MRI Image Quality Assessment
|
|
- Barnaby Scott
- 5 years ago
- Views:
Transcription
1 in partnership with MRI Image Quality Assessment David Collins CR-UK Cancer Imaging Centre, The Institute of Cancer Research Making the discoveries that defeat cancer
2 Overview Current Practice Quality Assurance and Acceptance testing Emerging applications Quantitative DCE-MRI, DWI Whole Body
3 Acceptance Testing Current Practice We measure 1, 2 : Resonance Frequency Signal to Noise Spatial Linearity Spatial Resolution Slice Thickness Slice Position/Separation Image Uniformity Artifacts; Phase and other ghosts Many others Dynamic and Quantitative Measurements, are the above sufficient? 1. Price et al Med Phys 17(2) Mar/Apr/ Och Med Phys 19(1) Jan/Feb 1992
4 Acceptance Testing Current Practice We rarely measure as specified 1, 2 : Eddy Currents (sequence programming or integrator circuit required) Slice Profiles (1D FT method) Nutation Angle calibration (vendor restrictions on user) SNR (Modern scanners have many Rx channels ) B0 (Phase images not produced), spectroscopy not always an option, vendors may provide field maps. B1 uniformity 1. Price et al Med Phys 17(2) Mar/Apr/ Och Med Phys 19(1) Jan/Feb 1992
5 Acceptance Testing Current Practice We rarely measure as specified 1, 2 : Eddy Currents (sequence programming or integrator circuit required) Slice Profiles (1D FT method) Nutation Angle calibration (vendor restrictions on user) SNR (Modern scanners have many Rx channels ) B0 (Phase images not produced), spectroscopy not always an option, vendors do provide field maps. B1 uniformity We rely on vendor installation tests and procedures, it is now vitally important to engage a site physicist in this process 1. Price et al Med Phys 17(2) Mar/Apr/ Och Med Phys 19(1) Jan/Feb 1992
6 Acceptance Testing Current Practice How long should we spend performing acceptance testing? At the Royal Marsden/ICR we allocate one week. At least two senior clinical scientists with trainees. We use a range of vendor, Eurospin test objects and home built test objects. Some sites (DGH) perform zero testing.
7 Examples of gradient linearity correction Previously gradient linearity required extensive evaluation and correction if MR images were used for R/T planning 1 Vendors now provide gradient linearity correction as either a default or selection option 1. Doran et al. Medical Physics 2005
8 NSA=2, Slice thickness=3mm Geometrical Distortions in Spin Echo Sequences Coronal Orientation NSA=1, Slice thickness=5mm NSA=2, Slice thickness=5mm NSA=2, Slice thickness=3mm, Without Distortion Correction Sequence 5mm, 1NSA 5mm, 2NSA 3mm, 2NSA 3mm, 2NSA, without Dist Cor 3mm, 2NSA, with Body Coil x 0.42 y 0.42 x 0.83 y 0.42 x 1.25 y 0.42 x 0.83 y 0 x 0.83 y 0 % deviation NSA=2, Slice thickness=3mm With Body Coil
9 Custom-built Quality Assurance equipment 1. 3T Achieva, without and with 3D distortion correction 2. Test object dimensions: 440mm LR, 400mm HF, 250mm AP 1.Tanner et al, Phys Med Biol 2000 Aug;45(8): Doran et al. Medical Physics 2005
10 Gradient Linearity QA Test Object Rapid 3D DCE-MRI protocol coronal acquisition No Gradient Correction Vendor Gradient Correction Dimensions 28 x 28cm Virtualscopics test object
11 Gradient Linearity QA Test Object Rapid 3D DCE-MRI protocol coronal acquisition No Gradient Correction Vendor Gradient Correction Dimensions 28 x 28cm Virtualscopics test object
12 SNR Well established tests for conventional Spin Echo. In practice single spin echo rarely used except acceptance testing Parallel imaging is now ubiquitous in clinical protocols. How do we assess SNR is a protocol specific way? Is it a practical measure with multiple channels and flexible coils?
13 Evaluation of SNR with G-noise Sum of Squares Adaptive Combine Mean 826 Mean 894 2D EPI, parallel imaging factor 2, different signal combination options All other measurement parameters fixed.
14 Evaluation of G-noise Sum of Squares Adaptive Combine Mean Mean snr = snr = Note the spatial dependence of the noise distribution Images acquired off resonance, alternatively transmitter at 0v.
15 Changing Role of MRI Moving from purely morphological to include functional quantitative imaging Moving from regional to whole body or large field of view imaging 32 Receiver channels 72 CP receiver elements in this configuration.
16 Functional Measures Functional MRI (fmri) minutes Dynamic susceptibility contrast (DSC-MRI) 1-2 minutes Dynamic Contrast Enhanced (DCE-MRI) 4-8 minutes Diffusion Weighted Imaging (DWI) 1-30 minutes
17 Functional Measures Functional MRI (fmri) minutes Dynamic susceptibility contrast (DSC-MRI) 1-2 minutes Dynamic Contrast Enhanced (DCE-MRI) 4-8 minutes Diffusion Weighted Imaging (DWI) 1-30 minutes 3 of the above typically use Echo Planar Imaging readout
18 Requirements for Functional Measures Temporal stability Rapid data acquisition, high parallel imaging factors Eddy Current compensation Uniformity of both B0 and B1 +
19 DCE-MRI Protocol Schematic MR images using different T1 contrast Dynamic Series Contrast agent injection Measurement of T 1 pre-injection Measurement of T 1 post-injection Schematic diagram of the data acquisition
20 Quality Control for DCE-MRI Acquisitions Dynamic contrast enhanced imaging requires an accurate and precise measurement of T1 changes to model pharmacokinetic behaviour. T1 measurements can be affected by many factors including Slice profile Scanner stability (RF and gradients) B1 variations K-space sampling Evaluate how these factors can effect T1 measurements with a series of phantom experiments at 1.5T and 3.0T.
