Hawaii Earthquake Oct Gemini Recovery Planning and Execution. Prepared By: Rolando Rogers

Size: px
Start display at page:

Download "Hawaii Earthquake Oct Gemini Recovery Planning and Execution. Prepared By: Rolando Rogers"

Transcription

1 Hawaii Earthquake Oct Gemini Recovery Planning and Execution Prepared By: Rolando Rogers

2 Initial Set of Actions On Sunday directorate decided to close the telescope for night work till complete system checkout. The following day, using the initial information provided by Steve and John, an earthquake recovery team was organized.

3 Organization of the Recovery Team (1) Team Leader: Rolando Rogers (Electronics & Instrumentation Group Manager) Define a group of engineers and technicians from different disciplines to work in the recovery tasks: Civil, Mechanical, Optical, Electrical, Network, Instrumentation, and Electronics. In total around 20 persons. Organize personnel in sub-groups to tackle different sub-systems and tasks. Prepare a comprehensive list of sub- systems to inspect and check. Special consideration to Safety.

4 Organization of the Recovery Team (2) Only the Team Leader could authorize putting in service any subsystem. Team Leader was the single point of contact for communication to directorate, science staff, and rest of engineering team. Organize daily coordination meeting (in the morning) with the recovery team to check activities for the day. Organize a daily meeting (at 4 pm) to report to the rest of the staff about progress. Circulate Daily Progress Report to all Gemini staff.

5 Gemini-North Post Earthquake Check List and Report Updated October Activity Responsible Report Date 1.0 Building Infrastructure SH 1.1 Building itself In general is ok. Shear type of action in the dome isolation foam 16-Oct 1.2 Elevator Initial Inspection ok. Contacted Elevator Contractor and elevator released for use 16-Oct 1.3 Structural Inspection of Dome from outside SH Inspection ok. 17-Oct 2.0 Piping CC 2.1 Glycol Circuit Fully inspected ok 16-Oct 2.2 Oil Circuit for HBS Fully inspected ok 16-Oct 3.0 Cooling system SH 3.1 Chillers Chiller 1 and 2 inspected. Chiller 1 failed. Chiller 2 is on and w orking nicely but it w ill be left off for tonight. 16-Oct Chiller 1 now w orking ok, problem w as a possible short circuit due to moisture. Chiller 1 and 2 are now running ok. 17-Oct 3.2 Exhaust Fans EF1 ready to be pow er up. EF2 have some issues 16-Oct EF2 now working ok. Found problem w ith a digital I/O signal that made the plant room PLC to trip it. This feature w as disabled for the time being. Tested ok afterw ards 17-Oct 3.3 Repair of w ater pump This is maintenance issue that w as pending before the earthquake. Repaired done. Needs more checkouts. 20-Oct Repair w ork finished, including inspection. Working fine now. 21-Oct 4.0 Electrical Pow er PC 4.1 Plant Room UPS, generator, panels, etc. 16-Oct 4.2 Slip Rings 5th floor Initial inspection ok. Found some old problems that w ill need repair at certain point. 16-Oct 4.3 Other floors Inspection ok 16-Oct 5.0 Telescope Mount CC, RR 5.1 Azimuth Tape Encoders Gap in az heads not even. 16-Oct Azimuth heads back off for HBS testing. Tape encoder looks fine, no visible Sample of the document used damage. 17-Oct Azimuth heads w ere put back. Pending alignment check. 18-Oct While moving telescope in azimuth, w e see that the tape encoder follow s GIS absolute encoder quite w ell. Nevertheless w e still need to verify alignment. process. 19-Oct process. This This document document was was Head #1 and #2 checked w ith oscilloscope. The quadrature signals of head #2 looks good w ith a updated level of 8 v peak to peak, centered around 0 volts, and proper phase. Also updated on the reference mark on a looks ok daily at daily basis the position basis checked. Head #1 quadrature signals looks 23-Oct Sample of the document used during during the the earthquake earthquake recovery recovery

6 Overall Recovery Strategy In general the work was organized sequentially with the target of moving the telescope in elevation to properly check the M2 tip-tilt system at the horizon. This was the main branch of the scheduling. To achieve the above goal we had to check and repair (not in order): cooling systems, HBS, PCS, MCS, Elevation and azimuth cable wraps, air supply, electrical power, Cass rotator, A&G, instruments, etc. Whenever possible and depending in manpower availability we worked in parallel checking other non-telescope systems.

7 Outcome of Strategy Used On October 20 th, 5 days after earthquake, we were able to tilt the telescope to start checking M2 system. Except for M2 tip-tilt system, the telescope was ready to go on the sky on October 21 st, 6 days after earthquake.

8 Gemini-North Post Earthquake Check List and Report Updated October Activity Responsible Date 1.0 Building Infrastructure SH 1.1 Building itself 16-Oct 1.2 Elevator 16-Oct 1.3 Structural Inspection of Dome from outside SH 17-Oct 2.0 Piping CC 2.1 Glycol Circuit 16-Oct 2.2 Oil Circuit for HBS 16-Oct 3.0 Cooling system SH 3.1 Chillers 16-Oct 3.2 Exhaust Fans 16-Oct 3.3 Repair of w ater pump 20-Oct 4.0 Electrical Pow er PC 4.1 Plant Room 16-Oct 4.2 Slip Rings 5th floor 16-Oct 4.3 Other floors 16-Oct

