information

Acknowledgments and financial disclosure

Acknowledgments and financial disclosure AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories

More information

Background 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J.

Background 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging Stephen J. Glick* J. Michael O Connor**, Clay Didier**, Mini Das*, * University of Massachusetts Medical

More information

Refraction Corrected Transmission Ultrasound Computed Tomography for Application in Breast Imaging

Refraction Corrected Transmission Ultrasound Computed Tomography for Application in Breast Imaging Refraction Corrected Transmission Ultrasound Computed Tomography for Application in Breast Imaging Joint Research With Trond Varslot Marcel Jackowski Shengying Li and Klaus Mueller Ultrasound Detection

More information

A fast scatter field estimator for Digital Breast Tomosynthesis

A fast scatter field estimator for Digital Breast Tomosynthesis A fast scatter field estimator for Digital Breast Tomosynthesis Oliver Díaz a, David R. Dance b,c, Kenneth C. Young b,c, Premkumar Elangovan a, Predrag R. Bakic d and Kevin Wells a a Centre for Vision,

More information

Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction

Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction Klaus Erhard a, Michael Grass a, Sebastian Hitziger b, Armin Iske b and Tim Nielsen a a Philips Research Europe Hamburg,

More information

Limitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods

Limitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods Stereoscopic Breast Imaging Andrew D. A. Maidment, Ph.D. Chief, Physics Section Department of Radiology University of Pennsylvania Limitations of Projection Radiography Mammography is a projection imaging

More information

Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors

Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors Elena Marimon 1,2*, Hammadi Nait-Charif 1, Asmar Khan 2, Philip A. Marsden 3, Oliver Diaz 4 1 Centre for

More information

Partial Volume Simulation in Software Breast Phantoms

Partial Volume Simulation in Software Breast Phantoms Partial Volume Simulation in Software Breast Phantoms Feiyu Chen, David Pokrajac, Xiquan Shi, Fengshan Liu, Andrew D.A. Maidment a, Predrag R. Bakic a Delaware State University, 1200 N DuPont Hwy, Dover

More information

Advanced Image Reconstruction Methods for Photoacoustic Tomography

Advanced Image Reconstruction Methods for Photoacoustic Tomography Advanced Image Reconstruction Methods for Photoacoustic Tomography Mark A. Anastasio, Kun Wang, and Robert Schoonover Department of Biomedical Engineering Washington University in St. Louis 1 Outline Photoacoustic/thermoacoustic

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

Digital Scatter Removal in Mammography to enable Patient Dose Reduction

Digital Scatter Removal in Mammography to enable Patient Dose Reduction Digital Scatter Removal in Mammography to enable Patient Dose Reduction Mary Cocker Radiation Physics and Protection Oxford University Hospitals NHS Trust Chris Tromans, Mike Brady University of Oxford

More information

Enhanced material contrast by dual-energy microct imaging

Enhanced material contrast by dual-energy microct imaging Enhanced material contrast by dual-energy microct imaging Method note Page 1 of 12 2 Method note: Dual-energy microct analysis 1. Introduction 1.1. The basis for dual energy imaging Micro-computed tomography

More information

Detection of microcalcification clusters by 2D-mammography and narrow and wide angle digital breast tomosynthesis

Detection of microcalcification clusters by 2D-mammography and narrow and wide angle digital breast tomosynthesis Detection of microcalcification clusters by 2D-mammography and narrow and wide angle digital breast tomosynthesis Andria Hadjipanteli a, Premkumar Elangovan b, Padraig T Looney a, Alistair Mackenzie a,

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

4/19/2016. Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT

4/19/2016. Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT 1 Three manufacturers approved for Tomo Hologic and GE, and Siemens Why 2D Digital Mammography 2D FFDM it appears to

More information

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

doi: /

doi: / Shuang Liu ; Mary Salvatore ; David F. Yankelevitz ; Claudia I. Henschke ; Anthony P. Reeves; Segmentation of the whole breast from low-dose chest CT images. Proc. SPIE 9414, Medical Imaging 2015: Computer-Aided