21 Phantoms What should be assessed Signal(t) T1(t) [Gd](t) in true dynamic phantom that emulates tissue enhancement and is well characterized? extremely difficult! Range of T1 / T2 values of well characterized materials in static phantom. feasible! Spatial uniformity of transmit (B1) field System SNR under test and clinical DCE scan conditions CNR under DCE scan conditions Repeatability and reproducibility
22 T1 Measurement Accuracy Phased array coil Sequence parameters of T1 measurement methods Inversion recovery turbo-flash method TR/TE/θ/NSA = ms/1.14 ms/2 /1 39 inversion times varying from ms Variable flip angle method 3D GRE sequence TR/TE/NSA = 4.36 ms/1.36 ms/3 3 varying θs = 2, 8 and 12 Measurements were acquired with GRAPPA (ipat = 2) and without GRAPPA Siemens, Avanto (1.5T) A quick, easy and accurate method of T1 measurement is the variable flip angle (VFA) method 1 where gradient echo MR images are acquired using different flip angles, αi and T1 is calculated from the gradient of the linear plot of Si/sin(αi) vs. Si/tan(αi) A range of [Gd] doped water phantoms with T1 values ranging from 55ms-2865ms were evaluated. No major difference in T1 values obtained with and without partial parallel imaging 1. Fram et al Magn Reson Imaging. 1987;5(3):201-8.
23 Accuracy of Variable Flip Angle Method at 1.5T The accuracy and dynamic stability of clinical DCE-MRI protocols are assessed with phantom measurements Eurospin TO5 with 12 gels (T1 s ranging from 200 to 1300ms) were measured. In each experiment the phantom temperature was equilibrated and recorded. Dynamic stability was evaluated GRE sequences frequently exhibited an initial (40 second to 2 minutes) signal drift in T1W acquisitions. T1 Map ( )ms) Eurospin test object The signal drift in T1W images translates to 8-12% change in calculated T1 values
24 Stability of DCE-MRI measurements Dynamic acquisitions may have a 2% T1 variation over the initial seconds. Calculated T1 values across a dynamic time series had different dynamic variability depending on the sampling scheme used. Standard deviations ranged from 11.5ms (elliptical encoding) to 4.7ms (elliptical encoding with phase and slice undersampling). 1.5T No signal drift has been observed at 3T and dynamic variation is comparable with experiments at 1.5T. The standard deviation of 3D GRE sequences was reduced from 12.4ms to 6.4ms with the use of phase stabilisation. 3.0T Eurospin test object is temperature sensitive ~3% change in T1 with 1 o C (20-23 o C)
25 GRE Slice profiles Slice profiles obtained from measurements at different nutation angles using a silicone oil test object Slice profile results can be used to provide correction terms within the post-processing software to correct these errors. Essential to ensure that the slice positioning is centred on the region of interest. Sample T1 = 740ms Brookes JA et al; J Magn Reson Imaging Feb;9(2):
26 Inhomogeneous Transmit RF (B 1 ) Field RF transmit coils produce non-uniform field strengths FAs vary over the field of view % dt 1 vs. T 1 40% error in α % dt 1 30% error in α 20% error in α dt 1 2 dα 10% error in α T 1 (ms) B 1 θ Correct δt1 by performing in-vivo B 1 mapping Sung et al Magn Reson Med Oct;70(4):954-61
27 B1 1 Typical value for breast at 3.0T ROI Fatty tissue ROI T1 = 77ms. Original T1 estimate nutation angles 3 o and 16 o 1. Linderman et al JMRI (40); Data courtesy of Elizabeth O Flynn
28 DCE-MRI QA Assessments Noise Factor evaluated from full DCE-MRI protocol Useful for comparing protocols Can be applied to pre-enhanced clinical data SNR evaluated from full DCE- MRI protocol Useful for comparing protocols Useful for routine QA Kurland RJ MRM 2; Imran J et al MRI 17;
29 What is diffusion MR? Signal is exponentially dependent on the ADC and the sequence parameters (b-value): b = 0 s mm -2 b = 50 s mm -2 b = 100 s mm -2 ADC map b = 250 s mm -2 b = 500 s mm -2 b = 750 s mm -2
30 Diffusion Weighted Imaging Quality Assurance Use a range of phantoms: ADC accuracy use Sucrose, PVP (temperature sensitive) Ice water, temperature insensitive (long preparation >60mins) Silicone oil, non diffusing material (artefact assessments) Flood phantom ADC linearity Fat/Water test object (evaluate fat suppression efficiency) Others; alkanes for ADC, standard geometrical test objects
31 Diffusion Encoding pulse schemes Unipolar Diffusion encoding scheme Stejskal-Tanner Scanning with shorter TE Bipolar Diffusion encoding scheme Reduction of eddy currents induced spatial distortions 1 1. Chan et al J Magn Reson Jul;244:74-84
32 EPI N/2 Ghosting Axial images of a PDMS phantom acquired using receive bandwidths (a) 1628 Hx/pixel, (b) 1776 Hx/pixel, (c) 1860 Hx/pixel, (d) 2170 Hx/pixel, (e) 2298 Hx/ pixel, (f) 2442 Hx/pixel, (g) 2790 Hx/pixel, (h) 3004 Hx/pixel b = 0 s mm -2 ). Images have been windowed to enhance the visibility of the ghosts, at the same window width and level. Target to reduce ghosting below 3%
33 EPI N/2 Ghosting Axial images of a PDMS phantom acquired using receive bandwidths (a) 1628 Hx/pixel, (b) 1776 Hx/pixel, (c) 1860 Hx/pixel, (d) 2170 Hx/pixel, (e) 2298 Hx/ pixel, (f) 2442 Hx/pixel, (g) 2790 Hx/pixel, (h) 3004 Hx/pixel b = 0 s mm -2 ). Images have been windowed to enhance the visibility of the ghosts, at the same window width and level. What bandwidth should we use for N/2 ghost assessment?
