1 Shane Foley: IPR, Chicago 2016 Iterative Reconstructions: The Impact on Dose and Image Quality Shane Foley, PhD Radiography and Diagnostic Imaging School of Medicine University College Dublin. Radagrafaíocht agus Iomhánna Fáthmheasach Scoil an Leighis Colaiste na hollscoile, BAC.

2 LEARNING OUTCOMES To understand the basic physical principles of iterative reconstruction techniques To become familiar with the various types of iterative reconstruction To review the published literature on the effects of IR use on both patient dose and image quality Shane Foley: IPR, Chicago 2016

3 OUTLINE CT image reconstructions The evolution of iterative reconstruction (hybrid model based) Clinical applications Potential disadvantages Shane Foley: IPR, Chicago 2016

4 C.T. CONTEXT Use has proliferated >80 million per year in USA 10% per year increase in most countries Radiation dose concerns Esp. for pediatrics Justification, optimisation, dose limitation Shane Foley: IPR, Chicago 2016

5 C.T. DOSE SURVEYS Large inter hospital discrepancies UK: fold 1 Norway: 8-20 fold 2 Ireland: 2-24 fold 3 Local site preferences and protocols largest influence on patient dose Goal of <1mSv imaging! 1. Shrimpton et al (1991). Survey of CT practice in the UK. Part 2: Dosimetric aspects. Chilton, NRPB-R249 (London, TSO) 2. Olerud HM. Analysis of factors influencing patient doses from CT in Norway. Radiation Protection Dosimetry, 71 (2), (1997) 3. Foley et al (2012). Establishment of CT diagnostic reference levels in Ireland. British Journal of Radiology.

6 RADIATION DOSE OPTIMISATION Operator dependent: Patient positioning / Protocol selection / Range limitation Technological features: 1994: AEC 2007: Dynamic beam collimators 2010: Tube voltage modulation 1999: ECG pulsing 2009: Iterative reconstruction 2010: Model based IR

7 C.T. IMAGE RECONSTRUCTION Filtered Back Projection (FBP) Used since CT was first developed Still primary method in use Main advantage = simple & quick Shane Foley: IPR, Chicago 2016

8 FILTERED BACK PROJECTION (FBP) Reconstructs an image by: a) Obtaining projection profiles b) Filtering each view Removes blurring seen in simple back projection Different filters for different tasks (sharp v smooth) c) Back projecting views Shane Foley: IPR, Chicago 2016

9 LIMITATIONS OF F.B.P. Any noise in raw data is propagated in image Traditionally countered by increasing exposure / thicker slices Higher spatial resolution comes with higher noise Based on many assumptions: focus spot is infinitely small that every detector element is small pencil beam without polychromatic spectrum. Each voxel has no shape or size Shane Foley: IPR, Chicago 2016

10 NOISE IN C.T. = principal limiting feature for low dose CT Degrades image quality: LCD But in CT: Quantum noise (detection of discrete photons) can be modelled by Poisson process Noise from other parts of the imaging chain can be statistically modelled Mathematic methods can be used to determine the most likely true signal from noisy projection data Edyvean S (2005). Impactscan.org

11 I.R. TECHNIQUES Not new available for 20+ years Used in SPECT and PET Introduction to CT slow - extensive computer power needed I.R. allows decoupling of spatial resolution & noise Correction loop introduced to reduce noise Iteration = the act of repeating a process Shane Foley: IPR, Chicago 2016

12 I.R. BASICS 1. Starts with an estimate of truth! (often FBP image) 2. Compares initial projection to a model projection 3. Initial projection is updated and process repeated (multiple iterations) until differences become acceptably small 4. By modelling noise can generate images with lower noise IR: does NOT decrease radiation dose Allows lower exposure settings to produce equivalent noise OR improved image quality Shane Foley: IPR, Chicago 2016

13 I.R. TECHNIQUES McCollough et al (2012) Radiology. Aug; 264(2):

14 I.R. BENEFITS Reduce noise within CT images Improved CNRs, or Reduce artefacts Beam hardening / Metallic implants / scatter Improved spatial resolution Avoiding data filtration before back projection Shane Foley: IPR, Chicago 2016

15 I.R. OPTIONS Full (pure) IR Physical properties of the acquisition system taken into account Computationally very expensive: reconstruction = cumbersome & time consuming Hybrid IR modified (faster) IR technique Blends FBP with IR, performing majority of noise reduction in image space Does not improve spatial resolution Does not model the optics Shane Foley: IPR, Chicago 2016

16 I.R. TECHNQIUES Currently three different options IR performed using: Image (slice) data Projection (raw) & image data Projection data only Image space: noise reduction primarily performed by statistical process Projection space: artefact reduction and increased spatial resolution Shane Foley: IPR, Chicago 2016

17 I.R: IMAGE DATA Eg: IRIS Iterative reconstruction in image space (Siemens) Process: Raw data first reconstructed using FBP Then forward projected with multiple iterations according to modelling of the noise data Pro: Recon time only slightly longer than FBP Con: assumes ideal system (LCD / streak artefacts not sign. improved) 1 1. Bittencourt MS, Schmidt B, Seltmann M et al. Iterative reconstruction in image space (IRIS) in cardiac computed tomography: initial experience. Int J Cardiovasc Imaging 2011;27:1081-7