More information

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. 2-D D Dose-CT Mapping in Geant4 Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. Table of Contents Background & Purpose Materials Methods

More information

Medical Image Processing: Image Reconstruction and 3D Renderings

Medical Image Processing: Image Reconstruction and 3D Renderings Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

Scatter reduction for grid-less mammography using the convolutionbased image post-processing technique

Scatter reduction for grid-less mammography using the convolutionbased image post-processing technique Scatter reduction for grid-less mammography using the convolutionbased image post-processing technique Elena Marimón *a,b, Hammadi Nait-Charif a, Asmar Khan b, Philip A. Marsden c, Oliver Diaz d a CDE

More information

Tomographic Reconstruction

Tomographic Reconstruction Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)

More information

Digital breast tomosynthesis: comparison of different methods to calculate patient doses

Digital breast tomosynthesis: comparison of different methods to calculate patient doses Digital breast tomosynthesis: comparison of different methods to calculate patient doses Poster No.: C-2220 Congress: ECR 2011 Type: Scientific Paper Authors: A. Jacobs 1, L. Cockmartin 1, D. R. Dance

More information

A Method for Producing Simulated Mammograms: Observer Study

A Method for Producing Simulated Mammograms: Observer Study A Method for Producing Simulated Mammograms: Observer Study Payam Seifi M.Sc. Michael R. Chinander Ph.D. Robert M. Nishikawa Ph.D., FAAPM Carl J. Vyborny Translational Laboratory for Breast Imaging Research

More information

Acknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury

Acknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury A. Sisniega et al. (presented at RSNA 214) High Performance Cone-Beam CT of Acute Traumatic Brain Injury A. Sisniega 1 W. Zbijewski 1, H. Dang 1, J. Xu 1 J. W. Stayman 1, J. Yorkston 2, N. Aygun 3 V. Koliatsos

More information

Monte Carlo Reference Data Sets for Imaging Research

Monte Carlo Reference Data Sets for Imaging Research AAPM REPORT NO. 195 Monte Carlo Reference Data Sets for Imaging Research The Report of AAPM Task Group 195 September 2015 DISCLAIMER: This publication is based on sources and information believed to be

More information

CT vs. VolumeScope: image quality and dose comparison

CT vs. VolumeScope: image quality and dose comparison CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,

More information

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE / Radiographic Tomosynthesis: Image Quality and Artifacts Reduction Baojun Li, Ph.D Department of Radiology Boston University Medical Center 2012 AAPM Annual Meeting Background Linear Trajectory Tomosynthesis

More information

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic

More information

CHAPTER 6 DETECTION OF MASS USING NOVEL SEGMENTATION, GLCM AND NEURAL NETWORKS

CHAPTER 6 DETECTION OF MASS USING NOVEL SEGMENTATION, GLCM AND NEURAL NETWORKS 130 CHAPTER 6 DETECTION OF MASS USING NOVEL SEGMENTATION, GLCM AND NEURAL NETWORKS A mass is defined as a space-occupying lesion seen in more than one projection and it is described by its shapes and margin

More information

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT.

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT. Progress report: Development of assessment and predictive metrics for quantitative imaging in chest CT Subaward No: HHSN6801000050C (4a) PI: Ehsan Samei Reporting Period: month 1-18 Deliverables: 1. Deployment

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

Scattered radiation in DBT geometries with flexible breast compression paddles: A Monte Carlo simulation study

Scattered radiation in DBT geometries with flexible breast compression paddles: A Monte Carlo simulation study Scattered radiation in DBT geometries with flexible breast compression paddles: A Monte Carlo simulation study Oliver Diaz, Eloy García, Arnau Oliver, Joan Martí, Robert Martí a a VICOROB research institute,

More information

CIVA Computed Tomography Modeling

CIVA Computed Tomography Modeling CIVA Computed Tomography Modeling R. FERNANDEZ, EXTENDE, France S. LEGOUPIL, M. COSTIN, D. TISSEUR, A. LEVEQUE, CEA-LIST, France page 1 Summary Context From CIVA RT to CIVA CT Reconstruction Methods Applications

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms.

Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms. Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms. D.Leone, A.Häußler Intitute for Nuclear Waste Disposal, Karlsruhe Institute for Technology,

More information

A comparative study of limited-angle cone-beam reconstruction methods for breast tomosynthesis

A comparative study of limited-angle cone-beam reconstruction methods for breast tomosynthesis A comparative study of limited-angle cone-beam reconstruction methods for breast tomosynthesis Yiheng Zhang, a Heang-Ping Chan, Berkman Sahiner, Jun Wei, Mitchell M. Goodsitt, Lubomir M. Hadjiiski, Jun

More information

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid by Yingxuan Chen Graduate Program in Medical Physics Duke University Date: Approved: Lei Ren, Supervisor Fang Fang Yin, Chair

More information

Disclosure 7/24/2014. Validation of Monte Carlo Simulations For Medical Imaging Experimental validation and the AAPM Task Group 195 Report

Disclosure 7/24/2014. Validation of Monte Carlo Simulations For Medical Imaging Experimental validation and the AAPM Task Group 195 Report Validation of Monte Carlo Simulations For Medical Imaging Experimental validation and the AAPM Task Group 195 Report Ioannis Sechopoulos, Ph.D., DABR Diagnostic Imaging Physics Lab Department of Radiology

More information

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1 Modifications for P551 Fall 2013 Medical Physics Laboratory Introduction Following the introductory lab 0, this lab exercise the student through

More information

DUE to beam polychromacity in CT and the energy dependence

DUE to beam polychromacity in CT and the energy dependence 1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact

More information

Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc.

Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc. Unique Features of the GE Senoclaire Tomosynthesis System Tyler Fisher, M.S., DABR Therapy Physics, Inc. Conflict of Interest Disclosure I have no conflicts to disclose. Learning Objectives Overview of

More information

Michael Speiser, Ph.D.

Michael Speiser, Ph.D. IMPROVED CT-BASED VOXEL PHANTOM GENERATION FOR MCNP MONTE CARLO Michael Speiser, Ph.D. Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX September 1 st, 2012 CMPWG Workshop Medical

More information

Average glandular dose coefficients for pendant-geometry breast CT using realistic breast phantoms

Average glandular dose coefficients for pendant-geometry breast CT using realistic breast phantoms Average glandular dose coefficients for pendant-geometry breast CT using realistic breast phantoms Andrew M. Hernandez Department of Radiology, Biomedical Engineering Graduate Group, University of California

More information

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct J. DHAENE* 1, E. PAUWELS 1, T. DE SCHRYVER 1, A. DE MUYNCK 1, M. DIERICK 1, L. VAN HOOREBEKE 1 1 UGCT Dept. Physics

More information

Optimisation of Toshiba Aquilion ONE Volume Imaging

Optimisation of Toshiba Aquilion ONE Volume Imaging Optimisation of Toshiba Aquilion ONE Volume Imaging Jane Edwards, RPRSG Royal Free London NHS Foundation Trust Dr Mufudzi Maviki, Plymouth Hospitals NHS Trust Background In 2011/12 Radiology at RFH was

More information

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization

More information

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Marc Kachelrieß, a Katia Sourbelle, and Willi A. Kalender Institute of Medical Physics, University of

More information

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Stefan C. Saur 1, Hatem Alkadhi 2, Luca Regazzoni 1, Simon Eugster 1, Gábor Székely 1, Philippe Cattin 1,3 1 Computer Vision

More information

Scatter Correction Methods in Dimensional CT

Scatter Correction Methods in Dimensional CT Scatter Correction Methods in Dimensional CT Matthias Baer 1,2, Michael Hammer 3, Michael Knaup 1, Ingomar Schmidt 3, Ralf Christoph 3, Marc Kachelrieß 2 1 Institute of Medical Physics, Friedrich-Alexander-University

More information

ENGLISH. Digital Breast Tomosynthesis

ENGLISH. Digital Breast Tomosynthesis ENGLISH Digital Breast Tomosynthesis Medical imaging excellence Planmed is strongly committed to improving the early detection of breast cancer. Our mission is to provide top quality imaging systems for

