Standard Operating Procedure Inhouse Solar Simulator

Size: px
Start display at page:

Download "Standard Operating Procedure Inhouse Solar Simulator"

Transcription

1 Standard Operating Procedure Inhouse Solar Simulator By Sanchit Khatavkar and Guru Burkul National Centre for Photovoltaic Research and Education (NCPRE), Indian Institute of Technology-Bombay, Powai, Mumbai INDIA Contamination Category: Litho/Analytical Safety Precautions:

2 Do not look directly onto the light source. If under any circumstances you need to look in the light, please wear the UV protection goggles placed near the system. Do not bring or place any liquid especially water near the system. Please take all necessary cautions needed when working with electrical equipments. Operating Instructions: TURN ON Sequence: 1. Check whether the power supply cable is plugged in the switchboard and then turn the switchboard ON. 2. Turn the knob to 75% (Upto the mark on the power supply). 3. Only after this is done, Press the switch at the back which will power ON both the lamp and the fan. 4. Make sure that the units knob is set to Watts. 5. Wait for 5 mins, you will see the units increasing. 6. After 5 minutes, very gently, slightly rotate the knob so that the power supply unit shows a stable 148 Watts on the display before proceeding with the experiment. 7. This completes the lamp turn ON procedure. 8. Switch ON the power supply to the computer, Keithley 2400 SMU and the Vacuum pump. 9. Switch ON the Keithley 2400 SMU by pressing the Power switch located on the front panel. 10. Place your sample onto to the chuck. 11. For organic solar cells, place the 4.3cm chuck below the light source such that the sample is exactly at the center of the light beam (Dia - 4 cm). Connect the leads of the chuck to the Keithley SMU. 12. For silicon solar cells, place the vacuum chuck below the light source. i) Place the sample onto the chuck and turn ON the vacuum pump which is located below the table. ii) Ensure that the sample is at the center of the light beam. iii) Make front and back contacts using micromanipulator and connect the leads to Keithley SMU. 13. Switch ON the PC and click icon Lab Tracer You will be getting a screen as shown below:

3 15. Check the settings by clicking the SETUP 2401 link before performing the IV sweep. You will get the window shown below. 16. Click on the Source tab. Enter the start and stop values of voltage according to the measurement (dark and illuminated) in the Source tab as shown in the image above. More sweep delay is useful only for devices like organic solar cells. 17. After that clicking on Advanced tab, make the four wire measurement as shown below

4 18. After making sure the Settings are correct, click Run Test to start the sweep. 19. By default, the I-V plot will be displayed at the end of the sweep, if not, then click on the Graph tab. You can save the graph if needed. To save the data, click on the Datasheet tab. On this tab, there is a Save data button. On clicking that, you will be asked for the filename against which you want to save the data. Data can be saved in.txt format. Please save data in a directory having your name in E:\. Please do not save data on the desktop. 20. To do another measurement, if you are using Keithley 2401 Sourcemeter, goto File -> Exit as shown below. This will exit you from Labtracer software.

5 21. If you are using Keithley 2400 sourcemeter, then click on the Close datacenter button. You will exit from Datacenter window only. 22. The GPIB connects the Labtracer setup to the Keithley 2400 SMU. 23. On clicking Run Test, the Keithley 2400 SMU should start the sweep. If this does not happen, refer to Annexure I for troubleshooting. 24. To Measure DARK IV, please close the shutter that blocks illumination and do the measurement. 25. During dark I-V measurement Si solar cell please keep the compliance 100 ma. TURN OFF Sequence: 26. Turn the knob slowly to zero. 27. Wait for 15mins for the lamp to cool down. 28. Turn OFF the lamp and fan switch at the back. 29. Turn off the computer and the Keithley Sourcemeter. 30. Turn OFF the switchboard. 31. For Solar cells with top and back metal contacts like Silicon Solar cells, use the probing manipulator to probe the top contact and the vacuum chuck will act as the back contact as shown below.

6 For Solar Cells with both contacts on top and light falling on the top, use two probing manipulators to probe both the contacts. Annexure I Troubleshooting for Keithley 2400 issue. 1. Click Menu. In the Main Menu enter Communication. 2. In Communications setup, click GPIB. 3. Make sure that the Address is seen as 24. If 24 is seen blinking, then it is OK. 4. Press ENTER and you will get GPIB Protocol. If is blinking, then it is OK. 5. Press EXIT to come out of the Main Menu

Simulator System SUN ABET

Simulator System SUN ABET SoP of I V Simulator System Part 2 SUN 3000 SOLAR SIMULATOR ABET Technologies Standard operating procedure: Flowchart of sequence to be followed: System switch ON sequence Stage and device probing Measuring

More information

Operating Procedure PerkinElmer LAMBDA 950 (Integrating sphere Mode)

Operating Procedure PerkinElmer LAMBDA 950 (Integrating sphere Mode) Operating Procedure PerkinElmer LAMBDA 950 (Integrating sphere Mode) National Centre for Photovoltaic Research and Education (NCPRE) Indian Institute of Technology- Bombay Powai, Mumbai-400076 INDIA Operating

More information

Title: Heidelberg DWL66+ Semiconductor & Microsystems Fabrication Laboratory Revision: B Rev Date: 05/03/2017

Title: Heidelberg DWL66+ Semiconductor & Microsystems Fabrication Laboratory Revision: B Rev Date: 05/03/2017 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Heidelberg DWL66+. All users are expected to have read and understood this document.

