Moving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region
|
|
- Brooke Garrison
- 5 years ago
- Views:
Transcription
1 Moving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region Andreas Hahn 1,2, Sebastian Sauppe 1,2, Michael Knaup 1, and Marc Kachelrieß 1,2 1 German Cancer Research Center (DKFZ), Heidelberg, Germany 2 University of Heidelberg, Heidelberg, Germany
2 Metal Artifacts beam hardening 2
3 Metal Artifacts beam hardening + scatter 3
4 Metal Artifacts beam hardening + scatter + directed noise 4
5 Metal Artifacts sharp edges + motion 5
6 Illustration of Conventional Metal Artifact Reduction: Normalized MAR (NMAR) Interpolation Normalized sinogram Interpol. & norm. Normalization Denormalization Input Original sinogram Metal projections Sinogram of tern. im. Corrected sinogram Uncorrected image Metal image Ternary image Corrected image Thresholding Output Meyer, Raupach, Lell, Schmidt, and Kachelrieß, Normalized metal artifact reduction (NMAR) in computed tomography, Med. Phys. 37(10): ,
7 Moving Metal Image-guided radiation therapy (IGRT) CBCT imaging unit mounted on gantry of a LINAC treatment system Slow gantry rotation speed of 6 per second (60 s/360 ) Much slower than clinical CT devices (0.25 s /360 ) Breathing about 10 to 30 rpm (respirations per minute) and thus per scan Linear Accelerator kv Source Detector Projections Motion due to breathing Gantry Rotation Metal subject to respiratory motion 7
8 Problem of Conventional MAR Conventional inpainting-based methods like NMAR do not account for motion. Threshold-based segmentation yields a static metal insert. Metal segmentation in volume Metal mask in projections Reconstructed volume after interpolation 8
9 Problem of Conventional MAR Conventional inpainting-based methods like NMAR do not account for motion. Threshold-based segmentation yields a static metal insert. Metal segmentation in volume Metal mask in projections Reconstructed volume after interpolation 9
10 Accounting for Metal Motion Most MAR methods do not account for motion. Several publications on raw data-based segmentation: Zhang et al.: Reducing metal artifacts in cone-beam CT images by preprocessing projection data, Int J Radiat Oncol Biol Phys 67(3): , March 2007 requires user input Veldkamp et al.: Development and validation of segmentation and interpolation techniques in sinograms for metal artifact suppression in CT, Med Phys 37(2): , February 2010 lacks robustness New approach aims for a combination of image and raw data-based segmentation: Toftegaard et al.: Moving metal artifact reduction in cone-beam CT scans with implanted cylindrical gold markers, Med Phys 41, December 2014 prior knowledge needed (shape of inserts) 10
11 Accounting for Metal Motion Combination of volume and raw data-based metal segmentation: Moving metal artifact reduction (MMAR) Brehm et al.: Moving metal artifact reduction (MMAR): A metal artifact reduction algorithm for flat 280 detector cone-beam CT scans with metal subject to respiratory motion, RSNA Conference Proc., November 2011 Metal segmentation in volume Metal mask in projections Reconstructed volume after interpolation MMAR refined segmentation 11
12 MoCoMAR Aim: Removal of metal artifacts in 3D CBCT volumes. Idea: For NMAR 1 static metal is segmented in a 3D volume. For our new approach (MoCoMAR), NMAR is improved by segmenting the metal in a 4D volume. The 4D volume is reconstructed using the acmoco 2 algorithm. Projections 4D Volume 4D Met. Mask in Volume 3D Met. Mask in Projections acmoco Thresholding Forward projection according to motion bin Corrected Volume NMAR 1: Meyer, Raupach, Lell, Schmidt, and Kachelrieß, Normalized metal artifact reduction (NMAR) in computed tomography, Med. Phys. 37(10): , : Brehm, Paysan, Oelhafen, and Kachelrieß, Artifact-resistant motion estimation with a patient-specific artifact model for motion-compensated cone-beam CT, Med. Phys. 40(10),
13 sagittal axial Results Patient 1 Standard NMAR MMAR MoCoMAR 13