34 Optimization of protocol critical to success Eddy currents: image distortion b0 b0 b1000 : b0 (background) 754 Hz/px 1450 Hz/px 2012 Hz/px 2416 Hz/px
35 Optimization of protocol critical to success Eddy currents: image distortion b0 b0 b1000 : b0 (background) 754 Hz/px 1450 Hz/px 2012 Hz/px 2416 Hz/px
36 (a) monopolar (b) DSE Axial images of a PDMS phantom (a, b) b = 0 s mm -2, (c, d) b = 1000 s mm -2 and (e, f) subtraction images. Images (a, c) acquired using a monopolar sequence and (b, d) using DSE Subtraction images show the b = 0 s mm -2 image subtracted from the b = 1000 s mm -2 image. (c) (d) (e) (f)
37 Quality Assurance ADC Accuracy ADC o C = mm 2 /s Malyarenko D et al J Magn Reson Imaging May;37(5):
38 Polyvinylpyrrolidone (PVP) phantom The cylindrical ice water phantom was used for the experiment, containing the vials with the different PVP concentrations. # PVP (% w/w) ROI 1 ROI 2 ROI 3 ROI 4 ROI 5 ROI 6 ROI min before the scanning the phantom was refilled with ice-water, in order to obtain a temperature of 00C. Dr Marianthi-Vasiliki Papoutsaki
39 ADC estimates from PVP phantom C PVP (% w/w) ADC median (*10-3 mm 2 /s) 0 - ROI ROI ROI ROI ROI ROI ROI ADC map Linear relationship between PVP concentration and R 2 ADC. = Boss et al ISMRM 2013 Dr. Marianthi-Vasiliki Papoutsaki
40 ADC Uniformity Assessment ADC ADC profile Evaluation of ADC uniformity is essential for quantitative WBDWI Malyarenko et al. Magn Reson Med May 13. Courtesy Dr. Jessica Winfield
41 ADC profile ADC profile Non -Uniform Gradient ~2% ADC ADC ADC uniformity is image acquisition, reconstruction and post-processing dependent Courtesy Dr. Jessica Winfield
42 ADC profiles along x and z axis of bipolar and unipolar diffusion pulse x- axis Median ADC value +/- 2.5% Centred at isocentre ADC map bipolar z- axis Median ADC value +/- 2.5% Centred at isocentre 53% deviation from the median value in bipolar Dr Marianthi-Vasiliki Papoutsaki
43 Fat-Water Test-object Inner cylinder: water/nacl/cuso mm coronal 400 mm x 400 mm FOV Annulus: Corn oil Gradient echo localiser images. Test object placed at 45 degrees to z-axis to create large ellipse in axial slices. 185 mm 140 mm axial 400 mm x 400 mm FOV Winfield et al Phys Med Biol May 7;59(9):
44 Upfield and downfield fat signals PE direction = AP Test object: No fat suppression, b=900 Volunteer: No fat suppression, b=900 upfield fat (1.3 ppm) shifted in anterior direction downfield fat (5.3 ppm) shifted in posterior direction Courtesy Dr. Jessica Winfield
45 Unsuppressed upfield fat at edges of FOV using SPAIR at 1.5 T Test object: SPAIR, b=900 Volunteer: SPAIR, b=900 Unsuppressed upfield fat at edges of FOV in test object and volunteer SPAIR leaves downfield fat signal unsuppressed Courtesy Dr Jessica Winfield
46 Whole Body Dixon (Fat/Water) Whole Body Dixon registered and fused with WBDWI a) WBDWI, b) Water, c) Fused a)+b), d) Dixon T1 map, e) Water ratio
47 Whole Body Dixon (Fat/Water) Whole Body Dual Contrast Dixon registered and fused with WBDWI a) WBDWI, b) Water, c) Fused a)+b), d) Dixon T1 map, e) Water ratio Three quantitative whole body metrics ADC, T1 and Fat/Water ratio Blackledge et al ISMRM 2009
48 Semi-automatic segmentation + b = 0 s/mm 2 Acquired data ADC map Select computed b-value and initial disease threshold Smoothing of regions using Markov random field model (MRF) Visualization/quantification of disease User modifiable regions of interest (ROI) Blackledge et al PLoS One Apr 7;9(4):e91779.
