Body Diffusion MRI: Basics and Beyond

Size: px
Start display at page:

Download "Body Diffusion MRI: Basics and Beyond"

Transcription

1 What is Diffusion? The process of Brownian motion Body Diffusion MRI: Basics and Beyond Xiaodong Zhong, PhD Senior R&D Expert, MR R&D Collaborations, Siemens Healthcare Adjunct Assistant Professor, Radiologyand Imaging Science, Emory University Atlanta, GA Siemens GmbH 2016 All rights reserved. Answers for life. Source: Body Diffusion MRI: Basics and Beyond Outline What is Diffusion? Basics of diffusion MRI and body diffusion Physics Pulse sequence Apparent diffusion coefficient (ADC) Common artifacts and parameter optimization Diffusion in Isotropic Medium e.g. water E1 E2 E3 Advanced topics New diffusion sequence techniques Motion correction / registration Advanced acceleration technique Diffusion in Oriented Tissue E3 E2 E1 Body Diffusion MRI: Basics and Beyond Outline What is Diffusion Imaging? 180 o Basics of diffusion MRI and body diffusion Physics Pulse sequence Apparent diffusion coefficient (ADC) Common artifacts and parameter optimization more mobility 90 o Phase dispersion time Advanced topics New diffusion sequence techniques Motion correction / registration Advanced acceleration technique restricted mobility ` Signal difference No differentiation between bound and free proton Handout 1

2 Body Diffusion Diffusion Sequence Card Parameters Basic Ideas (2) DW images (b-value images) Reflect the estimate of the water diffusion rate at that pixel The greater the b-value, the stronger the diffusion weighting, and the higher the contrast (hyperintense) in pathogenic regions (reduced diffusion) Primary application for early detection of cerebral ischemic stroke, more sensitive to early changes after a stroke than T1 or T2 weighted images* b = 0, not DW (T2 weighted image) b = 500 s/mm 2 b = 1000 s/mm 2 * S. Warach et al., J. Cereb. Blood Flow Metab. 16 (1996) Diffusion Encoding Schemes and Sequence Designs Basic Ideas (3) RF ADC Gdiff Gr Gp Bipolar diffusion encoding scheme. Benefits: better eddy current compensation => less spatial distortions better for WB-DWI Apparent diffusion coefficient (ADC) A measure of the strength (velocity) of diffusion in tissue Free of the influence of T1 and T2 effects The stronger the diffusion, the greater the diffusion coefficient Exhibits darker contrast (hypointense) in pathogenic regions (reduced diffusion) RF ADC Gdiff Monopolar (Stejskal-Tanner) diffusion encoding scheme. Compare G r Benefits: shorter TE s possible ADC map B = 1000 s/mm 2 Gp higher SNR => less averages needed better for DWI in breast, abdomen, pelvis Basic Ideas (1) Diffusion Weighted Imaging (DWI) Fundamental Equation Diffusion weighting factor, b-value Sets the measurement sensitivity to diffusion Determined by strength of diffusion gradient (G), duration of gradient ( ) and duration between the two gradients ( ) Direction of sensitivity can be altered by changing diffusion gradient direction diffusion gradient G diffusion gradient G S(b) S(0) time b Handout 2

3 Diffusion Weighted Imaging (DWI) Fitting the ADC Generally, more b-values give better fitting Basic Ideas (3) Anisotropy: In tissue water diffusion is limited by tissue boundaries Diffusion encoding can be applied to any directions, but typically three main orthogonal axes If the same b-value is used 2-pt: / -> partly due to the perfusion effect 3-pt: / 6-pt: 0.7 / : Isotropic DW image (geometric averaging), or trace-weighted image (TraceW) Averaged ADC map Reflect diffusion weight independent of diffusion orientation. ADC b-value filter for ADC maps calculation Diffusion Mode Options Selectable b-value range for ADC calculation eliminate perfusion effects by excluding lower b-values (organ specific e.g. < 150 s/mm 2 ) from ADC calculation ADC calc. from b 0,100,600 ADC calc. from b 100, 600 Can be used to replace 3- Scan Trace. Will add SNR Used to images in most but applications will also and increase is the time standard for all diffusion sequences Used in Whole body diffusion. Reduces TE and overall scan time. Allows for larger amounts of averaging. Note: No Trace weighted Images Diffusion Weighted Imaging (DWI) What is the optimal b-value? DWI Examples Free breathing EP2D-DIFF 3-Scan Trace b=50 s/mm 2 b=400 s/mm 2 b=800 s/mm 2 Siemens recommended b-values Anatomy b-value1 b-value 2 b-value 3 Liver Prostate Cervix Kidney Pancreas Whole Body Averaged ADC Handout 3

4 Basic Ideas (4) DWI b-value Image Extrapolation Multiple measurements/repetitions can be used to improve SNR Calculated b-value Multiple b-value images can be acquired to improve the ADC calculation Different averages can be set to different b-values to efficiently use the scan time The DW images can be inverted to mimic a PET-like appearance S(b) = image signal at b-value b b = b - b 0 b-value Averaging DWI Examples b-value Image Extrapolation Example liver DWI b = 0 s/mm 2 b = 750 s/mm 2 b = 2000 s/mm 2 Extrapolating images to b = 2000 s/mm 2 reveals the presence of water restriction within a peri-tumoralmargin in liver metastases (red arrow). Furthermore, it is observed that the contrast between metastases and background tissue is improved (green arrows). Courtesy Courtesy University Royal of Marsden Homburg/Saar, Hospital Germany / UK DWI Examples 3-Scan Trace vs 4-Scan Trace vs Increased Averages b-value Image Extrapolation & Inverted Grayscale Increase in SNR from 3 scan trace to 4 scan trace Time penalty isn t that severe in order to improve overall SNR 3 scan trace B 50(1avg), 400(2avg), 800 (4avg) 4 scan trace B 50(1avg), 400(2avg), 800 (4avg) 3 scan trace B 50(1avg), 400(2avg), 800 (7avg) Adding averages to the later b-value improves SNR to an already signal deficient image TA: 3:15 TA 4:34 4:15 Handout 4