18 I.R: PROJECTION & IMAGE DATA Eg: ASIR / AIDR 3D / SAFIRE / idose 4 Process: Projection data first reconstructed with FBP Compared with ideal noise model based on statistics (accounting for photon and electronic noise) Multiple iterations performed that compare each updated voxel with the ideal noise model until the algorithm converges Pro: recon times only slightly longer than FBP Con: assumes ideal system / limited reduction of streak artefact and spatial resolution improvement 1 1. Katsura M, Sato J, Akahane M et al. Comparison of pure and hybrid iterative reconstruction techniques with conventional filtered back projection: image quality assessment in the cervicothoracic region. Eur J Radiol. 2013;82:356-60

19 I.R: PROJECTION DATA ONLY Eg: MBIR / FIRST / IMR Models entire x-ray beam and system optics Pro: less noise 1 / improved spatial resolution and low contrast detectability, fewer streak artefacts 2 Con: Recon time significantly longer than FBP / Different look & feel to FBP, users need to adapt 1. Katsura M, Matsuda I, Akahane M et al. Model-based iterative reconstruction technique for radiation dose reduction in chest CT: comparison with the adaptive statistical iterative reconstruction technique. Eur Radiol. 2012;22: Thibault JB, Sauer KD, Bouman CA et al. A three dimensional statistical approach to improved image quality for multislice helical CT. Med Phys 2007;34:

20 MODEL BASED I.R. Includes details of the whole CT system & optics Physics modelling (scatter / cross talk) Focal spot size Detector element size The patient System geometry Beam energy Cone angle Electronic noise from the system Goal is to improve noise & spatial resolution at the same time Shane Foley: IPR, Chicago 2016

21 WHAT THE MANUFACTURERS SAY I.R. DOES Siemens: IRIS and SAFIRE* allows for up to 60% radiation dose reduction in routine clinical use. 1 Philips: 60-80% lower dose with 43-80% increased low contrast detectability and 70-83% lower noise 2 Toshiba: >75% dose reduction 3 GE: <1mSv scanning with higher resolution 4 1. Siemens Guide to Low dose. Available at 2. Philips (2016). Iterative Model based reconstruction. Available at 3. Toshiba (2014). White paper on AIDR 3D 4. GE (2015). Veo product

22 WHAT THE LITERATURE SAYS I.R. DOES 45% dose reduction for trunk CT Less noise (better lesion detection) & better spatial resolution than ASIR (30%) Ideally suited to pediatrics (better lesion detection & delineation) 92% dose reduction (depending on clinical indication) Improved visibility of small structures on coronal plane 1. Smith et al (2014). Model-based Iterative Reconstruction: Effect on Patient Radiation Dose and Image Quality in Pediatric Body CT. Radiology270: Number 2 February Miéville et al (2013). Model-based iterative reconstruction in pediatric chest CT: assessment of image quality in a prospective study of children with cystic fibrosis. Pediatr Radiol 43:

23 CLINICAL APPLICATIONS: CHEST Most widely investigated body part for pediatric IR High inherent contrast & low attenuation enable tolerance of image noise Studies to date all agree: IR allows dose reductions (31-92%) with improved image quality Dose reductions esp. useful due to radiosensitive organs in chest 1. Miéville et al. Paediatric cardiac CT examinations: impact of the iterative reconstruction method ASIR on image quality preliminary findings. Pediatr Radiol 2011; 41: Miéville et al (2013). Model-based iterative reconstruction in pediatric chest CT: assessment of image quality in a prospective study of children with cystic fibrosis. Pediatr Radiol 43:

24 Padole et al (2015). 2015;AJR 204: W384-W392. FBP: 7mGy SafeCT: 1.6mGy ASIR: 1.6mGy MBIR: 1.6mGy

25 den Harder et al (2015). AJR 204: CHEST: 11yo 80kVp / 70mAs / 38 DLP FBP: Noise 113 CNR 0.6 idose4: Noise 79 CNR 0.7 IMR 3: Noise 15 CNR 3.0, - Streak artefact

26 CHEST Smoothing effect of IR can potentially result in small structures not being visible Studies to date reported MBIR improved visualisation of small structures (lung fissures & small vessels) 1,2 Further studies needed 1. Koc et al (2014). Computed tomography depiction of small pediatric vessels with model-based iterative reconstruction. Pediatr Radiol 2014; 44: [ 2. Miéville et al (2013). Model-based iterative reconstruction in pediatric chest CT: assessment of image quality in a prospective study of children with cystic fibrosis. Pediatr Radiol 43:

27 Han et al (2011). JCCT May CARDIAC CT: 80 kv, DLP 12 SAFIRE: reduced noise 35%, improved SNR/CNR 50%

28 1. Vorona et al (2011). Reducing abdominal CT radiation dose with the adaptive statistical iterative reconstruction technique in children: a feasibility study. Pediatr Radiol 2011; 41: Singh et al (2012). Radiation Dose Reduction with Hybrid Iterative Reconstruction for Pediatric CT. Radiology Young et al (2016). Pediatr Radiol (2016) 46: ABDOMEN More challenging due to lower organ contrast Studies show varying % dose reductions according to vendor and application ASIR: 33% dose reduction (40% ASIR) while maintaining IQ % dose reduction 2 MBIR: 76% reduction vs FBP 3

29 Padole et al (2015). 2015;AJR 204: W384-W392. ABDOMEN: 7y.o. weight loss, diarrhea FBP: 4.3mGy FBP: 1.3mGy ASIR : 1.3mGy MBIR: 1.3mGy

30 LESION DETECTION (14 yo lymphoma) MBIR images were rated as superior to 100% ASIR images MBIR 100% ASIR FBP 1. Smith et al (2014). Model-based Iterative Reconstruction: Effect on Patient Radiation Dose and Image Quality in Pediatric Body CT. Radiology270: Number 2 February 2014

31 HEAD C.T. Most data from adult studies Improved SNR, CNR in ASIR vs FBP ASIR: 30% dose reductions vs FBP with similar IQ + lesion conspicuity Intracranial haemorrhage better seen on SAFIRE at reduced dose vs FBP 0.9mSv MBIR: improves IQ vs ASIR with dose reduction Paediatric studies: ASIR: 28% dose reduction for 3- to 12-year-old patients and 48% in reduction at 30 mgy >12 years vs FBP 1 MBIR: 50% dose reduction with similar IQ vs ASIR 2 1. McKnight et al Pediatr Radiol Aug;44(8): Smith et al (2014). Model-based Iterative Reconstruction: Effect on Patient Radiation Dose and Image Quality in Pediatric Body CT. Radiology270: Number 2 February 2014

32 den Harder et al (2015). AJR 204: HEAD C.T: 8yo 120kV, 198mAs, DLP 527 Better grey white matter discrimination with IMR FBP: Noise 7.7 CNR 0.3 IMR 1: Noise 3.2 CNR 3.4 IMR 2: Noise 2.8 CNR 3.9 IMR 3: Noise 2.4 CNR 4.5

33 SPINAL C.T. No studies to date on paediatrics Adults: 40% dose reductions (Safire) with better image quality for IV disks, neural foramina & ligaments BUT IQ for non spinal soft tissues declined slightly Need more studies Shane Foley: IPR, Chicago 2016

34 IMPACT ON IMAGE QUALITY Noise Spatial resolution (MTF Smith et al Radiology) Texture Artefacts Shane Foley: IPR, Chicago 2016

35 IMAGE QUALITY: NOISE Obvious reduction in image noise w/o affecting signal BUT conventional relationship between noise and dose may not be valid anymore. Noise is much less sensitive to the dose change. 50% dose reduction results in only 15-19% increase in noise. (vs 40% in FBP) 1 This relationship needs re-assessment under different dose levels 1. Dong (2014). AAPM Shane Foley: IPR, Chicago 2016

36 IMAGE QUALITY: SPATIAL RESOLUTION 1. Smith et al (2014). Model-based Iterative Reconstruction: Effect on Patient Radiation Dose and Image Quality in Pediatric Body CT. Radiology270: Number 2 February 2014

37 IMAGE QUALITY: TEXTURE MBIR: unusually smooth Less conspicuous on MPR images Radiologists need to adapt Blotchy / pixillated appearance reported in adults NOT paediatrics Lack of data / radiologists higher acceptance of noise/artifacts in paeds Earlier versions of IR algorithms Shane Foley: IPR, Chicago 2016

38 ARTEFACTS IR: potential to reduce beam hardening & photon starvation MBIR: subtle staircase effect on bony interfaces 1. Deak et al (2013). Radiology 266 1

39 ARTEFACTS MBIR: skin surfaces Small blacked out pixels on axial? Clinically significant 1. Deak et al (2013). Radiology 266 1

40 RING ARTEFACTS Wagner-Bartak et al (2013). CT model based iterative reconstruction technique: Pearls, Pitfalls and Practical solutions. RSNA

41 RING ARTEFACTS Ali Khawaja et al (2015). European Journal of Radiology, Volume 84, Issue 1, 2015, 2 10

42 BREAST SHIELD ARTEFACT Some evidence that shields may cause further artefacts Wagner-Bartak et al (2013). CT model based iterative reconstruction technique: Pearls, Pitfalls and Practical solutions. RSNA

43 DISADVANTAGES I.R. Computationally time consuming Expensive Compatible with only latest scanners? Need for vendor neutral options: e.g. MedicVision, Clarity, Saphenia Vendor specific algorithms may not be applicable to those of others Artefacts Shane Foley: IPR, Chicago 2016

44 1. Available from 2. Jensen et al (2013). Model-based Iterative Reconstruction (MBIR) Implementation: Process Description and Lessons Learned. RSNA 3. Vardhanabhuti et al (2013). Image quality assessment of standard and low dose CT using FBP, ASIR and MBIR, AJR 200, TIME CONSUMING Pure IR much longer FBP: 26 images per second (ips) SAFIRE: 20 ips idose4: 18 ips VEO: 0.09 ips (35-40mins per abdomen/pelvis) Mieville (2013): mins chest CT depending on patient size & DFOV? Suitable for emergency dept patients or in patients?