9 5.0 Telescope Mount CC, RR 5.1 Azimuth Tape Encoders 16-Oct 5.2 Elevation Tape Encoders 16-Oct 5.2 Hydraulic Bearing Systems (HBS) 16-Oct 5.3 Az w rap 16-Oct 5.4 El w rap 16-Oct 5.5 Miscellaneous 16-Oct 5.6 Elevation Zenith-End-of-Travel sw itch needs adjustment 23-Oct 5.7 Test of telescope motion. 23-Oct 6.0 PCS SK, GG, PG 6.1 Oil circuit inside mirror cell 16-Oct 6.2 Glycol circuit inside mirror cell 16-Oct 6.3 Mirror Cover Inspection 19-Oct 6.4 Mirror Inspection 16-Oct 6.5 M1 Baffle 19-Oct 6.6 Control system 19-Oct 7.0 SCS Ccarter, PG 16-Oct 7.1 Tip tilt mechanisms inspection 18-Oct 7.2 Mirror Inspection PG 18-Oct 7.3 Secondary support structure (SSS) CC 20-Oct

10 8.0 Electronics JW 8.1 Thermal Enclosures in 5th floor 16-Oct 8.2 ECS Drives cabinets and cabling 16-Oct 9.0 Instrum ent JW 9.1 GMOS 16-Oct 9.2 Altair 16-Oct 9.3 GCAL 16-Oct 9.4 NIRI 16-Oct 9.5 Michelle 9.6 NIFS 10.0 OtherTelescope Systems 10.1 A&G JW, PG 16-Oct 10.2 CRCS 19-Oct 11.0 ECS SH 11.1 Dome bogie inspection 11.2 Top and Bottom shutter 17-Oct 11.3 Vent Gates 11.4 Cranes 17-Oct 12.0 Miscellaneous 12.1 Cherry Picker SH 16-Oct 12.2 Central Section and LSE SH, CC 18-Oct 12.3 Dome light covers 21-Oct 12.4 P1 alignment PG 19-Oct 13. Laser systems PG 13.1 Laser bench 17-Oct 13.2 BTO - BTOOB - LLT 20-Oct

11 List of System That Suffered Damage M2 Tip-Tilt system HBS (Hydraulic Bearing systems) Dome Azimuth Encoder System Shutters Azimuth over speed sensor Azimuth Wrap limit protection and position sensor Chiller 1

12 M2 Tip-tilt System Recovery From October 23th onwards our main task was to recover M2 tip-tilt system. The team leader for this task was Mike Sheehan (Deputy Chief Engineer). The M2 Tip-Tilt system was ready for use in November 10th, our first night on the sky after the earthquake. In summary: 26 nights were lost at Gemini North due to the Hawaii earthquake of October 15 th 2006.

13 Key to Success Good support from Director and upper management Good communication with all Gemini staff Good support from science staff (very patience) Good knowledge of all our telescope and instrument systems. And on top of all that a great Engineering Team!!.

Canada-France-Hawaii Telescope. CFHT - the seismometer (and the earthquakes of Oct 15, 2006)

Canada-France-Hawaii Telescope. CFHT - the seismometer (and the earthquakes of Oct 15, 2006) Canada- CFHT - the seismometer (and the earthquakes of Oct 15, 2006) Summary Earthquake overview Events at Waimea headquarters Telescope construction and earthquake design criteria Events at the summit

More information

Conceptual-Level Cost Estimate for Point Design of a 30-m Giant Segmented Mirror Telescope (GSMT)

Conceptual-Level Cost Estimate for Point Design of a 30-m Giant Segmented Mirror Telescope (GSMT) Conceptual-Level Cost Estimate for Point Design of a 30-m Giant Segmented Mirror Telescope (GSMT) ALL INFORMATION HEREIN IS COPYRIGHTED AND PROVIDED TO NOAO, AURA FOR ITS USE FOR NON-COMMERCIAL RESEARCH

More information

M2 Reference Sensor Failure & Recovery. Chris Carter Senior Controls Engineer Gemini Observatory

M2 Reference Sensor Failure & Recovery. Chris Carter Senior Controls Engineer Gemini Observatory M2 Reference Sensor Failure & Recovery Chris Carter Senior Controls Engineer Gemini Observatory Overview Gemini s Secondary Mirror Tip/tilt System (M2TS) was damaged in the earthquake, A key component

More information

Environmental Control and Monitoring (Draft)

Environmental Control and Monitoring (Draft) Next Generation Adaptive Optics System Environmental Control and Monitoring (Draft) April 10, 2009 VersionV1.0 Prepared By Jason Chin 2 of 11 REVISION HISTORY Revision Date Author (s) Reason for revision

More information

ACE Desktop: Dome, Telescope, Instruments

ACE Desktop: Dome, Telescope, Instruments Page 1 of 9 ACE Desktop: Dome, Telescope, Instruments (Region 1) On the left hand side of the ACE Desktop the real-time states of the dome, telescope and instruments is displayed. Listed below are the

More information

Security and Disaster Preparedness: New York s Experience. Presented by: New York Department of Public Service. December 7, James T.

Security and Disaster Preparedness: New York s Experience. Presented by: New York Department of Public Service. December 7, James T. Security and Disaster Preparedness: New York s Experience Presented by: James T. Gallagher New York Department of Public Service james_gallagher@dps.state.ny.us December 7, 2001 Rome, Italy 1 Security

More information

50 Years of Facility Operations or Rebuilding from the inside out

50 Years of Facility Operations or Rebuilding from the inside out 50 Years of Facility Operations or Rebuilding from the inside out National Optical Astronomy Observatory (NOAO) NORTH Kitt Peak National Observatory and Tucson Headquarters Facility Presented by John Dunlop

More information

Fair City Plaza. Task: Diagnose problems, recommend improvements for small municipal district cooling system

Fair City Plaza. Task: Diagnose problems, recommend improvements for small municipal district cooling system Fair City Plaza Task: Diagnose problems, recommend improvements for small municipal district cooling system The Fair City Plaza district cooling system provides chilled water to five buildings in a downtown

More information

Position Title: H V A C T E C H N I C I A N. G.E.D. or High School Diploma required, an Associate s Degree is preferred.