More information

Series 370 S SoftwSystem are Qui Switch/Mult ck Start G imeter uide Quick Start Guide

Series 370 S SoftwSystem are Qui Switch/Mult ck Start G imeter uide Quick Start Guide Series ACS 3700A Software System Quick Switch/Multimeter Start Guide Quick Start Guide Safety precautions Observe the following safety precautions before using this product and any associated instrumentation.

More information

Series 2600A System SourceMeter

Series 2600A System SourceMeter www.keithley.com Series 2600A System SourceMeter Quick Start Guide 2600AS-903-01 Rev. A / September 2008 A G R E A T E R M E A S U R E O F C O N F I D E N C E WARRANTY Keithley Instruments, Inc. warrants

More information

Four-Point Probe System

Four-Point Probe System Manual version: 2.0.E Product code: T2001A2 Product version: 2.0 Software version: 1.1 enabling materials science ossila.com Contents 1. Safety... 3 1.1 Warning... 3 1.2 Caution... 3 2. Introduction...

More information

FluxGage. FluxGage. LED Luminaire Measurement System User Manual

FluxGage. FluxGage. LED Luminaire Measurement System User Manual FluxGage FluxGage LED Luminaire Measurement System User Manual 1 Acronyms... 3 2 Introduction... 4 2.1 Operation principle... 4 3 Specifications... 5 4 Mechanical and Electrical Installation... 9 4.1 Unpacking...

More information

2.2. Facilities Requirements

2.2. Facilities Requirements 2.2. Facilities Requirements Facilities requirements for the alpha-se system are listed in Table 2-1 and the system dimensions are given in Figure 2-1. As shown in Figure 2-2, the preferred clear work

More information

Models 2601B, 2602B, and 2604B System SourceMeter Instruments Quick Start Guide

Models 2601B, 2602B, and 2604B System SourceMeter Instruments Quick Start Guide Models 2601B, 2602B, and 2604B System SourceMeter Instruments Quick Start Guide Safety precautions Observe the following safety precautions before using this product and any associated instrumentation.

More information

FC 1800 #1 Operation Manual University of Notre Dame

FC 1800 #1 Operation Manual University of Notre Dame \ FC 1800 #1 Operation Manual University of Notre Dame For Training Please contact: Dave Heemstra 8 June 2015 Page 1 of 12 General Precautions Contacts For problems, clarification of procedures, or general

More information

Condensed AFM operating instructions:

Condensed AFM operating instructions: Condensed AFM operating instructions: 1. Log onto system at access controller 2. Take the parts you need to mount a probe out of the drawers. You need the appropriate probe holder, tweezers (these are

More information

RTT98 VINTAGE TURNTABLE

RTT98 VINTAGE TURNTABLE RTT98 VINTAGE TURNTABLE WARNING: TO PREVENT FIRE OR SHOCK HAZARD, DO NOT EXPOSE THIS APPLIANCE TO RAIN OR MOISTURE. DO NOT REMOVE COVER. PILOT LAMPS SOLDERED IN PLACE. NO USER SERVICEABLE PARTS INSIDE.

More information

Keithley Instruments Customer Documentation Series 2600A UL Safety Supplement

Keithley Instruments Customer Documentation Series 2600A UL Safety Supplement Keithley Instruments, Inc. 28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY www.keithley.com Keithley Instruments Customer Documentation Series 2600A UL Safety Supplement 1. Introduction The Keithley

More information

PANalytical X-Ray Diffractometer with PW3830 Generator. Standard Operating Procedure. Revised on 02/15/2007

PANalytical X-Ray Diffractometer with PW3830 Generator. Standard Operating Procedure. Revised on 02/15/2007 PANalytical X-Ray Diffractometer with PW3830 Generator Standard Operating Procedure Revised on 02/15/2007 Booking of XRD time 1. Please book the XRD schedule first come/first serve in advance on the sheets

More information

Prism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis

Prism Starter Guide 1.0 Hoskins Lab Last Modified 03/14/2017 Chris DeCiantis Start Up: Upon entering the laser room turn on the wall mounted Laser Power Button by pulling it away from the wall. Turn on Shutter controllers (toggle switch on back of unit). There should be a U in

More information

Operating Procedure for Horiba Raman Microscope

Operating Procedure for Horiba Raman Microscope Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch

More information

K2400 Electroplating program Users Manuals - updated Aug

K2400 Electroplating program Users Manuals - updated Aug K2400 Electroplating program Users Manuals - updated Aug 25 2011 Introduction: Thank you for purchasing BTInnoations K2400 Electroplating program. This program is designed to be used with a Keithley 2400

More information

IMPORTANT SAFETY INSTRUCTIONS

IMPORTANT SAFETY INSTRUCTIONS IMPORTANT SAFETY INSTRUCTIONS When using this electronic device, basic precautions should always be taken, including the following: 1. Read all instructions before using the product. 2. Do not use this

More information

Model 2600B-PM V Protection Module with 1 A Clamp. Description / April 2015 *PPA * 1

Model 2600B-PM V Protection Module with 1 A Clamp. Description / April 2015 *PPA * 1 Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.keithley.com Model 2600B-PM-1 200 V Protection Module with 1 A Clamp Description The Model 2600B-PM-1 200 V Protection

More information

Automation of wafer handling

Automation of wafer handling Project Report Automation of wafer handling ELECTRONIC DESIGN LAB DEPARTMENT OF ELECTRICAL ENGINEERING IIT BOMBAY Group Members: SHREYANS GANDHI 06D07005 ISHAN GAIKWAD 06D07013 PRADEEP GAIKWAD 06D07010

More information

Installation and Getting Started Guide

Installation and Getting Started Guide Installation and Getting Started Guide Metrics ICV Version 4.1.0 Copyright 1999-2015, Metrics Technology, Inc. All rights reserved. Table of Contents Installing the ICV Software... 4 Obtain a Codeword...

More information

TIF785 Photo Tachometer

TIF785 Photo Tachometer TIF785 Photo Tachometer Owner s Manual 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Safety Precautions CAUTION: This is a Class

More information

UNIVERSITY OF CENTRAL FLORIDA DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE HEC 406 NANO/MEMS LABORATORY

UNIVERSITY OF CENTRAL FLORIDA DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE HEC 406 NANO/MEMS LABORATORY UNIVERSITY OF CENTRAL FLORIDA DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE HEC 406 NANO/MEMS LABORATORY HIGH TEMPERATURE & HIGH FREQUENCY TESTING SYSTEM LABORATORY USER MANUAL 4/29/2011 Overview

More information

Title: Karl Suss MA150 Semiconductor & Microsystems Fabrication Laboratory Revision: E Rev Date: 03/01/06

Title: Karl Suss MA150 Semiconductor & Microsystems Fabrication Laboratory Revision: E Rev Date: 03/01/06 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Karl Suss MA150 Mask Aligner. All users are expected to have read and understood

More information

Photo Emission Tech., Inc. 760 Calle Plano, Camarillo, CA USA Tel: (805) Fax: (805)

Photo Emission Tech., Inc. 760 Calle Plano, Camarillo, CA USA Tel: (805) Fax: (805) IQE1800 SYSTEM INTRODUCTION Quantum Efficiency (QE) measurements provide a primary characterization method for completed photovoltaic devices. QE system provides electronics and software designed for fully

More information

KIP Fold 1000 Offline User Manual

KIP Fold 1000 Offline User Manual KIP Fold 1000 Offline User Manual Safety Instructions For your protection, please read these safety instructions and the manual completely before operating the machine and keep this manual for future reference

More information

Model: VA/230W UPS Backup System

Model: VA/230W UPS Backup System IMPORTANT SAFETY INSTRUCTIONS (SAVE THESE INSTRUCTIONS) This manual contains important safety instructions. Please read and follow all instructions carefully during installation and operation of unit.

More information

Bent Pins Inside the Sample Chamber

Bent Pins Inside the Sample Chamber PPMS Service Note 1070-311 Bent Pins Inside the Sample Chamber It has come to the attention of Quantum Design that there has been an increase in the instances of sample chambers with bent pins. We would

More information

Equipment Operating Procedure. Model Number: Dage-Series-4000PA Machine Name: Shear Tester Date: 8/17/2013 Serial Number:

Equipment Operating Procedure. Model Number: Dage-Series-4000PA Machine Name: Shear Tester Date: 8/17/2013 Serial Number: Equipment Operating Procedure Manufacturer: Dage Precision Location: PRC Assembly Laboratory Industries Model Number: Dage-Series-4000PA Machine Name: Shear Tester Date: 8/17/2013 Serial Number: 1922708

More information

Raman Lab - Laser 1 operation

Raman Lab - Laser 1 operation Raman Lab - Laser 1 operation Cover limit switches cooling water supply optimate panel light control laser remote control sample stage Remote Shutter Control laser power supply WARNING! Never ever try

More information

Model 2657A-LIM-3 LO Interconnect Module

Model 2657A-LIM-3 LO Interconnect Module Keithley Instruments, Inc. 28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY http://www.keithley.com Model 2657A-LIM-3 LO Interconnect Module User's Guide Description The Model 2657A-LIM-3 LO Interconnect