45 SUMMARY Excellent potential for dose reduction Standard IR: reduces noise + some artefacts Hybrid IR: significant dose reductions in acceptable time Model based IR: potential for further dose reductions with improvements in image quality Still testing limits and practical opportunities Shane Foley: IPR, Chicago 2016

46 Questions.?

47 REFERENCES den Harder et al. (2015) Hybrid and Model-Based Iterative Reconstruction Techniques for Pediatric CT. AJR; 204: Karmazyn et al (2013). Optimization of Hybrid Iterative Reconstruction Level in Pediatric Body CT. AJR 2014; 202: Mehta et al. (2013). ITERATIVE MODEL RECONSTRUCTION: SIMULTANEOUSLY LOWERED COMPUTED TOMOGRAPHY RADIATION DOSE AND IMPROVED IMAGE QUALITY. MEDICAL PHYSICS INTERNATIONAL Journal Miéville et al (2013). Model-based iterative reconstruction in pediatric chest CT: assessment of image quality in a prospective study of children with cystic fibrosis. Pediatr Radiol 43: Liu (2014). Model based iterative reconstruction: A promising algorithm for Todays Computed Tomography Imaging. Journal of Medical Imaging Sciences 45: Padole et al (2015). CT Radiation Dose and Iterative Reconstruction Techniques. AJR Smith et al (2013). Model-based iterative reconstruction: Effect on Patient Radiation Dose and Image Quality in Pediatric Body CT. Radiology Tricarico et al (2013). Cardiovascular CT angiography in neonates and children. Image quality and potential for radiation dose reduction with iterative image reconstruction techniques. Eur Radiol 23, Wagner-Bartak et al (2013). CT model based iterative reconstruction technique: Pearls, Pitfalls and Practical solutions. RSNA Young et al (2016) Knowledge-based iterative model reconstruction: comparative image quality and radiation dose with a pediatric computed tomography phantom. Pediatr Radiol 46: Shane Foley: IPR, Chicago 2016

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

CT Iterative Reconstruction Techniques

CT Iterative Reconstruction Techniques RECENT ADVANCES IN CT RADIATION DOSE REDUCTION TECHNIQUES CT Iterative Reconstruction Techniques Kalpana Kanal, PhD, FSCBTMR, FACR, FAAPM University of Washington Seattle, WA SCBT-MR 2017 Nashville, Tennessee

More information

Optimisation of Toshiba Aquilion ONE Volume Imaging

Optimisation of Toshiba Aquilion ONE Volume Imaging Optimisation of Toshiba Aquilion ONE Volume Imaging Jane Edwards, RPRSG Royal Free London NHS Foundation Trust Dr Mufudzi Maviki, Plymouth Hospitals NHS Trust Background In 2011/12 Radiology at RFH was

More information

AIDR 3D Iterative Reconstruction:

AIDR 3D Iterative Reconstruction: Iterative Reconstruction: Integrated, Automated and Adaptive Dose Reduction Erin Angel, PhD Manager, Clinical Sciences, CT Canon Medical Systems USA Iterative Reconstruction 1 Since the introduction of

More information

Limits of Dose Reduction in CT: How low is too low? Acknowledgement. Disclosures 8/2/2012

Limits of Dose Reduction in CT: How low is too low? Acknowledgement. Disclosures 8/2/2012 8/2/22 Limits of Dose Reduction in CT: How low is too low? Ehsan Samei Duke University Acknowledgement Disclosures Research grant: NIH Research grant: DHS Research grant: RSNA, QIBA Research grant: General

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004 Spiral ASSR Std p = 1.0 Spiral EPBP Std p = 1.0 Kachelrieß et al., Med. Phys. 31(6): 1623-1641, 2004 256 slices (0/300) Advantages of Cone-Beam Spiral CT Image quality nearly independent of pitch Increase

More information

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland Metal Artifact Reduction CT Techniques R S S S Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland N. 1 v o 4 1 0 2. Postoperative CT Metal Implants CT is accurate for assessment

More information

Performance Evaluation of CT Systems

Performance Evaluation of CT Systems 8/2/206 Performance Evaluation of CT Systems AAPM Task Group 233 Foundations, Methods, and Prospects Ehsan Samei Duke University Medical Center Disclosures Research grant: NIH R0 EB00838 Research grant:

More information

AAPM Standard of Practice: CT Protocol Review Physicist

AAPM Standard of Practice: CT Protocol Review Physicist AAPM Standard of Practice: CT Protocol Review Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center September 11, 2014 2014 Texas Radiation Regulatory Conference Goals Understand purpose

