Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications 1

Size: px
Start display at page:

Download "Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications 1"

Transcription

1 Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at Cynthia H. McCollough, PhD Shuai Leng, PhD Lifeng Yu, PhD Joel G. Fletcher, MD Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications 1 In x-ray computed tomography (CT), materials having different elemental compositions can be represented by identical pixel values on a CT image (ie, CT numbers), depending on the mass density of the material. Thus, the differentiation and classification of different tissue types and contrast agents can be extremely challenging. In dual-energy CT, an additional attenuation measurement is obtained with a second x-ray spectrum (ie, a second energy ), allowing the differentiation of multiple materials. Alternatively, this allows quantification of the mass density of two or three materials in a mixture with known elemental composition. Recent advances in the use of energy-resolving, photon-counting detectors for CT imaging suggest the ability to acquire data in multiple energy bins, which is expected to further improve the signal-tonoise ratio for material-specific imaging. In this review, the underlying motivation and physical principles of dualor multi-energy CT are reviewed and each of the current technical approaches is described. In addition, current and evolving clinical applications are introduced. Reviews and Commentary n Review q RSNA, From the Department of Radiology, Mayo Clinic, 200 First St SW, Rochester, MN Received November 12, 2014; revision requested January 2, 2015; revision received February 3; accepted February 19; final version accepted February 24. Address correspondence to C.H.M. ( mccollough.cynthia@mayo.edu). The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health. q RSNA, 2015 Radiology: Volume 276: Number 3 September 2015 n radiology.rsna.org 637

2 In computed tomographic (CT) imaging, materials having different elemental compositions can be represented by the same, or very similar, CT numbers, making the differentiation and classification of different types of tissues extremely challenging. A classic example is the difficulty in differentiating between calcified plaques and iodine-containing blood. Although these materials differ considerably in atomic number, depending on the respective Essentials nn Materials having different elemental compositions can be represented by the same CT numbers, making the differentiation and classification of different types of tissues extremely challenging. nn In dual-energy CT, attenuation measurements obtained at a second energy allow the decomposition of a mixture of two or three materials into its constituent materials. nn A number of technical approaches exist for acquiring dualenergy data, including sequential acquisition of two different scans, rapid tube potential switching, multilayer detectors, and dual x-ray sources. nn Energy-resolving, photon-counting detectors represent an emerging approach to acquiring more than two energy measurements; this may provide new applications such as K-edge energy subtraction techniques. nn A range of current and emerging clinical applications of dualenergy CT exist, including virtual monoenergetic imaging, automated bone removal in CT angiography, perfused blood volume imaging, virtual noncontrast material enhanced imaging, plaque removal, virtual noncalcium imaging, urinary stone characterization, imaging of crystalline arthropathies, and detection of silicone from breast implants. mass density or iodine concentration, calcified plaque or adjacent bone may appear identical to iodinated blood on a CT scan. In addition to the difficulty in differentiating and classifying tissue types, the accuracy with which material concentration can be measured is degraded by the presence of multiple tissue types. For example, when measuring the amount of iodine enhancement of a soft-tissue lesion, the measured mean CT number over the lesion reflects not only the enhancement due to iodine, but also the CT number of the underlying tissue. The reason for these difficulties in differentiating and quantifying different tissue types is that the measured CT number of a voxel is related to its linear attenuation coefficient m(e), which is not unique for any given material, but is a function of the material composition, the photon energies interacting with the material, and the mass density of the material. As can be seen in Figure 1, the same linear attenuation coefficient values can be measured for two different materials (eg, iodine and bone) at a given energy, depending on the mass density. In dual-energy CT, an additional attenuation measurement is obtained at a second energy, allowing the differentiation of the two materials (Fig 1). Assuming the use of monoenergetic x-rays, at approximately 100 kev the same linear attenuation coefficients can be measured for bone and iodine. Data acquired at approximately 50 kev would allow the differentiation of the two materials. Although medical x-ray tubes generate a polyenergetic spectrum, the general principle remains valid. Thus, dual-energy CT can be defined as the use of attenuation measurements acquired with different energy spectra, along with use of the known changes in attenuation between the two spectra, to differentiate and quantify material composition. It was initially explored and described by Godfrey Hounsfield, who stated in 1973, Two pictures are taken of the same slice, one at 100 kv and the other at 140 kv so that areas of high atomic numbers can be enhanced. Tests carried out to date have shown that iodine (z = 53) can be readily differentiated from calcium (z = 20) (1). Technical Approaches to Dual-Energy CT Dual-energy methods for CT were subsequently investigated by Alvarez and Macovski in 1976 (2,3). They demonstrated that even with polyenergetic x- ray spectra, one could still separate the measured attenuation coefficients into their contributions from the photoelectric effect and Compton scattering processes. Since this early work, a number of technical approaches have been developed for acquiring the dual-energy data set (Table). Two Temporally Sequential Scans Initial applications focused primarily on the characterization of lung, liver, and soft-tissue composition (2,3). Two temporally sequential scans were performed to acquire the data at each of the two tube potentials (Fig 2a). Because the data were not acquired simultaneously, patient motion occurring between the two scans caused severe degradation of the resultant images and material composition information. To minimize the time delay inherent in two consecutive scans of the entire anatomic volume of interest, a modified approach has been suggested, where one axial scan (ie, one tube rotation) is performed at each tube potential prior to table incrementation (Fig 2b). This reduces the interscan delay between the low- and high-energy images. The best temporal resolution can be obtained for this scenario with use of partial scan reconstruction techniques. In this Published online /radiol Radiology 2015; 276: Content code: Funding: This research was supported by the National Institutes of Health (grants DK083007, DK100227, and EB016966). Conflicts of interest are listed at the end of this article. 638 radiology.rsna.org n Radiology: Volume 276: Number 3 September 2015

3 Figure 1 Figure 1: Graph of linear attenuation coefficients for bone (assuming r = 1 g/cm 3 ), iodine (assuming r = 1 g/cm 3 ), and iodine with lower density (assuming r = 0.1 g/cm 3 ) as a function of energy (in kiloelectron volts). The plotted linear attenuation coefficients (in reciprocal centimeters) were generated by using the energy-dependent mass attenuation coefficients from the National Institute of Standards, which were multiplied by the assigned density values. The result is that the same linear attenuation value (m[e]) can be attained although the materials (iodine and bone) are different (arrow). Measuring attenuation at a second energy allows the two materials to be differentiated (arrowhead). approach, 180 plus the fan angle of projection data are acquired at each tube potential, with a small delay between acquisitions to allow for switching of the tube potential and table incrementation. This approach may be appropriate for relatively static organs or tissues. However, because the time delay between the two scans is still relatively long, the susceptibility to motion misregistration between the low- and high-energy data sets may limit the value of this approach for dual-energy imaging of vascular processes, or tissues and organs susceptible to motion. For scanners having relatively narrow z-axis coverage, the increase in total scan time may also be a limiting factor. Rapid Switching of X-Ray Tube Potential In the 1980s, a modified commercial CT system (Somatom DR; Siemens Healthcare, Erlangen, Germany) used fast tube potential switching to allow alternate projection measurements to be acquired at low and high tube potentials (Fig 3) (4,5). Clinical application of this technique focused primarily on bone densitometry measurements. However, the tube current could not be increased quickly enough for the low tube potential measurements to achieve comparable noise levels in both the low and high tube potential data sets. This difference in noise limited extension of the technique beyond bone densitometry applications. Switching the tube potential between consecutive views requires that the transition time from low to high tube potential be less than a millisecond. In addition, the transition must be as abrupt as possible to maximize the energy separation of the measured data, although difficulty in rapidly modulating the x-ray tube current may cause high noise levels in the lowenergy data or excessive dose from the high-energy projections. This problem can be addressed, however, by using asymmetric sampling for the low- and high-energy projections. In this manner, without rapidly changing the tube current, the needed increase in tube current-time product for the low-energy projections is obtained by using a longer sampling interval for the low-energy data. This approach has been successfully implemented (HD 750; GE Healthcare, Waukesha, Wis). However, because the same x- ray tube is used for both the low- and high-energy data set, it is technically difficult to optimize the spectral filtration for both the low- and high-energy images. Finally, very fast data sampling is needed to avoid decreasing the in-plane spatial resolution subsequent to allocating a fraction of the acquired samples to each energy s data set. The very short time interval between the low- and high-energy views (less than 1 msec) provides near-simultaneous data acquisition of the low- and highenergy data set. The one-view misregistration requires a correction to be applied to avoid streak artifacts; however, this allows dual-energy material decomposition algorithms to be implemented by using either projection data or reconstructed images. The ability to use projection data assists in reducing beam-hardening artifacts in calculated virtual monoenergetic images. Multilayer Detector A third mechanism for acquiring dual-energy CT projection data uses a single high tube potential beam and layered or sandwich scintillation detectors (Fig 4). The low-energy data are collected from the front or innermost detector layer and the high-energy data are collected from the back or outermost detector layer (Fig 5). This is analogous to the use of multilayer detectors for dual-energy radiography (6 8). To achieve comparable noise in the low- and high-energy images, different detector thicknesses are used. Preliminary clinical work using a prototype multilayer detector system and a 64 detector row CT scanner (Brilliance-64; Philips Medical Systems, Bothell, Wash) has been reported to be capable of successfully differentiating iodine from bone. An advantage to this approach is that the low- and high-energy data sets are acquired simultaneously, and the data from the inner and outer detector layers are Radiology: Volume 276: Number 3 September 2015 n radiology.rsna.org 639