Position Title: H V A C T E C H N I C I A N. G.E.D. or High School Diploma required, an Associate s Degree is preferred. Location Beeville-Primary; All Campuses Hours 40 hours/week minimum Salary $40,000-$50,000 annual, based on KSAs Job Classification Exempt Security Sensitive Y e s Division Physical Plant Department Physical

More information

Mission Critical Facilities & Technology Conference November 3, 2011 Cooling 101. Nick Gangemi Regional Sales Manager Data Aire

Mission Critical Facilities & Technology Conference November 3, 2011 Cooling 101. Nick Gangemi Regional Sales Manager Data Aire Mission Critical Facilities & Technology Conference November 3, 2011 Cooling 101 Nick Gangemi Regional Sales Manager Data Aire Agenda Exactly what are we talking about ASHRAE TC 9.9 Design best practices

More information

Engineering Requirements & Industrial Opportunities. Miquel Barceló LST Project Manager STFC Daresbury Laboratory

Engineering Requirements & Industrial Opportunities. Miquel Barceló LST Project Manager STFC Daresbury Laboratory Engineering Requirements & Industrial Opportunities Miquel Barceló LST Project Manager STFC Daresbury Laboratory CTA Observatory Sites Infrastructure Buildings Telescopes Mechanics Optics Cameras Off-site

More information

Theodolite Survey of the Dome from the AO9 Monument August 2001

Theodolite Survey of the Dome from the AO9 Monument August 2001 Theodolite Survey of the Dome from the AO9 Monument August 2001 The Gregorian dome was surveyed from the AO9 monument at the base of the main reflector in August 2001. The following is an edited and reformatted

More information

The Gamma-ray Cherenkov Telescope for CTA

The Gamma-ray Cherenkov Telescope for CTA The Gamma-ray Cherenkov Telescope for CTA Progress with GCT structure and design changes for preproduction. Status of CHEC-M. GCT tests: March in Paris. CHEC-S status: Sensors. TARGET ASICs and modules.

More information

1.1 ANNUAL DATA CENTER HEALTH CHECK

1.1 ANNUAL DATA CENTER HEALTH CHECK Statement of Work Maintenance Service 1.0 Executive Summary - 1 - Service Table of Contents 1.0 Executive Summary 2.0 Features & Benefits 3.0 Details of Service 4.0 Assumptions 5.0 Scope of Responsibility

More information

AMG Series. Motorized Position and Rate Gimbals. Continuous 360 rotation of azimuth and elevation including built-in slip ring

AMG Series. Motorized Position and Rate Gimbals. Continuous 360 rotation of azimuth and elevation including built-in slip ring AMG Series Optical Mounts AMG Series Motorized Position and Rate Gimbals Continuous rotation of azimuth and elevation including built-in slip ring High accuracy angular position and rate capability Direct-drive

More information

Observatory Automation Project

Observatory Automation Project Observatory Automation Project Detail Design Review Control of Electric Dome Drive System Grant Matsushige/Steven Bauman Version 1.0 17 May 2010 1. Overview After the preliminary design review, it s been

More information

Data Center Facility Analysis Project. Final Presentation Wednesday, July 22, :30 3:30 PM

Data Center Facility Analysis Project. Final Presentation Wednesday, July 22, :30 3:30 PM Data Center Facility Analysis Project Final Presentation Wednesday, July 22, 2009 1:30 3:30 PM 1 Commonwealth Technology Center Conference Rooms 1 & 2 Enterprise Server Farm (ESF) Facility Data: 1 Technology

More information

Estimating Dome Seeing for LSST Jacques Sebag a and Konstantinos Vogiatzis

Estimating Dome Seeing for LSST Jacques Sebag a and Konstantinos Vogiatzis Estimating Dome Seeing for LSST Jacques Sebag a and Konstantinos Vogiatzis a LSST Project, National Optical Astronomy Observatory, 950 N. Cherry Ave., Tucson, AZ, 85719 USA ABSTRACT Begin Dome seeing is

More information

High Performance, Full-Digital Control on the LMT and KVN Telescopes

High Performance, Full-Digital Control on the LMT and KVN Telescopes High Performance, Full-Digital Control on the LMT and KVN Telescopes David R. Smith (MERLAB), alpha@merlab.com Kamal Souccar (UMass), David Gale (INAOE), F. Peter Schloerb (Umass), David Hughes (INAOE)

More information

Phoenix Install 4m 12/12

Phoenix Install 4m 12/12 Phoenix Install 4m 12/12 Place Instrument Change in Progress sign on console Follow instructions for removing the cage bottom Once removed, add or remove the appropriate rocket weights If necessary, follow

More information

Nathan Loewen AMEC Dynamic Structures January 17, AMEC Corporate Profile. AMEC Dynamic Structures Ltd:

Nathan Loewen AMEC Dynamic Structures January 17, AMEC Corporate Profile. AMEC Dynamic Structures Ltd: CCAT Enclosure Nathan Loewen AMEC Dynamic Structures January 17, 2005 AMEC Corporate Profile AMEC Dynamic Structures Ltd: Located in Vancouver, Canada Design/build steel fabricating firm Specialize in

More information

Wind Vibration Analysis Giant Magellan Telescope

Wind Vibration Analysis Giant Magellan Telescope Wind Vibration Analysis Giant Magellan Telescope prepared for Carnegie Observatories 813 Santa Barbara Street Pasadena, CA 91101 prepared by Simpson Gumpertz & Heger Inc. 41 Seyon Street, Building 1, Suite

More information

Starrett-Bytewise Measurement Systems Profile In-Line Pipe Diameter Measurement System Technical Proposal May 8, 2014

Starrett-Bytewise Measurement Systems Profile In-Line Pipe Diameter Measurement System Technical Proposal May 8, 2014 Overview Profile360 TM is an in-line, real-time measurement system for continuously monitoring OD, OOR, and circumference for round pipe and profiles. Profile360 can be installed in two locations, in-line

More information

Passive RDHx as a Cost Effective Alternative to CRAH Air Cooling. Jeremiah Stikeleather Applications Engineer

