Tech Note: Measurement of New Ridge Technologies PMD Sources with Adaptif s A2000
|
|
- Katrina Higgins
- 6 years ago
- Views:
Transcription
1 Tech Note: adaptif A2000 set up to measure NRT-40133A DGD states Tunable Laser adaptif A2000 PMD measurement NRT PMDS (DUT) Settings at NRT-40133A 1. Calibrate at desired wavelength (e.g nm) 2. Set to PMD state to (0,0) New Ridge Technologies, LLC Page 1 adaptif PHOTONICS GmbH
2 Settings at adaptif A2000 polarization NAVIGATOR 3. Double-click on PDL/PMD measurement from the Application folder at left of main screen 4. Set Sweep Rate to 5.0 nm/sec 5. Start Wavelength = below the NRT-40133A calibration wavelength (e.g if calibrated to nm) 6. End Wavelength = above the NRT-40133A calibration wavelength (e.g if calibrated to nm) 7. Run measurement by pressing on Single Measurement adaptif A2000 Post Processing 8. Set Resolution (nm) to or For rapidly varying DGD spectra smaller resolution may be necessary to capture all of the features. However, smaller resolution may also increase the measurement noise. This often results in incorrect SOPMD, PDC and/or Depolarization outputs as they are calculated from derivatives of the noisier signal. New Ridge Technologies, LLC Page 2 adaptif PHOTONICS GmbH
3 Comparison of NRT-40133A DGD state with of ADAPTIF A2000 Measurement Set the NRT-40133A to the desired state, and then measure with the adaptif A2000 ( Single Measurement ). After scanning the PMD state set Y-axis on adaptif A2000 and measure the DGD max and min across the scan. Repeat this for the SOPMD, Depolarization and the PCD on the A2000. The scaling can be changed using the arrow-buttons (automatic scaling) or the tool-button (manual scaling) on the toolbar on the left of the screen: This should match the DGD spectrum on the PMDS (PMDS Tools Spectral and PSP Viewer). New Ridge Technologies, LLC Page 3 adaptif PHOTONICS GmbH
4 adaptif A2000 set up to measure NRT-10033A DGD states 1. Calibrate at desired wavelength (e.g nm) 2. Set to PMD state to (0,0). Settings at adaptif A2000 polarization NAVIGATOR 3. Double-click on PDL/PMD measureme nt from the Application folder at left of main screen 4. Set Sweep Rate to 5.0 nm/sec 5. Start Wavelength = below the NRT-40133A calibration wavelength (e.g if calibrated to nm) 6. End Wavelength = above the NRT-40133A calibration wavelength (e.g if calibrated to nm) New Ridge Technologies, LLC Page 4 adaptif PHOTONICS GmbH
5 7. Run measurement by pressing on Single Measurement adaptif A2000 Post Processing 8. Set Resolution (nm) to or For rapidly varying DGD spectra smaller resolution may be necessary to capture all of the features. However, smaller resolution may also increase the measurement noise. This often results in incorrect SOPMD, PDC and/or Depolarization outputs as they are calculated from derivatives of the noisier signal. Comparison of NRT-10033A DGD state with of ADAPTIF A2000 Measurement Set the NRT-10033A to the desired state, and then measure with the adaptif A2000 ( Single Measurement ). After scanning the PMD state set Y-axis on adaptif A2000 and measure the DGD max and min across the scan. Repeat this for the SOPMD, Depolarization and the PCD on the A2000. The scaling can be changed using the arrow-buttons (automatic scaling) or the tool-button (manual scaling) on the toolbar on the left of the screen: New Ridge Technologies, LLC Page 5 adaptif PHOTONICS GmbH
6 This should match the DGD spectrum on the PMDS (PMDS Tools Spectral and PSP Viewer), as shown with the NRT-10033A example above. Below is an example and comparison of measurements made on the PMD state (0, 1700) generated by an NRT-10033A. The green data points connected with the green traces were measured employing the Jones Matrix Eigen-analysis measurement technique with an Agilent 8509B. This measurement is time consuming and cumbersome although quite accurate. The blue traces were measured with the Adaptif A2000. The A2000 is capable of much wider PMD spectral scans, in far less time, and provides the same results. On the left side are superimposed JME measurements. The wider blue traces were made with the A2000. On the right side is the NRT-10033A spectral and PSP viewer for the state corresponding to the set state (0, 1700). The A2000 accurately measures the PMD set PMD state. Note the difference between the measured PCD and the set PCD is a factor of 1. After researching into the definition of PCD both Adaptif and NRT agree on the PCD standard shown at the right. New Ridge Technologies, LLC Page 6 adaptif PHOTONICS GmbH
MTS/T-BERD 8000 Platform Optical Spectrum Analyzer Module