More information

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation

More information

COMPREHENSIVE QUALITY CONTROL OF NMR TOMOGRAPHY USING 3D PRINTED PHANTOM

COMPREHENSIVE QUALITY CONTROL OF NMR TOMOGRAPHY USING 3D PRINTED PHANTOM COMPREHENSIVE QUALITY CONTROL OF NMR TOMOGRAPHY USING 3D PRINTED PHANTOM Mažena MACIUSOVIČ *, Marius BURKANAS *, Jonas VENIUS *, ** * Medical Physics Department, National Cancer Institute, Vilnius, Lithuania

More information

Automated Lesion Detection Methods for 2D and 3D Chest X-Ray Images

Automated Lesion Detection Methods for 2D and 3D Chest X-Ray Images Automated Lesion Detection Methods for 2D and 3D Chest X-Ray Images Takeshi Hara, Hiroshi Fujita,Yongbum Lee, Hitoshi Yoshimura* and Shoji Kido** Department of Information Science, Gifu University Yanagido

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

Quality control phantoms and protocol for a tomography system

Quality control phantoms and protocol for a tomography system Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing

More information

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Tecniques of imaging in Overview Introduction Tecniques of imaging

More information

Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application

Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application Physics in Medicine & Biology Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application To cite this article: Polad M Shikhaliev and Shannon G Fritz 2011

More information

Università degli Studi di Napoli Federico II

Università degli Studi di Napoli Federico II Università degli Studi di Napoli Federico II DOTTORATO DI RICERCA IN FISICA Ciclo XXIX Coordinatore: Prof. Salvatore Capozziello Dose and image quality in X-ray phase contrast breast imaging Settore Scientifico

More information

Validation of Cooper s Ligament Thickness in Software Breast Phantoms

Validation of Cooper s Ligament Thickness in Software Breast Phantoms Validation of Cooper s Ligament Thickness in Software Breast Phantoms Adam J. Kuperavage* a, Abdullah-Al-Zubaer Imran a, Predrag R. Bakic b, Andrew Maidment b, David D. Pokrajac a a Delaware State University,

More information

Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations

Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations Electron Dose Kernels (EDK) for Secondary Particle Transport in Deterministic Simulations A. Al-Basheer, G. Sjoden, M. Ghita Computational Medical Physics Team Nuclear & Radiological Engineering University

More information

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Joonki Noh, Jeffrey A. Fessler EECS Department, The University of Michigan Paul E. Kinahan Radiology Department,

More information

Computer-Tomography II: Image reconstruction and applications

Computer-Tomography II: Image reconstruction and applications Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de

More information

10/9/2018. Deborah Thames BSRS RT (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT

10/9/2018. Deborah Thames BSRS RT (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT Deborah Thames BSRS RT (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT 1 Mammography Five FFDM approved for Tomo Hologic, Ge Senoclair and GE Pristina, Siemens, and Fujifilm 2 Why 2D Digital

More information

MIXTURE MODELING FOR DIGITAL MAMMOGRAM DISPLAY AND ANALYSIS 1

MIXTURE MODELING FOR DIGITAL MAMMOGRAM DISPLAY AND ANALYSIS 1 MIXTURE MODELING FOR DIGITAL MAMMOGRAM DISPLAY AND ANALYSIS 1 Stephen R. Aylward, Bradley M. Hemminger, and Etta D. Pisano Department of Radiology Medical Image Display and Analysis Group University of

More information

Investigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography

Investigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography Investigating the influence of workpiece placement on the uncertainty of measurements in industrial computed tomography Natalia Grozmani 1, Andrea Buratti 1, Robert H. Schmitt 1 More info about this article:

More information

ICRP Symposium on Radiological Protection Dosimetry

ICRP Symposium on Radiological Protection Dosimetry ICRP Symposium on Radiological Protection Dosimetry Tokyo February 18 th, 2016 M. Zankl a, J. Becker a, K.F. Eckerman b, W.E. Bolch c, R. Behrens d a Helmholtz Zentrum München, Neuherberg, Germany b Oak