4 Technical Approaches for Acquiring Dual-Energy Data Set Technology Strengths Weaknesses Temporally sequential scanning of the entire scan volume Temporally sequential scanning of a single axial rotation Rapid switching of the x-ray tube potential Multilayer detector Dual x-ray sources Photon-counting detectors Can be performed on any CT scanner (no special hardware required). Can be performed on any CT scanner (no special hardware required). Reduced interscan delay between the low- and high-energy images. Near-simultaneous data acquisition of the low- and high-energy data set. Allows dual-energy material-decomposition algorithms to be implemented by using either projection data or reconstructed images. Reduces beam-hardening artifacts in calculated virtual monoenergetic images. Simultaneous data acquisition of the low- and high-energy data set. All image data are acquired in a manner that supports material-specific imaging. Tube current and tube filtration can be optimized for each tube potential independently. Relatively low degree of spectral overlap, which improves contrast-to-noise ratios in material-specific images. Beam-hardening corrections are applied prior to image reconstruction, allowing material-specific images to be created in the image domain. Uses energy-specific measurements and energy thresholds to reject electronic noise. Facilitates new imaging approaches, such as k-edge subtraction. Any patient motion occurring between the two scans may cause severe degradation of the resultant images and material composition information. Best temporal resolution obtained with partial-scan reconstructions, which are more susceptible to errors in CT numbers; this might degrade material composition accuracy. The susceptibility to motion misregistration between the low- and high-energy single axial rotation may limit the value of this approach for imaging of vascular processes or tissues and organs susceptible to motion. For scanners having relatively narrow z-axis coverage, the increase in total scan time associated with an axial acquisition mode may be a limiting factor. Requires specialized hardware. Relatively high overlap of the energy spectra. Requires specialized hardware. Relatively high overlap of the energy spectra. Noise level may differ between low- and high-energy images. Requires specialized hardware. A 90 phase shift between low- and high-energy data. Simultaneous use of both x-ray sources allows scattered radiation whose original primary photon came from one tube to be detected by the detector of the other tube, requiring specialized scatter correction. Requires specialized hardware, which is not anticipated to be commercially available for some time, if at all. recorded at all times (9). This allows dual-energy analysis to be performed on every data set acquired. That is, the system is always operating in a dual-energy mode. Dual X-Ray Sources Dual-source CT is a CT system where two x-ray sources and two data acquisition systems are mounted on the same gantry, positioned orthogonally to one another (Fig 6) (10,11). The first dual-source scanner was introduced in 2006 (Somatom Definition DS); the second, in 2009 (Somatom Definition Flash); and a third, in 2013 (Somatom Force), all from Siemens Healthcare (Forchheim, Germany). Each x-ray source is equipped with its own high-voltage generator, allowing independent control of both the x-ray tube potential and the tube current. The projection data are approximately 90 out of phase. Images reconstructed from each of the tubedetector pairs are used to perform material decomposition analyses in the image domain. Because each tube is operated at different tube potentials and different tube current values, the noise levels in the respective images are adjustable. Because both tubes are simultaneously energized, scattered radiation whose original primary photon came from one tube may be detected by the detector for the other tube, and vice versa. This degrades spectral separation and requires the implementation of an appropriate scatter-correction algorithm (12). An advantage of the dual-source approach is the ability to independently optimize the spectral filtration for each tube-detector pair, which enables increased spectral 640 radiology.rsna.org n Radiology: Volume 276: Number 3 September 2015

5 Figure 2 Figure 3 Figure 2: Slow kilovoltage (kv) switching: (a) Consecutive scans of the entire scan volume are obtained by using back-to-back scans at low and high tube potentials. Scans can be acquired in axial or spiral modes. The interscan delay is relatively long (several seconds) and is composed of the time to scan the complete volume of interest, as well as the time to reposition the table to the beginning of the scan volume. (b) Consecutive scans of one anatomic section are obtained with decreased interscan delay by switching the tube potential between axial scans at each anatomic section level. The interscan delay is reduced to one rotation time plus the small delay to switch tube potential and increment the table. Figure 3: Fast kilovoltage (kv) switching: The x-ray tube potential is switched between successive views in either axial or spiral mode. Dual-energy processing can be performed by using projection or image data, and the temporal resolution of each image and the entire examination remains essentially unchanged. For successful technical implementation and to maintain current levels of image quality and temporal resolution, very fast detector materials and electronics are needed. Additionally, the x-ray generator must be capable of very rapid transitions between the low and high tube potentials. Figure 4 separation and an increased signalto-noise ratio in the material-specific images (13,14). Technical Approaches to Multi-Energy CT Photon-counting Detectors The use of energy-resolving photoncounting detectors may offer the most robust solution to dual-energy, or multi-energy, data acquisition. Such detectors, which are capable of counting discrete photon interactions, are the subject of ongoing research and development and are not commercially available in CT systems, although they are used in nuclear medicine and spectral mammography. Based on the choice of energy thresholds and the associated energy of each photon, counts are placed into specific energy threshold data sets (Fig 7) (15 17). Data associated with specific energy windows are created by subtracting different energy threshold data. For the special case with n = 2 energy windows, photon-counting detectors can be used for dual-energy imaging. By using pulse-height discrimination, the transmitted x-ray spectrum is divided into a number of different energy bins (Fig 8), the number of which depends on the design of the application-specific integrated circuit that is bonded to the energy-resolving detector. Cadmium telluride (CdTe) and cadmium zinc telluride are considered the most likely candidates for the material used to convert the energy of an absorbed x-ray into an electrical signal, the magnitude of which is proportional to the incident photon s energy (18). Although such detectors are currently available and used in other disciplines, challenges Figure 4: A single x-ray source and single tube potential value are used in combination with a duallayer scintillating detector. kv = kilovoltage. exist in dealing with the exposure rates used in CT imaging. At the high peak x-ray flux used in CT imaging (approximately 10 9 counts/sec/mm 2 ) (19), currently available detectors experience a loss of counts due to pulse pile-up effects and can eventually Radiology: Volume 276: Number 3 September 2015 n radiology.rsna.org 641