More information

Short instructions. Olympus FV1000

Short instructions. Olympus FV1000 Olympus FV1000-1- Short instructions Short instructions. Olympus FV1000 Version 140416 This short instruction will guide you through the turn on and off sequence of the microscope system. For a detailed

More information

IMAGING GROUP x 1300 imaging array; 20mm x 20mm pixels Scientific grade CCD All-metal, hermetic vacuum seals; Lifetime vacuum guarantee

IMAGING GROUP x 1300 imaging array; 20mm x 20mm pixels Scientific grade CCD All-metal, hermetic vacuum seals; Lifetime vacuum guarantee Now Powered by LightField PIXIS: 13 The PIXIS: 13 series of cameras from Princeton Instruments (PI) are fully integrated, low noise cameras designed expressly for quantitative scientific imaging applications.

More information

Startup Software Quick Start Guide

Startup Software Quick Start Guide Startup Software Quick Start Guide Safety precautions Observe the following safety precautions before using this product and any associated instrumentation. Although some instruments and accessories would

More information

HIGH TEMPERATURE PROBE STATION STANDARD OPERATION PROCEDURE

HIGH TEMPERATURE PROBE STATION STANDARD OPERATION PROCEDURE Arizona State University NanoFab HIGH TEMPERATURE PROBE STATION STANDARD OPERATION PROCEDURE Rev A Table of Contents Contents Table of Contents... 1 1. Purpose / Scope... 2 2. Reference Documents... 2

More information

TMA Standard Operating Procedure [Updated April 30, 2015]

TMA Standard Operating Procedure [Updated April 30, 2015] Thermomechanical Analysis Q400 Location of Machine: Composites Lab, RFM 1218 Location of SOP and Machine Operating & Safety Manual: Composites Lab website under resources; Composites Lab TRACS site; and

More information

Lam AutoEtch 490 Plasma Etch System Operations and Maintenance Manual

Lam AutoEtch 490 Plasma Etch System Operations and Maintenance Manual Fabrication Laboratory Revision: E - 02/10/2009 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Lam AutoEtch 490 Plasma Etch System

More information

Computer. Test Fixture SPA. Key board

Computer. Test Fixture SPA. Key board Semiconductor Parameter Analyzer (Operating instructions) A PDF copy of this document should be available on the Windows Desktop of the SPA computer Read this document fully before attempting to operate

More information

Model 2460-KIT. Screw Terminal Connector Kit. Description / September 2014 *P * 1

Model 2460-KIT. Screw Terminal Connector Kit. Description / September 2014 *P * 1 Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.keithley.com Model 2460-KIT Screw Terminal Connector Kit Description The Model 2460-KIT Screw Terminal Connector Kit

More information

CVU-3K-KIT. 3 kv Bias Tee Kit. Description. Parts list / October 2014 *P * 1

CVU-3K-KIT. 3 kv Bias Tee Kit. Description. Parts list / October 2014 *P * 1 Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.keithley.com CVU-3K-KIT 3 kv Bias Tee Kit Description The CVU-3K-KIT Bias Tee Kit consists of three bias tees for

More information

SOLAR LIGHTLINE SERIES

SOLAR LIGHTLINE SERIES Features Meets Class AAA ASTM standards Fiberized output for flexible illumination. Easy-to-use touchscreen controls Light intensity control is available Remote control by PC Cost-effective for researchers

More information

Model 8020-KHV. Kelvin Keithley Triaxial Connector Card. Description / October 2014 *P * 1

Model 8020-KHV. Kelvin Keithley Triaxial Connector Card. Description / October 2014 *P * 1 Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.keithley.com Model 8020-KHV Kelvin Keithley Triaxial Connector Card Description The Model 8020-KHV Keithley HV Connector

More information

Title: Bruce Furnaces Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date: 02/21/06

Title: Bruce Furnaces Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date: 02/21/06 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Bruce Furnaces. All users are expected to have read and understood this document.

More information

IMAGING GROUP x 2048 imaging array; 13.5mm x 13.5mm pixels Scientific grade CCD All-metal, hermetic vacuum seals; Lifetime vacuum guarantee

IMAGING GROUP x 2048 imaging array; 13.5mm x 13.5mm pixels Scientific grade CCD All-metal, hermetic vacuum seals; Lifetime vacuum guarantee Now Powered by LightField PIXIS: 248 The PIXIS: 248 series of cameras from Princeton Instruments (PI) are fully integrated, low noise cameras designed expressly for quantitative scientific imaging applications.

More information

Premium Continuous Ink Supply Systems.