49 ADC stats (x10-3 mm2/s) Pretreatment Posttreatment Mean Variance Skewness Kurtosis
50 Discussion Routine Acceptance testing and QA informs only on the important basic functionality Broader range of test objects required How well do phantom measurements translate into meaningful measures (eg ADC repeatabilty, DCE-MRI noise factors)? QA has to move from the generic to the assessment of specific clinical protocols
51 in partnership with Acknowledgements Dr James d Arcy Dr Jessica Winfield Dr Mihaela Rata Dr Marianthi-Vasiliki Papoutsaki Dr Matthew Blackledge
52
Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.
Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing
More informationCOBRE Scan Information
COBRE Scan Information Below is more information on the directory structure for the COBRE imaging data. Also below are the imaging parameters for each series. Directory structure: var/www/html/dropbox/1139_anonymized/human:
More informationHST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008
MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.
More informationSIEMENS MAGNETOM Skyra syngo MR D13
Page 1 of 8 SIEMENS MAGNETOM Skyra syngo MR D13 \\USER\CIND\StudyProtocols\PTSA\*dm_ep2d_mono70_b0_p2_iso2.0 TA:1:05 PAT:2 Voxel size:2.0 2.0 2.0 mm Rel. SNR:1.00 :epse Properties Routine Prio Recon Load
More informationHST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008
MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.
More informationMRI Physics II: Gradients, Imaging
MRI Physics II: Gradients, Imaging Douglas C., Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor Magnetic Fields in MRI B 0 The main magnetic field. Always on (0.5-7 T) Magnetizes
More informationHST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006
MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.
More informationLab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street
Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street MRI is located in the sub basement of CC wing. From Queen or Victoria, follow the baby blue arrows and ride the CC south
More informationModule 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE
MRES 7005 - Fast Imaging Techniques Module 4 K-Space Symmetry Review K-Space Review K-Space Symmetry Partial or Fractional Echo Half or Partial Fourier HASTE Conditions for successful reconstruction Interpolation
More informationRole of Parallel Imaging in High Field Functional MRI
Role of Parallel Imaging in High Field Functional MRI Douglas C. Noll & Bradley P. Sutton Department of Biomedical Engineering, University of Michigan Supported by NIH Grant DA15410 & The Whitaker Foundation
More informationDiffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford
Diffusion MRI Acquisition Karla Miller FMRIB Centre, University of Oxford karla@fmrib.ox.ac.uk Diffusion Imaging How is diffusion weighting achieved? How is the image acquired? What are the limitations,
More informationParallel Imaging. Marcin.
Parallel Imaging Marcin m.jankiewicz@gmail.com Parallel Imaging initial thoughts Over the last 15 years, great progress in the development of pmri methods has taken place, thereby producing a multitude
More informationM R I Physics Course
M R I Physics Course Multichannel Technology & Parallel Imaging Nathan Yanasak, Ph.D. Jerry Allison Ph.D. Tom Lavin, B.S. Department of Radiology Medical College of Georgia References: 1) The Physics of
More informationMRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health
MRI image formation Indiana University School of Medicine and Indiana University Health Disclosure No conflict of interest for this presentation 2 Outlines Data acquisition Spatial (Slice/Slab) selection
More informationSIEMENS MAGNETOM TrioTim syngo MR B17
\\USER\KNARRGROUP\MultiBand\LavretskyMultiBand\trufi localizer 3-plane TA: 5.1 s PAT: Voxel size: 1.2 1.2 5. Rel. SNR: 1.00 SIEMENS: trufi Load to stamp Slice group 1 Slices 1 Dist. factor 20 % Phase enc.