6 Figure 5 Figure 6 Figure 5: In a dual-layer detector, low-energy quanta are predominantly collected in the front layer. X-rays capable of passing through the front layer are predominantly of higher energy and are collected in the back detector layer. kev = kiloelectron volt. become fully paralyzed (20,21). The accuracy of the recorded energy can be degraded due to the distribution of a photon s energy across multiple detector pixels (known as charge sharing [22 24]) or the reemission of a characteristic x-ray (known as K-escape). To reduce the effects of charge sharing, a charge-summing mode has been proposed in which communication between nearby detector pixels is established with use of hardware circuits. Charges from the neighboring pixels of coincident events are summed up and the total charge is assigned to the pixel that collects the most charge. Substantial improvement of spectral performance has been observed using this technique (25,26). A number of potential advantages, including improved spectral separation and increased dose efficiency (27), are driving significant research and development in this field. In particular, photon-counting detectors have a higher geometric efficiency than do energy-integrating detectors (by approximately 30%), and the application of an energy threshold enables the rejection of counts due only to measured electronic noise. Electronic noise measured above the threshold will affect only the measured energy of each photon; it will not change photon counts. This detector technology also provides the ability to perform K-edge imaging. Material Decomposition Algorithms The ability of dual- or multi-energy CT to decompose a material into its constituent elements relies on the energy- and element-dependent nature of x-ray attenuation. In the diagnostic energy range (E, 150 kev), the attenuation of x-rays by matter is primarily due to the photoelectric effect and Compton scattering processes. This is a monotonic and smoothly varying function for elements without K- or L-edges within the energy range that is useful for diagnostic x-ray imaging (2). In the absence of a K- or L-edges, the attenuation coefficient of a given material can be modeled as a linear combination of the photoelectric and Compton interactions. By modeling the dependence of the photoelectric and Compton interaction processes on material mass density (r) and atomic number (Z), a rz map can be generated and material-specific information obtained (2). Since the x-ray attenuation depends primarily on these two interaction mechanisms, any material s attenuation coefficient can be expressed as a linear combination of the attenuation coefficients of two underlying basis materials (28). Early investigations into dualenergy CT material decomposition focused on methods implemented in prereconstruction (projection) space, Figure 6: Dual-x-ray-source geometry: Independent x-ray tubes, detectors, and generators allow simultaneous collection of dual-energy data. Each tube can be operated by using the optimal tube current setting and with optimum spectral filtration. Due to the 90 offset between the low- and high-energy views, dual-energy processing is implemented by using already reconstructed image data. Both axial and spiral acquisition modes are possible, and the temporal resolution remains unchanged. Scatter originating from one tube can be detected by the orthogonal detector, which can reduce spectral separation. The field of view over which dual energy data are acquired is currently limited to 26, 33 or 35 cm, depending on the specific scanner model. kv = kilovoltage. creating either rz image pairs (2) or basis material image pairs (5). The use of projection data has the theoretical advantage of eliminating beam-hardening artifacts in the reconstructed images. In practice, however, beam-hardening artifacts may not be completely removed due to imperfect system calibrations (29). This demonstrates that regardless of the algorithm domain (projection or image space), accurate system calibrations, which establish the relationship between the projection measurements and known densities of basis materials, or CT numbers and known thicknesses of basis materials, are essential. Material decomposition can also be performed in postreconstruction (image) space, by using either rz decomposition (30) or basis material decomposition (31). Image space decomposition is relatively easy to 642 radiology.rsna.org n Radiology: Volume 276: Number 3 September 2015

7 Figure 7 Figure 7: Schematic diagram of a photon-counting (energy-resolving) x-ray detector and signal-processing components. The semiconductor detector (eg, CdTe) directly converts the absorbed x-ray energy into electrical charge, which is accelerated across a potential difference and collected by discreet signal electrodes. Unlike current scintillating CT detectors, there are no septa within the semiconductor material; this maximizes geometric efficiency. The signal collected is proportional to the absorbed photon energy, and pulse height analysis is used to bin the signal from discreet photon interactions into two or more energy windows, depending on the capability of the application-specific integrated circuit coupled to the semiconductor detector. kev = kiloelectron-voltage. Figure 8 Figure 8: Illustration of the binning of detected x-rays into six energy windows (w1 w6). This illustration neglects nonideal properties such as pulse pile-up, which occurs when the count rate is too high, and charge sharing or K-escape phenomena, which occur when the discrete spacing of the signal electrodes is too small. To achieve both an accurate photon count and spectral measurement in CT, both hardware and software advances are necessary. kev = kiloelectron volt. implement and is used in systems where spatial or temporal inconsistency exists between the projection data acquired at the two beam energies, for example, in dual-source CT (10,11). Iterative beam hardening correction in image space can be used in this scenario to minimize the dependence of quantitative measurements on patient size or composition (32). The use of photo- counting detectors in CT may provide more spectral information than is available from dual-energy CT systems. The energydiscriminating capability of photoncounting detectors enables binning of the detected photons into N energy bins, where N depends on the number of energy thresholds allowed by the application-specific integrated circuit design. The choice of N is affected by the energy resolution since a better energy resolution allows more accurate spectral information to be recorded in each energy bin. Energy resolution on the order of several kiloelectron volts is currently typical. The additional number of energy measurements (ie, bins) provides more discrete information about the transmitted spectra and enables new approaches for material decomposition, such as K-edge imaging, to detect specific, or even multiple, contrast agents (15,33 37). The optimal configuration of energy bins, that is, the number of bins and their upper and lower thresholds, has a critical impact on the material decomposition of spectral imaging, and a number of reports have focused on optimization strategies (2,38 40). Also, material decomposition algorithms for photon-counting detectorbased CT have been investigated for both projection- and image-space approaches (35,36,41,42). With use of material decomposition techniques, one can obtain mass density, effective atomic number, or other material-specific information. Clinical applications that make use of these capabilities can be classified into two main categories, the first of which is to quantify the concentration of a certain component of a mixture. The mixture usually involves two or Radiology: Volume 276: Number 3 September 2015 n radiology.rsna.org 643