14 sagittal axial Results Patient 1 Standard NMAR MMAR MoCoMAR 14
15 Conclusions MMAR and MoCoMAR outperform NMAR when it comes to moving metal inserts. The proposed method (MoCoMAR) was able to achieve similar results to MMAR. It is, however, computationally more expensive. More patients have to be evaluated to see if there are cases where one of the methods that account for motion outperforms the other. 15
16 Thank You! This presentation will soon be available at Job opportunities through DKFZ s international PhD or Postdoctoral Fellowship programs ( or directly through Marc Kachelriess (marc.kachelriess@dkfz.de). This study was supported by DFG KA 1678/13. Parts of the reconstruction software were provided by RayConStruct GmbH, Nürnberg, Germany. 16
Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT
Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT Marcus Brehm 1,2, Thorsten Heußer 1, Pascal Paysan 3, Markus Oehlhafen 3, and Marc Kachelrieß 1,2 1 German Cancer Research Center
More informationAutomatic Intrinsic Cardiac and Respiratory Gating from Cone-Beam CT Scans of the Thorax Region
Automatic Intrinsic Cardiac and Respiratory Gating from Cone-Beam CT Scans of the Thorax Region Andreas Hahn 1, Sebastian Sauppe 1, Michael Lell 2, and Marc Kachelrieß 1 1 German Cancer Research Center
More informationDeep Scatter Estimation (DSE): Feasibility of using a Deep Convolutional Neural Network for Real-Time X-Ray Scatter Prediction in Cone-Beam CT
Deep Scatter Estimation (DSE): Feasibility of using a Deep Convolutional Neural Network for Real-Time X-Ray Scatter Prediction in Cone-Beam CT Joscha Maier 1,2, Yannick Berker 1, Stefan Sawall 1,2 and
More informationDeep Scatter Estimation (DSE): Accurate Real-Time Scatter Estimation for X-Ray CT using a Deep Convolutional Neural Network
Deep Scatter Estimation (DSE): Accurate Real-Time Scatter Estimation for X-Ray CT using a Deep Convolutional Neural Network Joscha Maier 1,2, Stefan Sawall 1,2, and Marc Kachelrieß 1,2 1 German Cancer
More informationRespiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data
Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data Christopher M Rank 1, Thorsten Heußer 1, Andreas Wetscherek 1, and Marc Kachelrieß 1 1 German Cancer
More informationMotion Compensation from Short-Scan Data in Cardiac CT
Motion Compensation from Short-Scan Data in Cardiac CT Juliane Hahn 1,2, Thomas Allmendinger 1, Herbert Bruder 1, and Marc Kachelrieß 2 1 Siemens Healthcare GmbH, Forchheim, Germany 2 German Cancer Research
More informationFrequency split metal artifact reduction (FSMAR) in computed tomography
The Johns Hopkins University Advanced Computer Integrated Surgery Group 4 Metal Artifact Removal in C-arm Cone-Beam CT Paper Seminar Critical Review of Frequency split metal artifact reduction (FSMAR)
More informationThe Near Future in Cardiac CT Image Reconstruction
SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de
More informationOptimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago
Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic
More informationScatter Correction Methods in Dimensional CT
Scatter Correction Methods in Dimensional CT Matthias Baer 1,2, Michael Hammer 3, Michael Knaup 1, Ingomar Schmidt 3, Ralf Christoph 3, Marc Kachelrieß 2 1 Institute of Medical Physics, Friedrich-Alexander-University
More informationState-of-the-Art IGRT
in partnership with State-of-the-Art IGRT Exploring the Potential of High-Precision Dose Delivery and Real-Time Knowledge of the Target Volume Location Antje-Christin Knopf IOP Medical Physics Group Scientific
More informationCarestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2)
Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2) Author: Levon Vogelsang Introduction Cone beam computed tomography (CBCT), or cone beam CT technology, offers considerable promise
More information7/31/ D Cone-Beam CT: Developments and Applications. Disclosure. Outline. I have received research funding from NIH and Varian Medical System.
