T1/T2*-weighted MRI provides clinically relevant pseudo-ct density data for the pelvic bones in MRI-only based radiotherapy treatment planning

Size: px
Start display at page:

Download "T1/T2*-weighted MRI provides clinically relevant pseudo-ct density data for the pelvic bones in MRI-only based radiotherapy treatment planning"

Transcription

1 Acta Oncologica, 2013; 52: original article T1/T2*-weighted MRI provides clinically relevant pseudo-ct density data for the pelvic bones in MRI-only based radiotherapy treatment planning Mika Kapanen & Mikko Tenhunen Department of Oncology, Helsinki University Central Hospital, HUS, Finland Acta Oncol Downloaded from informahealthcare.com by on 05/20/14 Abstract Background and purpose. In radiotherapy (RT), target soft tissues are best defined on MR images. In several cases, CT imaging is needed only for dose calculation and generation of digitally reconstructed radiographs (DRRs). Image co-registration errors between MRI and CT can be avoided by using MRI-only based treatment planning, especially in the pelvis. Since electron density information can not be directly derived from the MRI, a method is needed to convert MRI data into CT like data. We investigated whether there is a relationship between MRI intensity and Hounsfield unit (HU) values for the pelvic bones. The aim was to generate a method to convert bone MRI intensity into HU data surrogate for RT treatment planning. Material and methods. The HU conversion model was generated for 10 randomly chosen prostate cancer patients and independent validation was performed in another 10 patients. Data consisted of 800 image voxels chosen within the pelvic bones in both T1/T2 * -weighted gradient echo and CT images. Relation between MRI intensity and electron density was derived from calibrated HU-values. The proposed method was tested by constructing five pseudo -CT series. Results. We found that the MRI intensity is related to the HU value within a HU range from 0 to 1400 within the pelvic bones. The mean prediction error of the conversion model was 135 HU. Dose calculation based on the pseudo-ct images was accurate and the generated DRRs were of good quality. Conclusions. The proposed method enables generation of clinically relevant pseudo-ct data for the pelvic bones from one MRI series. It is simpler than previously reported approaches which require either acquisition of several MRI series or T2 * maps with special imaging sequences. The method can be applied with commercial clinical image processing software. The application requires segmentation of the bones in the MR images. Magnetic resonance imaging (MRI) is widely used in radiotherapy (RT) treatment planning due to its superior soft tissue contrast when compared with computed tomography (CT). This means that MR and CT images should be co-registered to enable dose calculation and the generation of the digitally reconstructed radiographs (DRRs). Because of the different patient setup in these images, this increases the geometrical uncertainty of the position of the target and normal tissues [1]. MRI simulators are MRI devices which have a wide or open bore, flat imaging table top and a laser system suitable for skin marking. They can be used for MRI-only based RT treatment planning since they allow using the same fixation devices and patient setup that are used in RT treatment [2,3]. The advantage is that image coregistration errors are omitted. This approach, however, may result in dose calculation errors of 1 2% in the pelvic region due to the lack of electron density information which is derived directly from the CT images. Hounsfield unit (HU) information has been assigned for the MR images in literature. The simplest approach used a bulk density for the whole bone volume and assumed everything else as water equivalent tissue [2,4,5]. In the more elaborated techniques, density values or bone characteristics have been derived from T2 * (apparent transverse relaxation time) maps [6], by determining T2 (transverse relaxation time) spectrum of bone [7] or by combining information from several MRI series [8]. These approaches have utilized several image series and/or special imaging sequences based on ultra short echo times (TE 0.1 ms). These sequences are still prototypes and their application requires extra imaging time. The essential characteristics of bones Correspondence: M. Kapanen, Department of Oncology, Helsinki University Central Hospital, POB-180, HUS, Finland. mika.kapanen@hus.fi (Received 28 December 2011; accepted 4 May 2012) ISSN X print/issn X online 2013 Informa Healthcare DOI: / X

2 have also been evaluated simply by using anatomical T1/T2 * -weighted gradient echo MRI [9]. In this study, we investigated whether there is a relation between MRI intensity and the HU-value within the pelvic bones. The aim was to find a method that could provide HU data from the MRI to generate a pseudo-ct series which could be then used in clinical RT treatment planning. T1/T2 * -weighted anatomical gradient echo imaging was chosen because it provides essential characteristics of the bones and especially because T2 * time correlates with bone density [9,10]. Segmentation of the bones seemed possible in these images because the intensity range in the bones was clearly different from that of the adjacent muscles. The relation between MRI intensity and the relative electron density was obtained from calibrated HU-values. The results were validated in an independent patient group. We introduced the method in a form that can be applied with commercial clinical image processing software with limited processing capability. This was possible because the conversion model was constructed and validated by choosing the voxels randomly by hand and because the pseudo CT series were generated based on discrete MRI intensity values. The pseudo-ct series were constructed for five patients to test the feasibility of the method for RT treatment planning. We investigated the achieved dose calculation accuracy and constructed sample DRRs. Material and methods Equipment and imaging parameters The MRI was performed with a 1.5 T scanner GE Optima MR450w (GE Medical Systems, Milwaukee, WI, USA). The scanner has a flat table top and a wide bore (diameter 70 cm) enabling the use of similar patient immobilization devices as in external RT. A T1/T2 * -weighted (spoiled steady state) threedimensional (3D) gradient echo sequence LAVA Flex Ò was chosen. This is a dual echo sequence which provides out-phase and in-phase images for water and fat signals acquired at echo times (TE) of 2.1 and 4.2 ms, respectively. Fat and water images can be calculated from these images. Repetition time (TR) was 6.8 ms, flip angle 15 o, reconstruction slice thickness 2.4 mm (interpolated to 1.2 mm by zerofilling) and matrix size 428x448 (FOV cm). Bandwidth (BW) was 90.9 khz for the entire FOV. Only in-phase images were used (except for segmentation of bone) because bone marrow contains both fat and water. The images were corrected for intensity inhomogeneity using PURE Ò (based on calibration series) and for geometrical distortion using Grapwarp 3 Ò both of these software provided by the MRI provides pseudo-ct data for bones 613 vendor. The imaging area in the superior-inferior direction covered the prostate and seminal vesicles with margins ranging from 2 to 5 cm. The coverage in the anterior-posterior and left-right directions covered the whole body contour to enable dose calculation. Imaging was carried out using an eight channel phased array cardiac coil with the posterior part of the coil placed below the flat table top and the anterior part supported by two plastic semi arcs constructed in our hospital. This system setup enables the use of the same patient setup as in treatment. A drawback is a lower signal to noise ratio when compared with purely diagnostic imaging due to the greater distance between the coil and patient skin. For this kind of approach, however, the image quality has been considered sufficient [11]. The CT imaging was carried out with GE Light- Speed RT scanner (4-slice, GE Medical Systems, Waukesha, WI) at 120 kvp with slice thickness of 2.5 mm. The HU-values were calibrated to electron density values using the RMI-phantom (Gammex, Middleton, WI). The image processing was performed using MIM 5 Ò workstation (MIM Software Inc., Cleveland, OH, USA). Treatment planning was carried out with a grid size of 2.5 mm using Eclipse Ò workstation (Varian Medical Systems, Palo Alto, CA, USA). Patients and generation of data points Twenty randomly chosen prostate cancer patients (age 59 to 82 years, mean of 70 years) prescribed to external RT of the prostate and seminal vesicles were MRI and CT imaged according to our clinical routine procedure. The patients were informed about the study and verbal agreement was obtained for the inclusion. The study did not have any effect on the treatment of these patients but one or two extra MRI series were acquired (extra imaging time 6 minutes per series). Five patients were imaged also with slightly different values of the TE 5.2 ms, TR 8.3 ms and BW 62.5 khz to estimate the effect of imaging parameters. The axial MR and CT images were first co-registered according to bony landmarks. Ten patients were included in the generation of the model and another 10 for independent validation. We chose 40 voxels randomly within the pelvic bone area for each patient. The corresponding HU and MRI intensity values of a voxel gave a data point for the generation and validation of the model. This kind of model construction is simple and it is possible with clinically accepted image processing software with limited processing capability. The superior-inferior (S-I) coverage was from the femoral bone/trochanter major to the middle of the iliac bone. The voxels were chosen uniformly within the segmented bone volume and it was confirmed that they cover the whole