8 Figure 9 Figure 9: The CT numbers of three known materials in the low- and high-energy images can be plotted on the y- and x-axis, respectively. Unknown materials are then mapped onto this plot to determine the percent composition of each of the three basis materials. more materials with known elemental compositions, such as iodine, soft tissue, and fat. This could be applied, for example, to quantify contrast material uptake in the liver. The second type of clinical application is to classify materials into predefined groups, such as uric acid versus non uric acid urinary stones. The classification can be based on the effective atomic number information from the material decomposition or on densityindependent measurements, such as the ratio of the CT number measured in the low-energy image to that measured in the high-energy image. This ratio is referred to in various publications as either the dual-energy ratio, or DER, or the CT number ratio, or CTR (14,43). A three-material decomposition algorithm can also be used to quantify the concentration of three materials with known elemental compositions in a mixture (11,13,14). In the absence of the ability to perform K-edge imaging, the ability to quantify three materials from only two unique measurements relies on the introduction of a third piece of information, which is usually the assumption of volume or mass conservation (31). Conceptually, the CT numbers of three materials of known composition and density are plotted on a graph, where the y-axis is the CT number at low energy (eg, 80 kv) and the x-axis is the CT number at the high energy (eg, 140 kv) (Fig 9). Ideally, the three basis materials should be sufficiently different as to create a broad triangle in this plot. Thereafter, corresponding CT number pairs from the low- and high-energy images are mapped onto the calibration diagram. Depending on their position in the diagram, the material or percent composition of a certain material is estimated. The voxels can be color coded according to material type or the percent composition of the basis materials reported. Alternatively, one or more of the basis materials can be either suppressed or enhanced, depending on the desired clinical application. The noise level in the low- and high-energy images and the magnitude of difference in effective atomic number between basis materials each affect the accuracy of the material discrimination (13,14). Energy Domain Noise Reduction Whether in dual- or multi-energy CT, the number of photons in any given energy bin is less than the total number exiting the patient. This has important consequences for multi-energy CT, where the appropriate selection of energy bins (ie, the number of energy bins and the width of each energy bin) has a critical impact on the outcome of spectral imaging. A narrow energy bin has better energy resolution compared with a wider energy bin, and hence enables better material differentiation. For example, referring to the spectra in Figure 8, use of bins w2 and w5 would result in improved material discrimination than would use of wider bins (w2 + w3) and (w4 + w5). However, an obvious limitation of using narrow bins is that the number of photons available in each energy bin is much smaller than the total number of photons detected, and noise levels within each bin are dramatically increased. Thus, a critical problem occurs to maximize material separation, energy bins should be narrow and widely separated; however, for narrow energy bins, the noise may be so high as to make it impossible to detect small differences in material composition. Moreover, when only narrow energy bins are used, a large portion of the dose delivered to the patient in other energy bins is unutilized. Thus, the requirements for increasing the differences in material-specific attenuation are in direct conflict with the requirements for decreasing noise in the energy-specific data and in any material decomposition images derived from them. Similar observations exist for the selection of the total number of energy bins. For a given x-ray spectrum, more energy bins provide more measurements of energy-dependent information. With multiple data points available along the attenuation-energy curve, better curve fitting and consequently better material differentiation is achieved. However, more bins imply narrower bin widths and fewer photons in each bin. For example, for a system with six energy bins of equal width, the number of photons in each energy bin is only one-sixth of the total photons delivered (Fig 8). Noise in each energy bin is then substantially increased relative to the situation using fewer energy bins. Therefore, an intrinsic trade-off exists in the selection of energy bins (number, width, and placement) for multi-energy CT, resulting in the described trade-off between energyspecific signal (material differentiation information) and noise. To address this constraint, Leng et al (44) developed an energy-domain noise-reduction technique to reduce image noise in spectral CT. Simulations, phantom experiments, and patient studies demonstrated the ability to achieve noise levels in each energy-selective image similar to that of the composite (nonenergy-selective) image that makes use of all detected photons, both for clinical dual-energy CT systems and photon-counting detector-based spectral CT systems (44). This was achieved without any loss of spatial resolution. This approach allows flexible selection of energy bin width and number to maximize material decomposition 644 radiology.rsna.org n Radiology: Volume 276: Number 3 September 2015

9 Figure 10 Figure 11 Figure 11: Contrast-enhanced, dual-energy axial CT image in a 31-year-old man with a pulmonary embolism in the right lower lobe. Iodine signal is identified and color coded in red within the segmented lung. The iodine overlay image is superimposed on a gray-scale mixed image. The dark regions show a perfusion defect secondary to the embolism. Blue circle marks the diameter of the second tube on the dual-source CT scanner. Dual-energy data are acquired only within this circle. Figure 10: Coronal CT angiogram in a 62-year-old woman. With dual-energy techniques, bone anatomy can be automatically separated from the vascular anatomy and highly enhancing kidneys. in dual- and multi-energy CT (53,54). Finally, Gao et al have described a compressive sensing approach based on a prior rank, intensity and sparsity model, or PRISM (55). capabilities, without incurring increased image noise or requiring increased patient dose. Energy domain noise reduction has been successfully applied to the full range of dual- or multi-energy processing techniques, including the generation of blended images (linear or nonlinear combination of the low- and high-energy images), virtual monoenergetic images, or material-specific images (44 48). It can be applied before material decomposition on the individual energy-specific images and after material decomposition on the material-specific or virtual monoenergetic images. As noise reduction is applied in the energy domain, it offers additive benefit to spatial- or temporal-domain noise reduction techniques, such as image- or projection-based denoising or iterative reconstruction. Several other dual-energy noisereduction techniques have been previously investigated to reduce noise in material-specific images. Macovski et al proposed a method using measurement-dependent filtering, which essentially combines low-frequency information from the material-specific image and high-frequency information from the nonmaterial-specific image (49,50). Kalender et al (51) developed a method to reduce noise using the negative correlation in material-specific images from dual-energy CT. McCollough et al (52) developed a correlated noisereduction technique in dual-energy digital subtraction angiography. Compressed sensing-based methods, such as prior image constrained compressed sensing (PICCS) by Chen et al and nonconvex PICCS by Ramírez-Giraldo et al can also be used to reduce image noise Current and Emerging Clinical Applications of Dual- and Multi-Energy CT Although each of the major CT manufacturers is investigating dual-energy acquisition and reconstruction techniques, at present, a greater range of clinical applications and software tools has been developed for use with the dual-source system. In the remaining sections, examples from a dual-source CT system will be used to introduce several clinical applications of dual-energy CT. Patient examinations shown in this section were acquired under institutional review board approval. Virtual Monoenergetic Imaging In dual-energy CT, besides the materialspecific information, one may also synthesize monoenergetic images at different energies, which can be used for routine Radiology: Volume 276: Number 3 September 2015 n radiology.rsna.org 645

10 Figure 12 Figure 12: Axial contrast-enhanced, dual-energy scan in a 67-year-old man. (a, b) Mixed images show subcapsular hematoma of the liver ( ) and two similar-appearing low-attenuation liver lesions (arrow). (c, d) Iodine overlay images demonstrate iodine within one lesion (c), indicating it is a metastasis (arrow), while the other image (d) contains no iodine, indicating it is a hematoma (arrow) associated with prior wedge resection. Figure 13 Figure 13: Coronal maximum intensity projection images from CT angiographic study of the aortic bifurcation in a 70-year-old man. (a) Calcified plaque obscures the vessel lumen. (b) By using dual-energy material decomposition, the calcified plaque is identified and subtracted, giving a clear indication of vessel patency and areas of stenoses (arrows). diagnosis similar to conventional polyenergetic images acquired at a single x-ray tube potential (56 61). These images decrease beam-hardening artifacts, even when image-based methods are used. Because of the abrupt increase in iodine attenuation above 33 kev, iodine signal increases for photon energies that are just above 33 kev. The optimal energy at which to synthesize the virtual monoenergetic image depends on several factors, including patient size and the use of energy-domain noise reduction methods to address the increase in noise that would otherwise occur at lower energy settings. In practice, the optimal contrast-to-noise ratio for iodine occurs at virtual monoenergetic energies of kev. Automated Bone Removal in CT Angiography One of the most appealing applications of dual-energy CT is direct CT angiography (62 65). In this approach, the dual-energy algorithm identifies and removes bone, allowing direct visualization of iodinated vessels (Fig 10). On the most current dual-source CT system, the field of view of the second x-ray source limits dual-energy bone removal to a 35-cm field of view; singleenergy bone removal algorithms are used outside the 35-cm field of view on that system. As Hounsfield predicted in 1973, the differences in atomic number do indeed make calcium and iodine readily separable with use of dual-energy techniques. Perfused Blood Volume (Blood Pool Imaging) In addition to removing bone to see iodine, the identification of iodine voxels 646 radiology.rsna.org n Radiology: Volume 276: Number 3 September 2015