Passive RDHx as a Cost Effective Alternative to CRAH Air Cooling. Jeremiah Stikeleather Applications Engineer Passive RDHx as a Cost Effective Alternative to CRAH Air Cooling Jeremiah Stikeleather Applications Engineer Passive RDHx as a Cost Effective Alternative to CRAH Air Cooling While CRAH cooling has been

More information

Model 90. thermwood. Thermwood CNC Machining Centers. Machine Features Shown

Model 90. thermwood. Thermwood CNC Machining Centers. Machine Features Shown 22017 MADE IN USA 5 Axis Model 90 Thermwood, the industry leader in 5 Axis applications with more five-axis systems in operation than any other company. These 5 Axis CNC routers have become the accepted

More information

Starrett-Bytewise Measurement Systems Profile In-Line Pipe Diameter Measurement System Technical Proposal May 8, 2014

Starrett-Bytewise Measurement Systems Profile In-Line Pipe Diameter Measurement System Technical Proposal May 8, 2014 Overview Profile360 TM is an in-line, real-time measurement system for continuously monitoring OD, OOR, and circumference for round pipe and profiles. Profile360 can be installed in two locations, in-line

More information

KECK TELESCOPE CONTROL SYSTEM UPGRADE (TCSU) PROJECT STATUS [MOCOAAB05]

KECK TELESCOPE CONTROL SYSTEM UPGRADE (TCSU) PROJECT STATUS [MOCOAAB05] KECK TELESCOPE CONTROL SYSTEM UPGRADE (TCSU) PROJECT STATUS [MOCOAAB05] J.Johnson, K.Tsubota, J.Mader W.M. Keck Observatory ICALEPCS 2013 1 The Telescope Control System Upgrade The Keck telescopes, located

More information

Cooling. Highly efficient cooling products for any IT application. For More Information: (866) DATA CENTER SOLUTIONS

Cooling. Highly efficient cooling products for any IT application. For More Information: (866) DATA CENTER SOLUTIONS Cooling > Highly efficient cooling products for any IT application. DATA CENTER SOLUTIONS For More Information: (866) 787-3271 Sales@PTSdcs.com End-to-End Cooling Solutions from the Closet to the Data

More information

Liebert DCW Family Chilled Water-Based High Density Cooling For The Data Center. Precision Cooling For Business-Critical Continuity

Liebert DCW Family Chilled Water-Based High Density Cooling For The Data Center. Precision Cooling For Business-Critical Continuity Liebert DCW Family Chilled Water-Based High Density Cooling For The Data Center Precision Cooling For Business-Critical Continuity Flexible Approaches For Energy Efficient Cooling Save up to 70% in cooling

More information

Brad Nacke Enterprise Account Leader Data Centers

Brad Nacke Enterprise Account Leader Data Centers Brad Nacke Enterprise Account Leader Data Centers United Technologies Makes Modern Life Possible Commercial Aerospace Employees: ~56,000 Sales: $16.9B Employees: ~67,000 Sales: $12.0B Employees: ~35,000

More information

Ten Cooling Solutions to Support High-density Server Deployment. Schneider Electric Data Center Science Center White Paper #42

Ten Cooling Solutions to Support High-density Server Deployment. Schneider Electric Data Center Science Center White Paper #42 Ten Cooling Solutions to Support High-density Server Deployment Schneider Electric Data Center Science Center White Paper #42 1 Introduction 2 Introduction This paper provides 10 approaches for increasing

More information

T66 Central Utility Building (CUB) Control System Reliability, Optimization & Demand Management

T66 Central Utility Building (CUB) Control System Reliability, Optimization & Demand Management T66 Central Utility Building (CUB) Control System Reliability, Optimization & Demand Management Name Doug Demian Title Application Engineer Date November 7 & 8, 2012 Rev 5058-CO900C Copyright 2012 Rockwell

More information

WO P-NF-L-M-407c Spec Submittal

WO P-NF-L-M-407c Spec Submittal WO2-2-10000-2P-NF-L-M-407c Spec Submittal General Specifications WO2-2-10000(801304) Width: IN/CM 136.25/346.07 Height: IN/CM 84.5/214.63 Depth: IN/CM 41.7/105.92 Dry weight: LB/KG 4000/1814.37 Operational

More information

MICRO GROW GREENHOUSE SYSTEMS, INC

MICRO GROW GREENHOUSE SYSTEMS, INC MICRO GROW GREENHOUSE SYSTEMS, INC 6111 YNEZ RD., SUITE C-4, TEMECULA, CA 9591 PHONE (909)-96-3340 FAX (909)-96-3350 Revision 1.1, 11-09-01 Growstat Control System INSTALLATION PROCEDURES GROWSTAT SERIES

More information

Chiller Replacement and System Strategy. Colin Berry Account Manager

Chiller Replacement and System Strategy. Colin Berry Account Manager Chiller Replacement and System Strategy Colin Berry Account Manager Current Situation 4 Existing RTAA chillers All units 23 years old Limited service history One unit MUST be replaced immediately One unit

More information

GENERATOR ADDENDUM TO MAGNOLIA GARDENS ASSISTED LIVING COMPREHENSIVE EMERGENCY MANAGEMENT PLAN

GENERATOR ADDENDUM TO MAGNOLIA GARDENS ASSISTED LIVING COMPREHENSIVE EMERGENCY MANAGEMENT PLAN Contact Person: Philip McCollum - Administrator Magnolia Gardens Assisted Living AL Lic# 10314 3800 62 nd Avenue North Pinellas Park, FL 33781 Phone: 727.489.6440 GENERATOR ADDENDUM TO MAGNOLIA GARDENS

More information

DigitalHome Software Requirements Specification HomeOwner Inc. DigitalHomeOwner Division Version 1.3 October 27, 2010