5 DWI Examples b-value Image Extrapolation & Inverted Grayscale Example WB-DWI patient with multiple metastases along the spine from a primary prostatetumour The contrastin the computed image is improved compared to the acquired image as signalfrom tissues such as the kidneys, testes and salivary glands (red,green and blue arrows respectively)has been reduced. Typical Artifacts in Diffusion (1) Chemical Shift Artifacts in EPI Single shot: chemical shift differences have more time to evolve Low bandwidth along phase encode chemical shift of > 30 pixels kphase kread Courtesy Royal Marsden Hospital / UK acquired b = 900 s/mm 2 computed b = 1500 s/mm 2 Whole Body DWI Pediatric tumor staging Typical Artifacts in Diffusion (1) Chemical Shift Artifacts in EPI 4-years-old female patient with suspicion of Wilms tumor MIP images based on high b-value (800 s/mm 2 ) 5 steps composed Courtesy University of Homburg/Saar, Germany Solution: Fat suppression. Whole Body DWI Therapy Control of a patient with Hodgkin s Disease Typical Artifacts in Diffusion (2) N/2 ghosting in EPI Eddy currents and other imperfections cause phase differences between even and odd lines N/2 Shifted by N/2 from main image Phase correction not always perfect pre. and post 8 days after chemo cycle Before phase correction After phase correction Courtesy: Anwar Padhani, Mount Vernon Cancer Centre, UK Handout 5

6 Typical Artifacts in Diffusion (2) N/2 ghosting in EPI Typical Artifacts in Diffusion (5) Free Breathing vs. Navigator Possible reasons and solutions for N/2 EPI ghosts Mechanical resonance of scanner components Use appropriate echo spacing (T esp ), for example, avoid ms for Tim Trio ipat reconstruction artifacts Increase the number of reference lines (36-42), at the cost of recon time but not acquisition time b=50 T esp Note the improved sharpness in the sequence Navigator Navigator is longer but uses less averages to reduce scan time Navigator will produce more consistent image quality Free breathing is a great option for most routine exams Free Breathing TA: 3:43 Navigator Triggered TA: 5:00 Typical Artifacts in Diffusion (3) Long Echo Spacing and TE Typical Artifacts in Diffusion (6) ADC Thresholding Leads to longer TE More distortion More ghosting Decreased SNR TE 115 Echo Spacing 1.3 Noise Level = 10 TE 60 Echo Spacing 0.5 Possible solutions: Shorten echo spacing and TE as possible. Thresholding the ADC can lead to removal of anatomy Noise Level = 60 Typical Artifacts in Diffusion (4) Off-Center Patient Positioning Typical Artifacts in Diffusion (7) Image Resolution Interpolation ON Critical to position the patient in the center of the magnet to ensure consistent image quality!! Stair-step artifact, likely from low spatial resolution May need to switch ON interpolation when 128 matrix is used Handout 6

7 Basics of diffusion MRI and body diffusion Key Take Aways Three Body Diffusion Sequences Pros and Cons Basic diffusion concepts and principles b-value Diffusion weighted images (b-value images) ADC Very wide applications Liver, prostate, rectum, breast, etc Whole-body Therapy monitoring, tumor staging, etc Common artifacts in diffusion and important parameters for optimization Ghosting, failed fat suppression, etc Echo Spacing, TE, ipat, etc Sequence Applications Advantages Disadvantages Single-shot EPI Liver, Prostate, Whole Body, Prostate, Pelvis, Lung, Breast, Rectal CA, Pancreas, Kidney Most Commonly used diffusion in the field Can be used in all body applications More distortion than some techniques Prostate, Pelvis, Breast Reduced distortion Longer scan times Unable to use for free breathing applications ZOOMit Prostate, Rectal CA, Pancreas, Kidney, Breast Reduced distortion Zoomed Field of Views Only available on ptx systems (Parallel Transmit) Not useful for whole body diffusion Body Diffusion MRI: Basics and Beyond Outline Principle Basics of diffusion MRI and body diffusion Physics Pulse sequence Apparent diffusion coefficient (ADC) Common artifacts and parameter optimization Advanced topics New diffusion sequence techniques Motion correction / registration Advanced acceleration technique Readout-segmented, multi-shot diffusionweighted EPI High-quality, highresolution DWI and DTI Reduced susceptibility and blurring artefacts due to reduced TE and echo spacing Insensitivity to motioninduced phase errors Reduced SAR in comparison to TSE-based methods Phase-encode direction (ky) 1st shot 2nd shot 3rd shot 4th shot 5th shot Readout direction (kx) k-space trajectory Conventional single shot epi Single shot k-space trajectory Long TE Long Echo Spacing Porter DA & Heidermann RM (2009). High resolution diffusion weighted imaging using readout-segmented echo-planar imaging, parallel imaging and a twodimensional navigator-based reacquisition. Magn Reson Med, 62(2), Body Diffusion New Sequences in Diffusion Imaging Better delineation of tumor boundaries for prostate carcinoma RS-EPI DW images showed improved image quality compared to SS-EPI technique at 3T and is a feasible technique in the pelvis for producing highresolution DWI. Conventional b0 Conventional b800 Conventional ADC syngo syngo ZOOMit Thian YL. (2014). Readout-segmented echoplanar imaging for diffusion-weighted imaging in the pelvis at 3T A feasibility study. Acad Radiol, T2 TSE 21, T2 TSE b0 b800 ADC National University Hospital, Singapore *The statements by Siemens' customers described herein are based on results that were achieved in the customer's unique setting. Since there is no "typical hospital and many variables exist (e.g., hospital size, case mix, level of IT adoption) there can be no guarantee that other customers will achieve the same results. Handout 7