More information

Tomographic Reconstruction

Tomographic Reconstruction Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

ImPACT. CT dosimetry and a data base for CTDI values. EFOMP Workshop at ECR Imaging Performance Assessment of CT Scanners

ImPACT. CT dosimetry and a data base for CTDI values. EFOMP Workshop at ECR Imaging Performance Assessment of CT Scanners ImPACT Imaging Performance Assessment of CT Scanners A Medical Devices Agency Evaluation Group CT dosimetry and a data base for CTDI values EFOMP Workshop at ECR 2002 S. Edyvean, ImPACT St George s Hospital,

More information

Acknowledgments and financial disclosure

Acknowledgments and financial disclosure AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories

More information

A closer look at CT scanning

A closer look at CT scanning Vet Times The website for the veterinary profession https://www.vettimes.co.uk A closer look at CT scanning Author : Charissa Lee, Natalie Webster Categories : General, Vets Date : April 3, 2017 A basic

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

GPU implementation for rapid iterative image reconstruction algorithm

GPU implementation for rapid iterative image reconstruction algorithm GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial

More information

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist CT Protocol Review: Practical Tips for the Imaging Physicist Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center August 8, 2013 AAPM Annual Meeting Goals Understand purpose and importance

More information

RADIOLOGY AND DIAGNOSTIC IMAGING

RADIOLOGY AND DIAGNOSTIC IMAGING Day 2 part 2 RADIOLOGY AND DIAGNOSTIC IMAGING Dr hab. Zbigniew Serafin, MD, PhD serafin@cm.umk.pl 2 3 4 5 CT technique CT technique 6 CT system Kanal K: RSNA/AAPM web module: CT Systems & CT Image Quality

More information

Disclosures 7/21/2014. AAPM 2014 Challenges and opportunities in assessment of image quality

Disclosures 7/21/2014. AAPM 2014 Challenges and opportunities in assessment of image quality 7/2/24 AAPM 24 Challenges and opportunities in assessment of image quality Ehsan Samei Department of Radiology Duke University Medical Center Disclosures Research grant: NIH R EB838 Research grant: General

More information

Computer-Tomography II: Image reconstruction and applications

Computer-Tomography II: Image reconstruction and applications Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de

More information

CT vs. VolumeScope: image quality and dose comparison

CT vs. VolumeScope: image quality and dose comparison CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,

More information

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA Computed Tomography Principles, Design, Artifacts, and Recent Advances THIRD EDITION Jiang Hsieh SPIE PRESS Bellingham, Washington USA Table of Contents Preface Nomenclature and Abbreviations xi xv 1 Introduction

More information

Medical Image Processing: Image Reconstruction and 3D Renderings

Medical Image Processing: Image Reconstruction and 3D Renderings Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,

More information

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT Assessment of 3D performance metrics D and 3D Metrics of Performance Towards Quality Index: Volumetric imaging systems X-ray based Volumetric imaging systems: CBCT/CT Tomosynthesis Samuel Richard and Ehsan

More information

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare

More information

Fundamentals of CT imaging

Fundamentals of CT imaging SECTION 1 Fundamentals of CT imaging I History In the early 1970s Sir Godfrey Hounsfield s research produced the first clinically useful CT scans. Original scanners took approximately 6 minutes to perform

More information

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic

More information

7/13/2015 EVALUATION OF NONLINEAR RECONSTRUCTION METHODS. Outline. This is a decades-old challenge

7/13/2015 EVALUATION OF NONLINEAR RECONSTRUCTION METHODS. Outline. This is a decades-old challenge EVALUATION OF NONLINEAR RECONSTRUCTION METHODS Kyle J. Myers, Ph.D. Director, Division of Imaging, Diagnostics, and Software Reliability Office of Science and Engineering Laboratories, CDRH, FDA 2 Outline

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

Precision of Iodine Quantification in Hepatic CT: Effects of Iterative Reconstruction With Various Imaging Parameters

Precision of Iodine Quantification in Hepatic CT: Effects of Iterative Reconstruction With Various Imaging Parameters Medical Physics and Informatics Original Research Iterative Reconstruction Algorithms in Hepatic CT Medical Physics and Informatics Original Research Baiyu Chen 1,2 Daniele Marin 3 Samuel Richard 2,3,4

More information

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE / Radiographic Tomosynthesis: Image Quality and Artifacts Reduction Baojun Li, Ph.D Department of Radiology Boston University Medical Center 2012 AAPM Annual Meeting Background Linear Trajectory Tomosynthesis

More information

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead.

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead. 1 CT Basics Principles of Spiral CT Dose 2 Who invented CT? 1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections Sir Godfrey Hounsfield worked for the Central

More information

Computed Tomography. What a progress

Computed Tomography. What a progress IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, 2017 Computed Tomography (a little bit of all) Dean Pekarovič UMC Ljubljana, Institute of Radiology Quality and Safety office

More information

A noise power spectrum study of a new model-based iterative reconstruction system: Veo 3.0

A noise power spectrum study of a new model-based iterative reconstruction system: Veo 3.0 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 5, 2016 A noise power spectrum study of a new model-based iterative reconstruction system: Veo 3.0 Guang Li, 1 Xinming Liu, 1 Cristina T.