DigitalHome Software Requirements Specification HomeOwner Inc. DigitalHomeOwner Division Version 1.3 October 27, 2010 DigitalHome Software Requirements Specification HomeOwner Inc. DigitalHomeOwner Division Version 1.3 October 27, 2010 Change History *Added (A), Modified (M), Deleted (D) Version Date AMD* Author Description

More information

National Service Center

National Service Center UNITED TECHNOLOGIES CARRIER CORPORATION National Service Center MAXIMIZING PERFORMANCE AND LONGEVITY OF INDUSTRIAL CHILLERS Protection, Performance and Peace of Mind Because industrial chillers serve large

More information

IXcellerate Moscow One Datacentre - Phase 1 & 2 Overview

IXcellerate Moscow One Datacentre - Phase 1 & 2 Overview Contents 1. Document Purpose... 2 2. Facility Overview... 2 2.1 Technical Space... 2 2.2 Load Density... 2 2.3 Resilience... 2 2.4 Engineering Plant Maintenance & Service Restoration... 3 3. Engineering

More information

Assembled. HTE Auditing Services. Programming Support. Joint Engineering. Custom Fabrication of Hydraulic Power Units & Lubrication Systems

Assembled. HTE Auditing Services. Programming Support. Joint Engineering. Custom Fabrication of Hydraulic Power Units & Lubrication Systems Industrial Motion & Control Customer Compressed Air Assembled Group HTE Auditing Services Programming Support Joint Engineering Custom Fabrication of Hydraulic Power Units & Lubrication Systems Machine

More information

Remote Monitoring. for Building HVAC Systems. A unique combination of human expertise and technology helps ensure optimum operation.

Remote Monitoring. for Building HVAC Systems. A unique combination of human expertise and technology helps ensure optimum operation. Remote Monitoring SM for Building HVAC Systems A unique combination of human expertise and technology helps ensure optimum operation. For peace of mind: ProTech SM monitors HVAC systems 24 hours a day

More information

Drives Motors and PLCs

Drives Motors and PLCs Drives Motors and PLCs WHAT YOU WILL LEARN: The electrical drives and other rotating equipment connected to motors like pumps, fans, blowers, compressors etc. consumes about 65% of the total electrical

More information

OWL: FIRST STEPS TOWARDS DESIGNING THE MECHANICAL STRUCTURE

OWL: FIRST STEPS TOWARDS DESIGNING THE MECHANICAL STRUCTURE OWL: FIRST STEPS TOWARDS DESIGNING THE MECHANICAL STRUCTURE E. Brunetto, F. Koch, M. Quattri European Southern Observatory Abstract The design, fabrication and assembly of the supporting structure of a

More information

Introducing the first PATENT PENDING solid state selfregulating

Introducing the first PATENT PENDING solid state selfregulating Introducing the first PATENT PENDING solid state selfregulating THERMOELECTRIC GENERATOR. Covered by 4 Patent applications. This novel thermoelectric TEG system will provide continuous power as long as

More information

Gregory Walsh, Ph.D. San Ramon, CA January 25, 2011

Gregory Walsh, Ph.D. San Ramon, CA January 25, 2011 Leica ScanStation:: Calibration and QA Gregory Walsh, Ph.D. San Ramon, CA January 25, 2011 1. Summary Leica Geosystems, in creating the Leica Scanstation family of products, has designed and conducted

More information

SICK AG WHITEPAPER SAFETY LASER SCANNERS VS. SAFETY MATS WHICH ONE DO I CHOOSE?

SICK AG WHITEPAPER SAFETY LASER SCANNERS VS. SAFETY MATS WHICH ONE DO I CHOOSE? SICK AG WHITEPAPER 2014-10 AUTHOR Steve Aamodt Safety Product Manager at SICK Inc. in Bloomington ABSTRACT Since 1990, safety mats are facing an increasing competition from safety laser scanners. Because

More information

or

or Infinite possibilities for your data center The F-Series TeraFrame T M Cabinet System The F-Series TeraFrame Cabinet System from Chatsworth Products, Inc. (CPI) provides configurable thermal management

More information

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide Model P4017 Single Channel USB Oscilloscope Quick Start Guide General Warranty BNC warrants that the product will be free from defects in materials and workmanship for 3 years from the date of purchase

More information

PATENT PENDING RABBIT EARS THERMOELECTRIC GENERATOR

PATENT PENDING RABBIT EARS THERMOELECTRIC GENERATOR Introduction: The Rabbit Ears CHP (Combined heat & power) is a Thermoelectric wood stove Generator Nominally rated at 100 watts. It has the ability to supply reliable solid-state power. The ONLY moving

More information

Installation Service. 1.0 Executive Summary. Factory Trained Technician Train Customer Support Staff On Basic Operation Of The Equipment

Installation Service. 1.0 Executive Summary. Factory Trained Technician Train Customer Support Staff On Basic Operation Of The Equipment Installation Service Statement of Work Start-up Service for Cooling Start-up - 1 - Service for Cooling Service 1.0 Executive Summary Table of Contents 1.0 Executive Summary 2.0 Features & Benefits 3.0

More information

FastCam Outdoor Speed Dome Camera

FastCam Outdoor Speed Dome Camera FastCam Outdoor Speed Dome Camera CC-9760 & CC-9760-I USER INSTRUCTION MANUAL Version: V1.2 Doc No.: 20051114001 1 A BOUT THIS GUIDE Conventions used in this guide To make sure that you perform certain

More information

PRODUCT OVERVIEW. Well Inspection Cameras & Systems. See what you re missing. Proudly made in the U.S.A.

PRODUCT OVERVIEW. Well Inspection Cameras & Systems. See what you re missing. Proudly made in the U.S.A. PRODUCT OVERVIEW Well Inspection Cameras & Systems See what you re missing. Proudly made in the U.S.A. See what you re missing. Aries provides the equipment you need for efficient well inspection to ensure

More information

PRODUCT OVERVIEW. Well Inspection Cameras & Systems. See what you re missing. Proudly made in the U.S.A.