4D Cone-Beam CT: Developments and Applications Lei Ren, PhD, DABR Department of Radiation Oncology Duke University Medical Center Disclosure I have received research funding from NIH and Varian Medical
More informationSpiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004
Spiral ASSR Std p = 1.0 Spiral EPBP Std p = 1.0 Kachelrieß et al., Med. Phys. 31(6): 1623-1641, 2004 256 slices (0/300) Advantages of Cone-Beam Spiral CT Image quality nearly independent of pitch Increase
More informationImage Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD
Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image
More informationFinancial disclosure. Onboard imaging modality for IGRT
Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty
More informationDevelopment of a multi-axis X-ray CT for highly accurate inspection of electronic devices
Development of a multi-axis X-ray CT for highly accurate inspection of electronic devices Toru Kano 1, Michihiko Koseki 2 More info about this article: http://www.ndt.net/?id=20843 1 Tokyo University of
More informationDesign and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon
Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization
More informationMeasurement of inter and intra fraction organ motion in radiotherapy using cone-beam CT projection images
Measurement of inter and intra fraction organ motion in radiotherapy using cone-beam CT projection images T E Marchant, A M Amer 1 and C J Moore North Western Medical Physics, Christie Hospital NHS Foundation
More informationImage Guidance and Beam Level Imaging in Digital Linacs
Image Guidance and Beam Level Imaging in Digital Linacs Ruijiang Li, Ph.D. Department of Radiation Oncology Stanford University School of Medicine 2014 AAPM Therapy Educational Course Disclosure Research
More informationEmpirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography
Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Marc Kachelrieß, a Katia Sourbelle, and Willi A. Kalender Institute of Medical Physics, University of
More information8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition
Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare
More informationBrilliance CT Big Bore.
1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.
More informationA Curvelet based Sinogram Correction Method for Metal Artifact Reduction
A based Sinogram Correction Method for Metal Artifact Reduction Kiwan Jeon 1 and Hyoung Suk Park 1 More info about this article: http://www.ndt.net/?id=3715 1 National Institute for Mathematical Sciences,
More informationMulti-slice CT Image Reconstruction Jiang Hsieh, Ph.D.
Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced
More informationRespiratory Motion Compensation for C-arm CT Liver Imaging
Respiratory Motion Compensation for C-arm CT Liver Imaging Aline Sindel 1, Marco Bögel 1,2, Andreas Maier 1,2, Rebecca Fahrig 3, Joachim Hornegger 1,2, Arnd Dörfler 4 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg
More informationComputer-Tomography II: Image reconstruction and applications
Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de
More informationRespiratory Motion Estimation using a 3D Diaphragm Model
Respiratory Motion Estimation using a 3D Diaphragm Model Marco Bögel 1,2, Christian Riess 1,2, Andreas Maier 1, Joachim Hornegger 1, Rebecca Fahrig 2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg 2
More informationEstimating 3D Respiratory Motion from Orbiting Views
Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng, Jeffrey A. Fessler, James M. Balter The University of Michigan Oct. 2005 Funding provided by NIH Grant P01 CA59827 Motivation Free-breathing
More informationTumor motion during liver SBRT
Tumor motion during liver SBRT - projects at Aarhus University Hospital - Per Poulsen, Esben Worm, Walther Fledelius, Morten Høyer Aarhus University Hospital, Denmark SBRT: Stereotactic Body Radiation
More information7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT
CLINICAL COMMISSIONING AND ACCEPTANCE TESTING OF A 3D SURFACE MATCHING SYSTEM Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago Learning Objective Describe acceptance
More informationArtifacts in Spiral X-ray CT Scanners: Problems and Solutions
Artifacts in Spiral X-ray CT Scanners: Problems and Solutions Mehran Yazdi, and Luc Beaulieu International Science Index, Electrical and Computer Engineering waset.org/publication/12966 Abstract Artifact
More informationDEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE
Rajesh et al. : Proceedings of the National Seminar & Exhibition on Non-Destructive Evaluation DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh V Acharya, Umesh Kumar, Gursharan
More informationImprovement and Evaluation of a Time-of-Flight-based Patient Positioning System
Improvement and Evaluation of a Time-of-Flight-based Patient Positioning System Simon Placht, Christian Schaller, Michael Balda, André Adelt, Christian Ulrich, Joachim Hornegger Pattern Recognition Lab,
More informationComputed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA
Computed Tomography Principles, Design, Artifacts, and Recent Advances THIRD EDITION Jiang Hsieh SPIE PRESS Bellingham, Washington USA Table of Contents Preface Nomenclature and Abbreviations xi xv 1 Introduction
More informationNew Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC
New Technology in Radiation Oncology James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC Technology s s everywhere From the imaging chain To the planning system To the linac To QA..it..it
More informationA new calibration-free beam hardening reduction method for industrial CT
A new calibration-free beam hardening reduction method for industrial CT ECC 2 for industrial CT Tobias Würfl 1, Nicole Maaß 2, Frank Dennerlein 2, Andreas K. Maier 1 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg;
More informationConvolution-Based Truncation Correction for C-Arm CT using Scattered Radiation
Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation Bastian Bier 1, Chris Schwemmer 1,2, Andreas Maier 1,3, Hannes G. Hofmann 1, Yan Xia 1, Joachim Hornegger 1,2, Tobias Struffert
More informationRetrospective Spiral Respiratory Correlated Imaging with Varian RPM
Retrospective Spiral Respiratory Correlated Imaging with Varian RPM This is a Quick Step Guide for Retrospective Spiral Respiratory Correlated Imaging (4D CT) using the Varian RPM device v1.7 with the
More informationHelical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice Guido Hilgers, a Tonnis Nuver, André Minken Department
More informationPurely Data-Driven Respiratory Motion Compensation Methods for 4D-CBCT Image Registration and Reconstruction
2015 ANNUAL MAC-AAPM CONFERENCE: Purely Data-Driven Respiratory Motion Compensation Methods for 4D-CBCT Image Registration and Reconstruction M J Riblett 1, E Weiss 1, G E Christensen 2, and G D Hugo 1
More information8/3/2016. Image Guidance Technologies. Introduction. Outline
8/3/26 Session: Image Guidance Technologies and Management Strategies Image Guidance Technologies Jenghwa Chang, Ph.D.,2 Department of Radiation Medicine, Northwell Health 2 Hofstra Northwell School of
More informationDUAL energy CT (DECT) is a modality where one and. Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography
Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography Marc Kachelrieß, Member, IEEE, Timo Berkus, Philip Stenner, Willi A. Kalender Abstract Material selective imaging using dual energy
More informationSpatial-temporal Total Variation Regularization (STTVR) for 4D-CT Reconstruction
Spatial-temporal Total Variation Regularization (STTVR) for 4D-CT Reconstruction Haibo Wu a, b, Andreas Maier a, Rebecca Fahrig c, and Joachim Hornegger a, b a Pattern Recognition Lab (LME), Department
More informationDUE to beam polychromacity in CT and the energy dependence
1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact
More informationAn investigation of temporal resolution parameters in cine-mode four-dimensional computed tomography acquisition
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 An investigation of temporal resolution parameters in cine-mode four-dimensional computed tomography acquisition Yildirim D. Mutaf
More informationDigital Image Processing
Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture
More informationFast iterative beam hardening correction based on frequency splitting in computed tomography
Fast iterative beam hardening correction based on frequency splitting in computed tomography Qiao Yang a,b, Matthias Elter b, Ingo Schasiepen b, Nicole Maass b and Joachim Hornegger a,c a Pattern Recognition
More informationManifold Learning-based Data Sampling for Model Training
Manifold Learning-based Data Sampling for Model Training Shuqing Chen 1, Sabrina Dorn 2, Michael Lell 3, Marc Kachelrieß 2,Andreas Maier 1 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg 2 German Cancer
More informationProjection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT
Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Benedikt Lorch 1, Martin Berger 1,2, Joachim Hornegger 1,2, Andreas Maier 1,2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg
More informationShadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS
Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic
More informationReduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy
Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy T. Rohlfing a, D. Zerfowski b, J. Beier a, P. Wust a, N. Hosten a, R. Felix a a Department of
More informationScatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University
Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid by Yingxuan Chen Graduate Program in Medical Physics Duke University Date: Approved: Lei Ren, Supervisor Fang Fang Yin, Chair
More informationMotion artifact detection in four-dimensional computed tomography images
Motion artifact detection in four-dimensional computed tomography images G Bouilhol 1,, M Ayadi, R Pinho, S Rit 1, and D Sarrut 1, 1 University of Lyon, CREATIS; CNRS UMR 5; Inserm U144; INSA-Lyon; University
More informationHIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY
HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY David K. Lehmann 1, Kathleen Brockdorf 1 and Dirk Neuber 2 1 phoenix x-ray Systems + Services Inc. St. Petersburg, FL, USA 2 phoenix x-ray Systems + Services
More informationView Aliasing Artifacts Reduction Method Based on 4D Cone-Beam CT Reconstruction with Joint Projection Data
June 2017, Xi'an 1 View Aliasing Artifacts Reduction Method Based on 4D Cone-Beam CT Reconstruction with Joint Projection Data Shaohua Zhi, Xuanqin Mou Abstract Although the quality of the phase-resolved
More informationAn approximate cone beam reconstruction algorithm for gantry-tilted CT
An approximate cone beam reconstruction algorithm for gantry-tilted CT Ming Yan a, Cishen Zhang ab, Hongzhu Liang a a School of Electrical & Electronic Engineering, Nanyang Technological University, Singapore;
More informationReduction of motion artefacts in on-board cone beam CT by warping of projection images
Reduction of motion artefacts in on-board cone beam CT by warping of projection images T E Marchant 1, G J Price 1, B J Matuszewski 2 and C J Moore 1. 1 North Western Medical Physics, The Christie NHS
More informationImage-based Compensation for Involuntary Motion in Weight-bearing C-arm CBCT Scanning of Knees
Image-based Compensation for Involuntary Motion in Weight-bearing C-arm CBCT Scanning of Knees Mathias Unberath, Jang-Hwan Choi, Martin Berger, Andreas Maier, Rebecca Fahrig February, 24. 2015 Pattern
More informationDigital phantoms for the evaluation of a software used for an automatic analysis of the Winston-Lutz test in image guided radiation therapy
Author manuscript, published in "Medical Imaging 008: Physics of Medical Imaging, San Diego, CA, USA : United States (008)" DOI : 10.1117/1.768668 Digital phantoms for the evaluation of a software used
More informationTomographic Reconstruction
Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)
More informationML reconstruction for CT
ML reconstruction for CT derivation of MLTR rigid motion correction resolution modeling polychromatic ML model dual energy ML model Bruno De Man, Katrien Van Slambrouck, Maarten Depypere, Frederik Maes,
More informationMutual information based CT registration of the lung at exhale and inhale breathing states using thin-plate splines
Mutual information based CT registration of the lung at exhale and inhale breathing states using thin-plate splines Martha M. Coselmon, a) James M. Balter, Daniel L. McShan, and Marc L. Kessler Department
More informationLag correction model and ghosting analysis for an indirect-conversion flat-panel imager
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 8, NUMBER 3, SUMMER 2007 Lag correction model and ghosting analysis for an indirect-conversion flat-panel imager Noor Mail, Peter O Brien, and Geordi
More informationMulti-spectral (W, Mo, Cu, and Ag) XCT measurements
Multi-spectral (W, Mo, Cu, and Ag) XCT measurements Marko Katić 1, Damir Markučič 1, Gorana Baršić 1 Faculty of mechanical engineering and naval architecture, Ivana Lucica 5, Zagreb, Croatia, marko.katic@fsb.hr
More informationMetal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland
Metal Artifact Reduction CT Techniques R S S S Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland N. 1 v o 4 1 0 2. Postoperative CT Metal Implants CT is accurate for assessment
More informationSimple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 Simple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field Hideharu Miura, 1a Shuichi