3 614 M. Kapanen & M. Tenhunen observed HU range from about 0 up to The voxels were located at different kind of bone structures such as cortical bone, trabecular bone and bone marrow. The intensity in MR images may be inhomogeneous due to positional variations in the sensitivity of the surface coil and magnetic field imperfections caused by the system or the patient. Also the skin distance of a surface coil and gains in signal acquisition may differ between the patients. Even though the inhomogeneity correction (PURE Ò ) equalizes intensity variations within a patient, overall intensity levels may still differ between patients. To minimize the effect of this, the MRI intensity values of the voxels were normalized to the average intensity of a muscle that is located closest to the bone (see Figure 1). For this purpose, eight small ROIs were drawn in different image slices excluding contribution of fat. The ROIs were uniformly located within same S-I coverage as the voxels. The normalized data was multiplied by 150 (mean normalization factor) to achieve realistic MRI intensity level. Modeling of the relation between MR intensity and HU-value inside the bone Small solid fragments of the bone generate susceptibility gradients within pores that are located next to them. These pores contain both water and fat. The gradients have an effect on MRI intensity through changes in T2 *. The relationship is very complex and depends on the volume fraction of these solid parts, their number, thickness and separation [12,13]. We generated a simplified model to give a theoretical basis for the relation between the MRI intensity and HU-value (or electron density). Signal intensity s in gradient echo imaging is proportional to s A(T1) P exp(-te R2 * ), where R2 * 1/T2 *, P is proton (spin) density and A is a T1-dependent factor determined by the flip angle, TR and level of spoiling of transversal magnetization. The visible MRI signal comes from different soft tissue components (such as collagen, collagenbound water and pore water) and is the integral over different values of T1, P and T2 * within a voxel. When normalizing to the signal from a muscle (water) we obtain The aim was to investigate the overall relation between MRI intensity and HU-values and to find out whether the Equation 1 can be applied in a simple way and with sufficient accuracy for clinical RT planning. In practice, the inverse form of this formula is preferable (i.e. HU-values should be determined from the MRI intensity values). Since mathematic forms of all the factors in Equation 1 are not known, the inversion was done by a series expansion which results in a simple polynomial model of 2 HU( s) A Bs Cs... (2) where A, B and C are the fitting parameters and s is MRI intensity (normalized or unnormalized). The fitting parameters are complex functions of the distributions of T1, T2 *, P and tissue component fractions within a voxel. The order for the polynomial fit was chosen based on minimum value of Akaike information criterion AIC [14] calculated as 2 aic= Nln[ ( y) ] 2P (3) where P is the number of fitting parameters, N is the number of data points and S(Dy) 2 is the sum of the squares of the fitting errors [Dy HU measured - HU(s)]. Each fitting parameter value is given as the obtained value ( 1 SD). A 95% confidence limit for a parameter is 2 SD. The Pearson correlation coefficient (r) was used to quantify linear correlation between two sets of data and to approximate overall linear correlation for slightly non-linear relations. Generation of pseudo-cts Five pseudo-ct series were generated to demonstrate that the obtained model could provide appropriate data for dose calculation and DRR generation. Five patients were chosen randomly from the validation group. Pelvic bones were manually contoured according to the LAVA Flex in-phase, out-phase and water-images. The pseudo-cts were generated with clinically accepted software with limited processing s = A x yz TE ( R2 (x,y,z) R2 ( T1(,, )) P( x, yz, ) muscle ) e r( x, y, z) dx dy dz A muscle P muscle (1) where r(x,y,z) dx dy dz is the fraction of a component having T1, P and T2 * within a voxel and x, y and z are space coordinates. capability and, therefore, a voxel-based conversion was not possible. Instead, the voxels within the segmented bone structure were divided into 16 subgroups based

