Computed Tomography. What a progress

Size: px
Start display at page:

Download "Computed Tomography. What a progress"

Transcription

1 IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, 2017 Computed Tomography (a little bit of all) Dean Pekarovič UMC Ljubljana, Institute of Radiology Quality and Safety office What a progress

2 CT tecnical evolution Future Technology FPD, more then 2 x-ray tubes, IT implementation of AI 320 Slice CT = 320 slice per revolution 256 Slice CT = 256 slice per revolution Dual Source CT = 2 x ray tubes coupled with 2 detector arrays 64 Slice CT = 64 slice per revolution 8-16, Slice CT = 8, 16, 32, 40 slice per revolution Four Slice CT = 4 slice per revolution Dual Slice CT = 2 slice per revolution Single Slice Spiral/Helical CT = 1 slice per revolution First Clinical CT 1. Generation: Rotate/Translate, Pencil Beam Pencil beam shape -pinhole collimator to ensure only a single beam of x-rays small detectors rarely detected any scattered radiation 1-2detectors- positioned in x ray tube direction 4.5 minute for capturing data and for one slice only X ray tube rotated only 180 degrees 2

3 2. Generation Rotate/Translate, Narrow Fan Beam Narrow Fan beam shape ( 10 ) 30 or more detector elements 20 sec/slice and up to 10 minutes for 40 slice reconstruction 180 degree rotation x-ray tube Very long reconstruction times Detectors were exposed to more scattered radiation Only for head and brain due to its size and limited movement linear movement of the x-ray tube and detectors at each projection angle 3. Generation Rotate/Rotate, Wide Fan Beam Wide Fan beam shape (40 60 ) Next data is collected when x ray tube and detectors changed direction detector elements opposite to x-ray tube 360 degree rotation; rotation/table movement, 1 Gantry rot.= 1 slice Acquisition time to less than 20 seconds Reconstruction time for 1 slice is 1 second Not only for head, but for body scans too : ring artefacts! 3

4 4. Generation Rotate/Stationary Were developed minimize the ring artifacts Fixed ring with detector elements, around patient (~4800) X ray tube is rotating less radiation is needed Detector are capable to detect more radiation By removing the detectors from the rotating gantry and putting them in a stationary ring around the patient, detectors were able to maintain calibration 1 minute for more slices 5. Generation EBCT Stationary/Stationary Wide Fan beam shape CT scanner is composed of no moving parts. Like one Huge x-ray tube. Behind the patient, there is an electron beam, which ejects electrons. This electron beam is electronically deflected down, away from the patient, and makes contact with a large, half-circle tungsten target ring that encircles the patient. The interaction of the electrons with the target ring generates an x-ray beam, which travels through the patient s chest and is detected by a detector ring on the opposite side. Scan time is about 50 msec 4

5 6. Generation Spiral/Helical Slip Ring technology eliminate the time necessary for braking, and reversal of spooled cables X-ray source and detector array rotate continuously as the patient table is moved progressively through the scanner. Faster, helical (spiral or volume) CT, which offered greater spatial and temporal resolution Use of principles of the 3. and 4. generations with the slip ring technology 7. Generation Multiple Detector Array Cone Beam System can acquire an outstanding amount of information in a very short time span, requiring a much higher level of sophistication in the reconstruction process (slice competition between vendors z coverage). Spiral MDCT - submillimetre resolution during one breath hold. 5

6 CT generations overview Source Detector Source detector Generation Source Detector Advantages Disadvantages Collimation collimation movement Scattered Single x-ray Move linearly and Slow 1 G Pencil beam Single None energy is tube rotate unison ( 2.5 min) undetected Fan beam, not Lower efficacy Single x-ray Cannot collimate Move linearly and 2 G enough to cover Multiple Faster then 1G and larger noise tube detectors rotate unison FOV (10 sec) More expensive Single x-ray Fan beam, enough Cannot collimate Faster then 2G 3 G Many Rotate in synchrony then 2 G, low tube to cover FOV detectors,slip RING efficacy (0,5 sec) High scattering - Single x-ray Fan beam covers Stationary ring of Cannot collimate Fixed detectors, Higher 4 G detector are not tube FOV detectors detectors source rotates efficacy collimated (1 sec) Many Tungsten High cost, Stationary ring of Cannot collimate 5 G anodes in a Fan Beam No moving parts Extremely fast difficult to detectors detectors single large tube calibrate 3G/4G plus 6 G A bit more 3G/4G 3G/4G 3G/4G 3G/4G LINEAR Table Fast 3D images Spiral/Helical expensive MOTION 7 G Single x-ray Multiple arrays of Collimated to source Cone -Beam 3G/4G/6G motion Fast 3D images Expensive Multislice tube detectors direction CT components 6

7 CT needs In-plane resolution: mm Nominal slice thickness: S = mm Effective slice thickness: Seff = mm Tube (max. values): 120 kw, 150 kv, 1300 ma Effective tube current: maseff = 10 mas 1000 mas Rotation time: Trot = s Simultaneously acquired slices: M = Table increment per rotation: d = mm Pitch value: p = (up to 3.2 for DSCT) Scan speed: up to 73 cm/s Temporal resolution: ms Ref: Kachelrieß, Basics of Tomography 1 X ray Tube 7

8 X Ray Tube Characteristics X Ray Tube Tube voltages from 70 to 150 kv High instantaneous power levels (typ. 50 to 120 kw) ma High cooling rates (typ. > 1 MHU/minute) Must with stand centrifugal force Fast Tube Cooling Rates. Centrifugal force at 550 mm with 0.5 s: F = 9 g with 0.4 s: F = 14 g with 0.3 s: F = 25 g with 0.2 s: F = 55 g 8

9 State of The art CT performance GE Revolution Philips IQ Spectral Siemesn Force Toshiba One Vision Ref: CT scanners for, King's College London Detector needs Available as multi-row arrays Very fast sampling (typ. 300 μs) Favourable temporal characteristics (decay time < 10 μs) High absorption efficiency High geometrical efficiency - narrow gaps between active elements High count rate (up to 109 cps*) Small physical size Ref: Kachelrieß, Basics of Tomography 1 9

10 Detectors Properties: High detection efficiency(~90%). High geometrical efficiency (~80%). Small physical size of detector elements. Most commonly used detectors in new scanners. Hybrid, Matrix, Phased Array Matrix Phased Array Hybrid Hybrid 1. CT Generation 2. CT Generation 10

11 Detector Configuration 2 Hybrid Matrix 32 SLICE CT 64 SLICE CT Detector array configurations for state of the art MDCT 11

12 Detector Configuration Represents number of data channels along patient length (or z-axis) multiplied by the effective detector row thickness of individual data channel. 4 x x x 1.5 GE Philips Siemens Toshiba Detector configuration Collimation N X T (mm) Detector Configuration Detector Configuration Detector - collimation Siemens Sensation 64 Possible Configurations at spiral mode: 64 x 0.6* (19.2 mm nominal beam width ) Modes: 0.6, 0.75, 1, 1.5, 2, 3, 4, 5, 6, 7, 8, 10 mm Thinner slices, less volume per rotation, longer time for same volume. 24 x 1.2 (28.8 mm nominal beam width ) Modes : 1.2, 1.5, 2, 3, 4, 5, 6, 7, 8, 10 mm Bigger volume per scam, thinner slice is 1,2 mm. *Z-flying focal spot: In real 32 x 0.6 mm 12

13 Detector configuration- Impact on dose! 15 % 27 % Fixed mas AEC Off What is important? High detection capability : Detector must percept big amount of radiation in change it to light. Faster it is better the detector is. Afterglow : When radiation is converted to light, detector starts to glow. When detector is black" again, it can accumulate new amount of radiation. High Image Resolution: Number of pixels, number of channels,, number of rows in size of element below millimetre define ImQ. 13