8 Excellent correspondence to anatomy for rectal carcinoma ZoomIt Principle Utilizes 2 orthogonal pulses for slice selection Conventional DWI No need for oversampling Reduced distortion T1 3D VIBE FatSat, post contrast b0 b1000 ADC map Utilized in Prostate, Pelvis, Kidney, Pancreas T2 TSE fused with b1000 b0 b1000 ADC map Only available currently on ptx systems (Skyra, Prisma) National University Hospital, Singapore are.si emens.com/magnetic-res onanc e-imaging/options-and-upgrades/clinic al-applicati ons /s yngo-z oomit/features Superior delineation of breast tumor boundaries ZoomIt in Prostate Less distortions, better tumor delineation Rs-EPI was superior to ss-epi [ ] for anatomical structure distinction, ghosting artifact and overall image quality [ ]. Rs-EPI was superior to ss-epi in SNR and CNR. Conventional DWI, b750 and ADC map, matrix 192, SL 4 mm, TA 2:47 min ZOOMit DWI, matrix 58 x 98, FoV 71 x 120, SL 3 mm, TA 4:21 b100 b400 b800 b1200 ADC, b750 and ADC map, matrix 192, SL 4 mm, TA 4:10 min Kim, HJ, Kim SH et al. (2014). Readoutsegmented echo-planar imaging in diffusion-weighted MR imaging in breast cancer: Comparison w ithsingle-shot echo-planar imaging in image quality. Korean J Radiol, 15(4), Seoul St. Mary`s Hospital, Seoul, South Korea Conventional DWI, matrix 96 x 128, FoV 190 x 190, SL 3 mm, TA 4:28 Kantonsspital Aarau, Aarau, Switzerland *The statements by Siemens' customers described herein are based on results that were achieved in the customer's unique setting. Since there is no "typical hospital and many variables exist (e.g., hospital size, case mix, level of IT adoption) there can be no guarantee that other customers will achieve the same results. Syngo ZoomIt New Sequence ZoomIt in Kidney Higher resolution, better differentiation of cortex & medulla syngo ZOOMit Shape your image Conventional DWI 2.1 mm x 2.1 mm ZOOMit DWI 1.4 mm x 1.4 mm Handout 8

9 ZoomIt in Rectum Better delineation of rectal carcinoma with less distortions Emerging New Improvements in Body Diffusion (1) Motion-compensation* b=50, 6 acqs b=400, 9 acqs b=800, 15 acqs Original Conventional DWI b=50, ADC 3D T2 SPACE - morphology ZOOMit Moco University Hospital IKRN, Mannheim, Germany Emerging New Improvements in Body Diffusion (1) Motion-compensation* Emerging New Improvements in Body Diffusion (1) Motion-compensation* A series of correction/refinement algorithms can be applied on b-value images Inplane-registration of images Filtering of images Denoising of images Rescaling of images to compensate signal loss due to motion Original Moco Improvements Correction for the left lobe signal loss Correction for the misregistration of diffusion weighted images for ADC calculation Original Moco ADC~ 610 ADC~ 90 Emerging New Improvements in Body Diffusion (1) Motion-compensation* b=50, 6 acqs b=400, 9 acqs b=800, 15 acqs Emerging New Improvements in Body Diffusion (2) Simultaneous Multi-Slice* Simultaneous excitation of multiple slices with blipped CAIPIRINHA 1 Original Moco 1 Setsompop, K. (2012). Blipped-controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging w ith reduced f-factor penalty. Magn Reson Med, 67, Handout 9

10 Emerging New Improvements in Body Diffusion (2) About 40% time reduction without compromise 2 b0 b1000 ADC Conventional TA 2:20 SMS 2 TA 1:21 1 Thank you for your attention! (42% reduction in scan time) NYU School of Medicine, MAGNETOM Skyra, Head/Neck 20 1 Note that scan time reduction may not be an exact factor of slice acceleration as scan times depends on the TR value specified and also because of a fast reference scan required for slice separation. 2 Young, MG, Shepherd TM et al. Multiband sequence reduces scan time for diffusion MRI and tractography in clinical patients. RSNA, Siemens GmbH 2016 All rights reserved. Page 58 Answers for life. Magnetic Resonance Advanced topics Key Take Aways Understand three diffusion sequences and when to utilize each pulse sequence Single-shot EPI ZoomIt Emerging new improvements Motion correction / registration Advanced acceleration techniques: Simultaneous multi-slice acquisition (Multiband) Body Diffusion MRI: Basics and Beyond Acknowledgement Peter Kollasch, PhD Marcel Dominik Nickel, PhD Elisabeth Weiland, PhD Vibhas Deshpande, PhD Brian Dale, PhD Handout 10

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street

Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street Lab Location: MRI, B2, Cardinal Carter Wing, St. Michael s Hospital, 30 Bond Street MRI is located in the sub basement of CC wing. From Queen or Victoria, follow the baby blue arrows and ride the CC south

More information

COBRE Scan Information

COBRE Scan Information COBRE Scan Information Below is more information on the directory structure for the COBRE imaging data. Also below are the imaging parameters for each series. Directory structure: var/www/html/dropbox/1139_anonymized/human:

More information

M R I Physics Course

M R I Physics Course M R I Physics Course Multichannel Technology & Parallel Imaging Nathan Yanasak, Ph.D. Jerry Allison Ph.D. Tom Lavin, B.S. Department of Radiology Medical College of Georgia References: 1) The Physics of

More information

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE

Module 4. K-Space Symmetry. Review. K-Space Review. K-Space Symmetry. Partial or Fractional Echo. Half or Partial Fourier HASTE MRES 7005 - Fast Imaging Techniques Module 4 K-Space Symmetry Review K-Space Review K-Space Symmetry Partial or Fractional Echo Half or Partial Fourier HASTE Conditions for successful reconstruction Interpolation

More information

Fast Imaging UCLA. Class Business. Class Business. Daniel B. Ennis, Ph.D. Magnetic Resonance Research Labs. Tuesday (3/7) from 6-9pm HW #1 HW #2

Fast Imaging UCLA. Class Business. Class Business. Daniel B. Ennis, Ph.D. Magnetic Resonance Research Labs. Tuesday (3/7) from 6-9pm HW #1 HW #2 Fast Imaging Daniel B. Ennis, Ph.D. Magnetic Resonance Research Labs Class Business Tuesday (3/7) from 6-9pm 6:00-7:30pm Groups Avanto Sara Said, Yara Azar, April Pan Skyra Timothy Marcum, Diana Lopez,

More information

Diffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford

Diffusion MRI Acquisition. Karla Miller FMRIB Centre, University of Oxford Diffusion MRI Acquisition Karla Miller FMRIB Centre, University of Oxford karla@fmrib.ox.ac.uk Diffusion Imaging How is diffusion weighting achieved? How is the image acquired? What are the limitations,

More information

Role of Parallel Imaging in High Field Functional MRI

Role of Parallel Imaging in High Field Functional MRI Role of Parallel Imaging in High Field Functional MRI Douglas C. Noll & Bradley P. Sutton Department of Biomedical Engineering, University of Michigan Supported by NIH Grant DA15410 & The Whitaker Foundation

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

MRI Physics II: Gradients, Imaging

MRI Physics II: Gradients, Imaging MRI Physics II: Gradients, Imaging Douglas C., Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor Magnetic Fields in MRI B 0 The main magnetic field. Always on (0.5-7 T) Magnetizes