More informationSIEMENS MAGNETOM Avanto syngo MR B15
\\USER\INVESTIGATORS\Ravi\ADNI-Subject\Localizer TA: 0:10 PAT: Voxel size: 1.9 1.5 8.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer Inline movie
More informationSIEMENS MAGNETOM Avanto syngo MR B15
\\USER\INVESTIGATORS\Ravi\ADNI-phantom\QC Phantom-Localizer TA: 0:10 PAT: Voxel size: 1.9 1.5 8.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer
More informationSingle Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction
Single Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction Felix Lugauer 1,3, Jens Wetzl 1, Christoph Forman 2, Manuel Schneider 1, Berthold Kiefer
More informationLucy Phantom MR Grid Evaluation
Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is
More informationSupplementary methods
Supplementary methods This section provides additional technical details on the sample, the applied imaging and analysis steps and methods. Structural imaging Trained radiographers placed all participants
More information(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions
EE225E/BIOE265 Spring 2013 Principles of MRI Miki Lustig Assignment 8 Solutions 1. Nishimura 7.1 P)jc%z 5 ivcr3. 1. I Due Wednesday April 10th, 2013 (a Scrhon5 R2iwd b 0 ZOms Xn,s r cx > qs 4-4 8ni6 4
More informationSparse sampling in MRI: From basic theory to clinical application. R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology
Sparse sampling in MRI: From basic theory to clinical application R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology Objective Provide an intuitive overview of compressed sensing
More informationFast Imaging Trajectories: Non-Cartesian Sampling (1)
Fast Imaging Trajectories: Non-Cartesian Sampling (1) M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.05.03 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business
More informationSIEMENS MAGNETOM Verio syngo MR B15V
\\USER\ZAHID_RESEARCH\MS\No Name\3D SWI TA: 6:39 PAT: 2 Voxel size: 1.0 0.5 2.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer Inline movie Auto
More informationSIEMENS MAGNETOM Verio syngo MR B17
\\USER\Dr. Behrmann\routine\Ilan\ep2d_bold_PMU_resting TA: 8:06 PAT: Voxel size: 3.03.03.0 mm Rel. SNR: 1.00 USER: ep2d_bold_pmu Properties Special sat. Prio Recon System Before measurement Body After
More informationCHAPTER 9: Magnetic Susceptibility Effects in High Field MRI
Figure 1. In the brain, the gray matter has substantially more blood vessels and capillaries than white matter. The magnified image on the right displays the rich vasculature in gray matter forming porous,
More informationField Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005
Field Maps John Pauly October 5, 25 The acquisition and reconstruction of frequency, or field, maps is important for both the acquisition of MRI data, and for its reconstruction. Many of the imaging methods
More informationAbbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology
Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Outline High Spectral and Spatial Resolution MR Imaging (HiSS) What it is How to do it Ways to use it HiSS for Radiation
More informationRemoval of EPI Nyquist Ghost Artifacts With Two- Dimensional Phase Correction
Removal of EPI Nyquist Ghost Artifacts With Two- Dimensional Phase Correction Nan-kuei Chen 1,5 and Alice M. Wyrwicz 4 * Magnetic Resonance in Medicine 51:147 153 (004) Odd even echo inconsistencies result
More informationDynamic Autocalibrated Parallel Imaging Using Temporal GRAPPA (TGRAPPA)
Magnetic Resonance in Medicine 53:981 985 (2005) Dynamic Autocalibrated Parallel Imaging Using Temporal GRAPPA (TGRAPPA) Felix A. Breuer, 1 * Peter Kellman, 2 Mark A. Griswold, 1 and Peter M. Jakob 1 Current
More informationNew Technology Allows Multiple Image Contrasts in a Single Scan
These images were acquired with an investigational device. PD T2 T2 FLAIR T1 MAP T1 FLAIR PSIR T1 New Technology Allows Multiple Image Contrasts in a Single Scan MR exams can be time consuming. A typical
More informationOrthopedic MRI Protocols. Philips Panorama HFO
Orthopedic MRI Protocols Philips Panorama HFO 1 2 Prepared in collaboration with Dr. John F. Feller, Medical Director of Desert Medical Imaging, Palm Springs, CA. Desert Medical Imaging will provide the
More informationAutomatic Gradient Preemphasis Adjustment: A 15-Minute Journey to Improved Diffusion-Weighted Echo-Planar Imaging
Automatic Gradient Preemphasis Adjustment: A 15-Minute Journey to Improved Diffusion-Weighted Echo-Planar Imaging Vincent J. Schmithorst* and Bernard J. Dardzinski Magnetic Resonance in Medicine 47:208
More informationFast Imaging UCLA. Class Business. Class Business. Daniel B. Ennis, Ph.D. Magnetic Resonance Research Labs. Tuesday (3/7) from 6-9pm HW #1 HW #2
Fast Imaging Daniel B. Ennis, Ph.D. Magnetic Resonance Research Labs Class Business Tuesday (3/7) from 6-9pm 6:00-7:30pm Groups Avanto Sara Said, Yara Azar, April Pan Skyra Timothy Marcum, Diana Lopez,
More informationBreast MRI Accreditation Program Clinical Image Quality Guide
Breast MRI Accreditation Program Clinical Image Quality Guide Introduction This document provides guidance on breast MRI clinical image quality and describes the criteria used by the ACR Breast MRI Accreditation
More informationCOMPREHENSIVE QUALITY CONTROL OF NMR TOMOGRAPHY USING 3D PRINTED PHANTOM
COMPREHENSIVE QUALITY CONTROL OF NMR TOMOGRAPHY USING 3D PRINTED PHANTOM Mažena MACIUSOVIČ *, Marius BURKANAS *, Jonas VENIUS *, ** * Medical Physics Department, National Cancer Institute, Vilnius, Lithuania
More informationClinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis
Clinical Importance Rapid cardiovascular flow quantitation using sliceselective Fourier velocity encoding with spiral readouts Valve disease affects 10% of patients with heart disease in the U.S. Most
More informationHead motion in diffusion MRI
Head motion in diffusion MRI Anastasia Yendiki HMS/MGH/MIT Athinoula A. Martinos Center for Biomedical Imaging 11/06/13 Head motion in diffusion MRI 0/33 Diffusion contrast Basic principle of diffusion
More informationSources of Distortion in Functional MRI Data
Human Brain Mapping 8:80 85(1999) Sources of Distortion in Functional MRI Data Peter Jezzard* and Stuart Clare FMRIB Centre, Department of Clinical Neurology, University of Oxford, Oxford, UK Abstract:
More informationThe SIMRI project A versatile and interactive MRI simulator *
COST B21 Meeting, Lodz, 6-9 Oct. 2005 The SIMRI project A versatile and interactive MRI simulator * H. Benoit-Cattin 1, G. Collewet 2, B. Belaroussi 1, H. Saint-Jalmes 3, C. Odet 1 1 CREATIS, UMR CNRS
More informationA Model-Independent, Multi-Image Approach to MR Inhomogeneity Correction
Tina Memo No. 2007-003 Published in Proc. MIUA 2007 A Model-Independent, Multi-Image Approach to MR Inhomogeneity Correction P. A. Bromiley and N.A. Thacker Last updated 13 / 4 / 2007 Imaging Science and
More informationParallel Magnetic Resonance Imaging (pmri): How Does it Work, and What is it Good For?