11 Figure 14 Figure 14: Dual-energy CT virtual noncalcium images acquired 11 days after injury demonstrate bone bruises in the lateral femoral condyle of a 29-year-old male soccer player. (a) Coronal and (d) axial views of the virtual noncalcium images demonstrate increased signal intensity due to edema subsequent to a knee injury (ie, bone bruises, arrow). (b, e) Corresponding T2-weighted MR images show high signal intensity in the same region (arrow). (c) Coronal and (f) axial views of the mixed images, where calcium signal has not been removed, show no evidence of bone injury. allows for color enhancement of iodinated areas. One clinical use is visualization of the perfused blood volume, also referred to as blood pool imaging. Applications have been developed to detect perfusion deficits in the lung secondary to a pulmonary embolism (66 70) (Fig 11) and to detect areas of ischemia in the myocardium (71 75). Virtual Noncontrast enhanced Images Another potential application for dual-energy CT is, after identifying iodine voxels, to remove the iodine component of the CT number to create a virtual noncontrast image, that is, an image without contrast material enhancement (76 82). If the virtual noncontrast images are of sufficient quality to be a viable substitute for true noncontrast images, the noncontrast scan of multiphase examinations may be able to be omitted. An example of a clinical application where this possibility is desirable is in the CT evaluation of the kidney, ureters, and bladder, where a complete examination includes both a noncontrast and a contrast-enhanced scan. The noncontrast scan is required only for stone detection, because the presence of iodinated contrast material can obscure visualization of urinary stones. However, at present, urinary stones smaller than approximately 2 mm are not routinely visible in the virtual noncontrast images (81,83). In examinations where only a contrast-enhanced scan was performed and a subtle increased attenuation is observed, it would be helpful to have a noncontrast examination to determine if the increased attenuation is due to contrast enhancement or simply due to the presence of a slightly more attenuating tissue type (77,78). The mixed image is not a materialspecific image; it includes signal from soft tissue, bone, and iodinated contrast media. The tissue-only virtual noncontrast image suppresses the Radiology: Volume 276: Number 3 September 2015 n radiology.rsna.org 647

12 iodine signal, and the iodine information is shown as a color-coded map, which is overlaid on top of the mixed or virtual noncontrast image. These iodine maps are helpful in assessing whether lesions represent a malignant or benign process (Fig 12). Conversely, iodine identification can be helpful when only a noncontrast-enhanced scan is performed and a subtle increase in attenuation is observed. For example, after thrombolytic therapy for stroke or after complicated intracranial neurologic procedures, iodine classification can be helpful in differentiating between blood, tumor, or retained iodinated contrast material (76,80). Atherosclerotic Plaque Removal An extension of automated bone removal is to remove not just large bone anatomy, but also discrete hard plaques, potentially allowing clearer visualization of patent lumens in maximum intensity projections (Fig 13) (84 86). Figure 15 Figure 15: Axial images in (a) 20-year-old man and (b) 32-year-old man. A commercially available software application can be used to discriminate between uric acid and non uric acid renal stones. Figure 16 Virtual Noncalcium Images Similar to the virtual noncontrast application, where iodine is identified and removed to more clearly appreciate the underlying soft tissues, the same can be done with calcium (87). This increases the ability to detect bone marrow edema, which would otherwise be difficult to appreciate in the presence of the much larger calcium signal (Fig 14). Urinary Stone Characterization In vitro and in vivo experiments have demonstrated the ability of dual-energy material decomposition to accurately discriminate between uric acid and non uric acid stones (43,88). Clinically, this is an important application since in vivo determination that a stone is made of uric acid would allow immediate initiation of urinary alkalization, potentially avoiding the need for further medical testing or stone removal procedures. The accuracy of this approach has been validated in phantom models and approaches 100% (88). Figure 15 demonstrates the results of one commercial software application, which color codes uric Figure 16: Axial CT images. By using custom software that takes advantage of the increased in-spectral separation provided by a tin filter on the high-energy beam of a dual-source CT system, uric acid stones (red) can be distinguished from three groups of non uric acid stones: cystine (yellow), calcium oxalate/ brushite/struvite (green), and apatite (blue). (a) Uric acid stone in a 71-year-old man. (b) Cystine stone in a 63-year-old man. (c) Calcium oxalate stone in a 21-year-old woman. (d) Apatite stone in a 34-year-old woman. 648 radiology.rsna.org n Radiology: Volume 276: Number 3 September 2015

13 acid stones in red and non uric acid stones in blue. The various types of non uric acid stones (cystine, hydroxyapatite, calcium oxalate, etc) contain higher atomic number elements and so distinguish themselves from the uric acid stones, which do not. By using the increased spectral separation that results from filtering the high-energy beam of a dual-source scanner with a high-z material, such as tin, differentiation between several non uric acid stone types is also possible (14,89 92) (Fig 16). Sequential scanning techniques have also been shown to successfully identify uric acid and non uric acid stone types after application of a nonrigid registration technique, which aligns identified stones in the low- and high-energy data sets that otherwise were misaligned by several millimeters. Use of such registration techniques is critical if there is any motion present; even a 1-mm offset can alter the results in a small (approximately 2-mm) stone (93). Figure 17 Crystalline Arthropathies The ability of dual-energy CT to differentiate uric acid from calcium can be used to differentiate gout from pseudogout (calcium pyrophosphate deposition disease). Examples from two patients are shown in Figure 17 (94). Detection of Silicone from Breast Implant Leaks When the integrity of a silicone breast implant is compromised, small amounts of silicone can leak into the surrounding tissue and may travel to the surrounding lymph nodes, causing inflammatory or immune responses. Occasionally, the silicone is adjacent to similarly attenuating anatomy and difficult to detect. Dual-energy CT can be used to clearly differentiate these regions of silicone from the surrounding soft tissue (Fig 18) (95). Conclusion In summary, dual-energy and multienergy CT represent an emerging field Figure 17: Three-dimensional volume-rendered CT images of (a) gout (green) in a 56-year-old man and (b) calcium pyrophosphate crystals (purple) in the triangular fibrocartilage (arrow) in a 78-year-old woman, where composition of intra- and peri-articular crystals was automatically determined by using dual-energy CT. of clinical CT imaging. The ability of dual-energy and multi-energy CT to differentiate materials of different effective atomic numbers makes possible several new and clinically relevant CT applications. In approximately , these were limited primarily to automated bone and plaque removal in CT angiography, removal of iodine signal from contrast-enhanced scans, single-phase imaging of perfused blood volume, and identification of tissues having specific elemental compositions (11,96 98). Since that time, additional applications have been developed and prospective clinical trials performed to evaluate the clinical efficacy of these techniques (99). Active research and development in the field is ongoing and is yielding additional or improved clinical applications. The result is that dual- or multi-energy CT is moving firmly into the mainstream of CT imaging. Dualenergy CT may have disappeared from clinical consideration in the 1980s, Radiology: Volume 276: Number 3 September 2015 n radiology.rsna.org 649