More information

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT Qiao Yang 1,4, Meng Wu 2, Andreas Maier 1,3,4, Joachim Hornegger 1,3,4, Rebecca Fahrig

More information

William P. Shuman, MD Keith T. Chan, MD Janet M. Busey, MS Lee M. Mitsumori, MD Eunice Choi, BS Kent M. Koprowicz, MS Kalpana M. Kanal, PhD.

William P. Shuman, MD Keith T. Chan, MD Janet M. Busey, MS Lee M. Mitsumori, MD Eunice Choi, BS Kent M. Koprowicz, MS Kalpana M. Kanal, PhD. Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. William P. Shuman,

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

Biomedical Imaging. Computed Tomography. Patrícia Figueiredo IST

Biomedical Imaging. Computed Tomography. Patrícia Figueiredo IST Biomedical Imaging Computed Tomography Patrícia Figueiredo IST 2013-2014 Overview Basic principles X ray attenuation projection Slice selection and line projections Projection reconstruction Instrumentation

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

ML reconstruction for CT

ML reconstruction for CT ML reconstruction for CT derivation of MLTR rigid motion correction resolution modeling polychromatic ML model dual energy ML model Bruno De Man, Katrien Van Slambrouck, Maarten Depypere, Frederik Maes,

More information

F3-A5: Toward Model-Based Reconstruction in Scanned Baggage Security Applications

F3-A5: Toward Model-Based Reconstruction in Scanned Baggage Security Applications F3-A5: Toward Model-Based Reconstruction in Scanned Baggage Security Applications Abstract While traditional direct reconstruction algorithms such as filtered back projection depend on the analytic inversion

More information

8/2/2016. Acknowledgement. Common Clinical Questions. Presumption Images are Good Enough to accurately answer clinical questions

8/2/2016. Acknowledgement. Common Clinical Questions. Presumption Images are Good Enough to accurately answer clinical questions Image Quality Assessment using Model Observers: Clinical Implementation and Practical Considerations Shuai Leng, PhD Associate Professor, Department of Radiology Mayo Clinic, Rochester MN Acknowledgement

More information

F3-A5: Model-Based Iterative Reconstruction for Advanced Baggage Screening

F3-A5: Model-Based Iterative Reconstruction for Advanced Baggage Screening F3-A5: Model-Based Iterative Reconstruction for Advanced Baggage Screening Abstract While traditional direct reconstruction algorithms such as filtered back projection (FBP) and direct Fourier method (DFM)

More information

Some reference material

Some reference material Some reference material Physics reference book on medical imaging: A good one is The Essential Physics of Medical Imaging, 3 rd Ed. by Bushberg et al. ($170! new). However, there are several similar books

More information

The Near Future in Cardiac CT Image Reconstruction

The Near Future in Cardiac CT Image Reconstruction SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de

More information

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,

More information

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed

More information

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Stefan C. Saur 1, Hatem Alkadhi 2, Luca Regazzoni 1, Simon Eugster 1, Gábor Székely 1, Philippe Cattin 1,3 1 Computer Vision

More information

Cardiac Dual Energy CT: Technique

Cardiac Dual Energy CT: Technique RSNA 2013, VSCA51-01, Chicago, Dec. 5, 2013 Cardiac Radiology Series Cardiac Dual Energy CT: Technique Willi A. Kalender, Ph.D. Institute of Medical Physics University of Erlangen www.imp.uni-erlangen.de

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Limitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods

Limitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods Stereoscopic Breast Imaging Andrew D. A. Maidment, Ph.D. Chief, Physics Section Department of Radiology University of Pennsylvania Limitations of Projection Radiography Mammography is a projection imaging

More information

Agenda : Lung Density Breakout Session

Agenda : Lung Density Breakout Session Agenda : Lung Density Breakout Session 1. : Mathew Fuld and Bernice Hoppel 2. Automatic Exposure Control (AEC) Evaluation : Sean Fain Round 4 Project 3. Dose reduction effects on emphysema metrics : Philip

More information

Introduction to Positron Emission Tomography

Introduction to Positron Emission Tomography Planar and SPECT Cameras Summary Introduction to Positron Emission Tomography, Ph.D. Nuclear Medicine Basic Science Lectures srbowen@uw.edu System components: Collimator Detector Electronics Collimator

More information

Medical Physics and Informatics Original Research

Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Shuman et al. Three Reconstruction Algorithms for 64-MDCT of Liver Lesions Medical Physics and Informatics Original Research William P. Shuman 1 Doug E.

More information

Interaction map. X-Ray Computed Tomography Measures Tissue Properties from Macro to Micro. Outline

Interaction map. X-Ray Computed Tomography Measures Tissue Properties from Macro to Micro. Outline X-Ray Computed Tomography Measures Tissue Properties from Macro to Micro Michael Andre, Ph.D. Department of Radiology University of California, San Diego San Diego VA Healthcare System Outline Perspective

More information

PURE. ViSION Edition PET/CT. Patient Comfort Put First.