COMMUNICATIONS TEST & MEASUREMENT SOLUTIONS MTS/T-BERD 8000 Platform Optical Spectrum Analyzer Module MTS/T-BERD 8000 platform Key Features New in-band OSA version for measuring the true OSNR in ROADM
More informationConfiguration of Light Sources
UseCase.0012 (1.0) Configuration of Light Sources Keywords: LED, laser, harmonic fields, harmonic fields set, spectra, pulses, polarization Description This use case explains how light sources are configured
More informationUsing OPUS to Process Evolved Gas Data (8/12/15 edits highlighted)
Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted) Once FTIR data has been acquired for the gases evolved during your DSC/TGA run, you will process using the OPUS software package. Select
More informationIntroduction: Experiment 1: Wave Properties of Light
Natural Order Properties of Light Lab Introduction: In this lab we will explore the wave and particle nature of light. In the first experiment we will measure the diffraction pattern of light as it passes
More informationThe LLCP series follows the Mueller Matrix design while the LSDP series uses conventional designs
Labnics Equipment offers the scientific community a range of polarimeters to meet various application requirements, while at the same time keeping the budget of the customer in mind. The LLCP series follows
More informationJasco FP-6500 Spectrofluorimeter Updated November 14, 2017
Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017 Instrument instructions can be found at: http://academic.bowdoin.edu/chemistry/resources/instructions.shtml If you have any problems with the
More informationRaman Spectrometer Installation Manual
RI Raman Spectrometer Instruction Manual Application software (included in standard configuration) Connection RI RAMAN is recognized by Windows operational system as standard HID device. That is why there
More informationOr in the U.S., call Agilent Technologies at (8 am - 8 pm EST)
This manual contains references to Adaptif Photonics GmbH. Please note that Adaptif Photonics GmbH is now part of Agilent Technologies. This manual describes products or services now available through
More informationMonoVista CRS+ Raman Microscopes
MonoVista CRS+ Benefits Deep UV to NIR wavelength range Up to 4 integrated multi-line lasers plus port for large external lasers Dual beam path for UV and VIS/NIR Motorized Laser selection Auto Alignment
More informationOIW-EX 1000 Oil in Water Monitors
OIW-EX 1000 Oil in Water Monitors Spectrometer Handbook Document code: OIW-HBO-0005 Version: EX-002 www.advancedsensors.co.uk Tel: +44(0)28 9332 8922. FAX +44(0)28 9332 8669 Page 1 of 33 Document History
More informationPOL-001 Fiber-Optic In-Line Polarizer. Operation Manual
POL-001 Fiber-Optic In-Line Polarizer Operation Manual Feb. 18, 2002 General Photonics Corp. Ph: (909) 590-5473 5228 Edison Ave. Fax: (909) 902-5536 Chino, CA 91710 USA www.generalphotonics.com Document
More informationQuick Value Statistics in Presentation
Quick Value Statistics in Presentation Presentation can show you a simple time series and distribution for any measured value, based on all loaded measurements. In the Data Contents tree, navigate to the
More informationSPECTRAL IMAGING VIEWER SOFTWARE MANUAL
103 Quality Circle, Suite 215 Huntsville, AL 35806 Phone: (256) 704-3332 Fax: (256) 971-2073 E-Mail: info@axometrics.com Website: http://www.axometrics.com SPECTRAL IMAGING VIEWER SOFTWARE MANUAL 2012
More informationOperating instructions - Cary 100 Bio UV-visible Spectrophotometer
Operating instructions - Cary 100 Bio UV-visible Spectrophotometer This document describes how to power up the spectrophotometer, set its measurement parameters, insert one or more of samples for analysis
More information40 and 100 Gigabit Ethernet Consortium Clause 86 40GBASE-SR4 and 100GBASE-SR10 PMD Test Suite v0.1 Technical Document
40 and 100 Gigabit Ethernet Consortium Clause 86 40GBASE-SR4 and 100GBASE-SR10 PMD Test Suite v0.1 Technical Document Last Updated: March 26, 2013 10:00am 40 and 100 Gigabit Ethernet Consortium 121 Technology
More informationOperating Procedure for Horiba Raman Microscope
Operating Procedure for Horiba Raman Microscope SAFETY Be aware of Laser radiation at all times! Do not remove the covers of the instrument. Components are supplied with 110V electric source. Do not touch
More informationCS260 Contractor Series OTDR
CS260 Contractor Series OTDR Quick Reference Guide www.aflglobal.com or (800) 321-5298, (603) 528-7780 Functional Keys KEY NAME KEY FUNCTION Power Press and hold (~1 second) to turn the CS260 on or off.