Premium Continuous Ink Supply Systems. Premium Continuous Ink Supply Systems. Instruction Manual For Epson Printers R200, R210, R230, R300, R310, RX650 RX510, RX630 157 Union Road, Ascot Vale Victoria 3032 Phone: (+613) 9005 5555 2 Installation

More information

` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE

` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE ` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE Version: 1.0 April-2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research Center (NRC) TABLE OF CONTENTS 1. Introduction...2 1.1

More information

Model kw Pulse Mode Interactive

Model kw Pulse Mode Interactive Model 2461 1 kw Pulse Mode Interactive SourceMeter Instrument User s Manual 2461-900-01 Rev. A / November 2015 Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

More information

Moving-Minds.com Treadmill Desk Operation

Moving-Minds.com Treadmill Desk Operation LifeSpan Treadmill Desk - Owner s Manual Models DT-5/DT-7 Desk Assembly DT-5 Desk........................................................................................ Assembly Instructions..........................................................................

More information

SCHOTT FLM 4 Fiber Lighting Modul. Operating Instructions

SCHOTT FLM 4 Fiber Lighting Modul. Operating Instructions SCHOTT FLM 4 Fiber Lighting Modul Operating Instructions Contents 1. Important information 3 2. Scope of Delivery 3 3. Intended use 3 4. Safety information 4 5. Operation 5 5.1. LED Optics Module 5 5.2.

More information

Standard Operating Procedure of Triboindenter (Hysitron TI 950)

Standard Operating Procedure of Triboindenter (Hysitron TI 950) Standard Operating Procedure of Triboindenter (Hysitron TI 950) I Sample Loading and Preparation DO NOT TOUCH the bottom of transducer and optical microscope. Always place the tall samples on the most

More information

High-Temperature Susceptometer (Kappabridge)

High-Temperature Susceptometer (Kappabridge) High-Temperature Susceptometer (Kappabridge) [Kappa] Room 262 (Kappa) Sensitivity 4 10 8 SI; Frequency: 920 Hz Temperatures: ambient to 1000 K Several hours/sample ( 200 mg chip or powder) for complete

More information

Int883 INSTRUCTION MANUAL. IR-Preheating System BASIC TROUBLESHOOTING GUIDE. 12 Contact the vendor.

Int883 INSTRUCTION MANUAL. IR-Preheating System BASIC TROUBLESHOOTING GUIDE. 12 Contact the vendor. BASIC TROUBLESHOOTING GUIDE PROBLEM 1: THE UNIT HAS NO POWER 1. Check if the unit is switched ON. 2. Check the fuse. Replace with the same type if fuse is blown. 3. Check the power cord and ensure there

More information

4 Instrument Instructions for CE

4 Instrument Instructions for CE 4 Instrument Instructions for CE Overview: the vial trays There are three instrumental steps in the experiment: a) Condition the capillary; b) Set up and, if necessary, optimize your method for a single

More information

PIXIS: 512 IMAGING GROUP

PIXIS: 512 IMAGING GROUP PIXIS: 512 Now Powered with LightField The PIXIS: 512 series of cameras from Princeton Instruments (PI) are fully integrated, low noise cameras designed expressly for quantitative scientific imaging applications.

More information

Model 8020-STC. Kelvin Standard Triaxial Connector Card. Description / October 2014 *P * 1

Model 8020-STC. Kelvin Standard Triaxial Connector Card. Description / October 2014 *P * 1 Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.keithley.com Model 8020-STC Kelvin Standard Triaxial Connector Card Description The Model 8020-STC Kelvin Standard

More information

Contents. - i - Ver.:2

Contents. - i - Ver.:2 Contents 1 Accessories of D-Point 2... 2 2 Connecting Projector to PC... 4 3 Pen Action... 4 4 Driver Installation... 5 5 Configuration Setting... 8 6 D-Point 2 Pull-up Menu... 12 7 DT02 Multi-touch Interactive

More information

Co-Lase Users Guide. DM00001 Rev A (DCN 031) These instructions are protected by copyright. All rights reserved.

Co-Lase Users Guide. DM00001 Rev A (DCN 031) These instructions are protected by copyright. All rights reserved. Co-Lase Users Guide These instructions are protected by copyright. All rights reserved., Inc. http://www.propel-labs.com 131 East Lincoln Avenue, Suite 200 Fort Collins, CO 80524 1 Table of Contents Chapter

More information

The Microscope: A Tool of the Scientist

The Microscope: A Tool of the Scientist Name Class Date Microscope Lab The Microscope: A Tool of the Scientist Background Information (You may refer to page 61 of your textbook.) Problem One of the most important tools of a life scientist is

More information

Instruction Manual Extreme Temperature (XT) Data Logger

Instruction Manual Extreme Temperature (XT) Data Logger Instruction Manual Extreme Temperature (XT) Data Logger 15 Columbia Drive Amherst, NH 03031 USA Phone: (603) 883-3390 Fax: (603) 886-3300 E-mail: support@monarchinstrument.com Website: www.monarchinstrument.com