14 Figure Alvarez RE, Macovski A. Energy-selective reconstructions in x-ray computerized tomography. Phys Med Biol 1976;21(5): Macovski A, Alvarez RE, Chan JL, Stonestrom JP, Zatz LM. Energy dependent reconstruction in x-ray computerized tomography. Comput Biol Med 1976;6(4): Kalender WA, Klotz E, Suess C. Vertebral bone mineral analysis: an integrated approach with CT. Radiology 1987;164(2): Kalender WA, Perman WH, Vetter JR, Klotz E. Evaluation of a prototype dualenergy computed tomographic apparatus. I. Phantom studies. Med Phys 1986;13(3): Ergun DL, Mistretta CA, Brown DE, et al. Single-exposure dual-energy computed radiography: improved detection and processing. Radiology 1990;174(1): Ishigaki T, Sakuma S, Ikeda M. One-shot dual-energy subtraction chest imaging with computed radiography: clinical evaluation of film images. Radiology 1988;168(1): Figure 18: Evaluation of the integrity of silicone breast implants by using unenhanced dual-energy CT in a 65-year-old woman with silicone implants placed in the right breast in 1976 and replaced in the left breast in (a, b) Axial and (c, d) coronal views demonstrate that silicone and soft tissue have very similar CT numbers in the mixed images (a, c), making it potentially difficult to differentiate a small extracapsular silicone leak from fibroglandular breast tissue. In the material-specific images (b, d), silicone is color coded in red, making it easy to differentiate silicone from soft tissue. Dense or high-atomic-number materials also take on a red tint (eg, cartilage and ribs) in the color-coded image. but with the tremendous technical advances that have taken place since that time, modern CT technology is now capable of taking full advantage of the fundamental physics and clinical promise that Godfrey Hounsfield envisioned in Acknowledgments: The authors are grateful to Sally J. Reinhart, Xinhui Duan, PhD, and Zhoubo Li, MS, for their assistance in manuscript preparation. The authors are grateful to the many radiologists and physicians at the Mayo Clinic who have contributed to advancing the field of dual-energy CT through their clinical collaborations, and to Terri Vrtiska, MD, Amy Kotsenas, MD, and Katrina Glazebrook, MD, for providing images used in this article. Disclosures of Conflicts of Interest: C.H.M. Activities related to the present article: disclosed no relevant relationships. Activities not related to the present article: reports grants from Sie- mens Healthcare. Other relationships: has a patent pending related to noise reduction in spectral CT. S.L. Activities related to the present article: disclosed no relevant relationships. Activities not related to the present article: disclosed no relevant relationships. Other relationships: has a patent pending related to noise reduction in spectral CT. L.Y. Activities related to the present article: disclosed no relevant relationships. Activities not related to the present article: disclosed no relevant relationships. Other relationships: has a patent pending related to noise reduction in spectral CT. J.G.F. Activities related to the present article: disclosed no relevant relationships. Activities not related to the present article: received grants from Siemens Healthcare. Other relationships: has a patent pending related to noise reduction in spectral CT. References 1. Hounsfield GN. Computerized transverse axial scanning (tomography). I. Description of system. Br J Radiol 1973;46(552): Stewart BK, Huang HK. Single-exposure dual-energy computed radiography. Med Phys 1990;17(5): Carmi R, Naveh G, Altman A. Material separation with dual-layer CT. IEEE Nuclear Science Symposium Conference Record 2005;4: Flohr TG, McCollough CH, Bruder H, et al. First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 2006;16(2): Johnson TR, Krauss B, Sedlmair M, et al. Material differentiation by dual energy CT: initial experience. Eur Radiol 2007; 17(6): Petersilka M, Stierstorfer K, Bruder H, Flohr T. Strategies for scatter correction in dual source CT. Med Phys 2010;37(11): Kelcz F, Joseph PM, Hilal SK. Noise considerations in dual energy CT scanning. Med Phys 1979;6(5): Primak AN, Ramirez Giraldo JC, Liu X, Yu L, McCollough CH. Improved dual-energy material discrimination for dual-source CT by means of additional spectral filtration. Med Phys 2009;36(4): Schlomka JP, Roessl E, Dorscheid R, et al. Experimental feasibility of multi-energy photon-counting K-edge imaging in preclinical computed tomography. Phys Med Biol 2008;53(15): radiology.rsna.org n Radiology: Volume 276: Number 3 September 2015

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare

More information

Cardiac Dual Energy CT: Technique

Cardiac Dual Energy CT: Technique RSNA 2013, VSCA51-01, Chicago, Dec. 5, 2013 Cardiac Radiology Series Cardiac Dual Energy CT: Technique Willi A. Kalender, Ph.D. Institute of Medical Physics University of Erlangen www.imp.uni-erlangen.de

More information

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Stefan C. Saur 1, Hatem Alkadhi 2, Luca Regazzoni 1, Simon Eugster 1, Gábor Székely 1, Philippe Cattin 1,3 1 Computer Vision

More information

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology.

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology. Photon Counting Detectors and Their Applications in Medical Imaging Principles, Pitfalls and Progress in Photon-Counting-Detector Technology Taly Gilat Schmidt, PhD Associate Professor Department of Biomedical

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

DICOM. Supplement 188 Multi-Energy CT Imaging. DICOM Working Group 21 Computed Tomography

DICOM. Supplement 188 Multi-Energy CT Imaging. DICOM Working Group 21 Computed Tomography DICOM Supplement 188 Multi-Energy CT Imaging DICOM Working Group 21 Computed Tomography Rationale Short introduction of Multi Energy (ME) s Overview: Imaging techniques, including scanning, reconstruction,

More information

The Near Future in Cardiac CT Image Reconstruction

The Near Future in Cardiac CT Image Reconstruction SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Tecniques of imaging in Overview Introduction Tecniques of imaging

More information

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland Metal Artifact Reduction CT Techniques R S S S Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland N. 1 v o 4 1 0 2. Postoperative CT Metal Implants CT is accurate for assessment

More information

Fundamentals of CT imaging

Fundamentals of CT imaging SECTION 1 Fundamentals of CT imaging I History In the early 1970s Sir Godfrey Hounsfield s research produced the first clinically useful CT scans. Original scanners took approximately 6 minutes to perform

More information

DUAL energy CT (DECT) is a modality where one and. Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography

DUAL energy CT (DECT) is a modality where one and. Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography Marc Kachelrieß, Member, IEEE, Timo Berkus, Philip Stenner, Willi A. Kalender Abstract Material selective imaging using dual energy

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Dual-Energy CT Based Monochromatic Imaging

Dual-Energy CT Based Monochromatic Imaging Dual-Energy CT Review Yu et al. Monochromatic Imaging Based on Dual-Energy CT Dual-Energy CT Review Lifeng Yu 1 Shuai Leng Cynthia H. McCollough Yu L, Leng S, McCollough CH Keywords: CT, dual-energy CT,

More information

RADIOLOGY AND DIAGNOSTIC IMAGING

RADIOLOGY AND DIAGNOSTIC IMAGING Day 2 part 2 RADIOLOGY AND DIAGNOSTIC IMAGING Dr hab. Zbigniew Serafin, MD, PhD serafin@cm.umk.pl 2 3 4 5 CT technique CT technique 6 CT system Kanal K: RSNA/AAPM web module: CT Systems & CT Image Quality

More information

Enhanced material contrast by dual-energy microct imaging

Enhanced material contrast by dual-energy microct imaging Enhanced material contrast by dual-energy microct imaging Method note Page 1 of 12 2 Method note: Dual-energy microct analysis 1. Introduction 1.1. The basis for dual energy imaging Micro-computed tomography

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Marc Kachelrieß, a Katia Sourbelle, and Willi A. Kalender Institute of Medical Physics, University of

More information

A New Approach for the Enhancement. of Dual-energy Computed Tomography Images. Kyung Kook Park

A New Approach for the Enhancement. of Dual-energy Computed Tomography Images. Kyung Kook Park A New Approach for the Enhancement of Dual-energy Computed Tomography Images by Kyung Kook Park A Dissertation Presented in Partial Fulfillment of the Requirements for the Degree Doctor of Philosophy Approved

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

Tomographic Reconstruction

Tomographic Reconstruction Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)

More information

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead.