More informationSTS Spectrometer Getting Started
STS Spectrometer Getting Started STS is a family of compact, low-cost spectrometers ideal for embedding into OEM devices. STS provides remarkable performance for a spectrometer of its size (just 40 mm
More information67 Series Spectrophotometer PC Software
67 Series Spectrophotometer PC Software Instruction Manual 670 004/Rev D/04-17 IMPORTANT Please ensure the SD Card is fitted into the socket as detailed in the image below. -------------------------------------------------------------------------------------------------------------------------
More informationPCD 005 PolaMIX. Operation Manual
PCD 005 PolaMIX All-Fiber Polarization Scrambler Module with Microprocessor Controller Operation Manual July 29, 2011 General Photonics Corp. Ph: (909) 590-5473 5228 Edison Ave. Fax: (909) 902-5536 Chino,
More informationRadiance, Irradiance and Reflectance
CEE 6100 Remote Sensing Fundamentals 1 Radiance, Irradiance and Reflectance When making field optical measurements we are generally interested in reflectance, a relative measurement. At a minimum, measurements
More informationConsiderations on X00 Gb/s 40-80km interfaces with appropriate support for DWDM systems. Peter Stassar HUAWEI TECHNOLOGIES CO., LTD.
Considerations on X00 Gb/s 40-80km interfaces with appropriate support for DWDM systems Peter Stassar www.huawei.com HUAWEI TECHNOLOGIES CO., LTD. Co-author(s) Pete Anslow (Ciena), providing valuable feedback
More informationLED MINISPECTROMETER LMS-R
LED MINISPECTROMETER LMS-R TECHNICAL DESCRIPTION rev. 290518 CONTENTS General information 3 Device and applications overview 3 Main features 3 Device main view and operation principle 4-5 Package content
More informationDLIT and FLIT Reconstruction of Sources
Concept Tech Note 5 DLIT and FLIT Reconstruction of Sources Diffuse Luminescence Imaging Tomography (DLIT) is a technique that analyzes images of the surface light emission from a living subject to generate
More informationSpectrometer Visible Light Spectrometer V4.4
Visible Light Spectrometer V4.4 Table of Contents Package Contents...3 Trademarks...4 Manual Driver and Application installation...5 Manual Application Installation...6 First Start of the Application...8
More informationUNIT 102-9: INTERFERENCE AND DIFFRACTION
Name St.No. - Date(YY/MM/DD) / / Section Group # UNIT 102-9: INTERFERENCE AND DIFFRACTION Patterns created by interference of light in a thin film. OBJECTIVES 1. Understand the creation of double-slit
More informationAKELA Vector Network Analyzer (VNA) Quick Start Guide
AKELA Vector Network Analyzer (VNA) Quick Start Guide Copyright AKELA, Inc. 2012, all rights reserved http:\\akelainc.com LabVIEW and LabWindows are registered trademarks of National Instruments Incorporated
More informationAlpy 600 Mixed Mode Calibration
Alpy 600 Mixed Mode Calibration Paul Luckas October, 2016 1 Introduction Alpy 600 Calibration using the mixed mode involves the following basic steps. 1. Use a sequence of spectra from a type A or B star
More informationOptical Spectrum Analyzer AQ6330. Full featured optical spectrum analyzer with compact and lightweight design
Optical Spectrum Analyzer AQ6330 Full featured optical spectrum analyzer with compact and lightweight design The AQ6330 is like no other optical spectrum analyzer you have ever seen. It is compact, lightweight
More informationVCSEL-Friendly 1310nm Serial PMD Specifications
VCSEL-Friendly 1310nm Serial PMD Specifications Jack Jewell jljewell@picolight.com IEEE 802.3 Plenary - November 6-9, 2000 1 of 12 Why Tweak the Specs for VCSELs? 1310nm serial specs intended to be VCSEL-friendly
More informationWHITE PAPER. How to Generate a Ray Set from an RSMX Source Model. Zemax A Radiant Zemax Company
How to Generate a Ray Set from an RSMX Source Model How to Generate a Ray Set from an RSMX Source Model Introduction The most general description of a complex source is given in a Radiant Source Model
More informationInteractive comment on Quantification and mitigation of the impact of scene inhomogeneity on Sentinel-4 UVN UV-VIS retrievals by S. Noël et al.