More informationFlying Focal Spot (FFS) in Cone-Beam CT Marc Kachelrieß, Member, IEEE, Michael Knaup, Christian Penßel, and Willi A. Kalender
1238 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 3, JUNE 2006 Flying Focal Spot (FFS) in Cone-Beam CT Marc Kachelrieß, Member, IEEE, Michael Knaup, Christian Penßel, and Willi A. Kalender Abstract
More informationFour-Dimensional MVCT Imaging and Spatial Translation of Helical Tomotherapy Dose Distributions Using Sinogram Modification
University of Tennessee, Knoxville Trace: Tennessee Research and Creative Exchange Masters Theses Graduate School 12-2006 Four-Dimensional MVCT Imaging and Spatial Translation of Helical Tomotherapy Dose
More informationEnhancement Image Quality of CT Using Single Slice Spiral Technique
Enhancement Image Quality of CT Using Single Slice Spiral Technique Doaa. N. Al Sheack 1 and Dr.Mohammed H. Ali Al Hayani 2 1 2 Electronic and Communications Engineering Department College of Engineering,
More informationTomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram
TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization
More informationComputed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system
Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system ECNDT 2014 Prague October 6-10, 2014 Dr. Eberhard Neuser Dr. Alexander Suppes Imagination
More informationComparison of Scatter Correction Methods for CBCT. Author(s): Suri, Roland E.; Virshup, Gary; Kaissl, Wolfgang; Zurkirchen, Luis
Research Collection Working Paper Comparison of Scatter Correction Methods for CBCT Author(s): Suri, Roland E.; Virshup, Gary; Kaissl, Wolfgang; Zurkirchen, Luis Publication Date: 2010 Permanent Link:
More informationCombination of Markerless Surrogates for Motion Estimation in Radiation Therapy
Combination of Markerless Surrogates for Motion Estimation in Radiation Therapy CARS 2016 T. Geimer, M. Unberath, O. Taubmann, C. Bert, A. Maier June 24, 2016 Pattern Recognition Lab (CS 5) FAU Erlangen-Nu
More informationAnalysis of Acquisition Parameters That Caused Artifacts in Four-dimensional (4D) CT Images of Targets Undergoing Regular Motion
Original Article PROGRESS in MEDICAL PHYSICS Vol. 24, No. 4, December, 2013 http://dx.doi.org/10.14316/pmp.2013.24.4.243 Analysis of Acquisition Parameters That Caused Artifacts in Four-dimensional (4D)
More informationMetal Streak Artifacts in X-ray Computed Tomography: A Simulation Study
Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study B. De Man, J. Nuyts, P. Dupont, G. Marchal and P. Suetens Medical Image Computing, ESAT-PSI, K.U.Leuven, B-3000 Leuven, Belgium Department
More informationMetal artifact reduction in CT using fusion based prior image.
Metal artifact reduction in CT using fusion based prior image. Jun Wang, Shijie Wang, Yang Chen, Jiasong Wu, Jean-Louis Coatrieux, Limin Luo To cite this version: Jun Wang, Shijie Wang, Yang Chen, Jiasong
More informationAutomated Quality Assurance for Image-Guided Radiation Therapy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 1, WINTER 2009 Automated Quality Assurance for Image-Guided Radiation Therapy Eduard Schreibmann, a Eric Elder, Tim Fox Department of Radiation
More informationPatient Set-ups and Tumor Localizations
Patient Set-ups and Tumor Localizations Amy S. Harrison Patient Positioning Prior to starting any localization or simulation procedure patients need to be positioned and immobilized Patients disease location
More informationRefinement of Imaging Processing of Scatter Correction and Beam Hardening Tools for Industrial Radiography and Cone Beam CT
Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18073 Refinement of Imaging Processing
More informationA new Concept for High-Speed atline and inlinect for up to 100% Mass Production Process Control
18th World Conference on Nondestructive Testing, 16-20 April 2012, Durban, South Africa A new Concept for High-Speed atline and inlinect for up to 100% Mass Production Process Control Oliver BRUNKE 1,
More informationAnalysis of 4D CT cine images for the characterization of organ motion due to breathing
-243- Analysis of 4D CT cine images for the characterization of organ motion due to breathing Maria Francesca Spadea 1,2, Marta Peroni 2, Marco Riboldi 2, Guido Baroni 2, George TY Chen 3, Gregory Sharp
More informationImad Ali 1 Nesreen Alsbou 2 Justin Jaskowiak 1 Salahuddin Ahmad 1. Abstract RADIATION ONCOLOGY PHYSICS