More information

Head motion in diffusion MRI

Head motion in diffusion MRI Head motion in diffusion MRI Anastasia Yendiki HMS/MGH/MIT Athinoula A. Martinos Center for Biomedical Imaging 11/06/13 Head motion in diffusion MRI 0/33 Diffusion contrast Basic principle of diffusion

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

SIEMENS MAGNETOM Skyra syngo MR D13

SIEMENS MAGNETOM Skyra syngo MR D13 Page 1 of 8 SIEMENS MAGNETOM Skyra syngo MR D13 \\USER\CIND\StudyProtocols\PTSA\*dm_ep2d_mono70_b0_p2_iso2.0 TA:1:05 PAT:2 Voxel size:2.0 2.0 2.0 mm Rel. SNR:1.00 :epse Properties Routine Prio Recon Load

More information

SIEMENS MAGNETOM TrioTim syngo MR B17

SIEMENS MAGNETOM TrioTim syngo MR B17 \\USER\KNARRGROUP\MultiBand\LavretskyMultiBand\trufi localizer 3-plane TA: 5.1 s PAT: Voxel size: 1.2 1.2 5. Rel. SNR: 1.00 SIEMENS: trufi Load to stamp Slice group 1 Slices 1 Dist. factor 20 % Phase enc.

More information

MRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health

MRI image formation 8/3/2016. Disclosure. Outlines. Chen Lin, PhD DABR 3. Indiana University School of Medicine and Indiana University Health MRI image formation Indiana University School of Medicine and Indiana University Health Disclosure No conflict of interest for this presentation 2 Outlines Data acquisition Spatial (Slice/Slab) selection

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

Motion Artifacts and Suppression in MRI At a Glance

Motion Artifacts and Suppression in MRI At a Glance Motion Artifacts and Suppression in MRI At a Glance Xiaodong Zhong, PhD MR R&D Collaborations Siemens Healthcare MRI Motion Artifacts and Suppression At a Glance Outline Background Physics Common Motion

More information

Fast Imaging Trajectories: Non-Cartesian Sampling (1)

Fast Imaging Trajectories: Non-Cartesian Sampling (1) Fast Imaging Trajectories: Non-Cartesian Sampling (1) M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.05.03 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business

More information

GE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals.

GE Healthcare CLINICAL GALLERY. Discovery * MR750w 3.0T. This brochure is intended for European healthcare professionals. GE Healthcare CLINICAL GALLERY Discovery * MR750w 3.0T This brochure is intended for European healthcare professionals. NEURO PROPELLER delivers high resolution, motion insensitive imaging in all planes.

More information

Siemens AG, Healthcare Sector. syngo MR D13 0. Supplement - Parameters and image text 0.

Siemens AG, Healthcare Sector. syngo MR D13 0. Supplement - Parameters and image text 0. Siemens AG, Healthcare Sector 0 0 n.a. English Cs2 syngo Neuro Operator MR-05014 630 05/2010 01 02 Informatik, Manual D11 Cape syngo MR D13 0.0 Supplement - Parameters and image text 0. syngo MR D13 0.

More information

Applications Guide for Interleaved

Applications Guide for Interleaved Applications Guide for Interleaved rephase/dephase MRAV Authors: Yongquan Ye, Ph.D. Dongmei Wu, MS. Tested MAGNETOM Systems : 7TZ, TRIO a Tim System, Verio MR B15A (N4_VB15A_LATEST_20070519) MR B17A (N4_VB17A_LATEST_20090307_P8)

More information

Module 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution.

Module 5: Dynamic Imaging and Phase Sharing. (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review. Improving Temporal Resolution. MRES 7005 - Fast Imaging Techniques Module 5: Dynamic Imaging and Phase Sharing (true-fisp, TRICKS, CAPR, DISTAL, DISCO, HYPR) Review Improving Temporal Resolution True-FISP (I) True-FISP (II) Keyhole

More information

Parallel Imaging. Marcin.

Parallel Imaging. Marcin. Parallel Imaging Marcin m.jankiewicz@gmail.com Parallel Imaging initial thoughts Over the last 15 years, great progress in the development of pmri methods has taken place, thereby producing a multitude

More information

Initial Experience of Applying TWIST Dixon with Flexible View Sharing in Breast DCE-MRI

Initial Experience of Applying TWIST Dixon with Flexible View Sharing in Breast DCE-MRI Initial Experience of Applying TWIST Dixon with Flexible View Sharing in Breast DCE-MRI Yuan Le PhD 1, Hal D. Kipfer MD 1, Dominik M. Nickel PhD 2, Randall Kroeker PhD 2, Brian Dale PhD 2, Stephanie P.

More information

Use of the temporal median and trimmed mean mitigates effects of respiratory motion in multipleacquisition abdominal diffusion imaging

Use of the temporal median and trimmed mean mitigates effects of respiratory motion in multipleacquisition abdominal diffusion imaging Physics in Medicine & Biology NOTE OPEN ACCESS Use of the temporal median and trimmed mean mitigates effects of respiratory motion in multipleacquisition abdominal diffusion imaging To cite this article:

More information

Advanced MRI Techniques (and Applications)

Advanced MRI Techniques (and Applications) Advanced MRI Techniques (and Applications) Jeffry R. Alger, PhD Department of Neurology Ahmanson-Lovelace Brain Mapping Center Brain Research Institute Jonsson Comprehensive Cancer Center University of

More information

Parallel Magnetic Resonance Imaging (pmri): How Does it Work, and What is it Good For?