More information

Line Interactive 1000VA/1400VA/2000VA Uninterruptible Power System

Line Interactive 1000VA/1400VA/2000VA Uninterruptible Power System USER MANUAL Line Interactive 1000VA/1400VA/2000VA Uninterruptible Power System 614-06762-00 IMPORTANT SAFETY INSTRUCTIONS SAVE THESE INSTRUCTIONS This manual contains important instructions for Line Interactive

More information

NVR Security System Quick Start Guide

NVR Security System Quick Start Guide NVR Security System Quick Start Guide 1 HDD Installation If your system does not include a pre-installed HDD, you can purchase your own and follow the steps below for installation. Make sure you purchase

More information

What is an SMU Instrument, and How Do You Decide Which One is Right for Your Application? Bona, Oh

What is an SMU Instrument, and How Do You Decide Which One is Right for Your Application? Bona, Oh What is an SMU Instrument, and How Do You Decide Which One is Right for Your Application? Bona, Oh SMU Instrument Basics SourceMeter SMU Instruments SMUs are precision instruments which are used for sourcing

More information

MT-150. RENOGY 150A Peak High Precision Watt Meter and Power Analyzer E. Philadelphia St., Ontario CA Version: 1.

MT-150. RENOGY 150A Peak High Precision Watt Meter and Power Analyzer E. Philadelphia St., Ontario CA Version: 1. MT-150 RENOGY 150A Peak High Precision Watt Meter and Power Analyzer 0 2775 E. Philadelphia St., Ontario CA 91761 1-800-330-8678 Version: 1.1 Important Safety Instructions Please save these instructions.

More information

kena Instruction Manual T-1050

kena Instruction Manual T-1050 kena Instruction Manual T-1050 2 Table of Contents Introduction / Diagram 4 Package Contents 5 Operational Requirements 5 Optional Accessories to Order 5 Specifications 6 Quick Start Guide 7 1. Install

More information

7 Digital Photo Frame

7 Digital Photo Frame 7 Digital Photo Frame Instruction manual L7DPF17 12 month manufacturer's warranty -------------------------------------------------------------------------- --------------------------------------------------------------------------

More information

Spectroscopic Ellipsometer --- J. A. Woollam alpha-se

Spectroscopic Ellipsometer --- J. A. Woollam alpha-se Spectroscopic Ellipsometer --- J. A. Woollam alpha-se Introduction Figure 1: J. A. Woollam alpha-se spectroscopic ellipsometer An ellipsometer measures the change in polarization as light reflects or transmits

More information

Prepulse Inhibition (PPI) Boxes

Prepulse Inhibition (PPI) Boxes Prepulse Inhibition (PPI) Boxes Always calibrate the PPI boxes prior to testing on each day See print manual for more information Calibration: 1) Turn on the boxes by switching on the surge protector on

More information

20W Tungsten-Halogen Light Source. Operation Manual

20W Tungsten-Halogen Light Source. Operation Manual 20W Tungsten-Halogen Light Source - ASB-W-020 - Operation Manual Important Safety Notices 1. Never look directly into the light beam, including through the cooling fan while light is on, as this can cause

More information

UHP-M Ultra-High Power UV-Visible LED Light Source Fluorescence Microscopy Ver. 07

UHP-M Ultra-High Power UV-Visible LED Light Source Fluorescence Microscopy Ver. 07 Introduction UHP-M Ultra-High Power UV-Visible LED Light Source Fluorescence Microscopy Ver. 07 Prizmatix UHP-M Dual-LED light engine is designed to substitute the Metal-Halide and Mercury lamps in routine

More information

1065AR1_M3_MAN_Argos_OmegaZenPipetteController_Manual.indd 2

1065AR1_M3_MAN_Argos_OmegaZenPipetteController_Manual.indd 2 Pipette Controller Operator's Manual En ligne; Manuel de l opérateur Online; Bedienungshandbuch En línea; Manual del operador In linea; Manuale d uso www.argos-tech.com RoHS 1065AR1_M3_MAN_Argos_OmegaZenPipetteController_Manual.indd

More information

CVU-200-KIT. 200 V Bias Tee Kit. Description. Parts list / October 2014 *P A* 1

CVU-200-KIT. 200 V Bias Tee Kit. Description. Parts list / October 2014 *P A* 1 Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 http://www.keithley.com CVU-200-KIT 200 V Bias Tee Kit Description The CVU-200-KIT Bias Tee Kit consists of three 2600-RBT-200

More information

Field Update Guide. for Raven Viper Pro

Field Update Guide. for Raven Viper Pro Field Update Guide for Raven Viper Pro Introduction The field update kit (P/N 117-0171-467) is designed to allow the Raven Viper Pro to utilize the automatic power down feature without returning the console

More information

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer Operating instructions - Cary 100 Bio UV-visible Spectrophotometer This document describes how to power up the spectrophotometer, set its measurement parameters, insert one or more of samples for analysis

More information

BGS (at min. setting) BGS (at max. setting)

BGS (at min. setting) BGS (at max. setting) Compact Photoelectric Sensor with Built-in Amplifier EZ-LS CSM_EZ-LS_DS_E_7_ Distance-settable Sensor Unaffected by Workpiece Color and Background Distance-settable triangulation model unaffected by color.