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead. 1 CT Basics Principles of Spiral CT Dose 2 Who invented CT? 1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections Sir Godfrey Hounsfield worked for the Central

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic

More information

Acknowledgments and financial disclosure

Acknowledgments and financial disclosure AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories

More information

A closer look at CT scanning

A closer look at CT scanning Vet Times The website for the veterinary profession https://www.vettimes.co.uk A closer look at CT scanning Author : Charissa Lee, Natalie Webster Categories : General, Vets Date : April 3, 2017 A basic

More information

DUE to beam polychromacity in CT and the energy dependence

DUE to beam polychromacity in CT and the energy dependence 1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact

More information

Simulation of Internal Backscatter Effects on MTF and SNR of Pixelated Photon-counting Detectors

Simulation of Internal Backscatter Effects on MTF and SNR of Pixelated Photon-counting Detectors Simulation of Internal Backscatter Effects on MTF and SNR of Pixelated Photon-counting Detectors Alexander Korn, Juergen Giersch a and Martin Hoheisel b a Physikalisches Institut Universitaet Erlangen-Nuernberg,

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

CT imaging using energy-sensitive photon-counting detectors. Disclosure. Vision 20/20 paper

CT imaging using energy-sensitive photon-counting detectors. Disclosure. Vision 20/20 paper CT imaging using energy-sensitive photon-counting detectors Katsuyuki Ken Taguchi, Ph.D. ktaguchi@jhmi.edu Division of Medical Imaging Physics Department of Radiology Johns Hopkins University No financial

More information

Medical Image Processing: Image Reconstruction and 3D Renderings

Medical Image Processing: Image Reconstruction and 3D Renderings Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

CT vs. VolumeScope: image quality and dose comparison

CT vs. VolumeScope: image quality and dose comparison CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,

More information

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Qiao Yang 1,4, Meng Wu 2, Andreas Maier 1,3,4, Joachim Hornegger 1,3,4, Rebecca Fahrig

More information

ML reconstruction for CT

ML reconstruction for CT ML reconstruction for CT derivation of MLTR rigid motion correction resolution modeling polychromatic ML model dual energy ML model Bruno De Man, Katrien Van Slambrouck, Maarten Depypere, Frederik Maes,

More information

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology Diagnostic imaging techniques Krasznai Zoltán University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology 1. Computer tomography (CT) 2. Gamma camera 3. Single Photon

More information

FINDING THE TRUE EDGE IN CTA

FINDING THE TRUE EDGE IN CTA FINDING THE TRUE EDGE IN CTA by: John A. Rumberger, PhD, MD, FACC Your patient has chest pain. The Cardiac CT Angiography shows plaque in the LAD. You adjust the viewing window trying to evaluate the stenosis

More information

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),

More information

Scatter Correction Methods in Dimensional CT

Scatter Correction Methods in Dimensional CT Scatter Correction Methods in Dimensional CT Matthias Baer 1,2, Michael Hammer 3, Michael Knaup 1, Ingomar Schmidt 3, Ralf Christoph 3, Marc Kachelrieß 2 1 Institute of Medical Physics, Friedrich-Alexander-University

More information

Michal E. Kulon, MD 1,2 1. Peter Komlosi, MD, PhD 3 3. Radiology Universe Institute, 2

Michal E. Kulon, MD 1,2 1. Peter Komlosi, MD, PhD 3 3. Radiology Universe Institute, 2 Blend+Proximity, a novel algorithm achieves high suppression of metallic streak artifacts and maximal preservation of contrast between soft tissues and iodinated contrast material on dual-energy CT scans

More information

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry

ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry ISOCS Characterization of Sodium Iodide Detectors for Gamma-Ray Spectrometry Sasha A. Philips, Frazier Bronson, Ram Venkataraman, Brian M. Young Abstract--Activity measurements require knowledge of the

More information

Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application

Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application Physics in Medicine & Biology Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application To cite this article: Polad M Shikhaliev and Shannon G Fritz 2011

More information

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

Artifact Mitigation in High Energy CT via Monte Carlo Simulation PIERS ONLINE, VOL. 7, NO. 8, 11 791 Artifact Mitigation in High Energy CT via Monte Carlo Simulation Xuemin Jin and Robert Y. Levine Spectral Sciences, Inc., USA Abstract The high energy (< 15 MeV) incident

More information

Financial disclosure. Onboard imaging modality for IGRT

Financial disclosure. Onboard imaging modality for IGRT Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty

More information

Optimization of scanner parameters for dual energy micro-ct

Optimization of scanner parameters for dual energy micro-ct Optimization of scanner parameters for dual energy micro-ct E. PAUWELS* 1, J. DHAENE 1, A. DE MUYNCK 1 E., M. DIERICK 1, L. VAN HOOREBEKE 1 1 UGCT Dept. Physics and Astronomy, Ghent University, Proeftuinstraat

More information

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube T. Donath* a, F. Pfeiffer a,b, O. Bunk a, W. Groot a, M. Bednarzik a, C. Grünzweig a, E. Hempel c, S. Popescu c, M. Hoheisel

More information

Medical Imaging BMEN Spring 2016

Medical Imaging BMEN Spring 2016 Name Medical Imaging BMEN 420-501 Spring 2016 Homework #4 and Nuclear Medicine Notes All questions are from the introductory Powerpoint (based on Chapter 7) and text Medical Imaging Signals and Systems,

More information

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO

REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO REMOVAL OF THE EFFECT OF COMPTON SCATTERING IN 3-D WHOLE BODY POSITRON EMISSION TOMOGRAPHY BY MONTE CARLO Abstract C.S. Levin, Y-C Tai, E.J. Hoffman, M. Dahlbom, T.H. Farquhar UCLA School of Medicine Division

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.

Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran. DXRaySMCS First User Friendly Interface Developed for Prediction of Diagnostic Radiology X-Ray Spectra Produced by Monte Carlo (MCNP-4C) Simulation in Iran M.T. Bahreyni Toosi a*, H. Moradi b, H. Zare

More information

AIDR 3D Iterative Reconstruction:

AIDR 3D Iterative Reconstruction: Iterative Reconstruction: Integrated, Automated and Adaptive Dose Reduction Erin Angel, PhD Manager, Clinical Sciences, CT Canon Medical Systems USA Iterative Reconstruction 1 Since the introduction of

More information

Correlation between Model and Human Observer Performance on a Lesion Shape Discrimination Task in CT

Correlation between Model and Human Observer Performance on a Lesion Shape Discrimination Task in CT Correlation between Model and Human Observer Performance on a Lesion Shape Discrimination Task in CT Yi Zhang, Shuai Leng, Lifeng Yu and Cynthia McCollough Department of Radiology Mayo Clinic, Rochester

More information

Some reference material

Some reference material Some reference material Physics reference book on medical imaging: A good one is The Essential Physics of Medical Imaging, 3 rd Ed. by Bushberg et al. ($170! new). However, there are several similar books

More information

Empirical dual energy calibration EDEC for cone-beam computed tomography

Empirical dual energy calibration EDEC for cone-beam computed tomography Empirical dual energy calibration EDEC for cone-beam computed tomography Philip Stenner, Timo Berkus, and Marc Kachelriess Institute of Medical Physics, University of Erlangen-Nürnberg, Henkestrasse 91,

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy T. Rohlfing a, D. Zerfowski b, J. Beier a, P. Wust a, N. Hosten a, R. Felix a a Department of

More information

Quality control phantoms and protocol for a tomography system

Quality control phantoms and protocol for a tomography system Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2)

Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2) Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2) Author: Levon Vogelsang Introduction Cone beam computed tomography (CBCT), or cone beam CT technology, offers considerable promise

More information

Advanced Multi Material Decomposition of Dual Energy in Computed Tomography Image

Advanced Multi Material Decomposition of Dual Energy in Computed Tomography Image Advanced Multi Material Decomposition of Dual Energy in Computed Tomography Image A.Prema 1, M.Priyadharshini 2, S.Renuga 3, K.Radha 4 UG Students, Department of CSE, Muthayammal Engineering College, Rasipuram,

More information

Limitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods

Limitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods Stereoscopic Breast Imaging Andrew D. A. Maidment, Ph.D. Chief, Physics Section Department of Radiology University of Pennsylvania Limitations of Projection Radiography Mammography is a projection imaging

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms.

Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms. Comparison of internal and external dose conversion factors using ICRP adult male and MEET Man voxel model phantoms. D.Leone, A.Häußler Intitute for Nuclear Waste Disposal, Karlsruhe Institute for Technology,

More information

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Benedikt Lorch 1, Martin Berger 1,2, Joachim Hornegger 1,2, Andreas Maier 1,2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg

More information

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT.