PURE. ViSION Edition PET/CT. Patient Comfort Put First. PURE ViSION Edition PET/CT Patient Comfort Put First. 2 System features that put patient comfort and safety first. Oncology patients deserve the highest levels of safety and comfort during scans. Our Celesteion

More information

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT.

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT. Progress report: Development of assessment and predictive metrics for quantitative imaging in chest CT Subaward No: HHSN6801000050C (4a) PI: Ehsan Samei Reporting Period: month 1-18 Deliverables: 1. Deployment

More information

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Marc Kachelrieß, a Katia Sourbelle, and Willi A. Kalender Institute of Medical Physics, University of

More information

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Wenjuan Sun and Stephen Brown 10 th May 2016 1 Possible factors influencing XCT measurements Components Influencing variables Possible

More information

CT issues in PET / CT scanning. ImPACT technology update no. 4

CT issues in PET / CT scanning. ImPACT technology update no. 4 abc David Platten Introduction ImPACT technology update no. 4 ImPACT October 2004 Since its introduction in the 1970s, X-ray computed tomography (CT) scanning has become a widespread three-dimensional

More information

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Tecniques of imaging in Overview Introduction Tecniques of imaging

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram TomoTherapy Related Projects An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram Development of A Novel Image Guidance Alternative for Patient Localization

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

Financial disclosure. Onboard imaging modality for IGRT

Financial disclosure. Onboard imaging modality for IGRT Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty

More information

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing ImPACT Information Leaflet No. 1: CT Scanner Acceptance Testing Version 1.02, 18/05/01 CONTENTS: 1. SCOPE OF LEAFLET 2. GENERAL PRINCIPLES OF ACCEPTANCE AND COMMISSIONING 2.1 PHANTOMS 2.2 EXPOSURE AND

More information

Philips SPECT/CT Systems

Philips SPECT/CT Systems Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET

More information

Computed Tomography Imaging: CT Protocol Management. Caveat 8/3/2017

Computed Tomography Imaging: CT Protocol Management. Caveat 8/3/2017 Computed Tomography Imaging: CT Protocol Management Mark P. Supanich, Ph.D., DABR AAPM Annual Meeting 3 rd August, 2017 Slides at: goo.gl/k8n8jf Rush is an academic health system comprising Rush University

More information

Scatter Correction Methods in Dimensional CT

Scatter Correction Methods in Dimensional CT Scatter Correction Methods in Dimensional CT Matthias Baer 1,2, Michael Hammer 3, Michael Knaup 1, Ingomar Schmidt 3, Ralf Christoph 3, Marc Kachelrieß 2 1 Institute of Medical Physics, Friedrich-Alexander-University

More information

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),

More information

Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2)

Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2) Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2) Author: Levon Vogelsang Introduction Cone beam computed tomography (CBCT), or cone beam CT technology, offers considerable promise

More information

Real World Experience: Developing Dose and Protocol Monitoring from Scratch

Real World Experience: Developing Dose and Protocol Monitoring from Scratch Real World Experience: Developing Dose and Protocol Monitoring from Scratch Ingrid Reiser, PhD DABR Department of Radiology The University of Chicago Outline CT protocol monitoring Let s build a protocol

More information

7/11/2015. Imaging Equipment Specification and Selection in Radiation Oncology Department. What Therapy Physicists Need to Know

7/11/2015. Imaging Equipment Specification and Selection in Radiation Oncology Department. What Therapy Physicists Need to Know Imaging Equipment Specification and Selection in Radiation Oncology Department Chiaho Hua St. Jude Children s Research Hospital Yue Cao University of Michigan, Ann Arbor Margie Hunt Memorial Sloan Kettering

More information

Computed tomography - outline

Computed tomography - outline Computed tomography - outline Computed Tomography Systems Jørgen Arendt Jensen and Mikael Jensen (DTU Nutech) October 6, 216 Center for Fast Ultrasound Imaging, Build 349 Department of Electrical Engineering

More information

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Benedikt Lorch 1, Martin Berger 1,2, Joachim Hornegger 1,2, Andreas Maier 1,2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg

More information

Frequency split metal artifact reduction (FSMAR) in computed tomography

Frequency split metal artifact reduction (FSMAR) in computed tomography The Johns Hopkins University Advanced Computer Integrated Surgery Group 4 Metal Artifact Removal in C-arm Cone-Beam CT Paper Seminar Critical Review of Frequency split metal artifact reduction (FSMAR)

More information

DUE to beam polychromacity in CT and the energy dependence

DUE to beam polychromacity in CT and the energy dependence 1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact

More information

Registration concepts for the just-in-time artefact correction by means of virtual computed tomography

Registration concepts for the just-in-time artefact correction by means of virtual computed tomography DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Registration concepts for the just-in-time artefact correction by means of virtual

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

Acknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury

Acknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury A. Sisniega et al. (presented at RSNA 214) High Performance Cone-Beam CT of Acute Traumatic Brain Injury A. Sisniega 1 W. Zbijewski 1, H. Dang 1, J. Xu 1 J. W. Stayman 1, J. Yorkston 2, N. Aygun 3 V. Koliatsos

More information

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration

Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Low-Dose Dual-Energy CT for PET Attenuation Correction with Statistical Sinogram Restoration Joonki Noh, Jeffrey A. Fessler EECS Department, The University of Michigan Paul E. Kinahan Radiology Department,

More information

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology.