Atmos. Meas. Tech. Discuss., www.atmos-meas-tech-discuss.net/5/c741/2012/ Author(s) 2012. This work is distributed under the Creative Commons Attribute 3.0 License. Atmospheric Measurement Techniques Discussions
More informationInteractive comment on Quantification and mitigation of the impact of scene inhomogeneity on Sentinel-4 UVN UV-VIS retrievals by S. Noël et al.
Atmos. Meas. Tech. Discuss., 5, C741 C750, 2012 www.atmos-meas-tech-discuss.net/5/c741/2012/ Author(s) 2012. This work is distributed under the Creative Commons Attribute 3.0 License. Atmospheric Measurement
More informationLeica 3D Disto PROJECTOR function for stair case applications
Leica 3D Disto PROJECTOR function for stair case applications Workflow description: Measurement Stair Design Set Out of fixing points Preparing the measurement 1 mark 1 point on each wall of the stair
More informationWelcome to noiselab Express 1.0
Welcome to noiselab Express 1.0 Basic Operation Download and install the program. The program automatically unzips and installs. If you don t have a license code, you can run the program in demo mode for
More informationStellarNet Single-Channel SpectraWiz for LabVIEW
StellarNet Single-Channel SpectraWiz for LabVIEW 18 October 2010 Overview The StellarNet SpectraWiz for LabVIEW spectrometry application provides an easy-to-use graphical user interface for any of up to
More informationFluorimeter User s Booklet
Fluorimeter User s Booklet 1) Verify that the instrument power and the computer are both off. Then turn on the lamp. The red indicator light will glow when the lamp is on. Turn on lamp switch here. This
More informationThis presentation takes you through Region of Interest (ROI) Analysis.
This presentation takes you through Region of Interest (ROI) Analysis. There are many reasons to pursue this time of manual selection of spectrum, it can include reducing variation and selecting peaks
More informationRoughness measuring systems m rail trolley m rail m wheel
Precision measuring systems for the determination of rail and wheel roughness Roughness measuring systems m rail trolley m rail m wheel www.muellerbbm-rail.com High-precision measuring systems for rough
More informationIntroduction to Raman spectroscopy measurement data processing using Igor Pro
Introduction to Raman spectroscopy measurement data processing using Igor Pro This introduction is intended to minimally guide beginners to processing Raman spectroscopy measurement data, which includes
More informationNon-scanning Hyperspectral Imaging and Spectral Video Technique for Object Documentation
Non-scanning Hyperspectral Imaging and Spectral Video Technique for Object Documentation András Jung, René Michels & Rainer Graser Founders of Cubert GmbH and Developers of the Hyperspectral Frame Camera
More informationSUPPLEMENTARY INFORMATION
Supplementary Information Compact spectrometer based on a disordered photonic chip Brandon Redding, Seng Fatt Liew, Raktim Sarma, Hui Cao* Department of Applied Physics, Yale University, New Haven, CT
More informationAgilent 81910A Photonic All-Parameter Analyzer User s Guide
Agilent 81910A Photonic All-Parameter Analyzer User s Guide Agilent Technologies Notices Agilent Technologies, Inc. 2002-2005 No part of this manual may be reproduced in any form or by any means (including