4 MRI provides pseudo-ct data for bones 615 bones were corrected to correspond to that of actual CT ( fat correction ). Acta Oncol Downloaded from informahealthcare.com by on 05/20/14 Figure 1. Example of a small normalization ROI (upper white circle) chosen within a muscle that is located next to the bone. The edge of the bone (lower white contour) is clearly visible enabling segmentation of the bone edge in the MR image (LAVA Flex â in-phase image). on their MRI intensity values using an intensity ranging tool. According to the conversion model, the MRI intensity ranges of the subgroups corresponded to the following HU ranges: below 0, ,..., , and above The volume that had HU values above 1400 was below 1 ml. Therefore, the upmost subgroup was above Since only very small negative HU values were observed for the patients (HU 210), the lowest subgroup was below 0. Each pseudo-ct was generated by converting mean MRI intensities of the subgroups into HUs by using the obtained model. The rest of the body was assumed water equivalent (HU 0). Rectum and skin contour in MRI and CT were corrected to be similar to minimize their effect on dose calculation. DRRs were generated in a treatment planning system from these pseudo-ct images. A simple four-field box treatment plan was calculated for 6 and 15 MV beams to demonstrate the effect of the method on dose calculation. PTV included the prostate and seminal vesicles. Dose calculation differences between a pseudo-ct and actual CT were determined for five different versions of a pseudo-ct: i) by assuming the whole body as water equivalent; ii) by using a bulk HU value for the bones which corresponded to the average obtained for the whole group of the patients; iii) by assuming an individually determined bulk HU values for the bones converted from the MRI; iv) by determining regional HU values converted from the MRI; and v) as in iv) but the HU values for the body outside the pelvic Results Relation between HU-value and relative electron density The electron density values relative to water of 1.000, 1.039, 1.116, 1.142, 1.285, and of the RMI phantom resulted in measured HU-values of 10, 12, 97, 181, 516, 931 and 1401, respectively. Pearson correlation coefficient of these values was r (p ). The AIC method suggested a linear model. The fitting parameters were A ( 0.013) and B ( ) 1/HU. The obtained linear model was used to convert HUvalues into relative electron density values. Modeling the relation between MRI-intensity and HU-value The correlation between MRI-intensity and HUvalue was r (p ) (total of 400 data points). The AIC method suggested a second order fit. Mean, SD and maximal error of the fit were 131, 115 and 512 HU, respectively. The fitting parameters were A 1947 ( 71) HU, B ( 0.79) HU/MR intensity unit and C ( ) HU/ MR 2 intensity unit. The slope of a linear fit was HU/MR intensity unit demonstrating a mean steepness of the non-linear fit. The model fit is illustrated in Figure 2. Quite similar model fit was obtained without data normalization but systematic errors for an individual patient raised from 85 HU up to 154 HU. Figure 2. Relation between MRI intensity (in-phase image) and HU number obtained for 400 randomly chosen voxels within the pelvic bones. The second order polynomial fit is expressed as a solid line.

5 616 M. Kapanen & M. Tenhunen Normalization to fat resulted in poorer fit and the correlation of data was below 0.4. The derived absolute MRI intensity values were lower for the different set of imaging parameters. This demonstrates that the parametrization for the relation between MRI intensity and HU depends on the imaging parameters. The correlation between the MRI intensities obtained for the two parameter sets was almost perfect (p ). Validation group The correlation between the calculated and measured HUs was r (p ) for the 400 data points. The mean, SD and maximal differences were 135, 121 and 578 HU, respectively. Prediction error of the HU did not correlate with HU-level (r 0.291) within 0 HU The linear fit of the calculated and measured HU values resulted in the parameter values of A 25 ( 18) HU and B 0.96 ( 0.03) suggesting that there was no systematic error between the measured and calculated values. The relationship between the measured and calculated HUs is presented in Figure 3. The range of mean HU values of the pelvic bones varied between 225 and 473 with a mean SD of HU. Generated pseudo-ct images An example of the generated DRR images is presented in Figure 4. The dose calculation results are summarized in Table I. The proposed method increased overall calculation accuracy for all of the five patients. The individual bulk HUs for the pelvic bones were calculated within 23 HU. Discussion We have demonstrated that there is relation between the intensity of T1/T2 * -weighted gradient echo MR image and HU number (or relative electron density) within the pelvic bones. This relation provides patient specific bone density information both for dose calculation and the generation of verification images (DRRs) for MRI-only based RT treatment planning. Such data have earlier been derived from T2 * maps [6], T2 spectrum of bone [7] or from several MRI series [8] based on special ultra short echo time (UTE) sequences. To the best of our knowledge, this is the first report of generation of bone density data from only one anatomical MRI series. The proposed method is simple and that it utilizes the same MRI Figure 3. Relation between the measured and calculated HUvalues for 400 data points obtained from the validation group. Straight solid line with slope equal to one illustrates ideal calculation. Figure 4. DRR image generated from A) actual CT image and B) pseudo-ct image with HU-values inside the bones converted from the MRI. The field size is equal in the both images. Lateral images were chosen to demonstrate clear visibility of bone edges through whole transversal section of the pelvic bones.