14 Collimators Before Patient Single array collimator determine slice thickness MDCT different solutions by vendors Behind patient( reduce scatter radiation to improve contrast). Cone-Beam Artefacts mm from iso center 4 slice CT/ no processing 16 slice CT/ no processing 16 slice CT/ with processing 14

15 Anti-scatter grids 1D Anti Scatter Grid 2D Anti Scatter Grid As multi-slice scanners with large-area detectors have advanced, so has the need to reduce x-ray scatter through the use of improved anti-scatter grids. Increases low contrast for larger patients Anti-scatter grids are aligned to the detector pixels. Anti-scatter grids reject scattered radiation. Bowtie filter 15

16 z-over-scanning Helical scanners must irradiate a larger volume of patient than is ultimately displayed in the image data set : Necessary evil to avoid having incomplete data for reconstruction of first and last slices of interest. Typically an extra rotation is required a beginning and end of imaged volume. Pitch Is the Table Feed per gantry rotation divided by the beam width/collimation. Pitch is the ratio of two distances and therefore has no units. Users should monitor other parameters when changing Pitch. The scanner may or may not automatically compensate for changes in Pitch (for example, by changing the tube current). Pitch > 1 Some view angles are not covered by the beam width at certain table positions Pitch = 1 No overlap of Beam Width at each view angle and no view angles not covered at certain table positions Pitch < 1 Beam Width has some overlap at each view angle from rotation to rotation 16

17 Pitch (helical) Pitch (IEC,2003) T = Detector width in z axis N = nu. Of detector elements per rotation (N>1 at MDCT), N x T full width of detector during acquisition I = table speed in z direction in one full rotation mm/rot N*T = collimated x-ray beam width I = 15mm/rot N = 4 data channels T = 2,5 mm detector width P = (15 mm/rot) / (4 x 2,5 mm) P = 15/10 =1,5 Collimator and detector pitch T = table movement in 360⁰ ration of gantry C = width od collimation D = width of detector Collimator pitch = T/C Detector pitch = T/D Collimator pitch = Detector pitch / N Ref.: The Essential Physics of Medical Imagining;Bushberg et.all N = no. of detectors E.G.: FOV 20 mm Table speed 20 mm/rot Collimation 4 x 5 mm Collimator pitch = 1 Detector pitch = 4 High or low pitch? 17

18 Resolutions Image Quality Spatial Resolution Contrast resolution Temporal resolutions Spatial Resolution (SR) Ability to observe or separate small objects Minimal observable size depends on object contrast (CNR) Spatial resolution is also called high contrast resolution Spatial resolution is described by In axial plane: point spread function or MTF In z direction: slice sensitivity profile Gives the actual width of the imaged slice Usually quantified as FWHM 18

19 Factor affecting SR Detector pitch : is centre-to centre spacing along the array. Detector aperture : is the width of active element of one detector. Smaller detectors increase cut off(nyquist) frequency of the image (improve SR). No. of views : more views less artefacts (aliasing). mas: big influence in same FOV Focal spot size (FS): larger FS cause more geometric unsharpness and reduce SR Object magnification : Increased magnification amplifies the blurring of the focal spot. Slice thickness : the slice thickness is equivalent to the detector aperture in CR-CA direction. Large slice thickness Helical pitch : greater pitch reduce SR (increased slice sensitivity profile) Reconstruction kernel : Bone filters have the best SR and soft filters lower SR Pixel Matrix : direct influence (256², 512² view 1024²) FOV : 10 cm FOV in 512² ;0,2 mm pixel size 35 cm FOV in 512² ;0,7 mm pixel size The limiting factor is noise Contrast Resolution (CR) Difference in signal between a structure and its environment The contrast resolution of CT is not intrinsically high, because the difference in x-ray attenuation between different tissues is generally small (vs. MRI) High nu. of projections vs. object overlapping e.g. chest x ray 25 mas 200 mas The limiting factor is Noise -Dose 19

20 Factor affecting CR mas : directly influence the number of x-ray photons used to produce image, thereby affecting the SNR and CR (double mas-41% better CR-but) Dose : increase linearly with mas per scan Pixel size (FOV) : if FOV increase, pixel dimension increase, and no. of x- rays passing through same pixel increase(scan parameters and patient size is same) Slice thickness : thicker slice use more photons and have better SNR Reconstruction filter : Bone filters has lower CR and soft filters improves CR Patient size : Larger patients attenuate more x-rays, resulting in detection of fewer x-rays, lower SNR and therefore CR Gantry rotation speed : compared 1 s with 0.5 s results in reduced mas to produce each CT image, reducing CR Temporal resolution Ability to resolve fast moving objects For image reconstruction at least 180 is needed (180 MLI) If in plane are no moving organs 360 State of the art CT 3-4 rotations/s Influences motion artefacts Influences motion induced blurring Particularly important for moving organs (heart) children, uncooperative patients Requires fast tube rotation (or 2 tubes ) 20

21 Temporal resolution CCTA Ref: CT scanners for, King's College London Reconstruction settings Reconstruction order ma prescription under TCM may be linked to specific recon Reconstruction kernel determines spatial resolution, image noise and texture filtered back-projection, iterative reconstruction selection may affect ma prescription under TCM Recon mode e.g. Recon Full, Recon Plus selection may affect ma prescription under TCM Ref: E. Castellano, IAEA CT Workshop 21

22 Reconstruction algorithm Soft Standard Smooth - soft tissuee Detail Lung Bone Edge Bone Plus Sharp - bone GE recon algorithms Reconstruction kernel Represents a feature on the scanner which influences the smoothness and sharpness of images in transverse plane B20 B31 B50 B70 Lower noise Poorer edge delineation Better contrast Higher noise level Better edge delineation Poor contrast Standard Details Lung GE Philips Siemens Toshiba Algorithm Reconstruction filter Kernel Filter Convolution 22

23 Kernels Head 1 vendor Siemens H 40s H (Head), B (Body), C (Child Head), S (Special Applications) s standard mode f- fast mode h- high resolution mode H21, H31, H41 are like H20, H30, H40 but with finer grain noise and a milder edge enhancement. H22, H32, H42 don t include iterative beam hardening correction (PFO). Reconstruction speed is faster, but the reconstructed images may contain significant beam hardening artifacts. H37 is comparable to GE Soft H45 is intermediate sharpness between H40 and H50 H47 is comparable to GE Standard H48 is like H47 but a bit sharper Reconstructed slice thickness 10 mm 5mm 2mm 1 mm 0,5 mm 23

24 Reconstruction Possibilities-AIP 1. 0,8 mm Coronal view/plane 2. 4,0 mm Coronal view/plane ImQ more soft, better Contrast Resolution, almost like on axial planes Average Intensity Projection MIP MIP (cca 10 % basic image info) AIP (2,5 mm) MIP (2,5 mm) MIP 5 mm MIP 10 mm MIP 15 mm Overlaping with structures with high HU (White, Bone, other vessels..). 24

25 MIP slice Thickness 7mm /5 mm 2. 4mm/ 2mm Image reconstruction 1972 CT discovery Geofrey Hounsfield (UK) Technical part Allan Cormack (SA) Mathematical reconstruction 25

26 Hounsfield Cormak Improved Radon transformation 26

27 Mathematics (grrrr) I 0 Intensity of x-ray beam entering object h material thickness μ μ attenuation coefficient (tissue characteristic) I Intensity of x-ray beam exiting object I I 0 e h More then one tissue I 0 I I 0 e h μ is average value for all tissues in path 27

28 Matrix can be used for calculation I 0 I I μ1 μ2 μ3 μ4 μ5 μ6 I I 0 0 e e ( x x x3 4x4 5x5 6x6 ( ) x ) Reconstruction = using mathematical algorithms to reconstruct an image from projections 28