PRODUCT OVERVIEW. Well Inspection Cameras & Systems. See what you re missing. Proudly made in the U.S.A. PRODUCT OVERVIEW Well Inspection Cameras & Systems See what you re missing. Proudly made in the U.S.A. See what you re missing. Aries provides the equipment you need for efficient well inspection to ensure

More information

GP SERIES Packaged Chillers

GP SERIES Packaged Chillers Blending Conveying Drying Process Cooling Heat Transfer Size Reduction GP SERIES Packaged Chillers Comprehensive Flexible Versatile Compact unitary designs available in a variety of configurations from

More information

Power System Vulnerabilities : AHCA 2016 Copyright (c) 2016 SSR, Inc. All rights reserved

Power System Vulnerabilities : AHCA 2016 Copyright (c) 2016 SSR, Inc. All rights reserved Training presentation for: AHCA Florida 2016 Annual Conference Finding & Managing Power System Vulnerabilities David Stymiest, CHFM, CHSP, FASHE, (PEin LA MS MA) Copyright 2016, Smith Seckman Reid, Inc.;

More information

THE ISAAC NEWTON GROUP OF TELESCOPES

THE ISAAC NEWTON GROUP OF TELESCOPES THE ISAAC NEWTON GROUP OF TELESCOPES INSTRUMENT CHANGE CHECKLIST PRIME FOCUS IMAGING PLATFORM (PFIP) USING THE MOSAIC CAMERA WITH UDAS Revision : 2.2 Description : Location : Weight : CCD Wide Field Imaging

More information

Digital Scroll Compressor Control

Digital Scroll Compressor Control INSTALLATION OPERATION SUPPLEMENT Digital Scroll Compressor Control PZA model shown PZAT model shown DO NOT TURN THE POWER ON TO YOUR NEW LEGACY CHILLER UNTIL THE ENCLOSED COMMISSIONING CHECKLIST HAS BEEN

More information

Compliant Baffle for Large Telescope Daylight Imaging. Stacie Williams Air Force Research Laboratory ABSTRACT

Compliant Baffle for Large Telescope Daylight Imaging. Stacie Williams Air Force Research Laboratory ABSTRACT Compliant Baffle for Large Telescope Daylight Imaging Steven Griffin, Andrew Whiting, Shawn Haar The Boeing Company Stacie Williams Air Force Research Laboratory ABSTRACT With the recent interest in daylight

More information

Site Preparation for ClearCube Installation. Rev. F

Site Preparation for ClearCube Installation. Rev. F Site Preparation for ClearCube Installation Rev. F EXECUTIVE SUMMARY Because of the centralized nature of the ClearCube architecture, installation of the system requires a certain degree of site preparation.

More information

HARP-NEF Front End Assembly

HARP-NEF Front End Assembly Science Fibers HARP-NEF Front End Assembly Calib Fibers HARPN Front End Assembly ( w/ Partial Cover Plates ) HARPN Front End Assembly 1 HARP-NEF Front End Assembly Roger Eng December 6, 2007 2 HARP-NEF

More information

7 Best Practices for Increasing Efficiency, Availability and Capacity. XXXX XXXXXXXX Liebert North America

7 Best Practices for Increasing Efficiency, Availability and Capacity. XXXX XXXXXXXX Liebert North America 7 Best Practices for Increasing Efficiency, Availability and Capacity XXXX XXXXXXXX Liebert North America Emerson Network Power: The global leader in enabling Business-Critical Continuity Automatic Transfer

More information

AP1000 European 1. Introduction and General Description of the Plant Design Control Document

AP1000 European 1. Introduction and General Description of the Plant Design Control Document AP1000 European 1. Introduction and General Description of the Plant Design Control Document 1.7 Drawings and Other Detailed Information 1.7.1 Electrical and Instrumentation and Control Drawings Instrument

More information

ACE SmartDome TM User Manual Version 1.0, March 2006

ACE SmartDome TM User Manual Version 1.0, March 2006 Page 1 of 20 ACE SmartDome TM User Manual Version 1.0, March 2006 1999. All rights reserved Introduction Astronomical Consultants & Equipment, Inc. 2901 W Sahuaro Drive Tel: 520-219-8722 Tucson, AZ 85742

More information

Factsheet. Power Generation Service Energy efficiency in power generation and water

Factsheet. Power Generation Service Energy efficiency in power generation and water Factsheet Power Generation Service Energy efficiency in power generation and water Opportunity identification Opportunity identification overview The opportunity identification phase of Industrial Energy

More information

Planning a Wireless Network. Marco Zennaro Ermanno Pietrosemoli

Planning a Wireless Network. Marco Zennaro Ermanno Pietrosemoli Planning a Wireless Network Marco Zennaro Ermanno Pietrosemoli Goals To understand why there is a need for planning a wireless network (or link) and how it can be performed effectively To know what kind

More information

SYSCO HOUSTON HOUSTON, TEXAS

SYSCO HOUSTON HOUSTON, TEXAS SYSCO HOUSTON HOUSTON, TEXAS October, 2018 Infrared Inspection Report SYSCO Houston Inc 10710 Greens Crossing Blvd. Houston, Texas, 77038 Company Contact: Steve Moulder Project: 04121018 2605 CHARTER OAK

More information

Disaster Recovery Committee. Learning Resource Center Specialist

Disaster Recovery Committee. Learning Resource Center Specialist This document is intended to provide operational procedures and serve as a reference for the Key Information Technology Personnel at Community Higher Ed Disaster Recovery Community Higher Ed Disaster Recovery

More information

STATE DATA CENTER DRAFT PROPOSAL. STATE OF OREGON DATA CENTER 28 October 2016

STATE DATA CENTER DRAFT PROPOSAL. STATE OF OREGON DATA CENTER 28 October 2016 STATE DATA CENTER DRAFT PROPOSAL 1 PROJECT TEAM 2 SERA Architects, Inc George Hager Patrick Sullivan PAE Engineers Dave Williams Doug Nachtrieb Michael Yee JMB Consulting Group Jon Bayles MEETING GOALS