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT. Progress report: Development of assessment and predictive metrics for quantitative imaging in chest CT Subaward No: HHSN6801000050C (4a) PI: Ehsan Samei Reporting Period: month 1-18 Deliverables: 1. Deployment

More information

Computer-Tomography II: Image reconstruction and applications

Computer-Tomography II: Image reconstruction and applications Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de

More information

Micro-CT Methodology Hasan Alsaid, PhD

Micro-CT Methodology Hasan Alsaid, PhD Micro-CT Methodology Hasan Alsaid, PhD Preclinical & Translational Imaging LAS, PTS, GlaxoSmithKline 20 April 2015 Provide basic understanding of technical aspects of the micro-ct Statement: All procedures

More information

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation 26 IEEE Nuclear Science Symposium Conference Record M6-349 Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation Habib Zaidi, Senior Member, IEEE and Mohammad

More information

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

Improvement of Efficiency and Flexibility in Multi-slice Helical CT J. Shanghai Jiaotong Univ. (Sci.), 2008, 13(4): 408 412 DOI: 10.1007/s12204-008-0408-x Improvement of Efficiency and Flexibility in Multi-slice Helical CT SUN Wen-wu 1 ( ), CHEN Si-ping 2 ( ), ZHUANG Tian-ge

More information

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET

Workshop on Quantitative SPECT and PET Brain Studies January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Workshop on Quantitative SPECT and PET Brain Studies 14-16 January, 2013 PUCRS, Porto Alegre, Brasil Corrections in SPECT and PET Físico João Alfredo Borges, Me. Corrections in SPECT and PET SPECT and

More information

Fast iterative beam hardening correction based on frequency splitting in computed tomography

Fast iterative beam hardening correction based on frequency splitting in computed tomography Fast iterative beam hardening correction based on frequency splitting in computed tomography Qiao Yang a,b, Matthias Elter b, Ingo Schasiepen b, Nicole Maass b and Joachim Hornegger a,c a Pattern Recognition

More information

Dual-Energy CT: General Principles

Dual-Energy CT: General Principles Dual-Energy CT Review Johnson General Principles of DECT Dual-Energy CT Review Thorsten R. C. Johnson 1 Johnson TRC Keywords: CT detector technology, dual-energy CT, dualsource CT, rapid voltage switching,

More information

CT: Physics Principles & Equipment Design

CT: Physics Principles & Equipment Design CT: Physics Principles & Equipment Design James Kofler, Ph.D Radiology Mayo Clinic Rochester, MN June 27, 2012 Disclosures Nothing to disclose Learning Objectives Understand fundamental concepts of - CT

More information

Physical bases of X-ray diagnostics

Physical bases of X-ray diagnostics Physical bases of X-ray diagnostics Dr. István Voszka Possibilities of X-ray production (X-ray is produced, when charged particles of high velocity are stopped) X-ray tube: Relatively low accelerating

More information

3-D Compounding of B-Scan Ultrasound Images

3-D Compounding of B-Scan Ultrasound Images 3-D Compounding of B-Scan Ultrasound Images Jochen F. Krücker, Charles R. Meyer, Theresa A. Tuthill, Gerald L. LeCarpentier, J. Brian Fowlkes, Paul L. Carson University of Michigan, Dept. of Radiology,

More information

Computed tomography - outline

Computed tomography - outline Computed tomography - outline Computed Tomography Systems Jørgen Arendt Jensen and Mikael Jensen (DTU Nutech) October 6, 216 Center for Fast Ultrasound Imaging, Build 349 Department of Electrical Engineering

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

RADIOMICS: potential role in the clinics and challenges

RADIOMICS: potential role in the clinics and challenges 27 giugno 2018 Dipartimento di Fisica Università degli Studi di Milano RADIOMICS: potential role in the clinics and challenges Dr. Francesca Botta Medical Physicist Istituto Europeo di Oncologia (Milano)

More information

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation

More information

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist CT Protocol Review: Practical Tips for the Imaging Physicist Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center August 8, 2013 AAPM Annual Meeting Goals Understand purpose and importance

More information

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

C a t p h a n / T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 Internationally recognized for measuring the maximum obtainable performance of axial, spiral and multi-slice

More information

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed

More information

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid by Yingxuan Chen Graduate Program in Medical Physics Duke University Date: Approved: Lei Ren, Supervisor Fang Fang Yin, Chair

More information

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION

FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION 2009 IEEE Nuclear Science Symposium Conference Record M03-7 FAST KVP-SWITCHING DUAL ENERGY CT FOR PET ATTENUATION CORRECTION Wonseok Huh, Jeffrey A. Fessler, Adam M. Alessio, and Paul E. Kinahan Department

More information

Edge-Preserving Denoising for Segmentation in CT-Images

Edge-Preserving Denoising for Segmentation in CT-Images Edge-Preserving Denoising for Segmentation in CT-Images Eva Eibenberger, Anja Borsdorf, Andreas Wimmer, Joachim Hornegger Lehrstuhl für Mustererkennung, Friedrich-Alexander-Universität Erlangen-Nürnberg

More information

Photon counting spectroscopic CT with dynamic beam attenuator

Photon counting spectroscopic CT with dynamic beam attenuator Photon counting spectroscopic CT with dynamic beam attenuator Haluk Atak 1) and Polad M. Shikhaliev a) 1) Department of Nuclear Engineering, Hacettepe University, Ankara, Turkey Abstract Purpose: Photon

More information

Ultrasonic Multi-Skip Tomography for Pipe Inspection

Ultrasonic Multi-Skip Tomography for Pipe Inspection 18 th World Conference on Non destructive Testing, 16-2 April 212, Durban, South Africa Ultrasonic Multi-Skip Tomography for Pipe Inspection Arno VOLKER 1, Rik VOS 1 Alan HUNTER 1 1 TNO, Stieltjesweg 1,

More information

Enhancement Image Quality of CT Using Single Slice Spiral Technique

Enhancement Image Quality of CT Using Single Slice Spiral Technique Enhancement Image Quality of CT Using Single Slice Spiral Technique Doaa. N. Al Sheack 1 and Dr.Mohammed H. Ali Al Hayani 2 1 2 Electronic and Communications Engineering Department College of Engineering,

More information

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography

More information

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA Computed Tomography Principles, Design, Artifacts, and Recent Advances THIRD EDITION Jiang Hsieh SPIE PRESS Bellingham, Washington USA Table of Contents Preface Nomenclature and Abbreviations xi xv 1 Introduction

More information

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS 23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS P. Helberg 1,

More information

Suitability of a new alignment correction method for industrial CT

Suitability of a new alignment correction method for industrial CT Suitability of a new alignment correction method for industrial CT Matthias Elter 1, Nicole Maass 1, Peter Koch 2 1 Siemens AG, Healthcare Sector, Erlangen, Germany, e-mail: matthias.elter@siemens.com,

More information

Physics 210 Medical Physics Midterm Exam Fall 2012 October 12, 2012

Physics 210 Medical Physics Midterm Exam Fall 2012 October 12, 2012 Physics 210 Medical Physics Midterm Exam Fall 2012 October 12, 2012 Name Problem 1 /32 Problem 2 /32 Problem 3 /24 Total /88 I affirm that I have carried out my academic endeavors with full academic honesty.

More information

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings)

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings) Marquette University e-publications@marquette Biomedical Engineering Faculty Research and Publications Engineering, College of 1-1-1999 An Iterative Approach to the Beam Hardening Correction in Cone Beam

More information

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information

Spectral analysis of non-stationary CT noise

Spectral analysis of non-stationary CT noise Spectral analysis of non-stationary CT noise Kenneth M. Hanson Los Alamos Scientific Laboratory Int. Symposium and Course on Computed Tomography, Las Vegas, April 7-11, 1980 This presentation available

More information

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 16 Chapter 3 Set Redundancy in Magnetic Resonance Brain Images 3.1 MRI (magnetic resonance imaging) MRI is a technique of measuring physical structure within the human anatomy. Our proposed research focuses

More information