Parallel Magnetic Resonance Imaging (pmri): How Does it Work, and What is it Good For? Parallel Magnetic Resonance Imaging (pmri): How Does it Work, and What is it Good For? Nathan Yanasak, Ph.D. Chair, AAPM TG118 Department of Radiology Georgia Regents University Overview Phased-array coils

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

SIEMENS MAGNETOM Avanto syngo MR B15

SIEMENS MAGNETOM Avanto syngo MR B15 \\USER\INVESTIGATORS\Ravi\ADNI-Subject\Localizer TA: 0:10 PAT: Voxel size: 1.9 1.5 8.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer Inline movie

More information

Breast MRI Accreditation Program Clinical Image Quality Guide

Breast MRI Accreditation Program Clinical Image Quality Guide Breast MRI Accreditation Program Clinical Image Quality Guide Introduction This document provides guidance on breast MRI clinical image quality and describes the criteria used by the ACR Breast MRI Accreditation

More information

What is pmri? Overview. The Need for Speed: A Technical and Clinical Primer for Parallel MR Imaging 8/1/2011

What is pmri? Overview. The Need for Speed: A Technical and Clinical Primer for Parallel MR Imaging 8/1/2011 The Need for Speed: A Technical and Clinical Primer for Parallel MR Imaging Nathan Yanasak, Ph.D. Chair, AAPM TG118 Assistant Professor Department of Radiology Director, Core Imaging Facility for Small

More information

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis

Clinical Importance. Aortic Stenosis. Aortic Regurgitation. Ultrasound vs. MRI. Carotid Artery Stenosis Clinical Importance Rapid cardiovascular flow quantitation using sliceselective Fourier velocity encoding with spiral readouts Valve disease affects 10% of patients with heart disease in the U.S. Most

More information

New Technology Allows Multiple Image Contrasts in a Single Scan

New Technology Allows Multiple Image Contrasts in a Single Scan These images were acquired with an investigational device. PD T2 T2 FLAIR T1 MAP T1 FLAIR PSIR T1 New Technology Allows Multiple Image Contrasts in a Single Scan MR exams can be time consuming. A typical

More information

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging

Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging 1 CS 9 Final Project Classification of Subject Motion for Improved Reconstruction of Dynamic Magnetic Resonance Imaging Feiyu Chen Department of Electrical Engineering ABSTRACT Subject motion is a significant

More information

Automatic Gradient Preemphasis Adjustment: A 15-Minute Journey to Improved Diffusion-Weighted Echo-Planar Imaging

Automatic Gradient Preemphasis Adjustment: A 15-Minute Journey to Improved Diffusion-Weighted Echo-Planar Imaging Automatic Gradient Preemphasis Adjustment: A 15-Minute Journey to Improved Diffusion-Weighted Echo-Planar Imaging Vincent J. Schmithorst* and Bernard J. Dardzinski Magnetic Resonance in Medicine 47:208

More information

Scan Acceleration with Rapid Gradient-Echo

Scan Acceleration with Rapid Gradient-Echo Scan Acceleration with Rapid Gradient-Echo Hsiao-Wen Chung ( 鍾孝文 ), Ph.D., Professor Dept. Electrical Engineering, National Taiwan Univ. Dept. Radiology, Tri-Service General Hospital 1 of 214 The Need

More information

(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions

(a Scrhon5 R2iwd b. P)jc%z 5. ivcr3. 1. I. ZOms Xn,s. 1E IDrAS boms. EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 8 Solutions EE225E/BIOE265 Spring 2013 Principles of MRI Miki Lustig Assignment 8 Solutions 1. Nishimura 7.1 P)jc%z 5 ivcr3. 1. I Due Wednesday April 10th, 2013 (a Scrhon5 R2iwd b 0 ZOms Xn,s r cx > qs 4-4 8ni6 4

More information

MRI Image Quality Assessment

MRI Image Quality Assessment in partnership with MRI Image Quality Assessment David Collins CR-UK Cancer Imaging Centre, The Institute of Cancer Research Making the discoveries that defeat cancer Overview Current Practice Quality

More information

K-Space Trajectories and Spiral Scan

K-Space Trajectories and Spiral Scan K-Space and Spiral Scan Presented by: Novena Rangwala nrangw2@uic.edu 1 Outline K-space Gridding Reconstruction Features of Spiral Sampling Pulse Sequences Mathematical Basis of Spiral Scanning Variations

More information

Super-Resolution Reconstruction of Diffusion-Weighted Images from Distortion Compensated Orthogonal Anisotropic Acquisitions.

Super-Resolution Reconstruction of Diffusion-Weighted Images from Distortion Compensated Orthogonal Anisotropic Acquisitions. Super-Resolution Reconstruction of Diffusion-Weighted Images from Distortion Compensated Orthogonal Anisotropic Acquisitions. Benoit Scherrer Ali Gholipour Simon K. Warfield Children s Hospital Boston,

More information

MRI. When to use What sequences. Outline 2012/09/19. Sequence: Definition. Basic Principles: Step 2. Basic Principles: Step 1. Govind Chavhan, MD

MRI. When to use What sequences. Outline 2012/09/19. Sequence: Definition. Basic Principles: Step 2. Basic Principles: Step 1. Govind Chavhan, MD MRI When to use What sequences Govind Chavhan, MD Assistant Professor and Staff Radiologist The Hospital For Sick Children, Toronto Planning Acquisition Post processing Interpretation Patient history and

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

Sources of Distortion in Functional MRI Data

Sources of Distortion in Functional MRI Data Human Brain Mapping 8:80 85(1999) Sources of Distortion in Functional MRI Data Peter Jezzard* and Stuart Clare FMRIB Centre, Department of Clinical Neurology, University of Oxford, Oxford, UK Abstract:

More information

High Fidelity Brain Connectivity Imaging

High Fidelity Brain Connectivity Imaging CNI Inauguration Workshop Stanford, March 22 nd, 2012 High Fidelity Brain Connectivity Imaging -Recent Progress on Diffusion Weighted MRI for High Resolution and Low Distortion Allen W. Song, PhD Brain

More information

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology

Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Abbie M. Diak, PhD Loyola University Medical Center Dept. of Radiation Oncology Outline High Spectral and Spatial Resolution MR Imaging (HiSS) What it is How to do it Ways to use it HiSS for Radiation

More information

8/11/2009. Common Areas of Motion Problem. Motion Compensation Techniques and Applications. Type of Motion. What s your problem

8/11/2009. Common Areas of Motion Problem. Motion Compensation Techniques and Applications. Type of Motion. What s your problem Common Areas of Motion Problem Motion Compensation Techniques and Applications Abdominal and cardiac imaging. Uncooperative patient, such as pediatric. Dynamic imaging and time series. Chen Lin, PhD Indiana

More information

Fiber Selection from Diffusion Tensor Data based on Boolean Operators

Fiber Selection from Diffusion Tensor Data based on Boolean Operators Fiber Selection from Diffusion Tensor Data based on Boolean Operators D. Merhof 1, G. Greiner 2, M. Buchfelder 3, C. Nimsky 4 1 Visual Computing, University of Konstanz, Konstanz, Germany 2 Computer Graphics