29 CT from X ray tube to Image RECON TYPE : FBP IR Noise Reduction X ray Tube: kv mas Bow Tie Filter Detector : Data Collection DAS: Detector Efficiency Multiple calculations

30 Reconstruction algorithms FBP Filtered Back projection IR Iterative reconstruction Filtered Back Projection Smeares measured projection data back to image More projections better image 30

31 FBP 2 projections OBJECT FBP M M ,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,4 1,4 1,4 1,4 1,4 1,4 1,4 1,4 1,4 1,4 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,0 1,1 1,1 1,1 1,1 1,1 1,1 1,1 1,1 1,1 1,1 1,7 1,7 1,7 1,7 1,7 1,7 1,7 1,7 1,7 1,7 2,2 2,2 2,2 2,2 2,2 2,2 2,2 2,2 2,2 2,2 1,7 1,7 1,7 1,7 1,7 1,7 1,7 1,7 1,7 1,7 1,1 1,1 1,1 1,1 1,1 1,1 1,1 1,1 1,1 1,1 1 st proj. 2 nd proj. 1,1 1,7 2,2 1,7 1,1 1,0 1,4 0,9 1,0 1,0 1,1 1,7 2,2 1,7 1,1 1,0 1,4 0,9 1,0 1,0 1,1 1,7 2,2 1,7 1,1 1,0 1,4 0,9 1,0 1,0 1,1 1,7 2,2 1,7 1,1 1,0 1,4 0,9 1,0 1,0 1,1 1,7 2,2 1,7 1,1 1,0 1,4 0,9 1,0 1,0 1,1 1,7 2,2 1,7 1,1 1,0 1,4 0,9 1,0 1,0 1,1 1,7 2,2 1,7 1,1 1,0 1,4 0,9 1,0 1,0 1,1 1,7 2,2 1,7 1,1 1,0 1,4 0,9 1,0 1,0 1,1 1,7 2,2 1,7 1,1 1,0 1,4 0,9 1,0 1,0 1,1 1,7 2,2 1,7 1,1 1,0 1,4 0,9 1,0 1,0 1,1 1,4 1,6 1,4 1,1 1,0 1,2 1,0 1,0 1,0 1,1 1,4 1,6 1,4 1,1 1,0 1,2 1,0 1,0 1,0 1,3 1,6 1,8 1,6 1,3 1,2 1,4 1,2 1,2 1,2 1,1 1,4 1,6 1,4 1,1 1,0 1,2 1,0 1,0 1,0 1,1 1,4 1,6 1,4 1,1 1,0 1,2 1,0 1,0 1,0 1,1 1,4 1,7 1,4 1,1 1,1 1,3 1,1 1,1 1,1 1,4 1,7 2,0 1,7 1,4 1,4 1,6 1,4 1,4 1,4 1,7 2,0 2,2 2,0 1,7 1,6 1,8 1,6 1,6 1,6 1,4 1,7 2,0 1,7 1,4 1,4 1,6 1,4 1,4 1,4 1,1 1,4 1,7 1,4 1,1 1,1 1,3 1,1 1,1 1,1 IMAGE FBP limitations noise,artefacts.. FBP M IMAGE M OBJECT 1,1 1,4 1,6 1,4 1,1 1,0 1,2 1,0 1,0 1,0 1,1 1,4 1,6 1,4 1,1 1,0 1,2 1,0 1,0 1,0 1,3 1,6 1,8 1,6 1,3 1,2 1,4 1,2 1,2 1,2 1,1 1,4 1,6 1,4 1,1 1,0 1,2 1,0 1,0 1,0 1,1 1,4 1,6 1,4 1,1 1,0 1,2 1,0 1,0 1,0 1,1 1,4 1,7 1,4 1,1 1,1 1,3 1,1 1,1 1,1 1,4 1,7 2,0 1,7 1,4 1,4 1,6 1,4 1,4 1,4 1,7 2,0 2,2 2,0 1,7 1,6 1,8 1,6 1,6 1,6 1,4 1,7 2,0 1,7 1,4 1,4 1,6 1,4 1,4 1,4 1,1 1,4 1,7 1,4 1,1 1,1 1,3 1,1 1,1 1,1 31

32 FBP projections FBP what is in calculations Point source Pencil Beam Point Voxel Point detector 32

33 What contribute to noise on image Technology upgrade Improvements in x ray tubes Bowtie filters Rotation time/ collimation Antiscatter grid Detector efficiency (spectral imagining) High pitch scanning modes Reconstruction algorithms Not possible to change Patient (positioning, metal implants, moving.) FBP way to.. Lowering the dose to get still diagnostic quality Research, evaluation years Adopting different noise level (age, indication orientated ). Goal Define lower limit in LCR, no decrease SR mas ,7 mgy Pekarovic, Ref:Ramirez-Giraldo, IAEA Sofia Simens USA Ref:Radiation dose from MDCT, Tack et all 33

34 FBP limitations - detailed Simplified Geometry Infinetly small: Focal spot Thin beam Point detector Physics Perfect sampling Simplified physics Scatter Absorption Beam filtration Noise In reconstrucions it use even measurements with very low response same good as perfect ones. High noise is compensating with increase in dose : kv mas Rot time Pitch etc.. Artifacts In this type of reconstruction we can`t avoid : Metal artifacts Streak artifact Beam Hardening Windmil HU are higher in tissue where artifacts are Next.. Maybe we got maximum output from FBP ImQ Dose Is Balance Achieved? Is iterative reconstruction a solution? 34

35 Iterative reconstructions (IR) WHY IR?? Dose - CT highly contribute to population dose Number of CT scans rapidly increase More indications Post processing power of CT, visibility. Fast and accurate Experience from NM Possibility to reduce noise Market needs Iterative reconstruction 1st generation using statistical model 2nd generation using mathematical model of imaging system Vendor 1st generation 2nd generation Philips idose(irt) IMR General Electric ASIR ASIR-V, Veo* Siemens IRIS SAFIRE, ADMIRE Toshiba QDS/Boost 3D AIDR 3D Not included SIEMENS Itrim (cardiac, TR) 35

36 Iterative reconstructions Forward projection instead of back Several steps: Assume an image (empty or FBP), Compute projections from the image, Compare to the original projection data and Update the image based on the difference between calculated and actual projections. SIR 1 st Generation SIR uses FBP data set as the building block for image reconstruction assuming the beam to be a perfect point source. Use this or raw data to compare to other data (e.g. forward projection data, noise smoothed image) --- compares, makes corrections, does that again, until?? It aims to improve image quality by focusing on noise reduction. 36

37 SIR Comprehensive noise model includes: Photon, Beam hardening, electronic noise Severe photon starvation (such as through metal), which creates star, streak other artefacts. Can use adjacent projections and voxels to reweight. Attempts to preserve edges. How Really Works To complicated to understand without appropriate background knowledge Ask Medical Physicist to explain it softly. Important to know : How it looks on Images 37

38 Difference in FBP IR Point source Pencil Beam Point Voxel Point detector Real source Fan Beam 3D Voxel shape Detector Geometry Tube and detector response Huge calculations SIR FBP all vendor is same principle IR different approaches like : Assumptions, calculations (physics and optics). Algorithms, kernels (filters). Blending options. Choices (levels, percentages). Guidance (kv, mas, etc.). Software, hardware evolving rapidly. 38

39 Advantages of IR Dose Noise Artefacts Speed ImQ IR statistical Denoising 39

40 IR Homogenous phantom FBP 120 kv, 300 mas, CTDI vol 12,5 IR -50 % (SIR)- half Dose More Noise less softening vs. IR 100 % IR -100 % (SIR) half Dose Less Noise- softening vs. FBP Ref: Hara et all, Iterative Reconstruction Technique for Reducing Body Radiation Dose at CT: Feasibility Study, AJR:193, September 2009 Validation? 50 mgy FBP Average: 35; Min/Max : % IR - 12,5 mgy Average: 31 Min/Max :7 51 In both cases: ROI in Center : HU ± 3 40