Parallel Magnetic Resonance Imaging (pmri): How Does it Work, and What is it Good For? Nathan Yanasak, Ph.D. Chair, AAPM TG118 Department of Radiology Georgia Regents University Overview Phased-array coils
More informationOutline: Contrast-enhanced MRA
Outline: Contrast-enhanced MRA Background Technique Clinical Indications Future Directions Disclosures: GE Health Care: Research support Consultant: Bracco, Bayer The Basics During rapid IV infusion, Gadolinium
More informationFunctional MRI in Clinical Research and Practice Preprocessing
Functional MRI in Clinical Research and Practice Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization
More informationFmri Spatial Processing
Educational Course: Fmri Spatial Processing Ray Razlighi Jun. 8, 2014 Spatial Processing Spatial Re-alignment Geometric distortion correction Spatial Normalization Smoothing Why, When, How, Which Why is
More informationHST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008
MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.
More informationGE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals.
GE Healthcare CLINICAL GALLERY Discovery * MR750w 3.0T This brochure is intended for European healthcare professionals. NEURO PROPELLER delivers high resolution, motion insensitive imaging in all planes.
More informationA Virtual MR Scanner for Education
A Virtual MR Scanner for Education Hackländer T, Schalla C, Trümper A, Mertens H, Hiltner J, Cramer BM Hospitals of the University Witten/Herdecke, Department of Radiology Wuppertal, Germany Purpose A
More informationMR Advance Techniques. Vascular Imaging. Class III
MR Advance Techniques Vascular Imaging Class III 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology
More informationWhat is pmri? Overview. The Need for Speed: A Technical and Clinical Primer for Parallel MR Imaging 8/1/2011
The Need for Speed: A Technical and Clinical Primer for Parallel MR Imaging Nathan Yanasak, Ph.D. Chair, AAPM TG118 Assistant Professor Department of Radiology Director, Core Imaging Facility for Small
More informationSPM8 for Basic and Clinical Investigators. Preprocessing
SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial
More informationRelease Notes. Multi-Band EPI C2P. Release HCP_v1 10 February 2014
Release Notes Multi-Band EPI C2P Release HCP_v1 10 February 2014 Installation 1. Restart the system (reboot host and MRIR) 2. Extract the.zip file to a temporary directory 3. Run the installer.bat file
More informationMRI. When to use What sequences. Outline 2012/09/19. Sequence: Definition. Basic Principles: Step 2. Basic Principles: Step 1. Govind Chavhan, MD
MRI When to use What sequences Govind Chavhan, MD Assistant Professor and Staff Radiologist The Hospital For Sick Children, Toronto Planning Acquisition Post processing Interpretation Patient history and
More informationSpiral keyhole imaging for MR fingerprinting
Spiral keyhole imaging for MR fingerprinting Guido Buonincontri 1, Laura Biagi 1,2, Pedro A Gómez 3,4, Rolf F Schulte 4, Michela Tosetti 1,2 1 IMAGO7 Research Center, Pisa, Italy 2 IRCCS Stella Maris,
More informationSPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing
SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial
More informationUse of MRI in Radiotherapy: Technical Consideration
Use of MRI in Radiotherapy: Technical Consideration Yanle Hu, PhD Department of Radiation Oncology, Mayo Clinic Arizona 04/07/2018 2015 MFMER slide-1 Conflict of Interest: None 2015 MFMER slide-2 Objectives
More informationHigh Fidelity Brain Connectivity Imaging
CNI Inauguration Workshop Stanford, March 22 nd, 2012 High Fidelity Brain Connectivity Imaging -Recent Progress on Diffusion Weighted MRI for High Resolution and Low Distortion Allen W. Song, PhD Brain
More informationCompressed Sensing for Rapid MR Imaging
Compressed Sensing for Rapid Imaging Michael Lustig1, Juan Santos1, David Donoho2 and John Pauly1 1 Electrical Engineering Department, Stanford University 2 Statistics Department, Stanford University rapid
More informationSPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.
SPECT QA and QC Bruce McBride St. Vincent s Hospital Sydney. SPECT QA and QC What is needed? Why? How often? Who says? QA and QC in Nuclear Medicine QA - collective term for all the efforts made to produce
More informationA Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data
A Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data Alexey Samsonov, Julia Velikina Departments of Radiology and Medical
More informationModule 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution.