More information

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space

White Pixel Artifact. Caused by a noise spike during acquisition Spike in K-space <--> sinusoid in image space White Pixel Artifact Caused by a noise spike during acquisition Spike in K-space sinusoid in image space Susceptibility Artifacts Off-resonance artifacts caused by adjacent regions with different

More information

Dynamic Contrast enhanced MRA

Dynamic Contrast enhanced MRA Dynamic Contrast enhanced MRA Speaker: Yung-Chieh Chang Date : 106.07.22 Department of Radiology, Taichung Veterans General Hospital, Taichung, Taiwan 1 Outline Basic and advanced principles of Diffusion

More information

Controlled Aliasing in Parallel Imaging Results in Higher Acceleration (CAIPIRINHA)

Controlled Aliasing in Parallel Imaging Results in Higher Acceleration (CAIPIRINHA) www.siemens.com/magnetom-world Controlled Aliasing in Parallel Imaging Results in Higher Acceleration (CAIPIRINHA) Felix Breuer; Martin Blaimer; Mark Griswold; Peter Jakob Answers for life. Controlled

More information

SIEMENS MAGNETOM Avanto syngo MR B15

SIEMENS MAGNETOM Avanto syngo MR B15 \\USER\INVESTIGATORS\Ravi\ADNI-phantom\QC Phantom-Localizer TA: 0:10 PAT: Voxel size: 1.9 1.5 8.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer

More information

Machine Learning for Medical Image Analysis. A. Criminisi

Machine Learning for Medical Image Analysis. A. Criminisi Machine Learning for Medical Image Analysis A. Criminisi Overview Introduction to machine learning Decision forests Applications in medical image analysis Anatomy localization in CT Scans Spine Detection

More information

Orthopedic MRI Protocols. Philips Panorama HFO

Orthopedic MRI Protocols. Philips Panorama HFO Orthopedic MRI Protocols Philips Panorama HFO 1 2 Prepared in collaboration with Dr. John F. Feller, Medical Director of Desert Medical Imaging, Palm Springs, CA. Desert Medical Imaging will provide the

More information

Release Notes. Multi-Band EPI C2P. Release HCP_v1 10 February 2014

Release Notes. Multi-Band EPI C2P. Release HCP_v1 10 February 2014 Release Notes Multi-Band EPI C2P Release HCP_v1 10 February 2014 Installation 1. Restart the system (reboot host and MRIR) 2. Extract the.zip file to a temporary directory 3. Run the installer.bat file

More information

Enhao Gong, PhD Candidate, Electrical Engineering, Stanford University Dr. John Pauly, Professor in Electrical Engineering, Stanford University Dr.

Enhao Gong, PhD Candidate, Electrical Engineering, Stanford University Dr. John Pauly, Professor in Electrical Engineering, Stanford University Dr. Enhao Gong, PhD Candidate, Electrical Engineering, Stanford University Dr. John Pauly, Professor in Electrical Engineering, Stanford University Dr. Greg Zaharchuk, Associate Professor in Radiology, Stanford

More information

A Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data

A Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data A Novel Iterative Thresholding Algorithm for Compressed Sensing Reconstruction of Quantitative MRI Parameters from Insufficient Data Alexey Samsonov, Julia Velikina Departments of Radiology and Medical

More information

MRI Hot Topics Motion Correction for MR Imaging. medical

MRI Hot Topics Motion Correction for MR Imaging. medical MRI Hot Topics Motion Correction for MR Imaging s medical Motion Correction for MR Imaging Kyle A. Salem, PhD Motion Artifacts in a Clinical Setting Patient motion is probably the most common cause of

More information

MR Advance Techniques. Vascular Imaging. Class III

MR Advance Techniques. Vascular Imaging. Class III MR Advance Techniques Vascular Imaging Class III 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology

More information

Use of MRI in Radiotherapy: Technical Consideration

Use of MRI in Radiotherapy: Technical Consideration Use of MRI in Radiotherapy: Technical Consideration Yanle Hu, PhD Department of Radiation Oncology, Mayo Clinic Arizona 04/07/2018 2015 MFMER slide-1 Conflict of Interest: None 2015 MFMER slide-2 Objectives

More information

G Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine. Compressed Sensing

G Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine. Compressed Sensing G16.4428 Practical Magnetic Resonance Imaging II Sackler Institute of Biomedical Sciences New York University School of Medicine Compressed Sensing Ricardo Otazo, PhD ricardo.otazo@nyumc.org Compressed

More information

Acquisition Methods for fmri at 7 Tesla

Acquisition Methods for fmri at 7 Tesla Acquisition Methods for fmri at 7 Tesla Peter J. Koopmans, Nuffield department for clinical neurosciences, University of Oxford, UK fmri at 7 Tesla is mostly used in the context of high spatial resolution

More information

Comprehensive Approach for Correction of Motion and Distortion in Diffusion-Weighted MRI

Comprehensive Approach for Correction of Motion and Distortion in Diffusion-Weighted MRI Magnetic Resonance in Medicine 51:103 114 (2004) Comprehensive Approach for Correction of Motion and Distortion in Diffusion-Weighted MRI G.K. Rohde, 1,3 * A.S. Barnett, 2 P.J. Basser, 1 S. Marenco, 2

More information

Improvements in Diffusion Weighted Imaging Through a Composite Body and Insert Gradient Coil System

Improvements in Diffusion Weighted Imaging Through a Composite Body and Insert Gradient Coil System Brigham Young University BYU ScholarsArchive All Theses and Dissertations 2013-07-10 Improvements in Diffusion Weighted Imaging Through a Composite Body and Insert Gradient Coil System Peter Austin Jepsen

More information

An Automatic Algorithm for Quality Assurance of MRI Scanners using a DWI Phantom

An Automatic Algorithm for Quality Assurance of MRI Scanners using a DWI Phantom An Automatic Algorithm for Quality Assurance of MRI Scanners using a DWI Phantom Aalborg University Biomedical Engineering and informatics School of Medicine and Health Master s Thesis By Andreas Ormstrup

More information

Information about presenter

Information about presenter Information about presenter 2013-now Engineer R&D ithera Medical GmbH 2011-2013 M.Sc. in Biomedical Computing (TU München) Thesis title: A General Reconstruction Framework for Constrained Optimisation

More information

Dynamic Autocalibrated Parallel Imaging Using Temporal GRAPPA (TGRAPPA)