More informationAON Agile Optical Networks Measuring the Optical Signal-to-Noise Ratio in Agile Optical
White Paper AON Agile Optical Networks Measuring the Optical Signal-to- Ratio in Agile Optical Introduction The mainstreaming of consumer broadband and the accelerated growth of enterprise traffic have
More informationSolar Module Quantum Efficiency Characterization System (LabVIEW Sheds Light on Solar Cell Quality Improvements)
Solar Module Quantum Efficiency Characterization System (LabVIEW Sheds Light on Solar Cell Quality Improvements) by Richard M. Brueggman President & CEO and Gregory C. Cala, Ph.D. Chief Technology Officer
More informationAgilent 8509A/B Lightwave Polarization Analyzer. User s Guide
Agilent 8509A/B Lightwave Polarization Analyzer User s Guide Copyright Agilent Technologies 1992, 1994, 2001 All Rights Reserved. Reproduction, adaptation, or translation without prior written permission
More informationOIW-EX Series of Oil In Water Analyzers MiView Handbook Document code: OIW-HBO-0013 Version: EX July 2013
OIW-EX Series of Oil In Water Analyzers MiView Handbook Document code: OIW-HBO-0013 Version: EX-004 17 July 2013 Tel: +44 (0) 28 9332 8922 Document History Change No. Version No. Description Month & Year
More informationRoHS COMPLIANT 2002/95/EC
Superlum Broadband Light Sources cblmd-series (2nd Generation) Compact Broadband Light Technical Product Specification Document Number SL.3328.00.000D3 June 2017 Revision 001 ATTENTION ELECTROSTATIC SENSITIVE
More informationOperation Manual of J.A. Woollam Ellipsometer
Operation Manual of J.A. Woollam Ellipsometer 1) Press both the Lamp power and lamp Ignition buttons to turn them on (The buttons are located on the front panel of the bottom control box). 2) Put your
More informationTime-Resolved measurements by FEL spontaneous emission: A proposal for sub-picosecond pumps & probe structural and spectrometric investigations
Time-Resolved measurements by FEL spontaneous emission: A proposal for sub-picosecond pumps & probe structural and spectrometric investigations V. Rossi Albertini, B. Paci & P. Perfetti Istituto di Struttura
More informationOptical Component Test Agilent Multi-Wavelength Meters
Optical Component Test Agilent Multi-Wavelength Meters High-performance wavelength measurements for characterizing next generation optical networks Optical Component Test The Agilent Family of Multi-Wavelength
More informationAxoStep TM Imaging Mueller Matrix Polarimeter
103 Quality Circle, Suite 215 Huntsville, AL 35806 Phone: (256) 704-3332 Fax: (256) 971-2073 E-Mail: info@axometrics.com Website: http://www.axometrics.com AxoStep TM Imaging Mueller Matrix Polarimeter
More informationIgorCD Procedures. Data -> Load CD Data
IgorCD Procedures IgorCD is a set of Igor Pro procedures for processing CD spectra and CD thermal melts. It correctly implements spectral smoothing functions and facilitates subtracting baselines (or constant
More informationMassHunter File Reader
MassHunter File Reader vers 1.0.0 2015 Quadtech Associates, Inc. All Rights Reserved Release date: November 18, 2015 www.quadtechassociates.com MassHunter File Reader Welcome to MassHunter File Reader.