41 Low Contrast Resolution 50 mgy FBP 50 % IR 12,5 mgy 4 x less dose same contrast- same ImQ IR and SR A : regular dose B : low dose C : low dose +IR 60% D : low dose + IR 100% Ref :O.Rapalino, S.Kamalian et all, Cranial CT with Adaptive Statistical Iterative Reconstruction: Improved Image Quality with Concomitant Radiation Dose Reduction, AJNR

42 Way to go Protocols can remain unchanged Keep the same dose better ImQ Lower dose same ImQ 120 kv 10mAs 0,5 rot. time 0,625 slice thickness 0,61 msv Image Courtesy, dr.barbau,cnn,france 0,61 msv DLP 36,02 On Images- Low Contrast Rresolution Low Dose CT, FBP: no IR 120kv, 3,75 mm CTDI 8 Low Dose CT, FBP: + IR 120kv, 3,75 mm CTDI 8 Routine CT, FBP: no IR 140kv, 3, mm CTDI 22 Same Dose Same ImQ Ref : Hara et all,iterative Reconstruction Technique for Reducing Body Radiation Dose at CT: Feasibility Study, AJR :September 2009, Volume 193, Number 3 42

43 On Images Spatial Resolution Low Dose CT, FBP: no IR 120kv, 3,75 mm CTDI 11 Low Dose CT, FBP: with IR 120kv, 3,75 mm CTDI 11 routine CT, FBP: no IR 140kv, 3, mm CTDI 20 Edge of cyst soft or sharp? 1 st generation FBP FBP low dose FBP AIDR processing ASIR 67 % dose reduction IRIS 60% dose reduction 43

44 IR and CR A : regular dose B : low dose C : low dose +IR 60% D : low dose + IR 100% Plasticy Waxy Blotchy Pixelated Ref :O.Rapalino, S.Kamalian et all, Cranial CT with Adaptive Statistical Iterative Reconstruction: Improved Image Quality with Concomitant Radiation Dose Reduction, AJNR 2012 Steps or % betwwen FBP and IR 44

45 2nd Generation Model Based IR Whilst SIR still relies on FBP data sets, model-based iterative reconstruction builds a forward projection using dedicated system optics, taking into account every x-ray projection (in its true three-dimensional domain), and produces an image based on the raw data. Multiple iterations are performed to correct the residual error between the forward projection and acquired image. Much more complicated algorithms These algorithms also incorporate statistical noise information in the reconstruction process. The combination of system optic modeling and statistical modeling helps in noise reduction and results in truer image characteristics compared to FBP and SIR. In addition, MBIR also accounts for noise from photon flux as well as system noise (e.g. electronic noise) from the CT system itself. AEC - effect on IR Courtesy of GE 45

46 IR and CR 100 mas 25 mas 12 mas Ref : P.Kinahan, The Future of Low Dose CT, Univ. Of Washington 1st and 2nd Sub msv FBP ASIR 50% VEO Low dose GE Analytical rec. idose IMR Images: courtesy of Kevin Brown, Philips 46

47 1st and 2nd Sub msv FBP IRIS SAFIRE IV Images: courtesy of Thomas Flohr and Rainer Raupach, Siemens Analytical rec. QDS/Boost AIDIR3D Images: courtesy of proff.sujith Seneviratne,Toshiba IR and patient size 40-cm phantom FBP vs IR Diameter % noise reduction FBP IR Mild 24 cm 23% 30 cm 36% 40 cm 51% IR Std. IR Strong Ref;:Kim M. Et all, Adaptive Iterative Dose Reduction Algorithm in CT: Effect on Image Quality Compared with Filtered Back Projection in Body Phantoms of Different Sizes, Korean J Radiol

48 Recipe Dose 100 % 50 % 25 % 120 kv and 100 Kv Different reconstruction algoritms Slice thickness 1mm and 3mm 3 mm FBP IR 2 IR 3 1 mm FBP IR 2 IR 4 IR may reduce CT patient dose depending on the clinical task, patient size, anatomical location, clinical practice But An aggressive dose reduction can reduce lesion detection So.. A consultation with a radiologist and a physicist should be made to determine the appropriate dose to obtain diagnostic image quality for the particular clinical task. 48

49 ImQ- SR Courtesy of Ge Healthcare IR + metal artefact reduction 49

50 2 nd Artefacts and bones WW/WL 350/50 WW/WL 2000/400 Ref : Intel Xeon Processors + GE, Whitepaper, New Levels of CT Image Performance and New Levels in Radiation Dose Management 3 D (VR +MIP) + IR Ref: Philips PM Ref :Improved Delineation of Arteries in the Posterior Fossa of the Brain by Model-Based Iterative Reconstruction in Volume-Rendered 3D CT Angiograph, Machida et all,

51 IR Impact on CTDIvol and DLP Ref : IR dose reduction factors Ref: Qiu and Seeram; Does Iterative Reconstruction Improve Image Quality and Reduce Dose in Computed Tomography? 51

52 DECT First article! Two pictures are taken of the same slice, one at 100 kv and the other at 140 kv so that areas of high atomic numbers can be enhanced. Tests carried out to date have shown that iodine (Z = 53) can be readily distinguished from calcium (Z =20). Hounsfield 4 decades ago DECT Basic Principle The principles of DECT are based largely on the photoelectric effect and can be achieved by exploiting the energy-dependent attenuation of materials when exposed to 2 different photon energy levels ( 2 basis materials). These physical principles can be exploited for in vivo human imaging, because DECT is based on dissimilar tissue characteristics with respect to their energy dependent x- ray attenuation. 52

53 DECT - basics predominant at higher x-ray energy levels Materials in human body hydrogen (Z=1) calcuim (Z=20) carbon (Z=6) iodine (Z=53) nitrogen (Z=7) oxygen (Z=8) higher at the lower x-ray energy levels Compton scattering is proportional to the electron density. Lower noise on image. Photoelectron absorption is directly proportional to the atomic number and electron density. Emission of characteristic radiation, Higher noise on image. Can we see more with DECT Materials can be chosen arbitrarily, as long as their K edges are sufficiently different (i.e., attenuation profiles), such as water and iodine. Material with an attenuation spectrum different than that of the chosen basis materials will be reflected as a combination of the 2 basis materials Ref: Danad et all, New Application of Cardiac CT 53

54 DECT + Tube current and tube filtration can be optimized for each tube potential independently. Relatively low degree of spectral overlap, which improves CNR in material-specific images. Beam-hardening corrections are applied prior to image reconstruction, allowing material-specific images to be created in the image domain. - A 90 phase shift between low- and high-energy data. Simultaneous use of both x-ray sources allows scattered radiation whose original primary photon came from one tube to be detected by the detector of the other tube, requiring specialized scatter correction. Ref: Danad et all, New Application of Cardiac CT DECT + Data acquisition of the low- and highenergy data set allows dual-energy material-decomposition algorithms to be implemented by using either projection data or reconstructed images. Reduces beam-hardening artefacts in calculated virtual monoenergetic images. Same HU. - There is no current modulation. Relatively high overlap of the energy spectra (80 kv and 140 kv). Gantry rot/time is 0,5 s between (80 kv and 140 kv) prolongs time of aquisotion Ref: Danad et all, New Application of Cardiac CT 54

55 DECT + Simultaneous data acquisition of the low- and high-energy data set. All image data are acquired in a manner that supports material-specific imaging. - Relatively high overlap of the energy spectra. Noise level may differ between low- and high-energy images. Ref: Danad et all, New Application of Cardiac CT DECT + Can be performed on any CT scanner - Single source-detector pair with tube voltage switching between sequential gantry rotations. Any patient motion occurring between the two scans may cause severe degradation of the resultant images and material composition information. a change in contrast opacification Ref: Danad et all, New Application of Cardiac CT 55