6 MRI provides pseudo-ct data for bones 617 Table I. Maximal dose calculation differences (in percentages) between pseudo2ct and actual CT for five randomly chosen patients (equal MU counts were applied for both images). HU information used inside the pelvic bones Energy Fraction of PTV Water only Bulk value of 337 corresponding group average Individual bulk value obtained from MRI Regional HU obtained from MRI Regional HU from MRI with body corrected a 6 MV V95% 1.8% 21.9% 21.1% 21.2% 20.6% V50% 2.0% 21.9% 20.9% 21.0% 20.4% V5% 1.7% 21.9% 21.2% 21.2% 20.5% all points within 3.9% 3.2% 2.2% 2.4% 1.3% 15 MV V95% 1.9% 21.4% 20.9% 20.9% 20.4% V50% 1.6% 21.4% 20.7% 20.8% 20.4% V5% 1.5% 21.2% 20.8% 20.8% 20.4% all points within 3.7% 2.7% 2.2% 2.2% 1.2% a HU value for the body outside the bones was corrected for fat content (on the average to about 220) to demonstrate differences related mainly to the conversion of HU values inside the bones. Acta Oncol Downloaded from informahealthcare.com by on 05/20/14 series that can be used for the segmentation of bones, dose calculation and DRR generation. There is no need for multiple MRI series to replace CT. The method can be applied with commercial clinical image processing systems with limited processing capability. The provided conversion model was validated inside the pelvic bones and it should not be adopted for use outside the segmented bone structures. The mean prediction error obtained is equal to value of 137 HU reported for a voxel based method utilizing four image series [8]. In this light, this study presents an equal alternative for HU conversion. The mean prediction and fitting errors, and dose calculations were considered acceptable for a HU range from 0 to The MRI signal from dense cortical bone (HU 1200) was dark [15] and just about 10 units above the noise level. Even though the effect of noise was most pronounced with the lowest MRI intensities, the model seemed to average the data properly. The large maximal prediction and fitting errors were most probably due to slight image coregistration errors, noise and partial volume effects. These factors may depend on the voxel size. The results were compared with the conventional bulk density approach since it utilizes data obtained purely for the bones and it does not adjust HU based on dose calculation or on density values of body organs outside the bones (e.g. fat and muscle). Similar bulk HU values have been reported for several groups of patients (density from 1.19 to 1.22 g/cm 3 corresponding to HU range from 288 to 333) [2,4,5]. Mean prediction error of the relative electron density was 0.06 (corresponding to 135 HU) which should not normally result in a dose calculation error greater than about 1% for photons. By minimizing the effect of fat we demonstrated that the uncertainty of dose calculation related to the HU conversion was acceptable. We are aiming to develop a method for regional fat correction based on the LAVA Flex Ò images. In addition, we are evaluating calculation accuracy achievable by using the presented method in detail for different treatment sites and techniques. The normalization of the MRI intensity was observed important to minimize systematic errors in HU conversion. This is a drawback but no better method was found to correct for variations in the overall intensity level between patients. The normalization was also tested to one ROI but poorer results were obtained than by normalizing to the average of several ROIs. This is most likely explained by image noise and small local intensity variations left after the inhomogeneity correction (PURE Ò ). Since PURE Ò is based on calibration images, also potential patient movement during the imaging may result in uncorrected intensity variations. The use of larger normalization ROIs can not solve the problem since the ROI area should be kept small in order to minimize the contribution of fat and other tissues adjacent to the muscle. As a conclusion, the normalization should be done to the average of several small ROIs (at least eight) drawn within the muscles closest to the bones on different image slices (see Figure 1). The MRI intensity values could not be directly converted into electron densities because of the lack of a MRI compatible phantom. The uncertainty related to conversion of electron density from HU is regarded clinically acceptable and, therefore, we conclude that the presented conversion of the MRI intensity into electron density through HUs is acceptable. The relation between the MRI intensity and HU seemed to exist at least with this type of MRI imaging sequence within a moderate range of the imaging parameter values. The presented model should be tuned for the individual MRI scanner, imaging sequence and parameters and also for the CT scanner and CT imaging protocol. Changes in system performance (e.g. magnetic field shimming)

7 618 M. Kapanen & M. Tenhunen due to aging may require updating the model and its validity should be established regularly. The presented method can be considered sensitive to bone segmentation errors. The intensity conversion, however, increased the intensity of cortical bone area (and thus enhancing the bone edge) and minimized the effect of such errors. The bone edge was observed to be consistent within about 1 mm in MRI and CT. In addition, the location of the bone edge was observed to be accurate within 1 mm (about voxel size) in an on-going phantom study by our clinic. On this ground, the generated DRRs are of good quality and accurately present the structures of pelvic bones essential for treatment localization. In T2-weighted images, bone edges had intensity values very similar to those of adjacent muscles. This may limit their use for automatic segmentation of the bones supporting our choice of the MRI sequence. Our next goal is to develop an automated segmentation method for the pelvic bones in the T1/T2 * -weighted MR images. The segmentation is aimed to be implemented into the same step with the segmentation of other tissues needed for the RT treatment planning. Acknowledgements Rita Janes, M.D., is greatly acknowledged for her valuable comments and language corrections that significantly improved the manuscript. The authors report no conflicts of interest. The authors alone are responsible for the content and writing of the paper. Declaration of interest: The authors report no conflicts of interest. The authors alone are responsible for the content and writing of the paper. References [1] Roberson P, McLaughlin P, Narayana V, Troyer S, Hixson GV, Kessler ML. Use and uncertainties of mutual information for computed tomography/magnetic resonance (CT/ MR) registration post permanent implant of the prostate. Med Phys 2005;32: [2] Lee YK, Bollet M, Charles-Edwards G, Flower MA, Leach MO, McNair H, et al. Radiotherapy treatment planning of prostate cancer using magnetic resonance imaging alone. Radiother Oncol 2003;66: [3] Karlsson M, Karlsson MG, Nyholm T, Amies C, Zackrisson B. Dedicated magnetic resonance imaging in the radiotherapy clinic. Int J Radiat Oncol Biol Phys 2009;74: [4] Jonsson JH, Karlsson MG, Karlsson M, Nyholm T. Treatment planning using MRI data: An analysis of the dose calculation accuracy for different treatment regions. Radiat Oncol 2010;5:62 9. [5] Lambert J, Greer PB, Menk F, Patterson J, Parker J, Dahl K, et al. MRI-guided prostate radiation therapy planning: Investigation of dosimetric accuracy of MRI-based dose planning. Radiother Oncol 2011;98: [6] Messiou C, Collins DJ, Morgan VA, Robson MD, debono JS, Bydder GM, et al. Quantifying sclerotic bone metastases with 2D ultra short TE MRI: A feasibility study. Canc Biomark 2010;7: [7] Horch RA, Gochberg DF, Nyman JS, Does MD. Noninvasive predictors of human cortical bone mechanical properties: T2-discriminated 1H NMR compared with high resolution X-ray. PLoS ONE 2011;6:e [8] Johansson A, Karlsson M, Nyholm T. CT substitute derived from MRI sequences with ultrashort echo time. Med Phys 2011;38: [9] Majumdar S, Genant HK, Grampp S, Newitt DC, Truong VH, Lin JC, et al. Correlation of trabecular bone structure with age, bone mineral density, and osteoporotic status: In vivo studies in the distal radius using high resolution magnetic resonance imaging. J Bone Miner Res 1997;12: [10] Majumdar S, Thomasson D, Shimakawa A, Genant HK. Quantitation of the susceptibility difference between tracebular bone and bone marrow: Experimental studies. Magn Reson Med 1991;22: [11] McJury M, O Neill A, Lawson M, McGrath C, Grey A, Page W, et al. Assessing the image quality of pelvic MR images acquired with a flat couch for radiotherapy treatment planning. Br J Radiol 2011;84: [12] Ford JC, Wehrli FW, Chung H. Magnetic field distribution in models of trabecular bone. Mag Reson Med 1993;30: [13] Link TM, Majumdar S, Augat P, Lin JC, Newitt D, Lane NE, et al. Proximal femur: Assessment for osteoporosis with T2 * decay characteristics at MR imaging. Radiology 1998;209: [14] Akaike H. A new look at the statistical model identification. IEEE Trans Automat Control 1974;19: [15] Keereman V, Fierens Y, Broux T, De Deene Y, Lonneux M, Vandenberghe S. MRI-based attenuation correction for PET/MRI using ultrashort echo time sequences. J Nucl Med 2010;51:812 8.