More informationCVI SPECTRAL PRODUCTS. 111 Highland Drive Putnam CT, (860) SM32Pro
CVI SPECTRAL PRODUCTS 111 Highland Drive Putnam CT, 06260 (860) 928-5834 SM32Pro 2.8.28 Table of Contents Warranty and Liability 1 Quick Start Installation Guide 2 System Requirements 3 Requirements for
More information1. Getting Started: Brief Step-By-Step Guide (PDF File)
Page 1 of 5 1. Getting Started: Brief Step-By-Step Guide (PDF File) Leica SP2 confocal microscope is controlled via software LCS (Leica Confocal Software). The latest version is V2.5. Bellowing is a brief
More informationProtein Deconvolution Quick Start Guide
Protein Deconvolution Quick Start Guide The electrospray ionization (ESI) of intact peptides and proteins produces mass spectra containing a series of multiply charged ions with associated mass-to-charge
More informationAQ6375. "LONG-WAVE" Optical Spectrum Analyzer. Telecom and Beyond QUALITY INNOVATION FORESIGHT
AQ6375 "LONG-WAVE" Telecom and Beyond WORLD CLASS PERFORMANCE Long wavelength measurement capability High wavelength accuracy & resolution High sensitivity High speed measurement USB for mouse, keyboard,
More informationPowerMax-Pro USB and RS Sensors
PowerMax-Pro USB and RS Sensors 200 mw to 150W PowerMax-Pro USB and RS sensors incorporate LabMax-Pro instrumentation directly within the sensor cable. Similar to other Coherent USB and RS sensors, this
More informationTutorial: Instantaneous Measurement of M 2 Beam Propagation Ratio in Real-Time
Tutorial: Instantaneous Measurement of M 2 Beam Propagation Ratio in Real-Time By Allen M. Cary, Jeffrey L. Guttman, Razvan Chirita, Derrick W. Peterman, Photon Inc A new instrument design allows the M
More informationAgilent N5393C PCI Express Electrical Performance and Compliance Software Release Notes
Agilent N5393C PCI Express Electrical Performance and Compliance Software Release Notes Agilent N5393C Software Version 03.34 Released Date: 19 May 2014 File Name: SetupInfPCIExpress0334.exe Improved algorithm
More informationHyperspectral Chemical Imaging: principles and Chemometrics.
Hyperspectral Chemical Imaging: principles and Chemometrics aoife.gowen@ucd.ie University College Dublin University College Dublin 1,596 PhD students 6,17 international students 8,54 graduate students
More informationQ Exactive HF-X Software Manual
Q Exactive HF-X Software Manual Exactive Series 2.9 BRE0012262 Revision A May 2017 Q Exactive HF-X Software Manual Exactive Series 2.9 BRE0012262 Revision A May 2017 Legal Notices 2017 Thermo Fisher Scientific
More informationOpen Excel by following the directions listed below: Click on Start, select Programs, and the click on Microsoft Excel.
Candy is Dandy Grading Rubric You have been hired to conduct some market research about M&M's. First, you had your team purchase 4 large bags and the results are given for the contents of those bags. You
More informationInstytut Fizyki Doświadczalnej Wydział Matematyki, Fizyki i Informatyki UNIWERSYTET GDAŃSKI
Instytut Fizyki Doświadczalnej Wydział Matematyki, Fizyki i Informatyki UNIWERSYTET GDAŃSKI I. Background theory. 1. Characteristics of the apparatus: prismatic, grating, interferometers. 2. Operating
More information7310 and 7315 Glossary of Icons. Operating Manual
7310 and 7315 Glossary of Icons Operating Manual Glossary of Icons Mode ICON Description Back key Tick icon - Done/yes Cross icon Cancel/no Printer icon - Print/open printer settings No results to send
More informationOptical Loss Budgets
CHAPTER 4 The optical loss budget is an important aspect in designing networks with the Cisco ONS 15540. The optical loss budget is the ultimate limiting factor in distances between nodes in a topology.
More informationPI Scanner User Guide
PI Scanner User Guide Table of Contents 1. Highlights 2. Overview 3. Installation 3.1. PI Scanner Software Installation 3.2. USB to Serial Interface Board Installation 3.3. Programming the PI Scanner IP
More informationMultivariate Calibration Quick Guide
Last Updated: 06.06.2007 Table Of Contents 1. HOW TO CREATE CALIBRATION MODELS...1 1.1. Introduction into Multivariate Calibration Modelling... 1 1.1.1. Preparing Data... 1 1.2. Step 1: Calibration Wizard
More informationA Walk Through the MSA Software Spectrum Analyzer Mode 12/12/09