More information

User Manual. Binocular Stereo Microscope. MicroscopeNet.com

User Manual. Binocular Stereo Microscope. MicroscopeNet.com User Manual Binocular Stereo Microscope Model K2213 Series MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 1 1. Components Illustration... 2 2. Installation... 3 3. Operation...

More information

Portable Document & Image Scanner. User s Manual

Portable Document & Image Scanner. User s Manual Portable Document & Image Scanner User s Manual Table of Contents 1. Precaution. 2 2. Preparation..5 Standard Accessories 5 Part Names & Functions..6 3. key features..... 7 4. Status Icons Explanations.....8

More information

BGA IRDA-WELDER User Manual. Model: T870A

BGA IRDA-WELDER User Manual. Model: T870A BGA IRDA-WELDER Model: T870A 0 CATALOGUE Features....2 Technical Parameter and Components..3 T-870A illustrated explaining...4 (1)The whole machine 4 (2)Front panel.. 5 (3)Rear panel.. 5 (4)Focus holder

More information

SUSS SB6E BONDER SOP. October 2013

SUSS SB6E BONDER SOP. October 2013 SUSS SB6E BONDER SOP October 2013 The SUSS SB6 bonder is capable of silicon fusion, anodic, and eutectic bonding. There are two separate heads; one for anodic bonding, and another for fusion bonding. Definitions

More information

Panalytical XPert Powder X-Ray Diffraction SOP

Panalytical XPert Powder X-Ray Diffraction SOP Revision 1 2/06/2018 Panalytical XPert Powder SOP Page 1 of 11 Panalytical XPert Powder X-Ray Diffraction SOP NOTE: Latest revisions are in Blue. Table of Contents 1.0 Safety 2.0 Training 3.0 Sample Preparation

More information

5.2.2 Pins on the alignment detector are fragile. Use caution when inserting and removing the detector. Report any broken pins to staff.

5.2.2 Pins on the alignment detector are fragile. Use caution when inserting and removing the detector. Report any broken pins to staff. Title: Woollam Vase Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Woollam VASE. All users are expected to have read and understood

More information

University of Minnesota NanoFabrication Center Standard Operating Procedure

University of Minnesota NanoFabrication Center Standard Operating Procedure Equipment Name: JEOL SEM Coral Name: jeol-sem Revision Number: 4 Model: JSM-6610LV Revisionist: K. Roberts Location: Area 3 Date: 9/17/2013 1 Description The JSM-6610LV is a scanning electron microscope

More information

Quick Start. Getting Started

Quick Start. Getting Started Quick Start To begin using HyGCHP immediately: 1. Open the program from the Start menu, click OK to pass the title screen. 2. Choose one of the eight system configurations to model. Click OK. 3. On the

More information

PerkinElmer Series 200 HPLC: Operating Procedure

PerkinElmer Series 200 HPLC: Operating Procedure PerkinElmer Series 200 HPLC: Operating Procedure Chem Eng 5503 Version 1.01 Last Update 8/9/07 This document outlines the basic operating procedure for the PE series 200 HPLC, UV/Vis and SEDEX Light Scattering

More information

Models 707B and 708B Switching Matrix

Models 707B and 708B Switching Matrix E C N E D I F N O C F O E R U S A E M R E T A E R G A Models 707B and 708B Switching Matrix User s Manual 707B-900-01 Rev. A / August 2010 Test Equipment Depot - 800.517.8431-99 Washington Street Melrose,

More information

User's Manual. New ipad Protection Case with detachable Bluetooth keyboard. Product number: SI54072

User's Manual. New ipad Protection Case with detachable Bluetooth keyboard. Product number: SI54072 User's Manual New ipad Protection Case with detachable Bluetooth keyboard Product number: SI54072 Thank you for purchasing our New ipad Protection Case with detachable Bluetooth-Keyboard We hold several

More information

Survival Laser SL-002PB Laser Parts Bundle w/g-1 Lens Assembly & Operation Instructions

Survival Laser SL-002PB Laser Parts Bundle w/g-1 Lens Assembly & Operation Instructions Survival Laser SL-002PB Laser Parts Bundle w/g-1 Lens Assembly & Operation Instructions WARNING: READ ALL INSTRUCTIONS AND THE ENCLOSED SAFETY PRECAUTIONS BEFORE ASSEMBLY AND USE Assemble and use these

More information

Survival Laser SL-005PB Laser Parts Bundle Assembly & Operation Instructions

Survival Laser SL-005PB Laser Parts Bundle Assembly & Operation Instructions Survival Laser SL-005PB Laser Parts Bundle Assembly & Operation Instructions WARNING: READ ALL INSTRUCTIONS AND THE ENCLOSED SAFETY PRECAUTIONS BEFORE ASSEMBLY AND USE Assemble and use these parts ONLY

More information

Ref ID D S/N. JP10D00123 Vintage Wafer size 200 mm Software rev. B or higher. Test Head Hp E3101A 19 Para.