MRES 7005 - Fast Imaging Techniques Module 5: Dynamic Imaging and Phase Sharing (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review Improving Temporal Resolution True-FISP (I) True-FISP (II) Keyhole
More informationApplications Guide for Interleaved
Applications Guide for Interleaved rephase/dephase MRAV Authors: Yongquan Ye, Ph.D. Dongmei Wu, MS. Tested MAGNETOM Systems : 7TZ, TRIO a Tim System, Verio MR B15A (N4_VB15A_LATEST_20070519) MR B17A (N4_VB17A_LATEST_20090307_P8)
More informationBasic fmri Design and Analysis. Preprocessing
Basic fmri Design and Analysis Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial filtering
More informationInformation about presenter
Information about presenter 2013-now Engineer R&D ithera Medical GmbH 2011-2013 M.Sc. in Biomedical Computing (TU München) Thesis title: A General Reconstruction Framework for Constrained Optimisation
More informationPhilips MRI Protocol Dump Created on Comment Software Stream
Page 1 of 5 Philips MRI Protocol Dump Created on 2/17/2011 4:11:01 PM Comment Created by ExamCard_to_XML with inputs: "J:\ADNI GO - ADNI 2 Phantom5.ExamCard" on system (BU SCHOOL OF MEDICINE :: 192.168.71.10)
More informationEPI Data Are Acquired Serially. EPI Data Are Acquired Serially 10/23/2011. Functional Connectivity Preprocessing. fmri Preprocessing
Functional Connectivity Preprocessing Geometric distortion Head motion Geometric distortion Head motion EPI Data Are Acquired Serially EPI Data Are Acquired Serially descending 1 EPI Data Are Acquired
More informationConstrained Reconstruction of Sparse Cardiac MR DTI Data
Constrained Reconstruction of Sparse Cardiac MR DTI Data Ganesh Adluru 1,3, Edward Hsu, and Edward V.R. DiBella,3 1 Electrical and Computer Engineering department, 50 S. Central Campus Dr., MEB, University
More informationWhite Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space
White Pixel Artifact Caused by a noise spike during acquisition Spike in K-space sinusoid in image space Susceptibility Artifacts Off-resonance artifacts caused by adjacent regions with different
More informationADNI, ADNI_QH, SURVEY. Geometry. connection
ADNI, ADNI_QH, SURVEY Geometry Coil selection = Head connection = d Multi coil Homogeneity correction ne FOV (mm) = 250.00 RFOV (%) = 100.00 Foldover suppression Matrix scan = 256 reconstruction = 256
More informationAdvanced Imaging Trajectories
Advanced Imaging Trajectories Cartesian EPI Spiral Radial Projection 1 Radial and Projection Imaging Sample spokes Radial out : from k=0 to kmax Projection: from -kmax to kmax Trajectory design considerations
More information2. Creating Field Maps Using the Field Map GUI (Version 2.0) in SPM5
1. Introduction This manual describes how to use the Field Map Toolbox Version 2.0 for creating unwrapped field maps that can be used to do geometric distortion correction of EPI images in SPM5. 1. 1.
More informationAutomated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator
Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer Centre Western General Hospital EDINBURGH EH4
More informationFunctional analysis with DTI and diffusion-neurography of cranial nerves
Functional analysis with DTI and diffusion-neurography of cranial nerves Poster No.: C-1942 Congress: ECR 2013 Type: Educational Exhibit Authors: J. P. Martínez Barbero, T. Martín Noguerol, A. Luna Alcalá;
More informationT 1 MAPPING FOR DCE-MRI
T 1 MAPPING FOR DCE-MRI A dissertation submitted to the Faculty of Medicine, University of Malaya in partial fulfillment of the requirements for the degree of Master of Medical Physics By NURUN NAJWA BINTI
More informationXI Signal-to-Noise (SNR)
XI Signal-to-Noise (SNR) Lecture notes by Assaf Tal n(t) t. Noise. Characterizing Noise Noise is a random signal that gets added to all of our measurements. In D it looks like this: while in D
More informationCT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION
CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology
More informationEstimating 3D Respiratory Motion from Orbiting Views
Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng, Jeffrey A. Fessler, James M. Balter The University of Michigan Oct. 2005 Funding provided by NIH Grant P01 CA59827 Motivation Free-breathing
More informationAn Introduction to Image Reconstruction, Processing, and their Effects in FMRI
An Introduction to Image Reconstruction, Processing, and their Effects in FMRI Daniel B. Rowe Program in Computational Sciences Department of Mathematics, Statistics, and Computer Science Marquette University
More information2.1 Signal Production. RF_Coil. Scanner. Phantom. Image. Image Production
An Extensible MRI Simulator for Post-Processing Evaluation Remi K.-S. Kwan?, Alan C. Evans, and G. Bruce Pike McConnell Brain Imaging Centre, Montreal Neurological Institute, McGill University, Montreal,
More informationSuper-Resolution Reconstruction of Diffusion-Weighted Images from Distortion Compensated Orthogonal Anisotropic Acquisitions.