A Walk Through the MSA Software Spectrum Analyzer Mode 12/12/09 This document is intended to familiarize you with the basic features of the MSA and its software, operating as a Spectrum Analyzer, without
More informationOPTICAL COHERENCE TOMOGRAPHY:SIGNAL PROCESSING AND ALGORITHM
OPTICAL COHERENCE TOMOGRAPHY:SIGNAL PROCESSING AND ALGORITHM OCT Medical imaging modality with 1-10 µ m resolutions and 1-2 mm penetration depths High-resolution, sub-surface non-invasive or minimally
More informationStereo DEM Extraction from Radar Imagery Geomatica 2014 Tutorial
The purpose of this tutorial is to provide the steps necessary to extract a stereo DEM model from Radar imagery. This method of DEM extraction is useful if you do not have high resolution optical imagery
More informationdscope Series III Applications Manual - Auto Sequence by Ian Heaton
dscope Series III by Ian Heaton This manual is also available as 'on-line help' from the dscope software. You can access the on-line help from the 'Help' menu. The on-line version is context-sensitive:
More informationOptics for nonlinear microscopy
Optics for nonlinear microscopy Nonlinear microscopy Dispersion management Compact housing In-line input/output apertures High throughput Robust mechanical design Latest generations of Dispersive Mirrors
More information1 Introduction. ThinPrint Client Installation Page 1
ThinPrint Client Installation Page 1 1 Introduction In order to reduce the amount of bandwidth used when printing and to improve the overall performance and online experience for the customers that we
More informationFTB-5240S/BP NETWORK TESTING OPTICAL
Optical Spectrum analyzers FTB-5240S/BP NETWORK TESTING OPTICAL Highly accurate OSAs for current and next-generation networks Best-in-class resolution bandwidth Truly portable spectral characterization
More informationInstruction Manual of Modelling of Trap Detectors Using Matlab
Aalto University School of Electrical Engineering Metrology Research Institute Mikko Puranen Tuomas Poikonen Timo Dönsberg Petri Kärhä Instruction Manual of Modelling of Trap Detectors Using Matlab Version
More informationAgilent 6400 Series Triple Quadrupole LC/MS System
Agilent 6400 Series Triple Quadrupole LC/MS System Quick Start Guide Where to find information 4 Getting Started 6 Step 1. Start the Data Acquisition software 7 Step 2. Prepare the LC modules 13 Step 3.
More informationRgBLase LLC P/N: FB 671 XXX Diode Pumped Solid State Laser Module
Key Features: 671nm output, TEM00 Fiber Coupled (optional) Remote control TTL Modulation (optional) ESD protection Plug & Play Low Noise (optional) Applications: Bio Technology Photo Finishing Semiconductor
More informationInstruction for Mass Spectra Viewer
Instruction for Mass Spectra Viewer 1. A general description Basic functions Setup and Pre-process View, browser and undo Grouping, highlighting and labeling 2. Setup and Pre-process These procedures must
More informationCoSMoS Starter Guide 1.4 Hoskins Lab Last Modified 03/03/2017 Chris DeCiantis
Contents Start Up:... 2 Turn on Shutter controllers... 2 Turn on Lasers... 2 Turn on other equipment:... 2 Load Slide onto Sample Holder:... 2 Turn on Computer/ Start Labview/Glimpse:... 3 Shutters Activate.vi
More informationAgilent MassHunter Qualitative Data Analysis
Agilent MassHunter Qualitative Data Analysis Qualitative Navigator B.08.00 Presenters: Kevin Costalunga Stephen Harnos With Matt Leyden & Kevin Costalunga 1 MassHunter Webinar Series MassHunter Qualitative
More information(0, 1, 1) (0, 1, 1) (0, 1, 0) What is light? What is color? Terminology
lecture 23 (0, 1, 1) (0, 0, 0) (0, 0, 1) (0, 1, 1) (1, 1, 1) (1, 1, 0) (0, 1, 0) hue - which ''? saturation - how pure? luminance (value) - intensity What is light? What is? Light consists of electromagnetic
More informationGeoTASO Project Summary and Relevance. Jim Leitch Ball Aerospace
GeoTASO Project Summary and Relevance Jim Leitch Ball Aerospace jleitch@ball.com 303-939-5280 4/23/2013 Sensor Concept Overview Airborne nadir-viewing wide-swath imaging spectrometer Two channel spectrometer
More informationSystemVue 2011 Fundamentals (version ) LAB EXERCISE 1. SystemVue Basics
SystemVue 2011 Fundamentals (version 1-2012) LAB EXERCISE 1 SystemVue Basics This lab exercise introduces the basic use model of SystemVue, including using the built-in templates, running analyses and