Proton dose calculation algorithms and configuration data

Proton dose calculation algorithms and configuration data Proton dose calculation algorithms and configuration data Barbara Schaffner PTCOG 46 Educational workshop in Wanjie, 20. May 2007 VARIAN Medical Systems Agenda Broad beam algorithms Concept of pencil beam

More information

Effects of the difference in tube voltage of the CT scanner on. dose calculation

Effects of the difference in tube voltage of the CT scanner on. dose calculation Effects of the difference in tube voltage of the CT scanner on dose calculation Dong Joo Rhee, Sung-woo Kim, Dong Hyeok Jeong Medical and Radiological Physics Laboratory, Dongnam Institute of Radiological

More information

ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma

ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma ANALYSIS OF PULMONARY FIBROSIS IN MRI, USING AN ELASTIC REGISTRATION TECHNIQUE IN A MODEL OF FIBROSIS: Scleroderma ORAL DEFENSE 8 th of September 2017 Charlotte MARTIN Supervisor: Pr. MP REVEL M2 Bio Medical

More information

MRI Visualisation by Digitally Reconstructed Radiographs

MRI Visualisation by Digitally Reconstructed Radiographs Antoine Serrurier *a1, Andrea Bönsch b, Robert Lau a, Thomas M. Deserno (né Lehmann) a a Department of Medical Informatics, Uniklinik RWTH Aachen, Germany; b Virtual Reality Group, RWTH Aachen University,

More information

Modeling of MR-guided HIFU for Breast and Brain Therapy

Modeling of MR-guided HIFU for Breast and Brain Therapy Modeling of MR-guided HIFU for Breast and Brain Therapy Douglas A. Christensen, Allison Payne, Nick Todd, Scott Almquist, Alexis Farrer and Dennis L. Parker University of Utah Salt Lake City, Utah Overview

More information

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4

More information

Influence of electron density spatial distribution and X-ray beam quality during CT simulation on dose calculation accuracy

Influence of electron density spatial distribution and X-ray beam quality during CT simulation on dose calculation accuracy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 3, summer 2011 Influence of electron density spatial distribution and X-ray beam quality during CT simulation on dose calculation accuracy

More information

GPU applications in Cancer Radiation Therapy at UCSD. Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC

GPU applications in Cancer Radiation Therapy at UCSD. Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC GPU applications in Cancer Radiation Therapy at UCSD Steve Jiang, UCSD Radiation Oncology Amit Majumdar, SDSC Dongju (DJ) Choi, SDSC Conventional Radiotherapy SIMULATION: Construciton, Dij Days PLANNING:

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Outline High Spectral and Spatial Resolution MR Imaging (HiSS) What it is How to do it Ways to use it HiSS for Radiation

More information

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers Krista Chytyk MSc student Supervisor: Dr. Boyd McCurdy Introduction The objective of cancer radiotherapy

More information

Conference Biomedical Engineering

Conference Biomedical Engineering Automatic Medical Image Analysis for Measuring Bone Thickness and Density M. Kovalovs *, A. Glazs Image Processing and Computer Graphics Department, Riga Technical University, Latvia * E-mail: mihails.kovalovs@rtu.lv

More information

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 16 Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 3.1 MRI (magnetic resonance imaging) MRI is a technique of measuring physical structure within the human anatomy. Our proposed research focuses

More information

Use of MRI in Radiotherapy: Technical Consideration

Use of MRI in Radiotherapy: Technical Consideration Use of MRI in Radiotherapy: Technical Consideration Yanle Hu, PhD Department of Radiation Oncology, Mayo Clinic Arizona 04/07/2018 2015 MFMER slide-1 Conflict of Interest: None 2015 MFMER slide-2 Objectives

More information

The MSKCC Approach to IMRT. Outline

The MSKCC Approach to IMRT. Outline The MSKCC Approach to IMRT Spiridon V. Spirou, PhD Department of Medical Physics Memorial Sloan-Kettering Cancer Center New York, NY Outline Optimization Field splitting Delivery Independent verification

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

Female pelvic synthetic CT generation based on joint intensity and shape analysis

Female pelvic synthetic CT generation based on joint intensity and shape analysis 2949 Institute of Physics and Engineering in Medicine Physics in Medicine & Biology https://doi.org/0.088/36-6560/62/8/2935 Female pelvic synthetic CT generation based on joint intensity and shape analysis

More information

A Radiometry Tolerant Method for Direct 3D/2D Registration of Computed Tomography Data to X-ray Images

A Radiometry Tolerant Method for Direct 3D/2D Registration of Computed Tomography Data to X-ray Images A Radiometry Tolerant Method for Direct 3D/2D Registration of Computed Tomography Data to X-ray Images Transfer Function Independent Registration Boris Peter Selby 1, Georgios Sakas 2, Stefan Walter 1,

More information

Motion artifact detection in four-dimensional computed tomography images

Motion artifact detection in four-dimensional computed tomography images Motion artifact detection in four-dimensional computed tomography images G Bouilhol 1,, M Ayadi, R Pinho, S Rit 1, and D Sarrut 1, 1 University of Lyon, CREATIS; CNRS UMR 5; Inserm U144; INSA-Lyon; University

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti Monte Carlo methods in proton beam radiation therapy Harald Paganetti Introduction: Proton Physics Electromagnetic energy loss of protons Distal distribution Dose [%] 120 100 80 60 40 p e p Ionization

More information

UNIVERSITY OF SOUTHAMPTON

UNIVERSITY OF SOUTHAMPTON UNIVERSITY OF SOUTHAMPTON PHYS2007W1 SEMESTER 2 EXAMINATION 2014-2015 MEDICAL PHYSICS Duration: 120 MINS (2 hours) This paper contains 10 questions. Answer all questions in Section A and only two questions

More information

Basics of treatment planning II

Basics of treatment planning II Basics of treatment planning II Sastry Vedam PhD DABR Introduction to Medical Physics III: Therapy Spring 2015 Dose calculation algorithms! Correction based! Model based 1 Dose calculation algorithms!