Dynamic Autocalibrated Parallel Imaging Using Temporal GRAPPA (TGRAPPA) Magnetic Resonance in Medicine 53:981 985 (2005) Dynamic Autocalibrated Parallel Imaging Using Temporal GRAPPA (TGRAPPA) Felix A. Breuer, 1 * Peter Kellman, 2 Mark A. Griswold, 1 and Peter M. Jakob 1 Current

More information

Basic fmri Design and Analysis. Preprocessing

Basic fmri Design and Analysis. Preprocessing Basic fmri Design and Analysis Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial filtering

More information

Single Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction

Single Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction Single Breath-hold Abdominal T 1 Mapping using 3-D Cartesian Sampling and Spatiotemporally Constrained Reconstruction Felix Lugauer 1,3, Jens Wetzl 1, Christoph Forman 2, Manuel Schneider 1, Berthold Kiefer

More information

Lucy Phantom MR Grid Evaluation

Lucy Phantom MR Grid Evaluation Lucy Phantom MR Grid Evaluation Anil Sethi, PhD Loyola University Medical Center, Maywood, IL 60153 November 2015 I. Introduction: The MR distortion grid, used as an insert with Lucy 3D QA phantom, is

More information

Exam 8N080 - Introduction MRI

Exam 8N080 - Introduction MRI Exam 8N080 - Introduction MRI Friday January 23 rd 2015, 13.30-16.30h For this exam you may use an ordinary calculator (not a graphical one). In total there are 6 assignments and a total of 65 points can

More information

3-D Gradient Shimming

3-D Gradient Shimming 3-D Gradient Shimming 3-D Gradient Shimming on imaging and micro-imaging systems Technical Overview Introduction Advantage statement The GE3DSHIM method is a 3-D gradient shimming procedure developed for

More information

syngo.mr Neuro 3D: Your All-In-One Post Processing, Visualization and Reporting Engine for BOLD Functional and Diffusion Tensor MR Imaging Datasets

syngo.mr Neuro 3D: Your All-In-One Post Processing, Visualization and Reporting Engine for BOLD Functional and Diffusion Tensor MR Imaging Datasets syngo.mr Neuro 3D: Your All-In-One Post Processing, Visualization and Reporting Engine for BOLD Functional and Diffusion Tensor MR Imaging Datasets Julien Gervais; Lisa Chuah Siemens Healthcare, Magnetic

More information

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005

Field Maps. 1 Field Map Acquisition. John Pauly. October 5, 2005 Field Maps John Pauly October 5, 25 The acquisition and reconstruction of frequency, or field, maps is important for both the acquisition of MRI data, and for its reconstruction. Many of the imaging methods

More information

3D imaging with SPACE vs 2D TSE in MR guided prostate biopsy

3D imaging with SPACE vs 2D TSE in MR guided prostate biopsy 3D imaging with SPACE vs 2D TSE in MR guided prostate biopsy Poster No.: C-2004 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper M. Garmer, S. Mateiescu, M. Busch, D. Groenemeyer; Bochum/

More information

MRI Imaging Options. Frank R. Korosec, Ph.D. Departments of Radiology and Medical Physics University of Wisconsin Madison

MRI Imaging Options. Frank R. Korosec, Ph.D. Departments of Radiology and Medical Physics University of Wisconsin Madison MRI Imaging Options Frank R. Korosec, Ph.D. Departments of Radiolog and Medical Phsics Universit of Wisconsin Madison f.korosec@hosp.wisc.edu As MR imaging becomes more developed, more imaging options

More information

2D spatially selective excitation pulse design and the artifact evaluation

2D spatially selective excitation pulse design and the artifact evaluation EE 591 Project 2D spatially selective excitation pulse design and the artifact evaluation 12/08/2004 Zungho Zun Two-dimensional spatially selective excitation is used to excite a volume such as pencil

More information

design as a constrained maximization problem. In principle, CODE seeks to maximize the b-value, defined as, where

design as a constrained maximization problem. In principle, CODE seeks to maximize the b-value, defined as, where Optimal design of motion-compensated diffusion gradient waveforms Óscar Peña-Nogales 1, Rodrigo de Luis-Garcia 1, Santiago Aja-Fernández 1,Yuxin Zhang 2,3, James H. Holmes 2,Diego Hernando 2,3 1 Laboratorio

More information

Compressed Sensing for Rapid MR Imaging

Compressed Sensing for Rapid MR Imaging Compressed Sensing for Rapid Imaging Michael Lustig1, Juan Santos1, David Donoho2 and John Pauly1 1 Electrical Engineering Department, Stanford University 2 Statistics Department, Stanford University rapid

More information

Functional analysis with DTI and diffusion-neurography of cranial nerves

Functional analysis with DTI and diffusion-neurography of cranial nerves Functional analysis with DTI and diffusion-neurography of cranial nerves Poster No.: C-1942 Congress: ECR 2013 Type: Educational Exhibit Authors: J. P. Martínez Barbero, T. Martín Noguerol, A. Luna Alcalá;

More information

Super-resolution Reconstruction of Fetal Brain MRI

Super-resolution Reconstruction of Fetal Brain MRI Super-resolution Reconstruction of Fetal Brain MRI Ali Gholipour and Simon K. Warfield Computational Radiology Laboratory Children s Hospital Boston, Harvard Medical School Worshop on Image Analysis for

More information

TOPICS 2/5/2006 8:17 PM. 2D Acquisition 3D Acquisition

TOPICS 2/5/2006 8:17 PM. 2D Acquisition 3D Acquisition TOPICS 2/5/2006 8:17 PM 2D Acquisition 3D Acquisition 2D Acquisition Involves two main steps : Slice Selection Slice selection is accomplished by spatially saturating (single or multi slice imaging) or

More information

SIEMENS MAGNETOM Verio syngo MR B15V

SIEMENS MAGNETOM Verio syngo MR B15V \\USER\ZAHID_RESEARCH\MS\No Name\3D SWI TA: 6:39 PAT: 2 Voxel size: 1.0 0.5 2.0 mm Rel. SNR: 1.00 SIEMENS: gre Properties Prio Recon Before measurement After measurement Load to viewer Inline movie Auto

More information

syngo MR E11 Operator Manual Ortho Answers for life.

syngo MR E11 Operator Manual Ortho Answers for life. www.siemens.com/healthcare syngo MR E11 Operator Manual Ortho Answers for life. syngo MR E11 Operator Manual Ortho Legend Indicates a hint Is used to provide information on how to avoid operating errors