More information

Oil & Gas. Control Control. every move. Pressure Temperature Level Flow

Oil & Gas. Control Control. every move. Pressure Temperature Level Flow Oil & Gas Control Control every move Pressure Temperature Level Flow Barksdale Oil and Gas - The Challenge today Oil & Gas Today Worldwide economic growth, depletion of oil reserves, and continued investment

More information

Set-up and programming guide

Set-up and programming guide Set-up and programming guide HV 2.015-4.220 software for compatibility to HV 2.1-3.22 V 2012/05A-C01.4 771079512 Manual C01.4 for HV 2.015-4.220 ENGLISH Index 1 Important safety instructions...2 2 General

More information

Ventilation Policy Type: Policy Register No: Status: Public. Developed in response to: Contributes to CQC Outcome number: Outcome 8 and 10

Ventilation Policy Type: Policy Register No: Status: Public. Developed in response to: Contributes to CQC Outcome number: Outcome 8 and 10 Ventilation Policy Type: Policy Register No: 11056 Status: Public Developed in response to: HTM03-01 Contributes to CQC Outcome number: Outcome 8 and 10 Consulted With Post/Committee/Group Date Louise

More information

Domes & Shutters Preventative/Predictive Maintenance Tools

Domes & Shutters Preventative/Predictive Maintenance Tools Maunakea Operations and Engineering Workshop Domes & Shutters Preventative/Predictive Maintenance Tools John Baldwin W.M. Keck Observatory Maunakea Operations and Engineering Workshop, 8 December 2016

More information

Energy Game Changers

Energy Game Changers Energy Game Changers Use of Data at Adventist Health Castle Richard Kirchner Director of Facilities SMARDT chiller performance curve Red line represents probable power line if only one chiller in operation

More information

Synopsis of Risley Prism Beam Pointer

Synopsis of Risley Prism Beam Pointer Synopsis of Risley Prism Beam Pointer Mark T. Sullivan Lockheed Martin Space Systems Advanced Technology Center, 3251 Hanover Street, Palo Alto, CA 94304 mark.t.sullivan@lmco.com 650/424-2722 SUMMARY This

More information

ANSI/BICSI Data Center Design and Implementation Best Practices

ANSI/BICSI Data Center Design and Implementation Best Practices ANSI/BICSI 002-2014 Data Center Design and Implementation Best Practices Risk, Reliability & Availability Selecting a Data Center Design Class Bob Camerino RCDD DCDC CT Vice Chair BICSI Mainland Europe

More information

Datasheet 2102 SERIES TWO-AXIS POSITIONING AND RATE TABLE SYSTEM

Datasheet 2102 SERIES TWO-AXIS POSITIONING AND RATE TABLE SYSTEM Datasheet 2102 SERIES TWO-AXIS POSITIONING AND RATE TABLE SYSTEM FEATURES Position Accuracy: ± 30 arc seconds (both axes) Rate Accuracy: ± 0.01% Max Rate (varies depending on axis configuration): Inner

More information

(2KN) SW2-3150APP (3KN)

(2KN) SW2-3150APP (3KN) Systems Dynamic Test 1-3 kn a ir c oo l e d Electrodynamic Vibration Test Systems SW2-1150APP (1KN) SW2-2150APP (2KN) SW2-3150APP (3KN) Technical subject to changes without prior notice! Index Page Serial

More information

Table of Contents 1 ABOUT THIS GUIDE CONTACT INFORMATION ANTENNA INSTALLATION... 4

Table of Contents 1 ABOUT THIS GUIDE CONTACT INFORMATION ANTENNA INSTALLATION... 4 Table of Contents 1 ABOUT THIS GUIDE... 3 1.1 CONTACT INFORMATION... 3 2 ANTENNA INSTALLATION... 4 2.1 GENERAL INFORMATION... 4 2.2 SPECIFIC MOUNTING EXAMPLES... 5 2.3 CONNECTOR MOISTURE PROTECTION...

More information

Temperature control unit

Temperature control unit Temperature control unit 5739 18/19 Installation manual Part. U3582A - 12/08-01 PC Contents 1 - Introduction 5 1.1 - Warnings and tips 5 1.2 - Contents of package 5 2 - Description of the Control unit

More information

ANSI/BICSI Data Center Design and Implementation Best Practices

ANSI/BICSI Data Center Design and Implementation Best Practices ANSI/BICSI 002-2014 Data Center Design and Implementation Best Practices Risk, Reliability & Availability Selecting a Data Center Design Class Bob Camerino RCDD DCDC CT Secretary BICSI Mainland Europe

More information

Controlling PI Positioners with External Zygo Interferometers

Controlling PI Positioners with External Zygo Interferometers Controlling PI Positioners with External Zygo Interferometers Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstr. 1, 76228 Karlsruhe, Germany Page 1 of 6 1 Introduction Developing high-precision

More information

RVL470. Heating controller. G2522en. Installation Instructions. 1 Installation. 2 Commissioning Wall mounting DIN rail mounting

RVL470. Heating controller. G2522en. Installation Instructions. 1 Installation. 2 Commissioning Wall mounting DIN rail mounting G2522en Heating controller Installation Instructions RVL470 1 Installation 1.1 Place of installation In a dry room, e.g. the boiler room Mounting choices: In a control panel (on the inner wall or on a

More information

REQUEST FOR PROPOSAL:

REQUEST FOR PROPOSAL: REQUEST FOR PROPOSAL: FIBER CABLING INSTALLATION TO CONNECT SIX ADDITIONAL VIEWING STATIONS AT THE EMERGENCY OPERATION CENTER (EOC) WITH THE REGIONAL VIDEO COMMAND CENTER (RVCC) VIDEO FEED I. DESCRIPTION