Ref ID D S/N. JP10D00123 Vintage Wafer size 200 mm Software rev. B or higher. Test Head Hp E3101A 19 Para. OEM / Model Agilent HP 4071 / E3101A Ref ID D214463 S/N JP10D00123 Vintage 1997 Process Parametric test Wafer size 200 mm Software rev. B.02.00 or higher Test Head Hp E3101A 19 Para. chuck pins Hp 8116A

More information

Sense Autonomous 2_11. All rights reserved.

Sense Autonomous 2_11. All rights reserved. Sense Autonomous Sense Autonomous 2_11 All rights reserved. The material in this book may not be copied, duplicated, printed, translated, re-edited or broadcast without prior agreement in writing. For

More information

Steps for LCMS-8040 Triple Quadrupole

Steps for LCMS-8040 Triple Quadrupole Steps for LCMS-8040 Triple Quadrupole June 2017 Version 2 A. Data Acquisition 1. Check lights on MS. There should be a total of three lights lit by LED. One of them is blinking. Power Blinking Gas Status

More information

K Service Source. Macintosh II/IIx/IIfx

K Service Source. Macintosh II/IIx/IIfx K Service Source Macintosh II/IIx/IIfx K Service Source Basics Macintosh II/IIx/IIfx Basics Overview - 1 Overview This manual includes complete repair procedures for the Macintosh II, Macintosh IIx, and

More information

For documentation, both instruments are available with a trinocular head which permits mounting of a video or digital camera system.

For documentation, both instruments are available with a trinocular head which permits mounting of a video or digital camera system. PriorLux 1 1. Introduction The PriorLab and PriorLux upright compound microscopes are high quality instruments equipped with high resolution, chromatically corrected optics for excellent image quality.

More information

1. PRODUCT INTRODUCTION Description 1-2. Feature...

1. PRODUCT INTRODUCTION Description 1-2. Feature... CONTENTS PAGE 1. PRODUCT INTRODUCTION... 1-1. Description 1-2. Feature... 1 1 1 2. TECHNICAL SPECIFICATIONS 3 3. PRECAUTIONS BEFORE OPERATION.... 3-1. Unpacking the Instrument... 3-2. Checking the Line

More information

Model 8101-PIV Test Fixture User s Guide. Overview. SMU Device holder (near latch) Two Leaded Part Plungers. PIV Device holder (near hinge)

Model 8101-PIV Test Fixture User s Guide. Overview. SMU Device holder (near latch) Two Leaded Part Plungers. PIV Device holder (near hinge) Keithley Instruments, Inc. 28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY www.keithley.com Model 8101-PIV Test Fixture User s uide Overview WARNIN The procedures contained in this User s uide are

More information

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips Section 1: Starting an experiment 1.1 Opening a file 1. Open the File menu and select Open Activity. 2. In the Open dialog box, navigate to

More information

X-3 LED. Owner s Manual ROBOHEAD

X-3 LED. Owner s Manual ROBOHEAD X-3 LED Owner s Manual ROBOHEAD Thanks a lot for purchasing our LED robohead. Please read this user manual carefully and follow the instructions to avoid danger or damage to the unit due to mishandling.

More information

Working Copy For Review Purposes Only!

Working Copy For Review Purposes Only! Working Copy For Review Purposes Only! Version #1.5 08/29/06 Aaron Holland Up-dated D.Turnbull 10-18-2006 (MarkerSOP3.doc) Up-dated D. Turnbull 12-11-06 (MarkerSOP4.doc) Up-dated D.Turnbull 05-12-2008

More information

kena TM Instruction Manual T-1050

kena TM Instruction Manual T-1050 kena TM Instruction Manual T-1050 2 Table of Contents Introduction / Package Contents 4 Diagram 4 Package Contents 5 Operational Requirements 5 Optional Accessories to Order 5 Specifications 6 Quick Start

More information

Bruker OminiFlex MALDI-TOF Mass Spectrometer Operation Quick Start

Bruker OminiFlex MALDI-TOF Mass Spectrometer Operation Quick Start Bruker OminiFlex MALDI-TOF Mass Spectrometer Operation Quick Start Sample preparation for the Bruker OminiFlex MALDI-TOF Mass Spectrometer is of crucial importance. Please take care when preparing your

More information

Installation and Getting Started Guide

Installation and Getting Started Guide Installation and Getting Started Guide Metrics Win4156 Version 4.1.0 Copyright 1999-2015, Metrics Technology, Inc. All rights reserved. Table of Contents Installing the Win4156 Software... 4 Obtain a Codeword...

More information