56 DECT DECT mix series 20 % - 80 % 80 kv 150 kv 60 % - 40 % 80 kv 150 kv 90 % - 10 % 80 kv 150 kv 56

57 DECT Mono energetic Monoenergetic 44 kev Monoenergetic 50 kev Monoenergetic 60 kev Monoenergetic 80 kev Mono-energetic reconstructions allow to reconstruct the images at any given virtual kev setting of a CT scan to optimal contrast needed for diagnostic purposes. Low kev reduced volume of CM -higher High kev artefacts/obese patient Monoenergetic 100 kev Monoenergetic 120 kev Monoenergetic 150 kev Monoenergetic 189 kev SECT DECT example 3 cm scan length Arterial phase 120 kv 80 kv 150 kv 80 kv mixed 57

58 SECT DECT example DECT postprocessing Iodine Map Metal artefact Table/bone removal Perfusion - PE Renal Stones 58

59 Flat Panel Detector- - CT Detector Detector Matrix: elementov (40cm x30cm). Effectiv esize (magnification) :25 x 25 x 18cm. Element size : µm. FOV in z axis/1 rotation: 18 cm. Spatial Resolution (x, y, z): µm. ANY ANGLE frames/second( CT 1000x/s). Speed big FOV in detector. Z axis Organs: Heart, Brain, liver in 1 rotation. Diagnostic Conventional modalities CT Angiography Fluoroscopy image in real time Flat Panel Detector- - CT Quick detection. Big detector instead rows (30x 40cm, volume). Bigger FOV in 1 rotation (in z-axis). More radiation contribute to image formation vs. Amorphous Siliceous. Less noise contribute to wider dynamic range (grey levels - attenuation). Faster readout(cardiac, hold breath, moving artefacts...). Smaller pixel size then in crystal(powder) structure. Smaller Pixel- higher Spatial resolution. Similar images like in Interventional and Cardio modalities. 59

60 Flat Panel Detector- CT Bad material became old (improvements?). Worse scintillation. Lower Contrast resolution (on thinner slices). Higher dose for equal SNR ratio in comparison with MDCT. Longer time of acquisition. SR and LCR Up to 30 lp/cm in entire FOV (npr. 18cm). Routine CT aprox LP/CM UHR up to 25 lp/cm limted by FOV size. MDCT ± 3 HU FPCT ± 5 HU or more 60

61 Basic difference All standard dose description are not usefull (CTDI, DLP), NEW ONE should be estimated. Some image characteristic are not equal (training for radiologist) between MDCT in FPCT. Image formation is different. Especially for Interventional (angio), Cardio, Skeletal and ORL. Automatics : MDCT vs FPCT Old CT: constant kv, mas and time. Modern CT : constant kv (2 sources, 2 kv), ma, constant time. FPCT: kvp, ma, constant time. Conclusion... Correct use of technology requires didactic education, hands on training and continuous re training of users. 61

62 Thank you! 62

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH 3/27/212 Advantages of SPECT SPECT / CT Basic Principles Dr John C. Dickson, Principal Physicist UCLH Institute of Nuclear Medicine, University College London Hospitals and University College London john.dickson@uclh.nhs.uk

More information

Ch. 4 Physical Principles of CT

Ch. 4 Physical Principles of CT Ch. 4 Physical Principles of CT CLRS 408: Intro to CT Department of Radiation Sciences Review: Why CT? Solution for radiography/tomography limitations Superimposition of structures Distinguishing between

More information

Cardiac Dual Energy CT: Technique

Cardiac Dual Energy CT: Technique RSNA 2013, VSCA51-01, Chicago, Dec. 5, 2013 Cardiac Radiology Series Cardiac Dual Energy CT: Technique Willi A. Kalender, Ph.D. Institute of Medical Physics University of Erlangen www.imp.uni-erlangen.de

More information

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA Spiral CT Protocol Optimization & Quality Assurance Protocol optimization

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

Computer-Tomography II: Image reconstruction and applications

Computer-Tomography II: Image reconstruction and applications Computer-Tomography II: Image reconstruction and applications Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de

More information

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition Quantitative Multi-Energy Computed Tomography: Imaging and Therapy Advancements Disclosures MECT Systems Overview and Quantitative Opportunities The speaker receives research funding from GE Healthcare

More information

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA Computed Tomography Principles, Design, Artifacts, and Recent Advances THIRD EDITION Jiang Hsieh SPIE PRESS Bellingham, Washington USA Table of Contents Preface Nomenclature and Abbreviations xi xv 1 Introduction

More information

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION Frank Dong, PhD, DABR Diagnostic Physicist, Imaging Institute Cleveland Clinic Foundation and Associate Professor of Radiology

More information

BME I5000: Biomedical Imaging

BME I5000: Biomedical Imaging 1 Lucas Parra, CCNY BME I5000: Biomedical Imaging Lecture 4 Computed Tomography Lucas C. Parra, parra@ccny.cuny.edu some slides inspired by lecture notes of Andreas H. Hilscher at Columbia University.

More information

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead.

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead. 1 CT Basics Principles of Spiral CT Dose 2 Who invented CT? 1963 - Alan Cormack developed a mathematical method of reconstructing images from x-ray projections Sir Godfrey Hounsfield worked for the Central

More information

Optimisation of Toshiba Aquilion ONE Volume Imaging

Optimisation of Toshiba Aquilion ONE Volume Imaging Optimisation of Toshiba Aquilion ONE Volume Imaging Jane Edwards, RPRSG Royal Free London NHS Foundation Trust Dr Mufudzi Maviki, Plymouth Hospitals NHS Trust Background In 2011/12 Radiology at RFH was

More information

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT Simulation Imaging Ingrid Reiser Dept. of Radiology The University of Chicago Optimization of CT imaging Goal: Achieve image quality that allows to perform the task at hand (diagnostic

More information

Image Acquisition Systems

Image Acquisition Systems Image Acquisition Systems Goals and Terminology Conventional Radiography Axial Tomography Computer Axial Tomography (CAT) Magnetic Resonance Imaging (MRI) PET, SPECT Ultrasound Microscopy Imaging ITCS

More information

Introduction to Biomedical Imaging

Introduction to Biomedical Imaging Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu X-ray Projection Imaging Computed Tomography Digital X-ray

More information

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Corso di laurea in Fisica A.A Fisica Medica 4 TC Corso di laurea in Fisica A.A. 2007-2008 Fisica Medica 4 TC Computed Tomography Principles 1. Projection measurement 2. Scanner systems 3. Scanning modes Basic Tomographic Principle The internal structure

More information

RADIOLOGY AND DIAGNOSTIC IMAGING

RADIOLOGY AND DIAGNOSTIC IMAGING Day 2 part 2 RADIOLOGY AND DIAGNOSTIC IMAGING Dr hab. Zbigniew Serafin, MD, PhD serafin@cm.umk.pl 2 3 4 5 CT technique CT technique 6 CT system Kanal K: RSNA/AAPM web module: CT Systems & CT Image Quality

More information

A closer look at CT scanning

A closer look at CT scanning Vet Times The website for the veterinary profession https://www.vettimes.co.uk A closer look at CT scanning Author : Charissa Lee, Natalie Webster Categories : General, Vets Date : April 3, 2017 A basic

More information

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Tecniques of imaging in Overview Introduction Tecniques of imaging

More information

Digital Image Processing

Digital Image Processing Digital Image Processing SPECIAL TOPICS CT IMAGES Hamid R. Rabiee Fall 2015 What is an image? 2 Are images only about visual concepts? We ve already seen that there are other kinds of image. In this lecture