More information

RADIOMICS: potential role in the clinics and challenges

RADIOMICS: potential role in the clinics and challenges 27 giugno 2018 Dipartimento di Fisica Università degli Studi di Milano RADIOMICS: potential role in the clinics and challenges Dr. Francesca Botta Medical Physicist Istituto Europeo di Oncologia (Milano)

More information

DUAL-ENERGY CT IN PROTON THERAPY

DUAL-ENERGY CT IN PROTON THERAPY 10/31/17 DUAL-ENERGY CT IN PROTON THERAPY Isabel Almeida, MAASTRO Clinic 7th NCS Lustrum Symposium 1 10/31/17 http://zonptc.bouwwebcam.nl https://www.youtube.com/watch?v=3vvvf5bqn7g Range uncertainties

More information

Determination of rotations in three dimensions using two-dimensional portal image registration

Determination of rotations in three dimensions using two-dimensional portal image registration Determination of rotations in three dimensions using two-dimensional portal image registration Anthony E. Lujan, a) James M. Balter, and Randall K. Ten Haken Department of Nuclear Engineering and Radiological

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

Michael Speiser, Ph.D.

Michael Speiser, Ph.D. IMPROVED CT-BASED VOXEL PHANTOM GENERATION FOR MCNP MONTE CARLO Michael Speiser, Ph.D. Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX September 1 st, 2012 CMPWG Workshop Medical

More information

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

Coverage based treatment planning to accommodate organ deformable motions and contouring uncertainties for prostate treatment. Huijun Xu, Ph.D.

Coverage based treatment planning to accommodate organ deformable motions and contouring uncertainties for prostate treatment. Huijun Xu, Ph.D. Coverage based treatment planning to accommodate organ deformable motions and contouring uncertainties for prostate treatment Huijun Xu, Ph.D. Acknowledgement and Disclosure Dr. Jeffrey Siebers Dr. DJ

More information

IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques

IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques Report of the Coordinated Research Project (CRP) on Development of Procedures

More information

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT CLINICAL COMMISSIONING AND ACCEPTANCE TESTING OF A 3D SURFACE MATCHING SYSTEM Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago Learning Objective Describe acceptance

More information

Fmri Spatial Processing

Fmri Spatial Processing Educational Course: Fmri Spatial Processing Ray Razlighi Jun. 8, 2014 Spatial Processing Spatial Re-alignment Geometric distortion correction Spatial Normalization Smoothing Why, When, How, Which Why is

More information

MR-guided radiotherapy: Vision, status and research at the UMC Utrecht. Dipl. Ing. Dr. Markus Glitzner

MR-guided radiotherapy: Vision, status and research at the UMC Utrecht. Dipl. Ing. Dr. Markus Glitzner MR-guided radiotherapy: Vision, status and research at the UMC Utrecht Dipl. Ing. Dr. Markus Glitzner About myself Training Medizintechnik TU Graz PhD UMC Utrecht Clinical work Software implementation

More information

COMPREHENSIVE QUALITY CONTROL OF NMR TOMOGRAPHY USING 3D PRINTED PHANTOM

COMPREHENSIVE QUALITY CONTROL OF NMR TOMOGRAPHY USING 3D PRINTED PHANTOM COMPREHENSIVE QUALITY CONTROL OF NMR TOMOGRAPHY USING 3D PRINTED PHANTOM Mažena MACIUSOVIČ *, Marius BURKANAS *, Jonas VENIUS *, ** * Medical Physics Department, National Cancer Institute, Vilnius, Lithuania

More information

IMRT site-specific procedure: Prostate (CHHiP)

IMRT site-specific procedure: Prostate (CHHiP) IMRT site-specific procedure: Prostate (CHHiP) Scope: To provide site specific instructions for the planning of CHHIP IMRT patients Responsibilities: Radiotherapy Physicists, HPC Registered Therapy Radiographers

More information

Quantitative imaging for clinical dosimetry

Quantitative imaging for clinical dosimetry Quantitative imaging for clinical dosimetry Irène Buvat Laboratoire d Imagerie Fonctionnelle U678 INSERM - UPMC CHU Pitié-Salpêtrière, Paris buvat@imed.jussieu.fr http://www.guillemet.org/irene Methodology

More information

2.1 Signal Production. RF_Coil. Scanner. Phantom. Image. Image Production

2.1 Signal Production. RF_Coil. Scanner. Phantom. Image. Image Production An Extensible MRI Simulator for Post-Processing Evaluation Remi K.-S. Kwan?, Alan C. Evans, and G. Bruce Pike McConnell Brain Imaging Centre, Montreal Neurological Institute, McGill University, Montreal,

More information

Auto-Segmentation Using Deformable Image Registration. Disclosure. Objectives 8/4/2011

Auto-Segmentation Using Deformable Image Registration. Disclosure. Objectives 8/4/2011 Auto-Segmentation Using Deformable Image Registration Lei Dong, Ph.D. Dept. of Radiation Physics University of Texas MD Anderson Cancer Center, Houston, Texas AAPM Therapy Educational Course Aug. 4th 2011

More information

A simple method to test geometrical reliability of digital reconstructed radiograph (DRR)

A simple method to test geometrical reliability of digital reconstructed radiograph (DRR) JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 1, WINTER 2010 A simple method to test geometrical reliability of digital reconstructed radiograph (DRR) Stefania Pallotta, a Marta Bucciolini

More information

Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs

Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Doug Dean Final Project Presentation ENGN 2500: Medical Image Analysis May 16, 2011 Outline Review of the

More information

Virtual Phantoms for IGRT QA

Virtual Phantoms for IGRT QA TM Virtual Phantoms for IGRT QA Why ImSimQA? ImSimQA was developed to overcome the limitations of physical phantoms for testing modern medical imaging and radiation therapy software systems, when there

More information

3-D Compounding of B-Scan Ultrasound Images

3-D Compounding of B-Scan Ultrasound Images 3-D Compounding of B-Scan Ultrasound Images Jochen F. Krücker, Charles R. Meyer, Theresa A. Tuthill, Gerald L. LeCarpentier, J. Brian Fowlkes, Paul L. Carson University of Michigan, Dept. of Radiology,

More information

VALIDATION OF DIR. Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology

VALIDATION OF DIR. Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology VALIDATION OF DIR Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology Overview Basics: Registration Framework, Theory Discuss Validation techniques Using Synthetic CT data & Phantoms What metrics to

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

CT Protocol Clinical Graphics Move Forward 3D motion analysis service

CT Protocol Clinical Graphics Move Forward 3D motion analysis service CT Protocol Clinical Graphics Move Forward 3D motion analysis service Version 1.4 Description of contents This document describes the CT protocol of scans for Clinical Graphics Move Forward 3D Motion Simulation

More information

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space White Pixel Artifact Caused by a noise spike during acquisition Spike in K-space sinusoid in image space Susceptibility Artifacts Off-resonance artifacts caused by adjacent regions with different

More information

Commissioning of MR-only simulation for radiotherapy planning

Commissioning of MR-only simulation for radiotherapy planning Radiation Oncology Ingenia MR-RT MR-only sim Commissioning of MR-only simulation for radiotherapy planning Gerald Schubert, Teuvo Vaara, Matti Lindström, Reko Kemppainen, and Marieke van Grootel-Rensen

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept.

Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. 2-D D Dose-CT Mapping in Geant4 Hidenobu Tachibana The Cancer Institute Hospital of JFCR, Radiology Dept. The Cancer Institute of JFCR, Physics Dept. Table of Contents Background & Purpose Materials Methods

More information

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street MRI is located in the sub basement of CC wing. From Queen or Victoria, follow the baby blue arrows and ride the CC south

More information

1. Learn to incorporate QA for surface imaging

1. Learn to incorporate QA for surface imaging Hania Al-Hallaq, Ph.D. Assistant Professor Radiation Oncology The University of Chicago ***No disclosures*** 1. Learn to incorporate QA for surface imaging into current QA procedures for IGRT. 2. Understand

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

Super-resolution Reconstruction of Fetal Brain MRI

Super-resolution Reconstruction of Fetal Brain MRI Super-resolution Reconstruction of Fetal Brain MRI Ali Gholipour and Simon K. Warfield Computational Radiology Laboratory Children s Hospital Boston, Harvard Medical School Worshop on Image Analysis for

More information

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE MRES 7005 - Fast Imaging Techniques Module 4 K-Space Symmetry Review K-Space Review K-Space Symmetry Partial or Fractional Echo Half or Partial Fourier HASTE Conditions for successful reconstruction Interpolation

More information

Reproducibility of interactive registration of 3D CT and MR pediatric treatment planning head images

Reproducibility of interactive registration of 3D CT and MR pediatric treatment planning head images JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 3, SUMMER 2001 Reproducibility of interactive registration of 3D CT and MR pediatric treatment planning head images Jaap Vaarkamp* Department

More information

Thresholding technique for accurate analysis of density and geometry in QCT, pqct and ÌCT images

Thresholding technique for accurate analysis of density and geometry in QCT, pqct and ÌCT images J Musculoskelet Neuronal Interact 2007; 7(1):9-16 2 nd International Conference on Osteoporosis and Bone Research, October 19-23, 2005 Hylonome Thresholding technique for accurate analysis of density and

More information

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland Metal Artifact Reduction CT Techniques R S S S Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland N. 1 v o 4 1 0 2. Postoperative CT Metal Implants CT is accurate for assessment

More information

New Technology Allows Multiple Image Contrasts in a Single Scan

New Technology Allows Multiple Image Contrasts in a Single Scan These images were acquired with an investigational device. PD T2 T2 FLAIR T1 MAP T1 FLAIR PSIR T1 New Technology Allows Multiple Image Contrasts in a Single Scan MR exams can be time consuming. A typical

More information

David Wagner, Kaan Divringi, Can Ozcan Ozen Engineering

David Wagner, Kaan Divringi, Can Ozcan Ozen Engineering Internal Forces of the Femur: An Automated Procedure for Applying Boundary Conditions Obtained From Inverse Dynamic Analysis to Finite Element Simulations David Wagner, Kaan Divringi, Can Ozcan Ozen Engineering

More information

Abstract. 1. Introduction

Abstract. 1. Introduction A New Automated Method for Three- Dimensional Registration of Medical Images* P. Kotsas, M. Strintzis, D.W. Piraino Department of Electrical and Computer Engineering, Aristotelian University, 54006 Thessaloniki,

More information

Automated Quality Assurance for Image-Guided Radiation Therapy

Automated Quality Assurance for Image-Guided Radiation Therapy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 1, WINTER 2009 Automated Quality Assurance for Image-Guided Radiation Therapy Eduard Schreibmann, a Eric Elder, Tim Fox Department of Radiation

More information

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

A software tool for the quantitative evaluation of 3D dose calculation algorithms

A software tool for the quantitative evaluation of 3D dose calculation algorithms A software tool for the quantitative evaluation of 3D dose calculation algorithms William B. Harms, Sr., Daniel A. Low, John W. Wong, a) and James A. Purdy Washington University School of Medicine, Mallinckrodt

More information

An approach for measuring the spatial orientations of a computed-tomography simulation system

An approach for measuring the spatial orientations of a computed-tomography simulation system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 2, 2014 An approach for measuring the spatial orientations of a computed-tomography simulation system Meng Chia Wu, a Ramani Ramaseshan Department

More information

Effect of CT number stopping power conversion uncertainties on dose distributions

Effect of CT number stopping power conversion uncertainties on dose distributions Effect of CT number stopping power conversion uncertainties on dose distributions Stella Flampouri Roelf Slopsema, Daniel Yeung, Zuofeng Li 23 May 2007 PTCOG Intro Treatment planning system dose calculations

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending

3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending 3DVH : On The Accuracy Of The Planned Dose Perturbation Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools *Patent Pending introduction State-of-the-art IMRT QA of static gantry

More information

Is deformable image registration a solved problem?

Is deformable image registration a solved problem? Is deformable image registration a solved problem? Marcel van Herk On behalf of the imaging group of the RT department of NKI/AVL Amsterdam, the Netherlands DIR 1 Image registration Find translation.deformation

More information

Tomographic Reconstruction

Tomographic Reconstruction Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)

More information

A Virtual MR Scanner for Education

A Virtual MR Scanner for Education A Virtual MR Scanner for Education Hackländer T, Schalla C, Trümper A, Mertens H, Hiltner J, Cramer BM Hospitals of the University Witten/Herdecke, Department of Radiology Wuppertal, Germany Purpose A

More information

M R I Physics Course

M R I Physics Course M R I Physics Course Multichannel Technology & Parallel Imaging Nathan Yanasak, Ph.D. Jerry Allison Ph.D. Tom Lavin, B.S. Department of Radiology Medical College of Georgia References: 1) The Physics of

More information

Advancements in molecular medicine

Advancements in molecular medicine Advancements in molecular medicine Philips Ingenuity TF PET/MR attenuation correction Z. Hu, 1 S. Renisch, 2 B. Schweizer, 3 T. Blaffert, 2 N. Ojha, 1 T. Guo, 1 J. Tang, 1 C. Tung, 1 J. Kaste, 1 V. Schulz,

More information

CT vs. VolumeScope: image quality and dose comparison

CT vs. VolumeScope: image quality and dose comparison CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,

More information

Acknowledgments and financial disclosure

Acknowledgments and financial disclosure AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories

More information

NIH Public Access Author Manuscript Med Phys. Author manuscript; available in PMC 2009 March 13.