Received: 9 May 2017 Revised: 2 October 2017 Accepted: 21 November 2017 DOI: 10.1002/acm2.12246 RADIATION ONCOLOGY PHYSICS Quantitative evaluation of the performance of different deformable image registration
More informationComputer-Tomography I: Principles, History, Technology
Computer-Tomography I: Principles, History, Technology Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de History
More informationScaling Calibration in the ATRACT Algorithm
Scaling Calibration in the ATRACT Algorithm Yan Xia 1, Andreas Maier 1, Frank Dennerlein 2, Hannes G. Hofmann 1, Joachim Hornegger 1,3 1 Pattern Recognition Lab (LME), Friedrich-Alexander-University Erlangen-Nuremberg,
More informationCBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona
CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona Introduction CBCT has become a routine procedure for
More informationDUAL-ENERGY CT IN PROTON THERAPY
10/31/17 DUAL-ENERGY CT IN PROTON THERAPY Isabel Almeida, MAASTRO Clinic 7th NCS Lustrum Symposium 1 10/31/17 http://zonptc.bouwwebcam.nl https://www.youtube.com/watch?v=3vvvf5bqn7g Range uncertainties
More informationAdvanced single slice rebinning in cone beam spiral CT: Theoretical considerations and medical applications
Advanced single slice rebinning in cone beam spiral CT: Theoretical considerations and medical applications Marc Kachelrieß a, Stefan Schaller b and Willi A. Kalender a a Institute of Medical Physics,
More informationVirtual Phantoms for IGRT QA
TM Virtual Phantoms for IGRT QA Why ImSimQA? ImSimQA was developed to overcome the limitations of physical phantoms for testing modern medical imaging and radiation therapy software systems, when there
More informationMetal artifacts reduction in x-ray CT based on segmentation and forward-projection
Bio-Medical Materials and Engineering 24 (2014) 3287 3293 DOI 10.3233/BME-141151 IOS Press 3287 Metal artifacts reduction in x-ray CT based on segmentation and forward-projection Shoukat Nawaz, Jian Fu
More informationReal Time Tumor Motion Tracking with CyberKnife
Real Time Tumor Motion Tracking with CyberKnife Martina Descovich, Ph.D University of California San Francisco July 16, 2015 Learning objectives Review the principles of real-time tumor motion tracking
More information2005 IEEE Nuclear Science Symposium Conference Record M10-2
25 IEEE Nuclear Science Symposium Conference Record M1-2 Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng 1, Jeffrey A. Fessler 1, and James M. Balter 2 rzeng@eecs.umich.edu, fessler@eecs.umich.edu,
More informationSuitability of a new alignment correction method for industrial CT
Suitability of a new alignment correction method for industrial CT Matthias Elter 1, Nicole Maass 1, Peter Koch 2 1 Siemens AG, Healthcare Sector, Erlangen, Germany, e-mail: matthias.elter@siemens.com,
More informationKey words: Cone-beam CT CBCT, cardiac imaging, 4-D reconstruction, image quality
Extended parallel backprojection for standard three-dimensional and phase-correlated four-dimensional axial and spiral cone-beam CT with arbitrary pitch, arbitrary cone-angle, and 100% dose usage Marc
More informationGPU-based Fast Cone Beam CT Reconstruction from Undersampled and Noisy Projection Data via Total Variation
GPU-based Fast Cone Beam CT Reconstruction from Undersampled and Noisy Projection Data via Total Variation 5 10 15 20 25 30 35 Xun Jia Department of Radiation Oncology, University of California San Diego,
More informationClinical Evaluation of GPU-Based Cone Beam Computed Tomography.
Clinical Evaluation of GPU-Based Cone Beam Computed Tomography. Peter B. Noël 1,2, Alan M. Walczak 2, Kenneth R. Hoffmann 1,2, Jinhui Xu 1, Jason J. Corso 1, and Sebastian Schafer 2 1 Department of Computer
More informationAn Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy
An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy Chenyang Xu 1, Siemens Corporate Research, Inc., Princeton, NJ, USA Xiaolei Huang,
More information3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH
3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk
More informationSeparate CT-Reconstruction for Orientation and Position Adaptive Wavelet Denoising
Separate CT-Reconstruction for Orientation and Position Adaptive Wavelet Denoising Anja Borsdorf 1,, Rainer Raupach, Joachim Hornegger 1 1 Chair for Pattern Recognition, Friedrich-Alexander-University
More informationUvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication
UvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication Citation for published version (APA): Kruis, M. F. (2014). Motion compensation for 4D PET/CT General
More information