More information

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook X-Ray Computed Tomography in Biomedical Engineering, by Robert Cierniak, Springer, 211 Computed Tomography

More information

AIDR 3D Iterative Reconstruction:

AIDR 3D Iterative Reconstruction: Iterative Reconstruction: Integrated, Automated and Adaptive Dose Reduction Erin Angel, PhD Manager, Clinical Sciences, CT Canon Medical Systems USA Iterative Reconstruction 1 Since the introduction of

More information

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING Reveal pathology Reveal the anatomic truth Steven R. Singer, DDS srs2@columbia.edu IDEAL DIAGNOSTIC IMAGING STUDY Provides desired diagnostic

More information

Fundamentals of CT imaging

Fundamentals of CT imaging SECTION 1 Fundamentals of CT imaging I History In the early 1970s Sir Godfrey Hounsfield s research produced the first clinically useful CT scans. Original scanners took approximately 6 minutes to perform

More information

Tomographic Reconstruction

Tomographic Reconstruction Tomographic Reconstruction 3D Image Processing Torsten Möller Reading Gonzales + Woods, Chapter 5.11 2 Overview Physics History Reconstruction basic idea Radon transform Fourier-Slice theorem (Parallel-beam)

More information

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D.

Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Multi-slice CT Image Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced

More information

ML reconstruction for CT

ML reconstruction for CT ML reconstruction for CT derivation of MLTR rigid motion correction resolution modeling polychromatic ML model dual energy ML model Bruno De Man, Katrien Van Slambrouck, Maarten Depypere, Frederik Maes,

More information

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004

Spiral ASSR Std p = 1.0. Spiral EPBP Std. 256 slices (0/300) Kachelrieß et al., Med. Phys. 31(6): , 2004 Spiral ASSR Std p = 1.0 Spiral EPBP Std p = 1.0 Kachelrieß et al., Med. Phys. 31(6): 1623-1641, 2004 256 slices (0/300) Advantages of Cone-Beam Spiral CT Image quality nearly independent of pitch Increase

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

CT Iterative Reconstruction Techniques

CT Iterative Reconstruction Techniques RECENT ADVANCES IN CT RADIATION DOSE REDUCTION TECHNIQUES CT Iterative Reconstruction Techniques Kalpana Kanal, PhD, FSCBTMR, FACR, FAAPM University of Washington Seattle, WA SCBT-MR 2017 Nashville, Tennessee

More information

PURE. ViSION Edition PET/CT. Patient Comfort Put First.

PURE. ViSION Edition PET/CT. Patient Comfort Put First. PURE ViSION Edition PET/CT Patient Comfort Put First. 2 System features that put patient comfort and safety first. Oncology patients deserve the highest levels of safety and comfort during scans. Our Celesteion

More information

Fits you like no other

Fits you like no other Fits you like no other BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase in room

More information

Philips SPECT/CT Systems

Philips SPECT/CT Systems Philips SPECT/CT Systems Ling Shao, PhD Director, Imaging Physics & System Analysis Nuclear Medicine, Philips Healthcare June 14, 2008 *Presented SNM08 Categorical Seminar - Quantitative SPECT and PET

More information

Some reference material

Some reference material Some reference material Physics reference book on medical imaging: A good one is The Essential Physics of Medical Imaging, 3 rd Ed. by Bushberg et al. ($170! new). However, there are several similar books

More information

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland Metal Artifact Reduction CT Techniques R S S S Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland N. 1 v o 4 1 0 2. Postoperative CT Metal Implants CT is accurate for assessment

More information

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon WHITE PAPER Introduction Introducing an image guidance system based on Cone Beam CT (CBCT) and a mask immobilization

More information

Computed Tomography. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Sep 26th/Oct 3rd, 2016

Computed Tomography. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Sep 26th/Oct 3rd, 2016 Computed Tomography Principles of Medical Imaging Prof. Dr. Philippe Cattin MIAC, University of Basel Contents Abstract 1 Computed Tomography Basics Introduction Computed Tomography Hounsfield's CT Prototype

More information

The Near Future in Cardiac CT Image Reconstruction

The Near Future in Cardiac CT Image Reconstruction SCCT 2010 The Near Future in Cardiac CT Image Reconstruction Marc Kachelrieß Institute of Medical Physics (IMP) Friedrich-Alexander Alexander-University Erlangen-Nürnberg rnberg www.imp.uni-erlangen.de

More information

Iterative Reconstructions: The Impact on Dose and Image Quality

Iterative Reconstructions: The Impact on Dose and Image Quality Shane Foley: IPR, Chicago 2016 Iterative Reconstructions: The Impact on Dose and Image Quality Shane Foley, PhD Radiography and Diagnostic Imaging School of Medicine University College Dublin. Radagrafaíocht

More information

Biomedical Imaging. Computed Tomography. Patrícia Figueiredo IST

Biomedical Imaging. Computed Tomography. Patrícia Figueiredo IST Biomedical Imaging Computed Tomography Patrícia Figueiredo IST 2013-2014 Overview Basic principles X ray attenuation projection Slice selection and line projections Projection reconstruction Instrumentation

More information

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 Revised 10/10 COURSE SYLLABUS TM 220 COMPUTED TOMOGRAPHY PHYSICS CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0 CATALOG COURSE DESCRIPTION: This course is one of a three course set in whole body Computed

More information

Fits you like no other

Fits you like no other Fits you like no other Philips BrightView X and XCT specifications The new BrightView X system is a fully featured variableangle camera that is field-upgradeable to BrightView XCT without any increase

More information

CT: Physics Principles & Equipment Design

CT: Physics Principles & Equipment Design CT: Physics Principles & Equipment Design James Kofler, Ph.D Radiology Mayo Clinic Rochester, MN June 27, 2012 Disclosures Nothing to disclose Learning Objectives Understand fundamental concepts of - CT

More information

Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage. 29 April - 3 May, 2013

Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage. 29 April - 3 May, 2013 2455-5 Joint ICTP-TWAS Workshop on Portable X-ray Analytical Instruments for Cultural Heritage 29 April - 3 May, 2013 Lecture NoteBasic principles of X-ray Computed Tomography Diego Dreossi Elettra, Trieste

More information

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography Marc Kachelrieß, a Katia Sourbelle, and Willi A. Kalender Institute of Medical Physics, University of

More information

Effect of Scattering on the Image. Reducing Compton Scatter with a Grid

Effect of Scattering on the Image. Reducing Compton Scatter with a Grid Effect of Scattering on the Image Increasing Compton scattering degrades image. Webb 21 Reducing Compton Scatter with a Grid Grids Parallel (focused at infinity) Linear Focused (see figure) Moving grids

More information

AAPM Standard of Practice: CT Protocol Review Physicist

AAPM Standard of Practice: CT Protocol Review Physicist AAPM Standard of Practice: CT Protocol Review Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center September 11, 2014 2014 Texas Radiation Regulatory Conference Goals Understand purpose

More information

CT vs. VolumeScope: image quality and dose comparison

CT vs. VolumeScope: image quality and dose comparison CT vs. VolumeScope: image quality and dose comparison V.N. Vasiliev *a, A.F. Gamaliy **b, M.Yu. Zaytsev b, K.V. Zaytseva ***b a Russian Sci. Center of Roentgenology & Radiology, 86, Profsoyuznaya, Moscow,

More information

Limits of Dose Reduction in CT: How low is too low? Acknowledgement. Disclosures 8/2/2012

Limits of Dose Reduction in CT: How low is too low? Acknowledgement. Disclosures 8/2/2012 8/2/22 Limits of Dose Reduction in CT: How low is too low? Ehsan Samei Duke University Acknowledgement Disclosures Research grant: NIH Research grant: DHS Research grant: RSNA, QIBA Research grant: General