More information

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers Krista Chytyk MSc student Supervisor: Dr. Boyd McCurdy Introduction The objective of cancer radiotherapy

More information

ENHANCEMENT OF MAMMOGRAPHIC IMAGES FOR DETECTION OF MICROCALCIFICATIONS

ENHANCEMENT OF MAMMOGRAPHIC IMAGES FOR DETECTION OF MICROCALCIFICATIONS ENHANCEMENT OF MAMMOGRAPHIC IMAGES FOR DETECTION OF MICROCALCIFICATIONS Damir Seršiæ and Sven Lonèariæ Department of Electronic Systems and Information Processing, Faculty of Electrical Engineering and

More information

Validation of a digital mammography image simulation chain with automated scoring of CDMAM images

Validation of a digital mammography image simulation chain with automated scoring of CDMAM images Validation of a digital mammography image simulation chain with automated scoring of CDMAM images Mary Yip 1, Abdulaziz Alsager 2, Emma Lewis 1, Kevin Wells 1, and Kenneth C. Young 3 1 Centre for Vision,

More information

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Effects of the difference in tube voltage of the CT scanner on. dose calculation Effects of the difference in tube voltage of the CT scanner on dose calculation Dong Joo Rhee, Sung-woo Kim, Dong Hyeok Jeong Medical and Radiological Physics Laboratory, Dongnam Institute of Radiological

More information

Fundamentals of CT imaging

Fundamentals of CT imaging SECTION 1 Fundamentals of CT imaging I History In the early 1970s Sir Godfrey Hounsfield s research produced the first clinically useful CT scans. Original scanners took approximately 6 minutes to perform

More information

45 µm polystyrene bead embedded in scattering tissue phantom. (a,b) raw images under oblique

45 µm polystyrene bead embedded in scattering tissue phantom. (a,b) raw images under oblique Phase gradient microscopy in thick tissue with oblique back-illumination Tim N Ford, Kengyeh K Chu & Jerome Mertz Supplementary Figure 1: Comparison of added versus subtracted raw OBM images 45 µm polystyrene

More information

3D Computed Tomography (CT) Its Application to Aerospace Industry

3D Computed Tomography (CT) Its Application to Aerospace Industry 3D Computed Tomography (CT) Its Application to Aerospace Industry C. Muralidhar, M. P. Subramanian, V. Ravi Shankar and G. Chandrasekhar Directorate of Non Destructive Evaluation, Defence Research & Development

More information

RADIOMICS: potential role in the clinics and challenges

RADIOMICS: potential role in the clinics and challenges 27 giugno 2018 Dipartimento di Fisica Università degli Studi di Milano RADIOMICS: potential role in the clinics and challenges Dr. Francesca Botta Medical Physicist Istituto Europeo di Oncologia (Milano)

More information

Design Considerations in Optimizing a Breast Tomosynthesis System

Design Considerations in Optimizing a Breast Tomosynthesis System Design Considerations in Optimizing a Breast Tomosynthesis System Andrew Smith, Ph.D., Vice President - Imaging Science, Hologic Introduction Breast tomosynthesis, also referred to as three-dimensional

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA Computed Tomography Principles, Design, Artifacts, and Recent Advances THIRD EDITION Jiang Hsieh SPIE PRESS Bellingham, Washington USA Table of Contents Preface Nomenclature and Abbreviations xi xv 1 Introduction

More information

SVM-based CBIR of Breast Masses on Mammograms

SVM-based CBIR of Breast Masses on Mammograms SVM-based CBIR of Breast Masses on Mammograms Lazaros Tsochatzidis, Konstantinos Zagoris, Michalis Savelonas, and Ioannis Pratikakis Visual Computing Group, Dept. of Electrical and Computer Engineering,

More information

Loma Linda University Medical Center Dept. of Radiation Medicine

Loma Linda University Medical Center Dept. of Radiation Medicine Loma Linda University Medical Center Dept. of Radiation Medicine and Northern Illinois University Dept. of Physics and Dept. of Computer Science Presented by George Coutrakon, PhD NIU Physics Dept. Collaborators