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology. Photon Counting Detectors and Their Applications in Medical Imaging Principles, Pitfalls and Progress in Photon-Counting-Detector Technology Taly Gilat Schmidt, PhD Associate Professor Department of Biomedical

More information

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy T. Rohlfing a, D. Zerfowski b, J. Beier a, P. Wust a, N. Hosten a, R. Felix a a Department of

More information

Enhancement Image Quality of CT Using Single Slice Spiral Technique

Enhancement Image Quality of CT Using Single Slice Spiral Technique Enhancement Image Quality of CT Using Single Slice Spiral Technique Doaa. N. Al Sheack 1 and Dr.Mohammed H. Ali Al Hayani 2 1 2 Electronic and Communications Engineering Department College of Engineering,

More information

PLANMECA PROMAX 3D MID CBCT UNIT

PLANMECA PROMAX 3D MID CBCT UNIT 1(5) PLANMECA PROMAX 3D MID CBCT UNIT Introduction The Planmeca ProMax 3D Mid X-ray unit uses cone beam computerized tomography (CBCT) to produce three-dimensional X-ray images. Panoramic and cephalometric

More information

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition

Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Validation of GEANT4 for Accurate Modeling of 111 In SPECT Acquisition Bernd Schweizer, Andreas Goedicke Philips Technology Research Laboratories, Aachen, Germany bernd.schweizer@philips.com Abstract.

More information

Whole Body Submillimeter Scan

Whole Body Submillimeter Scan 128 Whole Body Submillimeter Scan The real value of 64ch/128slice CT systems does not come from the ability to perfom cardiac scans but the capability to scan all parts of the body in high definition submillimeter

More information

Digital Scatter Removal in Mammography to enable Patient Dose Reduction

Digital Scatter Removal in Mammography to enable Patient Dose Reduction Digital Scatter Removal in Mammography to enable Patient Dose Reduction Mary Cocker Radiation Physics and Protection Oxford University Hospitals NHS Trust Chris Tromans, Mike Brady University of Oxford

More information

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD 1 Outline q Introduction q Imaging performances in 4D-CBCT Image

More information

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct J. DHAENE* 1, E. PAUWELS 1, T. DE SCHRYVER 1, A. DE MUYNCK 1, M. DIERICK 1, L. VAN HOOREBEKE 1 1 UGCT Dept. Physics

More information

CT imaging using energy-sensitive photon-counting detectors. Disclosure. Vision 20/20 paper

CT imaging using energy-sensitive photon-counting detectors. Disclosure. Vision 20/20 paper CT imaging using energy-sensitive photon-counting detectors Katsuyuki Ken Taguchi, Ph.D. ktaguchi@jhmi.edu Division of Medical Imaging Physics Department of Radiology Johns Hopkins University No financial

More information

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA

Review of PET Physics. Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Review of PET Physics Timothy Turkington, Ph.D. Radiology and Medical Physics Duke University Durham, North Carolina, USA Chart of Nuclides Z (protons) N (number of neutrons) Nuclear Data Evaluation Lab.

More information

As fl exible as your care requires

As fl exible as your care requires As fl exible as your care requires Philips Ingenuity Flex 32 CT Built on Ingenuity The Philips Ingenuity Flex 32 helps you provide excellent care with outstanding flexibility, now and in the future. Built

More information

DIPLOMA THESIS. Optimization in CT. Evaluation of dose saving potential in a thorax-abdomen/pelvis protocol using iterative reconstruction techniques

DIPLOMA THESIS. Optimization in CT. Evaluation of dose saving potential in a thorax-abdomen/pelvis protocol using iterative reconstruction techniques Optimization in CT Evaluation of dose saving potential in a thorax-abdomen/pelvis protocol using iterative reconstruction techniques Ingvild Dalehaug Master of Science in Physics and Mathematics Submission

More information

Michal E. Kulon, MD 1,2 1. Peter Komlosi, MD, PhD 3 3. Radiology Universe Institute, 2

Michal E. Kulon, MD 1,2 1. Peter Komlosi, MD, PhD 3 3. Radiology Universe Institute, 2 Blend+Proximity, a novel algorithm achieves high suppression of metallic streak artifacts and maximal preservation of contrast between soft tissues and iodinated contrast material on dual-energy CT scans

More information

Introduction to Emission Tomography

Introduction to Emission Tomography Introduction to Emission Tomography Gamma Camera Planar Imaging Robert Miyaoka, PhD University of Washington Department of Radiology rmiyaoka@u.washington.edu Gamma Camera: - collimator - detector (crystal

More information