More information

DUAL energy X-ray radiography [1] can be used to separate

DUAL energy X-ray radiography [1] can be used to separate IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 53, NO. 1, FEBRUARY 2006 133 A Scatter Correction Using Thickness Iteration in Dual-Energy Radiography S. K. Ahn, G. Cho, and H. Jeon Abstract In dual-energy

More information

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit Ke Li Assistant Professor Department of Medical Physics and Department of Radiology School of Medicine and Public Health, University of Wisconsin-Madison This work is partially supported by an NIH Grant

More information

Medical Image Processing: Image Reconstruction and 3D Renderings

Medical Image Processing: Image Reconstruction and 3D Renderings Medical Image Processing: Image Reconstruction and 3D Renderings 김보형 서울대학교컴퓨터공학부 Computer Graphics and Image Processing Lab. 2011. 3. 23 1 Computer Graphics & Image Processing Computer Graphics : Create,

More information

Enhancement Image Quality of CT Using Single Slice Spiral Technique

Enhancement Image Quality of CT Using Single Slice Spiral Technique Enhancement Image Quality of CT Using Single Slice Spiral Technique Doaa. N. Al Sheack 1 and Dr.Mohammed H. Ali Al Hayani 2 1 2 Electronic and Communications Engineering Department College of Engineering,

More information

DUE to beam polychromacity in CT and the energy dependence

DUE to beam polychromacity in CT and the energy dependence 1 Empirical Water Precorrection for Cone-Beam Computed Tomography Katia Sourbelle, Marc Kachelrieß, Member, IEEE, and Willi A. Kalender Abstract We propose an algorithm to correct for the cupping artifact

More information

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid by Yingxuan Chen Graduate Program in Medical Physics Duke University Date: Approved: Lei Ren, Supervisor Fang Fang Yin, Chair

More information

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing ImPACT Information Leaflet No. 1: CT Scanner Acceptance Testing Version 1.02, 18/05/01 CONTENTS: 1. SCOPE OF LEAFLET 2. GENERAL PRINCIPLES OF ACCEPTANCE AND COMMISSIONING 2.1 PHANTOMS 2.2 EXPOSURE AND

More information

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology

Diagnostic imaging techniques. Krasznai Zoltán. University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology Diagnostic imaging techniques Krasznai Zoltán University of Debrecen Medical and Health Science Centre Department of Biophysics and Cell Biology 1. Computer tomography (CT) 2. Gamma camera 3. Single Photon

More information

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems. Slide 1 Technical Aspects of Quality Control in Magnetic Resonance Imaging Slide 2 Compliance Testing of MRI Systems, Ph.D. Department of Radiology Henry Ford Hospital, Detroit, MI Slide 3 Compliance Testing

More information

As fl exible as your care requires

As fl exible as your care requires As fl exible as your care requires Philips Ingenuity Flex 32 CT Built on Ingenuity The Philips Ingenuity Flex 32 helps you provide excellent care with outstanding flexibility, now and in the future. Built

More information

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE / Radiographic Tomosynthesis: Image Quality and Artifacts Reduction Baojun Li, Ph.D Department of Radiology Boston University Medical Center 2012 AAPM Annual Meeting Background Linear Trajectory Tomosynthesis

More information

Equipment Specification

Equipment Specification MULTISLICE CT SCANNER ( 16 Slices ) Merk : Hitachi Japan Model : SUPRIA 5 MHU Price : Rp 6.512.360.215,27 No Equipment Specification 1 2 3 Scanner Gantry Object for scanning : Whole body including head

More information

Financial disclosure. Onboard imaging modality for IGRT

Financial disclosure. Onboard imaging modality for IGRT Tetrahedron Beam Computed Tomography Based On Multi-Pixel X- Ray Source and Its Application in Image Guided Radiotherapy Tiezhi Zhang, Ph.D. Advanced X-ray imaging Lab Financial disclosure Patent royalty

More information

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis Stefan C. Saur 1, Hatem Alkadhi 2, Luca Regazzoni 1, Simon Eugster 1, Gábor Székely 1, Philippe Cattin 1,3 1 Computer Vision

More information

Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application

Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application Physics in Medicine & Biology Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application To cite this article: Polad M Shikhaliev and Shannon G Fritz 2011

More information

GPU implementation for rapid iterative image reconstruction algorithm

GPU implementation for rapid iterative image reconstruction algorithm GPU implementation for rapid iterative image reconstruction algorithm and its applications in nuclear medicine Jakub Pietrzak Krzysztof Kacperski Department of Medical Physics, Maria Skłodowska-Curie Memorial

More information

Quality control phantoms and protocol for a tomography system

Quality control phantoms and protocol for a tomography system Quality control phantoms and protocol for a tomography system Lucía Franco 1 1 CT AIMEN, C/Relva 27A O Porriño Pontevedra, Spain, lfranco@aimen.es Abstract Tomography systems for non-destructive testing

More information

Computer-Tomography I: Principles, History, Technology

Computer-Tomography I: Principles, History, Technology Computer-Tomography I: Principles, History, Technology Prof. Dr. U. Oelfke DKFZ Heidelberg Department of Medical Physics (E040) Im Neuenheimer Feld 280 69120 Heidelberg, Germany u.oelfke@dkfz.de History

More information

Z-MOTION. Universal Digital Radiographic System Z-MOTION. Control-X Medical CONTROL-X MEDICAL

Z-MOTION. Universal Digital Radiographic System Z-MOTION. Control-X Medical CONTROL-X MEDICAL Control-X Medical Z-MOTION Compact design, low ceiling height requirement Motorized and manual movement capability Wide motion / SID range Best-in-class image quality Flexible connectivity to PACS systems

More information

Acknowledgments and financial disclosure

Acknowledgments and financial disclosure AAPM 2012 Annual Meeting Digital breast tomosynthesis: basic understanding of physics principles James T. Dobbins III, Ph.D., FAAPM Director, Medical Physics Graduate Program Ravin Advanced Imaging Laboratories

More information

Motion Compensation from Short-Scan Data in Cardiac CT

Motion Compensation from Short-Scan Data in Cardiac CT Motion Compensation from Short-Scan Data in Cardiac CT Juliane Hahn 1,2, Thomas Allmendinger 1, Herbert Bruder 1, and Marc Kachelrieß 2 1 Siemens Healthcare GmbH, Forchheim, Germany 2 German Cancer Research

More information

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

C a t p h a n / T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 T h e P h a n t o m L a b o r a t o r y C a t p h a n 5 0 0 / 6 0 0 Internationally recognized for measuring the maximum obtainable performance of axial, spiral and multi-slice

More information

Computed tomography - outline

Computed tomography - outline Computed tomography - outline Computed Tomography Systems Jørgen Arendt Jensen and Mikael Jensen (DTU Nutech) October 6, 216 Center for Fast Ultrasound Imaging, Build 349 Department of Electrical Engineering

More information

FINDING THE TRUE EDGE IN CTA

FINDING THE TRUE EDGE IN CTA FINDING THE TRUE EDGE IN CTA by: John A. Rumberger, PhD, MD, FACC Your patient has chest pain. The Cardiac CT Angiography shows plaque in the LAD. You adjust the viewing window trying to evaluate the stenosis

More information

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist CT Protocol Review: Practical Tips for the Imaging Physicist Physicist Dianna Cody, Ph.D., DABR, FAAPM U.T.M.D. Anderson Cancer Center August 8, 2013 AAPM Annual Meeting Goals Understand purpose and importance

More information

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Developments in Dimensional Metrology in X-ray Computed Tomography at NPL Wenjuan Sun and Stephen Brown 10 th May 2016 1 Possible factors influencing XCT measurements Components Influencing variables Possible