More information

VDL-004. Miniature Manual Variable Optical Delay Line. User Guide

VDL-004. Miniature Manual Variable Optical Delay Line. User Guide VDL-004 Miniature Manual Variable Optical Delay Line User Guide Version: 1.1 Date: July 24, 2018 General Photonics Corporation is located in Chino, California. For more information visit the company's

More information

At a Glance. HP Dynamic Smart Cooling. Overview. DSC Base Station

At a Glance. HP Dynamic Smart Cooling. Overview. DSC Base Station Overview DSC Base Station The new (DSC) technology helps you tackle one of the most critical issues in today's data centers - power and cooling. DSC will enable you to change data center energy costs from

More information

Observatory Automation

Observatory Automation Observatory Automation Project Plan Project ongo02e February 12, 2001 Client: ISU Physics Department Advisor: Dr. J. Basart Team Members: Shahril Abdullah Justin Doyle Derek Halverson Jeff Rastrelli Rob

More information

Digital Realty Data Center Solutions Digital Chicago Datacampus Franklin Park, Illinois Owner: Digital Realty Trust Engineer of Record: ESD Architect

Digital Realty Data Center Solutions Digital Chicago Datacampus Franklin Park, Illinois Owner: Digital Realty Trust Engineer of Record: ESD Architect Digital Realty Data Center Solutions Digital Chicago Datacampus Franklin Park, Illinois Owner: Digital Realty Trust Engineer of Record: ESD Architect of Record: SPARCH Project Overview The project is located

More information

Motion Solutions Newport Gimbal Systems

Motion Solutions Newport Gimbal Systems Motion Solutions Newport Gimbal Systems Introduction In the past 55 years Newport has designed and manufactured thousands of high accuracy positioning systems for Scientific Research, Defense and Security,

More information

City of Oxnard. Public Works Integrated Master Plan WATER PROJECT MEMORANDUM 2.8 SCADA ASSESSMENT. FINAL DRAFT December 2015

City of Oxnard. Public Works Integrated Master Plan WATER PROJECT MEMORANDUM 2.8 SCADA ASSESSMENT. FINAL DRAFT December 2015 This document is released for the purpose of information exchange review and planning only under the authority of Hugh Steve McDonald, December 2015, State of California, PE No. 44074 and Tracy Anne Clinton,

More information

All Rights Reserved - 1

All Rights Reserved -   1 The Most Effective Chiller Alan Garbers http://www.johnsoncontrols.com/buildings/services-and-support Agenda Predictive Diagnostics -- Overview Focus: Diagnostics & Chillers Online Connected Diagnostics

More information

BGA IRDA-WELDER User Manual. Model: T870A

BGA IRDA-WELDER User Manual. Model: T870A BGA IRDA-WELDER Model: T870A 0 CATALOGUE Features....2 Technical Parameter and Components..3 T-870A illustrated explaining...4 (1)The whole machine 4 (2)Front panel.. 5 (3)Rear panel.. 5 (4)Focus holder

More information

Liebert DCW CATALOGUE

Liebert DCW CATALOGUE Liebert DCW CATALOGUE Chilled Water-Based High Density Cooling For The Data Center LIEBERT DCW Flexible Approaches For Energy Efficient Cooling Save up to 70% in cooling energy usage Data center managers

More information

Cooling Solutions & Considerations for High Performance Computers

Cooling Solutions & Considerations for High Performance Computers Cooling Solutions & Considerations for High Performance Computers White Paper 6 Revision 1 By Rich Whitmore Executive summary The recent advances in high performance computing have driven innovations not

More information

THE ROLE OF INFRARED TESTING AT DATA CENTERS

THE ROLE OF INFRARED TESTING AT DATA CENTERS THE ROLE OF INFRARED TESTING AT DATA CENTERS By: Eric R. Stockton CompuScanIR Division Stockton Infrared Thermographic Services, Inc. www.compuscanir.com 800-248-7226 Introduction Data centers and other

More information

Overcoming the Challenges of Server Virtualisation

Overcoming the Challenges of Server Virtualisation Overcoming the Challenges of Server Virtualisation Maximise the benefits by optimising power & cooling in the server room Server rooms are unknowingly missing a great portion of their benefit entitlement

More information

Thermal Diffusivity Instruments DLF-2 with EM-1600

Thermal Diffusivity Instruments DLF-2 with EM-1600 Thermal Diffusivity Instruments DLF-2 with EM-1600 Site Preparation Guide Revision A Issued February 2017 Table of Contents Table of Contents... 1 Ideal Setup... 2-3 System Components... 3 Instrument Measurements...

More information

Our Power & Control line of Business offers solutions that safely distribute power to your home or business. Our product range includes Low/Medium

Our Power & Control line of Business offers solutions that safely distribute power to your home or business. Our product range includes Low/Medium Our Power & Control line of Business offers solutions that safely distribute power to your home or business. Our product range includes Low/Medium Voltage Panels and Enclosures designed and built with

More information

Precise temperature control and high reliability

Precise temperature control and high reliability LASER COOLING Precise temperature control and high reliability Heat sinks for actively and passively cooled light sources Thermoelectric recirculating chillers Compressor based recirculating chillers Cabinet

More information

New Receipt Point Pre Inspection Checklist

New Receipt Point Pre Inspection Checklist New Receipt Point Pre Inspection Checklist The following checklist items are required to be completed prior to booking an initial inspection. Once this checklist has been completed, please fax a copy,

More information

WHITE PAPER. Solutions OnDemand Hosting Overview

WHITE PAPER. Solutions OnDemand Hosting Overview WHITE PAPER SAS Title Solutions OnDemand Hosting Overview ii Contents Overview... 1 Cary 1 (US) Facility Specifications...2 Cary 2 (US) Facility Specifications (SAS New Cloud Computing Center)...3 Charlotte

More information