More informationStereo DEM Extraction from Radar Imagery Geomatica 2015 Tutorial
Stereo DEM Extraction from Radar Imagery Geomatica 2015 Tutorial The purpose of this tutorial is to provide the steps necessary to extract a stereo DEM model from Radar imagery. This method of DEM extraction
More informationPositioning system of a metrological AFM: design considerations
Positioning system of a metrological AFM: design considerations AFM workshop LNE, Trappes Jan Piot K.U.Leuven Division PMA Overview Introduction General layout metrological AFM Layout of the positioning
More informationLeica 3D Disto PROJECTOR in stair case
Leica 3D Disto PROJECTOR in stair case Workflow description: Measurement Stair Design Projection of fixing points Preparing the measurement 1 mark 1 point on each wall of the stair case Measurement of
More informationTraining Tracker 6. Release Windows Mobile Software for OPN computer directions, llc 3 Gamecock Avenue Suite 308-A Charleston SC 29407
Training Tracker 6 Release 6.1.1 Windows Mobile Software for OPN-2001 computer directions, llc 3 Gamecock Avenue Suite 308-A Charleston SC 29407 Phone 843-571-2108 Fax 843-766-0105 www.comdirections.com
More informationTheory and Measuring of Antireflection Coatings
Theory and Measuring of Antireflection Coatings Steffen Lorch The characterization of antireflection (AR) coatings is not trivial. A preferred measurement method is the Hakki-Paoli method. But for broad-area
More informationSpectroscopic Analysis: Peak Detector
Electronics and Instrumentation Laboratory Sacramento State Physics Department Spectroscopic Analysis: Peak Detector Purpose: The purpose of this experiment is a common sort of experiment in spectroscopy.
More informationSpectral Color and Radiometry
Spectral Color and Radiometry Louis Feng April 13, 2004 April 13, 2004 Realistic Image Synthesis (Spring 2004) 1 Topics Spectral Color Light and Color Spectrum Spectral Power Distribution Spectral Color
More informationTTR500 Series Vector Network Analyzers Demonstration Guide
xx ZZZ TTR500 Series Vector Network Analyzers Demonstration Guide *P071349301* 071-3493-01 xx ZZZ TTR500 Series Vector Network Analyzers Demonstration Guide Register now! Click the following link to protect
More informationCoSMoS Starter Guide 1.6 Hoskins Lab Last Modified 03/30/2017 Chris DeCiantis
Contents Start Up:... 2 Turn on Shutter controllers... 2 Turn on Lasers... 2 Turn on other equipment:... 2 Load Slide onto Sample Holder:... 2 Turn on Computer/ Start Labview/Glimpse:... 3 Shutters Activate.vi
More informationEDX Spectra. Used to set the amount of time for which the spectrum will be acquired (dwell), and to perform the following functions:
EDS Spectrum Mode The EDS Spectrum Toolbar EDS spectrum display and acquisition functions are controlled from the EDS Spectrum Toolbar. If the toolbar is not visible at the top of your screen, selecting
More information37 (15 pts) Apply Snell s law twice (external, then internal) to find it emerges at the same angle.
37 (15 pts) Apply Snell s law twice (external, then internal) to find it emerges at the same angle. 38. (4 pts) Review the section on phase changes on reflection in Pedrotti section 3-3 Solution (a) For
More informationCornell Spectrum Imager (CSI) Open Source Spectrum Analysis with ImageJ Tutorial
Cornell Spectrum Imager (CSI) Open Source Spectrum Analysis with ImageJ Tutorial Electron Microscopy Summer School 2017 Why CSI Current Software Black box Expensive Steep learning curve Cornell Spectrum
More informationNMR Training VNMRJ 4.2A/VNMRS 400
NMR Training VNMRJ 4.2A/VNMRS 400 The VNMRS/new 400 MHz NMR uses VNMRJ 4.2A. Although VNMRJ is different than VNMRalmost ALL typed commands that you know from VNMR should work. LOGIN: Your username is
More informationX-Lab Pro 5 Operator Manual
X-Lab Pro 5 Operator Manual Revision 4; 17/08/2011 Page 1 of 397 Table of Contents SOFTWARE HELP...10 1 Job Manager...10 1.1 Job Manager...10 1.1.1 Window overview...10 1.2 Job Window...10 1.3 Create New
More informationChromQuest 5.0 Quick Reference Guide
ChromQuest 5.0 Quick Reference Guide This guide contains an overview of the ChromQuest chromatography data system, with topics organized by workflow. For more information, refer to the ChromQuest User
More information