More information

O-ARM IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE

O-ARM IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE SYSTEM FEATURES CATEGORY PHYSICAL DIMENSIONS IMAGING MODALITY PERFORMANCE SPECIFICATION Length Width Height Weight Gantry Opening Bore Diameter Single Plane

More information

Performance Evaluation of CT Systems

Performance Evaluation of CT Systems 8/2/206 Performance Evaluation of CT Systems AAPM Task Group 233 Foundations, Methods, and Prospects Ehsan Samei Duke University Medical Center Disclosures Research grant: NIH R0 EB00838 Research grant:

More information

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT Benedikt Lorch 1, Martin Berger 1,2, Joachim Hornegger 1,2, Andreas Maier 1,2 1 Pattern Recognition Lab, FAU Erlangen-Nürnberg

More information

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh et al. : Proceedings of the National Seminar & Exhibition on Non-Destructive Evaluation DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE Rajesh V Acharya, Umesh Kumar, Gursharan

More information

Central Slice Theorem

Central Slice Theorem Central Slice Theorem Incident X-rays y f(x,y) R x r x Detected p(, x ) The thick line is described by xcos +ysin =R Properties of Fourier Transform F [ f ( x a)] F [ f ( x)] e j 2 a Spatial Domain Spatial

More information

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

Improvement of Efficiency and Flexibility in Multi-slice Helical CT J. Shanghai Jiaotong Univ. (Sci.), 2008, 13(4): 408 412 DOI: 10.1007/s12204-008-0408-x Improvement of Efficiency and Flexibility in Multi-slice Helical CT SUN Wen-wu 1 ( ), CHEN Si-ping 2 ( ), ZHUANG Tian-ge

More information

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS 23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN 16-20 APRIL 2007 HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS P. Helberg 1,

More information

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system

Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system Implementation and evaluation of a fully 3D OS-MLEM reconstruction algorithm accounting for the PSF of the PET imaging system 3 rd October 2008 11 th Topical Seminar on Innovative Particle and Radiation

More information

Fujifilm DR Solution. FDR AcSelerate. The new pinnacle in diagnostic imaging from Fujifilm ISS. CsI. Dynamic Visualization. Technology.

Fujifilm DR Solution. FDR AcSelerate. The new pinnacle in diagnostic imaging from Fujifilm ISS. CsI. Dynamic Visualization. Technology. Fujifilm DR Solution FDR AcSelerate The new pinnacle in diagnostic imaging from Fujifilm CsI Scintillator ISS Technology Dynamic Visualization Welcome to the X-ray room of the future! A streamlined solution

More information

Brilliance CT Big Bore.

Brilliance CT Big Bore. 1 2 2 There are two methods of RCCT acquisition in widespread clinical use: cine axial and helical. In RCCT with cine axial acquisition, repeat CT images are taken each couch position while recording respiration.

More information

Quest for the ultimate cardiac CT scanner

Quest for the ultimate cardiac CT scanner Quest for the ultimate cardiac CT scanner Paul FitzGerald a) Imaging, GE Global Research, Niskayuna, NY 12309, USA Peter Edic Imaging, GE Global Research, Niskayuna, NY 12309, USA Hewei Gao Radiation Sensing

More information

Principles of Computerized Tomographic Imaging

Principles of Computerized Tomographic Imaging Principles of Computerized Tomographic Imaging Parallel CT, Fanbeam CT, Helical CT and Multislice CT Marjolein van der Glas August 29, 2000 Abstract The total attenuation suffered by one beam of x-rays

More information

Enhanced material contrast by dual-energy microct imaging

Enhanced material contrast by dual-energy microct imaging Enhanced material contrast by dual-energy microct imaging Method note Page 1 of 12 2 Method note: Dual-energy microct analysis 1. Introduction 1.1. The basis for dual energy imaging Micro-computed tomography

More information

A prototype table-top inverse-geometry volumetric CT system

A prototype table-top inverse-geometry volumetric CT system A prototype table-top inverse-geometry volumetric CT system Taly Gilat Schmidt a Department of Radiology, Stanford University, Stanford, California 94305 Josh Star-Lack NexRay, Inc., Los Gatos, California

More information

Non-Stationary CT Image Noise Spectrum Analysis

Non-Stationary CT Image Noise Spectrum Analysis Non-Stationary CT Image Noise Spectrum Analysis Michael Balda, Björn J. Heismann,, Joachim Hornegger Pattern Recognition Lab, Friedrich-Alexander-Universität Erlangen Siemens Healthcare, Erlangen michael.balda@informatik.uni-erlangen.de

More information

PLANMECA PROMAX 3D MID CBCT UNIT

PLANMECA PROMAX 3D MID CBCT UNIT 1(5) PLANMECA PROMAX 3D MID CBCT UNIT Introduction The Planmeca ProMax 3D Mid X-ray unit uses cone beam computerized tomography (CBCT) to produce three-dimensional X-ray images. Panoramic and cephalometric

More information

Medical Imaging BMEN Spring 2016

Medical Imaging BMEN Spring 2016 Name Medical Imaging BMEN 420-501 Spring 2016 Homework #4 and Nuclear Medicine Notes All questions are from the introductory Powerpoint (based on Chapter 7) and text Medical Imaging Signals and Systems,

More information

Computational Medical Imaging Analysis

Computational Medical Imaging Analysis Computational Medical Imaging Analysis Chapter 2: Image Acquisition Systems Jun Zhang Laboratory for Computational Medical Imaging & Data Analysis Department of Computer Science University of Kentucky

More information

Symbia T Series System Specifications Answers for life.

Symbia T Series System Specifications Answers for life. www.siemens.com/mi System Specifications Answers for life. 2 Diagnostic Excellence Today s healthcare institutions have increasingly diverse patient needs, greater budget limitations and the need for higher

More information

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images Tao Han, Chris C. Shaw, Lingyun Chen, Chao-jen Lai, Xinming Liu, Tianpeng Wang Digital Imaging Research Laboratory (DIRL),

More information

Recognition and Measurement of Small Defects in ICT Testing

Recognition and Measurement of Small Defects in ICT Testing 19 th World Conference on Non-Destructive Testing 2016 Recognition and Measurement of Small Defects in ICT Testing Guo ZHIMIN, Ni PEIJUN, Zhang WEIGUO, Qi ZICHENG Inner Mongolia Metallic Materials Research

More information

Interaction map. X-Ray Computed Tomography Measures Tissue Properties from Macro to Micro. Outline

Interaction map. X-Ray Computed Tomography Measures Tissue Properties from Macro to Micro. Outline X-Ray Computed Tomography Measures Tissue Properties from Macro to Micro Michael Andre, Ph.D. Department of Radiology University of California, San Diego San Diego VA Healthcare System Outline Perspective

More information

COMPARATIVE STUDIES OF DIFFERENT SYSTEM MODELS FOR ITERATIVE CT IMAGE RECONSTRUCTION

COMPARATIVE STUDIES OF DIFFERENT SYSTEM MODELS FOR ITERATIVE CT IMAGE RECONSTRUCTION COMPARATIVE STUDIES OF DIFFERENT SYSTEM MODELS FOR ITERATIVE CT IMAGE RECONSTRUCTION BY CHUANG MIAO A Thesis Submitted to the Graduate Faculty of WAKE FOREST UNIVERSITY GRADUATE SCHOOL OF ARTS AND SCIENCES

More information

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies g Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies Presented by Adam Kesner, Ph.D., DABR Assistant Professor, Division of Radiological Sciences,

More information

Scatter Correction Methods in Dimensional CT

Scatter Correction Methods in Dimensional CT Scatter Correction Methods in Dimensional CT Matthias Baer 1,2, Michael Hammer 3, Michael Knaup 1, Ingomar Schmidt 3, Ralf Christoph 3, Marc Kachelrieß 2 1 Institute of Medical Physics, Friedrich-Alexander-University

More information