More information

Pre-processing of ASL data T CT

Pre-processing of ASL data T CT Wed October 2, 2013 Image Processing Pre-processing: motion correction, denoising, outlier detection Alessandra Bertoldo Pre-processing of ASL data T CT C T C Single TI ASL T T T T C CCC average Pre-processing

More information

Magnetic Resonance Elastography (MRE) of Liver Disease

Magnetic Resonance Elastography (MRE) of Liver Disease Magnetic Resonance Elastography (MRE) of Liver Disease Authored by: Jennifer Dolan Fox, PhD VirtualScopics Inc. jennifer_fox@virtualscopics.com 1-585-249-6231 1. Overview of MRE Imaging MRE is a magnetic

More information

Correction for EPI Distortions Using Multi-Echo Gradient-Echo Imaging

Correction for EPI Distortions Using Multi-Echo Gradient-Echo Imaging Correction for EPI Distortions Using Multi-Echo Gradient-Echo Imaging Nan-kuei Chen and Alice M. Wyrwicz* Magnetic Resonance in Medicine 41:1206 1213 (1999) A novel and effective technique is described

More information

Sparse sampling in MRI: From basic theory to clinical application. R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology

Sparse sampling in MRI: From basic theory to clinical application. R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology Sparse sampling in MRI: From basic theory to clinical application R. Marc Lebel, PhD Department of Electrical Engineering Department of Radiology Objective Provide an intuitive overview of compressed sensing

More information

Blipped-Controlled Aliasing in Parallel Imaging for Simultaneous Multislice Echo Planar Imaging With Reduced g-factor Penalty

Blipped-Controlled Aliasing in Parallel Imaging for Simultaneous Multislice Echo Planar Imaging With Reduced g-factor Penalty IMAGING METHODOLOGY - Full Papers Magnetic Resonance in Medicine 67:1210 1224 (2012) Blipped-Controlled Aliasing in Parallel Imaging for Simultaneous Multislice Echo Planar Imaging With Reduced g-factor

More information

A Virtual MR Scanner for Education

A Virtual MR Scanner for Education A Virtual MR Scanner for Education Hackländer T, Schalla C, Trümper A, Mertens H, Hiltner J, Cramer BM Hospitals of the University Witten/Herdecke, Department of Radiology Wuppertal, Germany Purpose A

More information

Evaluations of k-space Trajectories for Fast MR Imaging for project of the course EE591, Fall 2004

Evaluations of k-space Trajectories for Fast MR Imaging for project of the course EE591, Fall 2004 Evaluations of k-space Trajectories for Fast MR Imaging for project of the course EE591, Fall 24 1 Alec Chi-Wah Wong Department of Electrical Engineering University of Southern California 374 McClintock

More information

XI Signal-to-Noise (SNR)

XI Signal-to-Noise (SNR) XI Signal-to-Noise (SNR) Lecture notes by Assaf Tal n(t) t. Noise. Characterizing Noise Noise is a random signal that gets added to all of our measurements. In D it looks like this: while in D

More information

Referenceless Interleaved Echo-Planar Imaging

Referenceless Interleaved Echo-Planar Imaging Referenceless Interleaved Echo-Planar Imaging Magnetic Resonance in Medicine 41:87 94 (1999) Scott B. Reeder, 1 Ergin Atalar, * Anthony Z. Faranesh, 1 and Elliot R. McVeigh 1, Interleaved echo-planar imaging

More information

Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data

Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data Respiratory Motion Compensation for Simultaneous PET/MR Based on Strongly Undersampled Radial MR Data Christopher M Rank 1, Thorsten Heußer 1, Andreas Wetscherek 1, and Marc Kachelrieß 1 1 German Cancer

More information

Removal of EPI Nyquist Ghost Artifacts With Two- Dimensional Phase Correction

Removal of EPI Nyquist Ghost Artifacts With Two- Dimensional Phase Correction Removal of EPI Nyquist Ghost Artifacts With Two- Dimensional Phase Correction Nan-kuei Chen 1,5 and Alice M. Wyrwicz 4 * Magnetic Resonance in Medicine 51:147 153 (004) Odd even echo inconsistencies result

More information

Philips MRI Protocol Dump Created on Comment Software Stream

Philips MRI Protocol Dump Created on Comment Software Stream Page 1 of 5 Philips MRI Protocol Dump Created on 2/17/2011 4:11:01 PM Comment Created by ExamCard_to_XML with inputs: "J:\ADNI GO - ADNI 2 Phantom5.ExamCard" on system (BU SCHOOL OF MEDICINE :: 192.168.71.10)

More information

Diffusion model fitting and tractography: A primer

Diffusion model fitting and tractography: A primer Diffusion model fitting and tractography: A primer Anastasia Yendiki HMS/MGH/MIT Athinoula A. Martinos Center for Biomedical Imaging 03/18/10 Why n how Diffusion model fitting and tractography 0/18 Why

More information

UNIVERSITY OF SOUTHAMPTON

UNIVERSITY OF SOUTHAMPTON UNIVERSITY OF SOUTHAMPTON PHYS2007W1 SEMESTER 2 EXAMINATION 2014-2015 MEDICAL PHYSICS Duration: 120 MINS (2 hours) This paper contains 10 questions. Answer all questions in Section A and only two questions

More information

Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm

Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm Improved Spatial Localization in 3D MRSI with a Sequence Combining PSF-Choice, EPSI and a Resolution Enhancement Algorithm L.P. Panych 1,3, B. Madore 1,3, W.S. Hoge 1,3, R.V. Mulkern 2,3 1 Brigham and

More information

GPUs Open New Avenues in Medical MRI

GPUs Open New Avenues in Medical MRI GPUs Open New Avenues in Medical MRI Chris A. Cocosco D. Gallichan, F. Testud, M. Zaitsev, and J. Hennig Dept. of Radiology, Medical Physics, UNIVERSITY MEDICAL CENTER FREIBURG 1 Our research group: Biomedical

More information

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI

CHAPTER 9: Magnetic Susceptibility Effects in High Field MRI Figure 1. In the brain, the gray matter has substantially more blood vessels and capillaries than white matter. The magnified image on the right displays the rich vasculature in gray matter forming porous,

More information

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing

SPM8 for Basic and Clinical Investigators. Preprocessing. fmri Preprocessing SPM8 for Basic and Clinical Investigators Preprocessing fmri Preprocessing Slice timing correction Geometric distortion correction Head motion correction Temporal filtering Intensity normalization Spatial

More information