More information

Estimating 3D Respiratory Motion from Orbiting Views

Estimating 3D Respiratory Motion from Orbiting Views Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng, Jeffrey A. Fessler, James M. Balter The University of Michigan Oct. 2005 Funding provided by NIH Grant P01 CA59827 Motivation Free-breathing

More information

Modifications for P551 Fall 2014

Modifications for P551 Fall 2014 LAB DEMONSTRATION COMPUTED TOMOGRAPHY USING DESKCAT 1 Modifications for P551 Fall 2014 Introduction This lab demonstration explores the physics and technology of Computed Tomography (CT) and guides the

More information

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT More Info at Open Access Database www.ndt.net/?id=18754 Novel evaluation method of low contrast resolution performance of dimensional X-ray CT Makoto Abe 1, Hiroyuki Fujimoto 1, Osamu Sato 1, Kazuya Matsuzaki

More information

GPU applications in Cancer Radiation Therapy at UCSD. Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC

GPU applications in Cancer Radiation Therapy at UCSD. Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC GPU applications in Cancer Radiation Therapy at UCSD Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC Conventional Radiotherapy SIMULATION: Construciton, Dij Days PLANNING:

More information

Basic Radiation Oncology Physics

Basic Radiation Oncology Physics Basic Radiation Oncology Physics T. Ganesh, Ph.D., DABR Chief Medical Physicist Fortis Memorial Research Institute Gurgaon Acknowledgment: I gratefully acknowledge the IAEA resources of teaching slides

More information

Cardiac Dual Energy CT: Technique

Cardiac Dual Energy CT: Technique RSNA 2013, VSCA51-01, Chicago, Dec. 5, 2013 Cardiac Radiology Series Cardiac Dual Energy CT: Technique Willi A. Kalender, Ph.D. Institute of Medical Physics University of Erlangen www.imp.uni-erlangen.de

More information

ACMP 25th Annual Meeting

ACMP 25th Annual Meeting Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation ACMP 25th Annual Meeting Seattle, WA May 3, 2008 Melissa C. Martin, M.S., FACR Therapy Physics, Inc. 879 West 190th St., Ste 419,

More information

Task-based Assessment of X-ray Breast Imaging Systems Using Insilico

Task-based Assessment of X-ray Breast Imaging Systems Using Insilico 55 th : Virtual Tools for Validation of X-ray Breast Imaging Systems Task-based Assessment of X-ray Breast Imaging Systems Using Insilico Modeling Tools Rongping Zeng Division of Imaging and Applied Mathematics

More information

A Generation Methodology for Numerical Phantoms with Statistically Relevant Variability of Geometric and Physical Properties

A Generation Methodology for Numerical Phantoms with Statistically Relevant Variability of Geometric and Physical Properties A Generation Methodology for Numerical Phantoms with Statistically Relevant Variability of Geometric and Physical Properties Steven Dolly 1, Eric Ehler 1, Yang Lou 2, Mark Anastasio 2, Hua Li 2 (1) University

More information

Introduction to Medical Image Processing

Introduction to Medical Image Processing Introduction to Medical Image Processing Δ Essential environments of a medical imaging system Subject Image Analysis Energy Imaging System Images Image Processing Feature Images Image processing may be

More information

INVESTIGATION OF PHYSICAL PROCESSES IN DIGITAL X-RAY TOMOSYNTHESIS IMAGING OF THE BREAST. A Thesis Presented to The Academic Faculty

INVESTIGATION OF PHYSICAL PROCESSES IN DIGITAL X-RAY TOMOSYNTHESIS IMAGING OF THE BREAST. A Thesis Presented to The Academic Faculty INVESTIGATION OF PHYSICAL PROCESSES IN DIGITAL X-RAY TOMOSYNTHESIS IMAGING OF THE BREAST A Thesis Presented to The Academic Faculty By Ioannis Sechopoulos In Partial Fulfillment Of the Requirements for

More information