NIH Public Access Author Manuscript Med Phys. Author manuscript; available in PMC 2009 March 13. NIH Public Access Author Manuscript Published in final edited form as: Med Phys. 2008 February ; 35(2): 660 663. Prior image constrained compressed sensing (PICCS): A method to accurately reconstruct dynamic

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

Geometric distortion evaluation using a multi-orientated water-phantom at 0.2 T MRI

Geometric distortion evaluation using a multi-orientated water-phantom at 0.2 T MRI Bio-Medical Materials and Engineering 26 (2015) S1431 S1438 OI 10.3233/BME-151441 IOS Press S1431 Geometric distortion evaluation using a multi-orientated water-phantom at 0.2 T MRI Jochi Jao a, Mingye

More information

Fundamentals of CT imaging

Fundamentals of CT imaging SECTION 1 Fundamentals of CT imaging I History In the early 1970s Sir Godfrey Hounsfield s research produced the first clinically useful CT scans. Original scanners took approximately 6 minutes to perform

More information

REAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT

REAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT REAL-TIME ADAPTIVITY IN HEAD-AND-NECK AND LUNG CANCER RADIOTHERAPY IN A GPU ENVIRONMENT Anand P Santhanam Assistant Professor, Department of Radiation Oncology OUTLINE Adaptive radiotherapy for head and

More information

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Stefan C. Saur 1, Hatem Alkadhi 2, Luca Regazzoni 1, Simon Eugster 1, Gábor Székely 1, Philippe Cattin 1,3 1 Computer Vision

More information

Post-implant computed tomography magnetic resonance prostate image registration using feature line parallelization and normalized mutual information

Post-implant computed tomography magnetic resonance prostate image registration using feature line parallelization and normalized mutual information JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 8, NUMBER 1, WINTER 2007 Post-implant computed tomography magnetic resonance prostate image registration using feature line parallelization and normalized

More information

Sphere Extraction in MR Images with Application to Whole-Body MRI

Sphere Extraction in MR Images with Application to Whole-Body MRI Sphere Extraction in MR Images with Application to Whole-Body MRI Christian Wachinger a, Simon Baumann a, Jochen Zeltner b, Ben Glocker a, and Nassir Navab a a Computer Aided Medical Procedures (CAMP),

More information

Prostate Detection Using Principal Component Analysis

Prostate Detection Using Principal Component Analysis Prostate Detection Using Principal Component Analysis Aamir Virani (avirani@stanford.edu) CS 229 Machine Learning Stanford University 16 December 2005 Introduction During the past two decades, computed

More information

CHAPTER 6 MODIFIED FUZZY TECHNIQUES BASED IMAGE SEGMENTATION

CHAPTER 6 MODIFIED FUZZY TECHNIQUES BASED IMAGE SEGMENTATION CHAPTER 6 MODIFIED FUZZY TECHNIQUES BASED IMAGE SEGMENTATION 6.1 INTRODUCTION Fuzzy logic based computational techniques are becoming increasingly important in the medical image analysis arena. The significant

More information

Patient Set-ups and Tumor Localizations

Patient Set-ups and Tumor Localizations Patient Set-ups and Tumor Localizations Amy S. Harrison Patient Positioning Prior to starting any localization or simulation procedure patients need to be positioned and immobilized Patients disease location

More information

MRI Physics II: Gradients, Imaging

MRI Physics II: Gradients, Imaging MRI Physics II: Gradients, Imaging Douglas C., Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor Magnetic Fields in MRI B 0 The main magnetic field. Always on (0.5-7 T) Magnetizes

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions

(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions EE225E/BIOE265 Spring 2013 Principles of MRI Miki Lustig Assignment 8 Solutions 1. Nishimura 7.1 P)jc%z 5 ivcr3. 1. I Due Wednesday April 10th, 2013 (a Scrhon5 R2iwd b 0 ZOms Xn,s r cx > qs 4-4 8ni6 4

More information

An experimental investigation on the effect of beam angle optimization on the reduction of beam numbers in IMRT of head and neck tumors

An experimental investigation on the effect of beam angle optimization on the reduction of beam numbers in IMRT of head and neck tumors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 An experimental investigation on the effect of beam angle optimization on the reduction of beam numbers in IMRT of head and neck tumors

More information

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy T. Rohlfing a, D. Zerfowski b, J. Beier a, P. Wust a, N. Hosten a, R. Felix a a Department of

More information

Supplementary methods

Supplementary methods Supplementary methods This section provides additional technical details on the sample, the applied imaging and analysis steps and methods. Structural imaging Trained radiographers placed all participants

More information

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic

More information

An Investigation of a Model of Percentage Depth Dose for Irregularly Shaped Fields

An Investigation of a Model of Percentage Depth Dose for Irregularly Shaped Fields Int. J. Cancer (Radiat. Oncol. Invest): 96, 140 145 (2001) 2001 Wiley-Liss, Inc. Publication of the International Union Against Cancer An Investigation of a Model of Percentage Depth Dose for Irregularly

More information

Good Morning! Thank you for joining us

Good Morning! Thank you for joining us Good Morning! Thank you for joining us Deformable Registration, Contour Propagation and Dose Mapping: 101 and 201 Marc Kessler, PhD, FAAPM The University of Michigan Conflict of Interest I receive direct

More information

TG 132: Use of Image Registration and Fusion in RT

TG 132: Use of Image Registration and Fusion in RT TG 132: Use of Image Registration and Fusion in RT Kristy K Brock, PhD, DABR, FAAPM Associate Professor Department of Radiation Oncology, University of Michigan Chair, AAPM TG 132: Image Registration and

More information

Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI. Floris Jansen, GE Healthcare July, 2015

Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI. Floris Jansen, GE Healthcare July, 2015 Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI Floris Jansen, GE Healthcare July, 2015 PET/MR 101 : challenges Thermal Workflow & Apps RF interactions?!!

More information