Super-Resolution Reconstruction of Diffusion-Weighted Images from Distortion Compensated Orthogonal Anisotropic Acquisitions. Benoit Scherrer Ali Gholipour Simon K. Warfield Children s Hospital Boston,
More informationExam 8N080 - Introduction MRI
Exam 8N080 - Introduction MRI Friday January 23 rd 2015, 13.30-16.30h For this exam you may use an ordinary calculator (not a graphical one). In total there are 6 assignments and a total of 65 points can
More informationsurface Image reconstruction: 2D Fourier Transform
2/1/217 Chapter 2-3 K-space Intro to k-space sampling (chap 3) Frequenc encoding and Discrete sampling (chap 2) Point Spread Function K-space properties K-space sampling principles (chap 3) Basic Contrast
More informationMagnetic Resonance Angiography
Magnetic Resonance Angiography Course: Advance MRI (BIOE 594) Instructors: Dr Xiaohong Joe Zhou Dr. Shadi Othman By, Nayan Pasad Phase Contrast Angiography By Moran 1982, Bryan et. Al. 1984 and Moran et.
More informationTechnological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI. Floris Jansen, GE Healthcare July, 2015
Technological Advances and Challenges: Experience with Time-Of-Flight PET Combined with 3T MRI Floris Jansen, GE Healthcare July, 2015 PET/MR 101 : challenges Thermal Workflow & Apps RF interactions?!!
More informationPre-processing of ASL data T CT
Wed October 2, 2013 Image Processing Pre-processing: motion correction, denoising, outlier detection Alessandra Bertoldo Pre-processing of ASL data T CT C T C Single TI ASL T T T T C CCC average Pre-processing
More informationControlled Aliasing in Volumetric Parallel Imaging (2D CAIPIRINHA)
Magnetic Resonance in Medicine 55:549 556 (2006) Controlled Aliasing in Volumetric Parallel Imaging (2D CAIPIRINHA) Felix A. Breuer,* Martin Blaimer, Matthias F. Mueller, Nicole Seiberlich, Robin M. Heidemann,
More informationDynamic Contrast enhanced MRA
Dynamic Contrast enhanced MRA Speaker: Yung-Chieh Chang Date : 106.07.22 Department of Radiology, Taichung Veterans General Hospital, Taichung, Taiwan 1 Outline Basic and advanced principles of Diffusion
More informationHigh dynamic range magnetic resonance flow imaging in the abdomen
High dynamic range magnetic resonance flow imaging in the abdomen Christopher M. Sandino EE 367 Project Proposal 1 Motivation Time-resolved, volumetric phase-contrast magnetic resonance imaging (also known
More informationImaging Notes, Part IV
BME 483 MRI Notes 34 page 1 Imaging Notes, Part IV Slice Selective Excitation The most common approach for dealing with the 3 rd (z) dimension is to use slice selective excitation. This is done by applying
More informationClassification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging
1 CS 9 Final Project Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging Feiyu Chen Department of Electrical Engineering ABSTRACT Subject motion is a significant
More informationQualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs
Qualitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Doug Dean Final Project Presentation ENGN 2500: Medical Image Analysis May 16, 2011 Outline Review of the
More informationInitial Experience of Applying TWIST Dixon with Flexible View Sharing in Breast DCE-MRI
Initial Experience of Applying TWIST Dixon with Flexible View Sharing in Breast DCE-MRI Yuan Le PhD 1, Hal D. Kipfer MD 1, Dominik M. Nickel PhD 2, Randall Kroeker PhD 2, Brian Dale PhD 2, Stephanie P.
More informationImproved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm
Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm L.P. Panych 1,3, B. Madore 1,3, W.S. Hoge 1,3, R.V. Mulkern 2,3 1 Brigham and
More informationG Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine. Compressed Sensing
G16.4428 Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine Compressed Sensing Ricardo Otazo, PhD ricardo.otazo@nyumc.org Compressed
More informationMagnetic Resonance Elastography (MRE) of Liver Disease
Magnetic Resonance Elastography (MRE) of Liver Disease Authored by: Jennifer Dolan Fox, PhD VirtualScopics Inc. jennifer_fox@virtualscopics.com 1-585-249-6231 1. Overview of MRE Imaging MRE is a magnetic
More informationFollowing on from the two previous chapters, which considered the model of the
Chapter 5 Simulator validation Following on from the two previous chapters, which considered the model of the simulation process and how this model was implemented in software, this chapter is concerned
More informationMEDICAL IMAGE COMPUTING (CAP 5937) LECTURE 4: Pre-Processing Medical Images (II)
SPRING 2016 1 MEDICAL IMAGE COMPUTING (CAP 5937) LECTURE 4: Pre-Processing Medical Images (II) Dr. Ulas Bagci HEC 221, Center for Research in Computer Vision (CRCV), University of Central Florida (UCF),
More informationLorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc.
Lorad FFDM QC Procedures for Medical Physicist Tao Wu, Ph.D. Hologic, Inc. Lorad Selenia FFDM Tests Following Lorad QC Manual Collimation Assessment Artifact Evaluation System Resolution Phantom Image
More informationSiemens AG, Healthcare Sector. syngo MR D13 0. Supplement - Parameters and image text 0.
Siemens AG, Healthcare Sector 0 0 n.a. English Cs2 syngo Neuro Operator MR-05014 630 05/2010 01 02 Informatik, Manual D11 Cape syngo MR D13 0.0 Supplement - Parameters and image text 0. syngo MR D13 0.
More informationK-Space Trajectories and Spiral Scan
K-Space and Spiral Scan Presented by: Novena Rangwala nrangw2@uic.edu 1 Outline K-space Gridding Reconstruction Features of Spiral Sampling Pulse Sequences Mathematical Basis of Spiral Scanning Variations
More information